b40 - b20v1 Technical Reference Manual
b40 - b20v1 Technical Reference Manual
b40 - b20v1 Technical Reference Manual
0459
Revision G
26 Jun, 2014
Trademarks
Dash, Datex, Ohmeda, DINAMAP, EK-Pro, Trim Knob, Unity Network, CARESCAPE and TruSignal,
GE Healthcare, GE Medical system, General Electric Company are trademarks of GE Healthcare. All
other product and company names are property of their respective owners.
1 Introduction
About this manual 1
1 Overview 3
1.1 Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.2 Safety message signal words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.3 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.4 ESD precautionary procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.2.5 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3 Service information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3.1 Service requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3.2 Equipment identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2 System description 11
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2 Bus structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3 Distributed processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4 Module communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.5 Software loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.6 Parameter modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3 Frame functional description 14
3.1 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1.1 Keyboards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1.3 CPU board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.1.4 Power board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1.5 AC/DC unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.1.6 Batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2 Interfacing computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3 Connectors and signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.3.1 External connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4 Hemo-dynamic module introduction 23
4.1 Monitor software compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.1 Hemo-dynamic module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.2 NIBP board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.2.3 ECG board in 5-lead measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.2.4 ECG filtering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.2.5 STP board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.2.6 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.3 Connectors and signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.3.1 Front panel connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.4 Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.4.1 NIBP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.4.2 ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.4.3 Pulse oximetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.4.4 Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.4.5 Invasive blood pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.4.6 Respiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
i
Document no. 2050802-001
B40/B20 Patient Monitor
2 Hardware installation
1 Hardware installation 1
1.1 Unpacking instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Choosing location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 Mounting the monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.4 Connection to mains. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.4.1 Install the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.5 Connection to Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.5.1 Pre-installation requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.5.2 Network configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.6 Inserting and removing the E-miniC module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.7 Monitor connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.8 Visual indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.9 Installation checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3 Maintenance
1 Instructions 1
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Recommended tools. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 Electrical Safety Tests 3
2.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.3 Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.4 Power Outlet Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.5 Power cord and plug. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.6 Ground (Earth) Integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.6.1 Ground Continuity Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.6.2 Impedance of Protective Earth Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.7 Ground (earth) wire leakage current tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.8 Enclosure (Touch) leakage current test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.9 Patient (source) leakage current test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.10 Patient (sink) leakage current test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.11 Test completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 Installation checkout 13
3.1 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2.1 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.3 Frame unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.4 Parameters measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.5 Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.6 Network connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.7 Conclusion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4 Maintenance and checkout 16
4.1 Visual inspection/preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1.1 Before beginning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1.2 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
ii
Document no. 2050802-001
4.2 Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.3 Keyboard(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.4 Time and date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.5 Hemo Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.6 Loudspeaker. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2.7 Monitor software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2.8 Watchdog circuitry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2.9 Batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2.10 Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.2.11 Final cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5 Adjustments and calibrations 23
5.1 NIBP calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2 Temperature calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.3 Invasive pressure calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4 Troubleshooting
1 Introduction 1
1.1 General troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2 Software troubleshooting chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Frame troubleshooting 4
2.1 NET section troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3 Hemo Troubleshooting 7
3.1 NIBP troubleshooting flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 ECG troubleshooting flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.3 STP troubleshooting flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.4 NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.4.1 NIBP toubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.4.2 NIBP error code explanation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5 ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.6 Impedance respiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.7 Pulse oximetry (SpO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.8 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.9 Invasive blood pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5 Service Menu
1 Introduction 1
1.1 Service Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Service Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 SW Management 3
2.1 SW Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Active Inactive SW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3 Country Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
iii
Document no. 2050802-001
B40/B20 Patient Monitor
3 Frame 5
3.1 Country Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.2 Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2.1 Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2.2 Dri Config (in S/5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2.3 Dri Comm (in S/5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2.4 Unity Config (in Unity) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2.5 TCP/IP Config . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.3.1 WPM Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4 Keyboard 17
4.1 Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1.1 Keyboard Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5 Parameters 18
5.1 Gas Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.1.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.1.2 Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.2 ECG Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.2.1 ECG Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.3 STP Module (for GE SpO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.3.1 Calibrations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.4 NIBP Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.4.1 NIBP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.4.2 NIBP Safety Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.4.3 NIBP Pneumatics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.5 SpO2 (for Masimo/Nellcor SpO2). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6 Set/Test 32
7 Service Log 33
iv
Document no. 2050802-001
2.7.2 To check the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.7.3 Conditioning the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.8 To replace the fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.9 To download the software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7 Technical specification
1 General Specifications 1
1.1 Genenral specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Defibrillator synchronization connector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 Parameters specifications 3
2.1 ECG specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Impedance respiration specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 GE SpO2 specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.4 Nellcor SpO2 specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.5 Masimo SpO2 specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.6 NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.7 Invasive blood pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.8 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.9 Airway gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
8 E-MiniC Module
1 Product overview 1
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2.1 CO2 measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 Main components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3.1 Gas sampling system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.2 MiniC sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3.3 CPU board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4 Connectors and signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2 Maintenance and checkout 7
2.1 Replacement of planned maintenance parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.1 Required parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.2 Replacement procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2 Visual inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3 Functional checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.1 Test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.3 Test completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 Calibration and adjustments 13
3.1 Sample flow rate adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1.1 Calibration setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1.2 Sample flow rate adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Gas calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.1 Calibration setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
v
Document no. 2050802-001
B40/B20 Patient Monitor
4 Troubleshooting 16
4.1 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2 Troubleshooting checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.1 Gas sampling system troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.2 MiniC sensor troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.3 Troubleshooting charts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5 Disassembly and reassembly 19
5.1 Disassembly guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.1.1 Serviceable parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.1.2 Service limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.1.3 ESD precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.1.4 Before disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.1.5 Tools needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.2 Disassembly and reassembly procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.2.1 Replacing the pump unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.2.2 Replacing the miniCO2 assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
6 Service parts 22
6.1 Ordering parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6.2 Spare parts for E-miniC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
vi
Document no. 2050802-001
1 Introduction
About this manual
Intended audience
This Technical reference manual is meant for service representatives and technical personnel
who install, configure, maintain, administer, troubleshoot or repair B40/B20 monitor running
the software license VSP-A.
1-1
Product availability
Some of the product parts and accessories mentioned in this manual may not be available in
all countries.
Please, consult your local representative for the availability.
Related documentation
− Clinical aspects, basic methods of measurement and technical background: B40/B20
Parient Monitor User’s Reference Manual
− Options and selections of the software: B40/B20 Parient Monitor Default Configuration
Worksheet
− Compatible supplies and accessories: B40/B20 Parient Monitor Supplies and Accessories
− Other devices closely related to the monitor:
• iCentral and iCentral Client User's Reference Manual
• CIC Pro Clinical Information Center Operator's Manual
Conventions used
To help you find and interpret information easily, the manual uses consistent text formats:
Within this manual, special styles and formats are used to distinguish between terms viewed
on screen, a button you must press, or a list of menu commands you must select:
− Names of hardware keys on the keypad are written in bold typeface: NIBP
Start/Cancel.
− Menu items are written in bold italic typeface: Monitor Setup.
− Emphasized text is in italic typeface.
− When referring to different sections in this manual, section names are enclosed in double
quotes: “Cleaning and care”.
− The word “select” means choosing and confirming.
− Messages (alarm messages, informative messages) displayed on the screen are written
inside single quotes: 'Learning.'
− Note statements provide application tips or other useful information.
1-2
1-3
1.1 Symbols
Alternating current
1-4
Power ON/OFF.
Fuse. Replace the fuse only with one of the same type and rating
Gas inlet.
Gas outlet.
IP21
Degree of ingress protection.
SN,S/N Serial number
Date of manufacture. This symbol indicates the date of
manufacture of this device. The four digits identify the year.
Storage temperature
1-5
A blinking heart next to the heart rate or pulse rate value indicates
the beats detected.
1-6
Warnings
• Use only GE recommended power cords
• When disconnecting the system from the power line, remove the plug from the wall outlet
first
• Due to high voltage, use insulated screw driver
• High voltage on test body; do not touch it during the test
1-7
• To avoid the risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
• Always check that power cord and plug are intact and undamaged
• All system devices must be connected to the same power supply circuit
• Only interconnect devices when determined safe by qualified biomedical personnel
• Only devices that are specified compliant with IEC 60950-1 or IEC 60601-1 may be
connected to the Ethernet MC or IX ports
• Biomed must determine interconnected parts are safe.
• Do not use without manufacturer approved mounting
• If the software package is changed, all clinical settings will reset to factory defaults.
• Do not use with iCentral software V5.0.2 and earlier or Mobile Care Server software earlier
of V5.2.
• Verify compatibility of all system components prior to installation
• Use only approved accessories, including mounts, and defibrillator-proof cables and
invasive pressure transducers. For a list of approved accessories, see the supplies and
accessories list delivered with the monitor. Other cables, transducers and accessories
may cause a safety hazard, damage the equipment or system, result in increased
emissions or decreased immunity of the equipment or system or interfere with the
measurement.
• Regular preventive maintenance should be carried out annually.
• Do not use multiple modules with identical measurements in the same monitor.
• The user may only perform maintenance procedures specifically described in this
manual.
• Incorrect power line frequency setting could adversely affect ECG processing.
• Make sure patient is not being monitored while servicing the equipment.
• Don’t press power key when changing language or doing factory reset.
Cautions
• Set the time of a newly added network device as close as possible to the time of devices
already on the network
1-8
1.2.5 Disposal
Dispose of the whole device, parts of it and its packing material and manuals in accordance
with local environmental and waste disposal regulations.
1-9
1-10
2 System description
2.1 Introduction
The B40/B20 monitor build up a freely configurable modular system. The architecture is
designed to enable different module combinations so that the user is able to get the desirable
parameter and feature set. This modular approach makes it possible to add new features
when they are needed.
NAND
Ethernet
Flash
DATA BUS
AT91SA
USB HOSE RS485
M7s256
SDRAM
AT91 ARM IIC BUS Sound
Address bus
1-11
Marker Out
1-12
+5V +13...16V
Isolation VMOD
Analog transformer
electronics
MODULE BUS
Patient isolation
A/D
PATIENT
CPU
convert
Data
RAM
EEPROM
Peripheral RS485
drivers Opto isolation
drivers
1-13
3.1.2 Display
The B40 use 12.1” LCD display with SVGA 800 x 600 resolution has bright long life lamps and a
wide viewing angle.
The B20 use 10.4” LCD display with SVGA 800 x 600 resolution has bright long live lamps and a
wide viewing angle.
NOTE: The LCD display backlight circuit runs on a high voltage. Do not touch the inverter board
or the backlight tube leads when powered.
Backlights
The backlight lamp unit consists of two integrated cold cathode fluorescent lamps. The
backlight lamp unit is driven by a separate inverter board.
1-14
I/O connector
AC INLET Cable or wires
100-240 Vac
50/60 Hz Pin-to-pin connection
AC/DC
Unit
LCD
display
SVGA
Nurse call
Serial data CPU Backlight CCFL
inverter LAMPS
Power Board XY/CW Board
Module bus
CW
Trim Knob
Battery board User interface
board XY Keyboard/Memb Power
External interface rane switch indicator
board
SMBUS
SMBUS
BAT1
BAT2
1-15
The board is based on AT91 ARM microprocessor. Other functions include LVDS display driver,
10/100Mbps on board Ethernet, Alarm Light function, KEY board and rotor encoder control,
audio driver function, nurse call function, defibrillation function, module bus function.
The CPU section takes care of the central processing.
The main features are:
• AT91 ARM
• 266 MHz Main CPU clock
• 64MBytes SDRAM
• 4 MBytes minimum NAND flash memory
• 8 MBytes Data flash memory
• Main CPU Provides one standard UART communication
Connectors
Ethernet communication connector
Color LCD operation connector
Audio operation connector
Alarm Light indicator operation connector
Power board connection connector
Voltage supervision
There are two voltage supervision chips that control the system reset signals.
