0Hdvxuhphqw8Qfhuwdlqw/ Lq7Hvwlqj: 7Hfkqlfdo5Hsruw1R - XQH
0Hdvxuhphqw8Qfhuwdlqw/ Lq7Hvwlqj: 7Hfkqlfdo5Hsruw1R - XQH
7HFKQLFDO5HSRUW1R
7HFKQLFDO 5HSRUW
-XQH
0HDVXUHPHQW8QFHUWDLQW\
LQ7HVWLQJ
The document is in the progress state and it is intended, to add further examples
from non-chemical fields of testing. Proposals for such examples are very wel-
come.
We would like express our thanks to Adriaan M H. van der Veen, Christian Ran-
son, Matthias Rößlein, Michele Desenfant, Tomas Quintana and Vitor Ramos for
their very helpful comments. We are grateful to Holger Frenz for Example 2,
Joachim Abshagen and Janusz Morkowski for permission to present Example 3
and Matthias Rößlein for permission to use Example 4.
June 2002
Contents:
5. Conclusions 21
1
Vocabulary of Basic and General Terms in Metrology, ISO, Geneva, 1993, ISBN 92-67-10175-1
2
e.g. ILAC G8 [69]
3
e.g. ISO/IEC 17025 (5.4.6) citing ISO 5725 [29]
4
Precision data may be obtained under repeatability or reproducibility conditions (compare ISO 5725)
Step 5) Calculate the result of the measurement from the input quantities.
Step 2) Determine the input quantities x1,x2,x3,...,xN. Determine the values of the input
quantities, e.g. through measurement, from tables and from specifications.
u ( x j ) = s(q)
the standard deviation will be used directly as standard uncertainty
of the respective single component xj, measured n times
5
the required level of confidence will vary for different fields: while 95% may be sufficient in a technical field of
testing, 99.7% may be required e.g. in the health sector and for legal applications.
u( x j )
the standard uncertainty is derived from prior information
either directly or after consideration of the statistical distribution
Step 5) Calculate the result of the measurement from the input quantities
The result is: y = f(x1,x2,x3,...,xN)
Step 6) Calculate the combined uncertainty from the uncertainties of the single compo-
nents (as determined in 3), taking into account possible covariances.
2
(first order Taylor approximation) for non-correlated n
∂f 2
( y ) = ∑ u ( xi )
2
uncertainty components u combined
i =1 ∂xi
[For reasons of completeness the equation for n n
∂f ∂f
ucombined ( y ) = ∑∑ u ( xi , x j )
2
correlated uncertainty components is also given:]
i =1 j =1 ∂xi ∂x j
∂f
In those equations the partial derivative is called sensitivity coefficient.
∂xi
The equations described above look very complicated to most experimenters.
In practice they become much simpler for some simple mathematical relations (in
case that no correlations exist):
Y = y ± U = y ± k ⋅ u combined ( y )
6
if normal distribution may be assumed
7
Student- or t-distribution, see also GUM [1] G.3
Example 1
Chemical Testing
1.a) Estimation of the uncertainty of the measurement of 100 mg sulfate in waste wa-
ter determined with ion-chromatography from proficiency test results.
th
(Step 3) The laboratory has participated satisfactorily in the 4 all-German waste-water
proficiency testing. The standard deviation of all of the laboratories‘ results was
s = 4%.
(Step 6) s = u = 4 % ⇒ u = 4.0 mg/l for 100 mg/l sulfate
(Step 7) Uexpanded= k· u = 2· 4.0 mg/l = 8.0 mg/l
(Step 8) Sulfate: 100.0 mg/l or: Sulfate: 100.0 ± 8.0 mg/l
measurement uncertainty: 8.0 mg/l (k=2)
This measurement uncertainty was derived from interlaboratory comparison re-
sults. It is expressed as an expanded uncertainty and was obtained by multiplica-
tion of the combined uncertainty with the factor k = 2, relating to a level of confi-
dence of 95%.
Note: Some publications e.g. [35] promote the approach of combining repeatability and repro-
ducibility uncertainties: ucombhined = u repeat 2 + u reproduce 2 , which is too pessimistic. The above method
is to be preferred, and sufficient if the proficiency test covers all relevant uncertainty compo-
nents.
