RotaMASS 3 Series Coriolis Mass Flow & Density Meter PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 36

General

3 Series
Specifications Coriolis Mass Flow and Density Meter

GS 01R4B04-00E-E
01R04B04-00E-E

RCCT39/XR RCCT34 - 39/IR RCCF31 + RCCS30LR - 33 RCCR31


RCCS34 - 39/IR

FEATURES
Contents • ROTAMASS is a flowmeter for nearly all fluids, including
Features Page 1 high viscosity liquids, slurries and multi- phase media
Principle of measurement Page 1
with a certain gas content
Performance specifications Page 2
• Field-mount and rack-mount remote converter available
Normal operating conditions Page 4
Mechanical specifications Page 5
• Refined digital signal processing enables accurate and
Remote cable specification Page 5
stable measurement
Electrical specifications Page 6
• A special detector decoupling system makes the device
Hazardous area specifications Page 7
highly independent from external loads or vibrations.
Planning and installation hints Page 12 • Simple flow path means self-draining, simple cleaning
Dimensions Page 15 and suitable for use in hygenic industries.
Model-, suffix- and option- codes Page 23 • High accuracy and high stability over a wide flow range
• Accurate density measurement, up to +/- 0.0005 g/cm³
• Concentration measurement for solutions, suspensions
ROTAMASS is a Coriolis mass flowmeter with highly
and emulsions (e.g. water cut, net oil computing)
refined digital signal processing electronics, so that accurate
• Volume flow with reference density
and stable mass flow measurement is achieved.
• Temperature measurement
• Two analog outputs, two pulse outputs or status-out and
ROTAMASS employs a flame-proof type converter case
one status-in as standard I/O
suitable for use in the hazardous area together with it’s
• Available in explosion proof versions (ATEX, FM, IECEx,
intrinsically safety type detector.
EAC, INMETRO, NEPSI, KOSHA)
• European MID approval acc. OIML R-117-1 and American
ROTAMASS´s signal processing, housing protection and its
NTEP approval for Custody Transfer Measurement
detector´s special decoupling system against external loads and
(see GS 01R04B07-00E-E)
vibrations, realize high performance in real applications.
• Wide process temperature range –200°C to 350°C
• Microprocessor-based multifunction capability
• EEPROM protects parameter settings and totalized
PRINCIPLE OF MEASUREMENT values during power failure.
Mass flow measurement according to the Coriolis principle. • High visibly LCD display
For abrasive or highly corrosive fluids please contact your • HART® communication function is available
Yokogawa representative. • Optional Foundation™ Fieldbus communication is available
(see GS 01R04B05-00E-E)
• Optional MODBUS® communication is available
• Optional intrinsically safe outputs are available
• Various choice of tubes materials
• EN, ASME or JIS flanges as standard, others on request
• FMEDA report on request, SIL 2, SIL 3 levels can be
achieved if redundant meters are used.

Rota Yokogawa GmbH & Co. KG GS 01R04B04-00E-E


Rheinstr. 8 ©Copyright July 2003 (RYG)
D-79664 Wehr 24th edition, October 2016 (RYG)
Germany
2

PERFORMANCE SPECIFICATIONS Pressure Dependency


The stiffness of the ROTAMASS tubes is slightly line pressure
Model dependent. The static pressure effect of mass flow and density
- Remote detector RCCS30LR to 33: 2 tubes, low flow design can be corrected by setting the static pressure manually via menu.
- Remote detector RCCS34 to 39/XR : 2 tube design
- Remote field-mount converter RCCF31
- Remote rack-mount converter RCCR31 Table 3 : Static pressure effect on mass flow (if not corrected)
- Integral type RCCT34 to 39/XR: 2 tube design Model
RCCS30
RCCS30 RCCS31 RCCS32 RCCS33
LR
Fluid to be measured : Liquid, gas or slurry % of 0.00000 / 0.00000 / 0.00012 / 0.00246 / 0.0035 /
SH
rate (0.0) (0.0) (0.00174) (0.03568) (0.0508)
Measurement Items : Mass flow, density, temperature per bar
and derived from these values: / (psi) HC ---- ---- ---- ---- ----
concentration, volume flow and RCC39
net flow Model RCC34 RCC36 RCC38 RCC39
/IR

% of 0.00081 / 0.00346 / 0.00950 / 0.01058 / 0.0047 /


Mass Flow Measurement SL
(0.01175) (0.05018) (0.1378) (0.15345) (0.0682)
rate
Table 1: measuring range per bar 0.00084 / 0.00336 / 0.00896 / 0.00808 / 0.00287 /
RCCS30 / (psi) HC
Model RCCS30 RCCS31 RCCS32 RCCS33 (0.01218) (0.04873) (0.12995) (0.11719) (0.04163)
LR
t/h 0.04 0.094 0.3 0.6 1.5 RCC39
Qmax Model
lb/h 88 207 661 1322 3307 /XR
t/h 0.021 0.045 0.17 0.37 0.95 0.00740 /
Qnom % of SL
lb/h 46 99 374 815 2094 (0.1073)
rate
RCC39 RCC39 per bar
Model RCC34 RCC36 RCC38 RCC39 HC ----
/IR /XR / (psi)
t/h 5 17 50 170 300 600
Qmax
lb/h 11023 37478 110231 374785 661386 1322773
Density Measurement
t/h 3 10 32 100 250 500 Adjustment with water and air at calibration temperature.
Qnom
lb/h 6613 22046 70547 220462 551155 1102311 Measuring range:
RCCS30LR - 38 : 0.3 kg/l to 5 kg/l
Qnom is the water flow rate at about 1 bar pressure drop.
The flowmeter has a default low cut of 0.05% of Qnom. RCC39 -39/XR : 0.3 kg/l to 2 kg/l
No density measurement for gas applications.
Accuracy of mass flow (refer to table 2): With option /K4 thermal stabilization is acquired.
Liquid RCCS30LR: For further details about the option /K6 please refer to “special
± 0.15% of flow rate ± zero stability / flow rate *100% calibrations” on page 3.
Liquid RCCS30 - 39/XR: Calibration condition standard
± 0.1% of flow rate ± zero stability / flow rate *100% Density : 0.9 kg/l ≤ r ≤ 1.1 kg/l
Gas (option /GA): Temp. Fluid : 22.5°C ± 12.5°C
± 0.5% of flow rate ± zero stability / flow rate *100% Flow Rate : about 0.2 * Qnom as defined for each
model
Accuracy of volume flow :
SQRT ( (mass flow error in %)² + (density error in %)²) Calibration condition for /K6:
Density : 0.7 kg/l ≤ r ≤ 1.65 kg/l
Temp. Fluid : 20°C ≤ T ≤ 80°C
Accuracy based on the frequency output includes the combined Temp. Ambient : 20°C ± 3K
effects of repeatability, linearity and hysteresis. Flow Rate : about 0.2 * Qnom as defined for each
model
Repeatability of liquids:
± 0.05% ± (zero stability/2) / flow rate *100% Table 4: Accuracy (at calibration conditions):
1.0
0.9 Type Standard Option /K4 Option /K6
0.8
0.7
0.6 RCCS30LR 0.02 g/cm³ *) ------ ------
0.5
0.4
0.3 RCCS30 0.008 g/cm³ *) ------ ------
Error in %

0.2
0.1
0
RCCS31 0.004 g/cm³ 0.001 g/cm³ ------
-0.1
-0.2
-0.3
RCCS32 0.004 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
-0.4
-0.5
-0.6
RCCS33 0.004 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
-0.7
-0.8 RCC34 0.003 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
-0.9
-1.0
0 50 100 150 RCC36 0.0022 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
Flow in % of Qnom F10.EPS
RCC38 0.0015 g/cm³ 0.001 g/cm³ 0.0005 g/cm³

Table 2 : Zero stability RCC39 0.0015 g/cm³ 0.001 g/cm³ 0.0005 g/cm³

RCCS30 RCC39//IR 0.0015 g/cm³ ------ ------


Model RCCS30 RCCS31 RCCS32 RCCS33
LR
RCC39//XR 0.0015 g/cm³ ------ ------
kg/h 0.003 0.005 0.0085 0.019 0.048
*) 0.003 g/cm³ on request
lb/h 0.006 0.011 0.018 0.04 0.105
Repeatability:
RCCS32-33, RCC34-39/XR : ± 0.0005 g/cm³ (Std, /K4)
RCC39 RCC39 Static pressure effect:
Model RCC34 RCC36 RCC38 RCC39
/IR /XR Compensated if static pressure is set in the menu.
kg/h 0.15 0.5 1.6 5 13 25
lb/h 0.33 1.1 3.5 11 28.6 55

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
3

Specification of high performance density measurement Special Calibrations


option /K6: Density calibration - Mass-/Volume flow calibration with factory certificate (option /K2):
Density range : 0.3 to 2.5 kg/l Calibration with water at customer specified flow values
Ambient temp. range : -10°C to 50°C (14°F to 122°F) according calibration order sheet.
Process temp. range Standard: -50°C to 150°C (58°F to 302°F) - Mass-/Volume flow calibration with DAkkS certificate
Process temp. range /HT: 150°C to 350°C (302°F to 662°F) EN17025 (option /K5):
Minimum flow rate for specified accuracy: Calibration with water at customer specified flow values
- RCC36 to RCC39 : 700 kg/h (1543 lb/h) according calibration order sheet.
- RCC34 : 140 kg/h (308 lb/h) - Density calibration with factory certificate (option /K6):
- RCCS33 : 90 kg/h (198 lb/h) Adjustment and check with 3 different fluids, fluid temperature
- RCCS32 : 37 kg/h (81 lb/h) influence adjustment for low ambient temperature influence
Maximum flow rate : Qnom
and thermal treatment for long term density measurement
Repeatability : ±0.0002 g/cm³
stability, improved temperature measurement accuracy (see
Temperature measurement:
also page 12).
±(0.5°C+0.002*abs(Tmedium-20°C)) (not /HT)
±(0.5°C+0.008*abs(Tmedium-20°C)) (/HT)
Density accuracy : only for liquids, one phase Dual Seal Approval (option /DS):
Process temperature influence : - Conform with ANSI/ISA-12.27.01.
±0.000015 g/cm³ * abs(Tmedium-20°C) - To be ordered if compliance with ANSI/ISA 12.27.01 is required.
- Up to ANSI class 900 line pressure.
Temperature Measurement - Only with FM approval option.
Temperature measuring range of converter : - For liquid application the leakage detection is realized by
Standard, /LT, /MT : -200°C to 230°C (-328°F to 448°F) software in the converter.
Option /HT : 0°C to 350°C (32°F to 662°F) - For gas application options /GA and /RD (rupture disk) are
Accuracy: mandatory.
Standard (-70°C to 150°C / -94°F to 302°F) - Rupture disk is only for annunciation.
: ±(0.5°C+0.005*abs(Tmedium-20°C))
Option /LT (-200°C to 150°C / -328°F to 302°F)
: ±(1.0°C+0.008*abs(Tmedium-20°C))
Option /MT (-70°C to 260°C / -94°F to 500°F)
: ±(0.5°C+0.005*abs(Tmedium-20°C))
Option /HT (0°C to 350°C / 32°F to 662°F)
: ±(1.0°C+0.008*abs(Tmedium-20°C))
For process temperatures more than 80°C (176°F) higher
than ambient temperature the detector should be insulated to
maintain optimum accuracy.

Heat Tracing
Heating with heat carrier, insulation and protection housing.
Typically the max. surface temperature at the protection
housing from inner heating is 40°C (at Tamb = 20°C). Above
150°C (302°F) process temperature insulation from the
manufacturer is recommended. However up to 230°C (446°F)
process temperature the customer can insulate the detector
himself. For this case order option /S2.
Option /T1 : only insulation and protection
Option /T2 : insulation, protection and heating line
Option /T3 : like /T2 but with ventilation
Process connection for the heat carrier fluid (see table 10):
for D-type flanges : EN DN 15 PN 40 Form B1
for A-type flanges : ANSI ½ - 150 lbs.
for J-type flanges : JIS DN15 10K
Max. pressure : PN 40
Protection class : IP54, install roof protected
For fluid temperatures below -70°C select option /LT.

Calibration for Liquids and Gases :


The ROTAMASS flowmeters are always factory calibrated with
water. Calibration Conditions:
- Water : 22.5°C ± 12.5°C (72.5°F ± 22.5°F)
- Ambient temperature : 22.5°C ± 12.5°C (72.5°F ± 22.5°F)
- Process pressure : 1 to 2 bar abs
- Installation: RCCS30LR to RCC38 vertical
RCC39 to RCCS39/XR horizontal
All specifications are based on above mentioned calibration
reference conditions, a flow calibration protocol is attached to
each instrument.

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
4

NORMAL OPERATING CONDITIONS Process Pressure Limit


In dependence of the process connections s. table 9.
Ambient Temperature Ranges
On request following maximum pressure up to 27°C (RT=Room Temp.):
- Remote detector RCCS3:
Standard : -50°C to +80°C (-58°F to 176°F) Material wetted parts SH [bar] / SL [bar] / HC [bar] /
(psi) (psi) (psi)
Option /LT : -50°C to +80°C (-58°F to 176°F)
RCCS30LR 400 / (5801) ---- ----
Option /MT : -50°C to +80°C (-58°F to 176°F)
Option /HT : -50°C to +65°C (-58°F to 149°F) RCCS30 400 / (5801) ---- ----
(up to 280°C (536°F) medium temperature) RCCS31 350 / (5076) ---- ----
-50°C to +55°C (-58°F to 131°F) RCCS32 285 / (4183) ---- ----
(up to 350°C (662°F) medium temperature) RCCS33 285 / (4183) ---- ----
Terminal box temperature should not exceed 100°C RCCS34 / RCCT34 ---- 260 / (3770) 385 / (5583)
- Remote converter RCCF31, RCCR31 and integral type RCCS36 / RCCT36 ---- 210 / (3045) 315 / (4568)
RCCT3: RCCS38 / RCCT38 ---- 175 / (2538) 260 / (3770)
Display operating range RCCS39 / RCCT39 ---- 135 / (1958) 260 / (3770)
: -20°C to +55°C (-4°F to 131°F) RCCS39/IR / RCCT39/IR ---- 110 / (1595) 180 / (2610)
Electronic operating range RCCS39/XR / RCCT39/XR ---- 95 / (1377) ----
: -40°C to +55°C (-40°F to 131°F)
Cold start : above -30°C (-22°F) For higher medium temperatures maximum tube pressure
Where meters are mounted in direct sunlight, it is recom- needs to be derated as follows :
mended to install a sunshade. This is particularly important in up to 50°C (122°F) : 4% derating
countries with high ambient temperatures. 51 to 100°C (123.8°F to 212°F) : 11% derating
101 to 150°C (213.8°F to 302°F) : 20% derating
Ambient Humidity Range : 0 to 95% RH 151 to 230°C (303.8°F to 446°F) : 30% derating
231 to 350°C (447.8°F to 662°F) : 38% derating
Process Temperature Ranges Higher pressure on request.
The maximum process pressure of a single instrument is given
Detector : by the lower value either of the process connections (table 9)
- RCCS30LR to 33 : -50°C to 150°C (-58°F to 302°F) or tubes. The maximum temperature and process pressure
- RCCS30LR to 33 /MT : -50°C to 260°C (-58°F to 500°F) limits of an instrument are marked on the name plate as TS
- RCCS34 to 39/XR : -70°C to 150°C (-94°F to 302°F) and PS.
- RCCS34 to 39/XR /LT : -200°C to 150°C (-328°F to 302°F) The given temperature/pressure ranges are calculated and
- RCCS34 to 39/XR /MT : -70°C to 230°C (-94°F to 446°F) approved without corrosion or erosion effects. The customer is
fully responsible of selecting proper materials which withstand
(Range 150°C - 230°C (302°F to 446°F) corrosive or erosive conditions. In case of heavy corrosion
recommended with /T option) and/or erosion the instrument may not withstand the pressure
- RCCS34 to 39/IR /HT : 0°C to 350°C (32°F to 662°F) and an incident may happen with human and/or environmental
(only with /T option or with /S2 and harm. Yokogawa will not take any liability regarding damage
customer insulation) caused by corrosion / erosion. If corrosion / erosion may hap-
- RCCS39/XR /HT : 0°C to 350°C (32°F to 662°F) pen, the user has to check periodically if the necessary wall
(only with /S2 and customer insulation) thickness is still in place.
Integral type :
Gas Content Limits for Liquid / Gas Mixtures
- RCCT34 to 39/XR : -50°C to 150°C (-58°F to 302°F) Gas content limit is defined as the amount of gas in a liquid /
gas mixture which generates an error in the converter. The
For use in hazardous area see “Hazardous Area Specifications” gas content limit is dependent on viscosity, surface tension
and bubble size of the liquid / gas mixture.
Heat Carrier Fluid Temperature Ranges Furthermore it is highly flow rate dependent (the higher the
(Option /T2 or /T3 only for remote type RCCS30LR to 39/IR) flow rate, the lower the gas content limits). The stated values
- Standard : 0°C to 150°C (32°F to 302°F) are for a flow of 50% of Qnom and water / air without /HP:
- With option /MT (RCCS30LR to 33) : Model Gas content limit
0°C to 200°C (32°F to 392°F)
RCCS30LR to RCCS32 no limitation
- With option /MT (RCCS34 to 39/IR) :
0°C to 230°C (32°F to 446°F) RCCS33 non-Ex type no limitation
- With option /HT : 0°C to 350°C (32°F to 662°F) RCCS33 Ex type approx. 35%
RCC34 no limitation
RCC36 approx. 50%
RCC38 approx. 30%
RCC39 approx. 7%
RCC39/IR approx. 3%
RCC39/XR (with /HP) approx. 2%
With option /HP the gas content limits are improved.
With liquid/gas mixtures the specified mass flow accuracy will
not be achieved.
For short time aeration a function can be activated to keep
the current outputs constant during the aeration time.

