SunTech CT40 Manual
SunTech CT40 Manual
User Manual
Copyright Information
All content in this manual is the proprietary of SunTech Medical and is provided solely for purposes of operation,
maintenance or service of the SunTech CT40. This manual and the SunTech CT40 described herein are protected
under copyright law under which states they may not be copied, in whole or in part, without written consent of
SunTech Medical.
The information in this manual is furnished for guidance only, is subject to change without notice, and should not be
construed as a commitment by SunTech Medical. SunTech Medical assumes no liability for errors or inaccuracies
that may appear in this manual.
SunTech Medical, Inc. SunTech Medical, Ltd. SunTech Medical (Shenzhen) Co., Ltd.
507 Airport Blvd, #117 Oakfield Industrial Estate 105 HuanGuan South Road, Suite 15 2~3/F
Morrisville, NC 27560-8200 Stanton Harcourt Road DaHe Community Guanlan,
Phone: 1-919-654-2300 Eynsham, Oxfordshire OX29 4TS LongHua District, Shenzhen
1-800-421-8626 England GuangDong PRC 518110
Fax: 1-919-654-2301 Phone: + 44 (0) 1865-884-234 Tel: + 86-755-29588810
Fax: + 44 (0) 1865-884-235 + 86-755-29588986 (Sales)
+ 86-755-29588665 (Service)
Fax: + 86-755-29588829
The SunTech CT40 is designed to be simple and efficient to use and the CT40 features:
• mobility for improved and compact efficiency in a cost-effective package
• automatic BP checks
• averaging of multiple BP readings
• optional use of stethoscope for traditional auscultatory BP measurement
• robust memory
• connection to EMR system
• multiple options for pulse oximetry and temperature measurement
Note: For purposes of this manual, the SunTech CT40 (Model 260) may be referred to as “the SunTech CT40,” “the
CT40” or “the device”.
The SunTech CT40 is intended to be used on adult and pediatric patients using appropriately-sized SunTech One-
piece Durable (OPD) BP cuffs.
User Responsibility
Your SunTech CT40 is designed to perform in conformity with the description thereof contained in this operation
manual and accompanying labels and inserts, when assembled, operated, maintained and repaired in accordance
with the instructions provided.
Further, the user of this device bears sole responsibility for any malfunction that results from improper use, faulty
maintenance, improper repair, damage or alteration by anyone other than SunTech Medical or authorized service
personnel.
Observe the patient carefully during the measurement. Ensure pressure compatibility to all patients. If any
abnormality occurs, either in the unit or the patient, suspend the operation immediately and disconnect the BP cuff,
SpO2 sensor and thermometer (if applicable) from the patient.
Accuracy of any BP reading or oxygen saturation measurement may be affected by the position of the patient, their
physical condition and use outside of the operating instructions detailed in this guide. The interpretation of BP and
oxygen saturation measurements should only be made by a physician.
Safety and effectiveness when used with pregnant women, children under 3 years of age and neonates have not been
established.
ChipOx SpO2 Module: Use only pulse oximeter (SpO2) sensors supplied by SunTech Medical or original Nellcor® pulse
oximeter (SpO2) sensors supplied by Covidien® (except for forehead reflectance sensors).
Masimo® SpO2 Module: Use only original Masimo® pulse oximeter (SpO2) sensors and cables.
Wireless Networking
When connecting this device to a wireless network, it is important to use only the hardware specified by SunTech
Medical (see Accessories section for details). Unsupported USB accessories, including unsupported Wireless
adapters, have been intentionally disabled and will not function with this device.
Following the application of the BP cuff, petechia formation (a minute reddish or purplish spot containing blood that
appears in the skin) or Rumpel-Leede phenomenon (multiple petechia) may appear on the arm, which may lead to
idiopathic-thrombocytopenia (spontaneous persistent decrease in the number of platelets, associated with
hemorrhagic conditions) or phlebitis (inflammation of a vein).
WARNING: Do not attach the cuff to a limb being used for IV infusions or any other intravascular access, therapy or
an arteriovenous (A-V) shunt. The cuff inflation can temporarily block blood flow, potentially causing harm to the
patient.
WARNING: Pressurization of the cuff can temporarily cause loss of functionality of SpO2 if simultaneously using
device on the same limb.
WARNING: Do not apply the BP cuff to a limb being used for IV infusions as the cuff inflation can temporarily block
the infusion, causing harm to the patient.
WARNING: Check frequently by observing the limb that operation of the AUTOMATED SPHYGMOMANOMETER does
not result in prolonged impairment of the circulation of the patient.
WARNING: The cuff should not be applied over a wound as this can cause further injury.
WARNING: The cuff should not be placed on the arm on the side of a mastectomy. In the case of a double
mastectomy use the side of the least dominant arm.
WARNING: Do not use in the presence of flammable anesthetics or other flammable substances in combination with
air, oxygen-enriched environments, or nitrous oxide.
WARNING: Do not use the device if it has failed its diagnostics self test, or if it displays a greater than zero pressure
with no BP cuff attached or a value of oxygen saturation with no SpO2 sensor attached.
WARNING: Do not use if device is dropped and/or damaged. Have a qualified service representative check the unit
before using again.
WARNING: Do not remove unit covers. Doing so may cause electrical shock to the user. The device does not contain
any user serviceable components.
WARNING: Do not immerse the device in any fluid, place fluids on top, or attempt to clean the unit with any liquid
detergents, cleaning agents, or solvents. This may cause an electrical hazard. Refer to the cleaning section of this
guide for instructions on cleaning. If any of these situations apply, please contact SunTech Medical.
WARNING: Pulse rate measurement is based on the optical detection of a peripheral flow pulse and therefore may
not detect certain arrhythmias. The pulse oximeter should not be used as a replacement or substitute for ECG based
arrhythmia analysis.
WARNING: Do not use the device or any of its accessories during magnetic resonance imaging (MRI) scanning.
Induced current could potentially cause burns.
WARNING: Tissue damage can be caused by incorrect application or use of an SpO2 sensor, for example by wrapping
the sensor too tightly. Inspect the sensor site as directed in the sensor Directions for Use to ensure skin integrity and
correct positioning and adhesion of the sensor.
WARNING: Do not use high frequency surgical equipment with the CT40 as this may cause loss of stored data
WARNING: Federal (U.S.) law restricts this device to sale by or on the order of a physician.
CAUTION: A compressed or kinked connection hose may cause continuous cuff pressure resulting in blood flow
interference and potentially harmful injury to the patient.
CAUTION: Immediately replace parts that are broken, worn, missing, incomplete, damaged or contaminated.
CAUTION: Contact the nearest SunTech approved service center should repair or replacement become necessary. A
list of approved service centers appears in the guide or on our website at www.SunTechMed.com.
CAUTION: The reliability of the device depends upon conformance with the operation and service instructions, as
detailed in this manual.
CAUTION: Only replace battery with same type and model number.
CAUTION: To avoid the risk of electrical shock, this equipment must be only connected to supply mains with
protective earth.
CAUTION: Do not connect the device to equipment that does not meet EN60601-1. When the device is attached to a
patient, the device’s communication ports must only be connected to equipment that meets EN60601-1 standard.
CAUTION: Use only SunTech branded cuffs approved for use on the CT40 device.
CAUTION: Use only Masimo® oximetry sensors for SpO2 measurements with the Masimo® SpO2 module. Other
oxygen transducers (sensors) may cause improper performance.
CAUTION: Do not use damaged SpO2 or temperature sensors. Do not use a SpO2 sensor with exposed optical
components.
CAUTION: Do not immerse the SpO2 or temperature sensors in water, solvents, or cleaning solutions (the sensors and
connectors are not waterproof). Do not sterilize by irradiation, steam, or ethylene oxide. See the cleaning instructions
in the Sensor Directions for Use.
CAUTION: Do not use damaged patient cables. Do not immerse the patient cables in water, solvents, or cleaning
solutions (the patient cable connectors are not waterproof). Do not sterilize by irradiation, steam, or ethylene oxide.
See the cleaning instructions in the patient cable directions for use.
The following icons and symbols are used in this guide, on the SunTech CT40 equipment and packaging.
Warning message
Caution message
Manufacturer
Manufacture Date
Catalog Number
Serial Number
Keep dry
Temperature limit
Humidity limitation
Type B
USB-A or USB-B
Class II Equipment
Expiration Date
DIA Diastolic BP
SYS Systolic BP
Refer to the separate packing label stating which components you received based on the options you ordered with
your device.
If no temperature module will be added, then attach the right panel of the CT40, using the included Phillips-head
screwdriver. This panel will be on the right side, when the screen is facing the user.
For FILAC®:
Temperature unit connector: Insert the probe/well-assembly into the top of the FILAC® module.
For Touchless:
Plug in the touchless thermometer cable connector into the already inserted module.
Remove the left panel of the CT40, using a Phillips head screw driver. This panel will be on the left side, when the
screen is facing the user.
To add an SpO2 module, attach the module on the left side of the main unit. Fasten with the included Phillips-head
screwdriver.
For Masimo®:
For ChipOx:
Attach the ChipOx SpO2 sensor to the connector on the module while the retention clip is held upwards. Once the
sensor/cable connector is inserted, push the clip downward over the connector to hold it securely in the module. For
easy insertion, ensure module is parallel to the device.
Install Batteries
Install the rechargeable battery into the battery bay. Please read the battery instructions and the label on its surface
before use. Allow 8 to 12 hours for charging. All segments of the Battery Symbol will be lit when the device is fully
charged.
CAUTION: Fire, explosion and severe burn hazard. Replace only with SunTech part number: 98-0900-00.
If not installed in the device, the battery shall be kept away from heat, fire or other high temperature environments.
Keep the battery in a dry place stored at room temperature.
Do not disassemble, attempt to repair or use the battery for any other device or for any other purpose.
Do not place near any metal or use metal to shield the battery from physical damage as this may cause battery
overheating and/or a fire risk.
Do not short across the contacts of the battery or attempt to discharge the battery by shorting as a risk of fire or
explosion may result.
Do not expose or immerse the battery in water or attempt to clean with any cleaning agents. Only wipe battery with a
damp cloth if necessary.
Wash the affected area if electrolyte spills on skin or clothes. Leaking electrolyte may cause discomfort to the skin. If
it gets into the eyes, do not rob the eyes. Flush eyes immediately with water, and seek medical attention.
Battery Disposal
The SunTech CT40 device contains a lithium ion battery that contains materials which may be hazardous to
human health. Do NOT dispose of battery in domestic waste! Instead, please dispose of in an environmentally
responsible way, or return the battery to SunTech Medical. A prepaid return label can be obtained. Please see our
website for more information about our environmental policy at http://www.suntechmed.com/about-
suntech/environmental-policy.
Connect the BP cable, SpO2 sensor cable, and temperature sensor (if option is included) to the device.
Connect the barcode scanner and/or printer to the device if these options are included.
Use the Power Button on the left side of the SunTech CT40 to turn it on.
NOTE: The blue LED around the Selection Knob will be lit whenever the device is powered on.
System Configuration
The SunTech CT40 can be used out of the box with the factory settings. Any changes desired to the factory settings
can be made using the Advanced Configuration Application. The Advanced Configuration Application can be
accessed when the device is connected to a Windows device. Advanced configuration, such as connecting to a
network or EMR system, should be done by a qualified technician. See Advanced Configuration Application Guide
(SunTech document 80-0072-00-MO) for details.
Advanced device configuration, such as choosing either BP or SpO2 as the pulse rate source, or activating MAP
measurement (not available in the U.S.), is also done using the Advanced Configuration Application.
Patient ID
Memory icon
AC power connected
WARNING: User must take immediate action (see additional information in Section 12)
: Averaging mode
: Time before first measurement in minutes (Illuminates when Averaging Mode is selected)
: EMR icon prompts the user to confirm whether or not to send measurement data as message to EMR
system. (Only shown if EMR connectivity has been set up via Configuration Application.)
Press and hold toggle button for 4 seconds, until selection knob starts flashing. Release toggle button, and use the
selection knob to select the icon needed. Selected icon will blink after 1 second. Press the selection knob to confirm
the selection.
: Adult BP Mode icon. This icon is illuminated when the Adult BP mode is selected.
: Pediatric BP Mode icon. This icon is illuminated when the Pediatric BP mode is selected. The CT40 will remain in
the selected patient mode until a new mode is selected.
♥: Measured in BPM (Beats per minute), is illuminated when a heart rate is shown in the heart rate display.
MAP: Measured in mmHg is illuminated when a MAP value is shown (only if MAP functionality is enabled). NOTE:
This is a factory setting. MAP is not available in the U.S.
: Predictive mode
: Direct mode
: Human thermometry mode. Temporal (arrow on left side of head), Oral (arrow on right side of head), Axillary
(arrow under arm), or Rectal (arrow pointing to rectum).
: Object thermometry mode. Illuminates when object thermometry mode has been selected.
Memory Button
Press this button to enter Memory Mode and to store, view or delete saved measurements.
Save Current Reading: A quick press gives the user an option to save the patients reading, suggested to be done after
each reading.
Access Previous Reading: A long press (holding the button down for 4 seconds) gives the user the option to access
the bank of prior readings. Please see Section 4 to read more about the saving and accessing saved readings.
Press this button to toggle between three different BP measurement modes: Single Measurement, Averaging
Measurement and Auscultatory SphygMode.
A quick press gives the user an option to switch between BP measurements modes.
A long press (holding the button down for four seconds) gives the user the option to switch between pediatric or
adult mode. More information on switching from adult to pediatric mode can be found in Section 5.
Selection Knob
This round knob rotates clockwise and counterclockwise and can be pressed in to select different options on the
SunTech CT40. This button can be used to change parameters and to set the intended parameter.
Start/Stop Button
Home Button
Monitor Setup
Press the power button to start Initial Power-up. The display will gradually illuminate over a 3-second period. If this is
the first time the device has been turned on since it left the factory or since the device was set to factory defaults, the
user will be prompted to set the time and date in the following format: MM.DD.YYYY HH:MM:SS. The “MM” portion
of the field will flash first, prompting you to set the month. Turn the Select Knob to select the appropriate month and
press the Selection Knob. Then the next field, “DD” will flash, prompting you to select the day, and so on and so forth,
until you have set “SS” for seconds.
After setting the date and time, the device is ready to use! It is possible to make additional configuration changes to
the SunTech CT40 using the Advanced Configuration Application.
Depending on the out-of-range value, the display will indicate whether the value is out-of-range (hi) or out-of-range
(lo). If the out-of-range parameter is the systolic or diastolic pressure (or both), and MAP is enabled, MAP is not
displayed for that measurement. NOTE: MAP is not available in the U.S.
See Section 12, “Status Messages and Alarms,” for more details.
Nominal Power-up refers to every time the device is turned on after Initial Power-up. Simply press the power button
and the display will illuminate.
NOTE: The blue LED around the Selection Knob will be lit whenever the device is powered on.
Power-down occurs when you press the power button when the device is powered up. During power-down, all LCD
segments and icons illuminate for 1 second and then the entire display gradually darkens from normal brightness to
black and then one beep sounds. At this point, the device enters Sleep Mode, a low-power state, and is considered
off.
Automatic Power-down occurs when the device has not been used (i.e., no measurements taken or buttons pressed)
for 1 hour. The device will automatically execute the Power-down sequence and enter Sleep Mode.
Your SunTech CT40 offers flexibility to function in ways that best meet your work setting.
Documenting Measurements
Before you start taking measurements, it’s important to document the patient ID per your facility’s procedures. In
order to do so, the barcode scanner must be used to enter the patient ID into the CT40.
If your SunTech CT40 is connected to a barcode scanner, you can scan your patient’s barcode ID and it will show at
the top left of your screen. Then, the patient ID will remain attached to all measurements taken, saved in memory,
and/or sent to your facility’s EMR.
If you don’t require a barcode scanner, all measurement results will be displayed on the main screen and can be
saved to the device’s memory, written down in a chart, printed out (with optional thermal printer) or captured in
another way that meets your facility’s documentation procedures.
Printing
The SunTech CT40 can be connected to an optional SunTech thermal printer to easily print out measurement results.
See Section 11 for ordering information.
After connecting the optional CT40 thermal printer (see Section 2 and Section 3), ensure that the green LED on the
top of the printer is lit and unblinking before printing.
Reset Print
To print the information currently displayed on the CT40 screen, press the print button on the top of the printer (see
diagram above).
Saving Measurements
With EMR Connectivity
Technicians may have already connected the SunTech CT40 to your facility’s EMR system. If so, you can easily send
measurements directly to the EMR.
With the patient’s current measurements displayed on the screen, press the Memory button . The EMR icon
Press the Selection Knob to send the data to the EMR. If the measurements are delivered to the EMR successfully,
you will see the EMR Message icon and the Check Mark icon blink four times.
To store the displayed measurements in the device memory without sending to the EMR:
While the EMR and Check Icons are flashing, turn the Selection Knob until the X icon is displayed and then press
the Selection Knob. The measurements will be saved to the device memory only.
Without EMR
The SunTech CT40 does not have to be connected to an EMR system to save measurements. You can save them to
the device’s own memory. With the patient’s current measurements displayed on the screen:
All of the measurement values on the screen, the Check icon and the Memory icon will flash on and off 2
times.
Note: The CT40 does not autosave. The user must save the reading to memory after every reading.
To view results stored in memory, press the Memory Button for more than four seconds. You will see the
most recently saved measurements displayed on the screen, and can turn the Selection Knob counterclockwise to
see older measurements and then clockwise to see newer measurements.
TIP: When you send results to memory, make note of the “Memory Location” number to keep track of an individual
patient’s measurements. Then later on, you can match the Memory Locator number with a patient’s name if you pull
up or print out results for documentation.
The memory will hold up to 99 records. At 100 readings the device will continue to save, but will save by overwriting
the oldest records.
According to AHA guidelines, the patient should be seated for at least five minutes before taking a blood pressure
reading. The patient should be seated with feet flat on the floor and back supported. The patient’s upper arm should
be fully supported and resting at heart level. Encourage the patient to relax, and to be still and quiet.
Selecting the wrong size cuff will produce an inaccurate blood pressure measurement. When wrapped around the
patient’s arm, the Cuff Index Line should fall within the Range Markers printed on the cuff. If not, select a larger or
smaller cuff to ensure optimal BP measurement accuracy. The Artery Marker printed on the cuff must point to the
brachial artery, located inside the patient’s upper arm, between bicep and tricep. Wrap the cuff snugly around the
patient’s upper arm, so that the bottom edge of the cuff is approximately 1 inch above elbow joint.
Hold the Toggle Mode button down for 4 seconds until the adult and pediatric icons start blinking.
Turn the Selection Knob in either direction to select the desired icon
After making your selection, wait 2 seconds. The selected icon will be the one flashing.
Once you have selected the intended mode, press the Selection Knob to finalize this selection.
Note: Adult and Pediatric mode can be used in conjunction with any type of measurement mode.
Automatic Single BP: You can take a single, automatic measurement (Single Measurement Mode).
Averaged BP : Automatically take and average up to 5 multiple BP measurements (Averaging Measurement Mode).
Auscultatory/ Sphyg mode. Allows you to verify an automated oscillometric BP reading with a manual auscultatory
measurement (Auscultatory SphygMode).
Note: Use of Auscultatory SphygMode may be required to meet certain clinical trial protocols. It also can be used to
verify a BP reading if the automated BP results diverge significantly from a patient’s prior history, or if you as the
clinician deem it necessary. Auscultatory SphygMode mimics a traditional sphygmomanometer and does not measure
Systolic or Diastolic blood pressure.
Note: Averaging Measurement Mode can only be used for Automatic BP Measurements. It cannot be used to
automatically average measurements taken in Auscultatory SphygMode.
This is the device’s default mode. The SunTech CT40 will be in this mode each time it is powered up
in succession until these icons no longer show to return to Single Measurement Mode.
Press the Mode Toggle Button The Number of Measurements Icon Default values for Time Before First
will flash as will the Selection Measurement and Time
until you see the Knob. To keep the default of 3
Averaging Measurement Mode measurements, press Knob. Or turn Between Measurements will
Knob to choose a different number flash in sequence as the user is
Icon . . From single
and press Knob to select. prompted to set each value. To keep
measurement mode, that will only
defaults, press Knob. Or turn Knob to
be one button press.
choose new values and press Knob
to select.
Example of Averaging Measurement You can select 2 to 5 measurements On left is Time Before First Reading
Mode setup screen with default to average*. In this example, 4 will start. Select 0-5 min. Default = 0.
values. measurements are set to be On right is Time Between
averaged. Measurements. Select 15-120 sec.
Default = 15 sec.
*Note: In averaging mode, you are unable to see individual readings. Averaged Results will be displayed at the end of all
the readings. By default, all measurements taken (from 2 to 5) will be included in the averaging calculation. Using the
Advanced Configuration Application, the device can be programmed to always throw out the first measurement taken
and not include it in the averaging calculation.
**Note: All values selected during the Averaging Measurement Mode will be displayed the next time that this mode is
used, so that the same customized averaging protocol can be set and used each time that Averaging Measurement
Mode is used.
Example shows Auscultatory SphygMode setup display. Three-digit number is the target cuff inflation pressure setting.
The default is 160 mmHg.
Ensure values are cleared from When selecting this mode the Put stethoscope over artery.
display except for SYS value will flash to allow the user to decide
representing current pressure in how many reading they would like Press Start/Stop to
cuff. averaged. inflate cuff automatically. Similar to
Press Start/Stop Button Select number of readings by an analog sphyg gauge, the display
rotating the selection knob and shows current pressure in cuff.
. push the selection knob to set the When cuff reaches max pressure,
number of readings. deflation will begin automatically.
Inflate/deflate cycle proceeds
automatically until results are The minute number will flash and Listen for K-sounds while watching
displayed. the user can select how many displayed pressure. Note SYS and
minutes until the first reading starts DIA.
A long beep will sound to indicate
measurement is complete. by rotating and pushing the
selection knob. Press Start/Stop again to
The user can then select how often quickly dump remaining pressure
the readings will occur, and select from the cuff.
this by rotating the knob and
pushing the knob to selection
wanted.
Example of SYS, DIA and HR* after Example of SYS, DIA and HR* after Example of SphygMode display with
BP measurement in Single all measurements have been cuff pressure at zero. P-set value is
Measurement Mode. averaged. default 160 mmHg.
*Note: Pulse rate source, whether BP or SpO2, can be selected at any time using the Advanced Configuration
Application. The default setting uses BP as HR source. If used outside of the United States, the device also can be
configured to measure MAP using the Configuration Application. If set up to measure MAP, the MAP and HR
measurements will alternate in 3-second intervals. NOTE: MAP is not available in the U.S.
Pressing the Memory Button will clear all patient and measurement data from the display, in addition to
saving the patient and measurement information in memory. It is recommended that the display is cleared before
taking readings on a new patient.
The SunTech CT40 uses BP as the default HR source. HR source can be selected using the Advanced Configuration
Application.
Traditional pulse oximetry assumes that all pulsations in the light absorbance signal are caused by oscillations in the
arterial blood volume.
The traditional pulse oximeter calculates the ratio of pulsatile absorbance (AC) to the mean absorbance (DC) at each
of two wavelengths (typically one visible, the other infrared). The oximeter then calculates the ratio of these two
arterial absorbance signals. This value is used to find the saturation SpO2 in a standard table built into the oximeter
software. The values in the table are based upon human blood studies against a laboratory co-oximeter on healthy
adult volunteers in induced hypoxia studies. The resulting value is reported as oxygen saturation level (%).
WARNING: Pulse oximeter probes and sensors are designed for use with specific pulse oximeter devices. Only use
probes and extension/adapter cables that are specified for each SpO2 module.
WARNING: The responsible organization and/or operator must verify the compatibility of the oximeter, oximeter
probe, and extension/adapter cable before use. Otherwise, patient injury can result.
WARNING: Misapplication of a pulse oximeter probe with excessive pressure for prolonged periods can induce
pressure injury.
Incorrect sensor application or use. Proper sensor use and placement is critical for good performance.
Dyes or any substance containing dyes that affect light absorption may cause erroneous readings.
• Some nail polish colors (particularly dark shades) or artificial fingernails may reduce light transmission and
affect pulse oximetry accuracy. Remove any nail polish or artificial fingernails before using the SpO2 sensor.
Exposure to excessive illumination, such as surgical lamps (especially ones with a xenon light source), bilirubin
lamps, fluorescent lights, infrared heating lamps, or direct sunlight (exposure to excessive illumination can be
corrected by covering the sensor with a dark or opaque material)
Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line. Always place
the SpO2 sensor on the arm without a blood pressure cuff, arterial catheter, or intravascular line.
CAUTION: Pulse rate measurement is based on the optical detection of a peripheral flow pulse and therefore may not
detect certain arrhythmias. The pulse oximeter should not be used as a replacement or substitute for ECG based
arrhythmia analysis.
CAUTION: SpO2 is empirically calibrated to functional arterial oxygen saturation in healthy adult volunteers with
normal levels of carboxyhemoglobin (COHb) and methemoglobin (MetHb). A pulse oximeter cannot measure
elevated levels of COHb or MetHb. Increases in either COHb or MetHb will affect the accuracy of the SpO2
measurement.
CAUTION: COHb levels above normal tend to increase the level of SpO2. The level of increase is approximately equal
to the amount of COHb that is present.
CAUTION: High levels of COHb may occur with a seemingly normal SpO2. When elevated levels of COHb are
suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed.
CAUTION: For increased MetHb, the SpO2 may be decreased by levels of MetHb of up to approximately 10% to 15%.
At higher levels of MetHb, the SpO2 may tend to read in the low to mid 80s. When elevated levels of MetHb are
suspected, laboratory analysis (CO-Oximetry) of a blood sample should be performed.
CAUTION: Venous congestion may cause under reading of actual arterial oxygen saturation. Therefore, assure proper
venous outflow from monitored site. Sensor should not be below heart level (e.g. sensor on hand of a patient in a bed
with arm dangling to the floor).
CAUTION: Venous pulsations may cause erroneous low readings (e.g. tricuspid value regurgitation).
CAUTION: The pulsations from intra-aortic balloon support can be additive to the pulse rate on the oximeter pulse
rate display. Be sure to verify patient's pulse rate.
CAUTION: Elevated levels of Total Bilirubin may lead to inaccurate SpO2, measurements.
CAUTION: With very low perfusion at the monitored site, the readings may read lower than core arterial oxygen
saturation.
CAUTION: Do not immerse the sensor or patient cable in water or, solvents, or cleaning solutions (The sensors and
connectors are not waterproof).
CAUTION: Loss of pulse signal can occur in any of the following situations:
There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp, or sunlight.
A blood pressure cuff is inflated on the same extremity as the one with a SpO2 sensor attached.
NOTE: See SP02 Sensor Specifications Table on page 45, the range of the peak wavelengths and maximum optical
output power of the light emitted by the pulse oximeter probe, which can be especially useful to clinicians.
NOTE: This device does not include an alarm system that detects an SpO2 or pulse rate physiological alarm condition.
This device is not designed for long-term monitoring and should only be used for spot-check measurements.
System Description:
The SunTech CT40 spot check device with pulse oximetry module consists of a base CT40 device for the display of
SpO2 measurement data and other user information, either a ChipOx or Masimo SET SpO2 module which is attached
to the CT40 base unit as described in Section 2, and the specific reusable pulse oximeter probe and/or cable that
comes packaged with each module. Device setup is described in more detail in Section 2. Both the Masimo® SET and
ChipOx SpO2 modules are calibrated to display functional oxygen saturation (SpO2).
2. The fingernail should face the side with the red light. Make sure that long fingernails do not interfere with proper
finger position
3. An SpO2 reading will be displayed after a few seconds. Remove the finger from the SpO2 sensor, and the last
measurement will be displayed and flashed.
4. If the sensor is not removed from the finger, SpO2 readings will be automatically stopped after 10 minutes of
continuous measurement, and the last measurement will be displayed and flashed.
5. To start a new measurement, re-apply the SpO2 sensor to the finger as described in Step 1.
NOTE: If the signal quality from the sensor drops below 90%, no measurement will be displayed in the SpO2 section
, and the percent symbol will begin flashing. Check sensor for correct placement and usage.
NOTE: If the accuracy of any measurement does not seem reasonable, first check the patient’s vital signs by alternate
means and the check the MS board pulse oximeter for proper functioning.
NOTE: A functional tester cannot be used to assess the accuracy of a Masimo pulse oximeter sensor or pulse oximeter.
Using Clinical Dynamics SmartSat SpO2 simulator, the device can display the entire range of SpO2 values.
Masimo® Sensors:
Before use, carefully read the M-LNCSTM DCI/DCIP sensor directions for use, which includes cleaning and disinfection
instructions.
In addition to the General Principles of Operation mentioned above, the Masimo® SET® MS board pulse oximeter uses
the following principle:
The MS board decomposes the ratio of pulsatile absorbance (AC) to the mean absorbance (DC) at each of two
wavelengths, 660 nm and 905 nm (S(660) and S(905)) into an arterial signal plus a noise component and calculates
the ratio of the arterial signals without the noise:
S(660) = S1 + N1
S(905) = S2 + N2
R = S1/S2
Again, R is the ratio of two arterial pulse-added absorbance signals and its value is used to find the saturation SpO2 in
an empirically derived equation into the oximeter’s software. The values in the empirically derived equation are based
upon human blood studies against a laboratory co-oximeter on healthy adult volunteers in induced hypoxia studies.
The above equations are combined and a noise reference (N’) is determined:
N’ = S(660) - S(905) x R
If there is no noise N’ = 0: then S(660) = S(905) x R which is the same relationship for the traditional pulse oximeter.
The equation for the noise reference is based on the value of R, the value being sought to determine the SpO2. The
MS board software sweeps through possible values of R that correspond to SpO2 values between 1% and 100% and
generates an N’ value for each of these R-values. The S(660) and S(905) signals are processed with each possible N’
noise reference through an adaptive correlation canceler (ACC) which yields an output power for each possible value
of R (i.e., each possible SpO2 from 1% to 100%). The result is a Discrete Saturation Transform (DST™) plot of relative
output power versus possible SpO2 value as shown in the following figure where R corresponds to SpO2 = 97%:
The DST plot has two peaks: the peak corresponding to the higher saturation is selected as the SpO2 value. This
entire sequence is repeated once every two seconds on the most recent four seconds of raw data. The MS board
SpO2 therefore corresponds to a running average of arterial hemoglobin saturation that is updated every two
seconds.
The tested device meets the stated accuracy specification of RMS Error = 2.00% for the range 70% to 100% SaO2. The
accuracy is not specifed below 70% SaO2. In addition, here are detailed measured ARMS values over the specified
discreet ranges:
Range ARMS
90-100% 0.60%
80-90% 0.54%
70-80% 0.67%
NOTE: A functional tester cannot be used to assess the accuracy of a ChipOx pulse oximeter sensor or pulse oximeter.
Using FLUKE Biomedical Index 2 SpO2 simulator, the device can display the entire range of SpO2 values.
ChipOx Sensors:
Before using, carefully read the ChipOx sensor directions for use, which includes cleaning and disinfection
instructions.
ChipOx sensors are provided non-sterile.
Range ARMS
90-100% 1.54%
80-90% 2.10%
70-80% 2.56%
NOTE: Always apply a new probe cover before taking temperature. Do not reuse probe cover. Use only probe covers
designed for use with FILAC® 3000 probes. Using the incorrect probe cover can severely damage the probe and cause
measurement inaccuracies.
NOTE: To change between Celsius and Fahrenheit, press the °C/°F button on the FILAC® 3000 temperature module.
Press again as needed.
After you withdraw the probe from the isolation chamber and apply a probe cover, the thermometer begins working
automatically. You will see the Human Thermometry Mode icon illuminated on your display with the appropriate
arrow illuminated (i.e., pointing to the mouth on right side of head for oral temperature taking). Press the Site button
on your thermometer unit to switch between Oral and Axillary mode.
NOTE: Be sure to eject the used probe cover into an approved bio-hazard container before attempting to place the probe
back in the isolation chamber.
Quick Mode is an oral predictive measurement mode that provides a fast reading. Quick Mode allows you to rapidly
identify patients with normal body temperatures. If the patient temperature is outside of the “normal” range, the
FILAC® 3000 electronic thermometer will automatically switch into its standard predictive mode in order to provide a
more accurate reading. Quick Mode is indicated by a rabbit icon on the display . Quick Mode is not available
when in Cold Mode or in Direct Mode.
The FILAC® 3000 thermometer normally operates in predictive mode to provide fast and accurate temperature
measurements. However, in instances when no measurement site is detected or the temperature does not stabilize,
the thermometer will automatically switch to Direct Mode . Additionally, the FILAC® 3000 electronic thermometer
will automatically switch into Direct Mode if the ambient temperature is greater than 35°C (95°F). If this happens, you
will hear two short beeps and see the Direct Mode turtle icon appear . This icon will be continuously displayed
whenever the thermometer is functioning in Direct Mode. The Direct Mode auto feature is always functional for both
the Red or Blue isolation chamber/probe combinations.
Cold Mode can be selected if a patient’s body temperature is expected to be lower than normal, such as when he is
coming out of surgery. Cold Mode is activated by pressing the Site Selection button and °C/°F button simultaneously
on the temp module. When selected, the Cold Mode snowflake icon is displayed, and the probe will preheat to
33°C (91°F). The accuracy and measurement time of Cold Mode measurements are equivalent to standard prediction
measurements at the respective body sites.
On your SunTech CT40 display, you will see the Human Thermometry Mode icon selected with the arrow pointing to
the side of the head . Hold the thermometer probe 2 cm to 3 cm from the patient’s temple and press the START
button on the back of the thermometer. The reading is complete when you hear a beep, and the temperature
measurement will appear on your device display. Place the thermometer back in the device holster.
NOTE: Be sure to wipe away any sweat from the area around the temporal artery, and be sure to remove eyeglasses and
push hair away from measurement site before initiating a measurement.
NOTE: The area just behind the ear lobe can be used as an alternative measurement site to the temporal artery.
NOTE: The SunTech Touchless Thermometer should not be used near sources of heat or cold, such as a heater or air
conditioner vent. Also, do not use outdoors. Use this product in a room where the temperature is between 16°C and
40°C (61°F and 104°F).
NOTE: If patients are subject to a significant change in temperature from outside to inside, the patient should be
allowed to acclimate for approximately 15 minutes before using the touchless thermometer.
The SunTech Touchless Thermometer can also be used to measure the temperature of an object, such as a vessel
containing liquid, or food. To use Object Mode, ensure that the thermometer is turned on by pressing the START
button. Press the Mode Button briefly. The Object Mode icon will then appear in the device display. Place the
thermometer 2-3 cm from the desired measuring point on the object and press the START button. Hold thermometer
in place until you hear a long beep. The temperature will be displayed on the screen.
NOTE: Remember that surface temperature and inner temperature of an object can vary greatly.
*Touchless Thermometry module is not available in all countries. Contact SunTech to find out if this feature is available
in your area
Memory Mode
When you power up the unit, the memory identifier appears at the top of the display beside the Memory Icon .
This is the number that will be associated with the next set of measurement results that are sent to the device
memory.
Press the Memory Button on the front of the device to save results to memory. The Memory Icon, check
mark icon and all values that are displayed on the screen (at the time the button is pressed) will flash before the
results are saved. The following values can be saved to memory.
Systolic BP Numeric
Diastolic BP Numeric
SpO2 Numeric
Temperature Numeric
NOTE: If a parameter has no data showing at the time the memory button is pressed, then all values for that parameter
will be stored as ‘null’.
Press and hold the Memory Button for four seconds on the front of the device .
You will then be able to use the Selection knob to scroll through prior readings
Press the Memory Button or the Home Button to exit memory mode.
Press the Selection Knob while in memory mode. The Memory Delete icon and the Check icon will flash, along
with the Memory Identifier number.
Press the Selection Knob to delete that measurement. Or rotate the knob clockwise until the X icon flashes and
press the Selection Knob to cancel the deletion.
While in Memory Mode, press and hold the Memory Button for less than three seconds, and while holding, then press
the Selection Knob.
The Memory Delete icon and Check icon will flash in sync with the Selection Knob. All other values on the
screen will be blank.
Press the Selection Knob again to delete all measurements, or rotate the knob clockwise until the X icon flashes
and press the Selection Knob to cancel the deletion.
EMR Transmissions
Section 4 covers the basics of sending measurements to EMR. Here are more details.
Scan the patient’s barcode ID with your barcode scanner. If your SunTech CT40 is connected to an EMR, a verification
query will automatically be sent to the EMR to confirm the validity of the patient ID. If the EMR responds that the
patient ID is valid, then the patient ID icon, field values and Check Icon will flash four times.
If the EMR responds that the Patient ID is invalid, then the Patient icon field, field values and X icon will flash six times
and a message will appear in the Patient ID field indicating that the Patient ID entered is not valid.
NOTE: The Patient ID must be validated by the EMR before any vital signs measurements can be sent to the EMR.
Sending measurements to an EMR will work only if your SunTech CT40 has been configured to connect to an EMR.
After scanning a valid Patient ID, and taking vital signs measurements, press the Memory Button . The EMR
icon and Check Mark Icon will start flashing. Press the Selection Knob to send the data to the EMR. If
the measurements are delivered to the EMR successfully, you will see the EMR Message Icon and the Check
If the EMR has a problem receiving the measurements, the EMR Message icon and X icon will flash eight
times and you will hear 4 short beeps. You will see a message indicating the nature of the problem, such as:
“Rejected” or “Rejected for an error.” This message occurs when the device is connected to the network but the EMR
rejects the message. A corresponding message will appear in the Patient ID field during a three-second period when
“No response (time out).” This message occurs when the device is connected to the network but the EMR is not
responding. It also can occur if network connectivity is lost. If there is a network connection, the SunTech CT40 will
continue to try to transmit the message every 10 seconds. A message in the Patient ID field will alert you that another
attempt to send the message will be made in 10 seconds. If network connectivity is lost, you will see a corresponding
message in the Patient ID field, and the device will continue trying to reestablish a connection and send the
measurements at intervals set during EMR setup. If problems persist, you may wish to contact your facility’s
information technology department. As a backup, the patient’s measurements are saved in the device memory when
the memory button was pressed, and can be retrieved using the procedure detailed in the beginning of Section 9.
Note: After the memory button is pressed, all values to be stored to memory will flash off and on twice, along with the
check mark icon, memory icon and memory location identifier. After flashing, the display will revert to null values for the
patient ID and all measurement data, and the check mark icon will flash off and on twice more.
CAUTION: The SunTech CT40 is not sterilizable. Do not immerse the monitor in any fluid or attempt to clean
with any liquid detergents, cleaning agents, or solvents.
Dampen a soft cloth with mild medical grade disinfectant and wipe the device to remove surface dust and dirt. Dry
surface thoroughly before use.
CAUTION: Never immerse sensors, clips or thermometers in fluids. Do not pour or spray any liquids onto the
sensor or thermometers. Caustic or abrasive cleaners will cause permanent damage. Do not open the case of the
Sp02 sensor finger clip sensor more than 45° or the case will be damaged.
Clean the Sp02 sensor and thermometers with a soft cloth dampened with a mild medical grade disinfectant or
isopropyl alcohol. Remove all tape residues. Allow the sensor and thermometer to dry thoroughly before reusing.
The SunTech CT40 performs a range of system checks during normal operation. If the device detects a problem, it
will display an error code.
Replaceable Parts
On a routine basis, inspect the monitor, cuffs and hoses for cracks, fraying, or kinks. Immediately replace any
damaged part. Refer to the list of Accessories & Replacement Parts in this guide.
CAUTION: Fire, explosion and severe burn hazard. Replace only with SunTech part number: 98-0900-00
When the battery no longer charges or it needs excessive recharging, it may need to be replaced. See the “Installing
Battery” section above for proper installation and precautions to be taken when installing the rechargeable battery.
Please dispose of the old battery per the instructions below.
Battery Disposal
The SunTech CT40 device contains a lithium ion battery that contains materials which may be hazardous to
human health. Do NOT dispose of battery in domestic waste! Instead, please dispose of in an environmentally
responsible way, or return the battery to SunTech Medical. A prepaid return label can be obtained. Please see our
website for more information about our environmental policy at http://www.suntechmed.com/about-
suntech/environmental-policy
Product Disposal
Do not dispose of this product as unsorted municipal waste. Prepare this product for reuse or separate
collection as specified by Directive 2002/96/EC of the European Parliament and the Council of the European Union on
Waste Electronic and Electrical Equipment (WEEE).
Cuff Disposal
Do not return used cuffs. Used blood pressure cuffs may be contaminated medical waste and should be dealt
with in accordance to your local regulations.
Routine Calibration
Have the SunTech CT40 blood pressure function checked every two years to verify the accuracy of the pressure
transducers and indicators. Please have this done by a biomedical technician or SunTech Customer Service.
CAUTION: Calibration should be done by a biomedical technician or other person familiar with the SunTech
CT40 device.
If a biomedical technician performs this function, they should contact SunTech Medical for instructions to access
“Verify Calibration”. Instructions are also available in the SunTech CT40 service manual (SunTech Part 80-0068-XX-
SM).
For customers in Europe, the Middle East, Africa, Asia, and the Pacific:
Main Unit
Part Number Description Details
98-0900-00 CT40 Rechargeable Lithium Ion Battery (rated 7.2V, 6600mAh) (custom Each
(*) Masimo® M-LNCS DCI Adult SpO2 Reusable Sensor, 2501 Each
Stands/Printer/Scanner
Part Number Description Details
The icon will blink with the respective measurement display result;
Below are common error codes. Take action as directed on the screen, or as suggested in the table below.
Status Messages
Blood Pressure:
Check that the cuff is properly tightened and tubing properly connected.
Check that there is no excessive clothing between the arm and cuff.
85 & 89 Check that the hose has no sharp bends and is not pinched.
86 If cancelled reading was unintended, allow cuff to deflate & restart reading.
Temperature:
Make sure patient has been at rest for a few minutes. If patient has been in cold
environment, make sure patient has acclimated to indoor temperature
FILAC®:
Make sure patient has not had any warm beverages recently (oral), and then retry
measurement.
Retry reading.
SpO2:
121-124, 133-136, Remove sensor from the patient, and ensure proper positioning. Attempt new measurement
150-156 ensuring proper protocol as in section 7 of the User Manual.
137-149 & 157-192 If issue persists, replace module or call Customer Service
601-610 & 612,614 Check EMR Configuration, check wireless strength and proximity. Call biomedical engineer
or IT support.
CT40 System:
Powering Device:
If repower attempt does not work, plug in power source and power off and repower on CT40
again.
203-211 Clear the error and power off and repower on the CT40.
Replace battery.
Barcode Scanner:
701-706 Check and insure proper barcode and Patient ID was scanned
Rescan Patient ID
Printing:
383- 387 & 388-390 Check cables. Reset Printer. Replace Printer.
The SunTech CT40 will display temperature and SpO2 within the following ranges.
SpO2: 70 – 100%
Single measurement mode: The highest or lowest possible threshold range will blink in the appropriate BP
measurement field.
Averaging mode: will blink in addition to the status code “3” in the blood pressure measurement display
Temperature: depending on the out or range value, the display will indicate whether the value is out of range (hi) or
out of range (Lo).
SP02: results falling below the measurement range will display “Lo” in the results field.
Service Centers
For customers in the Americas, Asia, and the Pacific:
When batteries are fully charged, the SunTech CT40 is designed to run on internal battery power for 120
measurement cycles (BP, temperature and SpO2 measurements) before requiring recharge. When only the bottom
segment of the battery symbol is illuminated, there is between 11 percent and 40 percent battery power remaining.
When no measurements have been taken or buttons pressed for one hour, the device will automatically power-down
and enter the low-power state and is considered “off.”
If you abort a BP measurement by pressing the Start/Stop button, you will hear one short beep. Then after the
pressure has been completely dumped from the cuff, you will hear one long beep.
Can pulse oximetry, temperature measurement, a barcode scanner or printer be added later too?
Yes, you can add vital signs modules and accessories to the SunTech CT40 at any time.
Can EMR connectivity be established at a later date if not enabled as part of initial configuration?
Yes, EMR settings can be accessed via the Advanced Configuration Application at any time.
If the device is configured to use the Sp02 sensor as the pulse rate source, and the sensor is not attached to the
patient, there will not be heart rate data on the display. Only dashes will appear, or MAP results if MAP measurement
is enabled. Note that MAP is not available in the United States.
The SunTech CT40 can store 99 unique sets of measurements. After the 99th measurement has been saved, the next
measurement will have the Memory Location Identifier label “01” and will overwrite the most recent measurement
(measurement 99). After all results are cleared from the memory, the Memory Location Identifier also will return to
“01.”
Connect the equipment into an outlet on a circuit different from which the other device(s) are connected.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
WARNING: Use of accessories, transducers, and cables other than those specified may result in increased emissions
or decreased immunity of the CT40.
WARNING: The CT40 should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, the CT40 should be observed to verify normal operation in the configuration in which it will be used.
The CT40 is intended for use in the electromagnetic environment specified below. The customer or the user of the
CT40 should assure that it is used in such an environment.
RF emissions Group 1 The CT40 uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
CISPR 11
in nearby electronic equipment.
Voltage Complies
fluctuations/flicker
emissions IEC
61000-3-3
The CT40 is intended for use in the electromagnetic environment specified below. The customer or the user of the
CT40 should assure that it is used in such an environment. When the product is in an environment with static, the
product may show error SpO2 reading, it will recover automatically when the static is gone. Though it may exhibit
DEGRADATION of performance (e.g. deviation from MANUFACTURER’S specifications) that does not affect BASIC
SAFETY or ESSENTIAL PERFORMANCE.
Electrical fast ±2 kV for power ±2 kV for power Mains power quality should be that of a
typical commercial or hospital
transient/burst supply lines and supply lines and patient
environment.
patient coupled coupled lines
IEC 61000-4-4
lines
Surge ±1 kV line(s) and ±1 kV line(s) and neutral Mains power quality should be that of a
neutral typical commercial or hospital
IEC 61000-4-5
environment.
70 % UT 70 % UT
<5 % UT <5 % UT
for 5s for 5s
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level
3 V/m
Radiated RF 3 V/m
80 MHz to
IEC 61000-4-3
2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the CT40 is used exceeds the
applicable RF compliance level above, the CT40 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the CT40.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the CT40
The CT40 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the CT40 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the CT40 as
recommended below, according to the maximum output power of the communications equipment.
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
W
100 12 12 23
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Specifications, General
Operating: 10°C (50°F) to 40°C (104°F) 15 – 90% RH non-condensing - 700 kPa - 1060 kPa.
Operating the monitor in an environment at maximum temperature can produce temperatures
Conditions for
exceeding 41°C (41.6°C highest recorded) on a patient applied part. It is up to the operator to
Use:
determine if this temperature is too high based upon the condition of a patient and, if so, to
ensure the ambient temperature of the environment is 38°C or below.
-20°C (-4°F) to 55°C (131°F) 15 – 90% RH non-condensing - 500 kPa - 1060 kPa. Performance
Storage: can be affected if, used or stored outside the specified temperature, humidity, or altitude listed
in the ranges above.
Internal power supply. Input: 100-240 VAC @ 1.5A max, 50-60 Hz. Output +9VDC @ 5A IEC 320
Power:
type input connector.
Equipment Classification: Class IIa per MDD, Class II (Electrical Shock), Continuous mode of
Classifications:
operation, CE
± 2% ± 2%
Saturation Normal (no motion)
(70-100%Arms) (70-100%Arms)
± 3%
Saturation Motion N/A
(70-100%Arms)
For more detailed information, see the instructions for use included with each device.
36 to ~39°C ± 0.2°C
39.1 to ~42.5°C
± 0.3°C
102.4 to ~108.5 °F
Limited Warranty
SunTech Medical, Inc. provides to the original purchaser the following limited warranty from date of invoice.
SunTech Medical, Inc. warrants each instrument to be free from defects in material and workmanship. Liability under
this warranty covers servicing of the instrument when returned from the customer’s facility within the United States
prepaid to the factory. SunTech Medical, Inc. will repair any component(s) or part(s) that it finds to be defective
during the period of this limited warranty. Should a defect become apparent, the original purchaser should first notify
SunTech Medical, Inc. of the suspected defect. The device should be carefully packaged and shipped prepaid to:
The device will be repaired in the shortest possible time and returned prepaid by the same shipping method as
received by the factory. This limited warranty is void if the device has been damaged by accident, misuse, negligence,
act of God or serviced by any person not authorized by SunTech Medical, Inc. This limited warranty contains the
entire obligation of SunTech Medical, Inc. and no other warranties expressed, implied or statutory are given. No
representative or employee of SunTech Medical, Inc. is authorized to assume any further liability or grant any further
warranties except as herein.
Conflict Minerals
SunTech’s current Conflict Minerals Statement can be accessed at:
http://www.suntechmed.com/pricing-support/customer-support/conflict-minerals-statement