MiniElcor Eng
MiniElcor Eng
miniELCOR
Device Description
Operation Manual
Technical Description
Mounting instructions
Device settings
January 2010
Rev. 6g
Safety Measures
This measurement device can be operated only by an operator trained in
compliance with the technical terms, safety regulations, and standards. It is
necessary to consider any other legal and safety regulations stipulated for special
applications. Similar measures also apply for special applications. Similar measures
also apply for using the accessories. The operator training must be in compliance
with Decree no. 50.1978 Coll.
The information in this manual does not have the burden of a legal obligation from the
manufacturer’s side. The manufacturer reserves the right to implement changes. Any
changes in the manual or in the product itself can be performed at any time without any
previous alert, with the goal of improving the device or fixing any typographical or technical
mistakes.
TABLE OF CONTENTS
1 Introduction ............................................................................................. 1
1.1 Basic device description ........................................................................................ 1
1.2 Function principle .................................................................................................. 2
1.3 Device dimensions ................................................................................................. 7
2 Device technical description ..................................................................... 7
2.1 Device architecture ............................................................................................... 7
2.2 Device power supply ............................................................................................. 8
2.3 Security marks ..................................................................................................... 11
2.4 Product label ....................................................................................................... 13
3 Safety instructions ................................................................................... 14
3.1 General ............................................................................................................... 14
3.2 Use in potentially explosive atmospheres ........................................................... 14
3.3 Risks of usage ...................................................................................................... 14
3.4 Special conditions of use ..................................................................................... 15
3.5 Using device variants for different groups of gas ................................................. 15
4 Metrology characteristics ........................................................................ 16
4.1 Measuring temperature ...................................................................................... 16
4.2 Measuring pressure ............................................................................................. 16
4.3 Compressibility calculation .................................................................................. 17
4.4 Volume measuring and calculation ...................................................................... 18
5 Connecting inputs and outputs ................................................................ 21
5.1 Inputs .................................................................................................................. 21
5.2 Outputs ............................................................................................................... 25
5.3 Adding of another pressure or temperature transducer ...................................... 26
6 Communication with device .................................................................... 29
6.1 RS‐232 and RS‐485 interfaces .............................................................................. 29
6.2 Optical interface IEC‐1107 ................................................................................... 31
7 Description of function ............................................................................ 33
7.1 Measurand marking ............................................................................................ 33
7.2 Instantaneous values ........................................................................................... 33
7.3 Archives .............................................................................................................. 34
7.4 Device parameterization ..................................................................................... 37
7.5 Other device functions ........................................................................................ 38
7.6 Securing the device against a change of metrology parameters .......................... 38
8 Putting in operation ................................................................................ 44
9 Device operation ..................................................................................... 45
9.1 Keypad ................................................................................................................ 45
9.2 Menu system ....................................................................................................... 46
9.3 Main menu .......................................................................................................... 49
9.4 Instantaneous values menu ................................................................................. 50
9.5 Stored values menu ............................................................................................. 50
9.6 Device parameters menu ..................................................................................... 51
9.7 Parameter settings menu .................................................................................... 52
9.8 System data menu ............................................................................................... 53
9.9 Diagnostics menu ................................................................................................ 54
10 Mounting instructions ............................................................................. 57
10.1 Mechanical mounting of the device ..................................................................... 57
10.2 Cable connection, grounding ............................................................................... 61
11 Accessories .............................................................................................. 63
11.1 Assembly accessories .......................................................................................... 63
11.2 Intrinsically safe supply sources for external power supply ................................. 63
11.3 Separation and communication modules ............................................................ 63
11.4 GPRS communicators .......................................................................................... 63
11.5 Other accessories ................................................................................................ 63
12 Technical parameters .............................................................................. 64
13 Inexplosiveness parameters .................................................................... 70
14 Device setting .......................................................................................... 72
14.1 Standard device control after installation ............................................................ 72
14.2 Device connection with PC .................................................................................. 73
14.3 Setting of communication between device and PC .............................................. 73
14.4 Password in the device ........................................................................................ 84
15 Configuration examples ........................................................................... 86
15.1 Device parameters displaying .............................................................................. 86
15.2 Gas meter constant setting .................................................................................. 86
15.3 Pulse outputs setting ........................................................................................... 89
15.4 Analogue output setting ...................................................................................... 94
15.5 Setpoint setting – limit values of measured quantity ........................................... 97
15.6 Setting of external power supply failure ............................................................. 100
15.7 Setting of communication through MODBUS protocol ....................................... 101
16 Pressure and temperature sensor/transducer replacement .................. 105
16.1 Pressure and temperature sensor/transducer replacement procedure in
miniELCOR device ............................................................................................... 105
16.2 Software settings of device for proper communication with new temperature
sensor ................................................................................................................. 105
16.3 Software settings of device for proper communication with new pressure
transducer .......................................................................................................... 108
17 Software settings of the device for proper communication with external
digital temperature (EDT‐34) or pressure transducer (EDT‐23) .............. 109
17.1 Adding of digital transducer into device’s parameters ........................................ 109
17.2 Adding of quantity measured by digital transducer into device’s archives .......... 110
18 Final verification of the device after replacement of sensor/transducer or
adding of digital transducer ................................................................... 111
19 What to do if something does not work ................................................ 116
20 Literature .............................................................................................. 119
21 Relevant Literature ................................................................................ 119
22 Software ................................................................................................ 120
23 Used trade marks .................................................................................. 120
24 List of figures ......................................................................................... 121
25 List of Tables ......................................................................................... 123
miniELCOR
Used symbols and definitions
1
miniELCOR
Qm ... Flowrate at measurement conditions ( further primary m3/h
flowrate)
Qb ... Flowrate at base conditions m3/h
T ... Absolute temperature at measurement conditions (T = t + K
273.15)
t ... Gas temperature °C
Tb ... Absolute temperature at base conditions K
V ... Volume Vm or Vc
Vm ... Volume at measurement conditions (further primary m3
volume)
Vc ... Corrected volume at measurement conditions ( volume m3
corrected based on correction curve of gasmeter)
Vb ... Volume at base conditions (hereinafter also the m3
standardized volume)
Vbs ... Error volume at base conditions (hereinafter also the m3
error standardized volume)
Vs ... Error volume at measurement conditions (hereinafter m3
also the error operational volume)
Vd ... Difference of primary volume m3
Vbd ... Difference of base volume m3
Vf ... Tariff pulse counter of primary volume
Vbf ... Tariff pulse counter of base volume
Z ... Compression gas factor at measurement conditions
Zb ... Compression gas factor at base conditions
Notice :
This handbook issue describes device functions with firmware FW 4.xx which is
compatible with previous firmware version 2.xx. All different features will be
mentioned respectively.
Chapters describing new device features of FW ver. 4.xx are marked with (*).
2
miniELCOR
1 Introduction
1.1 Basic device description
The Gas-volume conversion device miniELCOR (hereinafter only “the device”)
is a measuring instrument designed for the conversion of the gas volume measure at
measurement conditions to volume at base conditions.
The information on the gas volume passing through is measured using the
impulse outputs of the gas meter. The gas temperature and pressure are measured
by integrated converters. The device calculates the ratio of compressibility factors of
gas using standard methods or a constant value is used.
The device has been constructed and approved pursuant to the EN 12405-1
standard as a conversion device type 1 (compact system) and can be supplied as a
T, PT, or PTZ conversion device.
From safety point of view device is constructed according to EN 60079-11 like
intrinsic safe.
It is manufactured and supplied in compliance with the following European
Parliament directives:
1994/9/EC Equipment and protective systems for use in potentially explosive
atmospheres
2004/108/EC Electromagnetic compatibility
2004/22/EC Directive on measuring instruments
Device is put onto market and into usage according to above mentioned
standards and is marked with CE mark.
