Secrest Thesis
Secrest Thesis
ASYMMETRIES IN −
→
e p SCATTERING AT
0.1 < Q2 < 0.4 (GeV/c)2
A Dissertation
Presented to
In Partial Fulfillment
Doctor of Philosophy
by
Jeffery A. Secrest
2005
APPROVAL SHEET
Doctor of Philosophy
Jeffery A. Secrest
ii
CONTENTS
ABSTRACT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xiii
CHAPTER
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 Physics Formalism . . . . . . . . . . . . . . . . . . . . . . . . . 19
Asymmetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3 Experimental Apparatus . . . . . . . . . . . . . . . . . . . . . . 39
iii
3.3 The G0 Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.5.4 Photocathode . . . . . . . . . . . . . . . . . . . . . . . . . 61
4 Data Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
4.3 Cuts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
iv
4.8.1 French Results . . . . . . . . . . . . . . . . . . . . . . . . . 96
5 Conclusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
APPENDIX A
G0 Abbreviation & Acronym Glossary (GAAG) . . . . . . . . . . 127
APPENDIX B
Detector Testing and Calibration . . . . . . . . . . . . . . . . . . . 131
B.1 Output signals of the North American Focal Plane Detectors . . . 131
APPENDIX C
Injector Studies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
BIBLIOGRAPHY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
VITA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
v
LIST OF TABLES
4.2 IHWP out false asymmetries. The data are from 49 asymmetry runs. 95
4.3 IHWP in false asymmetries. The data are from 58 asymmetry runs. . 95
4.5 The average inelastic dilution factors for the French detectors for the
cuts on the elastic peak which were ∼ 4 ns. . . . . . . . . . . . . . . . 99
4.6 The average inelastic dilution factors and the final absolute errors for
the North American detectors. . . . . . . . . . . . . . . . . . . . . . . 105
vi
5.1 Extracted North American and French asymmetries. . . . . . . . . . 118
B.2 Minimum energies left by the protons hitting the G 0 scintillators and
subsequent number of photo-electrons detected. . . . . . . . . . . . . 142
C.3 Electronic noise by taking a pedestal run and injecting a fake signal. . 150
C.4 The responses of the injector BPM’s position differences to the X and
Y motion of the PZT mirror on December 19, 2002. . . . . . . . . . 151
C.6 These results are the angle (in degrees) between the responses of the
PZT motion in X and in Y on December 19, 2002. . . . . . . . . . . . 151
C.7 The responses of the injector BPM’s position differences to the X and
Y motion of the PZT mirror on January 14, 2003. . . . . . . . . . . 152
C.9 These results are the angle (in degrees) between the responses of the
PZT motion in X and in Y on January 14, 2003. . . . . . . . . . . . . 152
C.10 The responses of the injector BPM’s position differences to the X and
Y motion of the PZT mirror on January 24, 2003. . . . . . . . . . . . 153
vii
C.12 These results are the angle (in degrees) between the responses of the
PZT motion in X and in Y on January 24, 2003. . . . . . . . . . . . . 153
viii
LIST OF FIGURES
1.1 The tree level electromagnetic and weak contributions to the scatter-
ing cross section in electron proton scattering. . . . . . . . . . . . . . 7
2.4 Plot of the dipole parameterizations of GpM , GpE , and GnM and the
Galster parameterization of GnE . . . . . . . . . . . . . . . . . . . . . . 24
2.6 Tree level Feynman diagram for electron-proton weak neutral current
scattering. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
ix
3.1 A schematic of the G0 magnet collimators, target, and detectors. . . . 40
3.8 Schematic of the CEBAF polarized source laser table in the injector. 54
3.10 Typical performance of the charge feedback system using the IA cell. 58
3.12 Schematic of strained GaAs band structure and energy level diagram. 61
3.13 Plot of normalized rates as a function of run number for the 2003 G 0
engineering run as measured by the Beam Halo monitor. . . . . . . . 65
3.14 Plot of Lumi asymmetry as a function of run number for the 2003 G 0
commissioning run. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
3.16 Schematic of the North American electronics for the G 0 forward run-
ning mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
x
4.5 Plots of the helicity-correlated beam properties. . . . . . . . . . . . . 93
4.8 An example of the stability of the inelastic dilution factor over 106
runs for detector 14 Octant 2. . . . . . . . . . . . . . . . . . . . . . . 97
4.11 An example of several white noise spectra for North American octant
3 detector 4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
4.13 The plots of a simulated North American ToF spectrum using the fit
parameters obtained from the French ToF and North American ToF
spectrum with DNL. . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
4.16 Stability of the North American inelastic dilution factors over ∼50
runs for a typical detector. . . . . . . . . . . . . . . . . . . . . . . . . 105
xi
4.22 Polarization measurements made during the engineering run. . . . . . 111
5.1 Plot of the extracted North American and French asymmetries vs.
momentum transfer. . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
5.4 Expected forward angle results from the G0 , HAPPEX II, and A4
along with result from the HAPPEX experiment. . . . . . . . . . . . 125
B.5 Ratio of the gain of the actual PMT attached to the light pipe to the
gain necessary to perfectly match the light pipe. . . . . . . . . . . . . 144
B.6 Dispersion of the ratio of the measured to the predicted cosmic am-
plitudes in the Clean Room. . . . . . . . . . . . . . . . . . . . . . . . 144
C.5 Charge asymmetry calibration constants (ppm/V) for the IA. . . . . 159
xii
ABSTRACT
This work is based on the first engineering run of the G 0 experiment from
October 2002 though January 2003 in Hall C at Jefferson Lab. The G 0 experiment
will be the first to measure the weak neutral form factors: G ZE (Q2 ), GZM (Q2 ) and
GeA (Q2 ) and to extract the proton’s strange form factors: GsE (Q2 ) and GsM (Q2 ) via
a Rosenbluth separation over a range of Q2 (0.1 − 1.0 (GeV/c)2 ). This will require
four sets of measurements: forward angle measurements with a proton target, and
three sets of backward angle measurements with a hydrogen and deuterium target.
The measurements are made of the parity-violating asymmetries in elastic electron
scattering.
The G0 experiment is a major installation at Jefferson Lab’s Hall C with a new
dedicated spectrometer. The superconducting magnet is made up of 8 coils with
a maximum field of 1.6 T·m. The scintillator detector array (detector solid angle
between 0.4 - 0.9 sr) detects recoiled protons in the forward angle measurement
(where θp = 70◦ ± 10◦ corresponding to scattered electron angles of a few degrees)
and to detect scattered electrons in the backward angle measurements. This detector
array is made up of a set of 16 pairs of scintillators arranged in 8 sectors around
the beam line. Custom electronics handle the high data rate (approximately 1 MHz
per detector). The target is a 20 cm long liquid hydrogen cryotarget. Besides the
check-out of the new hardware, G0 has stringent requirements on the performance
of the polarized electron beam in order to minimize false asymmetries. Further
complicating this fact, in the forward mode, was the requirement that the time
structure of the JLab beam had to be changed from 499 MHz to 31 MHz in order
to count the recoiled protons in the spectrometer. Data was collected over a 12 day
period at the end of the engineering run. These data were analyzed for a Q 2 range of
0.1 − 0.4 (GeV/c)2 corresponding to measured electroweak asymmetries that ranged
from (-4.4 ± 1.6 ± 1.6 ppm) to (-8.5 ± 2.8 ± 2.5 ppm).
xiii
CHAPTER 1
Introduction
The existence of the proton has been known since the early 20th century, yet
its structure is still not entirely understood. The current theory of the strong in-
up of three valence quarks (two up and one down quarks) within a complicated
“sea” of quarks, antiquarks and gluons. A troubling difficulty of QCD is that while
the theory is perturbative, the theory is more difficult to handle at low momentum
Strange and anti-strange quarks, which are the next lightest quarks after the
up and down, are found in the quark “sea” around the proton. Since strange quarks
have a comparable mass with the proton, the question can be asked, what role do
Strange quarks are the lightest quarks not to contribute to the proton’s valence
distribution. The strange quarks exist as only strange quark-antiquark pairs in the
1
2
quark sea surrounding the proton. The net strangeness of the proton is zero. This
might lead one to believe that strange quarks cannot contribute to the properties
of the proton. Experiments have indicated that strange quarks do, in fact, play a
One set of strange quark observables is related to the so-called strange quark
matrix elements of the nucleon. These matrix elements have the form
hP |s̄Γs|P i (1.1)
where |P i is the proton state, s̄Γs is an operator containing strange quark, s, fields
the proton.
One of the original indicators that strange quarks play a fundamental role in
the proton came from looking at the pion-nucleon sigma term. Strange quarks
contribute to the mass of the proton via the matrix element hP |s̄s|P i. This matrix
element can be inferred from the so-called “sigma term” in π-N scattering. The
¯ i
σπN = m̂hP |ūu + dd|P (1.2)
where m̂ = 12 (mu + md ), the average of the up and down quark masses. The proton
mass, under the SU(3) flavor assumption that one can neglect cc̄, b b̄ and tt̄, can then
be written as
Mp = M0 + σs + σ (1.3)
where Mp is the physical mass of the proton, M0 is the mass of the proton as the
quark masses go to zero (in the so-called chiral limit), σ s is the mass contribution
due to strangeness (if any) and σ is the mass contribution from non-strange quarks.
3
The pion-nucleon term σπN can be used to find the non-strangeness quark con-
tribution σ by extracting σπN at the unphysical Cheng-Daschen point (that is, where
the Mandelstam variables s = Mp2 and t = m2π ) by use of dispersion relations. The
standard value [1] for this result, after taking into account higher-order contribu-
tions, is σ = 45 ± 8 MeV.
If the nucleon is free of strange quarks, σ should equal the SU(3)-octet scaler
¯ − 2s̄s|P i.
σ̂ = m0 hP |ūu + dd (1.4)
where y is the strangeness content of the proton, which can then be written as
2hP |s̄s|P i
y= ¯ i. (1.6)
hP |ūu + dd|P
This leads to a value of y = 0.2 ± 0.2, indicating that strange quarks might
contribute to the mass of the proton. This implies that as much as 200 MeV of
the proton’s mass might be due to the strange quarks. This result also indicates
a violation of the OZI rule [7] which assumes that the nucleon is free of strange
quarks. This result must be accepted with some degree of skepticism since there is a
significant uncertainty due to the data and the extensive theory needed to interpret
this result.
Another piece of evidence for strange quark contributions to the nucleon come
from some p̄p annihilation channels [8, 9]. In these channels, an observed enhance-
Another indicator that strange quarks play a role in the proton comes from po-
larized deep inelastic scattering (DIS) experiments. This method allows for access to
the hP |s̄γ 5 s|P i matrix element. The focus of these experiments is the spin structure
of the nucleon. These spin structure experiments indicate that the fraction of the
proton spin carried by the valence quarks is Σ ∼ 0.3. This naturally leads to the
Each of the quark (antiquark) flavors can be described by a single quark (an-
tiquark) distribution function q(x) (q̄(x)) over a range in Bjorken x, where Bjorken
bution q(x) is the probability that a parton carries a fraction x of the momentum
of the proton. A quantity of interest is the net spin polarization, ∆q, of the quark
flavor q
Z1
∆q ≡ [q ↑ (x) − q ↓ (x) + q̄ ↑ (x) − q̄ ↓ (x)]dx. (1.7)
0
The g1 (x) structure function is the charge-weighted vector sum of the quark polar-
1X
g1 = Qq ∆q (1.8)
2 q
The first moment of the g1 (x) structure function, Γ1 , describes the total spin carried
by the quarks
Z1
Γ1 ≡ g1 (x)dx. (1.9)
0
The Ellis-Jaffe sum rule [10] connects the structure functions to the quark spin
distributions using Equations 1.8 and 1.9. The Ellis-Jaffe sum rule for the proton
5
can be written as
1 5 3F/D − 1
Γp1 = 1+ (1.10)
12 3 F/D + 1
where F and D are the universal weak decay constants. Ellis and Jaffe assumed
that the strange quarks do not contribute to the nucleon’s spin and SU(3) flavor
relation,
1 4 1
Γp1 = ∆u + ∆d = 0.17 (1.11)
2 9 9
which is in disagreement with the calculated results in Equation 1.11 where ∆s was
frozen out. This was a hint that strange quarks might play a role in the properties
of the proton.
Now with these results one can isolate the individual flavor components. The
The minus sign implies that the strange quarks and antiquarks are polarized nega-
tively with respect to the direction of the nucleon spin. This extracted value is to be
taken with care since the strange quark extraction is sensitive to SU(3) flavor break-
ing, and information from neutron beta decay and hyperon semi-leptonic decays had
neutrinos with nucleons to probe the ss̄ sea. The NuTeV experiment [12] at Fermilab
trinos and anti-neutrinos with nucleons in the deep inelastic region. These charmed
particles are produced in d − c and s − c transitions. The neutrinos interact with the
d and s quarks by raising their charge and producing a negative lepton. The d − c
the strange sea. By observing two muon-neutrino events, the NuTeV collaboration
at Q2 = 16 (GeV/c)2 .
From experiment, there has thus been evidence that the strange quark-antiquark
sea may play a significant role in the proton. Much work has gone into extracting
the π − N sigma term to find the strange quark mass contribution, but theoretical
uncertainties still exist. The strange quark-antiquark contribution to the spin of the
proton has been another subject of intense research, but once again the result suffers
ing will offer the best hope in measuring ∆s. This can be technically challenging
since there can be uncertainties in knowing the neutrino flux, the detector efficien-
cies, and nuclear target effects. With all these tantalizing hints, parity-violating
to provide a direct and clean measurement of the strange vector and axial currents
in the nucleon.
7
1.2 Parity-Violating Electron Scattering
In order to access the strange magnetic and electric form factors, parity-violating
electron scattering has been employed. The strange magnetic and electric form fac-
tors, represented by GsE and GsM , are physical observables related to the strange
ing, two different kinds of interactions are involved: the electromagnetic interaction
via the exchange of a photon and the weak interaction via Z 0 exchange. This can
σ∝
FIG. 1.1: Electromagnetic and weak contributions to the scattering cross section in elec-
tron proton scattering.
interaction violates parity, thus by using polarized electrons and forming the ratio of
the difference of polarized cross sections over the sum of the polarized cross sections,
an asymmetry can be formed, which is non-zero only due to the weak interaction.
This asymmetry can be written for longitudinally polarized electrons scattering from
(GeV/c)2 for the G0 experiment), σ+(−) (sometimes this is denoted as σr(l) ) is the
8
cross section for right (left) handed incident electrons scattering from a proton and
α is the fine structure constant. A right (left) handed particle is a particle whose
spin vector is parallel (anti-parallel) to its momentum vector. This is known as the
particle’s helicity. This asymmetry can then be related to the electromagnetic and
weak form factors. The form factors can then be written in terms of quark flavors.
Then by utilizing charge symmetry between the neutron and proton, the asymmetry
can be written in terms of known electromagnetic form factors for the proton and
Measuring asymmetries on the order of 10−6 with errors on the order of 10−7 is a
challenging feat. In order to make a measurement with this precision, a large number
events, ns is given by
dσ dσ
ns = ×L= × I × ρ × L × ∆Ω (1.17)
dΩ dΩ
dσ
where is the differential scattering cross section, I is the beam current, ρ is the
dΩ
density of the target, L is the target length, ∆Ω is the solid angle of the detector
to get as many events as possible. This in turn drives the characteristics common
- long targets
error such as small drifts in some experimental parameter. Forming the measured
systematic errors. Using the variables defined above, but now with a helicity-
correlated systematic error, NHC sys the measured asymmetry, Ameas , can be written
as
the ‘true’ AP V must be controlled since they form a false asymmetry which adds
In order to control systematic errors and increase the number of scattered events
a Pockels cell. By applying different voltages to the Pockels cell, left and right-
wave plate can be inserted and retracted from the laser beam allowing for passive
target. During the transport through the accelerator and experimental end sta-
the beam. Position monitors in an accelerator arc can be used to measure the
tector packages can then be arranged in the experimental end station to detect the
scattered particles.
The full measurement of the G0 experiment will access the strange quark con-
tributions to the magnetic and charge distributions of the proton, G sM (Q2 ), GsE (Q2 ),
11
and the electron-proton axial form factor, GeA (Q2 ). This will be done by measur-
-35 parts-per-million (ppm)1 for the forward angle measurements and the asymme-
be larger, ranging from -18 to -72 ppm. This will allow for a clean extraction of
GsM (Q2 ), GsE (Q2 ) and GeA (Q2 ) with few assumptions (such as charge symmetry).
GsM (Q2 ), GsE (Q2 ) and GeA (Q2 ) and to measure the evolution of these observables
ceptance spectrometer (see Figure 1.2) was built in order to perform the experimen-
coils in a common cryostat that generate up to a 1.6 T field. The diameter of this
energy stored in the magnet is 6.6 MJ. The solid angle is defined by collimators at
the inner diameter of the coils. This geometry allows a line-of-sight shield from the
detectors to the target. Elastically scattered particles of the same Q 2 are focused
onto individual focal plane detectors. Each detector is made of a pair of plastic
scintillators. There are 8 sets of 16 detector pairs called octants that are placed
For the measured asymmetries, which are on the order of parts-per-million, the
1
Due to the large number of abbreviations and acronyms, Appendix A is glossary of many terms
used in this experiment.
