RotaMASS 3 Series Coriolis Mass Flow & Density Meter PDF
RotaMASS 3 Series Coriolis Mass Flow & Density Meter PDF
RotaMASS 3 Series Coriolis Mass Flow & Density Meter PDF
3 Series
Specifications Coriolis Mass Flow and Density Meter
GS 01R4B04-00E-E
01R04B04-00E-E
FEATURES
Contents • ROTAMASS is a flowmeter for nearly all fluids, including
Features Page 1 high viscosity liquids, slurries and multi- phase media
Principle of measurement Page 1
with a certain gas content
Performance specifications Page 2
• Field-mount and rack-mount remote converter available
Normal operating conditions Page 4
Mechanical specifications Page 5
• Refined digital signal processing enables accurate and
Remote cable specification Page 5
stable measurement
Electrical specifications Page 6
• A special detector decoupling system makes the device
Hazardous area specifications Page 7
highly independent from external loads or vibrations.
Planning and installation hints Page 12 • Simple flow path means self-draining, simple cleaning
Dimensions Page 15 and suitable for use in hygenic industries.
Model-, suffix- and option- codes Page 23 • High accuracy and high stability over a wide flow range
• Accurate density measurement, up to +/- 0.0005 g/cm³
• Concentration measurement for solutions, suspensions
ROTAMASS is a Coriolis mass flowmeter with highly
and emulsions (e.g. water cut, net oil computing)
refined digital signal processing electronics, so that accurate
• Volume flow with reference density
and stable mass flow measurement is achieved.
• Temperature measurement
• Two analog outputs, two pulse outputs or status-out and
ROTAMASS employs a flame-proof type converter case
one status-in as standard I/O
suitable for use in the hazardous area together with it’s
• Available in explosion proof versions (ATEX, FM, IECEx,
intrinsically safety type detector.
EAC, INMETRO, NEPSI, KOSHA)
• European MID approval acc. OIML R-117-1 and American
ROTAMASS´s signal processing, housing protection and its
NTEP approval for Custody Transfer Measurement
detector´s special decoupling system against external loads and
(see GS 01R04B07-00E-E)
vibrations, realize high performance in real applications.
• Wide process temperature range –200°C to 350°C
• Microprocessor-based multifunction capability
• EEPROM protects parameter settings and totalized
PRINCIPLE OF MEASUREMENT values during power failure.
Mass flow measurement according to the Coriolis principle. • High visibly LCD display
For abrasive or highly corrosive fluids please contact your • HART® communication function is available
Yokogawa representative. • Optional Foundation™ Fieldbus communication is available
(see GS 01R04B05-00E-E)
• Optional MODBUS® communication is available
• Optional intrinsically safe outputs are available
• Various choice of tubes materials
• EN, ASME or JIS flanges as standard, others on request
• FMEDA report on request, SIL 2, SIL 3 levels can be
achieved if redundant meters are used.
0.2
0.1
0
RCCS31 0.004 g/cm³ 0.001 g/cm³ ------
-0.1
-0.2
-0.3
RCCS32 0.004 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
-0.4
-0.5
-0.6
RCCS33 0.004 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
-0.7
-0.8 RCC34 0.003 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
-0.9
-1.0
0 50 100 150 RCC36 0.0022 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
Flow in % of Qnom F10.EPS
RCC38 0.0015 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
Table 2 : Zero stability RCC39 0.0015 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
3
Heat Tracing
Heating with heat carrier, insulation and protection housing.
Typically the max. surface temperature at the protection
housing from inner heating is 40°C (at Tamb = 20°C). Above
150°C (302°F) process temperature insulation from the
manufacturer is recommended. However up to 230°C (446°F)
process temperature the customer can insulate the detector
himself. For this case order option /S2.
Option /T1 : only insulation and protection
Option /T2 : insulation, protection and heating line
Option /T3 : like /T2 but with ventilation
Process connection for the heat carrier fluid (see table 10):
for D-type flanges : EN DN 15 PN 40 Form B1
for A-type flanges : ANSI ½ - 150 lbs.
for J-type flanges : JIS DN15 10K
Max. pressure : PN 40
Protection class : IP54, install roof protected
For fluid temperatures below -70°C select option /LT.
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
4
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
5
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
6
Setting Functions
ELECTRICAL SPECIFICATIONS Parameter setting is possible by using the infrared switches
Power Supply on the display or with HART® communication. For comfortable
- AC- type : 90 V to 264 V AC setting we recommend to use Yokogawas Field Mate setting
- AC- Ex-type : 90 V to 250 V AC tool (DTM based).
- DC- type : 20.5 V to 28.8 V DC
The option /MB2 can be set via Modbus and display.
Power consumption : max. 25 VA / 10 W
The option /MB3 can be set via Modbus, HART® and display.
External circuit breaker rating : 5 A, 250 V (The converter
doesn’t feature an installed power switch).
Display Function
Fuse on Base Board - Up to 4 lines.
- AC- type : 2 A, T, breaking capacity 1500A - 4 selectable languages (English, German, French, Russian)
- DC- type : 2 A, T, breaking capacity 1500A - Option /MB only available English
I/O Signals, including /KF1, /EF1, /UF1, /NF1, /GF1 Damping Functions
- Two active current outputs: Iout1, Iout2 The damping function is adjustable from 0 to 200 seconds
4 to 20 mA DC, galvanic separated from other signals, and affects the display and outputs.
Load resistance : 20 Ω to 600 Ω
Ambient temperature effect : < 0.05% of span/10°C Isolation Resistance of Converter
Linearity : 0.008 mA = 0.05% of span When surge arrestors are removed
Setting range URV for liquids: - between power and ground terminal: >100 MΩ / 500 V DC
-Qmax to + Qmax , Abs(URV-LRV) ≥ 5% of Qnom - between power and I/O terminals : >20 MΩ / 100 V DC
Setting range URV for gases: - between I/O terminals and ground : >20 MΩ / 100 V DC
-Qmax to + Qmax , Abs(URV-LRV) ≥ 1% of Qnom
- Two pulse outputs / status outputs : Pout, Sout Dielectric Strength
Passive transistor contact output, 30 V DC, 200 mA When surge arrestors are removed
Output rate - between power and ground terminal : 1,500 V AC for 1 minute
Output 1 : 0 to 10000 pulses/s
Output 2 : 0 to 2000 pulses/s Lightning Protection
Option /NM : passive, according EN 60947-5-6 Arresters 2000 A are inside of the converter for power
Option /AP : active output, 12 V, 6 mA, RL> 10 kΩ supply lines.
Active pulse output is not isolated from current output 2
As frequency output Electromagnetic Compatibility
Output 1 : 20 Hz to 10000 Hz Acc. IEC 61326-1: Class A, Table 2
Output 2 : 20 Hz to 2000 Hz IEC 61326-2-3
- Status input : Voltage-free contact IEC 61000-3-2
Closed : < 200 Ω IEC 61000-3-3
Open : > 100 kΩ
Safety Requirement Standards
Option /KF5, /EF5, /UF5, /NF5, /GF5: 3 intrinsically safe
Acc. IEC 61010-1
outputs
IEC 61010-2-030
- Two passive current outputs (additional power supply
Overvoltage category II
needed) : Iout1, Iout2, 4 to 20 mA DC, galvanically separated
Pollution degree 2
from other signals.
Supply voltage : 10.5 V to 30 V DC (without HART®), 165 mA
Supply voltage : 16.75 V to 30 V DC (with HART®), 165 mA
Load resistance : 20 Ω ... 600 Ω
Ambient temperature effect : < 0.05% of span/10°C
- One pulse output / status output : Pout
Passive transistor contact output, 30 V DC, 100 mA
Output rate : 0 to 2000 pulses/s
As frequency output : 20 Hz to 2000 Hz
- No status input
Digital Communication
- HART® communication protocol rev. 5, superimposed on
4 -20 mA DC signal (Iout1)
- Load resistance : 230 Ω to 600 Ω (including cable)
- Power line spacing : >15 cm, avoid parallel wiring
- Cable length : ≤ 2 km if „CEV” cables are used
- FOUNDATION™ Fieldbus communication (/FB)
- see GS 01R04B05-00E
- MODBUS® communication (/MB2 + /MB3)
- Physical interface RS485 as two wire data bus according
EIA/TIA-485
- Maximal bus length is depending from bus topology and
communication speed.