The device is built in a casing with sturdy plastic with IP66 protection. It is
equipped with a graphic display and a 6-button keypad. Furthermore, it has impulse
inputs for the connection of a gas meter with LF or HF impulse output and binary
inputs. Device with FW version 4.xx and higher is suitable for connection via encoder
NAMUR or SCR. If applied encoder SCR the only miniELCOR SCR1 variant is
allowed. The binary inputs can work as check inputs to check the connection with a
gas meter or can have a different function, e.g. monitoring the conditions of safety
snap locks, doors, etc. The device has 4 available outputs. These can be configured
as impulse or binary outputs, or as data outputs for the CL-1 module. When using
this module, an analog current output can be realized.
The device is powered by a lithium battery. The life cycle of the battery is 6
years in the defined work mode. In the case of a battery power supply, one can also
use the impulse outputs. An external power supply source can be used in
applications with higher demands.
The device has a data archive of the measured values with an adjustable
structure and storing period. The binary archive stores changes on the binary inputs
and the occurrence of the monitored events (limits, etc.) Error conditions are stored
in an status archive. It is possible to program the storing of important quantities and
calculations and storage of some statistical values in the daily and monthly archive.
The archive has settings for service and metrology; in case of changes of settings,
the acts influencing the device parameters are recorded. The other logs are available
as well , see more in 7.3.
1
miniELCOR
For communication with its superior system, the device has a serial interface
RS-232 and RS-485. Various communication protocols installed in the device allow
easier connection to the SCADA systems. The device cooperates with common
phone, radio, GSM, and GPRS modems, and in case of an alarm condition, it can
initiate the connection.
The device can be enhanced by one non-metrology converter for measuring
pressure or temperature. This enhancement can be performed without breaking the
official mark on an already installed device.
- The same like at basic device variant ( see ad 1) however without possibility
of connection of digital transducer EDTxx
- 1x input for SCR encoder connection by means of extention board KP 065 09
The device can be configured using the supplied SW [22] for PCs. This SW also
allows the readout, display and archive of both the immediate measured values as
well as the contents of the internal device archives.
2
miniELCOR
Volume at measuring conditions (operational volume):
N
V=
kp
Ratio of compressibility factor:
Z
K=
Zb
Conversion factor:
p Tb 1
C= * *
pb (t + 273.15) K
Volume at base conditions (standardized volume):
Vb = V * C
Gas compressibility factor expresses the deviation of properties of natural gas
from the properties of an ideal gas. By setting the parameters, it is possible to choose
a specific method for calculation of the compressibility factor pursuant to the standard
(AGA NX-19 mod, AGA8-G1, AGA8-G2, SGERG-88 or AGA8-92DC). A constant
compressibility value can be used for other gases besides natural gas. If the
pressure or temperature value gets out of the limits of validity of the chosen standard
for calculation of compressibility, the device calculates using a default compressibility
value.
The device calculates the gas flow from the impulse frequency on the input in
real time using mathematical filtration from the input signal.
Operational flow:
Q = ∆V / ∆t [m3/h]
3
miniELCOR
1.2.3 Volume correction at measurement conditions
Device enables to compensate gasmeter error according to predefined
correction curve from gasmeter test certificate. This function and parameters Vc can
be activated only by manufacturer or by Acreditive service to ensure that used
gasmeter correction curve in dependance on flowrate Qm is valid within working
conditions.
Vc = Vm x f(Qm)
where
Vc ... Corrected volume at measurement conditions
Vm ... Primary volume
Qm ... Primary flowrate
Two other counters ( energy counter E and estimated energy counter Es) are
dedicated for measurement in configurable energy units: MJ, kWh, Btu.
Note :
No conversion of absolute counter value (E or Es) is accomplished after
change of units. Following increases are added already respecting new units.
4
miniELCOR
Combustion heat Hs
To get correct conversion it is necessary to enter correct value of combustion
heat and relative conditions. Then device will make new conversion of relative
temperature for defined relative conditions and final value will be used for energy
calculation. In case of AGA8-92DC method combustion heat is not entered but
calculated directly from gas composition according to EN ISO 6976. For the other
methods value Hs (MJ/m3) must be entered manually and always under those
relative conditions:
5
miniELCOR
6
miniELCOR
Note:
If SCR encoder is required it can be arranged only by manufacturer
or by authorised service center. Those two subjects will ensure
appropriate labelling placed on housing.
7
miniELCOR
The lid of the housing contains a processor board which is protected by a cover
and secured by an official mark. The board cover has an opening for access to the
service switch. The service switch can be use to enable/disable the setting of the
device parameters using a service SW.
8
miniELCOR
Defined mode with life cycle of the supply battery of more than 5 years:
• Archiving period of the data archive 1 hr
• Communication with device 2 min/day
• Showing on the display 2 min/day
• Period of input impulses ≤10 Hz
• Measuring period 15 s
• Surrounding temperature 25 °C
• expansion board KP 065 09 ( SCR encoder ) is not used
If the device is operated with higher consumption than in the defined mode, it
is necessary to count on a more frequent replacement of the battery or use a network
power source.
9
miniELCOR
Defined mode for life cycle of back-up battery of 10 years
• Storing, temperature 25 °C
• Backed-up inputs (DI1 – DI4) not connected or connected contacts
disconnected
• Does not depend on the presence of the supply battery
Self-discharging of batteries
The back-up and supply batteries are lithium. Their capacity drops due to self-
discharging. The recommended time frame for their replacement is 10 years, even if
the battery was never connected.
10
miniELCOR
11
miniELCOR
12
miniELCOR
13
miniELCOR
3 Safety instructions
3.1 General
The device has been approved pursuant to the guideline 94/9/CE and an EC
certificate on type verification (ATEX) has been issued for it’s use in potentially
explosive atmospheres. Respecting this guideline is included in the CE compliance
notation.
14
miniELCOR
• At hazardous zones device must not be installed at places where outer
conditions could create an electrostatic charge.
• Device may be cleaned by humid wiper.
Group of gas
15
miniELCOR
4 Metrology characteristics
4.1 Measuring temperature
This device uses the PT1000 temperature sensor to measure temperature. The
temperature sensor’s connection is two-wired. The influence of the length and the
characteristics of the cable used are considered during calibration and therefore do
not influence the accuracy of the temperature measuring.
The temperature measuring range is -25 °C to +60 °C. The measuring period is
common for both the measuring of temperature and pressure and it can be custom
set at a range from 1 s to 30 s. The temperature measuring units can be adjusted.
Replacement of the temperature sensor is protected by the security mark of the
manufacturer (metrology mark) and can be performed solely at an Accredited Service
center (ASC).
During device configuration, the user must enter the constant parameter
Default temperature value. This value will be used for the calculation of
compressibility instead of the measured temperature value in the following cases:
- The value of the measured temperature deviated from the measuring
range
- An error occurred when measuring the temperature
16
miniELCOR
Method
Pressure AGA NX-19 SGERG-88 AGA8-G1 AGA8-92DC
measuring range mod AGA8-G2
80 ÷ 520 kPa -25 ÷ +60 °C -25 ÷ +60 °C -25 ÷ +60 °C -25 ÷ +60 °C
200 ÷ 1000 kPa N/A -25 ÷ +60 °C -25 ÷ +60 °C -25 ÷ +60 °C
400 ÷ 2000 kPa N/A -25 ÷ +60 °C -25 ÷ +60 °C -25 ÷ +60 °C
700 ÷ 3500 kPa N/A -10 ÷ +60 °C -10 ÷ +60 °C -25 ÷ +60 °C
1400 ÷ 7000 kPa N/A -10 ÷ +60 °C -10 ÷ +60 °C -25 ÷ +60 °C
80 ÷ 1000 kPa N/A -25 ÷ +60 °C -25 ÷ +60 °C -25 ÷ +60 °C
400 ÷ 7000 kPa N/A -10 ÷ +60 °C -10 ÷ +60 °C -25 ÷ +60 °C
Table 1 Limitation of standard validity range of compressibility calculation
Note:
At device there is applied compressibily calculation method GOST NX-19 which
is not approved by ČMI certificate.
