Electronics Measurements and Instrumentation
Electronics Measurements and Instrumentation
AND INSTRUMENTATION
LECTURE NOTES
B.TECH
(III YEAR – II SEM)
(2017-18)
Prepared by:
Mrs. M.ANUSHA, Associate Professor
Ms.G.ANUSHA, Assistant Professor
REFERENCES:
1. Electronic Instrumentation and Measurements - David A. Bell, Oxford Uiv. Press, 1997.
2. Electronic Measurements and Instrumentation: B. M. Oliver, J. M. Cage TMH Reprint.
3. Measurement Systems - Emest O. Doebelin and Dhanesh N Manik, 6th Ed., TMH.
4. Electronic Measurements and Instrumentations by K. Lal Kishore, Pearson Education -
2010.
5. Industrial Instrumentation: T. R. Padmanabham Spiriger 2009.
OUTCOMES
Upon a successful completion of this course, the student will be able to:
1. Describe the fundamental concepts and principles of instrumentation
2. Explain the operation of various instruments required in measurements
3. Apply the measurement techniques for different types of tests
4. To select specific instruments for specific measurement function.
5. Understand principle of operation and working of different electronic instruments
Students will understand functioning, specification and application of signal analyzing
instruments
UNIT-I
Introduction:
The measurement of any quantity plays very important role not only in science but in
all branches of engineering, medicine and in almost all the human day to day activities.
The technology of measurement is the base of advancement of science. The role of science
and engineering is to discover the new phenomena, new relationships, the laws of nature
and to apply these discoveries to human as well as other scientific needs. The science and
engineering is also responsible for the design of new equipments. The operation, control and the
maintenance of such equipments and the processes is also one of the important functions of the
science and engineering branches. All these activities are based on the proper measurement and
recording of physical, chemical, mechanical, optical and many other types of parameters.
The measurement of a given parameter or quantity is the act or result of a quantitative
comparison between a predefined standard and an unknown quantity to be measured. The major
problem with any measuring instrument is the error. Hence, it is necessary to select the appropriate
measuring instrument and measurement procedure which minimises the error. The measuring
instrument should not affect the quantity to be measured.
An electronic instrument is the one which is based on electronic or electrical principles for
its measurement function. The measurement of any electronic or electrical quantity or variable is
termed as an electronic measurement.
Advantages of Electronic Measurement
The advantages of an electronic measurement are
1. Most of the quantities can be converted by transducers into the electrical or electronic signals.
2. An electrical or electronic signal can be amplified, filtered, multiplexed, sampled and measured.
3. The measurement can easily be obtained in or converted into digital form for automatic analysis
and recording.
4 The measured signals can be transmitted over long distances with the help of cables or radio
links, without any loss of information.
5. Many measurements can be carried either simultaneously or in rapid succession.
6. Electronic circuits can detect and amplify very weak signals and can measure the events of very
short duration as well.
7. Electronic measurement makes possible to build analog and digital signals. The digital signals
are very much required in computers. The modern development in science and technology are
totally based on computers.
8. Higher sensitivity, low power consumption and a higher degree of reliability are the important
features of electronic instruments and measurements. But, for any measurement, a well defined set
of standards and calibration units is essential. This chapter provides an introduction to different
types of errors in measurement, the characteristics of an instrument and different calibration
standards.
Functional elements of an instruments:
Any instrument or a measuring system can be described in general with the help of a block
diagram. While describing the general form of a measuring system, it is not necessary to go into the
details of the physical aspects of a specific instrument. The block diagram indicates the
necessaryelements and their functions in a general measuring system. The entire operation of an
instrument can be studied interms of these functional elements. The Fig. 1.1 shows the block
diagram showing the functional elements of an instrument.
Calibration:
Calibration is the process of making an adjustment or marking a scale so that the readings
of an instrument agree with the accepted and the certified standard.
The calibration offers a guarantee to the device or instrument that it is operating with
required accuracy, under the stipulated environmental conditions. It creates the confidence of using
the properly calibrated instrument, in user's mind. The periodic calibration of an instrument is very
much necessary.
The calibration characteristics can be determined by applying known values of quantities to
be measured and recording the corresponding output of the instrument. Such output values are then
compared with the input, to determine the error. Such a record obtained from calibration is called
calibration record. It is generally recorded in the tabular form. If it is represented in the graphical
form, it is called calibration curve. Such a calibration record or calibration curve is useful to obtain
the performance characteristics of an instrument. The performance of the instrument is not
guaranteed by the calibration. It only mdicates whether the performance of the instrument is
meeting the accuracy and range specification or not. If the device has been repaired, aged, adjusted
or modified, then recalibration is carried out.
Static characteristics:
As mentioned earlier, the static characteristics are defined for the instruments which
measure the quantities which do not vary with time. The various static characteristics are accuracy,
precision, resolution, error, sensitivity, threshold, reproducibility, zero drift, stability and linearity.
Accuracy:
It is the degree of closeness with which the instrument reading approaches the true value of
the quantity to be measured. It denotes the extent to which we approach the actual value of the
quantity. It indicates the ability of instrument to indicate the true value of the quantity. The
accuracy can be expressed in the following ways.
1) Accuracy as 'Percentage of Full Scale Reading : In case of instruments having uniform scale,
the accuracy can be expressed as percentage of full scale reading.
For example, the accuracy of an instrument having full scale reading of 50 units may be expressed
as ± 0.1% of full scale reading. From this accuracy indication, practically accuracy is expressed in
terms of limits of error. So for the accuracy limits specified above, there will be ± 0.05 units error
in any measurement. So for a reading of 50 units, there will be error of ± 0.05 units i.e. ± 0.1 %
while for a reading of 25 units, there will be error of ± 0.05 units in the reading i.e. ± 0.2%. Thus
as reading decreases, error in measurement is ± 0.05 units but net percentage error is more. Hence,
specification of accuracy in this manner is highly misleading.
2) Accuracy as 'Percentage of True Value' : This is the best method of specifying the accuracy. It is
to be specified in terms of the true value of quantity being measured. For example, it can be
specified as ± 0.1% of true value. This indicates that in such cases, as readings get smaller, error
also gets reduced. Hence accuracy of the instrument is better than the instrument for which it is
specified as percent of full scale reading.
Precision:
It is the measure of consistency or repeatability of measurements.
Let us see the basic difference between accuracy and precision. Consider an instrument on
which, readings upto 1/1000th of unit can be measured. But the instrument has large zero
adjustment error. Now every time reading is taken, it can be taken down upto '1000th of unit. So as
the readings agree with each other, we say that the instrument is highly precise. But, though the
readings are precise upto 10100th of unit, the readings are inaccurate due to large zero adjustment
error. Every reading will be inaccurate, due to such error. Thus a precise instrument may not be
accurate. Thus the precision means sharply or clearly defined and the readings agree among
themselves. But there is no guarantee that readings are accurate. An instrument having zero error,
if calibrated properly, can give accurate readings but in that case still, the readings can be obtained
down upto 1/10th of unit only. Thus accuracy can be improved by calibration but not the precision
of the instrument.
The precision is composed of two characteristics:
• Conformity and
• Number of significant figures.
Conformity:
Consider a resistor having true value as 2385692.0Ω, which is being measured by an ohmmeter.
Now, the meter is consistently measuring the true value of the resistor. But the reader, can read
consistently, a value as 2.4 MΩ due to nonavailability of proper scale. The value 2.4 MΩ is
estimated by the reader from the available scale. There are no deviations from the observed value.
The error created due to the limitation of the scale reading is a precision error.
The example illustrates that the conformity is a necessary, but not sufficient condition for
precision. Similarly, precision is necessary but not the sufficient condition for accuracy.
Significant Figures:
The precision of the measurement is obtained from the number of significant figures, in which the
reading is expressed. The significant figures convey the actual information about the magnitude
and the measurement precision of the quantity.
Example:
The table shows the set of 5 measurements.Calculate the precision of the 3rd measurement
Resolution:
It is the smallest increment of quantity being measured which can be detected with certainty by an
instrument.
