H-046-001653-00 BeneHeart D3 User Manual PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 186

BeneHeart D3

Defibrillator/Monitor

Operator’s Manual
© 2010 - 2013 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.
For this Operator’s Manual, the issue date is 2013-09.

I
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property
rights to this Mindray product and this manual. This manual may refer to information protected by copyright or
patents and does not convey any license under the patent rights or copyright of Mindray, or of others.
Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in
this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.

Release, amendment, reproduction, distribution, rental, adaptation, translation or any other derivative work of this
manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.

, , and are the trademarks, registered or otherwise, of

Mindray in China and other countries. All other trademarks that appear in this manual are used only for informational
or editorial purposes. They are the property of their respective owners.

II
Responsibility on the Manufacturer Party
Contents of this manual are subject to change without prior notice.

All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained
herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this
manual.

Mindray is responsible for the effects on safety, reliability and performance of this product, only if:

„ all installation operations, expansions, changes, modifications and repairs of this product are conducted by
Mindray authorized personnel;

„ the electrical installation of the relevant room complies with the applicable national and local requirements;and

„ the product is used in accordance with the instructions for use.

WARNING
z This equipment must be operated by skilled/trained clinical professionals.

z It is important for the hospital or organization that employs this equipment to carry out a reasonable
service/maintenance plan. Neglect of this may result in machine breakdown or personal injury.

III
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE.

Exemptions

Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability
for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or
the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized
personnel.

This warranty shall not extend to:

„ Malfunction or damage caused by improper use or man-made failure.

„ Malfunction or damage caused by unstable or out-of-range power input.

„ Malfunction or damage caused by force majeure such as fire and earthquake.

„ Malfunction or damage caused by improper operation or repair by unqualified orunauthorized service people.

„ Malfunction of the instrument or part whose serial number is not legible enough.

„ Others not caused by instrument or part itself.

Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building,Keji 12th Road South,Hi-tech industrial
park,Nanshan,Shenzhen 518057,P.R.China
Website: www.mindray.com
E-mail Address: [email protected]

Tel: +86 755 81888998


Fax: +86 755 26582680

EC-Representative: Shanghai International Holding Corp. GmbH (Europe)

Address: Eiffestraβe 80, 20537 Hamburg, Germany


Tel: 0049-40-2513175
Fax: 0049-40-255726

IV
Preface
Manual Purpose
This manual contains the instructions necessary to operate the product safely and in accordance with its function and
intended use. Observance of this manual is a prerequisite for proper product performance and correct operation and
ensures patient and operator safety.

This manual is based on the maximum configuration and therefore some contents may not apply to your product. If
you have any question, please contact us.

This manual is an integral part of the product. It should always be kept close to the equipment so that it can be
obtained conveniently when needed.

Intended Audience
This manual is geared for clinical professionals who are expected to have a working knowledge of medical procedures,
practices and terminology as required for monitoring of critically ill patients.

Illustrations
All illustrations in this manual serve as examples only. They may not necessarily reflect the setup or data displayed on
your equipment.

Conventions
„ Italic text is used in this manual to quote the referenced chapters or sections.

„ [ ] is used to enclose screen texts.

„ → is used to indicate operational procedures.

V
FOR YOUR NOTES

VI
Contents

1 Safety............................................................................................................................................................................ 1-1
1.1 Safety Information.............................................................................................................................................................................................1-1
1.1.1 Dangers ...................................................................................................................................................................................................1-2
1.1.2 Warnings .................................................................................................................................................................................................1-2
1.1.3 Cautions ..................................................................................................................................................................................................1-3
1.1.4 Notes ........................................................................................................................................................................................................1-3
1.2 Equipment Symbols..........................................................................................................................................................................................1-4

2 The Basics ..................................................................................................................................................................... 2-1


2.1 Overview...............................................................................................................................................................................................................2-1
2.2 Intended Use .......................................................................................................................................................................................................2-1
2.2.1 AED............................................................................................................................................................................................................2-2
2.2.2 Manual Defibrillation..........................................................................................................................................................................2-2
2.2.3 Noninvasive Pacing .............................................................................................................................................................................2-2
2.2.4 ECG............................................................................................................................................................................................................2-2
2.2.5 Resp ..........................................................................................................................................................................................................2-2
2.2.6 SpO2 ..........................................................................................................................................................................................................2-2
2.2.7 NIBP...........................................................................................................................................................................................................2-2
2.3 Main Unit ..............................................................................................................................................................................................................2-3
2.3.1 Front View...............................................................................................................................................................................................2-3
2.3.2 Side View.................................................................................................................................................................................................2-8
2.3.3 Rear View.................................................................................................................................................................................................2-9
2.3.4 External Paddles................................................................................................................................................................................ 2-10
2.4 Display Views.................................................................................................................................................................................................... 2-11

3 Basic Operations and Settings.................................................................................................................................... 3-1


3.1 Installation............................................................................................................................................................................................................3-1
3.1.1 Unpacking and Checking .................................................................................................................................................................3-1
3.1.2 Environmental Requirements .........................................................................................................................................................3-2
3.2 Basic Operation ..................................................................................................................................................................................................3-2
3.2.1 Turning Power On ................................................................................................................................................................................3-2
3.2.2 Starting Monitoring or Therapy......................................................................................................................................................3-2
3.2.3 Disconnecting from Power...............................................................................................................................................................3-3
3.2.4 Auto Restoring to Last Configuration...........................................................................................................................................3-3
3.3 Using the Main Menu .......................................................................................................................................................................................3-3
3.4 Changing General Settings ............................................................................................................................................................................3-3
3.4.1 Setting the Date and Time................................................................................................................................................................3-3
3.4.2 Adjusting the Screen Brightness....................................................................................................................................................3-4
3.4.3 Changing Key Volume........................................................................................................................................................................3-4
3.4.4 Selecting High Contrast Mode........................................................................................................................................................3-4

1
4 Managing Patients....................................................................................................................................................... 4-1
4.1 Overview .............................................................................................................................................................................................................. 4-1
4.2 Editing Patient Information ........................................................................................................................................................................... 4-1

5 Alarms........................................................................................................................................................................... 5-1
5.1 Alarm Categories ............................................................................................................................................................................................... 5-1
5.2 Alarm Levels ........................................................................................................................................................................................................ 5-2
5.3 Alarm Indicators................................................................................................................................................................................................. 5-2
5.3.1 Alarm Lamps ......................................................................................................................................................................................... 5-2
5.3.2 Audible Alarms ..................................................................................................................................................................................... 5-3
5.3.3 Alarm Message ..................................................................................................................................................................................... 5-3
5.3.4 Flashing Numeric................................................................................................................................................................................. 5-3
5.3.5 Alarm Status Symbols ........................................................................................................................................................................ 5-3
5.4 Alarm Tone Configuration .............................................................................................................................................................................. 5-4
5.4.1 Changing the Alarm Volume........................................................................................................................................................... 5-4
5.4.2 Setting the Interval between Alarm Sounds ............................................................................................................................. 5-4
5.5 Understanding the Alarm Setup Menu..................................................................................................................................................... 5-4
5.5.1 Setting Alarm Properties for All Parameters.............................................................................................................................. 5-5
5.5.2 Adjusting Alarm Limits Automatically ......................................................................................................................................... 5-6
5.6 Pausing Alarms................................................................................................................................................................................................... 5-7
5.7 Switching Alarms Off ....................................................................................................................................................................................... 5-7
5.8 Pausing Alarm Sounds..................................................................................................................................................................................... 5-8
5.9 Switching Off Alarm Sounds ......................................................................................................................................................................... 5-8
5.10 Reminder Tones ............................................................................................................................................................................................... 5-8
5.11 Latching Alarms............................................................................................................................................................................................... 5-9
5.12 Clearing Technical Alarms............................................................................................................................................................................ 5-9
5.13 When an Alarm Occurs................................................................................................................................................................................5-10

6 Monitoring ECG............................................................................................................................................................ 6-1


6.1 Overview .............................................................................................................................................................................................................. 6-1
6.2 Safety ..................................................................................................................................................................................................................... 6-1
6.3 Monitoring View ................................................................................................................................................................................................ 6-2
6.4 Preparing to Monitor ECG .............................................................................................................................................................................. 6-2
6.4.1 ECG Monitoring with Electrodes.................................................................................................................................................... 6-2
6.4.2 ECG Monitoring with Paddles/Pads.............................................................................................................................................. 6-4
6.4.3 Checking Paced Status ...................................................................................................................................................................... 6-5
6.5 ECG Display.......................................................................................................................................................................................................... 6-5
6.6 Changing ECG Settings ................................................................................................................................................................................... 6-6
6.6.1 Changing Lead Settings.................................................................................................................................................................... 6-6
6.6.2 Changing ECG Wave Settings.......................................................................................................................................................... 6-7
6.6.3 Switching the Notch Filter On or Off ............................................................................................................................................ 6-7
6.6.4 Adjusting Heartbeat Volume........................................................................................................................................................... 6-8
6.7 Arrhythmia Analysis.......................................................................................................................................................................................... 6-8
6.7.1 Understanding the Arrhythmia Events........................................................................................................................................ 6-8
6.7.2 Switching Arrhythmia Analysis On and Off ............................................................................................................................... 6-9
6.7.3 Changing Arrhythmia Alarm Settings........................................................................................................................................6-10

2
6.7.4 Changing Arrhythmia Threshold Settings ............................................................................................................................... 6-10
6.7.5 Initiating Arrhythmia Relearning Manually............................................................................................................................. 6-11
6.7.6 Automatic Arrhythmia Relearn.................................................................................................................................................... 6-11
6.8 Calibrating ECG................................................................................................................................................................................................ 6-11

7 AED ............................................................................................................................................................................... 7-1


7.1 Overview...............................................................................................................................................................................................................7-1
7.2 Safety......................................................................................................................................................................................................................7-1
7.3 AED View...............................................................................................................................................................................................................7-2
7.4 AED Procedure....................................................................................................................................................................................................7-3
7.5 Shock Advised.....................................................................................................................................................................................................7-4
7.6 No Shock Advised (NSA)..................................................................................................................................................................................7-4
7.7 CPR ..........................................................................................................................................................................................................................7-5
7.7.1 CPR Metronome ...................................................................................................................................................................................7-5
7.8 AED Sound Recording......................................................................................................................................................................................7-6
7.9 AED Setup.............................................................................................................................................................................................................7-6

8 Manual Defibrillation .................................................................................................................................................. 8-1


8.1 Overview...............................................................................................................................................................................................................8-1
8.2 Safety......................................................................................................................................................................................................................8-1
8.3 Manual Defibrillation View .............................................................................................................................................................................8-3
8.4 Manual Defibrillation Procedure ..................................................................................................................................................................8-3
8.4.1 Using Pediatric Paddles .....................................................................................................................................................................8-5
8.4.2 Using Internal Paddles .......................................................................................................................................................................8-5
8.5 Synchronized Cardioversion..........................................................................................................................................................................8-6
8.5.1 Performing Synchronized Cardioversion ....................................................................................................................................8-7
8.5.2 Delivering Additional Synchronized Shocks..............................................................................................................................8-7
8.5.3 Disabling the Sync Function............................................................................................................................................................8-7
8.6 Remote Synchronized Cardioversion .........................................................................................................................................................8-8
8.7 Contact Impedance Indicator .......................................................................................................................................................................8-9

9 Noninvasive Pacing ..................................................................................................................................................... 9-1


9.1 Overview...............................................................................................................................................................................................................9-1
9.2 Safety......................................................................................................................................................................................................................9-1
9.3 Pacing View ..........................................................................................................................................................................................................9-2
9.4 Demand Mode versus Fixed Mode..............................................................................................................................................................9-3
9.5 Preparing for Pacing .........................................................................................................................................................................................9-3
9.5.1 Demand Mode Pacing .......................................................................................................................................................................9-4
9.5.2 Fixed Mode Pacing ..............................................................................................................................................................................9-5

10 Monitoring Resp ...................................................................................................................................................... 10-1


10.1 Overview ......................................................................................................................................................................................................... 10-1
10.2 Safety ................................................................................................................................................................................................................ 10-1
10.3 Resp View ........................................................................................................................................................................................................ 10-1
10.4 Placing Resp Electrodes ............................................................................................................................................................................. 10-2
10.4.1 Optimizing Lead Placement for Resp...................................................................................................................................... 10-3

3
10.4.2 Changing Resp Wave Settings....................................................................................................................................................10-3

11 Monitoring PR .......................................................................................................................................................... 11-1


11.1 Overview ..........................................................................................................................................................................................................11-1
11.2 Adjusting Pulse Tone Volume ...................................................................................................................................................................11-1

12 Monitoring SpO2 ...................................................................................................................................................... 12-1


12.1 Introduction....................................................................................................................................................................................................12-1
12.2 Safety.................................................................................................................................................................................................................12-2
12.3 Identifying SpO2 Modules..........................................................................................................................................................................12-2
12.4 SpO2 Monitoring Procedure......................................................................................................................................................................12-2
12.5 Changing SpO2 Settings .............................................................................................................................................................................12-3
12.5.1 Setting SpO2 Sensitivity ................................................................................................................................................................12-3
12.5.2 Monitoring SpO2 and NIBP on the Same Limb.....................................................................................................................12-3
12.5.3 Changing Averaging Time ...........................................................................................................................................................12-3
12.5.4 Sat-Seconds Alarm Management .............................................................................................................................................12-4
12.5.5 Changing the Speed of the Pleth Wave ..................................................................................................................................12-5
12.6 SpO2 Desat Alarm..........................................................................................................................................................................................12-5
12.7 Pitch Tone.........................................................................................................................................................................................................12-5
12.8 Measurement Limitations..........................................................................................................................................................................12-5
12.9 Masimo Information ....................................................................................................................................................................................12-6
12.10 Nellcor Information....................................................................................................................................................................................12-6

13 NIBP .......................................................................................................................................................................... 13-1


13.1 Introduction....................................................................................................................................................................................................13-1
13.2 Safety.................................................................................................................................................................................................................13-1
13.3 Measurement Limitations..........................................................................................................................................................................13-2
13.4 Measurement Modes...................................................................................................................................................................................13-2
13.5 Measuring Procedure ..................................................................................................................................................................................13-2
13.5.1 Preparing for NIBP Measurement .............................................................................................................................................13-2
13.5.2 Starting and Stopping NIBP Measurements .........................................................................................................................13-3
13.5.3 Correcting the Measurement .....................................................................................................................................................13-3
13.5.4 Enabling NIBP Auto Cycling ........................................................................................................................................................13-3
13.5.5 Starting a STAT Measurement ....................................................................................................................................................13-3
13.6 Understanding the NIBP Numerics ........................................................................................................................................................13-4
13.7 Setting Initial Cuff Inflation Pressure .....................................................................................................................................................13-4
13.8 Setting Pressure Unit ...................................................................................................................................................................................13-4

14 Marking Events ........................................................................................................................................................ 14-1

15 Freezing Waveforms ................................................................................................................................................ 15-1


15.1 Freezing Waveforms.....................................................................................................................................................................................15-1
15.2 Reviewing Frozen Waveforms ..................................................................................................................................................................15-1
15.3 Unfreezing Waveforms................................................................................................................................................................................15-2
15.4 Recording Frozen Waveforms...................................................................................................................................................................15-2

4
16 Review ...................................................................................................................................................................... 16-1
16.1 Reviewing Events ......................................................................................................................................................................................... 16-1
16.2 Reviewing Tabular Trends.......................................................................................................................................................................... 16-2

17 Data Management................................................................................................................................................... 17-1


17.1 Introduction ................................................................................................................................................................................................... 17-1
17.2 Reviewing Patient Events .......................................................................................................................................................................... 17-2
17.3 Exporting Data .............................................................................................................................................................................................. 17-2

18 Recording ................................................................................................................................................................. 18-1


18.1 Using a Recorder........................................................................................................................................................................................... 18-1
18.2 Recording Types............................................................................................................................................................................................ 18-1
18.3 Starting and Stopping Recordings ........................................................................................................................................................ 18-1
18.4 Setting the Recorder ................................................................................................................................................................................... 18-2
18.4.1 Accessing the Record Setup Menu .......................................................................................................................................... 18-2
18.4.2 Selecting Waveforms for Recording ........................................................................................................................................ 18-2
18.4.3 Setting the Realtime Recording Length................................................................................................................................. 18-2
18.4.4 Changing the Recording Speed................................................................................................................................................ 18-3
18.4.5 Switching Gridlines On or Off .................................................................................................................................................... 18-3
18.5 Loading Paper................................................................................................................................................................................................ 18-3
18.6 Removing Paper Jam .................................................................................................................................................................................. 18-4
18.7 Cleaning the Recorder Print head .......................................................................................................................................................... 18-4

19 Configuration Management ................................................................................................................................... 19-1


19.1 Introduction ................................................................................................................................................................................................... 19-1
19.2 Password.......................................................................................................................................................................................................... 19-1
19.3 Accessing Configuration Management................................................................................................................................................ 19-1
19.3.1 General Setup Menu ..................................................................................................................................................................... 19-2
19.3.2 Manual Defib Setup Menu .......................................................................................................................................................... 19-3
19.3.3 AED Setup Menu ............................................................................................................................................................................ 19-3
19.3.4 Pacer Setup Menu .......................................................................................................................................................................... 19-4
19.3.5 ECG Setup Menu............................................................................................................................................................................. 19-4
19.3.6 Resp Setup Menu ........................................................................................................................................................................... 19-6
19.3.7 SpO2 Setup Menu ........................................................................................................................................................................... 19-6
19.3.8 PR Setup Menu................................................................................................................................................................................ 19-6
19.3.9 NIBP Setup Menu............................................................................................................................................................................ 19-7
19.3.10 Alarm Setup Menu....................................................................................................................................................................... 19-8
19.3.11 Waveform Setup Menu .............................................................................................................................................................. 19-8
19.3.12 Mark Event Setup Menu ............................................................................................................................................................ 19-9
19.3.13 Record Setup Menu .................................................................................................................................................................... 19-9
19.3.14 Data Management Setup Menu............................................................................................................................................. 19-9
19.3.15 User Test Setup Menu...............................................................................................................................................................19-10
19.3.16 Others Menu................................................................................................................................................................................19-10

20 Battery...................................................................................................................................................................... 20-1
20.1 Introduction ................................................................................................................................................................................................... 20-1

5
20.2 Installing the Batteries ................................................................................................................................................................................20-2
20.3 Battery Alarms................................................................................................................................................................................................20-2
20.3.1 No Battery Alarm.............................................................................................................................................................................20-2
20.3.2 Low Battery Alarm ..........................................................................................................................................................................20-2
20.3.3 Battery Aged Alarm........................................................................................................................................................................20-3
20.3.4 Battery Error Alarm.........................................................................................................................................................................20-3
20.4 Checking the Batteries................................................................................................................................................................................20-3
20.5 Charging batteries........................................................................................................................................................................................20-3
20.6 Storing Batteries............................................................................................................................................................................................20-4
20.7 Recycling the Batteries................................................................................................................................................................................20-4

21 Care and Cleaning.................................................................................................................................................... 21-1


21.1 General Points ................................................................................................................................................................................................21-1
21.2 Cleaning ...........................................................................................................................................................................................................21-2
21.3 Disinfecting .....................................................................................................................................................................................................21-2

22 Maintenance and Testing ........................................................................................................................................ 22-1


22.1 Overview ..........................................................................................................................................................................................................22-1
22.2 Maintenance and Testing Schedule .......................................................................................................................................................22-2
22.3 Carrying Out Maintenance and Testing ................................................................................................................................................22-2
22.3.1 Power-on Tests .................................................................................................................................................................................22-2
22.3.2 Shift Check.........................................................................................................................................................................................22-3
22.3.3 Automated Tests..............................................................................................................................................................................22-3
22.3.4 User Test .............................................................................................................................................................................................22-4
22.3.5 Recorder Inspection.......................................................................................................................................................................22-6
22.3.6 ECG Cable Test..................................................................................................................................................................................22-7
22.3.7 Manual Defibrillation Test ............................................................................................................................................................22-7
22.3.8 Pacing Test .........................................................................................................................................................................................22-9
22.3.9 Peforming Testing in Installation Mode..................................................................................................................................22-9
22.3.10 NIBP Overpressure Protection Test...................................................................................................................................... 22-12
22.3.11 Electrical Safety Tests ............................................................................................................................................................... 22-12

23 Accessories ............................................................................................................................................................... 23-1


23.1 ECG Accessories.............................................................................................................................................................................................23-1
23.2 SpO2 Accessories ...........................................................................................................................................................................................23-3
23.3 NIBP Accessories............................................................................................................................................................................................23-5
23.4 Therapy Accessories.....................................................................................................................................................................................23-5
23.5 Miscellaneous.................................................................................................................................................................................................23-6

A Specifications ..............................................................................................................................................................A-1
A.1 General Specifications.....................................................................................................................................................................................A-1
A.2 Defibrillator Specifications ............................................................................................................................................................................A-2
A.3 Pacer Specifications .........................................................................................................................................................................................A-5
A.4 Monitor Specifications ....................................................................................................................................................................................A-5
A.5 Power Supply Specifications.........................................................................................................................................................................A-9
A.6 Recorder Specifications ............................................................................................................................................................................... A-10

6
A.7 Alarm Specifications......................................................................................................................................................................................A-10
A.8 Data Management Specifications ............................................................................................................................................................A-11
A.9 Environmental Specifications ....................................................................................................................................................................A-11

B EMC............................................................................................................................................................................... B-1

C BeneHeart Defibrillator Shift Checklist ..................................................................................................................... C-1

D Alarm Messages ..........................................................................................................................................................D-1


D.1 Physiological Alarm Messages ....................................................................................................................................................................D-1
D.2 Technical Alarm Messages ............................................................................................................................................................................D-2

E Electrical Safety Inspection......................................................................................................................................... E-1


E.1 Power Cord Plug.................................................................................................................................................................................................E-1
E.2 Device Enclosure and Accessories...............................................................................................................................................................E-1
E.3 Device Labelling.................................................................................................................................................................................................E-2
E.4 Protective Earth Resistance............................................................................................................................................................................E-2
E.5 Earth Leakage Test.............................................................................................................................................................................................E-2
E.6 Patient Leakage Current..................................................................................................................................................................................E-3
E.7 Mains on Applied Part Leakage....................................................................................................................................................................E-3
E.8 Patient Auxiliary Current .................................................................................................................................................................................E-4

F Symbols and Abbreviations........................................................................................................................................ F-1


F.1 Units ........................................................................................................................................................................................................................F-1
F.2 Symbols..................................................................................................................................................................................................................F-2
F.3 Abbreviations and Acronyms.........................................................................................................................................................................F-2

G Device Tracking ...........................................................................................................................................................G-1

7
FOR YOUR NOTES

8
1 Safety
1.1 Safety Information

DANGER
z Indicates an imminent hazard that, if not avoided, will result in death or serious injury.

WARNING
z Indicates a potential hazard or unsafe practice that, if not avoided, could result in death or serious injury.

CAUTION
z Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or
product/property damage.

NOTE

z Provides application tips or other useful information to ensure that you get the most from your product.

1-1
1.1.1 Dangers

DANGER
z The equipment delivers up to 360 J of electrical energy. Unless properly used as described in these
Operating Instructions, this electrical energy may cause serious injury or death. Do not attempt to operate
this defibrillator unless thoroughly familiar with these operating instructions and the function of all
controls, indicators, connectors, and accessories.

z Defibrillation current can cause operator or bystander severe injury or even death. Keep distance with the
patient or metal devices connected to the patient during defibrillation.

z Do not disassemble the defibrillator. It contains no operator serviceable components and dangerous high
voltages may be present. Contact authorized service personnel for repair.

z To avoid explosion hazard, do not use the equipment in the presence of oxygen-rich atmospheres,
flammable anesthetics, or other flammable agents (such as gasoline). Keep the equipment and the
operating environment dry and clean.

1.1.2 Warnings

WARNING
z Before putting the system into operation, the operator must verify that the equipment, connecting cables
and accessories are in correct working order and operating condition.

z Make sure the synchronous input system is applied to this equipment and the input signal is correct if
necessary.

z The equipment must be connected to a properly installed power outlet with protective earth contacts only.
If the installation does not provide for a protective earth conductor, disconnect it from the power line and
operate it on smart lithium-ion batteries.

z This equipment is used for single patient at a time.

z Medical electrical equipment which does not incorporate defibrillator protection should be disconnected
during defibrillation.

z Do not defibrillate a patient who lies on the wet ground.

z Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment of alarm volume to
a low level or off may result in a hazard to the patient. Remember that alarm settings should be customized
according to different patient situations and always keeping the patient under close surveillance is the most
reliable way for safe patient monitoring.

z Do not perform any functional check if the equipment is connected with a patient; otherwise the patient
might be shocked.

z Remain attentive to the patient during applying therapy. Delay in delivering a shock may result in a rhythm
that was analyzed as shockable converting spontaneously to non-shockable and could result in
inappropriate delivery of a shock.

1-2
z For the treatment of patients with implantable pacemakers, place therapy pads or paddles away from
internal pacemaker generator if possible to help prevent damage to the pacemaker.

z To avoid inadvertent disconnection, route all cables in a way to prevent a stumbling hazard. Wrap and
secure excess cabling to reduce risk of entanglement or strangulation by patients or personnel.

z Do not touch device connectors, recorder print head, battery connector or other live equipment if in contact
with the patient; otherwise patient injury may result.

z To ensure patient safety, use only parts and accessories specified in this manual.

z Package material may contaminate the environment. Properly dispose of the package material according to
applicable waste control regulations and keep it out of children’s reach.

1.1.3 Cautions

CAUTION
z Use of Manual Therapy security password requires the clinician to know and remember the password.
Failure to enter correct password will prevent the delivery of manual defibrillation, synchronized
cardioversion and pacing therapy.

z At the end of its service life, the equipment, as well as its accessories, must be disposed of in compliance
with the guidelines regulating the disposal of such products to avoid contaminating the environment.

z Magnetic and electrical fields are capable of interfering with the proper performance of the equipment. For
this reason make sure that all external devices operated in the vicinity of the equipment comply with the
relevant EMC requirements. Mobile phones, X-ray equipment or MRI devices are a possible source of
interference as they may emit higher levels of electromagnetic radiation.

z Before connecting the equipment to the power line, check that the voltage and frequency ratings of the
power line are the same as those indicated on the equipment’s label or in this manual.

z Always install or carry the equipment properly to avoid damage caused by drop, impact, strong vibration or
other mechanical force.

z Dry the equipment immediately in case of rain.

z Do not frequently charge the equipment and deliver shock in non-clinical situations. Otherwise equipment
damage could occur.

1.1.4 Notes

NOTE
z Put the equipment in a location where you can easily see the screen and access the operating controls.

z Keep this manual in the vicinity of the equipment so that it can be obtained conveniently when needed.

z If the equipment is run on a DC power supply, a DC/AC adapter we supply should be used.

z This manual describes all features and options. Your equipment may not have all of them.

1-3
1.2 Equipment Symbols

Caution (Attention, consult


Status indicator
accompanying documents)

Alternating current Battery indicator

Audio paused Alarm off

Audio off Alarm paused

Marker Lead select

Gain select Event summary

NIBP start/stop key Graphical recorder

Menu Unlocking

Network connector Shock button

USB connector Input/Output

Equipotentiality Protective earth (ground)

Serial number Dangerous voltage

1-4
Fragile Right side up

Keep dry Maximum stacks

Manufacturer Date of manufacture

General symbol for recovery/recyclable

Electrostatic sensitive devices

Mark of conformity to European Medical Device Directive 93/42/EEC

Authorised representative in the European community

DEFIBRILLATION-PROOF TYPE CF APPLIED PART

DEFIBRILLATION-PROOF TYPE BF APPLIED PART

Dispose of in accordance to your country’s requirements

1-5
FOR YOUR NOTES

1-6
2 The Basics
2.1 Overview
The BeneHeart (hereinafter called the equipment) is a lightweight and portable defibrillator/monitor. It provides four
operating modes: Monitor, Manual Defib, AED and Pacer.

In Monitor Mode, the equipment is intended for monitoring, displaying, reviewing, storing and printing multiple
physiological parameters and waveforms including ECG, pulse oximetry (SpO2), respiration (Resp), and non-invasive
blood pressure (NIBP).

In AED mode, the equipment automatically analyzes the patient’s ECG rhythm and indicates whether or not a shockable
rhythm is detected. Voice prompts provide easy-to-follow instructions and patient information to guide you through the
defibrillation process. Messages and flashing buttons are also presented to reinforce the voice prompts.

In the Manual Defib Mode, the operator analyzes the patient’s ECG, and, if appropriate, follows this procedure:

1 Select the Manual Defib mode, adjust the energy level if necessary;

2 Charge; and

3 Deliver the shock.

Defibrillation may be performed through paddles or multifunction electrode pads. In Manual Defib Mode, you can also
perform synchronized cardioversion. If desired, use of Manual Defib Mode may be password protected.

The Pacer Mode offers non-invasive transcutaneous pacing therapy. Pace pulses are delivered through multifunction
electrode pads. Use of Pacer Mode may also be password protected.