- Addresses: 1 to 247
- Baud rates: 1200, 2400, 4800, 9600, 19200, 38400,
57600, 76800, 115200
- Data formats: E/1, O/1, N/1, N/2 (parity/stop bits)
- Modes: RTU, ASCII
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
7
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
8
Electrical data remote detector RCCS30LR ... 33 : FM (For USA and Canada)
- Driving circuit : terminals D+ and D
Ex ib IIC : Ui = 16 V; Ii = 53 mA; Pi = 0.212 W Remote detector RCCS30LR ... 39/XR (Option /FS1) :
Li = 4.2 mH; Ci = negligible small - Intrinsically safe
Ex ib IIB : Ui = 16 V; Ii = 153 mA; Pi = 0.612 W - AEx ia IIC, Class 1, Zone 0
Li = 4.2 mH; Ci = negligible small - IS Class I, Division 1, Groups A, B, C, D T6
- Sensor circuits: terminals S1+ and S1- or S2+ and S2- - DIP Class II / III, Division 1, Groups E, F, G
Ex ib IIC : Ui = 16 V; Ii = 80 mA; Pi = 0.32 W - IP67 / NEMA 4X
Li = 4.2 mH; Ci = negligible small - Ambient temperature range : -50°C to +80°C (-58°F to 176°F)
- Temperature sensor circuit : terminals TP1, TP2, TP3
Ex ib IIC : Ui = 16 V; Ii = 50 mA; Pi = 0.2 W Remote converter RCCF31 (Option /FF1, /FF5) :
Li = negligible small; Ci = negligible small - Housing explosion proof
- Provides Intrinsically safe detector circuits
Electrical data remote detector RCCS34 ... 39/XR : - AEx d [ia] IIC, Class I, Zone 1, T6
- Driving circuit : terminals D+ and D - AEx d [ia] IIB, Class I, Zone 1, T6 with option /HP
Ex ib IIC : Ui = 16 V; Ii = 53 mA; Pi = 0.212 W - Class I, Division 1, Groups A, B, C, D
Li = 3.2 mH; Ci = negligible small - Class I, Division 1, Groups C, D with option /HP
Ex ib IIB : Ui = 16 V; Ii = 153 mA; Pi = 0.612 W - Class II / III, Division 1, Groups E, F, G
Li = 3.2 mH; Ci = negligible small - AIS Class I / II / III, Division 1, Groups A, B, C, D, E, F, G
- Sensor circuits: terminals S1+ and S1- or S2+ and S2- - AIS Class I / II / III, Division 1, Groups C, D, E, F, G with /HP
Ex ib IIC : Ui = 16 V; Ii = 80 mA; Pi = 0.32 W
- IP67 / NEMA 4X
Li = 2.1 mH; Ci = negligible small
- Temperature sensor circuit : terminals TP1, TP2, TP3 - Ambient temperature range : -40°C to +50°C (-40°F to 122°F)
Ex ib IIC : Ui = 16 V; Ii = 50 mA; Pi = 0.2 W
Li = negligible small; Ci = negligible small Remote converter RCCR31 (Option /FS1) :
- Intrinsically safe associated apparatus
Electrical data remote converter RCCF31, RCCR31 and - Provides Intrinsically safe detector circuits
Converter of integral type RCCT3 : - [AEx ia] IIC, Class I, Zone 1
- Driving circuit : terminals D+ / D- - [AEx ia] IIB, Class I, Zone 1, T6 with option /HP
Ex [ib] IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W - AIS Class I, Division 1, Groups A, B, C, D
Lo = 15 mH; Co = 0.65 µF - AIS Class I, Division 1, Groups C, D with option /HP
Ex [ib] IIB : Uo = 11.7 V; Io = 124 mA; Po = 0.363 W - Ambient temperature range : -40°C to +50°C (-40°F to 122°F)
Lo = 8 mH; Co = 10.3 µF
- Sensor circuits: terminals S1+/ S1- or S2+ / S2- Integral type RCCT34 ... 39/XR (Option /FF1, /FF5) :
Ex [ib] IIB/IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W - Housing explosion proof
Ex [ib] IIC : Lo = 15 mH; Co = 0.65 µF - AEx d [ia] IIC, Class I, Zone 1, T6
Ex [ib] IIB : Lo = 60 mH; Co = 4.07 µF - AEx d [ia] IIB, Class I, Zone 1, T6 with option /HP
- Temperature sensor circuit : terminals TP1, TP2, TP3 - Class I, Division 1, Groups A, B, C, D
Ex [ib] IIB/IIC : Uo = 13.3 V; Io = 40 mA; Po = 0.133 W - Class I, Division 1, Groups C, D with option /HP
Ex [ib] IIC : Lo = 20 mH; Co = 0.91 µF - Class II / III, Division 1, Groups E, F, G
Ex [ib] IIB : Lo = 80 mH; Co = 5.6 µF - IP67 / NEMA 4X
- Current outputs (only option /KF5) : - Ambient temperature range : -40°C to +50°C (-40°F to 122°F)
Ex ia IIC : Ui = 30 V; Ii = 300 mA; Pi = 1.25 W
Li = 4 µH; Ci = 5.5 nF Process temperature range :
- Standard : -50°C to 150°C / (-58°F to 302°F)
- Pulse output (only option /KF5) : - with option /LT : -200°C to 150°C / (-328°F to 302°F)
Ex ia IIC : Ui = 30 V; Ii = 300 mA; Pi = 1.25 W - with option /MT (RCCS30LR...33)
Li = 4 µH; Ci = 13.1 nF : -50°C to 260°C / (-58°F to 500°F)
- with option /MT (RCC34...39/XR)
For temperature classification see table 7. : -50°C to 220°C / (-58°F to 428°F)
- with option /HT : 0°C to 350°C / (-32°F to 662°F)
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
9
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
10
Remote converter RCCF31 (Option /EF5) : Electrical data remote detector RCCS30LR ... 33 :
- Flame proof with intrinsically safe connection to detector (ib) - Driving circuit : terminals D+ and D
- Additional Intrinsically safe outputs. Ex ib IIC : Ui = 16 V; Ii = 53 mA; Pi = 0.212 W
- Ex d [ia Ga] [ib] IIC T6 Gb or Li = 4.2 mH; Ci = negligible small
Ex d e [ia Ga] [ib] IIC T6 Gb Ex ib IIB : Ui = 16 V; Ii = 153 mA; Pi = 0.612 W
- Ex d [ia IIC Ga] [ib] IIB T6 Gb or Li = 4.2 mH; Ci = negligible small
Ex d e [ia IIC Ga] [ib] IIB T6 Gb with option /HP - Sensor circuits: terminals S1+ and S1- or S2+ and S2-
- Ex tb [ia Da] [ib] IIIC T75°C Db Ex ib IIC : Ui = 16 V; Ii = 80 mA; Pi = 0.32 W
- Max. surface temperature : 75°C (167°F) Li = 4.2 mH; Ci = negligible small
- Degree of protection : IP66/67 - Temperature sensor circuit : terminals TP1, TP2, TP3
- Power supply : 90 to 250 V AC, 50/60 Hz or Ex ib IIC : Ui = 16 V; Ii = 50 mA; Pi = 0.2 W
20.5 to 28.8 V DC Li = negligible small; Ci = negligible small
- Power consumption : max. 25 VA / 10 W
- Ambient humidity : 0 to 95% RH Electrical data remote detector RCCS34 ... 39/XR :
- Ambient temperature range: -40°C to +55°C (-40°F to 131°F) - Driving circuit : terminals D+ and D
Ex ib IIC : Ui = 16 V; Ii = 53 mA; Pi = 0.212 W
Remote converter RCCR31 (Option /ES1) : Li = 3.2 mH; Ci = negligible small
- Associated apparatus with Intrinsically safe connection to Ex ib IIB : Ui = 16 V; Ii = 153 mA; Pi = 0.612 W
detector (ib) Li = 3.2 mH; Ci = negligible small
- [Ex ib Gb] IIC - Sensor circuits: terminals S1+ and S1- or S2+ and S2-
- [Ex ib Gb] IIB with option /HP Ex ib IIC : Ui = 16 V; Ii = 80 mA; Pi = 0.32 W
- [Ex ib Db] IIIC Li = 2.1 mH; Ci = negligible small
- Power supply : 90 to 250 V AC, 50/60 Hz or - Temperature sensor circuit : terminals TP1, TP2, TP3
20.5 to 28.8 V DC Ex ib IIC : Ui = 16 V; Ii = 50 mA; Pi = 0.2 W
- Power consumption : max. 25 VA / 10 W Li = negligible small; Ci = negligible small
- Ambient humidity : 0 to 95% RH
- Ambient temperature range: -40°C to +55°C (-40°F to 131°F) Electrical data remote converter RCCF31, RCCR31 and
converter of integral type RCCT3 :
- Driving circuit : terminals D+ / D-
WARNING Ex [ib] IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W
Remote rack-mount converter RCCR31 must be installed in Lo = 15 mH; Co = 0.65 µF
safe area ! Ex [ib] IIB : Uo = 11.7 V; Io = 124 mA; Po = 0.363 W
Lo = 8 mH; Co = 10.3 µF
Integral type RCCT34 ... 39/XR (Option /EF1) : - Sensor circuits: terminals S1+/ S1- or S2+ / S2-
- Flame proof with Intrinsically safe connection to detector (ib) Ex [ib] IIB/IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W
- Ex d ib IIC T6...T3 Gb or Ex [ib] IIC : Lo = 15 mH; Co = 0.65 µF
Ex d e ib IIC T6...T3 Gb Ex [ib] IIB : Lo = 60 mH; Co = 4.07 µF
- Ex d ib IIB T6...T3 Gb or - Temperature sensor circuit : terminals TP1, TP2, TP3
Ex d e ib IIB T6...T3 Gb with option /HP Ex [ib] IIB/IIC : Uo = 13.3 V; Io = 40 mA; Po = 0.133 W
- Ex ib tb IIIC T150°C Db Ex [ib] IIC : Lo = 20 mH; Co = 0.91 µF
- Max. surface temperature : 150°C (302°F) Ex [ib] IIB : Lo = 80 mH; Co = 5.6 µF
- Degree of protection : IP66/67 - Current output (only option /EF5) :
- Power supply : 90 to 250 V AC, 50/60 Hz or
20.5 to 28.8 V DC Ex ia IIC : Ui = 30 V; Ii = 300 mA; Pi = 1.25 W
- Power consumption : max. 25 VA / 10 W Li = 4 µH; Ci = 5.5 nF
- Ambient humidity : 0 to 95% RH - Pulse output (only option /EF5) :
- Ambient temperature range: -40°C to +55°C (-40°F to 131°F) Ex ia IIC : Uo = 30 V; Io = 300 mA; Po = 1.25 W
Lo = 4 µH; Co = 13.1 nF
Integral type RCCT34 ... 39/XR (option /EF5) :
- Flame proof with Intrinsically safe connection to detector (ib) For temperature classification see table 7.