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
5

Other 2 Phase Flow, liquid/solid and liquid/liquid Wetted Parts


Two phase flow can generate minus span errors. The errors - RCCS30LR to 33 :
are proportional to the difference in density between the 2 Measuring tubes : Ni-Alloy C-22/2.4602
phases and the amount of the second phase. If the particles Process connections : 316L/1.4404
or droplets are very small no errors will be generated. - RCC34 to 39/IR :
Measuring tubes and process connection
Secondary Containment : 316L/1.4404/1.4435 or
Model Typical rupture Option /J1
Measuring tubes and flange face
: Ni-Alloy C-22/2.4602
pressure pressure test *)
- RCC39/XR :
RCCS30LR-33 65 bar / (942 psi) --- Measuring tubes and process connection
RCC34-36 120 bar / (1740 psi) 60 bar / (870 psi) : 316L/1.4404/1.4435
RCC38 120 bar / (1740 psi) 40 bar / (580 psi) Table 5 : Diameter of measuring tubes
RCC39 80 bar / (1160 psi) 10 bar / (145 psi) RCCS30
Type RCCS30 RCCS31 RCCS32 RCCS33
LR
RCC39/IR 50 bar / (725 psi) ---
Inner mm 0.9 1.2 2.1 3 4.5
RCC39/XR on request --- diameter inches 0.035 0.047 0.083 0.118 0.177
RCC39/XR/HT 50 bar / (725 psi) --- Wall mm 0.15 0.2 0.25 0.25 0.4
thickness inches 0.006 0.008 0.009 0.009 0.016
*) Pressure test with safety factor S=1.1
If the detector housing is exposed to a pressure close to
the rupture pressure it will deform and measurement will be RCC39 RCC39
Type RCC34 RCC36 RCC38 RCC39
strongly influenced. Therefore the pressure test of the housing /IR /XR
(option /J1) can only be done as shown in above table. Inner mm 7.7 13.4 22.1 37.2 54.5 82.50
diameter inches 0.303 0.528 0.870 1.485 2.146 3.248
Wall mm 0.89 1.24 1.65 2.6 2.9 3.2
MECHANICAL SPECIFICATIONS thickness inches 0.035 0.049 0.065 0.102 0.114 0.126
Protection Class
- RCCT3 : IP66/67 Chemical composition of wetted materials
- RCCF31 : IP66/67 316L/1.4404/1.4435 and Ni-Alloy C-22/2.4602 conform with
- RCCS3 : IP66/67 ANSI / NACE-MR0175 / ISO15156-2
- RCCR31 : IP20 ANSI / NACE-MR0175 / ISO15156-3
NACE MR0103
Materials For details please see RYG´s declaration about NACE
- Detector housing : Stainless steel 304/1.4301 conformity 8660001
- Detector terminal box : 316L/1.4404
- Detector gas filling plug: 1.4305 Pressure Equipment Directive 2014/68/EU
- Detector insulation housing Detectors comply with Directive 2014/68/EU on Pressure
: Stainless steel 304/1.4301 Equipment for fluid group 1 and 2.
- Detector rupture disk (/RD) Note: PED-compliance is not applicable for Rotamass
: 316L assembled in the USA that are identifiable by nameplate
- Field- mount converter housing containing “Assembled in USA with Foreign and Domestic Parts”.
: Aluminium alloy with Polyurethane
corrosion-resistant coating or CRN : CRN 0F12074.5
epoxy coating (option /X1) Approved process connections
- Field- mount converter mounting bracket: see table 12
: Stainless steel 304/1.4301
- Rack- mount converter housing Vibration Test : Acc. IEC 60068-2-64
: Aluminium
- Name plates : Stainless steel 304/1.4301

Coating Color
- Field-mount converter case
: Mint green

REMOTE CABLE RCCY03 SPECIFICATION


3 x Coaxial + 1 x 3 AWG20, shielded, twisted; overall shielding; flame propagation acc. IEC 60332-1.
We recommend to use only the original cable provided by Yokogawa. In case of local cable purchasing, please contact Yokogawa
for specification.
Table 6 : Cable specifications
Model code Temperature Wire gauge Resistance of Capacitance Capacitance Inductance
range loop wire/wire wire/shield wire/wire
-50 to +105°C Coaxial 37 Ω/km 120 nF/km 132 nF/km 0.175 mH/km
RCCY031 AWG 20 70 Ω/km 145 nF/km 290 nF/km 0.70 mH/km
-58°F to 221°F
-50 to +105°C Coaxial 37 Ω/km 120 nF/km 132 nF/km 0.175 mH/km
RCCY032 AWG 20 70 Ω/km 145 nF/km 290 nF/km 0.70 mH/km
-58°F to 221°F
-50 to +105°C Coaxial 37 Ω/km 120 nF/km 132 nF/km 0.175 mH/km
RCCY033
-58°F to 221°F AWG 20 70 Ω/km 145 nF/km 290 nF/km 0.70 mH/km
-50 to +105°C Coaxial 37 Ω/km 120 nF/km 132 nF/km 0.175 mH/km
RCCY034
-58°F to 221°F AWG 20 70 Ω/km 145 nF/km 290 nF/km 0.70 mH/km

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
6

Setting Functions
ELECTRICAL SPECIFICATIONS Parameter setting is possible by using the infrared switches
Power Supply on the display or with HART® communication. For comfortable
- AC- type : 90 V to 264 V AC setting we recommend to use Yokogawas Field Mate setting
- AC- Ex-type : 90 V to 250 V AC tool (DTM based).
- DC- type : 20.5 V to 28.8 V DC
The option /MB2 can be set via Modbus and display.
Power consumption : max. 25 VA / 10 W
The option /MB3 can be set via Modbus, HART® and display.
External circuit breaker rating : 5 A, 250 V (The converter
doesn’t feature an installed power switch).
Display Function
Fuse on Base Board - Up to 4 lines.
- AC- type : 2 A, T, breaking capacity 1500A - 4 selectable languages (English, German, French, Russian)
- DC- type : 2 A, T, breaking capacity 1500A - Option /MB only available English

I/O Signals, including /KF1, /EF1, /UF1, /NF1, /GF1 Damping Functions
- Two active current outputs: Iout1, Iout2 The damping function is adjustable from 0 to 200 seconds
4 to 20 mA DC, galvanic separated from other signals, and affects the display and outputs.
Load resistance : 20 Ω to 600 Ω
Ambient temperature effect : < 0.05% of span/10°C Isolation Resistance of Converter
Linearity : 0.008 mA = 0.05% of span When surge arrestors are removed
Setting range URV for liquids: - between power and ground terminal: >100 MΩ / 500 V DC
-Qmax to + Qmax , Abs(URV-LRV) ≥ 5% of Qnom - between power and I/O terminals : >20 MΩ / 100 V DC
Setting range URV for gases: - between I/O terminals and ground : >20 MΩ / 100 V DC
-Qmax to + Qmax , Abs(URV-LRV) ≥ 1% of Qnom
- Two pulse outputs / status outputs : Pout, Sout Dielectric Strength
Passive transistor contact output, 30 V DC, 200 mA When surge arrestors are removed
Output rate - between power and ground terminal : 1,500 V AC for 1 minute
Output 1 : 0 to 10000 pulses/s
Output 2 : 0 to 2000 pulses/s Lightning Protection
Option /NM : passive, according EN 60947-5-6 Arresters 2000 A are inside of the converter for power
Option /AP : active output, 12 V, 6 mA, RL> 10 kΩ supply lines.
Active pulse output is not isolated from current output 2
As frequency output Electromagnetic Compatibility
Output 1 : 20 Hz to 10000 Hz Acc. IEC 61326-1: Class A, Table 2
Output 2 : 20 Hz to 2000 Hz IEC 61326-2-3
- Status input : Voltage-free contact IEC 61000-3-2
Closed : < 200 Ω IEC 61000-3-3
Open : > 100 kΩ
Safety Requirement Standards
Option /KF5, /EF5, /UF5, /NF5, /GF5: 3 intrinsically safe
Acc. IEC 61010-1
outputs
IEC 61010-2-030
- Two passive current outputs (additional power supply
Overvoltage category II
needed) : Iout1, Iout2, 4 to 20 mA DC, galvanically separated
Pollution degree 2
from other signals.
Supply voltage : 10.5 V to 30 V DC (without HART®), 165 mA
Supply voltage : 16.75 V to 30 V DC (with HART®), 165 mA
Load resistance : 20 Ω ... 600 Ω
Ambient temperature effect : < 0.05% of span/10°C
- One pulse output / status output : Pout
Passive transistor contact output, 30 V DC, 100 mA
Output rate : 0 to 2000 pulses/s
As frequency output : 20 Hz to 2000 Hz
- No status input

Digital Communication
- HART® communication protocol rev. 5, superimposed on
4 -20 mA DC signal (Iout1)
- Load resistance : 230 Ω to 600 Ω (including cable)
- Power line spacing : >15 cm, avoid parallel wiring
- Cable length : ≤ 2 km if „CEV” cables are used
- FOUNDATION™ Fieldbus communication (/FB)
- see GS 01R04B05-00E
- MODBUS® communication (/MB2 + /MB3)
- Physical interface RS485 as two wire data bus according
EIA/TIA-485
- Maximal bus length is depending from bus topology and
communication speed.
- Addresses: 1 to 247
- Baud rates: 1200, 2400, 4800, 9600, 19200, 38400,
57600, 76800, 115200
- Data formats: E/1, O/1, N/1, N/2 (parity/stop bits)
- Modes: RTU, ASCII

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
7

HAZARDOUS AREA SPECIFICATIONS Remote converter RCCF31 (Option /KF5) :


- KEMA 02ATEX 2183 X
- Flame proof with Intrinsically safe connection to detector (ib)
ATEX - Additional Intrinsically safe outputs.
- II 2 (1) G Ex d [ia Ga] [ib] IIC T6 Gb or
Remote detector RCCS30LR ... 33 (Option /KS1): Ex d e [ia Ga] [ib] IIC T6 Gb
- KEMA 01ATEX 1075 X - II 2 (1) G Ex d [ia IIC Ga] [ib] IIB T6 Gb or
- Intrinsically safe Ex d e [ia IIC Ga] [ib] IIB T6 Gb with option /HP
- II 2G Ex ib IIB/IIC T1 ... T6 Gb - II 2 (1) D Ex tb [ia Da] [ib] IIIC T75°C Db
- II 2D Ex ib IIIC Txxx Db - Max. surface temperature : 75°C (167°F)
(xxx = max. surface temperature see below) - Degree of protection : IP66/67
- Max. surface temperature : - Power supply : 90 to 250 V AC, 50/60 Hz or
Standard : 150°C (302°F) 20.5 to 28.8 V DC
/MT : 260°C (500°F) - Power consumption : max. 25 VA / 10 W
- Degree of protection : IP66/67 - Ambient humidity : 0 to 95% RH
- Ambient humidity : 0 to 95% RH - Ambient temperature range: -40°C to +55°C (-40°F to 131°F)
- Ambient temperature range
: -50°C to +80°C (-58°F to 176°F)
- Process temperature limits : Remote converter RCCR31 (Option /KS1) :
Standard : -50°C to 150°C (-58°F to 302°F) - KEMA 02ATEX 2183 X
Option /MT : -50°C to 260°C (-58°F to 500°F) - Associated apparatus with Intrinsically safe connection
- Heat carrier fluid temperature limits to detector (ib)
Standard : 0°C to 150°C (32°F to 302°F) - II (2) G [Ex ib Gb] IIC
Option /MT : 0°C to 200°C (32°F to 392°F) - II (2) G [Ex ib Gb] IIB with option /HP
- II (2) D [Ex ib Db] IIIC
Remote detector RCCS34 ... 39/XR (Option /KS1): - Power supply : 90 to 250 V AC, 50/60 Hz or
- KEMA 01ATEX 1075 X 20.5 to 28.8 V DC
- Intrinsically safe - Power consumption : max. 25 VA / 10 W
- II 2G Ex ib IIB/IIC T1 ... T6 Gb - Ambient humidity : 0 to 95% RH
- II 2D Ex ib IIIC Txxx Db - Ambient temperature range: -40°C to +55°C (-40°F to 131°F)
(xxx = max. surface temperature see below)
- Max. surface temperature :
Standard + /LT : 150°C (302°F) WARNING
/MT : 220°C (500°F) Remote rack-mount converter RCCR31 must be installed in
/HT : 350°C (662°F) safe area !
- Degree of protection : IP66/67
- Ambient humidity : 0 to 95% RH Integral type RCCT34 ... 39/XR (Option /KF1) :
- Ambient temperature range - KEMA 02ATEX 2183 X
Standard , option /LT and option /MT - Flame proof with Intrinsically safe connection to detector (ib)
: -50°C to +80°C (-58°F to 176°F) - II 2G Ex d ib IIC T6...T3 Gb or Ex d e ib IIC T6...T3 Gb
Option /HT (process temperature < 280°C (536°F) - II 2G Ex d ib IIB T6...T3 Gb or Ex d e ib IIB T6...T3 Gb
: -50°C to +65°C (-58°F to 149°F) with option /HP
Option /HT (process temperature < 350°C (662°F) - II 2D Ex ib tb IIIC T150°C Db
: -50°C to +55°C (-58°F to 131°F) - Max. surface temperature : 150°C (302°F)
- Process temperature limits : - Degree of protection : IP66/67
Standard : -50°C to 150°C (-58°F to 302°F) - Power supply : 90 to 250 V AC, 50/60 Hz or
Option /LT : -200°C to 150°C (-328°F to 302°F) 20.5 to 28.8 V DC
Option /MT : -50°C to 220°C (-58°F to 428°F) - Power consumption : max. 25 VA / 10 W
Option /HT : 0°C to 350°C (32°F to 662°F) - Ambient humidity : 0 to 95% RH
- Heat carrier fluid temperature limits - Ambient temperature range: -40°C to +55°C (-40°F to 131°F)
Standard : 0°C to 150°C -32°F to 302°F)
Option /MT : 0°C to 220°C (32°F to 428°F) Integral type RCCT34 ... 39/XR (Option /KF5) :
Option /HT : 0°C to 350°C (32°F to 662°F) - KEMA 02ATEX 2183 X
- Flame proof with Intrinsically safe connection to detector (ib)
Remote converter RCCF31 (Option /KF1) : - Additional Intrinsically safe outputs.
- KEMA 02ATEX 2183 X - II 2 (1) G Ex d ib [ia Ga] IIC T6...T3 Gb or
- Flame proof with Intrinsically safe connection to detector (ib) Ex d e ib [ia Ga] IIC T6...T3 Gb
- II 2G Ex d [ib] IIC T6 Gb or Ex d e [ib] IIC T6 Gb - II 2 (1) G Ex d ib [ia IIC Ga] IIB T6...T3 Gb or
- II 2G Ex d [ib] IIB T6 Gb or Ex d e [ib] IIB T6 Gb Ex d e ib [ia IIC Ga] IIB T6...T3 Gb with option /HP
with option /HP - II 2 (1) D Ex ib tb [ia Da] IIIC T150°C Db
- II 2D Ex tb [ib] IIIC T75°C Db - Max. surface temperature : 150°C (302°F)
- Max. surface temperature : 75°C (167°F) - Degree of protection : IP66/67
- Degree of protection : IP66/67 - Power supply : 90 to 250 V AC, 50/60 Hz or
- Power supply : 90 to 250 V AC, 50/60 Hz or 20.5 to 28.8 V DC
20.5 to 28.8 V DC - Power consumption : max. 25 VA / 10 W
- Power consumption : max. 25 VA / 10 W - Ambient humidity : 0 to 95% RH
- Ambient humidity : 0 to 95% RH - Ambient temperature range: -40°C to +55°C (-40°F to 131°F)
- Ambient temperature range: -40°C to +55°C (-40°F to 131°F) - Process temperature range : -50°C to 150°C (-58°F to 302°F)

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
8

Electrical data remote detector RCCS30LR ... 33 : FM (For USA and Canada)
- Driving circuit : terminals D+ and D
Ex ib IIC : Ui = 16 V; Ii = 53 mA; Pi = 0.212 W Remote detector RCCS30LR ... 39/XR (Option /FS1) :
Li = 4.2 mH; Ci = negligible small - Intrinsically safe
Ex ib IIB : Ui = 16 V; Ii = 153 mA; Pi = 0.612 W - AEx ia IIC, Class 1, Zone 0
Li = 4.2 mH; Ci = negligible small - IS Class I, Division 1, Groups A, B, C, D T6
- Sensor circuits: terminals S1+ and S1- or S2+ and S2- - DIP Class II / III, Division 1, Groups E, F, G
Ex ib IIC : Ui = 16 V; Ii = 80 mA; Pi = 0.32 W - IP67 / NEMA 4X
Li = 4.2 mH; Ci = negligible small - Ambient temperature range : -50°C to +80°C (-58°F to 176°F)
- Temperature sensor circuit : terminals TP1, TP2, TP3
Ex ib IIC : Ui = 16 V; Ii = 50 mA; Pi = 0.2 W Remote converter RCCF31 (Option /FF1, /FF5) :
Li = negligible small; Ci = negligible small - Housing explosion proof
- Provides Intrinsically safe detector circuits
Electrical data remote detector RCCS34 ... 39/XR : - AEx d [ia] IIC, Class I, Zone 1, T6
- Driving circuit : terminals D+ and D - AEx d [ia] IIB, Class I, Zone 1, T6 with option /HP
Ex ib IIC : Ui = 16 V; Ii = 53 mA; Pi = 0.212 W - Class I, Division 1, Groups A, B, C, D
Li = 3.2 mH; Ci = negligible small - Class I, Division 1, Groups C, D with option /HP
Ex ib IIB : Ui = 16 V; Ii = 153 mA; Pi = 0.612 W - Class II / III, Division 1, Groups E, F, G
Li = 3.2 mH; Ci = negligible small - AIS Class I / II / III, Division 1, Groups A, B, C, D, E, F, G
- Sensor circuits: terminals S1+ and S1- or S2+ and S2- - AIS Class I / II / III, Division 1, Groups C, D, E, F, G with /HP
Ex ib IIC : Ui = 16 V; Ii = 80 mA; Pi = 0.32 W
- IP67 / NEMA 4X
Li = 2.1 mH; Ci = negligible small
- Temperature sensor circuit : terminals TP1, TP2, TP3 - Ambient temperature range : -40°C to +50°C (-40°F to 122°F)
Ex ib IIC : Ui = 16 V; Ii = 50 mA; Pi = 0.2 W
Li = negligible small; Ci = negligible small Remote converter RCCR31 (Option /FS1) :
- Intrinsically safe associated apparatus
Electrical data remote converter RCCF31, RCCR31 and - Provides Intrinsically safe detector circuits
Converter of integral type RCCT3 : - [AEx ia] IIC, Class I, Zone 1
- Driving circuit : terminals D+ / D- - [AEx ia] IIB, Class I, Zone 1, T6 with option /HP
Ex [ib] IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W - AIS Class I, Division 1, Groups A, B, C, D
Lo = 15 mH; Co = 0.65 µF - AIS Class I, Division 1, Groups C, D with option /HP
Ex [ib] IIB : Uo = 11.7 V; Io = 124 mA; Po = 0.363 W - Ambient temperature range : -40°C to +50°C (-40°F to 122°F)
Lo = 8 mH; Co = 10.3 µF
- Sensor circuits: terminals S1+/ S1- or S2+ / S2- Integral type RCCT34 ... 39/XR (Option /FF1, /FF5) :
Ex [ib] IIB/IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W - Housing explosion proof
Ex [ib] IIC : Lo = 15 mH; Co = 0.65 µF - AEx d [ia] IIC, Class I, Zone 1, T6
Ex [ib] IIB : Lo = 60 mH; Co = 4.07 µF - AEx d [ia] IIB, Class I, Zone 1, T6 with option /HP
- Temperature sensor circuit : terminals TP1, TP2, TP3 - Class I, Division 1, Groups A, B, C, D
Ex [ib] IIB/IIC : Uo = 13.3 V; Io = 40 mA; Po = 0.133 W - Class I, Division 1, Groups C, D with option /HP
Ex [ib] IIC : Lo = 20 mH; Co = 0.91 µF - Class II / III, Division 1, Groups E, F, G
Ex [ib] IIB : Lo = 80 mH; Co = 5.6 µF - IP67 / NEMA 4X
- Current outputs (only option /KF5) : - Ambient temperature range : -40°C to +50°C (-40°F to 122°F)
Ex ia IIC : Ui = 30 V; Ii = 300 mA; Pi = 1.25 W
Li = 4 µH; Ci = 5.5 nF Process temperature range :
- Standard : -50°C to 150°C / (-58°F to 302°F)
- Pulse output (only option /KF5) : - with option /LT : -200°C to 150°C / (-328°F to 302°F)
Ex ia IIC : Ui = 30 V; Ii = 300 mA; Pi = 1.25 W - with option /MT (RCCS30LR...33)
Li = 4 µH; Ci = 13.1 nF : -50°C to 260°C / (-58°F to 500°F)
- with option /MT (RCC34...39/XR)
For temperature classification see table 7. : -50°C to 220°C / (-58°F to 428°F)
- with option /HT : 0°C to 350°C / (-32°F to 662°F)

