Fiber Optic Laryngoscope Handles Directions For Use
Fiber Optic Laryngoscope Handles Directions For Use
Fiber Optic Laryngoscope Handles Directions For Use
CAUTION: Only use lamp specified. Failure to follow these instructions may cause
damage or poor performance of the handle.
CAUTION: If the device will be unused for several months or longer, remove the
batteries prior to storing the device.
2 Directions for use Welch Allyn fiber optic laryngoscope handles
Reprocessing instructions
These reprocessing instructions refer to procedures for cleaning and intermediate level
disinfection. Fiber optic laryngoscope handles must be reprocessed prior to first use and
between each use using the following method as outlined in this document:
Welch Allyn has validated the above instruction as being capable of preparing these
laryngoscope handles for re-use. The user must ensure that the reprocessing as actually
performed by the user's personnel, with the user's equipment and materials, achieves the
desired result. This may require validation and routine monitoring of the user's actual process
NOTE: The main handle and bottom cap components are compatible with the disinfection
solution and autoclave methods identified which are provided for facilities who wish to perform
either method after cleaning and intermediate level disinfection.
Point of use: 1. Separate blade assembly from handle and place handle
into suitable containment for subsequent reprocessing
(per figure 1). Do not place handle with sharp devices.
2. Prevent the handle from drying (i.e. wrap/cover in moist
germicidal wipe).
Preparation for 1. Select an appropriate quaternary ammonium
decontamination: isopropanol based germicidal cleaner labeled suitable
for use on healthcare equipment and capable
intermediate level disinfection. Reference EPA-
registered disinfectants:
http://www.epa.gov/oppad001/chemregindex.htm
Outside of the U.S., please consult applicable
regulatory body for equivalent quaternary ammonium
isopropanol germicidal cleaner.
Autoclave instructions
NOTE: The main handle and bottom cap are compatible with these autoclave methods which
are provided for facilities who wish to autoclave after cleaning and intermediate level
disinfection.
Maintenance instructions
CAUTION: Use only Welch Allyn replacement lamps REF: 06000 to ensure
proper illumination alignment.
1. Unscrew bottom cap of handle counterclockwise and remove batteries per figure 3.
2. Remove lamp cartridge assembly from main handle by applying finger pressure in the
direction shown by the arrow per figure 4.
3. Remove shroud cap from shroud by rotating shroud cap counterclockwise per figures 5
and 6.
4. Rotate lamp counterclockwise to remove per figure 6.
5. Replace lamp and rotate shroud cap clockwise to tighten per figure 7.
6. To replace lamp cartridge assembly in the main handle, invert handle, then gently slide
the cartridge down the inside of handle, tipping it side to side until the shroud cap
exits opening on top.
7. Re-insert batteries and apply slight pressure to set the cartridge in place.
8. Replace and tighten bottom cap.
9. Reprocess repaired assembly as appropriate per these instructions.
Figure 3 Figure 4
Figure 5 Figure 6
1. Shroud cap
2. Lamp
3. Shroud
Figure 7
Directions for use Welch Allyn standard laryngoscope handles 7
Specifications
Electrical:
For information about electromagnetic compatibility (EMC) see Welch Allyn website:
http://www.welchallyn.com
Operating: Storage/Transport:
104 F 120 F
(40 C) (49 C)
32 F -4 F
(0 C) (-20 C)
Approvals:
Conforms to ASTM F 965 and ISO-7376-1, IEC/EN 60601-1, IEC/EN 60601-1-2
The CE mark on this product indicates that it has been tested to and conforms
with the provisions noted within the 93/42/EEC Medical Device Directive.
Complies with EMC Framework of Australia
Warranty:
One year
Service Information:
For Technical Support or to obtain information about any Welch Allyn product, contact
Welch Allyn Technical Support: www.welchallyn.com/support.