- Additional Intrinsically safe outputs.
- Ex d ib [ia Ga] IIC T6...T3 Gb or
Ex d e ib [ia Ga] IIC T6...T3 Gb
- Ex d ib [ia IIC Ga] IIB T6...T3 Gb or
Ex d e ib [ia IIC Ga] IIB T6...T3 Gb with option /HP
- Ex ib tb [ia Da] IIIC T150°C Db
- Max. surface temperature : 150°C (302°F)
- Degree of protection : IP66/67
- Power supply : 90 to 250 V AC, 50/60 Hz or
20.5 to 28.8 V DC
- Power consumption : max. 25 VA / 10 W
- Ambient humidity : 0 to 95% RH
- Ambient temperature range: -40°C to +55°C (-40°F to 131°F)
- Process temperature range : -50°C to 150°C (-58°F to 302°F)
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
11
Table 7 : Temperature classification for ATEX, FM, IECEx, INMETRO, NEPSI and KOSHA certified flowmeter
RCCS30LR to RCCS33 RCCS30LR to RCCS33
without insulation with factory
insulation
Temp. class Max. ambient Max. process Max. ambient Max. process
temperature temperature temperature temperature
T6 50°C / 122°F 60°C / 140°F 60°C / 140°F 60°C / 140°F
T5 50°C / 122°F 80°C / 176°F 80°C / 176°F 90°C / 194°F
80°C / 176°F 100°C / 212°F
T4 80°C / 176°F 130°C / 266°F
50°C / 122°F 120°C / 248°F
T3 80°C / 176°F 180°C / 356°F 80°C / 176°F 180°C / 356°F
T2 80°C / 176°F 260°C / 500°F 80°C / 176°F 260°C / 500°F
T1 80°C / 176°F 260°C / 500°F 80°C / 176°F 260°C / 500°F
INMETRO APPROVAL (For Brazil) EAC APPROVAL (For Russia, Kazakhstan, Belorussia)
Certificate TÜV 11.0419 X Certificate RU C-DE.ГБ08.B.00208
RCCS3 with option /US1.same as IECEx /ES1 RCCS3 with option /GS1
RCCT3 with options /UF1 or /UF5 same as IECEx /EF1 or /EF5 RCCT3 with options /GF1 or /GF5
RCCF31 with options /UF1 or /UF5 same as IECEx /EF1 or /EF5 RCCF31 with options /GF1 or /GF5
RCCR31 with option /US1 same as IECEx /ES1 RCCR31 with option /GS1
Same parameters and specifications as IECEx approval. Same parameters and specifications as IECEx approval.
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
12
Recalibration Service
Yokogawa offers full recalibration service, if necessary with a
certificate traceable to German national standards. Please
contact your Yokogawa affiliate or directly Rota Yokogawa,
Germany.
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
13
Explosion Proof Concept The specified mass flow accuracy applies if the batch process
The detector is intrinsically safe Ex ib, the converter RCCT is >1 minute. For shorter batch time (Dt in s) the accuracy
and RCCF31 are flame (explosion) proof. The converter decreases with the square root of 60/Dt.
RCCR31 is an intrinsically safe associated apparatus. For short batches the opening and closing times of the valves
The driving power from converter to detector is limited and have to be longer than 2 seconds.
protected by an intrinsically safe barrier, which is part of the
converter. The barrier is protecting the detector either for gas Concentration Measurement for Liquids
group IIC or IIB (option /HP). The Standard Concentration Measurement (option /CST)
Option /KF5 delivers two passive intrinsically safe current and is suitable for concentration measurement of emulsions or
one pulse output, however the converter is flame (explosion) suspensions, where the density of the solid is assumed to
proof. be fix. It can also be used for (mainly low concentration)
solutions if the two fluids are not strongly interacting. The
Option /HP density change of the liquid components due to temperature
With option /HP the detector driving power is higher which is can normally be described with a linear or quadratic function
benefit to 2 phase flow. This is also true for non hazardous with very high accuracy within the desired measurement
applications. range. The coefficients of these function (linear and quadratic
thermal expansion coefficients) must be either known or have
Gas Measurement to be determined prior to using this function.
For gas applications please choose the option /GA. For interacting liquids the Advanced Concentration
Density reading below 0.3 kg/l is not possible. Volume flow is Measurement options should be used, these options can
calculated by using the fix density value stored in „Reference density“. be ordered using the appropriate /C concentration
Based on the selection of the gas density, the following volume measurement option. For more information please see
flow rates can be calculated; standard, reference, normal. TI 01R04B04-04E-E “Concentration Measurement with
Besides, the corresponding volume flow rate units can be ROTAMASS”.
selected. Some functions are unavailable for gas measurement,
including concentration measurement, empty pipe, slug or corrosion
detection.
Good and stress free installation is mandatory for a stable Zero.
Attention to resonance phenomenons has to be taken if gas
compressors are used in the pipe. Flow noise has to be avoided.
Batch Process
Density Measurement
There are 3 levels of density measurement. The standard adjustment and /K4 delivers an accuracy up to 0.001 g/cm³, if the fluid den-
sity is around 1 kg/l. However, at elevated temperatures the density error may increase. For option /K4 the instrument is
preheated ensuring long term stability. However, if high density stability is needed at high temperatures option /HT is recommended.
Option /K6 includes preheating, a full calibration at 3 different densities, increased temperature measurement specification and individual
adjustment of the fluid temperature dependency. Multi-phase flow can generate higher deviations. The higher the density differences
of the single components are the more likely it is that a negative density error is generated. Aeration has to be avoided fully to receive
good density measurement.
For more information please see TI 01R04B04-05E ”Density Measurement with ROTAMASS”.
Note: Density specification under calibration condition only with flow direction “forward” according the arrow on the meter.
