Instruction Manual: 2700 Series Benchtop Meters
Instruction Manual: 2700 Series Benchtop Meters
Instruction Manual: 2700 Series Benchtop Meters
Each meter includes a convenient slide-out calibration card for quick reference. All
models (except for DO 2700) include an electrode arm and metal bracket which
can be easily attached to the left or right side of the meter according to your
preference.
The 2700 series benchtop meter replaces our popular 1100 and 2100 series
meters that were introduced in 2001.
Please note that this manual serves all 2700 models, so not all sections of the
manual will be relevant to each model. The chart below lists the parameters
served for the various models:
We take great pride in every instrument we manufacture and hope this one
serves you well.
1
2. Getting Started
Auto Read is enabled. When flashing, the instrument has detected a ‘Stable’
reading and locked the value. Press MEAS to resume live reading. See
Section 3—Auto Read to disable this feature, or go to
SETUP…SYSTEM…AUTO READ.
Based on the stability criteria settings in System Setup, the instrument has
detected a stable reading.
Password Protection: Enabled. Requires password for all
calibration and setup menus
Print Setting Timed interval is active; data is being output at regular intervals.
“High Alarm”, “Low Alarm”, or “Cal Due” alarm limit had been reached.
High/Low alarms also have an audible notification..
2
Navigation of Tabs
3
Meter Connections
pH 2700
PC 2700
ION 2700
DC Power supply
4
3. System Setup & Configuration
Use the System Setup to customize operation of your 2700 series meter.
Press from the measurement screen and then press ENTER when
SYSTEM is highlighted to access these settings.
The following settings can be customized for each model:
STABILITY PRINT SET
STABILITY
DATE & TIME
CRITERIA
AUTO READ PASSWORD
BACKLIGHT CLEAR DATALOG
DATALOG FACTORY RESET
PRINT SETUP CONTRAST
PRINT FORMAT
Stability
To aid with measurement and calibration, your meter has a unique feature that
provides an obvious visual indication when the reading becomes stable. If the
Stability feature is enabled, the digits on the primary reading will be faded when
unstable, becoming solid when stable. If this feature is disabled, the primary
reading will always be solid. The image below on the left is an unstable reading
during pH Measurement mode with Stability enabled. The image on the right
shows a stable reading.
5
Stability Criteria
The 2700 allows adjustment of the Stable indicator which relates to how fast and
frequent it appears. There are three levels of adjustment; SLOW, MEDIUM, &
FAST.
To display a ‘Stable’ reading more quickly and more often, use “FAST” setting.
MEDIUM or SLOW is recommended for most applications. When Auto Read is
enabled, a stability criteria setting of “SLOW” is recommended to for best results.
SLOW: The Stable indicator will take longer to appear and will appear less
frequently. Use this setting if you want the best results and don’t mind waiting
longer to get them.
MEDIUM: This is the factory default setting. It provides a balanced response
which works best for most applications.
FAST: The Stable indicator will appear quickest of any setting. If you find that the
Stable indicator frequently appears and disappears, you may want to select a
slower setting. This setting is not recommended when AUTO READ is enabled. It
is also not suitable for most non-refillable electrodes which are generally slower to
respond.
Auto Read
If AUTO READ is enabled, the measured value will automatically lock
when the Stable indicator appears. A flashing “AR” icon confirms this
condition. To resume a live reading again, it is necessary to press MEAS
each time Stable appears. Disable this feature to continuously view the active
reading.
Backlight
Select “PERMANENTLY ON” on to keep the backlight on at all times that the
meter is powered on. This will offer the best visibility but consumes the most
power.
Select “ON WITH KEY PRESS” to turn on the backlight for a specified time after
any key press. Select “0” to keep the backlight off at all times that the meter is
powered on.
6
Data Log (storing data to internal memory)
Select “SINGLE” to manually save one point at a time. Press the “MEM IN”
function key any time during measurement to save data.
Select “TIMED” to automatically save data at selectable intervals— ranging from
3 to 3600 seconds. Data will be collected until the data limit is reached or until
“SINGLE” data log is selected. This feature is useful for gathering data from a
single sample over time. Note that “MEM IN” is not available during TIMED
setting. As data is collected, the display will indicate how much memory has been
used.