12
laboration will extract GsE , GsM , and GeA at the values Q2 = 0.30, 0.50, and 0.80
(GeV/c)2 from the measured asymmetries. The projected errors from this extrac-
tion are shown in Table 1.1 and in Figures 1.3, 1.4, and 1.5. The calculation assumes
700 hours of data taking for the forward angle measurement on the hydrogen target
and for each of the backward angle measurements on the hydrogen and deuterium
inate the overall errors on GsE , GsM , and GeA . The polarization of the electron beam
and GeA . Errors coming from the uncertainties in the electromagnetic form factors
are: 20% for GnE , 3% for GnM , 2% for GpE , and 2% for GpM . A theoretical uncertainty
13
FIG. 1.3: Projected errors on GsE . Result from the SAMPLE experiment is shown along
with the expected errors from the HAPPEX II experiments. Also shown are theoretical
predictions from chiral perturbation theory, lattice QCD, and various pole (dispersion
analysis) models.
had to be included on the isoscaler part of the electron-proton axial form factor,
In the first phase of the G0 experiment, forward angle asymmetries are mea-
sured. This is done by detecting elastically scattered protons between 62 ◦ < θp < 78◦
The 20 cm liquid hydrogen target is based on the SAMPLE [15] design. The tar-
get’s main requirement is that the density remain constant as the beam deposits
up to 500 W of power. The target provides a high longitudinal flow of about 5-10
m/s in order to provide enough mixing by turbulent flow. The spectrometer has an
acceptance of ∼ 0.9 sr defined by collimators at the inner radius of the coils of the
magnet. The measured Q2 range is from 0.1 < Q2 < 1.0 (GeV/c)2 with an incident
beam energy of 3.0 GeV. The scattered particles are detected by pairs of plastic
14
FIG. 1.4: Projected errors on GsM . Result from the SAMPLE experiment is shown along
with the expected errors from the HAPPEX II experiments. Also shown are theoretical
predictions from chiral perturbation theory, lattice QCD, and various pole (dispersion
analysis) models.
scintillators known as the Focal Plane Detectors (FPDs). For each of the eight
detector number. The detectors are shaped into arcs of a circle in order to collect
reject some of the backgrounds. Since it takes about 20 ns for a proton to reach a
detector from the target, the CEBAF machine producing polarized electrons must
reduce the micro-structure of the beam bursts from 499 MHz to 31 MHz in order to
to provide electrons every 32 ns for this experiment. Custom electronics were built
data are recorded by shift registers that feed scalers. This time-of-flight measure-
by the spectrometer and separate elastic from inelastic contributions, allowing for
15
TABLE 1.1: Expected relative contributions of statistical and systematic errors on the
proposed full G0 experiment for GsE , GsM , and GeA . The total absolute error is noted
in the second line of the table. The statistical errors on the forward, backward, and
deuterium asymmetry measurements are denoted by Af , Ab , and Adeut respectively.
versed by changing polarity of the voltage applied to the Pockels cell. As mentioned
above, G0 requires a beam energy of 3.0 GeV, with high intensity (40 µA) and a
pulsed structure (31 MHz instead of the typical 499 MHz) in the forward angle phase
of the experiment. This mode produces a charge per bunch that is 16 times larger
than normal. This requires changes to be made to the beam optics. It is important
FIG. 1.5: Projected errors on GeA . Result from the SAMPLE experiment is shown. Also
shown are theoretical predictions.
be measured. This will be done by reversing the apparatus relative to the beam
line. Elastically scattered electrons will be detected at 110 ◦ from the same 20 cm
liquid hydrogen target. In this phase of the experiment, the background is expected
to be composed of electrons and pions from inelastic processes. In this case, the
time-of-flight cannot discriminate between the different reactions and thus the se-
lection between particles will be obtained from their different trajectories. To aid in
discrimination of elastic and inelastic electrons, another set of 9 detectors, the cryo-
stat exit detectors (CEDs) will be added near the exit window of the cryostat. The
Focal Plane Detector arrays will be reduced to a single layer of 16 scintillators. This
reduction of the FPDs is due to the fact that the CEDs act as the second scintillator.
The decision to keep the front FPD scintillator and not the back FPD scintillator is
17
ing for the back scintillator. The coincidence between the CED-FPD combination
allows for a rough measurement of the electron momentum and scattering angle. To
reject pion background from the entrance and exit foils of the targets, a Cerenkov
detector will be placed between the CEDs and the FPDs. In the backward mode,
programmable logic chips in the electronics will be employed to identify elastic events
from CED/FPD coincidences. The target will also be filled with liquid deuterium
in order to perform a third set of measurements for the extraction of the axial form
factor.
This thesis is based on the work of the First G0 Engineering Run that occurred
in October 2002 through the January 2003. The First G 0 Engineering Run was
a “proof of principle” run. The layout of this thesis is as follows: Chapter 2 de-
describes the analysis of the data from the raw asymmetries to the extracted physics
tic dilution factors, and in Chapter 5, the physics asymmetries are compared to the
Physics Formalism
ing the strange quark vector matrix elements. This is done by measuring the parity-
known as partons). This cluster of quarks and gluons that form the proton can be
approximately written as
p = uud
|{z} + uū
| + dd̄ {z
+ ss̄ + . .}. + g + g + . . .; . (2.1)
| {z }
valence quarks sea quarks gluons
19
20
FIG. 2.1: The proton is made up of three valence quarks, two up and one down quark.
Gluons, the mediator of the strong interaction, are exchanged between quarks to hold the
proton together. These gluons, by virtue of the Heisenberg uncertainty principle, bubble
into quark-antiquark pairs as illustrated by the strange quarks in this figure.
The first term in Equation 2.1 refers to the so-called “valence quarks”, where the
mu ≈ 5 MeV (2.2)
md ≈ 9 MeV (2.3)
The second term is related to the large number of relatively low-momentum quark-
pairs fall off inversely with the mass of the quark species being produced (this is
why the heavy cc̄, bb̄, and tt̄ are expected to play a relatively small role). The
quarks are all held together by the mediator of the strong interaction, gluons. In
fact, it is from these gluons that the q q̄ sea quarks are generated via the Heisenberg
demonstrated in Section 1.1, strange quarks may play a major role in the structure
of the proton and accessing the strange quarks says something about the pure sea
quarks. Sadly, perturbative QCD cannot light the way, since the mass of the strange
λQCD
quark, ms is comparable to the QCD scale, λQCD , where ∼ 1.
ms
21
FIG. 2.2: Tree level Feynman diagram for electron-proton electromagnetic scattering.
with the complex structure of the nucleon in elastic scattering (see the tree-level
4πα µ p
Mγ = − J J . (2.5)
q2 e µ
The electron is a point-like spin- 21 particle; the associated current for the electron is
where u is the Dirac 4-component spinor that describes the initial electron, u is
the Dirac 4-component adjoint spinor of the final electron, and γ µ is the Dirac
gamma matrix. In contrast, the proton, being an extended spin- 21 particle, yields a
more complicated transition current due to its complex structure. Since the electro-
κ = 1.79 nuclear magnetons is the anomalous part of the magnetic moment of the
proton (if the proton were structureless then κ = 0). The momentum transfer, Q 2 ,
is the only variable needed to describe the electromagnetic interaction vertex and
The Sachs form factors are more physically insightful combinations of the Dirac
and Pauli form factors, and allow for a more convenient formalism to be written
without form factor cross terms in the cross section for electron-nucleon scattering.
In the Breit frame, where the exchanged boson (in this case a virtual photon or Z 0 )
is purely space-like (Q2 = 0), this implies that the initial and final momentum of the
scattered particle is equal in magnitude but opposite in sign (~p 0 = p~). These form
factors are closely related to the proton charge and magnetic moment distributions.
The Sachs form factors are related to the Pauli and Dirac form factors by
Q2
where τ = 4MN2 is a measure of the relativistic recoil effects. At Q2 = 0 these form
q0 µp
GpE (Q2 = 0) = GpM (Q2 = 0) = (2.9)
e e/2Mp
where q0 is the proton’s electric charge and µp is the proton’s magnetic moment.
GpE and GpM can be determined from elastic electron-proton scattering experiments.
dσ
In these experiments, the differential cross sections, , are measured at different
dΩ
values of Q2 and lab angle, θ. The Rosenbluth formula [16] which describes the
a function of tan2 ( θ2 ) for constant Q2 . The data should lie along a straight line with
a slope of 2τ (GpM )2 and the extrapolation to τ = 0 will determine the electric form
2
τ GM
e =2
Slop
d Ω d ΩMott
dσ/ dσ
G2E + τ G 2M
Y intercept =
1+τ
tan2 θ
2
FIG. 2.3: Rosenbluth plot showing the linear relationship between (dσ/dΩ)/(dσ/dΩ) M ott
and tan2 (θ/2) at a fixed value of Q2 .
The proton form factors GpE and GpM have been determined by the Rosenbluth
mately follow the so-called empirical dipole form (see Figure 2.4)
−2
Q2
GpE ' GD = 1+ (2.11)
0.71 (GeV/c)2
GpM ' µp GD (2.12)
2.5
2
GpM
Form Factor Value (n.m.)
1.5
1
GpE
0.5
GnE
0
-0.5
-1
GnM
-1.5
-2
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1
2
Momentum Transfer (GeV/c)
FIG. 2.4: Plot of the dipole parameterizations of G pM , GpE , and GnM and the Galster
parameterization of GnE .
unpolarized proton scattering. This technique was originally suggested in the 1970’s
[22]. This polarization transfer method is less prone to some systematic uncertainties
polarized electron beam transfers its polarization to the recoil proton with two non-
zero components, Pt , perpendicular to, and Pl , parallel to, the proton momentum
in the scattering plane. These components are related to the electric and magnetic
form factors
FIG. 2.5: Plot of µp GpE /GpM vs. Q2 for Rosenbluth separation technique (triangle sym-
bols) [23–28], and polarization transfer measurements (blue circles and red squares)
[30, 31] results. The systematic uncertainties of the JLab data are shown at the bottom
of the plot.
disagree with the Rosenbluth separation [23–28] as shown in Figure 2.5, especially
2 2 2 GpE
for Q larger than 1.0 (GeV/c) . At low Q , p ∼ 1 as measured with both
GM
26
techniques, but the ratio falls with increasing Q 2 using the polarization transfer data.
Theoretical work is ongoing to understand the source of this discrepancy [32, 33].
Since G0 operates at momentum transfers less than 1.0 (GeV/c)2 this discrepancy
The neutron form factors are defined in an analogous way. They are more
difficult to measure since there are no free neutron targets. Typically deuterium
and helium targets in quasi-elastic scattering are used and the theoretical and/or
Figure 2.4)
where µn is the neutron’s magnetic moment. GnE is a more difficult situation since the
neutron’s net electric charge is zero. Therefore, in the static limit, G nE (Q2 → 0) = 0.
The small value of GnE at Q2 = 0 compared to GnM makes the Rosenbluth separation
[34] challenging for the neutron. It has been found that G nE is approximated by the
−µn τ GD
GnE = . (2.16)
1 + 5.6τ
Recent experiments using recoil polarimetry [36–38] and polarization transfer [39–41]
without using the Rosenbluth separation. Interference between the magnetic and
related to a ratio of GnE and GnM . These techniques have several advantages, one of
the most important being that many of the systematic errors cancel in these ratios.
Comparing the Rosenbluth results and these recent experiments, they seem to be
in agreement.
27
FIG. 2.6: Tree level Feynman diagram for electron-proton weak neutral current scattering.
scattering has a mixture component that can occur via the weak neutral current
GF
MZ = √ j Zµ JµZ . (2.17)
2
Since the weak interaction is a V-A theory [42], the proton current is more compli-
The weak neutral form factors F1Z and F2Z are analogous to the electromagnetic
form factors F1γ and F2γ . Besides the vector form factors, in weak scattering there
factor, (GZA ). This form factor can be written in general as (see Figure 2.7)
where
γ
PV
Z0 γ
0
Z
FIG. 2.7: Feynman diagrams for contributions to the electron-nucleon axial coupling.
The first diagram describes a single Z 0 exchange. The second diagram describes the
parity-violating photon exchange which contributes to the nucleon anapole moment. The
third diagram is an example of radiative corrections that must be taken into account.
the proton interact weakly while interacting with the scattered photon. This term
Factors
in terms of individual quark flavors. In this way the strange quark content of the
basis) of the proton electromagnetic form factors can be written for each form fac-
tor. This is the sum of the individual quark flavor form factors weighted by the
2 u 1 1
Gγ,p
E = GE − GdE − GsE (2.22)
3 3 3
2 u 1 d 1
Gγ,p
M = GM − GM − GsM . (2.23)
3 3 3
By analogy the same can be done for the weak form factors, but this time the quark
flavor form factors are weighted by the weak charge of that flavor quark:
8 2 4 2 4 2
Z
GE = 1 − sin θW GE + −1 + sin θW GE + −1 + sin θW GsE (2.24)
u d
3 3 3
8 2 4 2 4 2
GM = 1 − sin θW GM + −1 + sin θW GM + −1 + sin θW GsM (2.25)
Z u d
3 3 3
where the parameter sin2 (θW ) is known with a high degree of accuracy. Its on-shell
Assuming charge symmetry between the up and down quarks in the neutron
and proton:
Gu,p d,n
E = GE Gu,p d,n
M = GM (2.27)
Gd,p u,n
E = GE Gd,p u,n
M = GM (2.28)
allows for the u and d quark contributions to be eliminated. The weak form factors
GsM = µs (2.31)
GsE = 0 (2.32)
where µs is the strange magnetic moment of the nucleon in terms of the nuclear mag-
neton. Relation 2.32 follows from the fact that the nucleon has no net strangeness.
the charge form factor, there must be a spatial polarization between strange and
anti-strange quarks. In order to contribute to the magnetic form factor, the strange
which allows one to write an expression for the SU(3) flavor singlet form factor
G0E,M (and hence the name of this experiment). This form factor characterizes the
difference between the corresponding electromagnetic and weak form factors for the
and 2.28) between the up and down quarks in the neutron and the proton demands
a closer look. Under this charge symmetry, the up and down quark wave functions
in the proton describe the down and up quark wavefunctions in the neutron. This
charge symmetry is broken by the mass differences between the up and down quarks
and from electromagnetic effects. This charge symmetry breaking manifests itself
gated using non-relativistic quark models [46]. In these non-relativistic quark mod-
els, the largest effect has been calculated to alter the values of the electromagnetic
Scattering
In order to measure the strange electric and magnetic form factors of the proton
it is necessary to measure the weak neutral form factors. Since the electromagnetic
interaction dominates over the weak interaction, direct measurement of the weak
neutral form factors is challenging. In order to observe the weak neutral form
the unpolarized nucleon. A polarized electron can come in two states: right handed
electrons (denoted by a ‘+’ or ‘r’) are electrons whose spin and momentum vectors
are parallel and left handed electrons (denoted by a ‘-’ or ‘l’) are electrons whose
spins and momentum vectors are anti-parallel. The scattering interaction can occur
via the exchange of a virtual photon or a virtual Z 0 . This gives rise to two scattering
M = M γ + MZ . (2.36)
Since the weak interaction violates parity, its scattering amplitude, M Z , will be
formed by taking the difference over the sum of helicity dependent scattering cross
sections, σR and σL
σR − σL
A= . (2.38)
σR + σL
where
1
= (2.42)
1 + 2(1 + τ )tan2 ( 2θ )
p
0 = τ (1 + τ )(1 − 2 ). (2.43)
The electromagnetic nucleon form factors have been measured and are fairly well
known (see Section 2.2) and these values can be used as inputs in the above equation.
of weak neutral form factors. Three independent measurements are needed for a
while backward angle measurements are most sensitive to large 0 and small . Note
33
that the axial term is suppressed relative to the vector electric and magnetic terms
because of the factor (1 − 4sin2 θW ) ≈ 0.08. For the third measurement to get to
off of deuterium.
as a linear superposition of the single nucleon observables. This allows one to write
the asymmetry as
σ p Ap + σ n An
AQE = (2.44)
σd
asymmetry on the proton Ap is given by Equation 2.41 and the asymmetry on the
neutron is given by
−GF Q2 Gγ,n Z,n γ,n Z,n 1 2 0 γ,n Z,n
E GE + τ GM GM − 2 (1 − 4sin θW ) GM GA
An = √ . (2.45)
4 2πα (Gγ,n 2 γ,n 2
E ) + (GM )
Effects associated with the deuteron wave function and different potential models
have been explored in [47] and were shown to be quite small. Corrections for final
state interactions and exchange currents must be taken into account for a reliable
separation of the axial and magnetic form factors. These issues have been addressed
in [48].
Asymmetry
Using Equations 2.41 and 2.29, the parity-violating asymmetry, to leading or-
der, can be written in terms of strange (unknown), axial (unknown), proton (known)
34
and neutron (known) form factors. This asymmetry equation can be written as
−GF Q2 1
A = √ 2 γ,p (2.46)
4 2πα (G )E + τ (GM )2
γ,p
and GeA :
where
−GF Q2 1 γ,p 2 γ,p γ,n
η = √ γ,p 2 γ,p 2 GE ((1 − 4sin θW )GE − GE )
4 2πα (G ) + τ (G M )
+ τ GM ((1 − 4sin θW )GM − Gγ,n
γ,p 2 γ,p
M ) (2.48)
γ,p
−GF Q2 GE
ξ = √ γ,p (2.49)
4 2πα (G )2 + τ (GM )2
γ,p
−GF Q2 τ Gγ,p
M
χ = √ γ,p (2.50)
4 2πα (Gγ,p )2 + τ (GM )2
−GF Q2 (1 − 4sin2 θW )0 Gγ,p
ψ = √ γ,p . (2.51)
4 2πα (Gγ,p )2 + τ (GM )2
even if the strange quarks do not contribute to the properties of the nucleon. This
and axial form factors of the nucleon and by the electroweak parameter sin 2 (θW ).
HAPPEX I [14] was an experiment that ran in 1998 and 1999 at Jefferson Lab.
electron beam scattered off an unpolarized 15 cm long liquid hydrogen target. In the
first run, the beam current was 100 µA with an average beam polarization of 38.8 ±
2.7% from a bulk GaAs polarized source. In the second run, the beam current was 35
µA with a beam polarization of about 70% from a “strained” GaAs polarized source.