Table 8: Overview density-/volume- flow measurement:
Option Accuracy Certificate Description Typical Application
Standard ± 0.0015 g/cm³ to Standard (mass flow) - Standard adjustment with water and air - Process medium and environment are
± 0.02 g/cm³ factory calibration - Density constants given in mass flow approximately at room temperature, the density
certificate certificate range is 0.9 kg/l to 1.1 kg/l
Option /K4 ± 0.001 g/cm³ Standard (mass flow) - Thermal treatment of the sensor and - Improved volume flow accuracy
factory calibration special hardware design - Process medium up to 150°C, for higher
certificate - Standard adjustment with water and air temperature select option /MT or /HT
- Density constants given in mass flow - Density range is 0.9 kg/l to 1.1 kg/l
certificate
Option /K6 ± 0.0005 g/cm³ Separate factory - Thermal treatment of the sensor and - Density and concentration measurement in
density calibration special hardware design addition to the mass flow:
certificate - Density calibration with 3 different liquids - Process medium up to 150°C, for higher
- Individual adjustment of the fluid temperature select option /HT
temperature dependency - Density range 0.3 kg/l to 2.5 kg/l
- Best volume flow accuracy
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
14
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
15
DIMENSIONS
Integral Type RCCT34, RCCT36, RCCT38, RCCT39
Without Display 266 (10.5)
Ø123 (4.84)
(1.93)
49
With Display
266 (10.5)
147 (5.79)
122 (4.8) 144 (5.67)
85 (3.35)
32 (1.26) 50 (1.97)
30
(1.18)
Ø123 (4.84)
49 (1.93)
H3
H2
(2.05)
52
H4
H1
L3 W1
L2
L1 ± 5 (0.2)
W2
Note: The flange dimensions depend on size and pressure rating of the flange.
Model L1 L2 L3 H1 H2 H3 H4 W1 W2 Weight
mm 13-24 kg
RCCT34 see table 11 272 (10.7) 212 (8.35) 177 (6.97) 214 (8.43) 279 (11) 80 (3.15) 60 (2.36) 80 (3.15)
(inches) (29-53 lbs)
mm 18-38 kg
RCCT36 see table 11 400 (15.7) 266 (10.5) 230 (9.06) 214 (8.43) 279 (11) 80 (3.15) 76 (2.99) 90 (3.54)
(inches) (40-84 lbs)
mm 28-53 kg
RCCT38 see table 11 490 (19.3) 267 (10.5) 269 (10.6) 224 (8.82) 289 (11.4) 100 (3.94) 89 (3.5) 110 (4.33)
(inches) (62-117 lbs)
mm 63-106 kg
RCCT39 see table 11 850 (33.5) 379 (14.9) 370 (14.6) 242 (9.54) 307(12.1) 135 (5.31) 129 (5.08) 160 (6.3)
(inches) (139-233 lbs)
Dimensions in mm (inches)..Weights with smallest and biggest flanges.
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
16
(4.86) (5.74)
(5.87)
(4.86)
(1.94)
(2.56)
(3.41)
(4.84)
(1.94)
(12.87)
(10.31)
(2.05)
(6.93)
(21.9)
(18.43)
(27.19)
(6.63)
(35.12)
(9.44)
240
(14.96)
52
380
(11.02)
280
1280 (50.39)
R3 .81)
(11
00
230
(9.05)
(11.02)
280
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
17
17 (0.66)
Ø102 (4.01) 75 (2.95)
201(7.91)
(4.72)
91.6 (3.6)
120
Ø123(4.84)
(2.51)
(12)
305
122 (4.80)
64
29 32 29
(1.14) (1.25) (1.14)
85 (3.34)
124 122
(4.88) (4.80) 148 (5.82)
With Display
Ø102 (4.01) 17 (0.66) 75 (2.95)
201(7.91)
(4.72)
120
91.6 (3.6)
Ø123 (4.84)
(12)
305
(2.51)
122 (4.80)
64
49 32 29
(1.92) (1.25) (1.14)
85 (3.34)
144 (5.66) 122 (4.80) 148 (5.82)
Dimensions in mm (inches)
Dimensios in mm
19-inch rack, acc. IEC / DIN EN 60297-3-101
Weight: 1.5 kg
Dimensions in mm
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
18
180 25 20 29 25 57 46
150
L1 ± 3 s. table 11
98
Ø102
110
150
Ø8,5
Ø89
111 (RCCS30LR)
111 (RCCS30)
99 (RCCS31)
89 (RCCS32) Option /PD
55 (RCCS33) Weight 6...18kg (13...40lbs) Mounting set for DN50 tube
Dimensions in mm
(3.14)
80
98 (3.85)
H6
H5
H4
H1
L3 W1 Option /S2
L2
W2
Note: The flange dimensions depend on size and pressure rating of the flange.
Model L1 L2 L3 H1 W1 W2 H4 H5 H6 Weight
mm 10-21 kg
RCCS34 see table 11 272 (10.7) 212 (8.35) 177 (6.97) 60 (2.36) 80 (3.15) 80 (3.15) 138 (5.43) 218 (8.58)
(inches) (22-46 lbs)
mm 15-35 kg
RCCS36 see table 11 400 (15.7) 266 (10.5) 230 (9.06) 76 (2.99) 90 (3.54) 80 (3.15) 138 (5.43) 218 (8.58)
(inches) (33-77 lbs)
mm 25-50 kg
RCCS38 see table 11 490 (19.3) 267 (10.5) 269 (10.6) 89 (3.5) 110 (4.33) 100 (3.94) 148 (5.82) 228 (8.97)
(inches) (55-110 lbs)
mm 60-103 kg
RCCS39 see table 11 850 (33.5) 379 (14.9) 370 (14.6) 129 (5.08) 160 (6.3) 135 (5.31) 166 (6.53) 245 (9.65)
(inches) (132-227 lbs)
Dimensions in mm (inches)..Weights with smallest and biggest flanges.
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
19
(2.2.8)
(3.85)
(10.47)
(7.32)
(6.93)
(21.9)
(21.9)
(27.19)
(6.63) Option /S2
(35.12)
(4.02)
(6.93)
(2.28)
L1 ± 5 (0.2) (see table 11)
(3.86)
(12.6)
320
98
58
(9.37)
238
(11.02)
280
(35.43)
900
(35.43)
900
(4.01)
Ø102
(11.02)
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
20
Ø50
(7.0)
178
(2.68)
(12.48)
(1.96)
317
68
(2.28)
58
(11.02)
280
(35.07)
891
683 (26.88) 273
(10.74)
1140 (44.88)
Option /S2
Ø102 (4.01)
(11.02)
280
Weight 290...370 kg
(639...815 lbs)
Dimensions in mm (inches). Weights with smallest and biggest flanges
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
21
Remote Detector RCCS34, RCCS36, RCCS38, RCCS39, RCCS39/IR- /V2 with option /T
(Insulation / Heating)
L1 ± 5 (0.2)
Ø102 (4.01)
H6
H9
Process
connection
D1 ± 5 (0.2)
H7
D2 ± 5 (0.2)
Heating
H8
connection
option /T2 or /T3
L5
Ventilation
L4 ± 5 (0.2) option /T3
W3
Note: The flange dimensions depend on size and pressure rating of the flange.
Model L1 L4 L5 D1 D2 H6 H7 H8 H9 W3 Weight
mm see table 420 310 330 218 411 273 138 240 19-30 kg
RCCS34 200 (7.87)
(inches) 11 (16.53) (12.20) (12.99) (8.58) (16.81) (10.74) (5.43) (9.44) (42-66 lbs)
mm see table 540 439 250 380 218 464 326 138 260 28-48 kg
RCCS36
(inches) 11 (21.25) (17.28) (9.84) (14.96) (8.58) (18.26) (12.83) (5.43) (10.23) (62-106 lbs)
mm see table 640 530 250 430 228 524 376 148 260 40-65 kg
RCCS38
(inches) 11 (25.19) (20.86) (9.84) (16.92) (8.97) (20.62) (14.80) (5.82) (10.23) (88-144 lbs)
mm see table 1000 894 350 545 245 668 503 165 302 104-147 kg
RCCS39
(inches) 11 (39.37) (35.19) (13.77) (21.45) (9.65) (26.29) (19.80) (6.49) (11.88) (230-325 lbs)
mm see table 1050 944 350 677 266 944 625 193 342 147-201 kg
RCCS39/IR
(inches) 11 (41.34) (37.16) (13.77) (26.65) (10.47) (37.16) (24.61) (7.6) (13.46) (325-444 lbs)
Dimensions in mm (inches)..Weights with smallest and biggest flanges including insulation cover and heat tracing.
Standard heating connection according table 10.