Usage of method GOST NX-19 is limited only for temperature range from -23°C
to +60°C.
Default compressibility
For the set method during each calculation, it is checked whether the measured
pressure and temperature value are in the valid interval of the pertinent method. If
some of the values are outside the valid interval, the so called default compressibility
is used for the conversion. The value of the default compressibility must be entered
by the user during device configuration.
17
miniELCOR
18
miniELCOR
19
miniELCOR
- After returning back to correct direction counting will get blocked out into
apropriate counters (Vb, Vbs) only after reaching level of primary volume
where reversed flow was started up. Primary volume counter is equivalent
to gasmeter counter all the time.
20
miniELCOR
21
miniELCOR
but it does not ensure the counting of the impulses. The terminals for the HF NAMUR
inputs are marked HF+ and HF- (see Fig. 12).
Changing measuring units, setting the gas meter constant
The impulse inputs measuring units and the gas meter constant can be
adjusted using the service SW. The gas meter constant and S/N of gasmeter can be
also set from the device keyboard.
Number of places of counters of the hf impulse inputs
In the case of hf impulse inputs, the counter works with 9 digit places.
Encoder NAMUR
No special HW is required for NAMUR encoder usage. The only condition for
NAMUR encoder data processing is usage of IS external power supply JBZ-02 ( or
JBZ-01).
Encoder SCR
To process data from SCR encoder miniELCOR SCR type must be used
complemented with KP 065 09 board. This board must be complemented directly
only by manufacturer. Additional assembly at customer side is not allowed. Encoder
board is connected with input board via IS RS485 Bus ( there is used connector
dedicated for digital transducer connection ( see Pict.3 and Pict.4). This type of
encoder can be also used for only battery powered EVC but it is too much energy
consuming. Enduser should take care about it during parameterisation.
22
miniELCOR
23
miniELCOR
Manual setup of primary volume counter Vm is not allowed at encoder input..
24
miniELCOR
5.2 Outputs
The device has 4 digital outputs DO1 to DO4 which can be configured as
binary, impulse, or data. A data output serves for the realization of an analog output
4-20 mA using the CL-1 module which is connected to this output.
The outputs can be controlled by the device using the calculation equations
entered by the user in the device parameters (for example, it is possible to generate
outputs according to the volume of the gas flown through, indication of alarm
condition, exceeding the set limits of pressure or temperature, etc).
The device structure allows the generation of outputs even when the device is
powered solely by the battery with no effect on the battery life cycle. The outputs are
“open collector” type and are not galvanic separated. All four outputs have a joint
GND conductor.
The outputs are intrinsically safe, thus when connecting standard devices, the
devices must be connected via a safety barrier (e.g. DATCOM-K3, see Fig. 13).
Impulse outputs
The impulse outputs have adjustable width and impulse periods in folds of 0.1 s.
The debt of impulse outputs can reach max. 65535 pulses. An output constant can
also be realized in the setting equation of the output quantity.
Binary outputs
Output terminals are according to the output quantity in the connected or
disconnected condition. In the resting state, the output terminals are disconnected
(condition log.1).
Data output
The digital output configured as a data output serves for communication with
the CL-1 module. An analog output 4-20 mA can be realized using this module.
25
miniELCOR
Using the calculation equations, the value of the output can be parameterized as
proportional to pressure, flow, daily consumption, etc. The CL-1 module must be
connected to the converter via a safety barrier (DATCOM-K3).
26
miniELCOR
protocol. On account of intrisic safety must be connected transducer intrinsically safe
- “ia” type. Type of the transducer is necessary to specify in time of order.
For connection of additional digital transducer (EDT 23, EDT 34) must be
volume corrector equipped with expansion RS-485 module (KP 065 08) (see Fig. 15).
Expansion module RS-485 and additional digital transducer there are not part of
standard accesories and it is neccessary to order it separately. Module RS-485 can
be ordered additionally and by this way expand already installed device.
Digital transducer is connected to the RS-485 clamps of expansion module.
Only one digital transducer can be connected to the expansion module.
Connecting/disconnecting of transducer and also of RS-485 module can be done
only when power supply is disconnected.
Procedure of connecting expansion module RS-485 and digital transducer
1. Disconnect volume corrector from external power supply ( if present)
2. Open the device and remove battery
3. Unscrew plastic cover of input/output board in place of plugging
expansion module RS-485 (factory seal will be breached)
4. Insert expansion module in to the X4 board of inputs. After inserting of
the module there it is neccessary to check if some connector pin is not
out of the contact tube. All pins must be inserted in to the connector
properly
5. Apply cover delivered with the expansion module and screw the
expansion board with input/output board
6. Connect digital transducer. Cable of the transducer pull through the cable
bushing. Shielding of the cable attach with body of the bushing. Electrical
scheme of connecting expansion module RS-485 is shown on Fig. 14.
7. Check digital transducer connection
8. Connect device back to the power. It is inserting of battery and external
power supply connection (if present)
After installation of digital transducer it is important to add it in to the parameters
by the help of service software (see paragrapph 17).
27
miniELCOR
28
miniELCOR
29
miniELCOR
30
miniELCOR
MTL 5051 setting
Switches OFF ON meaning
SW1a X ‐‐‐ other modes
SW1b X ‐‐‐ 5V output
SW2a RS232 RS422 output interface
SW2b RS422 RS232 output interface
Note: Communication output from MTL5051 can be RS-232 or RS-422
31
miniELCOR
from the economy mode to the mode in which it is able to accept data. It remains in
this mode for 180 s from the last communication (timeout) or until the user takes the
optical head of the communication interface.
Warning:
After applying the head, the communication channel from the RS-232/RS-485
device to the optical interface. That means that the communication via the RS-232 or
RS-485 is discontinued until the moment the optical head is removed, or until the
mentioned timeout from the last communication expires.
The communication speed of the optical interface can be set in the device
parameters independently of the speed of the RS-232/RS-485 interface. The setting
of the communication protocol is combined for all three interfaces.
32
miniELCOR
7 Description of function
The options of the device regarding displaying the data on the display and
storing the quantities are extremely variable and customizable. The user has full
control over which quantities will be displayed in the instantaneous values and also
which quantities will be stored in the individual archives.
33
miniELCOR
• Error operational volume Vms
• Standardized volume Vb
• Error standardized volume Vbs
• Flow Qm
• Standardized flow Qb
• Conversion factor C
• Compressibility ratio K
• Device error
• External power supply presence
• Battery capacity
• Internal temperature
7.3 Archives
The values are arranged in the archives in time sections, a time data of the
section, and values of the individual quantities selected for archiving form a part of
each time section.
The measured and calculated quantity values can be stored in the following
archives:
• Monthly archive
• Daily archive
• Data archive
• Binary archive
• Limits archive
Besides the listed data archives, the device also contains the following archives:
• Event archive
• Billing archive
• Settings archive
• Gas composition archive
First stored in the available device memory are the archives with a fixed number
of records (monthly, daily, binary, and limits) and the data archive is placed in the
remaining memory (its length depends on the size of the remaining memory).
34
miniELCOR
Standardized volume – absolute condition yes yes yes
Error operational volume – absolute condition yes yes yes
Error standardized volume – absolute condition yes yes yes
1)
Max. daily consumption – operational volume Yes
1)
Max. daily consumption – standardized volume Yes
1) 1)
Max. hourly consumption – operational volume Yes Yes
1) 1)
Max. hourly consumption – standardized volume Yes Yes
Internal counter – absolute condition yes yes yes
Output impulses – impulse debt condition yes yes yes
Operational flow – mean value yes yes yes
Standardized flow – mean value yes yes yes
2)
Minimum/maximum flow yes yes yes
Conversion, ratio of compressibility factors
Conversion factor – mean value yes yes yes
Ratio of compressibility factors – mean value yes yes yes
2)
Minimum/maximum of conversion, of ratio of yes yes yes
compressibility factors
Binary quantities
Binary input - condition yes yes
Binary output - condition yes yes
Set points - condition yes yes
Device errors and communication with yes yes
converters
Internal binary yes yes
Other quantities
Counter/timer – absolute condition yes
Input code yes
Notes:
1) Hour or day is stored along with the value (or combination, whichever suitable).
