H-046-001653-00 BeneHeart D3 User Manual PDF
H-046-001653-00 BeneHeart D3 User Manual PDF
H-046-001653-00 BeneHeart D3 User Manual PDF
Defibrillator/Monitor
Operator’s Manual
© 2010 - 2013 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.
For this Operator’s Manual, the issue date is 2013-09.
I
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property
rights to this Mindray product and this manual. This manual may refer to information protected by copyright or
patents and does not convey any license under the patent rights or copyright of Mindray, or of others.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in
this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.
Release, amendment, reproduction, distribution, rental, adaptation, translation or any other derivative work of this
manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.
Mindray in China and other countries. All other trademarks that appear in this manual are used only for informational
or editorial purposes. They are the property of their respective owners.
II
Responsibility on the Manufacturer Party
Contents of this manual are subject to change without prior notice.
All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained
herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this
manual.
Mindray is responsible for the effects on safety, reliability and performance of this product, only if:
all installation operations, expansions, changes, modifications and repairs of this product are conducted by
Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national and local requirements;and
WARNING
z This equipment must be operated by skilled/trained clinical professionals.
z It is important for the hospital or organization that employs this equipment to carry out a reasonable
service/maintenance plan. Neglect of this may result in machine breakdown or personal injury.
III
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability
for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or
the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized
personnel.
Malfunction or damage caused by improper operation or repair by unqualified orunauthorized service people.
Malfunction of the instrument or part whose serial number is not legible enough.
Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building,Keji 12th Road South,Hi-tech industrial
park,Nanshan,Shenzhen 518057,P.R.China
Website: www.mindray.com
E-mail Address: [email protected]
IV
Preface
Manual Purpose
This manual contains the instructions necessary to operate the product safely and in accordance with its function and
intended use. Observance of this manual is a prerequisite for proper product performance and correct operation and
ensures patient and operator safety.
This manual is based on the maximum configuration and therefore some contents may not apply to your product. If
you have any question, please contact us.
This manual is an integral part of the product. It should always be kept close to the equipment so that it can be
obtained conveniently when needed.
Intended Audience
This manual is geared for clinical professionals who are expected to have a working knowledge of medical procedures,
practices and terminology as required for monitoring of critically ill patients.
Illustrations
All illustrations in this manual serve as examples only. They may not necessarily reflect the setup or data displayed on
your equipment.
Conventions
Italic text is used in this manual to quote the referenced chapters or sections.
V
FOR YOUR NOTES
VI
Contents
1 Safety............................................................................................................................................................................ 1-1
1.1 Safety Information.............................................................................................................................................................................................1-1
1.1.1 Dangers ...................................................................................................................................................................................................1-2
1.1.2 Warnings .................................................................................................................................................................................................1-2
1.1.3 Cautions ..................................................................................................................................................................................................1-3
1.1.4 Notes ........................................................................................................................................................................................................1-3
1.2 Equipment Symbols..........................................................................................................................................................................................1-4
1
4 Managing Patients....................................................................................................................................................... 4-1
4.1 Overview .............................................................................................................................................................................................................. 4-1
4.2 Editing Patient Information ........................................................................................................................................................................... 4-1
5 Alarms........................................................................................................................................................................... 5-1
5.1 Alarm Categories ............................................................................................................................................................................................... 5-1
5.2 Alarm Levels ........................................................................................................................................................................................................ 5-2
5.3 Alarm Indicators................................................................................................................................................................................................. 5-2
5.3.1 Alarm Lamps ......................................................................................................................................................................................... 5-2
5.3.2 Audible Alarms ..................................................................................................................................................................................... 5-3
5.3.3 Alarm Message ..................................................................................................................................................................................... 5-3
5.3.4 Flashing Numeric................................................................................................................................................................................. 5-3
5.3.5 Alarm Status Symbols ........................................................................................................................................................................ 5-3
5.4 Alarm Tone Configuration .............................................................................................................................................................................. 5-4
5.4.1 Changing the Alarm Volume........................................................................................................................................................... 5-4
5.4.2 Setting the Interval between Alarm Sounds ............................................................................................................................. 5-4
5.5 Understanding the Alarm Setup Menu..................................................................................................................................................... 5-4
5.5.1 Setting Alarm Properties for All Parameters.............................................................................................................................. 5-5
5.5.2 Adjusting Alarm Limits Automatically ......................................................................................................................................... 5-6
5.6 Pausing Alarms................................................................................................................................................................................................... 5-7
5.7 Switching Alarms Off ....................................................................................................................................................................................... 5-7
5.8 Pausing Alarm Sounds..................................................................................................................................................................................... 5-8
5.9 Switching Off Alarm Sounds ......................................................................................................................................................................... 5-8
5.10 Reminder Tones ............................................................................................................................................................................................... 5-8
5.11 Latching Alarms............................................................................................................................................................................................... 5-9
5.12 Clearing Technical Alarms............................................................................................................................................................................ 5-9
5.13 When an Alarm Occurs................................................................................................................................................................................5-10
2
6.7.4 Changing Arrhythmia Threshold Settings ............................................................................................................................... 6-10
6.7.5 Initiating Arrhythmia Relearning Manually............................................................................................................................. 6-11
6.7.6 Automatic Arrhythmia Relearn.................................................................................................................................................... 6-11
6.8 Calibrating ECG................................................................................................................................................................................................ 6-11
3
10.4.2 Changing Resp Wave Settings....................................................................................................................................................10-3
4
16 Review ...................................................................................................................................................................... 16-1
16.1 Reviewing Events ......................................................................................................................................................................................... 16-1
16.2 Reviewing Tabular Trends.......................................................................................................................................................................... 16-2
20 Battery...................................................................................................................................................................... 20-1
20.1 Introduction ................................................................................................................................................................................................... 20-1
5
20.2 Installing the Batteries ................................................................................................................................................................................20-2
20.3 Battery Alarms................................................................................................................................................................................................20-2
20.3.1 No Battery Alarm.............................................................................................................................................................................20-2
20.3.2 Low Battery Alarm ..........................................................................................................................................................................20-2
20.3.3 Battery Aged Alarm........................................................................................................................................................................20-3
20.3.4 Battery Error Alarm.........................................................................................................................................................................20-3
20.4 Checking the Batteries................................................................................................................................................................................20-3
20.5 Charging batteries........................................................................................................................................................................................20-3
20.6 Storing Batteries............................................................................................................................................................................................20-4
20.7 Recycling the Batteries................................................................................................................................................................................20-4
A Specifications ..............................................................................................................................................................A-1
A.1 General Specifications.....................................................................................................................................................................................A-1
A.2 Defibrillator Specifications ............................................................................................................................................................................A-2
A.3 Pacer Specifications .........................................................................................................................................................................................A-5
A.4 Monitor Specifications ....................................................................................................................................................................................A-5
A.5 Power Supply Specifications.........................................................................................................................................................................A-9
A.6 Recorder Specifications ............................................................................................................................................................................... A-10
6
A.7 Alarm Specifications......................................................................................................................................................................................A-10
A.8 Data Management Specifications ............................................................................................................................................................A-11
A.9 Environmental Specifications ....................................................................................................................................................................A-11
B EMC............................................................................................................................................................................... B-1
7
FOR YOUR NOTES
8
1 Safety
1.1 Safety Information
DANGER
z Indicates an imminent hazard that, if not avoided, will result in death or serious injury.
WARNING
z Indicates a potential hazard or unsafe practice that, if not avoided, could result in death or serious injury.
CAUTION
z Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or
product/property damage.
NOTE
z Provides application tips or other useful information to ensure that you get the most from your product.
1-1
1.1.1 Dangers
DANGER
z The equipment delivers up to 360 J of electrical energy. Unless properly used as described in these
Operating Instructions, this electrical energy may cause serious injury or death. Do not attempt to operate
this defibrillator unless thoroughly familiar with these operating instructions and the function of all
controls, indicators, connectors, and accessories.
z Defibrillation current can cause operator or bystander severe injury or even death. Keep distance with the
patient or metal devices connected to the patient during defibrillation.
z Do not disassemble the defibrillator. It contains no operator serviceable components and dangerous high
voltages may be present. Contact authorized service personnel for repair.
z To avoid explosion hazard, do not use the equipment in the presence of oxygen-rich atmospheres,
flammable anesthetics, or other flammable agents (such as gasoline). Keep the equipment and the
operating environment dry and clean.
1.1.2 Warnings
WARNING
z Before putting the system into operation, the operator must verify that the equipment, connecting cables
and accessories are in correct working order and operating condition.
z Make sure the synchronous input system is applied to this equipment and the input signal is correct if
necessary.
z The equipment must be connected to a properly installed power outlet with protective earth contacts only.
If the installation does not provide for a protective earth conductor, disconnect it from the power line and
operate it on smart lithium-ion batteries.
z Medical electrical equipment which does not incorporate defibrillator protection should be disconnected
during defibrillation.
z Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment of alarm volume to
a low level or off may result in a hazard to the patient. Remember that alarm settings should be customized
according to different patient situations and always keeping the patient under close surveillance is the most
reliable way for safe patient monitoring.
z Do not perform any functional check if the equipment is connected with a patient; otherwise the patient
might be shocked.
z Remain attentive to the patient during applying therapy. Delay in delivering a shock may result in a rhythm
that was analyzed as shockable converting spontaneously to non-shockable and could result in
inappropriate delivery of a shock.
1-2
z For the treatment of patients with implantable pacemakers, place therapy pads or paddles away from
internal pacemaker generator if possible to help prevent damage to the pacemaker.
z To avoid inadvertent disconnection, route all cables in a way to prevent a stumbling hazard. Wrap and
secure excess cabling to reduce risk of entanglement or strangulation by patients or personnel.
z Do not touch device connectors, recorder print head, battery connector or other live equipment if in contact
with the patient; otherwise patient injury may result.
z To ensure patient safety, use only parts and accessories specified in this manual.
z Package material may contaminate the environment. Properly dispose of the package material according to
applicable waste control regulations and keep it out of children’s reach.
1.1.3 Cautions
CAUTION
z Use of Manual Therapy security password requires the clinician to know and remember the password.
Failure to enter correct password will prevent the delivery of manual defibrillation, synchronized
cardioversion and pacing therapy.
z At the end of its service life, the equipment, as well as its accessories, must be disposed of in compliance
with the guidelines regulating the disposal of such products to avoid contaminating the environment.
z Magnetic and electrical fields are capable of interfering with the proper performance of the equipment. For
this reason make sure that all external devices operated in the vicinity of the equipment comply with the
relevant EMC requirements. Mobile phones, X-ray equipment or MRI devices are a possible source of
interference as they may emit higher levels of electromagnetic radiation.
z Before connecting the equipment to the power line, check that the voltage and frequency ratings of the
power line are the same as those indicated on the equipment’s label or in this manual.
z Always install or carry the equipment properly to avoid damage caused by drop, impact, strong vibration or
other mechanical force.
z Do not frequently charge the equipment and deliver shock in non-clinical situations. Otherwise equipment
damage could occur.
1.1.4 Notes
NOTE
z Put the equipment in a location where you can easily see the screen and access the operating controls.
z Keep this manual in the vicinity of the equipment so that it can be obtained conveniently when needed.
z If the equipment is run on a DC power supply, a DC/AC adapter we supply should be used.
z This manual describes all features and options. Your equipment may not have all of them.
1-3
1.2 Equipment Symbols
Menu Unlocking
1-4
Fragile Right side up
1-5
FOR YOUR NOTES
1-6
2 The Basics
2.1 Overview
The BeneHeart (hereinafter called the equipment) is a lightweight and portable defibrillator/monitor. It provides four
operating modes: Monitor, Manual Defib, AED and Pacer.
In Monitor Mode, the equipment is intended for monitoring, displaying, reviewing, storing and printing multiple
physiological parameters and waveforms including ECG, pulse oximetry (SpO2), respiration (Resp), and non-invasive
blood pressure (NIBP).
In AED mode, the equipment automatically analyzes the patient’s ECG rhythm and indicates whether or not a shockable
rhythm is detected. Voice prompts provide easy-to-follow instructions and patient information to guide you through the
defibrillation process. Messages and flashing buttons are also presented to reinforce the voice prompts.
In the Manual Defib Mode, the operator analyzes the patient’s ECG, and, if appropriate, follows this procedure:
1 Select the Manual Defib mode, adjust the energy level if necessary;
2 Charge; and
Defibrillation may be performed through paddles or multifunction electrode pads. In Manual Defib Mode, you can also
perform synchronized cardioversion. If desired, use of Manual Defib Mode may be password protected.
The Pacer Mode offers non-invasive transcutaneous pacing therapy. Pace pulses are delivered through multifunction
electrode pads. Use of Pacer Mode may also be password protected.
The equipment can be powered by smart lithium ion batteries which are rechargeable and maintenance-free. You can
easily determine the remaining battery charge by viewing the battery power gauge displayed on the screen or by
checking the indicator on the battery itself. An external AC mains or a DC power supply connected through an DC/AC
adapter may also be used as a power source and for continuous battery charging.
The equipment automatically stores patient data in an internal storage card. You can also export the data through the
USB port for viewing and editing on a PC through the data management software.
The equipment is for use in hospital and pre-hospital settings by qualified medical personnel trained in the operation of
the equipment and qualified by training in basic life support, advanced cardiac life support or defibrillation.
2-1
2.2.1 AED
The AED mode is to be used only on cardio arrest patients who are at least 8 years. The patients must be:
Unresponsive
2.2.4 ECG
The ECG monitoring function is used to monitor and/or record the patient’s ECG waveform and heart rate.
2.2.5 Resp
The respiration monitoring function is used to continuously monitor the patient’s respiration rate and respiration
waveform.
2.2.6 SpO2
The SpO2 function is intended to measure patient’s oxygen saturation in arterial blood.
2.2.7 NIBP
The NIBP function is intended for non-invasive measurement of a patient’s arterial blood pressure.
2-2
2.3 Main Unit
2.3.1 Front View
External paddle
Handle
Alarm
lamp
Area 1
Area 3
Microphone
Area 2
Speaker
2-3
Area 1
2 3 4 5 6
1. Display screen
2. AC power indicator
3. Battery indicator
Green: when the battery is fully charged or the equipment is run on battery.
Flashing: when a failure is detected or when battery is not installed if [No Battery] is configured as
[Status Indicator ON].
NOTE
z Both status indicators are off when AC mains is not connected if the equipment is turned off and no failure
is detected.
6. Soft keys
They are corresponding with the soft key labels located immediately above. The labels of the soft keys changes
according to the current operating mode.
2-4
Area 2
3 4 5 6 7
1. Lead Select button
Press this button to select the lead of the first ECG waveform.
Press this button to select the size of the first ECG waveform.
Press it to manually mark specified events. If a menu has been open, pressing this button will close the menu.
If no menu is displayed on the screen, pressing it will enter the main menu. If there is a menu displayed, pressing it
will close that menu.
7. Navigation knob
You can:
2-5
Area 3
Rotate this knob to select the operating mode or turn the equipment off.
3. Charge button
4. Shock button
2-6
Recorder
1 2
3 5
1. Start/Stop Key
2. Indicator
Flashes: when an error occurred to the recorder, or the recorder runs out of paper.
3. Paper outlet
4. Recorder door
5. Latch
2-7
2.3.2 Side View
Therapy port
2 4
1. Recorder
2-8
2.3.3 Rear View
6
2
4
1. Hook
2. Battery
When the defibrillator/monitor and other devices are to be used together, their equipotential grounding terminals
should be connected together to eliminate the potential difference between them.
It connects an AC power cord or a DC/AC adapter to run the equipment respectively on the external AC mains or
DC power supply.
5. Multifunctional connector
6. USB connector
7. Network connector
2-9
2.3.4 External Paddles
1 2 5
3
4
1. Shock button
3. Shock indicator
4. Charge button
5. Shock button
2-10
2.4 Display Views
A typical screen in Manual Defib Mode is shown below.
1 2 3 4 5
6 7
This area shows patient name, patient category, paced status, and current date and time.
This area shows physiological alarm messages. When multiple alarms occur, they will be displayed circularly.
This area shows technical alarm messages and prompt messages. When multiple messages come, they will be
displayed circularly.
2-11
6. Waveform area
This area shows measurement waveforms. The waveform label is displayed at the upper left corner of the
waveform.
7. Parameter area
This area shows measurement parameters. Each measurement module has a parameter block and the parameter
name is displayed at the upper left corner.
This area shows the selected defibrillation energy, shock counter as well as prompt related to manual defibrillation.
9. Runtime area
This area shows the equipment's operating time since it is turned on.
The three soft key labels correspond to the soft key buttons located immediately below. The labels of the soft keys
changes according to the current display view and function. Soft key labels appearing as blank indicate that the
soft key is inactive.
2-12
3 Basic Operations and Settings
3.1 Installation
WARNING
z The equipment shall be installed by personnel authorized by the manufacturer.
z The software copyright of the equipment is solely owned by the the manufacturer. No organization or
individual shall resort to juggling, copying, or exchanging it or to any other infringement on it in any form
or by any means without due permission.
z Devices connected to the equipment must meet the requirements of the applicable IEC standards (e.g. IEC
60950 safety standards for information technology equipment and IEC 60601-1 safety standards for medical
electrical equipment). The system configuration must meet the requirements of the IEC 60601-1-1 medical
electrical systems standard. Any personnel who connect devices to the equipment’s signal input/output
port is responsible for providing evidence that the safety certification of the devices has been performed in
accordance to the IEC 60601-1-1. If you have any question, please contact the the manufacturer.
z If it is not evident from the equipment specifications whether a particular combination is hazardous, for
example, due to summation of leakage currents, consult the manufacturers or else an expert in the field, to
ensure the necessary safety of all devices concerned will not be impaired by the proposed combination.
WARNING
z Package material may contaminate the environment. Properly dispose of the package material according to
applicable waste control regulations and keep it out of children’s reach.
z The equipment might be contaminated during storage and transport. Before use, please verify whether the
packages are intact, especially the packages of single use accessories. In case of any damage, do not apply it
to patients.
NOTE
z Save the packing case and packaging material as they can be used if the equipment must be reshipped.