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
22
Remote Detector RCCS30LR, RCCS30, RCCS31, RCCS32, RCCS33 with option /T (Insulation / Heating)
L1 ± 3 (0.1) s. table 11 176 (6.93)
Ø102 (4.01)
350 (13.77)
425 (16.73)
299 (11.77)
Heating
connection
option /T2 or /T3
134 (5.27)
Ventilation
option /T3
180 (7.08)
160 (6.29)
270 (10.62)
Dimensions in mm (inches)
Weight in kg (lbs) with flange Table 10: Heat tracing connection types depending on
Weight in kg (lbs) without flange
(01A1) process connection type
/T1 /T2 /T3 /T1 /T2 /T3
8.7 11.5 12.5 9.5 12.3 13.3 Process connection Standard heating connection *)
(19.2) (25.3) (27.5) (21) (27.1) (29.3) A ASME ½” - 150
D EN DN 15 PN 40
J JIS 10K DN15
S2 ; S4 EN DN 15 PN 40
S8 ASME ½” - 150
G9 EN DN 15 PN 40
T9 ASME ½”- 150
*) others on request
N1 N2 ØD B ØC L1 L2 L Weight
[mm (in.)] [mm (in.)] [mm (in.)] [[mm (in.)] [mm (in.)] [mm (in.)] [kg (lbs)]
1/2´´ NPT G1/2´´ 15 (0.59) 27 (1.06) 30 (1.18) 19 (0.74) 23 (0.90) 45 (1.77) 0.07 (0.003)
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
23
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
24
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
25
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
26
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
27
Hazardous Area Approvals 1) /KF1 ATEX Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC
/KF5 ATEX Flame proof converter + Intrinsically safe detector output + Intrinsically safe with /HP not for gas group IIC
outputs 2 passive analog output + 1 passive pulse output
/FF1 FM approval for USA+Canada, Flame proof converter + Intrinsically safe detector only with cable conduit “A”; with /HP not for groups A
output + 2 active analog outputs + 2 passive pulse outputs + 1 status input and B
/FF5 FM approval for USA+Canada, Flame proof converter + Intrinsically safe detector only with cable conduit “A”; with /HP not for groups A
output + 2 passive analog outputs + 1 passive pulse output and B
/EF1 IECEx Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC
/EF5 IECEx Flame proof converter + Intrinsically safe detector output + Intrinsically with /HP not for gas group IIC
safe outputs 2 passive analog output + 1 passive pulse output
/UF1 INMETRO Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC
/UF5 INMETRO Flame proof converter + Intrinsically safe detector output + Intrinsically with /HP not for gas group IIC
safe outputs 2 passive analog output + 1 passive pulse output
/NF1 NEPSI Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC; only with /CN
/NF5 NEPSI Flame proof converter + Intrinsically safe detector output + Intrinsically with /HP not for gas group IIC; only with /CN
safe outputs 2 passive analog output + 1 passive pulse output
/GF1 EAC Flame proof converter + Intrinsically safe detector output with /HP for gas group IIB; only with /VE
/GF5 EAC Flame proof converter + Intrinsically safe detector output + Intrinsically safe with /HP for gas group IIB; only with /VE
outputs 2 passive analog output + 1 passive pulse output
Custody Transfer Measurement /Q01 European MID Approval (fluids other than water) Refer to GS 01R04B07-00E
/Q20 NTEP approval 12-080A2, Accuracy class 0.3 acc. NIST Handbook 44 Refer to GS 01R04B07-00E
/Q21 NTEP approval 12-080A2, Accuracy class 0,3 acc. NIST Handbook 44, Heated Refer to GS 01R04B07-00E
Products
GOST 1) /QR2 Primary Calibration and Test Confirmation valid in Kazakhstan see page 11; not with /Q20, /Q21; only with /VE
/QR3 Primary Calibration and Test Confirmation valid in Uzbekistan see page 11; not with /Q20, /Q21
Marine Approval /MC2 Marine approval acc. DNV piping class 2 see GS 01R04B10-00E-E
/MC3 Marine approval acc. DNV piping class 3 see GS 01R04B10-00E-E
High Driving Power /HP High Driving Power, recommended for combination with RCCS36 to 39, strongly mandatory for combination with RCCS39/XR 5)
recommended for combination with RCCS39/IR, Please see „ Hazardous Area
Specifications“
Communication /FB Digital communication (FOUNDATION™ Fieldbus protocol) refer to GS 01R04B05-00E
/MB2 Modbus communication protocol for all parameters and values not with /KF5, /EF5, /UF5, /GF5, /NF5, /FF5, /Q01,
/Q20, /Q21, /AP, /NM, /MB3
/MB3 Modbus communication protocol for all parameters and values incl. HART not with /KF5, /EF5, /FF5, /UF5, /NF5, /GF5, /Q01, /MB2
Communication, analogue output, pulse/status output, status input
Active Pulse Output /AP One active pulse output not with /KF5, /EF5, /FF5, /UF5, /NF5, /GF5, /NM
NAMUR Switch /NM One pulse output acc. to EN 60947-5-6 (NAMUR) not with /AP, not with /KF5, /EF5, /UF5, /NF5, /GF5
Tag Number /BG With customer specified tag number on name plate max. 16 characters; the last 8 significant characters are
used for HART® Tag;
characters acc. Codepage 850 (Multi-lingual)
HART® Tag Number (Software Tag) /BT1 With customer specified tag number for HART® communication in converter 22 characters for long tag;
characters acc. Codepage 850 (Multi-lingual)
Gas Measurement /GA Gas measurement, special factory adjustments and settings select affiliated RCCS3 with /GA; not with /Q20, /Q21
Combination with RCCS39/XR /XR Special converter adjustment with RCCS39/XR mandatory for combination with RCCS39/XR
No Combination /NC No combination with detector
Customer Presetting /PS Presetting sheet with customer data has to be issued with the order
Epoxy Coating /X1 Epoxy coating of converter housing
Delivery to Japan /PJ For meters which are delivered to Japan with SI-unit preset
Delivery to Korea /KC With KC-mark for Korea
Eurasian Conformity /VE With EAC- mark combination with Hazardous Area Approval only with /GF
Delivery to China /CN For meters which are delivered to China combination with Hazardous Area Approval only with /NF
Cable glands /AD2 2 pcs ANSI 1/2“ NPT / G1/2 adapter only with cable conduit “A”
Concentration Measurement 3) /CST Standard concentration measurement not with /GA, /C
/C Advanced concentration measurement, details see table „Advanced Concentration not with /GA, /CST
Measurement Options“
Certificates /P2 Certificate of compliance with the order acc. to EN 10204:2004 -2.1
/P3 Test Report acc. to EN10204:2004-2.2(QIC) including the content of option /P2
Instruction Manuals /IE Quantity of instruction manuals in English = 1 to 3 selectable 4)
/ID Quantity of instruction manuals in German = 1 to 3 selectable 4)
/IF Quantity of instruction manuals in French = 1 to 3 selectable 4)
Quick Delivery /QD Delivery within 24 hours from factory Delivery scope please refer to “Quick Delivery Possibili-
ties” or contact Yokogawa sales office.
Special order /Z Special design, must be specified separately.
If /Z is selected, several Suffix of Model-Suffix Code can be changed to Z.
1)
Select affiliated RCCS3 with the same approval type (e.g. /KF with /KS1).
2)
Select affiliated RCCS3 with the same approval type (/Q01, /Q20 or /Q21).
3)
For detailed information please see TI 01R04B04-04E-E. Option /K6 of RCCS3 is recommended with concentration measurement.
4)
If no instruction manual is selected, only a DVD with instruction manuals is shipped with the instrument. More than 3 manuals of one language on request available.
5)
RCCF31 in combination with RCCS39/XR with gas application /GA can also be ordered without /HP.
