COMPRESSOR MINI
VENTILATION/ANESTHESIA
| Table of contents |
TABLE OF CONTENTS
1 Before use | 4
2 Device description | 7
3 Preparation/Connection | 12
4 Function check | 13
5 Cleaning | 15
6 Preventive maintenance / Storage | 17
7 Alarms | 18
8 Troubleshooting | 19
9 Technical data | 20
1 BEFORE USE
The function check must always be carried Service and repair must be carried out at
out before the unit is put into use, or once regular intervals by MAQUET-authorized
a week when the unit is in continuous use. service personnel as follows:
Make sure the mains power and gas fulfill - 5 000 hour preventive maintenance
requirements for medical equipment. - 10 000 hour preventive maintenance
The Compressor Mini must only be
Note: For 230 V, 60 Hz version, the
connected to a grounded outlet.
preventive maintenance intervals are
4 000 hours and 8 000 hours of operation
1.3 OPERATION respectively.
1.8 CAUTION
2 DEVICE DESCRIPTION
KMP-001X
61 psi
12345 hours
KMP-042US
On Standby Alarm
61 psi
12345 hours
KMP-045
On Standby Alarm
1. Green LED - On
- The green LED lights up when the
mains switch is on. Display message:
The pressure in the compressor and
elapsed operating hours (when green
LED only).
2. Yellow LED - Standby
- The yellow Standby LED will be on
while the central compressed air supply
is sufficient from the central air supply
to the ventilator via the compressor.
Display message: The central
compressed air pressure and the word
STANDBY.
3. Red LED - Alarms
- The red Alarm LED indicates different
alarm situations. Display message:
Shows if the alarm is related to gas
distribution, dryer condition or
overheating of the compressor. The
alarms are described in detail in the
Alarms section.
4. Audible alarm
- A buzzer for audible alarm signals (3
beeps/second). The buzzer sounds
when an alarm is activated.
2.5 CONNECTIONS
1. Inlet
2. Outlet
OUT
3. Drainage bottle
4. Condensation drain valve
- Empty the air reservoir by pressing the
IN pin on the drain valve while the
compressor is running.
IN
KMP-013aX
3 PREPARATION/CONNECTION
OUT
0 IN
KMP-032X IN
KMP-052X
4 inch (10 cm)
Place the compressor in a suitable location. Connect the compressed air hose from the
The air vent at the back of the compressor compressor air outlet to the ventilator.
must not be covered or blocked Plug in the compressor mains cable to a
- Do not place the compressor near a grounded outlet.
radiator or heating unit, since the
operating temperature might increase 3.3 START-UP
and activate the temperature alarm.
- When the compressor is used as a
A function check should be done before the
standby unit, connect the central
unit is put into use, and once a week when in
compressed air supply to the air inlet on
continuous use.
the compressor.
- Check that an empty drainage bottle is
fitted. CAUTION: If a fault is discovered during
- MAQUET Compressor Mini mounting kits the function check, the compressor must
are mandatory for both stand-alone not be used until it has been repaired by
arrangements and for mounting on carts. authorized service personnel.
The mounting kits ensure that there is
sufficient air circulation around the
compressor.
4 FUNCTION CHECK
KMP-035X
KMP-007X
To check the internal air reservoir for water, Make sure that air circulates through the
press the pin on the drain valve. ventilating holes.
Disconnect the inlet (if connected) and
outlet air connection to the compressor.
Start the compressor with the Mains switch.
KMP-029X
On Standb Alarm
63 psi
=
IN
STANDBY
KMP-017X
Connect the central air supply. Check the water level in the drainage bottle
- The compressor is automatically regularly.
switched off when the air pressure Empty the bottle if necessary; the
exceeds 44 psi (300 kPa/3.0 bar) for compressor does not need to be turned off.
more than 30 seconds.
IN
KMP-019X
KMP-046X
On Standby Alarm
5 CLEANING
5.1 DAILY CLEANING Wipe all parts with a soft lint-free cloth
moistened in soap and water or
If the compressor is used in a dusty detergent-based disinfectant.
environment, the dust filters must be cleaned
daily.
Note: In case of more contaminated
surfaces use ethylene alcohol or isopropyl
5.2 200 HOUR CLEANING alcohol.
KMP-049X
OUT
a IN
KMP-047X
Clean the dust filters (a+b) on the back and
the underside of the compressor if
necessary
Wash the drainage bottle
KMP-048X
KMP-050X
0
0
KMP-020X
7 ALARMS
7.1 ALARMS
Alarms Description
High temp alarm 1 The red LED and the alarm buzzer are activated if
the temperature inside the compressor unit exceeds
158 °F ±5 °F (70 °C ±3 °C). The alarm is activated
62 psi until the temperature drops below 131 °F ±11 °F
High temp 1
(55 °C ±6 °C), or the compressor is switched off.