Heat carrier fluid temperature range :


- Standard : 0°C to 150°C / 32°F to 302°F
- with option /MT (RCCS30LR...33)
: -50°C to 200°C (-58°F to 392°F)
- with option /MT (RCC34...39/XR)
: -50°C to 220°C (-58°F to 428°F)
- with option /HT : 0°C to 350°C (32°F to 662°F)

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
9

Electrical data remote converter RCCF31, RCCR31 and IECEx APPROVAL


Converter of Integral Type RCCT3 : Certificate: IECEx KEM 06.0031X
- Driving circuit : terminals D+ / D-
Uo = 14.5 V; Io = 47 mA; Po = 0.171 W Remote detector RCCS30LR ... 33 (Option /ES1):
Lo = 15 mH; Co = 0.65 µF - Intrinsically safe
- Driving circuit : terminals D+ / D- with option /HP - Ex ib IIB/IIC T1 ... T6 Gb
Uo = 11.7 V; Io = 124 mA; Po = 0.363 W - Ex ib IIIC Txxx Db
Lo = 8 mH; Co = 10.3 µF (xxx = max. surface temperature see below)
- Sensor circuits: terminals S1+/ S1- or S2+ / S2- - Max. surface temperature :
Uo = 14.5 V; Io = 47 mA; Po = 0.171 W Standard : 150°C (302°F)
Lo = 15 mH; Co = 0.65 µF /MT : 260°C (500°F)
- Temperature sensor circuit : terminals TP1, TP2, TP3 - Degree of protection : IP66/67
Uo = 13.3 V; Io = 40 mA; Po = 0.133 W - Ambient humidity : 0 to 95% RH
Lo = 20 mH; Co = 0.91 µF - Ambient temperature range
: -50°C to +80°C (-58°F to 176°F)
Electrical data remote detector RCCS30LR ... 33 : - Process temperature range :
- Driving circuit : terminals D+ and D Standard : -50°C to 150°C (-58°F to 302°F)
Groups A-D: Ui = 16 V; Ii = 53 mA; Pi = 0.212 W Option /MT : -50°C to 260°C (-58°F to 500°F)
Li = 4.2 mH; Ci = negligible small - Heat carrier fluid temperature rangev
Groups C,D: Ui = 16 V; Ii = 153 mA; Pi = 0.612 W Standard : 0°C to 150°C (32°F to 302°F)
Li = 4.2 mH; Ci = negligible small Option /MT : 0°C to 200°C (32°F to 392°F)
- Sensor circuits: terminals S1+ and S1- or S2+ and S2-
Ui = 16 V; Ii = 80 mA; Pi = 0.32 W Remote detector RCCS34 ... 39/XR (Option /ES1):
Li = 4.2 mH;Ci = negligible small - Intrinsically safe
- Temperature sensor circuit : terminals TP1, TP2, TP3 - Ex ib IIB/IIC T1 ... T6 Gb
Ui = 16 V; Ii = 50 mA; Pi = 0.2 W - Ex ib IIIC Txxx Db
Li = negligible small; Ci = negligible small (xxx = max. surface temperature see below)
-Max. surface temperature :
Electrical data remote detector RCCS34 ... 39/XR : Standard + /LT : 150°C (302°F)
- Driving circuit : terminals D+ and D /MT : 220°C (500°F)
Groups A-D: Ui = 16 V; Ii = 53 mA; Pi = 0.212 W /HT : 350°C (662°F)
Li = 3.2mH; Ci = negligible small - Degree of protection : IP66/67
Groups C,D: Ui = 16 V; Ii = 153 mA; Pi = 0.612 W - Ambient humidity : 0 to 95% RH
Li = 3.2mH; Ci = negligible small - Ambient temperature range
- Sensor circuits: terminals S1+ and S1- or S2+ and S2- Standard , option /LT and option /MT
Ui = 16 V; Ii = 80 mA; Pi = 0.32 W : -50°C to +80°C (-58°F to 176°F)
Li = 2.1 mH;Ci = negligible small Option /HT (process temperature < 280°C (536°F)
- Temperature sensor circuit : terminals TP1, TP2, TP3 : -50°C to +65°C (-58°F to 149°F)
Ui = 16 V; Ii = 50 mA; Pi = 0.2 W Option /HT (process temperature < 350°C (662°F)
Li = negligible small; Ci = negligible small : -50°C to +55°C (-58°F to 131°F)
- Process temperature range :
The remote converter RCCF31 has a T6 temperature class Standard : -50°C to 150°C (-58°F to 302°F)
rating for operation at ambient temperature up to +50°C / +122°F. Option /LT : -200°C to 150°C (-328°F to 302°F)
Option /MT : -50°C to 220°C (-58°F to 428°F)
Special conditions :
Option /HT : 0°C to 350°C (32°F to 662°F)
- ROTAMASS with FM approval is only available with ANSI - Heat carrier fluid temperature range
1/2” NPT cable conduit connection “A”. Standard : 0°C to 150°C (32°F to 302°F)
- The flowmeter must be connected to the potential Option /MT : 0°C to 220°C (32°F to 428°F)
equalization system. Option /HT : 0°C to 350°C (32°F to 662°F)
- For AC-version maximum power supply is 250V AC. Remote converter RCCF31 (Option /EF1) :
- Use conduit seals within 18 inches for power supply- and - Flame proof with Intrinsically safe connection to detector (ib)
IO- cable entries at RCCT3 / RCCF31. - Ex d [ib] IIC T6 Gb or Ex d e [ib] IIC T6 Gb
- Ex d [ib] IIB T6 Gb or Ex d e [ib] IIB T6 Gb
For temperature classification see table 7. with option /HP
- Ex tb [ib] IIIC T75°C Db
- Max. surface temperature : 75°C (167°F)
- Degree of protection : IP66/67
- Power supply : 90 to 250 V AC, 50/60 Hz or
20.5 to 28.8 V DC
- Power consumption : max. 25 VA / 10 W
- Ambient humidity : 0 to 95% RH
- Ambient temperature range: -40°C to +55°C (-40°F to 131°F)

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
10

Remote converter RCCF31 (Option /EF5) : Electrical data remote detector RCCS30LR ... 33 :
- Flame proof with intrinsically safe connection to detector (ib) - Driving circuit : terminals D+ and D
- Additional Intrinsically safe outputs. Ex ib IIC : Ui = 16 V; Ii = 53 mA; Pi = 0.212 W
- Ex d [ia Ga] [ib] IIC T6 Gb or Li = 4.2 mH; Ci = negligible small
Ex d e [ia Ga] [ib] IIC T6 Gb Ex ib IIB : Ui = 16 V; Ii = 153 mA; Pi = 0.612 W
- Ex d [ia IIC Ga] [ib] IIB T6 Gb or Li = 4.2 mH; Ci = negligible small
Ex d e [ia IIC Ga] [ib] IIB T6 Gb with option /HP - Sensor circuits: terminals S1+ and S1- or S2+ and S2-
- Ex tb [ia Da] [ib] IIIC T75°C Db Ex ib IIC : Ui = 16 V; Ii = 80 mA; Pi = 0.32 W
- Max. surface temperature : 75°C (167°F) Li = 4.2 mH; Ci = negligible small
- Degree of protection : IP66/67 - Temperature sensor circuit : terminals TP1, TP2, TP3
- Power supply : 90 to 250 V AC, 50/60 Hz or Ex ib IIC : Ui = 16 V; Ii = 50 mA; Pi = 0.2 W
20.5 to 28.8 V DC Li = negligible small; Ci = negligible small
- Power consumption : max. 25 VA / 10 W
- Ambient humidity : 0 to 95% RH Electrical data remote detector RCCS34 ... 39/XR :
- Ambient temperature range: -40°C to +55°C (-40°F to 131°F) - Driving circuit : terminals D+ and D
Ex ib IIC : Ui = 16 V; Ii = 53 mA; Pi = 0.212 W
Remote converter RCCR31 (Option /ES1) : Li = 3.2 mH; Ci = negligible small
- Associated apparatus with Intrinsically safe connection to Ex ib IIB : Ui = 16 V; Ii = 153 mA; Pi = 0.612 W
detector (ib) Li = 3.2 mH; Ci = negligible small
- [Ex ib Gb] IIC - Sensor circuits: terminals S1+ and S1- or S2+ and S2-
- [Ex ib Gb] IIB with option /HP Ex ib IIC : Ui = 16 V; Ii = 80 mA; Pi = 0.32 W
- [Ex ib Db] IIIC Li = 2.1 mH; Ci = negligible small
- Power supply : 90 to 250 V AC, 50/60 Hz or - Temperature sensor circuit : terminals TP1, TP2, TP3
20.5 to 28.8 V DC Ex ib IIC : Ui = 16 V; Ii = 50 mA; Pi = 0.2 W
- Power consumption : max. 25 VA / 10 W Li = negligible small; Ci = negligible small
- Ambient humidity : 0 to 95% RH
- Ambient temperature range: -40°C to +55°C (-40°F to 131°F) Electrical data remote converter RCCF31, RCCR31 and
converter of integral type RCCT3 :
- Driving circuit : terminals D+ / D-
WARNING Ex [ib] IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W
Remote rack-mount converter RCCR31 must be installed in Lo = 15 mH; Co = 0.65 µF
safe area ! Ex [ib] IIB : Uo = 11.7 V; Io = 124 mA; Po = 0.363 W
Lo = 8 mH; Co = 10.3 µF
Integral type RCCT34 ... 39/XR (Option /EF1) : - Sensor circuits: terminals S1+/ S1- or S2+ / S2-
- Flame proof with Intrinsically safe connection to detector (ib) Ex [ib] IIB/IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W
- Ex d ib IIC T6...T3 Gb or Ex [ib] IIC : Lo = 15 mH; Co = 0.65 µF
Ex d e ib IIC T6...T3 Gb Ex [ib] IIB : Lo = 60 mH; Co = 4.07 µF
- Ex d ib IIB T6...T3 Gb or - Temperature sensor circuit : terminals TP1, TP2, TP3
Ex d e ib IIB T6...T3 Gb with option /HP Ex [ib] IIB/IIC : Uo = 13.3 V; Io = 40 mA; Po = 0.133 W
- Ex ib tb IIIC T150°C Db Ex [ib] IIC : Lo = 20 mH; Co = 0.91 µF
- Max. surface temperature : 150°C (302°F) Ex [ib] IIB : Lo = 80 mH; Co = 5.6 µF
- Degree of protection : IP66/67 - Current output (only option /EF5) :
- Power supply : 90 to 250 V AC, 50/60 Hz or
20.5 to 28.8 V DC Ex ia IIC : Ui = 30 V; Ii = 300 mA; Pi = 1.25 W
- Power consumption : max. 25 VA / 10 W Li = 4 µH; Ci = 5.5 nF
- Ambient humidity : 0 to 95% RH - Pulse output (only option /EF5) :
- Ambient temperature range: -40°C to +55°C (-40°F to 131°F) Ex ia IIC : Uo = 30 V; Io = 300 mA; Po = 1.25 W
Lo = 4 µH; Co = 13.1 nF
Integral type RCCT34 ... 39/XR (option /EF5) :
- Flame proof with Intrinsically safe connection to detector (ib) For temperature classification see table 7.
- Additional Intrinsically safe outputs.
- Ex d ib [ia Ga] IIC T6...T3 Gb or
Ex d e ib [ia Ga] IIC T6...T3 Gb
- Ex d ib [ia IIC Ga] IIB T6...T3 Gb or
Ex d e ib [ia IIC Ga] IIB T6...T3 Gb with option /HP
- Ex ib tb [ia Da] IIIC T150°C Db
- Max. surface temperature : 150°C (302°F)
- Degree of protection : IP66/67
- Power supply : 90 to 250 V AC, 50/60 Hz or
20.5 to 28.8 V DC
- Power consumption : max. 25 VA / 10 W
- Ambient humidity : 0 to 95% RH
- Ambient temperature range: -40°C to +55°C (-40°F to 131°F)
- Process temperature range : -50°C to 150°C (-58°F to 302°F)

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
11

Table 7 : Temperature classification for ATEX, FM, IECEx, INMETRO, NEPSI and KOSHA certified flowmeter
RCCS30LR to RCCS33 RCCS30LR to RCCS33
without insulation with factory
insulation
Temp. class Max. ambient Max. process Max. ambient Max. process
temperature temperature temperature temperature
T6 50°C / 122°F 60°C / 140°F 60°C / 140°F 60°C / 140°F
T5 50°C / 122°F 80°C / 176°F 80°C / 176°F 90°C / 194°F
80°C / 176°F 100°C / 212°F
T4 80°C / 176°F 130°C / 266°F
50°C / 122°F 120°C / 248°F
T3 80°C / 176°F 180°C / 356°F 80°C / 176°F 180°C / 356°F
T2 80°C / 176°F 260°C / 500°F 80°C / 176°F 260°C / 500°F
T1 80°C / 176°F 260°C / 500°F 80°C / 176°F 260°C / 500°F

RCCS34 to RCCS39/XR RCCS34 to RCCS39/XR RCCT34 to RCCT39/XR


without insulation with factory insulation
Temp. class Max. ambient Max. process Max. ambient Max. process Max. ambient Max. process
temperature temperature temperature temperature temperature temperature
T6 40°C / 104°F 40°C / 104°F 65°C / 149°F 65°C / 149°F 55°C / 122°F 65°C / 149°F
T5 55°C / 131°F 55°C / 131°F 75°C / 167°F 75°C / 167°F 55°C / 122°F 80°C / 176°F
80°C / 176°F 100°C / 212°F
T4 70°C / 158°F 115°C / 239°F 55°C / 122°F 115°C / 239°F
40°C / 104°F 120°C / 248°F
80°C / 176°F 160°C / 320°F
T3 70°C / 158°F 180°C / 356°F 55°C / 122°F 150°C / 302°F
40°C / 104°F 180°C / 356°F
T2 80°C / 176°F 220°C / 428°F 65°C /149°F 275°C / 527°F 55°C / 122°F 150°C / 302°F
T1 80°C / 176°F 220°C / 428°F 45°C / 113°F 350°C / 662°F 55°C / 122°F 150°C / 302°F

For customer insulation of RCCS30LR to 39/XR the following must be regarded :


The table "with factory insulation" is calculated with 80 mm insulation and k-factor = 0.4 W/m2K.
If your insulation data are worse than these use table "without insulation" !

INMETRO APPROVAL (For Brazil) EAC APPROVAL (For Russia, Kazakhstan, Belorussia)
Certificate TÜV 11.0419 X Certificate RU C-DE.ГБ08.B.00208
RCCS3 with option /US1.same as IECEx /ES1 RCCS3 with option /GS1
RCCT3 with options /UF1 or /UF5 same as IECEx /EF1 or /EF5 RCCT3 with options /GF1 or /GF5
RCCF31 with options /UF1 or /UF5 same as IECEx /EF1 or /EF5 RCCF31 with options /GF1 or /GF5
RCCR31 with option /US1 same as IECEx /ES1 RCCR31 with option /GS1
Same parameters and specifications as IECEx approval. Same parameters and specifications as IECEx approval.

NEPSI APPROVAL (For China) METROLOGICAL REGULATION IN CIS AND EAC


Certificate GYJ12.1381X COUNTRIES
RCCS3 with option /NS1, RCCT3 with options /NF1 + /NF5, Russia, Kazakhstan, Uzbekistan, Belorussia and Turkmenistan
RCCF31 with options /NF1 + /NF5, RCCR31 with option /NS1 are members of CIS.
Same parameters and specifications as IECEx approval Rotamass has “Pattern Approval Certificate of Measuring
except NEPSI has no dust proof certification. Instruments” and is registered as a measuring instrument in
Russia, Kazakhstan, Uzbekistan, Belorussia and Turkmenistan.
KOSHA APPROVAL (For Korea) Option /QR2 is for Kazakhstan.
Same parameters and specifications as IECEx approval. Option /QR3 is for Uzbekistan.
Meter with IECEx option must be ordered. For the Ukraine the test certificate of Rota Yokogawa is sufficient.
Therefore no special option exists.
TS APPROVAL (For Taiwan) Russia, Kazakhstan and Belorussia are covered by EAC.
Certificate ML0412007041H6 For export to CIS and EAC countries please contact your
Same parameters and specifications as IECEx approval. Yokogawa representative.
Meter with IECEx option must be ordered.
For export to Taiwan please contact your Yokogawa
representative regarding Taiwan Safety Mark.

PESO APPROVAL (For India)


Same parameters and specifications as ATEX approval.
Meter with ATEX option must be ordered.