To view stored data, use “MEM OUT”. Note: the more data that has been
acquired memory, the longer the data will take to display.
Print Setup
Adjust the RS-232 output settings for baud rate, parity, data bit, and stop bit to
match your printer or computer settings as necessary.
Print Format
When utilizing the RS-232 output; choose “TEXT/PRINTER” option to send the
data in an easily viewable format—best for printing.
Choose “CYBERCOMM” option to send the data as a comma separated value
(CSV)—best for exporting data into spreadsheet software.
Print Set
This is identical to the Data log function, but data is sent via RS-232 output
instead of internal memory.
Select “SINGLE” to manually output one value at a time—best for printing or
collecting data to your PC from many samples.
Select “TIMED” to automatically output live data at selectable intervals— ranging
from 3 to 3600 seconds. Data will be output until “SINGLE” Print set is selected.
This feature is most useful for gathering data for use with software. Unlike the
Data Log setting, data can be saved to memory using “MEM IN” during TIMED
setting.
7
Date & Time
Setting the correct date and time is required for GLP and will apply to power off,
measurement, data log, and print functions. Instrument has battery backup to
retain date/time settings upon power loss. Factory reset will not apply to date and
time setting once it has been set. Changes related to daylight savings time must
be manually entered.
Date Format: Select (MM DD YY) or (DD MM YY)
Time Format: Select 12Hrs (AM/PM) or 24Hrs
Password
Select “ENABLE” to restrict access to Calibration and Setup modes. When
password protection is enabled, password entry is required before performing any
calibration, or making changes to the setup mode. Setup parameters can be
viewed, but can not be changed without correct password entry. The password is
a user selectable number from 1 to 99999.
Select “DISABLE” if password protection is not desired.
The meter does not allow you to edit setup parameters or perform a new
calibration unless you enter the correct password. If an incorrect password is
entered 3 consecutive times, the meter returns to measurement mode.
In the event the password can not be recovered, a password can be provided via
a written request to Eutech Instruments/Oakton Instruments. The instrument
serial number and your contact information are required.
Clear Datalog
Select “YES” to delete all saved data that was manually or automatically stored
into memory. This step is most commonly required when the “TIMED” data log
interval is not turned off and as a result, fills the memory to capacity. When the
memory is full, deleting the saved data is required in order to save additional
data/ The 2700 will never erase old data in favor of new data when the memory is
full.
If you wish to view, print or send the saved data to a PC before deleting, use the
“MEM OUT” function from measurement mode.
8
Factory Reset
Select “YES” to reset the 2700 to the factory default settings except; Date & Time,
Temperature calibration, and data stored in memory.
Contrast Adjustment
Optimize the contrast setting of your 2700 display for best visibility in your
surrounding lighting conditions. Test various contrast settings for best results.
This setting will be applied to both backlight and non-backlight conditions.
9
4. Setup pH & mV
Use Setup pH or mV mode to customize these parameters. Note: mV setup offers
Sample ID and Alarm setting only.
Press from the measurement screen and then press ENTER when pH
or mV is selected.
Sample ID
This is a user selectable number from 1 to 99999.
Incorporating a sample ID to identify one or more
data points is useful to distinguish data that is
saved into memory or sent to a PC or printer. Use
the up/down arrow keys to adjust the values and
left/right arrow keys to move the cursor.
10
Cal Points
Specify the number of pH calibration points you intend to
calibrate with. Select the number of calibration points
from 1 to 5 with preset buffer groups or from 2 to 5
points with custom buffer group.
For example, if you have selected the USA buffer group
which contains 5 standards, but you will only be using 3
standards (pH 4, 7, and 10), selecting 3 calibration
points here will speed up the calibration process—the 2700 will automatically
recognize calibration is finished after the 3rd point has been completed.
Resolution
Select your desired pH resolution;
0.0, 0.00, or 0.000. Tip: Don’t choose more than you
really need—0.00 is the factory default setting since it is
most commonly used. Choosing 0.000 will lead to more
“unstable” conditions and cause a delay in obtaining the
final reading as the last digit (which is often not needed)
changes.
