[ FOR USER’S ]
NOTICE 1
1. Thank you for purchasing DIG SERIES(For 1-Tube), Dongmun High Frequency
Generator for X-ray radiographic systems.
2. This manual is designed to ensure correct use and operation of DIG-Series HFG.
Please read all the lines thoroughly before you use the generator.
4. Incorrect use and operation exceeding described conditions in this manual may
occur damage of the device and shorten its life. Particular attention must be paid to
all the warnings, cautions and notes incorporated herein.
NOTICE 2
1. This equipment should be used only by the legally qualified persons and practi-
tioners
-1-
USER’S MANUAL DONGMUN
6. If the device is broken, mark proper description and ask repairs to the engineer.
-2-
USER’S MANUAL DONGMUN
b. If you operate the device after long period of non-use, check the safety and
conditions of the machine.
9. Operate the machine in correct ways according to the manuals.
10. Be careful about following cautions
a. Wear the lead apron in order to prevent the x-ray irradiation.
b. Regular inspection should be executed about leakage dose and irradiation
meter.
c. Be careful about the x-ray exposures.
d. Do not exceed the maximum weight for the x-ray tables.
1. If you find any troubles about the conditions of patients or machine, stop the
operation and take appropriate measures.
2. For the need of emergency power cut-off, use the power switch on the operation
panel. Do not randomly cut off the power during operation.
3. Be careful about using the exposure and hand switches on the operation panel.
You can make exposures anytime under normal conditions.
2. Do not use the machine during the warning signs of the earthquake. As well, the
security of the machine must be checked after the occurrence of the earthquake.
3. In case any troubles happen to the machine, turn off the power and mark
appropriate sign such as “Do not use this machine” and ask repairs to the
engineers.
4. Regular inspection is required to the device and ask any questions to the sales
division of the Dongmun.
-3-
USER’S MANUAL DONGMUN
-4-
USER’S MANUAL DONGMUN
4. Each cable has enough length and flexibility but do not bend the cable forcibly.
.
・
DANGEROUS
-5-
USER’S MANUAL DONGMUN
The following symbols will be used in this manual: their meanings and applications
are described further below.
-6-
USER’S MANUAL DONGMUN
WAIVER CLAUSE
Poskom do not take responsibility for the following causes for damage and repairs.
3. Any troubles and defects by the use of other parts, which is not designated and
recommended by Dongmun.
6. Any defects and causes occurred by natural calamity such as fire, earthquake,
windfall and thunders.
7. Accident and damage happened by the use beyond the original purpose of the
machine.
8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do
not take responsible for the conditions and results of the medical prescriptions.
-7-
USER’S MANUAL DONGMUN
1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or
discrepancy occurred from normal use and operation of product.
3. Note that defects not related to the main functions of the product are not eligible
for this warranty.
6. Dongmun is not responsible for indirect damage related to the use of products.
7. Dongmun is not responsible for injury or damage after the end of warranty
period.
8. All detailed adaptability warranties and other warranties are subjected to this
statement.
-8-
USER’S MANUAL DONGMUN
SECTION 1.
COMPOSITION/SPECIFICATION
-9-
USER’S MANUAL DONGMUN
1.1. COMPOSITION
EM swich HF
Power AD/DC HV TANK
Supply Delay Convertor DC/AC Inverter
EM swich X-ray
TUBE
IGBT
Driver
Stater
OP
PANEL
HT CONT
Collimator
PWR
FILAMENT
IF
AEC ION
Chamber
- 10 -
USER’S MANUAL DONGMUN
Operation Panel
Main CPU
PWM kV FB PWM
kV, mA FB Amp
Filament
mA FB
Drive
circuit rms/DC CONV
IGBT AEC
FIL. INV.
HV Trans kV
FIL. Trans. mA
A C
FIL.
X-ray TUBE
ION Chamber
- 11 -
USER’S MANUAL DONGMUN
1-2. SPECIFICATION
DIG-325 32kW
DIG-525 40KW
1 Rating Output
DIG-650 50kW
DIG-1150 80KW
- 12 -
USER’S MANUAL DONGMUN
- 13 -
USER’S MANUAL DONGMUN
A. Avoid following places for the normal operation and safe storage.
a. Where the equipment is exposed to water vapor.
b. Where the equipment is exposed to direct sunlight.
c. Where the equipment is exposed to dust.
d. Where the equipment is exposed to high humidity.
e. Where there is a ventilation problem.
f. Where the equipment is exposed to salty atmosphere.
g. Where the equipment is exposed to chemicals or gas.