The +3.3V supervision chip outputs reset signals for +3.3V devices. 3.3V Reset Threshold will be
Falling: min 3.00 V; max 3.15 V.
The +1.2V supervision chip outputs reset signals for +1.2V devices. 1.2V Reset Threshold will be
Falling: min 1.08V; max 1.14 V.
1-16
1-17
MUX
+5V_PMC
Battery 2
Over Voltage
VSYS Enable
Protection Power Fail Alarm
Switch (Crowbar)
Over Current
15V Boost
Protection & Current VMOD
Regulator
Sense
LCD Backlight
Backlight_VCC
Enable Switch
1-18
AC/DC unit
The AC/DC unit is a compact medical power supply based on high-efficiency technology. It is
designed for 65 watt continuous output power, universal AC input and 15V output voltage.
3.1.6 Batteries
The B40/B20 has two lithium-ion batteries, located in the battery compartment. The power
board connects one of the batteries to be the power source, if no power is received from the
AC/DC unit.The battery charging is controlled by the power board.
The batteries can be charged separately, and screen symbols and monitor frame LED
indicators indicate their charging level and possible failure.
NOTE: When the monitor is battery powered, the green battery LED is on. When the monitor is
mains powered, the green mains LED is on.
1-19
Network connector
RJ45 connector Pin Signal
1 Tx +
2 Tx -
1 2 3 4 5 6 7 8
3 Rx +
4 N/C
5 N/C
6 Rx -
7 N/C
8 N/C
1-20
Pin 1
Pin 2,3
1-21
Serial port
9 pin female connector Pin Signal
1 GND
2 SERIAL_TXD
5 1 3 SERIAL_RXD
4 SERIAL+3.3V
6
9
5 GND
6 N/C
7 SERIAL_CTS
8 SERIAL_RTS
9 N/C
Main power
Mains connector Pin Signal
L Live
PE Protected earth
N Neutral
1-22
1. InvBP connector
1 2. Temperature connector
3. SpO2 connector
2
4. ECG connector
5. NIBP connector
3
1-23
The monitor displays waveforms and measurement readings, and handles the trending and
alarm management. The ECG (e.g. heart beat and arrhythmia detection) and the Impedance
Respiration algorithms are in the monitor software. The modules measure signals and send
them to the monitor. The NIBP, SpO2, Temperature and Invasive Pressure algorithms are in the
module.
There are four parameter circuit boards inside the hemo-dynamic module for processing the
measurement signals. Each processing board has a microcontroller with software.
The NIBP parameter measurement requires one signal processing board, pneumatic system,
valve and pump unit connected to NIBP parameter board.
The second parameter board is the optional board, for Nellcor or Masimo SpO2 measurement,
it’s Masimo MS-2011 board or Covidien NELL1GE-S board at different configuration.
The third parameter board is for GE SpO2, IBP and Temperature measurement including input
board. All these three parameter is optional, according to different configuration, using
different board: it’s STP board, TP board for Nellcor, TP board for Masimo, GE SpO2 board.
The fourth parameter board is for 3/5-lead ECG with the Impedance Respiration measurement
including ECG input unit connected to the ECG parameter board.
All parameter boards are connected together via module bus flex board connecting voltage
and module communication, the module communicates with frame through RS-485 bus.
Valves
connector
EEPROM
1024Bytes
connector
Pump
Main CPU Pneumatic
AT91SAM7S256 control
RS485
Module bus connector
interface
256KBytes Flash
64KBytes SRAM
10bits ADC
Safety CPU
MSP430F2013
2KB+256B Flash
128B RAM
16bits Sigma-
Delta ADC
1-24
Signal processing
Two signals from the pressure transducers are amplified and sent to the A/D converter. After
the converter, digitized signals are sent to the microprocessor for data processing.
The NIBP board is controlled with an ARM7 microprocessor at 16 MHz oscillator frequency.
Memory
The NIBP program memory (processor flash memory) size is 256k x 8. The processor has 64
kBytes RAM. The EEPROM size is 8K x 8 and it is used to store the calibration values for the
pressure transducers, the pulse valve constants gained during measurements, the PC board
identification, and the module serial number.
Software control
The software controls valves and a pump. In addition to the individual on/off signals for each
component there is a common power switch for the valves and the pump that can be used at
pump/valve failures.
Safety circuit
The NIBP board is equipped with an independent safety circuit to disconnect supply voltages
from the pump and the valves if the cuff has been pressurized longer than the preset
maximum measurement time, or if the pressure of the cuff is inflated over the specified
pressure limit. The maximum measurement time values and pressure limits for different
measurement modes have been specified in the technical specification section of this manual.
Pneumatics
1
5, 6
3
4 1
1-25
4. Safety valve/Dump valve; The Safety valve/Dump valve is intended to be used for
deflating the cuff in single fault case, i.e. to prevent too long a measurement time or too
high an inflation pressure of the cuff.
5. Main pressure sensor; for measuring the pressure of the blood pressure cuff and the
pressure fluctuations caused by arterial wall movement.
6. Second pressure sensor; for detecting the, cuff loose, cuff occlusion situations, etc. and
for recognizing the pressure sensor fault.
7. Cuff connector; for connection and hose identification.
1-26
ECG CABLE
- ECG LEAD SET
- ECG TRUNK CABLE
BASELINE
RESTORATION
ECG CPU
12_lead_ECG_meas_blck_dgrm.vsd
RS 485 POWER SUPPLY
COMMUNICATION NV
MEMORY
ISOLATION ISOLATION
ECG preamplifiers
The buffer amplifiers are used for each lead. The “Leads off” detection is implemented by
measuring the output level of the input buffer amplifiers with the A/D converter of the CPU. The
ECG signals are measured using differential amplifiers.
1-27
Pacemaker detection
Pacer detection has been made by using four slew rate detector circuits. The pacer detection
amplifiers have been realized at the front of the slew rate detectors independently of the ECG
measuring channels.
ECG CPU
The CPU is a 16 bit H8/3052 single-chip microcomputer. It contains 128 kbytes of flash memory
and 4 kbytes of RAM. The clock frequency is 16 MHz.
RS485 communication
The communication to the CPU board of the monitor uses RS485 protocol. The RS485 driver
circuits are optically isolated from the processor of the module.
Power supply
The ECG board has a driver-controlled half-bridge switching power supply with 5 kV isolation.
The supply voltages have been regulated with linear regulators.
1-28
1-29
A/D CONVERSION
STP CPU
STP_brd_blck_dgrm.vsd
COMMUNICATION NV
MEMORY
ISOLATION ISOLATION
1-30
NV MEMORY
RS 485
POWER SUPPLY COMMUNICATION
ISOLATION ISOLATION
Microprocessor unit
The CPU is a 16 bit H8/3052 single-chip microcomputer. It contains 128 kbytes of flash memory
and 4 kbytes of RAM. The clock frequency is 16 MHz.
High speed I/O is used to obtain a pulse control sequence necessary for pulse oximetry
measurement. Timing for the clock is from the oscillator.
1-31
constant
Defibrillation/ESD protection current
resistors and diodes source
Temperature
sensors T1, T2
R Ref1 Ref2
T1 T2
Differential
R amplifier
0C:7k36 d/dt 0
0 To A/D
15C:3k54
converter
R
25C:2k53 0
38C:1k30
45C: 984 0
PSM_temp_meas_principle.vsd
R
1-32
Vin
Current
measurement G
to AD converter
Pressure
transducer
Instrum entation
amplifier
PSM_pressure_meas_principle.vsd
Vout
Input filter G
to A D converter
1-33
LP Oximeter channel 1
Gain=7.5
DC-
suppression
LP Oximeter channel 2
Gain=7.5
LP Oximeter channel 3
Amplifier: Gain=7.5
Preamplifier: Gain = 2
Current-to-voltage type
Spo2_measurement_blck_diagr.vsd
Bipolar/single-ended modes DC-
Adjustable gain suppression
Oximeter channel 4
LP
DE-MUX Amplifiers
Analog Digital
1-34
Module BUS(RS-485)
Frame
1-35
1-36
1-37
4.4.2 ECG
Electrocardiography analyzes the electrical activity of the heart by measuring the electrical
potential produced with electrodes placed on the surface of the body.
ECG reflects:
• electrical activity of the heart
• normal/abnormal function of the heart
• effects of anesthesia on heart function
• effects of surgery on heart function
See the “User's Guide” or the “User’s Reference Manual” for electrodes’ positions and other
information.
1-38
HbO2
SaO2 frac =
HbO2 + Hb + Dyshemoglo bin Formula 1
or against functional saturation SaO2func;
HbO 2
SaO 2 func =
HbO 2 + Hb Formula 2
Functional saturation is more insensitive to changes of carboxyhemoglobin and
methemoglobin concentrations in blood.
The oxygen saturation percentage SpO2 measured by the Datex-Ohmeda module is calibrated
against functional saturation SaO2func. The advantage of this method is that the accuracy of
SpO2 measurement relative to SaO2func can be maintained even at rather high concentrations
of carboxyhemoglobin in blood. Independent of the calibration method, pulse oximeters are
not able to correctly measure oxygen content of the arterial blood at elevated
carboxyhemoglobin or methemoglobin levels.
Pulse rate
The pulse rate calculation is done by peak detection of the plethysmographic pulse wave. The
signals are filtered to reduce noise and checked to separate artifacts.
Intensity of
Imax (DC-component) Imax
transmitted
light AC-component
Imin
Arterial blood
Venous blood
absorption_of_light.vsd
Tissue
Time
No pulsation Pulsatile blood
Incident light
1-39
IRED
Emitter
RED
PSM_absorption_of_infrared.vsd
Detector
1-40
Test methods used to establish the accuracy claims during low perfusion
Nellcor Oximax Technology with Oximax sensors
Nellcor Oximax Technology with Oximax sensors have been validated for SpO2 low perfusion
accuracy in bench top testing using Nellcor's PS II simulator with signal strength setting of
0.03% modulation and oxygen saturation levels of 70 to 100% at a pulse rate of 90 BPM.
Nellcor Oximax Technology with Oximax sensors have been validated for low perfusion pulse
rate accuracy in bench top testing using Nellcor's PS II simulator at a simulated low perfusion
level of 0.10% modulation in the pulse rate range of 40 to 250 beats per minutes (BPM) at a
SpO2 of 95%.
Masimo SET Technology with LNOP and LNCS sensors
Masimo SET Technology with LNOP and LNCS sensors have been validated for low perfusion
accuracy in bench top testing against Biotek Index 2 Simulator and Masimo's simulator with
signal strength setting of greater than 0.02% and a % transmission of greater than 5% for
saturation ranging from 70%-100%. One percent was added to accuracies for neonatal/infant
sensors to account for accuracy variation due to properties of fetal hemoglobin.
1-41
4.4.4 Temperature
The temperature is measured by a probe whose resistance varies when the temperature
changes, called NTC (Negative Temperature Coefficient) resistor.
The resistance can be measured by two complementary methods:
• Applying a constant voltage across the resistor and measuring the current that flows
through it.
• Applying a constant current through the resistor and measuring the voltage that is
generated across it.
Hemo module uses the constant current method. The NTC-resistor is connected in series with a
normal resistor and a constant current is applied through them. The temperature dependent
voltage can be detected at the junction of the resistors, thus producing the temperature signal
from the patient. The signal is amplified by analog amplifiers and further processed by digital
electronics.
4.4.6 Respiration
Impedance respiration is measured across the thorax between ECG electrodes. The respiration
signal is made by supplying current between the electrodes and by measuring the differential
current from the electrodes. The signal measured is the impedance change caused by
breathing. The respiration rate is calculated from these impedance changes, and the
respiration waveform is displayed on the screen.
1-42
1 Hardware installation
Safety precautions
Warnings
• The monitor or its components should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the monitor and its components
should be observed to verify normal operation in the configuration in which it will be used.
• After transferring or reinstalling the monitor, always check that it is properly connected
and all parts are securely attached. Pay special attention to this in case of stacked
mounting.
• The monitor must not be used without a manufacturer approved mounting attached.
• Never install the monitor so that it is above the patient.
Cautions
• The monitor display is fragile. Ensure that it is not placed near a heat source or exposed to
mechanical shocks, pressure, moisture or direct sunlight.