Alarm
The 2700 offers a visual and audible alarm to alert you
when the High or Low values that have been set from
this menu are exceeded.
Cal Due
When enabled, the “Cal Due” indicator blinks if the
number of days since the last calibration has been
exceeded. Set the number of days from 1 to 31.
11
5. pH Calibration (with preset buffer group)
For best results, periodic calibration with known accurate standards is
recommended prior to measurement. Calibrate with standards that bracket your
intended measuring range while including a neutral point (7.00, 6.86, or 6.79). For
example, if you expect to measure samples from pH 6.2 to 9.5, calibration with
4.01, 7.00, and 10.01 will work well.
The 2700 series meters can be calibrated with up to 5 preset or custom buffers
(up to 6 with DIN buffer group). The non-volatile memory retains all calibration
values upon meter shut down.
The 2700 series automatically recognizes the following pH calibration buffers in
the groups listed below. See Section4—Buffer to select a different buffer group;
2) Rinse the pH and ATC electrodes with clean water then submerse in neutral
pH buffer—the 1st calibration point must be either 7.00, 6.86, or 6.79
depending on the buffer group used.
12
If the meter is password protected, you will be prompted to
enter a password. See Section 3—Password.
13
6. pH Calibration (with custom buffers)
Buffer Group pH Values
CUSTOM Any 2 - 5 values, ≥ 1.0 pH unit apart
2) Rinse the pH and ATC electrodes then submerse in your custom pH buffer—
any pH value can be used.
For best results, enter the value of the buffer that corresponds
to the measured temperature. For example, pH 10 buffer is
actually 10.06 at 20°C, so adjust to 10.06 instead of 10.00 if
the temperature is 20°C.
4) Press ENTER to accept the calibration value. When successful, the accepted
value will be visible inside of a beaker icon.
14
5) Rinse the pH and ATC electrodes then submerse in next custom pH buffer.
Use any pH value that is at least 1.0 pH unit from custom pH standards that
have already been calibrated.
As soon as the first calibration value is accepted during a new calibration, all prior
calibration values are erased.
15
7. Millivolt (mV) Calibration (Offset Adjustment)
Oxidization Reduction Potential (ORP or Redox) as measured by an ORP
electrode in mV units is not a precise measurement, but is useful as a relative
indicator. As such, mV offset adjustment is not meant to enhance accuracy, but
rather to make readings comparable to a reference.
1) Turn meter on, connect an ORP electrode, and press MODE if needed for mV
(or R.mV) measurement.
2) Dip the ORP electrode into a solution with a known mV value (i.e. Zobel,
Light’s, quinhydrone, or iodidetriiodide) with brief or slow stirring.
3) Press CAL when the reading is stable. The upper display shows the active
millivolt value (mV) while the lower display shows the factory default mV value
without calibration. The example below at right shows a mV calibration that
had previously been calibrated with a 10.0 mV offset.
4) Adjust the upper display using up/down arrows, press ENTER to accept the
calibration offset. The maximum adjustable value is ±150 mV from the factory
default mV value.
16
8. Temperature Setup
Use Setup Temperature mode to desired temperature units; º C or º F.
Press from the measurement screen and then press ENTER when
TEMPERATURE is selected.
Choose º C or º F temperature units of measure.
17
9. Temperature Calibration
The thermistor sensor used for automatic temperature compensation and
measurement is both accurate and stable, so require frequent calibration isn’t
required. Temperature calibration is recommended upon electrode replacement,
whenever the temperature reading is suspect, or if matching against a certified
thermometer is desired.
1) Connect the temperature probe to the meter and place into a solution with a
known accurate temperature such as a constant temperature bath.
The temperature probe may be separate or built into a pH, Conductivity, or
Oxygen electrode. The PC 2700 has two temperature sensors available—one
built into Conductivity, and another for use with the BNC connection.
2) When the reading is stable, press CAL from any mode, then press .