1b) Estimation of the uncertainty of the measurement of 100 mg sulfate in waste water
determined with ion-chromatography from control-charts.
(Step 3) The laboratory maintains an individual control chart for determination of sulfate.
The standard deviation from it is sCC = 3.8% .
While the control sample of 5 mg/l is measured directly, the sample of 100 mg/l
had to be diluted. This dilution step has to be taken into account for the uncer-
tainty budget. The dilution step with a dilutor was carried out and weighed 10
times and delivered a standard deviation sDil = 1%.
(Step 6) ucombined = u CC 2 + u Dil 2 = 3.8 2 + 12 = 3.9 %
ucombined = 3.9 mg/l for 100 mg/l sulfate
(Step 7) Uexpanded= k· ucombined = 2· 3.9 mg/l = 7.8 mg/l
(Step 8) Sulfate: 100.0 mg/l
measurement uncertainty: 7.8 mg/l (k=2)
1c) Estimation of the uncertainty of the measurement of 100 mg sulfate in waste water
determined with ion-chromatography from validation data in the method standard.
(Step 3) The validation data for determination of sulfate in wastewater with ionchroma-
tography are given in ISO 10304-2. In an interlaboratory comparison of an indus-
trial waste water for method validation the reproducibility standard deviation was
6.1%.
(Step 6) s = u = ucombined = 6.1 % ⇒ ucombined = 6.1 mg/l for 100 mg/l sulfate
(Step 7) Uexpanded= k· ucombined = 2· 6.1 mg/l = 12.2 mg/l
(Step 8) Sulfate: 100 mg/l
measurement uncertainty:12 mg/l (k=2)
This measurement uncertainty was derived from method validation data in accordance
to ISO 10304-2. ...
Note: The laboratory must have proved to be able to perform in accordance with this stan-
dard method.
Example 2
Mechanical Testing
Hardness testing according to Brinell
At the moment it is difficult to calculate the measurement uncertainty in the field of mechani-
cal testing. This is because of the lack of knowledge how to use the approaches developed
in various other fields of chemistry or calibration. Nevertheless there are possible way of
estimating the measurement uncertainty to comply with the demands of clients and test stan-
dards. The first area in mechanical testing for which a complete system was evaluated is the
following example which soon should be a part of the test standards.
This examples can be used to calculate the measurement uncertainty in hardness testing.
This is the first approach how to do this calculation. It is expected that this model will be
used in the ISO standards 6506, 6507 and 6508 in the near future. This model was verified
using the data of about 95 laboratories, participating in an European proficiency test.
The measurement uncertainty calculated is based on, hardness testing according to Brinell.
For this test it is measured how a ball can deform a sample plate.
This example was calculated on the basis of a hardness reference block (CRM) 246.8 ± 1.5
HBW 2.5/187.5 according to the test standard EN ISO 6506. The block was certified by the
MPA NRW Dortmund.
4 Standard uncertainty of CRM t * s X CRM t=1.15 (Student-factor) for n=5 1.15 ⋅ 0.77
uCRM uCRM = uCRM = = 0.39
n and 68.3% confidence level 5
5 Mean value and standard devia- 1 n
tion of the measurement on CRM H = ⋅ ∑ Hi Single measurements:
H , n I =1 H i acc. EN ISO 6506, part 2, 246.0–245.0–246.0–246.0-246.0
sH 1 n No. 5.6 H = 245.8 / s H = 0.45
sH = ∑
n − 1 i =1
(H i − H ) 2
15
Sources of uncertainty Abbre- Formula Literature/Certificate Sample calculation
viation
STEP
U = 5.1
10 Overall result steps 7 and 9
X = x ±U 95% X = 286.0 ± 5.1
ALTERNATIVE (in %)
11 Calculation of the relative ex- ~ U ~ 5.1
A panded measurement uncertainty ~ U= ⋅ 100% steps 1 to 9,
U= ⋅ 100% = 2.1%
U 95%
X CRM 246.8
12 Overall result steps 7 and 11A
A
X = x ±U 95% X = 286.0 HBW 2.5 / 187.5 ± 2.1%
16
Example 3
Determination of emission measurement and sampling
uncertainties estimated by well experienced experts
In 1998 an interlaboratory estimation was carried out among 26 institutes from Germany and
Switzerland measuring air polluting emissions according to VDI technical method standards.
