ISOTECH
ISOTECH
1064.180ºC
GOLD FREEZES
Silver Freeze
Point Cells 961.7800ºC
High Temperature Standard Platinum Resistance Thermometer 962 & 96178
Aluminium Freeze
Point Cells 660.3230ºC
ITL 17702B or ITL 17706
Zinc Freeze
Point Cells 419.5270ºC
Tin Freeze
ITL 17707 or 17702A
231.9280ºC
Standard Platinum Resistance Thermometer 909 & 670
Point Cells
465
Gold / Platinum Thermocouple
-259.3467ºC
TRIPLE POINT OF HYDROGEN
-273.1500ºC
Isotech ITS-90 ABSOLUTE ZERO
ITS-90
Metrology Temperature Acceptable
Furnaces Scale Thermometer Ranges
For full details, contact Isothermal Technology Limited, Pine Grove, Southport, Merseyside PR9 9AG England
Telephone +44 (0)1704 543830 Fax +44 (0)1704 544799 Email [email protected] Web www.isotech.co.uk
Contents
Metrology Instruments..........................................Inside Front TTI-22 True Temperature Indicator........................ 48 - 49
Introduction................................................................... 4 - 5 milliK Precision Thermometer............................... 50 - 52
millisKanner ................................................................ 53
Primary ITS-90 Standards
Model SRA & SRB AC/DC Standard Resistors............... 54
The Jarrett-Isotech Water Triple Point Cells................ 6 - 8
Model 455 Resistor Maintenance Bath.......................... 55
Maintenance Bath - Water Triple Point............................. 9
Model 456 Temperature Controlled AC/DC Resistors..... 56
Ice Mantle Maker - Water Triple Point............................ 10
Gallium Cell - Sostmann-Isotech........................... 11 - 12 The Best Comparison Calibration Equipment............ 57 - 58
Gallium Apparatus........................................................ 13 Nitrogen Boiling Point Apparatus................................... 59
The Isotech Mercury Cell.............................................. 14 Model 461 Simple Liquid N2 Apparatus......................... 60
Mercury Triple Point Apparatus..................................... 15 Model 459 Cryostat...................................................... 61
Isis -100°C Dry Block........................................... 62 - 63
The Best Primary Standards for your Laboratory.............. 16 915 Parallel Tube Stirred Liquid Bath..................... 64 - 65
UKAS Calibration Service for SPRTs.............................. 17 785 Parallel Tube Stirred Liquid Bath..................... 66 - 68
Fixed Point Cells - Sealed and Open....................... 18 -19 Orion 796 Stirred Liquid Baths.............................. 69 - 71
Hydra 798 Stirred Liquid Baths............................. 72 - 74
International Traceability.................................................. 20
813 Stirred Ice / Water Bath.......................................... 75
Furnace Selction Guide....................................................... 21
Model 820 Large Volume Calibration Bath..................... 76
Dual and Heat Pipe Furnaces................................ 22 - 23
Liquid Selection Guide........................................................ 77
Low Temperature Furnace............................................. 24
Model 875 Fluidized Calibration Bath..................... 78 - 79
Medium Temperature Furnace....................................... 25
Model 877 Thermocouple Calibration Furnace....... 80 - 81
3 Zone High Temperature Furnace................................. 26
Annealing Furnace........................................................ 27 Points on the Temperature Scale (POTTS) Introduction.... 82
ISOTower............................................................. 84 - 87
Introduction to SPRTs and Standard Thermocouples............ 28
Slim Fixed Point Cells - Sealed.............................. 88 - 89
670 Family Primary SPRT..................................... 29 - 31
Europa-6.............................................................. 90 - 91
Model 96178 Silver Point SPRT............................ 32 - 33
Oceanus-6........................................................... 92 - 93
909 Family Working SPRTs................................... 34 - 35
Medusa 510 and Medusa 511.............................. 94 - 95
Introduction to Standard Thermocouples . .................... 36
Oberon......................................................................... 96
Platinum / Palladium Thermocouple.............................. 37
POTTS Ease of Use............................................................ 97
Platinum / Platinum Rhodium Thermocouple................. 38
Platinum / Gold Thermocouple...................................... 39 Introduction to Temperature Calibration Software............ 98
Icarus Calibration Software........................................... 99
Guide to Precision Thermometers and accessories.... 40 - 41
Cal NotePad Calibration Software.................................100
microK Precision Thermometer............................. 42 - 44
I-Cal Easy Calibration Software......................... 101 - 102
microsKanner Channel Expander................................... 45
Calibration Services......................................................... 103
Resistance Bridge Calibrator................................. 46 - 47
The company is always willing to give technical advice and assistance where appropriate.
Equally because of the program of continual development and improvement, we reserve the right to amend or
alter characteristics and design without prior notice.
...our knowledge, product quality and integrity are the reasons for buying from Isotech
In our catalogues, datasheets and Selecting the best equipment Equipment Required
website you will not find unsubstantiated for a Primary Temperature Can be grouped under four headings:
claims by our own scientists (although
they are as good as the best), you will
Laboratory 1. An accurate electrical measuring
find the results of independent system.
The Purpose 2. A number of ITS-90 fixed point cells.
international intercomparisons done by
such organisations as BIPM and NIST. The purpose of a primary laboratory 3. Apparatus to melt, freeze and
realising part of ITS-90 is to calibrate maintain the cells.
Where we have to summarise we SPRTs at fixed temperatures, by 4. Stable thermometers to monitor the
provide references to the appropriate comparing their resistance to a fixed cells.
document for you to download and and known resistance.
read in full. When selecting the equipment for the
Uncertainty Limits laboratory look at performance,
Our uncertainties of calibration and uncertainties and price.
procedures have all been scrutinised For fixed resistors the limit is about
and agreed with UKAS, who are party ±0.05ppm.
to the ILAC-MRA. For 6N purity metallic cells the limits
The Electrical Measuring
are (ref CCT/2000-13 p.5) System
Each year we are audited by UKAS
who check our electrical and thermal Hg ±0.2mK
standards, examine and witness our
The Bridges
Ga ±0.1mK
procedures and question our A ratio bridge that compares a
In ±0.5mK temperature controlled stable fixed
uncertainty spread sheets.
Sn ±0.3mK resistor to the SPRT under test is ideal.
This ensures a continuous improvement Zn ±0.5mK
in our services to you, our customer. Bridge linearity, noise and stability are
Al ±0.7mK
the main things to consider. The
Our electrical standards are certified by Ag ±1.1mK uncertainty should be of the order of
NPL, the UK’s National Laboratory. 0.1ppm.
For SPRTs ±0.5mK (non-uniqueness of
Our fixed points from argon through to The Isotech microK 70 is the ideal
ITS-90 between fixed point of water
silver have been intercompared at NIST solution, it contributes just 0.07ppm to
and gallium), ±0.5mK (2 oxide state).
in the US to their references. the measurement uncertainties with
When selecting equipment the noise and linearity and uniquely it is
Our UKAS certificates are presented in
uncertainties should match the above. drift-free. It has input 3 channels
a form which makes them instantly
This is because uncertainties are extendable to over 90 with
useful. Our staff of experts are
combined using root sum of squares.
available to help with your enquiries
So if an uncertainty is much larger
and problems.
than the above it will dominate
the rest, if it is very
small it becomes
insignificant.
http://www.isotech.co.uk
microsKanners and with no mechanical g Open quartz g Open metal clad Procedures and
switches, relays, potentionmeters or g Sealed quartz Sealed metal clad
g
Uncertainties
moving parts is the most reliable
You will find full details on the following
product available. One of the great things about thermal
pages.
metrology is that there is no approved
Fixed Resistors See Isotech’s technical library on the way to calibrate. One of our staff was
website for a comparison of our UKAS asked by a UKAS Auditor how long it
The best resistors are made for us by
results of quartz clad and metal clad took to calibrate an SPRT... “Oh!” she
Tinsley and we can UKAS certify them
cells compared to CCT/2000-13. said without hesitation “until its right!”.
to 0.1ppm or better.
