Automatic Cupping Tester
Automatic Cupping Tester
SP4300 / SP4305
PRODUCT DESCRIPTION
The test is either used as a “pass / fail” test by testing to a specified depth or
defining the minimum depth at which a coating fails by gradually increasing the
indentation.
The ISO1520 standard requires panels to be slowly deformed at a steady rate between 0,1 mm/s and 0,3 mm/s
without interruption. Especially with thicker steel panels hand-operated testers not always allow an
uninterrupted deformation.
The TQC Automatic Cupping Tester is driven by a micro-step controlled electro motor which allows precise
and steady deformation with 0,01 mm steps. Operation is intuitive by means of a jog-dial switch and a multi-
lingual operating menu on a large illuminated display.
An integrated LED powered sample illumination system comforts examining the coating under test. To
guarantee maximum visibility of all possible types of surface including high gloss, matte or colored samples
the angle of the LED light can be set. Choose light from just one or all directions. The strength of the LED’s is
adjustable but also the colors can be changed to achieve maximum contrast.
BUSINESS
STANDARDS
ISO 1520, DIN EN NF 1520, DIN 53156, DIN 53232, BS 3900-E4, NBN T22-104, NFT 30 019.
Look up the appropriate standard for a correct execution of the test.
FEATURES
• easy-to-use
• user calibration possible
• integrated multi colour led lighting
• Manual and preset mode
• Multiple languages
• Panel holder adjustable in angle
SCOPE OF SUPPLY
ART NO.
ACCESSORIES
SPECIFICATIONS
Technical Data
Indenter Speed: 0.01 – 0.70 mm/s
Stroke length: 0 - 12 mm
Max panel width: Max 100mm
Max. panel thickness steel: max. 0.8 mm
Max. panel thickness aluminium: max. 1.2 mm
Basic Unit
Power Supply: 115 – 230 V, 50 - 60 Hz
Power consumption: max. 80 Watt
Display: Blue Illuminated, graphic 100 x 35 mm, 193x64 pixels
Safety: Emergency Button, integrated Acoustic Alarm
Function: Jog Shuttle knob by Rotation / Pushing
Accuracy
Indenter Speed accuracy: +/- 1% of set speed
Stroke length accuracy: +/- 0.01 mm
USE
The SP4300 / SP4305 TQC Cupping Test has a menu-driven interface and an integrated calibration function.
Check the manual for full details.
SPECIAL CARE
• Though robust in design, this instrument is precision-machined. Never drop it or knock it over
• Always clean the instrument after use.
• Clean the instrument using a soft dry cloth. Never clean the instrument by any mechanical means such as
a wire brush or abrasive paper. This may cause, just like the use of aggressive cleaning agents, permanent
damage.
• Do not use compressed air to clean the instrument.
• Never perform repairs or service to the instrument yourself. This should be done by TQC or selected
distributors.
SAFETY PRECAUTIONS
DISCLAIMER
The information given in this sheet is not intended to be exhaustive and any person using the product for any
purpose other than that specifically recommended in this sheet without first obtaining written confirmation
from us as to the suitability of the product for the intended purpose does so at his own risk. Whilst we
endeavour to ensure that all advice we give about the product (whether in this sheet or otherwise) is correct
we have no control over either the quality or condition of the product or the many factors affecting the use
and application of the product. Therefore, unless we specifically agree in writing to do so, we do not accept
any liability whatsoever or howsoever arising for the performance of the product or for any loss or damage
(other than death or personal injury resulting from our negligence) arising out of the use of the product. The
information contained in this sheet is liable to modification from time to time in the light of experience and
our policy of continuous product development.