3-1
3.1.2 Environmental Requirements
The operating environment of the equipment must meet the requirements specified in this manual.
The environment where the equipment is used shall be reasonably free from noises, vibration, dust, corrosive,
flammable and explosive substances. If the equipment is installed in a cabinet, sufficient space in front and behind shall
be left for convenient operation, maintenance and repair. Moreover, to maintain good ventilation, the equipment shall
be at least 2 inches (5 cm) away from around the cabinet.
When the equipment is moved from one place to another, condensation may occur as a result of temperature or
humidity difference. In this case, never start the system before the condensation disappears.
NOTE
z Make sure that the operating environment of the equipment meets the specific requirements. Otherwise
unexpected consequences, e.g. damage to the equipment, could result.
1. Before turning on the equipment, check for any mechanical damage and make sure that all external cables,
plug-ins and accessories are properly connected.
2. Plug the power cord into the AC power source. If you run the equipment on battery power, ensure that the battery
is sufficiently charged. If you run the equipment on DC power supply, an DC/AC adapter we supply should be used.
3. Turn the Mode Select knob to enter the desired working mode. After the start-up screen is displayed, the system
gives a beep, and meanwhile, the alarm lamp is illuminated in yellow, and then turns red, and then turns off.
WARNING
z Do not use the equipment for any monitoring or therapy procedure on a patient if you suspect it is not
working properly, or if it is mechanically damaged. Contact your service personnel or us.
4. Enter the appropriate operating mode and check that the settings are proper for your patient.
Refer to the appropriate sections for details of how to perform patient monitoring and therapy.
3-2
3.2.3 Disconnecting from Power
To disconnect the equipment from the AC power source, follow this procedure:
4. Turn the Mode Select Knob to Off. After 10 seconds, the equipment is shut down.
Other menus are similar to the main menu and contain the following parts:
1. Heading
2. Main body: displays options, buttons, prompt messages, etc. Pressing the menu button with “>>’’ enters a
submenu to reveal more options or information.
3. Exit button
3-3
You can also set system time by selecting [Configuration >>]→[View Config]→[General Setup >>]. However, you
cannot select date format and time format in this case. After the completion of setting system time, exit the
configuration mode, and then the system will restart.
2. Set [Brightness] to an appropriate level: 10 is the brightest, and 1 is the least bright.
You can also change screen brightness by entering configuration mode and selecting [Others] from the Configuration
Main menu.
2. Select [Key Volume] and then select an appropriate value. 0 means key volume off and 10 is the maximum volume.
You can also change key volume by entering configuration mode and selecting [Others] from the Configuration Main
menu.
In the Monitor, Manual Defib and Pacer mode, press the Main Menu button on the front panel, and then select
[High Contrast]. To disable high contrast display, select [Full Color] in the Main Menu.
In AED mode, press the [High Contrast] soft key. To disable high contrast display, press the [Full Color] soft key.
Once High Contrast is selected, the system remains in high contract mode when you change the operating mode.
However, the setting will not be saved if the equipment is turned off.
3-4
4 Managing Patients
4.1 Overview
Patient information management function enables you to edit and manage information of the current patient.
1. Press the Main Menu button on the front panel, and then select [Patient Demographics >>] and then make
changes as desired.
Archive ID is created automatically when the equipment is turned on. It is unchangeable. When the equipment is turned
off, the current patient is discharged and the archive ID turns to be historical archive ID
For a new patient, if patient category is changed, the system will restore the default alarm settings of this patient
category; if patient category is not changed, the alarm settings remain unchanged. If you restart the equipment after
normal power-off, the default alarm settings will be loaded.
4-1
FOR YOUR NOTES
4-2
5 Alarms
Alarms, triggered by a vital sign that appears abnormal or by technical problems of the equipment, are indicated to the
user by visual and audible alarm indications.
WARNING
z A potential hazard exists if different alarm presets are used for the same or similar device in any single area,
e.g. an intensive care unit or cardiac operating room.
1. Physiological alarms
Physiological alarms, also called patient status alarms, are triggered by a monitored parameter value that violates
set alarm limits or by an abnormal patient condition. Physiological alarm messages are displayed in the
physiological alarm area. In AED mode, no physiological alarm will be presented.
2. Technical alarms
Technical alarms, also called system status alarms, are triggered by a device malfunction or a patient data distortion
due to improper operation or system failure. Technical alarm messages are displayed in the technical alarm area.
3. Prompt messages
As a matter of fact, prompt messages are not alarm messages. Apart from the physiological and technical alarms,
the equipment also shows some messages indicating system status. Messages of this kind are usually displayed in
the prompt area. Therapy-related prompts are shown in corresponding information area. Some special prompts are
shown in dialog boxes.
5-1
5.2 Alarm Levels
By severity, alarms can be classified into three categories: high level alarms, medium level alarms and low level alarms.
Alarm lamp
Alarm tones
Alarm message
Flashing numeric
NOTE
z When multiple alarms of different levels occur simultaneously, the equipment will select the alarm of the
highest level and give visual and audible alarm indications accordingly. Alarm messages will be displayed
circularly.
z Some physiological alarms, such as Asystole, are exclusive. They have identical alarm tones and alarm lights
with normal high level physiological alarms, but their alarm messages are displayed exclusively. That is to
say, when an exclusive physiological alarm and a normal high level physiological alarms are triggered
simultaneously, only alarm message of the exclusive physiological alarm is displayed.
5-2
5.3.2 Audible Alarms
The equipment uses different alarm tone patterns to match the alarm level:
Additionally, the alarm message has different background color which matches the alarm level.
For physiological alarms
5-3
5.4 Alarm Tone Configuration
5.4.1 Changing the Alarm Volume
1. Press the Main Menu button on the front panel, and then select [Alarm Setup >>]→[Alm Volume >>].
If [Audio Off] is enabled, alarm volume can be set to a value between 0 and 10, in which 0 means audio off and
10 the maximum volume level.
If [Audio Off] is disabled, alarm volume can be set to a value between 1 and 10, in which 1 is the minimum
volume level and 10 the maximum.
The setting of alarm volume will not be saved when the system is power off.
You can also set alarm volume in configuration mode. In this case, the setting will be saved.
NOTE
z You cannot adjust alarm volume when an alarm is switched off.
3. Respectively set [High Alarm Interval (s)], [Med Alarm Interval (s)] and [Low Alarm Interval (s)].
WARNING
z Do not rely exclusively on audible alarm system. Setting alarm volume to a low level may result in a hazard
to the patient. Always keep the patient under close surveillance.
5-4
5.5.1 Setting Alarm Properties for All Parameters
In the main menu, select [Alarm Setup >>]→[Para. Alarm >>] to enter the Para. Alarm setup menu, where you can
review and set alarm limits, alarm switches, alarm level and alarm recordings for all parameters.
When a parameter alarm is switched on, the equipment gives alarm indications in accordance with the preset alarm
level and stores related waveforms and parameter values.
When a parameter alarm is switched off, the alarm off symbol “ ” is displayed in the parameter window.
If the measurement’s [On/Off] and [Record] are set to [On], automatic recording of all the measurement numerics
and related waveforms is possible when a measurement alarm occurs.
NOTE
z You cannot simultaneously switch on HR and PR alarms. In the case that PR alarm is on, switching on HR
alarm will automatically turn off PR alarm, and vice versa.
You can also set parameter alarm properties by selecting a parameter window and select [Para. Alarm >>] in the
pop-up menu.
WARNING
z Make sure that the alarm limits settings are appropriate for your patient before patient monitoring.
z Setting the alarm limit to an extreme value may cause the alarm system to be ineffective. For example, high
oxygen levels may predispose a premature infant to retrolental fibroplasia. If this is a consideration, do NOT
set the SpO2 high alarm limit to 100%, which is equivalent to switching the alarm off.
5-5
5.5.2 Adjusting Alarm Limits Automatically
The defibrillator/monitor can automatically adjust the patient’s alarm limits according to the measured vital signs. When
[Auto Limits] is selected, the equipment automatically calculates alarm limits based on the latest measured parameter
values.
To enable auto alarm limits, press the Main Menu button on the equipment’s front panel, and then select [Alarm Setup
>>]→[Para. Alarm>>]→[Auto Limits].
You can also access [Auto Limits] by select a parameter window to enter the [Para. Alarm] menu.
When auto alarm limits have been applied, you can manually adjust the alarm limits through the Para. Alarm Setup
menu so that they are appropriate for your patient.
The defibrillator/monitor calculates the auto limits based on the following rules.
Low alarm limit High alarm limit
Auto alarm limits
Module Parameter Adult/ Adult/
Neonate Neonate range
pediatric pediatric
HR×0.8 or 40 bpm (HR-30) or 90 bpm HR×1.25 or 240 (HR + 40) or 200 Adult/pediatric: 35 to
ECG HR (whichever is (whichever is bpm whichever is bpm whichever is 240
greater) greater) lower) lower) Neonate: 55 to 225
RR×0.5 or 6/min (RR-10) or 30/min RR×1.5 or 30/min (RR+25) or 85/min Adult/pediatric:
Resp RR (whichever is (whichever is (whichever is (whichever is 6 to 55
greater) greater) lower) lower) Neonate: 10 to 90
Same as the Same as the Same as the Same as the default Same as the
SpO2
default alarm limit default alarm limit default alarm limit alarm limit measurement range
SpO2 PR×0.8 or 40 bpm (PR-30) or 90 bpm PR×1.25 or 240 (PR + 40) or 200 Adult/pediatric: 35 to
PR (whichever is (whichever is bpm whichever is bpm whichever is 240
greater) greater) lower) lower) Neonate: 55 to 225
(SYS-15) or 45 (SYS + 15) or 105 Adult: 45 to 270
SYS × 0.68 + 10 SYS × 0.86 + 38
NIBP-S mmHg (whichever mmHg (whichever Pediatric: 45 to 185
mmHg mmHg
is greater) is lower) Neonate: 35 to 115
(Dia-15) or 20 (Dia + 15) or 80 Adult: 25 to 225
Dia × 0.68 + 6 Dia × 0.86 + 32
NIBP NIBP-D mmHg (whichever mmHg (whichever Pediatric: 25 to 150
mmHg mmHg
is greater) is lower) Neonate: 20 to 90
(Mean-15) or 35 (Mean + 15) or 95 Adult: 30 to 245
Mean × 0.68 + 8 Mean × 0.86 + 35
NIBP-M mmHg (whichever mmHg (whichever Pediatric: 30 to 180
mmHg mmHg
is greater) is lower) Neonate: 25 to 105
NOTE
z You can enable auto alarm limits only when the current parameter measurement is within the auto alarm
limits range.
5-6
5.6 Pausing Alarms
You can temporarily disable alarm indicators by pressing the hardkey on the equipment’s front. When alarms are
paused:
For physiological alarms, no alarm indication is shown. New physiological alarm will not be presented.
The remaining alarm pause time is displayed in the physiological alarm area.
For technical alarms, alarm sounds are paused, but alarm lamps and alarm messages remain presented.
is displayed in the sound symbol area. If a new technical alarm is triggered in the alarm paused period, the
When the alarm pause time expires, the alarm paused status is automatically deactivated. You can also cancel the alarm
The default alarm pause time is 2 minutes. To change alarm pause time,
1. Press the Main Menu button on the front panel, and then select [Others >>]→ [Configuration >>]→enter the
required password.
2. Select [Alarm Setup >>]→[Alarm Pause Time] and then select an appropriate value.
The equipment exit synchronous defibrillation when operating in the Manual Defib mode.
5-7
5.8 Pausing Alarm Sounds
You can press the [Audio Pause] softkey to pause alarm tones. In this case, the symbol will be displayed in the
sound symbol area indicating all system sounds are silenced temporarily. In the audio paused status, all alarm indicators
except audible alarm tones works properly. If alarms are switched off or paused, the [ Audio Pause] soft key is not
shown.
mode, Manual Defib mode or Pacer mode. In the audio off status, appears in the sound symbol area. In this case,
the alarm status is the same with that when alarm tones are paused.
The hardkey is pressed. In this case, the equipment enters the alarm paused status and the alarm volume is
reset to the default level. The symbol is displayed in the sound symbol area.
The [Audio Pause] soft key is pressed. In this case, the equipment enters the audio paused status and the alarm
volume is reset to the default level. The symbol is displayed in the sound symbol area.
Operating mode is switched. Then the equipment enters the default alarm status of corresponding operating mode.
Or
The reminder tone is switched off by default. You can switch it on by selecting [Alarm Setup]→[Reminder Tone]
through the Configuration Management Main menu. You can also change reminder volume. The default reminder
volume is [Med].
5-8
5.11 Latching Alarms
The latching setting for physiological alarms defines how alarm indicators behave when you do not acknowledge them.
If an alarm is latched, alarm indications remain presented even though alarm conditions end, except that:
The time when the alarm is last triggered is displayed behind the alarm message.
If an alarm is not latched, the alarm indications disappear as soon as the alarm conditions end.
1. Press the Main Menu button on the front panel, and then select [Others >>]→ [Configuration >>]→enter the
required password.
prompt messages after the hardkey or [Audio Pause] soft key is pressed. After the equipment restores the normal
alarm status, it can give alarm indications correctly in case these alarms are triggered again.
For some technical alarms, all their alarm indications are cleared after the hardkey or [Audio Pause] soft key is
pressed. After the equipment restores the normal alarm status, it can give alarm indications correctly in case these
alarms are triggered again.
For others, their alarm tones are cleared but the alarm lamp flashing and alarm messages remain after the hardkey
or [Audio Pause] soft key is pressed. After the equipment restores the normal alarm status, all the alarm indications will
continue if the alarm conditions still present.
5-9
5.13 When an Alarm Occurs
When an alarm occurs, observe the following steps and take proper actions:
For actions taken with regard to specific alarms, see Appendix D Alarm Messages.
5-10
6 Monitoring ECG
6.1 Overview
The electrocardiogram (ECG) measures the electrical activity of the heart and displays it as waveforms and numerics. The
equipment enables ECG monitoring through 3-, 5- lead ECG sets, external paddles and multifunction electrode pads. If
both ECG sets and paddles/pads are connected, the configured ECG waveforms are displayed in the waveform area.
6.2 Safety
WARNING
z Periodically inspect the electrode application site to ensure skin quality. If the skin quality changes, replace
the electrodes or change the application site.
z When monitoring a patient implanted with a pacemaker, be sure to select correct paced status. Otherwise,
the pacing pulses may be counted in the case of cardiac arrest or some arrhythmias. Do not completely rely
on the heart rate reading or the heart rate alarms. Always keep paced patients under close surveillance.
NOTE
z When connecting electrodes and/or patient cables, make sure that the connectors never come into contact
with other conductive parts, or with earth. Particularly make sure that all of the ECG electrodes are attached
to the patient.
z Interference from a non-grounded instrument near the patient and electrosurgery interference may cause
problems with the waveform.
z If selected lead cannot provide valid ECG signals, a dash line is shown in the ECG waveform area.
z Use the same type of ECG electrodes when monitoring ECG through ECG lead set.
6-1
6.3 Monitoring View
You can access Monitor mode by switching the Mode Select knob to the Monitor position. When operating in Monitor
mode, the equipment displays up to two ECG waveforms, the heart rate reading, other available parameter values and
active alarm settings.
Gently rub skin surface at application sites to remove dead skin cells.
Thoroughly clean the sites with mild soap and water. We do not recommend using ether or pure alcohol,
because this dries the skin and increases the resistance.
4. Attach the lead wires to the ECG trunk cable and then plug the trunk cable into the equipment’s ECG connector.
3-Lead Placement
The following is a typical AHA electrode placement for a 3-lead ECG set:
RA placement: directly below the clavicle and near the right shoulder.
LA placement: directly below the clavicle and near the left shoulder.
6-2
5-Lead Placement
The following is a typical AHA electrode placement for a 5-lead ECG set:
RA placement: directly below the clavicle and near the right shoulder.
LA placement: directly below the clavicle and near the left shoulder.
The chest (V) electrode can be placed on one of the following positions:
V5 placement: on the left anterior axillary line, horizontal with the V4 electrode
position.
V7 placement: on posterior chest at the left posterior axillary line in the fifth intercostal space.
V7R placement: on posterior chest at the left posterior axillary line in the fifth intercostal space.
The surgical site should be taken into consideration when placing electrodes on a surgical patient, e.g. for open-chest
surgery, the chest electrodes can be placed on the lateral chest or back. To reduce artifacts and interference from
electrosurgical units, you can place the limb electrodes close to the shoulders and lower abdomen and the chest
electrodes on the left side of the mid-chest. Do not place the electrodes on the upper arm. Otherwise, the ECG
waveform will be very small.
WARNING
z When using electrosurgical units (ESU), place ECG electrodes between the ESU and its grounding plate to
prevent unwanted burns. Never entangle ESU cable and ECG cable together.
z When using electrosurgical units (ESU), never place ECG electrodes near to the grounding plate of the ESU,
as this can cause a lot of interference on the ECG signal.
6-3
6.4.2 ECG Monitoring with Paddles/Pads
1. Prepare the patient’s skin.
If multifunction electrode pads are used, apply pads according to the instructions for use indicated on pads
package. Use anterior-lateral placement.
If external paddles are used, remove the paddle set from the paddle tray by grasping the handles and pulling
them straight up. Apply conductive gel to paddle electrodes. Place the paddles to the patient’s chest using the
anterior-lateral placement.
3. If multifunction electrode pads are used, connect the pads to the pads cable.
1. Place the RA pad or Sternum paddle on the patient’s upper right torso, lateral to the sternum and below the clavicle,
as shown below.
2. Place the LL pad, or Apex paddle to the patient’s left nipple in the midaxillary line, with the center of the electrode
in the midaxillary line, if possible. See the figure below.
Sternum
RA
paddle
LL
Apex paddle
NOTE
z Anterior - lateral placement is the only placement that can be used for ECG monitoring with paddles/pads
accessories.
6-4
6.4.3 Checking Paced Status
It is important to set the paced status correctly when you start monitoring ECG. The paced symbol is displayed in
the patient information area when the [Paced] is set to [Yes]. The pace pulse markers “︱” are shown on the ECG wave
when the patient has a paced signal.