The results are given in [86] “Verlauf und Ergebnisse einer geordneten Ringschätzung der
Unsicherheiten von Emissionsmessungen” (Process and results of an well-ordered interlabo-
ratory estimation of the uncertainties from emission measurements) by Janusz S. Morkowski,
Umweltbundesamt Texte 54/99, Berlin, 1999.
Five popular and well known VDI technical methods of emission measurement had been
chosen. Each of them consist of several procedural steps that may influence the final result
of the measurement. The methods chosen were:
1) VDI 2066, Blatt 2: Filterhead with stuffing tube for dust determination (gravimetrically)
2) VDI 2066 Blatt 7: Filterhead with a plane filter for dust determination
3) VDI 2456 Blatt 10: Total nitrogenoxides
4) VDI 3480 Blatt 1: HCl with different analytical techniques
5) VDI 3481 Blatt 2: Total organic carbon by sorption method.
The laboratories received a questionnaire specifying the procedure of the estimation and
were instructed by the organiser.
Uncertainty components for the different steps of the procedures (methods) were to be esti-
mated according to an estimation form. Those steps included planning, preparation, sam-
pling, determination of the gas volume, sample transportation, preparation of the sample,
measurement and analysis in the laboratory, evaluation of the result.
These single uncertainty components of the procedure were combined according to the un-
certainty propagation law of Gauss to the total measurement uncertainty UT.
Estimations were made both for the “standard case”, S-case, which can be expected for
about two thirds of the tests and for the “problematical case” X-case, which should also
cover most problematic cases (e.g. evil conditions, sampling not representative, ...) with a
probability of 95% and more.
75 sampling personnel and 20 employees of the laboratories participated in the estimation test
being completely independent from each other. The following table gives the results of the
estimation test in % (for the S-case). For methods 3 and 4 only the results for the photometri-
cal analysis are given here. The entire tables of the results can be found in [86].
As it can be seen from the table the estimated total uncertainties for the 5 different methods
vary from 13% to 21% with the sampling step being one of the major components.
The total uncertainties are calculated first for all steps of the procedure UT(I) and secondly
also for the case where the planning step is not included in the calculation UT(II). In the latter
case the uncertainties are reduced especially for the filter tests. The relations of X-case to S-
case are approximately 2.7.
For one of the methods, the dust determination with plane filters, also practical interlabora-
tory comparison data were available. Laboratory intercomparisons at the Emission-Simu-
lation-Plant at the HLfU at Kassel from 1994 to 1996 give a mean standard deviation of
15.4% for small concentrations of dust in emissions.
Because planning, which is usual in regular emission measurements is not needed in this
case, the estimated uncertainty data UT(II) of the dust determination with plane filters without
the planning step can be used for comparison. The value estimated in the interlaboratory
estimation is 15.5%, which is nearly equal to the value obtained in the HLfU-Plant.
17
Method standard VDI 2066 VDI 2066 VDI 2456 VDI3480 VDI 3481
Bl.2 Bl.7 Bl. 10 Bl. 1 Bl.2
Measurand Dust Dust NO+NO2 HCl TOC
Analytical technique: gravimetric gravimetric photometric photometric sorption
Planning 11.06 8.27 3.59 3.32 3.26
Preparation of sampling 6.44 6.85 3.56 5.43
Sampling 12.55 11.46 10.11 6.21 15.97
Transport and sample preparation 2.34 2.61 8.15 3.31 5.38
Preparation of sample / treatment of 3.61 8.09 10.06
the filter /sorption material
Determination of the volume 4.33 4.21 4.33 3.87
Additional parameters for volume 3.68 4.32 2.83 3.73
determination
Other influences 0.46 0.67 0.41 0.75
Analytical measurement 8.37 8.11
Other influences 1.12
Total uncertainty UT(I) 19.3 17.6 17.3 12.9 21.3
From these mean values of the
single uncertainty components
Total uncertainty UT(II) 15.8 15.5 16.9 12.5 21.1
as above without the planning step
Relation X-case/S-case 3.04 2.93 2.39 2.64 2.62
Though the single estimation results (which are not given here) are distributed widely, the
results of the estimation (average and the limits of a rather narrow range of confidentiality)
may be a useful approach for the determination of measurement uncertainties in this field.
