Astm Astm d2887 PDF
Astm Astm d2887 PDF
Astm Astm d2887 PDF
Designation: 406
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
3.3.1 A common abbreviation of hydrocarbon compounds is boiling temperature of 538°C (1000°F) before reaching the
to designate the number of carbon atoms in the compound. A upper end of the temperature program. The programming rate
prefix is used to indicate the carbon chain form, while a must be sufficiently reproducible to obtain retention time
subscripted suffix denotes the number of carbon atoms (for repeatability of 0.1 min (6 s) for each component in the
example, normal decane = n-C10; isotetradecane = i-C14). calibration mixture described in 7.8.
6.1.3 Cryogenic Column Cooling—Column starting tem-
4. Summary of Test Method peratures below ambient will be required if samples with IBPs
4.1 The boiling range distribution determination by distilla- of less than 93°C (200°F) are to be analyzed. This is typically
tion is simulated by the use of gas chromatography. A nonpolar provided by adding a source of either liquid carbon dioxide or
packed or open tubular (capillary) gas chromatographic col- liquid nitrogen, controlled through the oven temperature cir-
umn is used to elute the hydrocarbon components of the sample cuitry. Excessively low initial column temperature must be
in order of increasing boiling point. The column temperature is avoided to ensure that the stationary phase remains liquid. The
raised at a reproducible linear rate and the area under the initial temperature of the column should be only low enough to
chromatogram is recorded throughout the analysis. Boiling obtain a calibration curve meeting the specifications of the
points are assigned to the time axis from a calibration curve method.
obtained under the same chromatographic conditions by ana- 6.1.4 Sample Inlet System—The sample inlet system must
lyzing a known mixture of hydrocarbons covering the boiling be capable of operating continuously at a temperature equiva-
range expected in the sample. From these data, the boiling lent to the maximum column temperature employed, or provide
range distribution can be obtained. for on-column injection with some means of programming the
entire column, including the point of sample introduction, up to
5. Significance and Use the maximum temperature required. Connection of the column
5.1 The boiling range distribution of petroleum fractions to the sample inlet system must be such that no temperature
provides an insight into the composition of feedstocks and below the column temperature exists.
products related to petroleum refining processes. The gas 6.1.5 Flow Controllers—The gas chromatograph must be
chromatographic simulation of this determination can be used equipped with mass flow controllers capable of maintaining
to replace conventional distillation methods for control of carrier gas flow constant to 61 % over the full operating
refining operations. This test method can be used for product temperature range of the column. The inlet pressure of the
specification testing with the mutual agreement of interested carrier gas supplied to the gas chromatograph must be suffi-
parties. ciently high to compensate for the increase in column back-
5.2 Boiling range distributions obtained by this test method pressure as the column temperature is raised. An inlet pressure
are essentially equivalent to those obtained by true boiling of 550 kPa (80 psig) has been found satisfactory with the
point (TBP) distillation (see Test Method D2892). They are not packed columns described in Table 1. For open tubular
equivalent to results from low efficiency distillations such as columns, inlet pressures from 10 to 70 kPa (1.5 to 10 psig)
those obtained with Test Method D86 or D1160. have been found to be suitable.
profile to be stored in memory. This profile is automatically subtracted listed by the American Chemical Society, see Annual Standards for Laboratory
from the detector signal on subsequent sample analyses to compensate for Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
any baseline offset. Some integration systems also store and automatically and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,
subtract a blank analysis from subsequent analytical determinations. MD.
9.2.1.8 Remove the cap from the detector end of the column analysis (see 7.7 and Section 10), and is illustrated in Fig. 1.
and turn the carrier gas back on. Resolution (R) should be at least three, using the identical
9.2.1.9 Program the column temperature up to the maxi- conditions employed for sample analyses:
mum several times with normal carrier gas flow. Connect the R 5 2~t2 2 t1! / [1.699~w2 1 w1!# (1)
free end of the column to the detector.
9.2.1.10 An alternative method of column conditioning that where:
has been found effective for columns with an initial loading of R = resolution,
10 % liquid phase consists of purging the column with carrier t1 = time(s) for the n-C16 peak maximum,
gas at the normal flow rate while holding the column at the t2 = time(s) for the n-C18 peak maximum,
maximum operating temperature for 12 to 16 h, while detached w1 = peak width(s), at half height, of the n-C16 peak, and
from the detector. w2 = peak width(s), at half height, of the n-C18 peak.