Two spectrometers in Hall A with a small acceptance ∆Ω = 5.5 msr detected the
scattered electrons at this extreme forward angle. The detected signals were inte-
grated from a lead/lucite scintillator calorimeter which only accepted elastically scat-
ppm. Due to limited kinematics, HAPPEX I was unable to perform the Rosenbluth
The SAMPLE experiment [13] was performed at the Bates Linear Accelerator
(deuterium) target. The beam polarization was about 37% from a bulk GaAs po-
larized electron source. The scattered electrons were detected in a large solid angle
130◦ to 170◦ . Since measurements were made at backward angles, the asymmetry
is most sensitive to GsM and GeA . The scattered electrons with an average Q2 ∼ 0.1
(GeV/c)2 were detected by Cerenkov light produced in air absorbed by ten photo-
multiplier tubes via ten mirrors positioned around the beam axis. The measured
36
FIG. 2.8: SAMPLE uncertainty bands of GsM vs. GsE at Q2 = 0.1 (GeV/c)2 . Also
shown is the uncertainty in a theoretical calculation of G eA by Zhu [49] at the same
momentum transfer. The smaller ellipse (yellow) corresponds to a 1σ overlap between
the SAMPLE results on hydrogen and the theoretical calculation. The larger ellipse (ma-
genta) corresponds to a one sigma overlap between the SAMPLE hydrogen and deuterium
results.
asymmetry for the proton was A = −5.61 ± 0.67 ± 0.88 ppm. Making measurements
the electron-proton axial form factor, GeA = −0.83 ± 0.26. This in turn, at the
SAMPLE kinematics, allows one to isolate the strange magnetic form factor. The
FIG. 2.9: The PVA4 (represented as ’A4’ in the plot) results are shown as a solid line
representing all possible combinations of GsE + 0.225GsM at Q2 = 0.230 (GeV/c)2 . The
densely hatched region represents the PVA4 uncertainty. The HAPPEX result is shown
as a dashed line representing all possible combinations of G sE + 0.395GsM at Q2 = 0.477
(GeV/c)2 . The less densely hatched region represents the HAPPEX uncertainty.
The PVA4 experiment [51] was performed at MAMI at Mainz. It measured the
electrons between 30◦ and 40◦ were detected by a large acceptance (∆Ω = 0.62
sr) calorimeter. This calorimeter consists of 512 PbF2 crystals. The beam current
polarized electron source. The polarization was measured by both Compton and
38
Moller polarimeters. The measured asymmetry was A = −5.44 ± 0.54(stat) ±
0.26(syst). The extracted linear combination of the strange and magnetic form
This result is 1.2σ away from zero. It should be noted, with caution, that [51]
averages the PVA4 result (Q2 =0.230 (GeV/c)2 ) with the HAPPEX result (Q2 =0.477
(GeV/c)2 ) at different kinematics. They use a value for GsM based on theoretical
estimates and SAMPLE’s result to obtain GsM (0.1 < Q2 < 0.5 (GeV/c)2 ) = 0.066 ±
0.26 which leads to a non-zero contribution of strange quarks to the strange electric
form factor.
CHAPTER 3
Experimental Apparatus
The G0 engineering run was performed in Hall C at Jefferson Lab in the Fall of
2002 through January 2003. A second engineering run is scheduled for the Fall 2003
with the forward production run occurring in early 2004. This chapter will describe
and polarimetry.
apparatus. The purpose of the magnet is to bend elastically charged particles from
the target of the same momentum onto the focal plane detectors, independently of
the interaction point along the target length (see Figure 3.1). The magnet produces,
coils in a single cryostat. Each coil is made from 144 turns of integrated supercon-
ductor. The superconducting coils in series are cooled by four parallel liquid helium
convection circuits. Two additional parallel cooling paths are used to cool the su-
perconducting electrical buss through which power is supplied to the coils. The
39
40
FIG. 3.1: Elastically scattered protons, of the same momentum, are focused onto the
Focal Plane Detectors (FPD). This is independent of where along the target the scatter-
ing occurs. This is shown schematically by looking at three different momentum values
(denoted by red, green, and blue) at opposite ends of the target being focused on the de-
tectors. The collimators are used to reduce the background and to set the acceptance of
the detectors.
coils, electrical buss, and collimators (which define the spectrometer acceptance and
provide shielding for the target and detectors) make up what is called the cold mass.
This sits inside a stainless steel shell and makes up the bulk of the magnet. A liquid
nitrogen shield surrounds the cold mass. The cryogens, lN 2 and lHe are fed into
the magnet via a manifold at the bottom of the magnet from reservoirs located in
the control dewar at the top of the magnet. The cryogens percolate back up to
the reservoirs through the cooling circuit, losing density and absorbing power along
the way. Aluminum end caps cover both the front and back of the main shell that
houses the cold mass. There are eight trapezoidal holes, 0.51 m 2 in area, on the
downstream end cap that are covered by an 0.020 inch titanium plate. These are
the exit windows that provide a path of low energy loss and multiple scattering from
particles emanating from the target to the detectors. Mounted to the beam line at
the upstream magnet end cap is the target service module that contains the target
and its positioning mechanism. Valves are connected upstream of the target service
41
module and at the downstream end of the exit beam line to be isolated. When these
valves are closed, the entire experimental apparatus can be disconnected and moved
to the left of the beam line. This is necessary since the G 0 experimental program
will take several years to complete and will not be the only experiment running in
Hall C, thus the G0 apparatus must be able to be removed from the Hall C beam
line.
FIG. 3.2: A schematic of the G0 target. The G0 target has been designed to minimize
false asymmetries. The target cell is 20 cm long and filled with liquid hydrogen. The
target operates at 450 watts.
The G0 experiment uses a liquid hydrogen target (see Figure 3.2) based on the
SAMPLE experiment’s design [15]. The target is optimized to reduce energy loss
along the scattered particles’ path and to minimize density fluctuations, caused by
cryogenic loop to recirculate and cool the liquid hydrogen. The target and cryogenic
loop sit inside the 77K liquid nitrogen shield of the SMS. The hydrogen cell is 20
cm long and 5 cm in diameter. The radius of curvature of the endcap of the target
is 7.6 cm. The outer wall and endcap are 7.0 ± 0.5 mils of aluminum. A manifold
42
inside the lH2 cell has been designed to direct the fluid flow down the center of the
target cell and back near the cell walls. The target is fronted by a helium cell that
by matching the radius of curvature of the entrance and exit windows of the
hydrogen cell.
• It extends the hydrogen cell beyond the part of the cryoloop which is not sym-
metric about the beam axis allowing the target-beam interaction region to be
axially symmetric.
The heat exchanger uses gaseous helium with an inlet temperature of about 15 K
and pressure of 20 atm, supplied by the Jefferson Lab End Station Refrigerator.
With a beam current of 40 µA impinging on it, the target requires a flow of 17 g/s
of 15 K coolant. The coolant flows inside the finned tube and liquid hydrogen flows
over the fins on the outside of the tubing. The heat exchanger has 2 layers of finned
tubing through which the liquid hydrogen flows in parallel. The heat exchanger
removes about 450 W of heat from the target with 250 W coming from 40 µA beam
heating, 100 W from the pump motor and 100 W from connections to the outside
world.
Two internal heaters in the G0 cryogenic target are used to maintain a constant
heat load and/or temperature in the target. The high-power heater consists of
three heater coils in parallel. The high-power heater is a feedback loop that reads a
signal proportional to the beam current and calculates the heat load of the beam,
and then changes the heater such that there is a constant heat load on the target
and thus ensures a constant temperature in the cell and coolant loop. A second
constant target temperature against small time-dependent drifts from other heat
43
load sources (such as the motor, coolant inefficiencies and radiant heating). The
Target Cell
FIG. 3.3: Schematic of the dummy target looking downstream of the beam. There are
three dummy targets located above the target cell. Two of these dummy targets (large and
small hole) are used in studying beam halo. The third dummy target is a 12 C radiator
target.
target is mounted onto an aluminum frame that is 3.05 mm thick. On this aluminum
frame there are three dummy targets: two blank hole targets for studying beam halo
12
and one C target (see Figure 3.3). The big hole target has an inner diameter the
same as the hydrogen target but with an outer diameter of 19.05 mm and a thickness
of 3.912 mm to match the radiation length of 1 inch of aluminum. The small hole is
12
5.46 mm diameter. The C target sits behind a hole of 9.562 mm diameter in the
12
aluminum frame. Five slabs of ∼ 1 mm thick carbon make up the 4.9 mm thick C
target. The entire target can be removed from the beam trajectory for diagnostic
One issue with the target that must be addressed in a parity-violation experi-
ment is that of target boiling. If the target boils (e.g. density fluctuations due to a
high amount of energy being deposited on the target by the electron beam) this will
add noise to the experimental system. This noise will then increase the statistical
error of the experiment. This additional noise from target boiling, (σ b2 ), is added in
44
Asym_width_raw (ppm) σ0 : 299.371 +/- 7.183, σρ:612.091 +/- 10.855, exp: 1.614 +/- 0.067
σ0 : 326.193 +/- 3.343, σρ:580.686 +/- 10.024, exp: 2
800
600
400
1 1.5 2 2.5 3
Raster (mm)
FIG. 3.4: Fluctuations in the statistical width (ppm) as a function of raster size for target
boiling studies. These data come from measurements with the LUMI detector.
where the overall statistical width is given by σ 2 . The beam’s noise σb2 is parame-
σρ2
terized by rx
where r is the raster size and x is the exponent which can either be
fixed at 2 or allowed to be variable. The LUMI monitors (see Section 3.7) are used
to measure the target boiling because they are at extreme small angles relative to
the beam line. This means that the LUMIs see an electroweak asymmetry of zero
and their detected rate is very high allowing one to make faster measurements than
with the Focal Plane Detectors. A test that is performed is to boil the target by
depositing more energy on the target by changing the raster (area) of the electron
beam (see Figure 3.2). The LUMIs are then used to measure the contribution to
the statistical width from the target. Two separate fits were performed to the data
and the noise contribution at the G0 raster size of 3 mm was found to be about
45
300 ppm. This is small compared to the statistical width due to counting statistics
(found to be about 1400 ppm averaged over all the detectors). Taking the ratio
of the measured statistical width (σmeas ) compared to the statistical width from
counting statistics (σstat ) and using the fact that the noise contribution from the
This implies that the error has been increased by 5% past counting statistics (and
The G0 detector system consists of eight octants surrounding the beam line
(see Figure 3.5). Four octants were designed and mounted by French collaborators
and four by North American collaborators (see Figure 3.6). Each octant encloses
an array of 16 scintillator pairs along the focal plane of the G 0 magnet. Each end
These detectors are named the focal plane detectors (FPDs). The FPDs are the
only detectors necessary for the forward angle measurement in the G 0 experiment.
In this configuration, the scintillators detect elastically recoiled protons. The output
PMT signals are used to measure the Time-of-Flight of the particles between the
target and the focal plane of the magnet, providing particle identification. Each
FIG. 3.5: The G0 detector system, supported by an aluminum support known as the
‘Ferris Wheel‘, as seen upstream on the left picture and downstream on the right picture.
The octants are numbered 1 though 8 starting from the top and moving clock-wise. Odd
(even) numbered Octants are North American (French). The G0 target and magnet are
not pictured here; they were placed directly upstream of the Ferris Wheel.
Detector System
plane of the G0 magnet. They are labeled from 1 to 16 where the larger scintillator
number corresponds to the scintillators located further away from the beam line and
further from the target. Larger scintillator number also corresponds to the larger
scintillator size and, in general, the larger measured value of Q 2 (with detectors 14
it is used to measure background). The FPDs have been divided into 16 detectors
in order to keep the individual count rate for each detector around 1 MHz. The
arc shape of the scintillator was defined to follow the iso-Q 2 phase space. The
47
FIG. 3.6: Pictured on the left is the North American Octant 7. On the right is pictured
the French Octant 2. Both octants are shown without their light tight box.
dimensions and areas of the scintillators are given in Table 3.1. The scintillators are
paired together to increase background rejection, moreover two scintillators from the
same pair are separated by a thin material (aluminum for French and polycarbonate
for NA). Each FPD is a pair of two light pipes attached to the aluminum support
of one octant. Each light pipe is the assembly of a scintillator and two light guides
glued at each of the ends of the scintillator with photo-multiplier tubes attached to
each end of the light guides. Optical fibers are air coupled to each end (right and
left) of the scintillator. The fibers shine UV light to the scintillator and allow gain
monitoring of the PMTs. The whole assembly is enclosed inside a light-tight box.
Scintillator 1 2 3 4 5 6
Length (cm) 60.1 61.3 62.1 62.6 64.6 69.2
Length (ns) 4.4 4.3 4.3 4.4 4.3 4.8
Area (cm2 ) 177.8 237.1 254.0 238.9 293.5 353.9
Thickness (cm) 0.5 0.5 0.5 0.5(f)-1.0(b) 1.0 1.0
Scintillator 7 8 9 10 11 12
Length (cm) 74.7 80.8 88.6 95.8 104.5 112.3
Length (ns) 5.6 5.1 6.0 6.4 7.0
Area (cm2 ) 441.6 441.8 581.2 689.5 878.2 864.5
Thickness (cm) 1.0 1.0 1.0 1.0 1.0 1.0
Scintillator 13 14 15 16
Length (cm) 120.1 136.8 135.0 136.0
Length (ns) 8.0 9.5 9.3 9.3
Area (cm2 ) 871.4 1146.7 1209.9 1218.8
Thickness (cm) 1.0 1.0 1.0 1.0
TABLE 3.1: Lengths and areas of the NA G0 scintillators. The scintillators are arc
shaped. The length (in centimeters) refers to the perimeter of the inner arc. The statis-
tical precision of the measurement (in nanoseconds) is 0.3 ns. Note that the size of the
scintillator is not the same for French and North American octants. Also note that the
thickness of scintillator 1 to 3 is 0.5 cm. The thickness of scintillator 4 back (b) is 0.5
cm, the thickness of scintillator 4 front (f ) is 1 cm, and the thickness of scintillator 5 to
16 is 1 cm.
49
Light Pipe
The North American light guides are made out of Lucite. The French light
guides are made of Polymethyl Matacrylate (PMMA). The North American de-
tectors used Philips XP2262B (12 stages) PMTs powered by custom made Zener-
resistor bases [52]. The French detectors used Photonis XP2282B04 (8 stages) PMTs
powered by a custom made Zener base with an amplifier with a gain of 20. The
light pipes are wrapped in µ-metal to reduce any remaining fringe magnetic field
Light-Tight Box
The North American light pipes are individually wrapped in aluminized mylar.
The French light pipes are wrapped in aluminum foil. These light pipes are not
light-tight. For operating the PMTs, the assemblies are enclosed in a light-tight box
consists of aluminum plates on four sides with a black cover on the remaining fifth
side. The inside of the octant support as well as the aluminum plates have been
covered by black Tedlar to minimize reflection of light leaks. The front fifth side of a
No access is possible to the inside of the octant without destroying the integrity of
The North American and French octants “plug in” to a detector superstruc-
ture known as the “Ferris Wheel” (see Figure 3.5). Within the Ferris Wheel, the
North American octants are located at the 12 o’clock, 3 o’clock, 6 o’clock and 9
o’clock positions. The French octants are located in between the North American
50
octants. Each octant is attached to the Ferris Wheel by means of three bolts on the
downstream face of the structure. The Ferris Wheel is about 7 m above the floor
at its highest point with the symmetry axis (the beam line) 4 m from the floor.
The detectors are shielded from the beam line by 9.525 cm of lead and 15.716 cm
of poly-boron. A rail system permits the Ferris Wheel to be retracted from the
Superconducting Magnet System. This is done so that the G 0 apparatus can easily
be moved out of the beam line so other experiments in Hall C can be performed.
CEBAF 499 MHz. This produces a charge bunch that is 16 times larger than the
normal operating mode bunch. These high charge bunches require special beam
optics due to space charge effects. Each micropulse contains on average 1.28 pC,
In order to measure the small asymmetry between the two helicity states the
same experimental conditions must exist for the two different helicity states. Devi-
ations in the experimental conditions between the two helicity states can induce a
false asymmetry.
the electron-proton scattering cross section. This cross section is sensitive to energy,
beam position, and beam angle. Any systematic difference between these properties
for the two helicity states can manifest itself as a false asymmetry.
the circularly polarized laser light emerges from the Pockels cell. The beam will not
be perfectly circularly polarized but will be elliptically polarized due to some residual
linear component; this, coupled to the analyzing power of the cathode, produces
51
TABLE 3.2: The measured run averaged helicity-correlations for January 18-26, com-
pared to the G0 specifications for 700 hours of data taking.
to spatial variations in the quantum efficiency of the cathode will also produce
Charge asymmetries1 must also be minimized since charge asymmetries can in-
accelerator.
The G0 experiment has specified how large these helicity-correlated beam dif-
ferences can be allowed to be (see Table 3.2). This is because the size of these
measured. This in turn determines the errors on the correction procedures to cor-
pulsed at 31.2 MHz rather than the typical CEBAF 499 MHz. This translates
into buckets of polarized electrons arriving every 32 ns at the target. The helicity
of the beam is flipped at 30 Hz. One illustration of why this time scale is chosen
1 Q+ −Q−
The charge asymmetry is defined as Aq = Q+ +Q−
52
A Quartet(+--+) A Quartet(+--+)
+ - - + - + + -
1/30 Hz = 33 ms 500 µ s
FIG. 3.7: The G0 beam structure. The G0 beam is pulsed at 31.2 MHz. The G0 beam
flips helicity every 30 Hz, known as a macropulse (MPS). The G0 beam is divided into
quartets of four MPSs. The first MPS is chosen pseudo-randomly with the next two MPSs
the complement of the first MPS and the fourth MPS is the same as the first MPS.
to oscillate up and down at 60 Hz. Then during one half of the cycle more counts
would be collected than in the other cycle. If the spin flipping was not a multiple
of the 60 Hz period, there would be non-statistical fluctuations from one data col-
lection period to the next. By holding each helicity state for 1/30 s, the experiment
averages over any changes in the experiment caused by the 60 Hz noise and any
frequencies typically populated by noise from electronics and lower frequencies sen-
as a macropulse (MPS). The helicity is reversed during a 500 µs interval (see Figure
3.7). The macropulses are grouped into quartets with the sign of the first MPS
being chosen pseudo-randomly. The next two MPSs are the helicity complement of
the first MPS. The last MPS helicity in the quartet is the same as the first MPS
helicity. The beam helicity is flipped quickly to insure that short-lived changes in
Since there is a 500 µs settle time for the Pockels cell, this scheme allows the
53
phase of each new helicity states to slip with respect to the 60 Hz power cycle.
This phase slip is beneficial to the experiment. This will allow the helicity states to
precess through all the phases and thus sample different aspects of the 60 Hz line
noise.