The ECG parameter window to enter the [ECG Setup] menu, and then, select [Others >>]→[Paced],
Warning
z For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the patient monitor could
mistake a pace pulse for a QRS and fail to alarm when the ECG signal is too weak. Do not rely entirely on rate
meter alarms when monitoring patients with pacemakers. Always keep these patients under close
surveillance.
z For non-paced patients, you must set [Paced] to [No]. If it is incorrectly set to [Yes], the equipment may be
unable to detect premature ventricular beats (including PVCs).
Heartbeat icon
Alarm limits
PVC values
HR value
PVCs values is shown only when arrhythmia analysis is switched on. When external paddles or multifunctional electrode
pads are used for ECG monitoring, the PVCs values is shown as “---”.
6-5
6.6 Changing ECG Settings
6.6.1 Changing Lead Settings
6.6.1.1 Selecting Lead Type
1. Select the ECG parameter area to enter the [ECG Setup] menu.
1. Press the Menu button on the equipment’s front panel. In the Main Menu, select [Others >>]→[Configuration
>>]→enter the required password to enter the Configuration Main menu.
2. Select [ECG Setup]→[Lead Set] and toggle between [3-lead] and [5-lead].
2. In the Configuration Main menu, select [ECG Setup]→[ECG Standard], and then select [AHA] or [IEC] according to
the standard that is applied to your hospital.
The QRS should be either completely above or below the baseline and it should not be biphasic.
6-6
6.6.2 Changing ECG Wave Settings
You can select the ECG parameter area to enter the [ECG Setup] menu to set ECG cascade and wave speed. You can also
select the hot keys above the ECG waveform to change ECG lead, size and filter.
You can press the Lead Select button on the equipment’s front panel or use the Navigation knob to select the
lead hot key above the first ECG waveform to select a lead.
If the wave is too small or clipped, you can change its size by adjusting the gain select button on the front panel or
selecting the Size hot key above the ECG waveform.
When monitoring ECG through ECG lead set, filter mode is displayed above the first ECG waveform. The available
filter mode settings are [Monitor], [Therapy], and [Diagnostic]. The default is [Therapy]. To change filter mode,
select the filter mode hot key using the Navigation knob.
When monitoring ECG through pads/paddles, filter mode is always “Therapy” and is not shown..
In the [ECG Setup] menu, select [Sweep], and then choose an appropriate value. The faster the wave sweeps, the
wider the wave is. You can also access [Sweep] to adjust wave speed from the [ECG Setup] menu in the
Configuration mode.
In the [ECG Setup] menu, select [ECG Cascade] and toggle between [On] and [Off].
NOTE
z For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the system may mistake an
internal pace pulse for a QRS or fail to alarm when the pacer is broken.
To switch on or off notch filter, select the ECG parameter area to enter the [ECG Setup] menu, and then select [Others
>>]→[Notch Filter] and toggle between [On] and [Off].
Switching on the notch filter is recommended when there is interference with the waveform.
Set notch frequency according to the electric power frequency of your country. Follow this procedure:
1. Press the Main Menu button on the front panel, and then select [Others >>]→ [Configuration >>]→enter the
required password.
2. Select [ECG Setup]→[Notch Filter] and then select [50Hz] or [60Hz] according to the power frequency.
6-7
6.6.4 Adjusting Heartbeat Volume
In the case that ECG alarm is switched on, or both ECG alarm and PR alarm are switched off, heartbeat tone is issued.
Select the ECG parameter window to enter the [ECG Setup] menu, and then select [Others >>]→[QRS Volume], or
From the Configuration Main menu, select [ECG Setup] →[QRS Volume],
and then select an appropriate setting. The heartbeat volume can be set between 0 and 10, in which 0 means off, and 10
is the maximum volume.
When a valid SpO2 value exists, the system will adjust the pitch of the heartbeat tone according to the SpO2 value.
NOTE
z Arrhythmia analysis program is intended to detect ventricular arrhythmias. It is not designed to detect atrial
or supraventricular arrhythmias. It may incorrectly identify the presence or absence of an arrhythmia.
Therefore, a physician must analyze the arrhythmia information with other clinical findings.
Asystole No QRS complex for 4 consecutive seconds (in absence of ventricular fibrillation Lethal
or chaotic signals). arrhythmia
V-Fib/V-Tach A fibrillatory wave for 4 consecutive seconds.
A dominant rhythm of adjacent Vs and a HR > the V-Tach Heart Rate Limit.
Vtac The consecutive PVCs > Vtac PVCs limit, and the HR > the Vtac HR limit.
Vent. Brady The consecutive PVCs ≥ the Vbrd threshold and the ventricular HR < the Vbrd
Rate threshold.
Extreme Tachy The heart rate is greater than the extreme tachycardia limit.
Extreme Brady The heart rate is less than the extreme bradycardia limit.
PVCs/min PVCs/min exceeds high limit Nonlethal
PNP** No pace pulse detected for 1.75 x average R-to-R intervals following a QRS arrhythmia
complex (for paced patients only).
PNC** No QRS complex detected for 300 milliseconds following a pace pulse (for
paced patients only).
PVC One PVC detected in normal heartbeats.
Couplet Paired PVCs are detected.
VT>2 More than 2 consecutive PVCs within the last minute.
6-8
Arrhythmia event Description Category
When multifunctional electrode pads are used for ECG monitoring, the equipment provides only 4 arrhythmia
alarms, including asystole, ventricular fibrillation/ventricular tachycardia, PNP, and PNC.
When paddles are used, the equipment provides only 3 arrhythmia alarms, including ventricular
fibrillation/ventricular tachycardia, PNP, and PNC.
1. Select the ECG parameter area to enter the [ECG Setup] menu. Select [Arrhythmia >>].
You can also switch arrhythmia analysis on or off in the Configuration mode. Follow this procedure: Enter the
Configuration Main menu. Select [ECG Setup]→[Arrhythmia] and toggle between [On] and [Off].
6-9
6.7.3 Changing Arrhythmia Alarm Settings
To change arrhythmia alarm settings, select the ECG parameter area to enter the [ECG Setup] menu, and then select
[Arrhythmia >>]→[Arrh. Alarm] menu, where you can set alarm switch, alarm level and alarm record switch for all the
arrhythmia events..
You can also set arrhythmia alarm properties in the Configuration mode through the ECG Setup menu.
NOTE
z The alarm level for asystole, ventricular fibrillation, ventricular tachycardia, ventricular bradycardia,
extreme bradycardia, and extreme tachycardia alarms is always high and unchangeable. These alarms are
always on. As long as the alarm condition occurs, corresponding alarm will be triggered whether arrhythmia
analysis is switched on or off.
In case an arrhythmia violates its threshold, an alarm will be triggered. The setting of Asystole Delay is linked to ARR
relearning. When HR is less than 30 bpm, it is recommended to set Asystole Delay to 10 seconds.
You can also set arrhythmia threshold from the ECG Setup menu in the configuration mode.
6-10
6.7.5 Initiating Arrhythmia Relearning Manually
Normally arrhythmia relearning allows the equipment to learn new ECG patterns to correct arrhythmia alarms and heart
rate value. We suggest you to manually initiate arrhythmia relearning when you suspect the result of arrhythmia
analysis.
To initiate relearning manually, select the ECG parameter window to enter the [ECG Setup] menu, select [Arrhythmia
>>]→[Relearn Arrh.]. When the equipment is learning, the message “Learning ECG” is displayed in the technical alarm
area.
NOTE
z Arrhythmia relearning in the case of ventricular tachycardia may affect correct arrhythmia alarm.
1. Select ECG filter hot key. Turn the navigation knob to select [Diagnostic] mode.
2. Select the ECG parameter area to enter the [ECG Setup] menu.
3. Select [Others >>]→[Calibrate]. In this case, a square wave appears on the screen and the message “Calibrating
ECG” is displayed.
4. Compare the amplitude of the square wave with the 1 mV wave scale. The difference should be within 5%.
You can print out the waveform and wave scale and then measure the difference between them if necessary. If the
difference exceeds 5%, contact your service personnel.
6-11
FOR YOUR NOTES
6-12
7 AED
7.1 Overview
This chapter describes how to operate the equipment in AED Mode. While operating in AED Mode, the equipment
analyses the patient’s ECG waveforms and guides you through the defibrillation process.
The equipment starts analyzing the patient’s heart rhythm immediately after entering AED mode. When a shockable
rhythm is detected, the equipment gives a prompt and automatically starts charging. If a shockable rhythm is not
detected, a “No shock advised” prompt is given. Smart defibrillation analysis goes through automated external
defibrillation until the equipment enters CPR or abnormal pads connection occurs.
While operating in AED Mode, the capabilities of the device are limited to those essential to the performance of
semi-automated external defibrillation. Only ECG signals acquired through pads are displayed. Previously set alarms and
scheduled measurements are indefinitely paused and entry of patient information is disabled. Additionally, the Lead
Select, Alarm Pause, and Main Menu buttons are inactive.
7.2 Safety
DANGER
z Defibrillation current can cause operator or bystander severe injury or even death. Never touch the patient
or any equipment connected to the patient (including the bed or gurney) during defibrillation.
z Avoid contact between parts of the patient’s body such as exposed skin of head or limbs, conductive fluids
such as gel, blood, or saline, and metal objects such as a bed frame or a stretcher which may provide
unwanted pathways for the defibrillating current.
z Do not allow multifunction electrode pads to touch each other or to touch other ECG monitoring electrodes,
lead wires, dressings, etc. Contact with metal objects may cause electrical arcing and patient skin burns
during defibrillation and may divert current away from the heart.
z To avoid explosion hazard, do not use the equipment in the presence of oxygen-rich atmospheres,
flammable anesthetics, or other flammable agents (such as gasoline). Keep the equipment and the
operating environment dry and clean.
7-1
WARNING
z During defibrillation, air pockets between the skin and multifunction electrode pads can cause patient skin
burns. To help prevent air pockets, make sure defibrillation pads are completely adhered to the skin.
z In AED mode, this equipment is not designed to administer energy at pediatric joule settings. The American
Heart Association recommends AEDs be used only on patients who are more than eight years old.
CAUTION
z Aggressive handling of multifunction electrode pads in storage or prior to use can damage the pads. Discard
the pads if they become damaged.
z For patients with implantable pacemaker, the sensitivity and specificity of AED algorithm may be impaired.
In AED mode, HR numeric and one ECG waveform acquired from the multifunction electrode pads are displayed. Below
the ECG is the information area which displays the defibrillation mode, prompt message, contact impedance indicator,,
selected energy and a shock counter.
For details about the contact impedance indicator, refer to 8.7 Contact Impedance Indicator.
7-2
7.4 AED Procedure
Confirm that the patient is unresponsive, not breathing or not breathing normally.
Then:
1. Remove clothing from the patient’s chest. Wipe moisture from the patient’s chest and, if necessary, clip or shave
excessive chest hair.
2. Apply multifunction electrode pads to the patient as directed on the pads package. Use anterior-lateral placement.
3. Connect the pads with pads cable, and then plug the pads cable in the equipment’s therapy port.
When the equipment enters AED mode, it checks to see if the pads and pads cable are properly connected. If not, the
message “Connect Pads Cable” or “Apply Pads” will appear in the AED information area until corrective action has been
taken.
Once an ECG is detected through the multifunction electrode pads, the equipment automatically analyzes the patient’s
heart rhythm and warns you not to touch the patient. If a shockable rhythm is detected, the equipment charges
automatically.
You can switch on/off the voice prompt by accessing configuration management or adjust the volume of the voice
prompts by pressing the voice volume soft key.
Once charging is complete, the equipment gives prompt “Do Not Touch Patient! Press Shock Button”. Make sure no one is
touching the patient, bed or any equipment connected to the patient. Call out clearly and loudly “Stay Clear”. Then press
the Shock button on the front panel to deliver a shock to the patient.
Delivery of the shock is confirmed by the voice and screen prompt "Shock Delivered" and the shock counter on the
display is updated to reflect the number of shocks given. If the configured [Shock Series] is greater than one, the
equipment resumes analyzing the patient’s rhythm after the shock is delivered to see if the shock was successful. Voice
and text prompts continue to guide you through additional shocks.
NOTE
z Do not use anterior-posterior pads placement (multifunction electrode pads placed on the patient’s chest
and back). The AED algorithm used by the equipment has not been validated using this placement.
z Motion artifact may delay analysis or affect the ECG signal resulting in an inappropriate shock or no shock
advised message. Keep the patient still during ECG rhythm analysis.
z The Shock button must be pressed to deliver a shock. The equipment will not automatically deliver a shock.
z Impedance is the resistance between the defibrillator’s pads or paddles that the defibrillator must
7-3
overcome to deliver an effective discharge of energy. The degree of impedance differs from patient to
patient and is affected by several factors including the presence of chest hair, moisture, and lotions or
powders on the skin. If the “Impedance too high. Shock not delivered” message appears, make sure that the
patient’s skin has been washed and dried and that any chest hair has been clipped. If the message persists,
change the pads and/or the pads cable.
Heart rhythm analysis continues while the equipment charges. If a rhythm change is detected before the shock is
delivered and a shock is no longer appropriate, the stored energy is removed internally.
Once you are prompted "Do Not Touch Patient! Press Shock Button", if you do not do so within the configured Auto
Disarm time interval, the equipment disarms itself and resumes analyzing.
When the equipment is being charged or have been fully charged, you can remove the charged energy at any time by
pressing the [Pause for CPR] soft key.
The equipment enters CPR status and you will see and hear “No Shock Advised! Paused. If Needed, Begin CPR.” The
remaining pause time is displayed as shown below. The pause period can be defined by setting [CPR Time]
through configuration management.
Analysis resumes at the completion of the pause period or when you press the [Resume Analyzing] soft key in CPR
status.
The equipment continues to monitor the ECG and automatically resumes analysis if a potentially shockable rhythm
is detected. You will hear “No Shock Advised! If needed press pause for CPR”. The message "No Shock Advised!" and
“Monitoring” are shown circularly in the AED information area.
You can define the frequency of these prompts by adjusting [Voice Prompts Interval] through configuration
management. You may press the [Pause for CPR] soft key to suspend monitoring and administer CPR. The pause
period can be defined by setting [CPR Time] through configuration management.
7-4
7.7 CPR
If [Initial CPR Time] is not configured as Off, the system enters initial CPR if AED mode is entered. You can set [Initial
CPR Time] to an appropriate time or switch it off through configuration management.
After the shock series, ECG analysis pauses and the equipment enters the CPR status. Analysis resumes at the completion
of the pause period or when you press the [Resume Analyzing] soft key in CPR status.
In current shock series, the equipment enters the CPR status if you press the [Pause for CPR] soft key after a shock is
delivered. The CPR pause time is defined by setting [CPR Time] through configuration management.
NOTE
z You can start analyzing patient’s heart rhythm again at any time by pressing the [Resume Analyzing] soft
key in CPR status.
When activated, the metronome sounds 120 times per minute and give voice prompts to indicate the rescuers to
perform ventilation at configured compression/ventilation rate.
1. Move the Mode Select knob to Monitor, Manual Defib or Pacer. Press the Main Menu button on the front panel, and
then select→ [Others >>]→[Configuration >>]→Enter the required password.
2. Select [AED Setup] →[CPR Metronome], and toggle between [On] and [Off].
The CPR metronome is activated by default. When CPR metronome is on, you can set the compression/ventilation rate
by selecting [CPR Mode]. The default rate is 30:2.
7-5
Warning
z The CPR metronome sounds do not indicate information regarding the patient’s condition. Because patient
status can change in a short time, the patient should be assessed at all times. Do not perform CPR on a
patient who is responsive or is breathing normally.
NOTE
z CPR metronome and its volume is affected by the settings of [Voice Prompt] and [Voice Volume] in the AED
Setup menu.
1. Move the Mode Select knob to Monitor, Manual Defib or Pacer. Press the Main Menu button on the front panel, and
then select [Others >>]→[Configuration >>]→Enter the required password.
2. Select [AED Setup] →[Voice Recording], and toggle between [On] and [Off].
The symbol is shown at the top right corner of the AED information area if the sounding recording function is
enabled.
The equipment can store up to 180 minutes, maximum 60 minutes for one patient, of sound recording.
2. Select [AED Setup >>] to enter the AED Setup menu, and then change AED settings as desired.
7-6
8 Manual Defibrillation
8.1 Overview
This chapter explains how to prepare for and perform asynchronous defibrillation and synchronous cardioversion using
multifunction electrode pads and external paddles.
In Manual Defib Mode, you must assess the ECG waveforms, decide if defibrillation or cardioversion is indicated, select
appropriate energy setting, charge the equipment, and deliver the shock. Text messages and a contact impedance
indicator on the screen provide relevant information to guide your throughout the defibrillation process.
While operating manual defibrillation, besides ECG you can select to monitor SpO2..
Alarms are turned off automatically when you enter Manual Defib mode. Pressing the Alarm Pause button can turn on
the alarms.
8.2 Safety
DANGER
z Defibrillation current can cause operator or bystander severe injury or even death. Never touch the patient
or any equipment connected to the patient (including the bed or gurney) during defibrillation.
z Avoid contact between parts of the patient’s body such as exposed skin of head or limbs, conductive fluids
such as gel, blood, or saline, and metal objects such as a bed frame or a stretcher which may provide
unwanted pathways for the defibrillating current.
z Do not allow multifunction electrode pads and paddles to touch each other or to touch other ECG
monitoring electrodes, lead wires, dressings, etc. Contact with metal objects may cause electrical arcing and
patient skin burns during defibrillation and may divert current away from the heart.
z During manual defibrillation, make sure your hands are dry and free from conductive gel to avoid shock
hazard.
z Use care when operating this equipment close to oxygen sources (such as bag-valve-mask devices or
ventilator tubing). Turn off gas source or move source away from patient during defibrillation. This can
cause an explosion hazard.
8-1
WARNING
z During synchronous cardioversion, if monitoring patient’s ECG through external paddles, artifact
introduced by paddle movement may resemble an R-wave and trigger a defibrillation shock.
z Do not use conductive liquid. Use only conductive gel specified by the equipment manufacturer.
z If external paddles are used for defibrillation, apply the paddles tightly and evenly to the patient’s chest to
ensure good skin contact.
z Clinicians must select an appropriate energy level for defibrillation of pediatric patients.