B. Keep away from the place with strong vibration and maintain proper environment
and conditions.
*Operation Environment
Temperature range 10°C ~ 40°C,
Relative Humidity Range 30% ~ 75% RH
Others No Direct Sunlight
No combustible and corrosive gases
C. For safe storage and transportation, you must keep following range of
temperature, humidity and atmosphere.
- 14 -
USER’S MANUAL DONGMUN
SECTION 2
DISPLAY/FUNTION
- 15 -
USER’S MANUAL DONGMUN
- 16 -
USER’S MANUAL DONGMUN
- 17 -
USER’S MANUAL DONGMUN
400. 0 mm
200. 0 mm
36. 0 mm
42. 0 mm
- 18 -
USER’S MANUAL DONGMUN
22
10 12
16
9 3 4 11 5 1 2 6
23 24
13 14 15 18 19 20 17 21 25 7 8
- 19 -
USER’S MANUAL DONGMUN
kV
mA
decreases mAs/sec. If you keep the button pressed, the value of mAs/sec will
be increased or decreased continuously.
- 20 -
USER’S MANUAL DONGMUN
mA s
sec
It circulates the rotor and controls the heat of the filament that makes ready
actions for the x-ray exposure.
Press the exposure switch after the ready sign is lighted on.
- 21 -
USER’S MANUAL DONGMUN
It operates the selections of patient size. The selected size will be displayed on
the segment. Save the changed APR conditions by pressing the size switches.
*You can set the exposure conditions (kV, mA and mAs/sec) by selecting the
exposure regions, direction and patient size. Each condition can be changed
and stored by the exposure size switch.
If the x-ray exposure is closed early by the AEC backup timer, the AEC mode
LED will blink off and the buzzer sounds. The next exposure is forbidden until
the reset switch is pushed.
-3 0 +3
It selects the density level of the exposure and increases or decreases the
mAs/sec under the APR mode.
- 22 -
USER’S MANUAL DONGMUN
It selects the field on the AEC mode. Each LED represents the physical location
of irradiation detector and the combination of designated field is available.
- 23 -
USER’S MANUAL DONGMUN
SECTION 3
OPERATION/EXPOSURE
- 24 -
USER’S MANUAL DONGMUN
1. Power-on Switch
2. Power-off Switch
b. Please turn off the power supply except the time of use.
- 25 -
USER’S MANUAL DONGMUN
- 26 -
USER’S MANUAL DONGMUN
- 27 -
USER’S MANUAL DONGMUN
- 28 -
USER’S MANUAL DONGMUN
Choose APR mode with the (13) APR selection switch and the LED displaying APR
mode will be lighted on the upper left of the OP panel.
- 29 -
USER’S MANUAL DONGMUN
3) The selection of density in the AEC mode can adjust the level of x-ray exposure
levels.
4) “Ready” action is not operated if the bucky and field are not chosen after the AEC
mode selection.
3-6. RADIOGRAPHY
1. Photography Preparation
After setting up radiography conditions, assemble the cassette equipped with the
film on the x-ray table.
If the X-ray irradiation field lamp switch attached to the variable X-ray beam-
limiting device is pushed, the lamp will be lighted up. X-ray irradiation field is
adjusted with a radiography position, and radiography preparation is ended.
- 30 -
USER’S MANUAL DONGMUN
4) If the X-ray exposure LED is turned off or buzzer does not sound, it means
the radiography work is completed. Then, you can stop pressing the hand
switch.
5) If you touched out and stopped x-ray shooting during exposure, the check
code will be displayed.
- 31 -
USER’S MANUAL DONGMUN
SECTION 4
TROUBLES
- 32 -
USER’S MANUAL DONGMUN
- 33 -
USER’S MANUAL DONGMUN
In case any problems happen during the operation of the device, they will be
indicated on the kV/mA segment.
A. Error Code
The error code is displayed when problems, such as abnormalities, occur to the
equipment. While the error code is displayed, radiography operation is forbidden
for safety. A power supply should be switched off and on again and an error code
will be cleared. If the problem of equipment is not solved at the time of the power-
supply re-switched on, the error code will be displayed again.
Since it is the failure of equipment when an unusual state is not recovered, please
inform the service section of our company or our agent.
B. Check Code
Unlike an error code, a check code is the warning sign when wrong operations are
applied by the user. X-ray radiography operation is forbidden for the safety during
the check code. The way to release a check code is pressing the WR switch for 2
seconds. CH01 will be released if thermal switch is off during the re-supply of the
power.