This section describes the various pieces of equipment that are in the injector
and associated with the polarized source. A description of the helicity devices on
the polarized source laser table in the injector are covered along with a description
of the performance of the feedback systems that use these helicity devices. The G 0
TIGER laser used to excite the photo-cathode along with the GaAs photo-cathode
The G0 laser, known as the TIGER, is a Ti-Sapphire laser. Ti:Sa lasers can
achieve both high current and high polarization, unlike the typical diode lasers
used at CEBAF that can achieve either high current or high polarization but not
has passive mode-locking to reference the 31.2 MHz RF source. The TIGER laser
produced more than 300 mW of power at a wavelength of 840 nm. The TIGER
FIG. 3.8: Schematic of the CEBAF polarized source laser table in the injector. This
schematic shows the path the laser light produced by the G0 TIGER laser follows. Im-
portant optical elements to note are the IA cell used for controlling intensity differences
and the PZT used to control helicity-correlated position differences. There is a quadrant
photo-diode (QPD) that is used to understand the optical elements upstream of the QPD.
The insertable halfwave plate is a device to passively change the handedness of the laser
light and thus used as a diagnostic for understanding helicity-correlated noise in the sys-
tem. The helicity Pockels cell is used to flip and produce circularly polarized light before
striking the photo-cathode.
The Intensity Attenuator (IA) cell is the central device in the injector used
by the charge feedback system from Hall C to null the charge asymmetry without
modulator (see Appendix B for more details). This device modulates the laser light
broadband anti-reflective coating at 850 nm. The Pockels cell is between two parallel
linear polarizers followed by bare mica model WPUM-10-850 tenth-wave plate for
Linearly-polarized light from the laser is transmitted through the first linear
polarizer, then is transmitted through the tenth-wave plate. This produces slightly
low-voltage Pockels cell (operating at about 50 V in one helicity state). If the Pockels
the cell. The light is then transmitted through the second polarizer (parallel to the
first polarizer) and linear light is produced. If the IA Pockels cell is at some non-zero
voltage, more elliptically polarized light emerges from the Pockels cell. When this
elliptically polarized light is transmitted through the second parallel polarizer only
the linear component will survive, thus modulating the intensity of the laser beam.
The IA cell responded well during the commissioning, offering a large (∼ 400
ppm/V) intensity difference calibration, though it did generate large position dif-
ferences for reasons that were not always clear (see Table 3.3). The IA calibration
constants were fairly stable over time, but did require new measurements of this
constant roughly every couple of days, or whenever the insertable halfwave plate
was inserted or retracted. The main problem with the IA cell was variations in time
of the calibration constant (ppm/V). This was caused by the IA cell inducing po-
sition differences that created charge asymmetries at apertures in the injector (see
Figure 3.9). This occurred as the IA would steer the beam in a helicity-correlated
way causing the electron beam to clip along the sides of apertures in the injector.
IA Scan Comparison
IB=38 µA 12/15/02
1000
IB=20 µA 12/15/02
IB=20 µA 12/17/02
800
IB=40 µA 12/12/02
600
400
200
FIG. 3.9: The IA calibration constants (ppm/V) as a function of beam position monitor
in the injector to Hall C. A quadrant photo-diode (QPD) is placed on the laser table
to determine what is happening on the laser table. Note that the values from 1i02 to
1i06 are consistent with one another. Between 1i06 and 0L02 are two apertures. It is
hypothesized that, due to scraping on these apertures, the calibration constant did not
remain the same between the 1i06 and 0L02 regions.
Intensity Feedback
The automated intensity feedback uses the IA cell (see section 3.5.2). It was
used to vary the laser intensity in a helicity-correlated way to insure that the helicity-
Monitor) is kept small. BCM1 is used to measure beam currents above 20 µA and,
another BCM, BCM2, is used to measured beam currents from 5-20 µA. Typically
a change to this feedback system was made every 5 minutes (see Figure 3.10).
The nature of the feedback system is to cancel the contributions arising from the
statistical jitter of the charge asymmetry. In the absence of feedback, the statistical
reduction in the charge asymmetry goes like √1 where N is the number of 5 minute
N
measurements. With the charge asymmetry feedback active, the charge asymmetry
57
TABLE 3.3: Table of the IA calibration slopes measured in Hall C. Overall, the IA
feedback was moderately stable during the engineering run. Entries marked ’N/A’ do not
have data available.
1
convergence goes like N
.
The rotating halfwave plate (RHWP) is also used to minimize the intensity dif-
ference. The halfwave plate operates on the principle that the light from the Pockels
cell will not be perfectly circularly polarized, but will instead be elliptically polar-
ized. Elliptically polarized light can be decomposed into two components: circular
polarized and linearly polarized. The RHWP takes the residual linear component
of the light and rotates it with respect to the cathode depending on the adjustable
rotation angle (see Figure 3.11). Scans of the halfwave plate are done on the order
of days (versus the IA, which is used to minimize the intensity difference on the
order of minutes using an automated feedback system) to find the minimum inten-
sity difference (known as the “sweet spot”). From results in Hall C, the setting for
the sweet spot of the RHWP is very stable and the null setting only needed to be
FIG. 3.10: Typical performance of the charge feedback system using the IA cell. The
top panel shows the intensity difference as measured every 5 minutes. The bottom panel
shows the run-averaged intensity difference converging to zero as time goes on.
PZT
The PZT is a device used to correct for helicity-correlated beam position. This
device is a mirror mounted on a Thor Labs KC1-PZT kinematic mount. The PZT
can be set to different independent positions in both the X and Y directions. The
PZT oscillates between these set positions and the null position at the helicity-flip
frequency (each 1/30 s). This allows any helicity-correlated position differences in
the electron beam to be corrected by changing the angle that the laser beam strikes
15000
10000
Intensity Asymmetry (ppm)
5000
-5000
-10000
-15000
Another issue with the PZT was related to the orthogonality of the PZT X and
PZT Y motion. If this motion is not orthogonal, one may not be able to effectively
and independently correct for x and y position differences. For example, if the PZT
X and PZT Y motion was exactly at 180◦ then it would be impossible to correct for
both X and Y motion. Note that rotations are fine since some combination of PZT X
and PZT Y would still correct for X and Y beam motions. Hence the calibration of
the PZT requires not just measuring the calibration slope for PZT X in X (denoted
as XX) but also what affect PZT X has on Y (denoted at YX), motion and how
PZT Y not only affects Y (denoted at YY) motion but X motion (denoted as XY)
of the beam. Table 3.5.2 lists the calibration slopes of the PZT.
60
TABLE 3.4: Table of the limited number of PZT calibration slopes in nm/V. The PZT
did not perform well during the engineering run. XX (YX) denotes the affect of PZT X
on the X (Y) beam motion. YY (XY) denotes the effect of PZT Y on the Y (X) beam
motion.
Position Feedback
This feedback system was intended to be used to vary the position of the laser
position monitors G0 and G0B is kept small. The PZT’s calibration constants were
unstable on time scales of the order of minutes during the 2002-2003 G 0 commis-
sioning run. This meant that the position feedback to control the helicity-correlated
position differences was not used. Even though the position feedback was not used,
the measured mean position differences (∆X and ∆Y) were on the order of 50 nm
with a σ of 6 µm.
the helicity are located in the Injector Service Building above the polarized source.
The helicity of the electrons is determined by the helicity Pockels cell. The G 0
Helicity Controls adjusts the Pockels cell, thus controlling the helicity of the electrons
injected into the accelerator. The make-up of the G 0 Helicity Digital Controls
include a pseudo-random bit pattern stored in two 1Mbit EPROM’s and control
code, which includes multiplexers, registers, counters, and a state machine [53].
61
The output of the G0 Helicity Digital Controls is either 1 or 0. The helicity state of
the electrons in turn is based on whether the output from the G 0 Helicity Controls
helicity signal from the helicity electronics is fed to the G 0 electronics and DAQ in
Hall C via a fiber optic cable. Furthermore, this signal sent to Hall C is delayed by
eight helicity windows, thereby ensuring no in-time helicity correlation between the
3.5.4 Photocathode
FIG. 3.12: A schematic of the strained GaAs band structure and energy level diagram.
The circled numbers indicate the relative transition strengths.
The strained GaAs crystal acts as the photo-cathode for the polarized source.
GaAs is a direct band-gap crystal, which means that the valence band maximum
and the conduction band minimum are aligned in momentum space, allowing for
optical transitions between the energy bands that follow the angular momentum se-
lection rules for optical transitions in atoms. These transitions are shown in Figure
3.12. The strained layer GaAs photo-cathodes are produced by Bandwidth Semi-
an electron in the valence band (P3/2 and P1/2 levels) absorbs a circularly polarized
photon and is excited to the conduction band. The crystal has been specially treated
with Cesium Fluoride to increase the quantum efficiency of the cathode by reducing
the work function of the GaAs. This also produces a negative electron affinity to
allow the electrons to escape from the crystal. A typical quantum efficiency for the
photo-cathode is about 1% for light with a wavelength of 780 nm. The cathode is
held at a bias voltage of -100 kV in order to liberate the electrons from the cathode.
The maximum polarization of the strained GaAs crystal is about 80% compared
to typical polarizations of 40% for bulk GaAs crystal. This is done by growing a
thin layer of about 100 nm of GaAs on a substrate of GaAsP, which breaks the
An issue with the strained GaAs crystal is that a large quantum efficiency
anisotropy is produced, which in turn can produce a charge asymmetry. The light
emerging from the Pockels cell is not typically perfectly circularly, but has a resid-
ual linear component. The axis of the residual linear component of the light can
be different for the two helicity states. In this case there will be two different ori-
entations of the light’s polarization axis with respect to the “strain axis” of the
crystal. Because these axes are different, the number of electrons liberated from the
The crystal is located in the electron gun and is kept under very high vac-
uum. Still, residual gases can contaminate the crystal surface thereby lowering the
“bake-out”. This is when the temperature of the gun is raised until the contaminated
The Continuous Electron Beam Accelerator Facility at Jefferson Lab can deliver
up to 5.5 GeV of 200 µA polarized electron beam. A Wien filter in the injector sets
the launch angle of the polarization vector to compensate for g-2 precession, thereby
polarized electrons are injected at 45 MeV into the main accelerator. The electrons
are accelerated in two linacs that make up the straight portions of the accelerator
racetrack. The linacs are connected by recirculating arcs located at both ends of
frequency cavities that operate at 1497 MHz. The electron beam can take up to
5 passes around the accelerator before being sent to one of the three experimental
halls. This is done by using an electromagnetic “kicker” to send every third electron
bunch to the appropriate hall, resulting in a beam structure of 499 MHz in the halls.
Hall C is equipped with three different beam current monitors (BCMs). Two
of these BCMs, BCM1 and BCM2, are cylindrical resonant cavities [54, 55]. The
current is monitored by using the electron beam to excite the resonant modes in the
cylindrical waveguide. The cavity is sensitive, by design, to the TM 010 mode. Inside
the cavity is a loop antenna which couples to the resonant modes. The measured
signal is proportional to the beam current. In addition to the two cavity monitors,
Hall C is also equipped with a parametric current transformer (also known as the
Unser) [56]. The Unser has a very stable and well-measured gain but it suffers from
large unstable offsets. The Unser is not used to measure the beam current but due
to its stable gain it is used to calibrate the BCMs. The noise in BCM1 is found to
64
be about 300 ppm which is much smaller than the statistical width of the measured
The beam position monitors (BPMs) used during the first engineering run in
both the injector and Hall C were standard strip-line BPMs [57,58]. These BPMs are
The BPMs operate at 1500 MHz and inductively pick up the RF signals of the
electron beam as it passes through the device. The signals are then amplified and
down-converted to 1 MHz. The signals for beam position can then be computed
knowing that the signal in the antenna is proportional to the beam position and
beam intensity:
of each antenna may be different, αX is a measure of the different gain between the
The position (or position difference) calculated from the beam monitors has a
certain amount of noise associated with it. This noise is due both to beam noise
The instrumental noise can be found by using three BPMs along the beam line
without magnetic optics between the monitors. The first two BPMs can be used to
determine the position of the beam in the third monitor. This predicted behavior
can then be removed from the measured signal of the third BPM leaving behind
only the instrumental noise. The noise of the BPMs was found to be about 2 µm.
FIG. 3.13: Plot of normalized rates as a function of run number for the 2003 G 0 Engi-
neering run as measured by the Lucite Halo Monitor. The red (blue) points correspond
to the halo target ‘in’ (‘out’) data. The baseline value for ’good’ beam is approximately 5
Hz/µA)
66
Electrons outside the core of the electron beam are defined as the beam halo.
This halo could be produced in a variety of processes. The halo could be formed
in the beam line, etc. This halo can cause numerous troubles. These halo electrons
could cause an increase in the dead time by producing higher singles rates in the
G0 detectors from inelastic scattering and they could also contribute to the inelastic
background. This higher unwanted radiation could also cause radiation damage to
the PMTs. It is for these reasons that the G0 experiment requires a limit on the
beam halo of 1 ppm outside a 6 mm radius of the electron beam. The halo monitors
are used to detect beam halo and time-dependent halo behavior of the electron beam
in Hall C. This system was designed to be non-invasive, so that the beam halo could
be monitored continuously during data taking. This detector system was made up
of a bare Hamamatsu 931B low gain PMT, a Phillips XP2262 PMT coupled to a
glass brick that measured 4 cm × 4 cm × 43 cm. Multiple detector types were used
as a part of a study to determine which was the optimal design for the halo detector.
A 2 mm thick carbon target with a 6 mm diameter square hole was located at the
hall C pivot. The main portion of the electron beam would pass through the hole
but halo electrons would interact with the target. Besides the ability to read out
the beam halo rates, the halo target also had a Fast Shut Down output so that if
the beam halo rate became too high the electron beam would be turned off.
From the engineering run, it was found that not all the prototype halo monitors
were suitable for use. The bare PMT was damaged by work done at the Hall C target
pivot. The lead glass halo monitor had a large amount of noise associated with it
and was insensitive to the target ‘in’ and ‘out’ differences. The Lucite halo monitor
worked the best. It was fairly insensitive to low-energy background and was sensitive
to the position of the halo target. It was found that for good beam, the Lucite halo
67
monitor observed a baseline value of 5 Hz/µA (see Figure 3.13).
FIG. 3.14: Plot of Lumi asymmetry as a function of run number for the 2003 G 0 com-
missioning run. Due to the extreme forward angle of the Lumi monitors, measured
asymmetries may be due to beam boiling effects on helicity correlated properties of the
beam. The red circles (blue triangles )indicate when the insertable halfwave plate was in
the ’out’ (’in’) position.
ity. The luminosity is the product of the beam current and target density. This
detector package consists of a pair of bare PMTs, a pair of PMTs with Lucite
(Cerenkov light is produced in the Lucite and the signal is picked up by the PMT),
and a prototype water Cerenkov detector from Mainz experiment PVA4. After
dividing out the beam charge, target density fluctuations can be monitored along
are at extreme forward angles and thus the parity-violating asymmetry should be
68
near zero. If an asymmetry is measured by the detectors, this implies that a false
eter differences or electronics artifacts. Results from the LUMI monitors are shown
in Figure 3.14 indicating that the target was neither boiling nor were there large
FIG. 3.15: The Hall C Polarimeter. The top schematic shows the collimators and
quadrupole magnets used to select scattered electrons of interest. The bottom schematic
shows the scattered electrons focused onto the polarimeter detectors.
e + e ) polarimeter [59] which is located in the beam alcove upstream of Hall C. Ob-
69
serving differences in the scattering rates, depending on whether the beam and target
electrons are polarized parallel or anti-parallel with one another, provides a mea-
surement of the beam polarization. Since this is a QED process, the cross sections
4
has been calculated precisely (up to αQED ). The cross section for a longitudinally-
dσ dσ0 h k k
i
= 1 + Pt Pb AZZ (θ) (3.6)
dΩ dΩ
dσ0
where the unpolarized cross section dΩ
is given by
2
dσ0 α(4 − sin2 θ)
= (3.7)
dΩ 2me γ sin2 θ
One can measure the beam polarization by comparing the cross section asymmetry
for the beam and target spins aligned parallel and anti-parallel:
dσ k dσ ⊥
−
= dΩk dΩ = A (θ)P k P k .
ZZ t b (3.9)
dσ dσ ⊥
+
dΩ dΩ
k
Knowing the target polarization Pt allows one to isolate the beam polarization.
The target electrons are provided by atomic electrons associated with the iron
atoms in the target. Typically, about 2 electrons of the iron’s 26 electrons are po-
larized, leading to a target polarization ∼ 8%. The target is a thin foil of iron
ducting solenoid producing a 4 T field. The scattered electron and recoiled target
electron that emerge in the horizontal plane between 1.83 ◦ and 0.75◦ in the lab frame
are focused by a quadrupole magnet Q1. The desired scattering angles are set by
collimators. The electrons are then defocused using another quadrupole magnet,
70
Q2, and the electrons detected in coincidence using two symmetrically placed ho-
doscope counters and lead glass counters. This system of movable collimators and a
pair of quadrupoles allow this device to be tuned to operate at any beam momentum
The need to correct the elastic asymmetry for the polarization can be demon-
strated as follows from considering the scattering cross sections. One of these com-
helicity states. Then there is a parity-violating part, σ P V , which is caused by the in-
part has opposite signs depending upon the helicity, thus the contribution of (σ P V )
will scale with the beam polarization. The right-handed component of the cross
σ+ = σEM + Pb σP V (3.10)
σ− = σEM − Pb σP V . (3.11)
(σEM + Pb σP V ) − (σEM − Pb σP V ) σP V
Ael = ≈ Pb = Pb Aphysics. (3.12)
(σEM + Pb σP V ) − (σEM + Pb σP V ) σEM
The data acquisition system (DAQ) used by the G0 experiment is CODA (CE-
BAF Online Data Acquisition system) [60]. CODA was developed by Jefferson Lab.
The G0 DAQ runs on a 1 GHz Pentium III computer running Linux (kernel 2.4.18).