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
12

PLANNING AND INSTALLATION HINTS Heat Tracing and Insulation


Basically the detector can be insulated by the customer.
To be sure not to overheat the connection box choose one of /
Design Limits T options (insulation or heat tracing from Yokogawa) or /S2.
It is the responsibility of the user to use the instrument within For process temperatures between 150°C (302°F) and 230°C
the given design limits. Erosion and corrosion influence the (446°F) (RCCS34 - 39/XR) or 260°C (500°F) (RCCS30LR - 33)
accuracy and may restrict the temperature / pressure limits. choose /MT option and remote installation. If ROTAMASS detec-
Therefore corrosion and erosion should be avoided. tor with /MT or /HT is not insulated, the accuracy specification
can not be guaranteed. The converter should not be exceeded
Installation more than 50°C (122°F). Therefore never insulate the converter
The flowmeter can be installed vertically, horizontally or in any and keep the neck free from insulation too. Yokogawa will not
other position, as long as the measuring tubes are completely take any liability regarding customer insulation.
filled with the measured liquid during measurement.
Relations between Options /MT, /HT, /S2 and /T ( /T1, /T2, /T3)
Redundant Installation The meters with high temperature options (/MT, /HT) can
If two flowmeters of the same size are installed in series be insulated either by the customer by using option /S2
mutual interference called cross talk may take place. Cross (prolonged neck) or by the factory through options /T.
talk occurs due to the fact that both meters have the same The /T options already include the option /S2 so that the /S2
resonance frequency. If serial installation is planned please option can not be selected in case of the /T options.
contact your Yokogawa representative who can ensure that If the meter is not properly insulated by the customer, the
a frequency adjustment is made to one of the meters at the accuracy specification can not be guaranteed.
factory.
Installation above 100°C (212°F) Process Temperature
Sizing To provide enough cooling the instrument should be installed
The measuring range and accuracy are virtually independent vertically or horizontally with the converter down. This is
of fluid conditions and size of the connecting pipe. Select a recommended for size RCC36 and larger without /T option.
suitable nominal size from pressure loss calculation. Check
whether the measuring range and accuracy at minimal flow Installation below 0°C (32°F) Process Temperature
fit the application. The calculations of the pressure loss are The detector can be insulated to prevent ice capping either
based on newtonian fluids. For correct calculation please con- by the customer or by the manufacturer. Ask your Yokogawa
tact your local Yokogawa representative. representative for special insulation. If the customer wants to
insulate by themselves a closed cell foam as insulation mate-
Sanitary Applications rial is recommended to avoid water siphon. In this case option
For sanitary applications select process connection S2, S4 or /S2 should be selected. For temperatures below -70°C (-94°F)
S8. The wetted surface will be Ra ≤ 1.6 µm. However, if option /LT is recommended (on request).
option /SF is selected the surface roughness will be
Ra < 0.8µm and with /SF2 a certificate with a 3- point rough- Zero Adjustment Function
ness measurement certificate is delivered. The EHEDG certifi-
Zero point can be adjusted either by setting the switches on
cate shows that ROTAMASS conforms to the EHEDG criteria
display or with the HART® communication or with status
regarding the capability to be cleaned by a CIP process. The
input when the fluid flow is stopped and the detector filled.
evaluation does not include the process connections and
To ensure no flow conditions isolation valves should be
seals.
installed. To achieve the specified accuracy a zero should be
performed at process conditions (temperature, pressure).
Cavitation
To avoid cavitation keep the back pressure of the fluid suf-
ficiently above the vapor pressure of the fluid. For low viscous Pressure / Temperature Dependencies of Process Connections
fluids following condition should be fulfilled at the given See also process pressure limits in chapter ”Normal operation
temperature: conditions”.
pback > pvapor + 0.7*∆p
With ∆p = pressure loss Rupture Disk
The rupture disk is used as annunciation method in the case
Long Term Stability of tube rupture (Dual Seal) preferable for high pressure gas
To get stable deflection of the tubes by the Coriolis forces the service. Practically a tube rupture of ROTAMASS is not known
stiffness and therefore the wall thickness has to kept constant to the manufacturer. For large sizes it cannot be expected
during measuring. With corrosion or erosion the meter factor that the full line pressure can be released via the rupture disk.
is drifting with time and recalibration is necessary. Select the If this is requested please contact Yokogawa for a special
suitable resistant tube material for the process! execution.

Recalibration Service
Yokogawa offers full recalibration service, if necessary with a
certificate traceable to German national standards. Please
contact your Yokogawa affiliate or directly Rota Yokogawa,
Germany.

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
13

Explosion Proof Concept The specified mass flow accuracy applies if the batch process
The detector is intrinsically safe Ex ib, the converter RCCT is >1 minute. For shorter batch time (Dt in s) the accuracy
and RCCF31 are flame (explosion) proof. The converter decreases with the square root of 60/Dt.
RCCR31 is an intrinsically safe associated apparatus. For short batches the opening and closing times of the valves
The driving power from converter to detector is limited and have to be longer than 2 seconds.
protected by an intrinsically safe barrier, which is part of the
converter. The barrier is protecting the detector either for gas Concentration Measurement for Liquids
group IIC or IIB (option /HP). The Standard Concentration Measurement (option /CST)
Option /KF5 delivers two passive intrinsically safe current and is suitable for concentration measurement of emulsions or
one pulse output, however the converter is flame (explosion) suspensions, where the density of the solid is assumed to
proof. be fix. It can also be used for (mainly low concentration)
solutions if the two fluids are not strongly interacting. The
Option /HP density change of the liquid components due to temperature
With option /HP the detector driving power is higher which is can normally be described with a linear or quadratic function
benefit to 2 phase flow. This is also true for non hazardous with very high accuracy within the desired measurement
applications. range. The coefficients of these function (linear and quadratic
thermal expansion coefficients) must be either known or have
Gas Measurement to be determined prior to using this function.
For gas applications please choose the option /GA. For interacting liquids the Advanced Concentration
Density reading below 0.3 kg/l is not possible. Volume flow is Measurement options should be used, these options can
calculated by using the fix density value stored in „Reference density“. be ordered using the appropriate /C concentration
Based on the selection of the gas density, the following volume measurement option. For more information please see
flow rates can be calculated; standard, reference, normal. TI 01R04B04-04E-E “Concentration Measurement with
Besides, the corresponding volume flow rate units can be ROTAMASS”.
selected. Some functions are unavailable for gas measurement,
including concentration measurement, empty pipe, slug or corrosion
detection.
Good and stress free installation is mandatory for a stable Zero.
Attention to resonance phenomenons has to be taken if gas
compressors are used in the pipe. Flow noise has to be avoided.
Batch Process

Density Measurement
There are 3 levels of density measurement. The standard adjustment and /K4 delivers an accuracy up to 0.001 g/cm³, if the fluid den-
sity is around 1 kg/l. However, at elevated temperatures the density error may increase. For option /K4 the instrument is
preheated ensuring long term stability. However, if high density stability is needed at high temperatures option /HT is recommended.
Option /K6 includes preheating, a full calibration at 3 different densities, increased temperature measurement specification and individual
adjustment of the fluid temperature dependency. Multi-phase flow can generate higher deviations. The higher the density differences
of the single components are the more likely it is that a negative density error is generated. Aeration has to be avoided fully to receive
good density measurement.
For more information please see TI 01R04B04-05E ”Density Measurement with ROTAMASS”.
Note: Density specification under calibration condition only with flow direction “forward” according the arrow on the meter.
Table 8: Overview density-/volume- flow measurement:
Option Accuracy Certificate Description Typical Application
Standard ± 0.0015 g/cm³ to Standard (mass flow) - Standard adjustment with water and air - Process medium and environment are
± 0.02 g/cm³ factory calibration - Density constants given in mass flow approximately at room temperature, the density
certificate certificate range is 0.9 kg/l to 1.1 kg/l
Option /K4 ± 0.001 g/cm³ Standard (mass flow) - Thermal treatment of the sensor and - Improved volume flow accuracy
factory calibration special hardware design - Process medium up to 150°C, for higher
certificate - Standard adjustment with water and air temperature select option /MT or /HT
- Density constants given in mass flow - Density range is 0.9 kg/l to 1.1 kg/l
certificate
Option /K6 ± 0.0005 g/cm³ Separate factory - Thermal treatment of the sensor and - Density and concentration measurement in
density calibration special hardware design addition to the mass flow:
certificate - Density calibration with 3 different liquids - Process medium up to 150°C, for higher
- Individual adjustment of the fluid temperature select option /HT
temperature dependency - Density range 0.3 kg/l to 2.5 kg/l
- Best volume flow accuracy

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
14

Table 9 : Pressure rating


Process Temperature
Type of process connection
RT 3) 50°C 100°C 150°C 200°C 250°C 300°C 350°C
A1 2) Flange acc. ASME B16.5 Class 150 19 bar 18.4 bar 16.2 bar 14.8 bar 13.7 bar 12.1 bar 10.2 bar 8.4 bar
A2 2) Flange acc. ASME B16.5 Class 300 49.6 bar 48.1 bar 42.2 bar 38.5 bar 35.7 bar 33.4 bar 31.6 bar 30.3 bar
A3 2) Flange acc. ASME B16.5 Class 600 excl. RCC39/XR 99.3 bar 96.2 bar 84.4 bar 77 bar 71.3 bar 66.8 bar 63.2 bar 60.7 bar
A3 2) Flange acc. ASME B16.5 Class 600 for RCC39/XR 95 bar 89 bar 80 bar 73 bar 67 bar 62 bar 59 bar 58 bar
A4 2) Flange acc. ASME B16.5 Class 900 without /DS 148.9 bar 144.3 bar 126.6 bar 115.5 bar 107 bar 100.1 bar 94.9 bar 91 bar
A4 2) 4) Flange acc. ASME B16.5 Class 900 with /DS 130 bar 126 bar 110.5 bar 100.8 bar 93.4 bar 87.4 bar 82.9 bar 79.4 bar
A5 2) Flange acc. ASME B16.5 Class 1500 excl. RCC36 248.2 bar 240.6 bar 211 bar 192.5 bar 178.3 bar 166.9 bar 158.1 bar 151.6 bar
A5 2) Flange acc. ASME B16.5 Class 1500 for RCC36 210 bar 203 bar 176 bar 160 bar 148 bar 140 bar 133 bar 128 bar
D2 1) Flange acc. EN 1092-1 PN 16 16 bar 15.7 bar 15.1 bar 13.7 bar 12.7 bar 11.9 bar 11.0 bar 10.5 bar
D4 1) Flange acc. EN 1092-1 PN 40 40 bar 39.2 bar 37.9 bar 34.4 bar 31.8 bar 29.9 bar 27.6 bar 26.4 bar
D5 1) Flange acc. EN 1092-1 PN 63 63 bar 61.8 bar 59.7 bar 54.3 bar 50.1 bar 47.1 bar 43.5 bar 41.7 bar
D6 1) Flange acc. EN 1092-1 PN 100 100 bar 98 bar 94.7 bar 86.1 bar 79.5 bar 74.7 bar 69.0 bar 66.1 bar
D7 1) Flange acc. EN 1092-1 PN 160 160 bar 156.9 bar 151.6 bar 137.9 bar 127.2 bar 119.6 bar 110.4 bar 105.9 bar
G9 1) Internal thread RCCS30LR ... 33 285 bar 271 bar 247 bar 227 bar 208 bar 183 bar ----------------------------
T9 1) Internal thread NPT RCCS30LR ... 33 285 bar 271 bar 247 bar 227 bar 208 bar 183 bar ----------------------------
G9 1) 4) Internal thread RCCS34 260 bar 251 bar 231 bar 208 bar 190 bar 178 bar 167 bar 160 bar
T9 1) 4) Internal thread NPT RCCS34 260 bar 251 bar 231 bar 208 bar 190 bar 178 bar 167 bar 160 bar
Process Temperature
up to 120°C 220°C 300°C 350°C
J1 1) Flange acc. JIS B 2220 10K 14 bar 12 bar 10 bar ------
J2 1) Flange acc. JIS B 2220 20K 34 bar 31 bar 29 bar 26 bar
Process Temperature
up to 140°C *)
Pipe connection up to DN 40 40 bar
25 bar
*)
under the restriction using suitable gasket
S2 1) acc. DIN 11851 DN 50 to DN 100
materials
above DN 100 16 bar
Process Temperature
up to 150°C **)
Clamp connection up to DN 50 16 bar
S4 1)
acc. DIN 32676 above DN 50 10 bar
**)
under the restriction using suitable gasket
Clamp acc. Mini-Clamp up to 1/2” 16 bar
materials
S8 1)
Clamp acc. Tri-Clamp® up to 2” 16 bar
above 2” 10 bar
Process Temperature
Type of process connection
RT 3) 120°F 210°F 300°F 390°F 480°F 570°F 660°F
A1 2) Flange acc. ASME B16.5 Class 150 276 psi 267 psi 235 psi 215 psi 199 psi 175 psi 148 psi 122 psi
A2 2) Flange acc. ASME B16.5 Class 300 719 psi 698 psi 612 psi 558 psi 518 psi 484 psi 458 psi 439 psi
A3 2) Flange acc. ASME B16.5 Class 600 excl. RCC39/XR 1440 psi 1395 psi 1224 psi 1117 psi 1034 psi 969 psi 917 psi 880 psi
A3 2) Flange acc. ASME B16.5 Class 600 for RCC39/XR 1378 psi 1291 psi 1160 psi 1059 psi 972 psi 899 psi 856 psi 841 psi
A4 2) Flange acc. ASME B16.5 Class 900 without /DS 2160 psi 2093 psi 1836 psi 1675 psi 1552 psi 1452 psi 1376 psi 1320 psi
A4 2) 4) Flange acc. ASME B16.5 Class 900 with/DS 1885 psi 1827 psi 1602 psi 1461 psi 1354 psi 1267 psi 1202 psi 1151 psi
A5 2) Flange acc. ASME B16.5 Class 1500 excl. RCC36 3600 psi 3490 psi 3060 psi 2792 psi 2586 psi 2421 psi 2293 psi 2199 psi
A5 2) Flange acc. ASME B16.5 Class 1500 for RCC36 3046 psi 2944 psi 2553 psi 2321 psi 2147 psi 2031 psi 1929 psi 1856 psi
D2 1) Flange acc. EN 1092-1 PN 16 232 psi 227 psi 219 psi 198 psi 184 psi 172 psi 159 psi 152 psi
D4 1) Flange acc. EN 1092-1 PN 40 580 psi 568 psi 549 psi 499 psi 458 psi 433 psi 400 psi 382 psi
D5 1) Flange acc. EN 1092-1 PN 63 914 psi 896 psi 865 psi 787 psi 726 psi 683 psi 630 psi 604 psi
D6 1) Flange acc. EN 1092-1 PN 100 1450 psi 1421 psi 1373 psi 1248 psi 1152 psi 1083 psi 1000 psi 958 psi
D7 1) Flange acc. EN 1092-1 PN 160 2320 psi 2275 psi 2198 psi 1999 psi 1844 psi 1734 psi 1600 psi 1535 psi
G9 1) Internal thread RCCS30LR ... 33 4133 psi 3930 psi 3582 psi 3292 psi 3016 psi 2653 psi ----------------------------
T9 1) Internal thread NPT RCCS30LR ... 33 4133 psi 3930 psi 3582 psi 3292 psi 3016 psi 2653 psi ----------------------------
G9 1) 4) Internal thread RCCS34 3770 psi 3640 psi 3350 psi 3016 psi 2755 psi 2581 psi 2422 psi 2320 psi
T9 1) 4) Internal thread NPT RCCS34 3770 psi 3640 psi 3350 psi 3016 psi 2755 psi 2581 psi 2422 psi 2320 psi
Process Temperature
up to 248°F 428°F 572°F 662°F
J1 1) Flange acc. JIS B 2220 10K 1203 psi 174 psi 145 psi ------
J2 1) Flange acc. JIS B 2220 20K 493 psi 449 psi 420 psi 377 psi
Process Temperature
up to 284°F) *)
Pipe connection up to DN 40 580 psi
362 psi *)
under the restriction using suitable gasket
S2 1) acc. DIN 11851 DN 50 to DN 100
materials
above DN 100 232 psi
Process Temperature
up to 302°F **)
Clamp connection up to DN 50 232 psi
S4 1)
acc. DIN 32676 above DN 50 145 psi
**)
under the restriction using suitable gasket
Clamp acc. Mini-Clamp up to 1/2” 232 psi
materials
S8 1)
Clamp acc. Tri-Clamp® up to 2” 232 psi
above 2” 145 psi
1)
process connection material: 1.4404 / 1.4435 (equivalent to group 2.3 material AISI 316L acc. ASME B16.5)
2
) process connection material: 1.4401/1.4404 AISI 316/316L
3)
RT = Room Temperature; EN1092: -10°C to 50°C; ASME B16.5: -29°C to 38°C
4)
for option /DS max. pressure according A4. ASME class 900 13% derated

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
15

DIMENSIONS
Integral Type RCCT34, RCCT36, RCCT38, RCCT39
Without Display 266 (10.5)

147 (5.79) 32 (1.26) 30 (1.18)


30
(1.18)

Ø123 (4.84)

(1.93)
49
With Display
266 (10.5)
147 (5.79)
122 (4.8) 144 (5.67)
85 (3.35)
32 (1.26) 50 (1.97)
30
(1.18)

Ø123 (4.84)

49 (1.93)
H3

H2

(2.05)
52

H4
H1

L3 W1

L2

L1 ± 5 (0.2)
W2

Note: The flange dimensions depend on size and pressure rating of the flange.
Model L1 L2 L3 H1 H2 H3 H4 W1 W2 Weight
mm 13-24 kg
RCCT34 see table 11 272 (10.7) 212 (8.35) 177 (6.97) 214 (8.43) 279 (11) 80 (3.15) 60 (2.36) 80 (3.15)
(inches) (29-53 lbs)
mm 18-38 kg
RCCT36 see table 11 400 (15.7) 266 (10.5) 230 (9.06) 214 (8.43) 279 (11) 80 (3.15) 76 (2.99) 90 (3.54)
(inches) (40-84 lbs)
mm 28-53 kg
RCCT38 see table 11 490 (19.3) 267 (10.5) 269 (10.6) 224 (8.82) 289 (11.4) 100 (3.94) 89 (3.5) 110 (4.33)
(inches) (62-117 lbs)
mm 63-106 kg
RCCT39 see table 11 850 (33.5) 379 (14.9) 370 (14.6) 242 (9.54) 307(12.1) 135 (5.31) 129 (5.08) 160 (6.3)
(inches) (139-233 lbs)
Dimensions in mm (inches)..Weights with smallest and biggest flanges.

