Ant 5
Ant 5
ANT-5, 08.00...
User Manual
Acterna SDH Access Tester
ANT-5
© Copyright 2003 Acterna, LLC. All rights reserved. Acterna, The Keepers of
Communications, and its logo are trademarks of Acterna, LLC. All other
trademarks and registered trademarks are the property of their respective
owners.
Subject to change without notice.
Printed in the United Kingdom.
4565-8498.210 Revision 12/17.07.2003
Last saved on 17 July, 2003
Contents
1. Introduction ........................................................................................................................13
1.1 Intended use ..................................................................................................................13
1.2 Features of the SDH Access Tester application software ..............................................14
1.3 Acterna ANT-5 Software version summary....................................................................14
1.4 Acterna ANT-5 Hardware version summary ..................................................................15
1.5 Physical layout and connectors – BN4565/01................................................................15
1.6 Physical layout and connectors – BN 4565/02...............................................................17
1.7 Physical layout and connectors – BN 4565/03...............................................................19
1.8 Safety instructions .........................................................................................................20
1.8.1 Correct usage .........................................................................................................20
1.8.2 Connection to test circuits carrying dangerous voltages.........................................20
1.8.3 Laser safety ............................................................................................................21
1.8.4 Cleaning optical connectors....................................................................................21
1.8.5 Optical attenuators .................................................................................................21
1.8.6 Faults and damage .................................................................................................22
1.8.7 General cleaning ....................................................................................................22
1.9 Power supplies ..............................................................................................................22
1.9.1 Battery charging .....................................................................................................23
1.9.2 Low battery .............................................................................................................23
1.9.3 Displaying the battery charge indicator...................................................................23
1. Introduction
It has independent transmitters & receivers for connection to the following interfaces:
SDH STM-1 electrical line interface
Optional STM-1 optical interfaces, 1310nm and/or 1550nm wavelengths
Optional STM-4 optical interfaces, 1310nm and 1550nm wavelengths
Optional STM-16 optical interfaces, 1310nm and 1550nm wavelengths
PDH E4 electrical line interface, 140 Mbit/s
DS3 electrical line interface, 45 Mbit/s
E3 electrical line interface, 34 Mbit/s
E1 electrical line interface, 2048 kbit/s (balanced and unbalanced)
Optional T1 electrical line interface, 1544 kbit/s & ECL interface
The product may on occasions be used on a working public network and is designed
to be used by a skilled person, usually a technician, who is aware of the precautions
necessary when using measuring instruments. The product is intended only for
connection to circuits carrying Safety Extra Low Voltages (SELV).
Optional software for the ANT-5 is discussed in section 5. For more information on
purchasing any of these specific features please contact the nearest Regional Sales
Office as listed on the back of the manual.
Version Description
BN4565/03 STM-16 units. A T1 and ECL interface are located on the right
(label on reverse) side of the unit. Included in this build is optical power
measurement functionality, NRZ STM monitor port, E1 Hi-Z input
and a PDH T1 Interface. See section 1.7 and 4.22 for further
details.
BN4565/02 A CompactFlash card slot is situated on the right of the unit. See
(label on reverse) section 1.6 for further details.
BN4565/01 A PC Card slot situated on the left of the unit. See section 1.5 for
further details.
Ext Clk input BNC, 75Ω, Unbalanced electrical clock input for Tx at
nickel plated 2048 kbit/s or 2048 kHz
12V DC power 2.5mm 12V Power supply input. For use with type
PPS-1 only
ANT-5
Side view
RJ48 connector
Ext Clk input BNC, 75Ω, Unbalanced electrical clock input for Tx at
nickel plated 2048 kbit/s or 2048 kHz
19V DC power 2.1mm 19V Power supply input. For use with type
3A PPS-2 only
Ext Clk input BNC, 75Ω, Unbalanced electrical clock input for Tx at
nickel plated 2048 kbit/s or 2048 kHz
19V DC power 2.1mm 19V Power supply input. For use with type
3A PPS-2 only
ANT-5
Right-side view ECL input
Note: For reliable operation we recommend only using Acterna cables (e.g. K168,169)
WARNING: Section 1.8, Safety instructions, should be read before connecting
cables to the ANT-5.
Note 1: Optical connectors, where fitted, are provided with dust caps which should be
used whenever an optical lead is not connected.
Note 2: Care must be taken not to connect a long-haul (high power) transmitter
directly to the receiver, since this can overload or even damage some receiver
modules. By connecting an external optical attenuator (as supplied) this restriction can
be overcome.
The ANT-5 should only be used for the purpose and under the conditions for which it
is intended. Operation outside these conditions may be dangerous or may damage the
instrument. Refer to Section 1.1, Intended use, and Appendix A – Specifications
and ordering information for further details.
Telephone lines can have potentially dangerous voltages. Care should be taken when
connecting measuring leads to the line and to the ANT-5, and the instrument should
The BNC, bantam and RJ connectors must be connected to Safe Extra Low Voltage
(SELV) circuits only and must not be connected to any circuit carrying dangerous
voltages.
Optical radiation with wavelengths of 1310nm and/or 1550nm is present at the optical
connectors. Each interface is active if the yellow LED next to the connector is on. The
laser warning icon, , is also displayed with a yellow background when the ANT-5
laser source is active. Maintenance and repair work on the laser sources should only
be carried out by qualified service operatives familiar with the risks involved.
Laser light can cause irreparable damage to the eyes, particularly to the retina.
• To deactivate the ANT-5 laser source (Tx), clear the Laser On checkbox in the
Interface Setup Tx panel. See Section 4.4.2 for further details.
• Never look directly into the generator outputs or the connector end surfaces of the
connected cables (free ends) if the laser sources are activated.
• Never use a microscope to check generator outputs if laser sources are activated.
• Do not activate laser sources until all measurement connections have been made.
Care must be taken not to connect a long-haul (high power) transmitter directly to the
receiver, since this can overload or even damage some receiver modules. By
connecting an external optical attenuator this restriction can be overcome.
If it is considered that the instrument can no longer operate safely, it should be taken
out of service, labelled with the fault condition and secured against unintentional use.
The following are situations where this may be the case:
• if the instrument shows visible signs of damage
• if the instrument no longer operates correctly
• if the instrument has been subjected to any stresses (such as storage and
transport) where the permitted range limits were exceeded.
If the rechargeable battery pack develops a fault it can be replaced locally in an
appropriately equipped workshop. See Section 7.6.2, Replacing the ANT-5 battery
pack, for further details.
With the exception of the battery pack, there are no user serviceable parts in the
ANT-5 or its accessories. Do not attempt to open the case of the instrument – any
attempt to do so will invalidate the warranty.
The outer surfaces and bumpers of the ANT-5 may be cleaned with a cloth slightly
dampened with water or a mild detergent. Abrasives and solvents must not be used,
as they will damage the finish and may remove the labelling.
Switch off and disconnect all cables before cleaning and do not allow water or any
other substance to enter the instrument through the keyboard, connectors or card slot
locations.
The external power supply acts as a charger unit when the instrument is switched off.
When external power is connected, the battery LED, , above the screen will be
green. See Section 2.5, LED and on-screen indicators, for further details.
WARNING: The ANT-5 must only be operated from its internal battery or from
the included PPS power supply. Ensure that the correct power supply is
connected to the appropriate ANT-5. Refer to section 1.5, 1.6 and 1.7 for further
details.
The internal NiMH battery is charged automatically when the instrument is connected
to a live external power supply. An adequate charge will be reached within 3 hours if
the instrument remains switched off whilst charging. It will automatically switch to a
trickle charge mode when fully charged or when the instrument is switched on.
Note: If the temperature of the operating environment is too high, the instrument will
automatically select trickle charge mode to avoid overheating.
Before the instrument is used for the first time it is recommended that the battery is
conditioned by running through two full discharge/recharge cycles as described in
Section 7.6.1, Battery conditioning. This procedure will also establish the initial
conditions for the battery charge indicator.
When no external power supply is connected and the battery charge is getting low, the
battery LED, , above the screen will come on as a steady red light and the
instrument will emit an audible warning. As the battery gets very low the battery LED
will start to flash and further audible warnings will be emitted. An external power
supply should now be connected as the instrument will switch itself off approximately
90 seconds after the battery LED starts to flash.
Note: If the Battery Level indicator shows the warning ‘Reading may be inaccurate’,
the battery conditioning procedure described in Section 7.6.1 should be followed.
2. Getting started
In normal use the ANT-5 should be switched off by briefly pressing and releasing the
Off key. This allows the power-down sequence to be controlled by the ANT-5 system
software, which will ask the user to confirm or cancel the power-down if a test is
running.
Note: For ease of use it is not necessary to close applications before switching the
unit off.
Note: During the start-up process, clicking noises will be made by relays inside the
instrument. This is a design feature and is part of the normal start-up.
2.3 Menus
The operating characteristics of the
ANT-5 are controlled by sets of
hierarchical menus which can be
displayed or hidden by pressing
Alt+Menu (press Menu whist holding
down the Alt key). The highest level
of the menu structure is represented
by a bar at the top of the screen with
one choice highlighted. The System
menu is shown here as an example.
Use the direction keys to move the highlight around each menu and then press Enter
to either select an item or to display the next menu level in a similar way. Press Esc to
close the current menu and move up one level.
When the setup menus are hidden and then displayed again, the same level will be
shown if it is valid to do so. If a test is running, the application will not update the
displayed backdrop while a menu is displayed in the foreground.
The box will remain on the screen until a key associated with one of the buttons is
pressed or until one of the default keys Enter or Esc is pressed. Enter is used to
accept the highlighted choice and Esc is used to cancel the message box without
accepting any choice.
The top portion of the screen is used by the tester application to simulate LED
indicators on-screen in order to display information which is relevant to the current
setup of the application. The behaviour of the on-screen LEDs can be set to be
latching or non-latching in the LEDs section of the System menu, which can be
displayed by pressing the Alt + Menu key at any time. For further details See Section
4.24, On-Screen LED Indicators.