9.2.2 Open Tubular Columns—Open tubular columns with 9.3.2 Detector Response Calibration—This test method
cross-linked and bonded stationary phases are available from assumes that the detector response to petroleum hydrocarbons
many manufacturers and are usually pre-conditioned. These is proportional to the mass of individual components. This
columns have much lower column bleed than packed columns. must be verified when the system is put in service, and
Column conditioning is less critical with these columns but whenever any changes are made to the system or operational
some conditioning may be necessary. The column can be parameters. Analyze the calibration mixture using the identical
conditioned very rapidly and effectively using the following procedure to be used for the analysis of samples (see Section
procedure. 10). Calculate the relative response factor for each n-paraffin
9.2.2.1 Once the open tubular column has been properly (relative to n-decane) in accordance with Practice D4626 and
installed into the gas chromatograph and tested to be leak free, Eq 2:
set the column and detector gas flows. Before heating the Fn 5 ~Mn / An! / ~M10 / A10! (2)
column, allow the system to purge with carrier gas at ambient
temperature for at least 30 min. where:
9.2.2.2 Increase the oven temperature about 5 to 10°C per Fn = relative response factor,
minute to the final operating temperature and hold for about 30 Mn = mass of the n-paraffin in the mixture,
min. An = peak area of the n-paraffin in the mixture,
9.2.2.3 Cycle the gas chromatograph several times through M10 = mass of the n-decane in the mixture, and
A10 = peak area of the n-decane in the mixture.
its temperature program until a stable baseline is obtained.
9.3 System Performance Specification: The relative response factor (Fn) of each n-paraffin must not
9.3.1 Column Resolution—The column resolution, influ- deviate from unity (1) by more than 610 %.
enced by both the column physical parameters and operating 9.3.3 Column Elution Characteristics—The column mate-
conditions, affects the overall determination of boiling range rial, stationary phase, or other parameters can affect the elution
distribution. Resolution is therefore specified to maintain order of non-paraffinic sample components, resulting in devia-
equivalence between different systems (laboratories) employ- tions from a TBP versus retention time relationship. If station-
ing this test method. Resolution is determined using Eq 1 and ary phases other than those referenced in 7.3 are used, the
the C16 and C18 paraffins from a column resolution test mixture retention times of a few alkylbenzenes (for example, o-xylene,
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
FIG. 1 Column Resolution Parameters
n-butyl-benzene, 1,3,5-triisopropylbenzene, n-decyl-benzene, analysis except no injection is made. A blank analysis must be
and tetradecylbenzene) across the boiling range should be performed at least once per day. The blank analysis is neces-
analyzed to make certain that the column is separating in sary due to the usual occurrence of chromatographic baseline
accordance with the boiling point order (see Appendix X1). instability and is subtracted from sample analyses to remove
any nonsample slice area from the chromatographic data. The
10. Calibration and Standardization
blank analysis is typically performed prior to sample analyses,
10.1 Analysis Sequence Protocol—Define and use a prede- but may be useful if determined between samples or at the end
termined schedule of analysis events designed to achieve of a sample sequence to provide additional data regarding
maximum reproducibility for these determinations. The sched- instrument operation or residual sample carryover from previ-
ule will include cooling the column oven to the initial starting ous sample analyses. Attention must be given to all factors that
temperature, equilibration time, sample injection and system influence baseline stability, such as column bleed, septum
start, analysis, and final upper temperature hold time. bleed, detector temperature control, constancy of carrier gas
10.1.1 After chromatographic conditions have been set to flow, leaks, instrument drift, and so forth. Periodic baseline
meet performance requirements, program the column tempera- blank analyses should be made, following the analysis se-
ture upward to the maximum temperature to be used and hold quence protocol, to give an indication of baseline stability.
that temperature for the selected time. Following the analysis
sequence protocol, cool the column to the initial starting NOTE 6—If automatic baseline correction (see Note 4) is provided by
temperature. the gas chromatograph, further correction of area slices may not be
10.1.2 During the cool down and equilibration time, ready required. However, if an electronic offset is added to the signal after
baseline compensation, additional area slice correction may be required in
the integrator/computer system. If a retention time or detector the form of offset subtraction. Consult the specific instrumentation
response calibration is being performed, use the peak detection instructions to determine if an offset is applied to the signal. If the
mode. For samples and baseline compensation determinations, algorithm used is unclear, the slice area data can be examined to determine
use the area slice mode of integration. The recommended slice if further correction is necessary. Determine if any offset has been added
rate for this test method is 1.0 Hz (one slice per second). Other to the compensated signal by examining the corrected area slices of those
slice rates may be used if within the limits of 0.02 and 0.2 % time slices that precede the elution of any chromatographic unretained
substance. If these corrected area slices (representing the true baseline)
of the retention time of the final calibration component (C44).