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
16

Integral Type RCCT39/IR-/V2


(10.6)

(4.86) (5.74)
(5.87)
(4.86)
(1.94)
(2.56)

(3.41)

(4.84)

(1.94)
(12.87)

(10.31)

(2.05)

(6.93)
(21.9)

(18.43)

(27.19)
(6.63)
(35.12)

(0.2) (see table 11)


(6.93)

Weigth 98...152 kg (215... 334 lbs)

Dimensions in mm (inches). Weights with smallest and biggest flanges

Integral Type RCCT39/XR


L1 ± 5 (0.2) (see table 11) 269
(10.59)
(2.04)

(9.44)
240

(14.96)
52

380
(11.02)
280
1280 (50.39)

R3 .81)
(11
00

230
(9.05)
(11.02)
280

Weigth 293...374 kg (645-823 lbs)

Dimensions in mm (inches). Weights with smallest and biggest flanges

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
17

Remote field-mount Converter RCCF31


Without Display

17 (0.66)
Ø102 (4.01) 75 (2.95)

201(7.91)
(4.72)
91.6 (3.6)

120
Ø123(4.84)

(2.51)

(12)
305
122 (4.80)
64
29 32 29
(1.14) (1.25) (1.14)

85 (3.34)
124 122
(4.88) (4.80) 148 (5.82)

With Display
Ø102 (4.01) 17 (0.66) 75 (2.95)

201(7.91)
(4.72)
120
91.6 (3.6)

Ø123 (4.84)

(12)
305
(2.51)
122 (4.80)
64

49 32 29
(1.92) (1.25) (1.14)

85 (3.34)
144 (5.66) 122 (4.80) 148 (5.82)

Weight with bracket: 5.5 kg (12.1 lbs) (depends on type)

Dimensions in mm (inches)

Remote rack-mount Converter RCCR31


Terminal Board, back view Cassette, side view Cassette front view
128 (3HE)

22 228 107 (21TE)

Dimensios in mm
19-inch rack, acc. IEC / DIN EN 60297-3-101
Weight: 1.5 kg
Dimensions in mm

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
18

Remote Detector RCCS30LR, RCCS30, RCCS31, RCCS32, RCCS33


210 126

180 25 20 29 25 57 46
150

L1 ± 3 s. table 11

98
Ø102
110
150

Ø8,5
Ø89

111 (RCCS30LR)
111 (RCCS30)
99 (RCCS31)
89 (RCCS32) Option /PD
55 (RCCS33) Weight 6...18kg (13...40lbs) Mounting set for DN50 tube
Dimensions in mm

Remote Detector RCCS34, RCCS36, RCCS38, RCCS39


Ø102 (4.01)

L1 ± 5 (0.2) Ø102 (4.01)

(3.14)
80
98 (3.85)

H6
H5
H4

H1

L3 W1 Option /S2
L2
W2

Note: The flange dimensions depend on size and pressure rating of the flange.
Model L1 L2 L3 H1 W1 W2 H4 H5 H6 Weight
mm 10-21 kg
RCCS34 see table 11 272 (10.7) 212 (8.35) 177 (6.97) 60 (2.36) 80 (3.15) 80 (3.15) 138 (5.43) 218 (8.58)
(inches) (22-46 lbs)
mm 15-35 kg
RCCS36 see table 11 400 (15.7) 266 (10.5) 230 (9.06) 76 (2.99) 90 (3.54) 80 (3.15) 138 (5.43) 218 (8.58)
(inches) (33-77 lbs)
mm 25-50 kg
RCCS38 see table 11 490 (19.3) 267 (10.5) 269 (10.6) 89 (3.5) 110 (4.33) 100 (3.94) 148 (5.82) 228 (8.97)
(inches) (55-110 lbs)
mm 60-103 kg
RCCS39 see table 11 850 (33.5) 379 (14.9) 370 (14.6) 129 (5.08) 160 (6.3) 135 (5.31) 166 (6.53) 245 (9.65)
(inches) (132-227 lbs)
Dimensions in mm (inches)..Weights with smallest and biggest flanges.

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
19

Remote Detector RCCS39/IR- /V2


(4.02)
(0.2) (see table 11)

(2.2.8)
(3.85)

(10.47)
(7.32)
(6.93)

(21.9)

(21.9)
(27.19)
(6.63) Option /S2
(35.12)

(4.02)
(6.93)

Weight 95...149 kg (209...328 lbs)

Dimensions in mm (inches). Weights with smallest and biggest flanges

Remote Detector RCCS39/XR


Ø102 (4.01)

(2.28)
L1 ± 5 (0.2) (see table 11)
(3.86)

(12.6)
320
98

58
(9.37)
238
(11.02)
280

(35.43)
900
(35.43)
900

750 (29.52) 230


(9.05) Option /S2
1140 (44.88)

(4.01)
Ø102
(11.02)

Weight 290...371 kg (639...815 lbs)


280

Dimensions in mm (inches). Weights with smallest and biggest flanges

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
20

Remote Detector RCCS39/XR with option /HT


L1 ± 5 (0.2) (see table 11)

Ø50

(7.0)
178
(2.68)

(12.48)
(1.96)

317
68

(2.28)
58
(11.02)
280

(35.07)
891
683 (26.88) 273
(10.74)
1140 (44.88)
Option /S2

Ø102 (4.01)
(11.02)
280

Weight 290...370 kg
(639...815 lbs)
Dimensions in mm (inches). Weights with smallest and biggest flanges

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
21

Remote Detector RCCS34, RCCS36, RCCS38, RCCS39, RCCS39/IR- /V2 with option /T
(Insulation / Heating)

L1 ± 5 (0.2)

Ø102 (4.01)

H6

H9
Process
connection

D1 ± 5 (0.2)

H7
D2 ± 5 (0.2)
Heating

H8
connection
option /T2 or /T3

L5
Ventilation
L4 ± 5 (0.2) option /T3
W3

Note: The flange dimensions depend on size and pressure rating of the flange.
Model L1 L4 L5 D1 D2 H6 H7 H8 H9 W3 Weight
mm see table 420 310 330 218 411 273 138 240 19-30 kg
RCCS34 200 (7.87)
(inches) 11 (16.53) (12.20) (12.99) (8.58) (16.81) (10.74) (5.43) (9.44) (42-66 lbs)
mm see table 540 439 250 380 218 464 326 138 260 28-48 kg
RCCS36
(inches) 11 (21.25) (17.28) (9.84) (14.96) (8.58) (18.26) (12.83) (5.43) (10.23) (62-106 lbs)
mm see table 640 530 250 430 228 524 376 148 260 40-65 kg
RCCS38
(inches) 11 (25.19) (20.86) (9.84) (16.92) (8.97) (20.62) (14.80) (5.82) (10.23) (88-144 lbs)
mm see table 1000 894 350 545 245 668 503 165 302 104-147 kg
RCCS39
(inches) 11 (39.37) (35.19) (13.77) (21.45) (9.65) (26.29) (19.80) (6.49) (11.88) (230-325 lbs)
mm see table 1050 944 350 677 266 944 625 193 342 147-201 kg
RCCS39/IR
(inches) 11 (41.34) (37.16) (13.77) (26.65) (10.47) (37.16) (24.61) (7.6) (13.46) (325-444 lbs)
Dimensions in mm (inches)..Weights with smallest and biggest flanges including insulation cover and heat tracing.
Standard heating connection according table 10.

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
22

Remote Detector RCCS30LR, RCCS30, RCCS31, RCCS32, RCCS33 with option /T (Insulation / Heating)
L1 ± 3 (0.1) s. table 11 176 (6.93)

Ø102 (4.01)
350 (13.77)

425 (16.73)

299 (11.77)
Heating
connection
option /T2 or /T3

134 (5.27)
Ventilation
option /T3
180 (7.08)
160 (6.29)

270 (10.62)
Dimensions in mm (inches)
Weight in kg (lbs) with flange Table 10: Heat tracing connection types depending on
Weight in kg (lbs) without flange
(01A1) process connection type
/T1 /T2 /T3 /T1 /T2 /T3
8.7 11.5 12.5 9.5 12.3 13.3 Process connection Standard heating connection *)
(19.2) (25.3) (27.5) (21) (27.1) (29.3) A ASME ½” - 150
D EN DN 15 PN 40
J JIS 10K DN15
S2 ; S4 EN DN 15 PN 40
S8 ASME ½” - 150
G9 EN DN 15 PN 40
T9 ASME ½”- 150
*) others on request

Adapter 1/2” NPT - G1/2 (option /AD2)

N1 N2 ØD B ØC L1 L2 L Weight
[mm (in.)] [mm (in.)] [mm (in.)] [[mm (in.)] [mm (in.)] [mm (in.)] [kg (lbs)]
1/2´´ NPT G1/2´´ 15 (0.59) 27 (1.06) 30 (1.18) 19 (0.74) 23 (0.90) 45 (1.77) 0.07 (0.003)

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
23

MODEL, SUFFIX AND OPTION CODES


Integral Type RCCT3, Model and Suffix Code
Model Suffix Code Description Restrictions

RCCT34 Nominal Value : 3 t/h = 50 kg/min (110,2 lbs/min)


RCCT36 Nominal Value : 10 t/h = 170 kg/min (374.78 lbs/min)
RCCT38 Nominal Value : 32 t/h = 533 kg/min (1175 lbs/min)
RCCT39 Nominal Value : 100 t/h = 1670 kg/min (3681.72 lbs/min)
RCCT39/IR Nominal Value : 250 t/h = 4170 kg/min (9193.27 lbs/min) only with /V2
RCCT39/XR Nominal Value : 500 t/h = 8340 kg/min (18386.55 lbs/min) only with /HP
Power supply -A 90 - 264 V AC
-D 24 V DC
Indicator direction H1 Detector installation horizontal, tubes down, recom. for liquid service
H2 Detector installation horizontal, tubes up, recommended for gas service /GA
V0 Detector installation vertical
N0 Without indicator
Cable conduit connection M M20 x 1.5, female thread with cable glands not with /FF1, /FF5
A ANSI ½” NPT, female thread without cable glands
Process connection size 1)
23 ¾” see table 11
01 DN 15, ½” see table 11
02 DN 25, 1” see table 11
04 DN 40, 1½” see table 11
05 DN 50, 2” see table 11
06 DN 65, 2½” see table 11
08 DN 80, 3” see table 11
10 DN 100, 4” see table 11
12 DN 125, 5” see table 11
15 DN 150, 6” see table 11
20 DN 200, 8” see table 11
Process connection rating and style 1) A1 ASME flange class 150, process connection dim. + facing acc. ASME B16.5 see table 11
A2 ASME flange class 300, process connection dim. + facing acc. ASME B16.5 see table 11
A3 ASME flange class 600, process connection dim. + facing acc. ASME B16.5 see table 11
A4 ASME flange class 900, process connection dim. + facing acc. ASME B16.5 see table 11
A5 ASME flange class 1500, process connection dim. + facing acc. ASME B16.5 see table 11
D2 EN flange PN 16, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D4 EN flange PN 40, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D5 EN flange PN 63, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D6 EN flange PN 100, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D7 EN flange PN 160, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
J1 JIS flange 10K, JIS B 2220 see table 11
J2 JIS flange 20K, JIS B 2220 see table 11
S2 Thread acc. DIN 11851 see table 11
S4 Clamp, process connection dimensions acc. DIN 32676 see table 11
S8 Clamp, process connection dim. acc. Tri-Clover® (Tri-Clamp®) and ½” Mini Clamp see table 11
G9 G, female thread see table 11
T9 NPT female thread see table 11
Material of wetted parts 1)
SL Stainless steel 316L (1.4404)
HC Hastelloy C-22 (2.4602) only RCCT34 to 39/IR
1)
see selection table „Process connection and materials“ (table 11)

Integral Type RCCT3, Option Code :


Options Option Description Restrictions
code
Hazardous Area Approvals /KF1 ATEX Flame proof converter + Intrinsically safe detector with /HP not for gas group IIC
/KF5 ATEX Flame proof converter + Intrinsically safe detector + Intrinsically safe outputs with /HP not for gas group IIC
2 passive analog output + 1 passive pulse output
/FF1 FM approval for USA+Canada, Flame proof converter + Intrinsically safe detector + 2 only with cable conduit “A”; with /HP not for groups A
active analog outputs + 2 passive pulse outputs + 1 status input and B
/FF5 FM approval for USA+Canada, Flame proof converter + Intrinsically safe detector + 2 only with cable conduit “A”; with /HP not for groups A
passive analog outputs + 1 passive pulse output and B
/EF1 IECEx Flame proof converter + Intrinsically safe detector with /HP not for gas group IIC
/EF5 IECEx Flame proof converter + Intrinsically safe detector + Intrinsically safe outputs with /HP not for gas group IIC
2 passive analog output + 1 passive pulse output
/UF1 INMETRO Flame proof converter + Intrinsically safe detector with /HP not for gas group IIC
/UF5 INMETRO Flame proof converter + Intrinsically safe detector + Intrinsically safe with /HP not for gas group IIC
outputs 2 passive analog output + 1 passive pulse output
/NF1 NEPSI Flame proof converter + Intrinsically safe detector with /HP not for gas group IIC; only with /CN
/NF5 NEPSI Flame proof converter + Intrinsically safe detector + Intrinsically safe outputs 2 with /HP not for gas group IIC; only with /CN
passive analog output + 1 passive pulse output
/GF1 EAC Flame proof converter + Intrinsically safe detector with /HP not for gas group IIC; only with /VE
/GF5 EAC Flame proof converter + Intrinsically safe detector + Intrinsically safe outputs with /HP not for gas group IIC; only with /VE
2 passive analog output + 1 passive pulse output
Custody Transfer Measurement /Q01 European MID Approval (fluids other than water) Refer to GS 01R04B07-00E
/Q20 NTEP approval 12-080A2, Accuracy class 0.3 acc. NIST Handbook 44 Refer to GS 01R04B07-00E
/Q21 NTEP approval 12-080A2, Accuracy class 0,3 acc. NIST Handbook 44, Heated Refer to GS 01R04B07-00E
Products

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
24

Integral Type RCCT3, Option Code (continued)


Options Option Description Restrictions
code
GOST /QR2 Primary Calibration and Test Confirmation valid in Kazakhstan see page 11; not with /Q20, /Q21; only with /VE
/QR3 Primary Calibration and Test Confirmation valid in Uzbekistan see page 11; not with /Q20, /Q21
Dual Seal Approval /DS Dual Seal approval (conform with ANSI/ISA-12.27.01) not with process connection A5
/RD Rupture disk, rupture pressure 20 bar, nominal diameter 8 mm preferable with /GA, mandatory if /DS+/GA is selected
High Driving Power /HP High Driving Power; recommended for RCCT36 to 39, strongly recommended for mandatory for RCCT39/XR
RCCT39/IR 1) , Please see „ Hazardous Area Specifications“
Communication /FB Digital communication (FOUNDATION™ Fieldbus protocol) Refer to GS 01R04B05-00E
/MB2 Modbus communication protocol for all parameters and values not with /KF5, /EF5, /UF5, /GF5, /NF5, /FF5, /Q01,
/Q20, /Q21, /AP, /NM, /MB3
/MB3 Modbus communication protocol for all parameters and values incl. HART not with /KF5, /EF5, /FF5, /UF5, /NF5, /GF5, /Q01, /MB2
Communication, analogue output, pulse/status output, status input
Active Pulse Output /AP One active pulse output not with /KF5, /EF5, /FF5, /UF5, /NF5, /GF5, /NM
NAMUR Switch /NM One pulse output acc. to EN 60947-5-6 (NAMUR) not with /AP, not with /KF5, /EF5, /UF5, /NF5, /GF5
Tag Number /BG With customer specified tag number on name plate max. 16 characters; the last 8 significant characters are
used for HART® Tag;
characters acc. Codepage 850 (Multi-lingual)
HART® Tag Number (Software Tag) /BT1 With customer specified tag number for HART® communication in converter 22 characters for long tag;
characters acc. Codepage 850 (Multi-lingual)
Flange Facing /DN Flange with safety grooves acc. to EN 1092-1 form D only for D2 to D6; not HC
/EN Flange with spigot acc. to EN 1092-1 form E only for D2 to D6; not HC
/FN Flange with recess acc. to EN 1092-1 form F only for D2 to D6; not HC
/RJ Ring Type Joint Flanges only for A3, A4, A5; not HC
Gas Measurement /GA Gas measurement, special factory adjustments and settings to be conform with ANSI/ISA-12.27.01 select /RD; not
with /Q20, /Q21
Special Calibration /K2 2) Custom 5 pts mass-/volume-flow calibration using water with factory certificate
(traceable to German national standards)
/K4 Density adjustment + thermal treatment; (accuracy: 0.001 g/cm³) only RCCT34 to 39; not with /GA
/K5 2) Custom 10 pts mass-/volume-flow calibration using water with DAkkS certificate
(according EN-17025:2005)
/K6 Density calibration with 3 different fluids incl. individual temperature compensation only RCCT34 to 39; not with /GA
with certificate (accuracy: 0.0005 g/cm³)
Certificates /P2 Certificate of compliance with the order acc. to EN 10204:2004 -2.1
/P3 Test Report acc. to EN10204:2004-2.2(QIC) including the content of option /P2
/P6 Material certificate acc. to EN 10204: 2004 -3.1
/P8 Pressure test report measuring system
/H1 Oil and fat free for wetted surface acc. to ASTM G93-03 level C
/WP WPS acc. DIN EN ISO 15609-1 (Welding Procedure Specification) only for butt weld between process connection and
WPQR acc. DIN EN ISO 15614-1 (Welding Procedure Qualification Record) flow divider
WQC acc. DIN EN 287-1 (Welder Qualification Certificate)
/L2 Calibration certificate level 2: Declaration and the Calibration Equipment List are
issued
/L3 Calibration certificate level 3: Declaration and the Primary Standard List are issued
/L4 Calibration certificate level 4: Declaration and the Yokogawa Measuring
Instruments Control System are issued
Sanitary Type /SF1 Surface roughness Ra = 0.8 µm only RCCT34 to 39; not with RCCT34 /K4 and RCCT34 /
K6; only process connections S2, S4, S8;
/SF2 As /SF1 + Test report roughness of wetted parts as /SF1
/SA As /SF2 + 3A- declaration of conformity and 3A- mark as /SF1, but not with process connection S2
/SE As /SF2 + EHEDG certificate as /SF1, but not with process connection S2
Customer Presetting /PS Presetting sheet with customer data has to be issued with the order
Housing Pressure Test /J1 Rupture pressure proof test and certificate (see page 4) not for RCCT39/XR
X-Ray Examination /RT X-ray examination of flange welding RCCT34 with /K4 or /K6 only one-sided; not with HC
PMI Certificate /PM6 PMI Test on wetted parts
Dye Penetrant Test /PT Dye Penetrant Test acc. DIN EN ISO 3452-1 at the weldings of the process
connection, with certificate

Epoxy Coating /X1 Epoxy coating of converter housing


Concentration Measurement 3) /CST Standard concentration measurement not with /GA, /C
/C Advanced concentration measurement, details see table „Advanced Concentration not with /GA, /CST
Measurement Options“
Delivery to Japan /PJ For meters which are delivered to Japan, with SI-unit preset
Delivery to Korea /KC With KC-mark for Korea
Eurasian Conformity /VE With EAC- mark combination with Hazardous Area Approval only with /GF
Delivery to China /CN For delivery to China combination with Hazardous Area Approval only with /NF
Version IR /V2 Version 2 always with RCCT39/IR
Cable glands /AD2 2 pcs ANSI 1/2“ NPT / G1/2 adapter only with cable conduit “A”
Instruction Manuals /IE Quantity of instruction manuals in English  = 1 to 3 selectable 4)
/ID Quantity of instruction manuals in German  = 1 to 3 selectable 4)
/IF Quantity of instruction manuals in French  = 1 to 3 selectable 4)
Quick Delivery /QD Delivery within 24 hours from factory Delivery scope please refer to “Quick Delivery Possibili-
ties” or contact Yokogawa sales office.
Special order /Z Special design, must be specified separately.
If /Z is selected, several Suffix of Model-Suffix Code can be changed to Z.
1)
For gas application /GA RCCT39/XR can also be ordered without /HP.
2)
Calibration order sheet must be delivered with the order. This is available on the Flow Center Page at Coriolis/RCC3/Technical Information.
Volume calibration: Mass flow calibration converted by density to volume flow.
3)
For detailed information please see TI 01R04B04-04E-E. Concentration measurement is recommended with option /K6.
4)
If no instruction manual is selected, only a DVD with instruction manuals is shipped with the instrument. More than 3 manuals of one language on request available.