So if a nonzero input quantity is slowly increased, output reading will not increase until some
minimum change in the input takes place. This minimum change which causes the change in the
output is called resolution. The resolution of an instrument is also referred to as discrimination of
the instrument. The resolution can affect the accuracy of the measurement.
Errors:
Static error = measured value – true value
The most important static characteristics of an instrument is its accuracy, which is generally
expressed in terms of the error called static error.
Mathematically it can be expressed as, e = At - Am J
In this expression, the error denoted as e is also called absolute error. The absolute error does not
indicate precisely the accuracy of the measurements. For example, absolute error of ± 1 V is
negligible when the voltage to be measured is of the order of 1000 V but the same error of ±1 V
becomes significant when the voltage under measurement is 5 V or so. Hence, generally instead of
specifying absolute error, the relative or percentage error is specified.
Sensitivity:
If the calibration curve is not linear as shown in the Fig. 1.3 (b), then the sensitivity varies with the
input. The sensitivity is always expressed by the manufacturers as the ratio of the magnitude of
quantity being measured to the magnitude of the response. Actually, this definition is the
reciprocal of the sensitivity is called inverse sensitivity or deflection factor. But manufacturers call
this inverse sensitivity as a sensitivity.
The units of the sensitivity are millimeter per micro-ampere, millimeter per ohm, counts per volt,
etc. while the units of a deflection factor are micro-ampere per millimeter, ohm per millimeter,
volts per count, etc. The sensitivity of the instrument should be as high as possible and to achieve
this range of an instrument should not greatly exceed the value to be measured.
Drift : Gradual shift in the meassured value ,over an extended period, when there is no change in
input.
Threshold: The minimum value of input for which the device just starts to respond.
Range/Span: The minimum and maximum value of quantity so that the device is capable of
measuring.
Repeatability: A measure of how well the output returns to a given value when the same precise
input is applied several times. Or The ability of an instrument to reproduce a certain set of reading
within a given accuracy.
Linearity: Input output relationship of a device must be linear.
But practical systems shows small deviations from the linear shape ( allowed within the specified
limits)
Hysteresis: Input is increased from negative value, output increases as indicated by curve 1
• Then the input is steadily decreased , output does not follow the same path , but lag by a certain
value as indicated by curve 2 •
The difference between the two curves is called Hysterisis.
DYNAMIC CHARACTERISTICS:
The response of instruments or systems to dynamic I/P s are also functions of time.
Instruments rarely respond instantaneously to changes in the measured variables.
Instead, they exhibit slowness or sluggishness due to such things as mass, thermal capacitance, fluid
capacitance or electric capacitance.
• Speed of Response: It is the ability of a system to respond to a sudden changes in the input
signal/quantity
• Fidelity: It is the degree to which an instrument indicates the changes in the measured variable
without dynamic error ( Indication of how much faithfully system responds to the changes in input).
Lag: It is the retardation or delay in the response of an instrument to changes in the measured
variable. Two types : Process lag(process) and Control lag (Instrument)
Dynamic error:
It is the difference between the true value of the variable to be measured, changing with time and
the value indicated by the measurement system, assuming zero static error. The Fig. 1.13 shows
the dead time, i.e. time delay and the dynamic error.
Types of errors:
The static error is defined earlier as the difference between the true value of the variable and the
value indicated by the instrument. The static error may arise due to number of reasons. The static
errors are classified as:
1) Gross errors
2) Systematic errors
3) Random errors
Gross errors:
The gross errors mainly occur due to carelessness or lack of experience of a human being. These
cover human mistakes in readings, recordings and calculating results. These errors also occur due
to incorrect adjustments of instruments. These errors cannot be treated mathematically. These
errors are also called personal errors. Some gross errors are easily detected while others are very
difficult to detect.\
Systematic errors:
The systematic errors are mainly resulting due to the shortcomings of the instrument and the
characteristics of the material used in the instrument, such as defective or worn parts, ageing
effects, environmental effects, etc.
A constant uniform deviation of the operation of an instrument is known as a systematic
error. There are three types of systematic errors as
1) Instrumental errors 2) Environmental errors 3) Observational errors
Instrumental errors :
These errors are mainly due to following three reasons
• Short-comings of instrument
These are because of the mechanical structure of the instruments eg. Friction in the bearings of
various moving parts, irregular spring tensions, hysteresis, gear backlash, variation in air gap etc.
Misuse of instrument A good instrument if used in abnormal way gives misleading results. Poor
initial adjustments, Improper zero setting, Using leads of high resistance. Elimination: Use the
instrument intelligently & Correctly
• Loading effects Loading effects due to Improper way of using the instrument
• Ellimination.
– Selecting proper instrument and the transducer for the measurement.
– Recognize the effect of such errors and apply the proper correction factors.
– Calibrate the instrument carefully against standard.
Environmental Errors (due to the External Conditions)
• The various factors : Temperature changes, Pressure, vibratons, Thermal emf., stray capacitance,
cross capacitance, effect of External fields, Aging of equipments and Frequency sensitivity of an
instrument.
Elimination • Using proper correction factors and using the instrument Catalogue • Using
Temperature & Pressure control methods etc. • Reducing the effect of dust, humidity on the
components in the instruments. • The effects of external fields can be minimized by using the
magnetic or electrostatic shields of screens.
Observational Errors:
Error introduced by the observer
Few souces are:
• Parallax error while reading the meter,
• wrong scale selection,
• habits of individual obsever
• Elimination
Use the
• instrument with mirrors,
• instrument with knife edge pointers,
• Instrument having digital display
Random errors:
Some errors still result, though the systematic and instrumental errors are reduced or atleast
accounted for. The causes of such errors are unknown and hence, the errors are called random
errors. These errors cannot be determined in the ordinary process of taking the measurements.
This is also called fractional error. It is the ratio of the error to the specified magnitude of a
quantity.
D.C instruments:
a) Using shunt resistance, d.c. current can be measured. The instrument is d.c. microammeter,
milliammeter or ammeter.
b) Using series resistance called multiplier, d.c. voltage can be measured. The instrument is d.c.
millivoltmeter, voltmeter or kilovoltmeter.
c) Using a battery and resistive network, resistance can be measured. The instrument is ohmmeter.
A.C instruments:
a) Using a rectifier, a.c. voltages can be measured, at power and audio frequencies. The instrument
is a.c. voltmeter.
b) Using a thermocouple type meter radio frequency (RF) voltage or current can be measured.
c) Using a thermistor in a resistive bridge network, expanded scale for power line voltage can be
obtained.
Basic DC voltmeter:
The basic d.c. voltmeter is nothing but a permanent magnet moving coil (PMMC) 0' Arsonval
galvanometer. The resistance is required to be connected in series with the basic meter to use it as
a voltmeter. This series resistance is called a multiplier. The main function of the multiplier is to
limit the current through the basic meter so that the meter current does not exceed the full scale
deflection value. The voltmeter measures the voltage across the two points of a circuit or a voltage
across a circuit component. The basic d.c. voltmeter is shown in the Fig.
The voltmeter must be connected across the two points or a component, to measure the potential
difference, with the proper polarity.
The multiplier resistance can be calculated as:
The multiplying factor for multiplier is the ratio of full range voltage to be measured and the drop
across the basic meter.
Multirange voltmeters:
The range of the basic d.c. voltmeter can be extended by using number of multipliers clnd a
selector switch. Such a meter is called multirange voltmeter.
The R1, R2, R3 and R4 are the four series multipliers. When connected in series with the meter,
they can give four different voltage ranges as V1, V2, V3, and V4. The selector switch S is
multiposition switch by which the required multiplier can be selected in the circuit.
The mathematical analysis of basic d.c.voltmeter is equally applicable for such multirange
voltmeter. Thus,
Sensitivity of voltmeters:
In a multirange voltmeter, the ratio of the total resistance Rt to the voltage range remains same.
This ratio is nothing but the reciprocal of the full scale deflection current of the meter i.e. 1/101.
This value is called sensitivity of the voltmeter. Thus the sensitivity of the voltmeter is defined,
Loading effect:
While selecting a meter for a particular measurement, the sensitivity rating IS very important. A
low sensitive meter may give the accurate reading in low resistance circuit but will produce totally
inaccurate reading in high resistance circuit.
The voltmeter is always connected across the two points between which the potential difference is
to be measured. If it is connected across a low resistance then as voltmeter resistance is high, most
of the current will pass through a low resistance and will produce the voltage drop which will be
nothing but the true reading. But if the voltmeter is connected across the high resistance then due
to two high resistances in parallel, the current will divide almost equally through the two paths.
Thus the meter will record the voltage drop across the high resistance which will be much lower
than the true reading. Thus the low sensitivity instrument when used in high resistance circuit
'gives a lower than the true reading. This is called loading effect of the voltmeters. It is mainly
caused due to low sensitivity instruments.
The form factor is the ratio of r.m.s. value to the average value of an alternating quantity.
The diodes D1 and D2 are used for the rectifier circuit. The diodes show the nonlinear behaviour
for the low currents hence to increase the current through diode D1, the meter is shunted with a
resistance Rsh' This ensures high current through diode and its linear behaviour.
When the a.c. input is applied, for the positive half cycle, the diode 01 conducts and causes the
meter deflection proportional to the average value of that half cycle. In the negative cycle, the
diode D2 conducts and D1 is reverse biased. The current through the meter is in opposite direction
and hence meter movement is bypassed. Thus due to diodes, the rectifying action produces
pulsating d.c. and lile meter indicates the average value of the input.
The rectifier used in the circuit rectifies a.c. voltage into d.c. voltage for measurement of a.c.
voltage before current passes through the meter. The other diode is used for the protection purpose.
Use of multimeter for resistance measurement:
The Fig shows ohmmeter section of multimeter for a scale multiplication of 1. Before any
measurement is made, the instrument is short circuited and "zero adjust" control is varied until the
meter reads zero resistance i.e. it shows full scale current. Now the circuit takes the form of a
variation of the shunt type ohmmeter. Scale multiplications of 100 and 10,000 can also be used for
measuring high resistances. Voltages are applied the circuit with the help of battery.
Digital multimeters:
The digital multimeter is an instrument which is capable of measuring a.c. voltages, d.c. voltages,
a.c. and d.c. currents and resistances over several ranges. The basic circuit of a digital multimeter
is always a d.c. voltmeter as shown in the Fig
The current is converted to voltage by passing it through low shunt resistance. The a.c. quantities
are converted to d.c. by employing various rectifier and filtering circuits. While for the resistance
measurements the meter consists of a precision low current source that is applied across the
unknown resistance while gives d.c. voltage. All the quantities are digitized using analog to digital
converter and displayed in the digital form on the display.
The basic building blocks of digital multimeter are several AID converters, counting circuitry and an
attenuation circuit. Generally dual slope integration type ADC is prefprred in the multimeters. The
single attenuator circuit is used for both a.c. and d.c. measurements in many commercial
multimeters.
RMS value of the sinusoidal waveform is measured by theaverage reading voltmeter of which scale is
calibrated in terms of rms value. This method is quite simple and less expensive. But sometimes rms value of
the non-sinusoidal waveform is required to be measured. For such a measurement a true rms reading
voltmeter is required. True rms reading voltmeter gives a meter indication by sensing heating power of
waveform which is proportional to the square of the rms value of the voltage.
Thermo-couple is used to measure the heating power of the input waveform of which heater is supplied by
the amplified version of the input waveform. Output voltage of the thermocouple is proportional to the square
of the rms value of the input waveform. One more thermo-couple, called the balancing thermo-couple, is
used in the same thermal environment in order to overcome the difficulty arising out of non-linear behaviour
of the thermo-couple. Non-linearity of the input circuit thermo-couple is cancelled by the similar non-linear
effects of the balancing thermo-couple. These thermo-couples form part of a bridge in the input circuit of a dc
amplifier, as shown in block diagram.
AC waveform to be measured is applied to the heating element of the main thermocouple through an ac
amplifier. Under absence of any input waveform, output of both thermo-couples are equal so error signal,
which is input to dc amplifier, is zero and therefore indicating meter connected to the output of dc amplifier
reads zero. But on the application of input waveform, output of main thermo-couple upsets the balance and
an error signal is produced, which gets amplified by the dc amplifier and fedback to the heating element of
the balancing thermo-couple. This feedback current reduces the value of error signal and ultimately makes it
zero to obtain the balanced bridge condition. In this balanced condition, feedback current supplied by the dc
amplifier to the heating element of the balance thermo-couple is equal to the ac current flowing in the heating
element of main thermo-couple. Hence this direct current is directly proportional to the rms value of the input
ac voltage and is indicated by the meter connected in the output of the dc amplifier. The PMMC meter may
be calibrated to read the rms voltage directly.
By this method, rms value of any voltage waveform can be measured provided that the peak excursions of the
waveform do not exceed the dynamic range of the ac amplifier.
Recommended questions:
1. What is measurement? What are the two basic requirements of any measurement?
2. List the advantages of an electronic measurement.
3. Define and cxplain tilc term 'Calibration '.
4. How the performance characteristics of an instrumcnt are classified?
5. Definc and explain the following static characteristics of an instrulllcnt :
i) Accuracy ii) Precision iii) Static error iv) Resolution
v) SensitiVity v/) Threshold vii) Zero drift viii) Rcproducibility [jan 05,08 jul
07] ix) Lillearity and x) Stability
6. Explain how the accuracy can be specified for an instmment.
7. Distinguish clearly between aCCliracy and precision.
9. Explaill tile terms relative error and relative percentage error.
10. What is scale span of an instrument?
11. Define a dynamic response of an instrument.
12. Define the following terms,
i) Speed of response ii) Lag iii) Fidelity ivY Dynamic error.
13. Define and explain the types of errors possible in an instrument.
14. Define limiting errors. Derive the expression for relative limiting error.
15. A moving coil voltmeter has a uniform scale with 100 divisions, the full scale reading is 200 V
and 1/10 of scale division can be estimated with a fair degree of certainity. Determine the
resolution of the instrument in volt. [Ans. : 0.2 V]]
16. A digital voltmeter has a read out range from 0-9999 counts. Determine the resolution of the
instrument in volt when the full scale reading is 9.999 \I. [Ans. : 1 mV
17. A true value of voltage across resister is 50 V. The instrument reads 49 V. Calculate
i) absolute error ii) percentage error iii) percentage accuracy.
18. What is sensitil'ity of voltmcters & Explain.
19. What is a loading effect & Explain with the suitable example.
20. Explain the operation of basic d.c. voltmeter.
21. Explain the working of d.c. 11lultirange voltmeter.
22. State the requirements of a multiplier.
UNIT-2
AF Wave analyzer
The wave analyzer consists of a very narrow pass-band filter section which can Be tuned to
a particular frequency within the audible frequency range (20Hz to 20 KHz)). The block diagram of
a wave analyzer is as shown in fig 1.
The complex wave to be analyzed is passed through an adjustable attenuator which serves as
a range multiplier and permits a large range of signal amplitudes to be analyzed without loading the
amplifier.
The output of the attenuator is then fed to a selective amplifier, which amplifies the selected
requency. The driver amplifier applies the attenuated input signal to a high-Q active filter. This
high-Q filter is a low pass filter which allows the frequency which is selected to pass and reject all
others. The magnitude of this selected frequency is indicated by the meter and the filter section
identifies the frequency of the component. The filter circuit consists of a cascaded RC resonant
circuit and amplifiers. For selecting the frequency range, the capacitors generally used are of the
closed tolerance polystyrene type and the resistances used are precision potentiometers. The
capacitors are used for range changing and the potentiometer is used to change the frequency within
the selected pass-band, Hence this wave analyzer is also called a Frequency selective voltmeter. The
entire AF range is covered in decade steps by switching capacitors in the RC section.
The selected signal output from the final amplifier stage is applied to the meter circuit and to
an unturned buffer amplifier. The main function of the buffer amplifier is to drive output devices,
such as recorders or electronics counters.
The meter has several voltage ranges as well as decibel scales marked on it. It is driven by an
average reading rectifier type detector. The wave analyzer must have extremely low input distortion,
undetectable by the analyzer itself. The band width of the instrument is very narrow typically about
1% of the selective band given by the following response characteristics shows in fig.1.2
1. Electrical measurements
2. Sound measurements
3. Vibration measurements.
In industries there are heavy machineries which produce a lot of sound and vibrations, it is
very important to determine the amount of sound and vibrations because if it exceeds the
permissible level it would create a number of problems. The source of noise and vibrations is first
identified by wave analyzer and then it is reduced by further circuitry.
The block schematic of the wave analyser using the heterodyning principle is shown in fig.
above. The operating frequency range of this instrument is from 10 kHz to 18 MHz in 18
overlapping bands selected by the frequency range control of the local oscillator. The bandwidth is
controlled by an active filter and can be selected at 200, 1000, and 3000 Hz.
Wave analyzers have very important applications in the following fields:
1) Electrical measurements
2) Sound measurements and
3) Vibration measurements.
The wave analyzers are applied industrially in the field of reduction of sound and vibrations
generated by rotating electrical machines and apparatus. The source of noise and vibrations is first
identified by wave analyzers before it can be reduced or eliminated. A fine spectrum analysis with
the wave analyzer shows various discrete frequencies and resonances that can be related to the
motion of machines. Once, these sources of sound and vibrations are detected with the help of wave
analyzers, ways and means can be found to eliminate them.
Harmonic distortion :
The total harmonic distortion (THD) is a measurement of the harmonic distortion present in
a signal and is defined as the ratio of the sum of the powers of all harmonic components to the
power of the fundamental frequency. Distortion factor, a closely related term, is sometimes
used as a synonym.
where Vn is the RMS voltage of the nth harmonic and n = 1 is the fundamental frequency.
Spectrum Analyzer
The modern spectrum analyzers use a narrow band super heterodyne receiver.
Supereterodyne is nothing but mixing of frequencies in the super above audio range. The functional
block diagram of super heterodyne spectrum analyzer or RF spectrum nalyzer as shown in the
Figure
The RF input to be analyzed is applied to the input attenuator. After attenuating, the signal is
fed to low pass filter. The low pass filter suppresses high frequency components and allows low
frequency components to pass through it. The output of the low pass filter is given to the mixer,
where this signal is fixed with the signal coming from voltage controlled or voltage tuned
oscillator.
This oscillator is tuned over 2 to 3 GHz range. The output of the mixer includes two signals
whose amplitudes are proportional to the input signal but their frequencies are the sum and
difference of the input signal and the frequency of the local oscillator.
Since the frequency range of the oscillator is tuned over 2 to 3 GHz, the IF amplifier is tuned
to a narrow band of frequencies of about 2 GHz. Therefore only those signals which are separated
from the oscillator frequency by 2 GHz are converted to Intermediate Frequency (IF) band. This IF
signal is amplified by IF amplifier and then rectified by the detector. After completing amplification
and rectification the signal is applied to vertical plates of CRO to produce a vertical deflection on
the CRT screen. Thus, when the saw tooth signal sweeps, the oscillator also sweeps linearly from
minimum to maximum frequency range i.e., from 2 to 3 GHz.
Here the saw tooth signal is applied not only to the oscillator (to tune the oscillator) but also
to the horizontal plates of the CRO to get the frequency axis or horizontal deflection on the CRT
screen. On the CRT screen the vertical axis is calibrated in amplitude and the horizontal axis is
calibrated in frequency.
A spectrum analyzer, which uses computer algorithm and an analog to digital conversion
phenomenon and produces spectrum of a signal applied at its input is known as digital Fourier or
digital FFT or digital spectrum analyzer
Principle
When the analog signal to be analyzed is applied, the A/D converter digitizes the analog
signal (i.e., converts the analog signal into digital signal). The digitized signal, which is nothing but
the set of digital numbers indicating the amplitude of the analog signal as a function of time is
stored in the memory of the digital computer. From the stored digitized data, the spectrum of the
signal is computed by means of computer algorithm.
Description:
The block arrangement of a digital Fourier analyzer is illustrated in the figure above .The
analog signal to be ana1ysed is applied to the low pass filter, which passes only low frequency
signals and rejects high pass spurious signals. This filter section is used mainly, to prevent aliasing.
The output of low pass filter is given to the attenuator. The attenuator is a voltage dividing network
whose function is to set the input signal to the level of the A/D converter. The use of attenuator
prevents the converter from overloading. The function of A/D converter is to convert the samples of
analog data into digital i.e. ., to digitize the analog signal. When the output of A/D converter is
applied to the digital computer, the computer analyzes the digitized data and adjusts the attenuator
setting accordingly in order to obtain the maximum output from the inverter without any
overloading. As soon as the entire analog signal is sampled and digitized by the A/D converter)
computer performs calculations on the data according to the programmed algorithm and the
calculated spectral components are stored in the memory of the computer
If the spectral display is to be viewed on the oscilloscope, the digital values of spectral
components stored in the computer memory are converted into analog by using D/A converters and
then applied to the CRO. Thus the spectral display of the input waveform is obtained on the CRT
screen.
Advantages
1. The use of computer avoids most of the hardware circuitry such as electronic switches, Filters
and PLLs. The use of less hardware reduces the cost of the analyzer.
2. More mathematical calculations can be carried-out on the spectral display.
3. The rate of sampling analog signal can be modified in order to obtain better spectral display.
The modulation is indicated by a meter. The output signal can be Amplitude Modulated (AM) or
Frequency Modulated (FM). Modulation may be done by a sine wave, Square, rectangular, or a
pulse wave.
1) RF Osillator
(2) Wide band amplifier.
(3) External Osillator.
4) Modulation Osillator
(5) Out put attenuator.
The carrier frequency is generated by a very stable RF oscillator using an LC tank circuit,
having a constant output over any frequency range. The frequency of oscillations is indicated by
thefrequency range control and the venire dial setting. AM is provided by an internal sine
wavegenerator or from an external source.
The signal generator is called an oscillator. A Wien bridge oscillator is used in this
generator.The Wien bridge oscillator is the best of the audio frequency range. The frequency of
oscillations can be changed by varying the capacitance in the oscillator.
The frequency can also be changed in steps by switching the resistors of different values.
The output of the Wien bridge oscillator goes to the function switch.
The function switch directs the oscillator output either to the sine wave amplifier or to the
square wave shaper. At the output, we get either a square or sine wave.The output is varied by
means of an attenuator.
The instrument generates a frequency ranging from 10 Hz to 1 MHz continuously vV (rms).The
output is taker through a push-pull amplifier. For low output, the impedance is 6000. The square
wave amplitudes can be varied from 0 - 20 v (peak). It is possible to adjust the symmetry of the
square wave from 30 -70%. The instrument requires only 7W of power at 220V 50Hz.
l. Frequency selector: It selects the frequency in different ranges and varies it continuously in
a ratio of 1: 11. The scale is non-linear.
2. Frequency multiplier: It selects the frequency range over 5 decades from 10 Hz to 7 MHz
3. Amplitude multiplier: It attenuates the sine wave in 3 decades, x l x 0.1 and x 0.01.
4. Variable amplitude: It attenuates the sine wave amplitude continuously
5. Symmetry control: It varies the symmetry of the square wave from 30% to 70%.
6. Amplitude: It attenuates the square wave output continuously.
7. Function switch: It selects either sine wave or square output.
8. Output available: This provides sine wave or square wave output.
9. Sync: This terminal is used to provide synchronization of the internal signal with an external
signal.
10. On-Off Switch
Sweep Generator
It provides a sinusoidal output voltage whose frequency varies smoothly and continuously
over an entire frequency band, usually at an audio rate. The process of frequency modulation may be
accomplished electronically or mechanically. It is done electronically by using the modulating
voltage to vary the reactance of the oscillator tank circuit component, and mechanically by
means of a motor driven capacitor, as provided for in a modern laboratory type signal generator.
Figure shows a basic block diagram of a sweep generator. The frequency sweeper provides a
variable modulating voltage which causes the capacitance of the master oscillator to vary. A
representative sweep rate could be of the order of 20 sweeps/second. A manual control allows
independent adjustment of the oscillator resonant frequency. The frequency sweeper provides a
varying sweep voltage synchronization to drive the horizontal deflection plates of the CRO.
Thus the amplitude of the response of a test device will be locked and displayed on the screen.
To identify a frequency interval, a marker generator provides half sinusoidal waveforms at any
frequency within the sweep range. The marker voltage can be added to the sweep voltage of the
CRO during alternate cycles of the sweep voltage, and appears superimposed on the responsecurve.
The automatic level control circuit is a closed loop feedback system which monitors the
RF level at some point in the measurement system. This circuit holds the power delivered to the
load or test circuit constant and independent o frequency and impedance changes. A constant
power level prevents any source mismatch and also provides a constant readout calibration
with frequency.
These generators are used as measuring devices in combination with a CRO. They provide both
quantitative and qualitative information of the system under test. They are made use of in transient
response testing of amplifiers. The fundamental difference between a pulse generator and a square
wave generator is in the duty cycle.
1. The pulse should have minimum distortion, so that any distortion, in the display is solely due to
the circuit under test.
2. The basic characteristics of the pulse are rise time, overshoot, ringing, sag, and undershoot.
3. The pulse should have sufficient maximum amplitude, if appreciable output power is required by
the test circuit, e.g. for magnetic core memory. At the same time, the attenuation range shouldbe
adequate to produce small amplitude pulses to prevent over driving of some test circuit.
4. The range of frequency control of the pulse repetition rate (PRR) should meet the needs of
theexperiment. For example, a repetition frequency of 100 MHz is required for testing fast
circuits.Other generators have a pulse-burst feature which allows a train of pulses rather than a
continuous output.
5. Some pulse generators can be triggered by an externally applied trigger signal; conversely,
pulse generators can be used to produce trigger signals, when this output is passed through a
differentiator circuit.
6. The output impedance of the pulse generator is another important consideration. In a fast
pulsesystem, the generator should be matched to the cable and the cable to the test circuit. A
mismatch would cause energy to be reflected back to the generator by the test circuit, and this may
be rereflected by the generator, causing distortion of the pulses.
The frequency range of the instrument is covered in seven decade steps from 1Hz to 10
MHz, with a linearly calibrated dial for continuous adjustment on all ranges.
The duty cycle can be varied from 25 - 75%. Two independent outputs are available, a 50Ω
source that supplies pulses with a rise and fall time of 5 ns at 5V peak amplitude and a 600Ω source
which supplies pulses with a rise and fall tme of 70 ns at 30 V peak amplitude. The instrument can
be operated as a freerunning genenrator or, it can be synchronized with externalsignals.
The basic generating loop consists of the current sources, the ramp capacitor, the Schmitt
triggerand the current switching circuit as shown in the fig
The upper current source supplies a constant current to the capacitor and the capacitor
voltage increases linearly. When the positive slope of the ramp voltage reaches the upper limit set
by theinternal circuit components, the Schmitt trigger changes state. The trigger circuit output
becomesnegative and reverses the condition of the current switch. The capacitor discharges linearly,
controlled by the lower current source.
When the negative ramp reaches a predetermined lower level, the Schmitt trigger switches
back to its original state. The entire process is then repeated.
The ratio i1/i2 determines the duty cycle, and is controlled by symmetry control. The sum of
i1 and i2 determines the frequency. The size of the capacitor is selected by the multiplier switch.
The unit is powered by an intenal supply that provides regulated voltages for all stages of the
instrument.
A signal generator is an instrument, which can produce various types of wave forms such as
sine wave, square wave, triangular wave, saw tooth wave, pulse trains etc. As it can generate a
variety of waveforms it is widely used in applications like electronic troubleshooting anti
development, testing the performance of electronic equipments etc. In such applications a signal
generator is used to provide known test conditions (i.e., desired signals of known amplitude and
frequency
Hence, the following precautionary measures should be taken while using a signal generator
for an application.
1. The amplitude and frequency of the output of the signal generator should be made stable and well
known.
2. There should be provision for controlling the amplitude of signal generator output from
very small to relatively large values.
3. The output signal of generator should not contain any distortion and thus, it should possess very
low harmonic contents.
Capability of Phase Lock the function generator can be phase locked to an external source.
One function generator can be used to lock a second function generator, and the two output signals
can be displaced in phase by adjustable amount. In addition, the fundamental frequency of one
generator can be phase locked to a harmonic of another generator, by adjusting the amplitude and
phase of the harmonic; almost any waveform can be generated by addition.
The function generator can also be phase locked to a frequency standard and its
output waveforms will then have the same accuracy and stability as the standard source.
The block diagram of a function generator is illustrated in fig. Usually the frequency is
controlled by varying the capacitor in the LC or RC circuit. In the instrument the frequency is
controlled by varying the magnitude of current which drives the integrator. The instrument produces
sine, triangular and square waves with a frequency range of 0.01 Hz to 100 kHz.
The frequency controlled voltage regulates two current sources. The upper current source
supplies constant current to the integrator whose output voltage increases linearly with time,
according to the equation of the output signal voltage. An increase or decrease in the
currentincreases or decreases the slope of the output voltage and hence controls the frequency. The
voltage comparator multi-vibrator changes states at a pre-determined maximum level of
the integrator output voltage. This change cuts off the upper current supply and switches on
the lower current supply. The lower current source supplies a reverse current to the integrator, so
that its output decreases linearly with time. When the output reaches a pre-determined
minimum level, the voltage comparator again changes state and switches on the Lower current
source. The output of the integrator is a triangular waveform whose frequency is determined by
the magnitude of the current supplied by the constant current sources. The comparator
output delivers a square wave voltage of the same frequency.
e = - 1/C ∫ idt
The resistance diode network alters the slope of the triangular wave as its amplitude changes and
produces a sine wave with less than 1% distortion.
The waveforms produced by arbitrary waveform generators, AWGs can be either repetitive
or sometimes just a single-shot. If the AWG waveform is only a single shot, then a triggering
mechanism is needed to trigger the AWG and possibly the measuring instrument.
The AWG is able to generate an arbitrary waveform defined by a set of values, i.e.
"waypoints" entered to set the value of the waveform at specific times. They can make up a digital
or even an analogue waveform.
There are a number of ways of designing arbitrary waveform generators. They are based
around digital techniques, and their design falls into one of two main categories:
Direct Digital Synthesis, DDS: This type of arbitrary waveform generator is based around
the DDS types of frequency synthesizer, and sometimes it may be referred to as an Arbitrary
Function Generator, AFG.
Variable-clock arbitrary waveform generator The variable clock arbitrary function
generator is the more flexible form of arbitrary waveform generator. These arbitrary waveform
generators are generally more flexible, although they do have some limitations not possessed by
the DDS versions. Sometimes these generators are referred to as just arbitrary waveform
generators, AWGs rather than arbitrary function generators.
Combined arbitrary waveform generator This format of AWG combines both of the other
forms including the DDS and variable clock techniques. In this way the advantages of both
systems can be realised within a single item of test equipment.
Two of the main specifications for an arbitrary waveform generator are their resolution and
also the speed. These two parameters determine the precision with which the waveform can be
reproduced. They are governed by different elements within the arbitrary waveform generator
circuit.
The amplitude resolution is governed by the resolution of the digital to analogue converter
(D/A or D2A). This is described in terms of the number of bits. A 12 bit resolution provides 4096
amplitude steps.
The speed of the arbitrary waveform generator is also very important. The maximum
repetition rate for the waveform is governed by two factors: the length of the waveform in terms of
the number of samples required to simulate the waveform and the maximum clock frequency. For
example if the arbitrary waveform generator had a maximum clock frequency of 25 MHz and the
waveform had 1000 points, then the maximum repetition rate would be 25 kHz. If a higher
repetition rate was required, then it would be necessary to decrease the number of samples as it
would not be possible to increase the clock frequency in the arbitrary waveform generator!
Arbitrary waveform generator applications:
AWGs are used in many applications where specialised waveforms are required. These can
be within a whole variety of sectors of the electronics industry.
To give a view of some of the AWG applications, it is possible for DDS-based arbitrary
waveform generators is to create signals with precisely controlled phase offsets or ratio-related
frequencies. This enables the generation of signals like polyphase sine waves, I-Q constellations, or
simulation of signals from geared mechanical systems such as jet engines. Complex channel-
channel modulations are also possible.
The arbitrary waveform generator may not be the most widely used of items of test
instrumentation, but they can be immensely useful in a variety of applications. Modern arbitrary
waveform generators are very flexible and can be used to create very specific waveforms for use in
testing a variety of applications.
Direct digital synthesizer, DDS technology lends itself to being used within arbitrary
waveform generators, AWGs. Those AWGs that use DDS technology are often referred to as
arbitrary function generators, or AFGs.
The reason for being called arbitrary function generators is that they often appear as an
extension of the function generator test instruments that are available.
Arbitrary waveform generators using direct digital synthesis technology are able to benefit
from the technology, while not adding unwanted additional complexity and cost. DDS technology
has developed considerably in recent years and this makes them a very attractive option to form the
basis of a waveform generator. As a result arbitrary function generators are relatively widely used.
As mentioned, this type of arbitrary waveform generator is based around the DDS types of
frequency synthesizer, and sometimes it may be referred to as an Arbitrary Function generator,
AFG.
The arbitrary function generator uses integrated circuits intended for direct digital frequency
synthesizers, but enables an arbitrary waveform generator circuit to be created relatively easily and
for an economic price.
To look at how an arbitrary function generator works, it is necessary to look at the operation
of a direct digital synthesizer.
This circuit operates by storing the points of a waveform in digital format, and then recalling
them to generate the waveform. These points can be on any form of repetitive waveform that is
required. The rate at which the DDS completes one waveform governs the frequency. The basic
block diagram of the DDS based arbitrary waveform generator is shown below.
The operation of the DDS within the arbitrary function generator can be envisaged by
looking at the way that phase progresses over the course of one cycle of the waveform.
The phase is often depicted as a line of phasor rotating around a circle. As the phase
advances around the circle, this corresponds to advances in the waveform. The faster is progresses,
the sooner it completes a cycle and the hence the higher the frequency.
The direct digital synthesizer operates by storing various points of the required waveform in
digital format in a memory. These can then be recalled to generate the waveform as they are
required.
To simulate the phase advances a phase accumulator is used. This takes in phase increment
information, and clock pulses from a clock. For each clock pulse, the phase will advance a certain
amount. The greater the increment, the larger the phase advance, and hence the higher the frequency
generated.
At each clock pulse the phase information is presented to the memory and the relevant
location is accessed, proving the waveform information for that particular phase angle.
It can be seen that any waveform can be loaded into the memory; although a sine wave is
shown on the diagram, the actual waveform could be anything.
While it is possible to load certain preset waveforms into the memory, it is also possible to
load user generated ones in as well. These make the test instrument an arbitrary waveform generator
or arbitrary function generator rather than a standard function generator.
While the arbitrary function generator or DDS based version of the arbitrary waveform
generator, has many advantages, there are also some disadvantages that should also be taken into
account when choosing what type of signal generator to use.
Sub Hz frequency resolution: By using a long word length phase accumulator in the phase
accumulator of the DDS, it is possible to achieve sub-Hertz frequency resolution levels.
Down sampling: Waveforms are automatically truncated by sampling to allow repetition
rates above the clock frequency.
Digital modulation: It is possible to add digital modulation words to the phase accumulator
to provide a means of providing digital modulation.
Waveform jitter: Waveform jitter is an issue with arbitrary function generators because
frequencies are up-sampled or down-sampled and this results in missing samples and hence jitter.
Only frequencies equal to the clock frequency divided by the waveform length and its sub
multiples are not sampled and therefore they do not suffer from this problem
Single waveform capability: It is only possible to generate a single waveform at a time
because memory segmentation and waveform sequencing is not possible using a DDS arbitrary
function generator
The arbitrary function generator is the ideal instrument where a variety of programmed
waveforms are required without the added flexibility and complexity of the more expensive variable
clock arbitrary waveform generator. For most laboratory applications, the arbitrary function
generator is an ideal choice.
Unit 3
Oscilloscopes
Introduction:
In studying the various electronic, electrical networks and systems, signals which are
functions of time, are often encountered. Such signals may be periodic or non periodic in nature.
The device which allows, the amplitude of such signals, to be displayed primarily as " function of
time, is called cathode ray oscilloscope, commonly known as C.R.O. The CR.O gives the visual
representation of the time varying signals. The oscilloscope has become an universal instrument
and is probably most versatile tool for the development of electronic circuits and systems. It is an
integral part of electronic laboratories.
The oscilloscope is, in fact, a voltmeter. Instead of the mechanical deflection of a metallic
pointer as used in the normal voltmeters, the oscilloscope uses the movement of an electron beam
against a fluorescent screen, which produces the movement of a visible spot. The movement of
such spot on the screen is proportional to the varying magnitude of the signal, which is under
measurement.
Basic Principle:
The electron beam can be deflected in two directions : the horizontal or x-direction and the
vertical or y-direction. Thus an electron beam producing a spot can be used to produce two
dimensional displays, Thus CRO. can be regarded as a fast x-y plotter. The x-axis and y-axis can
be used to study the variation of one voltage as a function of another. Typically the x-axis of the
oscilloscope represents the time while the y-axis represents variation of the input voltage signal.
Thus if bhe input voltage signal applied to the y-axis of CRO. is sinusoidally varying and if x-axis
represents the time axis, then the spot moves sinusoidally, and the familiar sinusoidal waveform
can be seen on the screen of the oscilloscope. The oscilloscope is so fast device that it can display
the periodic signals whose time period is as small as microseconds and even nanoseconds. The
CRO. Basically operates on voltages, but it is possible to convert current, pressure, strain,
acceleration and other physical quantities into the voltage using transducers and obtain their visual
representations on the CRO.
Cathode Ray Tube (CRT):
The cathode ray tube (CRT) is the heart of the CR.O. the CRT generates the electron
beam, ,accelerates the beam, deflects the beam and also has a screen where beam becomes
visible ,as a spot. The main parts of the CRT are:
i) Electron gun ii) Deflection system iii) Fluorescent screen
iv) Glass tube or envelope v) Base
A schematic diagram of CRT, showing its structure and main components is shown
in the Fig.
Electron Gun:
The electron gun section of the cathode ray tube provides a sharply focused electron beam
directed :towards the fluorescent-coated screen. This section starts from theql1ally heated cathode,
limiting the electrons. The control grid is give!! negative potential with respect to cathode dc. This
grid controls the number of electrons in the beam, going to the screen.
The momentum of the electrons (their number x their speed) determines the intensity, or
brightness, of the light emitted from the fluorescent screen due to the electron bombclrdl1lent. The
light emitted is usually of the green colour. Because the electrons are negatively charged, a
repulsive force is created by applying a negative voltage to the control grid (in CRT, voltages
applied to various grids are stated with respect to cathode, which is taken as common point). This
negative control voltage can be made variable.
Deflection System:
When the electron beam is accelerated it passes through the deflection system, with which
beam can be positioned anywhere on the screen. The deflection system of the cathode-ray-tube
consists of two pairs of parallel plates, referred to as the vertical and horizontal deflection plates.
One of the plates' in each set is connected to ground (0 V), To the other plate of each set, the
external deflection voltage is applied through an internal adjustable gain amplifier stage, To apply
the deflection voltage externally, an external terminal, called the Y input or the X input, is
available.
As shown in the Fig. , the electron beam passes through these plates. A positive voltage
applied to the Y input terminal (Vy) Causes the beam to deflect vertically upward due to the
attraction forces, while a negative voltage applied to. the Y input terminal will cause the electron
beam to deflect vertically downward, due to the repulsion forces. When the voltages are applied
simultaneously to vertical and horizontcl1 deflecting plates, the electron beam is deflected due to
the resultant-of these two voltages.
Fluorescent Screen:
The light produced by the screen does not disappear immediately when bombardment by
electrons ceases, i.e., when the signal becomes zero. The time period for which the trace remains
on the screen after the signal becomes zero is known as "persistence". The persistence may be jS
short as a few microsecond, or as long as tens of seconds and minutes.
Long persistence traces are used in the study.. of transients. Long persistence helps in the
study of transients since the trace is still seen on the screen after the transient has disappeared.
Phosphor screen characteristics:
Many phosphor materials having different excitation times and colours as well as different
phosphorescence times are available. The type PI, P2, PI1 or P3I are the short persistence
phosphors and are used for the general purpose oscilloscope
Medical oscilloscopes require a longer phosphor decay and hence phosphors like P7 and
P39 are preferred for such applications. Very slow displays like radar require long persistence
phosphors to maintain sufficient flicker free picture. Such phosphors are P19, P26 and, P33.
The phosphors P19, P26, P33 have low burn resistance. The phosphors PI, P2, P4, P7, Pll
have medium burn resistance while PIS, P3I have high burn resistance.
Block diagram of simple oscilloscope:
CRT:
This is the cathode ray tube which is the heart of CR.O. It is' used to emit the rlectrons required to
strike the phosphor screen to produce the spot for the visual display of the signals.
Vertical Amplifier:
The input signals are generally not strong to provide the measurable deflection on the screen.
Hence the vertical amplifier. stage is used Jo amplify the input signals. The amplifier stages used
are generally wide band amplifiers so as to pass faithfully the entire band of frequencies to be
measured. Similarly it contains the attenuator stages as well. The attenuators are used when very
high voltage signals are to be examined, to bring the signals within the proper range of operation.
It consists of several stages with overall fixed sensltivity. The amplifier can be designed for
stability and required bandwidth very easily due to the fixed gain. The input stage colrtsists of an
attenuator followed by FET source follower. It has vel' high input impedance required to isolate
the amplifier from the attenuator. It is followed by BJT emitter follower to match the output
impedance of FET output With input of phase inverter. The phase inverter provides two antiphase
output signals which are required to operate the push pull output amplifier. The push pull operation
has advantages like better hum voltage cancellation, even harmonic suppression especially large
2nd harmonic, greater power output per tube and reduced number of defocusing and nonlinear
effects.
Delay line:
The delay line is used to delay the signal for some time in the verticClI sections. When the delay
line is not used, the part of the signal gets lost. Thus the input signal is not applied directly to the
vertical plates but is delClyed bv some time using a delay line cu-cuit as shown in the Fig.
If the trigger pulse is picked off at a time t = to after the signal has passed through the main
amplifier then signal is delayed by XI nanoseconds while sweep takes YI nanoseconds to reach.
The design of delay line is such that the delay time XI is higher than the time YI' Generally XI is
200. nsec while tl;1.eYI is 80 ns, thus the sweep starts well in time and no part of the signal is lost.
There are two types of delay lines used in CR.O. which are:
i) Lumped parameter delay line
ii) Distributed parameter delay line
Trigger circuit:
It is necessary that horizontal deflection starts at the same point of the input vertical signal, each
time it sweeps. Hence to synchronize horizontal deflection with vertical deflection a synchronizing
or triggering circuit is used. It converts the incoming signal into the triggering pulses, which are
used for the synchronization.
Time base generator:
The time base generator is used to generate the sawtooth voltage, required to deflect the beam in
the horizontal section. This voltage deflects the spot at a constant time dependent rate. Thus the x-
axis' on the screen can be represented as time, which, helps to display and analyse the time varying
signals.
LISSAJOUS FIGURES:
When both pairs of the deflection plates (horizontal deflection plates and vertical deflection
plates) of CRO (Cathode Ray Oscilloscope) are connected to two sinusoidal voltages, the patterns
appear at CRO screen are called the Lissajous pattern.
Shape of these Lissajous pattern changes with changes of phase difference between signal
and ration of frequencies applied to the deflection plates (traces) of CRO. Which makes these
Lissajous patterns very useful to analysis the signals applied to deflection plated of CRO. These
lissajous patterns have two Applications to analysis the signals. To calculate the phase difference
between two sinusoidal signals having same frequency. To determine the ratio frequencies of
sinusoidal signals applied to the vertical and horizontal deflecting plates.
Calculation of the phase difference between two Sinusoidal Signals having same frequency
When two sinusoidal signals of same frequency and magnitude are applied two both pairs of
deflecting plates of CRO, the Lissajous pattern changes with change of phase difference between
signals applied to the CRO.
For different value of phase differences, the shape of Lissajous patterns is shown in figure below,
There are two cases to determine the phase difference ø between two signals applied to the
horizontal & vertical plates,
Case - I: When, 0 < ø < 90o or 270o < ø < 360o : -
As we studied above it clear that when the angle is in the range of 0 < ø < 90o or 270o < ø <
360o, the Lissajous pattern is of the shape of Ellipse having major axis passing through origin from
first quadrant to third quadrant:
Let’s consider an example for 0 < ø < 90o or 270o < ø < 360o, as shown in figure below,
As we studied above it Clear that when the angle is in the range of 0o < ø < 90o or 270o < ø <
360o, the Lissajous Pattern is of the shape of Ellipse having major axis passing through origin from
second quadrant to fourth quadrant:
Let’s consider an example for When, 90o < ø < 180o or 180o < ø < 270o, as shown in figure
below, In this condition the phase difference will be,
Oscilloscope probes:
Oscilloscopes are widely used for test and repair of electronics equipment of all types. However it is
necessary to have a method of connecting the input of the oscilloscope to the point on the equipment
under test that needs monitoring.
To connect the scope to the point to be monitored it is necessary to use screened cable to prevent
any pick-up of unwanted signals and in addition to this the inputs to most oscilloscopes use coaxial
BNC connectors. While it is possible to use an odd length of coax cable with a BNC connector on
one end and open wires with crocodile / alligator clips on the other, this is not ideal and purpose
made oscilloscope probes provide a far more satisfactory solution.
Oscilloscope probes normally comprise a BNC connector, the coaxial cable (typically around a
metre in length) and what may be termed the probe itself. This comprises a mechanical clip
arrangement so that the probe can be attached to the appropriate test point, and an earth or ground
clip to be attached to the appropriate ground point on the circuit under test.
Care should be taken when using oscilloscope probes as they can break. Although they are robustly
manufactured, any electronics laboratory will consider oscilloscope probes almost as "life'd" items
that can be disposed of after a while when they are broken. Unfortunately the fact that they are
clipped on to leads of equipment puts a tremendous strain on the mechanical clip arrangement. This
is ultimately the part which breaks.
X1 and X10 oscilloscope probes
There are two main types of passive voltage scope probes. They are normally designated X1 and
X10, although 1X and 10X are sometimes seen. The designation refers to the factor by which the
impedance of the scope itelf is multiplied by the probe.
The X1 probes are suitable for many low frequency applications. They offer the same input
impedance of the oscilloscope which is normally 1 MΩ. However for applications where better
accuracy is needed and as frequencies start to rise, other test probes are needed.
To enable better accuracy to be achieved higher levels of impedance are required. To achieve this
attenuators are built into the end of the probe that connects with the circuit under test. The most
common type of probe with a built in attenuator gives an attenuation of ten, and it is known as a
X10 oscilloscope probe. The attenuation enables the impedance presented to the circuit under test to
be increased by a factor of ten, and this enables more accurate measurements to be made.
As the X10 probe attenuates the signal by a factor of ten, the signal entering the scope itself will be
reduced. This has to be taken into account. Some oscilloscopes automatically adjust the scales
according to the probe present, although not all are able to do this. It is worth checking before
making a reading.
The 10X scope probe uses a series resistor (9 M Ohms) to provide a 10 : 1 attenuation when it is
used with the 1 M Ohm input impedance of the scope itself. A 1 M Ohm impedance is the standard
impedance used for oscilloscope inputs and therefore this enables scope probes to be interchanged
between oscilloscopes of different manufacturers.
The scope probe circuit shown is a typical one that might be seen - other variants with the variable
compensation capacitor at the tip are just as common.
In addition to the X1 and X10 scope probes, X100 probes are also available. These oscilloscope
probes tend to be used where very low levels of circuit loading are required, and where the high
frequencies are present. The difficulty using the is the fact that the signal is attenuated by a factor of
100.