deviate from zero, subtract the average of these corrected area slices from
Larger slice rates may be used, as may be required for other each corrected area slice in the analysis.
reasons, if provision is made to accumulate (bunch) the slice
data to within these limits prior to determination of the boiling 10.3 Retention Time Versus Boiling Point Calibration—In
range distribution. order to analyze samples, a retention time versus boiling point
10.1.3 At the exact time set by the schedule, inject either the calibration must be performed. Inject an appropriate aliquot
calibration mixture or sample into the chromatograph; or make (0.2 to 2.0 µL) of the calibration mixture (see 7.8) into the
no injection (baseline blank). At the time of injection, start the chromatograph, using the analysis sequence protocol. Obtain a
chromatograph time cycle and the integrator/computer data normal (peak detection) data record in order to determine the
acquisition. Follow the analysis sequence protocol for all peak retention times and the peak areas for each component.
subsequent repetitive analyses or calibrations. Since complete Collect a time slice area record if a boiling range distribution
resolution of sample peaks is not expected, do not change the report is desired.
detector sensitivity setting during the analysis. 10.3.1 Inspect the chromatogram of the calibration mixture
10.2 Baseline Compensation Analysis—A baseline compen- for evidence of skewed (non-Gaussian shaped) peaks. Skew-
sation analysis, or baseline blank, is performed exactly like an ness is often an indication of overloading the sample capacity
of the column that will result in displacement of the peak apex difference in boiling point temperature. In general, the lower
relative to nonoverloaded peaks. Distortion in retention time the sample IBP, the lower will be the starting temperature of
measurement and hence errors in boiling point temperature the analysis. Although extrapolation of the curve at the upper
determination will be likely if column overloading occurs. The end is more accurate, calibration points must bracket the
column liquid phase loading has a direct bearing on acceptable boiling range of the sample at both the low and high ends.
sample size. Reanalyze the calibration mixture using a smaller 10.4 Reference Gas Oil Analysis—The Reference Gas Oil
sample size or a more dilute solution to avoid peak distortion. No. 1 sample is used to verify both the chromatographic and
10.3.2 Prepare a calibration table based upon the results of calculation processes involved in this test method. Perform an
the analysis of the calibration mixture by recording the time of analysis of the gas oil following the analysis sequence proto-
each peak maximum and the boiling point temperature in col. Collect the area slice data and provide a boiling point
degrees Celsius (or Fahrenheit) for every component in the distribution report as in Sections 12 and 13.
mixture. n-Paraffin boiling point temperatures are listed in
Table 2. 10.4.1 The results of this reference analysis must agree with
10.3.3 Plot the retention time of each peak versus the the values given in Table 3 within the range specified by the
corresponding normal boiling point temperature of that com- test method reproducibility (see 14.1.2). If it does not meet the
ponent in degrees Celsius (or Fahrenheit) as shown in Fig. 2. criteria in Table 3, check that all hardware is operating properly
10.3.4 Ideally, the retention time versus boiling point tem- and all instrument settings are as recommended by the manu-
perature calibration plot would be linear, but it is impractical to facturer. Rerun the retention boiling point calibration as de-
operate the chromatograph such that curvature is eliminated scribed in 10.3.
completely. The greatest potential for deviation from linearity 10.4.2 Perform this reference gas oil confirmation test at
will be associated with the lower boiling point paraffins. They least once per day or as often as required to establish
will elute from the column relatively fast and have the largest confidence in consistent compliance with 10.4.1.
5 6
Supporting data have been filed at ASTM International Headquarters and may Supporting data have been filed at ASTM International Headquarters and may
be obtained by requesting Research Report RR:D02-1477. be obtained by requesting Research Report RR:D02-1406.
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
TABLE 5 Reproducibility gas chromatography has no bias because the boiling range
NOTE—x = the average of the two results in °C and y = the average of distribution can only be defined in terms of a test method.
the two results in °F. 14.2.1 A rigorous, theoretical definition of the boiling range
Reproducibility
distribution of petroleum fractions is not possible due to the
% Off complexity of the mixture as well as the unquantifiable
°C °F
interactions among the components (for example, azeotropic
IBP 0.066 x 0.06 (y − 32)
5% 0.015 (x + 100) 0.015 (y + 148)
behavior). Any other means used to define the distribution
10–20 % 0.015 (x + 100) 0.015 (y + 148) would require the use of a physical process, such as a
30 % 0.013 (x + 100) 0.013 (y + 148) conventional distillation or gas chromatographic characteriza-
40 % 4.3 7.7
50–90 % 4.3 7.7
tion. This would therefore result in a method-dependent
95 % 5.0 9.0 definition and would not constitute a true value from which
FBP 11.8 21.2 bias can be calculated.
15. Keywords
15.1 boiling range distribution; correlation; distillation; gas
14.2 Bias—The procedure in Test Method D2887 for deter- chromatography; petroleum; petroleum fractions; petroleum
mining the boiling range distribution of petroleum fractions by products; simulated distillation
APPENDIXES
(Nonmandatory Information)
X1.1 There is an apparent discrepancy in the boiling point essentially the same. Deviations of simulated distillation boil-
of multiple ring-type compounds. When the retention times of ing points, as estimated from the curve, from actual boiling
these compounds are compared with n-paraffins of equivalent points for a few compounds are shown in Table X1.2. The
atmospheric boiling point, these ring compounds appear to be deviations obtained by plotting boiling points at 10 mm rather
eluted early from methyl silicone rubber columns. A plot than 760 mm are tabulated also. It is apparent that the deviation
showing 36 compounds other than n-paraffins plotted along the is much less at 10 mm pressure. This indicates that the
calibration curve for n-paraffins alone is shown in Fig. X1.1. distillation data produced by gas chromatography closely
The numbered dots are identified in Table X1.1. In this figure, approximates those obtained in reduced pressure distillation.
the atmospheric boiling points are plotted against the observed Since the vapor-pressure-temperature curves for multiple-ring
retention times. If columns containing different percentages of type compounds do not have the same slope or curvature as
stationary phase or different temperature programming rates those of n-paraffins, an apparent discrepancy would exist when
were used, the slope and curvature of the n-paraffin curve (solid
n-paraffin boiling points at atmospheric pressure are used.
line) would change, but the relative relationships would remain
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
FIG. X1.1 Boiling Point—Retention Time Relationships for Several High-Boiling Multiple-Ring Type Compounds (see Table X1.1)
X2.1 Test Method D2892 is the standard for conventional methods has been very good for petroleum products and
distillation of petroleum products. fractions within the scope of this test method.
X2.2 This test method has been compared with Test X2.3 The time required for analysis by this test method is
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
X3.1 Required Starting Elements X3.1.2.1 The analysis conditions for blank and sample must
X3.1.1 Sample Data Array, N Area Slices—The data must be identical through the point where sample analysis is
be collected at a minimum sampling frequency of 1 Hz (that is, terminated.
maximum slice width is 1 s). In addition, the slice width must X3.1.2.2 The number of slices in the blank array must be
be such that no sample or solvent elutes in the first five slices. equal to or greater than the number of slices in the sample
X3.1.2 Blank Data Array, N Area Slices—The slice width chromatogram. If the number of slices in the blank array is
for the blank and sample runs must be identical. (A blank data greater than the number of slices in the sample array, then drop
array is not necessary if electronic baseline compensation is the extra slices in the blank array. This situation could occur if
used.)
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
not give the same results as slice-by-slice blank subtraction. On 1E-7*total chromatogram area, then take slice N+1 as the start
some instruments using automatic baseline compensation, the of sample slice.
compensated baseline has been observed to exhibit anomalous X3.5.2 Print the retention time corresponding to the start of
features at (or near) the point in the chromatogram where the sample elution.
programmed oven temperature reaches maximum and is held
for some period of time. The anomalous feature appears as a X3.6 Determine End of Sample Elution Time
slow rise in baseline, followed by a relatively sharp decrease, X3.6.1 Starting at the last slice in the data array and working
followed by a level baseline. While the magnitudes of the toward the start of sample, determine where the rate of change
anomalies observed have been very small (only a few pico- per second between two consecutive bunched slices first
amps), the slope of the sharp decrease may be sufficient to meet exceeds 0.00001 % of the total chromatogram area (see
the criterion for determining the end of sample elution. In such X3.4.2).
event, this false triggering of the end of sample criterion will X3.6.1.1 For determining end of sample elution, the rate of
result in erroneously high values for the FBP and other percent change is calculated by subtracting the area of a slice from the
off values near the FBP. If false triggering occurs and can not area of the immediately preceding slice and dividing by the
be eliminated, the user should disable automatic baseline bunched slice width (see X3.3.4) in seconds.
compensation and perform blank subtractions as described in X3.6.1.2 If (<slice>N–1 – <slice>N)/(bunched slice width) >
this appendix.) 1E-7*total chromatogram area, then take slice N-1 as the end
NOTE X3.1—If the data was acquired on an instrument using automatic of sample slice.
baseline compensation, X3.2 must be skipped. In this case, the zeroed and X3.6.2 Print the retention time corresponding to end of
bunched sample data array (see X3.3.3) contains the corrected area slices sample elution.
to be used in subsequent calculations.
X3.7 Calculate Total Corrected Sample Area
X3.3 Zero and Bunch Data Slices
X3.7.1 Sum the corrected area slices from start of sample
X3.3.1 Calculate the average of the first five area slices of slice (see X3.5.1.2) to end of sample slice (see X3.6.1.2).
the blank-subtracted data array. X3.7.2 Designate this sum as the total corrected sample
X3.3.2 Subtract the average slice area (see X3.3.1) from area, and save for subsequent calculations.
each area slice in the blank-subtracted data array. Set negative
numbers to zero. X3.8 Normalize to Area Percent
X3.3.3 If the data sampling frequency was 1.5 Hz or greater, X3.8.1 Starting at the start of sample slice (see X3.5.1.2)
bunch (add together) an integral number of area slices to obtain and continuing to the end of sample slice (see X3.6.1.2), divide
a bunched slice width as close to 1 s as possible. Drop any each corrected area slice by the total corrected sample area (see
extra area slices at the end of the sample data array. X3.7.2) and multiply by 100.
NOTE X3.2—As an example, if data was collected at 3 Hz, add the areas X3.8.2 Save these normalized area percents in an array for
of slices 1, 2, and 3 and use the retention time of slice 3. Add the area of subsequent calculations.
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
X5.1 The resulting data obtained from carrying out an ai = ith coefficient from Table X5.1, and
analysis by Test Method D2887 can be used to obtain Test Tn = nth boiling point temperature of D2887.
Method D86 data via a correlation. The correlations used to
convert selected Test Method D2887 distillation points (per- X5.6 Basis
cent off) to Test Method D86 (percent off) are mathematical X5.6.1 This correlation model is based on data for 46 jet
equations. fuel samples and 39 diesel fuel samples analyzed in accordance
with both Test Method D86 and D2887. From these results, a
X5.2 A correlation model is presented here for the calcu- correlation model was determined using regression, specifying
lation of Test Method D86 correlated data from boiling range coefficients per recovery. A model of the remaining bias was
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
distribution analysis by gas chromatography according to Test determined by use of Practice D6708 on a dataset from the
Method D2887. This correlation model is only valid for diesel ASTM Interlaboratory Crosscheck Program of five jet fuels
and jet fuels, excluding biodiesels. and six diesels analyzed by 38 laboratories by Test Method
X5.3 This correlation model was validated by an analysis D2887 and 201 laboratories by Test Method D86.
of variance procedure in accordance with Practice D6708. X5.6.2 The bias correction model was used to correct the
results from the correlation model, resulting in a new correla-
X5.4 Significance and Use tion matrix given in Table X5.1.
X5.4.1 Valid data for conversion to Test Method D86 X5.6.3 Based on statistical significance tests, no sample
correlated data can be obtained by Test Method D2887. The specific biases were observed in the dataset used for the bias
model is only valid for diesel or jet fuel and samples must meet correction.
the specifications given in Test Method D2887. X5.6.4 Both methods were found sufficiently precise to
distinguish among the samples.
X5.5 Summary of the Procedure X5.6.5 Precision and Bias8—Reproducibility after conver-
X5.5.1 Test Method D86 correlated data is calculated from sion of Test Method D2887 data into Test Method D86 data is
Test Method D2887 data using Eq X5.1 and the coefficients equivalent to the reproducibility of Test Method D2887.
specified in Table X5.1. X5.6.6 Cross-method reproducibility after conversion of
tn 5 ao 1 a1 · Tn21 1 a2 · Tn 1 a3 · Tn11 (X5.1) Test Method D2887 data into Test Method D86 correlated data
is given in Table X5.2.
where:
tn = nth boiling point temperature of Test Method D86 8
Supporting data have been filed at ASTM International Headquarters, and may
correlated,
be obtained by requesting Research Reports RR:D02-1553 and D02–1564.
X6.1 This section has been added to the test method to TABLE X6.1 Typical Operating Parameters for Accelerated Test
provide guidance for carrying out Test Method D2887 in an Method D2887
accelerated mode. Without instrumental adaptation, it is pos- Column 1 Column 2
sible to significantly reduce analysis times. The term acceler- Column length (m) 10 5
ated is used here to distinguish this technique from fast Column ID (mm) 0.53 0.53
Stationary phase thickness (µm)A 0.88 2.65
chromatography, which requires direct heating of the column Carrier gas helium helium
only. Carrier gas flow rate (mL/min) 26 35
Initial column temperature (°C) 40 40
X6.2 The development of gas chromatographic ovens, Final column temperature (°C) 360 350
Oven programming rate (°C/min) 35 35
which allow reproducible heating rates from 10 to 35ºC°C/min, Detector FID FID
allows the chromatography to be carried out in shorter periods Detector temperature (°C) 360 360
of time, by decreasing the run time from about 20 min to 8 to Injector cold on-column cold on-column
Injector initial temperature (°C) 100 100
9 min. Two sets of typical operating conditions are provided in Injector programming rate (°C/min) 35 35
Table X6.1. Injector final temperature (°C) 360 350
Sample size (µL) 0.1 0.1
X6.3 Figs. X6.1 and X6.2 show chromatograms of the Dilution concentration neat neat
Analysis time (minutes) 8 7.8
retention time standard analyzed at an oven-heating rate of A
All columns contain a polydimethylsiloxane stationary phase.
35ºC/min with Columns 2 and 1, respectively.
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
FIG. X6.3 Analysis of the Reference Gas Oil Under Accelerated Conditions (Column 2)
FIG. X6.4 Analysis of the Reference Gas Oil Under Accelerated Conditions (Column 1)
X7.1 Subcommittee D02.04 has identified the location of (2) Corrected Footnote A in Table 2.
selected changes to this standard since the last issue (3) Removed Column Resolution requirement.
(D2887–01a) that may impact the use of this standard. (4) Added Appendix X3, Calculation Algorithm.
X7.1.1 Changes: (5) Corrected rate of change of chromatographic signal in
(1) Corrected selected Fahrenheit boiling point values in Note 8.
Table 2 and added a note explaining the derivation of the (6) Modified paragraph 6.1.5 to address inlet pressure for
corrected values. open tubular columns.
REFERENCES
(1) Green, L. E., Schumauch, L. J., and Worman, J. C., Analytical Properties of Petroleum Products From Gas Chromatographic Analy-
Chemistry., Vol 32, 1960, p. 904. ses,” Ed. by L. E. Green and D. K. Albert, ASTM STP 577, ASTM
(2) Hickerson, J. F., ASTM STP 577M, ASTM International, 1973, p. 71. International, 1975, pp. 20-30.
(3) Green, L. E., Chromatograph Gives Boiling Point, Hydrocarbon (5) Kennard, C., “Correlated ASTM Distillation Distribution Based on
Processing, May, 1976. Simulated Distillation (ASTM D2887) Data,” Hewlett Packard, Avon-
(4) Ford, D. C., Miller, W. H., Thren, R. C., and Wetzler, R., “Correlation dale, PA, Application Note AN230-5, April 1979.
of ASTM D2887-73 Boiling Range Distribution Data with ASTM (6) API Technical Data Book for Petroleum Refining, Chapter 3, Petro-
Method D86-67D86 Distillation Data” In “Calculation of Physical leum Fraction Distillation Interconversions, 1999.
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---
Subcommittee D02.04.0H has identified the location of selected changes to this standard since the last issue
(D2887–06a) that may impact the use of this standard.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/
COPYRIGHT/).
--```,``,`,,``,````,``,,,``,,,`,-`-`,,`,,`,`,,`---