Proceso de Mantenimiento Maquina Anestesia

Download as pdf or txt
Download as pdf or txt
You are on page 1of 20

Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 1 of 15 12/19/95 Number: 3.00

I. GENERAL INFORMATION

WARNING: This procedure contains limited warnings and cautions. Refer to the product
Service Manual for a full listing of precautions.

WARNING: Datex-Ohmeda employees must follow the guidelines explained in HRFS451,


Infection Control Procedures. Non Datex-Ohmeda users of this document must
follow the applicable institution policy for infection control.

II. PRE-INSPECTION PROCEDURES

A. Gas Machine and Ventilator Mechanical and Visual Checks

1. Check the condition and tightness of all cabinet/frame parts.


2. Check the casters for proper operation and mounting. DO NOT LUBRICATE THE
CASTERS. Check the brake for proper operation.
3. Check the storage drawers for smooth operation. Verify the end stops function correctly.
4. Check the condition of the O2 cartridge and sensor assembly. Reinstall the O2 sensor
assembly into the GMS O2 sensor port.
5. Check the condition of all front panels.
6. Check for proper mounting and smooth operation of all switches and rotary controls. Ensure
all associated knobs are secure.
7. Verify that all labels are in place and are clearly legible.
8. Check all external tubing, including rubber goods, hoses and pipeline hoses, and ensure
there is no deterioration and that they are securely attached.
9. Check all external electrical cabling. Be sure it is correctly connected and is not
deteriorating.
10. Check that the bellows housing is not cracked, chipped or damaged. Also check the
condition of the housing lettering.
11. Verify that the O&M Manual or Pre-operative checklist is present.

III. MACHINE AND ABSORBER LEAK TEST

A. Gas Machine High Pressure System/ High Pressure Leak Test

1. Ensure that the System Master Switch is in the STANDBY position. Remove all pipeline
connections.
2. Open all cylinder valves. Verify that there is at least 1000 psi (745 psi for N2O) or higher on
the cylinder pressure gauges.
3. Close all cylinder valves. Verify that the pressure does not visibly drop in 1 minute.
4. Loosen the "T" handles and pull the cylinders back to verify that the pin index and strainer
nipples are intact.

NOTE: It is normal to hear a momentary release of pressure.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 2 of 15 12/19/95 Number: 3.00

5. For each double cylinder yoke:


a. Connect one cylinder to the double yoke.
b. Open the cylinder. Listen for a continuous leak at the unused yoke position. If you hear
a leak, check the tightness of the check valve. Repair or replace as necessary.
c. Move the cylinder to the unused yoke position and repeat step b.
6. Tighten the "T" handle on each cylinder.

B. Common Gas Outlet Relief Valve.


1. Turn the System Master Switch to the ON position.
2. Open the oxygen cylinder.
3. Set oxygen flowmeter to 1 L/min.
4. Attach an approved pressure measuring device to the common gas outlet.
5. Ensure the common gas outlet relief valve relieves between 205 - 285 mmHg.
6. Remove the measuring device.
7. Close the oxygen cylinder.
8. Turn the System Master Switch to the STANDBY position.

C. Low Pressure Leak Test


Note: Annually, the Low Pressure Leak Test should be performed twice, once with all
vaporizers removed and once with vaporizers installed. At the annual check, replace the
vaporizer manifold and machining port O-rings.

1. Verify the integrity of low pressure leak test device.


2. Attach low pressure leak test device to the common gas outlet.
3. Check that each vaporizer is securely locked to the manifold and is in alignment with other
vaporizers.
4. Open all flow control valves two full turns.
5 Ensure all vaporizers are OFF.
6. Repeatedly squeeze and release the hand bulb of the device until it remains collapsed.
7. If the hand bulb inflates in less than 30 seconds, locate and correct the leak.
8. Turn a vaporizer ON to its maximum setting, checking the concentration dial for smooth
operation. Set the vaporizer to 1%. Repeat Steps 7 and 8 with the vaporizer ON.
9. Check the interlock system to ensure that only one vaporizer at a time may be turned ON.
10. Turn the vaporizer OFF. Repeat steps 9 - 11 for each mounted vaporizer.
11. Fully close all flow control valves to their minimum settings.
12. Remove the test device from the common gas outlet and reattach the common gas outlet
hose.
13. If a Blood Pressure Manometer is equipped, Check the calibration as follows:
a. Attach an approved test fixture to the fitting on the blood pressure inlet port.
b. Apply pressure until 200 mmHg ± 4 is displayed on the test device and verify that
200 mmHg ± 4 is indicated and maintained on the BP manometer.
c. Apply pressure until 100 mmHg ± 4 is displayed on the test device and verify that
100 mmHg ± 4 is indicated and maintained on the BP manometer.
d. Relieve the pressure and verify that 0 mmHg is indicated.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 3 of 15 12/19/95 Number: 3.00

D. Pressure Sensor/Low O2 Alarm

WARNING: Follow the appropriate agent evacuation/collection safety measures.


Use the hospital evacuation system with the service gas-evacuator tee. If
hospital evacuation system is not available, perform service in a
well-ventilated area.