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
25

Remote Detector RCCS3, Model and Suffix Code


Model Suffix Code Description Restrictions

RCCS30LR Nominal Value : 0.021 t/h = 0.35 kg/min (0.77 lbs/min)


RCCS30 Nominal Value : 0.045 t/h = 0.75 kg/min (1.65 lbs/min)
RCCS31 Nominal Value : 0.17 t/h = 2.8 kg/min (6.17 lbs/min)
RCCS32 Nominal Value : 0.37 t/h = 6.2 kg/min (13.66 lbs/min)
RCCS33 Nominal Value : 0.95 t/h = 16 kg/min (35.27 lbs/min)
RCCS34 Nominal Value : 3 t/h = 50 kg/min (110,2 lbs/min)
RCCS36 Nominal Value : 10 t/h = 170 kg/min (374.78 lbs/min)
RCCS38 Nominal Value : 32 t/h = 533 kg/min (1175 lbs/min)
RCCS39 Nominal Value : 100 t/h = 1670 kg/min (3681.72 lbs/min)
RCCS39/IR Nominal Value : 250 t/h = 4170 kg/min (9193.27 lbs/min) only with /V2
RCCS39/XR Nominal Value : 500 t/h = 8340 kg/min (18386.55 lbs/min) select affiliated RCCF31 or
RCCR31 with /HP
Cable conduit connection -M M20 x 1.5, female thread with cable glands not with /FS1
-A ANSI ½” NPT, female thread only with cable gland for detector connection
Process connection size 1)
41 ¼” see table 11
01 DN 15, ½” see table 11
23 ¾” see table 11
02 DN 25, 1” see table 11
04 DN 40, 1½” see table 11
05 DN 50, 2” see table 11
06 DN 65, 2½” see table 11
08 DN 80, 3” see table 11
10 DN 100, 4” see table 11
12 DN 125, 5” see table 11
15 DN 150, 6” see table 11
20 DN 200, 8” see table 11
Process connection rating and style 1) A1 ASME flange class 150, process connection dim. + facing acc. ASME B16.5 see table 11
A2 ASME flange class 300, process connection dim. + facing acc. ASME B16.5 see table 11
A3 ASME flange class 600, process connection dim. + facing acc. ASME B16.5 see table 11
A4 ASME flange class 900, process connection dim. + facing acc. ASME B16.5 see table 11
A5 ASME flange class 1500, process connection dim. + facing acc. ASME B16.5 see table 11
D2 EN flange PN 16, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D4 EN flange PN 40, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D5 EN flange PN 63, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D6 EN flange PN 100, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D7 EN flange PN 160, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
J1 JIS flange 10K, JIS B 2220 see table 11
J2 JIS flange 20K, JIS B 2220 see table 11
S2 Thread acc. DIN 11851 see table 11
S4 Clamp, process connection dimensions acc. DIN 32676 see table 11
S8 Clamp, process connection dim. acc. Tri-Clover® (Tri-Clamp®) and ½” Mini Clamp see table 11
G9 G female thread see table 11
T9 NPT female thread see table 11
Material of wetted parts 1)
SH 316L (1.4404) and Hastelloy C-22 (2.4602) for tube only RCCS30LR to 33
SL Stainless steel 316L (1.4404) only RCCS34 to 39/XR
HC Hastelloy C-22 (2.4602) only RCCS34 to 39/IR
1)
see selection table „Process connection and materials“ (table 11)

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
26

Remote Detector RCCS3, Option Code


Options Option Description Restrictions
code
Hazardous Area Approvals 1) /KS1 ATEX intrinsically safe approval
/FS1 FM intrinsically safe approval for USA + Canada only with cable conduit “A”
/ES1 IECEx intrinsically safe approval
/US1 INMETRO intrinsically safe approval for Brazil
/NS1 NEPSI intrinsically safe approval for China only with /CN
/GS1 EAC intrinsically safe approval only with /VE
Custody Transfer Measurement /Q01 European MID Approval (fluids other than water) Refer to GS 01R04B07-00E
/Q20 NTEP approval 12-080A2, Accuracy class 0.3 acc. NIST Handbook 44 Refer to GS 01R04B07-00E
/Q21 NTEP approval 12-080A2, Accuracy class 0,3 acc. NIST Handbook 44, Heated Refer to GS 01R04B07-00E
Products
GOST 1) /QR2 Primary Calibration and Test Confirmation valid in Kazakhstan see page 11; not with /Q20, /Q21; only with /VE
/QR3 Primary Calibration and Test Confirmation valid in Uzbekistan see page 11; not with /Q20, /Q21
Marine Approval /MC2 Marine approval acc. DNV piping class 2 see GS 01R04B10-00E-E
/MC3 Marine approval acc. DNV piping class 3 see GS 01R04B10-00E-E
Dual Seal Approval /DS Dual Seal approval (conform with ANSI/ISA-12.27.01) only RCCS34 to 39/XR; not with process connection A5
only RCCS34 to 39/XR, preferable with /GA, not with
/RD Rupture disk, rupture pressure 20 bar, nominal diameter 8 mm /T1, /T2, /T3, mandatory if /DS + /GA is selected
Tag Number /BG With customer specified tag number on name plate max. 16 digits
Flange Facing /DN Flange with safety grooves acc. to EN 1092-1 form D only for D2 to D6; not HC, for RCCS30LR only for 01D4
or 01D6
/EN Flange with spigot acc. to EN 1092-1 form E only for D2 to D6; not HC, for RCCS30LR only for 01D4
or 01D6
/FN Flange with recess acc. to EN 1092-1 form F only for D2 to D6; not HC, for RCCS30LR only for 01D4
or 01D6
/RJ Ring Type Joint Flanges only for A3, A4, A5; not HC, for RCCS30LR only for
01A3 or 01A5
Gas Measurement /GA Gas measurement, special factory adjustments and settings select affiliated RCCF31 or RCCR31 with /GA;
to be conform with ANSI/ISA-12.27.01 select /RD; not
with /Q20
Low temperature version /LT -200°C ≤ Tmedium ≤ 150°C (-328°F to 302°F) for RCCS34 to 39/XR; not with /MT, /HT, /Q01, /T1, /T2, /T3 ;
in combination with Hazardous Area Approval only with /S2
Extended temperature range /MT -70°C ≤ Tmedium ≤ 230°C (-94°F to 446°F) for RCCS34 to 39/XR; always with /S2 or /T1, /T2, /T3 ;
-50°C ≤ Tmedium ≤ 260°C (-58°F to 500°F) for RCCS30LR to 33, always with /S2 or /T1, /T2, /T3
High temperature version /HT Tmedium up to 350°C (662°F) RCCS34 to 39/XR; only with/T1, /T2, /T3 or /S2 (cus-
tomer insulation required); RCCS39/XR only with /S2;
Special Calibration /K2 3) Custom 5 pts mass-/volume-flow calibration using water with factory certificate only in combination with converter RCCF31/RCCR31
(traceable to German national standards)
/K4 Density adjustment + thermal treatment; (accuracy: 0.001 g/cm³) only RCCS31 to 39; not with /GA
/K5 3) Custom 10 pts mass-/volume-flow calibration using water with DAkkS certificate only in combination with converter RCCF31/RCCR31
(according EN-17025:2005)
/K6 Density calibration with 3 different fluids incl. individual temperature compensation only RCCS32 to 39; not with /GA; not with /LT;
with certificate (accuracy: 0.0005 g/cm³) not with /MT; only available if converter is also ordered
Certificates /P2 Certificate of compliance with the order acc. to EN 10204:2004 -2.1
/P3 Test Report acc. to EN10204:2004-2.2(QIC) including the content of option /P2
/P6 Material certificate acc to EN 10204: 2004 -3.1
/P8 Pressure test report measuring system
/H1 Oil and fat free for wetted surface acc. to ASTM G93-03 level C
/WP WPS acc. DIN EN ISO 15609-1 (Welding Procedure Specification) only for butt weld between process connection and
WPQR acc. DIN EN ISO 15614-1 (Welding Procedure Qualification Record) flow divider
WQC acc. DIN EN 287-1 (Welder Qualification Certificate)
/L2 Calibration certificate level 2: Declaration and the Calibration Equipment List are
issued
/L3 Calibration certificate level 3: Declaration and the Primary Standard List are issued
/L4 Calibration certificate level 4: Declaration and the Yokogawa Measuring
Instruments Control System are issued
Sanitary Type /SF1 Surface roughness Ra = 0.8 µm only RCCS34 to 39; not with RCCS34 /K4, RCCS34 / K6
and RCCS34 /LT ; only process connections S2, S4, S8;
/SF2 As /SF1 + Test report roughness of wetted parts as /SF1
/SA As /SF2 + 3A- declaration of conformity and 3A- mark as /SF1, but not with process connection S2
/SE As /SF2 + EHEDG- certificate as /SF1, but not with process connection S2
Mounting set /PD 2 inch pipe mounting set, recommended for RCCS30LR and RCCS30 only RCCS30LR to 33; not with /T
Housing Pressure Test /J1 Rupture pressure proof test and certificate (see page 4) not for RCCS30LR to 33 + RCCS39/XR
Customer insulation / Heating /S2 Terminal box on extension for high or low process temperature not with /T1, /T2, /T3
Factory Insulating / Heating /T1 Insulation not for RCCS39/XR
/T2 Insulation + Heat carrier heating not for RCCS39/XR
/T3 Insulation + Heat carrier heating with ventilation (purge) not for RCCS39/XR
X-Ray Examination /RT X-ray examination of flange welding RCCS30LR to 33 and RCCS34 with /K4, /K6 or /LT
only one-sided; not with HC
PMI Certificate /PM4 PMI Test on wetted parts RCCS30LR to 33 4) not RCCS34 to 39/XR
/PM6 PMI Test on wetted parts RCCS34 to 39/XR not RCCS30LR to 33
Dye Penetrant Test /PT Dye Penetrant Test acc. DIN EN ISO 3452-1 at the weldings of the process
connection, with certificate
Stainless steel cable gland /BS Cable gland stainless steel
Version IR /V2 Version 2 always with RCCS39/IR
Delivery to Japan /PJ For meters which are delivered to Japan
Delivery to Korea /KC With KC-mark for Korea
Eurasian Conformity /VE With EAC- mark combination with Hazardous Area Approval only with /GS1
Delivery to China /CN For meters which are delivered to China combination with Hazardous Area Approval only with /NS1
Quick Delivery /QD Delivery within 24 hours from factory Delivery scope please refer to “Quick Delivery Possibili-
ties” or contact Yokogawa sales office.
Special order /Z Special design, must be specified separately.
If /Z is selected, several Suffix of Model-Suffix Code can be changed to Z.
1)
Select affiliated converter RCCF31/RCCR31 with the same approval type (e.g. ATEX).
2)
Select affiliated converter RCCF31 with the same approval type (/Q01, /Q20 or /Q21).
3)
Calibration order sheet must be delivered with the order. This is available on the Flow Center Page at Coriolis/RCC3/Technical Information.
Volume calibration: Mass flow calibration converted by density to volume flow.
4)
Measuring tube PMI test is performed per delivery batch.

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
27

Remote field-mount Converter RCCF31, Model, Suffix and Option Code


Model Suffix Code Option Description Restrictions
Code
RCCF31 Remote field-mount converter to be connected to RCCS3;
when ordered without detector combination option /NC must be selected
Power supply -A 90 - 264 V AC
-D 24 V DC
Indicator direction H2 With indicator
N0 Without indicator
Cable conduit connection M M20 x 1.5, female thread with cable glands not with /FF1, /FF5
A ANSI ½” NPT, female thread, only cable gland for detector connection

Hazardous Area Approvals 1) /KF1 ATEX Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC
/KF5 ATEX Flame proof converter + Intrinsically safe detector output + Intrinsically safe with /HP not for gas group IIC
outputs 2 passive analog output + 1 passive pulse output
/FF1 FM approval for USA+Canada, Flame proof converter + Intrinsically safe detector only with cable conduit “A”; with /HP not for groups A
output + 2 active analog outputs + 2 passive pulse outputs + 1 status input and B
/FF5 FM approval for USA+Canada, Flame proof converter + Intrinsically safe detector only with cable conduit “A”; with /HP not for groups A
output + 2 passive analog outputs + 1 passive pulse output and B
/EF1 IECEx Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC
/EF5 IECEx Flame proof converter + Intrinsically safe detector output + Intrinsically with /HP not for gas group IIC
safe outputs 2 passive analog output + 1 passive pulse output
/UF1 INMETRO Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC
/UF5 INMETRO Flame proof converter + Intrinsically safe detector output + Intrinsically with /HP not for gas group IIC
safe outputs 2 passive analog output + 1 passive pulse output
/NF1 NEPSI Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC; only with /CN
/NF5 NEPSI Flame proof converter + Intrinsically safe detector output + Intrinsically with /HP not for gas group IIC; only with /CN
safe outputs 2 passive analog output + 1 passive pulse output
/GF1 EAC Flame proof converter + Intrinsically safe detector output with /HP for gas group IIB; only with /VE
/GF5 EAC Flame proof converter + Intrinsically safe detector output + Intrinsically safe with /HP for gas group IIB; only with /VE
outputs 2 passive analog output + 1 passive pulse output
Custody Transfer Measurement /Q01 European MID Approval (fluids other than water) Refer to GS 01R04B07-00E
/Q20 NTEP approval 12-080A2, Accuracy class 0.3 acc. NIST Handbook 44 Refer to GS 01R04B07-00E
/Q21 NTEP approval 12-080A2, Accuracy class 0,3 acc. NIST Handbook 44, Heated Refer to GS 01R04B07-00E
Products
GOST 1) /QR2 Primary Calibration and Test Confirmation valid in Kazakhstan see page 11; not with /Q20, /Q21; only with /VE
/QR3 Primary Calibration and Test Confirmation valid in Uzbekistan see page 11; not with /Q20, /Q21
Marine Approval /MC2 Marine approval acc. DNV piping class 2 see GS 01R04B10-00E-E
/MC3 Marine approval acc. DNV piping class 3 see GS 01R04B10-00E-E
High Driving Power /HP High Driving Power, recommended for combination with RCCS36 to 39, strongly mandatory for combination with RCCS39/XR 5)
recommended for combination with RCCS39/IR, Please see „ Hazardous Area
Specifications“
Communication /FB Digital communication (FOUNDATION™ Fieldbus protocol) refer to GS 01R04B05-00E

/MB2 Modbus communication protocol for all parameters and values not with /KF5, /EF5, /UF5, /GF5, /NF5, /FF5, /Q01,
/Q20, /Q21, /AP, /NM, /MB3
/MB3 Modbus communication protocol for all parameters and values incl. HART not with /KF5, /EF5, /FF5, /UF5, /NF5, /GF5, /Q01, /MB2
Communication, analogue output, pulse/status output, status input
Active Pulse Output /AP One active pulse output not with /KF5, /EF5, /FF5, /UF5, /NF5, /GF5, /NM
NAMUR Switch /NM One pulse output acc. to EN 60947-5-6 (NAMUR) not with /AP, not with /KF5, /EF5, /UF5, /NF5, /GF5
Tag Number /BG With customer specified tag number on name plate max. 16 characters; the last 8 significant characters are
used for HART® Tag;
characters acc. Codepage 850 (Multi-lingual)
HART® Tag Number (Software Tag) /BT1 With customer specified tag number for HART® communication in converter 22 characters for long tag;
characters acc. Codepage 850 (Multi-lingual)
Gas Measurement /GA Gas measurement, special factory adjustments and settings select affiliated RCCS3  with /GA; not with /Q20, /Q21
Combination with RCCS39/XR /XR Special converter adjustment with RCCS39/XR mandatory for combination with RCCS39/XR
No Combination /NC No combination with detector
Customer Presetting /PS Presetting sheet with customer data has to be issued with the order
Epoxy Coating /X1 Epoxy coating of converter housing
Delivery to Japan /PJ For meters which are delivered to Japan with SI-unit preset
Delivery to Korea /KC With KC-mark for Korea
Eurasian Conformity /VE With EAC- mark combination with Hazardous Area Approval only with /GF
Delivery to China /CN For meters which are delivered to China combination with Hazardous Area Approval only with /NF
Cable glands /AD2 2 pcs ANSI 1/2“ NPT / G1/2 adapter only with cable conduit “A”
Concentration Measurement 3) /CST Standard concentration measurement not with /GA, /C
/C Advanced concentration measurement, details see table „Advanced Concentration not with /GA, /CST
Measurement Options“
Certificates /P2 Certificate of compliance with the order acc. to EN 10204:2004 -2.1
/P3 Test Report acc. to EN10204:2004-2.2(QIC) including the content of option /P2
Instruction Manuals /IE Quantity of instruction manuals in English  = 1 to 3 selectable 4)
/ID Quantity of instruction manuals in German  = 1 to 3 selectable 4)
/IF Quantity of instruction manuals in French  = 1 to 3 selectable 4)
Quick Delivery /QD Delivery within 24 hours from factory Delivery scope please refer to “Quick Delivery Possibili-
ties” or contact Yokogawa sales office.
Special order /Z Special design, must be specified separately.
If /Z is selected, several Suffix of Model-Suffix Code can be changed to Z.
1)
Select affiliated RCCS3 with the same approval type (e.g. /KF with /KS1).
2)
Select affiliated RCCS3 with the same approval type (/Q01, /Q20 or /Q21).
3)
For detailed information please see TI 01R04B04-04E-E. Option /K6 of RCCS3 is recommended with concentration measurement.
4)
If no instruction manual is selected, only a DVD with instruction manuals is shipped with the instrument. More than 3 manuals of one language on request available.
5)
RCCF31 in combination with RCCS39/XR with gas application /GA can also be ordered without /HP.