2) Date and time or achieving the minimum/maximum is stored along with the value.
Table 3 Options of archiving the individual quantities
35
miniELCOR
The record with date 01.06 thus means statistical values of quantities in interval
1.05. 6:00 to 1.06. 6:00.
36
miniELCOR
7.3.7 Settings archive
Archive capacity: An average of 500 records (depends on length/type of
records)
The settings archive stores changes of parameters, especially if they have
effect on metrological features of the device. The archive also stores the identification
of the employee who performed the change. The record contains a time mark,
employee identification, description of his/her activity, and eventually the new and old
values of the parameters which were changed.
This archive, similarly as the event archive and unlike the other archives, does
not rewind, i.e. after filling the archive up, one can not add to it and other changes of
parameters are disabled. This archive can not be displayed on the display, and the
content can only be displayed using a PC.
37
miniELCOR
• Gas composition (individual components of the gas pursuant to the set
calculation method)
• Date and time in the device
• volume parameters like setup of gasmeter constant, Vm, Vms, Vb, Vbs, S/N
of gasmeter
38
miniELCOR
7.6.1 Switch protection
There are two switches located inside the device – the metrology switch and the
service switch.
Switch
Position Description
meaning
Complete OFF Writing parameters in the device is disabled.
1)
OFF
The position of the switch does not matter; it is possible to
none
write in the device. Protection using the switch is disabled.
ON
1)
This meaning is preset by the manufacturer (default setting)
39
miniELCOR
40
miniELCOR
User level
Position of
Allowing activity when using
Activity the service
passwords
switch
Data readout
2
) The effect of the turned-on passwords can be suppressed by using the HW key WGQOI, „service“
version.
41
miniELCOR
- upgrade firmware
- Change of the metrology
approval option (NMi, ČMI,
MID, etc.)
- Setting a reference
Metrology changes
temperature
- Setting a reference pressure
- Setting the Vb, Vbs counters
- Configuration of metrology Using HW key marked
quantities (C, K, V, Vb, Vs, ON WGQOI, “Accredited service“
Vbs) option.
- Replacement of the
converter
- One-point or two-point
addition to converter
- Zeroing settings archive and
status archive
- measurand marking
exchange by user
42
miniELCOR
43
miniELCOR
8 Putting in operation
Device is delivered either in operation condition with connected battery or
switched out with disconnected battery.
Device is delivered in switched-out position ( no displayed information after
pushing of any button) and battery is placed at battery holder. There is placed
blocking foil strip between battery and holder contact. Putting in operation is
arranged by removing of this foil strip. This operation is also allowed at hazardous
zone.
44
miniELCOR
9 Device operation
The device is not equipped with a power switch; if a supply battery is inserted in
the device, the device is automatically on (the device also registers LF pulses if the
battery is taken out).
A 6-button keypad serves for the operation of the device and displaying the
measured and other values. The values are displayed on a graphic display with a
resolution of 128 x 64 points. During battery operation, the display shuts down after
20 s from the last time you pressed any key. The display lights automatically once
you press any key. In case the device is powered by an external source, the display
is permanently lit.
You can select the displayed data using the device menu. Displaying the menu
items depends on the set parameters of the device. Content of some menu items can
be custom configured.
Display features
Automatic update of data changing with period 1 s
Autorepeat – when holding a key, the key pressing is automatically generated,
can be used for e.g. viewing archives
Displaying without diacritical marks
In compliance with the EN 12405-1 standard par. 6.3.1.5, the display goes in
the basic display Using parameters, one can choose a time period after which
the device should go back to the basic display
To simplify the operation for an untrained user, there is an option to display
gradually the instantaneous values by pressing the Enter key First, it is
necessary to first get out to the highest menu level by pressing the Esc key for
several times.
To conserve energy, the device display shuts down after 20 s during battery
operation; It lights up again once you press a key.
9.1 Keypad
• In archive display, transition to another
quantity in the set time cross-section.
• In archives, transition to the previous
item in the same time cross section
Movement in time in archives, movement
in a menu
Movement in time in archives, movement
in a menu
45
miniELCOR
• Allows transition in menu to a lower level
• When displaying the instantaneous values, Enter causes scrolling
through the screen for gradual display of all quantities
The first line displays icons in the right upper corner informing on the basic
condition of the device.
46
miniELCOR
2 Service switch Service switch is in OFF position
condition
Service switch is in ON position
3 Battery Battery is charged 100 %
condition
Battery is charged 50 %
Battery is charged 25 %
4 Device condition Device works flawlessly
(sum state – see There is an error in the device
art. 9.9)
The device generated a warning message
Table 7 Display status icons
47
miniELCOR
48
miniELCOR
49
miniELCOR
Movement in time
forward
Movement in time
backwards
50
miniELCOR
The binary archive is displayed in the same way, with the only difference that
the records are not stored in the archive with an archiving period, but in times when
the status of one of the stored quantities changed.
You can scroll through the data on the display using the buttons and
. The end of the data is marked on the display with the symbol “<”.
51
miniELCOR
The Input parameters menu displays the following data:
• Measuring pressure p - pressure range
- serial no. of the pressure converter
• Measuring temperature t - measuring range
- serial no. of the temperature sensor
• Impulse input V - input clamps indication
- gas meter constant kp [imp/m3]
- gas meter serial no.
52
miniELCOR
Billing archive
Binary archive
Extreme archive
Save Saving of parameters in to device
parameters
You must set the parameter you want to edit to the first line of the display using
Saving parameters
After completion of the parameter editing, the performed edits must be written in
the device. The recording in the device is performed by selecting this option.
Successful recording of parameters in the device is confirmed by message “Data
valid”.
Counters value and actual time are written immediately after new value setting.
Before storing of new value into appropriate register there is shown confirmation
message
Device test
After selecting this menu item, the device will test its internal status and will list
on the display the errors found and warning messages. The initiated test of the
device takes approx. several seconds and has no effect on the measuring and
archiving activities of the device.
53
miniELCOR
A warning is displayed on the display during the test. The indicated errors are
marked with prefix “E” and identification number; in case of a warning message the
“W” prefix is used. For a complete list of errors and warning messages see par. 9.9.
Device reset
After choosing a device reset, the software jumps to the starting address and
performs a repeated initialization of the entire measuring system. The contents of all
archives and the statuses of all the V and Vb gas volume counters do not change
during this operation. All the other set parameters also do not change.
Communication
This menu displays the set communication interface (i.e. “RS-232/485 line”,
“infra IEC-1107” or communication via modem). In case of communication via
GSM/GPRS mode, some diagnostic information is displayed.
Value reading
This option freezes the displayed instantaneous values. Use this option in case
you need to manually copy the measured data.
54
miniELCOR
The short form of the summary diagnostics is displayed in the right corner of the
highest menu level in the form of abbreviations OK, Err or Wrn (see par. 9.2). This is
abbreviated for a summary of the individual statuses; the abbreviation with the
highest priority is always displayed. The order of the priorities from the highest: Err,
Wrn, OK. More detailed display of the diagnostic information can be done via the
service SW [22].
55
miniELCOR
56
miniELCOR
10 Mounting instructions
Gas volume conversion device miniElcor is compact device inbuilt in sturdy
housing made from plastic and it is corresponding with IP66 protection. Device is
designed for mounting in hazardous area Zone 1 and Zone 2.
Inside the housing there is next to the completely covered evaluation electronics
also placed battery and analog pressure transducer with input thread M12x1.5
according to DIN W 3861 for attachment of pressure piping.