1. Connect all pipeline supplies and cylinder-only gas supplies (for example, Heliox).
a. Open all cylinder valves to ensure that the pneumatic circuits are pressurized.
b. Close cylinder valves for gases which also have pipeline supplies.
2. Turn the System Master Switch to the ON position.
3. Set all flowmeters for 3 L/min.
4. Verify that the cylinder gauges for pipeline supplied gases have not dropped to 0 psi. Such
a drop would indicate that the flowmeters are being supplied by the cylinders.
5. Remove oxygen supply.
6. Verify that the following events occur in the following sequence:
a. Audio alarm activates for at least 7 seconds and Pneumatic indicator is red or orange, if
equipped.
b. Flowmeter floats for all gases except O2 fall to zero.
c. O2 flowmeter float falls to zero .
7. Turn the Ventilator ON. Ensure that the Low O2 Supply Pressure LED is ON and the audio
alarm sounds (older units may not have an audio alarm).
8. Turn the Ventilator OFF.
9. Turn all flowmeters to their minimum settings.
10. Re-establish O2 supply.
11. Verify that the audio alarm silences and the Pneumatic indicator is green or black, if
equipped.

E. Scavenging System Checks

NOTE: The old style relief valve may allow a negative pressure indication up to –
2 cmH2O. If you find an old style relief valve, replace it with the new style valve
assembly.

1. Check and/or replace as necessary:


a. 1/4 inch vacuum hose.
b. All unused evacuation ports must be capped.
c. Evacuation hoses.
d. Three (3) liter evacuation bag.
e. 19 mm scavenger tee for gas monitors.
f. Ensure that the double 19 mm connector, if fitted, is located beneath the scavenger
valve.
2. Inspect and Clean
a. Remove positive and negative relief valve components.
b. Clean inside of manifold and beneath negative pressure button.
c. Clean needle valve and seat assembly. Lubricate the needle valve threads w/Krytox, as
necessary.
d. Replace positive and negative pressure gaskets as needed.
e. Reassemble the valve.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 4 of 15 12/19/95 Number: 3.00

F. GMS Inspection

NOTE: For units manufactured prior to 1986, verify that the Inhalation Port pressure
sensing nipple, a raised head gasket, a drain valve for adjustable height bag
arms, and check valve disk retainers with stand offs are installed.

1. (EVERY 24 MONTHS) Manifold Inspection


a. Remove the access screen. Replace the cup seals every 24 months. Looking through
the access screen opening, inspect the GMS manifold for contamination. If
contamination is present, remove the manifold and clean or replace components as
necessary.
2. Periodic Inspection
a. Gauge
1) Remove the gauge assembly and inspect the O-ring; replace as necessary. Insert
the gauge assembly into an approved test device (see tools appendix).
2) Verify the accuracy of the gauge at:
0 ± 1 cmH20 (0 ± 0.7 mmHg)
40 ± 1 cmH20 (29.4 ± 0.7 mmHg)
-5 ± 3 cmH20 (-3.7 ± 2.2 mmHg)
3) If 5500 equipped, ensure that the patient pressure sensing line is attached to the
test device. Verify that the 5500's sub-atmospheric pressure alarm activates at
-10 ± 5 cmH20 (- 7.4 ± 3.7 mmHg).
4) Return the gauge assembly to the absorber.
5) Ensure that the pressure sensing line is attached to the absorber.
b. Check the integrity of the sensor port cap and O-ring. Replace as necessary.
c. Check the condition of the Inhalation and Exhalation domes, O-rings, check valves and
check disc retainers. Replace as necessary.
d. Looking through the inhalation check valve seat and O2 sensor port, inspect the GMS
manifold for contamination. If contamination is detected, remove the GMS manifold and
clean.
e. Check the condition of the corrugated return hose. Replace as necessary.
f. Check the condition of the canister gaskets, seals and upper gaskets. Replace as
necessary.
g. Inspect the canister dish; clean as necessary.
h. For elevating arms, check the Bag Arm Tube Drain for cleanliness.
i. Check for correct firm operation of the Locking Lever.
j. Lubricate the GMS Mounting Pin, as necessary. Install a mounting pin gasket if one is
not present.
G. Inhalation Check Valve
1. Remove the bag and the patient circuit, if installed.
2. Disconnect the gas supply hose from the common gas outlet.
3. Detach the bellows assembly from the GMS or, if remotely mounted, the hose assembly.
4. Connect the corrugated hose from the common gas outlet to the inhalation valve.
5. Set the O2 flow to 200 ml/min.
6. Verify that the GMS gauge increases from 0 to at least 5 cmH2O in 30 seconds.
7. Remove the corrugated hose from the common gas outlet.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 5 of 15 12/19/95 Number: 3.00

H. GMS Selector Valve and Ventilator Port


1. Re-connect the gas supply hose to the common gas outlet.
2. Attach the free end of the corrugated hose to the 22 mm ventilator port on the back of the
absorber.
3. Check for correct switching action of the selector valve by slowly switching it three times
between Ventilator and BAG/APL. A single click indicates correct switching action. If other
than a single click, replace the selector valve.
4. Set the BAG/VENT switch to VENT.
5. Flow O2 until 40 cmH2O is observed on the GMS gauge. Reduce the O2 flow to 200 ml/min.
6. Verify that the GMS gauge reading does not drop in 10 seconds.
7. Open the drain plug and verify that the pressure on the GMS gauge depletes to zero.
8. Close the drain valve.