KMP-036US
On Standby Alarm
High temp alarm 2 The red LED and the alarm buzzer are activated if
the temperature difference between ambient air
and the temperature in the dryer is less than approx.
62 psi 4 °F (2 °C) after 15 minutes operating time. The
High temp 2
alarm is activated until the temperature difference
KMP-037US
High temp alarm 3 The red LED and the alarm buzzer are activated if
the temperature in the dryer exceeds 131 °F ±11 °F
(55 °C ±6 °C). The alarm is activated until the
62 psi temperature inside the dryer drops below 122 °F
High temp 3
±11 °F (50°C ±6 °C).
KMP-038US
On Standby Alarm
Low pressure alarm The red LED and alarm buzzer are activated if the
compressed air pressure drops and stays below
29 psi (200 kPa/2 bar) for more than 5 seconds.
24 psi The alarm is activated until the air pressure rises
Low pressure
above approx. 36 psi (250 kPa/2.5 bar) for more
KMP-043US
High Pressure alarm The red LED and alarm buzzer are activated if the
compressed air pressure rises and stays above
94 psi (650 kPa/6.5 bar) for more than 0.5 second.
100 psi The alarm is active until the air pressure drops and
High pressure
stays below 87 psi (600 kPa/6 bar) for more than
KMP-044US
Power failure alarm Power failure. The alarm buzzer is activated for
about 12 seconds.
8 TROUBLESHOOTING
8.1 TROUBLESHOOTING
9 TECHNICAL DATA
General
Dimension W 17", D 13", H 10" inches (W 430, D 330,
H 250 mm)
Weight approx. 57 lbs (26 kg)
Classification Class I, continuous operation equipment
according to IEC/EN 60601-1 (GB9706.1)
Power consumption 550 W at nominal voltage
Noise level at 1 m distance approx. 50 dB(A)
Power supply 115 V AC, 60 Hz (single phase). 220–240 V
AC, 50/60 Hz (single phase).
IP classification IP 20
Compressor capacity
Continuous flow at normal atmospheric 30 l/min (expanded to ambient air pressure)
pressure (approx. 1013 hPa) at 50 psi (350 kPa/3.5 bar)
Operating pressure (at normal setting of the 50–64 psi (350–450 kPa/3.5 – 4.5 bar)
internal pressure regulator)
Dew point suppression at 30 l/min >9 °F (5 °C) at 5 psi (350 kPa/3.5 bar)
Built-in air reservoir 1.25 litres
Environmental conditions – Operation
Ambient temperature:
Total working range
Full capacity range (30 l/min) 59 °F to 86 °F (15 °C to 30 °C)
Reduced capacity range* 50 °F to 59 °F and 86 °F to 104 °F (10 °C to
15 °C and 30 °C to 40 °C)
Ambient relative humidity:
Total working range
Full capacity range (30 l/min) 15% to 90%
Reduced capacity range* >90%
Ambient atmospheric pressure:
Total working range
Full capacity range (30 l/min) 950 to 1060 hPa (712 to 795 mmHg)
Reduced capacity range* 660 to 950 and 1060 to 1200 hPa (495 to 712
and 795 to 900 mmHg)
*Reduced capacity:
Expected flow and dew point suppression 24 l/min, 5 °F (3 °C) at 50 psi (350 kPa/3.5 bar).
This reduced capacity applies if one or more
of the limitations above apply.
Note: At a continuous flow of >30 l/min (normal for an adult patient is 8–10 l/min) a
surrounding temperature of approximately 104 °F (40 °C), and a relative humidity of >90%
RH, the pressure dew point could be less than 5 °F (3 °C).
This product complies with environmental protection use period as defined in People`s Republic
of China Electronic Industry Standard SJ/T11364-2006.
The following table shows the names and contents of toxic or hazardous substances in this
product as defined in People`s Republic of China Electronic Industry Standard SJ/T11364-2006.
Hazardous substances
6+
Parts Pb Hg Cd Cr PBB PBDE
Metal parts 0 0 0 0 0 0
Plastic and polymeric parts 0 0 0 0 0 0
Electrical components X 0 0 0 0 0
LCD display 0 0 0 0 0 0
0: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for
this part is below the limit and meets the requirement in SJ/T11363-2006.
X: Indicates that this toxic or hazardous substance contained in at least one of the homogeneous
materials used for this part is above the limit requirement in SJ/T11363-2006.