2-1
WARNING The power cord may only be connected to a three-wire, grounded, hospital
grade receptacle
2-2
MC Network
• The MC Network infrastructure shall be installed according to the "CARESCAPE Network
Configuration Guide".
• The installation site of the patient monitor shall have a wall jack and a network patch
cable for the MC Network.
• If the monitor with main software version earilier than 1.01 want to connect to the MC
network, please contact authorized service personnel to upgrade the software to the
latest version
S/5 Network
• The S/5 Network shall be installed according to the "S/5 Network Installation Guide". Refer
to the "iCentral and iCentral Client Service Manual" for iCentral installation instructions.
• The installation site of the patient monitor shall have a wall jack and a network patch
cable for the S/5 Network.
2-3
• Bed Name:
It is used for setting the bed name in the monitor. The default bed name is the last
five characters of the MAC address, excluding the colon delimiter characters.
• Save Changes:
It is gray if the changes have not been done to the menu information.
NOTE: When save changes the monitor will restart automatically.
5. Set up the TCP/IP configuration
TCP/IP Config - TCP/IP Config
• Configure the monitor’s IP address, subnet mask, default gateway.
NOTE: If the IP address or both of unit name and bed name duplicate, the alarm will
be given to the new added monitors.
• Save the configuration after configuring the TCP/IP. When save changes the
monitor will restart automatically.
WARNING If only B40/B20 monitors and Dash monitors in the Unity network, do not set
up B40/B20 monitors with the highest IP address. For this may cause the
Unity time sync function can’t work with Dash monitors.
6. If the monitor does not connect to network correctly, you may also need to set up the
speed and duplex configuration.
TCP/IP Config - TCP/IP Config - Speed and Duplex
Choose the suitable settings according to your network environment. The default setting
is AUTO.
2-4
To use the E-miniC module, your monitor need pre-configure the extension rack from
manufacture.
To insert module:
1. Align the module with the insertion guides
2. Push the module into the monitor frame until it clicks and stops.
3. Pull the module outwards to insure the module is firmly seated.
To remove module:
1. Pressing the release latch, on the bottom of the module.
2. Grasp the module firmly and pull out of the Frame. Make sure not to drop it when it comes
out.
WARNING When detaching modules, be careful not to drop them. Always support with
one hand while pulling out with the other.
2-5
2-6
2-7
2-8
1 Instructions
1.1 Introduction
These instructions include procedures of system maintenance for the B40/B20. It’s include four
sections:
• Electrical safety tests.
• Installation checkout, which should be performed after installation and service
configuration.
• Maintenance and checkout, which should be performed every 12 months.
• Adjustments and calibrations
NOTE: Please complete the check form when performing the corresponding procedures.
NOTE: For the E-MiniC maintenance and calibration, please refer to “8. E-MiniC Module“ or “S/5
Single-width Airway Module, E-miniC Technical Reference Manual Slot” M1027829-1.
The symbol in the instructions means that the procedure performed should be signed in
the check form.
The procedures should be performed in ascending order, bypassing those that are not
applicable for a particular monitor.
To enter the service menus, you need following passwords:
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8)
In case you evaluate the measurement accuracy with a patient simulator, add the simulator’s
accuracy specification to the one for the monitor.
CAUTION Failure on the part of all responsible individuals, hospitals or institutions,
employing the use of this device, to implement the recommended
maintenance schedule may cause equipment failure. The manufacturer does
not, in any manner, assume the responsibility for performing the
recommended maintenance schedule, unless an equipment maintenance
agreement exists. The sole responsibility rests with the individuals, hospitals, or
institutions utilizing the device.
CAUTION Only trained personnel with appropriate equipment should perform the tests
and repairs outlined in this section. Unauthorized service may void warranty of
the unit.
CAUTION Wear a static control wrist strap when handling PCB boards. Electrostatic
discharge may damage components on the board.
3-1
Accessories
A rigid cylinder or pipe
NIBP cuff
Adult NIBP cuff hose with cuff ID
Infant NIBP cuff hose with cuff ID
Tubing parts to connect a manometer and a pump to
the NIPB cuff and hose.
Dual invasive pressure adapter cable
ECG accessories, IEC or AHA
- Multi-link 3-lead integrated cable and leadwire
- Multi-link 5-leadwire set
- Multi-link 3/5-lead ECG trunk cable
SpO2 finger probe
SpO2 Interconnect Cable
Temperature dual cable
CO2 Sampling line 3m/10 ft
Tool
A multiparameter patient simulator with IBP, Temp
adpter cables
Screwdrivers PH1, PH2
NOTE: For details on recommended accessories see “Supplies and Accessories“ catalog.
3-2
2.2 Recommendations
GE recommends that you perform all safety tests presented in this chapter.
• Upon receipt of the device (monitor and its associated equipment).
• Every twelve months thereafter planned maintenance
• Each time the main enclosure is disassembled or a circuit board is removed, tested,
repaired, or replaced
These instructions are intended for every component in the system. GE recommends that the
qualified personnel performing the tests.
Perform electrical safety tests using an electrical safety analyzer per IEC 60601-1, UL 60601-1,
EN 60601-1 or CSA C22.2 No. 601.1. The schematics in the section provide a general
understanding of the test equipment. Actual configuration of test equipment may vary.
The patient monitor being tested should be placed on an insulating surface.
3-3
WARNING Use only AC power cords recommended or manufactured by GE.
NOTE: The measuring device (MD) represents the network and voltage measuring instrument
and its frequency characteristics per IEC 60601-1.
3-4
Acceptance criteria:
• For equipment without a power supply cord, the impedance between the protective earth
terminal and any accessible metal part which is protectively earthed shall not exceed 0.1
ohms.
• For equipment with a power supply cord, the impedance between the protective earth
pin in the mains plug and any accessible metal part which is protectively earthed shall
not exceed 0.2 ohms.
Acceptance criteria
• For equipment without a power supply cord, the impedance between the protective earth
terminal any accessible metal part which is protectively earthed shall not exceed 0.1
Ohms.
• For equipment with a power supply cord, the impedance between the protective earth
pin in the mains plug and any accessible metal part which is protectively earthed shall
not exceed 0.2 ohms.
3-5
Leakage Tester
Power Cord
HIGH NORM Power Cord
LOW
GND
GND
RVS Device
Under
Test
0.15µF
DMM 1K
10
NOTE: The measuring device (MD) represents the network and voltage measuring instrument
and its frequency characteristics per IEC 60601-1.
(1) Configure the safety analyzer as follows (NC):
• Polarity - NORMAL
• Neutral - CLOSED
(2) Power on the device under test.
(3) Read and record the current leakage indicated on the tester.
(4) Configure the safety analyzer as follows (SFC):
• Polarity - NORMAL
• Neutral - OPEN
(5) Read and record the current leakage indicated on the tester.
(6) Configure the safety analyzer as follows (SFC):
• Polarity - REVERSED
• Neutral - OPEN
(7) Read and record the current leakage indicated on the tester.
(8) Configure the safety analyzer as follows (NC):
• Polarity - REVERSED
• Neutral - CLOSED
(9) Read and record the current leakage indicated on the tester.
(10) Power off the device under test.
If measured reading is greater than the appropriate specification below, the device under test
fails. Contact GE Technical Support.
3-6
• (USA only) 300 µA, and the device under test is powered from a center-tapped 200-240
V/50-60 Hz, single phase circuit (UL Split Phase Exemption)
• 500 µA, and the device under test is powered from a non-centertapped, 100-240 V/50-60
Hz, single-phase circuit
Perform the test in Normal Condition (NC) and in two different Single Fault Conditions (SFC): 1)
earth open and 2) one of the supply conductors open at a time. Perform the test with normal
and reverse polarity.
NOTE: Refer to the instructions delivered with the safety analyzer to perform this test.
Leakage Tester
Power Cord
RVS
Device
Open
Under
Closed GND Test
0.15µF
DMM 1K
10
Probe to exposed conductive chassis
NOTE: The MD represents the network and voltage measuring instrument and its frequency
characteristics per IEC 60601-1.
(1) Configure the safety analyzer as follows (NC):
• Polarity - NORMAL
• Neutral - CLOSED
• GND (Earth) - CLOSED
(2) Power on device under test.
(3) Read and record the current leakage indicated on tester.
(4) Configure the safety analyzer as follows (SFC):
• Polarity - NORMAL
3-7
• Neutral - OPEN
• GND (Earth) - CLOSED
(5) Read and record the current leakage indicated on the tester.
(6) Configure the safety analyzer as follows (SFC):
• Polarity - NORMAL
• Neutral - CLOSED
• GND (Earth) - OPEN
(7) Read and record the current leakage indicated on the tester.
(8) Configure the safety analyzer as follows (SFC):
• Polarity - REVERSED
• Neutral - CLOSED
• GND (Earth) - OPEN
(9) Read and record the current leakage indicated on the tester.
(10) Configure the safety analyzer as follows (SFC):
• Polarity - REVERSED
• Neutral - OPEN
• GND (Earth) - CLOSED
(11) Read and record the current leakage indicated on the tester.
(12) Configure the safety analyzer as follows (NC):
• Polarity - REVERSED
• Neutral - CLOSED
• GND (Earth) - CLOSED
(13) Read and record the current leakage indicated on the tester.
If measured reading is greater than the appropriate specification below, the device under test
fails. Contact GE Technical Support.
Acceptance criteria NC
• 100 µA, and the device under test is powered from 100-240 V/50-60 Hz
Acceptance criteria SFC - earth open or one of the supply conductors open
at a time
• (USA only) 300 µA, and the device under test is powered from 100-120 V/50-60 Hz
• (USA only) 300 µA, and the device under test is powered from a center-tapped 200-240
V/50-60 Hz, single phase circuit (UL Split Phase Exemption)
• 500 µA, and the device under test is powered from a non-centertapped, 110-240 V/50-60
Hz, single-phase circuit
3-8
Center-tapped and non-center-tapped supply circuits produce different leakage currents and
the UL and IEC limits are different.
Leakage Tester
Power Cord
RVS
Device
Under
Closed GND
ECG Test Body Test
or SPO2 Test Body
0.15µF
DMM 1K
10
NOTE: The MD represents the network and voltage measuring instrument and its frequency
characteristics per IEC 60601-1.
(1) Connect the ECG/RESP Test Body to the green connector of the device under test.
(2) Configure the safety analyzer as follows (NC):
• Polarity - NORMAL
• Neutral - CLOSED
• GND (Earth) - CLOSED
(3) Power on the device under test.
(4) Read and record the current leakage indicated on the tester.
(5) Configure the safety analyzer as follows (SFC):
• Polarity - NORMAL
• Neutral - OPEN
• GND (Earth) - CLOSED
(6) Read and record the current leakage indicated on the tester.
(7) Configure the safety analyzer as follows (SFC):
• Polarity - NORMAL
• Neutral - CLOSED
• GND (Earth) - OPEN
3-9
(8) Read and record the current leakage indicated on the tester.
(9) Configure the safety analyzer as follows (SFC):
• Polarity - REVERSED
• Neutral - CLOSED
• GND (Earth) - OPEN
(10) Read and record the current leakage indicated on the tester.
(11) Configure the safety analyzer as follows (SFC):
• Polarity - REVERSED
• Neutral - OPEN
• GND (Earth) - CLOSED
(12) Read and record the current leakage indicated on the tester.
(13) Configure the safety analyzer as follows (NC):
• Polarity - REVERSED
• Neutral - CLOSED
• GND (Earth) - CLOSED
(14) Read and record the current leakage indicated on the tester.
(15) Power off the device under test.
(16) Repeat the steps in this procedure using the appropriate SPO2 Test Body. Connect the
SPO2 Test Body to the blue SPO2 connector of the device under test.
If measured reading is greater than the appropriate specification below, the device under test
fails. Contact GE Technical Support.
Acceptance criteria NC
• 10 µA
Acceptance criteria SFC - earth open or one of the supply conductors open
at a time
• 50 µA
3-10
Leakage Tester
Power Cord
HIGH NORM Power Cord
LOW
GND
RVS
Device
120K
ECG Test Body Under
Closed GND
or ECG Cable Test
or SPO2 Test Body
0.15µF
DMM 1K
10 (Keep cable length as
short as possible.)