3) The upper display shows the active temperature while the lower display shows
the factory default temperature without calibration. The example below at right
shows a temperature calibration that had previously been calibrated with a 3.0
ºC offset.
3) Adjust the upper display using up/down arrows, press ENTER to accept the
calibration temperature. The maximum adjustable value is ±5 ºC (or ± 9 ºF)
from the factory default temperature value.
18
10. Conductivity, TDS, Salinity, & Resistivity Setup
Use Setup mode to customize Conductivity, TDS, Salinity, & Resistivity
parameters.
Press from the measurement screen and then press ENTER when
Conductivity, TDS, Salinity, or Resistivity is selected.
Sample ID
This is a user selectable number from 1 to 99999.
Incorporating a sample ID to identify one or more
data points is useful to distinguish data that is
saved into memory or sent to a PC or printer. Use
the up/down arrow keys to adjust the values and
left/right arrow keys to move the cursor.
If you will only use one or more of the (4) calibration standards listed here,
19
automatic calibration is recommended. Otherwise, manual calibration should be
selected.
The factory default is automatic conductivity calibration. Note that automatic
calibration is not available for conductivity range 1.
Cal Points
Use SINGLE for Single-Point Calibration (SPC) to
apply a single calibration value across all ranges.
Use MULTI for Multi-Point Calibration (MPC) to
calibrate each range individually. This will restrict
an individual calibration so that it is applied to one
range only. When using multi-point calibration,
perform a calibration in each range that you
expect to use for best results.
The factory default is SPC. SPC or MPC will be shown with the calibration details
in measurement mode.
Cell Constant
The 2700 includes a probe with a nominal cell
constant (k) of 1.0. Use probes with k = 0.1 and 10
(sold separately) for improved performance in
extreme sample ranges. The factory default is 1.0
to match the included probe. Cell constant can be
adjusted from 0.010 to 10.000.
21
Alarm
The 2700 offers a visual and audible alarm to alert
you when the High or Low values that have been
set from this menu are exceeded.
Cal Due
When enabled, the “Cal Due” indicator blinks if the
number of days since the last calibration has been
exceeded. Set the number of days from 1 to 31.
22
11. Conductivity Calibration (automatic)
For best results, periodic calibration with known accurate standards is
recommended prior to measurement. Calibrate with standards that are close in
value to your intended sample(s).
When using multi-point calibration, perform a calibration in each range that you
expect to use for best results.
If the conductivity electrode has been stored dry for some time, soaking in alcohol
or clean water for a few minutes can help performance.
1) Turn meter on, press MODE if needed for conductivity measurement mode.
2) Press CAL if needed to enter calibration mode. Rinse the electrode with
clean water then submerse into your conductivity standard— either 84 µS, 1413
µS, 12.88 mS, or 111.8 mS. Provide stirring for best results.
3) Use the up/down arrows to adjust the cell constant manually or press NEXT
to keep the nominal cell constant value and proceed with a typical automatic
calibration (recommended). If adjusting the cell constant manually, press
ENTER after ‘Stable’ appears to confirm the adjustment.
24
12. Conductivity, TDS, Salinity, & Resistivity
Calibration (manual adjustment)
For best results, periodic calibration with known accurate standards is
recommended prior to measurement. Calibrate with standards that are close in
value to your intended sample(s).
When using multi-point calibration, perform a calibration in each range that you
expect to use for best results.
If the electrode has been stored dry for some time, soaking in alcohol or clean
water for a few minutes can help performance.
1) Turn meter on, press MODE if needed for the appropriate mode.
2) Press CAL if needed to enter calibration mode. Rinse the electrode with
clean water then submerse into your calibration standard.
3) Use the up/down arrows to adjust the cell constant manually or press NEXT
to keep the nominal cell constant value and proceed with a typical manual
calibration (recommended). If adjusting the cell constant manually, press
ENTER after ‘Stable’ appears to confirm the adjustment.
25
Before and after a manual calibration
5) For multi-point calibration repeat with additional standards. Press ESC to
save calibration or press NEXT to view the calibration report. Calibrate one
point per range, up to 5 points.
26
13. Ion Setup
Use Setup Ion mode to customize this parameter. Press from the
measurement screen and then press ENTER when ION is selected.
Sample ID
This is a user selectable number from 1 to 99999.
Incorporating a sample ID to identify one or more
data points is useful to distinguish data that is saved
into memory or sent to a PC or printer. Use the
up/down arrow keys to adjust the values and
left/right arrow keys to move the cursor.
Measure Unit
Select ppm, molar, or mg/L ion concentration units.
Alarm
The 2700 offers a visual and audible alarm to alert
you when the High or Low values that have been
set from this menu are exceeded.
27
Cal Due
When enabled, the “Cal Due” indicator blinks if the
number of days since the last calibration has been
exceeded. Set the number of days from 1 to 31.
28
14. Ion Calibration
The ION 2700 can measure ion concentration such as ammonia or fluoride when
using an ion selective electrode (ISE) for the specific ion of interest.
Ion calibration is required with at least two standards from (8) available values;
1.0.001, 0.01, 0.1, 1, 10, 100, 1000, and 10000.
Prepare ion calibration standards that bracket your measurement range. Follow
the ion selective electrode instruction manual for the appropriate ionic strength
adjustment and sample preparation.
For best results always begin with your lowest calibration standard value,
followed by the next lowest, and so on.
1) Connect the ISE and press MODE as needed for concentration mode.
3) The lower display is the mV reading of the ISE. Press ENTER when this
value is stable to accept the calibration point.
4) Rinse the ISE with clean water and dIp into your next lowest calibration
standard and stire. Press ENTER after “Stable” appears to accept the
next calibration point.
29
Before and after a two point ion calibration using 0.001 and 0.01 standards
6) Repeat steps 4 & 5 with additional ISE standards or press ESC to escape
and save your calibration.
At left; a three point ion calibration, at right; the Ion calibration report
To view the Ion Calibration Report at any time from the measure screen, press
REPORT.
30
15. Dissolved Oxygen Setup
Use Setup Dissolved Oxygen mode to customize the parameter. Press
from the measurement screen and then press ENTER when DO% or DO (mg/L)
is selected for percent saturation or concentration modes respectively. Note that
setup parameters differ slightly depending on which is selected.
Sample ID
This is a user selectable number from 1 to 99999.
Incorporating a sample ID to identify one or more
data points is useful to distinguish data that is
saved into memory or sent to a PC or printer. Use
the up/down arrow keys to adjust the values and
left/right arrow keys to move the cursor.
31
Pressure Cal (% saturation mode only)
The DO 2700 includes a built-in barometer that is
factory calibrated. While frequent pressure
adjustment is not required, adjustments can be
made in the Adjusted Pressure field. The
Measured Pressure is the active pressure value
measured by the DO 2700.
Note: if adjusting, “true” barometric pressure—that
has not been corrected to sea level, must be used.
Pressure reported by weather services are almost always corrected and therefore
are not appropriate for entry in the adjusted pressure field here.
32
Salinity Value (concentration mode only)
Enter the salinity value of your samples in parts per
thousand (ppt) units. Select from 0 to 50.0 ppt.
Alarm
The 2700 offers a visual and audible alarm to alert
you when the High or Low values that have been
set from this menu are exceeded.
Cal Due
When enabled, the “Cal Due” indicator blinks if the
number of days since the last calibration has been
exceeded. Set the number of days from 1 to 31.
33
16. Dissolved Oxygen Calibration
The DO 2700 utilizes a self-stirring polargraphic electrode with a built-in
temperature sensor. The electrode is designed for use in BOD bottles.
Always rinse the probe with clean water before and after each calibration/sample
measurement. When performing air calibration, gently shake remaining water
droplets from the probe’s membrane.
100 % Calibration
1) Power meter on with DO electrode attached for 5-15 minute warm
up.
2) Rinse the probe well with clean water then place into a BOD bottle
partially filled with water (100 mL or so).
34
0% Calibration
7) After completing steps 1 thru 6 for 100% saturation calibration, place your
electrode in zero oxygen solution (0.08M sodium sulfite). Continuous stirring
is not required.
3) Rinse the probe well with clean water then place into your solution with a
known dissolved oxygen concentration value (i.e. determined by titration or
another instrument).
4) Press CAL and switch the stirrer on. The upper reading is the current
measured value. The lower reading is the factory default reading for
comparison, along with salinity and temperature correction shown.
35
Additional dissolved oxygen calibration notes:
• Do not submerse the probe past the immersion limit (see below).
36
17. Transferring and Printing Data
To send data directly to a printer or PC, connect the 30X427301 / 35420-01 cable
to the RS-232 output port of the 2700. To connect to a USB port, an additional
adapter cable is required; see Section 25—Replacements & Accessories.
The 2700 can be used to:
• Print/send one or more data points directly to a dedicated printer or to
a PC. Press PRINT from measurement mode or from the MEM OUT
screen.
• Print/send the calibration report to a printer or PC. Press REPORT
from measurement mode.
• Send live data to a PC at intervals 3 seconds or higher. In the
System Setup, select a “TIMED” Print Set.
Choose the “TEXT/PRINTER” Print Format option to transfer the data in an easily
viewable format—best for printing.
Choose the “CYBERCOMM” Print Format option to send the data as a comma
separated value (CSV)—best for exporting data into software
To make changes to the printer, datalog, and output settings, see Section 3—
System Setup.
Certain 2700 models will include a complimentary software
application called CyberComm 2700. For details on
installation, connection and usage of this software, please
read the manual that is included on the software CD. If you
did not receive a CD or lost the CD that was included with
your instrument, check our website. You can download the
latest version at no charge when it becomes available.
The 2700 is also compatible with 3rd party capture software
such as the popular Microsoft Windows® HyperTerminal®
which is pre-installed with many PC’s in use today.
37
18. Calibration Report
Calibration report provides detailed information on the most recent calibration. It
includes date & time, calibration standard(s), offset, temperature, number of days
calibration is over due slope information, etc.
The calibration report is automatically displayed after a complete calibration.
To view the calibration report at any other time, press from the
desired measurement screen.
38
19. Dissolved Oxygen Theory
Dissolved Oxygen (DO) refers to the volume of oxygen that is contained in water.
There are two main sources of DO in water; atmosphere and photosynthesis.
Waves and tumbling water mix air into the water where oxygen readily dissolves
until saturation occurs. Oxygen is also produced by aquatic plants and algae
during photosynthesis.
The amount of DO that can be held by water depends on 3 factors:
1. TEMPERATURE:
DO increases with decreasing temperature
(colder water holds more oxygen)
2. SALINITY:
DO increases with decreasing salinity
(freshwater holds more oxygen than saltwater does)
3. ATMOSPHERIC PRESSURE:
DO decreases with decreasing atmospheric pressure
(amount of DO absorbed in water decreases as altitude increases)
14
12
Solubility mg/L
10
0
0 5 10 15 20 25 30 35 40 45 50
Temperature °C
40
Applications
Oxygen is essential for fish, invertebrate, plant, and aerobic bacteria respiration.
DO levels below 3 ppm are stressful to most aquatic organisms. Levels below 2
or 1 ppm will not support fish. Fish growth and activity usually require 5 to 6 ppm
of DO, an important consideration for Aqua-culture industry.
Low DO indicates a demand on the oxygen of the system. Natural organic
material such as leaves accumulate in the stream and create an oxygen demand
as it is decomposed. Organic materials from human activities also create an
oxygen demand in the system. Micro-organisms consume oxygen as they
decompose sewage, urban and agricultural run-off, and discharge from food-
processing plants, meat-packing plants and diaries. There is an optimum DO
level for this process and if the DO level falls too low, the micro-organisms die
and the decomposition ceases. When DO level is too high, this indicates too
much power is used than necessary for aeration resulting in an inefficient
process.
In boiler water application, presence of oxygen in the water will increase corrosion
and helps build up boiler scale that inhibits heat transfer. In such instance it is
critical to keep DO concentration to a minimum.
Some pollutants such as acid mine drainage produce direct chemical demands
on oxygen in the water. DO is consumed in the oxidation-reduction reactions of
introduced chemical compounds such as nitrate (NO31-) and ammonia (NH41+),
sulfate (SO42-), and sulfite (SO32-) and ferrous (Fe2+) and ferric (Fe3+) ions.
These are important consideration for water and wastewater treatment industry.
41
20. Calculating Temperature Coefficients
To determine the temperature coefficient of your sample solution use this formula:
Where:
tc = Temperature coefficient 25 = 25 ºC
CT1 = Conductivity at Temp 1 CT2 = Conductivity at Temp 2
T1 = Temp 1 T2 = Temp 2
1. Immerse the probe into a sample of your solution and adjust the temperature
coefficient to 0% (that is, no compensation) by following instructions as
described in Section 10—Temperature Coefficient.
2. Wait for 5 minutes. Note T1 and CT1 (conductivity at T1).
3. Condition the sample solution and probe to a temperature (T2) that is about 5
ºC to 10 ºC different from T1, and note the conductivity reading CT2.
Record your results for future reference. Ideally T1 and T2 should bracket your
measurement temperature, and should not differ by more than 5 ºC.
4. Calculate the temperature coefficient of your solution according to the formula
shown above.
5. Enter the calculated temperature coefficient into the meter.
The calculated temperature coefficient will now be applied to all the meter
readings.
42
21. Calculating TDS Conversion Factor
You can calibrate TDS using the value of the calibration standard solution at a
standard temperature such as 25 ºC. To determine the conductivity-to-TDS
conversion factor use the following formula:
Factor = Actual TDS ÷ Actual Conductivity @ 25 ºC
• Actual TDS: Value from the solution bottle label or as a standard
made using high purity water and precisely weighed salts.
ppm = milligram of salt(s) per liter of water = mg/L
ppt = gram of salt(s) per liter of water = g/L
Both the Actual TDS and the Actual Conductivity values must be in the same
magnitude of units. For example, if the TDS value is in ppm the conductivity value
must be in µS; if the TDS value is in ppt the conductivity value must be in mS.
Check your factor by multiplying the conductivity reading by the factor in the
above formula. The result should be in TDS value.
When the TDS factor is set to 1.0, Conductivity = TDS.
43
22. DO Electrode Maintenance
Changing the Membrane Cap
The DO 2700 electrode utilizes a cap attached to a pre-installed membrane that
is easily replaced. Replace the membrane cap if it becomes damaged or worn
over time as needed. When calibration can not be completed, look to replace the
electrolyte first as this may help. Replace the membrane cap and electrolyte
together using the following procedure.
1) Unscrew the old membrane cap from the probe. Before installing a new
membrane, clean the probe tip with deionized water to remove any
contaminants.
2) With the membrane cap removed fill it approximately 50% full using the
electrolyte solution. Screw the membrane cap on moderately tight (do not
over-tighten). A small amount of electrolyte should overflow—indicating that
all of the air has been removed.
Electrode Cleaning
Cleaning the BOD probe is only necessary after erratic
readings result; usually after many months of use.
Avoid excessive and un-necessary cleaning, as it can
reduce the life of the probe.
Gold Cathode
The gold cathode can become tarnished or plate with
silver after extended use. It is cleaned with the
adhesive backed sanding disc in the Membrane Kit
provided with the probe. Stick the disc to a flat object,
like a bottle cap, and gently sand the gold cathode with a twisting motion about 3
times or until all silver deposits are removed. The gold will have a matte finish.
Silver Anode
It is normal for a dark layer of silver chloride to cover the silver anode. After
prolonged use it may become necessary to clean the anode. Soak the probe in
14% ammonium hydroxide for 2 to 3 minutes or overnight in a 3% ammonium
hydroxide solution. Rinse with deionized water, recharge the electrode, and install
a new membrane.
44
23. Troubleshooting
Parameter Cause Solution
pH Won’t accept 1st New calibrations must begin with
calibration point neutral standard (7.00 or 6.86)
Conductivity Can’t adjust / can’t AUTO calibration values (84 µS, 1413
calibrate µS, 12.88 mS, or 111.8 mS) are not
used. Change Cal Method to MANUAL.