The DAQ reads out the time-encoding scaler data (see Section 3.10) at a rate of
30 Hz during the 500 µs window for the Pockels cell to settle down after flipping
71
helicity. The time encoding data is read from the North American Scalers and from
the French DSP concentrator on the DMCH-16X boards (see Section 3.10). Dur-
other types of data events are interleaved in the data stream. FASTBUS ADC and
TDC data (see Section 3.10) were collected at a rate of 1/30 Hz. These FASTBUS
data are useful for monitoring and calibrating the detector system. In addition,
“slow control” EPICS [61] events are taken at 30 Hz; these events record data from
the beam position monitors, beam current monitors, temperature and pressure of
the target, temperature and current in the SMS, etc. The DAQ computer running
CODA communicates with each of the electronic crates via a single board computer
on each crate called a ROC (Read Out Controller). There are six ROCs: ROC1,
ROC2, and ROC3 all contain North American scalers, ROC3 contains the French
and TDCs, and the last ROC known as TS0 is the trigger supervisor that also reads
out beam and slow control electronics. The CODA datafiles are copied to the tape
silo system and to a group of three computers where G0Analysis, the replay engine,
produces ntuples and histograms to be read by ROOT and fills a MySQL database.
The signals from the G0 detectors in Hall C are routed upstairs to be processed
by the G0 electronics. The signals coming from the North American (French) detec-
tors are handled by a North American (French) subset of the electronics. Having two
different sub-systems allows for a powerful cross check between the North American
Both North American and French electronics can be described as falling into two
(ToF) histograms accumulated over a macropulse. The other set of electronics are
used to monitor detector efficiencies, to calibrate the gains of the detectors, etc.
Common to both sets of electronics is the need to minimize both dead time and
helicity-correlated systematics.
ceives 256 input signals from the North American (French) octants 1,3,5,7 (2,4,6,8)
and the implementation of the electronics follow two different philosophies with the
North American electronics being highly modular and the French electronics being
highly compact.
ing sections describe in further detail the signal’s journey from the PMT signal
to accumulated data. Signals from the PMTs first go to a patch panel in Hall C
through 36 meters of RG58 cables. They are then sent up to the electronic counting
room in 107 meter long RG8 cables for reduced attenuation. Due to the high rate,
must then be chosen to reduce the time resolution of the signals arriving at the
Time Encoding Electronics. The nominal 1 ns wide bins are determined by a clock
logic pulse, and meantimers are employed to average the pulse times of the PMTs
.
73
FIG. 3.16: Schematic of the North American electronics for the G 0 forward running
mode. Custom built electronics are denoted by dark boxes.
Splitter
There are 16 custom built splitter modules. Each splitter module asymmet-
rically (70:30) and passively splits 16 PMT signals to provide 16 inputs to the
The high counting rates on the detectors implies that the signals must be of
low amplitude in order to reduce the instantaneous and integrated currents on the
PMTs. With these low amplitude signals, a good precision on the elastic proton
time information is required and time-walk could be a problem, hence the use of
74
the constant fraction discriminators. There are 16 Lecroy 3420 Constant Fraction
Discriminators (CFDs). These CFDs take the PMT analog signal from the splitter
and produce a logic signal if the signal is above some threshold, which was typically
about 35 mV. This is done by generating the logic signal at some constant fraction
of the peak height to produce a nearly walk-free signal. The output signals from
the CFDs are then sent to the FASTBUS TDCs and to the TEE meantimers.
Meantimer
There are 8 custom built modules with 16 meantimer channels in each module.
Each meantimer (MT) has inputs from two CFDs. One input is for each PMT
signal from each end of a detector. The output signal is the mean time of the two
PMT input signals. The mean timing of the input signals is performed by custom
This is done because the proton can hit anywhere along the scintillator. This
and t2 be the time of the signal from each PMT, d1 and d2 the distance from where
the particle hits along the scintillator to the PMT, and d = d 1 + d2 . By taking the
t1 + t2 d1 + d2 d
= =
2 2 2c
In total there are 128 meantimed signals that are sent to the FASTBUS TDCs
spectra for data rates of several MHz. There are 32 custom built LTD modules.
The LTDs have a cycle time of 32 ns. Twelve clock pulses are used to clock a shift
register. The LTDs take 2 signals from the front and back meantimers of a detector
in coincidence and latches for a single beam burst. The time since the beam sync
signal of the latched input is determined by using two 16 bit shift registers which
are clocked 180◦ out of phase with respect to one another. The status of the shift
registers are latched into another set of registers at the end of the beam pulse. These
registers then drive 24 VME scaler channels (see below) which count how many times
the bits were set. The depth of penetration of the input signal into the shift register
during the shifting sequence then depends upon the time of the coincidence within
the 32 ns cycle. This depth of penetration of the signal thus encodes the time of
the coincidence. The LTDs also monitor the quality of the clock train for errors of
“too many” or “too few” clock pulses. These gated clock trains for the LTDs are
• Clocking Gating Board: The custom built Clocking Gating Board (KGB) takes
the 499 MHz clock signal from the CEBAF accelerator master oscillator and the
“YO” signal provided by a beam pickoff monitor upstream of the G 0 target. This
• Signal Duplication Boards: The custom built Signal Duplication Boards (SDBs)
take a single copy of the gated clock train and a copy of the sync signal and
The scalers which capture the time spectra from the LTDs are custom built
1/30 s macropulse. During the settle time of 500 µs between macropulses, the scaler
data is latched into on-board memory and the scaler channels are cleared. The G 0
DAQ system then reads out the latched data for a macropulse while the scalers are
Splitter
There are 4 custom built splitter modules. Each splitter module symmetrically
(1:1) and actively splits 64 PMT signals to provide inputs to the Constant Fraction
DMCH-16X
The heart of the French time encoding electronics are 8 custom built boards
togramming, 16 channels within the vXi standard). Each board receives 32 PMT
analog signals and builds 8 Time-of-Flight histograms associated with the detectors
of two quarters of two octants. Most of the board’s settings such as the CFD thresh-
olds and differential non-linearity (DNL) for the TDCs are controlled via software.
check CFD thresholds and MT outputs. The S-DMCH keeps track of individual
counts for the CFDs and MTs. Each DMCH-16X motherboard holds 2 TDCs which
distribute data through FiFo buffers to 4 front-end digital signal processors (DSP)
77
where the ToF spectra are accumulated in different memory registers associated
with each detector. At the end of an MPS, the 4 front-end DSPs and the DSP on
the scaler S-DMCH send their data to another DSP known as the DSP concentrator,
which transfers the block of data associated with one DMCH-16X board over the
Data Analysis
The first G0 engineering run took place between October 2002-January 2003.
The data analysis from this run was performed in several steps. First the raw
measured asymmetries must be formed from the detector yields. Then the dead time
must be calculated and corrected. There is a correction for the background dilution
factor and the background has an asymmetry associated with it that must be taken
into account. Then the physics asymmetry can be extracted after correcting for the
The detector signals are accumulated during a helicity state (MPS). When the
helicity is reversed by the Pockels cell that flips the helicity at 30 Hz, the data
is read out by the electronics. The detectors count the desired elastically scattered
protons as well as particles from other processes such as inelastic scattering from the
aluminum windows of the target [62] and production of pions in the hydrogen target.
contain either 24 (NA electronics) or 128 (French electronics) time bins. Other types
of data are also recorded during this spin-flip period, such as the integrated charge
by the beam current monitors and integrated positions from the beam position
monitors. For these ToF histograms the yield (Y ) is the number of events (N)
measured in each time bin normalized to the beam charge (Q) accumulated during
that MPS. This charge normalization is needed since the event rate is a function of
×10
3
8000
Elastic
7000 Protons
6000
5000
evts
4000 Inelastic
Pions Protons Elastic Cut
3000
2000
1000
0
0 20 40 60 80 100 120
ToF(1/4ns)
FIG. 4.1: An example of a typical French Time-of-Flight spectrum from detector 8 with
data taken during a one hour run..
From the quartet, an asymmetry can then be computed for events within the
elastic cut window, which is about 4 ns wide. This asymmetry for each detector
80
measured over a quartet can be written as:
+ −
N1 N2+ N N2−
+ +
Q1 Q2+ − Q1− + Q−
Ameas,det,qrt = + 1− 2
(4.1)
N1 N2+ N N2−
Q+
+ Q+ + Q1− + Q−
1 2 1 2
where Nis is the number of counts recorded for the ith MPS of this quartet with beam
helicity of sign s, and Qsi is the beam charge incident on the target during that MPS.
Since the yield is the number of counts normalized to the charge, Equation 4.1 can
be written as:
The measured asymmetry from all quartets within a run are averaged in each de-
root mean square of the standard deviation of the quartet asymmetry distribution
The next step is to take the weighted average and error of the measured asym-
However, this is not the full story. Within the elastic proton peak cut window,
there is some contamination, since not all the events are from elastic protons. Some
of these events are inelastic protons produced in scattering within the target. This
contamination must be removed in order to evaluate A physics for the elastic protons.
The measured asymmetry for each detector must be corrected to subtract the
inelastic contribution (note the quartet helicity ordering denoted at “1” and “2”
81
above will now be suppressed for readability):
right(left) handed elastic (inelastic) events. One can define the so-called ‘inelastic
Yinel
dilution factor’ as d = and solve for the elastic asymmetry for each detector:
Yel
After correcting for the inelastic contributions, other corrections must be applied to
the elastic asymmetry to obtain the physics asymmetry, Aphy . The physics asym-
metry must be corrected for not having 100% beam polarization, radiative effects,
The error on the elastic asymmetry has several contributions, as can be seen
in Equation 4.8. For the G0 forward angle production run, an error on the elastic
1
σstat = p (4.9)
Nproton
where Nproton is the number of detected protons in one quartet. Thus if the measure-
ment is purely statistical, the width of the distribution of the measured asymmetries,
82
1400
1200
1000
Number
800
600
400
200
0
0.9 0.95 1 1.05 1.1 1.15 1.2 1.25
σmeas/ σstat
FIG. 4.2: The ratio of the measured quartet asymmetry distribution widths to the width
σmeas
expected solely from counting statistics for the detectors . Deviations from 1.0
σstat
indicate the presence of instrumental noise in the measurements.
σmeas , is:
1
σmeas = p + (4.10)
N1 + N2+ + N1− + N2−
for each helicity period.
noise can broaden the width of the measured distribution. The number of counts
used in this calculation has been corrected for dead time (see Section 4.6). The
measured statistical widths due to the number of counts for each detector is on the
order of 3800 ppm. This makes sense, since the detector rate is about 0.5 MHz, the
Obtaining the number of events in a quartet requires the result from Equation 4.11 to
be multiplied by 4 for four helicity windows in one quartet. Using Equation 4.2 yields
83
σmeas
Ameas,qrt,det ∼ 3800 ppm. All detectors were measured to have ∼ 1.06 (see
σstat
Figure 4.2) implying that all measurements are dominated by counting statistics.
4.3 Cuts
The measured raw asymmetry had several cuts applied event by event in or-
der to improve the statistical properties of the results. These were applied by the
G0Analysis replay engine which produces the histograms and NTuples, to be read by
ROOT, from the raw CODA data files. The first check is to be sure that an event in
the MPS make up a good quartet. The code will check the reported helicity (which
has been delayed by 8 helicity windows) against an algorithm for helicity prediction
in G0Analysis. These cuts defined a good MPS. A good quartet requires that all
MPSs in a quartet are good and have the correct helicity pattern. Asymmetries are
calculated for good quartets only. After this cut, two more sets of cuts were applied:
- Beam Cuts:
If the beam current read out by the beam current monitor (see Section 3.5.6)
is less than 5µA, that MPS along with the next 2000 MPS after beam recovery
are removed. This is because it has been noted, by looking at the luminosity
monitors, that the target takes approximately a minute to settle down after
beam is restored.
- Detector Cuts
Both North American and French electronics have error indicators in the data
stream. The North American electronics (see Section 3.10.1) have error indica-
tors on the LTD boards to note if the LTD is seeing “too many” or “too few”
micropulses within the 30 Hz helicity window. The French electronics (see Sec-
tion 3.10.2) are armed with “alerts” which count the number of overflow words in
84
a DMCH module. These counters are looked over first, then data from a detector
takes the first 100 MPS of each run for each detector and calculates the mean
and width of the yield distribution, for each detector and for each ToF bin, as a
reference value. A ±10σ cut around the reference value is then applied to each
ToF bin and a detector is tagged as “bad” if any ToF bin fails the cut.
The raw asymmetry is calculated for each quartet for each detector from the
normalized yield
The normalized yield is the yield for one MPS divided by the beam charge accumu-
(dσ/dΩ)L∆ΩTM P S
Y+MorP −
S=1 or 2
= (4.13)
QM P S
= NM P S /QM P S (4.14)
the length of time for one macropulse. Figure 4.3 is a plot of the measured raw
FIG. 4.3: North American and French (out-in) raw asymmetries. The insertable halfwave
plate reverse the sign of the raw asymmetry. By subtracting the ’in’ asymmetries from
the ’out’ asymmetries, one can combine the two sets of results. The error bar is found
by adding the ’in’ and ’out’ errors in quadrature
violation experiment it is important to demonstrate that the results are being pro-
electronic effects. Passively reversing the sign of the measured raw asymmetry is a
source, the helicity of the circularly polarized light produced by the G 0 TIGER laser
(see Section 3.5.1) is passively reversed. This in turn reverses the electron helicity
pattern without changing any other device systematics. If the analysis software does
not take into account the insertion of the halfwave plate, the opposite sign for the
86
asymmetry will be calculated.
p0 0.3339 ± 1.027
10
-5
-10
-15
0 2 4 6 8 10 12 14 16
Ring
p0 0.6342 ± 0.7087
10
-5
-10
-15
0 2 4 6 8 10 12 14 16
Ring
FIG. 4.4: North American (NA) and French (FR) (in+out) raw asymmetries. Under
the raw asymmetry sign reversal, due to the insertable halfwave plate, adding the ‘in’ and
‘out’ states should yield a zero result.
It is clear from Figure 4.3 and Table 4.1 that there is a correlation between the
presence of the halfwave plate and the sign of the asymmetry. From Figure 4.4, it
can be seen that adding the ‘in’ and ‘out’ data results in zero, as expected. When
the halfwave plate is inserted, the sign of the asymmetry is corrected when combined
The asymmetry had to be corrected to take into account electronic dead time.
The electronic dead time is caused when the electronics are still busy after/during
the processing of an event. The G0 custom electronics have been designed to measure
high rates on the order of 1 MHz with a controlled dead time. Dead time from the
North American electronics can come from the LTDs and the CFDs. Both North
American and French electronics are set to neutralize the next pulse after an event
of dead time. The probability of being dead, f , is proportional to the rates of events
triggering the electronics. This will cause the measured yield Y meas , to differ from
Af alse = −f × Aq (4.16)
The North American LTDs and French time encoding electronics record only a four-
fold coincidence and therefore cannot be used to correct for single events, where
singles occur when one to three (but not all four) PMTs fire. The singles are mainly
pions that come before the protons in the time spectrum, and so a singles hit will
then cause the loss of the proton count. This means that on average the electronics
In the North American electronics, the dead time was dominated by the CFDs.
The CFDs should work by taking the original signal and producing two more signals.
One signal is a duplicate of the original but delayed in time and the other signal
is a fraction of the original signal amplitude. The CFD then will fire when the
original signal is above a certain threshold and the two duplicate signals intersect.
What happened during the 2002-2003 engineering run was that sometimes a smaller
detector signal would precede a larger detector signal, and this would cause the
duplicate signals to intersect before the original signal would cross threshold. This
produces an apparent dead time in the electronics. This effective dead time was
about 70 ns for the North American electronics. This problem will be corrected in
future running by adjusting the delay and fraction in the CFDs to prevent these
sub-threshold particle signals from affecting the larger particle detector signals.
The overall dead time fraction for the NA detectors was found to range from 3%
89
to 11% for detectors 1 through 14. The technique for extracting the dead time
fraction is to plot the detector yield versus beam current or detector asymmetry
versus charge asymmetry. The overall dead time fraction is the sum of the dead
where Rcoinc (Rsingles ) is the rate due to the coincidence(singles), and τ coinc (τsingles )
is the dead time due to coincidence(singles). The dead time due to coincidences is
given by
where j is the timebin of the detected particle. The dead time due to singles was
found to be 70 ns (as explained above) and this dominates the dead time, thus the
where Rcoinc is determined from the LTDs and Rsingles is determined from FASTBUS
data.
The French electronics singles dead time from the CFDs was found to be 35 ns.
Properties
energy and charge can cause false asymmetries to appear in the data. This can be
seen if one of the beam parameters mentioned above is on average, different between
90
the two different helicity states. If this is the case, then the measured yield will be
different between the two different helicity states, thus producing a false asymme-
try. These fluctuations are the results of reversing the voltage on the helicity Pockels
cell. Besides producing circularly polarized light, this voltage reversing might pro-
duce some helicity-correlated systematic, e.g. the angle of the emerging laser light
might be different in one state versus another. This systematic then translates to
slightly different types of helicity-correlated laser light hitting the GaAs crystal dif-
the electron beam between helicity states (see Figure 4.5). These helicity-correlated
pear in the data. This false asymmetry requires a systematic correction to the
measured asymmetry.
yield Y and the beam parameter xi , where the beam parameter can be any of the
six mentioned above. The yield due to beam fluctuations can be written as
Y = αxi . (4.21)
+,−
The measured yield, Ymeas , is then a combination of the parity-violating yield, Y +,− ,
and the yield due to the correlation with the beam parameter
−
δxi = x+
i − xi (4.23)
Y+−Y−
Ameas = (4.24)
Y++Y−
91
substituting in the yield from Equation 4.22, one obtains
−
Y + − Y − + α(x+
i − xi )
Ameas = + − . (4.25)
Y + Y − + α(x+ i + xi )
Y + − Y − + αδx
Ameas =
Y++Y−
Y −Y−
+
δx
= −
+α +
+
Y +Y Y +Y−
δx
= Acorr + α
2<Y >
(4.26)
where the average yield is given by < Y >= 21 (Y + + Y − ), and α is identified as the
1 X ∂Y
Ameas = Acorr + δxi . (4.27)
2 < Y > i=1,6 ∂xi
∂Y
The detector yield response to the various beam parameters, , must be
∂xi
extracted from the data. This can be done by looking at the natural motion of the
X ∂Y
Ycorr = Ymeas − δxi (4.28)
i=1,6
∂xi
where Ymeas is the raw yield of the detector, xi is one of the six beam parameters (x
and y position at target, x and y angle at the target, charge at the target and the
energy). The helicity-correlated beam position and angle are calculated using two
BPMs closest to the G0 target. The energy difference is measured using a BPM in
a dispersive region in the arc. The beam charge is measured from the standard Hall
C BCMs.