The results also show that it is important that the estimator knows the testing procedure well,
that he is very experienced in the field concerned and that the details on the conditions of
the measurement as a basis for the estimation are provided.
The estimation of the uncertainties for the sampling step of the different methods provide
examples which may serve as a basis also for other methods.
As it can be seen from the table the estimated total uncertainties for the 5 different methods
vary from 13% to 21% with the sampling step being one of the major components.
The total uncertainties are calculated first for all steps UT(I) and secondly also for the case
where the planning step is not included UT(II). In the latter case the uncertainties are re-
duced especially for the filter tests. The relations of X-case to S-case are approximately 2.7.
For one of the methods, the dust determination with plane filters, also practical interlabora-
tory comparison data were available. Laboratory intercomparisons at the Emission-
Simulation-Plant at the HlfU at Kassel from 1994 to 1996 give a mean standard deviation of
15.4% for small concentrations of dust in emissions.
Because planning is not needed in this case, the estimated uncertainty data UT(II) of the dust
determination with plane filters without the planning step can be used for comparison. The
value estimated in the interlaboratory estimation is 15.5%, which is well comparable to the
practically obtained value.
Though the single estimation results (which are not given here) vary widely, the results show
that estimation may be a useful approach for determination of measurement uncertainties.
The results also show that it is important that the estimator knows the testing procedure well
and is provided details on the conditions as a basis for the estimation.
The estimation of the uncertainties for the sampling step of the different methods provide
examples which may serve as a basis for other methods.
18
Example 4
Determining the measurement uncertainty in preparing a calibra-
tion standard by the mathematical analytical approach
Useful examples for measurement uncertainty from the single components, closer following
the strict mathematical analytical way, can be found in references [2] and [3].
The following example leans on Example A1 of the EURACHEM /CITAC Guide ”Quantifying
Uncertainty in Analytical Measurement” [3]. It has been chosen to clarify the 8 steps of the
procedure.
The aim is to prepare a calibration standard of Cd in HNO3 for AAS. To obtain the Cd cali-
bration standard of ca. 1000mg/l the following procedure was applied:
1. Weighing a piece of metal (surface cleaned).
2. Dissolving the metal in a 100 ml flask by adding 1 ml HNO3 and filling with deionised
water.
19
Volume Purity
Temperature
Calibration
Repeatability
Concentration
cCd
Readability Readability
m (gross)
Linearity Linearity
Repeatability Repeatability
Sensitivity Sensitivity
Calibration Calibration
Step 5) Calculate the result of the measurement from the input quantities
Determination of the concentration
1000 ⋅ m ⋅ P 1000 ⋅ 100.28 ⋅ 0.9999
cCd = [mg / l ] = mg / l = 1002.7 mg / l
V 100
The concentration of the calibration standard is 1002.7 mg/l.
21
5. Conclusions
There are many possibilities and ways for uncertainty estimation of methods or components
thereof, based on experience and general knowledge gathered from practice.
Estimates can also be based on a list of the main influencing factors. Such a list is helpful in
any case, as it facilitates final checking of the estimation method for covering all relevant
components.
The laboratory’s expertise is needed to decide whether the respective data can be used for
the uncertainty estimation with view to: observed range, whether all major components are
included, e.g. preparation steps, dilution, .... It may be difficult to put all these components
and the available data together into a model. The assessment of measurement uncertainties
can only be based on the present state of the art, however. Laboratories should use the data
available for the evaluation of measurement uncertainty.
The degree of rigor and detail of mathematical models differs widely among fields of meas-
urement. In particular in the testing area, it is often not possible to draw up a very detailed
model, as the method does not allow it, or the method does not require it. In those cases, a
much simpler approach may be applied (ISO/IEC 17025, 5.4.6.2). Furthermore, the evalua-
tion of measurement uncertainty is still evolving, and as a result there are great differences
in the progress made in the various fields of measurement and testing. Therefore in different
fields uncertainty estimations will be at a different stage of development, and the realisation
may need different amounts of time and may need to be implemented stepwise.