CAUTION
z Use of Manual Defib mode may be password protected. Make sure the operator knows and remembers the
password as defined in Configuration. Failure to enter correct password will prevent the delivery of manual
defibrillation therapy.
z Clear the conductive gel from the external paddles at the completion of the therapy to prevent the paddles
from being corroded.
z Prior to using this defibrillator, disconnect from the patient all equipment that is not defibrillator-protected.
z Do not charge the equipment and deliver shock frequently in non-clinical situations. Otherwise equipment
damage could occur.
NOTE
z Impedance is the resistance between the defibrillator’s pads or paddles that the defibrillator must
overcome to deliver an effective discharge of energy. The degree of impedance differs from patient to
patient and is affected by several factors including the presence of chest hair, moisture, and lotions or
powders on the skin. If the “Impedance too high. Shock not delivered” message appears, make sure that the
patient’s skin has been washed and dried and that any chest hair has been clipped. If the message persists,
change the pads and/or the pads cable.
z Alarms are switched off automatically and the “Alarm Off” message is displayed when the equipment enters
the asynchronous defibrillation mode. Alarms remain off until toggled on by pressing the Alarm Pause
button, the Sync mode is entered, or the Mode Select knob is moved to Monitor or Pacer.
8-2
8.3 Manual Defibrillation View
A typical screen in Manual Defib Mode is shown below.
In the enlarged ECG area, an ECG waveform and related parameters are displayed. In the middle of the screen,
defibrillation mode, synchronous icon, prompt message, selected energy, contact impedance indicator, and a shock
counter are displayed.
2. Connect the therapy cable to the therapy port. Push until you hear it click into place.
If multifunction electrode pads are used, apply pads according to the instructions for use indicated on pads
package. Use anterior-lateral or anterior-posterior placement.
If external paddles are used, remove the paddle set from the paddle tray by grasping the handles and pulling
them straight up. Apply conductive gel to the electrode surface of each paddle. Place the paddles to the
patient’s chest using the anterior-lateral placement.
You can access manual therapy directly, by confirmation or by password, which can be defined through
configuration management. The default setting is [Direct].
If [Manual Therapy Access] is set to [Direct], the equipment directly enters Manual Defib mode when the
Mode Select knob is switched to Manual Defib.
If [Manual Therapy Access] is set to [Confirmed], a dialog box pops up when the Mode Select knob is
switched to Manual Defib. You have to confirm whether to enter Manual Defib mode or not.
If [Manual Therapy Access] is set to [Password], a dialog box pops up, requiring the user to enter a password,
when the Mode Select knob is switched to Manual Defib. The equipment enters Manual Defib mode only after
correct password is entered.
5. Select energy.
You can select desired energy level by the adjusting the Energy Select buttons on the equipment’s front panel or
the Energy Select buttons on external paddles if external paddles are used.
Your current energy selection is shown in the defibrillation information area as shown below.
6. Charge
Press the Charge button on the front panel. If external paddles are used, the Charge button on the paddles may be
used instead. As the equipment charges, a progress bar is shown in the defibrillation information area. A charging
tone sounds until desired energy level is reached, when you will hear a charge done tone.
If you have to increase or decrease the selected energy during charging or after charging is complete, adjust the Energy
Select button to select the desired energy level as explained above. Then press the charge button again to restart
charging.
To remove the energy, press the [Disarm] soft key. If the Shock button is not pressed within the specified time period,
the equipment disarms automatically. You can define [Time to Auto Disarm] through configuration management.
7. Shock
Confirm that a shock is still indicated and that the equipment has charged to the selected energy level. Make sure
no one is touching the patient, bed or any equipment connected to the patient. Call out loudly and clearly, “Stay
Clear!”
If pads are used, press the flashing Shock button on the front panel to deliver a shock to the patient.
If external paddles are used, simultaneously press the Shock buttons located on the paddles to deliver a shock
to the patient.
8-4
NOTE
z Defibrillation is always performed through paddles or pads. However, during defibrillation you may choose
to monitor the ECG using an alternate ECG source (3- or 5-lead monitoring electrodes). If an alternate ECG
source is connected, any available lead may be displayed.
z When external paddles are used, the Shock button on the equipment’s front panel is disabled.
To defibrillate, refer to the procedure as described in Section 8.4 Manual Defibrillation Procedure.
3. Connect the paddles to the defibrillator by aligning the white pointer on the paddles cable with the arrow on the
therapy port. Push until you hear it click into place.
4. Select energy by pressing the Energy Select key on the equipment's front panel.
5. Place the conductive surface of paddle electrodes against the patient's right atrium and left ventricle, as shown in
the figure below:
6. Charge the defibrillator by pressing the Charge key on the front panel.
7. Make sure no one is touching the patient or anything connected to the patient.
Using internal paddles for synchronized cardioversion requires that the patient’s ECG be acquired through a standard
ECG cable. The patient’s ECG acquired through the internal paddles may be unreliable for synchronized cardioversion
due to excessive noise or artifact causing inappropriate R-wave detection.
8-5
NOTE
z When internal paddles are used for defibrillation, the energy selection is automatically limited to 50 joules
because of possible cardiac damage from higher energies.
z Sterilize the internal paddles before each use. Otherwise, severe infection may result.
External paddles
To use synchronized cardioversion, press the [Enter Sync] soft key in the asynchronous defibrillation mode. Then “Sync”
appears in the manual Defibrillation information area and a marker appears above each R-wave, see the figure below:
Sync marker
R-wave marker
You can monitor ECG through multifunction electric pads, or external paddles, or electrodes connected to a 3- or 5-lead
ECG cable. Shock is delivered through either pads or paddles.For synchronized cardioversion, we recommend to acquire
patient’s ECG through ECG lead set
NOTE
z When you enter synchronous cardioversion, monitoring alarms is reactivated autonomously.
8-6
8.5.1 Performing Synchronized Cardioversion
1. Connect the therapy cable and apply the multifunction electrode pads or external paddles to the patient. If ECG set
is used for ECG monitoring, connect the ECG trunk cable and apply the ECG electrodes to the patient, referring to 6
Monitoring ECG.
2. With the Mode Select knob in the Manual Defib position, press the [Enter Sync] soft key to activate the
synchronous cardioversion function.
3. Select a lead. The selected lead should have a clear signal and a large QRS complex.
4. Verify that a white R-wave marker appears above each R-wave. If the R-wave markers do not appear or do not
coincide with the R-waves, for example above the T-waves, select another lead.
5. Verify that the equipment enters the Sync mode, as indicated by the SYNC mark shown in the defibrillation
information area.
7. Press the Charge button on the equipment’s front panel or, if using external paddles, the Charge button located on
the handle of Apex paddle.
8. Confirm that a shock is still indicated and that the equipment has charged to the selected energy level. Make sure
no one is touching the patient, bed or any equipment connected to the patient. Call out loudly and clearly, “Stay
Clear!”.
9. Press and hold the Shock button on the equipment or, if using external paddles, the Shock buttons on both
paddles. The shock will be delivered when the next R-wave is detected.
NOTE
z During synchronized cardioversion, it is important to continue to hold the shock button (or the paddle’s
Shock buttons) until the shock is delivered. The equipment shocks with the next detected R-wave.
1. Make sure the equipment is still in Sync mode, as indicated by the presence of the Sync message in the
defibrillation information area.
If [Sync after Shock] is set to [Yes], the equipment remains in the sync mode after a shock is delivered; if set to [No], the
equipment exits the sync mode and enters the asynchronous defibrillation mode after a shock.
8-7
8.6 Remote Synchronized Cardioversion
The equipment can be configured to receive an ECG source from a remote patient monitor (such as a bedside patient
monitor) to perform synchronized cardioversion. To do so, the remote patient monitor shall have a sync out connector
and shall be connected to the defibrillator/monitor’s multifunctional connector with a synchronous cable.
To switch on remote synchronization, select [Manual Defib Setup]→[Remote Sync]→[On] through the Configuration
Main Menu.
1. Use a sync cable to connect the defibrillator/monitor with a bedside monitor through the defibrillator/monitor’s
multifunctional connector.
3. Press the [Enter Sync] soft key. In this case the “Select Sync Mode” dialog pops up.
4. Select [Remote] to enter remote synchronization. Then the message “Remote Sync” is presented.
5. Confirm a square wave on the defibrillator/monitor blinks with each detected R wave on the remote monitor as
shown below, indicating a sync pulse is received.
6. Connect the therapy cable to the therapy port. Push until you hear it click into place.
8. Follow Steps 6 through 9 from 8.5.1 Performing Synchronized Cardioversion as described previously.
8-8
NOTE
z During remote synchronous defibrillation, the local defibrillator/monitor does not display the ECG
waveform. To view the patient’s ECG, check the remote monitor.
z When you use an remote monitor as the ECG source, a biomedical technician must verify that the remote
monitor and the defibrillator/monitor combination will deliver a synchronized shock within 60 ms of the
peak of the R-wave.
indicates the impedance is too high for the defibrillation or the pads or paddles are not
1. In Monitor, Manual Defib or Pacer mode, press the Main Menu button on the front panel, and then select→
[Others >>]→[Configuration >>]→Enter the required password.
WARNING
z It is recommended to perform defibrillation on a patient when the contact impedance is in normal range.
However, you can also perform defibrillation when the contact impedance is slightly higher.
8-9
FOR YOUR NOTES
8-10
9 Noninvasive Pacing
9.1 Overview
In pacer mode, the patient’s ECG is monitored through ECG lead set and pace pulses are delivered through multifunction
electrode pads. The pads cannot be used to monitoring ECG rhythm and deliver pacing current at the same time.
A white pacing marker is shown on the ECG waveform each time a pacer pulse is delivered to the patient. If pacing in
demand mode, white R-wave marker also appears on the ECG waveform until capture occurs.
During pacing, both ECG and SpO2 parameters continue to be monitored and parameter alarms remain active.
In demand mode pacing, a 3-lead or 5-lead ECG cable and electrodes are required to acquire ECG signal. Pace pulses are
delivered through multifunction electrode pads. However, the pads cannot be used to monitor the ECG and deliver pace
pulses simultaneously.
NOTE
z In Pacer mode, arrhythmia analysis is supported and available arrhythmia alarms are asystole, ventricular
vibrillation and ventricular tachycardia.
9.2 Safety
WARNING
z Heart rate displays and alarms function during pacing, but they can be unreliable. Observe the patient
closely while pacing. Do not rely on the indicated heart rate or heart rate alarms as a measure of the
patient’s perfusion status.
z To avoid explosion hazard when pacing a patient who is receiving oxygen delivery, properly route the
oxygen delivery tube. Do not keep it close to the multifunction electrode pads.
9-1
CAUTION
z Use of Pacer mode may be password protected. Make sure the operator knows and remembers the
password as defined in Configuration. Failure to enter correct password will prevent the delivery of pacing
therapy.
z Prolonged noninvasive pacing may cause patient skin irritation and burns. Periodically inspect the
underlying skin and change ECG electrodes and multifunction electrode pads.
NOTE
z If pacing is interrupted for any reason, the [Start Pacing] soft key must be pressed to resume pacing.
z In pacer mode, you cannot change the patient’s internal paced status from the ECG Setup menu.
z In the case that pads poorly contact the patient, the alarm “Pacer Stopped Abnormally” and “Pads Off” may
be presented.
z Pads are not an available choice for the source of ECG waveform in Pacer mode.
In pacer mode, an ECG waveform, related parameters and waveforms are displayed. The pacer information area shows
the pacer mode, pacer rate and output, as well as pacer-related prompts and alarms. Soft keys available for setting
pacing functions are also displayed.
9-2
9.4 Demand Mode versus Fixed Mode
The equipment can deliver paced pulses in either demand or fixed mode.
In demand mode, the pacer only delivers paced pulses when the patient’s heart rate is lower than the selected
pacing rate.
In fixed mode, the pacer delivers paced pulses at the selected rate.
During pacing, you can change pacer mode. Then the equipment continues to deliver pace pulses at selected pacer rate
and pacer output.
CAUTION
z Use demand mode pacing whenever possible. Use fixed mode pacing if noise or artifact interferes with
proper sensing of R-wave or when monitoring electrodes are not available.
z During fixed mode pacing, R-wave markers do not appear on the paced beats.
z During demand mode pacing, spontaneous beats may be presented which are not associated with the
delivery of pace pulse. If the patient’s heart rate is above the pacer rate, pace pulses are not delivered and,
therefore, pacing markers do not appear.
2. Make sure the multifunction electrode pads packaging is intact and within the expiration date shown.
5. If pacing in demand mode, apply monitoring electrodes, and connect the ECG cable to the equipment. To get the
best monitoring signal, make sure there is adequate space between ECG electrodes and therapy electrodes. Refer
to 6.4.1 ECG Monitoring with Electrodes.
9-3
9.5.1 Demand Mode Pacing
To pace in demand mode:
1. Turn the Mode Select knob to the Pacer position. Thus the pacing function is enabled in demand mode
automatically. ECG waveform of Lead II is displayed in the waveform area by default.
You can access manual therapy directly, by confirmation or by password, which can be defined through
configuration management. The default setting is [Direct].
If [Manual Therapy Access] is set to [Direct], the equipment directly enters Pacer mode when the Mode Select
knob is switched to Pacer.
If [Manual Therapy Access] is set to [Confirmed], a dialog box pops up when the Mode Select knob is
switched to Pacer. You have to confirm whether to enter Pacer mode or not.
If [Manual Therapy Access] is set to [Password], a dialog box pops up, requiring the user to enter a password,
when the Mode Select knob is switched to Pacer. The equipment enters Pacer mode only after correct
password is entered.
3. Verify that white R-wave markers appear above the R-waves, as shown in the figure below. If the R-wave markers do
not appear or do not coincide with the R-waves, for example above the T-waves, select another lead.
4. Select pacer rate. If necessary, select initial pacer output. To select pacer rate or pacer output, rotate the Navigation
knob to select an appropriate value and press it to confirm the selection. Do not forget to press the Navigation
knob to exit the setting after desired value is selected.
5. Press the [Start Pacing] soft key to start pacing. The message “Pacing” appears in the pacer information area. .
NOTE
z Pacing will not start if there is a problem with the pads cable connection, pad patient connection, or ECG
monitoring electrodes connection. If either situation occurs, a message will appear in the pacer information
area to alert you that a lead is disconnected or that the pads have a poor connection.
6. Verify that white pacing markers appear on the ECG waveform, as shown below:
Pacing markers
7. Adjust pacer output: increase pacer output until cardiac capture occurs (capture is indicated by the appearance of a
QRS complex after each pacing marker), and then decrease the output to the lowest level that still maintains
capture.
9-4
8. Verify the presence of a peripheral pulse.
You can temporarily withhold pacing pulse and observe the patient’s underlying rhythm by pressing and holding the
[4:1] soft key. This causes pacing pulse to be delivered at 1/4 of the defined pacer rate. To resume pacing at set rate,
release this key.
To stop pacing, press the [Stop Pacing] soft key. Pressing the [Start Pacing] soft key can resume pacing.
CAUTION
z Routinely assess the patient’s cardiac output.
2. Switch the pacer to the Fixed mode. To do so, move the cursor to the Pacer Mode hot key and rotate the Navigation
knob to select [Fix Mode]; then press it to confirm the selection, see the figure below:
Pacer Mode
hot key
3. If ECG electrodes are applied, use the Lead Select button to select the desired lead for viewing.
4. Select pacer rate. If necessary, select pacer output. To select pacer rate or pacer output, rotate the Navigation knob
to select an appropriate value and press it to confirm the selection.
5. Start pacing.
Press the [Start Pacing] soft key to start pacing. The message “Pacing” appears in the pacer information area. .
7. Adjust pacer output: increase pacer output until cardiac capture occurs (capture is indicated by the appearance of a
QRS complex after each pacing marker), and then decrease the output to the lowest level that still maintains
capture.
You can temporarily withhold pacing pulse and observe the patient’s underlying rhythm by pressing and holding the
[4:1] soft key. This causes pacing pulse to be delivered at 1/4 of the defined pacer rate. To resume pacing at set rate,
release this key.
9-5
WARNING
z Use care when handling the multifunction electrode pads on the patient to avoid shock hazard during
pacing.
z If you are using the pacing function with battery power and the Low Battery alarm is presented, connect the
equipment to external power or install a fully charged battery.
CAUTION
z The monitoring or pacing function may be unstable in the presence of ESU or other electronic devices.
9-6
10 Monitoring Resp
10.1 Overview
Impedance respiration is measured across the thorax. When the patient is breathing or ventilated, the volume of air
changes in the lungs, resulting in impedance changes between the electrodes. Respiration rate (RR) is calculated from
these impedance changes, and a respiration waveform appears on the equipment screen.
10.2 Safety
WARNING
z When monitoring the patient’s respiration, do not use ESU-proof ECG cables.
z The respiration measurement does not recognize obstructive and mixed apneas: it only indicates an alarm
when a pre-adjusted time had elapsed since the last detected breath. The safety and effectiveness of the
respiration measurement method in the detection of apnea, especially the apnea of prematurity and apnea
of infancy, has not been established.
RR unit
RR alarm
high limit
RR alarm RR reading
low limit
10-1
10.4 Placing Resp Electrodes
As the skin is a poor conductor of electricity, preparing the skin is necessary for a good Respiration signal. You can refer
to the ECG section for how to prepare the skin.
As the Respiration measurement adopts the standard ECG electrode placement, you can use different ECG cables (3-lead
or 5-lead). Since the respiration signal is measured between two ECG electrodes, if a standard ECG electrode placement
is applied, the two electrodes should be RA and LA of ECG Lead I, or RA and LL of ECG Lead II.
NOTE
z To optimize the respiration waveform, place the RA and LA electrodes horizontally when monitoring
respiration with ECG Lead I; place the RA and LL electrodes diagonally when monitoring respiration with
ECG Lead II.