On the bottom side of the housing there is placed 7 metal glands PG7(IP68
protection) and there are used for connecting of input and output signals with
possibility of conductive connection of cable shielding.
On the front of the device there is placed foil label with display, optical interface
for infrared head communication (HIE-01, 03, 04) and keypad.
57
miniELCOR
58
miniELCOR
59
miniELCOR
User mark
Temperature sensor
Safety nut
Thermowell
Direct weldolet
60
miniELCOR
For cable connection (size of conductor 0,5 – 1,5 mm2) there are used clamps
mounted in the device, in whose nearness there is printed on the board description of
signal which is assigned to each clamp (see Fig. 12). Before connecting cables it is
neccessary first to apply wire-end ferrule on the stripped end of the wire and force on
them by pliers which are supplied by the manufacturer of wire-end ferrules. Wires
ended by ferrules it is possible to plug in to the clamps without need of special tool.
During wire dissconnecting it is neccessary to slightly push nose of the clamp and
carefuly take out the wire.
61
miniELCOR
12 35
62
miniELCOR
11 Accessories
11.1 Assembly accessories
1 pc assembly board (metal)
2 pcs stirrup with splice
for assembly of board on pipes (for pipes Ø 50 mm, Ø 100 mm, Ø 150 mm
– must be specified when ordered)
1 pc temperature well (length 54 mm, 100 mm, 160 mm – must be specified
when ordered)
1 pc of welded-on piece for the temperature well (direct or slant – must be
specified when ordered)
1 pc three-way tap PN 100
63
miniELCOR
12 Technical parameters
Mechanical parameters
- mechanical dimensions (w x h x d) … 193 x 160 x 73
- weight … 1.2 kg
- casing material … polycarbonate
- terminals – conductor cross section … 0.5 mm2 – 1.5 mm2
- mechanical class ... M2
- electromagnetic environment ... E2
Environment
- protection … IP66, pursuant to EN 60529
- working temperature … -25 °C - +70 °C
Readability of display in ambient
temperature below -20 °C is not guaranteed.
- storing temperature … -40 °C - +85 °C
- working position … vertical 3)
- humidity … max. 95%, non-concentrating vapors
- protection against dangerous touch of live … Small voltage
and non-live parts
Power supply
- supply battery type … Lithium 3.6V/17Ah (size D)
- supply battery life time … 6 years 4)
- supply battery voltage … 2.8 ÷ 3.6 V
- measuring the supply battery life time … Yes, alert 90 days before discharge
- back-up battery type … Lithium 3.6V/1Ah (size ½ AA)
- back-up battery life time … 10 years
3
) Recommended working position. In case of a working environment where humidity condensation
avoidance is ensured, the device can also be installed in horizontal position.
4
) The life time of the supply battery depends on the set mode, and the life time of the back-up battery
depends on the method of using the device without the charging battery
5
) Even simpler options of conversion can be configured. Supported options are PTZ, PT, TZ and T.
64
miniELCOR
- type approval mark … TCM 143/06 – 4664
(in accordance certification under MID)
… 80 ÷ 1000 kPa 8)
… 400 ÷ 7000 kPa 9)
- measuring error … < 0.25 % of the measured value
- long-term stability … < 0.1 % for each year of the measured value
… < 0.2 % for each year of the range
- without MID certification
- measuring ranges … 80 ÷ 520 kPa
… 80 ÷ 1000 kPa
… 80 ÷ 2000 kPa
… 80 ÷ 3500 kPa
… 80 ÷ 7000 kPa
- measuring error … < 0.20 % of the range 7)
- long-term stability … < 0.2 % for each year of the range 7)
6
) In case of device version without MID certification there is a measurement error specified in
percentage from the range according to american usage.
7)
Selected calculation method of compressibility can involve temperature range. See Table 1.
8
) Enhanced range for extra charge. Can not be combined with enhanced accuracy.
65
miniELCOR
- Connecting pressure … tube ∅ 6 mm, screwing ERMETO M12 x 1.5
- design … Internal
... External, standard cable length 2.5 m max. 5
m
Measuring temperature
- Number of inputs … 1
- Sensor … Pt 1000, platinum resistor detector
- measuring range … -25 ÷ +60 ℃
- measuring error … ±0.2 ℃
- long-term stability … < 0.02 % for year (relative error in K)
- sensor design … tube ∅ 5.7 mm, length 120 mm with
integrated cable
- length of external sensor cable … Standard 2.5 m, max. 10 m
9
) Will damage pressure converter, the gas-tightness remains the same.
66
miniELCOR
Binary input - NAMUR 10) … Terminals DI1 (HF+/-), DI2 (HF+/-)
- Max. number of inputs … 2
- input type … NAMUR (DIN 19234)
- min. length of pulse / pause … 200 ms
- open circuit voltage … UPWR
- Internal resistance … 1 kΩ
10
) The device must be charged from an external source JBZ-02.
11
) The device must be charged from an external source JBZ-02.
12)
Optional accessories – influences gas group and category of intrinsically safety
67
miniELCOR
Analog output DO1, DO2, DO3, DO4 *) terminals
- Max. number of outputs … 4
- Output type … Current output 4-20 mA (realized by moduls
CL-1)
Interface for communication with … All three interfaces share the same
superior system communication channel – can not be
operated simultaneously
Metallic interfaces
- galvanic separation yes
RS-485 or RS-232 (not possible
- Interface of serial communication … simultaneously)
- Communication protocol … Optional, according to the firmware version
- Communication speed … 9600 Bd – 57600 Bd, adjustable
- Byte format … 8 bits, 1 stop, without parity
IEC-1107 interface
- Communication speed … 9600 Bd to 38400 Bd
68
miniELCOR
Binary Impulse Data
output output output*)
DO1 YES YES YES
DO2 YES YES YES
DO3 YES YES YES
DO4 YES YES YES
*) necessary to connect an external module CL-1 using the JB separator (e.g. Datcom-K3)
69
miniELCOR
13 Inexplosiveness parameters
HF inputs NAMUR DI1, DI2: HF+, HF- (INPUTS) Terminals
Uo = 10V
Io = 11mA
Po = 27mW
IIC IIB
Co 2,8μF 18μF
Lo 200mH 700mH
LF inputs and binary inputs DI1, DI2, DI3, and DI4: LF+/-, DI3+/-, DI4+/- (INPUTS)
terminals
Uo = 6.5V
Io = 8mA
Po = 15mW
IIC IIB
Co 2,8μF 18μF
Lo 200mH 700mH
RS485 communication line – internal bus (optional): Terminals GND, U+, D-, D+
Uo = 6.5V
Io = 1A
Po = 1.1W
IIC IIB
ΣCo 3,5μF 250μF
ΣLo 30μH 120μH
Digital outputs DO1 to DO4: Terminals GND, DO1, DO2, DO3, DO4 (OUTPUTS)
Ui = 15V
ΣPi = 1W
Ci = 500nF
Li = 0
70
miniELCOR
Li = 0
* Note: Sum of outputs is defined jointly for both interfaces, i.e. sum of outputs on
RS485 and RS232 must not exceed 0.33W except MTL5051.
71
miniELCOR
14 Device setting
After assembly and connection of the device in the measuring place is
necessary to set several device parameters (gas meter serial number, gas meter
constant and station identification, etc.). Setting of the device is made with help of
service SW. The installation is made with starting up of the file SETUP.EXE
For parameters setting is necessary to have the service switch in position “On”.
Recommended setting and control progress are necessary for correct device
function:
It is necessary to adjust this progress for other functions and controls (output
settings, external power supply, etc.) during concrete installation. Archives clearing
and summary status clearing must be completed at the end of control. After ending
of device setting is necessary to switch the service switch in „Off“ position.
Notes:
1. Metrological device parameters are set from the manufacture so that the device
will work according to customer’s requirements. Metrological values setting are
protected by HW key and metrological switch. This switch is secure with
authorized mark.
2. The possibility to set other device parameters is protected with service switch,
perchance with the password.