I. Exhalation Valve Assembly/APL Assembly


1. Bi-Annual APL Assembly Inspection (every 24 months only)
a. Remove the APL knob.
b. Remove the calibration screw, the spring and the APL needle.
c. Inspect the interior of the APL chamber for contamination.
d. Inspect the APL needle for nicks, wear or contamination.
e. Check the integrity of the calibration screw O-rings; replace as necessary.
f. Reinstall the APL needle, spring and calibration screw.
g. Remove the corrugated hose from the ventilator port and attach it to the bag arm.
h. Set the BAG/VENT switch to the bag position and turn the APL valve fully clockwise.
i. increase flows to 3 L/min.
j. Adjust the calibration screw for a maximum pressure of between 65 - 76 cmH2O.
k. Turn the APL valve fully counter clockwise.
l. Verify that the gauge indicates 1 - 5 cmH2O.
m. Reinstall the APL knob.
n. Proceed to step 3 of the Exhalation Valve/APL Assembly.
2. Remove the corrugated tubing from the ventilator port and attach it to the bag arm. Set the
BAG/VENT switch to BAG and close the APL valve fully clockwise.
3. If 5500 equipped, set the 5500 high pressure alarm to 50 cmH20 and sustained pressure
alarm to 30 cmH20.
4. Flow O2 until 40 cmH2O is observed on the GMS Gauge.
5. Reduce O2 flow to 200 ml/min.
6. Verify that the GMS gauge does not drop in 10 seconds.
7. If 5500 equipped:
a. maintain pressure and verify that the sustained pressure audio and visual alarms occur
within 15 seconds.
b. silence the sustained pressure alarm.
c. slowly increase the O2 flow to 50 cmH20.
d. verify that the high pressure alarm activates at 50 ± 5 cmH2O. Verify that the bar graph
matches the applied pressure.
8. If the bi-annual (24 month) GMS APL assembly inspection was performed at this visit, skip
this test. Increase the O2 flow to 3 L/min. Verify that the pressure relieves at between
65 and 76 cmH2O.
9. Slowly turn the APL knob counter-clockwise in 1/4 turn increments.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 6 of 15 12/19/95 Number: 3.00

10. Verify the smooth operation of the GMS gauge as the pressure drops with each turn of the
APL valve.
11. Verify that the tactile indication ceases at approximately 30 cmH2O and that the Sustained
Pressure alarm resets.
12. Continue turning the APL knob until it is fully counter-clockwise.
13. If the bi-annual (24 month) GMS APL assembly inspection was performed at this visit, skip
this test. Verify that the gauge indicates 1-5 cmH2O.
14. Reconnect the bellows assembly to the absorber.

J. Waste Gas Interface Valve

1. Functional Check
a. Note the position of the interface needle valve.
b. Close the interface needle valve.
c. Adjust the O2 flow to 6 L/min.

NOTE: For passive systems, cap the evacuation output port then proceed.

d. Observe for the following:


1) In less than 1 minute, the evacuation bag should fill and the positive pressure valve
should relieve.
2) The absorber gauge should not indicate a pressure rise of more than 4 cmH2O.
e. Reduce O2 to minimum.

NOTE: For a passive system, reattach evacuation output tubing. Step f need not
be performed.

f. Open the needle valve one full turn.


1) Occlude the absorber exhalation port using the test plug.
2) Negative pressures should not show on the absorber gauge.
3) Verify that the evacuation bag is completely collapsed, the negative relief valve
has relieved, and the absorber gauge is at 0.
g. Return the needle valve to the position noted in step a.

K. PEEP Valve - New style - if so equipped

New style PEEP valves started with serial number AAXS01000. If you find an old style valve, call
the Anesthesia Complaint Coordinator in Madison, at 1-800-521-0086, for disposition.

1. Pre-Operative Checkout
a. Check that the scale and knob labels are legible. If not, replace PEEP valve assembly.
b. Check that the instruction label is legible.
2. Inspection
a. Remove the PEEP Valve from the absorber.
b. Disassemble the lower section of the PEEP Valve and remove the blue check disk.
c. Verify that the pin moves side to side and freely in and out. Any restriction of movement
indicates that cleaning is required.
d. Reassemble the valve and reinstall it on the absorber.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 7 of 15 12/19/95 Number: 3.00

3. Testing
With the patient circuit attached, the ventilator turned "OFF", and the absorber switched to
the ventilator mode, perform the following:
a. Remove the gas supply hose from the common gas outlet of the machine.
b. Attach one end of a 22 mm corrugated hose to the common gas outlet, attach the other
end to the patient outlet of the "Y" connector in the patient circuit.
c. Turn the PEEP Valve fully clockwise to the maximum PEEP setting.
d. Press the flush button while observing the absorber pressure gauge. Typically, the
indicated pressure will be approximately 40 cmH2O. If the pressure exceeds 60 cmH2O,
the valve must be serviced or replaced.
e. Perform the following test:
1) Flow 3 L/min of oxygen with the PEEP Valve knob turned fully clockwise.

