EX-100/1000
SIDE STREAM OIL IN WATER ANALYZER Ultrasonics Fluorescence Spectroscopy
The EX-100/1000 is a side stream Oil in Water analyzer that uses fluorescence to provide
continuous accurate measurements of oil concentrations in water. Reliable real-time data
enables operators to take accurate discharge measurements and to improve efficiency of
separation processes enabling cost reductions.
In addition to the EX-100 features, the EX-1000 model offers spectral analysis.
FEATURES
• Patented ultrasonic cleaning
• Laser Induced Fluorescence (LIF)
• Periodic homogenisation of sample
• Sample point
• Configurable measurement ranges
(0-10 ppm, 0-100 ppm [...] up to 0-20,000 ppm)
• Measurement repeatability ±1% of full scale
• Remote management and diagnostics
• Easy to install (no sample conditioning required)
• Multiple communications options - 4-20 mA, HART, Modbus,
Extended Ethernet
• Optional integrated spectrometer
• Operates using Microsoft Windows 7
• Auto tuning functionality
• Viewing window of sample chamber
• Digital input & output
BENEFITS
• Robust and reliable
• Easy to use
• Low Cost Of Ownership (COO) with no routine maintenance
required
• No degradation of signal or recalibration required
• Side stream format offers localized sample control
• Droplet size compensation with homogenized samples
• Sample point facilitates laboratory correlation
• Remote control and monitoring (suitable for un-manned locations
and remote process monitoring)
• Instantaneous measurements
Measurement Performance
Measurement principle Laser Induced Fluorescence (LIF)
Cleaning Ultrasonic (automatic)
Range 0-20,000 ppm*
Repeatability ±1% of full scale range
Response time 1 Second, continuous results
Operating Conditions
Process temperature Up to 200°C
Process pressure Up to 100 barg
Process flow 5-25 l/min
Operational ambient temperature -20°C to 55°C
Utilities
Power supply 110 or 230 VAC (Pre-configured)
Power frequency 50 or 60 Hz
Power consumption 60 W normal, 300 W peak
Instrument air 5.5-7 barg (for pneumatic valve; electric valve option available)
(air must be filtered to <= 5 um)
Certification
Ingress protection IP66
Enclosure material 316L SS (Aluminium optional)
Analyzer ATEX / IECEx: EXII 2G d/de IIB T3/T4 Gb
Canada + USA: Class 1 Division 1 Groups C & D T3/T4
Class 1 Division 2 Groups A, B, C, D, T3/ T4
Class 1 Zone 2 AEx d/de IIB T3/T4
IMO MEPC-107 (49)
Communications
4-20 mA (1) Passive, Configurable for measurement readings/temperature
Digital Input (1) Start/Stop cycle control
Digital Output (s) Configurable as alarm contacts
Remote access Windows Remote Desktop
Internal data storage >10 years
Security 2 level password protection
Optional Communications
Second 4-20mA Passive, Configurable for measurement readings/temperature
HART Yes
Modbus RTU Implemented via HART to Modbus converter
Extended Ethernet 2 wire connection, capable of 1.6 Km distance
Additional Information
Flange fitting 1” ANSI RF (optional flange, sizes available)
Wetted parts 316L SS (other materials available on request)
Manual sample take off point Integral to analyzer
Viewing window Provided as standard
Ultrasonic Homogenisation Automatic oil droplet compensation
* dependent on sample matrix & instrument configuration. User may select any desired measurement from 0-10 ppm, 0-100 ppm [...] up to 20,000 ppm