Lead I Lead II
10-2
10.4.1 Optimizing Lead Placement for Resp
If you want to measure Resp and you are already measuring ECG, you may need to optimize the placement of the two
electrodes between which Resp will be measured. Repositioning ECG electrodes from standard positions results in
changes in the ECG waveform and may influence ST and arrhythmia interpretation.
Cardiac activity that affects the Resp waveform is called cardiac overlay. It happens when the Resp electrodes pick
up impedance changes caused by the rhythmic blood flow. Correct electrodes placement can help to reduce
cardiac overlay. Avoid the liver area and the ventricles of the heart in the line between the respiratory electrodes.
This is particularly important for neonates.
Some patients with restricted movement breathe mainly abdominally. In these cases, you may need to place the
left leg electrode on the left abdomen at the point of maximum abdominal expansion to optimize the respiratory
wave.
In clinical applications, some patients (especially neonates) expand their chests laterally, causing a negative
intrathoracic pressure. In these cases, it is better to place the two respiration electrodes in the right midaxillary and
the left lateral chest areas at the patient’s maximum point of the breathing movement to optimize the respiratory
waveform.
NOTE
z Respiration monitoring is not for use on the patients who are very active, as this will cause false alarms.
Select [Gain] and then choose an appropriate setting. The bigger the gain is, the larger the wave amplitude is.
Select [Sweep] and then choose an appropriate setting. The faster the wave sweeps, the wider the wave is.
In the [Resp Setup] menu, you can also change the [Apnea Alarm Delay] as desired.
10-3
FOR YOUR NOTES
10-4
11 Monitoring PR
11.1 Overview
The pulse numeric counts the arterial pulsations that result from the mechanical activity of the heart. You can display a
pulse from SpO2. The displayed pulse numeric is color-coded to match SpO2 parameter.
PR unit
11-1
FOR YOUR NOTES
11-2
12 Monitoring SpO 2
12.1 Introduction
SpO2 monitoring is a non-invasive technique used to measure the amount of oxygenated haemoglobin and pulse rate
by measuring the absorption of selected wavelengths of light. The light generated in the probe passes through the
tissue and is converted into electrical signals by the photodetector in the probe. The SpO2 module processes the
electrical signal and displays a waveform and digital values for SpO2 and pulse rate.
This device is calibrated to display functional oxygen saturation. It provides the following measurements.
5 6
1. Pleth wave: visual indication of patient’s pulse. The waveform is not normalized.
2. SpO2 unit
5. Oxygen saturation of arterial blood (SpO2): percentage of oxygenated hemoglobin in relation to the sum of
oxyhemoglobin and deoxyhemoglobin.
6. Perfusion indicator: the pulsatile portion of the measured signal caused by arterial pulsation.
12-1
12.2 Safety
WARNING
z Use only SpO2 sensors specified in this manual. Follow the SpO2 sensor’s instructions for use and adhere to
all warnings and cautions.
z When a trend toward patient deoxygenation is indicated, blood samples should be analyzed by a laboratory
co-oximeter to completely understand the patient’s condition.
z Do not use SpO2 sensors during magnetic resonance imaging (MRI). Induced current could potentially cause
burns.
z Prolonged continuous monitoring may increase the risk of undesirable changes in skin characteristics, such
as irritation, reddening, blistering or burns. Inspect the sensor site every two hours and move the sensor if
the skin quality changes. Change the application site every four hours. For neonates or patients with poor
peripheral blood circulation or sensitive skin, inspect the sensor site more frequently.
For Masimo or Nellcor SpO2 modules, you will see corresponding logo on the equipment.
2. Clean the application site, e.g. removing colored nail polish from the application site.
4. Select an appropriate adapter cable according to the connector type and connect it with the equipment.
NOTE
z The SpO2 simulator can be used to check if the SpO2 sensor works properly. However, it cannot be used to
verify the accuracy of the SpO2 sensor.
12-2
12.5 Changing SpO2 Settings
You can access the [SpO2 Setup] menu by selecting the SpO2 parameter area
When the [Sensitivity] is set to [High], the equipment is more sensitive to the changes to SpO2 values, but the
measurement accuracy is relatively low. When monitoring critically ill patients whose pulsations are very weak, it is
strongly recommended that the sensitivity is set to [High].
When the [Sensitivity] is set to [Low], the equipment responds slowly to the changes of SpO2 value, but the
measurement accuracy is relatively high.
For Mindray SpO2 module, set [Sensitivity] to [High], [Med] or [Low] from the [SpO2 Setup] menu, which
respectively correspond to 7 s, 9 s and 11 s.
For Masimo SpO2 module, set [Averaging] to [2-4 s], [4-6 s], [8 s], [10 s], [12 s], [14 s] or [16 s] in the [SpO2 Setup]
menu.
12-3
12.5.4 Sat-Seconds Alarm Management
With traditional alarm management, high and low alarm limits are set for monitoring oxygen saturation. During
monitoring, as soon as an alarm limit is violated, an audible alarm immediately sounds. When the patient’s SpO2 value
fluctuates near an alarm limit, the alarm sounds each time the limit is violated. Such frequent alarm can be distracting.
The Sat-Seconds feature is available with the Nellcor SpO2 module to decrease the likelihood of false alarms caused by
motion artifacts. The Sat-Seconds limit controls the amount of time that SpO2 saturation may be outside the set limits
before an alarm sounds. The method of calculation is as follows: the number of percentage points that the SpO2
saturation falls outside the alarm limit is multiplied by the number of seconds that it remains outside the limit. This can
be stated as the equation: Sat-Seconds= Points × Seconds
Only when the Sat-Seconds limit is reached, the monitor gives a Sat-Seconds alarm. The figure below demonstrates the
alarm response time with a Sat-Seconds limit set at 50 and a low SpO2 limit set at 90%. In this example, the patient’s
SpO2 drops to 88% (2 points) and remains there for 2 seconds. Then it drops to 86% (4 points) for 3 seconds, and then to
84% (6 points) for 6 seconds. The resulting Sat-Seconds are:
Points Seconds Sat-Seconds
2× 2= 4
4× 3= 12
6× 6= 36
Total Sat-Seconds = 52
After approximately 10.9 seconds, a Sat-Second alarm would sound, because the limit of 50 Sat-Seconds would have
been exceeded.
SpO2%
Second
Saturation levels may fluctuate rather than remaining steady for a period of several seconds. Normally, the patient’s SpO2
may fluctuate above and below an alarm limit, re-entering the non-alarm range several times. During such fluctuation,
the system sums the number of SpO2 points, both positive and negative, until either the Sat-Seconds limit is reached, or
the patient’s SpO2 re-enters the non-alarm range and remains there.
12-4
12.5.5 Changing the Speed of the Pleth Wave
In the [SpO2 Setup] menu, select [Sweep] and then select the appropriate setting. The faster the wave sweeps, the
wider the wave is.
You can Press the Main Menu button on the front panel, and then select [Alarm Setup >>]→ [Para. Alarm >>]→[Desat],
or select a parameter window and then select [Para. Alarm >>]→ [Desat] to set the Desat alarm.
NOTE
z In the case that the SpO2 low limit alarm value is set below the Desat limit, the the SpO2 low limit is
automatically adjusted to the Desat value.
There are two pitch tone modes. The select pitch tone mode, select [SpO2 Setup]→[Pitch Tone] from the Configuration
Main menu and toggle between [Mode 1] and [Mode 2].
If the SpO2 is disabled, the pitch tone function will be disabled also.
Ambient light
Diagnostic testing
Low perfusion
Electrosurgical units
12-5
Inappropriate positioning of the SpO2 sensor, or use of incorrect SpO2
Drop of arterial blood flow to immeaurable level caused by shock, anemia, low temperature or vasoconstrictor.
Masimo Patents
This device may be covered by one or more of the following US patents and foreign equivalents: 5,758,644, 6,011,986,
6,699,194, 7,215,986, 7,254,433, 7,530,955.
No Implied License
Possession or purchase of this device does not convey any express or implied license to use the device with
unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or
more of the patents relating to this device.
Nellcor Patents
This device may be covered by one or more of the following US patents and foreign equivalents: 5,485,847, 5,676,141,
5,743,263, 6,035,223, 6,226,539, 6,411,833, 6,463,310, 6,591,123, 6,708,049, 7,016,715, 7,039,538, 7,120,479, 7,120,480,
7,142,142, 7,162,288, 7,190,985, 7,194,293, 7,209,774, 7,212,847, 7,400,919.
No Implied License
Possession or purchase of this device does not convey any express or implied license to use the device with
unauthorized replacement parts which would, alone, or in combination with this device, fall within the scope of one or
more of the patents relating to this device.
12-6
13 NIBP
13.1 Introduction
Automatic non-invasive blood pressure monitoring uses the oscillometric method of measurement. It is intended for
adult, pediatric and neonatal patients. To understand how this method works, we’ll compare it to the auscultative
method.
With auscultation, the clinician listens to the blood pressure and determines the systolic and diastolic pressures. The
mean pressure can then be calculated with reference to these pressures as long as the arterial pressure curve is normal.
Since the equipment cannot hear the blood pressure, it measures cuff pressure oscillation amplitudes. Oscillations are
caused by blood pressure pulses against the cuff. The oscillation with the greatest amplitude is the mean pressure. Once
the mean pressure is determined, the systolic and diastolic pressures are calculated with reference to the mean.
Simply stated, auscultation measures systolic and diastolic pressures and the mean pressure is calculated. The
oscillometric method measures the mean pressure and determines the systolic and diastolic pressures.
NIBP diagnostic significance must be decided by the doctor who performs the measurement.
NOTE
z Blood pressure measurements determined with this device are equivalent to those obtained by a trained
observer using the cuff/stethoscope auscultatory method or an intra-arterial blood pressure measurement
device, within the limits prescribed by the American National Standard, Manual, electronic, or automated
sphygmomanometers.
13.2 Safety
WARNING
z Be sure to select the correct patient category setting for your patient before measurement. Do not apply the
higher adult settings for pediatric or neonatal patients. Otherwise it may present a safety hazard.
z Do not measure NIBP on patients with sickle-cell disease or any condition where skin damage has occurred
or is expected.
z Use clinical judgment to determine whether to perform frequent unattended blood pressure measurements
on patients with severe blood clotting disorders because of the risk of hematoma in the limb fitted with the
cuff.
13-1
z Do not use the NIBP cuff on a limb with an intravenous infusion or arterial catheter in place. This could cause
tissue damage around the catheter when the infusion is slowed or blocked during cuff inflation.
z If you doubt the NIBP readings, determines the patient’s vital signs by alternative means and then verify
that the equipment is working correctly.
Obesity, where a thick layer of fat surrounding a limb dampens the oscillations coming from the artery.
STAT: continually rapid series of measurements over a five minute period, then return to the previous mode.
3. Select a correct sized cuff by referring to the limb circumference marked on the cuff. The width of the cuff should
be 40% (50% for neonates) of the limb circumference, or 2/3 of the upper arm’s length. The inflatable part of the
cuff should be long enough to encircle at least 50% to 80% of the limb.
4. Apply the cuff to an upper arm or thigh of the patient and make sure the Φ or marking on the cuff matches the
artery location. Do not wrap the cuff too tightly around the limb. It may cause discoloration, and ischemia of the
extremities. Make sure that the cuff edge falls within the marked range. If it does not, use a cuff that fits better.
5. Connect the cuff to the air tubing. Avoid compression or restriction of pressure tubes. Air must pass unrestricted
through the tubing.
13-2
Warning
z Continuous CUFF pressure due to connection tubing kinking may cause blood flow interference and
resulting harmful injury to the patient.
2. Select [Interval] and then select a desired time interval. Selecting [Manual] switches to manual mode.
3. Start the measurement manually. The equipment will then automatically repeat NIBP measurements at the set time
interval.
Warning
z Continuous non-invasive blood pressure measurements may cause purpura, ischemia and neuropathy in
the limb with the cuff. Inspect the application site regularly to ensure skin quality and inspect the extremity
of the cuffed limb for normal color, warmth and sensitivity. If any abnormity occurs, move the cuff to
another site or stop the blood pressure measurements immediately.
2. Select [NIBP STAT]. The STAT mode initiates a 5-minute continuous, automatic NIBP measurement.
13-3
13.6 Understanding the NIBP Numerics
The NIBP display shows numerics only as below. Your display may be configured to look slightly different.
1 2
3 7
4
8
5
6
1. Measurement mode
6. Systolic pressure
7. Diastolic pressure
8. Mean pressure at the completion of measurement, or cuff pressure during the measurement
The table below list the setting range of initial cuff inflation pressure:
Patient category Range (mmHg) Default (mmHg) Accuracy (mmHg)
Adult 80 to 280 160 10
Pediatric 80 to 210 140
Neonate 60 to 140 90
NOTE
z Setting initial cuff inflation pressure is disabled during NIBP measurement.
z The initial cuff inflation pressure is restored to the default setting if NIBP module has been reset or patient
category has been changed.
13-4
14 Marking Events
During patient monitoring or therapy, some events may exert effects on the patient and as a result change related
waveforms and parameter values. To help analysing the waveforms or numerics at that time, you can mark these events.
Before marking an event, you can define events A to H, for example, define event D as injecting Atropine. You can only
define an event through configuration management. Event A is always [Generic] and cannot be changed.
To mark an event,
1. In Monitor mode, Manual Defib mode, or Pacer mode, press the [Event] button on the front panel to access the
[Mark Event] menu as shown below.
2. Select an event you want to mark from [A] to [H], or select [Exit] to return to the main screen.
In AED mode, pressing the [Event] button directly marks Event A “Generic”.
When you mark an event, the event name and the time when the event is triggered will be displayed at the prompt
information area. This information disappears automatically after a period of 5 seconds.
14-1
FOR YOUR NOTES
14-2
15 Freezing Waveforms
During patient monitoring, the freeze feature allows you to freeze the currently displayed waveforms on the screen so
that you can have a close examination of the patient’s status. Besides, you can select any frozen waveform for recording.
Waveforms can be frozen only in the Monitor Mode.
At the lower right corner of the bottommost waveform, there is an upward arrow. The frozen time is displayed below the
arrow. With each step or click, the frozen time changes at intervals of 1 second. The time can be applied to all the
waveforms on the screen.
15-1
15.3 Unfreezing Waveforms
To unfreeze the frozen waveforms, you can either:
Perform any other action that causes the screen to be readjusted or opens a menu, such as plugging in or out a
module, pressing the [Lead Select] or [Main Menu] button, etc.
2. Select the [Record] button. The selected waveforms and all numerics at the frozen time will be printed out by the
recorder.
15-2
16 Review
16.1 Reviewing Events
The equipment can automatically record and save patient events. You can review patient events following this
procedure:
In the Monitor, Manual Defib or Pacer mode, press the Main Menu button on the front panel, and then select
[Review>>]→[Review Events >>] to enter the [Review Events] menu, or
In the Monitor mode, repeatedly press the [Trends] soft key to enter the [Review Events] menu.
Select [Event Type], and then select [Operator Initiated], [Phys. Alarm], [Arrhythmia], [Tech. Alarm] or [All] to
review events as desired.
Select [Index] to enter the [Index] menu. In this menu, you can set the time span in which the events happened.
Patient events will be saved as archived events when the equipment is turned off. In case of power failure, the saved
patient events will not be cleared or lost; they will be turned into archived events instead.
16-1
16.2 Reviewing Tabular Trends
In the Monitor, Manual Defib o Pacer mode, press the Main Menu button on the front panel; select
[Review>>]→[Trends>>] or, if you are operating in the Monitor mode, select the [Trends] soft key to enter the tabular
trends menu, as shown below:
16-2
17 Data Management
17.1 Introduction
The data management function enables you to:
To access data management, press the Main Menu button on the front panel to enter the Main Menu, and then select
[Others>>]→ [Archives>>]. A prompt pops up as shown below:
Only in Monitor, Manual Defib and Pacer mode can you access archive mode. When you enter the Archives Main screen,
patient monitoring and therapy automatically end and the patient last admitted will be saved as the latest archived
patient.
17-1
17.2 Reviewing Patient Events
To view patient events, select a patient in the Archives Main screen, and then press the navigation knob to confirm the
selection. In this case, you can select the [Return] soft key to return to the Archives Main screen.
To edit patient information, select the [Patient Info] button and change the patient information as desired. Then you
can select the [Review Events] button to return to the Review Events screen or the [Return] soft key to return to the
Archives Main screen.
1. Select [Export Data] to enter the Export Data screen, in which select [USB Memory]. Then, the system starts
searching for USB memory and enters the data export screen if the memory is found.
2. Select the data you want to export and then press the [Export] button.
During data export, the message “Exporting Data. Please Wait...” appears in the prompt information area and a
progress bar is displayed. If an exception happens, data export stops automatically and the reason for interruption is
presented in the prompt information area.
After the data is completely exported, you can select the [Return] soft key to return to the Archives Main screen.
NOTE
z Do not remove the USB flash memory from the equipment before data is completely exported.
17-2
18 Recording
18.1 Using a Recorder
The thermal recorder records patient information, measurement numerics and up to three waveforms.
2. Event-triggered recordings.
Event recording
Check report
Configuration recording
At the completion of recording, two columns of “*” marks will be printed to indicate the end of recording.
18-1
Automatic recordings will be triggered in the following conditions:
If both [Alarm] and [Alm Rec] for a measurement are switched on, an alarm recording will be triggered
automatically as an alarm occurs.
A recording is completed.
NOTE
z If you change the ECG Lead, Gain or Filter during recording, the recorded ECG waveform changes
accordingly, but the label of Lead, Gain or Filter recorded remains unchanged.
2. Select [Wave Length] and toggle between [8 s], [16 s] and [32 s].
[8 s]: record a waveform 4 seconds before and 4 seconds after the current moment;
[16 s]: record a waveform 8 seconds before and 8 seconds after the current moment;
[32 s]: record a waveform 16 seconds before and 16 seconds after the current moment;
18-2
18.4.4 Changing the Recording Speed
1. Enter the [Record Setup] menu.
2. Select [Paper Speed] and toggle among [6.25 mm/s], [12.5 mm/s], [25 mm/s] and [50 mm/s].
4. Check if paper is loaded correctly and the paper end is feeding from the top.
Recorder
CAUTION
z Use only specified thermal paper. Otherwise, it may cause damage to the recorder’s print head, the recorder
may be unable to print, or poor print quality may result.
z Never pull the recorder paper with force when a recording is in process. Otherwise, it may cause damage to
the recorder.
z Do not leave the recorder door open unless you have to reload paper or remove troubles.