NOTE: If the pressure reading is not in the recommended range, the spring may
need replacing.
2) Observe the pressure on the absorber pressure gauge. It should be between
16 and 27 cmH2O.
3) Return the PEEP Valve to the "OFF" (fully counterclockwise) position and reduce
the O2 flow to minimum.
4) Remove the 22 mm corrugated hose from the common gas outlet and the "Y"
connector. Reconnect the gas supply hose and any circuit connections
previously removed.

IV. 7000 VENTILATOR TESTS

A. Pop Off Valve/ Bellows Retention Test


1. Connect the corrugated hose between the absorber inhalation and exhalation ports.
2. Set the BAG/VENT switch to VENT (GMS and 21 Absorber with selector switch).
3. Fill the bellows using oxygen flush.
4. Increase the O2 flow from 200 ml/min to 10 L/min and ensure that the circuit pressure
remains between 1 cmH2O and 5 cmH2O.
5. With the O2 flow set to 10 L/min, verify that you cannot turn the O2 flow control knob more
than 2 additional turns counterclockwise.
6. Decrease O2 flow to minimum.
7. Remove and occlude the bellows exhaust outlet tubing at the waste gas scavenging
interface valve connection.
8. Use the O2 flow control to obtain a steady pressure of 15 cmH2O on the absorber pressure
gauge.
9. Check the bellows for proper retention on the bellows base. Replace the bellows as
necessary or if 15 cmH2O cannot be obtained.
10. Remove the occlusion from the exhaust outlet and reconnect the hose.

B. Bellows Drop Test


1. Ensure the bellows is still at the top of the housing.
2. Set the BAG/VENT switch to the BAG position (For absorbers without a selector switch,
turn the O2 flow off).
3. The bellows must not drop more than 100 ml in one minute. For ABA Bellows, 100 ml is the
top of the tapered white arrow printed on the bellows housing.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 8 of 15 12/19/95 Number: 3.00

C. Breathing System Pressure Leak Test (Absorber, bellows assembly, manifold, and drive
gas hose)

NOTE: Before performing this test, the bellows must be at the bottom of the housing.

1. Set the BAG/VENT switch to VENTILATOR position (GMS and 21 Absorber with selector
switch).
2. Disconnect the 17 mm Drive Gas hose at the back of the ventilator and occlude the hose
with a test plug.
3. Observe the pressure gauge while slowly increasing the O2 flow until the pressure reaches
30 cmH2O. Quickly reduce the flow until the pressure stabilizes.
4. The O2 flowmeter reading is the total system leak rate. Verify that the leak rate is not
greater than 225 ml/min.

NOTE: If this test is performed with a patient circuit with an occluded 'Y' piece, the
allowable leak rate is 300 ml/min.

5. Remove the corrugated hose, the test plug from the drive gas hose, and occlusions on any
gas sampling connections. Reconnect the drive gas hose.

V. ADDITIONAL SYSTEM CHECKS


A. O2 Monitor Calibration, if equipped
1. Calibration/Alarms Testing
a. Note the Low and High O2 alarm settings.
b. Connect the O2 sensor assembly and test device to the Common Gas Outlet.
c. Verify that all O2 display segments are working. Verify that the backlighting is ON (5120
only).
d. Set the high O2 Alarm to 95%.
e. Introduce the O2 sensor assembly to 100% concentration of oxygen and allow the
display to stabilize.
f. Calibrate to 100% O2 and verify that the High O2 alarm activates (audio and visual).
g. Remove the O2 sensor assembly from 100% Oxygen and expose it room air. Verify that
the High O2 alarm cancels.
h. Set the Low O2 alarm to 25%. Verify that the Low O2 alarm activates, and verify that the
monitor stabiizes at 21% ± 3% within 3 minutes.
i. Press alarm silence and verify that the audible alarm is silenced.
j. Set the Low and High O2 alarm to their original setting.

B. Proportioning System
WARNING: Use the hospital evacuation system with the service gas-evacuator tee. If
hospital evacuation system is not available, perform service in a well-ventilated
area.

1. Every 12 months, perform flowmeter verification on O2 at the following flows: 500 ml/min,
1 L/min, and 5 L/min.
2. Use the N2O Flow Control Valve and perform the tests in Table A progressively from low to
high flows. Do not overshoot any setting. Allow the System O2 monitor to stabilize between
readings.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 9 of 15 12/19/95 Number: 3.00

TABLE A

Set the N2O The oxygen flowmeter


Flow to (L/min) must read between the
following readings:

MIN (L/Min) Max System O2


(L/min) Monitor

0.9 0.24 0.36 20-30%


6.0 1.58 2.44 20-30%
9.0 2.37 3.88 20-30%

3. Set the N2O flow to 10 L/min. Verify that you cannot turn the N2O flow control knob more
than 2 additional turns counterclockwise.
4. Increase O2 to 6 L/min.
5. Use the O2 Flow Control Valve and perform the tests in Table B progressively from high to
low flows. Do not overshoot any setting.

TABLE B
Set the O2 Flow Cont. The N2O flow must then read between
Valve to (L/Min) the following readings:

Min N2O Max N2O Oxygen


(L/Min) (L/Min) Monitor

3.0 7.36 11.41 20-30%

1.0 2.46 3.80 20-30%

0.5 1.23 1.90 20-30%

0.3 0.74 1.14 20-30%

6. Decrease N2O to minimum, the O2 float should not decrease at this time.
7. For optional gas(es) use the appropriate table shown below.

