L1050102 - Frontier IR Dual-Range Systems User's Guide

Download as pdf or txt
Download as pdf or txt
You are on page 1of 106

MOLECULAR SPECTROSCOPY

FRONTIER IR DUAL-RANGE SYSTEMS

User’s Guide
Release History

Part Number Release Publication Date

L1050102 B February 2012

Any comments about the documentation for this product should be addressed to:

User Assistance
PerkinElmer Ltd
Chalfont Road
Seer Green
Beaconsfield
Bucks HP9 2FX
United Kingdom

Or emailed to: info@perkinelmer.com

Notices

The information contained in this document is subject to change without notice.


Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no
warranty of any kind with regard to this document, including, but not limited to, the
implied warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in
connection with furnishing, performance or use of this material.

Copyright Information

This document contains proprietary information that is protected by copyright.


All rights are reserved. No part of this publication may be reproduced in any form whatsoever or
translated into any language without the prior, written permission of PerkinElmer, Inc.

Copyright © 2012 PerkinElmer, Inc.

Trademarks

Registered names, trademarks, etc. used in this document, even when not specifically marked as such,
are protected by law.
PerkinElmer is a registered trademark of PerkinElmer, Inc.
Spectrum, AssureID and Frontier are trademarks of PerkinElmer, Inc.
Contents
Introduction ............................................................................................... 5
About This Manual ......................................................................................... 6
Conventions Used in this Manual ..................................................................... 7
Notes, Cautions and Warnings .................................................................. 7
Warnings and Safety Information ........................................................... 11
Safety Summary .......................................................................................... 12
General Safety ............................................................................................ 13
Location and Ventilation ............................................................................... 14
To allow for adequate cooling ................................................................. 14
To allow any nitrogen used to disperse .................................................... 14
Use of flammable solvents and samples ................................................... 15
Electrical Safety ........................................................................................... 16
Laser Safety Regulations .............................................................................. 17
Radiation Hazards and their Classification ....................................................... 18
Laser radiation ...................................................................................... 18
Radiation emitted by an NIR source ........................................................ 18
Labels ........................................................................................................ 19
Warning Signs on the Instrument .................................................................. 21
EMC Compliance .......................................................................................... 22
European EMC Directive ......................................................................... 22
FCC rules and regulations ....................................................................... 22
An Overview of the Frontier IR Dual-Range Spectrometers ................... 23
A Guided Tour of the Frontier IR Dual-Range Spectrometers ............................ 24
Configuration changes ........................................................................... 25
Optical system ...................................................................................... 25
Top panel controls ................................................................................. 26
The sample compartment ....................................................................... 27
Internal accessories ............................................................................... 28
External accessories .............................................................................. 28
The Spotlight 400 Series Imaging System ................................................ 29
Sample compartment windows ............................................................... 30
Power switch and communications ports .................................................. 30
Accessories for the Frontier IR Spectrometer .................................................. 31
Frontier IR Sytems Upgrades ........................................................................ 33
Unpacking and Installation ...................................................................... 35
Requirements .............................................................................................. 36
Electrical requirements........................................................................... 36
Environment ......................................................................................... 36
Unpacking the spectrometer ......................................................................... 37
Opening the shipping container............................................................... 37
The Desiccant Indicator ................................................................................ 39
Connecting up the Spectrometer ................................................................... 40
Connecting to the PC ............................................................................. 40
Other connectors .................................................................................. 41
Connecting the spectrometer to the electrical supply ................................. 42
Installing the Instrument in the Software ....................................................... 44
Installing the software ........................................................................... 44
The Instrument Install Wizard................................................................. 44
Using the Spectrometer with Spectrum .................................................. 45
Basics of Software Control ............................................................................ 46
Starting Spectrum software .................................................................... 46
Scanning samples.................................................................................. 47
Changing the instrument range ............................................................... 48
Changing the beampath ......................................................................... 49
Working with the instrument display and Go button .................................. 49
Using the Spectrum on-screen Help system .............................................. 50
Atmospheric (CO2/H2O) Suppression .............................................................. 51
What is atmospheric suppression? ........................................................... 51
What does atmospheric suppression do? .................................................. 51
AVI Correction ............................................................................................. 53
What is AVI correction? .......................................................................... 53
What does AVI correction do? ................................................................. 53
Look Ahead ................................................................................................. 55
What is Look Ahead? ............................................................................. 55
What does Look Ahead do? .................................................................... 55
Quality Checks ............................................................................................ 56
What are Quality Checks? ....................................................................... 56
What do Quality Checks do? ................................................................... 56
Routine Maintenance ............................................................................... 57
Cleaning the spectrometer ............................................................................ 58
Cleaning the display............................................................................... 58
Moving the Spectrometer .............................................................................. 59
Condensation ........................................................................................ 60
The Set Temperature utility .................................................................... 60
Replacing the Frontier FT-IR/FT-FIR Sample Compartment Windows ................... 61
The Desiccant Indicator in Detail ................................................................... 64
Changing the Desiccant ................................................................................ 65
Renewing the instrument desiccant ......................................................... 66
Installing rechargeable desiccant in the instrument ................................... 67
Renewing the desiccant in the spare window compartment ........................ 68
Purging the Spectrometer ............................................................................. 70
Changing the External Fuse .......................................................................... 73
Cooling the MCT Detector (If Fitted) .............................................................. 75
Advanced Maintenance ............................................................................ 79
Opening the Main Cover ............................................................................... 80
Replacing a Source ...................................................................................... 83
Installing/Replacing Beam Port and Sample Compartment Windows .................. 85
Installing Filters in the Filter Wheel ................................................................ 87
Replacing the Laser and Power Supply ........................................................... 91
Appendices ............................................................................................... 95
Appendix 1: Changing the Sample Area Accessory ........................................... 96
Appendix 2: Instrument Self-Checks .............................................................. 98
Appendix 3: Instrument Performance Validation ............................................. 100
Appendix 4: Decontamination and Cleaning ................................................... 101
Appendix 5: WEEE Instructions for PerkinElmer Products ................................ 102
Index ........................................................................................................ 103
Introduction
6 . Frontier IR Dual-Range Systems User's Guide

About This Manual

This manual contains the following sections:

• Introduction

• Warnings and Safety Information

• An Overview of the Frontier-IR Dual-Range Spectrometers

• Unpacking and Installation

• Using the Spectrometer with Spectrum

• Routine Maintenance

• Advanced Maintenance

• Appendices.

For further information on collecting, viewing and processing spectra using Spectrum
software, refer to the help file that you can access from the Help menu, or by clicking Help
on a dialog. The Help menu also includes links to on-screen tutorials.

NOTE: This manual shows details for using your instrument with the Spectrum software
package (version 10 or later). If you have Spectrum ES or AssureID software, please
refer to the Administrator’s Guide for your software which can be found on the
IR & Raman Manuals CD (L1050002), or refer to the on-screen Help.
Introduction . 7

Conventions Used in this Manual


Normal text is used to provide information and instructions.
Bold text refers to text that is displayed on the screen.
UPPERCASE text, for example ENTER or ALT, refers to keys on the PC keyboard. '+' is
used to show that you have to press two keys at the same time, for example, ALT+F.
All eight digit numbers are PerkinElmer part numbers unless stated otherwise.
The term “instrument” refers to a Frontier IR spectrometer, and any sampling accessory
fitted. Frontier IR Dual-Range spectrometer refers to the Frontier FT-IR/FT-NIR and the
Frontier FT-IR/FT-FIR spectrometers.

Notes, Cautions and Warnings


Three terms, in the following standard formats, are also used to highlight special
circumstances and warnings.
NOTE: A note indicates additional, significant information that is provided with some
procedures.
8 . Frontier IR Dual-Range Systems User's Guide

We use the term CAUTION to inform you about situations that could
CAUTION result in serious damage to the instrument or other equipment. Details
about these circumstances are in a box like this one.

Caution (Achtung)
Bedeutet, daß die genannte Anleitung genau befolgt werden muß, um
einen Geräteschaden zu vermeiden.

Caution (Bemærk)
Dette betyder, at den nævnte vejledning skal overholdes nøje for at undgå
en beskadigelse af apparatet.

Caution (Advertencia)
Utilizamos el término CAUTION (ADVERTENCIA) para advertir sobre
situaciones que pueden provocar averías graves en este equipo o en otros.
En recuadros éste se proporciona información sobre este tipo de
circunstancias.

Caution (Attention)
Nous utilisons le terme CAUTION (ATTENTION) pour signaler les
situations susceptibles de provoquer de graves détériorations de
l'instrument ou d'autre matériel. Les détails sur ces circonstances figurent
dans un encadré semblable à celui-ci.

Caution (Attenzione)
Con il termine CAUTION (ATTENZIONE) vengono segnalate situazioni
che potrebbero arrecare gravi danni allo strumento o ad altra
apparecchiatura. Troverete informazioni su tali circostanze in un
riquadro come questo.

Caution (Opgelet)
Betekent dat de genoemde handleiding nauwkeurig moet worden
opgevolgd, om beschadiging van het instrument te voorkomen.

Caution (Atenção)
Significa que a instrução referida tem de ser respeitada para evitar a
danificação do aparelho.

Caution (小心)
我们使用“小心”这一术语来通知您有关可能会对
本仪器或其它设备造成严重损害的情况。
有关这些情况的详细信息可在此类方框中找到。
Caution (注意)
分光器や他の機材等に深刻なダメージを与える恐れがある場合は、
この様なボックスの中に表示しています。
Introduction . 9

We use the term WARNING to inform you about situations that could result in
personal injury to yourself or other persons. Details about these
circumstances are in a box like this one.
WARNING

Warning (Warnung)
Bedeutet, daß es bei Nichtbeachten der genannten Anweisung zu einer
Verletzung des Benutzers kommen kann.

Warning (Advarsel)
Betyder, at brugeren kan blive kvæstet, hvis anvisningen ikke overholdes.

Warning (Peligro)
Utilizamos el término WARNING (PELIGRO) para informarle sobre
situaciones que pueden provocar daños personales a usted o a otras
personas. En los recuadros como éste se proporciona información sobre este
tipo de circunstancias.

Warning (Danger)
Nous utilisons la formule WARNING (DANGER) pour avertir des situations
pouvant occasionner des dommages corporels à l'utilisateur ou à d'autres
personnes. Les détails sur ces circonstances sont données dans un encadré
semblable à celui-ci.

Warning (Pericolo)
Con il termine WARNING (PERICOLO) vengono segnalate situazioni che
potrebbero provocare incidenti alle persone. Troverete informazioni su tali
circostanze in un riquadro come questo.

Warning (Waarschuwing)
Betekent dat, wanneer de genoemde aanwijzing niet in acht wordt genomen,
dit kan leiden tot verwondingen van de gebruiker.

Warning (Aviso)
Significa que a não observância da instrução referida poderá causar um
ferimento ao usuário.

Warning (警告)
我们使用“警告”这一术语来通知您有关可能会对您自己或他人造成人
身伤害的情况。
有关这些情况的详细信息可在此类方框中找到。
Warning (警告)
使用者及びその他周辺に危害が及ぶ恐れがある場合は、
この様なボックスの中に注意事項が表示されています。
10 . Frontier IR Dual-Range Systems User's Guide
Warnings and Safety
Information
12 . Frontier IR Dual-Range Systems User's Guide

Safety Summary

The Frontier IR spectrometers have been designed to comply with a wide variety of
international standards governing the safety of laboratory equipment. In routine use, the
instruments pose virtually no risk to you. If you take some simple, common-sense
precautions, you can make sure that you maintain the continued safe operation of your
instrument:

DO make sure that the instrument is properly connected to the electrical supply; in particular
make sure that the ground (earth) is securely connected.

DO disconnect the electrical power cable before opening the main cover of the instrument.

DO keep the instrument dry. Avoid spilling liquid into the instrument. Clean all external spills
immediately. If anything that is spilled enters the main body of the instrument, switch off the
power and contact a PerkinElmer Service Engineer.

DO NOT stare into the internal laser beam under the instrument cover. The instrument
contains a low power, visible (red) laser; momentary exposure to the beam is not dangerous,
but deliberate, direct viewing of the beam along its axis could damage your eye.

DO NOT use a flammable gas to purge the instrument. The instrument contains a hot source,
and a fire or explosion will result. Only use clean, dry, oil-free nitrogen or air to purge the
instrument.