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
28

Remote rack-mount Converter RCCR31, Model, Suffix and Option Code


Model Suffix Code Option Description Restrictions
Code
RCCR31 Remote converter for 19” rack mounting to be connected to RCCS3
Power supply -A 90 - 264 V AC
-D 24 V DC

Hazardous Area Approvals 1) /KS1 ATEX associated apparatus for intrinsically safe detector connection for gas group IIC with /HP not for gas group IIC
/FS1 FM associated apparatus for intrinsically safe detector output with /HP not for gas groups A and B
/ES1 IECEx associated apparatus for intrinsically safe detector output for gas group IIC with /HP not for gas group IIC
/US1 INMETRO associated apparatus for intrinsically safe detector output for gas group IIC with /HP not for gas group IIC
/NS1 NEPSI associated apparatus for intrinsically safe detector output for gas group IIC with /HP not for gas group IIC; only with /CN
/GS1 EAC associated apparatus for intrinsically safe detector connection for gas group IIC with /HP not for gas group IIC; only with /VE
GOST 1) /QR2 Primary Calibration and Test Confirmation valid in Kazakhstan see page 11; only with /VE
Marine Approval /MC2 Marine approval acc. DNV piping class 2 see GS 01R04B10-00E-E
/MC3 Marine approval acc. DNV piping class 3 see GS 01R04B10-00E-E
High Driving Power /HP High Driving Power, recommended for combination with RCCS36 to 39, strongly
recommended for combination with RCCS39/IR, Please see „ Hazardous Area mandatory for combination with RCCS39/XR
Specifications“
Active Pulse Output /AP One active pulse output not with /NM
NAMUR Switch /NM One pulse output acc. EN 60947-5-6 (NAMUR) not with /AP
Tag Number /BG With customer specified tag number on name plate max. 16 characters; the last 8 significant characters are
used for HART® Tag;
characters acc. Codepage 850 (Multi-lingual)
HART® Tag Number (Software Tag) /BT1 With customer specified tag number for HART® communication in converter 22 characters for long tag;
characters acc. Codepage 850 (Multi-lingual)
Communication /MB2 Modbus communication protocol for all parameters and values not with /AP, /NM, /MB3
/MB3 Modbus communication protocol for all parameters and values incl. HART not with /MB2
Communication, analogue output, pulse/status output, status input
Gas Measurement /GA Gas measurement, special factory adjustments and settings select affiliated RCCS3 with /GA
Combination with RCCS39/XR /XR Special converter adjustment with RCCS39/XR mandatory for combination with RCCS39/XR
No Combination /NC No combination with detector
Customer Presetting /PS Presetting sheet with customer data has to be issued with the order
Concentration Measurement 2) /CST Standard concentration measurement not with /GA, /C
/C Advanced concentration measurement, details see table „Advanced Concentration not with /GA, /CST
Measurement Options“
Subrack /SR2 Subrack for 2 converter RCCR31 with mounting
/SR4 Subrack for 4 converter RCCR31 with mounting
Delivery to Korea /KC With KC-mark for Korea
Eurasian Conformity /VE With EAC- mark combination with Hazardous Area Approval only with /GS1
Delivery to China /CN For meters which are delivered to China combination with Hazardous Area Approval only with /NS1
Certificates /P2 Certificate of compliance with the order acc. to EN 10204:2004 -2.1
/P3 Test Report acc. to EN10204:2004-2.2(QIC) including the content of option /P2
Instruction Manuals /IE Quantity of instruction manuals in English  = 1 to 3 selectable 3)
/ID Quantity of instruction manuals in German  = 1 to 3 selectable 3)
/IF Quantity of instruction manuals in French  = 1 to 3 selectable 3)
Special order /Z Special design, must be specified separately.
If /Z is selected, several Suffix of Model-Suffix Code can be changed to Z.
1)
Select affiliated RCCS3 with the same approval type (e.g. /KS1 with /KS1).
2)
For detailed information please see TI 01R04B04-04E-E. Option /K6 of RCCS3 is recommended with concentration measurement.
3)
If no instruction manual is selected, only a DVD with instruction manuals is shipped with the instrument. More than 3 manuals of one language on request available.
4)
RCCR31 in combination with RCCS39/XR with gas application /GA can also be ordered without /HP.

Remote Cable RCCY03, Model, Suffix and Option Code


Model Suffix Code Option Description Restrictions
Code
RCCY031 Length in “meter” max. ambient temperature 70°C (158°F); with /FF or /FS1: 50°C (122°F)
RCCY032 Length in “feet” max. ambient temperature 70°C (158°F); with /FF or /FS1: 50°C (122°F)
RCCY033 Length in “meter”
max. ambient temperature 105°C (221°F); with /FF or /FS1: 85°C (185°F)
RCCY034 Length in “feet”
max. ambient temperature 105°C (221°F); with /FF or /FS1: 85°C (185°F)
Cable ends -0 No termination, with one termination kit
-1 Terminated
Cable length L Enter the length max. 300m / 999ft,
the following lengths can be ordered (e.g. 3m = L003):
RCCY031-0: 3m, 5m, 10m, 15m, 30m, 50m, 100m, 150m, 200m, 250m, 300m
RCCY031-1: 3m, 5m, 10m, 15m, 30m, 50m
RCCY032-0: 10ft, 15ft, 30ft, 50ft, 100ft, 150ft, 300ft, 500ft, 1000ft
RCCY032-1: 10ft, 15ft, 30ft, 50ft, 100ft, 150ft
RCCY033-0: 3m, 5m, 10m, 15m, 30m, 50m, 100m, 150m, 300m
RCCY033-1: 3m, 5m, 10m, 15m, 30m, 50m
RCCY034-0: 10ft, 15ft, 30ft, 50ft, 100ft, 150ft, 300ft, 500ft, 1000ft
RCCY034-1: 10ft, 15ft, 30ft, 50ft, 100ft, 150ft
Options:
Hazardous area installation /KS1 Blue cable for Ex-i indication
/NS1 Blue cable for Ex-i indication (China)
Termination kits /TK Quantity of additional termination kits  = 01 to 99
Quick delivery /QD Delivery within 24 hours from factory only L003, L005, L010

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
29

Advanced Concentration Measurement Options (others on request), recommended with Option /K6
Option Display Components Concentration Temp. range Source of concentration- / density table
range
/C00 --- --- --- --- Advanced concentration measurement function.
There are no pre-defined concentration setups (coefficients are set to zero).
The concentration coefficients will be set by customer making use of either
FieldMate or the stand- alone concentration tool and manual entry by HHT.
For more information, please contact your regional Yokogawa office.
/C01 °Brix Sugar / Water 0 - 85 °Brix 0 - 80°C PTB- Messages 100 5/90:
(32 - 176°F) „The density of watery Saccarose solutions after the introduction of the
international temperature scale of 1990 (ITS1990)“ Table 5
/C02 WT% NaOH / Water 2 - 50 WT% 0 - 100°C D´Ans-Lax, Handbook for chemists and physicists Vol.1, 3rd edition, 1967
(32 - 212°F)
/C03 WT% KOH / Water 0 - 60 WT% 54 - 100°C D´Ans-Lax, Handbook for chemists and physicists Vol.1, 3rd edition, 1967
(129 - 212°F)
/C04 WT% NH4NO3 / Water 1 - 50 WT% 0 - 80°C Data table on request
(32 - 176°F)
/C05 WT% NH4NO3 / Water 20 - 70 WT% 20 - 100°C Data table on request
(68 - 212°F)
/C06 1) WT% HCl / Water 22 - 34 WT% 20 - 40°C D´Ans-Lax, Handbook for chemists and physicists Vol.1, 3rd edition, 1967
(68 - 104°F)
/C07 WT% HNO3 / Water 50 - 67 WT% 10 - 60°C Data table on request
(50 - 140°F)
/C09 WT% H2O2 / Water 30 - 75 WT% 4 - 44°C Data table on request
(39.2 - 111.2°F)
/C10 WT% Ethylene Glycol / 10 - 50 WT% -20 - 40°C Data table on request
Water (-4 - 104°F)
/C11 WT% Amylum = starch / Water 33 - 43 WT% 35 - 45°C Data table on request
(95 - 113°F)
/C12 WT% Methanol / Water 35 - 60 WT% 0 - 40°C Data table on request
(32 - 104°F)
/C20 VOL% Alcohol / Water 55 - 100 VOL% 10 - 40°C Data table on request
(50 - 104°F)
/C21 °Brix Sugar / Water 40 - 80 °Brix 75 - 100°C Data table on request
(167 - 212°F)
/C30 WT% Alcohol / Water 66 - 100 WT% 15 - 40°C Standard Copersucar 1967
(59 - 104°F)
/C37 WT% Alcohol / Water 66 - 100 WT% 10 - 40°C Brazilian Standard ABNT
(50 - 104°F)
/C38 VOL% Alcohol / Water 73 - 100 VOL% 10 - 40°C Brazilian Standard ABNT
(50 - 104°F)
1)
only with material HC

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
30

Table 11 : Selection Table Process Connection and Materials, Installation Length in mm


RCCS RCCS RCCS34 RCCS36 RCCS38 RCCS39 RCCS39/IR RCCS39/XR
30LR 30-33 RCCT34 RCCT36 RCCT38 RCCT39 RCCT39/IR RCCT39/XR
SH SH SL HC SL HC SL HC SL HC SL HC SL HC
01A1 ½“-150 240 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A2 ½“-300 240 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A3 ½“-600 250 250 380 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A4 ½“-900 ---- ---- 400 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A5 ½“-900/1500 270 270 400 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02A1 1“-150 ---- 240 370 390 500 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A2 1“-300 ---- 240 370 390 500 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A3 1“-600 ---- 260 390 390 520 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A4 1“-900 ---- ---- 450 400 540 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A5 1“-900/1500 ---- 320 450 400 540 ---- ---- ---- ---- ---- ---- ---- ---- ----
04A1 1½“-150 ---- 250 380 390 500 520 600 ---- ---- ---- ---- ---- ---- ----
04A2 1½“-300 ---- 250 380 390 510 520 600 ---- ---- ---- ---- ---- ---- ----
04A3 1½“-600 ---- 270 400 400 530 530 620 ---- ---- ---- ---- ---- ---- ----
04A4 1½“-900 ---- ---- 470 ---- 600 ---- 640 ---- ---- ---- ---- ---- ---- ----
04A5 1½“-900/1500 ---- 340 470 ---- 600 ---- ---- ---- ---- ---- ---- ---- ---- ----
Flanges according to ASME B16.5

05A1 2“-150 ---- ---- ---- 390 510 520 600 620 ---- ---- ---- ---- ---- ----
05A2 2“-300 ---- ---- ---- 390 510 520 600 620 ---- ---- ---- ---- ---- ----
05A3 2“-600 ---- ---- ---- 400 540 540 630 630 ---- ---- ---- ---- ---- ----
05A4 2“-900 ---- ---- ---- ---- 660 ---- 720 ---- ---- ---- ---- ---- ---- ----
05A5 2“-900/1500 ---- ---- ---- ---- 660 ---- ---- ---- ---- ---- ---- ---- ---- ----
06A1 2½“-150 ---- ---- ---- ---- ---- ---- 610 620 ---- ---- ---- ---- ---- ----
06A2 2½“-300 ---- ---- ---- ---- ---- ---- 610 620 ---- ---- ---- ---- ---- ----
06A3 2½“-600 ---- ---- ---- ---- ---- ---- 640 640 ---- ---- ---- ---- ---- ----
06A4 2½“-900 ---- ---- ---- ---- ---- ---- 760 ---- ---- ---- ---- ---- ---- ----
08A1 3“-150 ---- ---- ---- ---- ---- ---- 610 620 1000 1020 ---- ---- ---- ----
08A2 3“-300 ---- ---- ---- ---- ---- ---- 620 620 1000 1020 ---- ---- ---- ----
08A3 3“-600 ---- ---- ---- ---- ---- ---- 640 640 1000 1025 ---- ---- ---- ----
08A4 3“-900 ---- ---- ---- ---- ---- ---- 760 ---- ---- ---- ---- ---- ---- ----
10A1 4“-150 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 ---- ---- ----
10A2 4“-300 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 ---- ---- ----
10A3 4“-600 ---- ---- ---- ---- ---- ---- ---- ---- 1030 1030 1100 ---- ---- ----
12A1 5“-150 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
12A2 5“-300 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
12A3 5“-600 ---- ---- ---- ---- ---- ---- ---- ---- 1040 1040 1160 1110 ---- ----
15A1 6“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 1350 ----
15A2 6“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 1350 ----
15A3 6“-600 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1200 1120 1390 ----
20A1 8“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1140 1100 1350 ----
20A2 8“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1140 1100 1350 ----
20A3 8“-600 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1440 ----
01D4 DN 15 PN 40 240 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01D6 DN 15 PN 100 250 250 380 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02D4 DN 25 PN 40 ---- 240 370 390 500 520 ---- ---- ---- ---- ---- ---- ---- ----
02D6 DN 25 PN 100 ---- 260 390 ---- 520 ---- ---- ---- ---- ---- ---- ---- ---- ----
04D4 DN 40 PN 40 ---- 240 370 390 500 520 600 ---- ---- ---- ---- ---- ---- ----
04D6 DN 40 PN 100 ---- 320 450 ---- 560 ---- 620 ---- ---- ---- ---- ---- ---- ----
05D4 DN 50 PN 40 ---- ---- ---- ---- 500 520 600 620 ---- ---- ---- ---- ---- ----
05D5 DN 50 PN 63 ---- ---- ---- ---- 520 ---- 620 ---- ---- ---- ---- ---- ---- ----
05D6 DN 50 PN 100 ---- ---- ---- ---- 590 ---- 660 ---- ---- ---- ---- ---- ---- ----
05D7 DN 50 PN 160 ---- ---- ---- ---- 590 ---- 660 ---- ---- ---- ---- ---- ---- ----
Flanges according to EN 1092-1

08D4 DN 80 PN 40 ---- ---- ---- ---- ---- ---- 610 620 1000 1020 ---- ---- ---- ----
08D5 DN 80 PN 63 ---- ---- ---- ---- ---- ---- 620 ---- 1000 ---- ---- ---- ---- ----
08D6 DN 80 PN 100 ---- ---- ---- ---- ---- ---- 730 ---- 1000 ---- ---- ---- ---- ----
10D2 DN 100 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- 1100 ---- ---- ----
10D4 DN 100 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 ---- ---- ----
10D5 DN 100 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- 1100 ---- ---- ----
10D6 DN 100 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- 1050 ---- 1100 ---- ---- ----
12D2 DN 125 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
12D4 DN 125 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
12D5 DN 125 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- 1100 1100 ---- ----
12D6 DN 125 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- 1100 ---- 1140 1100 ---- ----
12D7 DN 125 PN 160 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 ---- ----
15D2 DN 150 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 1350 ----
15D4 DN 150 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 1350 ----
15D5 DN 150 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1140 ---- 1350 ----
15D6 DN 150 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1180 ---- ---- ----
20D2 DN 200 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1350 ----
20D4 DN 200 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1350 ----
20D5 DN 200 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1350 ----

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
31

Table 11 : Selection Table Process Connection and Materials, Installation Length in mm (continued)

RCCS RCCS RCCS34 RCCS36 RCCS38 RCCS39 RCCS39 RCCS39


30LR 30-33 RCCT34 RCCT36 RCCT38 RCCT39 RCCT39/IR RCCT39/XR
SH SH SL HC SL HC SL HC SL HC SL HC SL HC
01J1 DN 15 10K 240 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01J2 DN 15 20K 240 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02J1 DN 25 10K ---- 240 370 390 500 ---- ---- ---- ---- ---- ---- ---- ---- ----
Flanges according to JIS B 2220

02J2 DN 25 20K ---- 240 370 390 500 ---- ---- ---- ---- ---- ---- ---- ---- ----
04J1 DN 40 10K ---- 240 370 390 500 520 600 ---- ---- ---- ---- ---- ---- ----
04J2 DN 40 20K ---- 240 370 390 500 520 600 ---- ---- ---- ---- ---- ---- ----
05J1 DN 50 10K ---- ---- ---- ---- 500 520 600 620 ---- ---- ---- ---- ---- ----
05J2 DN 50 20K ---- ---- ---- ---- 500 520 600 620 ---- ---- ---- ---- ---- ----
08J1 DN 80 10K ---- ---- ---- ---- ---- ---- 600 620 1000 1020 ---- ---- ---- ----
08J2 DN 80 20K ---- ---- ---- ---- ---- ---- 610 620 1000 1020 ---- ---- ---- ----
10J1 DN 100 10K ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 ---- ---- ----
10J2 DN 100 20K ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 ---- ---- ----
12J1 DN 125 10K ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
12J2 DN 125 20K ---- ---- ---- ---- ---- ---- ---- ---- 1000 1020 1100 1100 ---- ----
15J1 DN 150 10K ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 ---- ----
15J2 DN 150 20K ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 1100 1100 ---- ----
01S4 DN 15 240 240 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02S4 DN 25 ---- 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Clamp DIN

04S4 DN 40 ---- 240 370 ---- 500 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S4 DN 50 ---- ---- ---- ---- 500 ---- 600 ---- ---- ---- ---- ---- ---- ----
06S4 DN 65 ---- ---- ---- ---- ---- ---- 600 ---- ---- ---- ---- ---- ---- ----
10S4 DN 100 ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- ---- ---- ---- ----
01S8 ½“ 240 240 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02S8 1“ ---- 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Tri-Clamp