TABLE C

Oxygen He Oxygen
N2 Monitor Range

4 L/min 2 L/min 61-71%


4 L/min 4 L/min 45-55%
4 L/min 8 L/min 28-38%

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 10 of 15 12/19/95 Number: 3.00

TABLE D

Oxygen Air Oxygen


Monitor Range

4 L/min 3 L/min 61-71%


3.5 L/min 6 L/min 45-55%
1.5 L/min 8 L/min 28-38%

TABLE E

Oxygen Heliox (25% Oxygen, Oxygen


75% Helium) Monitor Range

4.0 L/min 3.0 L/min 63-73%


3.5 L/min 6.0 L/min 48-58%
1.5 L/min 8.0 L/min 32-42%

8. Set all gas flows to minimum. Verify that the O2 flowmeter indicates 200 ml ± 25.

C. Vaporizer Test

WARNING: Follow the appropriate agent evacuation/collection safety measures. Use


the hospital evacuation system with the service gas-evacuator tee. Use
the canister when appropriate vacuum systems are unavailable (Vapor
Absorption Kit).

1. Set the oxygen flow rate to 6 L/min.


2. Turn on a vaporizer. Slowly advance the vaporizer knob through a percentage range of 0 to
1.0% while observing the oxygen flowmeter.
3. The oxygen flow rate should not drop more than 1 L/min at any position.
4. Repeat Steps 2 through 3 for each of the vaporizer positions on the manifold.
5. Set the O2 flow to minimum.

D. O2 Flush Valve Test


1. Ensure that the flush valve button guard is secure.
2. Depress O2 flush valve.
3. Ensure a high rate of flow is heard.
4. Ensure the low oxygen supply alarm does not activate.
5. Remove all test equipment.
6. Reattach the fresh gas hose to the common gas outlet.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 11 of 15 12/19/95 Number: 3.00

E. Bi-annual Ventilator Internal Component Check - Every 24 months only


1. Replace the battery. Write the replacement date on the battery.
2. Verify the integrity of the AC wiring.
3. Check the condition of the internal tubing and the Exhalation Valve; replace as necessary.
4. Verify that the correct fuse is installed. Units with a 3/4 amp label should have a 3/4 amp
fast acting fuse. Units with no fuse label should have a 3/8 amp time delay fuse.

NOTE: The 3/8 Amp fast acting should not be used for any future replacements.

NOTE: The replacement fuse for the 7000 Multi-Voltage Ventilator is a 1 Amp 250 Volt
5X20mm fast acting fuse.

F. Ventilator Flow Delivery Tes


1. Disconnect the ventilator drive gas hose from the bellows base assembly. Connect a
respirometer to the free end of the drive gas hose. (Use adapters as required)
2. Pressurize the patient pressure sensing line (distal sense) connected to the ventilator to
20 cmH2O and maintain pressure. (This eliminates a low airway pressure alarm during the
test).
3. Turn the ventilator power switch ON.
4. Set the ventilator controls per Table F. Allow three cycles to occur. Ensure that the volume
is the same each cycle by re-zeroing respirometer between each cycle.
5. Compare the respirometer readings with the tidal volume limits in Table F.

TABLE F

CONTROL SET RESPIROMETER VOLUME


READINGS

MV R 1:E MIN NOMINAL MAX

5 10 1:2 360 ml 500 ml 690 ml

10 20 1:3 430 ml 540 ml 670 ml

30 30 1:1 940 ml 1090 ml 1280 ml

6 30 1:2 160 ml 200 ml 250 ml

15 10 1:2 1290 ml 1620 ml 2070 ml

NOTE: Respirometer readings have been adjusted to compensate for respirometer response
curve.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 12 of 15 12/19/95 Number: 3.00

G. Alarm Tests
1. Push the lamp test button. All the lamps must light and the audio alarm must sound.

2. Set the controls to: MV-15, R-15, I:E-1:2.5


Vary the MV control as indicated; verify the alarm condition(s).

MV Alarm Condition(s)
15 L/Min None
15-20 L/Min Actual I:E Less Than Dial Setting
20-25 L/Min Actual I:E Less Than Dial Setting and
Set Volume Not Delivered.
15 L/Min None

3. Relieve pressure on the patient pressure sensing line.


4. The "Low Airway Pressure" alarm must sound within 3 breath cycles.

H. Ventilator Electronic High Pressure Relief System.

NOTE 1: The electronic high pressure relief system should be present in all 7000
ventilators. If it is not present, retrofit the ventilator using the high pressure
relief kit. Charge the kit to warranty.

NOTE 2: There are two parts to this procedure, the first is verifying that the alarm
threshold is triggered at the correct pressure. The second test, which is done
later in conjunction with other alarm tests, checks that the alarm is actually
triggered at the patient.

Alarm Threshold Test


1. Apply a pressure greater than 65 cmH2O (approx. 48 mmHg) to the patient pressure (distal
sense) inlet.
2. When the inlet pressure is 65-75 cmH2O (48-55 mmHg), verify that the "Set Volume Not
Delivered" alarm lights and an audio alarm sounds.
3. With over-pressure present during the inspiration phase, the driving O2 is vented from the
control unit base until the inspiratory phase is complete.
4. Maintain the over-pressure condition and ensure that after 2-3 breaths, the ventilator goes
into the "Ventilator Failure" mode and will continue to vent drive gas out of the control unit
base during the inspiratory phase.
5. Remove pressure from the inlet. Verify that the "Set Volume Not Delivered" alarm resets but
that the "Ventilator Failure" alarm remains latched.
6. Turn the unit OFF and then ON to reset the unit.

I. Mechanical Relief Valve Test


1. Set the controls to MV-2, R-10, I:E-1:1
2. Apply 20 cmH2O to the patient pressure sensing line.
3. Remove the respirometer and attach an approved pressure measuring device.
4. Verify that the pressure starts relieving between 45-65 cmH2O (33 - 48 mmHg)during
inspiration.
5. Remove the test device. Turn the ventilator OFF and reconnect the drive gas hose.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 13 of 15 12/19/95 Number: 3.00

J. Manual Ventilation Test


1. Connect a complete patient circuit, with a test lung, to the absorber. Maintain at least
2 L/min fresh gas flow so that the bellows remains fully extended.
2. Reattach the patient pressure sensing line to the absorber.
3. Turn the ventilator ON. Push the manual cycle button during the expiratory cycle to verify
that the ventilator cycles.

K. Sigh Test
1. Set the controls to: MV-15, R-40, I:E-1:2
2. Turn the sigh switch ON and start stop watch. Count breaths beginning with the first
inspiratory cycle after the sigh is switched ON. On the 64th breath, ±2 breaths
(approximately 96 seconds) a larger breath, (approximately 150% of set volume) must be
delivered.
3. Reset the all controls to its their original positions.

VII. VOLUME MONITOR, if equipped

A. Pre-Inspection
1. Remove the TVX cartridge and sensor clip and inspect for cracks and physical damage.
2. Inspect the battery and compartment for damage. Replace the battery as necessary, or
every 24 months. Record the replacement date on the battery.
3. Turn Monitor ON. Verify all segments are lit and the backlighting is on.
4. Verify the function of Adult/Ped switch. Switch to Ped. and verify that PEd is displayed and
the green LED is on. Switch to Adult and verify Adu is displayed.

B. Functional Check
1. High Minute Volume Alarm Test
a. Set Ventilator controls to MV 6, Rate 10, I:E 1:1.
b. Set High Minute Volume Alarm to 4.
c. Verify the High Minute Volume alarm activates.
d. Reset High Minute Volume alarm to its original position.
2. Low Minute Volume Alarm Test
a. Set the Low Minute Volume alarm to 7.
b. Verify the Low Minute Volume alarm activates.
c. Press alarm silence to silence the alarm.
d. Reset the Low Minute Volume alarm to its original position.
3. Apnea Alarm
a. Turn the Ventilator OFF.
b. Verify the following occurs:
1) After 30 sec., APNEA message appears on the display, LED begins to flash, one
low-high tone sounds.
2) After 60 sec., two low-high tones sound.
3) After 90 sec., three low-high tones sound.
4) After 120 sec., continuous low-high tones sound.
c. Turn the monitor OFF.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 14 of 15 12/19/95 Number: 3.00

4. Flow Test
a. Connect the TVX cartridge and sensor clip to common gas outlet with the gas flow arrow
pointing away from the machine.
b. Set the O2 flow to 6 L/min.
c. Turn the monitor ON.
d. Verify that the volume displayed is 6L/min ± 2L/min.

5. Reverse Flow Test


a. Reverse the direction of the sensor clip on the TVX cartridge.
b. Verify that the reverse flow alarm activates.
c. Return the sensor clip to it's original position on the common gas outlet.
d. Simulate 3 breaths by occluding the TVX cartridge.
e. Verify that the reverse flow alarm cancels.
f. Turn the monitor OFF and reinstall the TVX cartridge and sensor clip on the absorber.
g. Reduce the O2 flow to minimum.
h. Remove the test lung.

VIII. ELECTRICAL SYSTEM

A. Monitor Battery Checks - Excel w/ Pod Only


1. Ensure power is off before replacing batteries.
2. Replace the 7.2 Vdc battery located on the sensor interface board every 24 months. Record
date on battery.
3. If Paw monitor equipped, replace battery every 24 months, record date on battery.
4 Turn the system master switch ON. Disconnect power cord.
5. Turn ventilator and all battery backed monitors ON.
6. Verify that after 2 minutes
a. 7000 Ventilator - The audio alarm should remain fully audible for the complete 2 minute
period.
b. Excel disconnect audio alarm cycles on and off twice and red LED on the pod is
flashing.
c. Volume, Oxygen and Paw monitors are operational, if equipped.
7. Turn ventilator and all monitors OFF.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


07/2004
AA Excel or Excel SE w/ 7000 Ventilator 15 of 15 12/19/95 Number: 3.00

IX. ELECTRICAL SAFETY CHECKS

A. Electrical Safety Tests

NOTE: Perform electrical Safety Tests every 6 months, or if the covers to a machine(s) are
removed.