∂Ymeas
The slopes, , are found by inverting the covariance matrix in the follow-
∂xi
ing equation that relates the slopes to the mean correlation between the yield and
92
the various beam parameters
X
∂Y
(< δY · δxi >) = (< δxi · δxj >) (4.29)
i=1,6
∂xi
the covariance of the yield and beam parameters. This variation in the yield to
beam properties. Figure 4.6 shows the percent yield on detector 1 for all the octants
as the beam properties are being changed. In general, diametrically opposing octants
should have opposing sensitivities and thus opposing signs. It is due to this behavior
that the sensitivities due to changes in beam position and angle are largely canceled
out between diametrically opposing octants when one takes the mean.
93
FIG. 4.5: Plots of the helicity-correlated beam properties. The position differences and
angle differences are the values projected onto the target from the BPMs G0 and G0B.
The energy difference is measured from BPM 3C12.
94
The resulting octant averaged slopes as well as the associated asymmetries are
shown in Table 4.2 for the insertable half wave plate out state and in Table 4.4
for the insertable halfwave plate in state. The false asymmetry A f alse due to the
found to be:
by
The French time-of-flight (ToF) spectra with its fine 128 binning (versus the
North American ToF with its more coarse and non-equal 24 binning) is the first
data set to be looked at for extracting the inelastic dilution factors for each detector.
Each spectrum was fit with one Gaussian for each of the three particle distributions:
pion, inelastic proton and elastic proton (see Figure 4.7). Numerical integration was
employed to find the contributions due to the elastic and inelastic protons within
the 4 ns wide elastic cut window. A total of 106 runs were fit. The inelastic dilution
factors of each detector were stable to 3% over the set of runs assuming the cut does
Elastic
Protons
Events
Elastic
Inelastic Cut
Pions
Protons
Time (.25ns/bin)
FIG. 4.7: An example of a 3 Gaussian fit to French Detector 9 Octant 2. Each par-
ticle distribution has been fit to a Gaussian. The vertical lines around the elastic peak
constitute the elastic cut for this spectrum.
Inelastic Dilution Factor
Cut Changed
Run Number
FIG. 4.8: An example of the stability of the inelastic dilution factor over 106 runs for
detector 14 Octant 2. The abrupt change near run 60 is due to the elastic cuts being
changed.
The contributions to the percent errors on the extracted inelastic dilution fac-
tors (which are added in quadrature to produce the final percent error) are :
– Statistical precision of the fit: the dispersion between the inelastic dilution
factors for the same detector and same octant, with the same cuts for different
• Error on the shape of the fit: the dispersion between using different plausible
mathematical equations to describe the same spectra. This was found to be 1%.
98
• Error on the average over the octants: the dispersion between the dilution factors
for different octants. Having different cuts for the same detector but for a different
Since there was a 5% error between dilution factors for each detector between
the octants, the inelastic dilution factors for each detector was averaged over the 4
FIG. 4.9: French inelastic dilution factors as a function of detector number, averaged
over 4 French octants, using the 3 Gaussian fit.
The North American (NA) time-of-flight (ToF) spectra were plagued by differ-
ential non-linearity (DNL) (see Figure 4.10). The NA ToF spectra was expected to
have time bins that were 1 ns wide (except for the first bin which was 6.5 ns wide
by design). The DNL manifested itself as deviations from the 1 ns width. The DNL
99
Det Inelastic Dilution Factor Error
1 0.148 0.007
2 0.171 0.009
3 0.180 0.009
4 0.160 0.008
5 0.194 0.010
6 0.202 0.010
7 0.216 0.011
8 0.200 0.010
9 0.210 0.010
10 0.222 0.011
11 0.259 0.013
12 0.267 0.013
13 0.307 0.015
14 0.401 0.020
TABLE 4.5: The average inelastic dilution factors for the French detectors for the cuts
on the elastic peak which were ∼ 4 ns. The error represents the final absolute error on
the dilution factor for each detector.
created time bins ranging from ∼ 0.5-2 ns. The DNL was significantly larger than
expected, and the cause(s) for this large DNL are only imperfectly understood at
the present, but appear to be at least partly due to the clock signal.
The North American electronics dictate the width of the bins of the North
American ToF spectra by forming the beginning and ending of the bins on the
rising and falling edge of a clock signal supplied by accelerator. If the clock train is
made up of perfect square waves, then the bins in the North American ToF spectra
will be 1 ns wide. If the clock train is asymmetric, then the bins will deviate from
1 ns, as observed during the 2002-2003 G0 commissioning run. This makes fitting
the North American ToF spectra to find the inelastic dilution factors more difficult
The DNL can be measured with “white noise” runs. A white noise run con-
sists of an LED shining on a photo-multiplier tube whose signal is then fed into the
North American electronics. For bins of equal width, a flat ToF spectrum should
be produced for bins 2 though 24. Since the North American ToF does suffer from
100
FIG. 4.10: An example of a time-of-flight spectra from a North American detector. The
jagged binning is due to the differential non-linearity that causes the bins not to be of
equal time widths.
the DNL problem, the bins are not of equal widths and by normalizing each of the
bins, the width of each bin can be found (see Figure 4.11). The qualitative pattern
of the DNL tends to alternate between ∼ 0.5 and ∼ 1.5 ns wide bins although there
are significant deviations from this pattern. Another difficulty with the DNL is that
it changes with time, on time scales of the order of a day, as can be seen in both
To evaluate what effect the DNL had on extracting the inelastic dilution fac-
tors, a simulation of the ToF spectra without DNL for each detector was produced.
Using the fit parameters obtained from the French ToF for the elastic and inelastic
proton distributions, the ToF spectra for detectors 1-14 were simulated with very
Three white noise spectra were taken during the G0 commissioning run and
three white noise runs were taken later after the commissioning run; with these six
101
FIG. 4.11: An example of several white noise spectra for North American octant 3 detector
4. The ordinate axis is the width of each bin along the abscissa axis. Note that the bins
deviate from the expected 1 ns width. This deviation is the so-called ’differential non-
linearity’ or DNL. Also note that the DNL changes with time.
runs, a total of 212 measurements1 of the DNL exist. The simulated spectra were
then recast into North American binning and the 212 measured DNLs were applied
These spectra were then fit to 2 Gaussians (one for the inelastic proton and one
for the elastic proton distributions) and numerical integration was applied to find
the inelastic dilution factors. When the results for each detector were histogrammed,
a bimodal distribution was found (see Figure 4.14). This bimodal distribution was
due to how the DNL was applied to the ToF spectra. The DNL creates a bias
weighting within the elastic cut (the first bin after the cut is typically either 0.5 ns
wide or 1.5 ns wide). The actual inelastic dilution factor 2 was distributed around
only one value (see Figure 4.14). This implied that the DNL must be corrected to
FIG. 4.12: An example of one bin’s variation over 8 measurements over 4 months for
detector 7 octant 1. Note that on April 17th the bin width changed by nearly a factor of
two from the day before and the day after.
Events
Time (ns)
FIG. 4.13: The plot on the left is a simulated North American ToF spectrum using
the fit parameters obtained from the French ToF. The plot on the right is a simulated
North American ToF spectra with DNL. Each simulated North American ToF spectra
was produced 212 times with different DNLs.
The DNL was corrected by using variable binning and normalizing each bin to
the variable bin width. Only the three sets of North American white noise runs taken
during the 2002-2003 G0 commissioning run were utilized to fix the North American
Since the DNL changes with time, one set of white noise measurements will
not fix the whole North American data set. Figure 4.17 shows a ToF spectra for
the same detector with the same DNL correction but for different runs where one
run passes the subjective test (the ToF spectrum is “smoothed” out) and the other
103
Number
FIG. 4.14: The top histogram is an example of the simulated extracted inelastic dilution
factors for a North American detector. The bottom histogram are the results of the “true”
inelastic dilution factor from the simulated corrected North American ToF spectra.
fails (the ToF spectrum still appears “jagged”). To further complicate this, since
the DNL can change one day to the next day, one white noise spectra might fix the
DNL for many runs while failing for some of the intermediate runs within the run
set (see Figure 4.16). Looking at the relative minimum in relation to all other runs,
the sum of the differences between the fit and the data (as shown versus run number
in Figure 4.18) was a useful (though not definitive) guide in determining which runs
With only three sets of white noise spectra which correct the asymmetry runs
near it in time to the white noise run, only 46 runs out of 124 were able to be
corrected. There was an overlap of 8 runs between the 3 sets of data corrected by
After the DNL was corrected on the 46 runs, a three Gaussian fit to each
particle distribution and the standard 4 ns database cuts were applied. Numerical
The contributions to the percent errors (added in quadrature to obtain the final
104
• Error on the fit: the dispersion of the fit on a detector over 46 runs. This is
important since the dilution factors are sensitive to the size of the cut window
which is known to change over the data set. This was found to be 14%.
• Error on the knowledge of the DNL: the dispersion between the results from the
8 runs that are corrected by the 3 white spectra. This was found to be ∼ 1%.
• Error on the shape of the fit: the dispersion between different mathematical
• Error on the average over octants: the dispersion between the dilution factors for
different octants. Having different cuts for the same detector but for a different
octant or having different set of thresholds might cause this dispersion. This was
an inelastic dilution factor for each detector for each octant, one can quote the
inelastic dilution factor for each detector averaged over the octants.
105
Run Number
FIG. 4.16: Stability of the North American inelastic dilution factors over ∼50 runs for
a typical detector. Note that for ∼5 runs the inelastic dilution factors have changed in
a similar fashion as noted in Figure 4.12. This is presumably due to the DNL changing
with time. ToF spectra for these runs resemble the failing ToF spectra in Figure 4.17.
These failing runs were not used in calculating the inelastic dilution factors.
The results of the extraction of the North American inelastic dilution factors
Events
Time (ns)
Events
Time(ns)
FIG. 4.17: Both ToF spectra have been corrected using the same white noise run. The
top ToF is taken to be a “passing” corrected spectrum. The bottom ToF spectrum is not
as smooth as the top ToF spectrum and it “fails” under the assumption that the DNL
has changed.
The North American and French inelastic dilution factors have been extracted
(see Tables 4.5 and 4.6) for the 2002-2003 G0 commissioning run. The errors as-
sociated with the inelastic dilution factors allow for the extraction of the elastic
asymmetry with an error that is tolerable. The North American errors on the
inelastic dilution factors are considerably larger than the French results. This is
presumably, in part, due to the width of the elastic cut window which is affected by
the DNL.
The DNL that plagued the North American ToF spectra should be reduced in
the second G0 commissioning run with the replacement of the RF translator board.
White noise runs will be taken on a frequent basis (∼ 1/day) until the DNL is under
Fit-Data
Sum of
Run Number
FIG. 4.18: In deciding which runs are fixed by which white noise spectra, looking by eye
at the DNL corrected ToFs in conjunction with looking for minimums of the sum of the
fit-data was used. This quantitative method, though useful, was not definitive. Notice
fluctuations between runs 43 and 78 might be considered failing but when examined by
eye these ToFs appeared corrected.
FIG. 4.19: North American inelastic dilution factors as a function of detector number,
averaged over the North American octants.
108
4.9 Background Inelastic Asymmetry
Asymmetry(ppm)
Bin
FIG. 4.20: Example of the coarser asymmetry binning to find the background inelastic
asymmetry for North American detector 4. The Time-of-flight asymmetries have been
broken down into 7 bins. The elastic proton bin is denoted in red. The asymmetry bins
above and below the elastic proton bin were interpolated to find the background inelastic
asymmetry under the elastic proton peak.
must be known, since inelastic events overlap with the elastic peaks and dilute the
where Ael is the elastic asymmetry, Acorr is the corrected measured asymmetry from
asymmetry, and d is the inelastic dilution factor. The main contribution to the back-
ground comes from processes involved in scattering from the downstream aluminum
window of the target. This background represents 13-25% of the events within the
elastic cut window. The background fraction rises with higher detector number (and
with it, on the order of 10 ppm. This background is thought to be due mainly to
FIG. 4.21: North American and French extracted background asymmetries. These asym-
metries were obtained by dividing the ToF asymmetries into 7 bins and interpolating the
bin above and below the elastic cut window.
extracted by dividing the Time-of-Flight spectra into 7 bins with the 5th bin being
the elastic proton cut (see Figure 4.20). A linear interpolation was made between
the side band bins below and above the elastic proton bin. The results can be seen
In the next commissioning run, the downstream window thickness will be re-
duced from 11 mils to 3 mils, which should reduce the background by nearly 60%.
The background asymmetries and yields will be directly measured during the next
run with dedicated dummy target runs with a 30 mil aluminum foil dummy target
(known as the “flyswatter”) and a 3.4 mil tungsten radiator. The purpose of the
flyswatter and radiator will be to confirm the expected fraction of events from the
TABLE 4.7: The extrapolated background inelastic asymmetries for North American and
French detectors. These asymmetries were interpolated from averaging the N bins above
and N bins below the elastic cut window. These results are reported by detector number,
where the results for a detector number were averaged over the 4 detectors from the NA
(Fr) Octants, e.g. results reported for Fr Ainel Det 1 are the weighted average of detectors
1 from French Octants 2,4,6,and 8.
4.10 Polarimetry
The Moller polarimeter described in Section 3.8 was used to correct the physics
Ael
Aphy = (4.35)
Pb
A limited number of measurements were made during the month of data taking
as can be seen in Figure 4.22. The average beam polarization was found to be (77.3
± 0.4)%.
111
FIG. 4.22: Polarization measurements made during the engineering run. Polarization
measurements made with the insertable halfwave plate ‘in‘ must be multiplied by -1 to
compare to the insertable halfwave plate ‘out‘ measurements.
Since the statistical error bars of this work are so large, the data have not been
corrected for electromagnetic radiative effects. This is because the radiative effects
are expected to be small, on the order a few percent of the measured asymmetry.
In order to carry out a complete analysis, radiative corrections should be taken into
account. In order to do this the following references are invaluable [63, 64].
tions and internal Bremsstrahlung corrections. External corrections are when the
beam electrons lose energy by bremsstrahlung from the target aluminum entrance
window or in the hydrogen itself before scattering off of a second proton and into
the detector. Internal corrections are when a beam electron interacts by more than
one photon with the proton. These internal corrections are further divided into
”real” and “virtual” processes. In real processes the photon is a real photon that
112
is emitted during the scattering. In virtual processes, during the scattering virtual
These higher-order interactions have two effects on the measured parity violat-
ing asymmetry
• The electron energy is reduced leading to a lower value of Q 2 and asymmetry for
These effects will reduce the measured asymmetry with respect to the tree
level asymmetry. Emission of the photons after the parity-violating interaction will
reduce the energy of the scattered proton leading to a reduction in the detector
signal. The effect of the internal and external bremsstrahlung is to remove protons
from the elastic peak and put them into a long tail.
two different parity-violating asymmetries: At , the tree level asymmetry from single
At (Q2 )
Rc = . (4.36)
AR (Q2 )
4.12 Q2 Determination
Knowing the elastic electromagnetic form factors from other experiments allows
for the extraction of the strange electric and magnetic form factors. In order to
perform this extraction, the Q2 must be known. The total error on the extracted
strange form factors should be smaller than 10%; this requires that Q 2 be known
to the 1% level. In order to reach this precision, one should know the absolute ToF
detector ToF at different magnetic fields with a Monte Carlo simulation [65]. As
the magnetic field is varied, the pion peak will remain stationary while the proton
trajectory will shift and may even reach another detector system.
As mentioned above, this study relied on the results from simulation (G0GEANT,
G0TRACE, and GRAAL). The simulation takes into account the electronics, the
effects of the spectrometer magnetic field and the detector positions. These results
allow one to determine the value of all these parameters directly from the measured
ToF. Using these values allows one to find the < Q2 > for each detector.
When the magnetic field varies, the particle trajectories are modified; they
might even reach another detector. The elastic proton peak is modified by different
field strengths though the pion peak remains unchanged. The idea is to study
the magnetic field variation using the relative position between the pion and elastic
proton peak. From simulation one can see how the ToF should change with magnetic
With the French electronics, the proton peak may be determined within a few
ps (since the French electronics has 250 ps bins). Unfortunately, in the case of the
North American electronics, it is not possible to know the peak positions to better
than 60 ps for the Time Encoding Electronics; this was further complicated by the
TABLE 4.8: < Q2 > values determined by comparing Time-of-Flight differences between
pions and the elastic protons at various magnetic fields. Only data from the French
detectors were used to determine < Q2 > due to the fine time binning of the French
electronics (0.25 ns).
CHAPTER 5
Conclusions
ing of polarized electrons from the proton at forward angles. The asymmetries are
models for predicting strange quark contributions to the proton are discussed for
The first G0 engineering run (from October 2002 through January 2003) was
Many of the challenges associated with generating and maintaining the unique
beam properties were met during the first engineering run. The time structure for
the electron beam was 32 ns which differs from CEBAF’s typical beam structure of 2
ns. This produces a higher bunch charge, due to having 16 times as many electrons
in one bunch, which in turn produces space-charge effects that complicate beam
transport through the injector. Most of the critical beam properties were delivered
in January 2003:
115
116
• beam current of 40 µA,
tested. The charge feedback system worked but the position feedback system re-
quires some investigation into its unstable behavior (the calibration slopes for the
PZT mirror seemed to change very quickly over a 3 hour period). The false asymme-
tries due to helicity-correlated beam properties were small and kept under control.
The G0 detectors performed well. The high voltages for the PMTs were set at
values that allowed for high detection efficiency and the PMTs were able to stand
rates at the nominal beam current of 40 µA. The gains were matched and their
stability was monitored and deemed satisfactory over time and for different beam
currents. The discriminator’s thresholds were adjusted to eliminate noise and low-
energy background while not rejecting the elastic proton signals. Typical detector
rates were on the order of 1-2 MHz with a typical dead time of 10%. This induces a
effect of ∼ 15%; after correction ∼ 1%. The G0 detectors observed yield sensitivities
to six beam properties: helicity-correlated x and y beam motion, x and y beam angle,
There was higher background in Hall C than what was expected. This problem
had to be taken care of early on since the projected anode currents of the PMTs in
the higher numbered (larger Q2 ) detectors would have been too high at the nominal
current of 40 µA with the nominal gain settings. This higher background was due to
neutrals coming from the downstream beam pipe. This was taken care of by adding
The G0 magnet ran at full design current at 5000A for the first time on Decem-
117
ber 18, 2002 and then throughout the January 2003 running.