18-3
18.6 Removing Paper Jam
If the recorder works incorrectly or produces unusual sounds, check if there is a paper jam first. If a paper jam is detected,
follow this procedure to remove it:
2. Take out the paper and tear off the draped part.
1. Take measures against the static electricity, such as disposable wrist strap, for the work.
3. Gently wipe around the print head using cotton swabs dampened with alcohol.
4. After the alcohol has completely been dried, reload the paper and close the recorder door.
CAUTION
z Do not use anything that may destroy the thermal element.
18-4
19 Configuration Management
19.1 Introduction
Configurations management enables you to customize you equipment to best meet your needs. With this function, you
can:
After the system configurations have been changed, the equipment restarts and new configuration settings take effect
immediately.
19.2 Password
Accessing configuration management is password protected. The required password is set to 315666 before the
equipment leaves the factory.
1. Press the Main Menu button on the front panel of the equipment. Select [Others >>]→ [Configuration >>], a
dialog box pops up as shown below:
To view the configurations or change system time, select [View Config]. In this case, password is not required.
To close the dialog box and return to normal operating mode, select [Cancel].
19-1
2. Enter the required password and then select [OK] to enter the Configuration Main menu as shown below:
Selecting [Factory Config] and confirming the selection restores all the current settings to factory default settings:
WARNING
z Never connect the equipment with the patient while performing configuration management.
Pads Default 2, 5, 10, 50, 100, 150, 170, 200, 300 J 200 J /
19-4
Menu Item Options/Range Default Remark
Arrhythmia On, Off Off /
ARR Alarm On, Off On /
ARR Alm Lev PVCs/min High, Med, Low Med /
R ON T
VT>2
Couplet
PVC
Vent. Rhythm
Bigeminy
Trigeminy
Tachy
Brady
PNP
PNC
Missed Beat
Multif. PVCs
Nonsus. Vtac
Pause
Irr. Rhythim
Asystole Delay 3 to 10 5 /
V-Tach Rate 100 to 200 130 /
V-Tach PVCs 3 to 99 6 /
Vbrd Rate 15 to 60 40 /
Vbrd PVCs 3 to 99 5 /
Extreme Adu 60 to 300 160 /
Tachy Ped 60 to 300 180 /
Extreme Adu 15 to 120 35 /
Brady Ped 15 to 120 50 /
PVCs High 1 to 10 10 /
Adu 60 to 300 120 /
Tachy
Ped 60 to 300 160 /
Adu 15 to 120 50 /
Brady
Ped 15 to 120 75 /
Multif. PVCs Window 3 to 31 15 /
Pause Time 1.5, 2.0, 2.5 2.0 /
19-5
19.3.6 Resp Setup Menu
Menu Item Options/Range Default Remark
RR Alarm On, Off Off /
RR Alm Lev High, Med, Low Med /
Sweep 6.25/12.5/25 mm/s 6.25mm/s /
Adu 30 /
(Low+2) to 100
RR High Ped 30 /
Neo (Low+2) to 150 100 /
Adu 8 /
RR Low Ped 6 to (High-2) 8 /
Neo 30 /
Apnea Time 10/15/20/25/30/35/40s 20s /
19-6
Menu Item Options/Range Default Remark
Adu 50
PR Low Ped 25 to (High-2) 75
Neo 100
This setting is linked with the [QRS
QRS Volume 0 to 10 2 Volume] setting in the [ECG Setup]
menu
NIBP-Mean
Alm Lev High, Med Med /
Adu (Low + 5) to 270 160
Sys High Ped (Low + 5) to 200 120 /
Neo (Low + 5) to 135 90
Adu 90 /
Sys Low Ped 40 to (High - 5) 70 /
Neo 40 /
Adu (Low + 5) to 230 110 /
Mean High Ped (Low + 5) to 165 90 /
Neo (Low + 5) to 110 70 /
Adu 60 /
Mean Low Ped 20 to (High - 5) 50 /
Neo 25 /
Adu (Low + 5) to 210 90 /
Dia High Ped (Low + 5) to 150 70 /
Neo (Low + 5) to 100 60 /
Adu 50 /
Dia Low Ped 10 to (High - 5) 40 /
Neo 20 /
19-7
19.3.10 Alarm Setup Menu
Menu Item Options/Range Default
19-8
19.3.12 Mark Event Setup Menu
Menu Item Options/Range Default Remark
Event B Adrenalin
Adrenalin, Lidocaine, Atropine, Nitroglycerin, Event names that have been
Event C Lidocaine
Morphine, Intubation, IV Access, Adenosine, selected by previous events will
Event D Atropine
Amiodarone, Vasopressin, Isoprenaline, not be included in the options of
Event E Nitroglycerin
Dopamine, Aspirin, Oxygen, CPR later events.
Event F Morphine
Options: entering event name using the
Custom Event 1 After being defined, custom event
keyboard included in the [Mark Event Setup]
/ is refreshed in real-time in the
menu.
Custom Event 2 Mark Event list.
Range: 1 to 20 characters
19-9
19.3.15 User Test Setup Menu
Menu Item Options/Range Default
User Test Prompt On, Off Off
NIBP
19-10
20 Battery
20.1 Introduction
The equipment is designed to operate on battery power when external power supply is not available. The battery is
charged whenever the equipment is connected to AC mains or the DC power supply through an external DC/AC adapter,
regardless of whether or not the equipment is currently turned on. In case of power failure, the equipment will
automatically run power from internal batteries. So we recommend you always install a fully charged battery in the
equipment.
The equipment is configured with one smart lithium ion batteries which are free of maintenance.
On-screen battery symbols indicate the current battery charge status, taking Battery 1 for an example:
≤ 20% of capacity
If the battery charge is too low, a technical alarm will be triggered and the “Low Battery” message displayed in the
Technical Alarm Area. At this moment, change the battery or apply external power to the equipment.
WARNING
z Keep the batteries out of children’s reach.
z The batteries should be charged in this equipment or in a device approved by the equipment manufacturer.
NOTE
z Always connect the equipment to AC mains whenever it is possible.
z After long term use, the power capacity indicated by the battery symbol may be different from the actual
capacity. Always observe the alarm information displayed on the screen.
20-1
20.2 Installing the Batteries
To install the batteries:
2. Insert the battery, and press until you hear it click into the place.
To replace a battery, press the latch on the battery and push the battery to the right until you remove it. Insert a new
battery into the battery compartment.
Press the Main Menu button on the front panel of the equipment. Select [Others >>]→ [Configuration >>]→enter the
required password. In the Configuration main menu, select [Alarm Setup]→ [No Battery], and toggle between [Status
Indicator On] and [Status Indicator Off]. The default setting is [Status Indicator Off].
If [Status Indicator On] is selected, the red cross status indicator flashes in the case that battery is not installed,
meanwhile a message “No Battery” is presented in the Technical Alarm Area.
If [Status Indicator Off] is selected, the red cross status indicator will not be illuminate in the case that battery
is not installed; the message “No Battery” will be presented in the Technical Alarm Area.
If the battery is almost depleted, a prompt “Battery depleted! System will shut down imminently. Connect to AC
mains or replace battery” pops up. In monitoring, manual defibrillation, and pacing mode, additional alarm lights and
alarm tones are provided. In this case, take appropriate actions immediately. This prompt will not disappear until the
battery is replaced or the equipment is connected to the external power supply. The equipment automatically shuts
down if no action is taken within a period of about 3 minutes.
NOTE
z The Low Battery alarm means that the battery is beginning to weaken and should be replaced at the first
opportunity. At least 20 minutes of monitoring and six full energy shocks can be performed when the Low
Battery alarm is activated. Replace the battery or connect the equipment to AC mains as soon as possible.
20-2
20.3.3 Battery Aged Alarm
If the battery runtime is significantly shorter than the specification, a low level technological alarm “Battery Aged” will
be presented. We recommend you to contact our company and replace it with a new one.
1. Connect the equipment to the external power supply and allow the batteries to charge uninterruptedly till it is fully
charged.
2. Remove the external power supply and allow the equipment to run from the batteries until it shuts off.
The operating time of the batteries reflects their performance directly. If the operating time of the batteries is noticeably
shorter than that stated in the specifications, discard the batteries or contact your service personnel.
NOTE
z Life expectancy of a battery depends on how frequent and how long it is used. When properly cared for, the
lithium-ion battery has a useful life of approximately 2 years. For improper use models, life expectancy can
be less. We recommend replacing lithium-ion batteries every 2 years.
z To optimize performance, a depleted, or nearly depleted battery should be charged as soon as possible.
Batteries should be charged at temperatures between 0°C (32°F) to 45°C (113°F). To optimize performance, a fully (or
nearly fully) discharged battery should be charged as soon as possible.
20-3
20.6 Storing Batteries
When storing batteries, make sure that the battery terminals do not come into contact with metallic objects. If batteries
are stored for an extended period of time, they should be placed in a cool place with a partial charge of 40% to 60%
capacity (3 LEDs illuminated). Storing batteries in a cool place slows the aging process. The idea storage temperature is
15°C (60°F). Batteries should not be stored at temperature outside the range of -20°C (-4°F) to 60°C (140°F).
Remove the battery from the equipment if the equipment is not used for a prolonged time. Otherwise the battery may
be over discharged and it will take significantly longer time to charge the battery. Stored batteries should be charged
every 2 months to 40% to 60% of full capacity. They should be charged to full capacity prior to use.
NOTE
z Do not store a battery in the equipment if the equipment is not used for an extended period of time.
z Storing batteries at temperature above 38°C(100°F) for an extended period of time significantly shorten the
life expectancy of a battery
WARNING
z Do not disassemble, puncture or incinerate batteries. Do not short the battery terminals. They may ignite,
explode, or leak, causing personal injury.
20-4
21 Care and Cleaning
Use only the substances approved by the equipment manufacturer and methods listed in this chapter to clean or
disinfect your equipment. Warranty does not cover damages caused by unapproved cleaning and disinfection
substances or methods.
We make no claims regarding the efficacy of the listed chemicals or methods as a means for controlling infection. For the
method to control infection, consult your hospital’s infection control officer or epidemiologist.
In this chapter we only describe cleaning and disinfection of the main unit. For the cleaning and disinfection of external
paddles and other reusable accessories, refer to instructions for use of corresponding accessories.
Always dilute according the manufacturer’s instructions or use lowest possible concentration.
Keep the paddles clean, Before user checks or after each use, thoroughly clean the paddles and paddle tray.
Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone or
acetone-based cleaners).
WARNING
z Be sure to shut down the system, disconnect the power cord and other cables, and remove the
batteries before cleaning the equipment.
CAUTION
z Contact your service personnel in case of spilling liquid on the equipment or accessories.
NOTE
z To clean or disinfect reusable accessories, refer to the instructions for use delivered with the
accessories.
21-1
21.2 Cleaning
Your equipment should be cleaned on a regular basis. If there is heavy pollution or lots of dust and sand in your place,
the equipment should be cleaned more frequently. Before cleaning the equipment, consult your hospital’s regulations
for cleaning the equipment.
Ethanol (70%)
Isopropanol (70%)
1. Shut down the equipment, disconnect the power cord and other cables, and remove the batteries.
2. Clean the display screen using a soft, clean cloth dampened with a glass cleaner.
3. Clean the exterior surface of the equipment using a soft, clean cloth dampened with a glass cleaner.
4. Clean the paddle tray using a soft, clean cloth dampened with a glass cleaner.
5. Wipe off all the cleaning solution with a dry cloth after cleaning if necessary.
21.3 Disinfecting
Disinfection may cause damage to your equipment and is therefore not recommended unless otherwise indicated in
your hospital’s servicing schedule. Cleaning equipment before disinfecting is recommended.
The recommended disinfectants include: ethanol 70%, isopropanol 70%, and Perform* classic concentrate OCY (KHSO4
solution).
21-2
22 Maintenance and Testing
WARNING
z Failure for the responsible individual, hospital or institution employing this equipment to implement a
satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
z The safety checks or maintenance involving any disassembly of the equipment should be performed by
professional servicing personnel. Otherwise, undue equipment failure and possible health hazards could
result.
z If you find a problem with any of the equipment, contact your service personnel or the manufacturer.
22.1 Overview
Before every use, each shift or once a week, checks shall be done to ensure that the equipment is ready for operation at
any time. After the equipment has been used for 12 months, or whenever the equipment is repaired or upgraded, a
thorough inspection should be performed by qualified service personnel to ensure the reliability.
Shift check
Recorder inspection
User test
Pacing test
Paddles and cables are critical parts for defibrillation but damageable. We recommend you to check the appearance and
performance of these parts every day and replace them every 3 years.
The ECG cables are critical parts for data acquisition and analysis but damageable. We recommend you to inspect the
cable as described in the Appendix C BeneHeart Defibrillator Shift Checklist.
In case of any damage or abnormity, remove the equipment from use. Contact the hospital’s biomedical engineers or
your service personnel immediately.
22-1
22.2 Maintenance and Testing Schedule
The following tests, except recorder check and user test, shall be carried out by the service personnel only. Contact your
service personnel if any maintenance is required. Make sure to clean and disinfect the equipment before any test and
maintenance.
Routine Test ×
Controls test ×
Recorder check ×
Charge/discharge
Manual defibrillation
Energy disarming ×
tests
Synchronous defibrillation
Pacing test ×
Accuracy test
NIBP tests ×
Leakage test
Power module,
Therapy module,
Monitoring function.
You need to power on the equipment every day, or when the equipment is first installed, or following any maintenance
or the replacement of any main unit parts to verify that the equipment can be turned on properly.
1. Place the external paddles on paddle tray, insert the battery in the battery compartment, and then connect the
equipment with AC mains. In this case, both the AC indicator and battery indicator shall light.
2. Turn the Mode Select knob to Monitor. Check that the equipment passes the self test and is turned on properly.
22-2
3. Check the display of technical alarm area, prompt area and battery status indicator on the upper right corner of the
main screen to judge whether the equipment runs normally.
The following table provides the description of automated test and lists the frequency with which each is performed.
Automated Routine Test can be initiated between 0:00 am to 5:00 am. To set auto test time, enter the Configuration Main
men, and select [Test Setup]→ [Auto Test Time]. The default setting is 3:00 am.
At the completion of Auto Test, a report is saved automatically. Enter the Configuration Main menu→ select [Record
Setup]→ [Auto Test Report] →select [On], [Off], or [Only if Failed], you can choose to print the Auto Test Report or not
after the tests are completed, or print it only if Auto Test Failed.
You can review the result of Auto Test by selecting the [History] softkey from the User Test Main menu.
CAUTION
z With power off, auto Test is performed only when AC mains is connected.
z Thoroughly clean the paddles and properly place them in the paddle tray after each use. Automated test
passes only when paddles properly contact the metal parts of the paddle tray.
z Install the battery and properly place the external paddles in the paddle tray or connect the pads cable and
50 Ω test load. Otherwise the Auto Test will fail.
22-3
22.3.4 User Test
WARNING
z Do not perform user test when a patient is connected to the equipment.
Routine Test,
Controls test.
NOTE
z Before user test or after each use, thoroughly clean the paddles and properly place them in the paddle tray.
User test passes only when paddles properly contact the metal parts of the paddle tray.
z Install the battery and properly place the external paddles in the paddle tray or connect the pads cable and
50 Ω test load. Otherwise the User Test will fail.
To access user test, press the Main Menu button and select [User Test >>]. Then a dialog box pops up, prompting you
that entering user test will end patient monitoring. Select [Yes] to enter the User Test Main menu.
Check the test items you want to perform and select [Start] to start user test. The message “Test completed” will be
presented when selected tests have been finished. Then you can press the [Record] soft key to print the test result.
22-4
22.3.4.2 Routine Test
Batteries,
Mainboard,
Monitor function
If any of above items fails, the red status indicator will flash. If mainboard, Defib/Pacer function, or monitor function fails,
a low level technical alarm “Last User Test Failed” will be displayed in the technical alarm area when the equipment is
restarted. We recommend you to perform a successful User Test to clear this alarm.
Energy delivery test delivers a 200J external discharge to check the defibrillation circuit.
If the test fails, the red status indicator will flash and a low level technical alarm “Last User Test Failed” will be displayed in
the technical alarm area when the equipment is restarted. We recommend you to perform a successful User Test to clear
this alarm.
Display test
NOTE
z The tested controls are indicated in green during controls test.
If any of above items fails, the red status indicator will be illuminated and a low level technical alarm “Last User Test
Failed” will be displayed in the technical alarm area when the equipment is restarted. We recommend you to perform a
successful User Test to clear this alarm.
NOTE
z The“Off” position of the Mode Select knob is not tested during the Controls Test. If you turn the knob to
“Off” for more than 3 seconds. The equipment is turned off.
22-5
22.3.4.5 Test Summaries
The results of User Test are automatically saved as summaries. You can select the [History] button from the User Test
Main menu to review the test summaries.
The equipment can store up to 300 historical test summaries which are listed in the sequence of time, with the latest on
the top. You can use the knob to select a test result and press the knob to check the detailed test report.
We recommend that Routine Test and Energy Delivery Test are executed once per week and Controls Test once every 12
months.
Each time when the equipment is turned on, the time to last Routine Test, Energy Delivery Test and Controls Test is
checked automatically. The equipment can be configured to give a “User Test Due” message to remind you that the
equipment is due for User Test.
The [User Test Prompt] is switched off by default. You can switch it on by selecting [User Test Setup]→[User Test
Prompt]→[On] through the configuration mode.
NOTE
z In case that automated Routine Test is not performed, we recommend you to run Routine Test once per day
through User Test.
2. Start recording to verify that the recorder works properly and the printout is legible and correct.
3. Simulate errors, such as removing the paper roll and losing the latch, correct information shall be shown in the
prompt area. The recorder shall work properly after the faults are corrected.
22-6
22.3.6 ECG Cable Test
It is recommended to perform ECG cable test once a year.