Conductivity 2nd Calibration point Only one point per range can be
replaces the 1st calibrated. Re-consider your SINGLE or
MULTI Cal Points setting in Section 10.
46
24. Specifications
We reserve the right to make changes, improvements and modifications to the
specifications listed here.
47
Conductivity CON 2700, PC 2700 only
Range 0.050 µS to 500.0 mS
Resolution 0.01 / 0.1 µS; 0.001 / 0.01 / 0.1mS
Accuracy ±1% full scale
Automatic (4 points); Maximum 1 per range
Cal. Points
Manual (5 points); Maximum 1 per range
Cell Constant 0.010 to 10.000
Coefficient (Per ºC) Linear & Pure; 0.000 to 10.000%
Normalization 15.0 to 30.0 ºC / 59.0 to 86.0 ºF
Compensation Automatic with supplied cell or Manual
0.0 to 100 ºC / 32.0 to 212.0 ºF (0.0 to 80 ºC / 32.0
Temp Compensation
to 176.0 ºF with supplied cell)
48
Dissolved Oxygen DO2700
Range 0 to 50.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.5% full scale
Offset Adjustment ±10.0%
Cal. Points (%) 2 points; 0%, 100%
1 point; ± 50% from the factory default value.
Cal. Points (conc)
Minimum reading allowed is 2.00.
Salinity Correction 0 to 50.0 ppt (manual input)
Barometric Pressure 450 to 825 mmHg (automatic with built in sensor);
Correction adjustable ±150 mmHg
Temperature 0.0 to 50.0 ºC / 32.0 to 122.0 ºF
49
25. Replacements and Accessories
Part number Ordering Code
Item Description Eutech Oakton
Instruments Instruments
pH 2700 meter with pH electrode (ECFG7370101B), ATC probe
ECPH270042GS
(PH5TEMB01P), integral electrode holder, 100/240 VAC adapter & 35420-20
01X543902
pH electrode refill solution
pH 2700 meter with integral electrode holder & 100/240 VAC ECPH270040S
35420-22
adapter 01X543901
ION 2700 meter with integral electrode holder & 100/240 VAC ECION270040S
35421-02
adapter 01X543903
CONSEN9501D
Conductivity/ATC electrode, 2-cell Ultem/stainless steel, k = 1.0 35608-74
01X466601
PH5TEMB01P
ATC Probe, Stainless Steel, 117 x 3 mm 35613-13
01X210303
EC620SSP
Dissolved Oxygen/BOD/ATC self-stirring electrode —
01X295704
50
pH 6.86 buffer solution, 480 mL bottle ECBU686BT 00654-03
pH 4.01, 7.00, & 10.01 buffer solution pack, 480 mL bottles — 05942-10
01X211297 35803-74
Double-junction pH electrode refilling solution, saturated KCl
60 mL 125 mL
100/240 VAC, SMPS, 9V, 6W power adapter, USA/UK/EUR Plug 60X426401 35615-07
Serial impact micro printer: dot-matrix, portable with 25-pin, female ECMICROPRNTR01
01X230001 —
connector. Includes roll of paper and 110/120 VAC power adapter
Serial impact micro printer: dot-matrix, portable with 25-pin, female ECMICROPRNTR02
01X230002 —
connector. Includes roll of paper and 220/230 VAC power adapter
ECADPTRM9M25
RS232C cable for micro printer, 9-pin (M) To 25-pin (M) —
09X305002
52
27. Return of Items
Authorization must be obtained from our Customer Service
Department or authorized distributor before returning items for any
reason. A “Return Material Authorization” (RMA) form is available
through our authorized distributor. Please include data regarding
the reason the items are to be returned. For your protection, items
must be carefully packed to prevent damage in shipment and
insured against possible damage or loss. Eutech Instruments will
not be responsible for damage resulting from careless or insufficient
packing. A restocking charge will be made on all unauthorized
returns.
NOTE: Eutech Instruments Pte Ltd /Oakton Instruments reserve
the right to make improvements in design, construction, and
appearance of products without notice.
53
54
55
For more information on our products, please contact our channel partner or visit
our websites listed below: