AFA Guide
AFA Guide
Calibration Guide
Airflow Monitor
AFA1000/1 MK2
Startup Calibration
The AFA1000 must be field-calibrated once the 1. Press Enter from the “Requires set up” screen or if the monitor is in
room air supply and exhaust is balanced. When the Run screen Press and Hold the Enter button for 5 seconds until
the unit is powered up, the following sequence the Main Menu is displayed.
of events occurs:
2. Using the + / - buttons select SET UP, then select CALIBRATION,
1. The alarm performs a self-test of its then enter the password (factory default password is 0-0-0-0, press
functions, LEDs and audible alarm Enter to continue.
(approximately 3 seconds)
3. Open the sash to the normal operating height and measure the
2. At the end of the delay, the unit will do one face velocity using a calibrated instrument. Using the + / -buttons
of two things: enter the measured face velocity then press Enter, the monitor will
a. If the monitor has been calibrated, the then sample the airflow for 5 seconds.
unit enters normal operating mode (displays
4. If the airflow sample is unstable the monitor will display “Deviations
airflow velocity, LED’s and audible alarm
too High”, follow the instructions to repeat the sample or quit the
enabled).
calibration.
b. If the unit has not been calibrated, the
monitor will display “ Requires set up, press 5. If the sample is accepted, lower the sash by half and measure the
Enter to continue” face velocity using a calibrated instrument. Using the + / - buttons
enter the measured face velocity and press Enter, the monitor will
then sample the airflow for 5 seconds.
6. If the calibration is successful the monitor will return the Main
Menu, select RUN to go to normal operating mode.
7. If the airflow sample is unstable the monitor will display “Deviations
too High”, follow the instructions to repeat the sample or quit the
calibration.
continued over...
Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK
T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com
Installation &
Calibration Guide
8. The monitor will display “Increase higher airflow sample” 4. Avoid movement in front of the Fume Cupboard whilst
if the second sample value is too close to the first value calibrating the monitor.
entered, close the sash a little and repeat the higher
5. The Low and High air samples must be at least 0.3m/sec
sample. The minimum difference between the samples
(60 fpm) apart to calibrate the monitor. This is to avoid
that the monitor will accept is 0.3m/sec (60fpm).
inaccuracy in the calibration sue to insufficient difference
9. The monitor will display “Sensor diff too low” if the between the samples. The minimum difference can
monitor doesn’t detect any difference in the sensor be changed in the “Low High diff” parameter in the Cal
output between the 2 airflow samples, check that the Config menu.
sensor hose is connected and repeat the calibration.
6. Do not use fully open and fully closed sash positions for
the calibration points, the recommended positions are
normal operating height (e.g. 500mm 20”) for the first
Calibration Tips sample and approximately half the sash opening (e.g.
250mm 10”) for the second sample.
1. Ensure that the sensor hose is connected correctly at the
side wall and rear of the sensor. 7. If the Fume Cupboard is under separate VAV control
then a different set up may be required to achieve
2. Ensure that the extract fan is running and the Fume the airflow difference required for calibration. Normal
Cupboard is balanced before calibrating the monitor. extract for the low sample and Purge for the high sample
3. Take extra time (at least 15 to 20 seconds) to wait for the may be used.
airflow to settle before capturing the airflow samples.
This will minimize the chance of a calibration error due
to turbulence or fluctuations.
Troubleshooting
Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK
T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com
Installation &
Calibration Guide
Connection Details
Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK
T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com
Installation &
Calibration Guide
1.57” 1.57”
(40mm) (40mm)
3.46” 3.15”
(88mm) (80mm)
SM6 Sensor Dimensions (rear view) SM6 Panel Cut Out Dimensions
0.47” 0.47”
(12mm) (12mm)
1.18” (30mm) Dia. Sensor
Hole for Vent Tube
0.157” (4mm)
0.157” (4mm)
Tapped
Tapped
Fixing Hole
Fixing Hole
1.45”
(37mm)
2.91”
(74mm)
0.39” 0.79”
(10mm) (20mm)
1.45”
(37mm)
0.67”
(17mm) 1” 1”
(25mm) (25mm)
1.06” 1.06”
1.45” 1.45” (27mm) (27mm)
(37mm) (37mm) 0.47” (12mm) Dia.
Hole for SM6 cable
2.91”
(74mm)
Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK
T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com
Installation &
Calibration Guide
Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK
T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com