3. The device is delivered without active passwords.
72
miniELCOR
The device can be connected with PC either with help of cable by force of serial
interface RS-232 (eventually RS-485), infrared-head (HIE-01) or eventually via
modem.
For device setting in terrain is best to use infrared-head or cable RS-232 (see
Fig. 18). In case of communication through infrared-head is to this communication
assigns the highest priority and the device will automatically choose this
communication.
Warning:
If is the PC used in the explosive surround (ZONE 1 or ZONE 2), then must be
during connection between device and PC accomplished all protecting norms.
73
miniELCOR
Tel. no. – fill up only, when is the communication via modems. Fill up
telephone number of modem, which is connected with device.
IP address:port – fill up only in case of LAN network or GPRS connection
(data is possible to receive from administrator)
UDP - choose No.
Name of comm. channel – after double-click choose communication channel
(description – see paragraph 14.3.2)
Addr.1, Addr.2 – if you have only one connected device fill up zeros – in
case of placing in communication network must be filled up its actual address
Protocol – Choose the same which is set in the device. For establishing of
first communication is in the device set communication protocol ELGAS (ver.2)
74
miniELCOR
If are all data in the line set in right way, then is consumption place set and
communication way between PC and device is ready. The user may test
communication with help of actual values reading (menu choice Readout -> Actual
values).
Basically if user wants to edit the table, change is made with click on item.
75
miniELCOR
In the device are real time clock with calendar. It is possible to display actual
date and time on device display in menu System data or with reading out of actual
values with help of service SW. The item Setup -> Date and time in programme
menu allows changing these values.
The user may readout parameters from the device with help of item from menu
Readout -> Parameters.
After parameters readout are data from the device displayed in single window.
The use may with help of icon choose between two types of display mopde:
76
miniELCOR
77
miniELCOR
b) Full mode display
All parameters are displayed at tree order. This type of display is
determined for advanced users.
Warning:
In some cases (a, b) may change of the setting effect clearing of some archives.
78
miniELCOR
14.3.5 Device parametrization with assistant
79
miniELCOR
Value range of parameter Address1 is form 0 to 65535, for Address2 is from 0
to 255. For MODBUS protocol is range of parameter Address1 limited from 1 to 247;
Adresa2 is not used.
After pressing the button Next will be displayed following screen, which is
determined for setting of Gas composition. Firstly must be chosen Standard of
compressibility. According to chosen method is in the device preset gas
composition from the manufacturer. User may set gas composition according to
actual values.
80
miniELCOR
Setting of default compressibility constant is set in the line C01
Convers. factor C1 and is marked Default const. compressibility. This
compressibility value is used in calculation only if there is no compressibility
calculated according to mathematical methods.
With pressing the button Next follows analogous screen Setting 2. channel.
Note:
After closing of wizzard for parameterization there are parameters prepared in
computer pattern for writing into device. Therefore do not forget write into device
using button „ Save into device“ before closing of this page. For successful
parameters writing into device is necessary to have the service switch in position
„On“. After parameters saving switch over the service switch in „Off“ position.
81
miniELCOR
device parameters (Fig. 40 – simply parameters display) choose Actual value and in
the column Value write required data.
82
miniELCOR
„Init. sum. stat.“ – serves for initialization (clearing) of summary status.
b) In SW on PC (Fig. 41)
Menu choice „Setup -> Diagnostics (status) of the device -> From
device“ will be read out from connected device summary status and status
after last device test. (For displaying of diagnostics from saved data choose
From file). It is possible to find out current status of the device with pressing
the button Device test.
In column „Summary status“ there are captured again all errors from the last
reset of summary status. The central button at bottom part of window can be used
for reset.
Table with description of errors and warning messages and pertinent
procedures how to remove it are shown in charter 16.
83
miniELCOR
14.3.8 Archives clearing
This operation may be made only with help of service SW. Choice in menu (see
Fig. 42 ) is possible to clear archives selectively or all archives with one order (except
Setup archive).
Warning:
Deleted data in archives cannot be refreshed.
84
miniELCOR
- With password for reading is possible to readout data from the device.
- Password for full access allows to readout data and writes data into device.
Password for full access merges both rights of passwords for reading and for
full access. Service SW remember password till its restart, so it is not necessary to
enter the passwords again during reading or writing. Passwords may be set in menu
Setup –> Password for reading; Password for full access.
85
miniELCOR
15 Configuration examples
In this chapter there are the most common device configuration examples. User
may deduce standard progresses of device setting according to user’s requirements.
Device configuration is made out by force of service SW.
Firstly must be displayed file of device parameters:
1) Parameters displaying of the device which is connected to PC:
menu Readout -> Parameters
2) Displaying of parameters which were readout form the device in the past:
menu Setup -> Parameters (select saved file with suffix *.par)
Notes:
1) Changing of parameters is made in computing memory of PC. It is saved into
device after pressing the button Save into device.
2) Described examples of configuration may do (except point out exceptions)
users without hardware key. For successful parameter saving is necessary to
have service switch in „On“ position. After finishing of parameters adjusting is
necessary to switch the service switch back to „Off“ position.
86
miniELCOR
15.2.1 Simple mode display
Value of constant by item V01 is automatically copied into item Q01. The same
rule stands for V02 a Q02, because this constant is used for counting of both
quantities.
87
miniELCOR
b) – counting of Flow Q1 (or Q2).
In this display mode is constant chosen for one parameter (V) not used
automatically for second parameter (Q).
Warning:
Practically it means that we can have for V and Q two different constants.
88
miniELCOR
For gas meter with HF output HF pulse (NAMUR) must be ticked on. Fig. 46
shows setting of gas meter with HF output constant on value 82.5564 pulse/m3 and
serial number of gas meter 987654321 for primary volume V1. Here is also
necessary to make same setting for Flown Q1.
Fig. 46 Setting of gas meter with HF output constant in full mode display
89
miniELCOR
To set this type of output it is possible only in case, that this output was already
configured in full mode display. Then is possible (in simple mode display) to switch
on or switch off by force of the button Output pulses, binary (see Fig. 47).
In case that this output was not configured it is necessary to activate it and
assign a quantity (representing pulses). It is necessary to set technical parameters of
output pulses as well.
90
miniELCOR
Then card of Pulse output counter will be created e.g. with indicated V03o (see Fig.
49).
91
miniELCOR
Now is activated hardware pulse output. The device offers possibility of four digital
outputs. These outputs are lead out on clamps (DO1 to DO4) of OUTPUTS terminal
block inside the device. Item Output No. indicates on which clamps of OUTPUT
terminal block is this output connected:
Output No. 1 – clamp DO1
Output No. 2 – clamp DO2
Output No. 3 – clamp DO3
Output No. 4 – clamp DO4
The programme assigns automatically free output clamp (in our example DO2). But it
is possible to swap it.
If necessary apply this output on output constant, then the expression is filled
with multiply of requested constant. E.g. if constant is 6.53, then it is necessary to
customize the mathematical expression on:
(dVb1+dVbs1)*6.53
92
miniELCOR
93
miniELCOR
After this setting related equation is by the V03 parameter displayed in the
frame Source of value. To make some changes in equation press the button Skip
on source (see Fig. 52).
In simple mode this output cannot be configured. Only in case that this output
was already configured in full mode display. Then it is possible to switch on or switch
off this output with ticking off/on (see Fig. 47).
94
miniELCOR
15.4.2 Full mode display
This procedure is similar with the procedure for pulse output. From item
Hardware (in left upper window) must be chosen Insert output measurand and
then Data output for CL-1 (see Fig. 53).
On card Analogue output (see Fig. 54, parameter A05o) Output number
gives appropriate clamp (from the range of the clamps from DO1 to DO4) where
output signal will be sent to or where module CL-1 will be connected to.