NOTE: The MD represents the network and voltage measuring instrument and its frequency
characteristics per IEC 60601-1.
NOTE: Per IEC 60601-1, the impedance to protect the circuitry and the person performing the
test, but low enough to accept currents higher than the allowable values of the LEAKAGE
CURRENT to be measured.
WARNING Shock hazard. The following step causes high voltage at the test body. Do
not touch the test body.
(1) Connect the ECG/RESP Test Body to the green connector of the device under test.
(2) Configure the safety analyzer as follows:
• Polarity - NORMAL
• Neutral - CLOSED
• GND (Earth) - CLOSED
(3) Power on the device under test.
(4) Read and record the current leakage indicated on the tester.
(5) Configure the safety analyzer as follows:
• Polarity - REVERSED
• Neutral - CLOSED
• GND (Earth) - CLOSED
(6) Read and record the current leakage indicated on the tester.
(7) Power off the device under test.
(8) Repeat the steps in this procedure using the appropriate SPO2 Test Body. Connect the
SPO2 Test Body to the blue SPO2 connector of the device under test.
If measured reading is greater than the appropriate specification below, the device under test
fails. Contact GE Technical Support.
Acceptance criteria
• 50 µA
3-11
3-12
3 Installation checkout
These instructions include procedures for a installation checkout for B40/B20 monitor. It is
recommended to be performed after monitor installation.
Skip the tests that are not applicable for the installed monitor.
These instructions include a “Installation and checkout form, B40/B20” to be filled in when
performing the procedures.
An electrical safety check and a leakage current test should be performed prior to the monitor
installation.
3-13
3.2.2 Display
1. Verify that all text is readable and all images are clear.
2. Verify that the brightness is good. Adjust if necessary.
3.2.3 Frame unit
1. Check that the clock on the screen shows correct time. Adjust the time and date, if
necessary.
NOTE: The B40/B20 can’t be set as the TIME MASTER in network. You should adjust the
time and date from the central station.
3.2.4 Parameters measurements
Connect the accessories (no need connect simulator/patient), check the following
phenomenon will appeared.
• ECG: After connecting ECG cable, ‘leads off’ will display in the Waveform Field
• NIBP: After connecting NIBP hose to module, ‘Adult/Pediatric’ or ‘Neonatal’ will display in
NIBP Digital Field for several seconds
• SpO2: After connecting SpO2 cable and sensor, SpO2 sensor will be lit.
• Temperature: After connecting Temp cable and sensor, ‘Performing temp test:’ will display
in Temp Digital Field for several seconds.
• IBP: After connecting IBP cable and transducer, ‘InvBP’s not Zeroed’ will display in
Message Field.
• CO2: After installing the E-miniC module, ‘Calibrating gas sensor’ will display in CO2
waveform field for about 1 minutes
3.2.5 Recorder
1. Press the Recorder Start/Stop key and check that the module starts recording the
selected waveforms. Press the Recorder Start/Stop key again to stop recording.
2. Check that the quality of the recordings is acceptable.
3.2.6 Network connection
1. Check that the CAT-5 cable connector is clean and intact, then connect it to the Network
connector on the backside of the monitor.
3-14
Check that the monitor connects to the network, i.e. the network symbol appears on the
upper right-hand corner of the screen.
3.2.7 Conclusion
• Power off the monitor
• Perform final cleaning
• Fill in all necessary documents
3-15
4.1.2 General
1. Check that all parts are intact and that the cables and screws are connected and
tightened properly. Check the connector pins are intact. Check the front cover and the
front panel sticker are intact.
2. Check that the modules go in smoothly and lock up properly.
3. Check the ventilation holes of the monitor and clean if necessary.
4. Check the module and applied parts are clean.
NOTE: Batteries are recommended to be conditioned every six months.
3-16
4.2.2 Display
1. Check that the image quality on the screen is correct
2. Check the display brightness, select
Monitor Setup - Display brightness
Push the Trim Knob and check that the display brightness follows the selected brightness.
4.2.3 Keyboard(s)
1. Tests with the keyboard/membrane switch:
− Enter the service menu:
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) -
Keyboard
− Check the keys one by one, turn trim knob one full turn clockwise and one full turn
counter clockwise.
4.2.4 Time and date
− Check that the clock on the screen shows correct time.
Readjust the time and date, if necessary.
Monitor Setup - Time and Date
NOTE: The B40/B20 can’t be set as the TIME MASTER in network. You should adjust the
time and date from the central station.
4.2.5 Hemo Module
ECG measurement
• Configure ECG settings in monitor:
Connect 5 lead ECG cable to monitor and simulator.
Monitor Setup -> Screen Setup -> Waveform Fields ->
Field 1: ECG1; Field 2: ECG2; Field 3: ECG3
ECG -> ECG1 Lead: II; ECG2 Lead: V1; ECG3 Lead: aVL; ECG Size: 1.0
-> ECG Setup -> Beat Sound Volume: 1 or greater; Pacemaker: Show;
HR Source: AUTO
• Configure ECG settings in simulator:
ECG rhythm: a normal sinus rhythm
heart rate: 80 bpm
Amplitude: 1 mV
1. Normal Sinus Rhythm
− Check that the monitor displays the ECG leads II, V1 & aVL and the waveforms are
noise-free. The monitor shall display a 80 ± 5 bpm heart rate and an audible QRS
tone sounds with each QRS complex.
3-17
2. Pacemaker Detection
− Configure the simulator to output "Asynchronous Pacemaker Pulse"
− Check that pacemaker spikes are shown on the ECG waveform.
3. Asystole Detection
− Configure the simulator to output "Asystole".
− Check that the 'Asystole' alarm appears to the monitor screen.
− Configure the simulator to show "80 beats per minute, Normal Sinus Rhythm".
4. Leads Off Detection
− Detach the RA/R leadwire from the simulator.
− Check that the Lead II waveform disappears from the ECG1 waveform field and a
message 'RA/R lead off'' is shown momentarily.
− Check that Lead II is replaced by Lead III in the ECG1 waveform field after a while
and a message 'Lead changed' is followed by a message 'Learning'.
− Reconnect the RA/R leadwire to the simulator.
− Check that Lead III is replaced back to Lead II in the ECG waveform field.
Respiration measurement
• Configure RESP settings in monitor:
Set up the Resp waveform field to the monitor screen. And
Others -> Resp Setup -> Resp Rate Source: Imped.
-> Measurement: ON
• Configure RESP settings in simulator:
Baseline impedance: 1000
Amplitude: 1
Respiration rate: 20 breaths per minute
Lead selection: II (or LL)
5. Respiration Rate
− Check that the RESP waveform is shown and the RR value is 20 (±5).
− Configure the simulator's Apnea Simulation to "32 sec".
6. Apnea Detection
− Check that the monitor activates the Apnea alarm.
− Configure the simulator's Apnea Simulation to "OFF"
3-18
Temperature measurement
• Configure the “T1+T2” digit field to the monitor screen.
• Configure the simulator’s temperature channels as follows: Temperature : 37 °C/98.6 °F
7. Temperature detection
− Check the corresponding temperature value appears and that no error messages
are shown on the monitor screen.
NOTE: If the deviation on a temperature reading on the screen is more than 0.1°C,
calibrate the temperature channels according to the instructions in chapter
"Temperature calibration".
Invasive blood pressure measurement
• Configure the simulator’s IBP channels as follows:
Sensitivity: 5 µV/V/mmHg
InvBP outputs: "0 mmHg static pressure" or "atmosphere"
8. Zeroing
− Press IBP Zero All key.
− Check that a message "Zeroing" followed by a message "Zeroed" is shown in the IBP
parameter window.
9. Static Pressure
− Configure the simulator's InvBP output to "200 mmHg static pressure".
− Check that the flat pressure line appears on the related waveform field. The reading
in the parameter window shall be 200 ±10 mmHg.
NOTE: Recalibration is required, if the measured value is not within the specification.
Calibrate the invasive pressure channels according to the instructions in "Invasive
pressure calibration".
10. Pressure Waveforms
− Configure the simulator's InvBP output to "Arterial 120/80".
− Check that the pressure waveform for tested invasive pressure channel appears in
the IBP waveform field and the Sys/Dia (Mean) pressure values are shown in the
related parameter window.
SpO2 measurement
11. Test measurement
Connect the SpO2 probe onto your finger. Check that the reading of 90-100 and SpO2
waveform appears. Check that the HR value is calculated from SpO2 when ECG and InvBP
(ABP or Art) are not measured.
NOTE: a functional tester cannot be used to assess the accuracy of a pulse oximeter probe or a
pulse oximeter monitor.
3-19
Non Invasive Blood Pressure measurement
• Connect and set up the NIBP integrated tool as following picture, ensure all the
connections made are leak-proof.
13. NIBP calibration
− Calibrate the Non-invasive blood pressure (NIBP) channel according to the
instructions in "NIBP calibrations" on page 3-23.
14. NIBP hose detection
− Disconnect the calibration test equipment.
− Make sure NIBP Setup - Inflation Limits is set to Auto.
− Attach an A/P NIBP cuff hose without cuff identification.
− Press the NIBP Start/Cancel key. After a few moments the monitor automatically
opens the selections NIBP Setup - Inflation Limits for user to manually select the
inflation limits.
3-20
4.2.6 Loudspeaker
Check the loudspeaker by setting the alarm volume:
Alarms Setup - Alarm Volume
Test the whole volume scale from 1 to 10 by turning the Trim Knob and check that the
alarm volume changes correspondingly. The alarm sound should be clear and audible
with all the settings.
4.2.7 Monitor software
Enter the service menu:
Monitor Setup - Install Service (Password 16-4-34) - Service (Password 26-23-8)
Take down the information regarding Monitor software.
4.2.8 Watchdog circuitry
Select:
Monitor Setup - Install Service (Password 16-4-34) - Service (Password 26-23-8) -
Set/Test
Perform the tests Watchdog and WD by Overload. Check that the monitor restarts in
each case.
NOTE: When selecting Watchdog and WD by Overload, auto restarting should take place
approximately after 10 seconds.
4.2.9 Batteries
1. Battery operation
Disconnect the power cord (without switching the monitor to standby).
Check that the monitoring continues normally. Check also the keyboard/membrane
switch LEDs:
- the green mains power LED is off
- the green battery LED is on
- the orange battery LED is off
2. Battery charging
Reconnect the power cord.
Check that the monitoring continues without problems. Check also the
keyboard/membrane switch LEDs:
- the green mains power LED is on
- the green battery LED is off
- the orange battery LED is on or off continuously (not blinking)
3-21
4.2.10 Network
− Connection to network
Check that the CAT-5 cable connector is clean and intact, then connect it to the monitor.
Check that the monitor connects to the network, i.e. the network symbol appears in the
screen.
4.2.11 Final cleaning
Switch off the monitor and perform final cleaning.
Fill in all necessary documents.
3-22
Calibration check
• Refer to the "Non Invasive Blood Pressure measurement" on page 3-20 in maintenance
checkout section for NIBP integrated tool set up.
1. Enter Calibration menu:
Monitor Setup - Install/Service (password 16-4-34) -Service (26-23-8) - Parameters -
NIBP - Calibrations
Calibration
Active Leak Test OFF
Calibration Check OFF
Protection OFF
Calibrate
Previous Menu
3-23
There is a reading of B1 and B2 in help filed under NIBP calibration menu. If the error of
pressure channel B1 and B2 is larger than specified above, the module should be recalibrated.
The error of B2 has no effect on blood pressure measurement accuracy.
NOTE: If the monitor pass the calibration check, no need to do NIBP calibration.
Calibration
1. Enter Calibration menu.
2. Disconnect the hoses from the NIBP connector to enable proper zeroing.
3. Select Calibrate. If it is not available, perform the steps a and b.
a. Select Protection OFF in the Calibration menu and push the Trim Knob.
b. Menu selection Calibrate is now enabled.
• Start calibration by pushing the Trim Knob. Messages ‘Zeroing’ and ‘Zeroed’ will be
displayed in the NIBP message field.
• Connect an external mercury manometer with a pump to the module through the both
tubes of the hose - both transducers B1 and B2 will be calibrated simultaneously. Pump
up to a pressure of about 200 mmHg according to the manometer. Calibration is possible
in the range of 150 to 250 mmHg.
• Verify that both pressure values in the prompt field match the manometer reading. If not,
adjust by turning the Trim Knob. When the values of the pressure bar and the manometer
are equal, push the Trim Knob to confirm the calibration. The message ‘Calibrating’ and
‘Calibrated’ will be displayed on the NIBP digit field after a few seconds.
• To set the protection on:
Select Protection ON and push the Trim Knob.
NOTE: If a long time not choose ON manually, this menu will automatically return to
Protection ON.
3-24
Using transducer
1. Connect the IBP accessories to the monitor, use a pressure manometer with a pressure
pump
2. Enter STP service menu.
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) -
Parameters - STP.
3. Enter Calibrations menu.
4. Choose Protection OFF in protect mode.
5. Connect a pressure transducer with a pressure manometer to the P1/P2 connector.
Choose Calibrate P1 or Calibrate P2. Leave the transducer to room air pressure.
6. Push the Trim Knob to start zeroing.
7. Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended
pressure is 200 mmHg.
8. Set the pressure on the display to match the pressure reading on the manometer and
push the Trim Knob. A tolerance of ±1 mmHg is allowed.
9. The message ‘Calibrated’ will be displayed on the display.
• Check Protection ON in protect mode.
NOTE: If a long time not choose ON manually, this menu will automatically return to
Protection ON.
3-25
Using simulator
1. Connect the IBP accessories to the monitor and simulator.
2. Enter STP service menu.
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) -
Parameters - STP.
3. Enter Calibrations menu.
4. Choose Protection OFF in protect mode.
5. Choose Calibrate P1 or Calibrate P2. Set the P1 or P2 channel to 0 mmHg on the
simulator.
6. Push the Trim Knob to start zeroing.
7. Set a pressure of 100 mmHg to 300 mmHg on the simulator. The recommended pressure
is 200 mmHg.
8. Set the pressure on the display to match the pressure reading on the simulator and push
the Trim Knob. A tolerance of ±1 mmHg is allowed.
9. The message ‘Calibrated’ will be displayed on the display.
10. Check Protection ON in protect mode.
3-26
1 Introduction
If a problem occurs during the functional examination, check the components of the monitor
according to the below troubleshooting table. If the problem persists, please refer to the
following detail troubleshootings.
Problem What to do
Nothing functions Unplug and re-plug the Power Cord. Also confirm that the cable is intact.
Confirm that the fuses are intact.
Hemo module does not Remove and replace the Hemo module.
function Confirm that the desired parameters are configured to be displayed.
E-miniC module does Confirm that ‘Calibrating Gas Sensor’ messages are not displayed.
not function Confirm that a D-fend water trap and a sample tube are attached.
Confirm that the desired parameters are configured to be displayed.
Remove and replace the module.
Network does not Refer to 1.5. Connection to Network section in Installation chapter about
function network’s configuration.
4-1
Mains power
LED lit?
No
No
Yes
Disconnect the mains
Possible AC/DC unit
power cord and batteries.
failure. Replace the Power board failure
Reconnect and press the
AC/DC unit.
ON/OFF key.
Yes
Mains power
No
LED lit?
4-2
Start-up with GE
logo image?
Yes
OK?
Turn the power on.
No
Try with another
Wait for 120
seconds. software CD to
No
download again. Please
read the note below.
Start-up display
No
appears?
OK? Yes
Yes
No
NOTE:
The software CD may be
defective.
4-3
2 Frame troubleshooting
Problem Cause What to do
Monitor is not starting. 1. The batteries are empty. 1. Connect the power cord.
2. Fuses may be blown. 2. Replace fuses.
3. If power cord connected, AC/DC 3. Replace the AC/DC power unit.
unit may be faulty.
4. If AC/DC unit is working, the 4. Replace the power board unit.
power board may be faulty.
5. On/Stby key may be faulty. 5. Replace the keyboard/membrane
switch panel or the
interconnection cable.
Monitor is not starting. 1. The connection between power 1. Check connection between power
board and CPU board may be board and CPU board.
faulty.
2. Faulty CPU board. 2. Replace the CPU board.
The monitor starts (alarm beep 1. The LCD display cables are loose. 1. Check the LCD display connection
is heard), but the display board connectors.
remains black. 2. The backlights are not lit. 2. Check inverter cable.
Backlight Inverter may be faulty.
Replace the Inverter board.
Display and monitor operating 1. Loudspeaker connector or wires 1. Check loudspeaker connector and
but no audible beep in start-up. loose or faulty. wires.
Display is too dim. 1. Incorrect brightness adjustment. 1. Adjust display brightness higher.
2. Backlight faulty. 2. Check inverter cable.
3. Backlight inverter faulty. 3. Replace backlight.
Stripes or white areas on screen. 1. Loose faulty display connection 1. Check display connection cable in
cable in CPU and display. CPU and display.
Module data disappears from Parameter module current (in Detach and change parameter
the screen. ‘Module power module bus) too high. module.
supply overload’ message.
Module data disappears. 1. Module bus voltage or signals 1. Replace module interface board.
path broken. 2. Replace the cable between
module interface board and
Power board.
3. Change the parameter module.
‘Replace Battery’ message on 1. Problem in communication 1. Replace battery.
the screen. between battery and power
board.
2. Battery too old or defected. 2. Replace battery.
3. power board may be faulty. 3. Replace the power board.
4. Problem in communication
between power board and CPU
'Frame temperature high' The temperature inside the frame is Check monitor ventilation holes.
message. too high.
‘Battery temperature high’ Battery SMBus temperature is too Check monitor ventilation holes.
message on the screen. high. Replace battery.
4-4
4-5
4-6
3 Hemo Troubleshooting
3.1 NIBP troubleshooting flowchart
NIBP module not
working
Insert Hemo
module to a good
monitor
NIBP
Replace NIBP
No parameter display
board
on the screen?
Yes
Check if NIBP
NIBP Cuff loose connector spring
ENd Yes Start NIBP
function well? message appears No contact well with
without hose
on screen? the button of cuff
ID board
Yes
Pump check
Check pump
in Service Menu No
connector.
OK?
NOTE: Please refer to 5.4. NIBP Module in Chapter 5 for Service Menu instruction.
4-7
Does NIBP
Refer to NIBP
module work No
flowcharts
well?
Yes
Replace ECG
board
ECG function
No End
well?
No
4-8
Does NIBP
Refer to NIBP
module work No
flowcharts
well?
Yes
Go to service
menu
Is STP parameter
Select the right
configuration No
configuration
correct?
Yes
Replace STP/TP
board
End
4-9
3.4 NIBP
3.4.1 NIBP toubleshooting
4-10
4-11
4-12
3.5 ECG
Problem Cause What to do
HR numerical display No heart rate available. If no ECG waveform, check LEADS OFF
shows ‘---’ message and connect the leads.
If ECG waveform exists, check heart rate
source e.g. in the ECG Setup menu behind
ECG key.
Unacceptable ECG Poor electrode or poor electrode skin Electrodes from different manufacturers
waveform contact. are used. /Too much/little gel is used.
Poor electrode condition. Electrodes are dried out.
Improper site of electrodes. Check that electrodes are not placed over
bones, active muscles, or layers of fat.
Improper skin preparation. Remove body hair. Clean attachment site
carefully with alcohol.
Improper bandwidth filter. Check filter.
Dirty ECG cable. Clean the cable.
Faulty ECG cable Change the cable.
No ECG trace Waveform not selected on screen. Press the Monitor Setup key and make
adjustments.
Module not plugged in correctly. Check the hemo module’s installation.
Noise-message High frequency or 50/60 Hz noise. Isolate noise source.
4-13
4-14
4-15
3.8 Temperature
Problem Cause What to do
No temperature displayed Wrong type of probe. Use correct probe.
Temperature out of measurable range. The range is between 10 and 45 °C.
Temperature calibration not protected. Set the protection ON in the Service
Menu.
The monitor not configure TEMP. Check the monitor’s configuration.
4-16
1 Introduction
The monitor has a Service Menu, which is a useful tool to examine monitor functions and
troubleshoot in case a fault occurs.
Service Menu
SW Download
Dri Config
Unity Config
Frame Network Dri Comm
TCP/IP Config
TCP/IP Config
STP Calibrations
NIBP Calibrations
Safety Valve
Service Log
5-1
5-2
2 SW Management
The SW Management menu includes
Software-specific service menus. Service Menu Sw version / Unit id
SW Management
----------------------------
SW Download
----------------------------
Active Inactive SW
Country Settings ----------------------------
----------------------------
----------------------------
----------------------------
number:-----------------
200
----------------------------
5-3
2.1 SW Download
The SW Download submenu show IP Address and
enable software download. SW Download
IP Address: Show the IP Address of the monitor.
IP Address 172.16.1.23
Enable SW Download: Select YES can enable to
download main software. Enable SW Download
Previous Menu
5-4
3 Frame
The Frame menu includes frame-specific service
menus. Service Menu Sw version / Unit id
Frame
----------------------------
Country Settings
----------------------------
Network
----------------------------
----------------------------
----------------------------
number:-----------------
200
----------------------------
5-5
Languages
Language
Country Settings
Select a language to be used during monitoring.
NOTE: Service pages will always appear in English despite Languages
of this selection.
Language ENG.LNG
NOTE: For language codes, see the table below.
Delete Language Delete Language ENG.LNG
Delete a language file from the permanent memory of the Previous Menu POR.LNG
monitor. Cancel
5-6
Abbreviation Language
CHI Chinese
CZE Czech
DAN Danish
ENG English
FRE French
GER German
HUN Hungarian
ITA Italian
NLB Dutch
NOR Norwegian
POL Polish
POR Portuguese
RUS Russian
SPA Spanish
SWE Swedish
TUR Turkish
5-7
3.2 Network
3.2.1 Network
For setting the monitor’s network mode: S5/Unity
Frame Frame
Previous Menu
5-8
Ethernet
The Ethernet Status view shows the
Ethernet Ethernet Status
general status of the ethernet network Previous Menu Driver DP83907
communication. Cable Connected
Driver: Ethernet chip name. Ethernet Addt 00:40:97:0b:01:fb
Speed (bits/s) 0
Cable: Indicates if the ethernet cable is
connected. Statistics In Out
EthernetAddr: Monitor’s ethernet address. Packets 2527 11327
Bytes 297776 9837268
Speed: Indicates the current ethernet
communication speed. Data errors
The service data related to the ethernet CRC Frame Transm. BER
status view is described in the following 0 0 0 0
table. Hardware errors
Intern. Missed FIFO Overrun
0 0 0 0
5-9
5-10
Logins 0
5-11
TCP/IP config
IP address: For setting the static IP address of monitor.
Subnet Mask: For setting the static subnet mask of
monitor.
Default Gateway: For setting the static default gateway
of monitor.
Speed and Duplex: For setting speed and dupliex of
monitor.
Save Changes: Close menu and save the current menu
information to permanent memory.
Cancel Changes: Close menu without saving changes
Ping
IP Address: For configuring the destination IP address for the ping command.
Ping: Enable the ping command by sending Internet Control Message Protocol (ICMP) echo request packets to
the target host and wait for an ICMP response. The response should be show in the Ping Status view.
5-12
5-13
5-14
Smart Batt1
Battx information from SMBus (this information Batt1 information from SMBus
is received from the battery via SMBus) WPM Battery Temperature 30.75 C
Temperature: Battery temperature Batts Voltage 11903 mV
Current 0 mA
Voltage: Battery voltage Smart Batt1 Avg. Current 0 mA
Rel. State of Charge 97 %
Current: Battery current (discharge or charge) Smart Batt2
Abs. State of Charge 72 %
Clear Temp Maxs Remaining Capacity 2532 mAh
Avg. Current: Rolling average of the battery Full Charge Capacity 2604 mAh
current Previous Menu Cvcle Count 33
Design Capacity 3520 mAh
Rel. State of Charge: Relative state of charge of Design Voltage 11100 mV
the battery. Expressed as a percentage of Rem.
Manufacture Date(DD:MM:YY) 20/12/24
Cap. (mAh) / Full Cap.(mAh).
Abs. State of Charge: Absolute state of charge. Manufacture Name NPC A07A90
Device Name SM201-6
Expressed as a percentage of Rem. Cap. (mAh)/ Device Chemistry LION
Design Capacity (mAh). Therm. Status from charger
5-15
Smart Batt2
If Smart Battx is selected for a battery NOT
Batt2 information from Memory and PMC
connected to the SMBus, the menu contains the
WPM Battery Rel. State of Charge 0%
following:
Batts Remaining Capacity 0 mAh
Battx information from Memory and PMC (This Full Charge Capacity 0 mAh
Smart Batt1 Design Capacity 0 mAh
information comes from the power Board memory Design Voltage 0 mV
or is measured by the power Board. SMBus data in Smart Batt2
Manufacture Date (DD:MM:YY) 00/00/00
this menu is not real time, because this battery is Clear Temp Maxs
Manufacturer Name
not connected to the SMBus). Previous Menu Device Name
NOTE: This page may not contain information if Device Chemistry LION
SMBus has been connected only to the other DC/DC Board A/D
battery. Page can be updated by battery Batt1 Voltage 12.00 V
Batt2 Voltage 12.34 V
disconnection and reconnection, if desired. Batt Current -0.01 A
Rel. State Charge: Relative state of charge of the Batt Not CHG Temperature 32.24 C
battery. Expressed as a percentage of Rem. Cap.
(mAh) / Full Cap.(mAh).
Remaining Capacity: Remaining battery capacity
(mAh).
Full Charge Capacity: Capacity of the battery
when it is fully charged.
Design Capacity: Theoretical capacity of a new
battery.
Design Voltage: Theoretical value for nominal voltage of a new battery.
Manufacturer Date (DD:MM:YY): The date the battery pack was manufactured
Manufacturer Name: Acronym of the battery pack manufacturer name.
Device Name: Battery pack model name.
Device Chemistry: Battery chemistry of the cells used.
Batt1 Voltage: Battery A voltage measured at the power Board.
Batt2 Voltage: Battery B voltage measured at the power Board.
Batt Current: Battery current (discharge or charge) for the battery connected to SMBus.
Measured at the power Board.
Batt Not CHG Temperature: This is subject to change.
5-16
4 Keyboard
4.1 Keyboard Keyboard Service Data
The service menu for testing the command board Message count 0
Red Led Leds red OFF yellow OFF cyan OFF
functions.
Yellow Led Direct action keys
Red Led is for testing the red alarm light on the Cyan Led Admit/ Pt. Data SpO2 ECG
monitor. When the text is highlighted, the red alarm Dischar & Trends
Dummy Press
light can be turned on and off by pressing the Trim
Keyboard Log NIBP Invasive Normal
Knob. Pressures Screen
Yellow Led is for testing the yellow alarm light on the Previous Menu
Monitor Print/
monitor. When the text is highlighted, the yellow Setup Record
alarm light can be turned on and off by pressing the Airway Others Silence Alarms
Trim Knob. Gas Alarms Setup
Cyan Led is for testing the cyan alarm light on the Zero Auto Start Start
monitor. When the text is highlighted, the cyan alarm All On/Off Cancel Stop
light can be turned on and off by pressing the Trim
Knob. Control wheel
Press 0
Dummy Press is for testing the Trim Knob. When the Clokwise 0 Counterclokwise 0
text is highlighted, pressing the Trim Knob creates a
sound from the loudspeaker and the corresponding
number on the service data field increases.
Service Data
Message Count counts the number of messages that are sent out to the main CPU board.
5-17
5 Parameters
NOTE: Parameter values in Service Data fields are only for reference in this section.
----------------------------
----------------------------
2, 2004-03-15
----------------------------
ode not set, 200
----------------------------
5-18
5-19
5.1.2 Gases
ID Not in use
ID unrel. Not in use
MAC Not in use
Sample Flow is calculated from differential pressure and adjusted by the module. Zero value as
measured during initialization when the pump is off. Gain: sample flow measurement can
be calibrated by adjusting the gain.
Ambient Ambient pressure is measured continuously. Amb-Work: ambient pressure - sampling
system internal pressure.
Fall time Not in use
CO2-O2 Delay Not in use
Pump Can be toggled ON/OFF. PWM output 0-100% is shown. Pump voltage is also shown.
Lamp The state, PWM control, and current of the lamp are shown.
Fan Not in use
Zero Valve Can be toggled between the measurement state (MEAS) and the zeroing state (ZERO).
Occl Valve Not in use
Temp Temperatures measured by the module from TPX, CPU, and OM.
Time after power on In minutes after power on.
5-20
5-21
Bad c-s by mod is a cumulative number that indicates how many communication errors the
module has detected.
The monitor starts counting these items at power up and resets to zero at power off. The values
may also be reset when a module is attached to the monitor frame and be set to 32769 or
continuous counting may be started when the module is removed from the monitor frame.
The nonzero values do not indicate a failure, but the continuous counting (more than 5 per
second) or value 32769 indicates either a serial communication failure or a module not in place.
Also failures in other modules may cause these numbers to rise or be set to 32769.
RAM indicates the state of the RAM memory.
ROM indicates whether the checksum at the EPROM is in accordance with the one the software
has calculated.
EEPROM indicates if the values stored in the permanent memory are valid.
The state is either OK, Fail or ? (module not in place or a communication error).
LeadsOff indicates whether the monitor can measure ECG even if one or more leadwires are off:
1 = measurement is not possible, 0 = measurement can be done.
MissedPkgs indicates the number of packages missed.
5-22
5-23
Temp error shows the status of the temperature test. No errors found show the
status (OFF) and errors found (ON).
Protect key and Protect mode show normally ON but can be turned to OFF for
the temperature calibration in Calibration Menu.
Configuration shows the chosen module configuration: T, P, TP, ST, or STP.
Timeouts is a cumulative number that indicates how many times the module
has not responded to the monitor’s inquiry. Bad checksums is a cumulative
number that indicates how many times communication from the module to the
monitor has broken down.
Bad c-s by mod is a cumulative number that indicates how many
communication errors the module has detected.
The monitor starts counting these items at power up and resets to zero at
power off. The nonzero values do not indicate a failure, but the continuous
counting (more than 5 per second) indicates either a serial communication
failure, or a module not in place. Also other modules can cause communication
errors that cause these numbers rise.
RAM indicates the state of the RAM memory.
ROM indicates whether the checksum at the EPROM is in accordance with the
one the software has calculated.
EEPROM indicates if the values stored in the permanent memory are valid.
The state is either OK, Fail or ? (module not in place or a communication error).
5-24
5.3.1 Calibrations
5-25
5-26
How to calibrate
NOTE: Perform NIBP Calibration Check first to evaluate if calibration is needed or not.
NOTE: Both transducers B1 and B2 will be calibrated simultaneously.
NOTE: The module must be in the frame during the whole procedure.
NOTE: Calibration selection is available only when protection is OFF.
1. Change the protection setting from ON to OFF to enable the Calibrate selection - the
color of the Calibrate selection changes from grey to white.
2. Zeroing:
− Disconnect the NIBP hose from the module connector.
− Select Calibrate and push the Trim Knob.
NOTE: Messages ‘Zeroing’ and ‘Zeroed’ is shown in the NIBP message field and next to the
Calibrate selection momentarily. After this, a pressure bar will appear beside the menu.
3. Calibration:
− Connect the NIBP hose to the module connector.
− Connect an external manometer with a pump to both tubes of the hose.
− Pump about 200 mmHg pressure.
5-27
− Verify that both pressure values, B1 and B2, shown in the prompt field of the
calibration menu match the manometer reading. If not, adjust the by turning the
Trim KnobTrim Knob.
− Press Trim Knob to complete the calibration.
NOTE: Messages ‘Calibrating’ and ‘Calibrated’ are shown in the NIBP message field and next to
the Calibrate selection.
NOTE: When calibrating NIBP, always change the displayed pressure value slightly with the
Trim Knob, even in cases where the value would be correct. For example, change the value one
step higher and then back one step lower. The ‘Calibrated’ text should appear in the display.
This ensures that the calibration procedure is correctly registered and stored by the module.
4. Change the Protection setting from OFF to ON to disable Calibrate selection - the color of
the Calibrate selection changes from white to grey.
NOTE: If a long time not choose ON manually, this menu will automatically return to
Protection ON.
5-28
Start test is for starting and Stop test for stopping the
Safety Valve test. Safety Valve Safety Valve Data
B1 B2
Pressure 000000 000000
Zero 000000 000000
Safety Valve Data: ADULT
Start Test St1 0000 AD0 -16
For information on general items Pressure, Zero, St1 to St2 0000 AD1 6
Previous Menu
St4, AD0 to AD7 as well as Timeouts etc., see service St3 0400 AD2 -1
St4 0000 AD3 1502
data descriptions in section 4.4 NIBP Module. AD4 2
Max. press and 2 s after stop show the measured AD5 -1643
AD6 4
values at Safety Valve test. AD7 -1505
B1 B2
Max press 0 0
2 s after stop 0 0
Timeouts 2 RAM OK
Bad checksums 0 ROM OK
Bad c-s by mod 0 EEPROM OK
5-29
5-30
5-31
6 Set/Test
The system contains a watchdog circuitry, which
needs refreshment at every 10 seconds. If the
Service Menu Sw version / Unit id
refreshment did not occur, the watchdog will reset Set / Test
the main CPU. ----------------------------
The purpose of the watchdog is to restart the Watchdog
monitor, if there was a serious malfunction. This ----------------------------
WD by Over load
feature is useful in two cases: when the main CPU is
Factory Reset ----------------------------
not able to control the monitor, or when the CPU
controls the monitor but detects a serious Previous Menu
----------------------------
malfunction. Watchdog tests check proper
functionality of the watchdog in various conditions. ----------------------------
Watchdog test ensures directly that the watchdog ----------------------------
functions properly. Choosing this test prevents the
watchdog from refreshing and shows running ----------------------------
seconds with an accuracy of 0.1 seconds. 2, 2004-03-15
number:-----------------
The test should have the following result when the
watchdog is working properly: The monitor will ----------------------------
restart after 10 seconds from the start of the test. In
malfunction: ‘>20 s’ is displayed. In this case, the
fault is in the watchdog.
WD by Overload test ensures the functionality of a feature, where the
software controls the monitor, but detects an overload situation in the main
CPU.
The test should have the following result when the feature is working properly:
The monitor will restart after 10 seconds from the start of the test.
Factory Reset restores the factory default settings and clears the data
memories. Factory reset should be run if the monitor software is replaced or if
the Timekeeper battery is replaced.
5-32
7 Service Log
Error, event and alarm data is stored in the Service
Log. Service Log Error History
The service log contains information about the
occurred monitor errors, events and alarms since the Error Log
Last errors:
last factory reset or service log reset. Alarm Log
5-33
5-34
1 Spare part
The following B40/B20 parts will be available as field replaceable spare parts.
6-1
6-2
1.3 Frame
6-3
6-4
NOTE: Refer to Technical Bulletin DOC1303926 for B20 new LED display parts choose and
replace.
NOTE: 2053489-019 FRU B40B20 keyboard/membrane switch includes multi-languages labels.
When service people replace front cover, please order 2053489-023 FRU B20 FRONT COVER or
2053518-008 FRU B20 FRONT COVER, and 2053489-019.
6-5
2 Disassembly
WARNING A short circuit may cause internal damage. Do not touch any exposed
wiring or conductive surface inside, this may cause an electric shock.
CAUTION Perform leakage current measurement whenever service or repair has been
done in the monitor.
NOTE: The backlight circuit runs on high voltage.
Do not touch the inverter board or the backlight tube leads when powered.
NOTE: Field repairs are recommended to the field replaceable unit (FRU) only. Attempting a field
repair on a factory sealed component or assembly could jeopardize the safe and effective
operation of the Monitor.
NOTE: GE recommends that you assemble the monitor using the NEW fasteners (screws,
washers, etc.) provided in the FRU kits. Some fasteners, like the screws with a thread locking
coating, are NOT intended to be re-used more than three times.
6-6
3. From the back side of the monitor, pull the extension rack
out of frame, grasp it firmly.
6-7
7. Use the screwdriver to prize, pulling out the multi I/O at the
same time.
6-8
10. Open the cover in the back side of the frame, then remove 2
screws inside.
6-9
To remove the alarm light board After opening the front cover
To remove the user interface board After opening the front cover
6-10
To separate the LCD display NOTE: Refer to Technical Bulletin DOC1303926 for B20 new LED
display parts choose and replace.
- Remove 4 screws from beside the LCD
6-11
6-12
6-13
6-14
6-15
6-16
6-17
2.7 Batteries
2.7.1 Battery indicators
The B40/B20 messages, screen symbols and the LED indicators tell the user about the status of
the batteries. For screen symbols, see “Part 1 Symbols”. For LED indicators, consult the table
below and for messages, see section “Troubleshooting.” in User’s Guide.
The Green main LED will indicate that monitor is powered by Main. The Green battery LED will
indicate that monitor is powered by battery. The Orange battery LED lit will indicate that the
battery is in charging. The Orange battery LED flashing will indicate that the battery is defect. If
the battery is fully charged and it’s main powered. the orange and Green battery LED will both
dark.
Table 12 Battery indicators
6-18
6-19
6-20
1 General Specifications
1.1 Genenral specifications
Size
Monitor
Without extension 312±5 mm (H) * 312±5 mm (W) * 158±5 mm (D)
modules
With extension modules 312±5 mm (H) * 352±5 mm (W) * 178±5 mm (D)
Weight
With extension module, 6 kg
recorder and CO2
Environment
Operating temperature Normal operation: +5 to +40°C (41 to 104°F)
Charging batteries: +5 to +35°C (41 to 95°F)
Storage and transport -20 to +60°C (-4 to 140°F)
temperature
Operating humidity 20 to 90% noncondensing
Storage and transport 20 to 90% noncondensing
humidity
7-1
7-2
2 Parameters specifications
2.1 ECG specifications
Leads available 3-lead configuration: I, II, III
5-lead configuration: I, II, III, aVR, aVL, aVF and VA
QRS detection range 0.5 to 5mV
QRS detection width (Q to 40 to 120 ms
S)
Defibrillation protection 5000 V, 360 J
Recovery time <5 s
Input impedance Common mode > 10 M Ω@ 50/60 Hz
Differential > 2.5 M Ω from 0.67 to 40 Hz
Common mode rejection 90 dB minimum at 50 Hz
Tall T wave rejection >1.4 mV
ECG leads off detection Active patient electrode: <30 nA
Reference electrode: <120 nA
Filter modes
60 Hz
Monitoring filter 0.5 to 40 Hz
ST filter 0.05 to 40 Hz
Dagnostic filter 0.05 to 150 Hz
50 Hz
Monitoring filter 0.5 to 35 Hz
ST filter 0.05 to 35 Hz
Dagnostic filter 0.05 to 150 Hz
Heart rate
Measurement range 30 to 300 bpm
Measurement accuracy ±5 % or ±5 bpm, whichever is greater
resolution 1 bpm
Heart rate response time (ANSI/AAMI EC13-2002 Section 4.1.2.1f)
Step increase from 80 to average 6.9 s (6.5 to 7.5 s)
120 bpm
Step decrease from 80 to average 8.2 s, (7.6 to 10.0 s)
40 bpm
7-3
NOTE: Pacemaker detector may not operate correctly during the use of high-frequency (HF)
surgical equipment. The disturbances of HF surgical equipment typically cause false positive
pacer detection.
7-4
7-5
* Because SpO2 measurements are statistically distributed, only about 2/3 of the measurements can be expected to fall
within ±1 Arms of the value measured by a CO-oximeter
Test methods used to establish SpO2 accuracy: GE TruSignal SpO2 measurement have been validated for no motion and
motion accuracy in a controlled hypoxia studies with healthy non-smoking adult volunteers over the specified SpO2 rage.
SpO2 readings were compared to SaO2 values of drawn blood samples measured by CO-oximetry. Subjects comprised
both adult men and women and spanned a range of skin pigmentation.
** GE TruSignal technology have been validated for low perfusion SpO2 accuracy over the specified range in a bench top
testing against BioTek Index 2 patient simulator with 0.3% signal amplitude.
Table 2 The table below shows Arms values measured using GE SpO2 sensors
with GE CARESCAPETM V100 in a clinical study.*
* The sensors were clinically tested for accuracy with the following sensors: OXY-E (equivalent to OXY-E-UN, TS-E-D,
TS-E2-GE, TS-E4-GE), OXY-SE (equivalent to OXY-SE-3, TS-SE-3), OXY-F (equivalent to OXY-F-UN, TS-F-D, TS-F2-GE,
TS-F4-GE, TS-SA-D, TS-SA4-GE), OXY-W (equivalent to OXY-W-UN, TS-W-D), OXY-AP (equivalent to OXY-AP-10, OXY-AP-25,
TS-AP-10, TS-AP-25), OXY-AF (equivalent to OXY-AF-10, TS-AF-10, TS-AF-25).
7-6
Measurement accuracy
Without motion Adult/Pediatric 100 to 70% ±2 digits
Neonate 100 to 70% ±3 digits
With motion Adult/Ped/Neo 100 to 70% ±3 digits
Low perfusion 100 to 70% ±2 digits
0~69% unspecified
Display resolution 1% of SpO2
Display averaging 2 to 16 seconds
Pulse rate
Measurement and display 25 to 240 bpm
range
Display resolution 1 bpm
Measurement accuracy Without motion ±3 bpm
With motion ±5 bpm
7-7
2.6 NIBP
Measurement technique Oscillometric with step deflation
Supported modes Manual, automatic and stat
Measurement time Adult/Pediatric inflate duration time less than 120 s
Neonate cycle time less than 85 s
Measurement ranges
Systolic Adult/Pediatric: 30 to 290 mmHg
Neonate: 30 to 140 mmHg
MAP Adult/Pediatric: 20 to 260 mmHg
Neonate: 20 to 125 mmHg
Diastolic Adult/Pediatric: 10 to 220 mmHg
Neonate: 10 to 110 mmHg
Accuracy According to AAMI SP10-2002 4.4.5.2 B, accuracy of
NIBP parameter was validated against the
intra-arterial method 1.
Default initial inflation Adult/Pediatric: 135 ± 15 mmHg
pressure Neonate: 100 ± 15 mmHg
Over pressure allowed by Adult/Pediatric: 300 to 330 mmHg
independent safety Neonate: 150 to 165 mmHg
controller
1. Blood pressure measurements determined with this device are equivalent to those
obtained by an intra-arterial blood pressure measurement device, within the limits
prescribed by the American National Standard, Manual, electronic, or automated
sphygmomanometers
7-8
2.8 Temperature
Measurement units ° Fahrenheit (F)
° Celsius (C)
Measurement range 10 to 45°C (50 to 113°F)
Measurement accuracy ± 0.1°C without temperature sensor
Display resolution ± 0.1°C at 25 to 45 °C with reusable probes
Probe types supported Use only GE Healthcare recommend temperature
YSI probes.
Temperature self-check At start-up and then every 10 minutes
Probe type time response
Reusable skin temperature 3 s
probe:
Reusable adult central 6s
temperature probe:
Reusable pediatric central 4s
temperature probe:
Disposable skin 3 to 6 s
temperature probe:
Disposable central 5 to 8 s
temperature probe, 12F:
Disposable central 5 to 8 s
temperature probe, 9F:
7-9
Resolution 0.1%
Measurement rise time < 300 ms with nominal flow
Valid for respiration rate < 40 breaths/min at I:E ratio of 1:1. (Relative error is
typically 10% for respiration rate 80 breaths/min at I:E ratio of 1:1.) The accuracy
is specified in simulated ventilation. With higher respiration rates and with varying
ventilation methods the specifications may not be met.
Respiration rate
Breath detection 1% change in CO2 level
Measurement range 4 to 80 breaths/min
Accuracy ±1 breaths/min in the range 4 to 20
breaths/min ±5% in the range 20 to 80 breaths/min
7-10
Resolution 1 breaths/min
Non-disturbing gases are Ethanol C2H5OH (<0.3%)
those with a maximum Acetone (<0.1%)
effect on the CO2 reading
Methane CH4(<0.2%)
at 5.0 vol% < 0.2 vol%. The
effect is valid for specific Nitrogen N2 (0 to 100%)
concentrations shown in Water vapor (0 to 100%)
parentheses of the
Dichlorofluoromethane (<1%)
non-disturbing gas:
Tetrafluoroethane (<1%)
Disturbing gases and their Halothane (4%) increases < 0.3 vol%
effect on the CO2 reading Isoflurane(5%) increases < 0.4 vol%
at 5.0 vol% CO2 are shown Enflurane(5%) increases < 0.4 vol%
below. Errors listed reflect
Desflurane(24%) increases < 1.2 vol%
the effect of specific
concentrations (shown in Sevoflurane(6%) increases < 0.4 vol%
parentheses) of an If O2 compensation is not activated:
individual disturbing gas O2 (40 to 95%) decreases < 0.3 vol%
and should be combined
when estimating the effect If O2 compensation is activated:
of gas mixtures: O2 (40 to 95%) error < 0.15 vol%
If N2O compensation is not activated:
N2O (40%) increases < 0.4 vol%
If N2O compensation is activated:
N2O (40 to 80%) error < 0.3 vol%
NOTE: CO2 measurement is intended for patients weighing over 5 kg (11 lb).
7-11
7-12
1 Product overview
1.1 Introduction
This chapter provides information for the maintenance and service of the E-miniC module. The
module is a single-width plug-in module designed for use with the GE modular monitors.
The E-miniC module provides airway measurements.
Letter C in the module name stands for CO2.
NOTE: E-miniC is intended for patients weighing over 5kg (11lb).
8-1
8-2
Document no. 2050802-001
Product overview
Mini D-fendTM
The sample is drawn through the sampling line. The gas then enters the module through the
water trap, where it is divided into two flows, a main flow and a side flow. The main flow goes
into the analyzer. This flow is separated from the patient side by a hydrophobic filter. The side
flow creates a slight subatmospheric pressure within the Mini D-fend water trap which causes
fluid removed by the hydrophobic filter to collect in the bottle.
Zero valve
The main flow passes through a magnetic valve before proceeding to the analyzer. This valve is
activated to establish the zero point for the MiniC measuring unit. When the valve is activated,
room air is drawn through a filter into the internal system and the gas sensor.
8-3
NafionTM tube 1)
A Nafion tube is used between the water trap and the zero valve to balance the sample gas
humidity with that of ambient air. The tube prevents errors caused by the effect of water vapor
on gas partial pressure when humid gases are measured after calibration with dry gases.
Gas analyzers
After the zero valve and Nafion tube, the gas passes through the miniC sensor.
Sampling pump
The gas sampling pump is a membrane pump run by a DC-motor. Sample flow is measured
with a differential pressure transducer across a known restriction. The motor is automatically
controlled to maintain a constant flow even when the D-fend water trap ages and starts to get
occluded. It also enables the use of sample tubes with varying lengths and diameters.
NOTE: In no occasion is the flow reversed towards the patient.
Zero
valve
MiniC
Diff. Abs.
Pressure Pressure
Sensor Sensor
Diff.
Pressure
Sensor
minic_fem_gastubing.vsd
Pump
Gas out
8-4
Document no. 2050802-001
Product overview
8-5
13 1
1 I RESET RS485
25 14
2 I -15 VDC (not used)
3 I +15 VDIRTY
4 I +15VDC (not used)
5 I/O -DATA RS485
6 I/O DATA RS485
7 Ground and Shield
8 I -RESET RS485
9 n/c
10 n/c
11 n/c
12 n/c
13 Ground and Shield
14 I +24/+32 VDIRTY depends on power supply (not
used)
15 I Ground DIRTY
16 n/c
17 n/c
18 n/c
19 n/c
20 I GASFR (not used)
21 I CTSD (not used)
22 I TXDD (not used)
23 O RXDD (not used)
24 I +5 VDC (not used)
25 I +5 VDC DIRTY (not used)
8-6
Document no. 2050802-001
Maintenance and checkout
Planned maintenance
Service personnel shall perform the following checkout procedure completely every 12 months
after installation:
1. "Replacement of planned maintenance parts"
2. "Visual inspections"
3. "Functional checkout"
The instructions include a check form ("APPENDIX E") which should be filled in when performing
the procedures.
The symbol in the instructions means that the check form should be signed after
performing the procedure.
8-7
It is also recommended to replace the Mini D-fend water trap and the gas sampling line as part
of the planned maintenance procedure.
NOTE: See the supplies and accessories document delivered with the manual for compatible
accessories.
Nafion tube
Zero line air filter
2. Replace the special tube (Nafion) and check the condition of the internal tubing.
• Replace the 300 mm nafion tube in the sample gas line between the Mini D-fend
water trap and the zero valve unit.
• Check that the tubing inside the module is not contaminated. Any contamination
inside the tubing may indicate that the valve or sensor is contaminated too. This can
increase a risk of faulty operation in valve or sensor. The gas sensor is not possible
to clean in the field. Therefore, replace the whole miniC assembly with a new one.
NOTE: The nafion tube do not include the silicon fittings they connect to. Use the original silicon
fittings unless they are not damaged or leaking.
3. Replace the Mini D-fend O-rings:
• Detach the Mini D-fend.
• Detach the old rubber O-rings that are around the metal Mini D-fend connectors e.g.
using a small flat blade screwdriver. Pay special attention not to scratch the metal
Mini D-fend connectors and thus causing leaking.
• Set the new rubber O-rings into place and attach a new Mini D-fend.
8-8
Document no. 2050802-001
Maintenance and checkout
2.3 Functional checkout
2.3.1 Test setup
Required tools
− A barometer
− A mass flowmeter for measuring air flow, minimum measurement range 100-300 ml/min,
accuracy 5% or better in the 100-300ml/min range.
− P/N: 755534-HEL Calibration Gas Regulator
− P/N M1006864 Calibration Gas Regulator, (US only)
− P/N: 755580 Calibration gas 5% CO2 and air, package of 4 cans
− P/N 755587 QUICK CAL calibration gas, (US only)
− 3 m / 10 ft gas sampling line
− A pressure manometer with either an integrated or a separate pressure pump
NOTE: See the supplies and accessories document delivered with the manual for compatible
accessories.
Connections
Disconnect the module from the monitor, if connected.
Monitor configuration
1. Configure the CO2 waveform field to the monitor screen.
2. check the CO2 unit: Airway Gas > CO2 Setup > Unit, make sure the unit is mmHg.
2.3.2 Procedure
1. Gas sampling system leak test
Check the gas sampling system for possible leakages.
• Disconnect the module from the monitor.
• Connect a new Mini D-fend water trap to the module.
8-9
• Connect a new gas sampling line to the sampling line connector in the water trap.
• Connect the other end of the gas sampling line to a pressure manometer and a
pressure pump.
• Block the "Sample Gas Out" connector.
• Pump 100 mmHg ± 20 mmHg (133 ± 26 mbar) pressure to the gas sampling system.
Let the pressure stabilize for approximately 10 seconds.
• Check that the pressure reading does not drop more than 6 mmHg (8 mbar) during 1
minute.
NOTE: The gas module shall be disconnected from the monitor during the leak test.
2. Sample flow check
Check the sample flow rate.
NOTE: Gas measurement is not available during the first 1 minute after the module is
connected due to warm-up. Message 'Calibrating gas sensor' is shown in the waveform field.
Wait until warm-up is completed before proceeding with the next steps.
• Connect the module with the gas sampling line to the monitor.
• Connect the gas sampling line to the sampling line connector in the water trap.
• Connect the other end of the gas sampling line to a flowmeter.
• Check the sample flow rate reading from the flowmeter. The flow rate shall be within
the specification limit 150 ± 25 ml/min.
NOTE: Readjustment is needed, if the measured value is not within the specification limits.
Adjust the sample gas flow rate according to the instructions in section “3.1. Sample flow rate
adjustment”.
NOTE: If the sampling pump is noisy it indicates possible problems with motor bearing. In this
case it is recommended to replace the noisy sampling pump with a new one.
3. Zero valve operation
Test the zero valve functionality.
• Connect the gas regulator to the calibration gas container.
• Connect the end of the gas sampling line to the regulator on the gas container.
Leave the regulator overflow port open to room air.
• Enter to gas service menu: select Monitor Setup > Install/Service (16-4-34) >
Service (26-23-8) > Parameters > Gas Unit > Gases.
• Start feeding calibration gas. Wait until the CO2 value rises to approximately 5%.
• Open the zero valve to room air by selecting Zero valve ctrl.
• Check that the CO2 values drop back near 0%.
• Stop feeding the calibration gas.
8-10
Document no. 2050802-001
Maintenance and checkout
• Turn the zero valve back to the normal measurement position by selecting ZERO to
MEAS.
4. Gas calibration
Perform gas calibration according to the instructions in section “3.2. Gas calibration”.
5. Ambient pressure
Use a barometer to check the operation of the absolute pressure sensor.
• Check that the ambient pressure value shown in the Gas service menu does not
differ more than ± 13.33 mbar from the value shown by the barometer.
6. Occlusion detection
• Block the tip of the sampling line by your finger.
• Check that message 'Sample line blocked' appears to the parameter window within
30 seconds.
7. Air leak detection
• Detach the Mini D-fend water trap.
• Check that message 'Check D-Fend' appears to the parameter window within 30
seconds.
8. Airway gases
• Breathe a minimum of 5 times to the tip of the sampling line.
• Check that a normal CO2 waveform appears to the waveform field and the EtCO2
and FiCO2 values are updated to the parameter window.
9. Apnea detection
• Stop breathing to the gas sampling line.
• Check that an 'Apnea' alarm appears to the message field within 30 seconds.
8-11
8-12
Document no. 2050802-001
Calibration and adjustments
Connections
• Ensure that the module is connected to the monitor.
• Ensure that you have a new Mini D-fend water trap in use.
• Connect a new gas sampling line to the sampling line connector in the water trap.
• Connect the other end of the gas sampling line to the flow meter.
NOTE: Before checking or adjusting the sample flow rate, make sure there is no leakage in the
sampling system.
8-13
8-14
Document no. 2050802-001
Calibration and adjustments
3.2.2 Procedure
NOTE: Gas calibration is not available during the first 5 minutes after the module is connected.
For maximum accuracy, let the monitor to warm up for 30 minutes before starting calibration.
NOTE: Gas calibration is not available during a 'Sample line blocked', 'Check D-Fend' and
'Check sample gas out’ alarm condition. Resolve the alarm condition before starting
calibration.
1. Enter to gas calibration menu: Airway Gas > Gas Calibration
2. The monitor will start automatic zeroing of the gas sensors. Wait until the message
'Zeroing' is replaced by a message 'Zero OK' message.
3. Open the regulator after a message 'Feed gas' is shown. The measured gas
concentrations are shown in real-time in the gas calibration menu.
Continue feeding the calibration gas until the measured gas concentrations are stabilized
and a message 'Adjust' is shown. Close the regulator.
4. Use the Trim Knob to adjust the gas readings shown in the Calibration menu to match
with the gas readings in the labelling of the calibration gas container. Press Trim Knob to
accept the adjusted values when the gas readings match each other.
5. Wait until a message 'Ok' is shown. NOTE: A message 'Zero error' is shown in case the
zeroing fails.
NOTE: A message 'Calibr.error' is shown, if you do not start feeding gas within 1 minute after
the automatic zeroing is completed, or if the calibration fails due to too large gain adjustment.
NOTE: If zeroing or calibration failed, select the Recalibrate to restart the calibration procedure
from the beginning.
8-15
4 Troubleshooting
The problems and solutions in this chapter represent only a few of the faults that you may
encounter and are not intended to cover every possible problem that may occur.
This chapter focuses on troubleshooting technical problems. For clinical and operation issues
refer to the "User’s Guide" for troubleshooting.
NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout”
each time after you have opened the module casing.
If the problem remains, call technical support for service. To ensure accurate problem solving,
please be prepared to provide the following information:
• Problem description and the troubleshooting done so far.
• Device information.
• Error messages displayed, if any.
• Other information, as requested.
8-16
Document no. 2050802-001
Troubleshooting
8-17
ETCO2 over scale >20% • abnormally high ETCO2 (permissive • let the module run without a
Shown until 32%, hypercapnia) sampling line until the CO2
specified range 0...20% • CO2 sensor contaminated sensor has dried out
• D-fend malfunction
• change D-fend
ETCO2>PaCO2 • unit is mmHg or kPa and ETCO2 is close to • change to “wet gas” by using
arterial PCO2 install/service menu
• “dry gas” as default
Abnormally high (or low) • Pressure transducer failure. • Replace miniCO2 assembly.
response to CO2, or
sudden occlusion
message.
Unable to perform gas • Gas calibration is not available during the • Wait until warm-up is
calibration. first 5 minute after the module is completed.
connected.
• Gas calibration is not available during • Resolve the alarm condition
‘Sample line blocked’, ‘Check D-fend’, and before starting calibration.
‘Check sample gas out’ alarm condition.
8-18
Document no. 2050802-001
Disassembly and reassembly
8-19
- pozidrive screwdrivers
- flat blade screwdriver
- pincers
- antistatic wristband
8-20
Document no. 2050802-001
Disassembly and reassembly
8-21
6 Service parts
6.1 Ordering parts
To order parts, Contact GE Healthcare. Contact information is available at
www.gehealthcare.com. Make sure you have all necessary information at hand.
NOTE: Perform the checkout procedure described in chapter “2. Maintenance and checkout”
after you have disassembled and reassembled the module.
8-22
Document no. 2050802-001
Service parts
8-23
8-24
Document no. 2050802-001
Software download instruction
A-1
1.5 Workflow
The following sections describe the sequential workflow for the software installation:
Sections 1.6 through 1.8 describe how to prepare the patient monitor(s) and the service PC for
the software installation.
Sections 1.9 through 1.11 describe the software transfer process using the GE Healthcare
Software Transfer Utility.
Sections 1.12 and 1.13 describe the software activation process and the functional check
procedure after activating the new software.
A-2
A-3
A-4
A-5
c. Select File -> Save. Specify the destination and file name of this text file (.txt) so you can
easily navigate to it.
d. From the GE Healthcare Software Transfer Utility window, select File ->Import IP File.
Navigate to the .txt file containing the IP addresses. The IP addresses display in the Add
Target IP Addresses list.
3. Verify these IP address(es) are the destination(s) where you want to transfer the software.
NOTE: To remove an incorrect IP address from the displayed list, select the IP address
and click the up arrow button.
A-6
4. Verify the transfer status of all selected software packages indicates Success in PC.
NOTE: If the transfer status show the Failed information, please do not activate the
software. You should redo the software download.
5. Select Exit.
Software Transfer to destination monitor(s) is now completed. Next Step is to activate the
transferred software.
NOTE: Please do not restart the monitor manually. Activate the software first. The monitor will
automatically restart by activation.
A-7
A-8
B-1
<5% UT <5% UT
(>95% dip in UT) (>95% dip in UT)
for 5 sec for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic field
(50/60 Hz) magnetic should be at levels characteristic of a
field typical location in a typical
IEC 61000-4-8 commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
B-2
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicated theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the equipment is used exceeds the applicable RF compliance level above, the
equipment should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the equipment.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
B-3
B-4
Monitor Installation
C-1
Notes
3.2.1. Start-up
3.2.2. Display
3.2.5. Recorder
3.2.7. Conclusion
Notes
Signature
C-2
Monitor Installation
D-1
Notes
4.1.2. General
Notes
4.2.1. General
4.2.2. Display
4.2.3. Keyboard(s)
Notes
. ECG measurement
2. Pacemaker Detection
3. Asystole Detection
D-2
Notes
. Respiration measurement
5. Respiration Rate
6. Apnea Detection
Notes
. Temperature measurement
7. Temperature detection
Notes
8. Zeroing
9. Static Pressure
Notes
. SpO2 measurement
Notes
Notes
4.2.6. Loudspeaker
4.2.9. Batteries
D-3
4.2.10. Network
Notes
Notes
Signature
D-4
Notes
4. Gas calibration
5. Ambient pressure
6. Occlusion detection
8. Airway gases
E-1
Document no. 2050802-001
B40/B20 Patient Monitor
9. Apnea detection
Notes
Signature
E-2
0459