The four LEDs above the screen are used for general status information as follows:
Alarm LED The non-latching Alarm and Signal LEDs are used by the
tester application to summarise the test status, even if the
Signal LED
screen has been blanked to save power.
The Signal LED will be green if there are no alarms or errors
to display on the on-screen LEDs. In this case the Alarm
LED will not be lit.
The Alarm LED will be red if one or more on-screen LEDs
are shown in red to indicate alarm or error conditions. In this
case the Signal LED will not be lit.
Screen LED Steady green indicates that the screen has been blanked to
save battery power. Press any key to restore the screen.
See Section 2.13, Power saving features.
A 4 5 6
B PQRS TUV WXYZ
C 7 8 9
SPACE
0 #
Alt
Alt+ Beeper / Press with the Alt key held down to display a scale
+ Contrast key for adjusting the screen contrast.
Menu
Application Displays the application menu, or turns off the menu
Menu key if the menu is already displayed.
Alt
System Displays the system menu, or turns off the menu if
+ Menu key the menu is already displayed.
Menu
Function keys The function keys are used by the tester application.
Active keys are labelled at the bottom of the screen.
Alt
Alt+Escape Use Alt+Esc to switch between applications. See
+ Section 3.1.1, Switching .
Esc
Alt
Alt key Extends the functions of the standard keys (see
above).
Alt
Alt+Hash Use Alt+Hash to save the current screen in a file or
+ send it to a printer on the serial port, depending on
# the Screen Dump setting in the Printer section of
the System menu.
Backlight key Toggle the backlight between high and low level.
To adjust the screen contrast, press the Beeper/Contrast key whilst holding
the Alt key down.
The contrast is increased or decreased using the directional keys. A scale will be
displayed on the screen which reflects the contrast level. To remove the scale from
the screen, press the Enter key to accept the new setting or the Esc key to cancel it.
The ANT-5 will continue to operate during contrast adjustment but will not receive any
keypad input until the adjustment is completed.
The ANT-5 has an internal beeper which can be used to provide an audible
notification when defects and anomalies are detected. When the SDH
Access Tester application is running, this feature can be turned on or off by
pressing the Beeper/Contrast key.
The six unmarked keys at the top of the keyboard are function keys which
can be used in different ways by applications. Labels will appear along the
bottom edge of the screen to show that a function key is active and to
indicate its purpose.
The alphanumeric keypad is normally used for entering numbers in setup menus but it
can also be used for entering alphabetic and special characters. Entry boxes generally
have a character on their right hand side to indicate that a specific type of data is
expected, such as a ‘b’ for binary, ‘d’ for decimal or ‘h’ for hexadecimal.
The ABC key is used to enter and leave the extended alphabetic mode, in A
which strings of alphanumeric and special characters can be entered by B
C
using repeated key presses punctuated by short pauses. The current mode
is indicated in non-specific data entry boxes by an ‘A’ or a ‘1’.
Each key has a set of characters associated with it, and the sequence will repeat in a
cycle until either a different key is pressed or a pause of one second occurs.
For example, repeatedly pressing the 9 key will display the sequence W-X-Y-Z-9-W-X-
Y… until there is a pause or another key is pressed. The last character will then be
accepted and the screen cursor will advance to allow entry of another character. To
switch between upper and lower case letters, hold any letter key down for one second
or more. Use the Del key to backspace over any mis-typed characters.
The and # keys on either side of the 0 key are used for entering non-
alphanumeric characters. The sequence for the # key includes the following
characters: - # + ! , : ” ’ & ( ). The 0 key can be used to enter space characters.
The ANT-5 will sound a series of beeps when the screen has been saved or printed. If
the screen is being saved in a file, it is particularly important not to remove the card
before these beeps have indicated the end of the process.
page also has a rectangular border that delimits the printable extent of the paper. This
can be used to ensure that the paper size and unprintable area are correct. If the
border is truncated (i.e. a complete rectangle is not visible) then the unprintable area
may be set too small or the paper size too large. Conversely, if the border is too far
from the edge of the paper then the unprintable area may be set too large or the paper
size too small.
Note: The SDH Access Tester application also uses the Menu key to provide a drop
down facility which can be used to print a comprehensive results and setup report via
the serial port (See Section 4.23). To use this feature successfully, it is important to
have the correct Printer Driver and Serial Port settings rather than relying on the
simple test data mentioned in Sections 2.7.1 and 2.8.
See Appendix C – Serial port and adapter cables for a diagram of the serial port.
A remote operation option is available for ANT-5, (See Section 5.4), enabling the unit
to be operated remotely from a Windows PC. This can be via the telephone network
using modems to the V.24 port or over Ethernet.
Note: BN4565/01 units will require an external Ethernet PC Card. Only 3COM 589
chip-set PC cards are supported, model numbers 3CCE589EC and 3CCE589ET.
An internal Ethernet port is fitted from BN 4565/02 onwards, care should be taken not
to confuse the E1 120Ω (RJ48) and the Ethernet port -> (RJ45).
BN4565/01 units
BN 4565/02 onwards
Note1: The Comms menu contains a baud rate setting which should not be confused
with the baud rate setting of the serial port. This should be set up in the Serial Port
section of the System menu.
Note 2: When typing in the ANT-5 IP address into a PC environment care should be
taken to avoid any leading zeros. For example 141.169.126.076 would be replaced
with 141.169.126.76. However, all twelve characters are needed for the ANT-5.
Note 3: Edited IP addresses will be available after the ANT-5 has been re-booted.
The cards should be inserted with the label side uppermost. Press into the slot until
firmly located. The ANT-5 will beep to indicate it has recognised the card. If no beep is
heard refer to section 7.3.
Ethernet cards can be used in the PC Card slot to enable remote access. See Section
2.8.1 for further details.
Note: The instrument does not automatically adjust the time for daylight savings. You
must set or clear this checkbox manually.
The screen has a backlight which can be switched between high and low
level by pressing the Backlight key. Use the low level to save battery power.
When the Display Saver is selected, the screen will turn off after a set period of
keyboard inactivity in order to save battery power. When this happens the screen LED
( ) will come on to indicate that the screen has been blanked. The full screen
display can be restored by pressing any key.
The time delay before blanking occurs can be set up in the Power Saving section of
the Preferences menu.
This section can also be used to set up the ANT-5 to restore the screen automatically
when there is a change in the status of the on-screen simulated LEDs. The behaviour
of these LEDs is controlled by the current tester application.
When Auto Power-off is selected, the ANT-5 will switch itself off after a set period of
keyboard inactivity in order to save battery power. The time delay before this occurs
can be set up in the Power Saving section of the Preferences menu.
The ANT-5 will not switch itself off if a test is running. However, if the unit is
accidentally switched off, or there is no power left in the battery, results will be saved
in the internal: device of the File Manager. These will be saved as AUTO files, with
suffixes of .A5R, the time and date will also be present so that the results are easily
identified.
Note: The ANT-5 operating system will need to be re-started before this change can
take effect. This requires turning the instrument off and then turning it on again.
The file manager and application launcher are system applications which can be
displayed at any time, without closing the SDH Access Tester application, by pressing
Alt+Esc (press Esc whilst holding the Alt key down). This can be useful for changing
the ANT-5 setup parameters without interrupting a test.
Pressing Alt+Esc repeatedly will switch between the ANT-5 system and the tester
application in turn.
If a system application is using the screen whilst the tester application is running, the
tester application will continue to run in the background and will accumulate test
results as normal. In this state the test application will maintain its control of the Alarm
and Signal LEDs ( and ) and may also be able to maintain the on-screen
simulated LEDs (this will depend on the particular system application being displayed).
Note: The “internal:” device is the internal RAM storage. In normal circumstances the
ANT-5 operating system copies the data stored in this area into non-volatile memory
during its power-down sequence. Data may be lost from this device in exceptional
circumstances when the power-down is not controlled by the operating system, for
example if the ANT-5 is switched off by holding the Off key down for more than 6
seconds.
3.2.2 Directories
The File Manager in… message at the top of the screen shows the current position in
the directory structure. To display the contents of a subdirectory, position the highlight
over the directory name and press the Enter key or right directional key.
To move back to a higher directory, press the left directional key or position the
highlight over the <back one> entry and press the Enter key; the contents of the
parent directory will then be displayed.
To create a new directory, press Menu to display the File Manager menu then select
Create directory from the File section.
At various times the user may be required to specify a filename. Typically this will be
for saving and retrieving data to and from the card.
When saving a file, the directory structure is displayed and navigated in the same way
as in the File Manager utility (See Section 3.2.2, Directories). Before saving a file in a
directory, the directory must have already been created.
To save a file using an existing filename, highlight the required file and press Enter.
To save a file using a new filename position the highlight over the <new file> entry,
press the Enter key and enter the new filename.
Valid filenames can have up to 8 letters or numbers in their name. They can also
contain a file extension of up to 3 letters, which usually indicates the type of the file.
The extension is separated from the first part of the name by a dot. Filenames are
always displayed in upper case characters. The following are examples of valid
filenames:
ABC123
RESULTS1
TEST1.DAT
Note: The ANT-5 does not support long filenames. If a card containing files with long
filenames is inserted into the ANT-5, the long filenames may be lost.
A FIFO (First-In, First-Out) buffer is used to manage automatically saved files. Results
from the last run test are saved as AUTO1.A5R in the internal: device of the file
manager when the key is pressed. Only three files can be saved AUTO1,
AUTO2 and AUTO3, consequently the next time a file is saved as AUTO1, the
previous AUTO1 is renamed AUTO2, etc and file AUTO3 is discarded. Therefore, we
recommend that important files are copied or renamed.