95
miniELCOR
On the first screen choose Current output, on the second screen choose
quantity which will be watched (e.g. Base flow) and then assign values of minimal
and maximal limits for output current according to flow value. After wizard finishing
will be displayed generated equation for analogue output (parameter A05 see Fig.
56).
96
miniELCOR
Resulting form of generated equation is:
A05o = (Qb1*16+48000)/12000
Form equation is possible to see that for Qb1=0 is equation result A05o=4 mA
and for flow Qb1=12000 is the result A05o=20 mA.
Resulting setting for analogue output A05 is on Fig. 57.
97
miniELCOR
In this type of displaying cannot be the output added into configuration of the
device. But if the parameter was added in full mode display (see next) it is possible
to set size of value.
98
miniELCOR
Example:
If measured pressure p1 will pass over the value 120 (kPa) and this passing
over will be longer than 5s, so then will be generated alarm (see Fig. 59).
99
miniELCOR
When will arise failure of external power supply (see Fig. 61) then will be
generated alarm signal, but only in case that the failure will be longer then 3 s.
100
miniELCOR
MODBUS template:
− it is pre-defined table of MODBUS addresses with sorted device parameters.
In the there table are also information about type of the parameter – if it is
101
miniELCOR
determined only for reading or also for writing. The manufacturer offers
standard template for one-channelled or two-channelled device.
For device setting it is necessary to record this template into device.
As the first step of setting new object Address map of MODBUS is added into
device parameters. The progress is displayed on the Fig. 62. Adding is made with
click of right mouse-button on item Calculated measurands->Insert object of
parametrization. Then there is displayed dialog window for file opening. Choose
required MODBUS template (file with extension *.db).
Note:
Only worker with access authorization of ASC (authorized service centre) is
allowed to carry out this change.
102
miniELCOR
After this device switching (after parameters into device writing) will be the
communication interrupted (communication protocols on device and PC are
different). For another communication with device is necessary to make in
parameters setting for connection with consumption place switching on MODBUS
protocol. It is also possible set parameter Adr1 either on non-zero address (which is
set in the device) or on universal address Adr1 = 248 (see 14.3.1).
103
miniELCOR
104
miniELCOR
Warning:
In case of replacement of sensor/transducer which is measuring basic
quantity, it is a hit to metrological part of device. Before this operation it is
necessary to breach metrological seal and move metrological switch to ON
position.
Breaching of metrological seal means, that initial verification is not valid!
This operation can be proceed only by employee with rights of acreditted
service center (ASC). To proceed this operation it is necessary to switch
metrological switch into the ON position and also it is importnat to use
hardware key (key variant: Authorized service center).
105
miniELCOR
At this moment we have replaced temperature sensor with the new one,
battery is inserted back into device and now it is necessary to finish installation by the
help of service software.
Readout parameters from the device (menu Readout-> parameters). In next
step click in menu of parameters on „Hardware“ item (see Fig. 65).
In right part of the opened window click on item „Change temperature
transmitter“. In next window enter path to the data file *.txt with stored calibration
data, which was delivered together with temperature sensor (see Fig. 66). Select and
confirm by clicking on „Open“ button required calibration file. Calibration data are
stored in parameters now. You should see following window after loading of
calibration file (see Fig. 67).
To finish this operation save parameters into the device by clicking on „Save
into device“ button. After saving of parameters readout parameters from the device
in order to check if all changes were proceed succesfuly. Click on „Hardware“ item
(see description above) and check corectness of serial number placed on sensor
cable with serial number in parameters (see Fig. 68).
If serial numbers are complying, changing of sensor is sucessfuly finished. In
case that serial numbers are varying, check correctness of your steps during
changing of sensor, step by step, or check if was used correct calibration file.
Note: After replacement of temperature sensor it is possible (if needed) to make
one or two point calibration of sensor. (see Fig. 67).
Fig. 65
106
miniELCOR
Fig. 66
Fig. 67
107
miniELCOR
Fig. 68
At this moment we have replaced pressure transducer with the new one,
battery is inserted back into device and now it is necessary to finish installation by the
help of service software.
Readout parameters from the device (menu Readout-> parameters). In next
step click in menu of parameters on „Hardware“ item (see Fig. 65). In right side of
the window click now on „Parameters of module“ (see Fig. 68). Now the service
software will load calibration data from the transducer’s memory into parameters of
the device. At this time is software setting almost finished. Now it is necessary to
save parameters into the device by clicking on „Save into device“ button. After that
check compliance of serial number placed on transducer and in parameters.
Hereby is the tranducer replacement finished.
Note: After replacement of pressure transducer it is possible (if needed) to
make one or two point calibration of the pressure sensor. (see Fig. 67).
108
miniELCOR
109
miniELCOR
17.1.2 Newly installed digital transducer has different address from
address preset in device’s parameters
If we will found that transducer is not communicating, probably has a different
communication address than is preset in device’s parameters. To set correct address
click in device’s parameters on transducer item (in our case „Temperature EDT-34“
see Fig. 70). After that in right part of the window click on „New address“ button.
Service software will ask you to enter serial number of added transducer. Enter serial
number and click „OK“ (see Fig. 73). After that service software will ask you to enter
new communication address. Enter the same address which is stored in device’s
parameters – standartly it is the address no.1 (see Fig. 74). New address is stored in
transducer now.
By clicking on „Parameters of module“ button we will check if transducer is
communicating properly with the device. If everything is correct you will see window
with transducer’s parameter (see Fig. 71). Herewith we finished the change of
address and also installation of transducer is finished.
Note: After adding of digital transducer it is possible (if needed) to make one or
two point calibration of transducer. (see Fig. 72).
110
miniELCOR
Fig. 69
111
miniELCOR
Fig. 70
Fig. 71
112
miniELCOR
Fig. 72
Fig. 73
113
miniELCOR
Fig. 74
Fig. 75
114
miniELCOR
Fig. 76
115
miniELCOR
116
miniELCOR
E4 error of FLASH Error of device FLASH memory. Err
- Necessary repair in ASS
E5 full setup Full setup archive. Err
archive - Device is full operational but no parameteres are
changable. Erase setup archieve in ASS.
E6 transducer Accomplished replacement of transducer or modification Err
replacement of parameters.
- Set back device to original setup or arrange
veryfication in ASS.
E7 transducer Error in communication with transducers. Err
communication - Check connection of transducers for example setup
of correct communication address.
E8 transducer error Error of transducer. Err
- Measured value can be out of measured range, or
transducer is defective – necessary replace
transducer in ASS.
E9 battery voltage Battery voltage decreased under allowable level. Err
- Replace battery.
E10 compressibility Error of calculation in compressibility table due to input Err
table parameters.
- Correct gas composition
E11 compressib. Infeasible calculation of compressibility table due to range Err
restriction of used standard
Note:
ASS – autorized service centre
117
miniELCOR
118
miniELCOR
20 Literature
[1] 22/1997 Coll. Act on Technical Requirements for Products and on
Amendments to Some Acts
[2] EN 60079-0:2006 – Electrical apparatus for explosive gas atmospheres – Part
0: General requirements.
[3] EN 60079-11:2007 – Explosive atmospheres –Part 11: Equipment protection by
intrinsic safety “i”
[4] EN 60079-26 :2007 – Explosive atmospheres – Part 26: Equipment with
equipment protection level (EPL) Ga
[5] EN 12405-1:2006 – Gas meters – Conversion devices – Part 1: Volume
conversion
[6] EN 60079-14:2004 – Electrical apparatus for explosive gas atmospheres – Part
14: Electrical installation in hazardous areas (other than mines).
[7] EN 61000-4-2:1995+A1:1998+A2:2001 – Electromagnetic compatibility (EMC) -
Part 4: Testing and measurement techniques – Section 2: Electrostatic
discharge immunity test – Basic EMC Publication
[8] EN 61000-4-3:2006/A1:2008 Electromagnetic compatibility (EMC) – Part 4-3:
Testing and measurement techniques – Radiated, radio-frequency
electromagnetic field immunity test.