2. Connect the ECG cable to the defibrillator and the electrodes to the simulator.
4. Wait for a few seconds and check that the waveform is displayed normally and that no lead-off alarms reported in
the alarm information area.
Charge/discharge
1. Remove the batteries and connect the equipment with AC mains. Turn the Mode Select knob to Manual Defib.
2. Connect the external paddles to the equipment and place the paddles on the defibrillator/pacer analyzer.
3. Enter the Configuration Main screen. From the Record Setup menu set [Shock Event] to [On] so that shock events
can be recorded automatically if happened.
4. Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as 0 or blank.
6. Charge/discharge the equipment to verify the energies measured by the analyzer meet the following accuracy:
8. Disconnect the equipment from the AC mains. Run the equipment on fully charged battery. Move the Mode Select
knob to Manual Defib. Repeat Steps 2 to 7.
9. Verify that the equipment records the shock events automatically and correctly.
Energy Disarming
1. Run the equipment on fully charged battery. Move the Mode Select knob to Manual Defib.
2. Connect the external paddles to the equipment and place the paddles on the defibrillator/pacer analyzer.
3. Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as 0 or blank.
8. Verify that a prompt “Charge Removed” appears on the screen and the charge done tone stops.
10. Enter the Configuration Main menu, select [Manual Therapy Setup] and set [Time to Auto Disarm] to [60s].
12. Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as 0 or blank.
14. Charge the equipment. Count time after charging is completed. Verify that the prompt “Shock Removed” appears
on the equipment and the energy measured by the analyzer is 0J or blank after 60 seconds.
Synchronous Defibrillation
1. Connect the external paddles and ECG cable to the equipment. Place the paddles ECG electrodes on the
defibrillator/pacer analyzer.
2. Set the analyzer to Time Measurement Mode and output normal sinus rhythms, e.g. amplitude value 1mV and HR
60bpm.
3. Enter Configuration Management. In the [Manual Therapy Setup] menu, set [Sync After Shock] to [On].
5. Press the [Enter Sync] soft key to start synchronous defibrillation. If Remote Sync is switched on, press the [Enter
Sync] soft key and select [Local] to start synchronous defibrillation
7. When charging finishes, press and hold the “Shock” button to deliver a shock.
8. Verify that synchronous discharge succeeds and the delivered energy measured by the analyzer is 10J±2J.
9. Verify that the delay time of synchronous defibrillation measured by the analyzer is less than 60ms.
10. Verify that the synchronous discharge mark appears on the R wave.
11. Verify that the prompt messages are correct during testing.
12. Select lead II as ECG source and perform charging. Repeat steps 7 to 11.
22-8
22.3.8 Pacing Test
Test tools: defibrillator/pacer analyzer
1. Run the equipment on fully charged battery. Move the Mode Select knob to Pacer. Select Fixed mode..
2. Connect the pads cable to the equipment and properly place the pads on the defibrillator/pacer analyzer.
3. Set the analyzer to Pacing Measurement mode. Use test load of 50Ω.
4. On the equipment, set [Pacer rate] to [70ppm] and [Pacer Output] to [30mA].
5. Press the [Start Pacing] soft key. Verify that the pacer rate measured by the analyzer is 70 ppm±1ppm and the
pacer output measured is 30 mA±5mA.
6. Press the “Stop Pacing” soft key, and then set [Pacer rate] to [170ppm] and [Pacer Output] to [200mA].
7. Press the [Start Pacing] soft key. Verify that the pacer rate measured by the analyzer is 170 ppm±2ppm, and the
measured current is 200 mA±10mA.
Accessing installation mode is password protected. The required password is set to 888888 before the equipment leaves
the factory.
You can access Installation Mode while operating in the Monitor, Manual Defib or Pacer mode. Patient monitoring and
therapy automatically end when you enter Installation Mode.
To enter Installation Mode, press the Main Menu button on the front panel, and then select [Others >>]→[Maintenance
>>]→[Installation Mode >>]→enter the required password. The Installation Mode Main menu pops up as shown
below.
22-9
22.3.9.3 NIBP Accuracy Test
The NIBP accuracy test is required at least once every two years or whenever you doubt the NIBP reading.
Tools required:
T-shape connector
Tubing
Balloon pump
Manometer
Defibrillator/monitor
Connector for Tubing
NIBP cuff
2. Before inflation, the reading of the manometer should be 0. If not, disconnect the airway and reconnect it until the
readings is 0.
3. In the Installation Mode Main menu, select [Maintain NIBP]→ [Start Accuracy Test].
4. Compare the value of manometer with the value displayed on the equipment’s screen. The difference should be no
greater than 3 mmHg.
5. Raise the pressure in the metal vessel to 50 mmHg with the balloon pump. Repeat steps 3 and 4.
6. Raise the pressure in the metal vessel to 200 mmHg with the balloon pump. Repeat steps 3 and 4.
If the difference between the values of the manometer and defibrillator/monitor is greater than 3 mmHg, contact your
service personnel.
When you select the [Start Accuracy Test] button, it turns to be [Stop Accuracy Test]. Select [Stop Accuracy Test],
accuracy test stops and the button turns to be [Start Accuracy Test] again.
22-10
22.3.9.4 NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required at least once every two years or
whenever you doubt the NIBP reading.
Tools required:
An adult cuff
An air tubing
Defibrillator/monitor Cylinder
4. In the Installation Mode Main menu, select [Maintain NIBP]→ [Start Leakage Test].
After about 20 seconds, the equipment automatically deflates. This means the leakage test is started.
When the accuracy test is completed, the result will be displayed. If the message “NIBP Pneumatic Leak” is displayed, it
indicates that the NIBP airway may have leakages. Check the tubing and connections for leakages, and then perform a
leakage test again.
If the problem persists, contact your service personnel.
When you select the [Start Leakage Test] button, it turns to be [Stop Leakage Test]. Select [Stop Leakage Test],
leakage test stops and the button turns to be [Start Leakage Test] again.
To view version information, select [Version] in the Installation Mode Main menu. In the popup menu, you can view
system software version and module software version.
22-11
22.3.9.6 Formatting Storage Card
You can format the storage card if data in the card is useless, or if the card has a failure. To format the storage card, select
[Format Data Card]→ [Format] through the Installation Mode Main menu. Then a dialog box pops up as shown below:
If storage card is formatted successfully, a prompt “Formatting is completed!” appears. If there is a failure, the system
stops formatting and present a prompt “Formatting fails!”. Contact the service personnel if formatting fails for 3 times”.
1. Open the equipment’s enclosure, remove the multi-parameter monitoring module (M51A module), disconnect the
air tube from the pressure measuring sensor (component U26 on the digital board), and then block the tube.
3. Press the NIBP hard key on the equipment’s front panel to start NIBP measurement. When the pressure exceeds the
hardware overpressure protection value (300 to 330mmHg), verify that the valve opens to release the air and the
alarm “NIBP Cuff Overpress” is shown on the screen.
The test is passed if you hear the valve discharge the gas properly and the “NIBP Cuff Overpress” alarm is given.
Otherwise, contact the service personnel.
22-12
23 Accessories
WARNING
z Use accessories specified in this chapter. Using other accessories may cause damage to the equipment or
not meet the claimed specifications.
z Single-use accessories are not designed to be reused. Reuse may cause a risk of contamination and affect
the measurement accuracy.
z Check the accessories and their packages for any sign of damage. Do not use them if any damage is
detected.
z At the end of its service life, the equipment, as well as its accessories, must be disposed of in compliance
with the guidelines regulating the disposal of such products to avoid contaminating the environment.
z When using the accessories, consider the accessories’ operating temperature. Refer to corresponding
accessory’s instrution for use for details.
z The accessory material that contact the patients have undertaken the bio-compatibility test and is verified
to be in compliance with ISO 10993-1.
23-1
Lead Sets
Adapting Cable
23-2
23.2 SpO2 Accessories
Extension Cables
SpO2 Sensors
23-3
Masimo SpO2 Module
Type Model Applicable patient PN Remark
FPS-1901 Pediatric, neonate (wrap type) 0010-10-42626 LNCS-NeoPt-L
FPS-1862 Neonate (wrap type) 0010-10-42627 LNCS-Neo-L
Disposable FPS-1861 Infant (wrap type) 0010-10-42628 LNCS-Inf-L
FPS-1860 Pediatric (wrap type) 0010-10-42629 LNCS-Pdt
FPS-1859 Adult (wrap type) 0010-10-42630 LNCS-Adt
FPS-1863 Adult (finger clip) 0010-10-42600 LNCS DC-I
Reusable FPS-1864 Pediatric (finger clip) 0010-10-42634 LNCS-DCIP
2258 Adult, pediatric, neonate 0010-10-43016 LNCS YI
Wavelength of Mindray 518B, 512E, 512F, 512G and 512H SpO2 sensors: red light 660 nm, infrared light 905 nm.
Wavelength of Masimo SpO2 sensors: red light: 660 nm, infrared light: 940 nm.
Wavelength of Nellcor SpO2 sensors: red light: 660 nm, infrared light: 890 nm.
The maximum photic output consumption of the sensor is less than 18 mW.
The information about the wavelength range and maximum photic output consumption can be especially useful to
clinicians, for example, clinicians performing photodynamic therapy.
23-4
23.3 NIBP Accessories
Tubing
Cuff
23-5
23.5 Miscellaneous
Description Model PN
23-6
A Specifications
A.1 General Specifications
Class I, equipment energized from an external and internal electrical power source.
Type of protection against
If you suspect the integrity of the external protective earthing or the protective earthing
electrical shock
wire, you should run the equipment on internal electrical power supply (battery).
electric shock Type CF defibrillation proof for ECG, SpO2, NIBP and internal defibrillation
Size
Width×depth×height,
288×203×275 mm
without external paddles
Weight
Main Unit 4.7 kg
Battery package 0.5 kg
External paddle set 0.8 kg
Display
Type TFT Color LCD
Size 7 inch
Resolution 800×480 pixels
Viewed waveforms Max. 3
Wave viewing time Max. 16s (ECG)
Equipment connectors
USB connector Connects USB flash memory
Multifunctional connector Connects a cable for defibrillator synchronization.
RJ45 connector Connects standard network cable.
A-1
Audio Indicator
Gives alarm tones (45 to 85 dB), key tones, QRS tones;
Speaker Supports PITCH TONE and multi-level tone modulation;
Alarm tones comply with IEC60601-1-8.
Multifunctional connector
Standard Meets the requirements of EN60601-1 for short-circuit protection and leakage current
Output impedance Typically 50Ω
Synchronous input
Input signal range 0 to 5V
Input impedance ≥10 kΩ
Pulse width >5 ms
A-2
360 J defibrillation waveform into impedance of 25Ω, 50Ω, 75Ω, 100Ω, 125Ω, 150Ω, 175Ω
Voltage (V)
Time (MS)
A-3
Charge time (Note: at 20 °C of ambient temperature)
Manual Defib AED
From initiation of
From initial power From initial power on to
Charge time rhythm analysis to
on to charge done charge done
charge done
200J 360J 200J 360J 200J 360J 200J 360J
With a new, fully
<5 s <8 s <11 s <14 s <16 s <21 s <21 s <26 s
charged battery
With a new, fully
charged battery,
<6 s <9 s <12 s <15 s <17 s <22 s <23 s <27 s
depleted by 15 360 J
discharges
AED
Energy level: 100 to 360J, configurable;
Shock series Shocks: 1, 2, 3, configurable;
Meeting AHA guidelines 2010 by default.
Shockable rhythm VF, VT (HR>150bpm and QRS width>120ms)
A-4
A.3 Pacer Specifications
Pacing mode Demand, fixed
Monophasic square wave pulse
Output waveform pulse width 20 ms
Accuracy: ±5%
40ppm to 170ppm
Pacing rate Accuracy: ±1.5%
Resolution: 5 ppm
0mA to 200mA,
Pacing output Accuracy: ±5% or ±5mA, whichever is greater
Resolution: 5mA
200 to 300 ms (depending on pacing rate)
Refractory period
Accuracy: ±3%
4:1 pacing Pacing pulse frequency reduced by factor of 4 when this function is activated.
Output protection The equipment has no sign of damage after defibrillation-proof test.
Patient connection 3-lead ECG cable, 5-lead ECG cable, paddles or multifunction electrode pads
A-5
Neonate 15 to 350 bpm
HR measurement range Pediatric 15 to 350 bpm
Adult 15 to 300 bpm
HR accuracy ±1% or ±1bpm, which ever is greater
HR resolution 1 bpm
Measuring electrode: <0.1 μA
Lead-off detection current
Drive electrode: <1 μA
Baseline recovery time <5 s (after defibrillation, in monitor mode and therapy mode)
When the test is performed based on part 4.1.2.1 c) of ANSI/AAMI EC 13-2002,
the heart rate meter will reject all 100 ms QRS complexes with less than 1.2 mV
Tall T-wave rejection capability
of amplitude, and T waves with T-wave interval of 180 ms and those with Q-T
interval of 350 ms.
Meets the requirements of ANSI/AAMI EC13-2002: section 4.1.2.1 e). The heart
rate reading after 20 seconds of stabilization is:
A-6
Pace Pulse
Pace pulses meeting the following conditions are labelled with a PACE marker:
Amplitude: ±2 to ±700 mV
Pace pulse markers
Width: 0.1 to 2 ms
Rise time: 10 to 100 μs
Meets the requirements of ANSI/AAMI EC13-2002: section 4.1.4.1 and 4.1.4.3.
The following pulses will be rejected.
Amplitude: ±2 to ±700 mV
Pace pulse rejection
Width: 0.1 to 2 ms
Rise time: 10 to 100 μs
Minimum input slew rate: 10V/s RTI
Resp
Technique Trans-thoracic impedance
Adult: 0 to 120 rpm
Measurement range
Pediatric, neonate 0 to 150 rpm
7 to 150 rpm: ±2 rpm or ±2%, whichever is greater.
Accuracy
0 to 6 rpm: Not specified
Respiration excitation waveform <300 μA, sinusoid, 62.8 kHz (±10%)
Detectable respiration impedance range 0.3 to 5Ω
Reference impedance range 200 to 2500Ω, using ECG cable with 1 kΩ resistor
Difference input impedance >2.5 MΩ
Apnea alarm time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
A-7
PR
Measurement range 20 to 254 bpm
±3 bpm (measured without motion)
Accuracy
±5 bpm (measured with motion)
Refreshing rate 1s
PR
Measurement range 20 to 300 bpm
±3 bpm (20 to 250 bpm)
Accuracy
Not specified (251 to 300 bpm)
A-8
NIBP
Standards Meet standards of EN60601-2-30/IEC60601-2-30, EN1060-1, EN1060-3, EN1060-4 and SP10
Technique Oscillometry
Mode of operation Manual, Auto and STAT
Static pressure measurement
0kPa to 40.0kPa (0mmHg to 300mmHg)
range
Static pressure measurement
±0.4kPa (±3mmHg)
accuracy
120s for adult and pediatric patients
Maximum measurement time
90s for neonatal patients
Adult Pediatric Neonate
mmHg 40 to 270 40 to 200 40 to 135
Systolic
KPa 5.3 to 36.0 5.3 to 26.7 5.3 to 18.0
Measurement range mmHg 10 to 210 10 to 150 10 to 100
Diastolic
KPa 1.3 to 28.0 1.3 to 20.0 1.3 to 13.3
mmHg 20 to 230 20 to 165 20 to 110
Mean
KPa 2.7 to 30.7 2.7 to 22.0 2.7 to 14.7
Setting range Default Accuracy
A-9
Battery
14.8V/3Ah, smart lithium ion battery, rechargeable and free of maintenance, one battery
Battery type
can be configured
Less than 2 hours to 80% and less than 3 hours to 100% with equipment power off;
Charge time
Less than 3.5 hours to 80% and less than 4.5 hours to 100% with equipment power on.
New fully charged battery Testing condition
Without recording, typical ECG
Monitoring ≥2.5 h
monitoring, LCD brightness set to 1
50 Ω load impedance,
pacing ≥2 h pacing rate: 80bpm,
pacing output 60mA, without recording
Battery fuel gauge 5 LEDs indicating the current battery charge level
At least 20 minutes of monitoring and six 360J discharges (after the low battery alarm
Shutdown delay
occurs)
A-10
A.8 Data Management Specifications
Data Storage Internal CF card, 1G Bytes
Marking Events 16 types of events, user customized
Event recording Up to 1000 events for each patient.
Waveform storage Up to 24 hours of consecutive ECG waveform
Voice recording Max. 180 minutes in total; max. 60 minutes for each patient
Tabular Trends Max. 72 h of all measured parameters; resolution:1 min
Data Export Data can be export to a PC through a USB flash memory
Patient archives Up to 100
Storage environment
Shock
A-11
Vibration
Bump
Complies with the requirements of 6.3.4.2, EN1789.
Peak acceleration: 15g
Duration: 6ms
Number of impacts: 1000
Impact direction: vertical impacts are applied when the equipment under test is placed at normal operating position.
Free fall
A-12
B EMC
The equipment meets the requirements of IEC 60601-1-2.
NOTE
z Use of accessories, transducers, and cables other than those specified may result in increased emission
and/or decreased electromagnetic immunity of the defibrillator/monitor.
z The equipment or its components should not be used adjacent to or stacked with other devices. If adjacent
or stacked use is necessary, the equipment should be observed to verify normal operation in the
configuration in which it will be used.
z The equipment needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided below.
z Other devices may affect this monitor even though they meet the requirements of CISPR.
z When the inputted signal is below the minimum amplitude provided in technical specifications, erroneous
measurements could result.
B-1
Guidance and Declaration - Electromagnetic Immunity
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the
equipment should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment -
guidance
Electrostatic discharge ±6 kV contact ±6 kV contact Floors should be wood, concrete or
(ESD) ±8 kV air ±8 kV air ceramic tile. If floors are covered
IEC 61000-4-2 with synthetic material, the relative
humidity should be at least 30%.
Electrical fast ±2 kV for power supply lines ±2 kV for power supply lines Mains power quality should be that
transient/burst ±1 kV for input/output lines ±1 kV for input/output lines of a typical commercial or hospital
(EFT) (length greater than 3 m) (length greater than 3 m) environment.
IEC 61000-4-4
Surge ±1 kV differential mode ±1 kV differential mode
IEC 61000-4-5 ±2 kV common mode ±2 kV common mode
Voltage dips, short <5% UT (>95% dip in UT) for <5% UT (>95% dip in UT) Mains power quality should be that
interruptions and 0.5 cycle for 0.5 cycle of a typical commercial or hospital
voltage variations on environment. If the user of our
power supply input 40% UT (60% dip in UT) for 5 40% UT (60% dip in UT) for product requires continued
lines cycle 5 cycle operation during power mains
IEC 61000-4-11 interruptions, it is recommended
70% UT (30% dip in UT) for 70% UT (30% dip in UT) for that our product be powered from
25 cycle 25 cycle an uninterruptible power supply or
a battery.
<5% UT (>95% dip in UT) for <5% UT (>95% dip in UT)
5 seconds for 5 seconds
Power frequency 3 A/m 3 A/m Power frequency magnetic fields
(50/60 Hz) magnetic should be at levels characteristic of
field a typical location in a typical
IEC 61000-4-8 commercial or hospital
environment.
Note: UT is the A.C. mains voltage prior to application of the test level.
B-2
Guidance and Declaration - Electromagnetic Immunity
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the
equipment should assure that it is used in such an environment.
IEC 60601
Immunity test Test level Compliance level Electromagnetic environment - guidance
Conduced RF 3 Vrms 3 Vrms (V1) Portable and mobile RF communications equipment should
IEC 61000-4-6 150k to 80 MHz be used no closer to any part of the device, including cables,
Outside ISM than the recommended separation distance calculated from
bandsa the equation applicable to the frequency of the transmitter.
Recommended separation distance:
⎡ 3 .5 ⎤
d = ⎢ P
10 Vrms 10 Vrms (V2) ⎣ V 1 ⎥⎦
150kHz to ⎡ 12 ⎤
d = ⎢ P
80MHz ⎣ V 2 ⎥⎦
⎡ 12 ⎤
in ISM bandsa d = ⎢ P
(for life support ⎣ E 1 ⎥⎦ 80 MHz to 800 MHz
⎡ 23 ⎤
devices) d = ⎢ P
⎣ E 1 ⎥⎦ 800 MHz to 2.5 GHz
Radiated RF 10V/m 10 V/m (E1)
IEC 61000-4-3 80 MHz to 2.5 where P is the maximum output power rating of the
B-3
a The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to
13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
b Compliance level in the ISM frequency bands between 150 kHz to 80 MHz and in the frequency range 80 MHz to 2.5 GHz
are intended to decrease the likelihood that portable/ mobile communication equipment could cause interference if it is
inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the recommended
separation distance for transmitters in these frequency ranges.
c Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If
the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above,
the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the device.
d Over the frequency ranges 150 kHz to 80 MHz beyond ISM band, field strengths should be less than 3V/m; within ISM band,
field strengths should be less than 10V/m.
Recommended Separation Distances between Portable and Mobile RF, Communications Equipment and This
Equipment
The equipment is intended for use in an electromagnetic environment in which radiated RF disturbance are controlled. The
customer or the user of the equipment can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the equipment as recommended below,
according to the maximum output power of the communication equipment.
B-4
C BeneHeart Defibrillator Shift Checklist
Inspect the defibrillator/monitor at the change of every shift. Place a “√” in the “Pass/Fail” box as you check the item , or
place a “-” if not applicable. Describe the problem if there is any abnormity.
C-1
*: Perform this test only when automatic selftest is not performed using pads cable or when selftest fails.
**: Perform this test only when automatic selftest is not performed using paddles or when selftest fails.
NOTE
z Remember to disconnect the test load when the test is finished. Otherwise, delay could happen in patient
treatment.
C-2
D Alarm Messages
This chapter lists only the most important physiological and technical alarm messages. Some messages appearing on
your equipment may not be included.
In this chapter:
The “I” column indicates how indications of technological alarms are cleared after the hardkey or [Audio Pause]
softkey is pressed: “A” means all alarm indications are cleared; “B” indicates alarm light and alarm tones are cleared
and the alarm messages change to prompt messages; and “C” indicates only alarm tone is disabled, but alarm light
and alarm message remain presented.
The “L” column indicates the alarm level: “H” refers to high, “M” refers to medium, and “L” refers to low. “*” means
the alarm level is user-adjustable.
XX represents a measurement or parameter label, such as ECG, HR, PVCs, RR, SpO2, PR, NIBP, etc.
In the “Cause and solution” column, corresponding solutions are given instructing you to troubleshoot problems. If the
problem persists, contact your service personnel.
D-1
Measurement Alarm Message L Cause and solution
Nonsus. Vtac M*
Pause M*
Irr. Rhythm M*
PNP M* The pacer appears abnormal. Check the pacer.
PNC M*
Resp Resp Apnea H The respiration signal was so weak that the equipment cannot perform
respiration analysis. Check the patient’s condition and the Resp connections.
SpO2 SpO2 Desat H The SpO2 value has fallen below the desaturation alarm limit. Check the
patient’s condition and check if the alarm limit settings are correct.
No Pulse L The pulse signal was so weak that the equipment cannot perform pulse
analysis. Check the patient’s condition, SpO2 sensor and measurement site.
D-2
Measurement Alarm Message L I Cause and solution
SpO2 Too Much Light L C There is too much light on the SpO2 sensor. Move the sensor
to a place with lower level of ambient light or cover the
sensor to minimize the ambient light.
SpO2 Low Signal L C The SpO2 signal is too low or too weak. Check the patient’s
SpO2 Weak Signal L C condition and change the sensor application site. If the error
SpO2 Weak Pulse L C persists, replace the sensor.
SpO2 Low Perf L B
SpO2 Interference L C The SpO2 signal has been interfered. Check for any possible
sources of signal noise form the area around the sensor, and
check the patient for excessive motion.
SpO2 Non-Pulsatile L C
SpO2 Board Fault L C There is a problem with the SpO2 measurement board. Do not
use the module and contact your service personnel.
NIBP NIBP Loose Cuff L A The NIBP cuff is not properly connected, or there is a leak in
NIBP Air Leak L A the airway.
NIBP Pneumatic Leak L A Check the NIBP cuff and pump for leakages.
NIBP Cuff Type Wrong L A The cuff type applied mismatches the patient category. Verify
the patient category and replace the cuff.
NIBP Air Press. Err L A An error occurred to the air pressure. Verify that the
equipment application site meets the environmental
requirements and check if there is any source that affects the
air pressure.
NIBP Weak Signal L A The patient’s pulse is weak or the cuff is loose. Check the
patient’s condition and change the cuff application site. If the
problem persists, change the cuff.
NIBP Sig. Saturated L A The NIBP signal is saturated due to excess motion or other
sources.
NIBP Overrange L A The patient’s NIBP value may be beyond the specified
measurement range.
NIBP Excessive Motion L A Check the patient’s condition and reduce the patient motion.
NIBP Cuff Overpress L A The NIBP airway may be occluded. Check the airway and
measure again.
NIBP Equip Err H A An error occurred during NIBP measurement and therefore
NIBP Time Out L A the equipment cannot perform analysis correctly. Check the
NIBP Measure Failed L A patient’s condition and NIBP connections, or replace the cuff.
NIBP Reset For Err L A An illegal reset occurred during NIBP measurement. Check if
the airway is occluded.
Main control No Fan L C Make sure that the fan is connected.
system No Speaker L C Make sure that the speaker is connected.
No Data Card L C Make sure that the CF card is in place or format the CF card.
Power Board Comm Err H C An error occurred to the power board, or there is a problem
with the communications between the power board and the
host. Restart the equipment.
D-3
Measurement Alarm Message L I Cause and solution
Keyboard Comm Err L C An error occurred to the keypad board, or there is a problem
with the communications between the keypad board and the
host. Restart the equipment.
Therapy Module Comm Err S C An error occurred to the therapy module, or there is a
problem with the communications between the therapy
module and the host. Restart the equipment. If the problem
persists, contact your service personnel.
Main Control Selftest Err H C The main control voltage is abnormal. Contact your service
personnel.
RT Clock Need Reset L C Reset system time.
RT Clock Err H C An error occurred to the RTC chip, or the button cell is
depleted. Replace corresponding part.
Data Card Err L C There is a problem with the data card. Check or replace the
data card if necessary.
Power board Power System Selftest Err H C An error occurred to the system power supply. Restart the
Power Board Volt Err L C equipment.
No battery L C No battery is installed. Please install the battery.
Low Battery S C Change battery or connect the equipment to the AC power
source to charge the batteries.
Battery Err H There is a problem with the batteries. Check the batteries for
damage; verify that correct batteries are used. Replace the
batteries if necessary.
Battery Depleted! System S C Connect the equipment to AC mains.
will shut shown imminently.
Connect to AC Mains or
Replace Battery.
Battery Aged L C Replace the battery.
Battery failed charging M C Battery failure or power board hardware failure. Replace the
battery. If the problem persists, contact your service
personnel.
Therapy Therapy Equip selftest Err S C An error occurred during therapy module self test. Restart the
module equipment If the problem persists, contact your service
personnel.
Defib Malfunction S C The defibrillation function fails or both the defibrillation and
pacing functions fail. Restart the equipment and test
defibrillation function. If the problem persists, contact your
service personnel.
Pacer Malfunction! S C The pacing function fails. Restart the equipment and test
pacer function. If the problem persists, contact your service
personnel.
Disarming Failed H C There is a problem with the therapy module disarming circuit.
Contact your service personnel.
Last User Test Failed L C Run a successful user test.
Last Auto Test Failed L C Run a successful user test again.
D-4
Measurement Alarm Message L I Cause and solution
Monitoring Monitor Module Selftest Err H C An error occurred during MPM module power-on self test.
module Contact your service personnel.
Monitor Module Reset Err H C MPM module reset abnormally. In this case, the MPM module
restores to default configuration. You can ignore this
problem.
Monitor Module Voltage Err L C The voltage of MPM module is abnormal. Contact your service
personnel.
Recorder Recorder Init Err L A Restart the equipment.
Recordhead Overheated L A The recorder has been working for a prolonged time. Clear
the recording tasks and resume the recording till the
recorder’s print head cools down.
Recorder Overcurrent L A Re-load the recorder paper.
Others Config Err L A Check if the configuration is correct, or restore the factory
configuration.
Pacer Pads cable Off H C Check that pads cable is properly connected.
Pads Off H C Check that pads are properly connected.
ECG Lead Off H C Check that ECG leadwires are properly connected.
Pacer Stopped Abnormally H C Check paddles. Check that pads well contact with patient’s
skin. Make sure pads are properly applied, and then start
pacing again.
Note: In the “L” column “S” refers to special technological alarm. The special technological alarms cannot be paused or
silenced, and the alarm volume is unchangeable. These alarms stop only when the alarm condition is eliminated.
D-5
FOR YOUR NOTES
D-6
E Electrical Safety Inspection
The following electrical safety tests are recommended as part of a comprehensive preventive maintenance program.
They are a proven means of detecting abnormalities that, if undetected, could prove dangerous to either the patient or
the operator. Additional tests may be required according to local regulations.
All tests can be performed by using commercially available safety analyzer test equipment. These procedures assume
the use of a 601PROXL International Safety Analyzer or equivalent safety analyzer. Other popular testers complying with
IEC 60601-1 used in Europe, such as Fluke, Metron or Gerb, may require modifications to the procedure. Please follow the
instructions of the analyzer manufacturer.
The electrical safety inspection should be periodically performed per years .The safety analyzer also proves to be an
excellent troubleshooting tool to detect abnormalities of line voltage and grounding, as well as total current loads.
E-1
E.3 Device Labelling
Check the labels provided by the manufacturer or the healthcare facilities are present and legible.
LIMITS
The following outlet conditions apply when performing the Earth Leakage test:
LIMITS
For UL60601-1,
For IEC60601-1,
E-2
E.6 Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth. All measurements have a true
RMS only
The following outlet conditions apply when performing the Patient Leakage Current test.
LIMITS
The following outlet conditions apply when performing the Mains on Applied Part test.
Normal Polarity
Reversed Polarity
LIMITS
E-3
E.8 Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the remaining Applied Part
connector s. All measurements may have a true RMS only response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.
LIMITS
NOTE
z Make sure the safety analyzer is authorized comply with requirement of IEC61010-1.
E-4
F Symbols and Abbreviations
F.1 Units
μA microampere
μV microvolt
A ampere
Ah ampere hour
ºC centigrade
cc cubic centimeter
cm centimeter
dB decibel
DS dyne second
ºF fahrenheit
g gram
GHz gigahertz
GTT gutta
h hour
Hz hertz
in inch
J Joule
k kilo
kg kilogram
kPa kilopascal
L litre
lb pound
m meter
Mb mega byte
mcg microgram
mEq milli-equivalents
mg milligram
min minute
ml milliliter
mm millimeter
F-1
mmHg millimeters of mercury
ms millisecond
mV millivolt
mW milliwatt
MΩ megaohm
nm nanometer
s second
V volt
VA volt ampere
Ω ohm
W watt
F.2 Symbols
– negative, minus
% percent
/ per; divide; or
+ plus
= equal to
± plus or minus
× multiply
© copyright
F-2
Art arterial
aVF left foot augmented lead
aVL left arm augmented lead
aVR right arm augmented lead
awRR airway respiratory rate
BAP brachial aterial pressure
BIS bispectral index
BP blood pressure
BPSK binary phase shift keying
BSA body surface area
BT blood temperature
BTPS body temperature and pressure, saturated
C.I. cardiac index
C.O. cardiac output
CaO2 arterial oxygen content
CCO continuous cardiac output
CCU cardiac (coronary) care unit
CE Conformité Européenne
CIS clinical information system
CISPR International Special Committee on Radio Interference
CMOS complementary metal oxide semiconductor
CMS central monitoring system
C.O. cardiac output
CO2 carbon dioxide
COHb carboxyhemoglobin
CP cardiopulmonary
CPR Cardiopulmonary resuscitation
CVP central venous pressure
DC direct current
Defib defibrillation
Des desflurane
Dia diastolic
DPI dot per inch
DVI digital video interface
ECG electrocardiograph
EDV end-diastolic volume
EEC European Economic Community
EEG electroencephalogram
EMC electromagnetic compatibility
EMG electromyograph
EMI electromagnetic interference
Enf enflurane
ESU electrosurgical unit
Et end-tidal
EtCO2 end-tidal carbon dioxide
EtN2O end-tidal nitrous oxide
F-3
EtO ethylene oxide
EtO2 end-tidal oxygen
FAP femoral arterial pressure
FCC Federal Communication Commission
FDA Food and Drug Administration
FEV1.0% first second forced expiratory volume ratio
Fi fraction of inspired
FiCO2 fraction of inspired carbon oxygen
FiN2O fraction of inspired nitrous oxide
FiO2 fraction of inspired oxygen
FPGA field programmable gate array
FV flow-volume
Hal halothane
Hb hemoglobin
Hb-CO carbon mono-xide hemoglobin
HbO2 oxyhemoglobin
HIS hospital information system
HR heart rate
I:E inspiratory-expiratory ratio
IBP invasive brood pressure
ICG impedance cardiography
ICP intracranial pressure
ICT/B intracranial catheter tip pressure transducer
ICU intensive care unit
ID identification
IEC International Electrotechnical Commission
IEEE Institute of Electrical and Electronic Engineers
IP internet protocol
Iso isoflurane
IT injectate temperature
LA left arm
LAP left atrial pressure
Lat lateral
LCD liquid crystal display
LCW left cardiac work
LCWI left cardiac work index
LED light emitting diode
LL left leg
LVDS low voltage differential signal
F-4
MetHb methemoglobin
MRI magnetic resonance imaging
MVe expiratory minute volume
MVi inspiratory minute volume
N/A not applied
Neo neonate
NIBP noninvasive blood pressure
NIP negative inspiratory pressure
O2 oxygen
O2CI oxygen consumption index
O2R oxygen extraction ratio
OR operating room
oxyCRG oxygen cardio-respirogram
PA pulmonary artery
Paw airway pressure
PAWP pulmonary artery wedge pressure
PD photodetector
Ped pediatric
PEEP positive end expiratory pressure
PEF peak expiratory flow
PEP pre-ejection period
PIF peak inspiratory flow
PIP peak inspiratory pressure
Pleth plethysmogram
Pmean mean pressure
PNC pacer not captured
PNP pacer not paced
Pplat plateau pressure
PR pulse rate
PVC premature ventricular complex
PVR pulmonary vascular resistance
PVRI pulmonary vascular resistance index
R right
RA right arm
RAM random access memory
RAP right atrial pressure
Raw airway resistance
Rec record, recording
Resp respiration
RHb reduced hemoglobin
RL right leg
RM respiratory mechanics
RR respiration rate
RSBI rapid shallow breathing index
SaO2 arterial oxygen saturation
SEF spectral edge frequency
F-5
Sev sevoflurane
SFM self-maintenance
SI stroke index
SMR satellite module rack
SpO2 arterial oxygen saturation from pulse oximetry
SQI signal quality index
SR suppression ratio
STR systolic time ratio
SVR systemic vascular resistance
SVRI systemic vascular resistance index
Sync synchronization
Sys systolic pressure
Taxil axillary temperature
TD temperature difference
Temp temperature
TFC thoracic fluid content
TFI thoracic fluid index
TFT thin-film technology
Toral oral temperature
TP total power
Trect rectal temperature
TVe expiratory tidal volume
TVi inspiratory tidal volume
UAP umbilical arterial pressure
UPS uninterruptible power supply
USB universal serial bus
UVP umbilical venous pressure
VAC volts alternating current
VEPT volume of electrically participating tissue
VI velocity index
WOB work of breathing
F-6
G Device Tracking
In order to provide high quality product and perform better service, we are going to track our product. Please contact us
with the device tracking information when you have received your defibrillator/monitor:
Please fill the information in the next page, cut the table and fax it to +86 755 26582934. You can also email your
information to [email protected].
G-1
FOR YOUR NOTES
G-2
Device Tracking Information
User Information
Customer Name
Department name
Address
Contact Person
Email Address
Device Information