1. Select file manager (Press Alt + Menu and select Tools > File Manager),
4. Press the Menu key and select File to Copy or Rename the file.
To format a card, insert an appropriate card and run the File Manager utility. If the
card has already been formatted it should appear as a device named “card:”. To
examine the contents of the card prior to formatting, move the highlight to this device
and press the Enter key or right directional key.
Use the Menu key to display the File Manager menu, then
select Format (PC) Card from the Tools section. A warning
message will be displayed to request confirmation before the
data stored on the card is erased. ‘Press’ the OK button to
proceed.
The File Manager can be used to transfer data files from the ANT-5 to a PC using the
serial port. The following example assumes that Windows HyperTerminal and the
COM1 port are available for use on the PC:
1. Connect an appropriate serial cable, such as K1524 or K1619, from the ANT-5 to
the COM1 port on the PC/Laptop
2. Switch on the ANT-5 and make a note of the Serial Port settings from the System
menu.
3. Run Windows HyperTerminal on the PC and choose a name for the connection.
5. When the COM1 properties appear in the HyperTerminal window, choose port
settings to emulate those on the ANT-5 and set the flow control to Xon/Xoff.
6. In the HyperTerminal window click on Transfer, and then Capture Text. Give the
file a name, and click on Start.
7. Open the ANT-5 File Manager and highlight the file to be transmitted.
8. Press the ANT-5 Menu key and select Transmit… from the File menu. A dialog
box will appear on the ANT-5 screen confirming the transfer and the
HyperTerminal screen will fill with data.
Note: It is also possible to send a test pattern by selecting Send Test Pattern under
the Serial Port item in the ANT-5 System menu.
Tip: Master configuration files for an ANT-5 can be simply copied to other units by
transferring the .A5C files.
Help > SDH Select this option to view a schematic of the SDH physical layer.
Structure
Help > SONET Select this option to view a schematic of the SONET physical
Structure layer.
Note: The tabbed pages for Performance Analysis, Repetitive BERT, APS, Delay and
Tributary Scan will only become available if the measurement is selected from the
Measurement Selection panel. See section 4.3 for further details.
Each page can be accessed by using the directional keys to highlight its page tab.
Setup view allows the user to set up the interface and test parameters. If a test is not
running, the settings can be easily changed by moving around the screen with the
directional keys and selecting items with the Enter key. If a test is currently running
the setup pages can be viewed but not changed.
Results view allows the user to view the test results as a test is running or after it has
been stopped. Results view also provides facilities for injecting anomalies and defects
and selecting different displays of test results.
Some of the Results view pages are split into two halves. The left half of these pages
are used for displaying results and the right half is used for setting the various test
parameters.
Note: The results of a test are always displayed with regard to the test parameters
which were set up at the time the test was run. These may not match the current setup
parameters if any changes have been made since the test was run.
Display area
Page tabs
Function keys
• The top portion of the screen is used by the application to simulate LED indicators
on-screen in order to display information which is relevant to the current setup of
the application. See Section 4.24 The title of the current application is shown in the
application title bar.
• The status bar is used to display the status of the interface and the current test.
• The display area has a number of tabbed pages which can be made active by
selecting the tabs down the right side. The display area is used for setting up tests
and displaying results. For some pages the display area may be split into left and
right areas and may have multiple results pages which can be selected.
• The six unmarked keys at the top of the keyboard, , are function keys to
control certain features of the ANT-5. The function of each active key is identified by
an animated button on the screen just above the actual key.
4.1.1 Moving around the screen
Use the directional keys for moving around active parts of the display area,
such as buttons and list boxes, which are highlighted in turn. The page tabs
can be highlighted by moving up and down the right side of the display area.
Press the Enter key to operate a highlighted button or to select a highlighted list
item.
The status bar is used for displaying the status information for pattern sync, defect
notification and anomaly/defect injection. The elapsed time of the test in hrs:min:sec
format is shown in position k. If a test is not running, the elapsed time of the previous
test will be displayed. Position l shows how long the test will run. (See Section 4.6 to
set the duration of the test.)
Elapsed time
Status bar
a b c d e f g h j k l
Activity icons
Rx Icons
The pattern sync icon (position a) displays a steady synchronised digital
signal if the instrument is synchronised on a BERT pattern.
The defect detected icon (position a) flashes if any other defect condition
is currently being detected at the selected Rx input, such as LOS.
Assuming the ATM option has been purchased and enabled (see Section
6) this icon (position d) becomes animated when load ATM traffic is
monitored at the receiver (Rx).
Tx icons
The anomaly injection icon (position e) displays an animated syringe
when anomalies are being injected. The icon is blanked when anomaly
injection is not active.
These icons show if the beeper is turned on or off, (position h). Pressing
the beeper key on the keyboard controls this function.
These icons (position j) show the current status of the test. The green
light is displayed while a test is running and a red light is displayed when
a test has been stopped. The amber light is displayed when a test is due
to begin from a delayed start. See section 4.6.
The six unmarked keys at the top of the keyboard, , are function keys to control
certain features of the ANT-5. The function of each active key is identified by an
animated button on the screen just above the actual key. The state of each button
icon, whether it is shown pushed in or not, is also used to indicate current functional
states such as Setup view or Results view. The button icons are greyed when a
function is not available.
Firm key Function
Switch to Setup view.
Alt
Switches to the Signal Structure setup page
+
Alt
Switches to the Results summary page
+
or These keys are used from the results page when clock offset is
enabled. The transmitter offset can be adjusted in steps of
0.1ppm by pressing the required offset firm key, or offset in
increments of 1.0 by pressing and holding down the firm key.
or Start a test, or Stop the test if one is running. (Once a test has
been stopped, up to three files are automatically saved on the
internal device, see 3.2.5 for further details).
with the Enter key. If a test is running, some parameters on the setup pages can be
viewed but not changed and these will be greyed to indicate that they are disabled.
When making changes to the setup, any value selected for a parameter (by pressing
the Enter key for example), becomes effective immediately unless it is not a valid
value. In this case the previous value will be retained or a substitute value used.
Note: Whether a test is running or not, the transmitter will start to transmit
continuously as soon as a valid configuration is entered. The receiver will look for and
attempt to recognise a valid signal on its Rx input in readiness for the Start key being
pressed. When a test is not running, the LEDs and Rx icons will continuously indicate
errors and alarms but any other results displayed will be the results of a previous test.
Feature Section
Performance Analysis Setup 4.7
Performance Analysis Results 4.12
Repetitive BERT 4.17
Automatic Protection Switching 4.18
Delay 4.20
VC12 Tributary Scanning Setup 4.8
VC12 Tributary Scanning Results 4.16
Note: To select Tributary Scan ensure VC12 is selected as the Virtual Container for
the Transmitter (Tx) and the Receiver (Rx). (See Section 4.4.3 and 4.4.9).
Pattern
Timeslot
Interface
Pattern Hierarchy
Timeslot Termination
75ohm
Virtual
PDH
Container Channel Interface
The procedure for setting the Rx elements of the test structure is very similar to the
setting of the Tx elements except for the Interface Setup Rx window, which has a
different set of parameters.
The list of available parameters for each part of the structure may vary depending
upon the configuration selected in the other elements. For example, if a PDH interface
is selected, the Virtual Container setup dialog box will not appear, since this box is
only applicable to SDH.
The signal structure always appears with a test pattern source and test pattern
detector. If the incoming signal is expected to contain network traffic, the Traffic
pattern may be selected to disable the receiver’s pattern detection process and the
reporting of LSS defects.
When setting up the various parts of the signal structure for the ANT-5, it is advisable
to make changes to each element of the structure in the order that they are referred to
in this chapter.
Note 1: The optical Rate for STM-1, STM-4 and STM-16 are also selectable if the
options have been purchased and enabled.
Note 2: AU-3 Mapping will also become selectable if the option has been purchased
and enabled.
Clock Source The transmitter Clock Source can be set to Internal, From Rx,
Ext. Data 2048kbit/s* or Ext. Clock 2MHz*. (* SDH only).
For the Internal setting, the clock for the transmitter is derived
from the internal 2048 kbit/s clock. For the From Rx setting, the
clock for the transmitter is derived from the Rx connector.
The Ext. Data 2048kbit/s and Ext. Clock 2048kHz settings are
only available if an SDH structure has been selected. For these
settings, the clock for the transmitter is synchronised to the
signal on the Ext. Clk connector.
Clock Offset The frequency offset of the received signal is the difference
between the received frequency and the expected frequency. All
network equipment (NE) can tolerate a certain amount of offset.
However, if the limits are exceeded a NE will fail to synchronise
to the received signal which will result in network errors.
Skewing the frequency is a standard test during installation to
ensure that an installed unit can tolerate the frequency offsets
occurring within the network.
Line Code The Line Code setting automatically defaults to the setting
appropriate to the interface type:
Interface Line Code
SDH Optical NRZ
SDH Electrical CMI
E4 CMI
E1 & E3 HDB3
DS3 B3ZS
If testing within a PDH hierarchy is required, the PDH tributary should be selected.
The test pattern will then be injected into the selected SDH virtual container with the
selected PDH parameters.
Note: VC11/TU11 and VC11/TU12 will also become selectable if the option has been
purchased and enabled.
For example, if a VC-3 has been selected for a test, then only the TUG-3 parameter
can be adjusted, enabling selection of which TUG-3 within the VC-4 will carry the
VC-3 payload. Alternatively, if a VC-12 has been selected for a test, then the TUG-3,
TUG-2 and TU-12 can all be adjusted, enabling selection of which TU-12 will carry the
VC-12 payload.
numbers will automatically change to match the new VC. The table in Section 12.3,
SDH Tributary Numbering, gives the relationship between the VC numbers and TUG
numbers for the Tributary and Timeslot (ITU-T) numbering methods.
If Bulk has been selected as the Tributary in the SDH Virtual Container then the PDH
Setup Tx panel will not be available.
Note: If an SDH line interface has been selected, only bit rates and framing types
appropriate for the SDH configuration will be displayed as options within this panel.
PDH Low Rate is only applicable if Nx64 testing is required, See Section 4.21.
Note: The pattern element will be greyed out if the round trip delay measurement has
been activated. See 4.20 for further information.
Set the Test type to Inline monitor to monitor signals at the electrical monitor point
without the need for optical spitters. The inline monitor allows the unit to be connected
in circuit for monitoring live links. The received signal is re-transmitted unaltered and
simultaneously monitors for alarms and errors. This is useful when in-service quality
measurements are required on optical circuits where there is no access to optical
splitters.
Note 1: The optical Rate for STM-1, STM-4 and STM-16 are also selectable if the
options have been purchased and enabled.
Note 2 AU-3 Mapping will also become selectable if the option has been purchased
and enabled.
Line Code The Line Code setting automatically defaults to the setting
appropriate to the interface type:
Interface Line Code
SDH Optical NRZ
SDH Electrical CMI
E4 CMI
E1 & E3 HDB3
DS3 B3ZS
DS1 B8ZS
Sensitivity If the Electrical interface is selected, the sensitivity can be set
for ITU-T, PMP (Protected Monitor Point) or HIGH connections.
The PMP setting includes appropriate linear gain to compensate
for the attenuation at the protected monitor point. The electrical
interface will be terminated in either 75Ω (BNC) or 120Ω (RJ48).
See 9.4.3, for further information.
Pulse Shape HIGH or DSX can be selected as the pulse shape and is used
when testing PDH tributaries with a bit rate of 45M
Note: A HIGH pulse shape should not be confused with high
impedance.
If testing within a PDH hierarchy is required, the PDH tributary should be selected.
The test pattern will then be injected into the selected SDH virtual container with the
selected PDH parameters.
Note: VC11/TU11 and VC11/TU12 will also become selectable if the option has been
purchased and enabled.
Note: If an SDH line interface has been selected, only bit rates and framing types
appropriate for the SDH configuration will be displayed as options within this panel.
PDH Low Rate is only applicable if Nx64 testing is required, See Section 4.21.
The Traffic pattern should be selected if the incoming signal is expected to contain
network traffic. Selecting this option effectively disables the pattern detection process
and the reporting of LSS defects.
Where the system under test is symmetrical, and the Tx and Rx structures are
required to be identical, time can be saved by setting up either the Tx or the Rx
structure, and then pressing either the or keys as appropriate. The
structure will be copied from Tx to Rx, or vice versa, as selected.
The Tx and Rx structure can also be set up independently, so that, for example, the
Tx can be sending a PRBS of 215-1 in a VC-12 carried within an optical STM-1, whilst
the Rx can be expecting to receive a PRBS of 215-1 in a 2Mbit/s PDH signal.
Note: Previous mixed mode configurations saved prior to software version 08.00 will
no longer be valid.
If required the following control characters can be inserted into the trace label strings
for expected and transmitted bytes. Alphanumeric characters can be entered using the
keyboard keys.
Note: The transmitted and expected frames become visible if SDH or SONET are
selected as the hierarchy type for Tx and Rx.
Test start The timer can be set up for Manual start or Delayed start.
Start time / date If Delayed start is selected, further entry fields can be used for
setting the Start time and Start date. These can be used to
enter any time and date in the future.
Duration The test can be set to run for a specified Duration from 1
second up to 99 days. The time selected will be shown on the
status bar.
Notes: The Stop key will always be available to stop tests manually, regardless of the
Timer settings. The time and date can be set in the Time & Date section of the
System menu.
For each analysis type, a set of hierarchies are available, which are dependent upon
the Analysis type selected. Some of these are only applicable to SDH, and will be
greyed-out in the list of options if a PDH Rx interface is selected.
For each selection of Analysis and Hierarchy, selecting the Default button restores
the default settings to the available parameters.
The ability to switch tributary signals without fully de-multiplexing and re-multiplexing
the SDH signal is a very important feature of SDH. Using either an add/drop (ADM) or
cross-connect (DXC), networks can be re-configured to cater for network changes.
4.9 Results
The results screens are selected by pressing the Results function key. This displays a
range of tabbed pages, similar to the setup pages. Some of the results pages use a
split screen layout, with the large OK in the left half of the screen, and other details of
the test in the right half of the screen. Other results pages use the whole screen and
may have multiple pages which can be selected from the main screen.
Note: Assuming the previous test was successful, the large OK will be greyed out
whilst the current test is being configured.
Once the test has started the ANT-5 samples the received signal from the network
element then calculates and displays the result of the frequency offset. The ANT-5
displays the Rx offset to check a network element is transmitting correctly.
The transmitter offset can be adjusted in steps of 0.1ppm by pressing the required
offset firm key, or offset in increments of 1.0 by pressing and holding down the
appropriate firm key.
The range over which the offset can be skewed depends on the initial value set in the
Interface Setup Tx panel. For example, if the test starts with a negative value, its
associated range will be from –100.0 to –0.1. However, if the test starts with a positive
value, the range will be from 0.0 to 100.0. The following example shows the results
associated with clock offset.
Note: The offset firm keys will be greyed out when the value reaches the end of its
range.
4.11 Injection
A wide range of anomalies and defects can be injected into the transmitted signal,
depending upon the signal structure selected.
If an SDH line interface is selected, the relevant SDH anomalies and defects can be
injected. If a PDH test signal is injected into the SDH virtual container, the relevant
PDH anomalies and defects can also be injected. If, however, a PDH line interface is
selected, then only the PDH anomalies can be injected.
In addition to injecting into either the SDH or PDH structures, injection of TSEs (bit
errors) into the relevant payload is also available using the Transmit Pattern.
The number of anomalies to be injected is selected from the Rate list box and can be
set to Single or Continuous. The Continuous rates available are 1E-2, 1E-3, 1E-4,
1E-5, 1E-6, 1E-7, 1E-8, 1E-9 or 1E-10, the available rates depending upon which
anomaly is selected.
To inject a single anomaly into the Tx path, set the Rate to Single then press the
Inject firm key. The Inject firm key will now be depressed indicating that anomalies are
being injected. To inject anomalies at a Continuous rate, set the Rate, then press the
Inject firm key. The injection icon, , displays an animated syringe when anomalies
are being injected. To stop injecting, press the Inject firm key, the injection icon now
disappears.
The Inject firm key appears automatically in the Injection page. To allow injection
from any other page, users can view the Inject firm key by pressing the Alt button.
2 frames
Note: The value for the Repetition Period is always greater than the value of the
Frames. The ANT-5 will automatically adjust the values if necessary.
The Inject firm key appears automatically in the Injection page. To allow injection
from any other page, users can view the Inject firm key by pressing the Alt button.
The Performance Analysis results page enables the user to view the performance of
the device or system under test according to a range of standards. The standard
applied is selected in the Performance Analysis Setup page – See Section 4.7.
Note: The Performance Analysis results of any test are only valid in the context of
the setup which was used at the time that the test was run and the results were
recorded. It is not valid to use the Performance Analysis setup page to select a
different Analysis or Hierarchy type in order to view existing results without running a
new test.
Note: Defect panel LED’s turn yellow during a test to show historic alarms, that is,
defects that have been raised during a test but are not currently active. To reset the
LEDS press Alt and Menu and select System >LEDs >Reset LEDs.
4.13.4 Graphs
time units. The horizontal scroll bar can be used to scroll in steps of 10 or 60 time
units, as described in the example below.
The list of anomalies or defects is shown to the left of the chart and can also be
scrolled vertically if required.
A detailed zoom feature is also available in the Anomalies Graph. This allows error
rates to be shown down to one-second resolution, as shown in the example below.
‘—’ ‘—’ appears in Analysis mode to indicate bytes for which a value
cannot be determined.
‘TI’ ‘TI’ appears in Analysis mode to indicate bytes for which a correct trace
identifier value has been recognised.
‘xx’ ‘xx’ appears in Injection mode to indicate bytes which may not be
injected.
The POH table is displayed on the right of the window, and its contents will depend
upon the signal structure that has been selected.
In Analysis mode, the value of individual received overhead bytes may be examined
in detail using the Overhead Zoom feature. In Injection mode, the same feature can
be used to edit the value of individual transmitted overhead bytes in hexadecimal,
binary or ASCII format.
For an application note on this topic, visit our Web Site, www.acterna.com, or contact
the nearest Regional Sales Office as listed on the back of the manual.
Note: The received HP and/or LP paths should be activated before starting the test.
This can be done by selecting the HP and LP check boxes as required in the TCM
view of the SOH Setup page. If neither box is selected, the TCM results page will not
be available. For details See Section 4.5.4, Tandem Connection Monitoring.
A Repeat Period of I second is adequate for measuring high error rates, but this
period can be adjusted to a maximum of 99 seconds to test for lower rates.
Note: It is important not to confuse the results of the repetitive BERT test for a specific
interval with those on the results page which show the results for the duration of the
test. Further a repetitive BERT is not recommended from a delayed start.
The Switch time limit (t1) can be set within the range of 10 to 2000ms. The Gate
Time (t2) is used to monitor multiple switches and is achieved by continuing the
detection process for an extended period, with a range from 100 to 5000ms.
By checking the Repetitive box the test will repeat automatically once the switch has
been detected. A green tick shows that the test has passed. A test will fail (shown by a
red cross) if the switch time has not been fast enough or there has been any defect).
Note: The pointer analysis page will only be available when there is an SDH signal on
the Rx. The TU column will only be visible if there is a tributary path set within the
signal structure. It is advisable that From Rx is set as the Clock Source in the
Interface Setup Tx panel.
4.20 Delay
Note: The Delay tab will become available if selected from the Measurement
Selection page. See Section 4.3 for further details.
The Result shows the latest value, there is no history. If a good signal is received, the
units for the result will be in µs if the result is less than 1ms, or ms if it is greater.
Where there is no valid result an asterisk will be displayed, or alternatively the cause
of the problem will be displayed if the result is due to signal problems, e.g. LOS.
Note: When the Delay Measurement is turned on, the PRBS button will contain the
word Delay on the signal structure page which will be greyed out.
4.21 Nx64
The ANT-5 can perform BER testing on a single user selectable 64k timeslot; where
each timeslot has traditionally been used to carry a voice channel; alternatively
multiple timeslots can be tested, supporting data pipe services which can consist of
any permutation of timeslots.
Note: The timeslot page will only be available if 64k is set as the PDH Low Rate in
the PDH Setup panel. Choose 2M as the PDH High Rate to select up to 32 x 64kbit/s
timeslots, or alternatively choose 1.5M as the PDH High Rate to select up to 24 x
64kbit/s timeslots.
Selecting the Timeslot element for Tx or Rx opens the Timeslot Setup panel. (The
element is referred to Ch1 see the graphic in section 4.4).
Note: The ANT-5 will only be able to transmit and receive nx64 when E1 framing is
part of the hierarchy. Channel 0 is used for FAS/NFAS and will always be greyed out;
channel 16 (used for signalling purposes) will be greyed out if either PCM30 or
PCM30C are selected as part of the framing structure.
Remember: Connect an external optical attenuator to the optical Rx port as care must
be taken not to connect a long-haul (high power) transmitter directly to the receiver,
since this can overload or even damage some receiver modules.
File > Open > Select this option to load an instrument configuration into the
Configuration ANT-5, which has previously been saved either to a card or to the
ANT-5 RAM. The file window will list all valid configuration files on
either the card if it is inserted, or in the ANT-5 RAM, along with
their date and time of creation and a brief description.
File > Open > Select this option to load a set of test results from a previous test
Results into the ANT-5, which has previously been saved either to a card
or to the ANT-5 RAM. The file window will list all valid results files
on either the card if it is inserted, or in the ANT-5 RAM, along with
their date and time of creation and a brief description. The results
file also includes the corresponding configuration from the time of
the saved test, and will change the configuration of the instrument
when the results file is loaded.
File > Save > Select this option to save an instrument configuration from the
Configuration ANT-5 either to a card or to the ANT-5 RAM. A configuration can
either be saved to an existing file, or a new file can be created. If
a new file is required, the new file name and a brief description
can be manually entered using the ANT-5 keyboard. Assuming
the ANT-5 is networked the configuration can also be saved and
transferred to other units using FTP.
File > Save > Select this option to save a set of test results from the ANT-5
Results either to a card or to the ANT-5 RAM. The set of results can either
be saved to an existing file, or a new file can be created. If a new
file is required, the new file name and a brief description can be
manually entered using the ANT-5 keyboard. Results saved using
this option can only be read back into the ANT-5 instrument.
File > Export > Select this option to export an instrument configuration to a card
Configuration in CSV (comma separated values) format. Results exported using
this option can be read by other PC applications, but cannot be
imported back into the ANT-5 instrument.
File > Export > Select this option to export a summary of the test results to a card
Results in CSV (comma separated values) format. Results exported using
this option can be read by other PC applications, but cannot be
imported back into the ANT-5 instrument.
Note1: To use the Print Results feature successfully, it is important to have the
correct Printer Driver and Serial Port settings rather than relying on the simple serial
port tests mentioned in Sections 2.7.1 and 2.8.
Note2: Previous mixed mode configurations saved prior to software version 08.00 will
no longer be valid.
Note: Ensure that the necessary drivers are enabled and the card is located in the
correct slot. Contact your IT department for further assistance if required.
8. This file displays a host of information relevant to the results and the configuration of
the unit.
3. Use the directional keys to select either the General Preferences or the Printer
Header Defaults screen. The graphic above shows the General Preferences
screen which indicates that the laser, speaker and the ANT-5 application will
launch automatically. The name of the operator and user information can be
inserted from the Printer Default screen.
Note: Another icon will be present only when the GUI is used with the Remote
Operation option. Select the cable icon to change the IP Address of the ANT-5. If the
IP address is edited whilst running the application, close and restart to enable the
change.
Option BN Number
PDH Mux/DeMux 4565/00.58
Concatenated Mapping 4565/00.59
Remote Operation 4565/00.60
Remote Control 4565/00.61
SONET 4565/00.62
AU3 & VT* Mappings 4565/00.53
(*requires SONET option)
ATM (see section 6) 4565/95.54
Note: Options purchased with a factory built ANT-5 will already be installed.
Note: If an invalid key has been entered incorrectly three times, refer to the nearest
technical assistance centre for advice. (This will not prohibit the unit being used for
other purposes).
The generation feature allows the insertion of alarms and errors at all levels in the
chain (including E2).
Selecting NFAS setup… opens the NFAS Setup Tx panel. (This is available only
when the 2Mbit/s (E1) signal structure has been selected for transmission).
Note: The Tx hierarchy type should be set to PDH to activate the PDH Mux.
Note: The Rx hierarchy type should be set to PDH to activate the PDH de-Mux.
Note 1: Only 3COM 589 chip-set PC cards are supported, model numbers
3CCE589EC and 3CCE589ET. These cards are used only when an Ethernet
connection is made via the PC Card slot. An Ethernet port is fitted from BN 4565/02
onwards, care should be taken not to confuse the E1 120Ω (RJ48) and the Ethernet
port -> (RJ45).
Note 2: Remote operation currently supports the SDH Access Tester Application and
is available only in English.
Note 3: The V.24/PPP connection is available only for the BN4565/01. The V.24/PPP
connection is currently not available on the BN4565/02 or BN4565/03.
Ethernet Ethernet
ANT-5
Modem Modem
V24/PPP
ANT-5
Note: Ensure that the ANT-5 unit has been fully upgraded and that the corresponding
version of the client software has been installed and is used.
Follow steps 1-3 below if the remote operation client needs to be installed.
1. Insert the ANT-5 Upgrade CD-ROM into the CD drive of the PC.
2. From Windows click on Start > Run and type in:
D:\ANT-5 Remote Operation Client\setup.exe (replace D with the drive
letter for your CD-ROM drive if necessary).
3. Follow the on-line instructions to complete the installation.
Note 1: The next time the application is launched, the previous IP address will be
shown as default, click on OK to proceed. If required, the IP address can be edited.
Note 2: When typing in the ANT-5 IP address into a PC environment care should be
taken to avoid any leading zeros. For example 141.169.126.076 would be replaced
with 141.169.126.76. However, all twelve characters are needed for the ANT-5.
Note 3: Edited IP addresses will be available after the ANT-5 has been re-booted.
After a short while the ANT-5 graphical user interface will load on the PC, green LEDs
will flash in the Tx and Rx boxes of the ANT-5 indicating that the link has been
established.
Tip: Create a directory on your PC, e.g. ANT-5. Select Menu > File > Save >
Configuration to save configuration files to this directory which later can be opened
by selecting Menu > File > Open > Configuration. Alternatively files can be copied to
other units using FTP, See Section 3.3.2, File Transfer for further details.
5.4.2 Exporting
From the GUI faceplate select Menu > File > Export. From the standard Windows
dialog box, .csv files can be saved to any directory on your PC.
5.4.3 Printing
From the GUI faceplate select Menu > File > Print. A standard Windows dialog box
will allow users to select and configure a printer.
2. Check that IP address of the physical ANT-5 matches the one used from the GUI
faceplate. (See Sections 2.8.1 and 4.25)
3. Press i from the gui faceplate. If the M/M Version shows a series of zeros,
check the IP address and re-launch the application from the PC.
4. Errors may occur on re-launching the ANT-5 application if previously it has failed to
shutdown correctly. In this instance delete the Auto.A5C file from your C: drive
and re-launch the application from your PC.
5.6 SONET
5.6.1 Introduction
SONET and SDH share many common principles of operation and architecture. A
comparison of the transmission rates supported by ANT-5 is given below:
Electrical Optical
STS-3 OC-3 STM-1 155.52
Optical
OC-12 STM-4 622.08
Optical
OC-48 STM-16 2488.32
SONET can be chosen as the Hierarchy Type by selecting the Hierarchy element
from the Tx or Rx panel as required.
If the unit supports Concatenated Mapping (See Section 5.3) then a test pattern can
be injected throughout the contiguous concatenated payload in either transmit or
receive mode as required by:
Note: VT1.5 and VT2 mapping will also become selectable if the AU3 mapping option
has been purchased and enabled.
The channel for testing will depend on the parameters set for the hierarchy and
synchronous payload envelope.
OC 1 2 3 4 5 6 7 8 9 10 11 12
STS-N 1 1 1 2 2 2 3 3 3 4 4 4
AU-3 channel 1 2 3 1 2 3 1 2 3 1 2 3
Only bit rates and framing types appropriate to the SONET configuration will be
displayed as options within this panel.
x1 STS-48c 2396160
STS-48c kbit/s
SPE
x1 STS-12c 599040
STS-N STS-12c
SPE kbit/s*1
The following table lists the SONET Anomalies and Defects pertinent to ANT-5, their
meanings and detection criteria.
Anomalies Detection Criteria
LOS Loss of signal All-zero pattern for
2.3 µs ≤ T ≤ 100 µs
LOF Loss of frame If SEF persists for ≥ 3ms
AIS-L Alarm Indication Signal – Line K2 (bits 6,7,8) = 111 for ≥ 5
frames
2396160
x1 kbit/s
STM-16 AUG16 AU-4-16c VC-4-16c C-4-16c
x4 599040
x1 kbit/s*1
STM-4 AUG4 AU-4-4c VC-4-4c C-4-4c
x4 139264
kbit/s
x1 x1 C-4
STM-1 AUG1 AU-4 VC-4
x3
x1
x3 TUG-3 TU-3 VC-3 44736
34368
kbit/s
x1 C-3
AU-3 VC-3
x7 x7
TUG-2
2048
Pointer x3 kbit/s
AU-4 TU-12 VC-12 C-12
processing
Multiplexing 1544
x4 kbit/s*2
Aligning TU-11 VC-11 C-11
Mapping
Concatenated Mapping
*1 option only
SDH AU-3
*2 mapping option
The VT mapping function enables the testing of DS-1 and E1 tributaries mapped into
an STS-1 SPE via VT1.5 and VT2 SPE’s as shown in the following diagram. However,
this option also requires the SONET option BN 4565/00.62.
x1 STS-48c 2396160
STS-48c kbit/s
SPE
x1 STS-12c 599040
STS-N STS-12c
SPE kbit/s*1
In the sections that follow, the user interface has been described with reference to the
ATM model, namely, the Physical layer, the ATM layer and the Adaptation layer. The
following diagram shows a basic layer model for ATM.
Application
Higher-layer protocols
AAL Layer -
adaptation of data to cell structure
ATM Layer -
multiplexing and switching of cells
Physical Layer -
adaptation of cells to the tranmission medium
Transmission medium
6.2 Introduction
The ATM option expands the applications of the ANT-5 to cover use in ATM networks.
The measurement methods and user interface have been designed for applications
involving user-network interfaces (UNI) and network-network interfaces (NNI).
Note 1: This section assumes that you are familiar with the functionality of the graphic
user interface. However, before proceeding you may wish to review section 4, The
SDH Access Tester application.
In most cases the user interface for the Tx is the same as the user interface for the
Rx. An exception to this rule is the ATM Header Rx, where a VCI and/or a CLP filter
can be enabled. See section 6.4.2 for further details.
Where the system under test is symmetrical, and the Tx and Rx structures are
required to be identical, time can be saved by setting up either the Tx or the Rx
structure, and then pressing either the or keys as appropriate. The
structure will be copied from Tx to Rx, or vice versa, as selected.
You will notice that the signal structure page has changed, similar to the diagram
shown below.
Pattern
Hierarchy
PDH Virtual
Interface
Container
From the signal structure page move the directional keys over Traffic Off panel. The
Enter key can be used to switch Traffic On or Traffic Off as required.
ATM Service
ATM Header Contract Attributes
Note: Use the tab or icon to switch to the physical signal structure layer at any
time.
Three bits are available for the Payload Type Identifier. The PTI identifies cells
belonging to different types of flow and indicates impending congestion in the network.
See section 6.12.1, PTI coding for reference.
Cell Loss Priority CLP allows cells to be classified as high priority CLP-0 or low priority
CLP-1. In the event of congestion, cells that are set to CLP-1 will be discarded first.
Note: Cells sent at a rate close to 100% of the maximum cell rate will be rounded-up.
The function and performance of other payload types must be considered separately.
For example, the payload type for Channel B will be fixed.
The Test Cell Format standard is ’97 Format. For example, this format is backwardly
compatible with Acterna ANT-20 Broadband Analyzer/ Generator module. The ’95
Format is backwardly compatible with equipment that uses legacy applications.
Once the parameters are set, press the Inject firm key. The Inject firm key will now be
depressed and the injection icon, , displays an animated syringe when defects or
anomalies are injected. To stop injecting, press the Inject firm key, the injection icon
now disappears.
The Inject firm key appears automatically in the Injection page. To allow injection
from any other page, users can view the Inject firm key by pressing the Alt button.
The Link
• Total is the number of cells
received
• Load is the number or percentage of cells that contains user information, including
OAM cells
• Idle/Unassigned is the number or percentage of cells received
• CLP =1 Count of cells where CLP=1. If congestion causes a problem, the network
will discard these cells first
• OAM is the number or percentage of cells that are used for network management
purposes.
The Specific ATM Path/Channel under test
• VPI and VCI shows the path and channel of the cell under test
• Filtered shows the number or percentage of cells which match the selected
VCI/VPI. This value is with reference to the total value.
• CLP =1. Count of cells where CLP=1. If congestion causes a problem, the network
will discard these cells first which match the VCI/VPI
• OAM is the number or percentage of cells that have been used for network
management purposes that match the VCI/VPI. This value is with reference to the
number of filtered cells.
6.8.4 ATM Quality of Service
2. Cell delay variation CDV, is the degree of variation in the CTD of the virtual
connection. By defining the QoS it is possible to offer different levels of service,
for example, by offering different guaranteed maximum cell loss rates. This
gives service providers a means for structuring the changes made for the
service, but also, the service provider must be able to demonstrate the QoS to
ITU-T O.191.
Note: PLM, UNEQ and TIM alarms are associated with G832 framing in PDH E3 or
E4 mode. If these alarms are generated the HP soft led situated at the top of the
screen will light up.
6.9.4 Graphs
zoomed feature is it allows time to be shown in different units for maximum effect.
As per ITU-T recommendation I.361, three bits are used for PT indentification.
7. Troubleshooting
When the ANT-5 is switched on, it automatically runs a self-test procedure and lights
the , and LEDs above the screen in a sequence of colours which
indicates the progress through the self-test. If the tests fail at any point, there will be a
series of beeps and the static pattern on the LEDs will remain visible to assist in fault
diagnosis. The Off key will then be the only working key available. If a self-test failure
occurs, it is unlikely that the user will be able to correct the fault or continue to use the
ANT-5 successfully. If the progress bar stalls at any point please refer to Section 8,
Customer support, and contact the nearest Acterna Sales Office or Technical
Assistance Centre.
When the ANT-5 is switched on, it automatically runs a self-test sequence which runs
from left to right across the screen and indicates the progress through the test. If the
progress bar stalls at any point please refer to Section 8, Customer support, and
contact the nearest Acterna Sales Office or Technical Assistance Centre.
Note: When the battery charge is very low, it may be possible for the ANT-5 to switch
on but fail to complete its self-test procedure before automatically switching off due to
the low battery. This is not a self-test failure and can be corrected by charging the
battery.
If the screen shows no image with the backlight on, it is possible that the LCD contrast
needs adjustment. See Section 2.6.1, Adjusting the screen contrast.
• If a PC card has been used, ensure that it has been formatted with an MS-DOS
compatible file system. If the card can be read normally by a PC with a suitable slot,
then it is possible that there is a problem with the ANT-5. See Section 3.2.6.
If the failure continues please refer to Section 8, Customer support, and contact the
nearest Acterna Sales Office or Technical Assistance Centre.
If it is necessary to send the ANT-5 for repair, the storage card should be included if
possible.
Note: Read the safety information in Section 1.8.3, Laser safety, before proceeding.
1. Deactivate the ANT-5 laser source (Tx), by clearing the Laser On checkbox in the
Interface Setup Tx panel. See Section 4.4.2 for further details.
2. Deactivate the incoming laser source connected to the Rx connector of the ANT-5.
3. Unscrew each patch lead and lift it away to break the connection.
4. Clean the ANT-5 connectors and patch lead connectors by an approved method
such as those outlined in Section 1.8.4, Cleaning optical connectors.
5. Holding the ANT-5 in a vertical position with the rear connectors facing upwards,
carefully locate and screw on the patch lead connectors making sure that each
locating lug is in the correct position and the patch lead is held straight.
6. Re-activate the laser sources after both patch leads have been reconnected.
1. Disable the timed Auto Power-off feature in the Power Saving section of the
Preferences menu.
2. Discharge the battery by leaving the ANT-5 switched on with the external power
supply disconnected. The ANT-5 will switch itself off when the battery is fully
discharged.
3. Reconnect the external power supply and check that the battery LED, , is lit.
4. Leave the ANT-5 switched off whilst the battery is charging. An adequate charge
will be reached within 3 hours and the ANT-5 will automatically switch to a trickle
charge mode when fully charged.
Note: Replacing the ANT-5 battery pack may unavoidably cause the setup information
under the System and Preferences menus to return to their default settings.
However, data can be retained if an external power supply is connected.
1. Switch off the ANT-5 and disconnect all connections and the external mains power
supply.
2. Invert the ANT-5 so that the underside is uppermost and the front bumper is closest
to you. Take care to protect the screen and keyboard if the ANT-5 is placed on a
hard surface.
3. The battery compartment is located underneath the ANT-5, just behind the front
bumper. Access is provided by two screws at the rear edge of the battery
compartment cover.
4. Remove the two screws and open the cover to reveal the battery pack. Do not
remove any other screws.
5. Carefully lift out the battery pack and disconnect the battery cable at the plastic
connector where it emerges from inside the ANT-5.
6. Plug in the replacement battery pack and place it in the battery compartment.
8. The battery pack should now be charged and conditioned as described in Section
7.6.1, Battery conditioning. This procedure will establish the initial conditions for
the battery charge indicator.
7.7 Accessing test results if the unit has been powered down.
The ANT-5 will not switch itself off if a test is running. However, if the unit is
accidentally switched off, or there is no power left in the battery, results will be saved
in the internal: device of the File Manager. These will be saved as AUTO files, with
suffixes of .A5R; the time and date will also be present so that the results are easily
identified.
8. Customer support
To assist our service engineers in responding to your enquiry, the fault which occurred
should be fully described and a note made of the circumstances at the time and any
error messages seen. The version information listed in Section 8.4 should also be
included where possible.
If possible, the original packing should be used for returning the instrument, otherwise
suitable packaging should be used to avoid damage in transit. Insurance against loss
or damage in transit is also recommended.
• The version information for the SDH Access Tester application can be displayed by
pressing the i key from within the application. Press the Enter or Esc keys
to remove the About screen and, if you do not want to remain in the application,
press Alt+Esc to switch to the application launcher.
• The options currently installed can be displayed by pressing Menu and selecting
Options from the Tools menu. This function can also be used to install and
uninstall software options, which is generally carried out at an Acterna Sales
Office, or Technical Assistance Centre.
If required, the About screens can be saved in a file or sent to a printer on the serial
port by pressing Alt+#. See Section 2.7, Printing the current screen.
Software will usually be supplied on a CD ROM, although other formats may also be
employed in the future. To load any software into the ANT-5, refer to the instructions
supplied with the software media.
9.1 General
Serial port....................................................................... 9-way, male, D-type connector
PC Card slot (BN4565/01 only) .............................................. Single, standard PC Card
Supported PC Cards (BN4565/01 only) ..............................PCMCIA/JEIDA Types I & II
5V SRAM cards with grounding/EMC clips
MS-DOS compatible file system
CompactFlash card slot (BN 4565/02 onwards)... Single, standard CompactFlash Card
Supported CompactFlash cards .................................................................. Types I & II
BN4565/01 series
Dimensions (LxWxH)................................................ approx. 258mm x 199mm x 95mm
(10.1in. x 7.8in. x 3.7in.)
Weight ...................................................................................... approx. 2.35 kg (5.18 lb)
BN 4565/02 onwards
Dimensions (LxWxH)................................................ approx. 275mm x 197mm x 76mm
(10.95in. x 7.9in. x 3in.)
Weight ........................................................................................ approx. 2.2 kg (4.85 lb)
Note: The PPS external power supply is supplied with an integral DC cable and a
separate AC mains cable. The country of use must be specified when ordering the
mains cable so that the correct plug is supplied.
9.3 Environment
Storage temperature .................................................... -20°C to +60°C (-4°F to +140°F)
Battery charging temperature ........................................ +5°C to +30°C (41°F to +86°F)
Operating temperature................................................. +5°C to +45°C (41°F to +113°F)
Condensation..................................................... Occasional short-term condensation is
allowed but performance is not guaranteed
Humidity............................................................................. Continuous operation in hot,
humid climates is not guaranteed
**ANT-5 is fully compliant with ITU-T G772, sections 2.2 and 3.2 (PMP incorporated in
transmission equipment).
Compliance with sec 2.3 and 3.3 (PMP between transmission equipment) is achieved
subject to the following conditions:
Bit errors may be encountered when √f and jitter are both at their maximum permitted
values. In addition the following precautions should be taken:
Performance at E1, E3 and DS3 extends to –26dB (linear attenuation) for full
compatibility with older types of transmission equipment.
Test patterns may be executed on any of the provided bit rates either directly at the
SDH interface or within the STMn substructure. ITU-T & Non ITU-T.
PRBS ........................................................................................ 215-1, 220-1, 223-1, 231-1
Inverted PRBS........................................................................... 215-1, 220-1, 223-1, 231-1
Programmable 16-bit digital word
Traffic (Rx only) ...............................................................................................................
The Traffic pattern should be selected if the incoming signal is expected to contain
network traffic. Selecting this option effectively disables the pattern detection process
and the reporting of LSS defects.
The following bytes within SOH and POH may be analysed and injected:
C2, D1, D2, D3, D4, D5, D6, D7, D8, D9, D10, D11, D12, E1, E2, F1, F2, F3, G1, J0,
J1, J2, K1, K2, K3, K4, M1, N1, N2, S1, V5, Z1, Z2.
The following bytes within SOH and POH may be analysed but NOT injected:
A1, A2, B1, B2, B3, H1, H2, H3, H4, Y.
9.5.8 Performance Analysis
9.6.2 Cables
Note: If an Ethernet connection is made via the PC Card slot, only 3COM 589 chip-set
PC cards are supported, model numbers 3CCE589EC and 3CCE589ET.
9.6.5 T1 testing
Bit rates using the T1 input impedance of 100Ω can be tested using the E1 BNC Tx
and Rx interfaces on the ANT-5. An optional balum and cables can be purchased from
Acterna, (BN4565/00.78). Contact the nearest Acterna Sales Office or Technical
Assistance Centre for further details.
Note: When purchasing further items to expand existing hardware and software, some
identification details may be required. Refer to Section 8.4, Version information and
serial numbers, for further information.
10.2 Safety
The instrument complies with the safety objective of:
The Council Directive dated 19th February 1973 on the approximation of the laws of
the Member States concerning electrical operating equipment for use within certain
voltage limits (Low Voltage Directive – 73/23/EEC), amended by the Council Directive
dated 22nd July 1993 (93/68/EEC).
To test the compliance of the instrument the following standard was used:
EN 61010-1 (1993/A2:1995) Safety Regulations for Electrical Measuring, Control and
Laboratory Instruments – Part 1: General Requirements.
Serial port
connections (DTE)
DCD
1
RD
2
TD
3
DTR (+5V)
4
SGND
5
DSR
6
RTS
7
CTS
8
RI
9
DCD 1 4 RTS
RD 2 2 TD
TD 3 3 RD
CTS 8 20 DTR
SGND 5 7 SGND
RTS 7 8 DCD
DCD 1 8 DCD
RD 2 3 RD
TD 3 2 TD
DTR 4 20 DTR
9 pin female 25 pin male
Connects to SGND 5 7 SGND Connects to
ANT-5 (DTE) Modem (DCE)
DSR 6 6 DSR
RTS 7 4 RTS
CTS 8 5 CTS
RI 9 22 RI
This cable has a number of internal jumper settings which must be configured by the
user prior to use. These settings must match the settings in the Serial Port section of
the System menu. The jumpers are accessible by removing the four securing screws
in the headshell; further details are supplied with the cable. K1589 has been tested
with the HP Deskjet 320 printer in the following configuration:
Baud Rate 19200 Parity NONE Stop Bits ONE
Word Length 8 BITS Flow Control SOFT
RJ48 Connector
RXT
1
RXR
2
RXG
3
TXT
4
TXR
5
6 RXG
Not used
7
Not used
8
XT=Tip, XR=Ring and XG=Ground
x4 599040
x1 kbit/s*1
STM-4 AUG4 AU-4-4c VC-4-4c C-4-4c
x4 139264
kbit/s
x1 x1 C-4
STM-1 AUG1 AU-4 VC-4
x3
x1
x3 TUG-3 TU-3 VC-3 44736
34368
kbit/s
x1 C-3
AU-3 VC-3
x7 x7
TUG-2
2048
Pointer x3 kbit/s
AU-4 TU-12 VC-12 C-12
processing
Multiplexing 1544
x4 kbit/s*2
Aligning TU-11 VC-11 C-11
Mapping
Concatenated Mapping
*1 option only
SDH AU-3
*2 mapping option
LOS/LOF
AI
S
RS-TIM J0
BIP Error B1
MS-AIS K2
AI
S
MS-BIP Error B2
MS-REI M1
MS-RDI K2
AU-AIS
AI
S
AU-LOP
HP-UNEQ C2
AI
S
HP-TIM J1
HP-BIP Error B3
HP-REI G1
HP-RDI G1
TU-AIS
AI
S
TU-LOP
LOM H4
HP-PLM C2
LP-UNEQ V5
AI
S
LP-TIM J2
LP-BIP Error V5
LP-REI V5
LP-RDI V5
AI
S
LP-PLM V5
Detection Generation
B1
Method Method Method
AIS
Method Method Method
LSS
TSE
LOF
LOS
SDH
OOF
RS-TIM
VC#
VC#
VC#
TU-13
TU-13
TU-13
TU-12
TU-12
TU-12
TUG-2
TUG-3
TUG-2
TUG-2
TUG-3
TUG-2
TUG-2
TUG-3
TUG-2
TUG-12
TUG-12
TUG-12
1 1 1 1 1 1 1 1 1 2 22 2 1 1 1 1 3 43 3 1 1
2 1 1 2 1 1 2 2 1 2 23 2 1 2 2 1 3 44 3 1 2
Regenerator Section
3 1 1 3 1 1 3 3 1 2 24 2 1 3 3 1 3 45 3 1 3
1 2 1 4 1 2 1 1 2 2 25 2 2 1 1 2 3 46 3 2 1
2 2 1 5 1 2 2 2 2 2 26 2 2 2 2 2 3 47 3 2 2
Description
Out of frame
Loss of frame
12.3 SDH Tributary Numbering
Loss of signal
3 2 1 6 1 2 3 3 2 2 27 2 2 3 3 2 3 48 3 2 3
J0
B1
1 7 1 19 1 7 1 1 7 2 40 2 7 1 1 7 3 61 3 7 1
A1, A2
A1, A2
2 7 1 20 1 7 2 2 7 2 41 2 7 2 2 7 3 62 3 7 2
Acterna ANT-5
3 7 1 21 1 7 3 3 7 2 42 2 7 3 3 7 3 63 3 7 3
Overhead Byte
Page 133
Appendix D – SDH and PDH information
Appendix D – SDH and PDH information
Multiplex Section
MS-AIS Multiplex section AIS K2
MS-RDI Multiplex section remote defect indication K2
MS-REI Multiplex section remote error indication M1
B2 (24 bits) Multiplex section error monitoring B2
Administrative Unit
AU-LOP Loss of AU pointer H1, H2
AU-NDF New data flag AU pointer
AU-AIS Administrative unit AIS AU incl. H1, H2
AU+PJE AU positive pointer justification event H1, H2
AU-PJE AU negative pointer justification event H1, H2
High Order Path
HP-UNEQ HO path unequipped C2
HP-RDI HO remote defect indication G1
HP-REI HO remote error indication G1
HP-TIM HO path trace identifier mismatch J1
HP-PLM HO path payload label mismatch C2
B3 HO path error monitoring B3
Tributary Unit
TU-LOP Loss of TU pointer V1, V2
TU-NDF New data flag TU pointer
TU-AIS TU alarm indication signal TU incl. V1 to V4
TU-LOM TU loss of multiframe H4
TU+PJE TU positive pointer justification event H1, H2
TU-PJE TU negative pointer justification event H1, H2
Low Order Path
LP-UNEQ LO path unequipped V5
LP-RDI LO path remote defect indication V5
LP-REI LO path remote error indication V5
LP-RFI LO path remote failure indication V5
LP-TIM LO path trace identifier mismatch J2
LP-PLM LO path payload label mismatch V5
BIP-2 LO path error monitoring (VC-11/VC-12) V5
B3 LO path error monitoring (VC-3) B3
E1 Anomaly Description
FAS Frame Alignment Signal
CRC-4 Cyclic Redundancy Check Error
E-Bit Remote Cyclic Redundancy Check Error
TSE Test Sequence Error
CODE Code Error
OFF (none)
G.821 TSE
G.821 definitions:
• Errored second (ES): A one-second time interval in which one or more bit errors
occurs.
• Severely errored second (SES): A one-second time interval in which the bit error
ratio exceeds 10-3.
• Unavailable second (UAS): A circuit is considered to be unavailable from the first of
at least ten consecutive SES. The circuit is available form the first of at least ten
consecutive seconds which are not SES.
• Degraded minute (DM): A one-minute time interval in which the bit error ratio
exceeds 10-6.
• Error-free second (EFS): A one-second time interval in which no bit errors occur.
The disadvantage of G.821 is that it relies on the evaluation of bit errors and so the
test channel must be taken out of service to perform the measurement.
This Recommendation takes higher bit rates into account and allows in-service
measurement as it relies on the evaluation of block errors.
G.826 definitions:
• Errored second (ES): A one-second time interval containing one or more errored
blocks.
• Errored block (EB): A block containing one or more errored bits.
• Severely errored second (SES): A one-second time interval in which more than 30%
of the blocks are errored or which contains at least one severely disturbed period
(SDP).
• Background block error (BBE): An errored block that is not a SES.
• Unavailable second (UAS): see under G.821 above.
The results are referred to the measurement time. This gives the following error
parameters: Errored seconds ratio (ESR), severely errored seconds ratio (SESR), and
background block error ratio (BBER). The specified quality requirements refer too a
particular path.
Unlike G.826, recommendation G.828 provides a precise block length for each bit
rate. For example, the number of blocks monitored per second for bit rates from VC3
upwards remains constant at 8000.
This recommendation is very close to M.2100 in terms of purpose and format, but it
deals exclusively with SDH systems. Recommendation M.2101 for SDH systems can
be considered as a sister recommendation to M.2100 for PDH systems.
V24/PPP V24/PPP
ANT-5
Host Computer
Modem Modem
Note: Before setting up the modems make sure you are familiar with its features and
operation. This section is not exhaustive and other scenarios may be appropriate for
your operating system. Please ensure the serial cable is less than 3 metres from the
ANT-5 to the modem.
1. Check the Baud Rate on the ANT-5 is 57600, by pressing Alt + Menu and
selecting Comms from the System menu.
3. Run a serial terminal program (e.g. Windows HyperTerminal) and set the
communication parameters in File > Properties to the values shown, assuming
the designated com port is selected (e.g.COM1) and connected to the modem via
the serial cable.
From the terminal program, AT commands can by typed to ensure that the modem has
the correct settings. The modem needs to be configured for hardware flow control, (e.g.
AT&F) to ignore DTR, (e.g. AT&D0) and automatically answer on the second ring (e.g.
ATS0=2).
Note: Not all modems are Hayes AT command compatible. Refer to the modem’s user
guide for the relevant command set.
Once the modem is successfully configured, remove the serial cable from the PC and
connect this to the ANT-5 without disconnecting the power to the modem or switching it
off, so ensuring that the modem indicator lights remain on.
2. Type in the phone number of the remote ANT-5, select Next and select Finish in
the next frame.
3. From the newly created connection (e.g. ANT-5 Remote Connection) right click
and select Properties.
• In the Run dialog box, type msconfig.exe and click OK. The System
Configuration Utility appears.
• In the Run dialog box, type msconfig.exe and click OK. The System
Configuration Utility appears.
• Click New.
• In the new line, type SET TZ=GMT+n or SET TZ=GMT-n, depending on the
relative difference in time zones between your PC and the BN4565/01.
Examples:
10. From Dial-Up Networking double click on the icon for the ANT-5 remote
connection and select Connect.
11. The modem will then dial out to the remotely located ANT-5 modem and establish
a connection.
Note: A password is not mandatory since the password facility of Windows Dial-up
Networking is not used by ANT-5.
12. Refer to section 5.4.1, Software Installation and operation on using the GUI
faceplate.
1. From Start > Programs >Accessories > Dial-Up Networking click on New. Type
in a name for the connection and the phone number of the remote ANT-5. (It is
advisable to uncheck Use another port if busy).
3. Click on TCP/IP Settings… and again emulate the following settings. Click on OK
to proceed.
6. Click on the X.25 tab to check the default settings, click on OK to finish.
11. The modem will then dial out to the remotely located ANT-5 modem and establish
a connection.
Note: A password is not mandatory since the password facility of Windows Dial-up
Networking is not used by ANT-5.
13. Refer to section 5.4.1, Software Installation and operation on using the GUI
faceplate.
1. From the Start menu, select Settings > Network and Dial-up Connections >
Make New Connection.
4. Configure the phone number to dial as shown below, then click Next. (The modem
may have to be configured to access an external line via the telephone network).
5. If prompted, decide who is going to use this connection, then click on Next.
6. Name the connection, for example ANT-5 Remote Operation Unit then click
Finish.
7. You will now have a new dial-up network connection icon in the dial-up networking
window, called ANT-5 Remote Operation Unit.
8. Right click on the ANT-5 Remote Operation Unit, and select Properties from the
menu. The following dialog box appears:
10. Select the Security tab and emulate the following settings.
11. Select the Networking tab. Select PPP as the type of dial-up server. The list of
components used by the connection may differ for your Windows 2000
configuration. Ensure that the Internet Protocol (TCP/IP) box is checked. The
greyed options (Novell Distributed Print Services and Workstation Manager) below
are disabled by Windows 2000. (You may have some components like this, as this
is normal. Ensure that all other components are unchecked).
12. Click the Internet Protocol (TCP/IP) component, and then click Properties. The
following dialog box appears.
13. Click Advanced…, and from the General Tab, uncheck Use default gateway on
remote network and the click OK.
14. Click OK, close the Internet Protocol (TCP/IP) Settings dialog box.
15. If privileges allow select the Sharing tab, and emulate the following settings. Click
OK to close the ANT-5 Remote Operation Unit Properties dialog.
18. A connection to an external power supply is advisable. Reboot the ANT-5, launch
the REMOTE application and leave on.
19. From Network and Dial-up Connections double click on the icon for the ANT-5
remote connection. The modem will then dial out to the remotely located ANT-5
modem and establish a connection.
Refer to section 5.4.1, Software Installation and operation on using the GUI
faceplate.
Ethernet Ethernet
ANT-5
Host computer
Note: The following illustrates how remote access can be achieved using the network,
it is not exhaustive and other scenarios may be equally valid. Please ensure the cable
is less than 3 metres from the ANT-5 to the RJ45 socket.
2. In the Systems > Comms > Ethernet menu of the ANT-5, edit the IP Address,
Subnet Mask and Gateway to those specified by the IT department and the
Transceiver to the network connection type.
Note: As an example the connection to the network can be made via an Ethernet card
plugged into the PC Card slot. A static IP address will be required for the particular
MAC address of the PC card. Refer to section 2.8.1 for information on
communications configuration.
Refer to section 5.4.1, Software Installation and operation on using the GUI
faceplate.
F1 F2 F3 F4 F5 F6
Function keys
F9
Information key
C-n Container; n = 1, 2, 3, 4
DM Degraded Minutes
HP Higher-order Path
K1, K2 APS channels for APS sig. & back-up line switching (MSOH)
K3, K4 APS channels for APS sig. & back-up line switching (POH)
LO Lower Order
LP Lower-order Path
MI Management Information
MO Managed Object
MS Multiplexer Section
ms Milliseconds
NE Network Element
OH Overhead
OOF Out-Of-Frame
OOS Out-Of-Service
UI Unit Interval
UNEQ Unequipped
µs Microseconds
VP Virtual Path
VT Virtual Tributary
12410 Milestone Center Dr. North America Latin America Western Europe Acterna, LLC. Note: Specifications,
Germantown, Maryland 12410 Milestone Center Dr. Av. Eng. Luis Carlos Berrini Arbachtalstrasse 6 All rights reserved. terms and conditions
20876-7100 Germantown, Maryland 936/8° e 9° andares 72800 Eningen u.A. are subject to change
USA 20876-7100 04571-000 São Paulo Germany Acterna, The Keepers of without notice.
USA SP-Brazil Tel: +49 7121 86 2222 Communications, and its
Acterna is present in more Toll Free: 1 866 ACTERNA Tel: +55 11 5503 3800 Fax:+49 7121 86 1222 logo are trademarks of
than 80 countries. To find Toll Free: 1 866 228 3762 Fax:+55 11 5505 1598 Acterna, LLC. All other
your local sales office go Tel: +1 301 353 1560x2850 Eastern Europe, trademarks and
to: www.acterna.com Fax:+1 301 353 9216 Asia Pacific Middle East & Africa registered trademarks
42 Clarendon Street Elisabethstrasse 36 are the property of their
PO Box 141 2500 Baden respective owners. Major
South Melbourne Austria Acterna operations sites
Victoria 3205 Tel: +43 2252 85 521 0 are ISO 9001 registered.
Australia Fax:+43 2252 80 727
Tel: +61 3 9690 6700
Fax:+61 3 9690 6750 1st Neopalimovskiy Per.
15/7 (4th floor)
RF 119121 Moscow
Russia
Tel: +7 095 248 2508
Fax:+7 095 248 4189
Doc. 4565-8498.210
Rev. 12, 07/2003
English