[9] EN 61000-4-4:2004 - Electromagnetic compatibility (EMC) – Part 4-4: Testing
and measurement techniques – Electrical fast transient/burst immunity test.
[10] EN 61000-4-6:2007+Cor.:2008 – Electromagnetic compatibility (EMC) - Part
4: Testing and measurements techniques – Section 6: Immunity to conducted
disturbances, inducted by radio-frequency fields.
[11] EN 61000-6-2: 2005/Cor.:2005-09 - Electromagnetic compatibility (EMC) – Part
6-2: Generic standards – Immunity for industrial environments.
[12] EN 61000-6-4: 2007
[13] EN 62056-21:2002 Electricity metering - Data exchange for meter reading, tariff
and load control – Part 21:Direct local data exchange.
[14] IEC 60364-4-41: 2005 – Low voltage electrical installations – Part 4-41:
Protection for safety - Protection against electric shock.
[15] Modicon Modbus Protocol Reference Guide, Modicon Inc., Industrial
Automation Systems, 1996
[16] FTZÚ 08 ATEX 0324X – EC -Type Examination Certificate
[17] EN ISO 12213-3: 2009 - Natural gas – Calculation of compression factor – Part
3: Calculation using physical properties
[18] Directive 2006/66/EC of the European Parlament and of the Council on
batteries and accumulators and waste batteries and accumulators
21 Relevant Literature
[19] TELVES – Software description.. User manual. Elgas, s.r.o.
[20] EDT 23 – Pressure converter with Modus protocol. User manual. Elgas, s.r.o.
[21] EDT 34 – Temperature converter with Modus protocol. User manual. Elgas,
s.r.o.
119
miniELCOR
22 Software
[22] TELVES exe, Elgas, s.r.o., software supplied with device
[23] Reliance, GEOVAP Pardubice
120
miniELCOR
24 List of figures
Fig. 1 Volume and energy calculations - Scheme....................................................... 6
Fig. 2 Device dimensions............................................................................................ 7
Fig. 3 Main parts of the device.................................................................................... 8
Fig. 4 Examples of external power supply ................................................................ 11
Fig. 5 Security marks (device without SCR encoder) ............................................... 12
Fig. 6 Security marks of miniELCOR SCR............................................................... 12
Fig. 7 Product label English version ........................................................................ 13
Fig. 8 Product label – original certification for ZONE 1 ............................................ 13
Fig. 9 Storing impulses in counters ........................................................................... 19
Fig. 10 Processing of volumes during reversed flow ............................................... 20
Fig. 11 Encoder SCR board (without cover) ............................................................. 23
Fig. 12 Inputs and outputs terminals ......................................................................... 25
Fig. 13 Example of an impulse (binary) output and current output scheme .............. 26
Fig. 14 Connecting of digital transducer with expansion module RS-485 ................. 28
Fig. 15 Placing of expansion module RS-485 in the device ...................................... 28
Fig. 16 Safety separation of communication using RS-485 module DATCOM-K3 ... 30
Fig. 17 Safety separation of RS-232 communication via separator MTL 5051 ......... 31
Fig. 18 Communication cable wiring ......................................................................... 31
Fig. 19 Removable foil strip at battery holder .......................................................... 44
Fig. 20 Significance of keys ...................................................................................... 46
Fig. 21Basic navigation from the initial screen (FW 4.xx) ......................................... 47
Fig. 22 Basic navigation from the initial screen (FW 2.xx) ........................................ 48
Fig. 23 Device main menu and first submenu level .................................................. 49
Fig. 24 Example of displaying the instantaneous values .......................................... 50
Fig. 25 Navigation in archives (archiving period 15 min) .......................................... 50
Fig. 26 Mounting of miniElcor to the mounting plate ................................................. 58
Fig. 27 Mounting on the pipeline............................................................................... 58
Fig. 28 Temperature sensor mounting ...................................................................... 60
Fig. 29 Temperature sensor mounting by the help of direct weldolet ....................... 60
Fig. 30 Shielding connection in bushing ................................................................... 62
Fig. 31 Consumption place setting .......................................................................... 74
Fig. 32 Adding of new communication channel ........................................................ 76
Fig. 33 Base device parameters ............................................................................... 77
Fig. 34 Data archive ................................................................................................ 77
Fig. 35 Base parameters – Full mode display.......................................................... 78
Fig. 36 Assistant for parametrization ....................................................................... 79
Fig. 37 Parametrization of identification and communication ................................... 79
Fig. 38 Gas composition parametrization ................................................................ 80
121
miniELCOR
Fig. 39 Cahnnel setting ............................................................................................ 81
Fig. 40 Primary volume counter setting ................................................................... 82
Fig. 41 Displaying of device diagnostics .................................................................. 83
Fig. 42 Device archives clearing .............................................................................. 84
Fig. 43 Password setting ......................................................................................... 85
Fig. 44 Gas meter constant setting in simple mode display ..................................... 87
Fig. 45 LF gas meter constant setting in full mode display ...................................... 88
Fig. 46 Setting of gas meter with HF output constant in full mode display ............... 89
Fig. 47 Output pulses, switching on and off .............................................................. 90
Fig. 48 Pulse output inserting ................................................................................... 91
Fig. 49 Pulse output parametrization ........................................................................ 91
Fig. 50 Wizard for mathematical expression ............................................................ 93
Fig. 51 Equation for primary volume pulses and output connection ........................ 93
Fig. 52 Parameter V03 after equation assigment .................................................... 94
Fig. 53 Analogue output setting ............................................................................... 95
Fig. 54 Analogue output parametrization ................................................................. 95
Fig. 55 Wizard for analogue output equation making .............................................. 96
Fig. 56 Resulting equation and connecting of analogue output on counted quantity 96
Fig. 57 Resulting setting of analogue output on output clamp DO3 ......................... 97
Fig. 58 Setpoint setting ............................................................................................. 98
Fig. 59 Setpoint parametrization .............................................................................. 99
Fig. 60 Displaying of device external power supply ............................................... 100
Fig. 61 Monitoring setting of external power supply failure .................................... 101
Fig. 62 MODBUS map inserting ............................................................................ 102
Fig. 63 Parameter of MODBUS map address........................................................ 103
Fig. 64 Switch of communication protocol in the device ......................................... 104
Fig. 65 .................................................................................................................... 106
Fig. 66 .................................................................................................................... 107
Fig. 67 .................................................................................................................... 107
Fig. 68 .................................................................................................................... 108
Fig. 69 .................................................................................................................... 111
Fig. 70 .................................................................................................................... 112
Fig. 71 .................................................................................................................... 112
Fig. 72 .................................................................................................................... 113
Fig. 73 .................................................................................................................... 113
Fig. 74 .................................................................................................................... 114
Fig. 75 .................................................................................................................... 114
Fig. 76 .................................................................................................................... 115
122
miniELCOR
25 List of Tables
Table 1 Limitation of standard validity range of compressibility calculation ............. 17
Table 2 Digital inputs setting options ........................................................................ 21
Table 3 Options of archiving the individual quantities ............................................... 35
Table 4 Service switch settings ................................................................................ 40
Table 5 User access level (for “complete” meaning of the service switch) ............... 42
Table 6 ASC access level ......................................................................................... 42
Table 7 Display status icons ..................................................................................... 47
Table 8 List of events – error messages (Err indication) .......................................... 55
Table 9 List of events – warning messages (Wrn indication) .................................... 56
Tab. 10 Assigning of weldolets and thermowells according to pipeline diameter ..... 59
Tab. 11 Recommended cable types ......................................................................... 61
123
miniELCOR
Issued by: ELGAS, s.r.o. Phone: +420 466 414 500, 511
Ohrazenice 211 Fax: +420 466 411 190
533 53 Pardubice http://www.elgas.cz
Czech Republic e-mail: [email protected]
Issued on: Januaryr 2010
Rev. no.: Rev. 6g
H065EN_201001_R6g_miniELCOR
124