Calibration Methods For Vector Network Analyzer Calibration Methods
Calibration Methods For Vector Network Analyzer Calibration Methods
978−1−4799−4269−5/14/$31.00
2014
c IEEE 810
Solving measured S-parameters from Fig.s 1 and 2 [2] TABLE I
PORT 1 CALIBRATION SUMMARY
A. VNA Calibration
TABLE II
The calibration procedure consists in measuring 7 PORT 2 CALIBRATION SUMMARY
different reference standards (2 Opens, 2 Shorts, 2 Matches
and a Thru) with known reflection and/or transmission Reference
Error to be
Description
values from a TOSM calibration kit. In this paper reference corrected
standards are considered to have ideal values as follows Open2 e′11 Source Match (R)
Short2 e′23e′32 Reflection Tracking (R)
ΓOPEN = 1 (6)
Match2 e′33 Directivity (R)
ΓSHORT = −1 (7)
1) Port 1 Calibration: Making steps 1 to 3, an OSM 4) Calibration between Ports: When making step 8 both
calibration [2] to Port 1 is performed and the following Load Match and Transmission Tracking error terms are
forward error terms are calculated from (1) calculated from (1), (2), (3) and (4) as follows
811
where III. INCOMPLETE 2-PORT VNA CALIBRATION
Δe = e00 .e11 − e01e10 (22) In the past, VNAs had only a transmission/reflection
(T/R) test set. This allowed only forward parameters to be
Δe´= e´33 .e´22 −e´23 e´32 (23) measured, since Port 1 acted as a source and Port 2 as a load.
Then, calibration methods used were:
812
Similar considerations are applied for the reverse
transmission tracking term Similar considerations are applied for reverse parameters
′ e′23 = S12M
e01 (39) ′
S 22M ( DUT ) − e33 (44)
S 22 =
′ − Δe ′
S 22M ( DUT ) ⋅ e22
813
2) Disadvantages: As only Forward (or Reverse) S- A. S11 Deviation Results
parameters are measured, Load Match value can not be used Maximum deviations of |S11| for 1-P+N and ER methods
for correcting DUT S-parameters. respect to TOSM are shown in Fig. 3 for different values of
|S21|. Since both of these incomplete methods calculate S11
in the same way (see equations (42) and (46)), such
VII. ONE-PATH 2-PORT deviations with respect to TOSM are the same. These
deviations depend on Port 2 Load Match, i.e. e22, and the
Originally named One-Path Full 2-Port, this method was DUT´s attenuation, i.e. S21.
introduced to T/R VNAs in order to measure all four S- For example, if the DUT is a 6-dB attenuator and |e22| =
parameters. However, the DUT must be manually reversed 0.1, then the maximum deviation with respect to TOSM will
to measure Reverse S-parameters . be 0.026.
At present most VNAs have this calibration option, but If now the DUT consists of a coaxial cable with nominal
special care must be taken as some manufactures call the 0 dB attenuation, and |e22| remains in 0.1, then maximum
ER method as 1-P 2-P. deviation respect to TOSM method rises to 0.100.
A. VNA Calibration
This method considers Forward and Reverse models the
same as follows
′
e00 =e33 (50)
e11 = e′22 (51)
e10 e01 = e′23 e32
′ (52)
′ e′23
e10 e32 = e01 (53)
′
e22 =e11 (54)
′
e30 = e03 (55)
VIII. SIMULATIONS
A series of comparisons between incomplete calibration
methods with respect to TOSM were carried out. S11 and S21
measurements were simulated and maximum deviation
results are shown in Fig. 3 to 6. In these simulations, the
TOSM case represents an ideal measurement of the DUT, Fig. 4 |S21|dB maximum deviation for Transmission Response with respect
to TOSM method when measuring a S21 value of 0 dB.
since all the error terms are perfectly determined and used
to correct the S-parameters.
814
be used to estimate the maximum systematic errors for all
the cases.
In practice, if VNA´s error terms and/or DUT´s
mismatches are quite low, there will be no significant
deviation between incomplete and complete calibration
methods when measuring S11 or S21. However, if VNA´s
source and load match error terms are high enough and the
DUT is lossy, then incomplete methods may not be suitable
since significant deviations may occur.
REFERENCES
[1] AN 1287-3, “Applying error correction for VNAs”, 2nd. Ed., Santa
Clara, CA: Agilent Tech., 2002.
[2] D. Rytting, “Network analyzer error models and calibration
methods”, Palo Alto, CA, Hewlett Packard Inc., 1998.
Fig. 5 |S21|dB maximum deviation for Enhanced Response with respect to [3] B. Hall, “VNA error models: Comments on EURAMET/cg-2/v.01”,
TOSM method (in this case the deviation does not depend on the S21 value). ANAMET Report 051, Measurement Standards Laboratory of New
Zealand Lower Hutt, New Zealand, 2010.
For example, if |S11| = |S22| = 0.1, and |e11| = |e22| = 0.1, [4] J. Dunsmore, “Handbook of microwave component measurement”,
then the maximum deviation with respect to TOSM will be Agilent Tech., John Wiley & Sons, Ltd., UK, 2012.
0.17 dB for TR method and 0.09 dB for ER method. [5] M. Hiebel, “Fundamentals of vector network analysis”, Rohde &
Schwarz GmbH & Co. KG, 5th Ed., 2011.
According to Fig. 5 and 6, if now the same DUT has an
[6] G. Wübbeler, Clemens Elster, Thomas Reichel and Rolf Judaschke,
attenuation value of 6 dB and all other values remain the “Determination of the complex residual error parameters of a
same, then maximum deviation respect to TOSM method calibrated one-port vector network analyzer”, IEEE Transactions,
rises to 0.24 dB for TR method and remains in 0.09 dB for Vol. 58, No. 9, 2009.
ER method. [7] AN 1287-11, “Specifying calibration standards and kits for Agilent
vector network analyzers”, Agilent Tech., 2011.
IX. CONCLUSIONS
815