NOTE: If tests are conducted with power from an isolated electrical power system, conduct
the tests as detailed below and record "ISOLATED POWER SYSTEM" on the MSR
detail block. Do not use an electrical jumper plug to defeat the isolation of the
power system.

1. Disconnect all pipeline hoses from the medical gas outlets.


2. Disconnect all electrical monitors and accessories from the unit under test.
3. Plug the unit under test into the safety analyzer. Plug the analyzer into a 110 volt electrical
outlet.
4. Set the safety analyzer to measure line voltage. Verify normal polarity is displayed.
5. Connect the grounding clip assembly to the safety analyzer.
6. Ground Resistance Check
a. Perform a ground resistance check. Use the electrical safety analyzer to measure the
resistance between the ground pin on the line cord plug and exposed metal of the
common gas outlet and frame. The ground resistance must be equal to or less than
0.10 ohms.
7. Chassis Leakage Current:
Measure the chassis leakage current for the following configurations:
normal polarity - neutral closed - ground closed
normal polarity - neutral closed - ground open
reverse polarity - neutral closed - ground closed
reverse polarity - neutral closed - ground open
The chassis leakage current must be less than 100 microamps in all cases.
8. Repeat steps 3-7 for each Datex-Ohmeda monitor and accessory on contract removed in
step 2.

X. FINAL MACHINE CHECKS

A. Disconnect all test devices and reconnect pipeline hoses, monitors and accessories that were
previously removed. Ensure the following:
B. All vaporizers are OFF.
C. The fresh gas hose is connected.
D. All flow control valves are at the minimum stop.
E. The APL valve is Open (minimum setting).
F. Verify correct operation of lightbar (if equipped).
G. All panels and tray top are attached.
H. The pipeline hoses are connected.
I. All cylinder gauges are at zero.
J. All units are plugged into an appropriate power source.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


App A 07/2004
AA Excel or Excel SE w/ 7000 Ventilator 1 of 1 12/19/95 Number: 3.00

Appendix A
Required Tools

Item # Description Part Number Equipment


1 22mm "T" Manifold 0212-0763-100 Excel/5120
2 Cello-Seal 0220-5160-300 GMS
3 Digital Multimeter 7000-0000-023 Excel
4 Disposable Gloves (MD/LG) 7000-0000-010 All
5 Disposable Gloves (SM) 7000-0000-009 All
6 Electrical Analyzer BET 300AD / Dale 600 0175-2305-000 All but Abs.
7 Electronic Leak Detector 7000-0000-004 Excel
8 Flowmeter Verification Device 0309-1324-810 Excel
9 Evacuation Service Kit 0175-2278-000 Excel
10 Low Pressure Leak Test Device 0309-1319-800 Excel
11 O2 Monitor 0304-2178-800 Excel
12 Meriam Smart Manometer 0 - 100 psi Call Cal Lab / Purchase All
Locally
13 High Pressure Manometer Fitting Kit See MP 1011 All
14 Meriam Smart Manometer 0-1000 mmHg Call Cal Lab / Purchase All
Locally
15 Low Pressure Manometer Fitting Kit See MP 1011 All
16 Brush 1400-3007-000 GMS (PEEP)
17 Krytox 1001-3854-000 GMS
18 40 " Corrugated Tube, Rubber w/22mm 0211-9005-800 GMS, 5120
bushings
19 Sealant, Silicone Rubber 0220-5251-300 GMS
20 Ring Pliers 0175-2364-000 GMS
21 Test Plug 2900-0001-000 GMS, 7000
22 Torque Screwdriver 0175-2038-000 GMS
23 Torque Adapter - 5/32" hex bit 0175-2385-000 GMS
24 Torque Adapter - 3/16" hex bit 7000-0000-056 GMS
25 Drive Gas Hose, 10" 0211-0118-300 7000
26 Respirometer 1603-3000-000 7000
27 Stop Watch 0175-1629-000 7000, 5500
28 Test Lung 0219-7210-300 7000

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


App B 07/2004
AA Excel or Excel SE w/ 7000 Ventilator 1 of 3 12/19/95 Number: 3.00

Appendix B
Required Parts

Model Description Part Number Frequency

5120 O2 Monitor
1. Outer O-ring 0210-0503-300 As Required
2. Inner O-ring 0210-0499-300 As Required
The batteries and cartridge are user replaceable and not covered by an MA contract.

5400/5420 Volume Monitor


1. Battery 0690-1000-311 24 Months

5500 Pressure Monitor


1. Battery 6021-0000-005 24 Months

7000 Ventilator
1. Battery 1502-3016-000 24 months
2. Bellows, Adult ABA 1500-3378-000 As Required
3. Bellows, Pediatric Non ABA 0229-1018-700 As Required
4. Diaphragm and Seat Assy, ABA 1500-3377-000 As Required
5. Disk/ Ring /Bumper Assy, ABA 1500-3381-000 As Required
6. O-ring, large, pop-off valve Non ABA 1500-3267-000 As Required
7. O-ring, small, pop-off valve Non ABA 0210-0593-300 As Required
8. O-ring, drive gas nipple 0210-0593-300 As Required
9. Retaining ring, Ped Bellows Non ABA 0229-0036-300 As Required
10. Seal, ABA 1500-3359-000 As Required
11. Seat, Free Breathing Valve 0207-5590-100 As Required
12. U Cup Seal Non ABA 0210-0784-300 As Required