The G0 target was well behaved with density fluctuations at 40 µA being neg-
ligible. Various target density studies were carried out to extract the contribution
The background yields and asymmetries are needed to correct for the elastic
asymmetries. The inelastic yield under the elastic cut due to inelastic protons was
found to be about 13-25% of the signal. In order to cut down on the background
signal, in the forward angle run the downstream window thickness will be reduced
and an insertable dummy target will be added to help quantify the background.
1
Aphy = ((1 + d)Ameas − dAinel ). (5.1)
Pe
The errors are determined by Equation 4.8 along with considering the error on
the polarization. Explicitly writing these out: the contribution to the error from
the contribution to the error from the background inelastic asymmetry is given by
2
2 1
σAinel = d2 ∆A2inel , (5.4)
Pe
118
and the error from the polarization is given by
2
2 1
σ Pe = A2el ∆Pe2 . (5.5)
Pe
Table 5.2 contains a list of different error contributions to the physics asymmetry.
The data shows good agreement with the expected statistical properties. The
sign and change sign under the influence of the insertable halfwave plate. The results
are consistent between the North American and French sets of detectors/electronics.
It is important to keep in mind when looking at the results of this work that the
1
amount of data taken during the first engineering run represents only ∼ of the
16
expected final statistics from the final forward production run. The statistical error
bars should be about 4 times smaller for the forward production run asymmetries.
two Q2 points (0.45 < Q2 < 0.9 (GeV/c)2 ) in the elastic TOF spectrum. This
119
complicates extracting the inelastic background. Decector 16 contains no elastic
protons in the TOF spectrum. This detector is used as a background detector. For
these reasons, detectors 15 and 16 are missing from the extracted asymmetry results.
NA Fr NA Fr NA Fr NA Fr
Det σd2 σd2 σA2 meas σA2 meas σA2 inel σA2 inel σP2 e σP2 e
(ppm)2 (ppm)2 (ppm)2 (ppm)2 (ppm)2 (ppm)2 (ppm)2 (ppm)2
1 0.04 0.00 6.21 4.17 2.06 2.13 0.00 0.00
2 0.05 0.00 10.16 3.55 3.86 2.02 0.00 0.00
3 0.09 0.00 6.064 3.60 3.17 2.06 0.00 0.00
4 0.24 0.00 7.29 3.94 4.83 2.00 0.00 0.00
5 0.00 0.00 11.47 3.97 5.94 2.48 0.00 0.00
6 0.03 0.00 12.19 3.95 8.57 2.35 0.00 0.00
7 0.00 0.00 7.79 3.95 6.03 2.48 0.00 0.00
8 0.11 0.00 12.31 4.55 8.22 2.25 0.00 0.00
9 0.41 0.01 9.15 4.37 7.29 2.22 0.00 0.00
10 0.14 0.00 12.56 4.73 7.89 2.51 0.00 0.00
11 0.12 0.02 7.80 5.06 5.05 3.13 0.00 0.00
12 0.04 0.00 11.84 6.37 6.96 3.80 0.00 0.00
13 2.15 0.01 27.51 8.17 14.86 5.44 0.00 0.00
14 2.07 0.04 3.72 10.04 20.55 8.87 0.00 0.00
TABLE 5.2: Individual contributions to the errors given by Equations 5.2 through 5.5.
Ameas contains corrections due to helicity-correlated beam properties and dead time.
metry, Ath = η + ψGeA , taken from Equations 2.46 through 2.51, which is written
Deviations from AT h would imply the presence of strangeness in the proton. The
using the dipole parameterization of the proton’s electric (with an uncertainty of 2%)
and magnetic form factors (with an uncertainty of 2%) , the dipole parameterization
120
FIG. 5.1: Plot of the extracted North American and French asymmetries vs. momentum
transfer. The errors are statistical and systematic errors added in quadrature. The
dashed line represents the Standard Model calculation of the parity-violating asymmetry
assuming no contributions from the strange quarks.
of the neutron’s magnetic form factor (with an uncertainty of 3%) and the Galster
A proper description of the strange form factors should be based on QCD. The
problem is that the mass of the strange quark, ms ≈ 150 MeV is comparable to the
QCD scale factor λQCD , thus not easily allowing for a small expansion parameter as
used for the heavier quark calculations. This forces the theorist into the territory of
models and chiral perturbation theory. Several review articles on this subject can be
121
found in the literature [66–68]. Even the applicability of chiral perturbation theory
is called into question here since the strange quark mass may not be light enough
mK 1
to make the SU(3) chiral perturbation valid since λχ
∼ 2
which is not particularly
small. Even if the SU(3) chiral expansion is well behaved, there appears to be various
counter terms (low energy constants) that have not been measured in experiment
and must be extracted from various models. Typically the form factors associated
Loop Models
Proton Proton
into a q q̄ pair to form a meson and an intermediate baryon state. The meson
and baryon later recombine as the q q̄ pair annihilates and the original nucleon is
122
left in the ground state. Only diagrams involving kaons and strange baryon states
contribute to yield non-zero strangeness. Koepf [69] first evaluated µ s and rs2 but
did not include the so-called “seagull” diagrams. These diagrams are needed to
al. [70]. The predictions of rs2 in the kaon loop calculation tends to be smaller than
the pole-fit analysis. To reconcile this difference, the kaon loop model was merged
Lattice QCD
Lattice QCD computations can provide a means of obtaining values for the
low energy constants that have not been measured. These calculations are typically
carried out in the “quenched” approximation where the ss̄ pairs appear only via
one would like to have light quark masses that one can extrapolate to physical values
involves various inputs such as form factors and experimental scattering amplitudes.
The nucleon strangeness arises from the nucleus coupling to a strange meson. In
this case, the exchanged vector boson (Z 0 or γ) fluctuates into an isoscaler meson
(ω or φ) which interacts with the nucleon. The ω and φ are linear combinations of
Jaffe [71] was the first to make theoretical predictions of µ s and rs2 . Jaffe based
his analysis on the work of Hohler [72]. Hammer et al. [73] updated this analysis
φ/ω
Proton Proton
A noticeable point with these analyses is that they typically yield a different sign of
ρs + µp µs
GsE
where ρs is the strangeness radius defined as ρs =, µp is the proton magnetic
dτ
moment and µs is the strange magnetic moment. This will be done by measur-
expected physics asymmetry will be about 1.7 ppm. This measurement is com-
plementary to the SAMPLE [13] measurement at MIT Bates at the same Q 2 but
124
TABLE 5.3: Some predicted values of strangeness radius rs2 and strange magnetic moment
µs .
in the HAPPEX family: the HAPPEX-He [85] experiment will measure the parity-
from 4 He will be sensitive only to GsE and not GsM or GeA due to the fact that 4 He is a
0++ nucleus. With the 4 He measurement of GsE (Q2 → 0) = ρs and the HAPPEX-II
PVA4
electrons between 140◦ < θe < 150◦ . These measurements will be made at two
values of Q2 = 0.23 (see Figure 5.4) and 0.48 (GeV/c)2 in order to complement the
FIG. 5.4: Expected forward angle results from the G 0 , HAPPEX II [84], and A4 [51]
along with result from the HAPPEX [14] experiment. A linear combination of the strange
and electric form factors are accessible from forward angles. This linear combination is
of the form of GsE + α(Q2 )GsM where α(Q2 ) is dependent on kinematic factors.
Figure 5.4). This will allow for an extraction of the vector neutral weak form
factors. Combining the neutral weak form factors with the known electromagnetic
form factors allows for an extraction of the linear combination of strange electric,
GsE (Q2 ), and strange magnetic, GsM (Q2 ), form factors over this momentum transfer
range.
126
G0 Backward Angle
electrons from polarized electron scattering from hydrogen and deuterium targets.
netic, GsM (Q2 ), and the electron-nucleon axial, GeA (Q2 ), form factors. Three sets of
momentum transfers: 0.3, 0.5, and 0.8 (GeV/c)2 . Combining these linear combina-
tion of GeA (Q2 ) and GsM (Q2 ) with the G0 forward angle measurements, which extract
a linear combination of GsE (Q2 ) and GsM (Q2 ), will allow a separation of the strange
Glossary (GAAG)
Acorr : The measured elastic asymmetry that has been corrected for helicity
Ainel : The asymmetry for inelastic protons that must be removed from the cor-
Ameas : The experimentally measured raw elastic and inelastic proton elec-
troweak asymmetry.
127
128
BCM: The Beam Current Monitor (BCM) is a cylindrical cavity whose reso-
nant frequency is adjusted to 1497 MHz (the frequency of the typical CEBAF beam).
Inside the cavity is a loop antenna located where the electric field is minimum and
the magnetic field is at a maximum. The antenna is coupled to one of the resonant
modes of the cavity and the output signal is proportional to the beam current.
BPM: The Beam Position Monitor (BPM) consists of four metal strips sur-
rounding the beamline. When the electron bunches pass through the BPM, a signal
is produced by induction. The output signals from the four strip lines are then
CFD: The Constant Fraction Discriminator (CFD) are designed to produce ac-
curate timing information from analog signals of varying heights but with the same
rise time. This will reduce the “walk” of the output signal.
verter, Histogramming, 16 channels, and X is for VXI standard) is the French elec-
tronics.
FPD: The Focal Plane Detectors (FPD) are 16 iso-Q2 double layered scintilla-
GMS: The Gain Monitoring System (GMS) shines laser light onto the scintil-
IA: The Intensity Attenuator (IA) is a charge feedback device that controls the
IHWP: The Insertable Halfwave Plate (IHWP) is used to change the helicity
of the polarized light coming from the G0 laser on the laser table in the injector.
LTD: The Latching Time Digitizers (LTDs) are specialized electronics used to
bin detector signals into 1 ns time bins for inputs to the scalers.
MPS: A Macro-pulse (MPS) is one helicity state that lasts for 33 ms (1/30
second).
NPN: Next Pulse Neutralization (NPN) is the disabling of the LTDs for the
next beam burst (32 ns) after a hit has been recorded from a detector. This allows
the mean-timers to clear and allows for an exact calculation of the deadtime.
PZT: “PZT” is an acronym for Lean Zirconate Titante. This device is a mirror
attached to a piezo-electric mount that changes the angle of the laser beam before
entering the Pockels cell. This is used to minimize helicity correlated position dif-
ferences.
try can be formed. The helicity of the first macro-pulse is chosen pseudo-randomly
with the next two macro-pulse helicities the complement of the first macro-pulse.
The fourth macro-pulse is the same helicity as the first macro-pulse. This allows for
RHWP: The Rotatable Halfwave Plate (RHWP) is used to minimize the charge
asymmetry by rotating the residual linear component of the slightly elliptical light
SMS: The Superconducting Magnet System (SMS) is the 1.6 T·m magnet used
in the G0 experiment.
ToF: The Time-of-Flight (ToF) is amount of time it takes for a particle to reach
Plane Detectors
each of the four photo-multiplier tubes (PMT) attached to the light pipe pair are
where :
- gain(HV ) is the gain2 which increases as the HV applied to the PMT increases,
1
A photo-electron is an electron produced by photo-electric effect when a (scintillation) photon
hits the PMT photo-cathode.
2
The gain of a PMT is the factor of amplification of the photo-electrons through the dynodes
chain of the PMT.
131
132
- Acable is the attenuation of the signal through the cables between the PMT output
threshold, one may consider increasing the gain. This option seems the most efficient
although its long term effect in a radiation harsh environment should be considered.
A large gain could produce a large anode current and as a consequence an early
following section.
where :
- ∆E is the energy lost by the particle passing through the scintillator, and C is
the conversion factor between energy lost and photon produced. Those quantities
are characteristic of the scintillator type and the energy of the detected particle,
is the fraction of photons exiting the light guide end that actually hit the PMT
133
photo-cathode.
- Ascint is the attenuation of photons along the scintillator from the hit position to
the glue joint between the scintillator and the light guide. A global is the attenua-
tion of the photons going through the glue joint between the scintillator and the
light guide and the bulk attenuation along the light guide.
have been performed during the assembly of the detectors; they took place at JLab
in the so-called Clean Room. They are described in the following section. For
clarity, it has been decided to separate the characterization of the light pipe and the
characteristic of the PMT. Therefore, the following results are given for an arbitrary
but constant value of P M TQE and P MTcontact as described later. Ascint and Aglobal
can vary from one light pipe to the other as they are a function of the quality of the
scintillator surface and the length of the scintillator (A scint ), or the global quality of
the light guide and the gluing between the scintillator and the light guide (A global ).
As soon as the light pipe have been mounted on one octant, the ends of the
light pipe were equipped with PMTs and the assembly was rolled into a dark space.
The signals were produced either by a radioactive source placed on the scintillator, a
LED illuminating the faces of the PMTs, or cosmic rays. The systematic calibrations
pipes were :
front of the PMT face. The brightness of the LED was reduced by lowering the
voltage applied to the LED down to a point were only one photon at a time
was detected by the PMT. The ADC signal measured in that case was very small
and usually overlapped with the pedestal of the ADC. Two independent methods
were used to amplify the single photo-electron signal. In one hand the signal was
is ∼ 10%. In the other hand, the single photo-electron peak was measured with a
large HV applied to the PMT, the calibration was then extrapolated to “regular”
∼ 10% due to the extrapolation. The two methods agree within error bars.
For the tests performed in the Clean Room, only eight PMTs were used. Each
PMT was mounted in the same position and tested. A photo-electron calibration
was performed for each of those PMT before almost all measurements using the
analog amplifier method. The gain of each tube was found to be constant in
the 10% error bar during the course of the measurements (almost a year). The
results in term of photo-electrons presented later are computed with the daily
The number of photo-electrons detected at the end of the light pipe depends
on the quality of the PMT used for detection because of the quantum effi-
ciency (P M TQE ) of the cathode. It also depends on the quality of the con-
tact (P M Tcontact ) between the tube and the light-guide. The reproducibility of
P M Tcontact was measured in the following way. The radioactive source was placed
on a scintillator, and was not moved during the whole course of the test. The
signal produced by the source was measured in an ADC channel. The contact
Minimum Maximum
135
3
1
2
Point of Observation
FIG. B.1: Positions of interest along the G0 scintillator. A signal produced in position 2
will undergo a maximum attenuation in its travel to the point of observation. Position 3
is the geometrical middle of the scintillator, a signal produced at this position therefore
undergo an average attenuation to the point of observation. Signals produced in position
1 undergo a minimal attenuation into the scintillator before being detected.
between the PMT face and the light pipe was broken by removing the PMT
from its housing. The silicon cookie was unstuck from the face of the PMT. The
PMT and the cookie were then put back in the housing therefore creating a new
contact between the PMT face and the light pipe. The signal of the radioactive
source was re-measured. The process was repeated ten times and the ADC value
was found to be consistent within 5%. The measurement of the relative tube
quantum efficiencies (P M TQE ) was performed using the same protocol. Again,
the radioactive source stays fixed. The output signal is measured in terms of
photo-electrons for different tubes inserted in the housing. It was found that
the number of photo-electrons produced by four different tubes used during this
testing differed by 15%. In the following, the different relative quantum efficien-
cies are corrected to normalize the signal measured with different tubes between
themselves. Nevertheless, one should consider this uncertainty when quoting the
Measurements
least 40 photo-electrons, the signal in the worst case must be determined. Prelim-
136
inary measurements [87] and simulation [88] have shown that the worst case is for
the particle hitting the far end of the scintillator at the bottom of the scintillator
(position 2 in Figure B.1). The attenuation of the signal along the scintillator as
well as the attenuation along the light pipe or through the glue joints were also
of Figure B.1. Those data allow us to differentiate between bad glue joints and
a bad scintillator. The data of attenuation along the scintillator can be compared
to data after data-taking with beam to indicate possible yellowing of the scintillator.
along the scintillator. The ratio of the signal for position 2 and 3 (see Figure B.1)
gives the attenuation along the scintillator. The measurement in position 3 gives
another one. For this measurement, a Ru-106 source [89] was used. It emits betas
the betas to traverse both scintillators and the plastic spacer) and discriminator
threshold, one is able to select only the most energetic part of the beta spectrum.
The energy deposited by the beta is 1.97 MeV per cm. Those measurements have
been carried out for 41 scintillator pairs out of the total of 64.
This method is tedious and cannot be used once the detector is enclosed in
the light tight box. Part of the calibration of the light pipes were carried out using
cosmic (µ) rays with the octants oriented such that the scintillators were concave
towards the earth and the scintillator faces perpendicular to the ground. The cosmic
trigger required all four PMTs on one detector to fire and two of the PMTs from an
adjacent detector to fire. This method is quicker as one can test many light pipes
137
at a time. Also, this method can be used once the detectors are in the hall (and
enclosed in the light tight box). While the energy deposited (per cm) by cosmic
rays is known, the path the cosmic takes through the scintillators is not known.
For this reason, the signals produced by cosmic rays were calibrated by comparing
them with the signals produced by the Ru source located in position 3 on Figure B.1.
that one can use a constant multiplicative conversion factor between the cosmic and
B.2. The 10% dispersion in the data, is mostly due to the precision which which
the cosmic peak was located on the ADC distributions. No significant variation
of this ratio was found as a function of the size of the scintillator. Also, the data
for cosmic runs were taken with the same HV being applied on the PMT, thus the
this method, one can measure the average number of photo electrons produced by
the light pipes. It is possible to measure the attenuation of the signal along the
scintillator using the cosmic ray data. A careful measurement of the time arrival of
signals between each other allows us to locate the hit position. Though the principle
was demonstrated, the quality of the data taken at that time did not allow us to
extract this information. The cosmic method was applied to the 23 scintillator pairs
Raw results of the testing are presented in the upper plots of Figure B.3. The
Measurements
3.575 / 4
Constant 18.76 3.709
Mean 0.8256 0.1569E-01
20
Sigma 0.1031 0.1362E-01
17.5
15
12.5
10
7.5
2.5
0
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
Ru/Cosmics
FIG. B.2: Calibration of the signal produced by non-collimated cosmic rays crossing the
scintillator against the signal produced by the Ru source located in the middle of the
scintillator. Those data represent the calibration of the signal for all scintillators of
Octant 7.