Silver, gold and copper require special The auditor was so impressed he asked
Even fixed resistors have a tempco and if he could use the saying elsewhere.
handling and with many years of
so we offer a maintenance bath.
experience we can help you with the Your calibration procedures need to be
The maintenance bath needs to be temperature points. unique to you to suit your equipment,
monitored using an SPRT. It can be your staff and your customers.
Isotech can UKAS certify any of these
connected to the one channel of the
cells to the smallest uncertainties Likewise uncertainties, you will have a
microK to save cost.
outside NIST. unique combination of equipment and
You now have a world-class measuring procedures. These will dictate your
system. To maintain it the resistors will
Selecting Furnaces uncertainty.
need regular recalibration (every year
or two years). Although we cannot write your
For indium through copper we have a
procedures and calculate your
The SPRT can be recalibrated in your choice of furnaces. The Isotech Dual
uncertainties, we want to help, and so if
own fixed point cells at water and Furnaces use heat pipes for best
you go to our website you will find
gallium. performance, and include a second
some examples that may help you
independent furnace for annealing. For
The microK can be rechecked by you develop your own budgets.
those with existing annealing furnaces
using compliment checks with your we offer Heat Pipe only models. Three
fixed resistors. Zone furnaces can be used over very More Information
wide ranges. Our Furnace Selection
Fixed Point Cells Guide will help you select the A list of our recommended books is
appropriate models for your needs. available on the Isotech website and
ITS-90 gives 7 optional ranges, check
the Isotech journal of thermometry.
the graph on the inside front cover and
select the range for your laboratory. Monitoring Thermometers We have an e-learning course and we
run occasional 2 or 3 day courses here
Every primary laboratory needs water
Ideally each fixed point cell should in Southport.
triple point and the gallium melt point,
have its own designated SPRT. If this is
because to qualify as an World-wide, we have a network of 90
unaffordable one SPRT can be used
ITS-90 SPRT a agencies to help you locally. Contact
with a number of fixed point cells.
thermometer must us for your nearest.
have a W Gallium These SPRTs need to be very stable.
equal to or greater We recommend model
then 1.11807. 670SQ/25.5/480 for use with Hg, H20,
Ga, In and Sn.
Mercury triple point
is the next most useful 670SQ/25.5/650 for any cell from Hg to
temperature. and including Al.
and 96178/0.25 for Zn, Al and Ag.
Isotech is unique in
offering the other The above monitor thermometers can
commonly used fixed be supplied with UKAS certificates or
points of Indium, Tin, Zinc you can certify them in your own fixed
and Aluminium in choice point cells.
of 4 housings; In the primary lab, only the resistors and
the cells require external traceability.
http://www.isotech.co.uk
g Uncertainty 0.000070°C
g Fifty Years of Proven Use
g Quartz Glass and Isotopic Analysis Available
Quality
Isotopic Content * Technical annex for the International Temperature Scale of 1990 (ITS-90)
Adopted by the CCT on 10 June 2005
Jarrett-Isotech Cells use water with an isotopic content
Since early in 2000 Jarrett-Isotech Cells have been made
essentially similar to Standard Mean Ocean Water.
to this definition and our cells have been within +10 and
Following research by the international science
-40µK of it.
community into the make up of Standard Mean Ocean
Water, and V-SMOW the BIPM recommended in 2005* During 2005, subsequent to the CCT definition, the water
that in our triple point cells has been further enriched and our
latest cells meet ±20µK of the above definition.
“The triple point of water is now defined as the
equilibrium temperature of vapour, liquid and solid water, Further we can provide Isotopic Analysis giving the exact
with the liquid water having the isotopic composition composition with an uncertainty of ±3µK. Samples of the
defined by the following amount of substance ratios: actual water used in a particular cell can be supplied for
purity analysis.
Performance
Type A cells were designed by Dr. H. F. Stimson at Type B cells were designed at NRC of Canada.
NBS. A tubular glass extension at the top of the cell The thermometer well extends 100mm above the
serves as a convenient handle for lifting and carrying top of the cell. Heat transfer to the ice mantle may
the cell, as a hook for supporting it in an ice bath, be essentially eliminated by keeping these cells
and as an indicator of partial pressure of air in the packed in ice to the top of the well extension, or by
cell. immersing them sufficiently in a Water Triple Point
Maintenance Bath.
95 95
E E E E
14 14 70 70
A A A A
C CD D C CD D
TYPE A TYPE B B B
Nominal Dimensions in mm
B B
Model A B C D E Comments
A11-50-270 11 50 350 270 100 Highly recommended (1) (2) (1) Isotopic Analysis
is available.
A13-50-270 13 50 350 270 100 Large re-entrant tube (1)
B8-30-130 8 30 160 130 0 Slim Cell - Fits Isotech Dry Blocks (2) Available in
Quartz Glass.
B12-40-210 12 40 290 210 75 Replacement for NPL, UK Type 32 Cells
B12-46-210 12 46 290 210 75 Recommended for Isotech Oceanus
B11-50-270 11 50 350 270 100 Highly recommended (1) (2)
B11-65-270 11 65 350 270 100 NRCC’s favourite Cell (1) (2)
B13-65-270 13 65 350 270 100 Large re-entrant tube (1)
B16-65-270 16 65 350 270 100 Larger re-entrant tube (1)
It will maintain the water triple point for long periods of time
(many months, with occasional attention) so that this
fundamental fixed point of thermometry is continuously
available.
The Bath ensures total safety of the cells with dual electrical
safety circuits and the passive safety of the design, see
below.
Are you fed up with cold wet hands, and hours of frustration
when you produce an ice mantle in your Triple Point Cell?
Change your life and try the Isotech Ice Mantle Maker.
Background
Typical time 20 to 30 minutes using a Jarrett-Isotech 452 Ice Mantle Maker and one ice bridge prevention
to create a A11 Water Triple Point Cell.* collar
mantle
Please specify
* Longer or shorter times for larger or smaller cells. A) Cell Type or
B) Outside diameter of Water Triple Point Cell (mm)
Options C) Depth from shoulder of Cell to water level (mm)
Additional Ice Bridge Prevention Collar
g Uncertainty 0.000070°C
g Thirty Years of Proven Use
g Open and Sealed Models
Total Confidence
Quality
Melt 1 Day1
0.000025K 1hr
0.25mK
1.1810343
1.1810342
Melt 2 Day2
0.000025K 1hr
0.25mK
1.1810343
1.1810342 1.1810342
0.000025K 1hr
0.25mK
1 2 3 4
1.1810343
1.1810342
Intercomparison
of Water and Gallium Water Gallium
g Simple to Use
g Automatic Operation
g Totally Safe Operation
g Uncertainty 0.000220°C
g Rugged Stainless Steel Construction
g 7N Pure
Total Confidence
Operation
How to order
ITL-M-17724 Mercury Cell
The Cell is supplied with a Certifcate of Metal Purity
*We have a choice of UKAS calibration Services, the
stated uncertainty figure is for our Premium UKAS
Calibration Service
Uncertainty 0.22mK
(with cell)
How to order
ITL-M-17725 Mercury Triple Point Apparatus
Suitable only for Isotech 670SQ Models or other Primary Standard SPRTS of similar stability
Typical Uncertainties ±
Suitable for Primary and Working SPRTS - Isotech 670 & 909 families and other SPRTS of similar stability
FIXED POINT
Fixed Point °C Range Range Range Range Range Range
1 2 3 4 5 61
Typical Uncertainties ±
Uncertainties
Optimal Cells include a conformity certificate which includes a reference cell. This takes a minimum of 15 days of
a copy of the impurities analysis, a copy of the metal of laboratory time.
the cell evaluation freeze and melt curves. Where required
With our Standard Comparison service we perform one melt,
we can also provide UKAS calibration.
one freeze and one intercomparison, the time to calibrate
The uncertainty mentioned in the table is that which can be is less than the Premium Service and so the cost is lower.
offered with our optional UKAS Calibration service. Our The uncertainties are still small, and suitable for all but the
Premium Calibration service involves realizing three melt most demanding of Primary Laboratories.
plateau, three freeze plateau and two intercomparisons to
Cell Premium Calibration Service Standard Calibration Service The latest schedule
can be found on the
UKAS Schedule Note 4 UKAS Schedule Note 5 Isotech website
or at www.ukas.org.