GMS Absorber
1. APL Valve Service Kit 0216-6785-870 As Required
1 ea Instr Instal GMS Apl Repl Parts
1 ea Label Set ABS GMS
1 ea Shaft Needle
1 ea Seat Valve APL
2 ea O-Ring 1-1/16OD 15/16ID
1 ea Disk APL 1.03D BL Delrin GMS ABS
2 ea Washer .946ID 1.1960D .016T Brass
2 ea Ring Retainer 15/16Shaft Truarc #5103-93H ST ST
1 ea Grommet APL Valve
1 ea Ring Locking APL Valve
1 ea Spring, Com 12.70D 38.1L
2. Canister Gaskets 0210-1214-100 As Required
3. Canister Seals 0210-1218-300 As Required
4. Disk Valve - PEEP 0210-5297-100 As Required

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


App B 07/2004
AA Excel or Excel SE w/ 7000 Ventilator 2 of 3 12/19/95 Number: 3.00

Model Description Part Number Frequency

GMS Absorber
5. Disk Retainer - PEEP 0207-1642-542 As Required
6. Dome - PEEP 0207-1638-100 As Required
7. Instruction Tab - PEEP 1400-3006-000 As Required
8. Manifold Gasket with Risers 0229-2077-100 As Required
9. Washer Pin Gasket 0202-0094-300 As Required
10. O-Ring - Bag Arm Drain 0210-0483-300 As Required
11. O-Ring - Gauge 0210-0566-300 As Required
12. O-Ring - Lower Dish Drain, New Style 0210-0594-300 As Required
13. O-Ring - Lower Dish Drain, Old Style 0210-0481-300 As Required
14. O-Ring 0210-0574-300 As Required
15. Pivot Bracket 0229-2085-200 As Required
16. Retaining Ring - PEEP 0203-5249-300 As Required
17. Return Tube 1400-3009-000 As Required
18. Screen 0214-7167-535 As Required
19. In/Exhalation Check Valve Service Kit 1400-8011-000 As Required
2 ea O-Ring 2-1/16OD 1-7/8ID
1 ea O-Ring 9/16OD 7/16ID
2 ea Disc Chk Vlv Rvsbl
2 ea Pad Die Cut 2-1/2 X 1-1/4
2 ea Dome Valve Check
2 ea Cup Seal
1 ea O-Ring .437ID .625OD
20. 2 ea Cup Seal 0229-2007-100 24 Months
21. Service Kit - Selector Valve GMS 0216-6782-800 As Required
1 ea Bearing .130 DIA X. 250L
1 ea Bearing Nylon .250ID
1 ea Bushing Selector Valve
1 ea O-Ring .500 ID .625 OD .062W EPR 80 Duro
1 ea O-Ring 1/8X1/4X1/16 EPR
1 ea O-Ring .250ID .375OD .062W EPR 78 Duro
1 ea Roller
1 ea Sleeve Pivot Pin
1 ea Instr Instal Kit Selector Valve & Gauge Sensing
1 ea Spring Com
2 ea Ring Truarc .125 Shaft #5133-12H E-Ring
1 ea Adapter Barbed End 1/16ID Tube X 6-32 THD
1 ea Net, HT 6-32, Full MCH, SST
1 ea Pin Pivot

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


App B 07/2004
AA Excel or Excel SE w/ 7000 Ventilator 3 of 3 12/19/95 Number: 3.00

Model Description Part Number Frequency

GMS
1 ea Retainer Cover
1 ea Shaft Retaining Ring
1 ea Bracket Support
1 ea Shaft Selector Valve
1 ea Bracket Pivot
2 ea Washer .438OD X .128ID X .065T Delrin
1 ea Tube Connector Patient Press 1/16ID Silicone Med
1 ea Washer No 6 Flat SST
22. Spring - PEEP 1400-3023-000 As Required
23. Thrust Pin - PEEP 0207-1635-500 As Required
24. Upper Gasket 0210-1057-100 As Required
25. Washers - Switch Assy. 0402-1044-500 As Required

Excel
1. Battery 7.2V 1502-3016-000 24 Months
2. Gasket, Yoke Cylinder Valve 0210-5022-300 As Required
3. Index Pin 0201-0503-500 As Required
4. Inlet Strainer Nipple w/ Filter 9914-6380-000 As Required
5. O - Ring Tec Port Manifold 0210-0736-300 12 Months

Waste Gas Interface Valve


1. New Style Waste Gas Scavenging Valve Assembly 1001-8889-000 As Required

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures

Mfr. Description Page Date Issued: Revision Date:


App C 07/2004
AA Excel or Excel SE w/ 7000 Ventilator 1 of 1 12/19/95 Number: 3.00

Appendix C *
Maintenance Procedure Revision Log

Description of Changes Originator Revision Date New Revision Number


1) Revised Appendix A Required Tools M. Rogers 07/2004 3.00
2) Revised the confidentiality statement
3) Deleted 21 absorber appendix

* Note: This document has been under revision control, but a revision history has not
been required to be included with this procedure. From this revision going forward, this
log will be updated to inform the user of changes made to this document.

MP 1017 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSIDIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.

You might also like