Application Note - ASTM D7575
Application Note - ASTM D7575
Application Note - ASTM D7575
News
Quantitative Analysis of Oil and Grease in Water
Using FTIR Based on ASTM D7575
No. A544
The discharge of oil and grease into water environments <Contents of ClearShot Extraction Package>
such as oceans and rivers is a water quality control issue • ClearShotTM Extraction Technology cartridges
since it may lead to environmental problems including (ClearShot extractors)
adverse effects on ecological systems and malodor. • ClearShot Holding Card
A measurement method described in JIS K 0102 "Testing • Calibration Standard Devices (CSD) Set
Methods for Industrial Wastewater"*1 that involves • Drying System
determining the quantity of n-Hexane extract is one
method for quantifying oil and grease in water. However,
there are issues with this method including the necessity Calibration Curve Creation
of cumbersome pretreatment and the long time required After measuring the background with a new ClearShot
until results are obtained. extractor, the seven calibration standard devices (CSD) for
This article introduces a quantitative analysis method for calibration curve creation were measured using the
oil and grease in water that employs FTIR based on the transmittance mode. A calibration curve was created by
ASTM D7575 standard.*2 ASTM standards are set and setting a baseline between 2990 cm-1 and 2800 cm-1 in the
published by ASTM International, which is the world's obtained infrared spectra and determining the heights of
largest international standardization and standards- the top peaks from the baseline at 2920 cm-1. Table 1 lists
setting organization. ASTM D7575 enables quantitation of the measurement conditions, Fig. 2 shows an enlarged
oil and grease in water with a simple measurement view of peaks around 2920 cm-1 of the standard samples,
method that utilizes the absorption band of CH groups, and Fig. 3 shows the calibration curve and lists the
thereby eliminating the need for solvent extraction. standard sample concentrations.
R. Fuji Table 1 Measurement Conditions
0.6
0.4
0.2
0.0
ClearShot Extraction Package
3100 3080 3060 3040 3020 3000 2980 2960 2940 2920 2900 2880 2860 2840 2820 2800 2780 2760 2740 2720
<Measurement Instrument and Testing Equipment> cm-1
• Fourier transform infrared spectrophotometer Enlarged View of Peaks Around 2920 cm-1
• 1 L glass sample collection bottle
• Ultrasonic cleaner capable of heating to 40 °C and Sample Absorbance
Concentration
[ppm]
accommodating the 1 L glass sample collection bottle Height
The content of this publication shall not be reproduced, altered or sold for any commercial purpose without the written approval of Shimadzu.
Company names, product/service names and logos used in this publication are trademarks and trade names of Shimadzu Corporation or its
affiliates, whether or not they are used with trademark symbol “TM” or “£”. Third-party trademarks and trade names may be used in this
publication to refer to either the entities or their products/services. Shimadzu disclaims any proprietary interest in trademarks and trade names
www.shimadzu.com/an/ other than its own.
The information contained herein is provided to you "as is" without warranty of any kind including without limitation warranties as to its
accuracy or completeness. Shimadzu does not assume any responsibility or liability for any damage, whether direct or indirect, relating to the
use of this publication. This publication is based upon the information available to Shimadzu on or before the date of publication, and subject
to change without notice.
© Shimadzu Corporation, 2017