It is often argued, that customers are confused by uncertainty statements. Therefore
ISO/IEC 17025 (5.10.3) states, that measurement uncertainty does not always have to be
stated but only in those cases where it is relevant for the client.
Finally, it should be realised that measurement uncertainty is of added value, both for the
laboratory in terms of its quality management and for the customer as well.
22
6. Compilation of the main and some specific literature on the field of measurement uncertainty
Title Author Year Remarks Source
[2] Expression of the Uncertainty of Measurement in Calibration EA-4/02, EA Taskforce for Revision of 1999 Very good, comprehensible document. Many Can be downloaded from EA home-
WECC doc 19-1990 well explained examples from calibration field. page (www.european-accreditation.org)
[3] Quantifying Uncertainty in Analytical Measurement, EURACHEM Eurachem / CITAC 1995 / 2000 Most comprehensible document for meas- 120 p.; download from
/CITAC Guide urement uncertainties. Many very good exam- www.measurementuncertainty.org
ples. Good explanation of "Spreadsheet or www.eurachem.bam.de
method".
[4] The Expression of Uncertainty and Confidence in Measurement NAMAS NIS 3003 1997 Good description of the concepts of uncer- NAMAS, United Kingdom Accreditation
tainty determination, examples mainly from Service, 21-47 High Street, Feltham,
calibration area. Middlesex TW13 4UN
[5] Guidelines for evaluating and expressing uncertainty of NIST Barry N. Taylor and Chris E. Kuyatt, 1993 Good description of the concepts of uncer- Download from http://physics.nist.gov
measurement results NIST tainty determination
23
Implementation of uncertainty
[15] Uncertainty of test results ("Result Uncertainty") DAR-EM22, DAR ATF Ad-hoc group 1996 General notes on application of uncertainty. DAR-EM22, http://www.dar.bam.de/
"Uncertainty in testing" Good table of publications on uncertainty. under documents, ATF
[16] The Expression of Uncertainty in Quantitative Testing EA-3/02, (previously EAL-G23) 1996 Guidance for accreditors for implementation http://www.european-accreditation.org/
of uncertainty requirements
[17] 1. Important information to our customers concerning the quality SP, P.O. Box 857, SE-501, 15 Boras, 2000 Leaflet on measurement uncertainty in prac- SP, P.O. Box 857, SE-501, 15 Boras,
of measurements Sweden, Tel: +46-33-13-55-02 tice Sweden, Tel: +46-33-13-55-02
2. Measurement Uncertainty – Surveys about Customers’ Knowl- U. Örnemark, Magnus Holmgren 2001 Surveys about Customers’ Knowledge, Reac- [email protected]
edge, Reactions and Needs tions and Needs
[18] Measurement Uncertainty UKAS 2000 Short introduction to measurement uncer- www.ukas.com Accreditation topics:
tainty Measurement Uncertainty
[19] ILAC-G17:2002: Introducing the Concept of Uncertainty of Meas- ILAC and EA 2001 Important policy www.ilac.org
urement in Testing in Association with the Application of the
Standard ISO/IEC 17025
[20] APLAC Policy and Guidance on the estimation of Uncertainty of APLAC 2001 Good and useful document, gives sector
Measurement in Testing – Draft November 2001 oriented advice
[21] A2LA policy A2LA 2000 An implementation strategy into accreditation www.a2la.org
[22] Assessment of Uncertainties of Measurement for Electrical Test- Nata Australia 1992
ing
Standards associated with measurement uncertainty
[23] French Standardisation: Aid in the procedure for estimating and AFNOR Description how to use precision data for
using uncertainty in measurement and test results – FD X 07-021 uncertainty estimation