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
28
Hazardous Area Approvals 1) /KS1 ATEX associated apparatus for intrinsically safe detector connection for gas group IIC with /HP not for gas group IIC
/FS1 FM associated apparatus for intrinsically safe detector output with /HP not for gas groups A and B
/ES1 IECEx associated apparatus for intrinsically safe detector output for gas group IIC with /HP not for gas group IIC
/US1 INMETRO associated apparatus for intrinsically safe detector output for gas group IIC with /HP not for gas group IIC
/NS1 NEPSI associated apparatus for intrinsically safe detector output for gas group IIC with /HP not for gas group IIC; only with /CN
/GS1 EAC associated apparatus for intrinsically safe detector connection for gas group IIC with /HP not for gas group IIC; only with /VE
GOST 1) /QR2 Primary Calibration and Test Confirmation valid in Kazakhstan see page 11; only with /VE
Marine Approval /MC2 Marine approval acc. DNV piping class 2 see GS 01R04B10-00E-E
/MC3 Marine approval acc. DNV piping class 3 see GS 01R04B10-00E-E
High Driving Power /HP High Driving Power, recommended for combination with RCCS36 to 39, strongly
recommended for combination with RCCS39/IR, Please see „ Hazardous Area mandatory for combination with RCCS39/XR
Specifications“
Active Pulse Output /AP One active pulse output not with /NM
NAMUR Switch /NM One pulse output acc. EN 60947-5-6 (NAMUR) not with /AP
Tag Number /BG With customer specified tag number on name plate max. 16 characters; the last 8 significant characters are
used for HART® Tag;
characters acc. Codepage 850 (Multi-lingual)
HART® Tag Number (Software Tag) /BT1 With customer specified tag number for HART® communication in converter 22 characters for long tag;
characters acc. Codepage 850 (Multi-lingual)
Communication /MB2 Modbus communication protocol for all parameters and values not with /AP, /NM, /MB3
/MB3 Modbus communication protocol for all parameters and values incl. HART not with /MB2
Communication, analogue output, pulse/status output, status input
Gas Measurement /GA Gas measurement, special factory adjustments and settings select affiliated RCCS3 with /GA
Combination with RCCS39/XR /XR Special converter adjustment with RCCS39/XR mandatory for combination with RCCS39/XR
No Combination /NC No combination with detector
Customer Presetting /PS Presetting sheet with customer data has to be issued with the order
Concentration Measurement 2) /CST Standard concentration measurement not with /GA, /C
/C Advanced concentration measurement, details see table „Advanced Concentration not with /GA, /CST
Measurement Options“
Subrack /SR2 Subrack for 2 converter RCCR31 with mounting
/SR4 Subrack for 4 converter RCCR31 with mounting
Delivery to Korea /KC With KC-mark for Korea
Eurasian Conformity /VE With EAC- mark combination with Hazardous Area Approval only with /GS1
Delivery to China /CN For meters which are delivered to China combination with Hazardous Area Approval only with /NS1
Certificates /P2 Certificate of compliance with the order acc. to EN 10204:2004 -2.1
/P3 Test Report acc. to EN10204:2004-2.2(QIC) including the content of option /P2
Instruction Manuals /IE Quantity of instruction manuals in English = 1 to 3 selectable 3)
/ID Quantity of instruction manuals in German = 1 to 3 selectable 3)
/IF Quantity of instruction manuals in French = 1 to 3 selectable 3)
Special order /Z Special design, must be specified separately.
If /Z is selected, several Suffix of Model-Suffix Code can be changed to Z.
1)
Select affiliated RCCS3 with the same approval type (e.g. /KS1 with /KS1).
2)
For detailed information please see TI 01R04B04-04E-E. Option /K6 of RCCS3 is recommended with concentration measurement.
3)
If no instruction manual is selected, only a DVD with instruction manuals is shipped with the instrument. More than 3 manuals of one language on request available.
4)
RCCR31 in combination with RCCS39/XR with gas application /GA can also be ordered without /HP.
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
29
Advanced Concentration Measurement Options (others on request), recommended with Option /K6
Option Display Components Concentration Temp. range Source of concentration- / density table
range
/C00 --- --- --- --- Advanced concentration measurement function.
There are no pre-defined concentration setups (coefficients are set to zero).
The concentration coefficients will be set by customer making use of either
FieldMate or the stand- alone concentration tool and manual entry by HHT.
For more information, please contact your regional Yokogawa office.
/C01 °Brix Sugar / Water 0 - 85 °Brix 0 - 80°C PTB- Messages 100 5/90:
(32 - 176°F) „The density of watery Saccarose solutions after the introduction of the
international temperature scale of 1990 (ITS1990)“ Table 5
/C02 WT% NaOH / Water 2 - 50 WT% 0 - 100°C D´Ans-Lax, Handbook for chemists and physicists Vol.1, 3rd edition, 1967
(32 - 212°F)
/C03 WT% KOH / Water 0 - 60 WT% 54 - 100°C D´Ans-Lax, Handbook for chemists and physicists Vol.1, 3rd edition, 1967
(129 - 212°F)
/C04 WT% NH4NO3 / Water 1 - 50 WT% 0 - 80°C Data table on request
(32 - 176°F)
/C05 WT% NH4NO3 / Water 20 - 70 WT% 20 - 100°C Data table on request
(68 - 212°F)
/C06 1) WT% HCl / Water 22 - 34 WT% 20 - 40°C D´Ans-Lax, Handbook for chemists and physicists Vol.1, 3rd edition, 1967
(68 - 104°F)
/C07 WT% HNO3 / Water 50 - 67 WT% 10 - 60°C Data table on request
(50 - 140°F)
/C09 WT% H2O2 / Water 30 - 75 WT% 4 - 44°C Data table on request
(39.2 - 111.2°F)
/C10 WT% Ethylene Glycol / 10 - 50 WT% -20 - 40°C Data table on request
Water (-4 - 104°F)
/C11 WT% Amylum = starch / Water 33 - 43 WT% 35 - 45°C Data table on request
(95 - 113°F)
/C12 WT% Methanol / Water 35 - 60 WT% 0 - 40°C Data table on request
(32 - 104°F)
/C20 VOL% Alcohol / Water 55 - 100 VOL% 10 - 40°C Data table on request
(50 - 104°F)
/C21 °Brix Sugar / Water 40 - 80 °Brix 75 - 100°C Data table on request
(167 - 212°F)
/C30 WT% Alcohol / Water 66 - 100 WT% 15 - 40°C Standard Copersucar 1967
(59 - 104°F)
/C37 WT% Alcohol / Water 66 - 100 WT% 10 - 40°C Brazilian Standard ABNT
(50 - 104°F)
/C38 VOL% Alcohol / Water 73 - 100 VOL% 10 - 40°C Brazilian Standard ABNT
(50 - 104°F)
1)
only with material HC
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
30
05A1 2“-150 ---- ---- ---- 390 510 520 600 620 ---- ---- ---- ---- ---- ----
05A2 2“-300 ---- ---- ---- 390 510 520 600 620 ---- ---- ---- ---- ---- ----
05A3 2“-600 ---- ---- ---- 400 540 540 630 630 ---- ---- ---- ---- ---- ----
05A4 2“-900 ---- ---- ---- ---- 660 ---- 720 ---- ---- ---- ---- ---- ---- ----
05A5 2“-900/1500 ---- ---- ---- ---- 660 ---- ---- ---- ---- ---- ---- ---- ---- ----
06A1 2½“-150 ---- ---- ---- ---- ---- ---- 610 620 ---- ---- ---- ---- ---- ----
06A2 2½“-300 ---- ---- ---- ---- ---- ---- 610 620 ---- ---- ---- ---- ---- ----
06A3 2½“-600 ---- ---- ---- ---- ---- ---- 640 640 ---- ---- ---- ---- ---- ----
06A4 2½“-900 ---- ---- ---- ---- ---- ---- 760 ---- ---- ---- ---- ---- ---- ----
08A1 3“-150 ---- ---- ---- ---- ---- ---- 610 620 1000 1020 ---- ---- ---- ----
08A2 3“-300 ---- ---- ---- ---- ---- ---- 620 620 1000 1020 ---- ---- ---- ----
08A3 3“-600 ---- ---- ---- ---- ---- ---- 640 640 1000 1025 ---- ---- ---- ----
08A4 3“-900 ---- ---- ---- ---- ---- ---- 760 ---- ---- ---- ---- ---- ---- ----
10A1 4“-150 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 ---- ---- ----
10A2 4“-300 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 ---- ---- ----
10A3 4“-600 ---- ---- ---- ---- ---- ---- ---- ---- 1030 1030 1100 ---- ---- ----
12A1 5“-150 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
12A2 5“-300 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
12A3 5“-600 ---- ---- ---- ---- ---- ---- ---- ---- 1040 1040 1160 1110 ---- ----
15A1 6“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 1350 ----
15A2 6“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 1350 ----
15A3 6“-600 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1200 1120 1390 ----
20A1 8“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1140 1100 1350 ----
20A2 8“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1140 1100 1350 ----
20A3 8“-600 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1440 ----