The equipment can be powered by smart lithium ion batteries which are rechargeable and maintenance-free. You can
easily determine the remaining battery charge by viewing the battery power gauge displayed on the screen or by
checking the indicator on the battery itself. An external AC mains or a DC power supply connected through an DC/AC
adapter may also be used as a power source and for continuous battery charging.

The equipment automatically stores patient data in an internal storage card. You can also export the data through the
USB port for viewing and editing on a PC through the data management software.

2.2 Intended Use


The equipment is intended for external defibrillation, internal defibrillation, synchronized cardioversion and
semi-automatic defibrillation (AED). It can also be used for non-invasive external pacing as well as ECG, SpO2, Resp, and
NIBP monitoring.

The equipment is for use in hospital and pre-hospital settings by qualified medical personnel trained in the operation of
the equipment and qualified by training in basic life support, advanced cardiac life support or defibrillation.
2-1
2.2.1 AED
The AED mode is to be used only on cardio arrest patients who are at least 8 years. The patients must be:

„ Unresponsive

„ Not breathing or not breathing normally

2.2.2 Manual Defibrillation


Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that
are pulseless and unresponsive. Synchronous defibrillation is intended for termination of atrial fibrillation.

2.2.3 Noninvasive Pacing


Noninvasive pacing therapy is intended for patients with symptomatic bradycardia. It can also be helpful in patients
with asystole, if performed early.

2.2.4 ECG
The ECG monitoring function is used to monitor and/or record the patient’s ECG waveform and heart rate.

2.2.5 Resp
The respiration monitoring function is used to continuously monitor the patient’s respiration rate and respiration
waveform.

2.2.6 SpO2
The SpO2 function is intended to measure patient’s oxygen saturation in arterial blood.

2.2.7 NIBP
The NIBP function is intended for non-invasive measurement of a patient’s arterial blood pressure.

2-2
2.3 Main Unit
2.3.1 Front View

External paddle

Handle

Alarm
lamp

Area 1

Area 3
Microphone

Area 2
Speaker

2-3
Area 1

2 3 4 5 6
1. Display screen

2. AC power indicator

‹ Illuminated: when AC mains is connected.

‹ Off: when AC mains is not connected.

3. Battery indicator

‹ Yellow: when the battery is being charged.

‹ Green: when the battery is fully charged or the equipment is run on battery.

‹ Off: when no battery is installed or battery fails.

4. Status indicator (red cross)

‹ Flashing: when a failure is detected or when battery is not installed if [No Battery] is configured as
[Status Indicator ON].

5. Status indicator (green tick)

‹ Illuminated: when AC mains is connected, and the equipment operates properly.

NOTE

z Both status indicators are off when AC mains is not connected if the equipment is turned off and no failure
is detected.

6. Soft keys

They are corresponding with the soft key labels located immediately above. The labels of the soft keys changes
according to the current operating mode.

2-4
Area 2

3 4 5 6 7
1. Lead Select button

Press this button to select the lead of the first ECG waveform.

2. Gain select button

Press this button to select the size of the first ECG waveform.

3. NIBP button (for equipments configured with NIBP function)

Press this button to start or stop NIBP measurements.

Record button (for equipments without NIBP function)

Press this button to start a recording or stop the current recording.

4. Alarm Pause button

Press this button to pause, reactivate or switch off the alarms.

5. Mark Event button

Press it to manually mark specified events. If a menu has been open, pressing this button will close the menu.

6. Main Menu button

If no menu is displayed on the screen, pressing it will enter the main menu. If there is a menu displayed, pressing it
will close that menu.

7. Navigation knob

You can:

‹ Rotate it clockwise or counterclockwise to move the cursor, or

‹ Press it to confirm the selection.

2-5
Area 3

1. Mode Select knob

Rotate this knob to select the operating mode or turn the equipment off.

2. Energy Select button

In Manual Defib mode, press this button to select energy level.

3. Charge button

Press this button to charge the defibrillator.

4. Shock button

Press this button to deliver a shock to the patient.

2-6
Recorder

1 2

3 5

1. Start/Stop Key

Press this key to start a recording or stop the current recording.

2. Indicator

‹ Illuminated: when the recorder works correctly.

‹ Flashes: when an error occurred to the recorder, or the recorder runs out of paper.

3. Paper outlet

4. Recorder door

5. Latch

2-7
2.3.2 Side View

Therapy port

Therapy port is used to connect paddles cable or pads cable.

2 4

1. Recorder

2. ECG: ECG cable connector

3. SpO2: SpO2 sensor connector

4. NIBP: NIBP cuff connector

2-8
2.3.3 Rear View

6
2

4
1. Hook

2. Battery

3. Equipotential grounding terminal

When the defibrillator/monitor and other devices are to be used together, their equipotential grounding terminals
should be connected together to eliminate the potential difference between them.

4. External power input

It connects an AC power cord or a DC/AC adapter to run the equipment respectively on the external AC mains or
DC power supply.

5. Multifunctional connector

It provides ECG output and defib synchronization input.

6. USB connector

7. Network connector

It is a standard RJ45 connector.

2-9
2.3.4 External Paddles

1 2 5

3
4

Sternum paddle Apex paddle

1. Shock button

2. Energy Select button

3. Shock indicator

4. Charge button

5. Shock button

2-10
2.4 Display Views
A typical screen in Manual Defib Mode is shown below.

1 2 3 4 5

6 7

1. Patient Information area

This area shows patient name, patient category, paced status, and current date and time.

‹ : indicates that the patient has an implanted pacemaker.

2. Alarm status symbols

indicates alarms are paused;

indicates all alarms are turned off;

indicates alarm sounds are paused;

indicates alarm sounds are turned off.

3. Physiological Alarm area

This area shows physiological alarm messages. When multiple alarms occur, they will be displayed circularly.

4. Technical Alarm area

This area shows technical alarm messages and prompt messages. When multiple messages come, they will be
displayed circularly.

5. Battery Status indicator

It indicates battery status. Refer to 18 Battery for details.

2-11
6. Waveform area

This area shows measurement waveforms. The waveform label is displayed at the upper left corner of the
waveform.

7. Parameter area

This area shows measurement parameters. Each measurement module has a parameter block and the parameter
name is displayed at the upper left corner.

8. Manual Defib information area

This area shows the selected defibrillation energy, shock counter as well as prompt related to manual defibrillation.

9. Runtime area

This area shows the equipment's operating time since it is turned on.

10. Prompt area

This area shows the prompt information.

11. Soft Key area

The three soft key labels correspond to the soft key buttons located immediately below. The labels of the soft keys
changes according to the current display view and function. Soft key labels appearing as blank indicate that the
soft key is inactive.

2-12
3 Basic Operations and Settings
3.1 Installation

WARNING
z The equipment shall be installed by personnel authorized by the manufacturer.

z The software copyright of the equipment is solely owned by the the manufacturer. No organization or
individual shall resort to juggling, copying, or exchanging it or to any other infringement on it in any form
or by any means without due permission.

z Devices connected to the equipment must meet the requirements of the applicable IEC standards (e.g. IEC
60950 safety standards for information technology equipment and IEC 60601-1 safety standards for medical
electrical equipment). The system configuration must meet the requirements of the IEC 60601-1-1 medical
electrical systems standard. Any personnel who connect devices to the equipment’s signal input/output
port is responsible for providing evidence that the safety certification of the devices has been performed in
accordance to the IEC 60601-1-1. If you have any question, please contact the the manufacturer.

z If it is not evident from the equipment specifications whether a particular combination is hazardous, for
example, due to summation of leakage currents, consult the manufacturers or else an expert in the field, to
ensure the necessary safety of all devices concerned will not be impaired by the proposed combination.

3.1.1 Unpacking and Checking


Before unpacking, examine the packing case carefully for signs of damage. If any damage is detected, contact the carrier
or the manufacturer. If the packing case is intact, open the package and remove the equipment and accessories carefully.
Check all materials against the packing list and check for any mechanical damage. If you have any question, please
contact us.

WARNING
z Package material may contaminate the environment. Properly dispose of the package material according to
applicable waste control regulations and keep it out of children’s reach.

z The equipment might be contaminated during storage and transport. Before use, please verify whether the
packages are intact, especially the packages of single use accessories. In case of any damage, do not apply it
to patients.

NOTE
z Save the packing case and packaging material as they can be used if the equipment must be reshipped.

3-1
3.1.2 Environmental Requirements
The operating environment of the equipment must meet the requirements specified in this manual.

The environment where the equipment is used shall be reasonably free from noises, vibration, dust, corrosive,
flammable and explosive substances. If the equipment is installed in a cabinet, sufficient space in front and behind shall
be left for convenient operation, maintenance and repair. Moreover, to maintain good ventilation, the equipment shall
be at least 2 inches (5 cm) away from around the cabinet.

When the equipment is moved from one place to another, condensation may occur as a result of temperature or
humidity difference. In this case, never start the system before the condensation disappears.

NOTE
z Make sure that the operating environment of the equipment meets the specific requirements. Otherwise
unexpected consequences, e.g. damage to the equipment, could result.

3.2 Basic Operation


3.2.1 Turning Power On
Once the equipment is installed, you can get ready for monitoring and therapy:

1. Before turning on the equipment, check for any mechanical damage and make sure that all external cables,
plug-ins and accessories are properly connected.

2. Plug the power cord into the AC power source. If you run the equipment on battery power, ensure that the battery
is sufficiently charged. If you run the equipment on DC power supply, an DC/AC adapter we supply should be used.

3. Turn the Mode Select knob to enter the desired working mode. After the start-up screen is displayed, the system
gives a beep, and meanwhile, the alarm lamp is illuminated in yellow, and then turns red, and then turns off.

4. The equipment enters the screen of selected mode.

WARNING
z Do not use the equipment for any monitoring or therapy procedure on a patient if you suspect it is not
working properly, or if it is mechanically damaged. Contact your service personnel or us.

3.2.2 Starting Monitoring or Therapy


1. Decide which measurements or therapy you want to make.

2. Check that the patient cables and sensors are correct.

3. Connect the required patient cables and sensors.

4. Enter the appropriate operating mode and check that the settings are proper for your patient.

Refer to the appropriate sections for details of how to perform patient monitoring and therapy.

3-2
3.2.3 Disconnecting from Power
To disconnect the equipment from the AC power source, follow this procedure:

1. Confirm that the patient monitoring or therapy is completed.

2. Disconnect the patient cables and sensors from the patient.

3. Make sure to save or clear the patient data as required.

4. Turn the Mode Select Knob to Off. After 10 seconds, the equipment is shut down.

3.2.4 Auto Restoring to Last Configuration


During operation, you may make changes to some settings. However, these changes may not be saved as user
configuration. To prevent the changes from losing in case of sudden power failure, the equipment saves the settings in
real time. The saved settings are the latest configuration. In case of power failure, the equipment loads the latest
configuration if restarts within 60 seconds; it load the user configuration if restarts 120 seconds later after the power
failure; it may load either the latest configuration or the user configuration if restarts between 60 and 120 seconds after
the power failure.

3.3 Using the Main Menu


To enter the main menu, press the Main Menu button on the equipment’s front panel.

Other menus are similar to the main menu and contain the following parts:

1. Heading

2. Main body: displays options, buttons, prompt messages, etc. Pressing the menu button with “>>’’ enters a
submenu to reveal more options or information.

3. Exit button

3.4 Changing General Settings


3.4.1 Setting the Date and Time
1. Press the Main Menu button on the front panel, and then select [Others >>]→ [Configuration >>]→enter the
required password.

2. Select [General Setup >>].

3. Select [Date Format] from [yyyy-mm-dd], [mm-dd-yyyy] and [dd-mm-yyyy].

4. Select [Time Format] and toggle between [24h] and [12h].

5. Set [System Time].

3-3
You can also set system time by selecting [Configuration >>]→[View Config]→[General Setup >>]. However, you
cannot select date format and time format in this case. After the completion of setting system time, exit the
configuration mode, and then the system will restart.

3.4.2 Adjusting the Screen Brightness


1. Press the Main Menu button on the front panel, and then select [Others >>].

2. Set [Brightness] to an appropriate level: 10 is the brightest, and 1 is the least bright.

You can also change screen brightness by entering configuration mode and selecting [Others] from the Configuration
Main menu.

3.4.3 Changing Key Volume


1. Press the Main Menu button on the front panel, and then select [Others >>].

2. Select [Key Volume] and then select an appropriate value. 0 means key volume off and 10 is the maximum volume.

You can also change key volume by entering configuration mode and selecting [Others] from the Configuration Main
menu.

3.4.4 Selecting High Contrast Mode


The equipment has the function of high contrast display so that the user can view the display under high ambient
illumination.

To enable the High Contrast display,

„ In the Monitor, Manual Defib and Pacer mode, press the Main Menu button on the front panel, and then select
[High Contrast]. To disable high contrast display, select [Full Color] in the Main Menu.

„ In AED mode, press the [High Contrast] soft key. To disable high contrast display, press the [Full Color] soft key.

Once High Contrast is selected, the system remains in high contract mode when you change the operating mode.
However, the setting will not be saved if the equipment is turned off.

3-4
4 Managing Patients
4.1 Overview
Patient information management function enables you to edit and manage information of the current patient.

4.2 Editing Patient Information


You can edit patient information in Monitor, Manual Defib and Pacer mode.

To edit patient information,

1. Press the Main Menu button on the front panel, and then select [Patient Demographics >>] and then make
changes as desired.

2. Select [Others >>] to edit more patient information.

Archive ID is created automatically when the equipment is turned on. It is unchangeable. When the equipment is turned
off, the current patient is discharged and the archive ID turns to be historical archive ID

For a new patient, if patient category is changed, the system will restore the default alarm settings of this patient
category; if patient category is not changed, the alarm settings remain unchanged. If you restart the equipment after
normal power-off, the default alarm settings will be loaded.

4-1
FOR YOUR NOTES

4-2
5 Alarms
Alarms, triggered by a vital sign that appears abnormal or by technical problems of the equipment, are indicated to the
user by visual and audible alarm indications.

WARNING
z A potential hazard exists if different alarm presets are used for the same or similar device in any single area,
e.g. an intensive care unit or cardiac operating room.

5.1 Alarm Categories


By nature, the equipment’s alarms can be classified into three categories: physiological alarms, technical alarms and
prompt messages.

1. Physiological alarms

Physiological alarms, also called patient status alarms, are triggered by a monitored parameter value that violates
set alarm limits or by an abnormal patient condition. Physiological alarm messages are displayed in the
physiological alarm area. In AED mode, no physiological alarm will be presented.

2. Technical alarms

Technical alarms, also called system status alarms, are triggered by a device malfunction or a patient data distortion
due to improper operation or system failure. Technical alarm messages are displayed in the technical alarm area.

3. Prompt messages

As a matter of fact, prompt messages are not alarm messages. Apart from the physiological and technical alarms,
the equipment also shows some messages indicating system status. Messages of this kind are usually displayed in
the prompt area. Therapy-related prompts are shown in corresponding information area. Some special prompts are
shown in dialog boxes.

5-1
5.2 Alarm Levels
By severity, alarms can be classified into three categories: high level alarms, medium level alarms and low level alarms.

Physiological alarms Technical alarms


High level Indicate that your patient is in a life Indicate a severe device malfunction or an improper operation,
threatening situation, such as Asystole, which could make it possible that the equipment cannot detect
Vfib/Vtac and so forth, and an critical patient status or may cause therapy failed, and thus
emergency treatment is demanded. threaten the patient’s life, such as low battery.
Medium Indicate that your patient’s vital signs Indicate a device malfunction or an improper operation, which
level appear abnormal and an immediate may not threaten the patient’s life but may compromise patient
treatment is required. monitoring or therapy.
Low level Indicate that you patient’s vital signs Indicate a device malfunction or an improper operation, which
appear abnormal and an immediate may compromise a certain function but will not threaten the
treatment may be required. patient’s life.

5.3 Alarm Indicators


When an alarm occurs, the equipment indicates it to the user through visual or audible alarm indications.

„ Alarm lamp

„ Alarm tones

„ Alarm message

„ Flashing numeric

NOTE
z When multiple alarms of different levels occur simultaneously, the equipment will select the alarm of the
highest level and give visual and audible alarm indications accordingly. Alarm messages will be displayed
circularly.

z Some physiological alarms, such as Asystole, are exclusive. They have identical alarm tones and alarm lights
with normal high level physiological alarms, but their alarm messages are displayed exclusively. That is to
say, when an exclusive physiological alarm and a normal high level physiological alarms are triggered
simultaneously, only alarm message of the exclusive physiological alarm is displayed.

5.3.1 Alarm Lamps


If an alarm occurs, the alarm lamp will flash. The color and flashing frequency match the alarm level as follows:

„ High level alarms the lamp quickly flashes red.

„ Medium level alarms the lamp slowly flashes yellow.

„ Low level alarms the lamp lights yellow without flashing.

5-2
5.3.2 Audible Alarms
The equipment uses different alarm tone patterns to match the alarm level:

„ High level alarms triple+double+triple+double beeps.

„ Medium level alarms triple beeps.

„ Low level alarms single beep.

5.3.3 Alarm Message


When an alarm occurs, the alarm message will appear in the technical or physiological alarm area. For physiological
alarms, the asterisk symbols (*) before the alarm message match the alarm level as follows:

„ High level alarms ***

„ Medium level alarms **

„ Low level alarms *

Additionally, the alarm message has different background color which matches the alarm level.
For physiological alarms

„ High level alarms red

„ Medium level alarms yellow

„ Low level alarms yellow

For technical alarms

„ High level alarms: red

„ Medium level alarms: yellow

„ Low level alarms: blue

5.3.4 Flashing Numeric


If an alarm triggered by an alarm limit violation occurs, the numeric of the measurement in alarm will flash every second,
and corresponding alarm limit will also flash at the same frequency indicating the alarm limit is violated.

5.3.5 Alarm Status Symbols


Apart from the aforementioned alarm indicators, the equipment still uses the following symbols telling the alarm status:

„ indicates all alarms are paused.

„ indicates all system alarms are turned off.

„ indicates alarm sounds are paused.

„ indicates alarm sounds are turned off.

5-3
5.4 Alarm Tone Configuration
5.4.1 Changing the Alarm Volume
1. Press the Main Menu button on the front panel, and then select [Alarm Setup >>]→[Alm Volume >>].

2. Set [Alm Volume] to an appropriate level:

‹ If [Audio Off] is enabled, alarm volume can be set to a value between 0 and 10, in which 0 means audio off and
10 the maximum volume level.

‹ If [Audio Off] is disabled, alarm volume can be set to a value between 1 and 10, in which 1 is the minimum
volume level and 10 the maximum.

To disable or enable [Audio Off], access configuration management.

The setting of alarm volume will not be saved when the system is power off.
You can also set alarm volume in configuration mode. In this case, the setting will be saved.

NOTE
z You cannot adjust alarm volume when an alarm is switched off.

5.4.2 Setting the Interval between Alarm Sounds


1. Press the Main Menu button on the front panel, and then select [Others >>]→ [Configuration >>]→enter the
required password.

2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.

3. Respectively set [High Alarm Interval (s)], [Med Alarm Interval (s)] and [Low Alarm Interval (s)].

WARNING
z Do not rely exclusively on audible alarm system. Setting alarm volume to a low level may result in a hazard
to the patient. Always keep the patient under close surveillance.

5.5 Understanding the Alarm Setup Menu


Press the Main Menu button on the front panel, and then select [Alarm Setup >>] to enter the Alarm Setup menu.

5-4
5.5.1 Setting Alarm Properties for All Parameters
In the main menu, select [Alarm Setup >>]→[Para. Alarm >>] to enter the Para. Alarm setup menu, where you can
review and set alarm limits, alarm switches, alarm level and alarm recordings for all parameters.

„ When a parameter alarm is switched on, the equipment gives alarm indications in accordance with the preset alarm
level and stores related waveforms and parameter values.

„ When a parameter alarm is switched off, the alarm off symbol “ ” is displayed in the parameter window.

„ If the measurement’s [On/Off] and [Record] are set to [On], automatic recording of all the measurement numerics
and related waveforms is possible when a measurement alarm occurs.

To reset the default settings, select the [Defaults] button.

NOTE
z You cannot simultaneously switch on HR and PR alarms. In the case that PR alarm is on, switching on HR
alarm will automatically turn off PR alarm, and vice versa.

You can also set parameter alarm properties by selecting a parameter window and select [Para. Alarm >>] in the
pop-up menu.

WARNING
z Make sure that the alarm limits settings are appropriate for your patient before patient monitoring.

z Setting the alarm limit to an extreme value may cause the alarm system to be ineffective. For example, high
oxygen levels may predispose a premature infant to retrolental fibroplasia. If this is a consideration, do NOT
set the SpO2 high alarm limit to 100%, which is equivalent to switching the alarm off.

5-5
5.5.2 Adjusting Alarm Limits Automatically
The defibrillator/monitor can automatically adjust the patient’s alarm limits according to the measured vital signs. When
[Auto Limits] is selected, the equipment automatically calculates alarm limits based on the latest measured parameter
values.

To enable auto alarm limits, press the Main Menu button on the equipment’s front panel, and then select [Alarm Setup
>>]→[Para. Alarm>>]→[Auto Limits].
You can also access [Auto Limits] by select a parameter window to enter the [Para. Alarm] menu.

When auto alarm limits have been applied, you can manually adjust the alarm limits through the Para. Alarm Setup
menu so that they are appropriate for your patient.

The defibrillator/monitor calculates the auto limits based on the following rules.
Low alarm limit High alarm limit
Auto alarm limits
Module Parameter Adult/ Adult/
Neonate Neonate range
pediatric pediatric
HR×0.8 or 40 bpm (HR-30) or 90 bpm HR×1.25 or 240 (HR + 40) or 200 Adult/pediatric: 35 to
ECG HR (whichever is (whichever is bpm whichever is bpm whichever is 240
greater) greater) lower) lower) Neonate: 55 to 225
RR×0.5 or 6/min (RR-10) or 30/min RR×1.5 or 30/min (RR+25) or 85/min Adult/pediatric:
Resp RR (whichever is (whichever is (whichever is (whichever is 6 to 55
greater) greater) lower) lower) Neonate: 10 to 90
Same as the Same as the Same as the Same as the default Same as the
SpO2
default alarm limit default alarm limit default alarm limit alarm limit measurement range
SpO2 PR×0.8 or 40 bpm (PR-30) or 90 bpm PR×1.25 or 240 (PR + 40) or 200 Adult/pediatric: 35 to
PR (whichever is (whichever is bpm whichever is bpm whichever is 240
greater) greater) lower) lower) Neonate: 55 to 225
(SYS-15) or 45 (SYS + 15) or 105 Adult: 45 to 270
SYS × 0.68 + 10 SYS × 0.86 + 38
NIBP-S mmHg (whichever mmHg (whichever Pediatric: 45 to 185
mmHg mmHg
is greater) is lower) Neonate: 35 to 115
(Dia-15) or 20 (Dia + 15) or 80 Adult: 25 to 225
Dia × 0.68 + 6 Dia × 0.86 + 32
NIBP NIBP-D mmHg (whichever mmHg (whichever Pediatric: 25 to 150
mmHg mmHg
is greater) is lower) Neonate: 20 to 90
(Mean-15) or 35 (Mean + 15) or 95 Adult: 30 to 245
Mean × 0.68 + 8 Mean × 0.86 + 35
NIBP-M mmHg (whichever mmHg (whichever Pediatric: 30 to 180
mmHg mmHg
is greater) is lower) Neonate: 25 to 105

NOTE
z You can enable auto alarm limits only when the current parameter measurement is within the auto alarm
limits range.

5-6
5.6 Pausing Alarms
You can temporarily disable alarm indicators by pressing the hardkey on the equipment’s front. When alarms are
paused:

„ For physiological alarms, no alarm indication is shown. New physiological alarm will not be presented.

„ The remaining alarm pause time is displayed in the physiological alarm area.

„ For technical alarms, alarm sounds are paused, but alarm lamps and alarm messages remain presented.

„ is displayed in the sound symbol area. If a new technical alarm is triggered in the alarm paused period, the

alarm message will be displayed.

When the alarm pause time expires, the alarm paused status is automatically deactivated. You can also cancel the alarm

paused status by pressing the hardkey.

The default alarm pause time is 2 minutes. To change alarm pause time,

1. Press the Main Menu button on the front panel, and then select [Others >>]→ [Configuration >>]→enter the
required password.

2. Select [Alarm Setup >>]→[Alarm Pause Time] and then select an appropriate value.

5.7 Switching Alarms Off


When an alarm is switched off, the alarm status is the same with that when an alarm is paused.

Alarms are switched off when:

„ The hardkey is pressed if the [Alarm Pause Time] is set to [Permanent].

„ The equipment is switched to the Manual Defib mode. Or

„ The equipment exit synchronous defibrillation when operating in the Manual Defib mode.

The alarm off status is exited when,

„ The hardkey is pressed, or

„ Sync Defib is switched on in Manual Defib mode.

5-7
5.8 Pausing Alarm Sounds

You can press the [Audio Pause] softkey to pause alarm tones. In this case, the symbol will be displayed in the

sound symbol area indicating all system sounds are silenced temporarily. In the audio paused status, all alarm indicators
except audible alarm tones works properly. If alarms are switched off or paused, the [ Audio Pause] soft key is not
shown.

The audio paused status is exited when,

„ The [ Audio On] soft key is pressed.

„ Other alarms occur. Or

„ The hardkey is pressed to permanently or temporarily switch off the alarms.

5.9 Switching Off Alarm Sounds


In the event that [Audio Off] is enabled, to switch off the alarm tone, set [Alm Volume] to 0 while operating in Monitor

mode, Manual Defib mode or Pacer mode. In the audio off status, appears in the sound symbol area. In this case,

the alarm status is the same with that when alarm tones are paused.

The audio off status is exited when:

„ The hardkey is pressed. In this case, the equipment enters the alarm paused status and the alarm volume is

reset to the default level. The symbol is displayed in the sound symbol area.

„ The [Audio Pause] soft key is pressed. In this case, the equipment enters the audio paused status and the alarm

volume is reset to the default level. The symbol is displayed in the sound symbol area.

„ Operating mode is switched. Then the equipment enters the default alarm status of corresponding operating mode.
Or

„ Alarm volume is changed to a value between 1 and 10.

5.10 Reminder Tones


When alarms or alarm sounds are turned off, the equipment can give a reminder tone of a single beep every 60 seconds.

The reminder tone is switched off by default. You can switch it on by selecting [Alarm Setup]→[Reminder Tone]
through the Configuration Management Main menu. You can also change reminder volume. The default reminder
volume is [Med].

5-8
5.11 Latching Alarms
The latching setting for physiological alarms defines how alarm indicators behave when you do not acknowledge them.

„ If an alarm is latched, alarm indications remain presented even though alarm conditions end, except that:

‹ The parameter reading and violated alarm limit stop flashing.

‹ The time when the alarm is last triggered is displayed behind the alarm message.

„ If an alarm is not latched, the alarm indications disappear as soon as the alarm conditions end.

To latch a physiological alarm,

1. Press the Main Menu button on the front panel, and then select [Others >>]→ [Configuration >>]→enter the
required password.

2. Select [Alarm Setup] and set [Latching Alarms] to [Yes].

Only physiological alarms can be latched.

You can clear the latched alarms by pressing the hardkey.

5.12 Clearing Technical Alarms


For some technical alarms, their alarm lamp flashing and alarm tones are cleared and the alarm messages change to

prompt messages after the hardkey or [Audio Pause] soft key is pressed. After the equipment restores the normal

alarm status, it can give alarm indications correctly in case these alarms are triggered again.

For some technical alarms, all their alarm indications are cleared after the hardkey or [Audio Pause] soft key is

pressed. After the equipment restores the normal alarm status, it can give alarm indications correctly in case these
alarms are triggered again.

For others, their alarm tones are cleared but the alarm lamp flashing and alarm messages remain after the hardkey

or [Audio Pause] soft key is pressed. After the equipment restores the normal alarm status, all the alarm indications will
continue if the alarm conditions still present.

5-9
5.13 When an Alarm Occurs
When an alarm occurs, observe the following steps and take proper actions:

1. Check the patient’s condition.

2. Confirm the alarming parameter or alarm category.

3. Identify the alarm source.

4. Take proper action to eliminate the alarm condition.

5. Make sure the alarm condition is corrected.

For actions taken with regard to specific alarms, see Appendix D Alarm Messages.

5-10
6 Monitoring ECG
6.1 Overview
The electrocardiogram (ECG) measures the electrical activity of the heart and displays it as waveforms and numerics. The
equipment enables ECG monitoring through 3-, 5- lead ECG sets, external paddles and multifunction electrode pads. If
both ECG sets and paddles/pads are connected, the configured ECG waveforms are displayed in the waveform area.

6.2 Safety

WARNING
z Periodically inspect the electrode application site to ensure skin quality. If the skin quality changes, replace
the electrodes or change the application site.

z Use defibrillation-proof ECG cables during defibrillation.

z When monitoring a patient implanted with a pacemaker, be sure to select correct paced status. Otherwise,
the pacing pulses may be counted in the case of cardiac arrest or some arrhythmias. Do not completely rely
on the heart rate reading or the heart rate alarms. Always keep paced patients under close surveillance.