Mercury +/-0.22mK +/-1mK
Gallium +/-0.07mK +/-1mK
Indium +/-0.65mK +/-2mK
Tin +/-0.60mK +/-2mK
Zinc +/-0.90mK +/-2mK
Aluminium +/-1.1mK +/-6mK 0175
Silver +/-2mK +/-30mK
Isotech cells are of the highest purity available. Open cells conform to CCT/2000-13. Sealed cells are sealed to one
atmosphere with 6N pure argon at the freeze temperature.
Cell baskets, complete with appropriate heat shunts and reflectors are available seperately.
A carry case is included with Sealed Cells.
Nominal Dimensions
Outside Dia. 50mm Outside Dia. 50mm Outside Dia. 50mm Outside Dia. 50mm
Inside Dia. 8mm Inside Dia. 8mm Inside Dia. 8mm Inside Dia. 8mm
Height* 270mm Height* 275mm Height* 270mm Height*
Metal Depth 200mm Metal Depth 200mm (+350mm Tube) In, Sn, Zn - 520mm
Metal Depth 200mm Al, Ag, Cu - 610mm
Metal Depth 200mm
*The height is measured from the bottom of the flange wherever necessary, excludes sealing tip.
Isotech has always invested heavily ILAC was formed to harmonize Under these MRA’s UKAS calibration
in, and attached great importance to technical issues between certificates are recognised and
its calibration laboratory which was accreditation bodies worldwide with accepted in over 70 countries
first accredited in 1985. the goal of breaking down
For more information on ILAC visit
international barriers, resulting in
Today the laboratory, known as NTPL www.ilac.org
increased global trade.
is formally accredited by UKAS to
ISO / IEC 17025:2005. Since 2008 This was implemented through a
NTPL UKAS Certificates have been network of Mutual Recognition
licensed to bear the ILAC MRA mark. Arrangements (MRA) amongst
accreditation bodies throughout
ILAC International Laboratory
the world.
Accreditation Cooperation
N
OF CALIBRATIO
CERTIFICATE
GY LIMITED.
THER MAL TECHNOLO
ISSUED BY: ISO
: 2nd February 2009
DATE OF ISSUE
: 09-01-123
CERTIFICATE NO
Page 1of 6
Approved Signatory
Name:
Signature:
.
olumbus Calibration
CUSTOMER: C.C 2810034
ITL REFERENCE:
R: 1492
CUSTOMER ORDE
Cell.
Gallium Fixed Point
tech Model 17401
DESCRIPTION: Iso
N: Ga 454
IDENTIFICATIO
2009 to 26
th
January 2009
From 21 January
st
RATION:
DATE OF CALIB
ference
the Laboratory Re
l has bee n inter- compared to one of
RATION: The cel
BASIS OF CALIB ture Scale of 1990
ernational Tempera
cells.
le in use in this Laboratory is the Int
The temperature sca
(ITS 90). ndard
ed International Sta
e with the recognis
orator y is acc redited in accordanc str ates techn ica l competence for a
This lab n demon nt
5. This accreditatio ent system (refer joi
ISO/IEC 17025:200 y quality managem
pe and the op eration of a laborator
defined sco uary 2009 ).
mmuniqué dated Jan
ISO-ILAC-IAF Co g on the publication
s and
on the ILA C we bsite at www.ilac.or
ique is available
The Joint Commun
resources page.
Technology Limited
ory of Isothermal
Calibration Laborat
Laboratory is the ice. It provides
perature Primary m Accreditation Serv ory or other
The Northern Tem the United Kingdo orat
ry accr editation requirements of at the Nati onal Physical Lab issuing
ce with the laborato s and to units of measurement real
ized ten approval of the
issued in accordan pt with the prior writ
This certificate is onal standard r than in full, exce
of mea sure men t to recognized nati ificate may not be reproduced othe
traceability ries. This cert
standards laborato
recognized national
laboratory.
g Dual Furnaces - g Three Zone Furnaces NIST in the US like to work with long
the no compromise choice plateaus whereas according to our
All heatpipes have a limited operating
UKAS procedure we should calibrate
range, determined by fluid that flows
These furnaces use heatpipes to an SPRT 2 or 3 times using a new
inside the pipe. Furnaces without
provide an essentially gradient free plateau each time.
heatpipes can work over wider
environment to melt and freeze the temperature ranges. Isotech offer two Our apparatus has sufficient
ITS-90 fixed points. These furnaces models of Three Zone Furnaces, one performance that the length of the
meet all the requirement of from 50°C to 700°C and one from plateau is dictated mainly by how close
CCT/2000-13 and allow a uniformity of 200°C to 1200°C. These furnaces use the set point of the apparatus is to the
<10mK over the entire length of the top and bottom guard heaters to fixed point being realized.
fixed-point sample. minimise temperature gradients and
Plateau lengths at the silver point of
also meet the requirement of “Optimal
The second independent furnace is over 70 hours (3 days) have been
Realizations”.
used to pre warm and anneal the achieved using our furnaces. From a
thermometers being calibrated. This practical point we normally work with
concept of heatpipe and second g Single Zone Furnaces one working day long plateaus,
furnace for pre and post conditioning remelting the cell overnight ready for a
Finally the range includes an
the thermometers in a single apparatus new freeze the next day.
economical single zone furnace for
was developed from a concept of Dr Indium, Tin and Zinc Cells and an
Marcarino of IMGC, Italy. Annealing Furnace for pre and post g High Temperature Furnaces
conditioning thermometers.
To eliminate problems with noise and
g Heat Pipe Furnaces electrical safety from high earth
For those laboratories who already have g Plateau Lengths leakage currents our high temperature
furnaces for pre and post conditioning furnaces operate from 110VAC. For
CCT/2000-13 says that a plateau
SPRTs we offer the range of furnaces in operation from 230VAC the furnace can
length of 10 or more hours is suitable
single, heatpipe only version. be provided with an internally mounted
for optimal realizations.
isolating transformer.
Dual Furnaces
In addition to heat pipes described above the Dual
Furnaces incorporate a second furnace which, because
of its unique design, will safely (and without
contamination) pre- and post-condition the
thermometers.
To complete the apparatus, a further pre-warming tube
(with a temperature approximately equal to that of the
heat-pipe) made of a unique and gas-tight material, is
provided, together with a storage rack for 4
thermometers.
Equalizing Block
Accessories
Accessories include equalizing blocks, a fan assembly to
keep thermometer handles cool, a thermometer holder
and internal and external isolating transformers.
Cooling Coil
The Potassium and Sodium models have a cooling coil in
the lid with connections to circulate tap water to keep the
furnace lid cool protecting the SPRT and reducing heat
load into the lab.
The Isotech Medium Temperature Furnace is designed
specifically to realize and maintain the freeze plateaux of
Isotech Indium, Tin, Zinc and Aluminium Cells, for the
calibration of thermometers on ITS-90. It can also be
used with an insert, as an annealing facility of the highest
order for SPRTs or as a comparison calibration facility.
Equalizing Block
Thermocouples
In practise Standard Platinum Resistance Thermometers, Model 1600 Platinum / Platinum Rhodium
SPRTs, are constructed to cover sub ranges of the ITS-90 Available as Type R or Type S these thermocouples are
and SPRTs are available in different construction types. housed in a 99.7% recrystallized alumina sheath, 300 or
600 mm long and can be used to 1600°C
Isotech offer the 670 family as SPRTs recommended for
Primary Applications and the 909 Family for Secondary Platinum / Gold Thermocouple
Laboratories. This model offers smaller uncertainties than Type R or S
These families span from -200°C to 670°C, for higher using only pure metals in the construction. An economical
temperatures, up to 961.78°C the freezing point of Silver alternative to HTSPRTs.
Isotech offer the 96178 HTSPRT.
NPL Platinum / Palladium Thermocouple
Standard Thermocouples This model manufactured by the National Physical Laboratory
Whilst no longer a part of the temperature scale (NPL) was developed to operate reliably and accurately to
1500°C and offers superior stability to conventional platinum /
thermocouples are widely used in calibration laboratories.
platinum rhodium thermocouples. They now can be
Isotech can supply Standard Thermocouples to 1600°C,
purchased from Isotech.
either in platinum / platinum rhodium or platinum /gold
materials.
650.00 90.00
38.00
7.50
25.00
150.00 2.5 METRES OF CABLE
670SQ -200 to 670 25.5 Quartz 7.5mm 650mm 35mm Recommended for
or wide range use in the
480mm Primary Laboratory
670SQ -200 to 550 100 Quartz 7.5mm 650mm 35mm 100 Ohm version
or
480mm
670SH -80 to 670 25.5 Metal 6mm 650mm 35mm Metal sheathed high
or temperature model
480mm
670SH -80 to 550 100 Metal 6mm 650mm 35mm 100 Ohm version
or
480mm
670SL -200 to 165 25.5 Metal 6mm 480mm 35mm Metal sheathed low
temperature model
Suitable only for Quartz Sheath SPRTs such as the Isotech 670SQ Models or other Primary Standard SPRTS of similar
stability
Typical Uncertainties ±
Note: The 100Ω 670 has a maximum temperature of 550°C and so cannot be UKAS certified over Range 4.
Please contact Isotech if calibration is required above Zinc.
Table shows Premium Calibration Service, Standard Service is also available, see page 17.
Note1: It is necessary to make the former on which the Note2: Gases, like Iron, Chromium, Nickel under
platinum is wound of high-purity quartz. Even reducing conditions, can penetrate the quartz
quartz does not provide absolute isolation at the sheath and poison the platinum.
high temperature end of the range. The former,
It is necessary to purchase not only the 96178,
or mandrel, is thus a shunt resistance across the
but items such as the Dual Furnace to ensure
platinum winding, and because of the uncertainty
that your high temperature thermometer does
of the contacts between platinum and quartz, it is
not become contaminated.
uncertain and unstable in magnitude. The
practical solution is to reduce the element Only Isotech offers a comprehensive solution to
resistance so that the shunt resistance produces the measurement and use High Temperature
a smaller network effect. For example, for a Thermometry.
25.5Ω thermometer, suppose that the shunt
resistance were 20 MΩ. Then the network
resistance is 25.499967Ω. But we require Note3: Our know-how and expertise in the field of High
measurement assurance of better than 1 part per Temperature Thermometry has been written
million, so this won’t do, even if the shunt were a down and is available in the Isotech Journal of
constant (calibratable) value, which it is not. For Thermometry.
a 0.25Ω thermometer, a 20 MΩ shunt gives a
network resistance of 0.24999997Ω, which is
tolerable. The cost, and there is a cost, is
increased difficulty on the electrical
measurement side, particularly in the face of
noise, which is present at high temperatures.
650.00 90.00
40.00
7.50
25.00
150.00 2.5 METRES OF CABLE
The 909Q has a quartz sheath while the 909L and 909H
have metal sheaths. Whilst metal sheathed thermometers
appear more robust than the quartz glass models it
should be noted that ALL SPRTs are fragile devices and
must be handled with care.
Model 909
Three thermometer lengths are available, standard length
480mm, extra length 550mm or maximum length 600mm. RTPW 25.5Ω 100Ω
Quartz glass thermometers have the advantage that the Nominal Resistance 25.5Ω at 0°C 100Ω
internal components are visible and can be inspected Recommended
and continue to be our recommended models. The low Max. Current mA 1 0.5
temperature models have excellent immersion Nominal Sensitivity 0.1 Ω/°C 0.4 Ω/°C
characteristics and a significant cost saving when
compared to the higher temperature models. Resistance Ratio Wga>1.11807 as required by ITS-90
Self-heating 1mK / 25 microwatts
The Model 909 is supplied with a calibration certificate
giving RTPW and Wga. Alternatively we can provide a Stability Depends upon the temperature
complete UKAS calibration certificate, see table opposite. range of use. Typical annual stability,
For transportation and storage the Model 909 is supplied see the table on the next page.
in its own attractive carrying case. Internal leads 4 wire-platinum
External Leads Silver-plated multi-strand wires in a
low-loss insulation cable terminating
in gold-plated terminals.
How to Order
Model 909 Specify model, resistance and length.
State with UKAS Calibration or without UKAS Calibration.
at Fixed Points or by Comparison.
0175
Note: The 100Ω 909 has a maximum temperature of 550°C and so cannot be UKAS certified over Range 4.
Please contact Isotech if calibration is required above Zinc.
909Q -200 >1.11807 Quartz Sealed with 7.5mm 480mm 65mm Isotech
25.5Ω to dry oxygen 550mm recommended
670 / argon mix 600mm secondary
standard SPRT
909Q -200 >1.11807 Quartz Sealed with 7.5mm 480mm 65mm 100 Ohm
100Ω to dry oxygen 550mm secondary
550 / argon mix 600mm standard SPRT
909H -80 >1.11807 Metal Sealed 6mm 480mm 65mm Internal alumina
25.5Ω to 550mm tube protects
670 600mm sensor from
contamination
909H -80 >1.11807 Metal Sealed 6mm 480mm 65mm Internal alumina
100Ω to 550mm tube protects
550 600mm sensor from
contamination
909L -200 >1.11807 Metal Sealed with 6mm 480mm 65mm Optimised for low
25.5Ω to dry oxygen 550mm temperatures, less
165 / argon mix 600mm stem conduction
due to internal
construction
909L -200 >1.11807 Metal Sealed with 6mm 480mm 65mm Optimised for low
100Ω to dry oxygen 550mm temperatures, less
165 / argon mix 600mm stem conduction
due to internal
construction
n Calibration Options
Comparison Calibration
The Model 1600 includes UKAS calibration to 1100°C,
with options to extend this
http://www.isotech.co.uk
to 1200°C or 1300°C. The
thermocouples are calibrated
using comparison techniques
in the Isotech 877 Furnace.
Uncertainties for models with a
cold junction (recommended)
are ±0.7°C to 1100°C.
In addition to developing the novel structural design Applicable Standard IEC 62460 Edition 1.0
along with new annealing and construction procedures
NPL have established the world’s first ISO 17025 (UKAS)
accredited calibration services using metal-carbon How To Order
eutectic fixed point cells. Model Type: NPL Pt/Pd Thermocouple
We are pleased to be able to offer the NPL Pt/Pd Includes UKAS Calibration, specify range from either:
thermocouple with calibration alongside our own Pt/Pt-Rh
and Au/Pt Standards. 0°C to 1100°C Uncertainty of ±0.2°C
Calibrated at Fixed Points Zn, Ag and Cu
The covered noble metal thermocouple wire connects the Also available without the physical cold junction - Specify No Cold Junction (NCJ).
Thermocouple characteristics are sufficiently smooth to The standard thermocouple described can be supplied
allow interpolation of deviations from the reference table in the following noble metal combinations
to be carried out over fairly wide temperature spans TYPE R: Platinum vs Platinum 13% Rhodium
without introducing unacceptable errors. Isotech can TYPE S: Platinum vs Platinum 10% Rhodium
offer a 4-point UKAS calibration for temperatures up to How to order
1100°C (supplied as standard), a 6 point UKAS calibration Model 1600 Type R/300
up to 1300°C with the option of a table of millivolts to Model 1600 Type R/600
degrees Celsius in 10°C steps or, alternatively, arrange Model 1600 Type S/300
for an NPL calibration at fixed points, see page 36. Model 1600 Type S/600
Please contact Isotech to obtain current prices for If cold junction not required, specify NCJ.
calibration. UKAS calibration is included
Selection Guide
Values SRA: 1, 10, 25 and 100 Ohm 10, 25, 100, 1000, Standard Resistor
400 Ohm to special order 10,000 Ohm Maintenance Bath
Application Primary and Secondary Affordable Resistors with Precisely Maintains SRA
Lab microK inbuilt temperature control Primary Resistors at Fixed
Temperature
http://www.isotech.co.uk
When used with an external 0°C cold junction reference Parallel Processing Technology
unit (or by measuring the junction temperature with a The microK 70 and 125 incorporate additional technology
PRT on another channel) the microK can be used for low (compared to the micro 250 and 500) to deliver superior
uncertainty precision thermocouple measurements. The performance for the Primary Laboratory. A new technique
microK is designed for ALL the thermometer types used of Analogue Parallel Processing is used to lower noise to a
in a laboratory including Standard Thermocouples. The level that previously could only be achieved by the best
voltage uncertainty is 0.25µV, equivalent to 0.01°C for a AC Bridges. These models also feature an Ethernet port
Platinum / Gold thermocouple at 1000°C.
Notes: 1. Over whole range of SPRT, -200°C to 962°C. For Ro=0.25Ω increased by a factor of 2.5
2. E.g.: 25Ω SPRT with 25Ω standard resistor at water triple point or with direct comparison of similar SPRTs.
3. The microK uses a “substitution technique” in which the Device-Under-Test and the Reference are successively switched into the same position in the
measuring circuit. This means that the stability of resistance ratio measurements is immeasurably small.
4. Using external reference resistors.
g Expandable to 90 channels
g Supports PRTs, thermocouples & thermistors
g Plug-and-play operation
Keep-Warm Currents: The micosKanner has 10 Input connectors Cable Pod™ connector accepting:
individually programmable keep-warm current sources 4mm plugs, spades or bare wires
to maintain the power in PRTs when they are not being Contact material Gold plated tellurium copper
measured, eliminating uncertainty caused by power
coefficients. Interface RS232 (9600 baud)
Cable Pod™ Connector System: The connectors Operating conditions 15-30°C / 50-85°F, 10-90% RH
accept 4mm plugs, spades or bare wires. The standard (for full specification)
¾” separation is compatible with standard 4mm to BNC 0-50°C / 32-120°C, 0-99% RH
adaptors, so you can use thermometers with any standard (operational)
termination type. The Cable Pod™ connector system uses
gold-plated, tellurium-copper to give the lowest possible Power 88-264V (RMS), 47-63Hz (Universal)
thermal EMF and the best measurement uncertainty. 10W maximum, 1.2A (RMS) maximum
Reliable: Like the microK, the microsKanner uses the Size 520mm x 166mm x 300mm /
latest semiconductor technology for channel selection 20.5” x 6.6” x 11.9” (W x D x H)
and signal routing. This completely solid-state design
Weight 12.6kg / 28lb
therefore provides the highest possible reliability.
Operating principles
The problem: Temperature measurement is one of the most
demanding applications of resistance measurement. It
requires the measurement of resistance ratios to accuracies
of 1 part in 107 or better. While dc resistance standards are
sometimes available at this level, ac resistance standards Manual
are generally not. So how can we show our bridges
are accurate at this level, and that our resistance and
temperature measurements are traceable?
Automatic
The linearity check:
One simple method for checking a resistance bridge is to
measure a pair of resistors separately, and then measure the
two in series. Ideally the series measurement should equal
the sum of the two individual measurements. If not, then the
measurements give us a little bit of information about the
errors in the bridge readings. Note that we do not need to
know the values of the resistors to make this test work.
The RBC software in use: the data is input on the left and the Can you trust your bridge?
results of the analysis are seen in the right-hand windows.
In the paper “A Method for Calibrating Resistance
Thermometry Bridges” Rod White evaluated 38
Bridges, small but significant faults were found
with 15% of those tested, but “like the walking
wounded” they continued to provide a plausible
reading.
The RBC allows easy verification and calibration of
your bridge ensuring measurements are accurate
and traceable, use it to Restore Bridge Confidence
Built in statistics calculation can show you both the Inputs 2 channel
measured and average values along with the standard Pt100 (BS EN 60751 / IEC 751) or
deviation over previous measurements. 25.5/100Ω SPRT to ITS-90
Measuring 0.41mA
The Isotech TTI-22 is ideal as a reference standard Current
alongside liquid calibration baths, for the smallest
Self Heating 0.29mA (0.41mA / √2)
uncertainty calibration with Dry Blocks or for demanding
Test Current
stand alone measurement applications.
Measuring Time 1.44 seconds for both channels
Previously this level of performance was confined to
Measuring -250 to 960°C (0 to 440 Ohm)
specialist laboratories with expensive thermometry bridges;
Range
TTI-22 delivers 5 to 10 times the performance of
comparably priced instruments. Resolution Temperature: 0.0001°C, 0.1mK
Resistance: 0.00004Ω, 40 μΩ
g The TTI-22 uses the same patented measurement Uncertainty of Temperature: 0.001°C, 1mK 100 Ohm PRT
technique as the earlier TTI-2. Measurement Resistance: 0.4mΩ @ 20°C
Instrument only, uncertainty with sensor dependant
g Each measurement performs a zero point and gain on range and sensor type.
correction. Reference Internal 380Ω
g The switched polarity DC measuring current (0.4mA) Resistor TCR ±0.3ppm / °C
eliminates thermal EMFs. Stability ±5ppm / year
g
Interface RS232, Ethernet, built-in web server
Surface mount construction ensures long term
provides simple temperature display
reliability.
Ambient Temp. 10°C to 30°C
Range
Power Supply 7.5VDC, 250mA power adaptor or 2 x AA
batteries (typically >10 hours operating time)
Case Width: 190mm
Dimensions Height: 112mm
Depth: 240mm
Weight: 1.8kg
http://www.isotech.co.uk
The milliK has two input channels for sensors and a third
channel for current. It can be expanded to become a
measuring system with up to 33 channels reading SPRTs,
RTDs, Thermistors, or Thermocouples with the option to
control calibration baths and log readings accurately.
No Compromise Design
Benefiting You The design team have considered industrial users and
The milliK sets a new standard for value, versatility and applications in order to avoid measurement errors and
accuracy - < ±5ppm over range for PRTs, ±2µV for problems encountered in some instruments from other
Thermocouples and ±1µA for current transmitters, see table. manufacturers:
Supporting a wide range of sensors and functions it n Eliminates Thermal EMF Errors in PRTS
replaces individual devices making it a cost effective Fast current reversal technology and solid state switching
calibration solution. eliminate thermal EMF effects avoiding the errors that
occur with fixed DC instruments.
A robust design and operation from AC or DC power allows n Lead Wire Correction
the milliK to be used in the laboratory, test room or out in the PRT lead wire errors are eliminated for up to 30m of four
field. core screened cable. Also supports lead wire wire
correction for three wire PRTs.
The milliK can display in °C, °F, K, Ohms, mV and mA with
numeric and graphical display modes. The large back lit n Galvanic Isolation
display makes configuring the instrument and setting the Not only are the two sensor channels galvanically
scrolling strip charts intuitive. The USB port allows for the isolated, the 4 - 20mA input is also separately isolated.
use of a mouse, keyboard or USB Drive. The benefits of the advance design are no ground loops,
improved safety and noise immunity.
Built on World Leading Technology
In 2006 Isotech launched the microK range of thermometry High Resolution
bridges which quickly established themselves as the instrument The display resolution is 0.0001°C (0.1mK) made possible
of choice for National Metrology Institutes and Primary by using a powerful Sigma Delta Analogue to Digital
Laboratories with innovative features, accuracy and versatility. converter to achieve a true measuring resolution of just
28µΩ equivalent to 0.00007°C (0.07mK) for PRT inputs.
In response to industry demands for greater accuracy,
the milliK now brings the same design philosophy used in Expandable
the microK to those outside the Primary Laboratory. Users The millisKanner adds eight channels, and each can be
calibrating industrial sensors in the laboratory, pharmaceutical configured individually as a SPRT, PRT, Thermistor or
plants, food and beverage plants, aerospace, power Thermocouple input. A maximum of four millisKanners can
industries and service companies will welcome the milliK as be added, providing up to 32 channels - all controlled from
a solution to increase measurement confidence, ensure high the milliK touch screen or an RS232 connection.
accuracy traceable calibration, improve quality as well as
ensure safety and lower energy consumption.
1300°C
Accuracy Initial Over 1 year Display 89mm / 3.5” QVGA (320 x 240) colour
SPRTs/PRTs: 5ppm 7ppm TFT LCD with LED backlight
Thermistors: 50ppm 150ppm
Thermocouples: 2µV 4µV Operating Operating: 0-45°C / 32-113°F,
4-20mA: 0.01% 0.02% Conditions 0-99% humidity
Full Specification: 15-30°C / 50-85 °F,
Temperature Accuracy Initial Over 1 year 10-90% humidity
SPRTs/PRTs (at 0°C): 3mK 4mK
(over full range): 5mK 7mK Statistics In Addition to Instantaneous Display
Thermistors: 50ppm 150ppm user can select mean of 2 - 100
Thermocouples: measurements with Standard Deviation
Type B: ±0.23°C ±0.46°C
Type E: ±0.03°C ±0.06°C Measurement 950mS
Type J: ±0.04°C ±0.07°C Time
Type K: ±0.05°C ±0.10°C
Type L: ±0.04°C ±0.07°C Cable Length Limited to 10Ω per core and 10nF shunt
Type N: ±0.06°C ±0.12°C capacitance (equivalent to 100m of
Type R ±0.17°C ±0.34°C typical 4-core screened PTFE cable)
Type S: ±0.19°C ±0.38°C
Type T: ±0.05°C ±0.09°C Logging Capacity to store > 180 Days of time
Au-Pt: ±0.12°C ±0.23°C stamped measurements to internal
memory
Resolution Resistance (PRTs): 0.00001Ω
(Thermistors): 0.001Ω Recommended Isotech Semi Standard PRTs
Voltage: 0.00001mV Probes Isotech Model 909 SPRT
Current: 0.001mA
Temperature: 0.0001° Power 88-264V (RMS), 47-63Hz (universal),
6W maximum or 4 x AA cells
Temperature PRTs: IEC60751(2008),
Conversions Callendar-van Dimensions 255mm x 255mm x 114mm / 10” x 10” x
Dusen, ITS90 4.5” (W x D x H)
Thermocouples: IEC584-1 1995
(B,E,J,K,N,R,S,T), Weight 2.25kg / 5lb
L, Au-Pt
Optional 931-22-102
Thermistors: Steinhart-Hart,
Carring Case
polynomial
NOTE: Due to our program of continual development and improvement, we reserve the right to amend or alter characteristics and design without prior notice.
With no loss of accuracy and total flexibility a milliK and also from the control circuitry which allows for better
system can be configured to suit a wide range of measurements and lower noise.
reference thermometers and units under test. This The millisKanner is controlled from the milliK with plug
adaptable system saves on cost with no need for and play operation.
separate dedicated expansion modules and the flexibility
For use as a standalone switch for PRTs, the device
maximises the usefulness of each channel.
has UP / DOWN touch buttons or can be operated via
A solid state design avoids mechanical relays and RS232. The temperature of the input thermocouple
provides high reliability. The inputs are isolated with connectors can be read directly over RS232 to facilitate
galvanic isolation between the contacts and the PSU reference junction compensation.
Model 455
Accessories
932-19-72 35 litres of special oil
How to order
455 Standard Resistor Maintenance Bath
Evaluation reports describing the performance of Isotech equipment are freely available.
g Safe to use
g Economical
Method of Operation
A standard calibrated SPRT is placed in the equalizing
block together with the sensors to be calibrated. The
whole is allowed equilibriate.
The level is checked and Nitrogen added as necessary
and readings taken 10 minutes afterwards.
The Isotech Simple Liquid Nitrogen Apparatus is safe to
use, having no glass dewar flask internally to explode.
A comprehensive handbook accompanies the apparatus
which includes an article by Henry E. Sostmann on the
corrections required to convert the calibration to the ITS-
90 value of the Argon Triple Point.
Model 461
Technical Note: Simple Liquid Nitrogen Apparatus
Temperature Range -196°C Nominal
The Simple Liquid Nitrogen Apparatus, because there is
air access will slowly condense oxygen from the Uncertainty ±0.002°C
atmosphere increasing the temperature of the Boiling The temperature distribution across
Point. the block is typically 2mK. To this must
be added the uncertainty issued with
This is of small importance provided a calibrated SPRT is the calibration certificate from the
being used as the reference and simultaneous ratios of National Laboratory.
SPRT and unknown thermometers are being recorded, Extra uncertainties will also exist if
with a bridge such as the Isotech microK. dissimilar probes are compared.
Liquid Nitrogen is not supplied with the apparatus. Power NA
Dewar Dimensions Inside diameter 100mm
Depth 280mm
Volume 3 litres
Thermometer wells Four 8mm as standard, others to
special orders
How to order
461 Simple Liquid N2 Apparatus
http://www.isotech.co.uk
ALTERNATIVE INSERTS
Standard Insert B
13mm, 10mm,
8mm, 5mm
and 3.5mm
All 157 deep
Standard Insert C
8mm and
6 x 6.5mm
All 157 deep
The 915 is, no question, the finest stirred liquid parallel tube
bath produced.
Using best practise ratio methods of comparison calibration
uncertainties of less than 1mK can be obtained. The bath
can also be used to realise ITS-90 Fixed Point Cells (Mercury to
Tin Points) with absolute uncertainties to less than 1mK.
The Isotech Parallel Tube Liquid Bath is a development of the
previous 815 model and has many features which enhance
its performance and enable ease of operation. It is suitable
for the calibration of Liquid in Glass Thermometers, Industrial
Platinum Resistance thermometers, Standard Platinum
Resistance Thermometers, Thermocouples and Industrial
Temperature sensors. It can also be used with fixed point cells.
The temperature range of the standard 915 liquid calibration
bath (915H) is 40°C to 300°C. When the 915 is used in
conjunction with our external chiller the low temperature limit
can be extended to -65°C.
The 915 has a wide temperature range using silicon oils and
The standard temperature controller has resolution from 0.01°C
other suitable liquids. All components in contact with the
to 0.1°C which auto scales to suit the four digit display (all digits
liquid are of stainless steel and are insulated with materials
can be read with the Cal NotePad Software).
which are completely safe in use. The 915 used in conjunction
with a chiller, utilizes the latest ozone friendly gases. The standard controller can be replaced with a model offering a
Liquids are circulated by a propeller which mixes and forces display resolution of 0.01°C over the whole range. This High
the liquid through a specially designed orifice in the rear of Stability controller (Option 915/E) also offers better short term
the two parallel tubes. A variable speed motor optimizes the temperature stability, see table.
flow as the viscosity of the liquid changes. Below the orifice
With the high cost of some silicon oils the 915’s seven litre
plate liquid is circulated over a mineral insulated heater and
capacity makes it relatively inexpensive to fill compared to
temperature sensors which control the temperature of the bath.
many other baths.
The liquid flows up the calibration tube and weirs over the tube
into a collection tray where it returns to the rear tube for The changing of liquid is easily enabled by using the fitted drain.
recirculation. An angled side entry tube enables a refrigeration The design also allows for the expansion of liquids when being
cooling probe to be inserted in the rear of the two parallel tubes. raised to a particular calibration temperature.
Controller Note:
The standard and enhanced controllers include features
custom designed for Isotech by a world-leader in temperature
control technology. Power feedback is used to stabilise
against supply voltage changes, leading to greater
stability. A digital filter circuit ensures high integrity of
measurement correcting for drift, rejecting 50/60Hz
pick-up and filtering out other sources of input noise.
The High Stability Controller (Option 915/E) has a resolution of
better than 0.25µV which combined with powerful filtering
provides outstanding temperature stability.
915/11
915-05-38
915/05
(Block not
included)
915/03
915/02
915/01a
915/01b
See page 68
g Uncertainties to <1mK
g Physical separation of heating from Calibration Volume
g Three models covering -80 to 300°C
Isotech calibration baths employ sophisticated designs to
ensure calibration to the smallest of uncertainties. In 1986
Isotech purchased the interests of H N Irving & Sons
Temperature division which had specialised in providing high
quality baths to NPL and other international laboratories.
The knowledge and concepts from this scientific instrument
manufacturer were used in the development of the Isotech
815 and 915 Parallel Tube Liquid Baths. Whilst there is a
higher cost compared to simple designs the quality and
pedigree is unmatched.
The key feature is the physical separation of the chamber in
which the liquid is heated or cooled from the chamber that
contains the calibration volume in which the thermometers are
placed.
The geometry of parallel tube is also important, with the shape
and angle of the curves optimised for temperature uniformity.
The new Libra Model 785 achieves this with the same parallel
tube assembly as used in the 915 packaged into a different
case and now with internal refrigeration can be used to lower
Whilst the bath has a maximum immersion of 485mm the fluid
temperatures.
requirement is just 8.5 litres. This helps to lower operating
The cooling power varies dynamically with the benefit of costs as many other baths require a much larger volume of
giving greater stability at lower temperatures. fluid, which in turn can be costly.
915/01a Variable depth aluminium equalising block 915/05 Calibration Tube Cover. The cover consists
containing four drilled pockets 8mm of a square enclosure containing baffles
diameter by 160mm deep in which and fits over the calibration tube area
temperature sensors can be placed and is insulating the circulating liquid from
suitable for use with silicon oils. ambient air. Access for temperature probes
The equalising block is suspended centrally remains at the top of the enclosure. An
within the calibration tube and is easily equalising block may also be supported
removed. from this assembly. Maximum operating
temperature 180°C.
915/01b As an alternative to the above a copper
equalising block, dimensionally the same
as 915/01a, may be supplied. This block is
more suitable for use in water and other
liquids.
These deep stirred liquid baths are suitable for the low
uncertainty calibration of long thermometers including
liquid in glass and extended length SPRTS.
g Deep immersion depth (470mm / g Over and Under Temperature g Automatic Power Up - no manual
18.5”) with low filling costs Protection - protect expensive intervention needed - easier
Water Triple Point Cells as well as automation - energy saving
g Variable Speed Stirring - allowing
over temperature conditions
optimum performance for a wide g LiquidLevel to suit Liquid in Glass
range of liquid viscosities g Connections for easy drain and Thermometers
overflow - no level adaptors required
g Modular design for easier
maintenance
Calibration
90mm
Area
Equalizing Block 798-05-02A Aluminium Equalizing Overflow Adapter 796-02-10 The Orion liquid level is
Block, 4 pockets, within 10mm of the top
8mm diameter, lid allowing most liquid
160mm deep in glass thermometers
798-05-02B Copper Equalizing to be calibrated.
Block, 4 pockets,
8mm diameter, For applications where
160mm deep the liquid level needs to
798-05-02C Special Aluminium be higher than the lid
Equalizing Block the Overflow Adapter
To suit customer can be be fitted to the
requirements. Orion L and M
798-05-02D Special Copper This assembly includes
Equalizing Block a motor to pump the
To suit customer liquid and raises the
requirements. liquid level 20mm above
the top lid
Dual Cell Holder 796-05-01 Standard Dual
Cell Holder Includes Power Supply
Includes
interchangeable
Cell adaptors Monocular and Support 796-03 Monocular with x8
magnification to ease
reading liquid in glass
thermometers
Includes Tripod
Fixed Point Cells
17724 Mercury
Triple Point Cell
-38.8344°C
B12 Water
Triple Point Cell
0.01°C
17401 Gallium
Melt Point Cell
29.7646°C
17668ML Indium
Freeze Point Cell
156.5985°C
Over
Temperature Calibration
Controller Volume
Temperature
Controller
Overflow
Power
Set / Reset Drain
Cooling
(798H and
798EHT only)
Temp Range -80°C – 125°C -40°C – 125°C 30°C – 200°C 30°C – 300°C
Dimensions 405mm Wide, 610mm Deep, 940mm High (870mm to Top Panel)
100°C 100°C
75°C 75°C
50°C 50°C
TEMPERATURE (°C)
TEMPERATURE (°C)
25°C 25°C
0°C 0°C
-25°C -25°C
-50°C -50°C
-75°C -75°C
-100°C -100°C
0 30 60 90 120 150 180 0 20 40 60 80 100 120
TIME (minutes) TIME (minutes)
Calibration
Area 100mm
150mm
Accesories - Refer to ‘Liquid Selection Guide’ on page 77 for information on liquids and important Health and Safety Information
If calibrating over a limited range, say There are non toxic alternatives to The practise at Isotech is to use one
from 5°C to 90°C then water may be Methanol, including FlorinertTM from bath with a low temperature liquid,
used with is ideal in many ways - low 3MTM and Halocarbon 0.8 from and a second with oil for higher
cost, safe and readily available. Halocarbon. The disadvantage is temperatures and this is our
that these fluids are expensive recommendation. If an oil is used
For many labs though there will be a
compared to Methanol see www.3m. below the dew point water will
requirement to calibrate over much
com and www.halocarbon.com for condense into the open bath. If the
wider ranges.
further information. bath is then used above 100°C the
For low temperature calibration expansion of water present in the bath
For high temperature calibration the
alcohol or a mixture of alcohol and to steam can force the liquid from the
choice is usually that of a silicone oil.
water can be used. Methanol is bath creating a serious hazard. This
Silicone oils can be used at low
commonly chosen as it is easily is true for all open oil baths and the
temperatures too but icing can be a
obtained and relatively low cost. reason why Isotech do not
problem, particularly if a bath is used
Methanol is both toxic and flammable recommend using an oil both below
for long periods at low temperatures.
and must be used with care. the dew point, and then above 100°C
Model 875
Temp Range 50°C - 700°C
Volume (875/02) 67mm diameter, 475mm deep
Absolute Stability: ±0.005°C at 150°C
In Equalising Block ±0.005°C at 450°C
(875/09a) 10 Minutes ±0.020°C at 660°C
Vertical Uniformity ±0.005°C at 350°C
±0.005°C at 420°C
±0.026°C at 660°C
Heating time 50°C to 700°C < 240 minutes
Compressed Air Supply 1 BAR, 30 litres/minute at 100°C (50 L max)
Communications Includes Serial Interface, PC Cable and Software
Dimensions 580mm Wide, 640mm Deep, 1570mm High (880mm to Top Panel)
Weight 85kg (including 22Kg of Alumina Powder)
Safety Compliant to CE Regulations
Power 3kW (3 x 1kW heaters), 208-240V, 50/60Hz
How to Order Model 875 Fluidized Calibration Bath
Refer to Evaluation Report for Full Details
875/07
875/03
875/08
875/09A
875/06
Accessories
The following options are available:
875/02 Calibration Tube. As fitted as standard. This enables a calibration volume of 67mm inside diameter and 475mm
deep. This calibration tube is supplied and fitted as standard.
875/03 Oil free compressor and connecting tube. Used to supply air to the fluidized bed where an independent air
supply is not available within the laboratory.
875/04 Alumina powder specifically chosen for this application, supplied in a 25kg container.
875/06 Liquid in Glass Thermometer Support Gantry System consisting of a multi-tube probe holder with ten 10.80mm
by 470mm deep pockets, four support pillars and an adjustable stainless steel gantry with holes corresponding
to the probe holder, to locate up to ten thermometers simultaneously. Thermometer collars and ‘O’ rings (10 off)
are included with the assembly.
875/07 The multi-tube probe holder, described under 875/06 is available separately upon request.
875/08 Monocular and Support. Useful for viewing and magnifying the liquid column within the thermometer being
calibrated. This ancillary piece of equipment is used in conjunction with
Note: The bath is fully commissioned, tested and charged with the correct amount of aluminium oxide powder before despatch.
Accessories
877/01 Equalising Block 4 x 8mm + 4 x 6mm i.d.
(standard)
877/01S Equalising Block 16 x 6mm (special)
877/01E Equalising Block - Other configurations
(consult Isotech)
877/02 Platinum Foil Temperature Barrier
How to Order
877 Thermocouple Calibration Facility including Standard
Equalising Block.
Please specify voltage required
Note:
Spherical furnaces are normally supplied 240 Volts, 50 Hz, 3 kW with 8 sensor
insertion points, one of which is used to house the control thermocouple.
Eight tubes with a nominal internal diameter of 6 mm have been found to give
a very good and stable performance. It is the configuration around which the
specification has been written.
ISOTowers are intergrated devices including the ITS-90 Fixed Point Cell
g High accuracy
g Fast to temperature
g Simple to use
New
Conventional
Integrated Device
Isothermal
23 Hour Melt / 80% 1mK
Zone, T1
Ceramic Insulators
Reflectors
Air QFP
Immersion
Compensator
Q+
Q-
2 Hours
3mK Cell, T1
Heat Siphon
Cell
Q+
Q-
n Immersion Compensator*
Fully compensates for the
immersion characteristics of the
thermometer under test.
n Siphonic Cell*
A combined Fixed Point Cell and
Heat Pipe
The outer wall of the fixed point cell
is the inner wall of the heat pipe or
siphon, thus integrating the cell and
its apparatus, giving a guaranteed
performance from the cell.
http://www.isotech.co.uk/isotower_certificate.html
Isotech cells are of the highest purity available. Open cells conform to CCT/2000-13. Sealed cells are sealed to one
atmosphere with 6N pure argon at the freeze temperature.
Nominal Dimensions
Cell Outside Dia. Inside Dia. Height Material Depth
Slim Metal 37 8 220 160
Slim Quartz 38 8 226 160
Slim Mercury 36 9 235+140 130
Slim Gallium 35 10 200+45 140
Water B8 30 130 30 8 160 130
Water B12 40 210 40 12 365 210
Water B12 46 210 46 12 365 210
140
Heat up
120
100
Temperature (ºC )
80
60
40
20
-0
-20
Cool Down
-40
0 10 20 30 40 50 60
Time (minutes)
Accessories
Metal Block Bath Standard Insert 2 x 4.5 mm, 2 x 6.4 mm, 1 x 8 mm and 1 x 9.5 mm holes x 155 mm deep.
Non-standard Insert - please consult Isotech
Stirred Liquid Bath Stirred Liquid Container (For alcohols, water & oil)
C10 Oil (-35°C +140°C) 0.1 litre required
Liquid in Glass Thermometer Support kit
Stirred Ice/Water Bath Stirred Liquid Container (For alcohols, water & oil)
Liquid in Glass Thermometer Support kit
Blackbody Source Blackbody Target
Surface Sensor Calibrator Surface Sensor Calibrator Kit
ITS-90 Fixed Point ITL M 17724M Slim Triple Point of Mercury Cell
ITL M 17401M Slim Gallium Melting Point Cell
B8 Slim Triple Point of Water Cell
Semi-Standard Probe 935-14-82
Carrying Case 931-22-64
UKAS 5 Point Comparison Calibration
30
20
10
Temperature (ºC)
-10
-20
-30
0 1 2
Time (hours)
Gallium
Cell
Accessories
Metal Block Bath Standard Insert 6 x 8mm holes x 250mm
deep.
Adjustable Equalising Block
Non-standard Insert - please consult
Isotech
Stirred Liquid Bath Stirred Liquid Container (For alcohols,
water & oils)
C10 Oil (-35°C to 140°C) 1L required
Liquid in Glass Thermometer Support kit
Stirred Ice/Water Bath Stirred Liquid Container (For alcohols,
water & oils)
Liquid in Glass Thermometer Support kit
Blackbody Source Blackbody Target
Surface Sensor Calibrator Surface Sensor Calibrator Kit
ITS-90 Fixed Point B12 Water Triple Point Cell
ITL M 17401 Gallium Melting Point Cell
Semi-Standard Probe 935-14-85
Carrying Case 931-22-58
UKAS 5 Point Comparison Calibration
International Intercomparison
A set of metal clad slim cells from
mercury through aluminium were
tested at PTB by comparison to
PTB’s reference cells. All cells were
within PTB’s uncertainties.
Their results are summarised
Demonstration
versions of our software
are available to download
for evaluation:
http://www.isotech.co.uk/software.html
Accuracy Estimator
http://www.isotech.co.uk/icaleasy
milliK
SYSTEM CONNECTIONS
System Requirements
Computer Desktop or Notepad PC
Operating System Windows 98/2000/XP
Display 800 X 600 / 16 bit display
(1024 X 768 recommended) Note:
Available in different languages -
I-Cal - Supported Camera Check with Isotech for supported models Spanish and Chinese currently available.
Serial Ports A maximum of three ports are required, one for the dry
block, one for the TTI and one for the Switchboxes
http://www.isotech.co.uk/icaleasy
(Two Switchboxes can be operated from a single port)
Training
Our training course, “From Industry to ITS-90” has been
attended by several hundred delegates at venues across the
world. Many more have completed the content via our
E-Learning system. We also provide specialist courses
relating to the use of Primary Standards
http://www.isotech.co.uk/training.html
Journal of Thermometry
Published over ten years this remains a valuable resource for
temperature metrologists.
http://www.isotech.co.uk/journal.html