[24] ISO/DIS 10576-1 Statistical methods “Guidelines for the evaluati- ISO, TC 69 2001, draft
on of conformity with specified requirements”
[25] ISO/TS 14253-2:1999 Geometrical Product Specifications (GPS) - ISO 1999 ISO, www.iso.ch; CHF 164.00
- Inspection by measurement of workpieces and measuring
equipment -- Part 2: Guide to the estimation of uncertainty in GPS
measurement, in calibration of measuring equipment and in prod-
uct verification
[26] ISO 3951:1989 Sampling procedures and charts for inspection by ISO 1989 ISO, www.iso.ch; CHF 188.00
variables for percent nonconforming
[27] ISO 6974 Natural gas -- Determination of composition with de- ISO 2000
fined uncertainty by gas chromatography
[28] ISO 13752 Air quality -- Assessment of uncertainty of a meas- ISO 1998
urement method under field conditions using a second method as
reference
[29] ISO 7066 Assessment of uncertainty in calibration and use of flow ISO 1997
measurement devices
[30] ISO 5725-1-6: 1994/Cor. 2001 Accuracy (trueness and precision) ISO 1994 ISO, www.iso.ch
of measurement methods and results
[31] DIN 40080 Stichprobenprüfung anhand qualitativer Merkmale DIN 1993 Beuth Verlag
DIN ISO 2859-1, Ausgabe:1993-04
Annahmestichprobenprüfung anhand der Anzahl fehlerhafter
Einheiten oder Fehler (Attributprüfung)
24
[32] DIN 53804 T3/T4 Statistische Auswertungen, Ordinal- DIN 1982 / 1985 Beuth Verlag
/Attributmerkmale
[33] DIN 25424 Fehlerbaumanalye DIN 1990 Beuth Verlag
[34] DIN 1319 Teil 3 "Auswertung v. Messungen einer einzelnen Meß- DIN 1996 / 1999 Beuth Verlag
größe"; Meßunsicherheit; Teil 4 "Behandlung von Unsicherheiten
bei der Auswertung von Messungen"
[35] Statistical assessment of the uncertainty of measurement results: ISO TC69/SC6/WG7 2001 Draft
Guide to the use of repeatability, reproducibility and trueness
estimates in measurement uncertainty estimation
Software for Uncertainty calculation
[36] Tools for the test laboratory to implement measurement uncer- Sven Nytoft Rasmussen; Nordtest 1999 Very good comparison of the concepts of the Nordtest, P.O. Box 116, FIN-02151
tainty budgets Techn. Report 430 different computer programs Espoo, Finland, Phone +358-0-455-
4600
[37] MUSAC (Measurement Uncertainty in Analytical Chemistry) EMPA, Creasoft AG and others 2001 software system for application in the chemi- information at: www.musac.ch.; price
cal laboratory: e.g. titration, HPLC, GC, not jet known
GC/MS, ICP/OES, ICP/MS and AAS, first part
available September 2001
[38] AESoft Uncertainty Atkinson Engineering, Inc 1000 $, free demo version;
www.aesoft.com/Unc1.html
[39] GUM Workbench Metrodata GmbH, Grenzach-Wyhlen, 1100 $, free demo version; Metrodata
Germany GmbH; http://www.gum.dk;
http://www.metrodata.de/
[40] DFM-GUM Danish Institute of fundamental me- 300 $;
trology http://www.dfm.dtu.dk/en/consult/dfm-
gum.htm
[41] Uncertainty Analyzer Integrated Sciences Group. Uncer- 1994-96 700 $, USA, Bakersfield,
tainty Analyser 1.0 manual Ca.www.isgmax.com;
http://www.quametec.com/downloads.ht
m
[42] Expression Buddy James E. Presley and Daniel B. free software; download at
Presley http://www.jpresley.com/
[43] “Uncert” Project
Books
[44] Experimentation and Uncertainty Analysis for Engineers Hugh W.Coleman, W. Glenn Steel 1999 New York, Wiley & Sons Inc. ISBN 0-
471-12146-0
[45] Meßunsicherheit und Meßdatenauswertung Klaus Weise, Wolfgang Wöger 1999 Weinheim, Wiley-VCH; ISBN: 3-527-
29610-7
[46] Messunsicherheiten: Theorie und Praxis Franz Adunka 1998 Essen, Vulkan-Verlag, ISBN 3-8027-
2186-1