DO read the more detailed information on warnings and safety in the following pages to
ensure the safe operation of the instrument.
Warnings and Safety Information . 13

General Safety

The Frontier IR spectrometers have been designed and tested in accordance with
PerkinElmer specifications and in accordance with the safety requirements of the
International Electrotechnical Commission (IEC). The instruments conform to IEC publication
61010-1 (“Safety requirements for electrical equipment for measurement, control, and
laboratory use”) as it applies to IEC Class 1 (earthed) appliances and therefore meets the
requirements of EC low voltage Directive 2006/95/EC.

If the instrument is used in a manner not specified by the manufacturer, the protection
provided by the instrument may be impaired. Only use the instrument indoors and under the
following conditions:

Temperature 15 °C to 35 °C
Relative Humidity 80% maximum (non-condensing)

If possible, avoid any adjustment, maintenance and repair of the opened, operating
instrument. If any adjustment, maintenance and repair of the opened, operating instrument
is necessary, this must only be done by a skilled person who is aware of the hazard involved.

Whenever it is likely that the instrument is unsafe, make it inoperative. The instrument may
be unsafe if it:

• Shows visible damage

• Fails to perform the intended measurement

• Has been subjected to prolonged storage in unfavorable conditions

• Has been subjected to severe transport stresses.

If the equipment is used in a manner not specified herein the protection


provided by the equipment may be impaired.

WARNING

The instrument has been designed to be safe under the following environmental conditions:

• Indoor use

• Altitude up to 2000 m (above mean sea level)

• Ambient temperatures of 5 °C to 40 °C

• A maximum ambient relative humidity of 80% for temperatures up to 31 °C, decreasing


linearly to 50% relative humidity at 40 °C

• Mains supply fluctuations not exceeding ±10% of the nominal voltage.


14 . Frontier IR Dual-Range Systems User's Guide

Location and Ventilation

Make sure that the switch at the electrical supply inlet on the rear of the
instrument is not obstructed.

WARNING

To allow for adequate cooling


Do not site the instrument near to room heating equipment, for example, central heating
radiators.

During operation, there should be a minimum gap of:

• 15 cm (6 inches) between any surface and the cooling louvers at the rear of the
instrument.

• 7 cm (3 inches) between the instrument and adjacent equipment.

• 45 cm (18 inches) between any surface and the top surface of the closed sample area
lid (to allow for the lid to be opened fully).

To allow any nitrogen used to disperse

Do not site the instrument in a poorly ventilated area if nitrogen will be


used as a purge gas, or if liquid nitrogen will be used to cool a detector.

Oxygen depletion in an enclosed space does not trigger a gasping reflex,


WARNING
and errors of judgment, confusion, or unconsciousness can occur in
seconds and without warning.

The spectrometer includes a coupling that enables the instrument body, and another
coupling that enables the sample compartment, to be purged using clean, dry, oil-free air or
nitrogen. The recommended flow rate to each connector is 10 l/min, and both the instrument
and the sample compartment vent to their surroundings.

If the instrument is fitted with an MCT detector, this is cooled using liquid nitrogen. A risk
assessment should include personal protection protocols, and proper actions in the event of
accidental spillage. 1 l (2 US pints) of liquid nitrogen evolves to 700 l (approximately
25 cubic feet) of nitrogen gas, and the cold vapor can pool at floor level.
Warnings and Safety Information . 15

Use of flammable solvents and samples

The instrument contains a hot source and contact with flammable vapors
may cause an explosion. When working with flammable solvents or
samples, particularly during unattended operation with flow-cells, it is
recommended that the instrument optics area should be continuously
WARNING
purged with dry air or nitrogen to maintain a positive pressure and
prevent flammable vapor from entering the instrument.

If flammable solvents or samples are spilled on the instrument and there


is any possibility that they have entered the interior (by coming into
contact with cover gaskets, for example) then the instrument must be
switched off immediately and disconnected from the power supply. The
WARNING
optics area should then be thoroughly purged with dry air or nitrogen, or
the main cover should be opened to thoroughly ventilate the optics area
before proceeding.

Flammable solvents or samples should not be stored on or near the


instrument. Handling of such materials during preparation should be
performed in a safe area away from the instrument such as a fume
cabinet.
WARNING

Do not use a flammable gas to purge the Frontier IR instruments. Use


only clean, dry, oil-free nitrogen or air.

WARNING
16 . Frontier IR Dual-Range Systems User's Guide

Electrical Safety

• Connect the instrument to a power supply line that includes a switch or other means of
disconnection from the electricity supply.

• Only plug the instrument into an electricity-supply socket that is provided with a
protective earth connection.

• When fuses need replacing, use only those with the required current rating and of the
specified type. Do not use makeshift fuses and do not short-circuit fuse holders.

• When the instrument is connected to its electricity supply, terminals may be live and the
removal of covers other than those which can be removed by hand is likely to expose
live parts.

• Capacitors inside the instrument may still be charged even if the instrument has been
disconnected from all voltage sources.

• The instrument must be disconnected from all voltage sources before it is opened for
any adjustment, replacement, maintenance or repair.

Any interruption of the protective earth conductor inside or outside the


instrument or disconnection of the protective earth terminal can make the
instrument dangerous.
WARNING

The instrument has an IEC Insulation Class I rating for external circuits – only connect other
equipment that meets the requirements of IEC 61010-1, IEC 60950 or equivalent standards.
Warnings and Safety Information . 17

Laser Safety Regulations


The Frontier IR spectrometers are Class 1 laser products as defined by IEC 60825-1. The
optical module contains a Class 2 Helium Neon (HeNe) laser, which emits visible, continuous
wave radiation at a wavelength of 633 nm and has a maximum output power of 1 mW. Some
diffuse HeNe laser radiation, within Class 1 limits, emerges from:

• The window in the left side of the sample compartment when an internal beam path is
selected.

• An external beam port when the port cover is removed, no accessory is fitted at the
port, and the beampath to the port is selected.

Do not stare into any laser beam. Staring into a laser beam (intrabeam
viewing) can cause permanent damage to your eyes.

WARNING

The laser is automatically shut down when the main cover of the instrument is raised.

Do not attempt to override or modify the interlock system.

Use of controls or adjustments or performance of procedures other than


those specified herein may result in hazardous radiation exposure.
WARNING

The instrument complies with the following laser safety regulations:

1. 21 CFR Part 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50
dated 24 June 2007. Administered by the Center for Devices and Radiological Health,
U.S. Department of Health and Human Services.

2. European Standard EN 60825-1:2007 – “Safety of laser products – Part 1: Equipment


classification and requirements”.
18 . Frontier IR Dual-Range Systems User's Guide

Radiation Hazards and their Classification

Laser radiation
Indirect observation of the laser beam radiation in the optical path is not hazardous. Directly
viewing the laser beam along its axis (allowing the laser beam radiation to pass into the eye)
can be hazardous, depending upon the power of the beam, the length of time that the eye is
exposed to the beam and the optical efficiency of the exposed eye. Direct viewing of a laser
beam along its axis is termed intrabeam viewing.
Protection of the eye during accidental, momentary intrabeam viewing of a Class 2 laser
beam is normally given by the eye’s aversion response, including the blink reflex, which limits
exposure of the eye to less than 0.25 seconds.
Class 1 levels of laser radiation are not considered to be hazardous.

Radiation emitted by an NIR source


Your Frontier IR instrument can be fitted with an NIR source. The quartz halogen bulb, which
produces the near infrared beam, emits ultraviolet, visible and infrared radiation. The
majority of this radiation is in the infrared region. Do not stare into the beam produced by
this bulb.

Measurements of the infrared radiation emitted from the sample area show that exposure
limits recommended by the American Conference of Governmental Industrial Hygienists
(ACGIH) and International Commission on Non-Ionizing Radiation Protection (ICNIRP) will
not be exceeded during normal operation.
Warnings and Safety Information . 19

Labels

When this label is attached to an instrument it means “Caution, risk of


danger”. Refer to the manual to find out the nature of the potential
hazard and any actions that have to be taken.

The product identification label is on the front of the instrument. Other labels are fixed to the
Frontier IR spectrometers in the locations shown in Figure 1 and Figure 2:

Figure 1 Labels (rear of instrument)

NOTE: The label with a crossed-out wheeled bin symbol and a rectangular bar indicates that
the product is covered by the Waste Electrical and Electronic Equipment (WEEE)
Directive. Refer to Appendix 5: WEEE Instructions for PerkinElmer Products on
page 102.
20 . Frontier IR Dual-Range Systems User's Guide

Serial number, part


number and date of
manufacture

NOTE: There is a second copy of the CAUTION label inside the sample compartment cover.
Figure 2 Labels (sample compartment)
Warnings and Safety Information . 21

Warning Signs on the Instrument

Caution, hot surface.

Caution, risk of electric shock.

Caution, laser radiation hazard.

Caution, risk of danger.

Refer to accompanying documents in all cases where this symbol is used


to find out the nature of the potential hazard and any actions which have
to be taken.
22 . Frontier IR Dual-Range Systems User's Guide

EMC Compliance

European EMC Directive


The Frontier IR spectrometers have been designed and tested to meet the requirements of
the EMC Directive 2004/108/EC.

FCC rules and regulations


These products are classified as digital devices used exclusively as industrial, commercial, or
medical test equipment. They are exempt from the technical standards specified in Part 15 of
the FCC Rules and Regulations, based on Section 15.103(c).
An Overview of the
Frontier IR Dual-Range
Spectrometers
24 . Frontier IR Dual-Range Systems User's Guide

A Guided Tour of the Frontier IR Dual-Range


Spectrometers

PerkinElmer Frontier IR Dual-Range spectrometers are bench-top FT-IR instruments


(Figure 3). Fully optimized performance is achieved by the provision of dual beamsplitters,
dual sources (Frontier FT-IR/FT-NIR only) and dual detectors.

Figure 3 Frontier IR FT-IR/FT-NIR spectrometer

For the MIR configuration:

• The optical system enables you to collect data over a total range of 7800 to 350 cm−1
with a best resolution of 0.4 cm−1.

• A DTGS (deuterated triglycine sulfate) MIR detector or an MCT (mercury cadmium


telluride) detector is fitted.

• KBr windows are fitted in the sample compartment.

For the NIR configuration:

• The optical system enables you to collect data over a total range of 14700 to 1250 cm−1
with a best resolution of 0.4 cm−1.

• An NIR DTGS (deuterated triglycine sulfate) detector is fitted.

• KBr windows are fitted in the sample compartment.

For the FIR configuration:

• The optical system enables you to collect data over a total range of 700 to 30 cm−1 with
a best resolution of 0.4 cm−1.

• An FIR DTGS (deuterated triglycine sulfate) detector is fitted.

• A PE (polyethylene) window and a PE/D (polyethylene/diamond) window are supplied to


be fitted in the sample compartment for operation in the FIR range.

The instrument can operate in ratio, single-beam, or interferogram mode.


An Overview of the Frontier IR Dual-Range Spectrometers . 25

The instrument is connected to a PC, either point-to-point or over a network. The Spectrum
software package supplied enables you to control the instrument and to manipulate the
spectra that you collect.

Configuration changes
Spectrum software enables you to initiate an automated changeover between, for example,
the MIR and NIR ranges, changing the source, detector, and beamsplitter. You can also
individually select any combination of these elements.

NOTE: If you change between the MIR and FIR ranges, the software will prompt you to
change the sample compartment windows. For information about changing windows,
see Replacing the Frontier FT-IR/FT-FIR Sample Compartment Windows on page 61.

• When you change the source or detector, or swap between an external and internal
accessory, the infrared beam is redirected using the appropriate rotating mirror.

• Before you change to the MIR MCT detector, fit one of the attenuators supplied
(Attenuator Set – L1160560) to the window to the right of the sample compartment to
prevent the detector from overloading. If you have an open beam we recommend a
14%T attenuator. See Replacing the Frontier FT-IR/FT-FIR Sample Compartment
Windows on page 61 or Installing/Replacing Beam Port and Sample Compartment
Windows on page 85 for more information.

NOTE: If you have a Frontier FT-IR/FT-FIR spectrometer with an MCT detector and the
instrument was in the MIR range when it was switched off, we recommended that a
6%T attenuator is fitted to the sample compartment window before switching on the
instrument again, to prevent errors during the initialization of the FIR detector.

• When you change the detector, allow it to equilibrate before scanning. The MCT
detector (if fitted) will not perform correctly until it has been sufficiently cooled using
liquid nitrogen.

• When you select the alternate beamsplitter, there is a pause while the element is
changed and initialized. The performance of the beamsplitter is temperature sensitive,
so there is also an equilibration period.

Optical system
The optical system is under the main cover of the instrument. Usually the main cover of the
instrument is closed, but to perform most maintenance tasks, the cover has to be open.
When you do this, a safety interlock automatically switches off the power. Nevertheless, the
instrument should be disconnected from the mains before opening the main cover for
maintenance.

Consistent, reliable performance is achieved by having few adjustable parts, and by


extensive insulation of the optical system from the effects of humidity and vibration.
26 . Frontier IR Dual-Range Systems User's Guide

Stability of the optical system

The entire optical system is purged and sealed at the factory. A supply of desiccant placed
within the system removes any water vapor and carbon dioxide that may enter. A desiccant
indicator is fitted in the top cover, which warns you when the desiccant needs changing (an
internal humidity sensor is available as an option).

KBr windows (or polyethylene and polyethylene/diamond for the FIR range) separate the
sample compartment from the purged optical system. You can purge the sample
compartment with clean, dry, oil-free air or nitrogen. Either one removes water vapor;
however, nitrogen is preferable because it also removes atmospheric carbon dioxide.

Do not use a flammable gas to purge the Frontier IR instruments. The


instrument contains a hot source, and a fire or explosion will result. Only
use clean, dry, oil-free nitrogen or air to purge the instrument.
WARNING

The optics are kinematically mounted to ensure accurate positioning and to make them
rugged. The interferometer is enclosed and mounted on anti-vibration mounts to guard
against air and bench borne disruptions. The interferometer uses very low-friction point
bearings and a frictionless electromagnetic drive to ensure long life.

Top panel controls


The display on the front right of the top of your instrument (Figure 4) has two purposes:

• To display messages generated by the instrument’s firmware, such as those that


monitor initialization and diagnostics when the instrument is switched on.

• To display prompts and other messages generated by Spectrum software.

The instrument’s Go button is used in concert with the displayed prompts and other
messages generated by Spectrum software. It allows you to Scan or Halt data collection
while you are away from your computer.

Display

Go button

Figure 4 Frontier IR top panel controls


An Overview of the Frontier IR Dual-Range Spectrometers . 27

The sample compartment


The instrument has a large, purgeable sample compartment (Figure 5) located at the front of
the instrument.

Sample compartment lid

Baseplate

Figure 5 Frontier IR sample compartment

The windows in the sample compartment and external beam ports are
composed of KBr (for MIR and NIR configurations) that, although coated,
can be damaged by high levels of humidity.
CAUTION
If you spill a liquid in the sample compartment, wipe it up quickly. When
working with water-based samples open to the air, either purge the
sample compartment or leave the cover open.

A relative humidity higher than 80% can damage the windows of the
CAUTION sample compartment. If you expect the humidity to exceed 80%,
continually purge or desiccate the sample compartment.

Open the sample compartment by lifting the cover using the recess at the front.
The infrared beam enters the sample compartment through an aperture on the left. After
passing through the sample, it enters the detector area through an aperture on the right side
of the sample compartment.

Standard sampling accessories are mounted on a baseplate. The standard baseplate includes
a sample holder located by a knurled screw.
28 . Frontier IR Dual-Range Systems User's Guide

Internal accessories
A wide range of optional MIR and NIR accessories, such as the Universal ATR accessory, fit
in the sample compartment (Figure 6).

Figure 6 Frontier IR with Universal ATR accessory

It is easy to remove an internal accessory and to replace it with another. See Appendix 1:
Changing the Sample Area Accessory on page 96.

The instrument identifies the type of accessory fitted using a coded connector on its
baseplate that plugs into a socket on the rear wall of the sample compartment.

External accessories
The external beam port on the right side of the instrument is used, for example, with the
optional External Near Infra Red Accessory (External NIRA). Spectrum software enables you
to switch the instrument between this external accessory and another accessory fitted in the
sample compartment.

Figure 7 Frontier IR with External NIRA accessory

NOTE: To use your instrument with an external accessory, you require the optional External
Beam Pack. If you did not purchase the External Beam Pack with your spectrometer,
it is available as a field-fitted upgrade (L1280248).

For information about removing or refitting an external accessory, refer to its User’s Guide,
which is distributed on the IR & Raman Manuals CD (L1050002) as a .pdf file.
An Overview of the Frontier IR Dual-Range Spectrometers . 29

Do not attempt to remove the external accessory bracket mounted within


the handhold on the right of the instrument or to transfer the bracket to
CAUTION another spectrometer.
The external accessory bracket is fitted by a PerkinElmer Service
Engineer, along with any other internal components that may be required.

Do not attempt to lift the instrument, or to move it to another location,


when an external accessory is attached.
CAUTION
The optical alignment of the accessory may be disturbed.

For information about moving the instrument, see Moving the Spectrometer on page 59.

The Spotlight 400 Series Imaging System


The external port on the left side of the instrument is used with the Spotlight 400 Imaging
System (Figure 8) and SpectrumIMAGE software.

Figure 8 Spotlight 400 Series Imaging System

NOTE: To use your instrument with a microscope, you require the optional External Beam
Pack. If you did not purchase the External Beam Pack with your spectrometer, it is
available as a field-fitted upgrade (L1280248).
30 . Frontier IR Dual-Range Systems User's Guide

Sample compartment windows


The Frontier FT-IR/FT-NIR instrument has KBr windows fitted in the sample compartment. It
is possible to change the sample compartment windows if they are not suitable for your
application. For information about changing windows, see Installing/Replacing Beam Port and
Sample Compartment Windows on page 85.

The Frontier FT-IR/FT-FIR instrument has KBr windows fitted in the sample compartment for
the MIR range, but a PE (polyethylene) window and a PE/D (polyethylene/diamond) window
(supplied) should be fitted for the FIR range. For information about changing these windows,
see Replacing the Frontier FT-IR/FT-FIR Sample Compartment Windows on page 61.

Changing windows allows some CO2 and water vapor from the
surroundings to enter the instrument.
CAUTION
The instrument desiccant packs should remove these contaminants within
30 minutes.

A desiccated compartment is provided for storing spare windows.

NOTE: For long-term storage of spare windows we recommend that you use a desiccator.

Compartment
desiccant indicator
Spare window

Desiccated
compartment

Figure 9 Desiccated compartment for spare windows

Power switch and communications ports


The power switch, AC power cable connector, and communications ports are on the rear of
the instrument. The power switch is marked I/O (on/off).

NOTE: It can take the instrument about two hours to equilibrate when switched on after
being switched off overnight. To save time, we suggest that you leave the instrument
switched on at all times.
An Overview of the Frontier IR Dual-Range Spectrometers . 31

Accessories for the Frontier IR Spectrometer

A wide range of optional accessories, such as the Universal ATR accessory (UATR), fit in the
sample compartment. In addition, a number of external accessories are available, including
the External NIRA. An external accessory can be fitted in conjunction with another internal
accessory (Figure 10).

Figure 10 Frontier IR spectrometer with Universal ATR and External NIRA


accessories

There is a range of specialized sampling accessories available for the Frontier IR instruments,
as shown in Table 1.
Table 1 Sampling accessories for the Frontier IR instruments

Accessory Type MIR NIR FIR

Sample Shuttle Internal Yes Yes Yes

Horizontal ATR Internal Yes n/a n/a

Universal ATR Internal Yes n/a n/a

Diffuse Reflectance Internal Yes n/a Yes

NIRA Internal (limited utility) Yes n/a

External NIRA External (limited utility) Yes n/a

NIR Fiber Optic Probe Internal (limited utility) Yes n/a

NIR Tablet Autosampler* Internal n/a Yes n/a

Liquid Sipper* Internal Yes Yes n/a

TL 8000 EGA System (TG-IR) – Yes Yes No


32 . Frontier IR Dual-Range Systems User's Guide

Accessory Type MIR NIR FIR

General Purpose Optical Bench


(GPOB) and External MCT External Yes Yes n/a
Detector**

General Purpose Optical Bench


(GPOB) and External LiTaO3 External Yes Yes n/a
Detector**
*The NIR Tablet Autosampler accessory is supported by AssureID software and Spectrum software
version 6.x. It is not supported in Spectrum software version 10.0 or later.
** For more details of the GPOB options, contact your PerkinElmer Sales Representative.

Further information about the use of these accessories can be found on the IR & Raman
Manuals CD (L1050002).
An Overview of the Frontier IR Dual-Range Spectrometers . 33

Frontier IR Sytems Upgrades

The Frontier IR systems are designed to be flexible, upgradable instruments. The upgrades
available for Frontier IR Dual-Range Systems are listed in Table 2.
Table 2 Frontier IR Dual-Range Systems Upgrades

Part Number Description

L1280248 External Pack Beam Upgrade – Microscope, GPOB, NIRA and


external beam input port ready
The kit contains KBr windows, but other windows are available. Ask
your PerkinElmer Customer Service Representative for details

L1280252 Frontier FT-IR/FT-NIR Narrow band MCT Detector Upgrade


(replaces standard MIR detector)

L1280253 Frontier FT-IR/FT-NIR Medium band MCT Detector Upgrade


(replaces standard MIR detector)

L1280254 Frontier FT-IR/FT-NIR Wide band MCT Detector Upgrade


(replaces standard MIR detector)

L1280255 Frontier FT-IR/FT-FIR Narrow band MCT Detector Upgrade


(replaces standard MIR detector)

L1280256 Frontier FT-IR/FT-FIR Medium band MCT Detector Upgrade


(replaces standard MIR detector)

L1280257 Frontier FT-IR/FT-FIR Wide band MCT Detector Upgrade


(replaces standard MIR detector)

L1200391 Electronic Humidity Sensor


34 . Frontier IR Dual-Range Systems User's Guide
Unpacking and
Installation
36 . Frontier IR Dual-Range Systems User's Guide

Requirements

NOTE: Read the warnings and safety information at the start of this manual before you
install the instrument. They contain important information.

Electrical requirements
The Frontier IR instruments can operate on electricity supplies of 50 or 60 Hz and in the
100 V to 230 V range without any adjustment.

The nominal power consumption of the instrument is 120 VA.

The line supply must be within 10% of the nominal voltage.

If possible, do not connect the instrument to circuits that have heavy-duty equipment
connected, such as large motors.

If possible, do not use photocopiers, discharge lamps, radio transmitters, and other
equipment with large or frequent transient loads on the same supply circuit.

The primary fuse (2 AT, 250 V) is in the drawer on the rear of the instrument next to the
mains inlet: the spare fuse is in the same drawer. The primary fuse is connected in the live
line.

Environment
To obtain the best performance from your instrument:

• Place the instrument in an environment that is relatively dust-free.


• Make sure that the bench top is free from vibration or mechanical shocks.
• Do not place the instrument or the PC near to room-heating equipment, such as
central-heating radiators.
• Do not position the instrument in direct sunlight, as this may cause overheating.
• Leave at least 15 cm (6 inches) from any surface and the cooling louvers at the rear of
the instrument.
• Leave at least 7 cm (3 inches) from any vertical obstacle to the sides of the instrument,
to permit an adequate flow of cooling air.
• Make sure that there are no overhanging shelves, and no water pipes or faucets that
could leak onto the instrument.
• The area near the PC must be free of strong magnetic fields, direct sunlight, and
heating or cooling units or ducts.

The instrument has been designed for indoor use and operates correctly under the following
conditions:

Ambient temperature 15 °C to 35 °C
Ambient relative humidity 80% maximum (non-condensing).
Unpacking and Installation . 37

Unpacking the spectrometer

The Frontier IR spectrometer is a heavy precision instrument, so two


people are required for safe handling.

The instrument weighs approximately 36 kg unpacked (40 kg packed)


WARNING
and has a lifting recess on either side. Consult the local codes of practice
issued by safety advisors before attempting to lift it.

Take care not to injure yourself or others, or to drop the instrument.

Take great care when installing your Frontier IR spectrometer, and


CAUTION follow the procedures described in this manual. If you require assistance,
contact your local PerkinElmer Service Engineer.

Opening the shipping container


Your instrument is packed inside the box in a silver bag that protects it from condensation.
The beamsplitters are packed as a kit in a separate protective box.

1. First remove the manuals/software box, leads, and so on from the box, and check that
all the following parts are present:

Part Description
Number

L1250432 Quick Change Window Kit (Frontier FT-IR/FT-FIR only) 1

L1202057 Polystyrene Calibration Film 1

L1250405 NIR Performance Validation Kit (Frontier FT-IR/FT-NIR only) 1

04974265 Quick Release Purge Coupling 2

L1200466 Ethernet Crossover Cable 1

04970839 2A, 250V Time Lag Fuse 2

L9003693 8.0 mm Hexagonal Wrench 1

N0171159 Desiccant Kit (Frontier FT-IR/FT-FIR only) 1

L1050002 IR & Raman Manuals CD 1

LX108873 Spectrum Standard Software Kit 1

OR OR

LX108875 Spectrum ES Software Kit

L1250230 Spectrum Configuration Disk 1


38 . Frontier IR Dual-Range Systems User's Guide

Part Description
Number

L1240055 Cuvette Holder Assembly (Frontier FT-IR/FT-NIR only) 1

L1240056 Cell Holder Assembly and Disposable Cells (5) 1


(Frontier FT-IR/FT-NIR only)

L1160560 Attenuator Set: 1%T, 4%T, 6%T, 14%T and 32%T 1


(Frontier IR Systems with MCT MIR detector only)

L1250397 Frontier FT-IR/FT-NIR Beamsplitter Kit (as applicable) 1

OR OR

L1250438 Frontier FT-IR/FT-FIR Beamsplitter Kit (as applicable)

If any items are missing or damaged, contact your local PerkinElmer office.

2. Carefully remove the instrument from the shipping container, but not from the bag in
which it was shipped.

The spectrometer and beamsplitters must be allowed to reach the


CAUTION temperature of their surroundings. This means leaving them in their
packaging overnight if moved from a cold area, and for at least four hours
after removal from the shipping container.

3. When the spectrometer has been allowed to warm to the temperature of its
surroundings, remove it from the bag and place it on the bench where it is to be used.
Ensure that you can reach the rear of the spectrometer to connect the cables.

Do not open the beamsplitter kit, or attempt to fit the beamsplitters yourself.
CAUTION
The beamsplitters must be fitted by a PerkinElmer Service Engineer.

NOTE: Any accessories will be shipped in separate boxes.


Unpacking and Installation . 39

The Desiccant Indicator


The optical system of the spectrometer is purged at the factory. This protects the beam port
windows (and, once fitted, the beamsplitters) from being damaged by humidity. Replaceable
packs of desiccant keep the optics dry and free of CO2.

The top panel of the instrument includes a desiccant indicator, whose sectors change
sequentially from blue to pink as the desiccant becomes exhausted (Figure 11). Change the
desiccant in the instrument when the sector marked 10 becomes pink, but while the sectors
marked 15 and 20 are still blue. These numbers correspond to the approximate % Relative
Humidity in the instrument.

Refer to The Desiccant Indicator in Detail on page 61 for more information about the
appropriate action to take if the sectors marked 15 and 20 are pink.

Pink Blue

Blue

Figure 11 The desiccant indicator

If all three sectors of the desiccant indicator are pink, then you m ust
change the desiccant.
CAUTION
The instrument optics may be fogged. Do not switch the instrument
either ON or OFF until all sectors are blue.

NOTE: A similar desiccant indicator is fitted in the desiccated compartment (Figure 34 on


page 68). This is used for storing sample compartment windows when not in use.
40 . Frontier IR Dual-Range Systems User's Guide

Connecting up the Spectrometer

Connecting to the PC
NOTE: To control your instrument from a PC, use the crossover cable supplied.
To control your instrument over your network, use a standard Ethernet cable (not
supplied).

1. Plug one end of the cable into the connector port on the right-hand side of the
instrument (Figure 12).

PC or network
connection port

Figure 12 PC Connection port

2. Plug the other end of the cable into the network connection on your PC (if you are using
the crossover cable) or network hub (if you are using a standard Ethernet cable).

Connector details

Description Connector Voltage Maximum


Type Current

10/100 Base-T Ethernet connector. This is the Ethernet <5 V <100 mA


standard interface between the PC and the
instrument, or between a Local Area Network (LAN)
and the instrument. If connecting directly from the
PC to the instrument use the Category 5 UTP
Cross-Over cable supplied. If you are connecting to
your network, use a standard Ethernet cable.
DO NOT USE THE CROSS-OVER CABLE TO
CONNECT TO YOUR NETWORK.
Unpacking and Installation . 41

Other connectors
The communication ports for peripheral devices are shown in Figure 13.

Instrument purge

10101 Spt. Lt.

Sample area purge EXT.R

Figure 13 Communication ports

Legends and icons that identify the function of each of the ports are printed on the hinge
molding directly above the port.

Do not attempt to connect a monitor to either the EXT.R port or the


CAUTION MICROSCOPE port or you will cause serious damage to the instrument
when it is switched on.

Connector details

Icon Description Connector Voltages Maximum


Type Currents

10101 Serial port. Connects a PC to the 9-way +12 V <100 mA


instrument using an RS232 D-type
interface cable. This connector is
only used for service diagnostics. −12 V <100 mA

General Purpose Optical Bench 26-way +5 V <50 mA


(GPOB). Allows synchronization high density
between the instrument and an D-type
external accessory.
Spt.Lt Spotlight. Used to connect the 11-way mixed +5 V <100 mA
instrument to a Spotlight D-type
Imaging System. ±12 V 50 mA
42 . Frontier IR Dual-Range Systems User's Guide

Icon Description Connector Voltages Maximum


Type Currents

EXT.R Right external detector module. 15 way +12 V 0.65 A


Connects to an external detector high density
module. Usually the detector is D-type −12 V 0.65 A
located in an accessory placed on
the right of the instrument. +5 V 4A

Microscope external detector 15 way +12 V 0.65 A


module. Connects to a high density
PerkinElmer infrared microscope, D-type −12 V 0.65 A
placed on the left of the
instrument. +5 V 4A

Connecting the spectrometer to the electrical supply


The power cable for the electrical supply plugs into the rear of the instrument. It has a
molded socket at one end. If it is necessary to fit a plug on the power cable, use the wire
color code below:

Plug Pin Wire Color (100–120 V) Wire Color (220–240 V)

Ground (Earth) Green or Green/Yellow Green/Yellow

Live Black Brown

Neutral White Blue

To ensure safe and satisfactory operation of the instrument, it is


essential that the green or green/yellow ground (earth) wire of the
power cord is connected to a ground that complies with the regulations
of the local electricity supply authority (or equivalent body); ground
WARNING circuit continuity is essential for safe operation of the equipment.
Unpacking and Installation . 43

Frontier IR instruments can operate on electricity supplies of 50 or 60 Hz and in the 100 V to


230 V range without any adjustment.

1. Fit the molded socket of the power cable into the electrical supply inlet (Figure 14) of
the instrument.

Electricity
supply inlet

Figure 14 Electrical supply inlet

2. Switch on the power at the supply.

3. Switch on the instrument at the switch above the supply inlet.


I is on and O is off.

After a short diagnostic check, which takes about 2 minutes and is described in Appendix 2:
Instrument Self-Checks on page 98, your instrument will be ready to communicate with the
PC.
44 . Frontier IR Dual-Range Systems User's Guide

Installing the Instrument in the Software

Before you can use the instrument it must be set up in the software.

Installing the software


NOTE: If you are supplying your own computer, make sure that it meets the minimum
requirements for hardware and software set out in the “PC Requirements” section of
the Administrator’s Guide for your software, which can be found on the IR & Raman
Manuals CD (part number L1050002) supplied with your instrument.

 To install Spectrum software, insert the supplied DVD and follow the instructions on the
screen. Details of the installation program are given in the Administrator’s Guide.

The Instrument Install Wizard


After installing your software you will need to install your instrument.

 If you have Spectrum software or Spectrum ES software, select Add Instrument from
the Administration group on the Setup menu.
The Instrument Install Wizard starts.

OR

If you have AssureID software, select Configure Instruments from the Configure
Instruments and Accessories group on the Tools menu.
The Instrument Install Wizard starts.

Details of the Instrument Install Wizard can be found in the Administrator’s Guide for your
software, which can be found on the IR & Raman Manuals CD supplied with your instrument.
Using the Spectrometer
with Spectrum
46 . Frontier IR Dual-Range Systems User's Guide

Basics of Software Control

Starting Spectrum software


1. Switch on the power to the instrument using the switch on the rear of the instrument.
The instrument will initialize, which will take approximately 2 minutes.

NOTE: It can take the instrument about two hours to equilibrate when switched on after
being switched off overnight. To save time, we suggest that you leave the instrument
switched on at all times.

2. From the Start menu select Programs; the PerkinElmer Applications group; the
Spectrum sub-group and then the Spectrum application.

OR

Double-click the shortcut icon on the desktop.


The Spectrum start-up splash-screen is displayed, followed by a dialog that may require
your login details:

3. If required, enter your User name and Password, and then click OK.
If you already have an instrument set up in Spectrum or AssureID software for this user
on this PC, the Instrument Connection dialog may be displayed, unless Auto Connect is
set up for the instrument. If Auto Connect is enabled Spectrum will automatically
connect to the instrument.

4. Select the Instrument you want to use.

OR

If you want to work with data that has been collected previously, without connecting to
an instrument, select work offline.

NOTE: Working offline can free a networked instrument for use by another user.
Spectrum starts.
Using the Spectrometer with Spectrum . 47

Scanning samples
When you connect to your spectrometer the instrument settings will default to appropriate
values for your instrument type and accessory.

The configurable Scan toolbars at the top of the workspace include the tools you need to
collect a sample spectrum (Figure 15).

Figure 15 Frontier IR Dual-Range systems default scan toolbars

By default, the Measurement bar includes Scan, Halt, Background and Monitor buttons.
You can also select Scanalyze and then choose one of the scan and then process options
Scan and Compare, Scan and Search or Scan and Quant. All these commands can be
selected from the Measurement menu.

NOTE: The Scan toolbars are not displayed if you have chosen to work offline.

To scan a sample:

1. Check and set the instrument parameters.


Here you set the Start and End points of the scan range (by default in wavenumbers,
but the abscissa units can also be set to nanometers or microns on the Setup
Instrument Basic tab) and the Accumulations required, either as a number of scans,
or as a length of time. You can enter a unique Sample ID and Description for the
sample.
If required, you can also change the wavelength range of the spectrometer using the
Mode drop-down list. See Changing the instrument range on page 48 for more
information on configuring your Frontier Dual-Range spectrometer for different
wavelength ranges.

2. If a background scan is required, the Scan button includes a small background flag.
Clear the instrument beampath, or insert a suitable background material, and then click

to collect a background spectrum.


The background spectrum is displayed briefly, and then the Viewing Area is prepared
for data collection from your sample.
We recommended that you run a background scan before every sample.

3. Place your sample in the instrument or accessory and then click to begin scanning
your sample.
By default, during scanning the sample data is displayed on the Live tab in the Viewing
Area.
The completed spectrum is displayed on the Graph tab, and added to the current
Samples View in the Data Explorer. The results of the Quality Checks, selected by
default on the Setup Instrument Advanced tab, are displayed in the results for the
spectrum. If you have the Auto save option selected on the Setup Instrument Data
Collection tab, your spectrum will be saved automatically.
48 . Frontier IR Dual-Range Systems User's Guide

4. If, for any reason, you want to stop scanning your sample, click .

You can use the Sample Shuttle Accessory (L1200302) to set up an interleaved cycle scan so
that you do not have to open the sample compartment between the background and sample
scans. This helps to reduce changes in the concentration of carbon dioxide and water vapor
in the sample compartment. Set the Scan Type to Interleaved on the Setup Instrument
Basic tab. See the Spectrum on-screen Help for more information.

Changing the instrument range


You can change the instrument range from, for example MIR to NIR, using Spectrum
software. Changing ranges requires the instrument to change the detector, beamsplitter and,
for the Frontier FT-IR/FT-NIR only, source in concert.

NOTE: Depending on the beamsplitters and detectors fitted, the instrument may require
some time to equilibrate after changing ranges.

1. Select the Setup menu and then click Instrument.


The Setup Instrument tabs are displayed.

2. Select the Setup Instrument BeamPath tab.


The Mode field in the settings table, shown here as MIR, displays the current range.

NOTE: The Mode option is also available in the Instrument Settings toolbar.

3. Select the range you wish to change to from the drop-down list.
The options are MIR, NIR or Custom for the Frontier FT-IR/FT-NIR spectrometer, or
MIR, FIR or Custom for the Frontier FT-IR/FT-FIR spectrometer.
If you select MIR, NIR or FIR, the Source, Beamsplitter, Detector, Window and
Optimum Scan Range fields will be updated. If you select Custom, the Source,
Beamsplitter and Detector options can be selected independently. The Optimum Scan
Range and Window fields will update for the combination you have selected.

NOTE: Some combinations may damage the instrument, and will not be allowed.

4. Click the Apply button or click Scan to change the range.


If you have a Frontier FT-IR/FT-FIR spectrometer, you will be prompted to change the
windows. Change the windows (as described in Replacing the Frontier FT-IR/FT-FIR
Sample Compartment Windows on page 61) and then click Scan.
While the instrument is changing range, you will not be able to collect data. The range
change is monitored by messages in the Status Bar.
Using the Spectrometer with Spectrum . 49

Working with the instrument display and Go button

Display

Go button

Figure 16 Frontier IR top panel controls

The display on the front right of the top of your instrument is used to display prompts and
other messages generated by Spectrum software.

Simply follow the prompts in the instrument display and use the Go button located on the
instrument to complete a series of sample scans.

Changing the beampath


Hover over the icons on the Setup Instrument BeamPath tab to identify elements in the
instrument beampath. As you click on each item, the appropriate row in the Settings Table is
highlighted.

The accessory fitted in the sample compartment is identified by a graphic element in the
beampath schematic. If an external accessory is fitted at the beam port on the right of the
instrument, or a Spotlight Imaging System is fitted, the Setup Instrument Beampath tab
includes additional elements that enable you to redirect the beampath to the accessory you
want to use.

In Figure 17 the beampath is directed to the standard sample slide. In this example, the
additional interface elements for the External NIRA (including its filter wheel), and the
Spotlight Imager, are also shown.

Figure 17 Setup Instrument BeamPath tab with the beampath directed to the
Slide Holder
50 . Frontier IR Dual-Range Systems User's Guide

 To direct the beam to the External NIRA, select Right external from the Beam
Location drop-down list.

Figure 18 Setup Instrument BeamPath tab with the beampath directed to the
External NIRA (circled)

Using the Spectrum on-screen Help system


Use the Spectrum Help system to find further information about using Spectrum software to
control, set up and adjust your instrument.

To open the Help file, select Contents from the Help menu. This menu also includes links to
on-screen tutorials (Tutorials), and information about the software (About).
Using the Spectrometer with Spectrum . 51

Atmospheric (CO2/H2O) Suppression

Atmospheric suppression can be selected in the Setup Instrument Advanced tab (Figure 19).

Figure 19 Setup Instrument Advanced tab in Spectrum

What is atmospheric suppression?


This is an atmospheric correction routine. This routine is more powerful than simple
subtraction, overcoming the following issues:

• Non-linearity due to resolution

• The measured spectrum is temperature dependent

• Lineshape and calibration are affected by J-stop and sample or accessory.

What does atmospheric suppression do?


When CO2/H2O is switched on, the software uses a single reference spectrum derived from
high resolution data and our understanding of the instrument to model the lineshape, then
finds the current real instrument parameters by least squares fitting to the measured
spectrum.

−1
Figure 20 Correction of an MIR spectrum at 4 cm resolution
52 . Frontier IR Dual-Range Systems User's Guide

Figure 21 Atmospheric suppression


Using the Spectrometer with Spectrum . 53

AVI Correction

AVI correction can be selected in the Setup Instrument Advanced tab in Spectrum
(Figure 19).

NOTE: AVI correction is not available when in the FIR range.

AVI correction can only be performed if an AVI Calibration has been set up for the current
configuration and resolution. Select AVI from the Adjustments Toolbox, available from the
Setup Instrument Advanced Tab. Follow the instructions on-screen. For further information,
refer to the Spectrum on-screen Help file.

NOTE: AVI correction uses a methane gas cell installed in the filter wheel.

What is AVI correction?


The objectives of Absolute Virtual Instrument (AVI) correction are:

• Consistent performance over time and between instruments

• Traceability for all measurements.

Although FT-IR spectrometers use a reference laser, the wavenumber calibration and
lineshape are affected by differences in beam divergence and uniformity. This is true for all
FT-IR spectrometers. Differences can occur between instruments, when using different
sampling accessories and when components are changed. AVI allows calibration and
lineshape to be maintained.

The Absolute Virtual Instrument is an instrument with theoretical performance, such that the
result of measuring a known sample on such an instrument can be predicted. So, if we
measure with a real instrument and calculate the software transform to match the theoretical
result, we can apply this transform to future measurements.

The Absolute Virtual Instrument is defined by wavenumber calibration, instrument lineshape


and ordinate accuracy.

What does AVI correction do?


When AVI is switched on, the software measures the current instrument profile relative to an
absolute standard (the methane cell) and an ideal lineshape function, and applies a
correction (Figure 22). If you have an automated, internal filter wheel, this can provide
correction for any sampling configuration.
54 . Frontier IR Dual-Range Systems User's Guide

−1
Figure 22 Spectra of methane at 4 cm resolution as measured (top) and with
AVI (bottom)
Using the Spectrometer with Spectrum . 55

Look Ahead

Look Ahead can be selected in the Setup Instrument Advanced tab (Figure 19).

What is Look Ahead?


Look Ahead is a novel system where the spectrometer scans continuously and uses the
properties of the measured spectrum to identify changes corresponding to sample removal,
sample insertion, or sample change, automatically.

This information is used to identify the scans that are being collected, and to accumulate
sample scans. This can decrease the overall scanning time required for those samples that
require scan accumulation.

What does Look Ahead do?


If Look Ahead has been enabled, the spectrometer scans continuously. When you request a
sample scan, the scan collected is compared to the scans that have already been collected. If
identical scans are found, the number of scans requested is decreased by the number of
scans found, so decreasing the overall scanning time.
56 . Frontier IR Dual-Range Systems User's Guide

Quality Checks

Quality Checks can be selected in the Setup Instrument Advanced tab (Figure 19).

What are Quality Checks?


Quality checking is a tool for less experienced IR users that identifies possible problems in
the spectrum being collected and suggests ways of improving the measurement.

What do Quality Checks do?


Simply select the Quality Checks that you want to perform from the list in the Setup
Instrument Advanced tab and, if required, adjust the threshold range using the slider bars.
To display more information about an individual test, double-click on its name in the list.

When you collect your spectrum, the selected tests are performed as the data is collected
and a signal light (green , amber , or red ) indicates the result.

The results from the Quality Checks imply:

Okay ( ) – the quality of the spectra is satisfactory.

Caution ( ) – the software has identified a problem that you may want to investigate in
order to improve the quality of the spectra you are collecting.

Warning ( ) – the software has identified a serious problem that you should attempt to
cure before collecting further spectra.
Routine Maintenance
58 . Frontier IR Dual-Range Systems User's Guide

Cleaning the Spectrometer

Clean the outside of the instrument using a damp cloth. If necessary, a mild detergent may
be used. Before you clean the entire instrument, always perform a patch test on an
inconspicuous area.

Avoid spilling liquid into the instrument. Clean all external spills immediately. If anything that
is spilled enters the main body of the spectrometer, switch off the power and contact a
PerkinElmer Service Engineer.

Do not touch or attempt to clean any optical surface in the instrument,


because this will impair its performance and may easily damage the
CAUTION
component.

Cleaning the display


Clean grease and dirt off the display using a soft damp cloth and a mild detergent.
Routine Maintenance . 59

Moving the Spectrometer

The spectrometer is a heavy precision instrument, so two people are


required for safe handling.

Consult the local codes of practice issued by safety advisors before


attempting to lift it.
WARNING
Take care not to injure yourself or others, or to drop the spectrometer.

Before moving the spectrometer, switch off the power supply, wait
60 seconds, and disconnect the power cable.

WARNING

Do not attempt to lift the instrument, or to move it to another location,


when an external accessory is attached. The optical alignment of the
accessory may be disturbed.
CAUTION
The accessory is fitted to an external accessory bracket mounted within
the handhold on the right of the instrument (Figure 23). The two screws
that attach the external accessory to this bracket are accessible from the
right side of the accessory.

The spectrometer can be lifted using the shaped handholds on its sides, as shown in
Figure 23. Two people are needed to lift it because its basic weight is approximately 36 kg.

Lifting handhold,
(repeated on left of
the instrument)

Figure 23 Lifting a Frontier IR spectrometer


60 . Frontier IR Dual-Range Systems User's Guide

Condensation
Be aware that condensation caused by moving your spectrometer from a cooler environment
to a warmer one can damage the windows of the sample compartment. To prevent this
damage from occurring:

• Make sure that the windows are protected by placing fresh bags of desiccant in the
sample compartment.

• Leave the spectrometer to reach the temperature of the surroundings before removing
the desiccant.

The Set Temperature utility


A Frontier IR instrument is able to thermostat (that is, control the temperature of) its internal
optical bench. By default, this capability is enabled by the PerkinElmer Service Engineer when
the instrument is installed.

Usually, there is no need for adjustment. However, if the temperature range in your
laboratory changes, or if you move the instrument to another location, it may be necessary
to set a different temperature range to ensure optimal performance.

1. Browse to the C:\Program Files\PerkinElmer\ServiceIR or C:\Program Files


(x86)\PerkinElmer\ServiceIR directory, and then double-click the Set Temperature
shortcut.

The PerkinElmer Login dialog is displayed.

The ServiceIR folder contains a number of other utilities, many of which


CAUTION are intended solely for use by a PerkinElmer Service Engineer. If used
incorrectly, some utilities could damage the instrument.

2. Enter your login details.


You must be a Spectrum Administrator.
The Select Instrument dialog is displayed.

3. Select your instrument and then click OK.


The Set Temperature dialog is displayed.

4. Select the setting most appropriate for the range of temperature in your laboratory, and
then click Apply.
The Disabled setting is not recommended.
The temperature range setting is stored in the instrument, and the utility closes.

NOTE: This setting applies until the Set Temperature Utility is used again. It is not reset
when the instrument power is recycled (that is, switched OFF and then ON).
Routine Maintenance . 61

Replacing the Frontier FT-IR/FT-FIR Sample


Compartment Windows

NOTE: You can also change the sample compartment windows in the Frontier FT-IR/FT-NIR
spectrometer. For more information, see Installing/Replacing Beam Port and Sample
Compartment Windows on page 85.
The sample compartment windows in the Frontier FT-IR/FT-FIR spectrometer need to be
changed when moving between the MIR and FIR ranges.
KBr windows are used in the MIR range. In the FIR range, the PE window is installed in the
left side of the sample compartment and the PE/D window is installed in the right side of the
sample compartment.

1. Open the sample compartment cover.

2. Using the 3 mm screwdriver (part number L9003317) supplied with the Quick Change
Window Kit (part number L1250432), slacken the three screws securing the current
window to the side of the sample compartment.

3. Remove the current window with its seal in place.


Store the window in the desiccated spare window compartment (Figure 9).

Be careful not to get moisture on the optical surface of the window. Do


CAUTION not touch or breathe on the optical surface of the window.

4. Ensure the seal is fitted to the replacement window and is correctly seated (Figure 24).

Window

Seal

Figure 24 Seal fitted to window


62 . Frontier IR Dual-Range Systems User's Guide

5. Fit the window to the sample compartment.


Ensure the seal is fully seated and the key on the window lines up with the notch in the
main cover (Figure 25).

Sample compartment wall


Sample
compartment
wall

Screw Notch and


Screw Notch and key
key

Figure 25 Window installed in sample compartment

6. Secure the window to the sample compartment wall by tightening the screws using the
3 mm screwdriver (Figure 25).

NOTE: Do not over tighten the screws.

7. Repeat steps 2–6 for the other sample compartment window.

8. Close the sample compartment cover.

To store the windows:

1. Open the spare window compartment cover (Figure 26).

Compartment Spare window


desiccant indicator

Screw Screw

Figure 26 Windows situated in the spare window compartment


Routine Maintenance . 63

2. Place each window in the holder, with the seal in place, and tighten the screw using the
screwdriver provided.

NOTE: Do not over tighten the screw.

3. Close the spare window compartment cover.


If you have an MCT detector for the MIR range, you will be supplied with an Attenuator kit
(part number L1160560) containing 1%T, 4%T, 6%T, 14%T and 32%T attenuators. The
attenuators attach to the KBr window magnetically.

Before you change to the MIR MCT detector in Spectrum software, fit one of the attenuators
supplied to the window to the right of the sample compartment to prevent the detector from
overloading (Figure 27). If you have an open beam, a 14%T attenuator is recommended.

Figure 27 KBr window with 14%T attenuator

NOTE: If you have a Frontier FT-IR/FT-FIR spectrometer with an MCT detector and the
instrument was in the MIR range when it was switched off, it is recommended that a
6%T attenuator is fitted to the sample compartment window before switching on the
instrument again, to prevent errors during the initialization of the FIR detector.
64 . Frontier IR Dual-Range Systems User's Guide

The Desiccant Indicator in Detail


The optical system of the spectrometer is purged at the factory. This protects the sample
compartment windows, external port windows, or beamsplitters from being damaged by
humidity. Replaceable packs of desiccant maintain the purge.

The top panel of the instrument includes a desiccant indicator (Figure 28), whose sectors
change sequentially from blue to pink as the desiccant becomes exhausted. Change the
desiccant packs in the instrument when the sector marked 10 becomes pink, but while the
sectors marked 15 and 20 are still blue. These numbers correspond to the approximate
% Relative Humidity in the instrument.

NOTE: A similar desiccant indicator is fitted in the desiccated compartment (Figure 34 on


page 68). This is used for storing sample compartment windows when not in use.

Pink Blue

Blue

Figure 28 The desiccant indicator: when desiccant packs should be changed

If all three sectors of the desiccant indicator are pink, then you m ust
change the desiccant.
CAUTION
The instrument optics may be fogged. Do not switch the instrument
either ON or OFF until all sectors are blue.

Sector Action Required

All sectors blue None.

The humidity levels in the instrument are below approximately 10%


Relative Humidity.
Desiccant change is recommended.
Sector 10 pink
If the instrument has been switched OFF for an extended period, do not
switch ON until you have changed the desiccant and all sectors are blue.

Sector 10 & 15 Replace desiccant the immediately to avoid instrument damage.


pink
Do not switch the instrument ON or OFF until you have changed the
desiccant and all sectors are blue.
Routine Maintenance . 65

Changing the Desiccant

Expect to change the desiccant in the spectrometer every six months.


Old, used desiccant releases moisture.
CAUTION
In regions experiencing high humidity levels we recommend that you
change the desiccant more often.

The desiccant change interval is set in the Desiccant change due in (days) option on the
Setup Instrument BeamPath tab in Spectrum software (Figure 29).

Figure 29 Setup Instrument BeamPath tab in Spectrum

If an instrument desiccant change is overdue, a warning will be displayed in the Status Bar
and in the Setup Instrument BeamPath table.

The warning is displayed until you tell the software that the desiccant has been changed. To
reset the desiccant change interval, select the current value, click Changed to clear any
warning messages, and then enter the number of days before the next desiccant change is
due.
66 . Frontier IR Dual-Range Systems User's Guide

Renewing the instrument desiccant

Old, used desiccant releases moisture and can cause catastrophic failure
CAUTION of KBr optics.

Do not use damaged packs of desiccant. Make sure that the packs you
use have not been left in contact with the air.

You can purchase disposable desiccant kits from PerkinElmer (part number N0171159). A kit
contains two packs of desiccant, and three kits are required.

1. Inspect the humidity indicator card in the plastic bags in which the spare desiccant
packs are packed. If the card indicates humidity in the bag, discard the desiccant pack.

2. Remove the sample area cover, if fitted, by opening the cover, pressing the clip and
pulling the cover vertically to remove it (Figure 30).

Sample cover
release clip

Figure 30 Removing the sample compartment cover

3. Undo the two captive screws securing the desiccant cover (Figure 31).

Screw Screw

Figure 31 Captive screws securing the desiccant cover


Routine Maintenance . 67

4. Open the cover and remove all the exhausted desiccant packs (Figure 32), noting how
they are installed.

Desiccant
compartment
(empty)

Figure 32 Desiccant removed

5. Place the first three packs of desiccant upright in the recess on the right of the
desiccant holder then, one at a time, lay the three remaining packs flat in the upper
part of the holder (Figure 33).

Figure 33 Stowing the desiccant packs

Ensure that when the packs have been installed they do not rise above the level of the
black rubber purge seal.

6. Close the cover and tighten the screws.

7. Refit the sample area cover.

Installing rechargeable desiccant in the instrument


There is also a rechargeable desiccant kit available (part number L1250311).

To install rechargeable desiccant packs follow the procedure in Renewing the instrument
desiccant, starting on page 65.

The instrument requires two rechargeable desiccant packs; the packs are sized to fit into the
left (70 mm × 78 mm × 33 mm) or right (102 mm × 78 mm × 84 mm) of the desiccant
container.

The standard kit contains two pairs of rechargeable desiccant packs. This is so you can
immediately replace the desiccant in the instrument with re-activated packs.

The desiccant packs can be re-activated by baking them in an oven at 250 °C for
approximately 8 hours. They should be cooled in a dry atmosphere. For optimum results,
re-activate the packs immediately before re-use.
68 . Frontier IR Dual-Range Systems User's Guide

Renewing the desiccant in the spare window


compartment
The windows in the sample compartment and the external ports can be exchanged for
windows made from another material.

To avoid serious damage, spare windows must be kept in a dry atmosphere,


such as in the desiccated compartment on top of the instrument.
CAUTION
If more than one sector on the desiccant indicator inside the compartment is
pink, change the desiccant.

Installing disposable desiccant in the window compartment

The desiccated compartment contains two disposable desiccant packs. You can purchase
disposable desiccant kits from PerkinElmer (part number N0171159). A kit contains two
packs of desiccant, and one kit is required.

1. Inspect the humidity indicator card in the plastic bags in which the spare desiccant
packs are packed. If the card is pink (indicating humidity in the bag) discard the
desiccant packs.

2. Open the desiccated compartment, and then remove the two thumb screws securing
the desiccant pack cover (Figure 34).

Thumb screw Thumb screw

Figure 34 Desiccated compartment for spare windows

3. Lift off the desiccant pack cover and remove the desiccant packs noting how they were
installed.

4. Install two new disposable desiccant packs, or the rechargeable desiccant pack, and
then refit the desiccant pack cover.

5. Close the compartment.


The desiccant indicator should be completely blue, indicating satisfactory storage
conditions, within approximately half an hour.
Routine Maintenance . 69

Installing rechargeable desiccant in the window compartment

There is also a rechargeable desiccant kit available (part number L1251840).

To install rechargeable desiccant packs refer to the procedure for Installing disposable
desiccant in the window compartment on page 68.

The desiccated compartment requires a single rechargeable desiccant pack that is sized to fit
the desiccant container.

The rechargeable desiccant kit contains two rechargeable desiccant packs. This is so you can
immediately replace the desiccant in the instrument with a re-activated pack.

Re-activate a desiccant pack by baking it in an oven at 250 °C for approximately eight hours.
It should be cooled in a dry atmosphere. For optimum results, re-activate the pack
immediately before re-use.
70 . Frontier IR Dual-Range Systems User's Guide

Purging the Spectrometer

Under most circumstances, you do not need to purge the optical system. However, after
performing any maintenance tasks that involved opening the main cover, you may purge the
optical system to remove water vapor and CO2 that entered while the system was open.

Purge the sample compartment if you need to make sure that no residual water vapor or CO2
peaks are visible in the spectrum, particularly when operating in the FIR range.

There are two separate sets of connectors for purge gas lines, one for the sample
compartment and one for the optical system. You can purge with either dry air or nitrogen.
Both remove water vapor; however, nitrogen is preferable because it also removes
atmospheric carbon dioxide.

A typical cylinder of dry nitrogen (or dry air) stores 6.26 m3 at 1.4 × 104 kPa (220 ft3 at
2200 lbf/in2). Make sure that the gas is free of oil, water, or dirt particles larger than 25 µm
(0.001 inch).

Do not site the instrument in a poorly ventilated area if nitrogen will be


used as a purge gas.

Oxygen depletion in an enclosed space does not trigger a gasping reflex,


WARNING
and errors of judgment, confusion, or unconsciousness can occur in
seconds and without warning.

Do not use a flammable gas to purge the instrument. The spectrometer


contains a hot source, and a fire or explosion will result. Only use clean,
dry, oil-free nitrogen or air to purge the instrument.
WARNING

Never connect the purge tubing directly to a gas cylinder or other high
pressure supply; always use a pressure regulator and set the pressure to
a maximum of 6.9 kPa (1 lbf/in2) before you start the flow.
WARNING

A length of clear plastic tubing (4 mm internal diameter) will be required to transfer the gas
from the gas bottle to the instrument purge connectors.
Routine Maintenance . 71

1. The tubing is connected to the instrument by a universal pipe fitting which is secured to
the tubing with a nut. Two sets of fittings are supplied with the instrument. Assemble
the fittings to the tubing as illustrated in Figure 35.

Knurled nut

Beveled end
of connector Tubing

Figure 35 Purge fittings

2. Tighten the nut, ensuring it is not over tightened.

3. Push the connector on the end of the tubing into the purge connector of the
compartment to be purged (Figure 36).
If the connector will not push in, press the clip at the bottom of the port and try again.

Sample purge
Instrument purge

Figure 36 Connector fitted to purge port

4. Connect the free end of the tubing to the gas regulator.

NOTE: Ensure the regulator is set to a pressure not greater than 6.9 kPa (1 lbf/in2).

5. Open the gas flow valve on the regulator until a flow rate of 28 l/minute (1 ft3/minute)
is achieved.
The instrument compartment is purged in 10 minutes. The sample compartment is
purged in 2 minutes.

6. Once purging is complete turn off the gas supply.


72 . Frontier IR Dual-Range Systems User's Guide

7. Disconnect the fitting from the instrument by pushing the clip located at the bottom of
the purging port (Figure 37).

Locking clip

Figure 37 Disconnect the locking clip


Routine Maintenance . 73

Changing the External Fuse

Very occasionally, a fuse may fail with age and will need to be changed.

NOTE: If fuses fail repeatedly, there is an electrical fault: disconnect the power supply and
contact your PerkinElmer Service Engineer.

There is one spare fuse in the fuse drawer. If you need more, order 2 A, 250 V Time Lag
fuses (part number 04970839) from PerkinElmer. You must only replace the fuse with one of
this type and rating. Do not use makeshift fuses and do not short-circuit fuse holders.

Before you begin this procedure, switch off the power to the
spectrometer, wait for 60 seconds, and disconnect the power cable.

This will protect you from electrical shock.


WARNING

The external fuse drawer is located at the rear of the instrument between the power switch
and the power socket (Figure 38).

Fuse drawer

Figure 38 Fuse drawer closed


74 . Frontier IR Dual-Range Systems User's Guide

1. Lever out the fuse drawer so that it swings down over the power socket.
The fuse is in the right hand slot.

2. Remove the old fuse and discard it.

3. Fit the new fuse into the slot (Figure 39).

Fuse
Spare fuse

Fuse drawer

Figure 39 Fuse fitted; drawer open

4. Close the drawer.


The new fuse is now installed.

5. Reconnect the power cable and switch on the instrument.


Routine Maintenance . 75

Cooling the MCT Detector (If Fitted)


The instrument may be fitted with an MCT (mercury cadmium telluride) detector. When it is
fitted in the instrument it must be cooled to 77 K before you collect spectra. The detector is
mounted in a dewar that can be filled with liquid nitrogen. The detector is not enabled until it
has been sufficiently cooled.

Do not site the instrument in a poorly ventilated area if liquid nitrogen


will be used to cool a detector.

Oxygen depletion in an enclosed space does not trigger a gasping reflex,


WARNING
and errors of judgment, confusion, or unconsciousness can occur in
seconds and without warning.

The extremely low temperature of liquid nitrogen can burn skin and
eyes. Avoid exposure by wearing protective gloves and safety goggles
whenever you work with it.
WARNING

As liquid nitrogen cools the detector it evaporates so rapidly that the


resulting pressures can send a funnel or detector cap suddenly and
forcefully shooting upward from the top of the dewar.
WARNING

Be sure to wait the specified time when filling the funnel and before
replacing the plug. This enables the bubbling nitrogen to settle down
and the pressure to dissipate. In addition to wearing safety goggles at all
times, stand back from the instrument after each time you fill the funnel.
WARNING
76 . Frontier IR Dual-Range Systems User's Guide

1. Remove the MCT coolant cover (Figure 40).

MCT coolant cover

Figure 40 Location of MCT coolant cover

2. Remove the plug from the dewar.

3. Ensure that the funnel guide, which helps prevent overspill entering the instrument
body rather than the dewar, is in place.

4. Make sure the supplied plastic funnel is dry.

Stand where you can see the inside of the funnel as you pour the
CAUTION nitrogen in. Pour slowly, so that neither the funnel nor the dewar
overflows. Take care not to overfill the funnel and splash liquid nitrogen
onto the instrument covers.

5. Insert the plastic funnel into the funnel guide and slowly fill the detector dewar with
liquid nitrogen.

NOTE: Make sure that the detector dewar is filled to the top with liquid nitrogen before
selecting the detector. Prevent radiation from reaching the detector by obstructing
the beam path in the sample area during filling.

6. Stand back and wait two minutes.


This nitrogen also vaporizes as the dewar continues to cool. The two-minute wait allows
the bubbling to settle down and the pressure of the vaporizing nitrogen to dissipate.

7. Continue to pour liquid nitrogen into the funnel, adding a little more each time the
funnel empties.
The funnel takes longer to empty as the dewar fills. This happens after two or three
more funnels of liquid nitrogen. Because the dewar has now cooled, the liquid nitrogen
does not vaporize, but instead fills the dewar.

8. Remove the funnel and wait two minutes.


The liquid nitrogen settles down and bubbling slows.
Routine Maintenance . 77

9. When the nitrogen stops bubbling, refit the plug.


When the detector dewar has been filled, it will remain at its operating temperature for
approximately eight hours.

NOTE: Cooled detector dewars require pumping down after approximately 12 months
operation. When the boil-off rate of liquid nitrogen becomes excessive (that is, the
liquid nitrogen level is low after 3–4 hours of operation), consult your PerkinElmer
Service Engineer.
78 . Frontier IR Dual-Range Systems User's Guide
Advanced Maintenance
80 . Frontier IR Dual-Range Systems User's Guide

Opening the Main Cover

To perform most maintenance tasks, you have to open the main cover of the spectrometer.

Switch off the mains power supply to the spectrometer, wait 60 seconds
and disconnect the power cable before you open the cover of the
spectrometer. This makes sure that you are safe from electrical shock
and laser radiation.
WARNING

Do not open the main cover if the humidity is >80% relative humidity for
an instrument fitted with KBr optics.
CAUTION
The humidity sensitive optics will be damaged if subjected to a relative
humidity in excess of the specified level.

The main cover is hinged at the back of the instrument and opens from the front. Figure 41
identifies the locking bolts, which are on the front of the instrument.

Locking bolt Locking bolt

Figure 41 Main cover locking bolts

1. Switch off the power supply to the spectrometer.

2. Wait for 60 seconds.

3. Disconnect the power cable.


Advanced Maintenance . 81

4. Remove the sample compartment cover, if fitted; open the cover, press on the cover
retaining clip and then pull the cover vertically to remove it (Figure 42).

Sample cover
release clip

Figure 42 Removing the sample compartment cover

5. Remove the fitted sampling accessory, by pulling the release handle towards you and
sliding the accessory towards you (Figure 43).

Sampling accessory
release handle

Figure 43 Removing the sampling accessory

6. Undo the two locking bolts on the front of the instrument using the hexagonal wrench
supplied; insert the wrench, so that the free end is pointing horizontally away from the
instrument and then rotate the wrench 180° in the direction indicated by the arrows in
Figure 44.

Figure 44 Undoing the locking bolts

7. Lift the main cover from the front.


82 . Frontier IR Dual-Range Systems User's Guide

8. Ensure the cover is fully open and retained by the stay.

9. Perform the necessary maintenance inside the instrument.


Figure 45 illustrates the primary component parts of a Frontier IR Dual-Range
spectrometer.

MIR source Laser power supply MIR source Laser power supply

NIR source NIR source


Power supply
Power supply Laser
Laser

NIR
detector

NIR
detector
Beamsplitters; within MIR Beamsplitters; within MIR
interferometer and detector interferometer and detector
carrier carrier

Figure 45 Frontier IR FT-IR/NIR components

NOTE: The Frontier FT-IR/FT-FIR spectrometer has one source only, in the MIR source
position. The FIR detector is in the NIR detector position.

NOTE: If you have an MCT detector fitted for the MIR range, the MCT detector is fitted
toward the rear of the instrument and the NIR/FIR detector is fitted at the front of
the instrument.

10. Close the cover.

11. Lock the two locking bolts by inserting the wrench into each bolt in turn and rotating it
opposite to the direction indicated by the arrows in Figure 44.

12. If required, reinstall the sampling accessory by sliding it on the ledge and pushing it into
the sample compartment until it engages with the connector.

13. If required, refit the sample cover by inserting the hinge into the grooves and pushing
the cover downwards until the clip engages.

14. Reconnect the power cable and switch the power back on.

15. If required change the desiccant (see Changing the Desiccant on page 65), and/or
purge the instrument (see Purging the Spectrometer on page 70).
If the cover is left open for more than 1 hour, the desiccant should be replaced.
Advanced Maintenance . 83

Replacing a Source

Switch off the mains power supply to the spectrometer, wait 60 seconds
and disconnect the power cable before you open the cover of the
spectrometer. This makes sure that you are safe from electrical shock
and laser radiation.
WARNING

Due to its high operating temperature allow sufficient time for the source
to cool before attempting to remove it.

WARNING

1. Open the main cover of the instrument as detailed on page 80 and identify the source
(Figure 46).

NIR source NIR source

MIR source MIR source

Figure 46 Source location for Frontier FT-IR/FT-NIR

NOTE: The Frontier FT-IR/FT-FIR spectrometer has one source only. It is located in the MIR
source position.

Caution, hot surface.

The source housing and source itself will be hot if recently switched on.
Allow to cool before performing the following procedure.
84 . Frontier IR Dual-Range Systems User's Guide

The source element is brittle and can be broken if not handled properly.
Be careful when removing and installing the source.
CAUTION
Do not touch the glass bulb as this will degrade the performance of the
bulb and shorten its life.

2. Slacken the two terminal screws (Figure 47 or Figure 48) securing the wires leading to
the source. Disconnect the wires from the connectors.
For the Frontier FT-IR/FT-NIR spectrometer, the connecting wires for the MIR source
are white, and for the NIR source are red.
For the Frontier FT-IR/FT-FIR spectrometer, the connecting wires for the source are white.

3. By grasping the round metal source body, NOT the wires, remove the source from its
housing by pulling, noting its installed position.

4. Install the new source in its housing, by pushing in until it clicks into place.

5. Reconnect the two wires to the terminals and secure by tightening the two terminal
screws (Figure 47 or Figure 48).

NIR source MIR source

MIR source terminal

MIR source terminal

NIR source terminal

NIR source terminal

Figure 47 Source terminal screws (from instrument left) for Frontier FT-IR/FT-NIR

Source

Source terminal

Source terminal

Figure 48 Source terminal screws (from instrument left) for Frontier FT-IR/FT-FIR

6. Close and secure the main cover.


Refer to steps 11 to 15 on page 82.

7. Allow the source temperature to equilibrate for approximately 2 hours before using the
instrument.
Advanced Maintenance . 85

Installing/Replacing Beam Port and Sample


Compartment Windows
The following procedure describes how to install a window in one of the external beam ports.
It also applies to installing windows in the sample compartment of the Frontier FT-IR/FT-NIR
spectrometer.

NOTE: For information about changing the Frontier FT-IR/FT-FIR sample compartment
windows, see Replacing the Frontier FT-IR/FT-FIR Sample Compartment Windows on
page 61.

Switch off the mains power supply to the spectrometer, wait 60 seconds,
and disconnect the power cable before you open the cover of the
spectrometer. This makes sure that you are safe from electrical shock
WARNING
and laser radiation.

1. Open the main cover of the instrument, as detailed on page 80.


– Using the hexagonal wrench supplied with the new window kit, slacken and
remove the screws and washers securing the beam port cover or current window
to the main cover.
Remove the screws carefully, and ensure that the washers do not fall into the
instrument.
– Remove the beam port cover or current window and seal. Retain for further use.

Be careful not to get moisture on the optical surface of the window. Do


CAUTION not touch or breathe on the optical surface of the window.

2. Ensure the seal is fitted to the window and is correctly seated (Figure 49).

Window

Seal

Figure 49 Seal fitted to window


86 . Frontier IR Dual-Range Systems User's Guide

3. Fit the window to the main cover from the outside in.
Ensure the seal is fully seated and the key on the window lines up with the notch in the
main cover.

Screw and washers Screw and washers


Main cover

Main cover

Notch and key


Screw and washers Screw and washers

Notch and key

Figure 50 Window installed in main cover

4. Secure the window to the main cover by fitting the screws and washers and tightening
the screws using the hexagonal wrench (Figure 50).

5. Close and secure the main cover.


Refer to steps 11 to 15 on page 82.

If you have an MCT detector for the MIR range, you will be supplied with an Attenuator kit
(part number L1160560) containing 1%T, 4%T, 6%T, 14%T and 32%T attenuators. The
attenuators attach to the window magnetically. You will require the ferritic window surround
(part number C1201453) to attach the attenuators to the new windows.

Before you change to the MIR MCT detector in Spectrum software, fit one of the attenuators
supplied to the window to the right of the sample compartment to prevent the detector from
overloading. If you have an open beam, we recommend a 14%T attenuator.
Advanced Maintenance . 87

Installing Filters in the Filter Wheel


The following procedure describes how to install a filter in the filter wheel. The same
procedure should be followed when replacing a filter or the methane cell.
The filter wheel assembly is removed from its mounting and placed on the standard sample
slide baseplate. It therefore may be necessary to remove the sample slide.

Switch off the mains power supply to the spectrometer, wait 60 seconds,
and disconnect the power cable before you open the cover of the
spectrometer. This makes sure that you are safe from electrical shock
and laser radiation.
WARNING

1. Set the Filter Wheel to Position 1 on the Setup Instrument BeamPath tab in Spectrum
software.
Position 1 will be labelled as None 15000 – 0 cm-1 and be the first option in the
drop-down list.

2. Open the main cover of the instrument as detailed on page 80 and identify the filter
wheel (Figure 51).

Filter wheel Filter wheel

Figure 51 Filter wheel location


88 . Frontier IR Dual-Range Systems User's Guide

3. Install the standard sample holder baseplate, but remove the sample holder itself by
unscrewing the knurled screw (Figure 52).

Knurled screw

Figure 52 Knurled screw

4. To remove the filter wheel, loosen the set screw that secures it to its shaft using a
1.5 mm Allen key (Figure 53) and then gently ease the filter wheel off the shaft,
rotating the filter wheel as required to clear the filter wheel from the other instrument
components.
Take care not to remove the spacer ring from the spindle, as this may be lost.

NOTE: If the set screw is not accessible, you may not have moved the filter wheel to
Position 1 as described in Step 1.

Screw

Figure 53 Filter wheel securing screw

Avoid touching the filters.


CAUTION
Some of the filters are quite difficult to see.
Advanced Maintenance . 89

5. Place the assembly face down on the sample holder baseplate.

6. Working clockwise around the filter wheel, identify the next available filter position
(Figure 54).
Position 5 and Position 7 are available for your custom filters.
There is no filter in Position 6 in the Frontier FT-IR/FT-FIR spectrometer.

NOTE: Do not change the positions of the standard filters, or place a filter in Position 1.

Position 5:
Position 4:
for custom filter
Methane cell

Position 3: Position 6:
NG11 glass NIR polystyrene

Position 2: Position 7:
MIR polystyrene for custom filter

Opto flag
Position 1: None
Do not place filter here

Figure 54 Filter wheel positions

7. Release the filter holder by turning it about 30 degrees counter-clockwise, and then
remove the filter holder and foam spacer.

8. Place the filter in the hole, place the foam spacer on top of it, and then fit the filter
holder.
Figure 55 illustrates the assembly sequence.
Align the notches of the holder with the indents in the filter wheel. You must twist the
holder clockwise to lock it to the filter wheel.

Filter

Foam spacer

Filter holder

Figure 55 Filter, foam spacer and filter holder

When fitted correctly, the small hole in the filter holder is close to the outer edge of the
filter wheel.
90 . Frontier IR Dual-Range Systems User's Guide

9. Refit the filter wheel to its spindle and secure it with the set screw.

Make sure that you tighten the set screw onto the flat of the shaft.
CAUTION
If the set screw is tightened onto a round portion of the shaft the filter
wheel may loosen.

10. Refit the sample holder onto the baseplate with the knurled screw.

11. Close and secure the main cover.


Refer to steps 11 to 15 on page 82.

12. Select the filter you would like to use in the beam path from the Filter Wheel
drop-down list on the Setup Instrument BeamPath tab in Spectrum software.

NOTE: If you wish to run a background scan with a filter in place, you must set the Scan
Type to Background on the Setup Instrument Basic tab, rather than using the

Background toolbar button . See the Spectrum on-screen Help for more
information.
Advanced Maintenance . 91

Replacing the Laser and Power Supply

Switch off the main power supply to the spectrometer, wait 60 seconds
and disconnect the power cable before you open the cover of the
spectrometer. This makes sure that you are safe from electrical shock
and laser radiation.
WARNING

Do not attempt to separate the Laser from the power supply. The Laser
retains an electrical charge even when the main power is switched off.

WARNING

1. Open the main cover of the instrument as detailed on page 80 and identify the laser
and the laser power supply.

Laser Power supply Laser Power supply

Figure 56 Laser and power supply

2. Snip, and remove, the cable tie surrounding the power supply, noting how it secures the
laser cables (Figure 57).
The laser power supply fixings are accessible from the left of the instrument.
92 . Frontier IR Dual-Range Systems User's Guide

Screw

Cable tie

Screw

Figure 57 Laser power supply fixings (from left of instrument)


3. Slacken the two screws securing the laser power supply.

4. Unscrew and remove the cable clip that secures the thin black and red PSU laser cable
at the rear of the instrument, underneath the laser assembly (Figure 58).

Instrument cover rear hinge

Connector
Cable clip

Figure 58 Laser board connector and cable clip (from top of instrument)

5. Disconnect the black and red PSU laser cable from the board.

Be careful, the nearby source housing may be hot.

WARNING
Advanced Maintenance . 93

6. Unhook the spring at the rear of the laser, and remove the two screws on the retaining
bracket at the front of the laser (Figure 59).

Laser spring
retaining pin
Laser
Laser power supply

Laser bracket
retaining screw Baffle retaining
screw

Laser bracket
Laser end stop retaining screw

Figure 59 Laser and power supply installation

7. Remove the laser and power supply; noting the way the cabling is routed.

8. Install the new laser and power supply.


Ensure the laser is pushed fully home to the end stop at the front of the housing.
The laser must be installed so that the yellow laser warning label is facing directly
upwards.

9. Refit the laser retaining bracket and secure with the two securing screws.

10. Secure the spring on the retaining pin.

11. Reconnect the PSU laser cable to the board connector.

12. Secure the laser power supply with the two securing screws (Figure 57).

13. Refit the cable clip (Figure 58), so that it does not get trapped in the cover.

14. Secure the laser cabling to the laser power supply with a new cable tie.

15. Close and secure the main cover.


Refer to steps 11 to 15 on page 82.
94 . Frontier IR Dual-Range Systems User's Guide
Appendices
96 . Frontier IR Dual-Range Systems User's Guide

Appendix 1: Changing the Sample Area


Accessory

1. Remove the sample area cover if fitted by opening the cover, pressing the clip and
pulling the cover vertically to remove (Figure 60).

Sample
cover
release clip

Figure 60 Removing the sample area cover

2. Remove the current sampling accessory by pulling the release handle (under the
baseplate of the accessory) towards you and then sliding the accessory towards you
(Figure 61).

Release
handle

Figure 61 Removing a sampling accessory


Appendices . 97

3. Install the new accessory (Figure 62) by sliding it onto the ledge and pushing it into the
sample compartment until it engages with the connector.

Figure 62 Installing a sampling accessory

Spectrum automatically recognizes which accessory is installed, and displays the


accessory information on the Setup Instrument tabs.

Figure 63 Setup Instrument BeamPath tab with Universal ATR icon (circled)
98 . Frontier IR Dual-Range Systems User's Guide

Appendix 2: Instrument Self-Checks

NOTE: It can take the instrument about two hours to equilibrate when switched on after
being switched off overnight. To save time, we suggest that you leave the
spectrometer switched on at all times.

• When you switch on the instrument, the instrument display provides this sequence of
messages:
>Initializing
Initializing
Checking Program Integrity
Checking Instrument Configuration
Checking Source
Checking Laser
Checking Scan
Checking Calibration
Initialization Complete.

• When initialization is complete the last message fades, leaving the instrument identity
message:
PerkinElmer Frontier FT-IR

When connected to a PC, the instrument display adds Ready.

• When collecting data, the instrument display provides prompts generated by the
PerkinElmer software.
Appendices . 99

Initialization errors

If your instrument generates one of the following error messages, switch off the instrument,
then switch the instrument on again. Consult your PerkinElmer service representative if the
problem persists.

Error message Action

Program Integrity Failed The instrument’s firmware may be corrupt, so must be


reinstalled.

Instrument Configuration Failed Suggests that one or more internal devices is failing. If
the problem persists, note any associated error
messages and consult your PerkinElmer service
representative.

Scan Failed If the problem persists, note any associated error


messages and consult your PerkinElmer service
representative.

Laser Failed Consult your PerkinElmer service representative.


Replace Laser
The beamsplitters may be missing, the laser fringe pre-
amplifier may have failed, or the laser itself may have
failed.

Refer to Replacing the Laser and Power Supply on


page 90.

Source Failed Refer to Replacing a Source on page 83.


Replace Source

Calibration Failed Too little energy may be reaching the detector. Make
sure the sample compartment is clear.
100 . Frontier IR Dual-Range Systems User's Guide

Appendix 3: Instrument Performance


Validation

The Frontier FT-IR/FT-NIR spectrometer is supplied with NIR Performance Validation Kit
(part number L1250405), which contains a polystyrene reference sample.

To validate the MIR performance of your instrument, use the optional MIR Performance
Validation Kit (part number L1250404). This kit contains an MIR polystyrene reference
sample and an NG11 Schott Glass reference sample.

Macros are supplied with Spectrum software (version 10 or later) that can be used with the
Performance Validation kits to produce a report certifying your instrument.

NOTE: Instrument Performance Validation is not applicable for the FIR range.
Appendices . 101

Appendix 4: Decontamination and Cleaning

Before using any cleaning or decontamination methods except those specified by


PerkinElmer, users should check with PerkinElmer that the proposed method will not damage
the equipment.

Decontamination

Customers wishing to return instrumentation and/or associated materials to


PerkinElmer for repair, maintenance, warranty or trade-in purposes are advised
that all returned goods must be certified as clean and free from contamination.

The customer’s responsible body is required to follow the “Equipment Decontamination


Procedure” and complete the “Certificate of Decontamination”. These documents are
available on the PerkinElmer public website:
http://las.perkinelmer.com/OneSource/decontamination.htm

Alternatively, if you do not have access to the internet contact Customer Care:

Customer Care USA: 1-800-762-4000 (inside the USA)


(8:30 a.m. – 7 p.m. EST) (+1) 203-925-4602 (outside the USA)
Customer Care Canada: 800-561-4646
Customer Care EU: 0800 40 858 (Brussels)
0800 90 66 42 (Monza)

If you are located outside of these regions, please call your local PerkinElmer sales office for
more information.

Cleaning the instrument

Exterior surfaces may be cleaned with a soft cloth, dampened with a mild detergent and
water solution. Do not use abrasive cleaners or solvents.
102 . Frontier IR Dual-Range Systems User's Guide

Appendix 5: WEEE Instructions for PerkinElmer


Products

or

A label with a crossed-out wheeled bin symbol and a rectangular bar indicates that the
product is covered by the Waste Electrical and Electronic Equipment (WEEE) Directive and is
not to be disposed of as unsorted municipal waste. Any products marked with this symbol
must be collected separately, and in accordance with the regulatory guidelines in your area.

The objectives of this program are to preserve, protect and improve the quality of the
environment, protect human health, and utilize natural resources prudently and rationally.
Specific treatment of WEEE is indispensable in order to avoid the dispersion of pollutants into
the recycled material or waste stream. Such treatment is the most effective means of
protecting the customer’s environment.

The requirements for waste collection, reuse, recycling, and recovery programs are set by
the regulatory authority at your location. Contact your local responsible person (such as your
laboratory manager) or authorized representative for information regarding applicable
disposal regulations.

See the PerkinElmer web address below for information specific to PerkinElmer products, and
contact details for the Customer care department in your region.

http://las.perkinelmer.com/OneSource/Environmental-directives.htm

Products from other manufacturers may also form a part of your PerkinElmer system. These
other manufacturers are directly responsible for the collection and processing of their own
waste products under the terms of the WEEE Directive. Please contact these manufacturers
directly before discarding any of their products.

Consult the PerkinElmer web site (above) for manufacturer’s names and web addresses.
Index . 103

Index

A E
About this manual .......................... 6 Electrical supply inlet .................... 43
Accessories .................................. 31 EMC
External ................................... 28 Compliance ............................... 22
Internal .................................... 28 EC Directive .............................. 22
Accessory FCC rules .................................. 22
Changing ............................ 28, 96 External accessories ...................... 28
Release handle.......................... 96
Attenuators ................. 25, 38, 63, 86 F
FCC rules and regulations .............. 22
C
Filter wheel
Capacitors.................................... 16 Filter ........................................ 89
Changing configuration ................. 25 Filter holder .............................. 89
Cleaning ...................................... 58 Foam spacer ............................. 89
Display ..................................... 58 Installing filters ......................... 87
External ................................... 58 Location ................................... 87
Cleaning the instrument .............. 101 Retaining screws ....................... 88
Communications ports ............. 30, 41 Flammable solvents and samples.... 15
Condensation ............................... 60 Fuse
Configuration ............................... 25 Changing .................................. 73
Connecting the PC ........................ 40 Drawer ..................................... 73
Connecting to electrical supply ....... 42
Conventions G
Notes, Cautions and Warnings ...... 7
Text ........................................... 7 Go button .................................... 26
Cooling ........................................ 14 Guided tour ................................. 24

D H

Desiccant HTML help ..................................... 6


Changing ............................ 65, 82 Humidity...................................... 13
Disposable .......................... 66, 67
Indicator ............................ 39, 64 I
Rechargeable ............................ 67 Installing software ........................ 44
Removed .................................. 67 Installing the instrument ............... 44
Stowed..................................... 67 Instrument
Window compartment packs ....... 68 Changing range ................... 47, 48
Desiccant cover Changing the beampath ............. 49
Captive screws .......................... 66 Instrument controls
Desiccant indicator ................. 39, 64 Description ............................... 49
Display Display ..................................... 26
Cleaning ................................... 58 Go button ................................. 26
Display ........................................ 26 Instrument display and Go button .. 49
Instrument validation .................. 100
Internal accessories ...................... 28
IPV ........................................... 100
104 . Frontier IR Dual-Range Systems User's Guide

L R
Labels Requirements ............................... 36
Sample area ............................. 20 Electrical .................................. 36
Warning signs ........................... 21 Environment ............................. 36
Laser
Regulations ............................... 17 S
Replacing ................................. 91
Retaining bracket ...................... 93 Safety
Retaining pin............................. 93 Electrical .................................. 16
Securing screws ........................ 93 General .................................... 13
Laser and power supply Laser ....................................... 17
Laser radiation hazard................ 18
Cabling ..................................... 93
Installation ............................... 93 Summary .................................. 12
Laser power supply Use of flammable solvents and
Location ................................... 91 samples ................................ 15
Replacing ................................. 91 Use of nitrogen ......................... 14
Laser safety regulations ................. 17 Safety interlock ............................ 25
Lifting points ................................ 59 Sample area cover
Release clip .............................. 96
Locking bolts
Undoing ................................... 81 Removing ................................. 96
Sample compartment .................... 27
Sampling accessories
M
Diffuse Reflectance .................... 31
Main cover External NIRA ........................... 31
Locking bolts............................. 80 GPOB and LiTaO3 Detector .......... 32
Opening ................................... 80 GPOB and MCT Detector ............. 32
Mains supply HATR ....................................... 31
Fluctuations .............................. 13 Liquid Sipper ............................. 31
MCT detector NIR Fiber Optic Probe ................ 31
Cooling..................................... 75 NIR Tablet Autosampler ............. 31
Liquid nitrogen .......................... 75 NIRA ........................................ 31
WARNING ................................. 75 Sample Shuttle .......................... 31
Moving ........................................ 59 TL 8000 EGA System ................. 31
Universal ATR ........................... 31
N Sampling accessory
Changing ............................ 28, 96
Nitrogen ...................................... 14 Installing new ........................... 97
Removing ................................. 96
O Set Temperature utility .................. 60
Optical system.............................. 25 Shipping container
Stability .................................... 26 Opening ................................... 37
Software
P Installation ............................... 44
Software control
PC connection port........................ 40 Basics ...................................... 46
Performance validation ................ 100 Source
Power switch ................................ 30 High operating temperature ........ 83
Purge fittings ............................... 71 Replacing ........................... 83, 99
Purge ports .................................. 71
Purging ................................. 70, 82
Connectors ............................... 70
Gas regulator pressure ............... 71
Purge gas lines.......................... 70
Index . 105

W
T
WEEE Directive........................... 102
Temperature ................................ 13 Windows ..................................... 30
Top panel .................................... 26 Installing/Replacing ...25, 30, 61, 85
Main cover .......................... 62, 86
U Seal ................................... 61, 85
Windows Start menu ..................... 46
Unpacking ................................... 37
106 . Frontier IR Dual-Range Systems User's Guide

You might also like