04S8 1½“ ---- 240 370 ---- 500 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S8 2“ ---- ---- ---- ---- 500 ---- 600 ---- ---- ---- ---- ---- ---- ----
08S8 3“ ---- ---- ---- ---- ---- ---- 600 ---- ---- ---- ---- ---- ---- ----
10S8 4“ ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- ---- ---- ---- ----
02S2 DN 25 ---- 240 370 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
DIN11851

04S2 DN 40 ---- ---- ---- ---- 500 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S2 DN 50 ---- ---- ---- ---- ---- ---- 600 ---- ---- ---- ---- ---- ---- ----
10S2 DN 100 ---- ---- ---- ---- ---- ---- ---- ---- 1000 ---- ---- ---- ---- ----
41G9 G¼“ female 260 260 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01G9 G½“ female 260 260 390 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Thread

23G9 G¾“ female 260 260 390 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
41T9 NPT¼“ female 260 260 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01T9 NPT½“ female 260 260 390 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
23T9 NPT¾“ female 260 260 390 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
32

Table 11 : Selection Table Process Connection and Materials, Installation Length in inch
RCCS RCCS RCCS34 RCCS36 RCCS38 RCCS39 RCCS39/IR RCCS39/XR
30LR 30-33 RCCT34 RCCT36 RCCT38 RCCT39 RCCT39/IR RCCT39/XR
SH SH SL HC SL HC SL HC SL HC SL HC SL HC
01A1 ½“-150 9.45 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A2 ½“-300 9.45 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A3 ½“-600 9.84 9.84 14.96 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A4 ½“-900 ---- ---- 15.75 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A5 ½“-900/1500 10.63 10.63 15.75 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02A1 1“-150 ---- 9.45 14.57 15.35 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A2 1“-300 ---- 9.45 14.57 15.35 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A3 1“-600 ---- 10.24 15.35 15.35 20.47 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A4 1“-900 ---- ---- 17.72 15.75 21.26 ---- ---- ---- ---- ---- ---- ---- ---- ----
02A5 1“-900/1500 ---- 12.6 17.72 15.75 21.26 ---- ---- ---- ---- ---- ---- ---- ---- ----
04A1 1½“-150 ---- 9.84 14.96 15.35 19.69 20.47 23.62 ---- ---- ---- ---- ---- ---- ----
04A2 1½“-300 ---- 9.84 14.96 15.35 20.08 20.47 23.62 ---- ---- ---- ---- ---- ---- ----
04A3 1½“-600 ---- 10.63 15.75 15.75 20.87 20.87 24.21 ---- ---- ---- ---- ---- ---- ----
04A4 1½“-900 ---- ---- 18.5 ---- 23.62 ---- 25.2 ---- ---- ---- ---- ---- ---- ----
04A5 1½“-900/1500 ---- 13.39 18.5 ---- 23.62 ---- ---- ---- ---- ---- ---- ---- ---- ----
Flanges according to ASME B16.5

05A1 2“-150 ---- ---- ---- 15.35 20.08 20.47 23.62 24.21 ---- ---- ---- ---- ---- ----
05A2 2“-300 ---- ---- ---- 15.35 20.08 20.47 23.62 24.21 ---- ---- ---- ---- ---- ----
05A3 2“-600 ---- ---- ---- 15.75 21.26 21.26 24.8 24.8 ---- ---- ---- ---- ---- ----
05A4 2“-900 ---- ---- ---- ---- 25.98 ---- 28.35 ---- ---- ---- ---- ---- ---- ----
05A5 2“-900/1500 ---- ---- ---- ---- 25.98 ---- ---- ---- ---- ---- ---- ---- ---- ----
06A1 2½“-150 ---- ---- ---- ---- ---- ---- 24.02 24.21 ---- ---- ---- ---- ---- ----
06A2 2½“-300 ---- ---- ---- ---- ---- ---- 24.02 24.21 ---- ---- ---- ---- ---- ----
06A3 2½“-600 ---- ---- ---- ---- ---- ---- 25.2 25.2 ---- ---- ---- ---- ---- ----
06A4 2½“-900 ---- ---- ---- ---- ---- ---- 29.92 ---- ---- ---- ---- ---- ---- ----
08A1 3“-150 ---- ---- ---- ---- ---- ---- 24.02 24.21 39.37 40.16 ---- ---- ---- ----
08A2 3“-300 ---- ---- ---- ---- ---- ---- 24.21 24.21 39.37 40.16 ---- ---- ---- ----
08A3 3“-600 ---- ---- ---- ---- ---- ---- 25.2 25.2 39.37 40.35 ---- ---- ---- ----
08A4 3“-900 ---- ---- ---- ---- ---- ---- 29.92 ---- ---- ---- ---- ---- ---- ----
10A1 4“-150 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 ---- ---- ----
10A2 4“-300 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 ---- ---- ----
10A3 4“-600 ---- ---- ---- ---- ---- ---- ---- ---- 40.55 40.55 43.31 ---- ---- ----
12A1 5“-150 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
12A2 5“-300 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
12A3 5“-600 ---- ---- ---- ---- ---- ---- ---- ---- 40.94 40.94 45.67 43.31 ---- ----
15A1 6“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 53.15 ----
15A2 6“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 53.15 ----
15A3 6“-600 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 47.2 43.31 54.72 ----
20A1 8“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 44.9 43.31 53.15 ----
20A2 8“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 44.9 43.31 53.15 ----
20A3 8“-600 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 56.69 ----
01D4 DN 15 PN 40 9.45 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01D6 DN 15 PN 100 9.84 9.84 14.96 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02D4 DN 25 PN 40 ---- 9.45 14.57 15.35 19.69 20.47 ---- ---- ---- ---- ---- ---- ---- ----
02D6 DN 25 PN 100 ---- 10.24 15.35 ---- 20.47 ---- ---- ---- ---- ---- ---- ---- ---- ----
04D4 DN 40 PN 40 ---- 9.45 14.57 15.35 19.69 20.47 23.62 ---- ---- ---- ---- ---- ---- ----
04D6 DN 40 PN 100 ---- 12.6 17.72 ---- 22.05 ---- 24.41 ---- ---- ---- ---- ---- ---- ----
05D4 DN 50 PN 40 ---- ---- ---- ---- 19.69 20.47 23.62 24.41 ---- ---- ---- ---- ---- ----
05D5 DN 50 PN 63 ---- ---- ---- ---- 20.47 ---- 24.41 ---- ---- ---- ---- ---- ---- ----
05D6 DN 50 PN 100 ---- ---- ---- ---- 23.23 ---- 25.98 ---- ---- ---- ---- ---- ---- ----
05D7 DN 50 PN 160 ---- ---- ---- ---- 23.23 ---- 25.98 ---- ---- ---- ---- ---- ---- ----
Flanges according to EN 1092-1

08D4 DN 80 PN 40 ---- ---- ---- ---- ---- ---- 24.02 24.41 39.37 40.16 ---- ---- ---- ----
08D5 DN 80 PN 63 ---- ---- ---- ---- ---- ---- 24.41 ---- 39.37 ---- ---- ---- ---- ----
08D6 DN 80 PN 100 ---- ---- ---- ---- ---- ---- 28.74 ---- 39.37 ---- ---- ---- ---- ----
10D2 DN 100 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- 43.31 ---- ---- ----
10D4 DN 100 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 ---- ---- ----
10D5 DN 100 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- 43.31 ---- ---- ----
10D6 DN 100 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- 41.34 ---- 43.31 ---- ---- ----
12D2 DN 125 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
12D4 DN 125 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
12D5 DN 125 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- 43.31 43.31 ---- ----
12D6 DN 125 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- 43.31 ---- 44.88 43.31 ---- ----
12D7 DN 125 PN 160 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 ---- ----
15D2 DN 150 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 53.15 ----
15D4 DN 150 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 53.15 ----
15D5 DN 150 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 44.88 ---- 53.15 ----
15D6 DN 150 PN 100 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 46.46 ---- ---- ----
20D2 DN 200 PN 16 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 53.15 ----
20D4 DN 200 PN 40 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 53.15 ----
20D5 DN 200 PN 63 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 53.15 ----
GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
33

Table 11 : Selection Table Process Connection and Materials, Installation Length in inch (continued)

RCCS RCCS RCCS34 RCCS36 RCCS38 RCCS39 RCCS39 RCCS39


30LR 30-33 RCCT34 RCCT36 RCCT38 RCCT39 RCCT39/IR RCCT39/XR
SH SH SL HC SL HC SL HC SL HC SL HC SL HC
01J1 DN 15 10K 9.45 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01J2 DN 15 20K 9.45 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02J1 DN 25 10K ---- 9.45 14.57 15.35 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
Flanges according to JIS B 2220

02J2 DN 25 20K ---- 9.45 14.57 15.35 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
04J1 DN 40 10K ---- 9.45 14.57 15.35 19.69 20.47 23.62 ---- ---- ---- ---- ---- ---- ----
04J2 DN 40 20K ---- 9.45 14.57 15.35 19.69 20.47 23.62 ---- ---- ---- ---- ---- ---- ----
05J1 DN 50 10K ---- ---- ---- ---- 19.69 20.47 23.62 24.41 ---- ---- ---- ---- ---- ----
05J2 DN 50 20K ---- ---- ---- ---- 19.69 20.47 23.62 24.41 ---- ---- ---- ---- ---- ----
08J1 DN 80 10K ---- ---- ---- ---- ---- ---- 23.62 24.41 39.37 40.16 ---- ---- ---- ----
08J2 DN 80 20K ---- ---- ---- ---- ---- ---- 24.02 24.41 39.37 40.16 ---- ---- ---- ----
10J1 DN 100 10K ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 ---- ---- ----
10J2 DN 100 20K ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 ---- ---- ----
12J1 DN 125 10K ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
12J2 DN 125 20K ---- ---- ---- ---- ---- ---- ---- ---- 39.37 40.16 43.31 43.31 ---- ----
15J1 DN 150 10K ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 ---- ----
15J2 DN 150 20K ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- 43.31 43.31 ---- ----
01S4 DN 15 9.45 9.45 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02S4 DN 25 ---- 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Clamp DIN

04S4 DN 40 ---- 9.45 14.57 ---- 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S4 DN 50 ---- ---- ---- ---- 19.69 ---- 23.62 ---- ---- ---- ---- ---- ---- ----
06S4 DN 65 ---- ---- ---- ---- ---- ---- 23.62 ---- ---- ---- ---- ---- ---- ----
10S4 DN 100 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- ---- ---- ---- ----
01S8 ½“ 9.45 9.45 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02S8 1“ ---- 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Tri-Clamp

04S8 1½“ ---- 9.45 14.57 ---- 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S8 2“ ---- ---- ---- ---- 19.69 ---- 23.62 ---- ---- ---- ---- ---- ---- ----
08S8 3“ ---- ---- ---- ---- ---- ---- 23.62 ---- ---- ---- ---- ---- ---- ----
10S8 4“ ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- ---- ---- ---- ----
02S2 DN 25 ---- 9.45 14.57 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
DIN11851

04S2 DN 40 ---- ---- ---- ---- 19.69 ---- ---- ---- ---- ---- ---- ---- ---- ----
05S2 DN 50 ---- ---- ---- ---- ---- ---- 23.62 ---- ---- ---- ---- ---- ---- ----
10S2 DN 100 ---- ---- ---- ---- ---- ---- ---- ---- 39.37 ---- ---- ---- ---- ----
41G9 G¼“ female 10.24 10.24 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01G9 G½“ female 10.24 10.24 15.35 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
Thread

23G9 G¾“ female 10.24 10.24 15.35 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
41T9 NPT¼“ female 10.24 10.24 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01T9 NPT½“ female 10.24 10.24 15.35 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
23T9 NPT¾“ female 10.24 10.24 15.35 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ----

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
34

Table 12 : CRN approved process connections

RCCS RCCS34 RCCS36 RCCS38 RCCS39 RCCS39/IR RCCS39/XR


30 - 32 RCCT34 RCCT36 RCCT38 RCCT39 RCCT39/IR RCCT39/XR
SH SL HC SL HC SL HC SL HC SL HC SL
01A1 ½“-150 X X ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A2 ½“-300 X X ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A3 ½“-600 X X ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
01A5 ½“-900/1500 X X ---- ---- ---- ---- ---- ---- ---- ---- ---- ----
02A1 1“-150 X X X X ---- ---- ---- ---- ---- ---- ---- ----
02A2 1“-300 X X X X ---- ---- ---- ---- ---- ---- ---- ----
02A3 1“-600 X X ---- X ---- ---- ---- ---- ---- ---- ---- ----
02A5 1“-900/1500 X X ---- X ---- ---- ---- ---- ---- ---- ---- ----
04A1 1½“-150 X X X ---- X X ---- ---- ---- ---- ---- ----
04A2 1½“-300 X X X X X X ---- ---- ---- ---- ---- ----
04A3 1½“-600 X X ---- X ---- X ---- ---- ---- ---- ---- ----
04A4 1½“-900 ---- ---- ---- ---- ---- X ---- ---- ---- ---- ---- ----
04A5 1½“-900/1500 X X ---- X ---- ---- ---- ---- ---- ---- ---- ----
Flanges according to ASME B16.5

05A1 2“-150 ---- ---- ---- X X X X ---- ---- ---- ---- ----
05A2 2“-300 ---- ---- ---- X X X X ---- ---- ---- ---- ----
05A3 2“-600 ---- ---- ---- X ---- X ---- ---- ---- ---- ---- ----
05A4 2“-900 ---- ---- ---- ---- ---- X ---- ---- ---- ---- ---- ----
05A5 2“-900/1500 ---- ---- ---- X ---- ---- ---- ---- ---- ---- ---- ----
06A1 2½“-150 ---- ---- ---- ---- ---- X X ---- ---- ---- ---- ----
06A2 2½“-300 ---- ---- ---- ---- ---- X X ---- ---- ---- ---- ----
06A3 2½“-600 ---- ---- ---- ---- ---- X ---- ---- ---- ---- ---- ----
06A4 2½“-900 ---- ---- ---- ---- ---- X ---- ---- ---- ---- ---- ----
08A1 3“-150 ---- ---- ---- ---- ---- X X X X ---- ---- ----
08A2 3“-300 ---- ---- ---- ---- ---- X X X X ---- ---- ----
08A3 3“-600 ---- ---- ---- ---- ---- X ---- X ---- ---- ---- ----
08A4 3“-900 ---- ---- ---- ---- ---- X ---- ---- ---- ---- ---- ----
10A1 4“-150 ---- ---- ---- ---- ---- ---- ---- X X X X ----
10A2 4“-300 ---- ---- ---- ---- ---- ---- ---- X X X X ----
10A3 4“-600 ---- ---- ---- ---- ---- ---- ---- X ---- X ---- ----
12A1 5“-150 ---- ---- ---- ---- ---- ---- ---- X X X X ----
12A2 5“-300 ---- ---- ---- ---- ---- ---- ---- X X X X ----
12A3 5“-600 ---- ---- ---- ---- ---- ---- ---- X ---- X ---- ----
15A1 6“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- X X X
15A2 6“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- X X X
15A3 6“-600 ---- ---- ---- ---- ---- ---- ---- ---- ---- X ---- ----
20A1 8“-150 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- X
20A2 8“-300 ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- ---- X

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa
35

All Rights Reserved. Copyright © 2003, Rota Yokogawa GS 01R04B04-00E-E 24th edition October 27, 2016-00
36

REGISTERED TRADEMARKS
HART® : Registered trademark of HART Communication Foundation, Austin, TX, USA
MODBUS® : Registered trademark of MODBUS Organization
FOUNDATION™ Fieldbus : Registered trademark of the Fieldbus FOUNDATION, Austin, TX, USA
TRI-CLAMP® : Registered trademark of Ladish & Co., Inc., Kenosha, USA
FieldMate™ : Registered trademark of Yokogawa Electric Corporation

YOKOGAWA ELECTRIC CORPORATION


Headquarters
2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 JAPAN
Phone : 81-422-52-5555
Branch Sales Offices
Osaka, Nagoya, Hiroshima, Kurashiki, Fukuoka, Kitakyusyu

YOKOGAWA CORPORATION OF AMERICA


Head Office
12530 West Airport Blvd, Sugar Land, Texas 77478, USA
Phone : 1-281-340-3800 Fax : 1-281-340-3838
Georgia Office
2 Dart Road, Newnan, Georgia 30265, USA
Phone : 1-800-888-6400/ 1-770-253-7000 Fax : 1-770-254-0928

YOKOGAWA AMERICA DO SUL LTDA.


Praca Acapulco, 31 - Santo Amaro, Sáo Paulo/SP, BRAZIL, CEP-04675-190
Phone : 55-11-5681-2400 Fax : 55-11-5681-4434

YOKOGAWA EUROPE B. V.
Euroweg 2, 3825 HD Amersfoort, THE NETHERLANDS
Phone : 31-88-4641000 Fax : 31-88-4641111

YOKOGAWA ELECTRIC CIS LTD.


Grokholskiy per 13 Building 2, 4th Floor 129090, Moscow, RUSSIA
Phone : 7-495-737-7868 Fax : 7-495-737-7869

YOKOGAWA CHINA CO., LTD.


3F Tower D Cartelo Crocodile Building, No.568 West Tianshan Road,
Shanghai 200335, CHINA
Phone : 86-21-62396262 Fax : 86-21-62387866

YOKOGAWA ELECTRIC KOREA CO., LTD.


(Yokogawa B/D, Yangpyeong-dong 4-Ga), 21, Seonyu-ro 45-gil, Yeongdeungpo-gu,
Seoul, 150-866, KOREA
Phone : 82-2-2628-6000 Fax : 82-2-2628-6400

YOKOGAWA ENGINEERING ASIA PTE. LTD.


5 Bedok South Road, Singapore 469270, SINGAPORE
Phone : 65-6241-9933 Fax : 65-6241-2606

YOKOGAWA INDIA LTD.


Plot No.96, Electronic City Complex, Hosur Road, Bangalore - 560 100, INDIA
Phone : 91-80-4158-6000 Fax : 91-80-2852-1442

YOKOGAWA AUSTRALIA PTY. LTD.


Tower A, 112-118 Talavera Road, Macquarie Park NSW 2113, AUSTRALIA
Phone : 61-2-8870-1100 Fax : 61-2-8870-1111

YOKOGAWA MIDDLE EAST & AFRICA B.S.C.(C)


P.O. Box 10070, Manama, Building 577, Road 2516, Busaiteen 225, Muharraq,
Kingdom of BAHRAIN
Phone : 973-17358100 Fax : 973-17336100
Apr. '15

GS 01R04B04-00E-E 24th edition October 27, 2016-00 All Rights Reserved. Copyright © 2003, Rota Yokogawa

You might also like