- 34 -
USER’S MANUAL DONGMUN
Error
Code DESCRIPTION ACTION
Err 01 CHARGE MONITOR ERROR (HT-CONT)
Err 02 OVER TIME ERROR (OP PANEL) AT STAND BY RESUPPLY OF
Err 03 READY-OUT SIGNAL ERROR AT SAND BY THE POWER
Err 04 X-RAY OUT SIGNAL ERROR AT STAND BY
Err 05 READY SWITCH INPUT ERROR AT STAND BY
Err 06 X-RAY SWITCH INPUT ERROR AT STAND BY
Err 07 HAND SWITCH INPUT ERROR AT STAND BY
Err 08 HAND SWITCH X-RAY INPUT ERROR AT STAND BY
Err 09 ROTOR DETECTER SIGNAL AT STANDBY
Err 10 FILLAMENT CURRENT DETECTER ERROR AT STANDBY
Err 11 ‘kV FEEDBACK ERROR AT STANDBY
Err 12 ‘mA FEEDBACK ERROR AT STANDBY
Err 13 ROTOR DETECTER ERROR AT EXPOSURE
Err 14 FILLAMENT CUREENT DETECTER ERROR AT EXPOSURE
Err 15 ‘kV FEEDBACK ERROR DETECTER ERROR AT EXPOSURE
Err 16 ‘mA FEEDBACK DETECTIVE ERROR EXPOSURE
Err 17 NOT DEFINED
Err 18 X-RAY TIME OVER IN SEC MODE
Err 19 X-RAY TIME OVER IN mAs MODE
Err 20 X-RAY TIME OVER IN AEC MODE
Err 21 HV INTERLOCK ERROR
Err 22 IGBT FEEDBACK ERROR
Err 23 ‘kV FEEDBACK OVER DETECTER ERROR AT EXPOSURE
Err 24 ‘mA FEEDBACK OVER DETECTER ERROR AT EXPOSURE
Err 25 NO ZERO-CROSS OR NO INPUT
Err 26 ZERO-CROSS OVER OR WRONG INPUT FREQUENCY
Err 27 BUCKY 1 OK SIGNAL ERROR (HT-CONT)
Err 28 BUCKY 2 OK SIGNAL ERROR (HT-CONT)
Err 29 AEC A/D DATA-IN ERROR (HT-CONT)
Err 30 AEC A/D DATA-IN ERROR (HT-CONT) AT NON-EXPOSURE
Err 31 I/F BOARD CONDITION ERROR
Err 32 AEC BOARD CONDITION ERROR
Err 33 SMALL-FOCUS SELECTION ERROR (HT-CONT)
Err 34 LARGE-FOCUS SELECTION ERROR (HT-CONT)
- 35 -
USER’S MANUAL DONGMUN
* CH01 is will be released if the TUBE terminal switch is off, in case resupply of the
power.
- 36 -
USER’S MANUAL DONGMUN
SECTION 5
MAINTANENCE
- 37 -
USER’S MANUAL DONGMUN
Maintenance has check menus carried out daily by the user and periodic
inspection by technical service engineers who completed special training
programs on the repair of the generator.
User checks items based on " Daily Checking Items by the operator " and should
keep data records.
The following checks and maintenance procedures, together with the suggested
intervals, are the manufacturer’s recommendation for most effective Periodical
Maintenance schedule for this generator.
The first periodic maintenance service should be performed six (6) months after
installation, and the subsequent service every twelve (12) months.
When a trouble and a poor part are discovered, please stop use of equipments,
turn off a power supply, and contact a near service station.
- 38 -
USER’S MANUAL DONGMUN
- 39 -
USER’S MANUAL DONGMUN
Please check clear environment inside the operation room and x-ray room.
The system is designed to operate with normal conditions under following
environment. Please keep following temperature and humidity.
Operation Panel
Please check whether each switch of an operation panel lights up.
When you clean the surface of an operation panel, it should be wiped with the
soft and fine cloth. When using a cleaner, it is made to contain in cloth and is
used.
- 40 -
USER’S MANUAL DONGMUN
- 41 -
USER’S MANUAL DONGMUN
Type/Name :
X-ray Room No.:
Date :
Performed by :
Interval Result
Check item Daily Weekly Good/No Good
Air Conditioner ○
Temperature and Humidity ○
Operation Panel ○
Radiographic Condition ○
Memo:
- 42 -
USER’S MANUAL DONGMUN
S/N: Date:
S/S: Performed by:
- 43 -
USER’S MANUAL DONGMUN
Memo:
- 44 -
USER’S MANUAL DONGMUN
Scope
- 45 -