[47] Estimer l'incertitude, Mesures - Essais Christophe Perruchet, Marc Priel 2000
[48] Uncertainty, calibration, and probability; the statistics of scientific C. F. Dietrich 1991 Adam Hilger, Bristol; ISBN: 0-7503-
and industrial measurement 0060-4
[49] Uncertainty models for knowledge-based systems : a unified Irwin R. Goodman and Hung T. 1985 North-Holland, Amsterdam, ISBN: 0-
approach to the measurement of uncertainty Nguyen 444-87796-7
[50] Measurement uncertainty: methods and applications Ronald H. Dieck 1992, 1997 Instr. Soc. Of America, ISBN: 1-55617-
628-7
25
[51] Handbuch Validierung in der Analytik Stavros Kromidas 1999 Wiley-VCM ISBN: 3-527-28748-5
[52] Calculation & Reporting Uncertainties of Measurements in Testing N. Kukadia, Tenby Industries Limited, 1996
Birmingham, UK
Some examples of special topics
[53] Uncertainty of quantitative determinations derived by cultivation of
microorganisms, Centre for metrology and accreditation, Helsinki,
2002
[54] Guidelines for expressing the Uncertainty of Measuerement re- Philips, Eberhardt and Parry, NIST; 1997 Taking into account uncorrected bias download from
sults containing uncorrected Bias Journal of Research of the Nat. Inst. http://nvl.nist.gov/pub/nistpubs/jres/102/
Of Standards and Technology, 102, S. 5/j25phi.pdf
577
[55] Calculation of Measurement Uncertainty Using Prior Information S.D. Phillips, W.T. Estler, M.S. Leven- 1998 Treating bias in uncertainty evaluation download from
son, K.R. Eberhardt http://nvl.nist.gov/pub/nistpubs
[56] Uncertainty and Dimensional Calibrations Doiron and Stoup, Journal of Research 1997 Examples for variables contributing in calibra- download from
of the Nat. Inst. Of Standards and tion http://nvl.nist.gov/pub/nistpubs
Technology, 102, S. 647
[57] An Interpretation of the Guide to the expression of Uncertainty in R.N. Kacker, NIST Techni-Pubs SP 2000 20$, ordering at
Measurement 500-244 http://nvl.nist.gov/pub/nistpubs
[58] Observing validation, uncertainty determination and tracebility in Magnus Holmgren, NT Techn. Report 1998 www.vtt.fi/nordtest/tec403.htm
developing Nordtest test methods 403
[59] Uncertainty of pH Measurements H. Jensen, L. Nielsen, Nordtest Techn 1995 Nordtest, P.O. Box 116, FIN-02151
Report 284 Espoo, Finland, Phone +358-0-455-
4600
[60] Traceable calibration and uncertainty of measurements in scan- Halldur Gudmundsson, Thomas 1994 Nordtest, P.O. Box 116, FIN-02151
ning electron microscopy Runarsson, Nordtest Techn Report Espoo, Finland, Phone +358-0-455-
252 4600
[61] Traceable calibration and uncertainty of measurements and tests B. Steffen, H.Kallio, Freygardur Thor- 1994 Nordtest, P.O. Box 116, FIN-02151
steinsson, Nordtest Techn. Report 251 Espoo, Finland, Phone +358-0-455-
4600
[62] Calibration, traceability and uncertainty Nordtest-Seminar in Espoo, NT Techn. 1995
Report 305
[63] A view on the assessment of the technical competence of testing Jarl Forstén, Nordtest Techn Report 1991
laboratories 149
[64] The use of uncertainty estimates in testing Nordtest SP Report 1993:47 1993
[65] Nordtest’s Views on Measurement in Metrology and Testing Nordtest, Position Paper 1994 Nordtest, P.O. Box 116, FIN-02151
Espoo, Finland, Phone +358-0-455-
4600
[66] Uncertainty - to a certain level Nordtest Position paper 005 1998 Ways for assessing uncertainties for different Nordtest, P.O. Box 116, FIN-02151
situations of tests and measurement Espoo, Finland, Phone +358-0-455-
4600
[67] Estimating Measurement Uncertainty Erwin Achermann, Oscar Chinellato; 2000 Theoretical Scientific Computing: The "Maxi- ETH Zürich, Institute of Scientific Com-
Technical Report 346. ETH Zürich mum Likelihood (fitting of a) Functional Rela- puting;
tionship" (MLFR) http://www.inf.ethz.ch/research/wr/publi
cations/tr.html
[68] Instrument and Apparatus - Measurement Uncertainty, part 1 ANSI/ASME PTC 19.1-1985 (1990) 1985 / 1990 ASME, New York
[69] Measurement uncertainty of radiated emissions David A. Hill, Motohisa Kanda 1997
[70] The fitness for purpose of analytical methods Eurachem 1998 Method validation studies ISBN 0-948926-12-0
26
[71] ILAC-G13. Guidelines for the requirements for the Competence of ILAC 2000 http://www.ilac.org/
Providers of Proficiency Testing Schemes (uncertainty assess-
ment in the Annex)
[72] ILAC-G8. Guidelines on Assessment and Reporting of Compli- ILAC Rules for supplier and client concerning the http://www.ilac.org/
ance with Specifications compliance of a product
[73] NAMAS NIS 80 NAMAS NIS 80
[74] Uncertainties of measurement for NAMSA Electrical product NAMAS NIS 20 1989
testing laboratories NIS 20, NAMAS
[75] Estimation and Expression of Measurement Uncertainty in Chemi- Nordic Committee on Food Analysis, 1997
cal Analysis NMKL Procedure 5
[76] Measurement uncertainty considerations for co-ordinate measur- S.D. Phillips, B. Borchardt, G. Caskey 1993
ing machines
[77] Step by step estimate of uncertainty of test results J.S. Morowski, EMPA, Switzerland, at 1994 Examples of estimation of uncertainties from EUROLAB
2nd EUROLAB Symposium "Testing practice
for the year 2000", Florence, April
1994
[78] Assessment and Practical Use of Uncertainties in test results Hans Andersso, SP, Sweden at 2nd 1994 Calculating measurement uncertainties EUROLAB
EUROLAB Symposium "Testing for
the year 2000", Florence, April 1994
[79] Uncertainty in testing Norbert Müller, arsenal research, 1999 Important aspects of harmonisation of the use EUROLAB - in discussion
Austria of uncertainties
[80] Practical Experiences with Uncertainty Evaluation at VTT Chemi- Veikko Komppa, VTT, Finland at 2nd 1997 Examples from practice, graphics: measure- Eurachem/D / EUROLAB-D
cal Technology Workshop "Measurement Uncertainty ment uncertainty against concentration
in Chemical Analysis", Berlin Septem-
ber 1997
[81] Interlaboratory Study and Validation Data Stephen L.R. Ellison, Laboratory of the 1997 Use of interlaboratory comparisons for uncer- Eurachem/D / EUROLAB-D
government Chemist, UK at 2nd Work- tainty determination
shop "Measurement Uncertainty in
Chemical Analysis", September 1997
[82] Autosamplers - a major uncertainty factor in HPLC Analysis Pre- S. Küppers, B. Renger, V.R. Meyer, 2000
cision LC,GC Europe, 2/2000
[83] The evaluation of the uncertainty in knowing a directly measured I.H. Lira, W. Wöger, PTB, Braun- 1998
quantity schweig, Germany; Pontificia Universi-
dad Catolica de Chile, Santiago
Meas. Sci. Technol. 9, 1167-1173
[84] Is the estimation of measurement uncertainty a viable alternative S. Küppers, Accred. Qual. Assur. 3 1998
to validation? 1998 p. 412-415
[85] NIST list of recent and new documents NIST homepage, search option Good variety of publications, e.g. special http://nvl.nist.gov/pub/nistpubs/nistpubs
literature on measurement uncertainty, some .htm
can be downloaded
[86] Verlauf und Ergebnisse einer geordneten Ringschätzung der J.S. Morkowski, Umweltbundesamt, See Example 3 of this document 7,67 Euro;
Unsicherheiten von Emissionsmessungen (Process and results of Texte 54/99, Berlin, 1999 www.umweltbundesamt.de
an ordered interlaboratory estimation of the uncertainties from
emission measurements)
27