The X10 scope probe is effectively an attenuator and this enables it to load the circuit under test far
less. It does this by decreasing he resistive and capacitive loading on the circuit. It also has a much
higher bandwidth than a traditional X1 scope probe.
The x10 scope probe achieve a better high frequency response than a normal X1 probe for a variety
of reasons. It does this by decreasing the resistive and capacitive loading on the The X10 probe can
often be adjusted, or compensated, to improve the frequency response.
Probes that have only one adjustment, it is the LF compensation that is adjusted, sometimes the HF
compensation may be adjusted in the factory.
To achieve the correct compensation the probe is connected to a square wave generator in the scope
and the compensation trimmer is adjusted for the required response - a square wave.
A note of caution: many oscilloscope probes include a X1/X10 switch. This is convenient, but it
must be understood that the resistive and capacitive load on the circuit increase significantly in the
X1 position. It should also be remembered that the compensation capacitor has no effect when used
in this position.
As an example of the type of loading levels presented, a typical scope probe may present a load
resistance of 10MΩ along with a load capacitance of 15pF to the circuit in the X10 position. For the
X1 position the probe may have a capacitance of possibly 50pF plus the scope input capacitance.
This may end up being of the order of 70 to 80pF.
Apart from the standard 1X and 10X voltage probes a number of other types of scope probe are
available.
Using two scope probes in this way can give rise to a number of problems. The main one is that
single ended measurements of this nature do not give the required rejection of any common mode
signals (i.e. Common Mode Rejection Ratio, CMMR) and additional noise is likely to be present.
There may be a different cable length on each probe that may lead to a time differences and a
slight skewing between the signals.
To overcome this a differential probe may be used. This uses a differential amplifier at the
probing point to provide the required differential signal that is then passed along the scope probe
lead to the oscilloscope itself. This approach provides a far higher level of performance.
High voltage probes: Most standard oscilloscope voltage probes like the X1 or X10 are
only specified for operation up to voltages of a few hundred volts at most. For operation higher
than this a proper high voltage probe with specially insulated probe is required. It also will step
down the voltage for the input to the scope so that the test instrument is not damaged by the high
voltage. Often voltage probes may be X50 or X100.
Delay Lines:
Applications of CRO:
1. Measurement of voltage – Voltage waveform will be made on the oscilloscope screen.
From the screen of the CRO, the voltage can be measured by seeing its amplitude variation
on the screen.
2. Measurement of current – Current waveform will be read from the oscilloscope screen in
the similar way as told in above point. The peak to peak, maximum current value can be
measured from the screen.
3. Measurement of phase – Phase measurement in cro can be done by the help of Lissajous
pattern figures. Lissajous figures can tell us about the phase difference between two signals.
Frequency can also be measured by this pattern figure.
4. Measurement of frequency – Frequency measurement in cathode ray oscilloscope can be
made with the help of measuring the time period of the signal to be measured.
The comparison of two or more voltages is very much ,necessary in the analysis and study
of many electronic circuits and systems. This is possible by using more than one oscilloscope but
in such a case it is difficult to trigger the sweep of each oscilloscope precisely at the same time. A
common and less costly method to solve this problem is to use dual trace or multitrace
oscilloscopes. In this method, the same electron beam is used to generate two traces which can be
deflected from two independent vertical sources. The methods are used to generate two
independent traces which the alternate sweep method and other is chop method.
The block diagram of dual trace oscilloscope is shown in the Fig
There are two separate vertical input channels A and B. A separate preamplifier and -
attenuator stage exists for each channel. Hence amplitude of each input can be individually
controlled. After preamplifier stage, both the signals are fed to an electronic switch. The switch
has an ability to pass one channel at a time via delay line to the vertical amplifier. The time base
circuit uses a trigger selector switch 52 which allows the circuit to be triggered on either A or B
channel, on line frequency or on an external signal. The horizontal amplifier is fed from the sweep
generator or the B channel via switch 5! and 51. The X-Y mode means, the oscilloscope operates
from channel A as the vertical signal and the channel B as the horizontal signal. Thus in this mode
very accurate X-Y measurements can be done.
Sampling Time Base:
The time base circuit of the sampling oscilloscope is different than the conventional oscilloscope.
The time base of sampling oscilloscope has two functions:
i) To move the dots across the screen
ii) To generate the sampling command pulses for the sampling circuit.
It consists of synchronous circuit, which determines the sampling rate and establishes a
reference point in time with respect to the input signal. The time base generates a triggering pulse
which activates the oscillator to generate a ramp voltage. Similarly it generates a stair case
waveform. The ramp generation is based on the output of the synchronizing circuit.
Both the ramp as well as staircase waveforms are applied to a voltage comparator. This
comparator compares the two voltages and whenever these two voltages are equal, it generates a
samppng pulse. This pulse then momentarily bias the diodes of the sampling gate in the forward
direction and thus diode switch gets closed for short duration of time.
The capacitor charges but for short time hence, it can charge to only a small percentage of
the input signal value at that instant. This voltage is amplified by the vertical amplifier and then
applied to the vertical deflecting plates. This is nothing but a sample. At the same time, the
comparator gives a signal to the staircase generator to advance through one step. This is applied to
horizontal deflecting plates, thus during each step of the stair case waveform, the spot moves
across the screen. Thus the sampling time base is called a staircase-ramp generator in case of a
sampling oscillosope.
Block diagram of Sampling Oscilloscope:
The block diagram of sampling oscilloscope is shown in the Fig.
The input signal is applied to the diode sampling gate. At the start of each sampling cycle a trigger
inpu t pulse is generated which activates the blocking oscillator. The oscillator output is given to
the ramp generator which generates the linear ramp signal. Since the sampling must be
synchronized with the input signal freq\,lency, the signal is delayed in the vertical amplifier.
The staircase generator produces a staircase waveform which is applied to an attenuator.
The attenuator controls the magnitude of the staircase signal and then it is applied to a voltage
comparator. Another input to the voltage comparator is the output of the ramp generator. The
voltage comparator compares the two signals and produces the output pulse when the two voltages
are equal. This is nothing but a sampling pulse which is applied to sampling gate through the gate
control circuitry.
This pulse opens the diode gate and sample is taken in. This sampled signal is then applied
to the vertical amplifier and the vertical deflecting plates. The output of the staircase generator is
also applied to the horizontal deflecting plates.
During each step of staircase the spot moves on the screen. The comparator output
advances the staircase output through one step. After certain number of p\llses about thousand or
so, the staircase generator resets. The sm,lIler the size of the steps of the staircase generator, larger
is the number of samples and higher is the resolution of the image.
In this digital storage oscilloscope, the waveform to be stored is digitised ,md then stored in
a digital memory. The conventional cathode ray tube is used in this oscilloscope hencethe cost is
less. The power to be applied to memory is small and can be supplied by small battery. Due to this
the stored image can be displayed indefinitely as long ,15 power is supplied to memory. Once the
waveform is digitised then it can be further loaded into thecomputer and can be ana lysed in detail.
Block Diagram:
The block diagram of digital storage oscilloscope is shown in the Fig.
As done in all the oscilloscopes, the input signal is applied to the amplifier and attenuator
section. The oscilloscope uses same type of amplifier and attenuator circuitry as used in the
conventional oscilloscopes. The attenuated signal is then applied to the vertical amplifier.
The vertical input, after passing through the vertical amplifier, is digitised by an analog to
digital converter to create a data set that is stored in the memory. The data set is processed by the
microprocessor and then sent to the display.
To digitise the analog signal, analog to digital (A/D) converter is used. The output of the vertical
amplifier is applied to the AID converter section. The main requirement of A/D converter in the
digital storage oscilloscope is its speed, while in digital voltmeters accuracy and resolution were
the main requirements. The digitised output needed only in the binary form and not in BCD. The
successive approximation type of AID converter is most oftenly used in the digital storage
oscilloscopes.
Modes of operation:
The digital storage oscilloscope has three modes of operation:
1. Roll mode ii) Store mode iii) Hold or save mode.
Roll mode:
This mode is used to display very fast varying signals, clearly on the screen. The fast varying signal
is displayed as if it is changing slowly, on the screen. In this mode, the input signal is not triggered
at all.