01D4 DN 15 PN 40 240 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01D6 DN 15 PN 100 250 250 380 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02D4 DN 25 PN 40 ---- 240 370 390 500 520 ---- ---- ---- ---- ---- ---- ---- ----
02D6 DN 25 PN 100 ---- 260 390 ---- 520 ---- ---- ---- ---- ---- ---- ---- ---- ----
04D4 DN 40 PN 40 ---- 240 370 390 500 520 600 ---- ---- ---- ---- ---- ---- ----
04D6 DN 40 PN 100 ---- 320 450 ---- 560 ---- 620 ---- ---- ---- ---- ---- ---- ----
05D4 DN 50 PN 40 ---- ---- ---- ---- 500 520 600 620 ---- ---- ---- ---- ---- ----
05D5 DN 50 PN 63 ---- ---- ---- ---- 520 ---- 620 ---- ---- ---- ---- ---- ---- ----
05D6 DN 50 PN 100 ---- ---- ---- ---- 590 ---- 660 ---- ---- ---- ---- ---- ---- ----
05D7 DN 50 PN 160 ---- ---- ---- ---- 590 ---- 660 ---- ---- ---- ---- ---- ---- ----
Flanges according to EN 1092-1
08D4 DN 80 PN 40 ---- ---- ---- ---- ---- ---- 610 620 1000 1020 ---- ---- ---- ----
08D5 DN 80 PN 63 ---- ---- ---- ---- ---- ---- 620 ---- 1000 ---- ---- ---- ---- ----
08D6 DN 80 PN 100 ---- ---- ---- ---- ---- ---- 730 ---- 1000 ---- ---- ---- ---- ----
10D2 DN 100 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- 1100 ---- ---- ----
10D4 DN 100 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 ---- ---- ----
10D5 DN 100 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- 1100 ---- ---- ----
10D6 DN 100 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- 1050 ---- 1100 ---- ---- ----
12D2 DN 125 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
12D4 DN 125 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
12D5 DN 125 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- 1100 1100 ---- ----
12D6 DN 125 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- 1100 ---- 1140 1100 ---- ----
12D7 DN 125 PN 160 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 ---- ----
15D2 DN 150 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 1350 ----
15D4 DN 150 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 1350 ----
15D5 DN 150 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1140 ---- 1350 ----
15D6 DN 150 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1180 ---- ---- ----
20D2 DN 200 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1350 ----
20D4 DN 200 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1350 ----
20D5 DN 200 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1350 ----
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
31
Table 11 : Selection Table Process Connection and Materials, Installation Length in mm (continued)
02J2 DN 25 20K ---- 240 370 390 500 ---- ---- ---- ---- ---- ---- ---- ---- ----
04J1 DN 40 10K ---- 240 370 390 500 520 600 ---- ---- ---- ---- ---- ---- ----
04J2 DN 40 20K ---- 240 370 390 500 520 600 ---- ---- ---- ---- ---- ---- ----
05J1 DN 50 10K ---- ---- ---- ---- 500 520 600 620 ---- ---- ---- ---- ---- ----
05J2 DN 50 20K ---- ---- ---- ---- 500 520 600 620 ---- ---- ---- ---- ---- ----
08J1 DN 80 10K ---- ---- ---- ---- ---- ---- 600 620 1000 1020 ---- ---- ---- ----
08J2 DN 80 20K ---- ---- ---- ---- ---- ---- 610 620 1000 1020 ---- ---- ---- ----
10J1 DN 100 10K ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 ---- ---- ----
10J2 DN 100 20K ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 ---- ---- ----
12J1 DN 125 10K ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
12J2 DN 125 20K ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
15J1 DN 150 10K ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 ---- ----
15J2 DN 150 20K ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 ---- ----
01S4 DN 15 240 240 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02S4 DN 25 ---- 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Clamp DIN
04S4 DN 40 ---- 240 370 ---- 500 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S4 DN 50 ---- ---- ---- ---- 500 ---- 600 ---- ---- ---- ---- ---- ---- ----
06S4 DN 65 ---- ---- ---- ---- ---- ---- 600 ---- ---- ---- ---- ---- ---- ----
10S4 DN 100 ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- ---- ---- ---- ----
01S8 ½“ 240 240 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02S8 1“ ---- 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Tri-Clamp
04S8 1½“ ---- 240 370 ---- 500 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S8 2“ ---- ---- ---- ---- 500 ---- 600 ---- ---- ---- ---- ---- ---- ----
08S8 3“ ---- ---- ---- ---- ---- ---- 600 ---- ---- ---- ---- ---- ---- ----
10S8 4“ ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- ---- ---- ---- ----
02S2 DN 25 ---- 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
DIN11851
04S2 DN 40 ---- ---- ---- ---- 500 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S2 DN 50 ---- ---- ---- ---- ---- ---- 600 ---- ---- ---- ---- ---- ---- ----
10S2 DN 100 ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- ---- ---- ---- ----
41G9 G¼“ female 260 260 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01G9 G½“ female 260 260 390 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Thread
23G9 G¾“ female 260 260 390 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
41T9 NPT¼“ female 260 260 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01T9 NPT½“ female 260 260 390 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
23T9 NPT¾“ female 260 260 390 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
32
Table 11 : Selection Table Process Connection and Materials, Installation Length in inch
RCCS RCCS RCCS34 RCCS36 RCCS38 RCCS39 RCCS39/IR RCCS39/XR
30LR 30-33 RCCT34 RCCT36 RCCT38 RCCT39 RCCT39/IR RCCT39/XR
SH SH SL HC SL HC SL HC SL HC SL HC SL HC
01A1 ½“-150 9.45 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A2 ½“-300 9.45 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A3 ½“-600 9.84 9.84 14.96 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A4 ½“-900 ---- ---- 15.75 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A5 ½“-900/1500 10.63 10.63 15.75 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02A1 1“-150 ---- 9.45 14.57 15.35 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A2 1“-300 ---- 9.45 14.57 15.35 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A3 1“-600 ---- 10.24 15.35 15.35 20.47 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A4 1“-900 ---- ---- 17.72 15.75 21.26 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A5 1“-900/1500 ---- 12.6 17.72 15.75 21.26 ---- ---- ---- ---- ---- ---- ---- ---- ----
04A1 1½“-150 ---- 9.84 14.96 15.35 19.69 20.47 23.62 ---- ---- ---- ---- ---- ---- ----
04A2 1½“-300 ---- 9.84 14.96 15.35 20.08 20.47 23.62 ---- ---- ---- ---- ---- ---- ----
04A3 1½“-600 ---- 10.63 15.75 15.75 20.87 20.87 24.21 ---- ---- ---- ---- ---- ---- ----
04A4 1½“-900 ---- ---- 18.5 ---- 23.62 ---- 25.2 ---- ---- ---- ---- ---- ---- ----
04A5 1½“-900/1500 ---- 13.39 18.5 ---- 23.62 ---- ---- ---- ---- ---- ---- ---- ---- ----
Flanges according to ASME B16.5
05A1 2“-150 ---- ---- ---- 15.35 20.08 20.47 23.62 24.21 ---- ---- ---- ---- ---- ----
05A2 2“-300 ---- ---- ---- 15.35 20.08 20.47 23.62 24.21 ---- ---- ---- ---- ---- ----
05A3 2“-600 ---- ---- ---- 15.75 21.26 21.26 24.8 24.8 ---- ---- ---- ---- ---- ----
05A4 2“-900 ---- ---- ---- ---- 25.98 ---- 28.35 ---- ---- ---- ---- ---- ---- ----
05A5 2“-900/1500 ---- ---- ---- ---- 25.98 ---- ---- ---- ---- ---- ---- ---- ---- ----
06A1 2½“-150 ---- ---- ---- ---- ---- ---- 24.02 24.21 ---- ---- ---- ---- ---- ----
06A2 2½“-300 ---- ---- ---- ---- ---- ---- 24.02 24.21 ---- ---- ---- ---- ---- ----
06A3 2½“-600 ---- ---- ---- ---- ---- ---- 25.2 25.2 ---- ---- ---- ---- ---- ----
06A4 2½“-900 ---- ---- ---- ---- ---- ---- 29.92 ---- ---- ---- ---- ---- ---- ----
08A1 3“-150 ---- ---- ---- ---- ---- ---- 24.02 24.21 39.37 40.16 ---- ---- ---- ----
08A2 3“-300 ---- ---- ---- ---- ---- ---- 24.21 24.21 39.37 40.16 ---- ---- ---- ----
08A3 3“-600 ---- ---- ---- ---- ---- ---- 25.2 25.2 39.37 40.35 ---- ---- ---- ----
08A4 3“-900 ---- ---- ---- ---- ---- ---- 29.92 ---- ---- ---- ---- ---- ---- ----
10A1 4“-150 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 ---- ---- ----
10A2 4“-300 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 ---- ---- ----
10A3 4“-600 ---- ---- ---- ---- ---- ---- ---- ---- 40.55 40.55 43.31 ---- ---- ----
12A1 5“-150 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
12A2 5“-300 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