The analysis of the raw data shows that scintillators of the same size with the
same light-guide configuration3 gives in average similar results (see lower plots of
Figure B.3). Moreover the dispersion of 10% on Nγe can be interpreted as a result of
shows that this attenuation is independent of the scintillator size and is found to be
1.44 ± 0.05. As a result, the average values (see Figure B.3: middle plots and Table
The following equation describes the extrapolation of the Clean Room data to
the scintillators.
detected : ∆E(g0) = ∆E(Ru) is the energy lost by minimum ionizing particles. For
the G0 forward running, ∆E(g0) is lost by low-energy protons hitting the scintilla-
tors. The variation of energy lost by protons from one scintillator size to the other
one is significant [88]. For example the energy lost by the proton in scintillator 5
cm thickness) is on average 3.0 MeV. Moreover, this energy loss can vary by up to
25% across the face of a single scintillator; the minimum energy lost is for protons
crossing the scintillator on the top as they are the more energetic. This minimal
energy loss is considered for the computation of N γe (g0). This is the absolute worst
case scenario, as one is combining the least energy deposited (top of the scintillator)
with the worst transmission (bottom far end of the scintillator). The energies taken
into account for the extrapolation to the G0 forward running are given in Table B.2,
they have been computed using the Bethe-Bloch formula. The computation takes
into account the different materials crossed by the protons before hitting the scintil-
lators ; the main losses occur in the LH2 target, the air gap and, when relevant, the
140
Ascintfar
400 2.5
Nγe(mid)
2
300
1.5
200
1
100
0.5
0 0
0 5 10 15 0 5 10 15
scint. # scint. #
<Ascintfar>
400 2.5
<Nγe(mid)>
2
300
1.5
200
1
100 Front detectors
0.5
Back detectors
0 0
0 5 10 15 0 5 10 15
scint. # scint. #
100 100
events
events
8.254 / 10 5.889 / 5
Constant 61.99 5.296 Constant 64.86 7.219
Mean 0.9976 0.7218E-02 Mean 1.005 0.5681E-02
80 Sigma 0.1116 0.6241E-02
80 Sigma 0.6764E-01 0.5218E-02
60 60
40 40
20 20
0 0
0.5 1 1.5 0.5 1 1.5
Nγe(mid)/<Nγe(mid)> Ascintfar/<Ascintfar>
FIG. B.3: Results of the testing with minimum ionizing particles. The tests were per-
formed with the Ru radioactive source or with cosmic rays, following a procedure described
in Section B.1.1. The left plots refer to the average number of photo-electrons, while the
right ones refers to the attenuation of the signal from the far end of the detector to the
average signal. Upper plots show the raw data, middle plots show the average value on a
scintillator size to scintillator size basis, lower plots show the deviation of the actual data
(upper plot) to the average value (middle plot) again on a scintillator size to scintillator
size basis.
141
FIG. B.4: Minimum number of photo-electrons detected by the G 0 NA FPD in the case
of the proton forward angle running running. Error bars on these predictions are 20%.
Note that detector 4 is made up of 1 cm thick front scintillator and 0.5 cm thick back
scintillator. The horizontal line indicates the design minimum value of 40 photoelectrons.
produced in the NA-FPD light pipes in the cases of the G 0 running is presented
in Table B.2 as well as in Figure B.4. The error associated with this estimation
is the quadratic sum of the precision on the photo-electron calibration (10%), the
estimation on the stability (from one tube to another one) of the quantum efficiency
of the PMTs (15%) and the precision of the energy loss computation (10%). The
North American Focal Plan Detectors (NA FPD) is always larger than 40 photo-
running is always larger than 100. This minimum should allow for the time of flight
The next section explains how the PMT were matched to the light pipes.
142
Scintillator 9 10 11 12 13 14 15 16
< Nγe (mid) > front 240 213 198 207 232 166 170 166
< Nγe (mid) > back 189 187 152 165 200 145 148 147
< Afscint
ar
> 1.46 1.62 1.50 1.35 1.47 1.93 1.69 1.74
TABLE B.1: Results of the testing with minimum ionizing particles. N γe front,Nγe back
and Afscint
ar
(and their errors) are defined in the text.
Scintillator 9 10 11 12 13 14 15 16
∆E(g0) front 7.58 6.99 6.41 5.78 5.23 4.73 4.10 3.68
∆E(g0) back 8.31 7.53 6.79 6.03 5.41 4.84 4.16 3.71
Nγe (g0) front 633 467 430 450 419 207 209 178
Nγe (g0) back 546 441 349 374 374 185 185 159
TABLE B.2: Minimum energies left (in MeV) by the protons (G0 forward running) hit-
ting the G0 scintillators and subsequent number of photo-electrons detected. The energies
left take in account the variation of scintillator thickness.
143
B.1.2 PMT attributes
To understand the characteristics of the scintillator and light pipes, the previ-
ous measurements were performed using the same eight PMTs. The characteristics
of those eight PMT were very well known : their gains were tracked on a regular
basis, and their relative quantum efficiencies measured. After these tests had been
Gain Matching
The goal of the gain matching process was to pair the PMTs to specific light
pipes such that if one applies a given HV value to all of them, the output signal will
be roughly equivalent. In other words, one tried to compensate for the variation in
the number of photo-electrons (Nγe (g0)) produced by the G0 particles (see Figure
B.4) by carefully choosing the gain (gain(HV )) of the PMTs at a given HV. That
is :
For this computation, the PMT gains measurements performed at JLab [90]
were used. Those measurements were performed using a regular resistive basis and
not the final Zener-resistance G0 basis. The number of photo-electrons (Nγe (g0))
used for this pairing are the ones corresponding to the forward angle setup,in which
protons will be measured. There was nearly a perfect gain match for detectors 1
through 12 (see Figure B.5). The last three detectors required a gain that could not
of photo-electrons. The PMTs on these later detectors will have a higher voltage
1.6 Octant 1
Octant 3
1.4 Octant 5
Octant 7
1.2
0.8
0.6
0.4
0.2
0
0 2 4 6 8 10 12 14 16 18
Detector
FIG. B.5: Ratio of the gain of the actual PMT attached to the light pipe to the gain
necessary to perfectly match the light pipe. For detectors 13 through 16, not perfectly
matched PMT have to be used. This implies that the high voltage will be higher on these
detectors in comparison with the other detectors.
events
35 6.984 / 8
Constant 28.87
Mean 0.9661
Sigma 0.3013
30
25
20
15
10
0
0 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25 2.5
predicted gain / measured gain
FIG. B.6: Dispersion of the ratio of the measured to the predicted cosmic amplitudes in
the Clean Room. The 31% error is due to the gain measurements, the assumed quantum
efficiency, and to the fitting of the cosmic and pe peak for the Ru.
145
Gain Balancing
The optimization described in the previous section is based on the use of various
data sets. It also supposed that the quantum efficiency of all PMTs is equivalent.
In order to check the validity of the previous matching as well as the running of
the bases and the tubes, a cosmic measurement was performed on all the light pipe
equipped with their proper PMTs. In addition, cosmics were available in both the
Clean Room and down in Hall C. This meant that results in the Clean Room could
be compared with results after moving the octants into Hall C. The amplitude of
the signal produced by cosmics rays going through whatever scintillators is brought
Using the procedure described in Section B.1.2, one is able to compute the HV which
should be applied to the tube so that the cosmic peak will have a given amplitude 4 .
In this case one should take care of correction factors not considered in Equation
B.4, that is :
where P M TQE is the quantum efficiency of the tube and Acable is the attenuation
of the signal through the cable between the PMT and the ADC channel. For this
measurement Acable was known. The variation5 of the attenuation from one cable
Figure B.6 shows the ratio of the signal amplitude produced by the actual light
pipe with the amplitude expected. The ratio is distributed as a Gaussian of sigma
31%. The larger part of this dispersion can be explained by taking in account the
10%.
- The knowledge of the gain of each tube : 15%. This value is estimated by
comparing the gain of the eight tubes used for the initial calibration with the one
given in [90].
This yields an error of 25%. Thus the gain balancing process appears as a successful
cross check of our previous work and of the assumptions made during testing.
From these measurements the high voltage was then adjusted until the cosmic
peaks were aligned within 10%. From this data, the high voltage can also be adjusted
for the G0 proton case for the experiment. In that case, one should take in account
Injector Studies
The G0 commissioning run began in August 2002 until the end of January 2003.
This was an opportunity to test many of the systems under G0 running conditions.
This appendix discusses results of the commissioning run to understand the helicity
correlated devices at the source that were monitored by the injector DAQ system.
The BPMs used in the injector are the standard JLab stripline BPMs. These
BPMs are made up of four antennae situated symmetrically about the beam pipe.
When the electron beam passes through the stripline BPM, rf signals are picked up
by the monitors:
The position (or position difference) calculated from beam monitors have a
certain amount of noise associated with it. This noise is due to beam noise and to
FIG. C.1: The top two plots show the BPM resolution. The bottom two plots are the
profile plots of the top scatter plots. These profile plots can now be fitted and one axis
can be used to predict the position of one BPM from the measured position in another
BPM.
2 2 2
σmeasured = σbeam + σinstrumental .
The instrumental noise can be found by using three BPMs along the beamline
without magnetic optics between the monitors. The first two BPMs can be used to
determine the position of the beam in the third monitor. This predicted behavior
can then be removed from the measured signal of the third BPM leaving behind
only the instrumental noise. This analysis has been done when looking at absolute
positions and position differences. See Table C.2 and Table C.1 for results.
Another method for finding the instrumental noise of the BPM is to take a
run without beam but at a gain comparable to when beam is present. This is just
a typical pedestal run in ’Forced Gain Mode’, then in the analyzer code a typical
beam signal is injected into the BPM ntuple channels. Results of this analysis can
149
be found in Table C.3.
FIG. C.2: These are plots of the instrumental noise of the BPMs. After predicting the
position of the beam in a BPM from the above plots, the predicted position of the beam
can be subtracted from the BPM signal leaving behind only the instrumental noise of the
BPM.
It is not known why these results do not agree with one another. One reason
why these results might be inconsistent is that it takes two BPMs to predict the
location of the beam in a third BPM. This was not done in the above analysis since
there are magnetic elements between many of the BPMs in the injector.
150
From the 2001-2002 G0 commissioning run, the following PZT calibration slopes
The PZT was calibrated with only G0 beam in the injector on three days during
sample. The PZT response as measured in Hall C showed erratic behavior that was
not understood at the time. G0 will investigate further the response of the PZT on
the bench between the end of the February 2003 and the beginning of the second
FIG. C.3: These are plots of the helicity correlated position differences in X and Y as a
function of PZT X and Y motion and as a function of beam monitor in the injector.
155
Adiabatic Damping (Transverse Magnification)
If the energy of the electron beam increases much slower (adiabatically) than
the betatron oscillations, then the normalized emittance will remain constant while
the unnormalized emittance (that is the actual beam size) will shrink. This means
that the transverse size of the beam as measured in Hall C should be smaller than
what is measured in the injector. During the HAPPEX experiment there was so
much adiabatic damping that they did not need to run position feedback. The adia-
batic damping should be about a factor of 20 in suppression between the injector and
Hall C. This was demonstrated during the HAPPEX running. Besides benefiting
from the smaller position differences in the hall, adiabatic damping is an indicator
March 18,2002. At this time, the adiabatic damping was defined as:
Measurements were made with both the Hall A diode laser and a homemade
Ti:Sapphire laser, the following was observed for each laser respectively:
1100±100nm
• 24568±1373
∼ 0.045 ± 0.005 →∼ 22 ± 2
1200±70nm
• 34380±1584
∼ 0.035 ± 0.003 →∼ 29 ± 3.
sure the adiabatic damping between the injector and Hall C. The reason for so few
opportunities was due to the fact that other experimental halls were operational
meaning the other hall’s beam would be present in the injector making the mea-
surement difficult to do. From April 2002, BPM 0L02 was the reference BPM in the
injector due to the fact that this BPM had the largest response. This response has
156
changed during the commissioning run and all the BPMs in the injector that were
The method for observing the adiabatic damping had further been improved for
the commissioning run. Instead of only looking at the position differences between
BPMs in the injector and in the hall the G0 PZT mirror located on the laser table
was utilized. Now the ratio of the responses of the PZT mirror in X and Y are used
√
2 2
Adiabatic Damping in X = √XX +Y X
XX 2 +Y X 2
√
2 2
Adiabatic Damping in Y = √XY +Y Y
XY 2 +Y Y 2
where:
the factor of 20 that is expected. This might be due to the fact that 0L02 now sits at
157
FIG. C.4: Adiabatic damping is defined to be the ratio of position differences in the
injector (quadrant photodiode, BPM 1i02, 1i04, 1i06, and 0L02) to the hall monitor
(BPM G0B). The adiabatic damping as measured on the three dates indicated. Data
taken on December 19, 2002 was at a beam current of 20 µA. Data taken on January
16, 2003 was at a beam current of 5 µA. The data taken on January 24, 2003 was at a
beam current of 40 µA. The top plot is the adiabatic damping as one varies the x PZT.
The bottom plot is the adiabatic damping as one varies the y PZT.
the waist of the beta function of the machine when these measurements were made.
After a better tune of the machine can be found for the G 0 beam, work should be
done to find which BPM is most sensitive in the injector and use that as a reference.
158
C.1.3 Intensity Attenuator Cell
The IA responded well during the commissioning offering a large (∼ 400 ppm/V)
charge asymmetry calibration though it did generate large position differences. The
IA calibration constants were very stable over time requiring new measurements
of this constant every couple of days. Work to minimize the position differences
FIG. C.5: Charge asymmetry calibration constants (ppm/V) for the IA. Note that the
values at the QPD are consistent with one another over both months. Note the values
from 1I02 to 1I06 are consistent with one another. Between 1I06 and 0L02 are a variety
of apertures such as A1, A2, the Chopper, etc. It is due to scraping on these apertures
that
BIBLIOGRAPHY
[1] J. Gasser, H. Leutwyler and M.E. Sanio, Phys. Lett. B253, 252 (1991).
[6] G.P. Zeller et al., Phys. Rev. Lett. 88, 091802 (2002).
[7] S. Okubo, Phys. Lett. B5, 165 (1963); G. Zweig, CERN Report 8419/TH412
(1964); I. Iizuka, Prog. Theor. Phys. Suppl. 21, 37 (1966).
[15] E.J. Beise et al., Nucl. Instrum. Meth. A378, 383 (1996).
160
161
[17] N.F. Mott, Proc. Roy. Soc. A124, 425 (1929).
[22] A.I. Akhiezer and M.P. Rekalo, Sov. J. Part. Nucl. 3, 277 (1974); R. Arnold,
C. Carlson, and F. Gross, Phys. Rev. C23, 363 (1981).
[29] B. Milbrath et al., Phys. Rev. Lett. 80, 452 (1998); erratum Phys. Rev. Lett.
82, 221 (1999).
[30] M.K. Jones et al., Phys. Rev. Lett. 84, 1398 (2000).
[32] P.G. Blunden, W. Melnitchouk, J.A. Tjon, Phys. Rev. Lett. 91, (2003).
[42] F. Halzen and A.D. Martin, Quarks and Leptons(John Wiley & Sons, New
York, 1984).
[45] G.P. Ramsey et al., Int. J. Mod. Phys. A18 , 1211 (2003)
[47] E. Hadjimichael, G.I. Poulis, and T.W. Donnelly, Phys. Rev. C45, 2666 (1992).
[48] L.Diaconescu, R. Schiavilla, and U. van Kolck, Phys. Rev. C63, 4007 (2001).
[50] H.-W. Hammer, S.J. Puglia, M.J. Ramsey-Musolf, and S.-L. Zhu, Phys. Lett.
B562, 208 (2003).
[51] F.E. Maas et al., Phys. Rev. Lett. 93, 022002 (2004).
[54] G. Niculescu, Resonant Cavities used as Beam Current Monitors, CEBAF In-
ternal Report (unpublished).
[56] K.B. Unser, The Parameteric Current Transfer, a Beam Current Monitor De-
veloped for LEP, CERN SL/91-42 (unpublished).
163
[57] P. Gueye, Status of the Actual Beam Position Monitors in Hall C Beamline,
CEBAF Internal Report CEBAF-TN-93-004 (unpublished).
[58] G. Krafft and A. Hofler, How the LINAC Beam Position Monitors “Work”,
JLab Technical Note CEBAF-TN-93-004 (unpublished).
[61] S. Lewis, Overview of the Experimental Physics and Industrial Control System:
EPICS, LBNL Internal Report (2000).
[62] L. Hannelius, G0-Geant Simulation of the Inelastic Background from the Target
Windows, Internal report G0-03-044 (unpublished).
[63] L.W. Mo and S. Tsai, Rev. Mod. Phys. 41, 205 (1969).
[66] D.H. Beck and R.D. McKeown, Ann. Rev. Nucl. Part. Sci. 51, 189 (2001).
[67] D.H. Beck and B.R. Holstein, Int. J. Mod. Phys. E10, 1 (2001).
[87] J. Roche et al., Status Report of the G0 Detector 8 Prototype Studies, Internal
report G0-99-032.
[88] H. Breuer, Photo-electrons for all detectors, Tosca 4.5 shapes, V7, Internal
report G0-99-012.
[90] R. Woo and D. Armstrong, Spreadsheet of PMT Test Results, Internal report
G0-00-31.
VITA
Jeffery Allen Secrest was born in Cincinnati, Ohio on November 22nd 1973 to
Mike and Joy Secrest. He received a Bachelor of Science degree in physics at the
University of Cincinnati in 1997. In the fall of 1997 he begun his graduate career
1998 he won the Outstanding Teaching Assistant award. In the fall of 2000 he
entered the College of William and Mary as doctoral candidate in the Department
of Physics. Along the way, he earned a Masters of Science in 2001. This dissertation
was defended on December 17th 2004 at the College of William and Mary.
165