The TQC Automatic Cupping Tester is a device that performs cupping (Erichsen) tests on coated metal panels to test the adhesion of coatings like paint and varnish. It applies a gradual, precise indentation at a controlled speed, allowing observation of when and how the coating cracks or detaches. Key features include adjustable LED lighting and a user-friendly interface. It is used in coating industry labs to test compliance with standards like ISO 1520.

Uploaded by

Rodrigo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
117 views3 pages

Automatic Cupping Tester

The TQC Automatic Cupping Tester is a device that performs cupping (Erichsen) tests on coated metal panels to test the adhesion of coatings like paint and varnish. It applies a gradual, precise indentation at a controlled speed, allowing observation of when and how the coating cracks or detaches. Key features include adjustable LED lighting and a user-friendly interface. It is used in coating industry labs to test compliance with standards like ISO 1520.

Uploaded by

Rodrigo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

TQC AUTOMATIC CUPPING TESTER

SP4300 / SP4305

PRODUCT DESCRIPTION

TQC Automatic Cupping Tester to perform a cupping (Erichsen / Dent) test on


coated steel panels to define the resistance of paint, varnish or related products to
cracking and/or detachment from a metal substrate when subjected to gradual
deformation by indentation under standard conditions.

The test is either used as a “pass / fail” test by testing to a specified depth or
defining the minimum depth at which a coating fails by gradually increasing the
indentation.

The ISO1520 standard requires panels to be slowly deformed at a steady rate between 0,1 mm/s and 0,3 mm/s
without interruption. Especially with thicker steel panels hand-operated testers not always allow an
uninterrupted deformation.

The TQC Automatic Cupping Tester is driven by a micro-step controlled electro motor which allows precise
and steady deformation with 0,01 mm steps. Operation is intuitive by means of a jog-dial switch and a multi-
lingual operating menu on a large illuminated display.

An integrated LED powered sample illumination system comforts examining the coating under test. To
guarantee maximum visibility of all possible types of surface including high gloss, matte or colored samples
the angle of the LED light can be set. Choose light from just one or all directions. The strength of the LED’s is
adjustable but also the colors can be changed to achieve maximum contrast.

Mandatory test in Qualicoat and QIB accredited laboratories.

BUSINESS

Coating industry, industrial finishing, laboratory

STANDARDS

ISO 1520, DIN EN NF 1520, DIN 53156, DIN 53232, BS 3900-E4, NBN T22-104, NFT 30 019.
Look up the appropriate standard for a correct execution of the test.

FEATURES

• easy-to-use
• user calibration possible
• integrated multi colour led lighting
• Manual and preset mode
• Multiple languages
• Panel holder adjustable in angle
SCOPE OF SUPPLY

• Automatic Cupping Test


• Reference panel
• Optical tool support rod
• Optical tool mount
• Power cord
• manual

ART NO.

SP4300 TQC Automatic Cupping Test set for 240V


SP4305 TQC Automatic Cupping Test set for100V

ACCESSORIES

LD6181 TQC USB digital microscope ‘Full kit”

SPECIFICATIONS

Technical Data
Indenter Speed: 0.01 – 0.70 mm/s
Stroke length: 0 - 12 mm
Max panel width: Max 100mm
Max. panel thickness steel: max. 0.8 mm
Max. panel thickness aluminium: max. 1.2 mm

Dimensions and Weight


Depth: 450 mm
Width: 350 mm
Height: 600 mm
Net weight: approx. 31 kg

Basic Unit
Power Supply: 115 – 230 V, 50 - 60 Hz
Power consumption: max. 80 Watt
Display: Blue Illuminated, graphic 100 x 35 mm, 193x64 pixels
Safety: Emergency Button, integrated Acoustic Alarm
Function: Jog Shuttle knob by Rotation / Pushing

Accuracy
Indenter Speed accuracy: +/- 1% of set speed
Stroke length accuracy: +/- 0.01 mm
USE

The SP4300 / SP4305 TQC Cupping Test has a menu-driven interface and an integrated calibration function.
Check the manual for full details.

SPECIAL CARE

• Though robust in design, this instrument is precision-machined. Never drop it or knock it over
• Always clean the instrument after use.
• Clean the instrument using a soft dry cloth. Never clean the instrument by any mechanical means such as
a wire brush or abrasive paper. This may cause, just like the use of aggressive cleaning agents, permanent
damage.
• Do not use compressed air to clean the instrument.
• Never perform repairs or service to the instrument yourself. This should be done by TQC or selected
distributors.

SAFETY PRECAUTIONS

• Always make sure the instrument is connected to an earthed socket.


• Maintenance and inspection should be carried out at the correct intervals
• Operating personnel should be informed before starting with maintenance or repair work
• Always make sure the instruments power is turned off and the instrument is not connected to a socket
while adjusting any electrical component whenever maintenance, inspection or repair work is done.
• Do not open the instrument. In case of malfunction always consult the manufacturer.

DISCLAIMER

The right of technical modifications is reserved.

The information given in this sheet is not intended to be exhaustive and any person using the product for any
purpose other than that specifically recommended in this sheet without first obtaining written confirmation
from us as to the suitability of the product for the intended purpose does so at his own risk. Whilst we
endeavour to ensure that all advice we give about the product (whether in this sheet or otherwise) is correct
we have no control over either the quality or condition of the product or the many factors affecting the use
and application of the product. Therefore, unless we specifically agree in writing to do so, we do not accept
any liability whatsoever or howsoever arising for the performance of the product or for any loss or damage
(other than death or personal injury resulting from our negligence) arising out of the use of the product. The
information contained in this sheet is liable to modification from time to time in the light of experience and
our policy of continuous product development.

You might also like