12A3 5“-600 ---- ---- ---- ---- ---- ---- ---- ---- 40.94 40.94 45.67 43.31 ---- ----
15A1 6“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 53.15 ----
15A2 6“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 53.15 ----
15A3 6“-600 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 47.2 43.31 54.72 ----
20A1 8“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 44.9 43.31 53.15 ----
20A2 8“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 44.9 43.31 53.15 ----
20A3 8“-600 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 56.69 ----
01D4 DN 15 PN 40 9.45 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01D6 DN 15 PN 100 9.84 9.84 14.96 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02D4 DN 25 PN 40 ---- 9.45 14.57 15.35 19.69 20.47 ---- ---- ---- ---- ---- ---- ---- ----
02D6 DN 25 PN 100 ---- 10.24 15.35 ---- 20.47 ---- ---- ---- ---- ---- ---- ---- ---- ----
04D4 DN 40 PN 40 ---- 9.45 14.57 15.35 19.69 20.47 23.62 ---- ---- ---- ---- ---- ---- ----
04D6 DN 40 PN 100 ---- 12.6 17.72 ---- 22.05 ---- 24.41 ---- ---- ---- ---- ---- ---- ----
05D4 DN 50 PN 40 ---- ---- ---- ---- 19.69 20.47 23.62 24.41 ---- ---- ---- ---- ---- ----
05D5 DN 50 PN 63 ---- ---- ---- ---- 20.47 ---- 24.41 ---- ---- ---- ---- ---- ---- ----
05D6 DN 50 PN 100 ---- ---- ---- ---- 23.23 ---- 25.98 ---- ---- ---- ---- ---- ---- ----
05D7 DN 50 PN 160 ---- ---- ---- ---- 23.23 ---- 25.98 ---- ---- ---- ---- ---- ---- ----
Flanges according to EN 1092-1
08D4 DN 80 PN 40 ---- ---- ---- ---- ---- ---- 24.02 24.41 39.37 40.16 ---- ---- ---- ----
08D5 DN 80 PN 63 ---- ---- ---- ---- ---- ---- 24.41 ---- 39.37 ---- ---- ---- ---- ----
08D6 DN 80 PN 100 ---- ---- ---- ---- ---- ---- 28.74 ---- 39.37 ---- ---- ---- ---- ----
10D2 DN 100 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- 43.31 ---- ---- ----
10D4 DN 100 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 ---- ---- ----
10D5 DN 100 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- 43.31 ---- ---- ----
10D6 DN 100 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- 41.34 ---- 43.31 ---- ---- ----
12D2 DN 125 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
12D4 DN 125 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
12D5 DN 125 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- 43.31 43.31 ---- ----
12D6 DN 125 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- 43.31 ---- 44.88 43.31 ---- ----
12D7 DN 125 PN 160 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 ---- ----
15D2 DN 150 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 53.15 ----
15D4 DN 150 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 53.15 ----
15D5 DN 150 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 44.88 ---- 53.15 ----
15D6 DN 150 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 46.46 ---- ---- ----
20D2 DN 200 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 53.15 ----
20D4 DN 200 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 53.15 ----
20D5 DN 200 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 53.15 ----
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
33
Table 11 : Selection Table Process Connection and Materials, Installation Length in inch (continued)
02J2 DN 25 20K ---- 9.45 14.57 15.35 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
04J1 DN 40 10K ---- 9.45 14.57 15.35 19.69 20.47 23.62 ---- ---- ---- ---- ---- ---- ----
04J2 DN 40 20K ---- 9.45 14.57 15.35 19.69 20.47 23.62 ---- ---- ---- ---- ---- ---- ----
05J1 DN 50 10K ---- ---- ---- ---- 19.69 20.47 23.62 24.41 ---- ---- ---- ---- ---- ----
05J2 DN 50 20K ---- ---- ---- ---- 19.69 20.47 23.62 24.41 ---- ---- ---- ---- ---- ----
08J1 DN 80 10K ---- ---- ---- ---- ---- ---- 23.62 24.41 39.37 40.16 ---- ---- ---- ----
08J2 DN 80 20K ---- ---- ---- ---- ---- ---- 24.02 24.41 39.37 40.16 ---- ---- ---- ----
10J1 DN 100 10K ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 ---- ---- ----
10J2 DN 100 20K ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 ---- ---- ----
12J1 DN 125 10K ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
12J2 DN 125 20K ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
15J1 DN 150 10K ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 ---- ----
15J2 DN 150 20K ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 ---- ----
01S4 DN 15 9.45 9.45 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02S4 DN 25 ---- 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Clamp DIN
04S4 DN 40 ---- 9.45 14.57 ---- 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S4 DN 50 ---- ---- ---- ---- 19.69 ---- 23.62 ---- ---- ---- ---- ---- ---- ----
06S4 DN 65 ---- ---- ---- ---- ---- ---- 23.62 ---- ---- ---- ---- ---- ---- ----
10S4 DN 100 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- ---- ---- ---- ----
01S8 ½“ 9.45 9.45 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02S8 1“ ---- 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Tri-Clamp
04S8 1½“ ---- 9.45 14.57 ---- 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S8 2“ ---- ---- ---- ---- 19.69 ---- 23.62 ---- ---- ---- ---- ---- ---- ----
08S8 3“ ---- ---- ---- ---- ---- ---- 23.62 ---- ---- ---- ---- ---- ---- ----
10S8 4“ ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- ---- ---- ---- ----
02S2 DN 25 ---- 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
DIN11851
04S2 DN 40 ---- ---- ---- ---- 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S2 DN 50 ---- ---- ---- ---- ---- ---- 23.62 ---- ---- ---- ---- ---- ---- ----
10S2 DN 100 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- ---- ---- ---- ----
41G9 G¼“ female 10.24 10.24 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01G9 G½“ female 10.24 10.24 15.35 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Thread
23G9 G¾“ female 10.24 10.24 15.35 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
41T9 NPT¼“ female 10.24 10.24 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01T9 NPT½“ female 10.24 10.24 15.35 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
23T9 NPT¾“ female 10.24 10.24 15.35 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
34
05A1 2“-150 ---- ---- ---- X X X X ---- ---- ---- ---- ----
05A2 2“-300 ---- ---- ---- X X X X ---- ---- ---- ---- ----
05A3 2“-600 ---- ---- ---- X ---- X ---- ---- ---- ---- ---- ----
05A4 2“-900 ---- ---- ---- ---- ---- X ---- ---- ---- ---- ---- ----
05A5 2“-900/1500 ---- ---- ---- X ---- ---- ---- ---- ---- ---- ---- ----
06A1 2½“-150 ---- ---- ---- ---- ---- X X ---- ---- ---- ---- ----
06A2 2½“-300 ---- ---- ---- ---- ---- X X ---- ---- ---- ---- ----
06A3 2½“-600 ---- ---- ---- ---- ---- X ---- ---- ---- ---- ---- ----
06A4 2½“-900 ---- ---- ---- ---- ---- X ---- ---- ---- ---- ---- ----
08A1 3“-150 ---- ---- ---- ---- ---- X X X X ---- ---- ----
08A2 3“-300 ---- ---- ---- ---- ---- X X X X ---- ---- ----
08A3 3“-600 ---- ---- ---- ---- ---- X ---- X ---- ---- ---- ----
08A4 3“-900 ---- ---- ---- ---- ---- X ---- ---- ---- ---- ---- ----
10A1 4“-150 ---- ---- ---- ---- ---- ---- ---- X X X X ----
10A2 4“-300 ---- ---- ---- ---- ---- ---- ---- X X X X ----
10A3 4“-600 ---- ---- ---- ---- ---- ---- ---- X ---- X ---- ----
12A1 5“-150 ---- ---- ---- ---- ---- ---- ---- X X X X ----
12A2 5“-300 ---- ---- ---- ---- ---- ---- ---- X X X X ----
12A3 5“-600 ---- ---- ---- ---- ---- ---- ---- X ---- X ---- ----
15A1 6“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- X X X
15A2 6“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- X X X
15A3 6“-600 ---- ---- ---- ---- ---- ---- ---- ---- ---- X ---- ----
20A1 8“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- X
20A2 8“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- X
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
35
All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
36
REGISTERED TRADEMARKS
HART® : Registered trademark of HART Communication Foundation, Austin, TX, USA
MODBUS® : Registered trademark of MODBUS Organization
FOUNDATION™ Fieldbus : Registered trademark of the Fieldbus FOUNDATION, Austin, TX, USA
TRI-CLAMP® : Registered trademark of Ladish & Co., Inc., Kenosha, USA
FieldMate™ : Registered trademark of Yokogawa Electric Corporation
YOKOGAWA EUROPE B. V.
Euroweg 2, 3825 HD Amersfoort, THE NETHERLANDS
Phone : 31-88-4641000 Fax : 31-88-4641111
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa