Service Manual

Tektronix 4000 Series


Digital Phosphor Oscilloscopes
071-2137-00

Revision A

This document applies to firmware version 2.00


and above.

Warning
The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to all safety summaries prior to
performing service.

www.tektronix.com
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or
suppliers, and are protected by national copyright laws and international treaty provisions.

Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.

TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.

TekLink is a trademark of Tektronix, Inc.

Tektronix is an authorized licensee of the CompactFlash® trademark.

Contacting Tektronix
Tektronix, Inc.
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA

For product information, sales, service, and technical support:


H In North America, call 1-800-833-9200.
H Worldwide, visit www.tektronix.com to find contacts in your area.
Table of Contents

General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii


Service Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix
Manual Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix
Operating Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1--1
Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2--1
Block Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-- 1
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-- 2
I/O Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-- 2
Main Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-- 2
Front-- Panel Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-- 3
Adjustment Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3--1
Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-- 1
Overview of the Adjustment Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-- 2
Factory Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-- 4
Completing the Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-- 8
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4--1
Preventing ESD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-- 1
Inspection and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-- 1
General Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-- 2
Interior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-- 2
Exterior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-- 2
Overview of Removal Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-- 6
Removal Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-- 8
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-- 12
Troubleshooting Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-- 13
Unpacking and Repacking Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-- 16
Mechanical Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5--1
Parts Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-- 1
Using the Replaceable Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-- 2

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual i


Table of Contents

List of Tables

Table 4--1: External inspection check list . . . . . . . . . . . . . . . . . . . . . . 4--3


Table 4--2: Internal inspection check list . . . . . . . . . . . . . . . . . . . . . . 4--4
List of Figures

Figure 2--1: Oscilloscope module interconnections . . . . . . . . . . . . . . . 2--1


Figure 3--1: Rear panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3--4
Figure 3--2: Utility page menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3--5
Figure 3--3: Calibration selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3--6
Figure 3--4: Factory information dialog . . . . . . . . . . . . . . . . . . . . . . . 3--6
Figure 3--5: Digital channel input . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3--10
Figure 4--1: Locator for trim and cabinet removal . . . . . . . . . . . . . . . 4--7
Figure 4--2: Locator for internal modules . . . . . . . . . . . . . . . . . . . . . . 4--8
Figure 4--3: Primary troubleshooting procedure . . . . . . . . . . . . . . . . 4--13
Figure 4--4: AC power supply troubleshooting procedure . . . . . . . . . 4--14
Figure 4--5: Module isolation troubleshooting procedure . . . . . . . . . 4--15
Figure 5--1: Exploded view, case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5--5
Figure 5--2: Exploded view: DPO4000 front panel and display,
4-channel instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5--7
Figure 5--3: Exploded view: MSO4000 front panel and display,
4-channel instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5--9
Figure 5--4: Exploded view: DPO4000 front panel and display,
2-channel instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5--11
Figure 5--5: Exploded view: MSO4000 front panel and display,
2-channel instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5--13
Figure 5--6: Exploded view: DPO4000 Power Supply
and I/O board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5--15
Figure 5--7: Exploded view: MSO4000 Power Supply
and I/O board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5--17
Figure 5--8: Exploded view: DPO4000 Main board,
Inverter board, and side fans . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5--19
Figure 5--9: Exploded view: MSO4000 Main board,
Inverter board, and side fans . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5--21

ii Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.

To Avoid Fire or Use Proper Power Cord. Use only the power cord specified for this product and
Personal Injury certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
The inputs are not rated for connection to mains or Category II, III, or IV
circuits.
Connect the probe reference lead to earth ground only.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual iii


General Safety Summary

Terms in this Manual These terms may appear in this manual:

WARNING. Warning statements identify conditions or practices that could result


in injury or loss of life.

CAUTION. Caution statements identify conditions or practices that could result in


damage to this product or other property.

Symbols and Terms These terms may appear on the product:


on the Product
H DANGER indicates an injury hazard immediately accessible as you read the
marking.
H WARNING indicates an injury hazard not immediately accessible as you
read the marking.
H CAUTION indicates a hazard to property including the product.

The following symbols may appear on the product:

CAUTION Protective Ground Mains Connected


Refer to Manual (Earth) Terminal Chassis Ground Push button OUT
ON (Power)

Push button
IN

iv Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Service Safety Summary

Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power.
Use Care When Servicing With Power On. Dangerous voltages or currents may
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual v


Service Safety Summary

vi Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Environmental Considerations
This section provides information about the environmental impact of the
product.

Product End-of-Life Observe the following guidelines when recycling an instrument or component:
Handling
Equipment Recycling. Production of this equipment required the extraction and
use of natural resources. The equipment may contain substances that could be
harmful to the environment or human health if improperly handled at the
product’s end of life. In order to avoid release of such substances into the
environment and to reduce the use of natural resources, we encourage you to
recycle this product in an appropriate system that will ensure that most of the
materials are reused or recycled appropriately.

The symbol shown to the left indicates that this product


complies with the European Union’s requirements
according to Directive 2002/96/EC on waste electrical and
electronic equipment (WEEE). For information about
recycling options, check the Support/Service section of the
Tektronix Web site (www.tektronix.com).

Mercury Notification. This product uses an LCD backlight lamp that contains
mercury. Disposal may be regulated due to environmental considerations. Please
contact your local authorities or, within the United States, the Electronics
Industries Alliance (www.eiae.org) for disposal or recycling information.

Restriction of Hazardous This product has been classified as Monitoring and Control equipment, and is
Substances outside the scope of the 2002/95/EC RoHS Directive. This product is known to
contain lead, cadmium, mercury, and hexavalent chromium.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual vii


Environmental Considerations

viii Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Preface

This service manual provides information that you need to troubleshoot,


disassemble, and replace parts on the Tektronix 4000 Series Digital Phosphor
Oscilloscopes.

Manual Conventions
This manual uses certain conventions that you should become familiar with
before attempting service.

Modules Throughout this manual, any replaceable component, assembly, or part is


referred to by the term module. A module is composed of electrical and
mechanical assemblies, circuit boards, interconnecting cables, and
user-accessible controls.

Replaceable Parts This manual refers to any field-replaceable assembly or mechanical part
specifically by its name or generically as a replaceable part. In general, a
replaceable part is any circuit board or assembly, such as the hard disk drive, or a
mechanical part, such as the I/O port connectors, that is listed in the replaceable
parts list.

Safety Symbols and terms related to safety appear in the General Safety Summary found
at the beginning of this manual.

Related Documentation

To read about Use these documents


Installation and Operation Tektronix 4000 Series Digital Phosphor Oscilloscopes User
Manual (available in 11 languages)
Specifications and Performance Tektronix 4000 Series Technical Reference (PDF only)
Programmer Commands Tektronix 4000 Series Programmer Manual (PDF only)
Analysis and Connectivity Tools Getting Started with OpenChoice® Solutions Manual

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual ix


Preface

To read about Use these documents


Installing and testing applica- Tektronix 4000 Series Application Modules Installation
tions modules Instructions Manual
Oscilloscope calibrator Fluke Oscilloscope Calibrator Manual at
http://us.fluke.com

x Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Operating Information
Operating Information

For information on installing and operating your DPO4000 or your MSO4000


Series Digital Phosphor Oscilloscope, refer to the Tektronix 4000 Series Digital
Phosphor Oscilloscopes User Manual.

This manual is available, in 11 languages, on the Web at www.tektronix.com.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 1- 1


Operating Information

1- 2 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Theory of Operation
Theory of Operation

This chapter describes the electrical operation of the oscilloscope to the module
level. Figure 2--1 shows the oscilloscope module interconnections.

Block Diagrams

Display
back light LCD
XGA Trigger Display
Video inverter
Out power
port

J140 J170 J680 J930


AC Line
Power Front Panel
Supply J100
I/O Board Input ports for
J0480 J1
J440 J610 application
CompactFlash J420, J421 modules
Port Main board
J120
J130
J570
USB J571 Bezel
J510 3 Fans J850 buttons
Attenuator circuit
Ethernet
port

Ch 1 Ch 2 Ch 3 Ch 4 Aux in 4-channel oscilloscopes


Ch 1 Ch 2 Aux in 2-channel oscilloscopes

Figure 2- 1: Oscilloscope module interconnections

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 2- 1


Theory of Operation

Power Supply
The Power Supply board converts AC line voltage to +12 V to power for all
internal circuits.

I/O Board
The I/O board contains USB ports, a CompactFlash port, an Ethernet port
(LAN), an XGA Video port, and a Trigger Out BNC connector.

Main Board
The Main board module contains the following functions:

Acquisition System The Acquisition system begins with the analog signal path and ends with a
digitized signal in memory. The signal enters a channel input, and then passes
through an attenuator and preamplifier. The analog signal from each preamplifier
goes through a digitizer, a time base controller, and then into acquisition
memory. The analog signal from each preamplifier is also distributed to a trigger
circuit.

Trigger System The Trigger system processes the analog signals from the preamplifiers and
sends trigger information to the time--base controller. Advanced trigger functions
are enabled only when the appropriate application modules and supporting
software are installed.

Display System The Display system combines live waveform data from acquisition memory with
menus and text, and stores this information in display memory. It then uses this
data to refresh the XGA display module (LCD).

2- 2 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Theory of Operation

Processor System The Processor system contains a 440EP Power PC microprocessor that controls
the entire instrument. The processor system also contains FLASH ROM, system
RAM, and interfaces to USB ports and a CompactFlash port.

Power Converter The Power Converter receives +12 V from the main power supply and generates
voltages for the analog and digital circuitry on the Main board.

Front- Panel Board


The Front Panel board contains a microprocessor that reads the front-panel
buttons and controls, and then sends this information to the processor system on
the Main board. The Front Panel board also generates the probe compensation
output signal and provides an interface to the application modules.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 2- 3


Theory of Operation

2- 4 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Adjustment Procedures
Adjustment Procedures

This chapter contains the factory adjustment procedure for the DPO4000 Series
and the MSO4000 Series oscilloscopes. Only qualified personnel should perform
adjustment procedures. Read the Service Safety Summary and the General Safety
Summary before performing any service procedures.

NOTE. The voltage references inside the oscilloscope are very stable over time
and should not require routine adjustment. Before performing any procedure in
this chapter, do the Performance Verification procedures to check whether the
oscilloscope meets specifications. See the Tektronix 4000 Series Technical
Reference documentation.
Successful completion of this adjustment procedure automatically updates the
stored time and date of the latest successful adjustment. Completion of the
Performance Verification procedure does not update this date and time. If no
adjustment is needed, there is also no need to update the time and date of the
adjustment.

Required Equipment
The following equipment, or a suitable equivalent, is required to complete these
procedures.
The following table specifies required equipment for the MSO4000 Series
oscilloscope.

Description Minimum requirements Example


DC voltage source 50 mV to 70 V, ±0.1% amplitude Fluke or Wavetek 9500
accuracy Oscilloscope Calibrator with
five 9530 active heads.
Digital calibrator probe Tektronix part number:
174-- 4535-- XX

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 3- 1


Adjustment Procedures

The following table specifies required equipment for the DPO4000 Series
oscilloscope.

Description Minimum requirements Example


DC voltage source 50 mV to 70 V, ±0.1% amplitude Fluke or Wavetek 9500
accuracy Oscilloscope Calibrator with
9530 active heads
Sine Generator 1kHz1 GHz Wavetek 9500 Oscilloscope
Calibrator with one 9510
Edge Generator 1 kHz with < 50 ps ch-- ch skew Output Module
50 Ω BNC cable BNC male to BNC male, about Tektronix part number
10 in (25 cm) long 012-- 0208-- 00
Adjustment tool Non-- conducting shaft with Tektronix part number
diameter = 0.1 in (= 2.5 mm) 003-- 1433-- 00

Overview of the Adjustment Process


Before performing adjustment procedures, you must warm up the oscilloscope
for at least 30 minutes in an ambient temperature between 20 °C and 30 °C.
Adjustments performed before warm-up or outside this temperature range may
result in poor performance or calibration failure.
The factory adjustment procedure consists of a series of steps; as you move
through these steps, the oscilloscope display provides instructions that describe
the specific input signal requirements for each step. If the oscilloscope passes the
step, it moves on to the next step. If the oscilloscope fails, you can repeat the
step or choose to abort the procedure.
If you have difficulty completing the steps, refer to the Completing the
Procedure section for specific instructions and troubleshooting information.
To complete the calibration procedure, you must know how to operate the
oscilloscope calibrator. Please refer to the user manual, which can be found at
http://us.fluke.com.

3- 2 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Adjustment Procedures

NOTE. Do not turn any knobs or push any front-panel buttons other than the
Next Step or Previous Step buttons during the adjustment procedure. Doing so
will cause the oscilloscope to abort the adjustment procedure. The oscilloscope
uses the previous calibration constants if the adjustment procedure is aborted or
fails.
The screen does not display the actual oscilloscope settings (such as channel
input impedance, or vertical and horizontal settings) during the adjustment
procedure. The oscilloscope automatically sets the instrument settings, but these
settings may not read out correctly on the display.
During some steps, the instrument may appear to be idle for several minutes
while it is processing information internally.

If the oscilloscope completes all steps in the procedure successfully, a “Pass”


message is displayed and the new calibration constants take effect. If the
oscilloscope does not pass a step, you can repeat the step. To cancel the
calibration procedure, press the Menu Off front panel button on the oscilloscope.
This reverts the oscilloscope to the old calibration constants.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 3- 3


Adjustment Procedures

Factory Adjustment Procedure


To perform the factory adjustment procedure, complete these steps:
1. Connect the oscilloscope to an AC power source.

NOTE. You must connect the oscilloscope and the test equipment to the same AC
power circuit. Connect the oscilloscope and test instruments to a common power
strip if you are unsure of the AC power circuit distribution. Connecting the
oscilloscope and test instruments to separate AC power circuits can result in
offset voltages between the equipment, which can invalidate the adjustment
procedure.

2. Power on the oscilloscope and warm it up for 30 minutes.


3. Power on the oscilloscope calibrator and warm it up for 30 minutes.

4. Connect the active heads from the calibrator to the oscilloscope, ensuring
that Channel 1 on the calibrator connects to Channel 1 on the oscilloscope,
and so on.

NOTE. For the DPO4000, there will be either 2 or 4 channels to connect. For the
MSO4000, there will be either 2 or 4 channels to connect, plus an additional
channel connection for the digital probe. Plug the digital probe into the
Channel 5 active head on the calibrator, but do not plug the probe into the
oscilloscope until you begin a Deskew test.

5. Power off the oscilloscope. Locate the hole in the rear panel that provides
access to the CAL switch, as shown in Figure 3--1.

CAL

Figure 3- 1: Rear panel

3- 4 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Adjustment Procedures

6. Insert the adjustment tool into the CAL switch hole. Use the tool to push and
hold the CAL switch in while you power on the oscilloscope. Keep holding
the switch until the oscilloscope shows the “Power-on self--test passed”
message with the oscilloscope graticule in the background. This can take up
to a minute.
7. Push the Default Setup button on the front panel of the oscilloscope.
8. Push the Utility button on the front panel of the oscilloscope.
9. Push the Utility Page lower--bezel button. The utility page menu appears, as
shown in Figure 3--2.

Figure 3- 2: Utility page menu

10. Use the Multipurpose a knob to select Calibration, as shown in


Figure 3--3.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 3- 5


Adjustment Procedures

Figure 3- 3: Calibration selection

11. Push the Factory lower--bezel button. The information dialog appears, as
shown in Figure 3--4.

Figure 3- 4: Factory information dialog

3- 6 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Adjustment Procedures

12. Push the OK Do Factory Calibration side--bezel button, as shown in


Figure 3--4, to start the adjustment process. Prompts appear on the oscillo-
scope screen to indicate the signal type and the channels to which it should
be connected.

NOTE. The oscilloscope adjusts itself automatically using the calibrator signal
as a reference. You do not need to make any adjustments.
If you make an error, such as connecting the wrong input signal, you can repeat
the last step by pushing the Return to Previous Screen side bezel button.

13. For each step, refer to the next table to identify the signal type, and then
carefully follow the instructions for that signal type on the specified page.

Signal type Example prompt Refer to Parameters to set


DC Voltage Apply 0 V DC 50 Ω ter- Page 3-- 8 Waveform type, voltage,
mination signal to 3. termination, channel
Time Mark Apply 500 mV Pk-- Pk 625 Page 3-- 9 Waveform type, voltage,
Hz 50 Ω termination frequency, termination,
time mark signal to 3. channel
Deskew Apply 1.0 V Pk-- Pk 1 kHz Page 3-- 10 Waveform type, voltage,
50 Ω termination deskew frequency, termination,
signal to 3. channel
Sine Signal Apply 2.0 V Pk-- Pk 164M Page 3-- 11 Waveform type, voltage,
50 Ω termination sine frequency, termination,
signal to 3. channel

NOTE. The oscilloscope screen will not provide you with all of the information
that you need to complete the steps successfully, so it is critical that you follow
the steps outlined in the Completing the Procedure section. Each time a new
prompt appears, refer to the table above to identify the signal type, and then
ensure that you are following every step that is outlined for that specific signal
type.

14. Continue with the adjustment process until it is complete and you receive the
Pass notification. You can cancel the process at any time by pushing the
MENU OFF button.
15. Complete the performance verification tests to verify that the adjustment
procedure has correctly calibrated the oscilloscope. (See the Tektronix 4000
Series Technical Reference.) Incorrect use of calibration equipment can cause
the oscilloscope to pass the Calibration Procedure but fail Performance
Verification.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 3- 7


Adjustment Procedures

Completing the Procedure


Instructions for completing the DC Voltage, Time Mark, Deskew and Sine Signal
tests are below. For each step in the calibration procedure, start at the beginning
of the numbered instructions for the specific test type, and carefully complete
each step.
If the oscilloscope fails a step, consult the Troubleshooting section at the end of
this chapter, revert to the previous step and try again. If the step fails again,
return to the first step in the calibration procedure and try the entire procedure
again. If the procedure fails, there could be problem with the oscilloscope.
Consider seeking customer support.

DC Voltage The DC Voltage test is the first test to appear on the prompts, and it usually
makes up the majority of the calibration procedure tests. For each DC Voltage
test, complete all 8 steps listed below to best ensure that the test will pass.
1. Push the DC/Square button.
2. Push the WAVEFORM lower--bezel button, and then push the DC Positive
right--bezel button.
3. Push the CHANNEL SELECT lower--bezel button.
4. Push the LOAD lower--bezel button to highlight either 1M Ω or 50 M Ω.
5. Select the specified channels, by using the side--bezel buttons, and then click
the EXIT lower--bezel button.
6. If the amplitude field is not selected, use the front panel scroll buttons. Type
the specified voltage, and then push the appropriate unit in the right--bezel
menu.
H For 0 V: ground the signal by pushing the grounding lower--
bezel button. When this is active, you will not be able to make a channel
selection, so in some cases you might need to temporarily deselect the
grounding option while you make a channel selection. If the prompt
requires all channels and the signal is grounded, The calibrator screen
will default to Channel 1.
H For any non--zero voltage: Before you select the voltage, you must first
1 2
5
1.Ø
ensure that the calibrator is in direct mode. Push the lower--bezel
button until 1.0 is selected.
H For 6 V or --6 V:
a. Push the AUX front panel button.

3- 8 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Adjustment Procedures

b. Push the side--bezel button for a 6 V signal or push the


side--bezel button for a --6 V signal. The polarity will be
indicated only by the icon that appears in the upper--left corner of the
calibrator screen.
c. Set the amplitude to either 6 V or --6 V.

d. Set the Pulse/Energy setting to 7.2 Joules (the time should show
10s).
e. When all of the parameters are set and you ensure that the output is
ON, push the Trig Pulse side--bezel and IMMEDIATELY push the
OK Do Next Step side--bezel button on the oscilloscope. WARN-
ING Once you push the Trig Pulse button on the calibrator, you
have only 9 seconds to push the OK Do Next Step button on the
oscilloscope. If the instrument fails one of these tests, go all the way
back to the first 6 V test (6 V on channel 1) and start that sequence
over again.
7. Ensure that the red light is illuminated to indicate that the output is on. If
not, push the ON button, as shown below:

8. On the oscilloscope, push the OK Do Next Step side--bezel button.

Time Mark The Time Mark test is often the second test type to appear in the prompts. When
you receive a time mark prompt, complete the following steps:
1. Push the Markers front--panel button.
2. Push the WAVEFORM lower--bezel button, and then push the
side--bezel button.

NOTE. If a 625 Hz signal is required, set the time marker to 1.6 ms.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 3- 9


Adjustment Procedures

3. Select the Amplitude field using the front panel scroll buttons, and then
change the voltage by using the large front panel knob on the calibrator.
4. Ensure that the red light is illuminated to indicate that the output is on. If
not, push the ON button.
5. On the oscilloscope, push the OK Do Next Step right--bezel button.

Deskew (MSO Only) If you are using the MSO4000 Series oscilloscope, and you recieve a prompt for
a deskew test, complete the following steps:

1. Plug the probe, notch down, into the digital channel input, as shown in
Figure 3--5. Ensure that is snaps into place. Keep the probe plugged in for
the remainder of the calibration procedure.

Measure Search Test

Cursors

Select Fine Acquire Run / Stop

Intensity
Single

Set / Clear Autoset

M Menu

B1 1 2 3 4

B2 Set to 50%

B3
ForceTrig

B4
Menu
Off

D15 - D0
CompactFlash
Save Menu Default Utility
Setup

Figure 3- 5: Digital channel input

2. Push the Aux front--panel button, and then push the Zero skew
side--bezel button.
3. Set the specified voltage, frequency, and termination.
4. Push the CHANNEL SELECT button and ensure that all channels are
selected, including Channel 5.

5. Ensure that the red light is illuminated to indicate that the output is on. If
not, push the ON button.
6. On the oscilloscope, push the OK Do Next Step right--bezel button. In some
cases, this can take several minutes.

3- 10 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Adjustment Procedures

CAUTION. Keep the digital probe plugged in to the oscilloscope. Unplugging the
probe may result in aborting the test.

Sine Signal When a Sine Signal prompt appears, complete the following steps:
1. Push the Sine front panel button.
1 2
5
1.Ø
2. To ensure that the voltage is in direct mode, push the lower--bezel
button and select 1.0.
3. Push the CHANNEL SELECT lower--bezel button, and then ensure that
Channel 5 is selected.
4. Ensure that the red light is illuminated to indicate that the output is on. If
not, push the ON button.
5. On the oscilloscope, push the OK Do Next Step right--bezel button.

Troubleshooting Refer to the next table for common issues encountered during the calibration
procedure.

Problem Things to try


The instrument fails a test Are the channels hooked up properly, routed to the right
channels, and turned on?
Is the signal set to the correct waveform, frequency, and
termination (if applicable to the test)?
Is the output set to ON (the red light will be illuminated)?
If you make a correction or find that everything appears to be
correct, click the Go back a step button to try the test again.
I can’t set the amplitude to 0 V, Ground the signal by pushing the lower-- bezel
as specified in the prompt button.
I can’t find the parameters that I Is the waveform set correctly?
need to set
Is the grounding button on? When this is active, you will not
be able to make a channel selection, so in some cases you
might need to temporarily deselect the grounding option
while you make a channel selection.
If you can find the termination settings, click the CHANNEL
SELECT lower-- bezel button.
I can’t set the voltage to 625 Hz Set the time marker to 1.6 Ms. This will automatically set the
voltage to 625 Hz.
I can’t set the signal to all If you are unable to set the signal to all channels, try setting
channels the signal to Channel 1.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 3- 11


Adjustment Procedures

Problem Things to try


I’m having trouble with tests that Follow the steps for 6 V and - 6 V that appear in Step 6 of
require 6 V or - 6 V the DC Voltage test.
Ensure that the polarity is correct, by referring to the icon
that appears in the upper-- left corner of the calibrator screen.
If the instrument fails a test, go all the way back to the first 6
V test (6 V on Channel 1) and start the sequence over again.
The Sine Signal test failed If the prompt specified Channel 4, try setting the signal to
Channel 5, instead.

3- 12 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Maintenance
Maintenance

This section contains the information needed to do periodic and corrective


maintenance on the oscilloscrope, as well as repackaging instructions if you need
to return the oscilloscope to Tektronix for service.

Preventing ESD
Before servicing this product, read the Safety Summary and Introduction at the
front of the manual and the electrostatic discharge (ESD) information below.

CAUTION. Static discharge can damage any semiconductor component in this


oscilloscope

When performing any service that requires internal access to the oscilloscope,
adhere to the following precautions to avoid damaging internal modules and their
components due to electrostatic discharge.
1. Minimize handling of static-sensitive circuit boards and components.

2. Transport and store static-sensitive modules in their static protected


containers or on a metal rail. Label any package that contains static-sensitive
boards.
3. Discharge the static voltage from your body by wearing a grounded antistatic
wrist strap while handling these modules. Service static-sensitive modules
only at a static-free work station.
4. Do not place anything capable of generating or holding a static charge on the
work station surface.
5. Handle circuit boards by the edges when possible.
6. Do not slide the circuit boards over any surface.
7. Avoid handling circuit boards in areas that have a floor or work-surface
covering capable of generating a static charge.

Inspection and Cleaning


Inspection and cleaning are done as preventive maintenance. Preventive
maintenance, when done regularly, may prevent oscilloscope malfunction and
enhance its reliability.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 4- 1


Maintenance

Preventive maintenance consists of visually inspecting and cleaning the


oscilloscope and using general care when operating it.
How often you do maintenance depends on the severity of the environment in
which the oscilloscope is used. A proper time to perform preventive maintenance
is just before oscilloscope adjustment.

General Care The cabinet helps keep dust out of the oscilloscope and should normally be in
place when operating the oscilloscope.

Flat Panel Display The display is a soft plastic display and must be treated with care during
Cleaning cleaning.

CAUTION. Improper cleaning agents or methods can damage the flat panel
display.
Avoid using abrasive cleaners or commercial glass cleaners to clean the display
surface.
Avoid spraying liquids directly on the display surface.
Avoid scrubbing the display with excessive force.

Clean the flat panel display surface by gently rubbing the display with a
clean-room wipe (such as Wypall Medium Duty Wipes, #05701, available from
Kimberly-Clark Corporation).

Interior Cleaning Use a dry, low-velocity stream of air to clean the interior of the chassis. Use a
soft-bristle, non-static-producing brush for cleaning around components. If you
must use a liquid for minor interior cleaning, use a 75% isopropyl alcohol
solution and rinse with deionized water.

WARNING. To avoid injury, power down the instrument and disconnect it from
line voltage before performing any procedure that follows.

Exterior Cleaning Clean the exterior surfaces of the chassis with a dry lint-free cloth or a soft-
bristle brush. If any dirt remains, use a cloth or swab dipped in a 75% isopropyl
alcohol solution. Use a swab to clean narrow spaces around controls and
connectors. Do not use abrasive compounds on any part of the chassis that may
damage the chassis.

4- 2 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Maintenance

Clean the On/Standby switch using a dampened cleaning towel. Do not spray or
wet the switch directly.

CAUTION. Avoid the use of chemical cleaning agents, which might damage the
plastics used in this oscilloscope. Use only deionized water when cleaning the
menu buttons or front-panel buttons. Use a 75% isopropyl alcohol solution as a
cleaner and rinse with deionized water. Before using any other type of cleaner,
consult your Tektronix Service Center or representative.

Lubrication. There is no periodic lubrication required for this oscilloscope.

Exterior Inspection Inspect the outside of the oscilloscope for damage, wear, and missing parts,
using Table 4--1 as a guide. Immediately repair defects that could cause personal
injury or lead to further damage to the oscilloscope.

Table 4- 1: External inspection check list

Item Inspect for Repair action


Cabinet, front panel, Cracks, scratches, deformations, Repair or replace defective
and cover damaged hardware. module.
Front-panel knobs Missing, damaged, or loose Repair or replace missing or
knobs. defective knobs.
Connectors Broken shells, cracked insulation, Repair or replace defective
and deformed contacts. Dirt in modules. Clear or wash out dirt.
connectors.
Carrying handle, and Correct operation. Repair or replace defective
cabinet feet module.
Accessories Missing items or parts of items, Repair or replace damaged or
bent pins, broken or frayed missing items, frayed cables, and
cables, and damaged defective modules.
connectors.

If the display is very dirty, moisten the wipe with distilled water or a 75%
isopropyl alcohol solution and gently rub the display surface. Avoid using excess
force or you may damage the plastic display surface.

CAUTION. To prevent getting moisture inside the oscilloscope during external


cleaning, use only enough liquid to dampen the cloth or applicator.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 4- 3


Maintenance

Interior Inspection To access the inside of the oscilloscope for inspection and cleaning, refer to the
Removal Procedures in this section.
Inspect the internal portions of theoscilloscope for damage and wear, using Table
4--2 as a guide. Repair any defects immediately.
If any circuit board is repaired or replaced, check Table 4--2 to see if it is
necessary to adjust the oscilloscope.

CAUTION. To prevent damage from electrical arcing, ensure that circuit boards
and components are dry before applying power to the oscilloscope.

Table 4- 2: Internal inspection check list

Item Inspect for Repair action


Circuit boards Loose, broken, or corroded Remove and replace damaged
solder connections. Burned circuit board.
circuit boards. Burned, broken, or
cracked circuit-run plating.
Resistors Burned, cracked, broken, blis- Remove and replace damaged
tered condition. circuit board.
Solder connections Cold solder or rosin joints. Resolder joint and clean with
isopropyl alcohol.
Capacitors Damaged or leaking cases. Remove and replace damaged
Corroded solder on leads or circuit board.
terminals.
Wiring and cables Loose plugs or connectors. Firmly seat connectors. Repair or
Burned, broken, or frayed wiring. replace modules with defective
wires or cables.
Chassis Dents, deformations, and dam- Straighten, repair, or replace
aged hardware. defective hardware.

4- 4 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Maintenance

Cleaning Procedure — Interior. To clean the oscilloscope interior, do the following


steps:
1. Blow off dust with dry, low-pressure, deionized air (approximately 9 psi).
2. Remove any remaining dust with a lint-free cloth dampened in isopropyl
alcohol (75% solution) and rinsed with warm deionized water. (A cotton-
tipped applicator is useful for cleaning in narrow spaces and on circuit
boards.)

STOP. If, after doing steps 1 and 2, a module is clean upon inspection, skip the
following steps.

If there is still dust or dirt on the module, the oscilloscope may be spray washed
using a solution of 75% isopropyl alcohol by doing the following steps:
1. Gain access to the parts to be cleaned by removing easily accessible shields
and panels (see Removal Procedures).
2. Spray wash dirty parts with the isopropyl alcohol and wait 60 seconds for the
majority of the alcohol to evaporate.
3. Use hot (120 _F to 140 _F) deionized water to thoroughly rinse them.
4. Dry all parts with low-pressure, deionized air.
5. Dry all components and assemblies in an oven or drying compartment using
low-temperature (125 _F to 150 _F) circulating air.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 4- 5


Maintenance

Overview of Removal Procedures


Refer to Figures 4--1 and 4--2 for the location of modules that you are removing
or installing. If you are disassembling the instrument for cleaning, refer to the
Inspection and Cleaning procedure on page 4--1 for instructions.

List of Modules The Mechanical Parts List chapter provides a list of all replaceable modules.
Any replaceable component, assembly, or part is referred to as a module.

WARNING. Only qualified personnel should perform service procedures. Before


doing this or any other procedure in this manual, read the General Safety
Summary and Service Safety Summary located at the beginning of this manual.
Also, to prevent possible injury to service personnel or damage to electrical
components, read Preventing ESD on page 4--1.

Tools Required You will need the following tools to remove and replace all the modules in the
instrument.
H Torque-limiting screwdriver, 12 in--lb (1.3 N--m) range with Torx® T-15
tips. If you are using a magnetic screwdriver with interchangeable bits, the
T-15 bit must be a long bit, at least 5 cm (2 in) overall length.
H Screwdriver with 1@8 inch flat blade

H 3@ inch nut driver


16

H Needle-nose pliers

Trim, Cabinet, and Module Use the following tables to remove the trim, cabinet, and internal modules.
Removal
WARNING. Before doing any procedure in this subsection, disconnect the power
cord from the line voltage source. Failure to do so could cause serious injury or
death.

4- 6 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Maintenance

Rear case

Front foot
Front cover

Figure 4- 1: Locator for trim and cabinet removal

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 4- 7


Maintenance

Power supply
assembly
Inverter board
assembly Side fans
Display assembly
assembly

I/O board
assembly
Main board
Front Case
assembly
Front Panel
assembly

Figure 4- 2: Locator for internal modules

Removal Procedures
These procedures require that you have access to the module that you are
removing. Refer to the exploded drawings, Figures 5--1 through 5--9 on pages
5--5 through 5--21, and to Figures 4--1 and 4--2 on pages 4--7 and 4--8 to
determine which trim and/or modules to remove.

CAUTION. When removing or installing the keypad, do not touch the switch
contacts with your fingers. The oils on your fingers will degrade or damage the
switch contacts. To help prevent damage to the keypad, use cotton gloves when
removing or installing the keyboard pad.

4- 8 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Maintenance

CAUTION. To avoid damage to the Front Panel Standby/On switch assembly, do


not set the Display module assembly on a work surface. Sliding the oscilloscope
over the edge of the work surface could break off the On/Standby switch
assembly.

Rear Case Follow these steps to remove the Rear Case assembly. Use a Torx T-15 screw-
driver tip. See Figure 5--1 on page 5--5.
1. Remove the screw from the back and sides of the Rear Case assembly.
2. Grasp the case and pull outwards to remove it.

Power Supply Module Follow these steps to remove the Power Supply module. You need to have
previously removed the Rear Case assembly. Use a Torx T-15 screwdriver tip.
See Figure 5--6 on page 5--15.
1. Remove the 3 screws that secure the Power Supply module to the chassis.
2. Disconnect the power supply fan and connectors from the I/O board and the
Main board.
3. Lift the Power Supply module off the chassis.

Power Supply Fan. Follow these steps to remove the Power Supply fan assembly.
You need to have previously removed the Rear Case assembly and Power Supply
assembly. Use a Torx T-15 and a P2 screwdriver tip. See Figure 5--6 on page
5--15.
1. Remove the 3 screws that attach the power supply cover.
2. Lift the Power Supply cover off.
3. Use the P2 screwdriver tip and remove the 2 screws that attach the fan to the
cover.

Power Supply Board. Follow these steps to remove the Power Supply board. You
need to have previously removed the Rear Case assembly and Power Supply
assembly. Use a Torx T-15 screwdriver tip. See Figure 5--6 on page 5--15.

1. Remove the 3 screws that attach the Power Supply cover.


2. Lift off the Power Supply cover.
3. Remove the 4 screws that attach the Power Supply board to the Power
Supply bracket.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 4- 9


Maintenance

I/O Board Follow these steps to remove the I/O board. You need to have previously
removed the Rear Case assembly and Power Supply module. Use a Torx T-15
screwdriver tip. See Figure 5--6 on page 5--15.
1. Remove the 5 screws that hold the I/O board to the chassis.
2. Lift straight up to remove.

Main Fans Follow these steps to remove the side fan assembly. You need to have previously
removed the Rear Case assembly, Power Supply assembly, and I/O board. Use a
Torx T-15 screwdriver tip. See Figure 5--8 on page 5--19.
1. Remove the 6 screws that hold the side fans to the chassis.
2. Disconnect the two fan connectors from the Main board.
3. Slide the fan assembly out and away from the chassis.

Main Board Follow these steps to remove the Main board. You need to have previously
removed the Rear Case assembly, power supply, I/O board, and side fans. Use a
Torx T-15 screwdriver tip. See Figure 5--8 on page 5--19.
1. Disconnect the display cable from the Main board.
2. Disconnect the backlight cable from the Inverter board.
3. Remove 13 screws from the Main board that connect it to the chassis.
4. Lift the Main board up to remove.

Inverter board. Follow these steps to remove the Inverter board from the Main
board. You must have previously removed the Main board. Use a Torx T-15
screwdriver tip. See Figure 5--8 on page 5--19.
1. Disconnect the Inverter board cable from the Main board.
2. Remove the Inverter board from the plastic bracket.
3. Remove the 2 screws that hold the Inverter board bracket to the Main board.

4. Remove the Inverter board bracket from the Main board.

4- 10 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Maintenance

Front Case Assembly Follow these steps to remove the Front Case assembly. You must have previous-
ly removed the Rear Case assembly. See Figure 5--2 on page 5--7.
1. Unattach 8 snaps (3 top, 3 bottom, 1 left and 1 right). Work from a corner,
and keep tension on the case while moving to the next set of snaps
2. Remove the Front Case by pulling back the board snaps and lifting the lower
right corner of the Front Case.
3. Place the oscilloscope face up on a soft surface (such as an anti-static mat),
with the bottom facing you.
4. Disconnect the bezel button flex circuit by gently pulling the flex circuit out
of the connector on the Front Panel board.

Display Module Follow these steps to remove the display module. You need to have previously
removed the power supply assembly, I/O board, and Front Case assembly. Use a
Torx T-15 screwdriver tip. See Figure 5--2 on page 5--7.
1. Disconnect the display power cable from the Main board and the backlight
cable from the Inverter board.
2. Remove the 4 screws that hold the display to the front chassis.

3. Gently lift the display up and off of the chassis.

CAUTION. Be careful when removing and reinstalling the Display module cables.
If the connectors have bent pins or are installed incorrectly, the Display may be
destroyed.

Front Panel Module Follow these steps to remove the Front Panel module. You need to have
previously removed the Front Case assembly. Use a Torx T-15 screwdriver tip.
See Figure 5--2 on page 5--7.
1. Remove the 9 screws holding the Front Panel board to the chassis.
2. Lift the Front Panel board off of the chassis.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 4- 11


Maintenance

Troubleshooting
This section contains information and procedures to help you isolate a defective
module in the DPO4000 series or the MSO4000 series oscilloscopes.

WARNING. Before doing this or any other procedure in this manual, read the
General Safety Summary and Service Safety Summary found at the beginning of
this manual. Also, to prevent possible injury to service personnel or damage to
electrical components, read Preventing ESD on page 4--1.

Adjustment After Repair If you replace the Main board, you must adjust the instrument after repair. Refer
to the Adjustment Procedures section, beginning on page 3--1, for information
about the adjustment.

Required Tools and You need the following equipment to troubleshoot the instrument.
Equipment
Tools and Equipment Example
DMM 3.5 or above

4- 12 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Maintenance

Troubleshooting Procedure
Figures 4--3 through 4--5 are troubleshooting procedure flowcharts. Use them to
troubleshoot an instrument failure. Begin with Figure 4--3.

WARNING. Before removing the cabinet, disconnect the power cord from the line
voltage source. Failure to do so could cause serious injury or death.

Power on the
instrument.

Does
the instrument No Perform the AC Power Supply
display the power-up troubleshooting procedure
screen? (Figure 4-- 4).

Yes

Does Perform the Module Isolation


the instrument pass No troubleshooting procedure
all the (Figure 4-- 5).
diagnostics?

Yes

Done.

Figure 4- 3: Primary troubleshooting procedure

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 4- 13


Maintenance

Continued from
Figure 4-- 3.

¯ Power off oscilloscope.


¯ Disconnect AC power cord.
¯ Remove instrument case.
¯ Check that all cables are
connected and fully seated into
board connectors.

Are all cables No ¯ Reconnect loose cables.


connected? ¯ Return to top of Figure 4-- 3.

Yes

Connect AC power cord.

¯ Measure for +12 V DC


(± 0.3 V) on the connector pins.
(See figure at right)
WARNING!
Be careful when measuring near
the AC power supply; there are
line voltages present on exposed
parts.

Are there
+12 volts on the No Replace the AC Power
connector Supply module.
pins ?
Yes
AC Power Supply module
OK. Return the MSO4000
oscilloscope to the nearest
Tektronix Service Center for
repair.

Figure 4- 4: AC power supply troubleshooting procedure

4- 14 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Maintenance

Continued from
Figure 4-- 3.

From the oscilloscope


power-up diagnostics
message display.

Does
the power-up No
Oscilloscope is operating OK.
screen report any
errors?

Yes

¯ Push Utility.
¯ Push Utility page and then
select Self Test.
¯ Push OK Run Self Test.
¯ Determine failure from
Self Test Results.
¯ Use table to select module
to replace.

Failed Test Most Likely Cause

Acquisition Main board


CPU Main board
Display Main board
Front Panel Front Panel (or Main board)
I/O Comm module (or Main board)
ROM Main board
App Modules Front Panel (or App Keys)

Figure 4- 5: Module isolation troubleshooting procedure

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 4- 15


Maintenance

Unpacking and Repacking Instructions


This section contains the information needed to unpack the oscilloscope and
repack it for shipment or storage.

Unpacking The oscilloscope and its standard accessories are carefully packed at the factory
in a shipping carton. If, upon receipt, damage to the shipping carton is evident,
notify the shipper. Tektronix, Inc. is not responsible for damage caused during
shipping.
If you have not already done so, carefully remove the oscilloscope and its
accessories from the shipping carton and inspect them for damage. Save the
shipping carton for repacking or storage.

Repacking Use a corrugated cardboard shipping carton having a test strength of at least
275 pounds (125 kg) and with an inside dimension at least six inches (15.25 cm)
greater than the instrument dimensions.
If the instrument is being shipped to a Tektronix Service Center, enclose the
following information:
H The owner’s address
H Name and phone number of a contact person
H Type and serial number of the instrument
H Reason for returning
H A complete description of the service required
Seal the shipping carton with an industrial stapler or strapping tape.
Mark the address of the Tektronix Service Center and also your own return
address on the shipping carton in two prominent locations.
See www.tektronix.com/service to find a service center near you.

Storage The oscilloscope should be stored in a clean, dry environment. The following
environmental characteristics apply for both shipping and storage:

H Temperature range: --20 °C to +60 °C


H Altitude: To 15,000 m
See the Tektronix 4000 Series Technical Reference for a complete listing of the
environmental characteristics.

4- 16 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Mechanical Parts List
Mechanical Parts List

This chapter contains a list of the replaceable modules for the DPO4000 and
MSO4000 Series Oscilloscopes. Use this list to identify and order replacement
parts.

Parts Ordering Information


Replacement parts are available through your local Tektronix field office or
representative.

Changes to Tektronix products are sometimes made to accommodate improved


components as they become available and to give you the benefit of the latest
improvements. Therefore, when ordering parts, it is important to include the
following information in your order.
H Part number (see Part Number Revision Level below)
H Instrument type or model number
H Instrument serial number
H Instrument modification number, if applicable
If you order a part that has been replaced with a different or improved part, your
local Tektronix field office or representative will contact you concerning any
change in part number.
Change information, if any, is located at the rear of this manual.

Part Number Revision Tektronix part numbers contain two digits that show the revision level of the
Level part. For some parts in this manual, you will find the letters XX in place of the
revision level number.

Part Number Revision Level Revision Level May Show as XX

670-7918-03 670-7918-XX

When you order parts, Tektronix will provide you with the most current part for
your product type, serial number, and modification (if applicable). At the time of
your order, Tektronix will determine the part number revision level needed for
your product, based on the information you provide.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 5- 1


Mechanical Parts List

Module Servicing Modules can be serviced by selecting one of the next three options. Contact your
local Tektronix service center or representative for repair assistance.

Module Exchange. In some cases you may exchange your module for a reman-
ufactured module. These modules cost significantly less than new modules and
meet the same factory specifications. For more information about the module
exchange program, call 1-800-833--9200, extension 2.

Module Repair and Return. You may ship your module to us for repair, after which
we will return it to you.

New Modules. You may purchase replacement modules in the same way as other
replacement parts.

Using the Replaceable Parts List


This section contains a list of the mechanical and/or electrical components that
are replaceable for the instrument. Use this list to identify and order replacement
parts. The following table describes each column in the parts list.

Parts List Column Descriptions

Column Column Name Description


1 Figure & Index Number Items in this section are referenced by figure and index numbers to the exploded view
illustrations that precede the list.
2 Tektronix Part Number Use this part number when ordering replacement parts from Tektronix.
3 and 4 Serial Number Column three indicates the serial number at which the part was first effective. Column four
indicates the serial number at which the part was discontinued. No entries indicates the part is
good for all serial numbers.
5 Qty This indicates the quantity of parts used.
6 Name & Description An item name is separated from the description by a colon (:). Because of space limitations, an
item name may sometimes appear as incomplete. Use the U.S. Federal Catalog handbook
H6-1 for further item name identification.

Abbreviations Abbreviations conform to American National Standard ANSI Y1.1--1972.

5- 2 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Mechanical Parts List

Exploded Views Figures 5--1 through 5--8 on the following pages show the module-level exploded
views of the DPO4000 Series and the MSO4000 Series oscilloscopes. Each
exploded view is indexed by the numbers in the figure.

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 5- 3


Mechanical Parts List

Replaceable Parts List: DPO4000 and MSO4000 case

Fig. &
index Tektronix part Serial no. Serial no.
number number effective discont’d Qty Name & description
5-- 1 Exploded View, Case
-1 200-- 4908-- 00 1 COVER, FRONT, PROTECTIVE
-2 202-- 0441-- 00 1 CASE, REAR ASSEMBLY
-3 211-- 1174-- 00 4 SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING
MATERIAL; 6-- 32 X 0.312 L,PNH,STL CAD PLT,T15
-4 335-- 1186-- 00 1 MARKER,IDENT;LABEL,REAR PANEL
-5 348-- 1798-- 00 2 FOOT, FRONT ASSEMBLY

5- 4 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Mechanical Parts List

Figure 5- 1: Exploded view, case

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 5- 5


Mechanical Parts List

Replaceable Parts List: DPO4000 front panel and display, 4 channel

Fig. &
index Tektronix part Serial no. Serial no.
number number effective discont’d Qty Name & description
5-- 2 DPO4000 Front Panel and Display, 4 Channel
-1 202-- 0432-- 00 1 FRONT CASE ASSEMBLY
-2 335-- 1187-- 00 1 LABEL. FRONT PANEL, MODEL DPO4054
335-- 1191-- 00 1 LABEL. FRONT PANEL, MODEL DPO4104
335-- 1193-- 00 1 LABEL. FRONT PANEL, MODEL DPO4034
-3 366-- 0860-- 01 5 KNOB, .685 DIA, SOFT TOUCH
-4 366-- 0859-- 01 6 KNOB, .470 DIA, SOFT TOUCH
-5 366-- 0861-- 01 2 KNOB, .925 DIA, SOFT TOUCH
-6 366-- 0865-- 00 1 KNOB, JOG
-7 358-- 0883-- 00 1 BUSHING (JOG SHUTTLE KNOB)
-8 366-- 0867-- 01 1 KNOB, SHUTTLE
-9 335-- 1467-- 00 1 LABEL. FRONT PANEL, OVERLAY, RUSSIAN
335-- 1466-- 00 1 LABEL. FRONT PANEL, OVERLAY, KOREAN
335-- 1465-- 00 1 LABEL. FRONT PANEL, OVERLAY, STD CHINESE
335-- 1464-- 00 1 LABEL. FRONT PANEL, OVERLAY, SIM CHINESE
335-- 1463-- 00 1 LABEL. FRONT PANEL, OVERLAY, PORTUGUESE
335-- 1462-- 00 1 LABEL. FRONT PANEL, OVERLAY, JAPANESE
335-- 1461-- 00 1 LABEL. FRONT PANEL, OVERLAY, SPANISH
335-- 1460-- 00 1 LABEL. FRONT PANEL, OVERLAY, GERMAN
335-- 1459-- 00 1 LABEL. FRONT PANEL, OVERLAY, ITALIAN
335-- 1458-- 00 1 LABEL. FRONT PANEL, OVERLAY, FRENCH
10 335-- 1183-- 00 1 LABEL. FRONT PANEL, 4CH
11 200-- 4418-- 00 1 COVER; OPTION KEY DOOR,0.7270 X 1.2250 X 0.310,PC/
ABS,FR110,SILVER GRAY
12 260-- 2813-- 01 1 SWITCH,KEYPAD;ELASTOMERIC FRONT PANEL
13 211-- 1174-- 00 9 SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING
MATERIAL; 6-- 32 X 0.312 L,PNH,STL CAD PLT,T15
14 872-- 0005-- 00 1 FRONT PANEL BOARD, 4 CH
15 335-- 1194-- 00 1 MARKER, FRONT PANEL, I/O
16 348-- 1860-- 00 2 GASKET, ELEC; CONDUCTIVE URETHANE FOAM
17 441-- 2378-- 00 1 CHASSIS;FRONT; SAFETY CONTROLLED
18 650-- 4715-- 01 1 ASSEMBLY, LCD DISPLAY
19 174-- 5014-- 00 1 CABLE ASSY,SP;DISPLAY DATA,FLAT FLEX
20 211-- 1174-- 00 4 SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING
MATERIAL; 6-- 32 X 0.312 L,PNH,STL CAD PLT,T15

5- 6 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Mechanical Parts List

5 6

4 7

3 8

9
2

1 10

15

Flex Circuit

J13
11

20
13 12

13

19
14

16

18 17

Figure 5- 2: Exploded view: DPO4000 front panel and display, 4-channel instruments

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 5- 7


Mechanical Parts List

Replaceable Parts List: MSO4000 front panel and display, 4 channel

Fig. &
index Tektronix part Serial no. Serial no.
number number effective discont’d Qty Name & description
5-- 3 MSO 4000 Front Panel and Display, 4 Channel
-1 202-- 0414-- 00 1 FRONT CASE ASSEMBLY
-2 335-- 1642-- 00 1 LABEL. FRONT PANEL, MODEL DPO4054
335-- 1643-- 00 1 LABEL. FRONT PANEL, MODEL DPO4104
335-- 1641-- 00 1 LABEL. FRONT PANEL, MODEL DPO4034
-3 366-- 0860-- 01 5 KNOB, .685 DIA, SOFT TOUCH
-4 366-- 0859-- 01 6 KNOB, .470 DIA, SOFT TOUCH
-5 366-- 0861-- 01 2 KNOB, .925 DIA, SOFT TOUCH
-6 366-- 0865-- 00 1 KNOB, JOG
-7 358-- 0883-- 00 1 BUSHING (JOG SHUTTLE KNOB)
-8 366-- 0867-- 01 1 KNOB, SHUTTLE
-9 335-- 1653-- 00 1 LABEL. FRONT PANEL, OVERLAY, RUSSIAN
335-- 1652-- 00 1 LABEL. FRONT PANEL, OVERLAY, KOREAN
335-- 1651-- 00 1 LABEL. FRONT PANEL, OVERLAY, STD CHINESE
335-- 1650-- 00 1 LABEL. FRONT PANEL, OVERLAY, SIM CHINESE
335-- 1649-- 00 1 LABEL. FRONT PANEL, OVERLAY, PORTUGUESE
335-- 1648-- 00 1 LABEL. FRONT PANEL, OVERLAY, JAPANESE
335-- 1647-- 00 1 LABEL. FRONT PANEL, OVERLAY, SPANISH
335-- 1646-- 00 1 LABEL. FRONT PANEL, OVERLAY, GERMAN
335-- 1645-- 00 1 LABEL. FRONT PANEL, OVERLAY, ITALIAN
335-- 1644-- 00 1 LABEL. FRONT PANEL, OVERLAY, FRENCH
10 335-- 1395-- 00 1 LABEL. FRONT PANEL, 4CH
11 200-- 4418-- 00 1 COVER; OPTION KEY DOOR,0.7270 X 1.2250 X 0.310,PC/
ABS,FR110,SILVER GRAY
12 260-- 2813-- 01 1 SWITCH,KEYPAD;ELASTOMERIC FRONT PANEL
13 211-- 1174-- 00 9 SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING
MATERIAL; 6-- 32 X 0.312 L,PNH,STL CAD PLT,T15
14 872-- 0005-- 00 1 FRONT PANEL BOARD, 4 CH
15 335-- 1359-- 00 1 MARKER, FRONT PANEL, I/O
16 348-- 1860-- 00 2 GASKET, ELEC; CONDUCTIVE URETHANE FOAM
17 441-- 2378-- 00 1 CHASSIS;FRONT; SAFETY CONTROLLED
18 650-- 4715-- 01 1 ASSEMBLY, LCD DISPLAY
19 174-- 5014-- 00 1 CABLE ASSY,SP;DISPLAY DATA,FLAT FLEX
20 211-- 1174-- 00 4 SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING
MATERIAL; 6-- 32 X 0.312 L,PNH,STL CAD PLT,T15

5- 8 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Mechanical Parts List

5 6

4 7

3 8

9
2

1 10

15

Flex Circuit

J13
11

20 12
13

13

19
14

16

18 17

Figure 5- 3: Exploded view: MSO4000 front panel and display, 4-channel instruments

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 5- 9


Mechanical Parts List

Replaceable Parts List: DPO4000 front panel and display, 2 channel

Fig. &
index Tektronix part Serial no. Serial no.
number number effective discont’d Qty Name & description
5-- 4 DPO4000 Front Panel and Display, 2 Channel
-1 335-- 1189-- 00 1 LABEL. FRONT PANEL, MODEL DPO4032
-2 335-- 1185-- 00 1 MARKER,IDENT;LABEL,FRONT PANEL,2CH
-3 260-- 2812-- 01 1 SWITCH,KEYPAD;ELASTOMERIC FRONT PANEL
-4 872-- 0004-- 00 1 CIRCUIT BOARD; FRONT PANEL, 2 CHANNEL
-5 200-- 4906-- 00 2 COVER;BNC FILLER
-6 366-- 0860-- 01 3 KNOB, .685 DIA, SOFT TOUCH
-7 366-- 0859-- 01 4 KNOB, .470 DIA, SOFT TOUCH

5- 10 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Mechanical Parts List

6
7
2

Figure 5- 4: Exploded view: DPO4000 front panel and display, 2-channel instruments

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 5- 11


Mechanical Parts List

Replaceable Parts List: MSO4000 front panel and display, 2 channel

Fig. &
index Tektronix part Serial no. Serial no.
number number effective discont’d Qty Name & description
5-- 5 MSO4000 Front Panel and Display, 2 Channel
-1 335-- 1640-- 00 1 LABEL. FRONT PANEL, MODEL MSO4032
-2 335-- 1397-- 00 1 MARKER,IDENT;LABEL,FRONT PANEL,2CH
-3 260-- 2814-- 01 1 SWITCH,KEYPAD;ELASTOMERIC FRONT PANEL
-4 872-- 0004-- 00 1 CIRCUIT BOARD; FRONT PANEL;2 CHANNEL
-5 200-- 4906-- 00 2 COVER;BNC FILLER
-6 366-- 0860-- 01 3 KNOB, .685 DIA, SOFT TOUCH
-7 366-- 0859-- 01 4 KNOB, .470 DIA, SOFT TOUCH

5- 12 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Mechanical Parts List

6
7
2

Figure 5- 5: Exploded view: MSO4000 front panel and display, 2-channel instruments

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 5- 13


Mechanical Parts List

Replaceable Parts List: DPO4000 Power Supply and I/O board

Fig. &
index Tektronix part Serial no. Serial no.
number number effective discont’d Qty Name & description
5-- 6 DPO4000 Power Supply and I/O Board
-1 407-- 5019-- 01 1 BRACKET;POWER SUPPLY COVER
-2 211-- 1198-- 00 19 SCREW,MACHINE; 6-- 32 X 0.250,PNH,STL,CDPL,T-- 15 TORX
-3 335-- 1655-- 00 1 SERIAL NUMBER LABEL
-4 119-- 7037-- 00 1 FAN,TUBEAXIAL; 12VDC,0.09A,1.4W,19
CFM,4000RPM,31DBA,60MM X 60MM X 25MM,W/7IN LEADS
& CONNECTOR,TACH OUT
-5 650-- 4939-- 02 1 POWER SUPPLY SUB ASSEMBLY;DOMESTIC
-6 366-- 0844-- 00 1 PUSH BUTTON;POWER; SAFETY CONTROLLED
-7 672-- 1785-- 01 1 I/O BOARD
-8 335-- 1213-- 01 1 MARKER,IDENT;LABEL,REAR PANEL,LEFT
-9 335-- 1738-- 00 1 MARKER,IDENT;LABEL,REAR PANEL, RIGHT
10 337-- 4378-- 00 1 SHIELD;INSULATOR,ELECTRICAL
11 131-- 6643-- 00 1 GROUNDING CLIPS

5- 14 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Mechanical Parts List

2
1 3

11

10 4

8 5
9

2 I/O board assemly

J310

Main board

J780
Main board

J850
6

Figure 5- 6: Exploded view: DPO4000 Power Supply and I/O board

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 5- 15


Mechanical Parts List

Replaceable Parts List: MSO4000 Power Supply and I/O board

Fig. &
index Tektronix part Serial no. Serial no.
number number effective discont’d Qty Name & description
5-- 7 MSO4000 Power Supply and I/O Board
-1 407-- 5019-- 01 1 BRACKET;POWER SUPPLY COVER
-2 211-- 1198-- 00 19 SCREW,MACHINE; 6-- 32 X 0.250,PNH,STL,CDPL,T-- 15 TORX
-3 335-- 1655-- 00 1 SERIAL NUMBER LABEL
-4 119-- 7223-- 00 1 FAN,TUBEAXIAL; 12VDC,0.09A,1.4W,19
CFM,4000RPM,31DBA,60MM X 60MM X 25MM,W/7IN LEADS
& CONNECTOR,TACH OUT
-5 650-- 4939-- 02 1 POWER SUPPLY SUB ASSEMBLY;DOMESTIC
-6 366-- 0844-- 00 1 PUSH BUTTON;POWER; SAFETY CONTROLLED
-7 672-- 1785-- 01 1 I/O BOARD
-8 335-- 1213-- 01 1 MARKER,IDENT;LABEL,REAR PANEL,LEFT
-9 335-- 1738-- 00 1 MARKER,IDENT;LABEL,REAR PANEL, RIGHT
10 337-- 4378-- 00 1 SHIELD;INSULATOR,ELECTRICAL
11 131-- 6643-- 00 1 GROUNDING CLIPS

5- 16 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Mechanical Parts List

2
1 3

11

10 4

8 5
9

2 I/O board assemly

J310

Main board

J780
Main board

J850
6

Figure 5- 7: Exploded view: MSO4000 Power Supply and I/O board

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 5- 17


Mechanical Parts List

Replaceable Parts List: DPO4000 main board, inverter board, and side fans

Fig. &
index Tektronix part Serial no. Serial no.
number number effective discont’d Qty Name & description
5-- 8 DPO4000 Main board, Inverter Board, Side Fans
-1 650-- 4937-- 00 1 FAN TRAY ASSEMBLY
-2 119-- 6806-- 00 1 POWER SUPPLY; INVERTER,1200VAC NO-- LOAD & 600VAC
@ 85KOHM LOADED OUTPUT
-3 211-- 1143-- 00 2 SCREW,PT; K35-- 1.57,PAN HEAD,T-- 15
-4 407-- 5113-- 00 1 BRACKET,PCB;BACKLIGHT BOARD
-5 174-- 5015-- 00 1 CABLE ASSY,SP;INVERTER POWER,FLAT FLEX
-6 671-- 5843-- 00 1 CKT BD ASSY;ACQUISITION BOARD,4 CHANNEL, 1GHz
(DPO4104 ONLY)
671-- 5959-- 00 1 CKT BD ASSY;ACQUISITION BOARD,4 CHANNEL, 500MHZ
(DPO4054 ONLY)
671-- 5961-- 00 1 CKT BD ASSY;ACQUISITION BOARD,4 CHANNEL, 350MHZ
(DPO4034 ONLY)
671-- 5960-- 00 1 CKT BD ASSY;ACQUISITION BOARD,2 CHANNEL, 350MHZ
(DPO4032 ONLY)
-7 211-- 1198-- 00 6 SCREW,MACHINE; 6-- 32 X 0.250,PNH,STL,CDPL,T-- 15 TORX
DR
-8 174-- 5014-- 00 1 CABLE ASSY,SP;DISPLAY DATA,FLAT FLEX
-9 211-- 1174-- 00 13 SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING
MATERIAL; 6-- 32 X 0.312 L,PNH,STL CAD PLT,T15

5- 18 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Mechanical Parts List

1 Display assembly
7 to Inverter board
assembly
CN2

Main board 2 Inverter board


3 assembly

CN1
J570

J571
5

6
9

Main board

J930

8
I/O board
assemly

J610
Main board

Figure 5- 8: Exploded view: DPO4000 Main board, Inverter board, and side fans

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 5- 19


Mechanical Parts List

Replaceable Parts List: MSO4000 Main board, Inverter board, and side fans

Fig. &
index Tektronix part Serial no. Serial no.
number number effective discont’d Qty Name & description
5-- 9 MSO4000 Main board, Inverter board, side fans
-1 650-- 5077-- 00 1 FAN TRAY ASSEMBLY
-2 119-- 6806-- 00 1 POWER SUPPLY; INVERTER,1200VAC NO-- LOAD & 600VAC
@ 85KOHM LOADED OUTPUT
-3 211-- 1143-- 00 2 SCREW,PT; K35-- 1.57,PAN HEAD,T-- 15
-4 407-- 5113-- 00 1 BRACKET,PCB;BACKLIGHT BOARD
-5 174-- 5015-- 00 1 CABLE ASSY,SP;INVERTER POWER,FLAT FLEX
-6 671-- 5988-- 00 1 CIRCUIT BD ASSY; MAIN BOARD, (MSO4104)
671-- 6144-- 00 1 CIRCUIT BD ASSY;500MHZ 4 CHANNEL, (MSO4054)
671-- 6288-- 00 1 CIRCUIT BD ASSY;500MHZ 4 CHANNEL, MSO4034)
671-- 6145-- 00 1 CKT BD ASSY; 2 CHANNEL, (MSO4032)
-7 211-- 1198-- 00 6 SCREW,MACHINE; 6-- 32 X 0.250,PNH,STL,CDPL,T-- 15 TORX
DR
-8 174-- 5014-- 00 1 CABLE ASSY,SP;DISPLAY DATA,FLAT FLEX
-9 211-- 1174-- 00 13 SCREW,MACHINE; W/HEAVY PATCH THREADLOCKING
MATERIAL; 6-- 32 X 0.312 L,PNH,STL CAD PLT,T15

10 351-- 1115-- 00 1 PROBE GUIDE

5- 20 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual


Mechanical Parts List

1 Display assembly
7 to Inverter board
assembly
CN2

Main board 2 Inverter board


3 assembly

CN1
J570

J571
5

6
9

Main board
10

J930

8
I/O board
assemly

J610
Main board

Figure 5- 9: Exploded view: MSO4000 Main board, Inverter board, and side fans

Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual 5- 21


Mechanical Parts List

Replaceable Parts List: DPO4000 and MSO4000 Accessories

Fig. &
index Tektronix part Serial no. Serial no.
number number effective discont’d Qty Name & description

No Image Accessories
161-- 0066-- 09 1 CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M
(OPTION A1 - UNIVERSAL EURO)
161-- 0400-- 00 1 CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M
(OPTION A11 - INDIA)
161-- 0298-- 00 1 CA ASSY,PWR:3,1.0MM SQ,250V/10A,2.5 M
(OPTION A6 - JAPAN)
161-- 0066-- 00 1 CABLE ASSY,PWER,:3,18 AWG,92 L
(STANDARD CABLE; OPTION A0)
161-- 0066-- 10 1 CABLE ASSY,PWR; 3,1.0 MM SQ,250V/10A,2.5 ME-
TER,STR,IEC320,RCPT X 13A,FUSED UK PLUG(13A FUSE),
SAFETY CONTROLLED (OPTION A2 - UNITED KINGDOM)
161-- 0066-- 13 1 CABLE ASSY,PWR; 3,1.0 MM SQ,250V/10A,2.5 ME-
TER,STR,IEC320,RCPT,SAFETY CONTROLLED, INSULATED
PINS (OPTION A3 - AUSTRALIA)
161-- 0154-- 00 1 CA ASSY,PWR; 3,1.0MM SQ,250V/10A,2.5 ME-
TER,STR,IEC320,RCPT,SAFETY CONTROLLED
(OPTION A5 - SWITZERLAND)
161-- 0304-- 00 1 CABLE ASSY,PWR; 3,1.0MM SQ,250V/10A,2.5 ME-
TER,STR,IEC320,3C CERTIFICATION,RCPT,SAFETY
CONTROLLED (OPTION A10 - CHINA)
016-- 1967-- 00 1 POUCH, PLASTIC:POUCH
020-- 2662-- 00 1 MSO4000 ACCESSORY PACKAGE P6516

5- 22 Tektronix 4000 Series Digital Phosphor Oscilloscopes Service Manual