NOTE
z When connecting electrodes and/or patient cables, make sure that the connectors never come into contact
with other conductive parts, or with earth. Particularly make sure that all of the ECG electrodes are attached
to the patient.

z Interference from a non-grounded instrument near the patient and electrosurgery interference may cause
problems with the waveform.

z If selected lead cannot provide valid ECG signals, a dash line is shown in the ECG waveform area.

z Avoid using external paddles for ECG monitoring if possible.

z Use the same type of ECG electrodes when monitoring ECG through ECG lead set.

6-1
6.3 Monitoring View

You can access Monitor mode by switching the Mode Select knob to the Monitor position. When operating in Monitor
mode, the equipment displays up to two ECG waveforms, the heart rate reading, other available parameter values and
active alarm settings.

6.4 Preparing to Monitor ECG


6.4.1 ECG Monitoring with Electrodes
1. Prepare the patient’s skin. Proper skin preparation is necessary for good signal quality at the electrode, as the skin is
a poor conductor of electricity. To properly prepare the skin, choose flat areas and then follow this procedure:

‹ Shave hair from skin at chosen sites.

‹ Gently rub skin surface at application sites to remove dead skin cells.

‹ Thoroughly clean the sites with mild soap and water. We do not recommend using ether or pure alcohol,
because this dries the skin and increases the resistance.

‹ Dry the skin completely before applying the electrodes.

2. Attach the clips or snaps to the electrodes before placing them.

3. Place the electrodes on the patient.

4. Attach the lead wires to the ECG trunk cable and then plug the trunk cable into the equipment’s ECG connector.

5. Connect the ECG trunk cable to the equipment.

6. Switch the Mode Select knob to Monitor.

6.4.1.1 Placing Electrodes

3-Lead Placement

The following is a typical AHA electrode placement for a 3-lead ECG set:

„ RA placement: directly below the clavicle and near the right shoulder.

„ LA placement: directly below the clavicle and near the left shoulder.

„ LL placement: on the left lower abdomen.

6-2
5-Lead Placement

The following is a typical AHA electrode placement for a 5-lead ECG set:

„ RA placement: directly below the clavicle and near the right shoulder.

„ LA placement: directly below the clavicle and near the left shoulder.

„ RL placement: on the right lower abdomen.

„ LL placement: on the left lower abdomen.

„ V placement: on the chest.

The chest (V) electrode can be placed on one of the following positions:

„ V1 placement: on the fourth intercostal space at the right sternal border.

„ V2 placement: on the fourth intercostal space at the left sternal border

„ V3 placement: midway between the V2 and V4 electrode positions.

„ V4 placement: on the fifth intercostal space at the left midclavicular line.

„ V5 placement: on the left anterior axillary line, horizontal with the V4 electrode
position.

„ V6 placement: on the left midaxillary line, horizontal with the V4 electrode


position.

„ V3R-V6R placement: on the right side of the chest in positions corresponding to


those on the left.

„ VE placement: over the xiphoid process.

„ V7 placement: on posterior chest at the left posterior axillary line in the fifth intercostal space.

„ V7R placement: on posterior chest at the left posterior axillary line in the fifth intercostal space.

Electrode Placement for Surgical Patients

The surgical site should be taken into consideration when placing electrodes on a surgical patient, e.g. for open-chest
surgery, the chest electrodes can be placed on the lateral chest or back. To reduce artifacts and interference from
electrosurgical units, you can place the limb electrodes close to the shoulders and lower abdomen and the chest
electrodes on the left side of the mid-chest. Do not place the electrodes on the upper arm. Otherwise, the ECG
waveform will be very small.

WARNING
z When using electrosurgical units (ESU), place ECG electrodes between the ESU and its grounding plate to
prevent unwanted burns. Never entangle ESU cable and ECG cable together.

z When using electrosurgical units (ESU), never place ECG electrodes near to the grounding plate of the ESU,
as this can cause a lot of interference on the ECG signal.

6-3
6.4.2 ECG Monitoring with Paddles/Pads
1. Prepare the patient’s skin.

2. Apply the paddles/pads to the patient.

‹ If multifunction electrode pads are used, apply pads according to the instructions for use indicated on pads
package. Use anterior-lateral placement.

‹ If external paddles are used, remove the paddle set from the paddle tray by grasping the handles and pulling
them straight up. Apply conductive gel to paddle electrodes. Place the paddles to the patient’s chest using the
anterior-lateral placement.

3. If multifunction electrode pads are used, connect the pads to the pads cable.

4. Connect paddles/pads cable with the equipment if not connected.

Anterior-lateral Paddles/Pads placement

1. Place the RA pad or Sternum paddle on the patient’s upper right torso, lateral to the sternum and below the clavicle,
as shown below.

2. Place the LL pad, or Apex paddle to the patient’s left nipple in the midaxillary line, with the center of the electrode
in the midaxillary line, if possible. See the figure below.

Sternum
RA
paddle

LL
Apex paddle

NOTE
z Anterior - lateral placement is the only placement that can be used for ECG monitoring with paddles/pads
accessories.

6-4
6.4.3 Checking Paced Status

It is important to set the paced status correctly when you start monitoring ECG. The paced symbol is displayed in

the patient information area when the [Paced] is set to [Yes]. The pace pulse markers “︱” are shown on the ECG wave
when the patient has a paced signal.

To change the paced status, you can select either:

„ [Main Menu]→[Patient Demographics >>]→[Paced], or

„ The ECG parameter window to enter the [ECG Setup] menu, and then, select [Others >>]→[Paced],

and toggle between [Yes] and [No].

Warning
z For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the patient monitor could
mistake a pace pulse for a QRS and fail to alarm when the ECG signal is too weak. Do not rely entirely on rate
meter alarms when monitoring patients with pacemakers. Always keep these patients under close
surveillance.

z For non-paced patients, you must set [Paced] to [No]. If it is incorrectly set to [Yes], the equipment may be
unable to detect premature ventricular beats (including PVCs).

6.5 ECG Display


The figure below shows the ECG monitoring view in 5-lead mode. It is for reference only. Your display may be configured
to look slightly different.
Lead label Gain Filter mode

Heartbeat icon

Alarm limits

PVC values

HR value
PVCs values is shown only when arrhythmia analysis is switched on. When external paddles or multifunctional electrode
pads are used for ECG monitoring, the PVCs values is shown as “---”.
6-5
6.6 Changing ECG Settings
6.6.1 Changing Lead Settings
6.6.1.1 Selecting Lead Type

1. Select the ECG parameter area to enter the [ECG Setup] menu.

2. Select [Lead Set] and toggle between [3-lead] and [5-lead].

You can also set lead type in the configuration mode:

1. Press the Menu button on the equipment’s front panel. In the Main Menu, select [Others >>]→[Configuration
>>]→enter the required password to enter the Configuration Main menu.

2. Select [ECG Setup]→[Lead Set] and toggle between [3-lead] and [5-lead].

6.6.1.2 Choosing AHA or IEC Lead Placement

1. Select [Main Menu]→[Others >>]→[Configuration >>]→enter the required password.

2. In the Configuration Main menu, select [ECG Setup]→[ECG Standard], and then select [AHA] or [IEC] according to
the standard that is applied to your hospital.

6.6.1.3 Choosing the Heart Rate Source


To compute heart rate and to detect and analyse arrhythmia more accurately, you can choose a lead of best quality
signals as the HR calculation lead. To do so, select [ECG1] from the [ECG Setup] menu and then select the lead you want.

The selected lead should have the following characteristics:

„ The QRS should be either completely above or below the baseline and it should not be biphasic.

„ The QRS should be tall and narrow.

The P-waves and T-waves should be less than 0.2mV.

6-6
6.6.2 Changing ECG Wave Settings
You can select the ECG parameter area to enter the [ECG Setup] menu to set ECG cascade and wave speed. You can also
select the hot keys above the ECG waveform to change ECG lead, size and filter.

‹ You can press the Lead Select button on the equipment’s front panel or use the Navigation knob to select the
lead hot key above the first ECG waveform to select a lead.

„ If the wave is too small or clipped, you can change its size by adjusting the gain select button on the front panel or
selecting the Size hot key above the ECG waveform.

„ When monitoring ECG through ECG lead set, filter mode is displayed above the first ECG waveform. The available
filter mode settings are [Monitor], [Therapy], and [Diagnostic]. The default is [Therapy]. To change filter mode,
select the filter mode hot key using the Navigation knob.

‹ When monitoring ECG through pads/paddles, filter mode is always “Therapy” and is not shown..

„ In the [ECG Setup] menu, select [Sweep], and then choose an appropriate value. The faster the wave sweeps, the
wider the wave is. You can also access [Sweep] to adjust wave speed from the [ECG Setup] menu in the
Configuration mode.

„ In the [ECG Setup] menu, select [ECG Cascade] and toggle between [On] and [Off].

NOTE
z For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the system may mistake an
internal pace pulse for a QRS or fail to alarm when the pacer is broken.

6.6.3 Switching the Notch Filter On or Off


The notch filter removes the AC line noise. When [Filter] is set to [Monitor] or [Therapy], the notch filter is always on.
When [Filter] is set to [Diagnostic], you can switch the notch filter on or off as required.

To switch on or off notch filter, select the ECG parameter area to enter the [ECG Setup] menu, and then select [Others
>>]→[Notch Filter] and toggle between [On] and [Off].

Switching on the notch filter is recommended when there is interference with the waveform.

Set notch frequency according to the electric power frequency of your country. Follow this procedure:

1. Press the Main Menu button on the front panel, and then select [Others >>]→ [Configuration >>]→enter the
required password.

2. Select [ECG Setup]→[Notch Filter] and then select [50Hz] or [60Hz] according to the power frequency.

6-7
6.6.4 Adjusting Heartbeat Volume
In the case that ECG alarm is switched on, or both ECG alarm and PR alarm are switched off, heartbeat tone is issued.

To adjust the heartbeat volume,

„ Select the ECG parameter window to enter the [ECG Setup] menu, and then select [Others >>]→[QRS Volume], or

„ From the Configuration Main menu, select [ECG Setup] →[QRS Volume],

and then select an appropriate setting. The heartbeat volume can be set between 0 and 10, in which 0 means off, and 10
is the maximum volume.

When a valid SpO2 value exists, the system will adjust the pitch of the heartbeat tone according to the SpO2 value.

6.7 Arrhythmia Analysis


Arrhythmia analysis provides information about your patient’s condition, including heart rate and arrhythmia alarms.

NOTE
z Arrhythmia analysis program is intended to detect ventricular arrhythmias. It is not designed to detect atrial
or supraventricular arrhythmias. It may incorrectly identify the presence or absence of an arrhythmia.
Therefore, a physician must analyze the arrhythmia information with other clinical findings.

z The arrhythmia detection is not intended for neonatal patients.

6.7.1 Understanding the Arrhythmia Events

Arrhythmia event Description Category

Asystole No QRS complex for 4 consecutive seconds (in absence of ventricular fibrillation Lethal
or chaotic signals). arrhythmia
V-Fib/V-Tach A fibrillatory wave for 4 consecutive seconds.
A dominant rhythm of adjacent Vs and a HR > the V-Tach Heart Rate Limit.
Vtac The consecutive PVCs > Vtac PVCs limit, and the HR > the Vtac HR limit.
Vent. Brady The consecutive PVCs ≥ the Vbrd threshold and the ventricular HR < the Vbrd
Rate threshold.
Extreme Tachy The heart rate is greater than the extreme tachycardia limit.
Extreme Brady The heart rate is less than the extreme bradycardia limit.
PVCs/min PVCs/min exceeds high limit Nonlethal
PNP** No pace pulse detected for 1.75 x average R-to-R intervals following a QRS arrhythmia
complex (for paced patients only).
PNC** No QRS complex detected for 300 milliseconds following a pace pulse (for
paced patients only).
PVC One PVC detected in normal heartbeats.
Couplet Paired PVCs are detected.
VT>2 More than 2 consecutive PVCs within the last minute.

6-8
Arrhythmia event Description Category

Bigeminy A dominant rhythm of N, V,N, V, N, V.


Trigeminy A dominant rhythm of N, N, V,N, N, V, N, N, V.
R ON T R on T detected in normal heartbeats.
Missed Beats* No beat detected for 1.75x average R-R interval for HR <120, or No beat for 1
second with HR >120 (for non-paced patients only), or No beat detected for
more than the set pause threshold.
Brady The average heart rate is less than 60 bpm.
Tachy The average heart rate is greater than 100 bpm.
Vent. Rhythm The consecutive PVCs > the Vbrd PVCs limit, and the HR is between Vbrd Rate
limit and the Vtac Rate limit.
Multif. PVC Multiform PVCs detected in Multif. PVC's Window (which is adjustable).
Nonsus. Vtac The consecutive PVCs < the Vtac PVCs limit but > 2, and HR > the Vtac Rate
limit.
Pause* No QRS detected within the set time threshold of pause.
Irr. Rhythm Consistently irregular rhythm.
*: indicates that this arrhythmia alarm is not presented when [Paced] is set to [Yes].
**: indicates that this arrhythmia alarm is not presented when [Paced] is set to [No].

„ When multifunctional electrode pads are used for ECG monitoring, the equipment provides only 4 arrhythmia
alarms, including asystole, ventricular fibrillation/ventricular tachycardia, PNP, and PNC.

„ When paddles are used, the equipment provides only 3 arrhythmia alarms, including ventricular
fibrillation/ventricular tachycardia, PNP, and PNC.

6.7.2 Switching Arrhythmia Analysis On and Off


To switch arrhythmia analysis on or off:

1. Select the ECG parameter area to enter the [ECG Setup] menu. Select [Arrhythmia >>].

2. Select [Arrhthmia] and toggle between [On] and [Off].

You can also switch arrhythmia analysis on or off in the Configuration mode. Follow this procedure: Enter the
Configuration Main menu. Select [ECG Setup]→[Arrhythmia] and toggle between [On] and [Off].

6-9
6.7.3 Changing Arrhythmia Alarm Settings
To change arrhythmia alarm settings, select the ECG parameter area to enter the [ECG Setup] menu, and then select
[Arrhythmia >>]→[Arrh. Alarm] menu, where you can set alarm switch, alarm level and alarm record switch for all the
arrhythmia events..

You can also set arrhythmia alarm properties in the Configuration mode through the ECG Setup menu.

NOTE
z The alarm level for asystole, ventricular fibrillation, ventricular tachycardia, ventricular bradycardia,
extreme bradycardia, and extreme tachycardia alarms is always high and unchangeable. These alarms are
always on. As long as the alarm condition occurs, corresponding alarm will be triggered whether arrhythmia
analysis is switched on or off.

6.7.4 Changing Arrhythmia Threshold Settings


To change arrhythmia threshold settings, select the ECG parameter window→[Arrhythmia >>]→[Arrh. Threshold >>].

In case an arrhythmia violates its threshold, an alarm will be triggered. The setting of Asystole Delay is linked to ARR
relearning. When HR is less than 30 bpm, it is recommended to set Asystole Delay to 10 seconds.

Arrh. event Range Default Step Unit


PVCs High 1 to 10 10 1 /
Asystole. Delay 2 to 10 5 1 s
Tachy 60 to 300 Adult: 120 5 bpm
Pediatric: 160
Brady 15 to 120 Adult: 50 5 bpm
Pediatric: 75
Extreme Tachy 60 to 300 Adult: 160 5 bpm
Pediatric: 180
Extreme Brady 15 to 120 Adult: 35 5 bpm
Pediatric: 50
Multif. PVCs Window 3 to 31 15 1 Beats
V-Tach Rate 100 to 200 130 5 bpm
V-Tach PVCs 3 to 12 6 1 Beats
Pause Time 1.5, 2.0, 2.5 2.0 / s
Vbrd Rate 15 to 60 40 5 bpm
Vbrd PVCs 3 to 99 5 1 Beats

You can also set arrhythmia threshold from the ECG Setup menu in the configuration mode.

6-10
6.7.5 Initiating Arrhythmia Relearning Manually
Normally arrhythmia relearning allows the equipment to learn new ECG patterns to correct arrhythmia alarms and heart
rate value. We suggest you to manually initiate arrhythmia relearning when you suspect the result of arrhythmia
analysis.

To initiate relearning manually, select the ECG parameter window to enter the [ECG Setup] menu, select [Arrhythmia
>>]→[Relearn Arrh.]. When the equipment is learning, the message “Learning ECG” is displayed in the technical alarm
area.

NOTE
z Arrhythmia relearning in the case of ventricular tachycardia may affect correct arrhythmia alarm.

6.7.6 Automatic Arrhythmia Relearn


Arrhythmia relearning is initiated automatically whenever:

„ The ECG lead or lead label is changed

„ The ECG lead is re-connected

„ Patient category is changed

„ The paced status is changed,

„ Arrhythmia analysis is switched on

„ [Stop Calibrating] is selected after ECG calibration is completed.

6.8 Calibrating ECG


The ECG signal may be inaccurate due to hardware or software problems. As a result, the ECG wave amplitude becomes
greater or smaller. In that case, you can perform ECG calibration to check if the ECG wave amplitude is in normal range.

1. Select ECG filter hot key. Turn the navigation knob to select [Diagnostic] mode.

2. Select the ECG parameter area to enter the [ECG Setup] menu.

3. Select [Others >>]→[Calibrate]. In this case, a square wave appears on the screen and the message “Calibrating
ECG” is displayed.

4. Compare the amplitude of the square wave with the 1 mV wave scale. The difference should be within 5%.

5. After the calibration is completed, select [Stop Calibrating].

You can print out the waveform and wave scale and then measure the difference between them if necessary. If the
difference exceeds 5%, contact your service personnel.

6-11
FOR YOUR NOTES

6-12
7 AED
7.1 Overview
This chapter describes how to operate the equipment in AED Mode. While operating in AED Mode, the equipment
analyses the patient’s ECG waveforms and guides you through the defibrillation process.

The equipment starts analyzing the patient’s heart rhythm immediately after entering AED mode. When a shockable
rhythm is detected, the equipment gives a prompt and automatically starts charging. If a shockable rhythm is not
detected, a “No shock advised” prompt is given. Smart defibrillation analysis goes through automated external
defibrillation until the equipment enters CPR or abnormal pads connection occurs.

While operating in AED Mode, the capabilities of the device are limited to those essential to the performance of
semi-automated external defibrillation. Only ECG signals acquired through pads are displayed. Previously set alarms and
scheduled measurements are indefinitely paused and entry of patient information is disabled. Additionally, the Lead
Select, Alarm Pause, and Main Menu buttons are inactive.

7.2 Safety

DANGER
z Defibrillation current can cause operator or bystander severe injury or even death. Never touch the patient
or any equipment connected to the patient (including the bed or gurney) during defibrillation.

z Avoid contact between parts of the patient’s body such as exposed skin of head or limbs, conductive fluids
such as gel, blood, or saline, and metal objects such as a bed frame or a stretcher which may provide
unwanted pathways for the defibrillating current.

z Do not allow multifunction electrode pads to touch each other or to touch other ECG monitoring electrodes,
lead wires, dressings, etc. Contact with metal objects may cause electrical arcing and patient skin burns
during defibrillation and may divert current away from the heart.

z To avoid explosion hazard, do not use the equipment in the presence of oxygen-rich atmospheres,
flammable anesthetics, or other flammable agents (such as gasoline). Keep the equipment and the
operating environment dry and clean.

7-1
WARNING
z During defibrillation, air pockets between the skin and multifunction electrode pads can cause patient skin
burns. To help prevent air pockets, make sure defibrillation pads are completely adhered to the skin.

z Do not use dried-out pads.

z In AED mode, this equipment is not designed to administer energy at pediatric joule settings. The American
Heart Association recommends AEDs be used only on patients who are more than eight years old.

CAUTION

z Aggressive handling of multifunction electrode pads in storage or prior to use can damage the pads. Discard
the pads if they become damaged.

z For patients with implantable pacemaker, the sensitivity and specificity of AED algorithm may be impaired.

7.3 AED View


A typical screen in AED Mode is shown below.

In AED mode, HR numeric and one ECG waveform acquired from the multifunction electrode pads are displayed. Below
the ECG is the information area which displays the defibrillation mode, prompt message, contact impedance indicator,,
selected energy and a shock counter.

For details about the contact impedance indicator, refer to 8.7 Contact Impedance Indicator.

7-2
7.4 AED Procedure
Confirm that the patient is unresponsive, not breathing or not breathing normally.

Then:

1. Remove clothing from the patient’s chest. Wipe moisture from the patient’s chest and, if necessary, clip or shave
excessive chest hair.

2. Apply multifunction electrode pads to the patient as directed on the pads package. Use anterior-lateral placement.

3. Connect the pads with pads cable, and then plug the pads cable in the equipment’s therapy port.

4. Turn the Mode Select knob to AED.

When the equipment enters AED mode, it checks to see if the pads and pads cable are properly connected. If not, the
message “Connect Pads Cable” or “Apply Pads” will appear in the AED information area until corrective action has been
taken.

5. Follow the screen and voice prompts.

Once an ECG is detected through the multifunction electrode pads, the equipment automatically analyzes the patient’s
heart rhythm and warns you not to touch the patient. If a shockable rhythm is detected, the equipment charges
automatically.

You can switch on/off the voice prompt by accessing configuration management or adjust the volume of the voice
prompts by pressing the voice volume soft key.

6. Press the Shock button, if prompted.

Once charging is complete, the equipment gives prompt “Do Not Touch Patient! Press Shock Button”. Make sure no one is
touching the patient, bed or any equipment connected to the patient. Call out clearly and loudly “Stay Clear”. Then press
the Shock button on the front panel to deliver a shock to the patient.

Delivery of the shock is confirmed by the voice and screen prompt "Shock Delivered" and the shock counter on the
display is updated to reflect the number of shocks given. If the configured [Shock Series] is greater than one, the
equipment resumes analyzing the patient’s rhythm after the shock is delivered to see if the shock was successful. Voice
and text prompts continue to guide you through additional shocks.

NOTE
z Do not use anterior-posterior pads placement (multifunction electrode pads placed on the patient’s chest
and back). The AED algorithm used by the equipment has not been validated using this placement.

z Motion artifact may delay analysis or affect the ECG signal resulting in an inappropriate shock or no shock
advised message. Keep the patient still during ECG rhythm analysis.

z The Shock button must be pressed to deliver a shock. The equipment will not automatically deliver a shock.

z Impedance is the resistance between the defibrillator’s pads or paddles that the defibrillator must

7-3
overcome to deliver an effective discharge of energy. The degree of impedance differs from patient to
patient and is affected by several factors including the presence of chest hair, moisture, and lotions or
powders on the skin. If the “Impedance too high. Shock not delivered” message appears, make sure that the
patient’s skin has been washed and dried and that any chest hair has been clipped. If the message persists,
change the pads and/or the pads cable.

7.5 Shock Advised


If a shockable rhythm is detected, the equipment automatically charges to the pre-configured energy level. A charging
tone is sounded, and the Shock button flashes when the equipment is fully charged.

Heart rhythm analysis continues while the equipment charges. If a rhythm change is detected before the shock is
delivered and a shock is no longer appropriate, the stored energy is removed internally.

Once you are prompted "Do Not Touch Patient! Press Shock Button", if you do not do so within the configured Auto
Disarm time interval, the equipment disarms itself and resumes analyzing.

When the equipment is being charged or have been fully charged, you can remove the charged energy at any time by
pressing the [Pause for CPR] soft key.

For defibrillation of adult patients, recommended energy level is 200 Joules.

7.6 No Shock Advised (NSA)


If a shockable rhythm is not detected, the equipment will tell you "No Shock Advised!".

„ If the [NSA Action] is set to [CPR]

The equipment enters CPR status and you will see and hear “No Shock Advised! Paused. If Needed, Begin CPR.” The
remaining pause time is displayed as shown below. The pause period can be defined by setting [CPR Time]
through configuration management.

Analysis resumes at the completion of the pause period or when you press the [Resume Analyzing] soft key in CPR
status.

„ If the [NSA Action] is set to [Monitor]

The equipment continues to monitor the ECG and automatically resumes analysis if a potentially shockable rhythm
is detected. You will hear “No Shock Advised! If needed press pause for CPR”. The message "No Shock Advised!" and
“Monitoring” are shown circularly in the AED information area.

You can define the frequency of these prompts by adjusting [Voice Prompts Interval] through configuration
management. You may press the [Pause for CPR] soft key to suspend monitoring and administer CPR. The pause
period can be defined by setting [CPR Time] through configuration management.

7-4
7.7 CPR
If [Initial CPR Time] is not configured as Off, the system enters initial CPR if AED mode is entered. You can set [Initial
CPR Time] to an appropriate time or switch it off through configuration management.

After the shock series, ECG analysis pauses and the equipment enters the CPR status. Analysis resumes at the completion
of the pause period or when you press the [Resume Analyzing] soft key in CPR status.

In current shock series, the equipment enters the CPR status if you press the [Pause for CPR] soft key after a shock is
delivered. The CPR pause time is defined by setting [CPR Time] through configuration management.

NOTE
z You can start analyzing patient’s heart rhythm again at any time by pressing the [Resume Analyzing] soft
key in CPR status.

7.7.1 CPR Metronome


The equipment provides a CPR metronome feature that can be used to encourage rescuers to perform chest
compression and ventilation at AHA/ERC recommended rate.

When activated, the metronome sounds 120 times per minute and give voice prompts to indicate the rescuers to
perform ventilation at configured compression/ventilation rate.

To activate or deactivate CPR metronome,

1. Move the Mode Select knob to Monitor, Manual Defib or Pacer. Press the Main Menu button on the front panel, and
then select→ [Others >>]→[Configuration >>]→Enter the required password.

2. Select [AED Setup] →[CPR Metronome], and toggle between [On] and [Off].

The CPR metronome is activated by default. When CPR metronome is on, you can set the compression/ventilation rate
by selecting [CPR Mode]. The default rate is 30:2.

7-5
Warning
z The CPR metronome sounds do not indicate information regarding the patient’s condition. Because patient
status can change in a short time, the patient should be assessed at all times. Do not perform CPR on a
patient who is responsive or is breathing normally.

NOTE
z CPR metronome and its volume is affected by the settings of [Voice Prompt] and [Voice Volume] in the AED
Setup menu.

7.8 AED Sound Recording


The equipment include a sound recording fucntion that can record the voice information during AED therapy. The
sound recording function can be configured on or off.

To switch on or off the sounding recording,

1. Move the Mode Select knob to Monitor, Manual Defib or Pacer. Press the Main Menu button on the front panel, and
then select [Others >>]→[Configuration >>]→Enter the required password.

2. Select [AED Setup] →[Voice Recording], and toggle between [On] and [Off].

The symbol is shown at the top right corner of the AED information area if the sounding recording function is
enabled.

The equipment can store up to 180 minutes, maximum 60 minutes for one patient, of sound recording.

7.9 AED Setup


1. Move the Mode Select knob to Monitor, Manual Defib or Pacer. Press the Main Menu button on the front panel, and
then select [Others >>]→[Configuration >>]→Enter the required password.

2. Select [AED Setup >>] to enter the AED Setup menu, and then change AED settings as desired.

Refer to Section 19.3.3 AED Setup Menu for details.

7-6
8 Manual Defibrillation
8.1 Overview
This chapter explains how to prepare for and perform asynchronous defibrillation and synchronous cardioversion using
multifunction electrode pads and external paddles.

In Manual Defib Mode, you must assess the ECG waveforms, decide if defibrillation or cardioversion is indicated, select
appropriate energy setting, charge the equipment, and deliver the shock. Text messages and a contact impedance
indicator on the screen provide relevant information to guide your throughout the defibrillation process.

While operating manual defibrillation, besides ECG you can select to monitor SpO2..

Alarms are turned off automatically when you enter Manual Defib mode. Pressing the Alarm Pause button can turn on
the alarms.

8.2 Safety

DANGER
z Defibrillation current can cause operator or bystander severe injury or even death. Never touch the patient
or any equipment connected to the patient (including the bed or gurney) during defibrillation.

z Avoid contact between parts of the patient’s body such as exposed skin of head or limbs, conductive fluids
such as gel, blood, or saline, and metal objects such as a bed frame or a stretcher which may provide
unwanted pathways for the defibrillating current.

z Do not allow multifunction electrode pads and paddles to touch each other or to touch other ECG
monitoring electrodes, lead wires, dressings, etc. Contact with metal objects may cause electrical arcing and
patient skin burns during defibrillation and may divert current away from the heart.

z During manual defibrillation, make sure your hands are dry and free from conductive gel to avoid shock
hazard.

z Use care when operating this equipment close to oxygen sources (such as bag-valve-mask devices or
ventilator tubing). Turn off gas source or move source away from patient during defibrillation. This can
cause an explosion hazard.

8-1
WARNING
z During synchronous cardioversion, if monitoring patient’s ECG through external paddles, artifact
introduced by paddle movement may resemble an R-wave and trigger a defibrillation shock.

z Do not use conductive liquid. Use only conductive gel specified by the equipment manufacturer.

z If external paddles are used for defibrillation, apply the paddles tightly and evenly to the patient’s chest to
ensure good skin contact.

z Never apply the paddles to human body to verify paddle connection.

z Clinicians must select an appropriate energy level for defibrillation of pediatric patients.

CAUTION
z Use of Manual Defib mode may be password protected. Make sure the operator knows and remembers the
password as defined in Configuration. Failure to enter correct password will prevent the delivery of manual
defibrillation therapy.

z Clear the conductive gel from the external paddles at the completion of the therapy to prevent the paddles
from being corroded.

z Prior to using this defibrillator, disconnect from the patient all equipment that is not defibrillator-protected.

z Do not charge the equipment and deliver shock frequently in non-clinical situations. Otherwise equipment
damage could occur.

NOTE
z Impedance is the resistance between the defibrillator’s pads or paddles that the defibrillator must
overcome to deliver an effective discharge of energy. The degree of impedance differs from patient to
patient and is affected by several factors including the presence of chest hair, moisture, and lotions or
powders on the skin. If the “Impedance too high. Shock not delivered” message appears, make sure that the
patient’s skin has been washed and dried and that any chest hair has been clipped. If the message persists,
change the pads and/or the pads cable.

z Alarms are switched off automatically and the “Alarm Off” message is displayed when the equipment enters
the asynchronous defibrillation mode. Alarms remain off until toggled on by pressing the Alarm Pause
button, the Sync mode is entered, or the Mode Select knob is moved to Monitor or Pacer.

8-2
8.3 Manual Defibrillation View
A typical screen in Manual Defib Mode is shown below.

In the enlarged ECG area, an ECG waveform and related parameters are displayed. In the middle of the screen,
defibrillation mode, synchronous icon, prompt message, selected energy, contact impedance indicator, and a shock
counter are displayed.

8.4 Manual Defibrillation Procedure


1. Remove clothing from the patient’s chest. Wipe moisture from the patient’s chest and, if necessary, clip or shave
excessive chest hair.

2. Connect the therapy cable to the therapy port. Push until you hear it click into place.

3. Apply multifunction electrode pads or external paddles to the patient.

‹ If multifunction electrode pads are used, apply pads according to the instructions for use indicated on pads
package. Use anterior-lateral or anterior-posterior placement.

‹ If external paddles are used, remove the paddle set from the paddle tray by grasping the handles and pulling
them straight up. Apply conductive gel to the electrode surface of each paddle. Place the paddles to the
patient’s chest using the anterior-lateral placement.

Do not touch this surface and the


part below.
8-3
WARNING
z Hold only the insulating parts of the paddle handles to avoid shock hazard during charging or shock
delivery.

4. Turn the Mode Select knob to Manual Defib.

You can access manual therapy directly, by confirmation or by password, which can be defined through
configuration management. The default setting is [Direct].

‹ If [Manual Therapy Access] is set to [Direct], the equipment directly enters Manual Defib mode when the
Mode Select knob is switched to Manual Defib.

‹ If [Manual Therapy Access] is set to [Confirmed], a dialog box pops up when the Mode Select knob is
switched to Manual Defib. You have to confirm whether to enter Manual Defib mode or not.

‹ If [Manual Therapy Access] is set to [Password], a dialog box pops up, requiring the user to enter a password,
when the Mode Select knob is switched to Manual Defib. The equipment enters Manual Defib mode only after
correct password is entered.

5. Select energy.

You can select desired energy level by the adjusting the Energy Select buttons on the equipment’s front panel or
the Energy Select buttons on external paddles if external paddles are used.

Your current energy selection is shown in the defibrillation information area as shown below.

6. Charge

Press the Charge button on the front panel. If external paddles are used, the Charge button on the paddles may be
used instead. As the equipment charges, a progress bar is shown in the defibrillation information area. A charging
tone sounds until desired energy level is reached, when you will hear a charge done tone.

If you have to increase or decrease the selected energy during charging or after charging is complete, adjust the Energy
Select button to select the desired energy level as explained above. Then press the charge button again to restart
charging.

To remove the energy, press the [Disarm] soft key. If the Shock button is not pressed within the specified time period,
the equipment disarms automatically. You can define [Time to Auto Disarm] through configuration management.
7. Shock

Confirm that a shock is still indicated and that the equipment has charged to the selected energy level. Make sure
no one is touching the patient, bed or any equipment connected to the patient. Call out loudly and clearly, “Stay
Clear!”

‹ If pads are used, press the flashing Shock button on the front panel to deliver a shock to the patient.

‹ If external paddles are used, simultaneously press the Shock buttons located on the paddles to deliver a shock
to the patient.

8-4
NOTE
z Defibrillation is always performed through paddles or pads. However, during defibrillation you may choose
to monitor the ECG using an alternate ECG source (3- or 5-lead monitoring electrodes). If an alternate ECG
source is connected, any available lead may be displayed.

z When external paddles are used, the Shock button on the equipment’s front panel is disabled.

z For defibrillation of adult patients, recommended energy level is 200 Joules.

8.4.1 Using Pediatric Paddles


The external paddle set comes with pediatric paddles included. To use the pediatric paddle set, depress the latch at the
front of the external paddle set while pulling forward on the adult paddle electrodes.

To defibrillate, refer to the procedure as described in Section 8.4 Manual Defibrillation Procedure.

8.4.2 Using Internal Paddles


To defibrillate using internal paddles:

1. Turn on the defibrillator and enter the manual defib mode.

2. Select the appropriate paddle size.

3. Connect the paddles to the defibrillator by aligning the white pointer on the paddles cable with the arrow on the
therapy port. Push until you hear it click into place.

4. Select energy by pressing the Energy Select key on the equipment's front panel.

5. Place the conductive surface of paddle electrodes against the patient's right atrium and left ventricle, as shown in
the figure below:

6. Charge the defibrillator by pressing the Charge key on the front panel.

7. Make sure no one is touching the patient or anything connected to the patient.

8. Press the Shock key on the front panel.

Using internal paddles for synchronized cardioversion requires that the patient’s ECG be acquired through a standard
ECG cable. The patient’s ECG acquired through the internal paddles may be unreliable for synchronized cardioversion
due to excessive noise or artifact causing inappropriate R-wave detection.

8-5
NOTE
z When internal paddles are used for defibrillation, the energy selection is automatically limited to 50 joules
because of possible cardiac damage from higher energies.

z Sterilize the internal paddles before each use. Otherwise, severe infection may result.

z Clean the internal paddles after each use.

8.5 Synchronized Cardioversion


Synchronized Cardioversion allows you to synchronize delivery of the defibrillator shock with the R-wave of the ECG. You
may choose to perform synchronized cardioversion through either:

„ Multifunction electrode pads, or

„ External paddles

To use synchronized cardioversion, press the [Enter Sync] soft key in the asynchronous defibrillation mode. Then “Sync”
appears in the manual Defibrillation information area and a marker appears above each R-wave, see the figure below:

Sync marker

R-wave marker

You can monitor ECG through multifunction electric pads, or external paddles, or electrodes connected to a 3- or 5-lead
ECG cable. Shock is delivered through either pads or paddles.For synchronized cardioversion, we recommend to acquire
patient’s ECG through ECG lead set

NOTE
z When you enter synchronous cardioversion, monitoring alarms is reactivated autonomously.

8-6
8.5.1 Performing Synchronized Cardioversion
1. Connect the therapy cable and apply the multifunction electrode pads or external paddles to the patient. If ECG set
is used for ECG monitoring, connect the ECG trunk cable and apply the ECG electrodes to the patient, referring to 6
Monitoring ECG.

2. With the Mode Select knob in the Manual Defib position, press the [Enter Sync] soft key to activate the
synchronous cardioversion function.

3. Select a lead. The selected lead should have a clear signal and a large QRS complex.

4. Verify that a white R-wave marker appears above each R-wave. If the R-wave markers do not appear or do not
coincide with the R-waves, for example above the T-waves, select another lead.

5. Verify that the equipment enters the Sync mode, as indicated by the SYNC mark shown in the defibrillation
information area.

6. Press the Energy Select button to select a desired energy level.

7. Press the Charge button on the equipment’s front panel or, if using external paddles, the Charge button located on
the handle of Apex paddle.

8. Confirm that a shock is still indicated and that the equipment has charged to the selected energy level. Make sure
no one is touching the patient, bed or any equipment connected to the patient. Call out loudly and clearly, “Stay
Clear!”.

9. Press and hold the Shock button on the equipment or, if using external paddles, the Shock buttons on both
paddles. The shock will be delivered when the next R-wave is detected.

NOTE
z During synchronized cardioversion, it is important to continue to hold the shock button (or the paddle’s
Shock buttons) until the shock is delivered.  The equipment shocks with the next detected R-wave.

8.5.2 Delivering Additional Synchronized Shocks


If additional synchronized shocks are indicated, perform the following steps:

1. Make sure the equipment is still in Sync mode, as indicated by the presence of the Sync message in the
defibrillation information area.

2. Repeat Steps 4 to 9 as described above.

If [Sync after Shock] is set to [Yes], the equipment remains in the sync mode after a shock is delivered; if set to [No], the
equipment exits the sync mode and enters the asynchronous defibrillation mode after a shock.

8.5.3 Disabling the Sync Function


To switch off the Sync function, press the [Sync Off] soft key to enter the Manual Defib mode. .

8-7
8.6 Remote Synchronized Cardioversion
The equipment can be configured to receive an ECG source from a remote patient monitor (such as a bedside patient
monitor) to perform synchronized cardioversion. To do so, the remote patient monitor shall have a sync out connector
and shall be connected to the defibrillator/monitor’s multifunctional connector with a synchronous cable.

To switch on remote synchronization, select [Manual Defib Setup]→[Remote Sync]→[On] through the Configuration
Main Menu.

The remote synchronization procedure is as follows:

1. Use a sync cable to connect the defibrillator/monitor with a bedside monitor through the defibrillator/monitor’s
multifunctional connector.

2. Switch on the equipment and enter the Manual Defib mode.

3. Press the [Enter Sync] soft key. In this case the “Select Sync Mode” dialog pops up.

4. Select [Remote] to enter remote synchronization. Then the message “Remote Sync” is presented.

5. Confirm a square wave on the defibrillator/monitor blinks with each detected R wave on the remote monitor as
shown below, indicating a sync pulse is received.

6. Connect the therapy cable to the therapy port. Push until you hear it click into place.

7. Apply multifunction electrode pads or external paddles to the patient.

8. Follow Steps 6 through 9 from 8.5.1 Performing Synchronized Cardioversion as described previously.

8-8
NOTE
z During remote synchronous defibrillation, the local defibrillator/monitor does not display the ECG
waveform. To view the patient’s ECG, check the remote monitor.

z When you use an remote monitor as the ECG source, a biomedical technician must verify that the remote
monitor and the defibrillator/monitor combination will deliver a synchronized shock within 60 ms of the
peak of the R-wave.

8.7 Contact Impedance Indicator


Contact impedance indicator is used to indicate the impedance between the defibrillator’s pads or paddles in Manual
Defib or AED mode, as shown below:
Contact Impedance
Indicator

Different levels of the measured impedance are shown as follows:

„ indicates the impedance is in normal range.

„ indicates the impedance is slightly higher for the defibrillation.

„ indicates the impedance is too high for the defibrillation or the pads or paddles are not

properly connected to the patient.

„ indicates the therapy cable is not properly connected.

To switch on the contact impedance indicator:

1. In Monitor, Manual Defib or Pacer mode, press the Main Menu button on the front panel, and then select→
[Others >>]→[Configuration >>]→Enter the required password.

2. Select [Manual Defibrillation] and set [Contact Impedance Indicator] to [On].

WARNING

z It is recommended to perform defibrillation on a patient when the contact impedance is in normal range.
However, you can also perform defibrillation when the contact impedance is slightly higher.

8-9
FOR YOUR NOTES

8-10
9 Noninvasive Pacing
9.1 Overview
In pacer mode, the patient’s ECG is monitored through ECG lead set and pace pulses are delivered through multifunction
electrode pads. The pads cannot be used to monitoring ECG rhythm and deliver pacing current at the same time.

A white pacing marker is shown on the ECG waveform each time a pacer pulse is delivered to the patient. If pacing in
demand mode, white R-wave marker also appears on the ECG waveform until capture occurs.

During pacing, both ECG and SpO2 parameters continue to be monitored and parameter alarms remain active.

In demand mode pacing, a 3-lead or 5-lead ECG cable and electrodes are required to acquire ECG signal. Pace pulses are
delivered through multifunction electrode pads. However, the pads cannot be used to monitor the ECG and deliver pace
pulses simultaneously.

NOTE

z In Pacer mode, arrhythmia analysis is supported and available arrhythmia alarms are asystole, ventricular
vibrillation and ventricular tachycardia.

9.2 Safety

WARNING
z Heart rate displays and alarms function during pacing, but they can be unreliable. Observe the patient
closely while pacing. Do not rely on the indicated heart rate or heart rate alarms as a measure of the
patient’s perfusion status.

z To avoid explosion hazard when pacing a patient who is receiving oxygen delivery, properly route the
oxygen delivery tube. Do not keep it close to the multifunction electrode pads.

9-1
CAUTION
z Use of Pacer mode may be password protected. Make sure the operator knows and remembers the
password as defined in Configuration. Failure to enter correct password will prevent the delivery of pacing
therapy.

z For treatment of patients with implanted devices such as permanent pacemakers or


cardioverter-defibrillators, consult a physician and the instructions for use provided by the device’s
manufacturer

z Prolonged noninvasive pacing may cause patient skin irritation and burns. Periodically inspect the
underlying skin and change ECG electrodes and multifunction electrode pads.

NOTE

z If pacing is interrupted for any reason, the [Start Pacing] soft key must be pressed to resume pacing.

z In pacer mode, you cannot change the patient’s internal paced status from the ECG Setup menu.

z In the case that pads poorly contact the patient, the alarm “Pacer Stopped Abnormally” and “Pads Off” may
be presented.

z Pads are not an available choice for the source of ECG waveform in Pacer mode.

9.3 Pacing View


A typical screen in pacer Mode is shown below.

In pacer mode, an ECG waveform, related parameters and waveforms are displayed. The pacer information area shows
the pacer mode, pacer rate and output, as well as pacer-related prompts and alarms. Soft keys available for setting
pacing functions are also displayed.

9-2
9.4 Demand Mode versus Fixed Mode
The equipment can deliver paced pulses in either demand or fixed mode.

„ In demand mode, the pacer only delivers paced pulses when the patient’s heart rate is lower than the selected
pacing rate.

„ In fixed mode, the pacer delivers paced pulses at the selected rate.

During pacing, you can change pacer mode. Then the equipment continues to deliver pace pulses at selected pacer rate
and pacer output.

CAUTION
z Use demand mode pacing whenever possible. Use fixed mode pacing if noise or artifact interferes with
proper sensing of R-wave or when monitoring electrodes are not available.

z During fixed mode pacing, R-wave markers do not appear on the paced beats.

z During demand mode pacing, spontaneous beats may be presented which are not associated with the
delivery of pace pulse. If the patient’s heart rate is above the pacer rate, pace pulses are not delivered and,
therefore, pacing markers do not appear.

9.5 Preparing for Pacing


1. If not pre-connected, connect the pads cable to the equipment. Then push until you hear it click into place.

2. Make sure the multifunction electrode pads packaging is intact and within the expiration date shown.

3. Connect the pads connector to the pads cable.

4. Apply the pads to the patient using anterior-lateral or anterior-posterior placement.

5. If pacing in demand mode, apply monitoring electrodes, and connect the ECG cable to the equipment. To get the
best monitoring signal, make sure there is adequate space between ECG electrodes and therapy electrodes. Refer
to 6.4.1 ECG Monitoring with Electrodes.

9-3
9.5.1 Demand Mode Pacing
To pace in demand mode:

1. Turn the Mode Select knob to the Pacer position. Thus the pacing function is enabled in demand mode
automatically. ECG waveform of Lead II is displayed in the waveform area by default.

You can access manual therapy directly, by confirmation or by password, which can be defined through
configuration management. The default setting is [Direct].

‹ If [Manual Therapy Access] is set to [Direct], the equipment directly enters Pacer mode when the Mode Select
knob is switched to Pacer.

‹ If [Manual Therapy Access] is set to [Confirmed], a dialog box pops up when the Mode Select knob is
switched to Pacer. You have to confirm whether to enter Pacer mode or not.

‹ If [Manual Therapy Access] is set to [Password], a dialog box pops up, requiring the user to enter a password,
when the Mode Select knob is switched to Pacer. The equipment enters Pacer mode only after correct
password is entered.

2. Select a lead with an easily detectable R-wave.

3. Verify that white R-wave markers appear above the R-waves, as shown in the figure below. If the R-wave markers do
not appear or do not coincide with the R-waves, for example above the T-waves, select another lead.

4. Select pacer rate. If necessary, select initial pacer output. To select pacer rate or pacer output, rotate the Navigation
knob to select an appropriate value and press it to confirm the selection. Do not forget to press the Navigation
knob to exit the setting after desired value is selected.

5. Press the [Start Pacing] soft key to start pacing. The message “Pacing” appears in the pacer information area. .

NOTE

z Pacing will not start if there is a problem with the pads cable connection, pad patient connection, or ECG
monitoring electrodes connection. If either situation occurs, a message will appear in the pacer information
area to alert you that a lead is disconnected or that the pads have a poor connection.

6. Verify that white pacing markers appear on the ECG waveform, as shown below:

Pacing markers

7. Adjust pacer output: increase pacer output until cardiac capture occurs (capture is indicated by the appearance of a
QRS complex after each pacing marker), and then decrease the output to the lowest level that still maintains
capture.
9-4
8. Verify the presence of a peripheral pulse.

You can temporarily withhold pacing pulse and observe the patient’s underlying rhythm by pressing and holding the
[4:1] soft key. This causes pacing pulse to be delivered at 1/4 of the defined pacer rate. To resume pacing at set rate,
release this key.

To stop pacing, press the [Stop Pacing] soft key. Pressing the [Start Pacing] soft key can resume pacing.

CAUTION
z Routinely assess the patient’s cardiac output.

9.5.2 Fixed Mode Pacing


To pace in fixed mode:

1. Turn the Mode Select knob to the Pacer position.

2. Switch the pacer to the Fixed mode. To do so, move the cursor to the Pacer Mode hot key and rotate the Navigation
knob to select [Fix Mode]; then press it to confirm the selection, see the figure below:
Pacer Mode
hot key

3. If ECG electrodes are applied, use the Lead Select button to select the desired lead for viewing.

4. Select pacer rate. If necessary, select pacer output. To select pacer rate or pacer output, rotate the Navigation knob
to select an appropriate value and press it to confirm the selection.

5. Start pacing.

Press the [Start Pacing] soft key to start pacing. The message “Pacing” appears in the pacer information area. .

6. Verify that white pacing markers appear on the ECG waveform.

7. Adjust pacer output: increase pacer output until cardiac capture occurs (capture is indicated by the appearance of a
QRS complex after each pacing marker), and then decrease the output to the lowest level that still maintains
capture.

8. Verify the presence of a peripheral pulse.

You can temporarily withhold pacing pulse and observe the patient’s underlying rhythm by pressing and holding the
[4:1] soft key. This causes pacing pulse to be delivered at 1/4 of the defined pacer rate. To resume pacing at set rate,
release this key.

To stop pacing, press the [Stop Pacing] soft key.

9-5
WARNING
z Use care when handling the multifunction electrode pads on the patient to avoid shock hazard during
pacing.

z If you are using the pacing function with battery power and the Low Battery alarm is presented, connect the
equipment to external power or install a fully charged battery.

CAUTION
z The monitoring or pacing function may be unstable in the presence of ESU or other electronic devices.

9-6
10 Monitoring Resp
10.1 Overview
Impedance respiration is measured across the thorax. When the patient is breathing or ventilated, the volume of air
changes in the lungs, resulting in impedance changes between the electrodes. Respiration rate (RR) is calculated from
these impedance changes, and a respiration waveform appears on the equipment screen.

10.2 Safety

WARNING
z When monitoring the patient’s respiration, do not use ESU-proof ECG cables.

z The respiration measurement does not recognize obstructive and mixed apneas: it only indicates an alarm
when a pre-adjusted time had elapsed since the last detected breath. The safety and effectiveness of the
respiration measurement method in the detection of apnea, especially the apnea of prematurity and apnea
of infancy, has not been established.

10.3 Resp View

RR unit
RR alarm
high limit

RR alarm RR reading

low limit

10-1
10.4 Placing Resp Electrodes
As the skin is a poor conductor of electricity, preparing the skin is necessary for a good Respiration signal. You can refer
to the ECG section for how to prepare the skin.

As the Respiration measurement adopts the standard ECG electrode placement, you can use different ECG cables (3-lead
or 5-lead). Since the respiration signal is measured between two ECG electrodes, if a standard ECG electrode placement
is applied, the two electrodes should be RA and LA of ECG Lead I, or RA and LL of ECG Lead II.

NOTE
z To optimize the respiration waveform, place the RA and LA electrodes horizontally when monitoring
respiration with ECG Lead I; place the RA and LL electrodes diagonally when monitoring respiration with
ECG Lead II.

Lead I Lead II

10-2
10.4.1 Optimizing Lead Placement for Resp
If you want to measure Resp and you are already measuring ECG, you may need to optimize the placement of the two
electrodes between which Resp will be measured. Repositioning ECG electrodes from standard positions results in
changes in the ECG waveform and may influence ST and arrhythmia interpretation.

„ Cardiac activity that affects the Resp waveform is called cardiac overlay. It happens when the Resp electrodes pick
up impedance changes caused by the rhythmic blood flow. Correct electrodes placement can help to reduce
cardiac overlay. Avoid the liver area and the ventricles of the heart in the line between the respiratory electrodes.
This is particularly important for neonates.

„ Some patients with restricted movement breathe mainly abdominally. In these cases, you may need to place the
left leg electrode on the left abdomen at the point of maximum abdominal expansion to optimize the respiratory
wave.

„ In clinical applications, some patients (especially neonates) expand their chests laterally, causing a negative
intrathoracic pressure. In these cases, it is better to place the two respiration electrodes in the right midaxillary and
the left lateral chest areas at the patient’s maximum point of the breathing movement to optimize the respiratory
waveform.

NOTE
z Respiration monitoring is not for use on the patients who are very active, as this will cause false alarms.

10.4.2 Changing Resp Wave Settings


In the [Resp Setup] menu, you can:

„ Select [Gain] and then choose an appropriate setting. The bigger the gain is, the larger the wave amplitude is.

„ Select [Sweep] and then choose an appropriate setting. The faster the wave sweeps, the wider the wave is.

„ Select [Lead] to set lead for RESP monitoring.

In the [Resp Setup] menu, you can also change the [Apnea Alarm Delay] as desired.

10-3
FOR YOUR NOTES

10-4
11 Monitoring PR
11.1 Overview
The pulse numeric counts the arterial pulsations that result from the mechanical activity of the heart. You can display a
pulse from SpO2. The displayed pulse numeric is color-coded to match SpO2 parameter.

PR unit

PR alarm high limit PR value

PR alarm low limit

11.2 Adjusting Pulse Tone Volume


When PR alarm is switched on, the equipment gives out pulse tone. You can select the PR parameter area to enter the
[PR Setup] menu, and then adjust [QRS Volume] to change the volume of pulse tone .The default QRS volume is 2.
When a valid SpO2 value exists, the system will adjust the pitch of pulse tone according to the SpO2 value.

11-1
FOR YOUR NOTES

11-2
12 Monitoring SpO 2

12.1 Introduction
SpO2 monitoring is a non-invasive technique used to measure the amount of oxygenated haemoglobin and pulse rate
by measuring the absorption of selected wavelengths of light. The light generated in the probe passes through the
tissue and is converted into electrical signals by the photodetector in the probe. The SpO2 module processes the
electrical signal and displays a waveform and digital values for SpO2 and pulse rate.

This device is calibrated to display functional oxygen saturation. It provides the following measurements.

5 6

1. Pleth wave: visual indication of patient’s pulse. The waveform is not normalized.

2. SpO2 unit

3. SpO2 alarm high limit

4. SpO2 alarm low limit

5. Oxygen saturation of arterial blood (SpO2): percentage of oxygenated hemoglobin in relation to the sum of
oxyhemoglobin and deoxyhemoglobin.

6. Perfusion indicator: the pulsatile portion of the measured signal caused by arterial pulsation.

12-1
12.2 Safety

WARNING
z Use only SpO2 sensors specified in this manual. Follow the SpO2 sensor’s instructions for use and adhere to
all warnings and cautions.

z When a trend toward patient deoxygenation is indicated, blood samples should be analyzed by a laboratory
co-oximeter to completely understand the patient’s condition.

z Do not use SpO2 sensors during magnetic resonance imaging (MRI). Induced current could potentially cause
burns.

z Prolonged continuous monitoring may increase the risk of undesirable changes in skin characteristics, such
as irritation, reddening, blistering or burns. Inspect the sensor site every two hours and move the sensor if
the skin quality changes. Change the application site every four hours. For neonates or patients with poor
peripheral blood circulation or sensitive skin, inspect the sensor site more frequently.

12.3 Identifying SpO2 Modules


The equipment can be configured with any of the following SpO2 modules.

„ Mindray SpO2 module;

„ Masimo SpO2 module;

„ Nellcor SpO2 module.

For Masimo or Nellcor SpO2 modules, you will see corresponding logo on the equipment.

12.4 SpO2 Monitoring Procedure


1. Select an appropriate sensor according to the module type, patient category and weight.

2. Clean the application site, e.g. removing colored nail polish from the application site.

3. Apply the sensor to the patient.

4. Select an appropriate adapter cable according to the connector type and connect it with the equipment.

5. Connect the sensor cable to the adapter cable.

6. Switch the Mode Select knob to Monitor.

NOTE
z The SpO2 simulator can be used to check if the SpO2 sensor works properly. However, it cannot be used to
verify the accuracy of the SpO2 sensor.

12-2
12.5 Changing SpO2 Settings
You can access the [SpO2 Setup] menu by selecting the SpO2 parameter area

12.5.1 Setting SpO2 Sensitivity


For Mindray SpO2 module, you can set [Sensitivity] to [High], [Med] or [Low] from the [SpO2 Setup] menu. For Masimo
SpO2 module, you can set [Sensitivity] to [Normal] or [High], in which [Normal] is equivalent to [Med].

„ When the [Sensitivity] is set to [High], the equipment is more sensitive to the changes to SpO2 values, but the
measurement accuracy is relatively low. When monitoring critically ill patients whose pulsations are very weak, it is
strongly recommended that the sensitivity is set to [High].

„ When the [Sensitivity] is set to [Low], the equipment responds slowly to the changes of SpO2 value, but the
measurement accuracy is relatively high.

12.5.2 Monitoring SpO2 and NIBP on the Same Limb


When monitoring SpO2 and NIBP on the same limb simultaneously, you can switch [NIBP Simul] on from the [SpO2
Setup] menu to lock the SpO2 alarm status until the NIBP measurement ends. If you switch [NIBP Simul] off, low
perfusion caused by NIBP measurement may lead to inaccurate SpO2 readings and therefore cause false physiological
alarms.

12.5.3 Changing Averaging Time


The SpO2 value displayed on the screen is the average of data collected within a specific time. The shorter the averaging
time is, the quicker the equipment responds to changes in the patient’s oxygen saturation level, but the measurement
accuracy is relatively low. Contrarily, the longer the averaging time is, the slower the patient monitor responds to
changes in the patient’s oxygen saturation level, but the measurement accuracy will be improved. When monitoring
critically ill patients, selecting shorter averaging time will help understanding the patient’s condition.

To set the averaging time:

„ For Mindray SpO2 module, set [Sensitivity] to [High], [Med] or [Low] from the [SpO2 Setup] menu, which
respectively correspond to 7 s, 9 s and 11 s.

„ For Masimo SpO2 module, set [Averaging] to [2-4 s], [4-6 s], [8 s], [10 s], [12 s], [14 s] or [16 s] in the [SpO2 Setup]
menu.

12-3
12.5.4 Sat-Seconds Alarm Management
With traditional alarm management, high and low alarm limits are set for monitoring oxygen saturation. During
monitoring, as soon as an alarm limit is violated, an audible alarm immediately sounds. When the patient’s SpO2 value
fluctuates near an alarm limit, the alarm sounds each time the limit is violated. Such frequent alarm can be distracting.

The Sat-Seconds feature is available with the Nellcor SpO2 module to decrease the likelihood of false alarms caused by
motion artifacts. The Sat-Seconds limit controls the amount of time that SpO2 saturation may be outside the set limits
before an alarm sounds. The method of calculation is as follows: the number of percentage points that the SpO2
saturation falls outside the alarm limit is multiplied by the number of seconds that it remains outside the limit. This can
be stated as the equation: Sat-Seconds= Points × Seconds

Only when the Sat-Seconds limit is reached, the monitor gives a Sat-Seconds alarm. The figure below demonstrates the
alarm response time with a Sat-Seconds limit set at 50 and a low SpO2 limit set at 90%. In this example, the patient’s
SpO2 drops to 88% (2 points) and remains there for 2 seconds. Then it drops to 86% (4 points) for 3 seconds, and then to
84% (6 points) for 6 seconds. The resulting Sat-Seconds are:
Points Seconds Sat-Seconds
2× 2= 4
4× 3= 12
6× 6= 36
Total Sat-Seconds = 52

After approximately 10.9 seconds, a Sat-Second alarm would sound, because the limit of 50 Sat-Seconds would have
been exceeded.

SpO2%

Second
Saturation levels may fluctuate rather than remaining steady for a period of several seconds. Normally, the patient’s SpO2
may fluctuate above and below an alarm limit, re-entering the non-alarm range several times. During such fluctuation,
the system sums the number of SpO2 points, both positive and negative, until either the Sat-Seconds limit is reached, or
the patient’s SpO2 re-enters the non-alarm range and remains there.

12-4
12.5.5 Changing the Speed of the Pleth Wave
In the [SpO2 Setup] menu, select [Sweep] and then select the appropriate setting. The faster the wave sweeps, the
wider the wave is.

12.6 SpO2 Desat Alarm


The defibrillator/monitor provides an SpO2 Desat alarm. The SpO2 Desat alarm provides an additional limit setting below
the SpO2 low limit setting to notify you of potentially life threatening decreases in oxygen saturation.
The SpO2 Desat alarm is a high-level exclusive alarm. You cannot change its alarm level.

You can Press the Main Menu button on the front panel, and then select [Alarm Setup >>]→ [Para. Alarm >>]→[Desat],
or select a parameter window and then select [Para. Alarm >>]→ [Desat] to set the Desat alarm.

NOTE
z In the case that the SpO2 low limit alarm value is set below the Desat limit, the the SpO2 low limit is
automatically adjusted to the Desat value.

12.7 Pitch Tone


The pitch tone function enables the equipment to give variable pitches of heartbeat tone or pulse tone as the patient’s
SpO2 level changes. This equipment provides 22 levels of pitch tone. The pitch of heartbeat tone or pulse tone rises as
SpO2 level increases and falls as SpO2 level decreases.

There are two pitch tone modes. The select pitch tone mode, select [SpO2 Setup]→[Pitch Tone] from the Configuration
Main menu and toggle between [Mode 1] and [Mode 2].

If the SpO2 is disabled, the pitch tone function will be disabled also.

12.8 Measurement Limitations


If you doubt the SpO2 readings, check the patient’s vital signs first. Then check the equipment and SpO2 sensor. The
following factors may influence the accuracy of measurement:

„ Ambient light

„ Patient movement (active and imposed motion)

„ Diagnostic testing

„ Low perfusion

„ Electromagnetic interference, such as MRI environment

„ Electrosurgical units

„ Dysfunctional haemoglobin, such as carboxyhemoglobin (COHb)and methemoglobin (MetHb)

„ Presence of certain dyes, such as methylene and indigo carmine

12-5
„ Inappropriate positioning of the SpO2 sensor, or use of incorrect SpO2

„ Drop of arterial blood flow to immeaurable level caused by shock, anemia, low temperature or vasoconstrictor.

12.9 Masimo Information

„ Masimo Patents

This device may be covered by one or more of the following US patents and foreign equivalents: 5,758,644, 6,011,986,
6,699,194, 7,215,986, 7,254,433, 7,530,955.

„ No Implied License

Possession or purchase of this device does not convey any express or implied license to use the device with
unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or
more of the patents relating to this device.

12.10 Nellcor Information

„ Nellcor Patents

This device may be covered by one or more of the following US patents and foreign equivalents: 5,485,847, 5,676,141,
5,743,263, 6,035,223, 6,226,539, 6,411,833, 6,463,310, 6,591,123, 6,708,049, 7,016,715, 7,039,538, 7,120,479, 7,120,480,
7,142,142, 7,162,288, 7,190,985, 7,194,293, 7,209,774, 7,212,847, 7,400,919.

„ No Implied License

Possession or purchase of this device does not convey any express or implied license to use the device with
unauthorized replacement parts which would, alone, or in combination with this device, fall within the scope of one or
more of the patents relating to this device.

12-6
13 NIBP
13.1 Introduction
Automatic non-invasive blood pressure monitoring uses the oscillometric method of measurement. It is intended for
adult, pediatric and neonatal patients. To understand how this method works, we’ll compare it to the auscultative
method.

With auscultation, the clinician listens to the blood pressure and determines the systolic and diastolic pressures. The
mean pressure can then be calculated with reference to these pressures as long as the arterial pressure curve is normal.

Since the equipment cannot hear the blood pressure, it measures cuff pressure oscillation amplitudes. Oscillations are
caused by blood pressure pulses against the cuff. The oscillation with the greatest amplitude is the mean pressure. Once
the mean pressure is determined, the systolic and diastolic pressures are calculated with reference to the mean.

Simply stated, auscultation measures systolic and diastolic pressures and the mean pressure is calculated. The
oscillometric method measures the mean pressure and determines the systolic and diastolic pressures.

As specified in IEC 60601-2-30/EN60601-2-30, NIBP monitoring is allowed while an electrosurgical operation is in


progress or a defibrillation shock is being delivered.

NIBP diagnostic significance must be decided by the doctor who performs the measurement.

NOTE
z Blood pressure measurements determined with this device are equivalent to those obtained by a trained
observer using the cuff/stethoscope auscultatory method or an intra-arterial blood pressure measurement
device, within the limits prescribed by the American National Standard, Manual, electronic, or automated
sphygmomanometers.

13.2 Safety

WARNING
z Be sure to select the correct patient category setting for your patient before measurement. Do not apply the
higher adult settings for pediatric or neonatal patients. Otherwise it may present a safety hazard.

z Do not measure NIBP on patients with sickle-cell disease or any condition where skin damage has occurred
or is expected.

z Use clinical judgment to determine whether to perform frequent unattended blood pressure measurements
on patients with severe blood clotting disorders because of the risk of hematoma in the limb fitted with the
cuff.

13-1
z Do not use the NIBP cuff on a limb with an intravenous infusion or arterial catheter in place. This could cause
tissue damage around the catheter when the infusion is slowed or blocked during cuff inflation.

z If you doubt the NIBP readings, determines the patient’s vital signs by alternative means and then verify
that the equipment is working correctly.

13.3 Measurement Limitations


Measurements are impossible with heart rate extremes of less than 40 bpm or greater than 240 bpm, or if the patient is
on a heart-lung machine.

The measurement may be inaccurate or impossible,

„ If a regular arterial pressure pulse is hard to detect;

„ With excessive and continuous patient movement such as shivering or convulsions;

„ With cardiac arrhythmias;

„ Rapid blood pressure changes;

„ Severe shock or hypothermia that reduces blood flow to the peripheries;

„ Obesity, where a thick layer of fat surrounding a limb dampens the oscillations coming from the artery.

13.4 Measurement Modes


There are three modes of measuring NIBP:

„ Manual: measurement on demand.

„ Auto: continually repeated measurements at set intervals.

„ STAT: continually rapid series of measurements over a five minute period, then return to the previous mode.

13.5 Measuring Procedure


13.5.1 Preparing for NIBP Measurement
1. Verify that the patient category is correct. Change it if necessary.

2. Plug the air tubing into the equipment’s NIBP connector.

3. Select a correct sized cuff by referring to the limb circumference marked on the cuff. The width of the cuff should
be 40% (50% for neonates) of the limb circumference, or 2/3 of the upper arm’s length. The inflatable part of the
cuff should be long enough to encircle at least 50% to 80% of the limb.

4. Apply the cuff to an upper arm or thigh of the patient and make sure the Φ or marking on the cuff matches the
artery location. Do not wrap the cuff too tightly around the limb. It may cause discoloration, and ischemia of the
extremities. Make sure that the cuff edge falls within the marked range. If it does not, use a cuff that fits better.

5. Connect the cuff to the air tubing. Avoid compression or restriction of pressure tubes. Air must pass unrestricted
through the tubing.

6. Switch the Mode Select knob to Monitor.

13-2
Warning
z Continuous CUFF pressure due to connection tubing kinking may cause blood flow interference and
resulting harmful injury to the patient.

13.5.2 Starting and Stopping NIBP Measurements


You can start and stop NIBP measurements by using the hardkey on the equipment’s front panel.

13.5.3 Correcting the Measurement


The cuffed limb should be at the same level as the patient’s heart. If not, correct the measurement by:

„ Add 0.75 mmHg (0.10 kPa) for each centimeter higher, or

„ Deduct 0.75 mmHg (0.10 kPa) for each centimeter lower.

13.5.4 Enabling NIBP Auto Cycling


1. Select the NIBP parameter area to access the [NIBP Setup] menu.

2. Select [Interval] and then select a desired time interval. Selecting [Manual] switches to manual mode.

3. Start the measurement manually. The equipment will then automatically repeat NIBP measurements at the set time
interval.

Warning
z Continuous non-invasive blood pressure measurements may cause purpura, ischemia and neuropathy in
the limb with the cuff. Inspect the application site regularly to ensure skin quality and inspect the extremity
of the cuffed limb for normal color, warmth and sensitivity. If any abnormity occurs, move the cuff to
another site or stop the blood pressure measurements immediately.

13.5.5 Starting a STAT Measurement


1. Select the NIBP parameter area to access the [NIBP Setup] menu.

2. Select [NIBP STAT]. The STAT mode initiates a 5-minute continuous, automatic NIBP measurement.

13-3
13.6 Understanding the NIBP Numerics
The NIBP display shows numerics only as below. Your display may be configured to look slightly different.
1 2

3 7

4
8
5

6
1. Measurement mode

2. Pressure unit: mmHg or kPa

3. NIBP alarm high limit

4. NIBP alarm low limit

5. Time of last measurement

6. Systolic pressure

7. Diastolic pressure

8. Mean pressure at the completion of measurement, or cuff pressure during the measurement

13.7 Setting Initial Cuff Inflation Pressure


You can set the initial cuff inflation pressure manually. In the [NIBP Setup] menu, select [Initial Pressure] and then
select the appropriate setting. The cuff is inflated accordingly at the next NIBP measurement.

The table below list the setting range of initial cuff inflation pressure:
Patient category Range (mmHg) Default (mmHg) Accuracy (mmHg)
Adult 80 to 280 160 10
Pediatric 80 to 210 140
Neonate 60 to 140 90

NOTE
z Setting initial cuff inflation pressure is disabled during NIBP measurement.

z The initial cuff inflation pressure is restored to the default setting if NIBP module has been reset or patient
category has been changed.

13.8 Setting Pressure Unit


To set pressure unit, access configuration management. In the Configuration-Main menu, select [NIBP Setup]→[Press.
Unit] and toggle between [mmHg] and [kPa].
You can change NIBP unit only through configuration management.

13-4
14 Marking Events
During patient monitoring or therapy, some events may exert effects on the patient and as a result change related
waveforms and parameter values. To help analysing the waveforms or numerics at that time, you can mark these events.

Before marking an event, you can define events A to H, for example, define event D as injecting Atropine. You can only
define an event through configuration management. Event A is always [Generic] and cannot be changed.

To mark an event,

1. In Monitor mode, Manual Defib mode, or Pacer mode, press the [Event] button on the front panel to access the
[Mark Event] menu as shown below.

2. Select an event you want to mark from [A] to [H], or select [Exit] to return to the main screen.

In AED mode, pressing the [Event] button directly marks Event A “Generic”.

When you mark an event, the event name and the time when the event is triggered will be displayed at the prompt
information area. This information disappears automatically after a period of 5 seconds.

14-1
FOR YOUR NOTES

14-2
15 Freezing Waveforms
During patient monitoring, the freeze feature allows you to freeze the currently displayed waveforms on the screen so
that you can have a close examination of the patient’s status. Besides, you can select any frozen waveform for recording.
Waveforms can be frozen only in the Monitor Mode.

15.1 Freezing Waveforms


In the Monitor mode, select the [Freeze] soft key, then all waveforms on the screen stop refreshing or scrolling and the
[Freeze] menu pops up, as shown below. The [Freeze] soft key changes to [Unfreeze], while the parameter area remains
refreshing properly.

The equipment can freeze the waveforms for 120 seconds.

15.2 Reviewing Frozen Waveforms


When waveforms are frozen, you can view the waveforms by selecting the [Scroll] button and then rotating the
navigation knob clockwise or counterclockwise to move the frozen waveforms right or left.

At the lower right corner of the bottommost waveform, there is an upward arrow. The frozen time is displayed below the
arrow. With each step or click, the frozen time changes at intervals of 1 second. The time can be applied to all the
waveforms on the screen.

15-1
15.3 Unfreezing Waveforms
To unfreeze the frozen waveforms, you can either:

„ Press the [Unfreeze] soft key, or

„ Select [Exit] to return to the Main screen, or

„ Perform any other action that causes the screen to be readjusted or opens a menu, such as plugging in or out a
module, pressing the [Lead Select] or [Main Menu] button, etc.

15.4 Recording Frozen Waveforms


1. In the [Freeze] menu, select, in turn, [Wave 1], [Wave 2] and [Wave 3] and then select your desired waveforms.

2. Select the [Record] button. The selected waveforms and all numerics at the frozen time will be printed out by the
recorder.

15-2
16 Review
16.1 Reviewing Events
The equipment can automatically record and save patient events. You can review patient events following this
procedure:

To review events, you can:

„ In the Monitor, Manual Defib or Pacer mode, press the Main Menu button on the front panel, and then select
[Review>>]→[Review Events >>] to enter the [Review Events] menu, or

„ In the Monitor mode, repeatedly press the [Trends] soft key to enter the [Review Events] menu.

The following figure shows the Review Events menu.

In the [Review Events] menu, you can

„ Select [Event Type], and then select [Operator Initiated], [Phys. Alarm], [Arrhythmia], [Tech. Alarm] or [All] to
review events as desired.

„ Select [Prev/Next] to page up/down to view more events.

„ Select [Index] to enter the [Index] menu. In this menu, you can set the time span in which the events happened.

„ Select [Record] to print out the current event list.

„ Select [Previous Menu] to return to previous menu.

„ Select [Exit] to return to the Main screen.

Patient events will be saved as archived events when the equipment is turned off. In case of power failure, the saved
patient events will not be cleared or lost; they will be turned into archived events instead.

16-1
16.2 Reviewing Tabular Trends
In the Monitor, Manual Defib o Pacer mode, press the Main Menu button on the front panel; select
[Review>>]→[Trends>>] or, if you are operating in the Monitor mode, select the [Trends] soft key to enter the tabular
trends menu, as shown below:

16-2
17 Data Management
17.1 Introduction
The data management function enables you to:

„ Edit patient information;

„ Review patient events; and

„ Export patient data to USB memory.

To access data management, press the Main Menu button on the front panel to enter the Main Menu, and then select
[Others>>]→ [Archives>>]. A prompt pops up as shown below:

Select [Yes] to enter the Archives Main screen as shown below

Only in Monitor, Manual Defib and Pacer mode can you access archive mode. When you enter the Archives Main screen,
patient monitoring and therapy automatically end and the patient last admitted will be saved as the latest archived
patient.

17-1
17.2 Reviewing Patient Events
To view patient events, select a patient in the Archives Main screen, and then press the navigation knob to confirm the
selection. In this case, you can select the [Return] soft key to return to the Archives Main screen.

To edit patient information, select the [Patient Info] button and change the patient information as desired. Then you
can select the [Review Events] button to return to the Review Events screen or the [Return] soft key to return to the
Archives Main screen.

17.3 Exporting Data


In the Archives Main screen,

1. Select [Export Data] to enter the Export Data screen, in which select [USB Memory]. Then, the system starts
searching for USB memory and enters the data export screen if the memory is found.

2. Select the data you want to export and then press the [Export] button.

During data export, the message “Exporting Data. Please Wait...” appears in the prompt information area and a
progress bar is displayed. If an exception happens, data export stops automatically and the reason for interruption is
presented in the prompt information area.

After the data is completely exported, you can select the [Return] soft key to return to the Archives Main screen.

NOTE

z Do not remove the USB flash memory from the equipment before data is completely exported.

17-2
18 Recording
18.1 Using a Recorder
The thermal recorder records patient information, measurement numerics and up to three waveforms.

18.2 Recording Types


By the way recordings are triggered, they can be classified into the following categories:

1. Manually-triggered realtime waveform recordings.

2. Event-triggered recordings.

3. Alarm recordings triggered by an alarm limit violation or an arrhythmia event.

4. Manually-triggered, task-related recordings.

The task-related recordings include:

„ Frozen wave recording

„ Tabular trends recording

„ Event recording

„ Parameter alarm recording

„ Event review recording

„ Event Summary Report

„ Check report

„ Configuration recording

For details about alarm recording, refer to 5 Alarm.


For details about task-related recordings, refer to respective sections of this manual.

18.3 Starting and Stopping Recordings


To manually start a recording, you can either

„ Press the hardkey on the front of the recorder,


„ Select the [Record] button from the current menu or window.

At the completion of recording, two columns of “*” marks will be printed to indicate the end of recording.

18-1
Automatic recordings will be triggered in the following conditions:

„ If both [Alarm] and [Alm Rec] for a measurement are switched on, an alarm recording will be triggered
automatically as an alarm occurs.

„ When related event is triggered.

To manually stop a recording, you can press the hardkey again.

Recordings stop automatically when:

„ A recording is completed.

„ The recorder runs out of paper.

„ The recorder has a failure.

„ Operating mode is changed.

NOTE
z If you change the ECG Lead, Gain or Filter during recording, the recorded ECG waveform changes
accordingly, but the label of Lead, Gain or Filter recorded remains unchanged.

18.4 Setting the Recorder


18.4.1 Accessing the Record Setup Menu
To access the [Record Setup] menu, press the Main Menu button on the front panel and select [Others>>]→[Record
Setup >>].

18.4.2 Selecting Waveforms for Recording


The recorder can record up to 3 waveforms at a time. You can select, in turn, [Wave 1], [Wave 2] and [Wave 3] in the
[Record Setup] menu, and then select the waveforms you want to record. You can also turn off a waveform recording by
selecting [Off]. These settings are intended for realtime recordings.

18.4.3 Setting the Realtime Recording Length


After starting a realtime recording, the recording time depends on your equipment’s settings.

1. Enter the [Record Setup] menu.

2. Select [Wave Length] and toggle between [8 s], [16 s] and [32 s].

‹ [8 s]: record a waveform 4 seconds before and 4 seconds after the current moment;

‹ [16 s]: record a waveform 8 seconds before and 8 seconds after the current moment;

‹ [32 s]: record a waveform 16 seconds before and 16 seconds after the current moment;

18-2
18.4.4 Changing the Recording Speed
1. Enter the [Record Setup] menu.

2. Select [Paper Speed] and toggle among [6.25 mm/s], [12.5 mm/s], [25 mm/s] and [50 mm/s].

This setting is for all recordings containing waveforms.

18.4.5 Switching Gridlines On or Off


1. Enter the [Record Setup] menu.

2. Select [Gridlines] and toggle between [On] and [Off].

‹ [On]: show gridlines when recording waveforms..

‹ [Off]: hide gridlines when recording waveforms.

If [Gridlines] is switched on, gridlines will be printed at the waveform area.

18.5 Loading Paper


1. Use the latch at the upper right of the recorder door to pull the door open.

2. Insert a new paper roll into the compartment as shown below.

3. Close the recorder door.

4. Check if paper is loaded correctly and the paper end is feeding from the top.

Recorder

CAUTION
z Use only specified thermal paper. Otherwise, it may cause damage to the recorder’s print head, the recorder
may be unable to print, or poor print quality may result.

z Never pull the recorder paper with force when a recording is in process. Otherwise, it may cause damage to
the recorder.

z Do not leave the recorder door open unless you have to reload paper or remove troubles.

18-3
18.6 Removing Paper Jam
If the recorder works incorrectly or produces unusual sounds, check if there is a paper jam first. If a paper jam is detected,
follow this procedure to remove it:

1. Open the recorder door.

2. Take out the paper and tear off the draped part.

3. Reload the paper and close the recorder door.

18.7 Cleaning the Recorder Print head


If the recorder has been used for a long time, deposits of paper debris may collect on the print head compromising the
print quality and shortening the lifetime of the roller. Follow this procedure to clean the print head:

1. Take measures against the static electricity, such as disposable wrist strap, for the work.

2. Open the recorder door and take out the paper.

3. Gently wipe around the print head using cotton swabs dampened with alcohol.

4. After the alcohol has completely been dried, reload the paper and close the recorder door.

CAUTION
z Do not use anything that may destroy the thermal element.

z Do not add unnecessary force to the thermal head.

18-4
19 Configuration Management
19.1 Introduction
Configurations management enables you to customize you equipment to best meet your needs. With this function, you
can:

„ Change system configuration;

„ Record system configuration;

„ Restore the factory default configuration.

After the system configurations have been changed, the equipment restarts and new configuration settings take effect
immediately.

19.2 Password
Accessing configuration management is password protected. The required password is set to 315666 before the
equipment leaves the factory.

19.3 Accessing Configuration Management


To access configuration management, move the Mode Select knob to Monitor, Manual Defib or Pacer, and then press the
Main Menu button on the front panel. Patient monitoring and therapy automatically end when you enter the
Configuration Main menu.

1. Press the Main Menu button on the front panel of the equipment. Select [Others >>]→ [Configuration >>], a
dialog box pops up as shown below:

„ To view the configurations or change system time, select [View Config]. In this case, password is not required.

„ To close the dialog box and return to normal operating mode, select [Cancel].

19-1
2. Enter the required password and then select [OK] to enter the Configuration Main menu as shown below:

„ Selecting [Factory Config] and confirming the selection restores all the current settings to factory default settings:

„ Selecting [Record] records the settings of all system configurations.

„ Selecting [Exit] pops up a dialog box as shown below:

WARNING
z Never connect the equipment with the patient while performing configuration management.

19.3.1 General Setup Menu


Menu Item Options/Range Default Remark
Device Name 20 characters / The characters are included in the
Institution 20 characters / keyboard. Restoring factory
Department 20 characters / default configurations does not

Bed No. 20 characters / change these items.

Patient Cat. Adu, Ped, Neo Adu /


Height Unit cm, inch cm /
Weight unit kg, lb kg /
Chinese, English, French, German, Italian,
Language Polish, Spanish, Portuguese, Russian, Czech, / /
Turkish, Dutch, Hungarian
Data Format yyyy-mm-dd, mm-dd-yyyy, dd-mm-yyyy yyyy-mm-dd /
Time Format 12 h, 24 h 24 h /
19-2
Menu Item Options/Range Default Remark
Year 2007 to 2099 2007 /
Month 01 to 12 01 /
Day 01 to 31 01 /
System
24 h: 00 to 23 24 h: 00
Time Hour /
12 h: 12AM to 11PM 12 h: 12AM
Minute 00 to 59 00 /
Second 00 to 59 00 /

19.3.2 Manual Defib Setup Menu


Menu Item Options/Range Default Remark
Manual Therapy Access Direct, Confirmed, Password Direct /
(0000-9999). [Set Password] is active
Set Password 4 digits 0000 only when [Manual Therapy Access]
is set to [Password].

Pads Default 2, 5, 10, 50, 100, 150, 170, 200, 300 J 200 J /

Internal Default 2/5/10/20/30/50 J 10J /


Time to Auto Disarm 30s, 60s, 90s, 120s 60s /
Syn After Shock Yes, No No /
Remote Sync On, Off Off /
Monitor Para. SpO2, Off Off
Charge Tone Vlm High, Med, Low Med /
Contact Impedance Indicator On, Off Off /

19.3.3 AED Setup Menu


Menu Item Options/Range Default Remark
In AED mode, Iif Shock Series change,
defibrillation energies will change
accordingly. But if the working mode has
Shock Series 1, 2, 3 1
been switched, the Shock Series will be
cleared while the energy proceeds from
the energy of the last shock.
Energy 1 100, 150, 170, 200, 300, 360J 200 J ≤Energy 2
Energy 2 Energy 1 to 360J 300 J Cannot be less than Energy 1
Energy 3 Energy 2 to 360J 360 J Cannot be less than Energy 2
Time to Auto Disarm 30s, 60s, 90s, 120s 30s /
Initial CPR Time Off, 30s, 60s, 90s, 120s, 150s, 180s Off /
CPR Time 30, 60, 90, 120, 150, 180s 120s /
CPR Metronome On, Off On /
CPR Mode 30:2, 15:2, Compression_only 30:2 /
NSA Action Monitor, CPR CPR /
19-3
Menu Item Options/Range Default Remark
Voice Prompts On, Off On /
Voice Volume High, Med, Low High /
Voice Prompt Interval Off, 30s, 60s, 90s, 120s, 150s, 180s 30s /
Voice Recording On, Off Off /

19.3.4 Pacer Setup Menu


Menu Item Options/Range Default
Pacer Rate 40 to 170 ppm 70 ppm
Pacer Output 0 to 200 mA 30 mA
Default Pacer Mode Demand Mode, Fixed Mode Demand Mode

19.3.5 ECG Setup Menu


Menu Item Options/Range Default Remark

ECG Standard AHA, IEC AHA /


Notch Filter 50Hz, 60Hz 50Hz /
ECG Bandwidth Monitor, Therapy Therapy /
The default waveform position in
[Waveform Setup] menu is determined
Lead Set 3-Lead, 5-Lead 3-Lead
according to the setting of [Lead set]
defined in [ECG Setup] menu.
This setting is linked with the [Beat Vol]
QRS Volume 0 to 10 2
setting in the [SpO2 Setup] menu
3-Lead: II, I, III Available options are defined by the
ECG1 II
5-Lead: II, I, III, aVL, aVR, aVF, V current setting of [Lead set].

3-Lead: / / 1. If 3-lead is used, the default setting of


this item is blank and this item is
ECG2 inactivate.
5-Lead: II, I, III, aVL, aVR, aVF, V V
2. The options of ECG1 are not available
for ECG2.
Sweep 6.25/12.5/25/50 mm/s 25 mm/s Paper speed
HR Alarm On, Off On /
HR Alm Lev High, Med, Low Med /
Adu (Low+2) to 300 120
HR High Ped 160 /
(Low+2) to 350
Neo 200
Adu 50
HR Low Ped 15 to (High-2) 75 /
Neo 100

19-4
Menu Item Options/Range Default Remark
Arrhythmia On, Off Off /
ARR Alarm On, Off On /
ARR Alm Lev PVCs/min High, Med, Low Med /
R ON T
VT>2
Couplet
PVC
Vent. Rhythm
Bigeminy
Trigeminy
Tachy
Brady
PNP
PNC
Missed Beat
Multif. PVCs
Nonsus. Vtac
Pause
Irr. Rhythim
Asystole Delay 3 to 10 5 /
V-Tach Rate 100 to 200 130 /
V-Tach PVCs 3 to 99 6 /
Vbrd Rate 15 to 60 40 /
Vbrd PVCs 3 to 99 5 /
Extreme Adu 60 to 300 160 /
Tachy Ped 60 to 300 180 /
Extreme Adu 15 to 120 35 /
Brady Ped 15 to 120 50 /
PVCs High 1 to 10 10 /
Adu 60 to 300 120 /
Tachy
Ped 60 to 300 160 /
Adu 15 to 120 50 /
Brady
Ped 15 to 120 75 /
Multif. PVCs Window 3 to 31 15 /
Pause Time 1.5, 2.0, 2.5 2.0 /

19-5
19.3.6 Resp Setup Menu
Menu Item Options/Range Default Remark
RR Alarm On, Off Off /
RR Alm Lev High, Med, Low Med /
Sweep 6.25/12.5/25 mm/s 6.25mm/s /
Adu 30 /
(Low+2) to 100
RR High Ped 30 /
Neo (Low+2) to 150 100 /
Adu 8 /
RR Low Ped 6 to (High-2) 8 /
Neo 30 /
Apnea Time 10/15/20/25/30/35/40s 20s /

19.3.7 SpO2 Setup Menu


Menu Item Options/Range Default Remark
SpO2 Alarm On, Off On /
SpO2 Alm Lev High, Med Med /
Sweep 12.5mm/s, 25 mm/s 25 mm/s /
Adu 100
SpO2 High Ped (Low+1) to 100 100 /
Neo 95
Adu 90
SpO2 Low Ped Desat Limit to (High-1) 90 /
Neo 90
Adu 80
Desat Ped 50 to (High-1) 80 /
Neo 80
Averaging Masimo SpO2 2-4s, 4-6s, 8s, 10s, 12s, 14s, 16s 8s For Masimo SpO2 module only.
Mindray SpO2 High, Med, Low Med Different options are available to
Sensitivity
Masimo SpO2 Normal, Maximum Normal match the SpO2 module used.

Pitch Tone Mode 1, Mode 2 Mode 1 /


Sat-Seconds 0s, 10s, 25s, 50s, 100s 0s For Nellcor SpO2 module only.

19.3.8 PR Setup Menu


Menu Item Options/Range Default Remark
PR Alarm On, Off Off /
PR Alm Lev High, Med, Low Med /

Adu 120 PR ranges are different for different

PR High (Low+2) to 240 modules.


Ped 160
Neo 200

19-6
Menu Item Options/Range Default Remark
Adu 50
PR Low Ped 25 to (High-2) 75
Neo 100
This setting is linked with the [QRS
QRS Volume 0 to 10 2 Volume] setting in the [ECG Setup]
menu

19.3.9 NIBP Setup Menu


Menu Item Options/Range Default Remark
Manual, 1 min, 2 min, 2.5 min, 3 min, 5 /
Interval min, 10 min, 15 min, 20 min, 30 min, 1 h, Manual
1.5 h, 2 h, 3 h, 4 h, 8 h
Alarm limits are refreshed in real time
Press. Unit mmHg, kPa mmHg
when pressure unit is changed.
NIBP-Sys
Alarm NIBP-Dia On, Off On /

NIBP-Mean
Alm Lev High, Med Med /
Adu (Low + 5) to 270 160
Sys High Ped (Low + 5) to 200 120 /
Neo (Low + 5) to 135 90
Adu 90 /
Sys Low Ped 40 to (High - 5) 70 /
Neo 40 /
Adu (Low + 5) to 230 110 /
Mean High Ped (Low + 5) to 165 90 /
Neo (Low + 5) to 110 70 /
Adu 60 /
Mean Low Ped 20 to (High - 5) 50 /
Neo 25 /
Adu (Low + 5) to 210 90 /
Dia High Ped (Low + 5) to 150 70 /
Neo (Low + 5) to 100 60 /
Adu 50 /
Dia Low Ped 10 to (High - 5) 40 /
Neo 20 /

19-7
19.3.10 Alarm Setup Menu
Menu Item Options/Range Default

Alarm Pause Time 1, 2, 3, 5, 10, 15 min, Permanent 2 min

Audio Off Enabled, Disabled Disabled

0 to 10 (If Audio Off is enabled),


Alarm Volume 2
1 to 10 (If Audio Off is disabled)

Reminder Tone On, Off Off

Reminder Volume High, Med, Low Med

High Level Alarm 3 to 15s 10s


Alarm Tone
Med Level Alarm 3 to 30s 20s
Interval
Low Level Alarm 15 to 100s 20s

Latching Alarms Yes, No No

Displaying Alarm Limits Yes, No Yes

ECG Lead Off Lev High, Med, Low Low

SpO2 Sensor Off Lev High, Med, Low Low

No Battery Status Indicator On, Status Indicator Off Status Indicator On

19.3.11 Waveform Setup Menu


Menu Item Options Default Remark

Wave 1 / ECG1 Unchangeable

Wave 2 Pleth, Resp Pleth For optional parameters, options of


Pleth, Resp waveforms are available only when
Wave 3 Resp
the parameters are configured.
ECG Green
Parameter/ RESP Green, Yellow, Cyan, White, Red, Blue, Purple, Yellow
/
Wave Color SpO2 Orange Cyan
NIBP White

19-8
19.3.12 Mark Event Setup Menu
Menu Item Options/Range Default Remark

Event A Generic Generic Unchangeable

Event B Adrenalin
Adrenalin, Lidocaine, Atropine, Nitroglycerin, Event names that have been
Event C Lidocaine
Morphine, Intubation, IV Access, Adenosine, selected by previous events will
Event D Atropine
Amiodarone, Vasopressin, Isoprenaline, not be included in the options of
Event E Nitroglycerin
Dopamine, Aspirin, Oxygen, CPR later events.
Event F Morphine
Options: entering event name using the
Custom Event 1 After being defined, custom event
keyboard included in the [Mark Event Setup]
/ is refreshed in real-time in the
menu.
Custom Event 2 Mark Event list.
Range: 1 to 20 characters

19.3.13 Record Setup Menu


Menu Item Options/Range Default
Wave Length 8s, 16s, 32s 8s
Paper Speed 6.25 mm/s, 12.5 mm/s, 25mm/s, 50mm/s 25mm/s
Gridlines On, Off Off
Energy Delivered Output On, Off Off
Charge Event Off
Shock Event On, Off On
Auto Record
Marked Event Off

Auto Test Report On, Off, Only If Failed On


HR
ARR
PVCs
Alm Rec Resp On, Off Off
SpO2
PR
NIBP

19.3.14 Data Management Setup Menu


Menu Item Options/Range Default
Tabular Trends Interval 1, 2, 5, 10, 15, 30, 60 min 5 min
Event Wave Length 8s, 16s, 32s 16 s

19-9
19.3.15 User Test Setup Menu
Menu Item Options/Range Default
User Test Prompt On, Off Off

12h time format: 12:00 AM ~ 05:00 AM 03:00 AM


Auto Test Time
24h time format: 00:00 ~ 05:00 03:00

19.3.16 Others Menu


Menu Item Options/Range Default
Brightness 1 to 10 8
Key Volume 0 to 10 2
Wave Line Thick, Medium, Thin Medium
Draw wave Mono, Color Color
SPO2

Modules Resp On, Off On

NIBP

19-10
20 Battery
20.1 Introduction
The equipment is designed to operate on battery power when external power supply is not available. The battery is
charged whenever the equipment is connected to AC mains or the DC power supply through an external DC/AC adapter,
regardless of whether or not the equipment is currently turned on. In case of power failure, the equipment will
automatically run power from internal batteries. So we recommend you always install a fully charged battery in the
equipment.

The equipment is configured with one smart lithium ion batteries which are free of maintenance.

On-screen battery symbols indicate the current battery charge status, taking Battery 1 for an example:

„ ≤100%, but >80% of capacity

„ ≤80%, but >60% of capacity

„ ≤60%, but >40% of capacity

„ ≤40%, but >20% of capacity

„ ≤ 20% of capacity

„ Low battery and charging is required immediately

„ Battery is not installed


You can also check the battery’s charge status by pressing the fuel gauge button on the battery to illuminate the battery
gauge. The fuel gauge consisting of 5 LEDs, each LED represents a charge of approximately 20% of capacity.

If the battery charge is too low, a technical alarm will be triggered and the “Low Battery” message displayed in the
Technical Alarm Area. At this moment, change the battery or apply external power to the equipment.

WARNING
z Keep the batteries out of children’s reach.

z Use only specified batteries.

z The batteries should be charged in this equipment or in a device approved by the equipment manufacturer.

NOTE
z Always connect the equipment to AC mains whenever it is possible.

z Always install a fully charged battery in the equipment.

z After long term use, the power capacity indicated by the battery symbol may be different from the actual
capacity. Always observe the alarm information displayed on the screen.

20-1
20.2 Installing the Batteries
To install the batteries:

1. Align a battery with the battery compartment.

2. Insert the battery, and press until you hear it click into the place.

To replace a battery, press the latch on the battery and push the battery to the right until you remove it. Insert a new
battery into the battery compartment.

20.3 Battery Alarms


20.3.1 No Battery Alarm
You can configure the status of status indicator when no battery is installed in the configuration mode.

Press the Main Menu button on the front panel of the equipment. Select [Others >>]→ [Configuration >>]→enter the
required password. In the Configuration main menu, select [Alarm Setup]→ [No Battery], and toggle between [Status
Indicator On] and [Status Indicator Off]. The default setting is [Status Indicator Off].

‹ If [Status Indicator On] is selected, the red cross status indicator flashes in the case that battery is not installed,
meanwhile a message “No Battery” is presented in the Technical Alarm Area.

‹ If [Status Indicator Off] is selected, the red cross status indicator will not be illuminate in the case that battery
is not installed; the message “No Battery” will be presented in the Technical Alarm Area.

20.3.2 Low Battery Alarm


If the equipment is run on battery and the battery charge is low, a technical alarm “Low Battery” will be triggered. In this
case, replace the battery with a fully charged on or connect the equipment to external power supply immediately.

If the battery is almost depleted, a prompt “Battery depleted! System will shut down imminently. Connect to AC
mains or replace battery” pops up. In monitoring, manual defibrillation, and pacing mode, additional alarm lights and
alarm tones are provided. In this case, take appropriate actions immediately. This prompt will not disappear until the
battery is replaced or the equipment is connected to the external power supply. The equipment automatically shuts
down if no action is taken within a period of about 3 minutes.

NOTE

z The Low Battery alarm means that the battery is beginning to weaken and should be replaced at the first
opportunity. At least 20 minutes of monitoring and six full energy shocks can be performed when the Low
Battery alarm is activated. Replace the battery or connect the equipment to AC mains as soon as possible.

20-2
20.3.3 Battery Aged Alarm
If the battery runtime is significantly shorter than the specification, a low level technological alarm “Battery Aged” will
be presented. We recommend you to contact our company and replace it with a new one.

20.3.4 Battery Error Alarm


In the situation that the battery has a failure, a high level technological alarm “Battery Err” will be presented. In this case,
replace the battery or contact your service personnel.

20.4 Checking the Batteries


Battery capacity diminishes with use and age. To check the performance of the batteries, follow this procedure:

1. Connect the equipment to the external power supply and allow the batteries to charge uninterruptedly till it is fully
charged.

2. Remove the external power supply and allow the equipment to run from the batteries until it shuts off.

The operating time of the batteries reflects their performance directly. If the operating time of the batteries is noticeably
shorter than that stated in the specifications, discard the batteries or contact your service personnel.

NOTE

z Life expectancy of a battery depends on how frequent and how long it is used. When properly cared for, the
lithium-ion battery has a useful life of approximately 2 years. For improper use models, life expectancy can
be less. We recommend replacing lithium-ion batteries every 2 years.

z To optimize performance, a depleted, or nearly depleted battery should be charged as soon as possible.

z Battery operating time depends on the device configuration and operation.

20.5 Charging batteries


The batteries can be charged only when they are installed in the equipment or using an device approved by the
equipment manufacturer. With the equipment turned off and at a temperature of 25°C (77°F), a completely discharged
battery charges to 80% of its capacity in approximately 2 hours, and to 100% of its capacity in approximately 3 hours.
Batteries charge at a lower rate with the equipment turned on.

Batteries should be charged at temperatures between 0°C (32°F) to 45°C (113°F). To optimize performance, a fully (or
nearly fully) discharged battery should be charged as soon as possible.

20-3
20.6 Storing Batteries
When storing batteries, make sure that the battery terminals do not come into contact with metallic objects. If batteries
are stored for an extended period of time, they should be placed in a cool place with a partial charge of 40% to 60%
capacity (3 LEDs illuminated). Storing batteries in a cool place slows the aging process. The idea storage temperature is
15°C (60°F). Batteries should not be stored at temperature outside the range of -20°C (-4°F) to 60°C (140°F).

Remove the battery from the equipment if the equipment is not used for a prolonged time. Otherwise the battery may
be over discharged and it will take significantly longer time to charge the battery. Stored batteries should be charged
every 2 months to 40% to 60% of full capacity. They should be charged to full capacity prior to use.

NOTE

z Do not store a battery in the equipment if the equipment is not used for an extended period of time.

z Storing batteries at temperature above 38°C(100°F) for an extended period of time significantly shorten the
life expectancy of a battery

20.7 Recycling the Batteries


A battery should be discarded if there are visual signs of damage, the battery fails, the battery aged alarm is presented,
or the batteries has been used for more than two years. Properly dispose of batteries according to local regulations.

WARNING
z Do not disassemble, puncture or incinerate batteries. Do not short the battery terminals. They may ignite,
explode, or leak, causing personal injury.

20-4
21 Care and Cleaning
Use only the substances approved by the equipment manufacturer and methods listed in this chapter to clean or
disinfect your equipment. Warranty does not cover damages caused by unapproved cleaning and disinfection
substances or methods.

We make no claims regarding the efficacy of the listed chemicals or methods as a means for controlling infection. For the
method to control infection, consult your hospital’s infection control officer or epidemiologist.

In this chapter we only describe cleaning and disinfection of the main unit. For the cleaning and disinfection of external
paddles and other reusable accessories, refer to instructions for use of corresponding accessories.

21.1 General Points


Keep you equipment and accessories free of dust and dirt. To avoid damage to the equipment, follow these rules:

„ Always dilute according the manufacturer’s instructions or use lowest possible concentration.

„ Do not immerse part of the equipment into liquid.

„ Do not pour liquid onto the equipment or accessories.

„ Keep the paddles clean, Before user checks or after each use, thoroughly clean the paddles and paddle tray.

„ Do not allow liquid to enter the case.

„ Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone or
acetone-based cleaners).

WARNING
z Be sure to shut down the system, disconnect the power cord and other cables, and remove the
batteries before cleaning the equipment.

CAUTION
z Contact your service personnel in case of spilling liquid on the equipment or accessories.

NOTE
z To clean or disinfect reusable accessories, refer to the instructions for use delivered with the
accessories.

21-1
21.2 Cleaning
Your equipment should be cleaned on a regular basis. If there is heavy pollution or lots of dust and sand in your place,
the equipment should be cleaned more frequently. Before cleaning the equipment, consult your hospital’s regulations
for cleaning the equipment.

Recommended cleaning agents are:

„ sodium hypochlorite bleach (diluted)

„ Hydrogen peroxide (3%)

„ Ethanol (70%)

„ Isopropanol (70%)

To clean your equipment, follow these rules:

1. Shut down the equipment, disconnect the power cord and other cables, and remove the batteries.

2. Clean the display screen using a soft, clean cloth dampened with a glass cleaner.

3. Clean the exterior surface of the equipment using a soft, clean cloth dampened with a glass cleaner.

4. Clean the paddle tray using a soft, clean cloth dampened with a glass cleaner.

5. Wipe off all the cleaning solution with a dry cloth after cleaning if necessary.

6. Dry your equipment in a ventilated, cool place.

21.3 Disinfecting
Disinfection may cause damage to your equipment and is therefore not recommended unless otherwise indicated in
your hospital’s servicing schedule. Cleaning equipment before disinfecting is recommended.

The recommended disinfectants include: ethanol 70%, isopropanol 70%, and Perform* classic concentrate OCY (KHSO4
solution).

21-2
22 Maintenance and Testing
WARNING
z Failure for the responsible individual, hospital or institution employing this equipment to implement a
satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.

z The safety checks or maintenance involving any disassembly of the equipment should be performed by
professional servicing personnel. Otherwise, undue equipment failure and possible health hazards could
result.

z If you find a problem with any of the equipment, contact your service personnel or the manufacturer.

22.1 Overview
Before every use, each shift or once a week, checks shall be done to ensure that the equipment is ready for operation at
any time. After the equipment has been used for 12 months, or whenever the equipment is repaired or upgraded, a
thorough inspection should be performed by qualified service personnel to ensure the reliability.

Items to be checked are as follows:

„ Shift check

„ Recorder inspection

„ User test

„ Manual defibrillation test

„ Pacing test

„ Functional modules test

„ NIBP overpressure protection test

„ Electrical safety tests.

Paddles and cables are critical parts for defibrillation but damageable. We recommend you to check the appearance and
performance of these parts every day and replace them every 3 years.

The ECG cables are critical parts for data acquisition and analysis but damageable. We recommend you to inspect the
cable as described in the Appendix C BeneHeart Defibrillator Shift Checklist.

In case of any damage or abnormity, remove the equipment from use. Contact the hospital’s biomedical engineers or
your service personnel immediately.

22-1
22.2 Maintenance and Testing Schedule
The following tests, except recorder check and user test, shall be carried out by the service personnel only. Contact your
service personnel if any maintenance is required. Make sure to clean and disinfect the equipment before any test and
maintenance.

Test item Weekly After use 12 months 24 months


Cleaning equipment and accessories ×

Routine Test ×

User test Energy delivery test ×

Controls test ×

Recorder check ×

ECG cable test ×

Charge/discharge
Manual defibrillation
Energy disarming ×
tests
Synchronous defibrillation
Pacing test ×

Accuracy test
NIBP tests ×
Leakage test

NIBP overpressure protection test ×


Electrical safety tests as per IEC60601-1 ×

22.3 Carrying Out Maintenance and Testing


22.3.1 Power-on Tests
The equipment performs self test each time you turn it on or restart the equipment. If a failure is detect the red status
indicator will flash and a message is presented in the technological alarm area.

Power-on self tests will check the following items:

„ Power module,

„ Therapy module,

„ Main control system, and

„ Monitoring function.

You need to power on the equipment every day, or when the equipment is first installed, or following any maintenance
or the replacement of any main unit parts to verify that the equipment can be turned on properly.

1. Place the external paddles on paddle tray, insert the battery in the battery compartment, and then connect the
equipment with AC mains. In this case, both the AC indicator and battery indicator shall light.

2. Turn the Mode Select knob to Monitor. Check that the equipment passes the self test and is turned on properly.

22-2
3. Check the display of technical alarm area, prompt area and battery status indicator on the upper right corner of the
main screen to judge whether the equipment runs normally.

22.3.2 Shift Check


In order to ensure your defibrillator/monitor is ready when needed, we recommended you to inspect your equipment
and complete a check list at the change of every shift. See Appendix C BeneHeart Defibrillator Shift Checklist for
details.

22.3.3 Automated Tests


The equipment automatically performs Routine Test and weekly Energy Delivery Test while the equipment is off but AC
mains is connected to check the equipment’s operational performance and alert operators if a problem exists.

The following table provides the description of automated test and lists the frequency with which each is performed.

Test name Test items Frequency


Routine Test Batteries and therapy module (includes a 1 J Once per day, between 0:00 am to 5:00 am
internal discharge, and 10 J external discharge
through external paddles or pads cable.)
Energy Delivery Test Delivers a 200 J internal discharge Once per week, at the completion of Routine Test

Automated Routine Test can be initiated between 0:00 am to 5:00 am. To set auto test time, enter the Configuration Main
men, and select [Test Setup]→ [Auto Test Time]. The default setting is 3:00 am.

The equipment displays no information on the screen during Auto Test.


In the case that Auto Test fails, the red status indicator flashes and the equipment gives a beep periodically till the
equipment is restarted. Then a low level technical alarm “Last Auto Test Failed” will be displayed. The “Last Auto Test
Failed” alarm is cleared if Auto Test is passed for the next time or the failed Routine Test or Energy Delivery Test is passed
during User Test. We recommend you to perform the User Test if Auto Test failed.

At the completion of Auto Test, a report is saved automatically. Enter the Configuration Main menu→ select [Record
Setup]→ [Auto Test Report] →select [On], [Off], or [Only if Failed], you can choose to print the Auto Test Report or not
after the tests are completed, or print it only if Auto Test Failed.

You can review the result of Auto Test by selecting the [History] softkey from the User Test Main menu.

CAUTION

z With power off, auto Test is performed only when AC mains is connected.

z Thoroughly clean the paddles and properly place them in the paddle tray after each use. Automated test
passes only when paddles properly contact the metal parts of the paddle tray.

z Install the battery and properly place the external paddles in the paddle tray or connect the pads cable and
50 Ω test load. Otherwise the Auto Test will fail.

22-3
22.3.4 User Test

WARNING
z Do not perform user test when a patient is connected to the equipment.

User test covers the following items:

„ Routine Test,

„ Energy delivery test, and

„ Controls test.

NOTE

z Before user test or after each use, thoroughly clean the paddles and properly place them in the paddle tray.
User test passes only when paddles properly contact the metal parts of the paddle tray.

z Install the battery and properly place the external paddles in the paddle tray or connect the pads cable and
50 Ω test load. Otherwise the User Test will fail.

22.3.4.1 Entering the User Test Main Menu

To access user test, press the Main Menu button and select [User Test >>]. Then a dialog box pops up, prompting you
that entering user test will end patient monitoring. Select [Yes] to enter the User Test Main menu.

Check the test items you want to perform and select [Start] to start user test. The message “Test completed” will be
presented when selected tests have been finished. Then you can press the [Record] soft key to print the test result.

22-4
22.3.4.2 Routine Test

Routine Test covers the following items:

„ Batteries,

„ Mainboard,

„ Defib/Pacer function, and

„ Monitor function

If any of above items fails, the red status indicator will flash. If mainboard, Defib/Pacer function, or monitor function fails,
a low level technical alarm “Last User Test Failed” will be displayed in the technical alarm area when the equipment is
restarted. We recommend you to perform a successful User Test to clear this alarm.

22.3.4.3 Energy Delivery Test

Energy delivery test delivers a 200J external discharge to check the defibrillation circuit.

If the test fails, the red status indicator will flash and a low level technical alarm “Last User Test Failed” will be displayed in
the technical alarm area when the equipment is restarted. We recommend you to perform a successful User Test to clear
this alarm.

22.3.4.4 Controls Test

Controls test covers the following items:

„ Mode Select knob,

„ All hardkeys on the equipment’s front panel;

„ Audio test, and

„ Display test

Follow the displayed prompts to run the controls test.

NOTE
z The tested controls are indicated in green during controls test.

If any of above items fails, the red status indicator will be illuminated and a low level technical alarm “Last User Test
Failed” will be displayed in the technical alarm area when the equipment is restarted. We recommend you to perform a
successful User Test to clear this alarm.

NOTE
z The“Off” position of the Mode Select knob is not tested during the Controls Test. If you turn the knob to
“Off” for more than 3 seconds. The equipment is turned off.

22-5
22.3.4.5 Test Summaries

The results of User Test are automatically saved as summaries. You can select the [History] button from the User Test
Main menu to review the test summaries.

The equipment can store up to 300 historical test summaries which are listed in the sequence of time, with the latest on
the top. You can use the knob to select a test result and press the knob to check the detailed test report.

22.3.4.6 User Test Reminder

We recommend that Routine Test and Energy Delivery Test are executed once per week and Controls Test once every 12
months.

Each time when the equipment is turned on, the time to last Routine Test, Energy Delivery Test and Controls Test is
checked automatically. The equipment can be configured to give a “User Test Due” message to remind you that the
equipment is due for User Test.

The [User Test Prompt] is switched off by default. You can switch it on by selecting [User Test Setup]→[User Test
Prompt]→[On] through the configuration mode.

NOTE

z In case that automated Routine Test is not performed, we recommend you to run Routine Test once per day
through User Test.

22.3.5 Recorder Inspection


1. Turn the Mode Select knob to Monitor.

2. Start recording to verify that the recorder works properly and the printout is legible and correct.

3. Simulate errors, such as removing the paper roll and losing the latch, correct information shall be shown in the
prompt area. The recorder shall work properly after the faults are corrected.

22-6
22.3.6 ECG Cable Test
It is recommended to perform ECG cable test once a year.

Test tool: ECG simulator

Follow this procedure to perform ECG cable test:

1. Turn the Mode Select knob to “Monitor”.

2. Connect the ECG cable to the defibrillator and the electrodes to the simulator.

3. Turn on the simulator and select a normal ECG rhythm.

4. Wait for a few seconds and check that the waveform is displayed normally and that no lead-off alarms reported in
the alarm information area.

22.3.7 Manual Defibrillation Test


Test tools: defibrillator/pacer analyzer

Charge/discharge

1. Remove the batteries and connect the equipment with AC mains. Turn the Mode Select knob to Manual Defib.

2. Connect the external paddles to the equipment and place the paddles on the defibrillator/pacer analyzer.

3. Enter the Configuration Main screen. From the Record Setup menu set [Shock Event] to [On] so that shock events
can be recorded automatically if happened.

4. Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as 0 or blank.

5. Select the energy level to 1J.

6. Charge/discharge the equipment to verify the energies measured by the analyzer meet the following accuracy:

Selected Energy (J) Measured Value (J)


1 0 to 3
100 85 to 115
360 306 to 414

7. Set the energy to 100J and 360J respectively. Repeat Step 6.

8. Disconnect the equipment from the AC mains. Run the equipment on fully charged battery. Move the Mode Select
knob to Manual Defib. Repeat Steps 2 to 7.

9. Verify that the equipment records the shock events automatically and correctly.

10. Use multifunctional electrode pads. Repeat Step 3 to Step 9.

Energy Disarming

1. Run the equipment on fully charged battery. Move the Mode Select knob to Manual Defib.

2. Connect the external paddles to the equipment and place the paddles on the defibrillator/pacer analyzer.

3. Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as 0 or blank.

4. Select the energy level to 360J.

5. Charge the equipment.

6. Verify that the charge tone is issued during charging.


22-7
7. Press the “Disarm” soft key to discharge the energy internally.

8. Verify that a prompt “Charge Removed” appears on the screen and the charge done tone stops.

9. Verify that the value measured by the analyzer is 0J or blank.

10. Enter the Configuration Main menu, select [Manual Therapy Setup] and set [Time to Auto Disarm] to [60s].

11. Exit “Configuration Management”. The equipment restarts automatically.

12. Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as 0 or blank.

13. Select the energy level to 360J.

14. Charge the equipment. Count time after charging is completed. Verify that the prompt “Shock Removed” appears
on the equipment and the energy measured by the analyzer is 0J or blank after 60 seconds.

15. Use multifunctional electrode pads. Repeat Steps 3 to 14.

Synchronous Defibrillation

1. Connect the external paddles and ECG cable to the equipment. Place the paddles ECG electrodes on the
defibrillator/pacer analyzer.

2. Set the analyzer to Time Measurement Mode and output normal sinus rhythms, e.g. amplitude value 1mV and HR
60bpm.

3. Enter Configuration Management. In the [Manual Therapy Setup] menu, set [Sync After Shock] to [On].

4. Adjust the energy setting of the equipment to be 10J.

5. Press the [Enter Sync] soft key to start synchronous defibrillation. If Remote Sync is switched on, press the [Enter
Sync] soft key and select [Local] to start synchronous defibrillation

6. Select Pads as the ECG source and begin charging.

7. When charging finishes, press and hold the “Shock” button to deliver a shock.

8. Verify that synchronous discharge succeeds and the delivered energy measured by the analyzer is 10J±2J.

9. Verify that the delay time of synchronous defibrillation measured by the analyzer is less than 60ms.

10. Verify that the synchronous discharge mark appears on the R wave.

11. Verify that the prompt messages are correct during testing.

12. Select lead II as ECG source and perform charging. Repeat steps 7 to 11.

13. Use multifunctional electrode pads. Repeat steps 2 to 12.

22-8
22.3.8 Pacing Test
Test tools: defibrillator/pacer analyzer

1. Run the equipment on fully charged battery. Move the Mode Select knob to Pacer. Select Fixed mode..

2. Connect the pads cable to the equipment and properly place the pads on the defibrillator/pacer analyzer.

3. Set the analyzer to Pacing Measurement mode. Use test load of 50Ω.

4. On the equipment, set [Pacer rate] to [70ppm] and [Pacer Output] to [30mA].

5. Press the [Start Pacing] soft key. Verify that the pacer rate measured by the analyzer is 70 ppm±1ppm and the
pacer output measured is 30 mA±5mA.

6. Press the “Stop Pacing” soft key, and then set [Pacer rate] to [170ppm] and [Pacer Output] to [200mA].

7. Press the [Start Pacing] soft key. Verify that the pacer rate measured by the analyzer is 170 ppm±2ppm, and the
measured current is 200 mA±10mA.

22.3.9 Peforming Testing in Installation Mode


You can view software version and format storage card.by accessing the Installation Mode.

22.3.9.1 Password for Installation Mode

Accessing installation mode is password protected. The required password is set to 888888 before the equipment leaves
the factory.

22.3.9.2 Accessing Installation Mode

You can access Installation Mode while operating in the Monitor, Manual Defib or Pacer mode. Patient monitoring and
therapy automatically end when you enter Installation Mode.

To enter Installation Mode, press the Main Menu button on the front panel, and then select [Others >>]→[Maintenance
>>]→[Installation Mode >>]→enter the required password. The Installation Mode Main menu pops up as shown
below.

22-9
22.3.9.3 NIBP Accuracy Test

The NIBP accuracy test is required at least once every two years or whenever you doubt the NIBP reading.

Tools required:

„ T-shape connector

„ Tubing

„ Balloon pump

„ Metal Vessel, volume 500±25 ml

„ Calibrated manometer for reference, accuracy superior to 1 mmHg

Follow this procedure to perform the accuracy test:

1. Connect the equipment as shown below.

Manometer
Defibrillator/monitor
Connector for Tubing
NIBP cuff

Balloon pump Metal vessel

2. Before inflation, the reading of the manometer should be 0. If not, disconnect the airway and reconnect it until the
readings is 0.

3. In the Installation Mode Main menu, select [Maintain NIBP]→ [Start Accuracy Test].

4. Compare the value of manometer with the value displayed on the equipment’s screen. The difference should be no
greater than 3 mmHg.

5. Raise the pressure in the metal vessel to 50 mmHg with the balloon pump. Repeat steps 3 and 4.

6. Raise the pressure in the metal vessel to 200 mmHg with the balloon pump. Repeat steps 3 and 4.

When the accuracy test is completed, the result will be displayed.

If the difference between the values of the manometer and defibrillator/monitor is greater than 3 mmHg, contact your
service personnel.

When you select the [Start Accuracy Test] button, it turns to be [Stop Accuracy Test]. Select [Stop Accuracy Test],
accuracy test stops and the button turns to be [Start Accuracy Test] again.

22-10
22.3.9.4 NIBP Leakage Test

The NIBP leakage test checks the integrity of the system and of the valve. It is required at least once every two years or
whenever you doubt the NIBP reading.

Tools required:

„ An adult cuff

„ An air tubing

„ A correct sized cylinder

Follow this procedure to perform the leakage test:

1. Set the patient category to [Adu].

2. Connect the cuff to the equipment’s NIBP connector.

3. Wrap the cuff around the cylinder as shown below.

Defibrillator/monitor Cylinder

Connector for NIBP Tubing Cuff


cuff

4. In the Installation Mode Main menu, select [Maintain NIBP]→ [Start Leakage Test].

After about 20 seconds, the equipment automatically deflates. This means the leakage test is started.

When the accuracy test is completed, the result will be displayed. If the message “NIBP Pneumatic Leak” is displayed, it
indicates that the NIBP airway may have leakages. Check the tubing and connections for leakages, and then perform a
leakage test again.
If the problem persists, contact your service personnel.

When you select the [Start Leakage Test] button, it turns to be [Stop Leakage Test]. Select [Stop Leakage Test],
leakage test stops and the button turns to be [Start Leakage Test] again.

22.3.9.5 Checking Version Information

To view version information, select [Version] in the Installation Mode Main menu. In the popup menu, you can view
system software version and module software version.

22-11
22.3.9.6 Formatting Storage Card

You can format the storage card if data in the card is useless, or if the card has a failure. To format the storage card, select
[Format Data Card]→ [Format] through the Installation Mode Main menu. Then a dialog box pops up as shown below:

If storage card is formatted successfully, a prompt “Formatting is completed!” appears. If there is a failure, the system
stops formatting and present a prompt “Formatting fails!”. Contact the service personnel if formatting fails for 3 times”.

22.3.10 NIBP Overpressure Protection Test


We recommend this test to be carried out once every year. The guidelines for completing this test are as follows:

1. Open the equipment’s enclosure, remove the multi-parameter monitoring module (M51A module), disconnect the
air tube from the pressure measuring sensor (component U26 on the digital board), and then block the tube.

2. Connect the NIBP cuff.

3. Press the NIBP hard key on the equipment’s front panel to start NIBP measurement. When the pressure exceeds the
hardware overpressure protection value (300 to 330mmHg), verify that the valve opens to release the air and the
alarm “NIBP Cuff Overpress” is shown on the screen.

The test is passed if you hear the valve discharge the gas properly and the “NIBP Cuff Overpress” alarm is given.
Otherwise, contact the service personnel.

22.3.11 Electrical Safety Tests


For details about Electrical Safety Tests, refer to Appendix E Electrical Safety Inspection.

22-12
23 Accessories
WARNING
z Use accessories specified in this chapter. Using other accessories may cause damage to the equipment or
not meet the claimed specifications.

z Single-use accessories are not designed to be reused. Reuse may cause a risk of contamination and affect
the measurement accuracy.

z Check the accessories and their packages for any sign of damage. Do not use them if any damage is
detected.

z At the end of its service life, the equipment, as well as its accessories, must be disposed of in compliance
with the guidelines regulating the disposal of such products to avoid contaminating the environment.

z When using the accessories, consider the accessories’ operating temperature. Refer to corresponding
accessory’s instrution for use for details.

z The accessory material that contact the patients have undertaken the bio-compatibility test and is verified
to be in compliance with ISO 10993-1.

23.1 ECG Accessories


ECG Electrodes

Model Specification Applicable patient PN


31499224 10 pcs/pack Adult 0010-10-12304
2245 50 pcs/pack Pediatric 9000-10-07469
2258-3 3 pcs/pack Neonate 900E-10-04880

12-pin Trunk Cable

Leadwire supported Model Compatible with Type Applicable patient PN


3-lead EV 6202 AHA, IEC Defibrillation-proof Pediatric, neonate 0010-30-42720
3/5-lead EV 6201 AHA, IEC Defibrillation-proof Adult, pediatric 0010-30-42719

23-1
Lead Sets

3-Electrode Lead Sets


Type Compatible with Model Applicable patient PN Remark
EL6302A 0010-30-42725 /
Adult, pediatric
EL6304A 0010-30-42732 Long
IEC
EL6306A Neonate 0010-30-42897 /
EL6308A Pediatric 0010-30-42899 /
Clip
EL6301A 0010-30-42726 /
Adult, pediatric
EL6303A 0010-30-42731 Long
AHA
EL6305A Neonate 0010-30-42896 /
EL6307A Pediatric 0010-30-42898 /
EL6302B Adult, pediatric 0010-30-42733 /
IEC
EL6308B Pediatric 0010-30-42901 /
Snap
EL6301B Adult, pediatric 0010-30-42734 /
AHA
EL6307B Pediatric 0010-30-42900 /

5-Electrode Lead Sets


Type Compatible with Model Applicable patient PN Remark
IEC EL6502A 0010-30-42728 /
IEC EL6504A 0010-30-42730 Long
Clip
AHA EL6501A 0010-30-42727 /
Adult, pediatric
AHA EL6503A 0010-30-42729 Long
IEC EL6502B 0010-30-42736 /
Snap
AHA EL6501B 0010-30-42735 /

Adapting Cable

Description Compatible with Applicable patient PN


12-pin to 6 pin connector AHA, IEC Adult, pediatric, neonate 0010-30-43054

23-2
23.2 SpO2 Accessories
Extension Cables

Module type Applicable patient PN Remark


Mindray SpO2 module 0010-20-42710 /
040-000332-00 8 pins, purple connector
Masimo SpO2 module Adult, pediatric, neonate
0010-30-42738 7 pins, white connector
Nellcor SpO2 module 0010-20-42712 /

SpO2 Sensors

Mindray SpO2 module


Type Model Applicable patient PN
MAX-A Adult (>30 kg) 0010-10-12202
MAX-P Pediatric (10 to 50 kg) 0010-10-12203
Disposable
MAX-I Infant (3 to 20 kg) 0010-10-12204
MAX-N Neonate (<3 kg), Adult (>40 kg) 0010-10-12205
520A Adult 520A-30-64101
520P Pediatric 520P-30-64201
Single patient use
520I Infant 520I-30-64301
520N Neonate 520N-30-64401
DS-100A Adult 9000-10-05161
OXI-P/I Pediatric, infant 9000-10-07308
OXI-A/N Adult, neonate 9000-10-07336
518B Adult, pediatric, neonate (Multi-sites) 518B-30-72107
Reusable
512E 512E-30-90390
Adult (Finger type)
512F 512F-30-28263
512G 512G-30-90607
Pediatric (Finger type)
512H 512H-30-79061

23-3
Masimo SpO2 Module
Type Model Applicable patient PN Remark
FPS-1901 Pediatric, neonate (wrap type) 0010-10-42626 LNCS-NeoPt-L
FPS-1862 Neonate (wrap type) 0010-10-42627 LNCS-Neo-L
Disposable FPS-1861 Infant (wrap type) 0010-10-42628 LNCS-Inf-L
FPS-1860 Pediatric (wrap type) 0010-10-42629 LNCS-Pdt
FPS-1859 Adult (wrap type) 0010-10-42630 LNCS-Adt
FPS-1863 Adult (finger clip) 0010-10-42600 LNCS DC-I
Reusable FPS-1864 Pediatric (finger clip) 0010-10-42634 LNCS-DCIP
2258 Adult, pediatric, neonate 0010-10-43016 LNCS YI

Nellcor SpO2 Module


Type Model Applicable patient PN
MAX-A Adult (>30 kg) 0010-10-12202
MAX-P Pediatric (10 to 50 kg) 0010-10-12203
Disposable
MAX-I Infant (3 to 20 kg) 0010-10-12204
MAX-N Neonate (<3 kg), Adult (>40 kg) 0010-10-12205
DS-100A Adult 9000-10-05161
Reusable OXI-P/I Pediatric, infant 9000-10-07308
OXI-A/N Adult, neonate 9000-10-07336

Wavelength of Mindray 518B, 512E, 512F, 512G and 512H SpO2 sensors: red light 660 nm, infrared light 905 nm.
Wavelength of Masimo SpO2 sensors: red light: 660 nm, infrared light: 940 nm.
Wavelength of Nellcor SpO2 sensors: red light: 660 nm, infrared light: 890 nm.
The maximum photic output consumption of the sensor is less than 18 mW.
The information about the wavelength range and maximum photic output consumption can be especially useful to
clinicians, for example, clinicians performing photodynamic therapy.

23-4
23.3 NIBP Accessories
Tubing

Type Applicable patient PN

Adult, pediatric 6200-30-09688


Reusable
Neonate 6200-30-11560

Cuff

Applicable Limb Circumference Bladder


Type Model Applied site PN
patient (cm) Width (cm)
CM1201 Infant 10 to 19 9.2 0010-30-12157
Reusable

CM1202 Pediatric 18 to 26 12.2 0010-30-12158


Upper arm
CM1203 Adult 24 to 35 15.1 0010-30-12159
CM1204 Large adult 33 to 47 18.3 0010-30-12160
CM1205 Adult Thigh 46 to 66 22.5 0010-30-12161
CM1500A 3.1 to 5.7 2.2 001B-30-70692
Single patient

CM1500B 4.3 to 8 2.9 001B-30-70693


Neonate
CM1500C 5.8 to 10.9 3.8 001B-30-70694
CM1500D 7.1 to 13.1 4.8 001B-30-70695
Upper arm
CM1501 Infant 10 to 19 7.2 001B-30-70697
CM1502 Pediatric 18 to 26 9.8 001B-30-70698
CM1503 Adult 25 to 35 13.1 001B-30-70699
CM1504 Large adult 33 to 47 16.5 001B-30-70700
CM1505 Adult Thigh 46 to 66 20.5 001B-30-70701

23.4 Therapy Accessories


Description Model Applicable patient Remark PN
External paddles MR6601 Adult, pediatric Reusable 0651-30-77001
Multifunction MR60 Adult 0651-30-77007
Disposable (5 sets/pack)
electrode pads MR61 Pediatric 0651-30-77008
Pads cable MR6702 / Reusable 040-000545-00
Conductive gel 15-25 / Consumable 0000-10-10775

23-5
23.5 Miscellaneous
Description Model PN

Rechargeable lithium ion battery LI24I001A 022-000047-00

Test load MR6901 0651-20-77032


Test load MR6905 040-000413-00
Synchronous defibrillation input cable / 009-001404-00
Grounding cable UL1015/14AWG 1000-21-00122

DC/AC adapter / 0010-30-12471


Patient data management software kit / 0651-30-77145
Carrying case and shield cover / 115-018610-00
D3 back pouch / 115-008708-00
Conducting gel mount kit / 115-007857-00
Pothook kit / 115-007587-00
Charger Station kit (with international power cord) 115-009187-00
Charger Station kit (with US power cord) 115-009188-00
Charger Station kit (with Indian power cord) 115-009189-00
BatteryFeed 20
Charger Station kit (with EU power cord) 115-009190-00
Charger Station kit (with Brazilian power cord) 115-009191-00
Charger Station kit (with UK power cord) 115-009192-00

23-6
A Specifications
A.1 General Specifications
Class I, equipment energized from an external and internal electrical power source.
Type of protection against
If you suspect the integrity of the external protective earthing or the protective earthing
electrical shock
wire, you should run the equipment on internal electrical power supply (battery).

Degree of protection against Type BF defibrillation proof for external defibrillation.

electric shock Type CF defibrillation proof for ECG, SpO2, NIBP and internal defibrillation

Mode of operation Continuous


Degree of protection against
IP4X
harmful ingress of solid
Degree of protection against IPX4 (when running on battery)
harmful ingress of water IPX1 (when running on AC power supply)
Degree of mobility Portable

Size
Width×depth×height,
288×203×275 mm
without external paddles

Weight
Main Unit 4.7 kg
Battery package 0.5 kg
External paddle set 0.8 kg

Display
Type TFT Color LCD
Size 7 inch
Resolution 800×480 pixels
Viewed waveforms Max. 3
Wave viewing time Max. 16s (ECG)

Equipment connectors
USB connector Connects USB flash memory
Multifunctional connector Connects a cable for defibrillator synchronization.
RJ45 connector Connects standard network cable.

A-1
Audio Indicator
Gives alarm tones (45 to 85 dB), key tones, QRS tones;
Speaker Supports PITCH TONE and multi-level tone modulation;
Alarm tones comply with IEC60601-1-8.

Multifunctional connector
Standard Meets the requirements of EN60601-1 for short-circuit protection and leakage current
Output impedance Typically 50Ω
Synchronous input
Input signal range 0 to 5V
Input impedance ≥10 kΩ
Pulse width >5 ms

A.2 Defibrillator Specifications


Defibrillation mode Manual defib, synchronous cardioversion, AED
Biphasic truncated exponential (BTE) waveform, auto-compensation according to patient
Defibrillation waveform
impedance
Defibrillation electrodes External paddle set coming with pediatric paddles included, multifunction electrode pads.
Controls and indicators on
Charge button, Shock buttons, Energy Select buttons and charge done indicator
external paddles

Range of selected energy


External defibrillation 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 100, 150, 170, 200, 300, 360 J

Patient impedance range


External defibrillation 20 to 200 Ω

A-2
360 J defibrillation waveform into impedance of 25Ω, 50Ω, 75Ω, 100Ω, 125Ω, 150Ω, 175Ω
Voltage (V)

Time (MS)

Selected energy accuracy


Impedance
25Ω 50Ω 75Ω 100Ω 125Ω 150Ω 175Ω Accuracy
Energy
1J 1 1 1 0.9 0.9 0.9 0.8 ±2J
2J 2 2 2 1.9 1.8 1.7 1.6 ±2J
3J 2.9 3 2.9 2.8 2.7 2.6 2.4 ±2J
4J 3.9 4 3.9 3.7 3.6 3.4 3.2 ±2J
5J 4.9 5 4.9 4.7 4.5 4.3 4.1 ±2J
6J 5.8 6 5.8 5.6 5.3 5.1 4.9 ±2J
7J 6.8 7 6.8 6.6 6.3 6 5.7 ±2J
8J 7.8 8 7.8 7.4 7.1 6.8 6.5 ±2J
9J 8.8 9 8.8 8.4 8 7.7 7.3 ±2J
10 J 9.7 10 9.7 9.3 8.9 8.5 8.1 ±2J
15 J 15 15 15 14 13 13 12 ±15%
20 J 20 20 20 19 18 17 16 ±15%
30 J 29 30 29 28 27 25 24 ±15%
50 J 49 50 49 47 45 43 41 ±15%
70 J 68 70 68 65 62 60 57 ±15%
100 J 97 100 97 93 89 85 81 ±15%
150 J 146 150 146 140 134 128 122 ±15%
170 J 166 170 166 159 151 145 138 ±15%
200 J 195 200 195 187 178 170 163 ±15%
300 J 292 300 292 280 267 255 244 ±15%
360 J 351 360 350 336 321 306 293 ±15%

A-3
Charge time (Note: at 20 °C of ambient temperature)
Manual Defib AED
From initiation of
From initial power From initial power on to
Charge time rhythm analysis to
on to charge done charge done
charge done
200J 360J 200J 360J 200J 360J 200J 360J
With a new, fully
<5 s <8 s <11 s <14 s <16 s <21 s <21 s <26 s
charged battery
With a new, fully
charged battery,
<6 s <9 s <12 s <15 s <17 s <22 s <23 s <27 s
depleted by 15 360 J
discharges

With 90% to 100% rated


<5 s <8 s <11 s <14 s <16 s <19 s <22 s <24 s
mains voltage

AC mains: 100 to 240 VAC (±10%)

Synchronized discharge delay


Local synchronized discharge delay < 60ms (from the peak of R-wave)
Remote synchronized discharge delay < 25ms (from the rise edge of synchronous signal)

AED
Energy level: 100 to 360J, configurable;
Shock series Shocks: 1, 2, 3, configurable;
Meeting AHA guidelines 2010 by default.
Shockable rhythm VF, VT (HR>150bpm and QRS width>120ms)

AED ECG Analysis Performance

Rhythm Class Performance requirement Remark


Shockable rhythm Sensitivity > 90% Meets IEC 60601-2-4 and AAMI DF80
Ventricular fibrillation requirement and AHA recommendation
Shockable rhythm Sensitivity > 75% Meets IEC 60601-2-4 and AAMI DF80
Ventricular tachycardia requirement and AHA recommendation
Non-shockable rhythm Specificity > 99% Meets IEC 60601-2-4 and AAMI DF80
Normal sinus rhythm requirement and AHA recommendation
Non-shockable rhythm Specificity > 95% Meets IEC 60601-2-4 and AAMI DF80
Asystole requirement and AHA recommendation
Non-shockable rhythm Specificity > 95% Meets IEC 60601-2-4 and AAMI DF80
All other non-shockable rhythms requirement and AHA recommendation

A-4
A.3 Pacer Specifications
Pacing mode Demand, fixed
Monophasic square wave pulse
Output waveform pulse width 20 ms
Accuracy: ±5%
40ppm to 170ppm
Pacing rate Accuracy: ±1.5%
Resolution: 5 ppm
0mA to 200mA,
Pacing output Accuracy: ±5% or ±5mA, whichever is greater
Resolution: 5mA
200 to 300 ms (depending on pacing rate)
Refractory period
Accuracy: ±3%
4:1 pacing Pacing pulse frequency reduced by factor of 4 when this function is activated.
Output protection The equipment has no sign of damage after defibrillation-proof test.

A.4 Monitor Specifications


ECG

Patient connection 3-lead ECG cable, 5-lead ECG cable, paddles or multifunction electrode pads

Defibrillation electrodes: pads/paddles


ECG inputs 3-lead ECG set: I, II, III
5-lead ECG set: I, II, III, aVR, aVL, aVF, V,
2.5 mm/mV (×0.25), 5 mm/mV (×0.5), 10 mm/mV (×1), 20 mm/mV (×2),
Gain
40mm/mV (×4)
Paper speed 6.25mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Diagnostic mode: 0.05 to 150 Hz
Bandwidth
Monitor mode: 0.5 to 40 Hz
(-3dB, ECG lead set)
Therapy mode: 1 to 20 Hz
Bandwidth (-3dB, defibrillation
Therapy mode: 1 to 20 Hz
electrodes)
Diagnostic mode: >90 dB
Common mode rejection
Monitor mode: >105 dB
(ECG lead set)
Therapy mode: >105 dB
Common mode rejection
Therapy mode: >90 dB
(defibrillation electrodes)
50/60Hz,
Notch filter In Monitor and Therapy mode: notch filter turns on automatically
In Diagnostic mode: notch filter is turned on manually
ECG signal range ±8mV

Electrode offset potential tolerance ±500mV

A-5
Neonate 15 to 350 bpm
HR measurement range Pediatric 15 to 350 bpm
Adult 15 to 300 bpm
HR accuracy ±1% or ±1bpm, which ever is greater
HR resolution 1 bpm
Measuring electrode: <0.1 μA
Lead-off detection current
Drive electrode: <1 μA
Baseline recovery time <5 s (after defibrillation, in monitor mode and therapy mode)
When the test is performed based on part 4.1.2.1 c) of ANSI/AAMI EC 13-2002,
the heart rate meter will reject all 100 ms QRS complexes with less than 1.2 mV
Tall T-wave rejection capability
of amplitude, and T waves with T-wave interval of 180 ms and those with Q-T
interval of 350 ms.

Meets the requirements of ANSI/AAMI EC13-2002: section 4.1.2.1 e). The heart
rate reading after 20 seconds of stabilization is:

Ventricular bigeminy (3a): 80±1 bpm


Response to irregular rhythm
Slow alternating ventricular bigeminy (3b): 60±1 bpm
Rapid alternating ventricular bigeminy (3c): 120±1 bpm
Bidirectional systoles (3d): 90±2 bpm
Meets requirements of ANSI/AAMI EC13-2002: section 4.1.2.1 f).
Response to change in heart rate From 80 to 120 bpm: less than 11 s;
From 80 to 40 bpm: less than 11 s;
Meets the requirements of ANSI/AAMI EC13-2002: section 4.1.2.1 g).
Waveform
4ah - range: 11 s
4a - range: 11 s
Time to alarm for tachycardia
4ad - range: 11 s
4bh - range: 11 s
4b - range: 11 s
4bd - range: 11 s

In compliance with the requirements in Clause 4.1.2.1 d)of ANSI/AAMI


EC13-2002, the following method is used:
If the last 3 consecutive RR intervals are greater than 1200 ms, the 4 most recent
Heart rate averaging RR intervals are averaged to compute the HR. Otherwise, heart rate is computed
by subtracting the maximum and minimum ones from the most recent 12 RR
intervals and then averaging them.
The HR value displayed on the monitor screen is updated every second.

Asystole, V-Fib/V-Tach, Vtac, Vent. Brady, Extreme Tachy, Extreme Brady,


Arrhythmia Analysis Classifications PVCs/min, PVC, Couplet, VT>2, Bigeminy, Trigeminy, R on T, Tachy, Brady, Missed
Beats, PNP, PNC, Vent. Rhythm, Multif. PVC, Nonsus. Vtac, Pause, Irr. Rhythm

A-6
Pace Pulse
Pace pulses meeting the following conditions are labelled with a PACE marker:
Amplitude: ±2 to ±700 mV
Pace pulse markers
Width: 0.1 to 2 ms
Rise time: 10 to 100 μs
Meets the requirements of ANSI/AAMI EC13-2002: section 4.1.4.1 and 4.1.4.3.
The following pulses will be rejected.
Amplitude: ±2 to ±700 mV
Pace pulse rejection
Width: 0.1 to 2 ms
Rise time: 10 to 100 μs
Minimum input slew rate: 10V/s RTI

Resp
Technique Trans-thoracic impedance
Adult: 0 to 120 rpm
Measurement range
Pediatric, neonate 0 to 150 rpm
7 to 150 rpm: ±2 rpm or ±2%, whichever is greater.
Accuracy
0 to 6 rpm: Not specified
Respiration excitation waveform <300 μA, sinusoid, 62.8 kHz (±10%)
Detectable respiration impedance range 0.3 to 5Ω
Reference impedance range 200 to 2500Ω, using ECG cable with 1 kΩ resistor
Difference input impedance >2.5 MΩ
Apnea alarm time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s

Mindray SpO2 Module


*Measurement accuracy verification: The SpO2 accuracy has been verified in human experiments by comparing with arterial
blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed and about
two-thirds of the measurements are expected to come within the specified accuracy range compared to CO-oximeter
measurements.
Measurement range 0 to 100%
Resolution 1%
70 to 100%: ±2% (measured without motion in adult/pediatric mode)
70 to 100%: ±3% (measured without motion in neonate mode)
Accuracy
70 to 100%: ±3% (measured with motion)
0% to 69%: Not specified
Refreshing rate 1s
7 s (When the sensitivity is set to High)
SpO2 averaging time 9 s (When the sensitivity is set to Medium)
11 s (When the sensitivity is set to Low)
Data update period 7 to 9 s (When the sensitivity is set to High)

A-7
PR
Measurement range 20 to 254 bpm
±3 bpm (measured without motion)
Accuracy
±5 bpm (measured with motion)

Masimo SpO2 Module


Measurement range 1 to 100%
Resolution 1%
70 to 100%: ±2% (measured without motion in adult/pediatric mode)
70 to 100%: ±3% (measured without motion in neonate mode)
Accuracy
70 to 100%: ±3% (measured with motion)
1% to 69%: Not specified
Refreshing rate 1s
SpO2 averaging time 2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s
Pulse amplitude: >0.02%
Low perfusion conditions
Light penetration: >5%
Low perfusion SpO2
±2%
accuracy
PR
Measurement range 25 to 240 bpm
±3 bpm (measured without motion)
Accuracy
±5 bpm (measured with motion)
Low perfusion PR accuracy ±3 bpm

Nellcor SpO2 Module


Measurement range 0 to 100%
Resolution 1%
Sensor Range Accuracy*
MAX-A, MAX-AL, MAX-N 70 to 100% ±2%
MAX-P, MAX-I, MAX-FAST 0% to 69% Not specified
OxiCliq A, OxiCliq N 70 to 100% ±2.5%
Accuracy OxiCliq P, OxiCliq I 0% to 69% Not specified
70 to 100% ±3%
D-YS, DS-100A, OXI-A/N, OXI-P/I
0% to 69% Not specified
70 to 100% ±3.5%
MAX-R, D-YSE, D-YSPD
0% to 69% Not specified

Refreshing rate 1s

PR
Measurement range 20 to 300 bpm
±3 bpm (20 to 250 bpm)
Accuracy
Not specified (251 to 300 bpm)

A-8
NIBP
Standards Meet standards of EN60601-2-30/IEC60601-2-30, EN1060-1, EN1060-3, EN1060-4 and SP10
Technique Oscillometry
Mode of operation Manual, Auto and STAT
Static pressure measurement
0kPa to 40.0kPa (0mmHg to 300mmHg)
range
Static pressure measurement
±0.4kPa (±3mmHg)
accuracy
120s for adult and pediatric patients
Maximum measurement time
90s for neonatal patients
Adult Pediatric Neonate
mmHg 40 to 270 40 to 200 40 to 135
Systolic
KPa 5.3 to 36.0 5.3 to 26.7 5.3 to 18.0
Measurement range mmHg 10 to 210 10 to 150 10 to 100
Diastolic
KPa 1.3 to 28.0 1.3 to 20.0 1.3 to 13.3
mmHg 20 to 230 20 to 165 20 to 110
Mean
KPa 2.7 to 30.7 2.7 to 22.0 2.7 to 14.7
Setting range Default Accuracy

Adult: 80 to 280 mmHg 160 mmHg


Cuff inflation pressure
Pediatric 80 to 210 mmHg 140 mmHg 5 mmHg

Neonate 60 to 140 mmHg 90 mmHg

Adult: 297±3 mmHg


Software overpressure
Pediatric: 240±3 mmHg
protection
Neonate: 147±3 mmHg
Max mean error: ±5 mmHg
Measurement accuracy
Max standard deviation: 8 mmHg

A.5 Power Supply Specifications


Fuse Time-lag, 250V, T3.15A
AC power
Line voltage 100 to 240 VAC (±10%)
Current 1.8 to 0.8 A
Frequency 50 / 60Hz (±3Hz)
DC Power (with an external DC/AC adapter)
Input voltage 12VDC
Power consumption 190 W

A-9
Battery
14.8V/3Ah, smart lithium ion battery, rechargeable and free of maintenance, one battery
Battery type
can be configured
Less than 2 hours to 80% and less than 3 hours to 100% with equipment power off;
Charge time
Less than 3.5 hours to 80% and less than 4.5 hours to 100% with equipment power on.
New fully charged battery Testing condition
Without recording, typical ECG
Monitoring ≥2.5 h
monitoring, LCD brightness set to 1

360J discharges at intervals of one minute,


Run time Defibrillation ≥100 discharges
without recording

50 Ω load impedance,
pacing ≥2 h pacing rate: 80bpm,
pacing output 60mA, without recording
Battery fuel gauge 5 LEDs indicating the current battery charge level
At least 20 minutes of monitoring and six 360J discharges (after the low battery alarm
Shutdown delay
occurs)

Note: The specifications above base on a new battery, and at 20ºC.

A.6 Recorder Specifications


Method High-resolution thermal dot array
Number of waveforms Max. 3
Paper speed 25 mm/s, 50 mm/s
Paper width 50 mm
Grid lines The operator can choose to print grid lines or not
Charge events, shock events, marked events, auto test report, parameter alarms, ARR
Auto record
alarms, if configured on

A.7 Alarm Specifications


Alarm Levels High, medium, low level alarms, complying with IEC60601-1-8
Alarm Categories Physiological alarms, technical alarms; Latched alarms and unlatched alarms.
Alarm lamp Independent alarm LED
Alarm properties of all available parameters can be set simultaneously in the Para. Alarm
Parameter alarm setting
menu
Parameter alarm limits can be automatically adjusted according to currently measured
Auto alarm limits
vital signs

A-10
A.8 Data Management Specifications
Data Storage Internal CF card, 1G Bytes
Marking Events 16 types of events, user customized
Event recording Up to 1000 events for each patient.
Waveform storage Up to 24 hours of consecutive ECG waveform
Voice recording Max. 180 minutes in total; max. 60 minutes for each patient
Tabular Trends Max. 72 h of all measured parameters; resolution:1 min
Data Export Data can be export to a PC through a USB flash memory
Patient archives Up to 100

A.9 Environmental Specifications


Operating environment
0 to 45ºC (at least 60 minutes of working time when the temperature reduces from room
Operating temperature
temperature to – 20ºC)
Operating humidity 10 to 95%, non-condensation
Operating altitude -381m to +4575 m (-1250 ft to 15000 ft, or 106.2kPa to 57kPa)

Storage environment

Storage temperature -30 to 70ºC


Storage humidity 10 to 95%, non-condensation
Storage altitude -381m to +4575 m (-1250 ft to 15000 ft, or 106.2kPa to 57kPa)

Shock

Complies with requirements of 21.102, ISO9919:


Peak acceleration: 1000m/s2 (102g)
Duration: 6ms
Pulse shape: half-sine
Number of shocks: 3 shocks per direction per axis (18 total)

A-11
Vibration

Complies with requirements of 21.102, ISO9919.

Bump
Complies with the requirements of 6.3.4.2, EN1789.
Peak acceleration: 15g
Duration: 6ms
Number of impacts: 1000
Impact direction: vertical impacts are applied when the equipment under test is placed at normal operating position.

Free fall

Complies with the requirements of 6.3.4.3, EN1789.


Drop height: 0.75 m
Number of drops: once for each of the six surfaces

A-12
B EMC
The equipment meets the requirements of IEC 60601-1-2.

NOTE
z Use of accessories, transducers, and cables other than those specified may result in increased emission
and/or decreased electromagnetic immunity of the defibrillator/monitor.

z The equipment or its components should not be used adjacent to or stacked with other devices. If adjacent
or stacked use is necessary, the equipment should be observed to verify normal operation in the
configuration in which it will be used.

z The equipment needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided below.

z Other devices may affect this monitor even though they meet the requirements of CISPR.

z When the inputted signal is below the minimum amplitude provided in technical specifications, erroneous
measurements could result.

z Portable and mobile RF communications equipment may affect this equipment.

Guidance and Declaration - Electromagnetic Emissions


The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the
equipment should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions Group 1 The equipment uses RF energy only for its internal function.
CISPR 11 Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions Class B Meet the requirements of Class B
CISPR 11
Harmonic emissions Class A The equipment is suitable for use in all establishments,
IEC 60601-1-2 including domestic establishments and those directly
EN 61000-3-2 connected to the public low-voltage power supply network
Voltage fluctuations/flicker emissions Complies that supplies buildings used for domestic purposes.
IEC 60601-1-2
EN 61000-3-3

B-1
Guidance and Declaration - Electromagnetic Immunity
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the
equipment should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment -
guidance
Electrostatic discharge ±6 kV contact ±6 kV contact Floors should be wood, concrete or
(ESD) ±8 kV air ±8 kV air ceramic tile. If floors are covered
IEC 61000-4-2 with synthetic material, the relative
humidity should be at least 30%.
Electrical fast ±2 kV for power supply lines ±2 kV for power supply lines Mains power quality should be that
transient/burst ±1 kV for input/output lines ±1 kV for input/output lines of a typical commercial or hospital
(EFT) (length greater than 3 m) (length greater than 3 m) environment.
IEC 61000-4-4
Surge ±1 kV differential mode ±1 kV differential mode
IEC 61000-4-5 ±2 kV common mode ±2 kV common mode
Voltage dips, short <5% UT (>95% dip in UT) for <5% UT (>95% dip in UT) Mains power quality should be that
interruptions and 0.5 cycle for 0.5 cycle of a typical commercial or hospital
voltage variations on environment. If the user of our
power supply input 40% UT (60% dip in UT) for 5 40% UT (60% dip in UT) for product requires continued
lines cycle 5 cycle operation during power mains
IEC 61000-4-11 interruptions, it is recommended
70% UT (30% dip in UT) for 70% UT (30% dip in UT) for that our product be powered from
25 cycle 25 cycle an uninterruptible power supply or
a battery.
<5% UT (>95% dip in UT) for <5% UT (>95% dip in UT)
5 seconds for 5 seconds
Power frequency 3 A/m 3 A/m Power frequency magnetic fields
(50/60 Hz) magnetic should be at levels characteristic of
field a typical location in a typical
IEC 61000-4-8 commercial or hospital
environment.
Note: UT is the A.C. mains voltage prior to application of the test level.

B-2
Guidance and Declaration - Electromagnetic Immunity
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the
equipment should assure that it is used in such an environment.

IEC 60601
Immunity test Test level Compliance level Electromagnetic environment - guidance
Conduced RF 3 Vrms 3 Vrms (V1) Portable and mobile RF communications equipment should
IEC 61000-4-6 150k to 80 MHz be used no closer to any part of the device, including cables,
Outside ISM than the recommended separation distance calculated from
bandsa the equation applicable to the frequency of the transmitter.
Recommended separation distance:
⎡ 3 .5 ⎤
d = ⎢ P
10 Vrms 10 Vrms (V2) ⎣ V 1 ⎥⎦
150kHz to ⎡ 12 ⎤
d = ⎢ P
80MHz ⎣ V 2 ⎥⎦
⎡ 12 ⎤
in ISM bandsa d = ⎢ P
(for life support ⎣ E 1 ⎥⎦ 80 MHz to 800 MHz
⎡ 23 ⎤
devices) d = ⎢ P
⎣ E 1 ⎥⎦ 800 MHz to 2.5 GHz
Radiated RF 10V/m 10 V/m (E1)
IEC 61000-4-3 80 MHz to 2.5 where P is the maximum output power rating of the

GHz transmitter in watts (W) according to the transmitter

(for life support manufacturer and d is the recommended separation distance

devices) in meters (m)b.


Field strengths from fixed RF transmitters, as determined by
20V/m 20 V/m (E2)
an electromagnetic site surveyc, should be less than the
80 MHz to 2.5
compliance level in each frequency range d
GHz
Interference may occur in the vicinity of equipment marked
(ISO 9919)

with the following symbol: .


Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.

B-3
a The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to
13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
b Compliance level in the ISM frequency bands between 150 kHz to 80 MHz and in the frequency range 80 MHz to 2.5 GHz
are intended to decrease the likelihood that portable/ mobile communication equipment could cause interference if it is
inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the recommended
separation distance for transmitters in these frequency ranges.
c Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If
the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above,
the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the device.
d Over the frequency ranges 150 kHz to 80 MHz beyond ISM band, field strengths should be less than 3V/m; within ISM band,
field strengths should be less than 10V/m.
Recommended Separation Distances between Portable and Mobile RF, Communications Equipment and This
Equipment
The equipment is intended for use in an electromagnetic environment in which radiated RF disturbance are controlled. The
customer or the user of the equipment can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the equipment as recommended below,
according to the maximum output power of the communication equipment.

Rated Maximum Separation Distance According to Frequency of Transmitter (m)


Output power of 150 kHz to 80 MHz 150 kHz to 80 MHz in 80 MHz to 800 MHz 800 MHz to 2.5 GHz
Transmitter Watts outside ISM bands ISM bands
(W) ⎡ 3 .5 ⎤ ⎡ 12 ⎤ ⎡ 12 ⎤ ⎡ 23 ⎤
d =⎢ P d =⎢ P d =⎢ ⎥ P d =⎢ ⎥ P
⎣ V 1 ⎥⎦ ⎣ V 2 ⎥⎦ ⎣ E1 ⎦ ⎣ E1 ⎦
0.01 0.12 0.12 0.12 0.23
0.1 0.37 0.38 0.38 0.73
1 1.20 1.20 1.20 2.30
10 3.70 3.80 3.80 7.30
100 12.00 12.00 12.00 23.00
For transmitters at a maximum output power not listed above, the recommended separation distanced in meters (m) can be
determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553
MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the ISM
frequency bands between 150 kHz to 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease the likelihood that
portable/ mobile communication equipment could cause interference if it is inadvertently brought into patient areas.
Note 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.

B-4
C BeneHeart Defibrillator Shift Checklist
Inspect the defibrillator/monitor at the change of every shift. Place a “√” in the “Pass/Fail” box as you check the item , or
place a “-” if not applicable. Describe the problem if there is any abnormity.

Equipment Name: Serial Number: Department:

Item Requirements Pass/Fail Description of Abnormity

Equipment appearance Clean, no foreign substance, no crack


Batteries At least one battery installed, battery
indicator on, and more than 60% of
battery capacity left
AC power supply AC power supply connected, green AC
indicator on
Recorder paper Present and sufficient
Cables/connectors Cables not frayed, connectors and pins not
broken or loose
ECG cable set and electrodes Present. Cables not frayed, connectors and
(skip if not in use) pins not broken or loose
External paddles Cables not frayed, connectors and pins not
(skip if not in use) broken or loose ,Properly placed in paddle
tray and easily removed
Pads cable and pads Present, pads cable not damaged, pads
(skip if not in use) not expired. Cables not frayed, connectors
and pins not broken or loose
Pads cable shock test* Turn the mode select knob to “Manual
(skip if not in use) Defib”, connect the pads cable and test
load, charge to 200J, and press the “Shock”
button. Then the system prompts that
shock delivered normally. Disconnect the
test load after the test is finished.
Paddles shock test** Turn the mode select knob to “Manual
(skip if not in use) Defib”, connect the paddle cable and place
the paddles onto the paddle trays, charge
to 10J, and press the “Shock” button. Then
the system prompts that shock delivered
normally.
Service indicator Off

Checked by: Date:

C-1
*: Perform this test only when automatic selftest is not performed using pads cable or when selftest fails.
**: Perform this test only when automatic selftest is not performed using paddles or when selftest fails.

NOTE

z Remember to disconnect the test load when the test is finished. Otherwise, delay could happen in patient
treatment.

C-2
D Alarm Messages
This chapter lists only the most important physiological and technical alarm messages. Some messages appearing on
your equipment may not be included.

In this chapter:

„ The “I” column indicates how indications of technological alarms are cleared after the hardkey or [Audio Pause]

softkey is pressed: “A” means all alarm indications are cleared; “B” indicates alarm light and alarm tones are cleared
and the alarm messages change to prompt messages; and “C” indicates only alarm tone is disabled, but alarm light
and alarm message remain presented.

„ The “L” column indicates the alarm level: “H” refers to high, “M” refers to medium, and “L” refers to low. “*” means
the alarm level is user-adjustable.

„ XX represents a measurement or parameter label, such as ECG, HR, PVCs, RR, SpO2, PR, NIBP, etc.

In the “Cause and solution” column, corresponding solutions are given instructing you to troubleshoot problems. If the
problem persists, contact your service personnel.

D.1 Physiological Alarm Messages


Measurement Alarm Message L Cause and solution
XX XX Too High M* XX value has risen above the high alarm limit or fallen below the low alarm
XX Too Low M* limit. Check the patient’s condition and check if the patient category and
alarm limit settings are correct.
ECG Asystole H Arrhythmia has occurred to the patient. Check the patient’s condition and
V-Fib/ V-Tach H the ECG connections.
Vent. Brady H
Extreme Tachy H
Extreme Brady H
Tachy M*
Brady M*
PVC M*
R on T M*
VT>2 M*
Couplet M*
Bigeminy M*
Trigeminy M*
Missed Beats M*
Vent. Rhythm M*
Multif. PVC M*

D-1
Measurement Alarm Message L Cause and solution
Nonsus. Vtac M*
Pause M*
Irr. Rhythm M*
PNP M* The pacer appears abnormal. Check the pacer.
PNC M*
Resp Resp Apnea H The respiration signal was so weak that the equipment cannot perform
respiration analysis. Check the patient’s condition and the Resp connections.
SpO2 SpO2 Desat H The SpO2 value has fallen below the desaturation alarm limit. Check the
patient’s condition and check if the alarm limit settings are correct.
No Pulse L The pulse signal was so weak that the equipment cannot perform pulse
analysis. Check the patient’s condition, SpO2 sensor and measurement site.

D.2 Technical Alarm Messages


Measurement Alarm Message L I Cause and solution
XX XX SelfTest Err H C An error occurred to the XX module, or there is a problem
XX Init Err H C with the communications between the module and the host.
XX Comm Err L C Restart the equipment.
XX Comm Stop H C
XX Overrange L C The measured XX value is not within the specified range for
XX measurement. Contact your service personnel.
ECG ECG Lead Off L* B The ECG electrode has become detached from the patient or
ECG YY Lead Off L* B the lead wire has become disconnected from the trunk cable.
(YY represents the Check the connection of the electrodes and leadwires.
leadwires V, LL, LA, and RA,
as per AHA standard, or C, F,
L and R as per IEC standard.)
Pads/Paddles off L* B The pads/paddles have been detached from the patient or
the therapy cable is loose. Check that the pads/paddles and
therapy cable are properly connected.
ECG Noise L A The ECG signal is noisy. Check for any possible sources of
signal noise form the area around the cable and electrode,
and check the patient for excessive motion.
ECG Signal Invalid L A ECG amplitude is so low that ECG signal is undetectable.
Check for any possible source of interference from the area
around the cable and electrode; check the patient’s condition.
SpO2 SpO2 Sensor Off L* B The SpO2 sensor has become detached from the patient or
SpO2 Sensor Fault L C the module, or there is a fault with the SpO2 sensor, or an
SpO2 No Sensor L B unspecified SpO2 sensor has been used. Check the sensor
SpO2 Unknow Sensor L C application site and the sensor type, and make sure the sensor
SpO2 Sensor Incompatible L C is not damaged. Reconnect the sensor or use a new sensor.

D-2
Measurement Alarm Message L I Cause and solution
SpO2 Too Much Light L C There is too much light on the SpO2 sensor. Move the sensor
to a place with lower level of ambient light or cover the
sensor to minimize the ambient light.
SpO2 Low Signal L C The SpO2 signal is too low or too weak. Check the patient’s
SpO2 Weak Signal L C condition and change the sensor application site. If the error
SpO2 Weak Pulse L C persists, replace the sensor.
SpO2 Low Perf L B
SpO2 Interference L C The SpO2 signal has been interfered. Check for any possible
sources of signal noise form the area around the sensor, and
check the patient for excessive motion.
SpO2 Non-Pulsatile L C
SpO2 Board Fault L C There is a problem with the SpO2 measurement board. Do not
use the module and contact your service personnel.
NIBP NIBP Loose Cuff L A The NIBP cuff is not properly connected, or there is a leak in
NIBP Air Leak L A the airway.
NIBP Pneumatic Leak L A Check the NIBP cuff and pump for leakages.
NIBP Cuff Type Wrong L A The cuff type applied mismatches the patient category. Verify
the patient category and replace the cuff.
NIBP Air Press. Err L A An error occurred to the air pressure. Verify that the
equipment application site meets the environmental
requirements and check if there is any source that affects the
air pressure.
NIBP Weak Signal L A The patient’s pulse is weak or the cuff is loose. Check the
patient’s condition and change the cuff application site. If the
problem persists, change the cuff.
NIBP Sig. Saturated L A The NIBP signal is saturated due to excess motion or other
sources.
NIBP Overrange L A The patient’s NIBP value may be beyond the specified
measurement range.
NIBP Excessive Motion L A Check the patient’s condition and reduce the patient motion.
NIBP Cuff Overpress L A The NIBP airway may be occluded. Check the airway and
measure again.
NIBP Equip Err H A An error occurred during NIBP measurement and therefore
NIBP Time Out L A the equipment cannot perform analysis correctly. Check the
NIBP Measure Failed L A patient’s condition and NIBP connections, or replace the cuff.
NIBP Reset For Err L A An illegal reset occurred during NIBP measurement. Check if
the airway is occluded.
Main control No Fan L C Make sure that the fan is connected.
system No Speaker L C Make sure that the speaker is connected.
No Data Card L C Make sure that the CF card is in place or format the CF card.
Power Board Comm Err H C An error occurred to the power board, or there is a problem
with the communications between the power board and the
host. Restart the equipment.

D-3
Measurement Alarm Message L I Cause and solution
Keyboard Comm Err L C An error occurred to the keypad board, or there is a problem
with the communications between the keypad board and the
host. Restart the equipment.
Therapy Module Comm Err S C An error occurred to the therapy module, or there is a
problem with the communications between the therapy
module and the host. Restart the equipment. If the problem
persists, contact your service personnel.
Main Control Selftest Err H C The main control voltage is abnormal. Contact your service
personnel.
RT Clock Need Reset L C Reset system time.
RT Clock Err H C An error occurred to the RTC chip, or the button cell is
depleted. Replace corresponding part.
Data Card Err L C There is a problem with the data card. Check or replace the
data card if necessary.
Power board Power System Selftest Err H C An error occurred to the system power supply. Restart the
Power Board Volt Err L C equipment.
No battery L C No battery is installed. Please install the battery.
Low Battery S C Change battery or connect the equipment to the AC power
source to charge the batteries.
Battery Err H There is a problem with the batteries. Check the batteries for
damage; verify that correct batteries are used. Replace the
batteries if necessary.
Battery Depleted! System S C Connect the equipment to AC mains.
will shut shown imminently.
Connect to AC Mains or
Replace Battery.
Battery Aged L C Replace the battery.
Battery failed charging M C Battery failure or power board hardware failure. Replace the
battery. If the problem persists, contact your service
personnel.
Therapy Therapy Equip selftest Err S C An error occurred during therapy module self test. Restart the
module equipment If the problem persists, contact your service
personnel.
Defib Malfunction S C The defibrillation function fails or both the defibrillation and
pacing functions fail. Restart the equipment and test
defibrillation function. If the problem persists, contact your
service personnel.
Pacer Malfunction! S C The pacing function fails. Restart the equipment and test
pacer function. If the problem persists, contact your service
personnel.
Disarming Failed H C There is a problem with the therapy module disarming circuit.
Contact your service personnel.
Last User Test Failed L C Run a successful user test.
Last Auto Test Failed L C Run a successful user test again.

D-4
Measurement Alarm Message L I Cause and solution
Monitoring Monitor Module Selftest Err H C An error occurred during MPM module power-on self test.
module Contact your service personnel.
Monitor Module Reset Err H C MPM module reset abnormally. In this case, the MPM module
restores to default configuration. You can ignore this
problem.
Monitor Module Voltage Err L C The voltage of MPM module is abnormal. Contact your service
personnel.
Recorder Recorder Init Err L A Restart the equipment.
Recordhead Overheated L A The recorder has been working for a prolonged time. Clear
the recording tasks and resume the recording till the
recorder’s print head cools down.
Recorder Overcurrent L A Re-load the recorder paper.
Others Config Err L A Check if the configuration is correct, or restore the factory
configuration.
Pacer Pads cable Off H C Check that pads cable is properly connected.
Pads Off H C Check that pads are properly connected.
ECG Lead Off H C Check that ECG leadwires are properly connected.
Pacer Stopped Abnormally H C Check paddles. Check that pads well contact with patient’s
skin. Make sure pads are properly applied, and then start
pacing again.

Note: In the “L” column “S” refers to special technological alarm. The special technological alarms cannot be paused or
silenced, and the alarm volume is unchangeable. These alarms stop only when the alarm condition is eliminated.

D-5
FOR YOUR NOTES

D-6
E Electrical Safety Inspection
The following electrical safety tests are recommended as part of a comprehensive preventive maintenance program.
They are a proven means of detecting abnormalities that, if undetected, could prove dangerous to either the patient or
the operator. Additional tests may be required according to local regulations.

All tests can be performed by using commercially available safety analyzer test equipment. These procedures assume
the use of a 601PROXL International Safety Analyzer or equivalent safety analyzer. Other popular testers complying with
IEC 60601-1 used in Europe, such as Fluke, Metron or Gerb, may require modifications to the procedure. Please follow the
instructions of the analyzer manufacturer.

The electrical safety inspection should be periodically performed per years .The safety analyzer also proves to be an
excellent troubleshooting tool to detect abnormalities of line voltage and grounding, as well as total current loads.

E.1 Power Cord Plug


Test Item Acceptance Criteria
The power plug pins No broken or bent pin. No discolored pins.
The power The plug body No physical damage to the plug body.
plug The strain relief No physical damage to the strain relief. No plug warmth for device in use.
The power plug No loose connections.
No physical damage to the cord. No deterioration to the cord.
The power cord For devices with detachable power cords, inspect the connection at the device.
For devices with non-detachable power cords, inspect the strain relief at the device.

E.2 Device Enclosure and Accessories


E.2.1 Visual Inspection
Test Item Acceptance Criteria
No physical damage to the enclosure and accessories.
No physical damage to meters, switches, connectors, etc.
The enclosure and accessories
No residue of fluid spillage (e.g., water, coffee, chemicals, etc.).
No loose or missing parts (e.g., knobs, dials, terminals, etc.).

E.2.2 Contextual Inspection


Test Item Acceptance Criteria
No unusual noises (e.g., a rattle inside the case).
The enclosure and accessories No unusual smells (e.g., burning or smoky smells, particularly from ventilation holes).
No taped notes that may suggest device deficiencies or operator concerns.

E-1
E.3 Device Labelling
Check the labels provided by the manufacturer or the healthcare facilities are present and legible.

„ Main unit label

„ Integrated warning labels

E.4 Protective Earth Resistance


1. Plug the probes of the analyzer into the device’s protective earth terminal and protective earth terminal of the AC
power cord.

2. Test the earth resistance with a current of 25 A.

3. Verify the resistance is less than limits.

LIMITS

For all countries, R = 0.2 Ω Maximum

E.5 Earth Leakage Test


Run an Earth Leakage test on the device being tested before performing any other leakage tests.

The following outlet conditions apply when performing the Earth Leakage test:

„ normal polarity (Normal Condition)

„ reverse polarity (Normal Condition)

„ normal polarity with open neutral (Single Fault Condition)

„ reverse polarity with open neutral (Single Fault Condition)

LIMITS

For UL60601-1,

‹ 300 μA in Normal Condition

‹ 1000 μA in Single Fault Condition

For IEC60601-1,

‹ 500 μA in Normal Condition

‹ 1000 μA in Single Fault Condition

E-2
E.6 Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth. All measurements have a true
RMS only

The following outlet conditions apply when performing the Patient Leakage Current test.

„ normal polarity (Normal Condition)

„ reverse polarity (Normal Condition)

„ normal polarity with open neutral (Single Fault Condition)

„ reverse polarity with open neutral (Single Fault Condition)

„ normal polarity with open earth (Single Fault Condition)

„ reverse polarity with open earth (Single Fault Condition)

LIMITS

For CF applied parts


‹ 10μA in Normal Condition

‹ 50μA in Single Fault Condition

For BF applied parts


‹ 100μA in Normal Condition

‹ 500μA in Single Fault Condition

E.7 Mains on Applied Part Leakage


The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage, through a limiting resistance,
to selected applied part terminals. Current measurements are then taken between the selected applied part and earth.
Measurements are taken with the test voltage (110% of mains) to applied parts in the normal and reverse polarity
conditions

The following outlet conditions apply when performing the Mains on Applied Part test.

„ Normal Polarity

„ Reversed Polarity

LIMITS

„ For CF applied parts: 50 μA

„ For BF applied parts: 5000 μA

E-3
E.8 Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the remaining Applied Part
connector s. All measurements may have a true RMS only response.

The following outlet conditions apply when performing the Patient Auxiliary Current test.

„ normal polarity (Normal Condition)

„ reverse polarity (Normal Condition)

„ normal polarity with open neutral (Single Fault Condition)

„ reverse polarity with open neutral (Single Fault Condition)

„ normal polarity with open earth (Single Fault Condition)

„ reverse polarity with open earth (Single Fault Condition)

LIMITS

For CF applied parts,


‹ 10μA in Normal Condition

‹ 50μA in Single Fault Condition

For BF applied parts,


‹ 100μA in Normal Condition

‹ 500μA in Single Fault Condition

NOTE
z Make sure the safety analyzer is authorized comply with requirement of IEC61010-1.

z Follow the instructions of the analyzer manufacturer.

E-4
F Symbols and Abbreviations
F.1 Units
μA microampere

μV microvolt

A ampere

Ah ampere hour

bpm beat per minute

bps bit per second

ºC centigrade

cc cubic centimeter

cm centimeter

dB decibel

DS dyne second

ºF fahrenheit

g gram

GHz gigahertz

GTT gutta

h hour

Hz hertz

in inch

J Joule

k kilo
kg kilogram

kPa kilopascal

L litre

lb pound

m meter

mAh Milliampere hour

Mb mega byte

mcg microgram

mEq milli-equivalents

mg milligram

min minute

ml milliliter

mm millimeter

F-1
mmHg millimeters of mercury

ms millisecond

mV millivolt

mW milliwatt

MΩ megaohm

nm nanometer

rpm breaths per minute

s second

V volt

VA volt ampere

Ω ohm

W watt

F.2 Symbols
– negative, minus

% percent

/ per; divide; or

+ plus

= equal to

< less than

> greater than

≤ less than or equal to


≥ greater than or equal to

± plus or minus
× multiply

© copyright

F.3 Abbreviations and Acronyms


AaDO2 alveolar-arterial oxygen gradient
AAMI Association for Advancement of Medical Instrumentation
AC alternating current
ACI acceleration index
Adu adult
AG anaesthesia gas
AED Semi-automated external defibrillation
AHA American Heart Association
ANSI American National Standard Institute
Ao aortic pressure

F-2
Art arterial
aVF left foot augmented lead
aVL left arm augmented lead
aVR right arm augmented lead
awRR airway respiratory rate
BAP brachial aterial pressure
BIS bispectral index
BP blood pressure
BPSK binary phase shift keying
BSA body surface area
BT blood temperature
BTPS body temperature and pressure, saturated
C.I. cardiac index
C.O. cardiac output
CaO2 arterial oxygen content
CCO continuous cardiac output
CCU cardiac (coronary) care unit
CE Conformité Européenne
CIS clinical information system
CISPR International Special Committee on Radio Interference
CMOS complementary metal oxide semiconductor
CMS central monitoring system
C.O. cardiac output
CO2 carbon dioxide
COHb carboxyhemoglobin
CP cardiopulmonary
CPR Cardiopulmonary resuscitation
CVP central venous pressure
DC direct current
Defib defibrillation
Des desflurane
Dia diastolic
DPI dot per inch
DVI digital video interface
ECG electrocardiograph
EDV end-diastolic volume
EEC European Economic Community
EEG electroencephalogram
EMC electromagnetic compatibility
EMG electromyograph
EMI electromagnetic interference
Enf enflurane
ESU electrosurgical unit
Et end-tidal
EtCO2 end-tidal carbon dioxide
EtN2O end-tidal nitrous oxide

F-3
EtO ethylene oxide
EtO2 end-tidal oxygen
FAP femoral arterial pressure
FCC Federal Communication Commission
FDA Food and Drug Administration
FEV1.0% first second forced expiratory volume ratio
Fi fraction of inspired
FiCO2 fraction of inspired carbon oxygen
FiN2O fraction of inspired nitrous oxide
FiO2 fraction of inspired oxygen
FPGA field programmable gate array
FV flow-volume
Hal halothane
Hb hemoglobin
Hb-CO carbon mono-xide hemoglobin
HbO2 oxyhemoglobin
HIS hospital information system
HR heart rate
I:E inspiratory-expiratory ratio
IBP invasive brood pressure
ICG impedance cardiography
ICP intracranial pressure
ICT/B intracranial catheter tip pressure transducer
ICU intensive care unit
ID identification
IEC International Electrotechnical Commission
IEEE Institute of Electrical and Electronic Engineers
IP internet protocol
Iso isoflurane
IT injectate temperature
LA left arm
LAP left atrial pressure
Lat lateral
LCD liquid crystal display
LCW left cardiac work
LCWI left cardiac work index
LED light emitting diode
LL left leg
LVDS low voltage differential signal

LVET left ventricular ejection time

LVSW left ventricular stroke work


LVSWI left ventricular stroke work index
MAC minimal alveolar concentration
MAP mean arterial pressure
MDD Medical Device Directive

F-4
MetHb methemoglobin
MRI magnetic resonance imaging
MVe expiratory minute volume
MVi inspiratory minute volume
N/A not applied
Neo neonate
NIBP noninvasive blood pressure
NIP negative inspiratory pressure
O2 oxygen
O2CI oxygen consumption index
O2R oxygen extraction ratio
OR operating room
oxyCRG oxygen cardio-respirogram
PA pulmonary artery
Paw airway pressure
PAWP pulmonary artery wedge pressure
PD photodetector
Ped pediatric
PEEP positive end expiratory pressure
PEF peak expiratory flow
PEP pre-ejection period
PIF peak inspiratory flow
PIP peak inspiratory pressure
Pleth plethysmogram
Pmean mean pressure
PNC pacer not captured
PNP pacer not paced
Pplat plateau pressure
PR pulse rate
PVC premature ventricular complex
PVR pulmonary vascular resistance
PVRI pulmonary vascular resistance index
R right
RA right arm
RAM random access memory
RAP right atrial pressure
Raw airway resistance
Rec record, recording
Resp respiration
RHb reduced hemoglobin
RL right leg
RM respiratory mechanics
RR respiration rate
RSBI rapid shallow breathing index
SaO2 arterial oxygen saturation
SEF spectral edge frequency

F-5
Sev sevoflurane
SFM self-maintenance
SI stroke index
SMR satellite module rack
SpO2 arterial oxygen saturation from pulse oximetry
SQI signal quality index
SR suppression ratio
STR systolic time ratio
SVR systemic vascular resistance
SVRI systemic vascular resistance index
Sync synchronization
Sys systolic pressure
Taxil axillary temperature
TD temperature difference
Temp temperature
TFC thoracic fluid content
TFI thoracic fluid index
TFT thin-film technology
Toral oral temperature
TP total power
Trect rectal temperature
TVe expiratory tidal volume
TVi inspiratory tidal volume
UAP umbilical arterial pressure
UPS uninterruptible power supply
USB universal serial bus
UVP umbilical venous pressure
VAC volts alternating current
VEPT volume of electrically participating tissue
VI velocity index
WOB work of breathing

F-6
G Device Tracking
In order to provide high quality product and perform better service, we are going to track our product. Please contact us
with the device tracking information when you have received your defibrillator/monitor:

Please fill the information in the next page, cut the table and fax it to +86 755 26582934. You can also email your
information to [email protected].

G-1
FOR YOUR NOTES

G-2
Device Tracking Information
User Information

Customer Name

Department name

Address

City State Zip/Post Code Country

Contact Person

Tel No. Fax No.

Email Address

Device Information

Product name Serial number Model Installation Date


P/N: 046-001653-00 (6.0)

You might also like