EMC Pre Compliance Testing
EMC Pre Compliance Testing
What to do?
Besides applying best practice with respect to all EMC related aspects
of the design, it is essential to carry out pre-compliance testing
Products:
5µH LISN for conducted emission testing of DC powered equipment
TEM cell for radiated emission and immunity testing
EMC probe set to localize the origin of emissions or to localize the origin of immunity issues
The DUT shall be isolated and elevated from the groundplane. The spectrum
analyzer shall measure the conducted emissions on both supply lines. The RF output of
the unused LISN shall be terminated with 50 Ohm. If the power return line is less
than 200 mm, a set up with a single LISN is sufficient. For the exact details of the set up
and for the limits of conducted emissions refer to the CISPR 25 or relevant
applicable standard.
Conducted emission testing
Setup for conducted emission testing of AC-powered equipment (e.g.CISPR16)
The DUT shall be isolated and elevated from the groundplane. The spectrum
analyzer shall measure the conducted emissions on both line and neutral.
The value of the parallel combination of the capacitors is 12µF from line and neutral to ground.
This causes around 0.75A flowing into the earth connection and would trip the ground fault switch.
Hence, an insulation transformer is required and good grounding is essential for safety.
For the exact details of the set up and for the limits of conducted emissions
refer to the CISPR 16 or other relevant applicable standard. A built in limiter or
a limiter between the LISN RF output and spectrum analyzer is essential for the protection
of the equipment.
Radiated emission testing
Setup for radiated emission testing
Terminate any of the two ports of the TEM cell with the 50Ω load, connect the DC-Block
to the other port and connect it to the input of the spectrum analyzer. Place the DUT
under the septum and power it on.
Monitor the radiated spectrum.
As a rule of thumb, given that the PCB is positioned not much higher than 1-2cm
above the bottom wall, any spurious with amplitudes higher than 40dµV may
potentially cause a failed compliance test.
Use compliance tested DUTs and the corresponding test reports as a reference for
measurements in the TEM cell.
Localizing the source of emissions
Localize the source of unwanted emissions on the DUT PCB using near field probes
and a spectrum analyzer.
Implement counter measures and re-test the DUT inside the TEM cell to verify if the
modifications result in a reduction of the radiated emissions
RF immunity testing
Setup for RF immunity testing
Terminate any of the two ports of the TEM cell with the 50Ω load, and connect the
signal generator + power amplifier to the other port. The signal needs to be AM
modulated (80%) according to most standards. The power of the signal has to be
chosen according to the required field strength. Place the DUT under the septum and
power it on.
Sweep the signal generator and monitor the DUT for RF immunity issues.
The picture above shows a minimum setup. Inserting a directional coupler with
connected RF power meter in between power amplifier and TEM cell would be of
advantage to monitor power levels.
Localizing the origin of immunity issues