! Microarray Analysis
! Microarray Analysis
Chapter 1
I N T R O D U C T I O N T O M I C R O A R R AY A N A LY S I S
1.0 Introduction
Microarray analysis has emerged in the last few years as a flexible
method for analyzing large numbers of nucleic acid fragments in parallel.
Its origins can be traced to several different disciplines and techniques.
Microarrays can be seen as a continued development of molecular
biology hybridization methods, as an extension of the use of fluorescence
microscopy in cell biology, as well as a diagnostic assay using capture to
solid surface as a way to reduce the amount of analytes needed. The
convergence of ideas and principles utilized in these fields, together with
technological advancements in preparing miniaturized collections of
nucleic acids on solid supports, have all contributed to the emergence
of microarray and microchip technologies.
● 1
M I C R O A R R AY
hb 63-0048-49 ● 2
C H A P T E R 1 : I N T R O D U C T I O N T O M I C R O A R R AY A N A LY S I S
Labelled
probe
Array
spotter
Hybridization
Array scanner
Scanning and
data analysis
9 4 7
3 1
6
2 2
● 3
M I C R O A R R AY
1.1.2 Microarrays
Microarrays consist of a collection of nucleic acid sequences immobilized
onto a solid support so that each unique sequence forms a tiny feature,
called a ‘spot’ or ‘target’. These nucleic acids are obtained in numerous
ways, and there are different methods for depositing them onto microarray
slides (refer to chapter 3). The size of these spots varies from one system to
another, but it is usually less than two hundred micrometers in diameter. A
glass slide or glass wafer acts as the solid support onto which up to tens
of thousands of spots can be arrayed in a total area of a few square
centimeters (Fig 4).
Fluorescent dyes, and especially the cyanine dyes Cy3 and Cy5, have
been adopted as the predominant label in microarray analysis.
Fluorescence has the advantage of permitting the detection of two or
more different signals in one experiment. This has allowed investigators
to perform comparative analysis of two or more samples on one
microarray. It has also increased the accuracy and throughput of
microarray analysis over filter-based macroarrays, in which only one
radioactively labelled sample can be conveniently analyzed at a time.
hb 63-0048-49 ● 4
C H A P T E R 1 : I N T R O D U C T I O N T O M I C R O A R R AY A N A LY S I S
● 5
M I C R O A R R AY
hb 63-0048-49 ● 6
C H A P T E R 1 : I N T R O D U C T I O N T O M I C R O A R R AY A N A LY S I S
● 7
M I C R O A R R AY
hb 63-0048-49 ● 8
C H A P T E R 2 : G E N E T I C C O N T E N T O F M I C R O A R R AY S
Chapter 2
G E N E T I C C O N T E N T O F M I C R O A R R AY S
2.0 Introduction
The genetic content of microarrays resides in the immobilized nucleic
acid sequences on the microarray. The identity of these sequences
determines what information can be obtained from array experiments
and how reliable this information is. As microarrays enable simultaneous
interrogation of up to tens or hundreds of thousands of targets with one
or more labelled probes, generation of accurate data demands that only
specific interactions result in detectable signals. Several strategies for
preparing the immobilized target nucleic acids for microarrays exist
(Fig 6). These nucleic acids can be synthesized directly on the microarray
or they can be purified cDNA clones, other DNA fragments or
oligonucleotides, which are deposited onto the array by a printing
process. This flexibility of using either partially characterized sequences
or defined oligonucleotides as targets has improved the application of
microarray analysis to different biological problems in a number of
species.
A G T T C G A G A T T C C A
T C G A C G C A T G T G C A
● 9
M I C R O A R R AY
hb 63-0048-49 ● 10
C H A P T E R 2 : G E N E T I C C O N T E N T O F M I C R O A R R AY S
● 11
M I C R O A R R AY
hb 63-0048-49 ● 12
C H A P T E R 2 : G E N E T I C C O N T E N T O F M I C R O A R R AY S
% retention
100
90
80
Fig 8. Retention of microarray targets of 70
different lengths on aminosilane-treated 60
microarray slide. 50
40
30
20
10
0
200 300 400 500 600 700 800
Size of PCR insert
● 13
M I C R O A R R AY
A control system that consists of both control targets and RNA spikes
can monitor most aspects of the microarray process. A control strategy
adopted in the Lucidea™ Universal ScoreCard™ combines the use of
different types of control targets for spotting onto microarray slides and
spikes added to samples before labelling. Together these elements cover
aspects of slide printing, sample labelling, slide pretreatment, and
hybridization. In a typical experiment, up to 24 replicas of the ScoreCard
hb 63-0048-49 ● 14
C H A P T E R 2 : G E N E T I C C O N T E N T O F M I C R O A R R AY S
● 15
M I C R O A R R AY
hb 63-0048-49 ● 16
C H A P T E R 3 : M A N U F A C T U R I N G O F M I C R O A R R AY S L I D E S
Chapter 3
M A N U F A C T U R I N G O F M I C R O A R R AY S L I D E S
3.0 Introduction
Microarray analysis is invariably performed on a glass slide, which
enables the performance of high-throughput miniaturized hybridization
assays with fluorescently labelled samples—a significant improvement
over the use of membrane support.
■ production method
■ microarray slide
● 17
M I C R O A R R AY
This method (Fig 9) produces arrays of small features that are anchored
at their 3' ends to the array surface. Each feature is made up of
oligonucleotides that all have the same nucleotide sequence. These
arrays have a high density: an area of 1.6 cm can contain up to 400 000
features. Additionally, multiple arrays can be synthesized simultaneously
onto a large glass wafer, further automating the manufacturing process.
The wafers are then cut into individual arrays in preparation for use.
Mask Mask
C A T A T
Chemical
coupling A G C T G
O O O O O HO HO O O O T T O O O T T O O O T T C C O T T C C G
T– C– Repeat
3.1.2 Deposition
Using common deposition methods, purified nucleic acids are attached to
a modified glass slide. Typically, small volumes of nucleic acid solution—
nanoliters or picoliters—are transferred onto the glass slide. Deposition
methods are equally suitable for preparing microarrays containing
oligonucleotides, cDNA sequences, as well as genomic DNA. Deposition
methods are commonly used for preparing customized microarray slides.
hb 63-0048-49 ● 18
C H A P T E R 3 : M A N U F A C T U R I N G O F M I C R O A R R AY S L I D E S
Spot reproducibility
The spots should be of uniform size and shape, and they should contain
equal amounts of the target nucleic acids. This requires careful
calibration and matching of individual printing pens.
Environmental control
Environmental conditions can have a significant effect on the quality
of the spotted slides. Clean environment is important because dust
particles can become fixed onto slides, causing background signals
in microarray hybridization and spot finding problems during data
analysis. Controlling the humidity helps to avoid changes in sample
concentrations due to evaporation during printing and when spots are
drying. High humidity levels may cause spots to smear whereas low
humidity levels may cause evaporation from the sample plate. Under
high temperature conditions targets will dry rapidly at the outer edges
of the spot, thereby causing poor spot uniformity. This effect can cause a
donut-shaped spot morphology. A humidity between 10–70% has been
found to be most suitable for a microarray application.
Sample carryover
There must not be any carryover of previous target during the printing
process.
Throughput
The printing process should be fast to allow timely printing of slides. The
total time need for sample retrieval, printing, and washing of the printer
pens needs to be considered.
● 19
M I C R O A R R AY
Syringe
Pressure
hb 63-0048-49 ● 20
C H A P T E R 3 : M A N U F A C T U R I N G O F M I C R O A R R AY S L I D E S
Contact deposition requires less target nucleic acid solution than the
non-contact methods and also results in smaller spots that can be packed
more densely on the microarray surface.
● 21
M I C R O A R R AY
■ Patent pending, stainless steel capillary pens that conserve sample and
uniformly deposit picoliter volumes of target (Fig 13). From a single
sample uptake of less than 200 nl, up to 150 spots can be spotted in
duplicate, across each of 75 slides. The design of the pens (Fig 14)
minimizes clogging with target solution and simplifies washing after
each sample, resulting in no detectable carryover or mixing of samples
Fig 12. Lucidea Array Spotter.
during printing. To achieve good spot uniformity, the pens in each pen
set are quantitatively tested during manufacturing to ensure performance.
hb 63-0048-49 ● 22
C H A P T E R 3 : M A N U F A C T U R I N G O F M I C R O A R R AY S L I D E S
Uniform signals
Spots Scanned
The uniformity and thickness of the surface coating on the slide is critical
for good quality microarray results, as this will influence spot uniformity
and morphology, DNA binding, as well as background signals from
microarray hybridization. Variation in slide coating can contribute to the
variation in microarray signals and decrease the resolution of a
microarray experiment. Uneven slide coating can also lead to poor
attachment of deposited nucleic acid, which may come loose during
microarray hybridization.
● 23
M I C R O A R R AY
DNA
NH H NH
sodium
borohydride
hb 63-0048-49 ● 24
C H A P T E R 3 : M A N U F A C T U R I N G O F M I C R O A R R AY S L I D E S
Amine slides
Amine groups can be introduced onto microarray slides by treating
cleaned glass with aminosilane, such as 3-aminopropyltrimethoxysilane
(Fig 18). Vapor treatment of slides gives generally better results than
deposition by a dipping method (37). Unmodified DNA can be attached
to amine-modified slides, via interactions between negatively charged
phosphate groups on the DNA and the positively charged slide surface.
This interaction helps ensure denaturation of the DNA as well as increase
its binding affinity to the slide surface. UV treatment can be used to
further immobilize the DNA onto the slide surface. Attachment via
electrostatic interactions is suitable for binding DNA fragments that are
longer than 60–70 nucleotides. For attaching oligonucleotides to amine-
modified glass, chemical coupling methods must be used (2).
DNA
Fig 18. Attachment of unmodified DNA to an
amine-modified slide surface. The DNA binds
O O O O O O
to the surface of the slide via an electrostatic
P=O P=O P=O
interaction. The positive amines in the silane
coating will attract the negative phosphate O- O- O-
backbone of the DNA.
+ + +
NH3 NH3 NH3
Solid support
Poly-lysine slides
Treatment of the slide with poly-lysine creates a positively charged
surface to which unmodified DNA can bind via ionic interactions (38).
● 25
M I C R O A R R AY
Sample layer
Fig 19. Diagram of the structure of a Silane layer
reflective microarray slide. Silicon dioxide layer
Aluminium layer
Glass slide
sample
Fig 20. The enhancement of signal by a 1/4 Wavelength
reflective slide. Silicon dioxide layer
Aluminium layer
Glass slide
hb 63-0048-49 ● 26
C H A P T E R 3 : M A N U F A C T U R I N G O F M I C R O A R R AY S L I D E S
Purified + primers
The targets, once attached to the microarray surface, are only available
for hybridization when they are present in a denatured, single-stranded
form. This can be achieved by spotting the targets under denaturing
Fig 21. Purification of PCR-amplified
conditions. Typically, targets are dissolved in high salt solutions such
targets. cDNA targets were amplified
as 3 × SSC, or in denaturing solvents such as DMSO.
with PCR, then purified with column
chromatography. The indicated reagents
were added to the purified target DNA
before spotting. Significantly decreased
3.5 Critical success factors for microarray
hybridization signals were observed from preparation
all targets containing impurities as ■ Always handle microarray slides in a clean environment.
compared with purified targets.
■ Never use gloves that contain powder as the powder will invariably
get onto the slides and cause background signals.
■ Never touch the array surface, only handle slides from sides.
■ Make sure that spotting is carried out under known and controlled
temperature and humidity.
● 27
M I C R O A R R AY
hb 63-0048-49 ● 28
C H A P T E R 4 : F L U O R E S C E N T L A B E L S I N M I C R O A R R AY A N A LY S I S
Chapter 4
F L U O R E S C E N T L A B E L S I N M I C R O A R R AY A N A LY S I S
4.0 Introduction
Most researchers performing microarray analysis prefer to use
fluorescent dyes as labels in these experiments as their use offers high
sensitivity of detection and enables detection of different dyes
simultaneously. Furthermore, fluorescent dyes do not carry the hazards
associated with radioactive markers. In this chapter, the general
properties of fluorescence and CyDyeTM fluorophores are discussed.
● 29
M I C R O A R R AY
4.1.1 Excitation
Extinction coefficient (ε) is a measure for a fluorophore’s ability to
absorb light energy. When a photon of light energy (hvEX) of the
appropriate wavelength is absorbed by a fluorophore, an electron is
boosted to a higher, unstable excited energy state. The difference between
the ground state (S0) and the higher energy state (Sn) is a property of the
fluorophore; it is equivalent to the energy of the photon absorbed.
Because photons with shorter wavelengths have higher energy, the
shorter the wavelength of the absorbed photon, the higher the excited
energy state reached by the fluorophore. The wavelength at which the
fluorophore has maximum excitation is determined by the structural
properties of that fluorophore.
4.1.3 Emission
Once the fluorophore has reached the lowest vibrational energy level
Sn
Vibration decay
within the electronic excited state (S1), the electron falls from the excited
S1 state to the ground state (S0). It is at this point in the decay process that
light is emitted at a specific wavelength (hvEM). Because some energy
Absorption Emission between excitation and emission has already been lost, the emitted
photon has less energy than the original photon absorbed by the
fluorophore (Fig 23). Therefore, the emitted light has a longer
S0
wavelength. The difference between the maximum excitation wavelength
and the maximum emission wavelength is known as the Stokes shift
Fig 23. Diagram illustrating the energy (hvEX – hvEM).
levels of the fluorescence process.
hb 63-0048-49 ● 30
C H A P T E R 4 : F L U O R E S C E N T L A B E L S I N M I C R O A R R AY A N A LY S I S
4.1.4 Photobleaching
The fluorescent process is rapid (10-8 seconds) and cyclical, enabling the
fluorophore molecule to be excited repeatedly. It must be considered,
however, that the excited state of a fluorophore is generally more
chemically reactive than the ground state. In conditions of intense light,
the fluor may gradually lose its fluorescent properties, a phenomenon
known as "photobleaching". This results in lower fluorescent output
from the fluorophore after prolonged exposure to light or repeated
excitation. For more photosensitive dyes, such as fluorescein,
photobleaching may be a significant factor when using instrumentation
with high laser power. In contrast, as seen in Figure 24, CyDye fluors are
more resistant to photobleaching, which makes them more suitable for
multiple applications.
Normalized
fluorescence
100
90
80
70
Fig 24. Photostability of fluorophores 60
Cy3
under "natural" light conditions. 50
Cy3.5
40 Cy3B
30 Cy5
20 Fluorescein
10
0
0 10 20 30 40 50 60
● 31
M I C R O A R R AY
Another advantage of using fluorescent dyes as labels is that they are less
hazardous than radioactive compounds and offer significantly increased
stability, or longer shelf life. The availability of fluorophores conjugated
to many different chemical groups enables the labelling of nucleic acids,
proteins, lipids, and carbohydrate molecules. Furthermore, fluorescent
dyes can also be used as reporters to detect changes in pH, ion
concentrations, or dye environment.
Fluorescence
100
Cy3 excitation
90
Cy3 emission
80
Cy5 excitation
70 Cy5 emission
60
Fig 25. Excitation and emission spectra of 50
Cy3 and Cy5 dyes. 40
30
20
10
0
400 450 500 550 600 650 700 750 800
Wavelength (nm)
hb 63-0048-49 ● 32
C H A P T E R 4 : F L U O R E S C E N T L A B E L S I N M I C R O A R R AY A N A LY S I S
Quenching can occur when samples have been labelled too densely or
when too much labelled sample is used in hybridization. Over-labelling
not only results in the loss of linearity between fluorescent signal emitted
and the number of fluors present, but will, in extreme cases, reduce the
signal to levels that cannot be observed. In microarray analysis, quenching
may also occur when two probe strands come into close proximity of
each other. This is likely to be most apparent in the presence of highly
expressed transcripts, where a very large number of labelled molecules
are bound densely at a precise location. See chapter 6, section 6.2.4 for
further overview of this process.
Fluorescence
A C C C T A G G A T G C G T G C G C T A A G A A C A T G G G G A T T Interaction & absorption
A C C C T A G G A T G C G T G C G C T A A G A A C A T G G G G A T T
● 33
M I C R O A R R AY
Fluorescence
100 Cy2
90 FluorX
80 Cy3
Cy3.5
70
Cy5
60
Cy5.5
Fig 27. The emission spectra of some 50 Cy7
CyDye fluors. 40
30
20
10
0
450 500 550 600 650 700 750 800 8500
Wavelength (nm)
hb 63-0048-49 ● 34
C H A P T E R 4 : F L U O R E S C E N T L A B E L S I N M I C R O A R R AY A N A LY S I S
HO3S SO3H
N+ N
Fig 29. Structure of Cy3 NHS ester
(3 carbon bridge) and Cy5 NHS ester
(5 carbon bridge ).
Cy3 NHS ester O
Excitation max = 548 nm
Emission max = 562 nm O O N
O
HO3S SO3H
N+ N
SO3H SO3H
HO3S SO3H
+
N N
● 35
M I C R O A R R AY
■ The photostability of Cy3 and Cy5 is higher than that of other widely
used dyes (Fig 24), making the use of CyDye fluors more practical.
■ Cy3 and Cy5 are bright dyes that give strong fluorescent signals.
■ The good spectral separation of Cy3 and Cy5 means that each can be
excited at a different wavelength and their emissions can be detected
separately. This requires the use of two different lasers, typically a
532 nm and 633 nm laser for Cy3 and Cy5, respectively. By choosing
optical filters that only collect emitted light from part of the spectra,
Cy3 and Cy5 signals can be measured with minimal overlap.
hb 63-0048-49 ● 36
C H A P T E R 4 : F L U O R E S C E N T L A B E L S I N M I C R O A R R AY A N A LY S I S
● 37
M I C R O A R R AY
hb 63-0048-49 ● 38
C H A P T E R 4 : F L U O R E S C E N T L A B E L S I N M I C R O A R R AY A N A LY S I S
● 39
M I C R O A R R AY
hb 63-0048-49 ● 40
C H A P T E R 5 : P R E P A R AT I O N O F R N A S A M P L E S F O R M I C R O A R R AY A N A LY S I S
Chapter 5
P R E P A R AT I O N O F R N A S A M P L E S F O R M I C R O A R R AY
A N A LY S I S
5.0 Introduction
There are multiple steps involved in isolating and preparing RNA
samples to be used for microarray analysis. Discussed in this chapter
are factors affecting the quality of analyzed RNA, protecting RNA
samples from contamination and degradation, purifying RNA, and
finally, characterizing the purified RNA.
● 41
M I C R O A R R AY
■ Use only disposable plastic tubes and pipette tips for RNA work.
Clean plastic-ware should be baked at 120 °C for at least 20 min to
reduce RNAse (and other nuclease) contamination. Before baking,
pack centrifuge tubes into glass beakers and cover them with
aluminium foil. Additionally, wrap tip racks in foil to keep the
baked items free of contamination after treatment.
■ Set aside reagents for use in RNA work only. Always use baked or
disposable spatulas and weighing trays for measuring out reagents.
hb 63-0048-49 ● 42
C H A P T E R 5 : P R E P A R AT I O N O F R N A S A M P L E S F O R M I C R O A R R AY A N A LY S I S
■ The longer the time elapsed between collecting cells or tissue and
preparing a denatured homogenate, the more chance there is for RNA
degradation to take place. Furthermore, during this time, changes in
gene expression can also take place as cells react to the change in their
environment. Therefore it is good practice to work as quickly as
possible when preparing biological samples for RNA extraction.
● 43
M I C R O A R R AY
hb 63-0048-49 ● 44
C H A P T E R 5 : P R E P A R AT I O N O F R N A S A M P L E S F O R M I C R O A R R AY A N A LY S I S
● 45
M I C R O A R R AY
hb 63-0048-49 ● 46
C H A P T E R 5 : P R E P A R AT I O N O F R N A S A M P L E S F O R M I C R O A R R AY A N A LY S I S
■ Probes prepared from mRNA usually give higher signal to noise values
on arrays than probes prepared from similar amounts of
total RNA.
■ The yield of labelled cDNA is higher from mRNA than from total
RNA, because alternative priming strategies that use oligo(dT) can be
used.
● 47
M I C R O A R R AY
Fig 31. Schematic illustration of typical 5.4.1 Measuring the amount of RNA
results obtained from analyzing the The amount of RNA can be quantified by measuring the absorbance of
quality of RNA with a denaturing RNA solution at 260 nm. Pure RNA solution that contains 40 µg of
agarose gel. Denatured RNA samples
RNA per mL will give absorbance of 1 AU. This method works well with
were electrophoretically separated in
an agarose gel and visualized with UV clean RNA samples that are devoid of other contaminating substances.
light after ethidium bromide staining. Unfortunately, this is rarely the case with total RNA samples purified
Lane 1. Intact total RNA with both with simple methods. Proteins and DNA or other compounds, such as
ribosomal RNA bands present as sharp those released from some affinity chromatography columns, will absorb
and bright bands. High abundance
at 260 nm. This absorbance is indistinguishable from that of RNA and
transcripts can be discerned as distinct
will give an artificially high estimate for the amount of RNA in the
bands. Lane 2. Partially degraded total
RNA. Although ribosomal bands are sample. By measuring absorption spectra from 200 to 350 nm, some
still visible, the average size of RNA is conclusions on the purity of RNA can be made. However, the presence of
smaller, and no distinct transcript DNA in the sample can still go undetected. Therefore, it is important to
bands are visible. Lane 3. Badly use an RNA isolation method that specifically removes DNA from the
degraded total RNA. Most of the RNA
purified sample.
is shorter than 1 kb, and ribosomal
RNA bands appear as smears. Lane 4.
Intact mRNA. Fragments longer than 4
kb should be visible and abundant
5.4.2 Verifying the quality of RNA
transcripts should appear as distinct Agarose gel electrophoresis performed under denaturing conditions can
bands. Lane 5. Degraded mRNA. The be used to analyze the quality of RNA. Gels containing formaldehyde
transcripts appear as a fast migrating
have been traditionally used for this purpose, but denaturation of RNA
smear.
hairpins by glyoxal has gained popularity, as this method does not
involve the use of large quantities of harmful chemicals (42, 43, 46).
Both methods, followed by staining of the gels with nucleic acid
binding stains, such as ethidium bromide or Vistra Green™, are useful
for observing overall differences in RNA quality. It is relatively easy
to see if a sample is badly degraded as the ribosomal bands appear as
smears. However, it may be difficult to detect more subtle degradation of
transcripts unless large amounts of sample are used. Figure 31 shows a
schematic illustration of typical results from RNA gel electrophoresis.
Transferring the gel onto a membrane for Northern blotting analysis
can give more precise information, as the status of specific transcripts
in different samples can be analyzed.
hb 63-0048-49 ● 48
C H A P T E R 5 : P R E P A R AT I O N O F R N A S A M P L E S F O R M I C R O A R R AY A N A LY S I S
■ Prepare all the samples for microarray analysis with the same
protocol.
● 49
M I C R O A R R AY
hb 63-0048-49 ● 50
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
Chapter 6
S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
6.0 Introduction
In differential gene expression analysis two or more RNA samples are
compared to identify differences in the abundance and identity of the
transcripts they contain. In order to convert the information contained in
the transcript populations into a form that can be hybridized with
microarrays and subsequently detected, the transcript populations need
to be labelled. This can be achieved using different methods; an ideal
method retains both the information carried by the identity of the
transcripts as well as their relative abundance in the sample.
● 51
M I C R O A R R AY
hb 63-0048-49 ● 52
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
Fluorescence
A C C C T A G G A T G C G T G C G C T A A G A A C A T G G G G A T T Interaction & absorption
A C C C T A G G A T G C G T G C G C T A A G A A C A T G G G G A T T
● 53
M I C R O A R R AY
CyDye
Fluorescence
T G C C A C C C T A C G C C C C C G C T C C C T C T A C C C C T A A
Cy-dCTP Interaction & absorption
U G C C A C C C U A C G C C C C C G C U C C C T C U A C C C C U A
Cy-dUTP
hb 63-0048-49 ● 54
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
Increasing labelling density is not the best solution for increasing signal
from microarrays. Using a lower labelling density is more likely to result
in higher signal. As fluorescent molecules tend to be fairly large, high
labelling densities can lead to changes in the hybridization kinetics of
labelled molecules (47). The melting temperature (Tm) of highly
substituted nucleic acids are lower than those of unlabelled nucleic acids
and can result in lower hybridization signal.
Eukaryotic transcripts usually have a poly-dA tail at their 3' ends. This
property can be exploited by using a complementary poly-dT sequence
to capture polyadenylated transcripts away from other RNA species,
as well as from other molecules and impurities. Because of the additional
purification steps involved in the preparation of mRNA, mRNA samples
tend to give higher signal to noise values on microarrays than total
RNA samples.
● 55
M I C R O A R R AY
hb 63-0048-49 ● 56
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
C T T T T T T T T T T T
cDNA strand primed
with anchored oligo(dT) G A A A A A A A A A A A A A A
● 57
M I C R O A R R AY
mRNA or RNA
hb 63-0048-49 ● 58
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
● 59
M I C R O A R R AY
6.5.1 Principle
One of the simplest and most popular labelling strategies is to convert
mRNA population into a labelled first-strand cDNA population. This is
achieved by copying the transcripts into cDNA molecules with a reverse
transcriptase while incorporating a modified CyDye nucleotide. The
cDNA synthesis can be primed with a choice of primers including
random primers, anchored oligo(dT) as well as gene specific primers.
This allows the use of both mRNA and total RNA as sample.
Nucleotide,
CyDye nucleotide
and enzyme added
NaOH/Hepes added
cDNA synthesis
NaOH/Hepes added
RNA degradation
Purification columns
cDNA purification
hb 63-0048-49 ● 60
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
6.5.2 Optimization
The incorporation of fluorescently labelled nucleotides is the rate-
limiting step of this labelling method. This is because all polymerase
enzymes, both DNA and RNA dependent polymerases, incorporate
unlabelled nucleotides more efficiently than larger fluorescent dye
nucleotides. The incorporation kinetics are very much dependent on
the identity of the enzyme, and even homologous enzymes can have
very different properties. The ratio between the fluorescently labelled
nucleotide and the corresponding unlabelled nucleotide in the labelling
reaction determines the incorporation of fluors into cDNA. As different
fluorescent dyes and nucleotides have different structures, it is necessary
to optimize this ratio separately for each combination of dye and
nucleotide. Furthermore, optimized ratios determined for one enzyme
will not necessarily give optimal results with other enzymes.
● 61
M I C R O A R R AY
6.6.1 Principle
The shortcomings of the first-strand cDNA labelling method and the
availability of CyDye as reactive N-hydroxyl succinimidal dyes (NHS-dyes)
have led to the development of cDNA post-labelling method.
Annealing
Nucleotide,
CyDye nucleotide
and enzyme added
Coupling of CyDye
NaOH/Hepes added
to modified cDNA
cDNA synthesis
Hydroxylamine
quencher added
NaOH/Hepes added
RNA degradation
Purification columns
cDNA purification
Elimination of
free reactive dye
Purification columns
Purification of
CyDye labeled cDNA
hb 63-0048-49 ● 62
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
R′ = CyDye Conjugate
O O
O O
R NH2 + R + N OH
N R′
R′ N
O O
O H
Amide group on cDNA NHS ester dye Amide bond coupling NHS leaving group
CyDye to cDNA
● 63
M I C R O A R R AY
6.6.2 Benefits
The main benefits of this method over the first-strand cDNA labelling
method are derived from the more efficient incorporation of the smaller
aminoallyl nucleotide compared with the bulkier CyDye nucleotides. As
a consequence, the yield of cDNA is considerably higher than with the
first-strand labelling method. The cDNA fragments synthesized in the
post-labelling method are also longer. Further benefits of the post-labelling
method stem from the chemical labelling step itself. The number of
amine groups on cDNA is the main factor influencing labelling density.
The sequences of the cDNAs being labelled do not have a major impact
on labelling outcome. Because of this, more random attachment of labels
is achieved than with the first-strand cDNA labelling method. Furthermore,
as the labelling process is not dependent on the structure of different
fluorescent dyes, it is easier to achieve equal labelling with both Cy3 and
Cy5, and the labelling method introduces less variation into microarray
analysis. This is illustrated in tighter scatter plots derived from microarray
hybridizations in which the mRNA samples were labelled with the post-
labelling method (Fig 39). As the extent of experimental variation is
reduced, the detection of smaller changes in gene expression between two
samples is improved.
all probes.
100 100
100 1000 10000 100000 1000000 100 1000 10000 100000 1000000
hb 63-0048-49 ● 64
6.6.3 Chemical considerations
Coupling of CyDye NHS-ester to amine-modified cDNA requires mildly
alkaline pH, which is provided by the coupling buffer. However, if they
are not carefully removed, buffer components or acidic residues from
preceding steps can alter the pH of this buffer. Furthermore, any
amine groups present on other compounds will compete with the
amine-modified cDNA for CyDye incorporation. Hence the free
aminoallyl-dUTP, Tris-buffer, and reverse transcriptase enzyme must
be removed from the cDNA preparation before labelling. This can be
best achieved with affinity column chromatography methods, such as
GFX columns, in which amine-modified cDNA is bound to the matrix,
other compounds are washed away, and cDNA is finally eluted with
water. Alternatively, standard ethanol precipitation also works well
and provides the added benefit of concentrating the cDNA ready for
CyDye coupling.
NHS-esters are readily hydrolyzed with water, even with the small
amount of moisture present in laboratory air. Because of this, aliquots
of CyDye NHS-esters must always be stored desiccated and protected
from light. Storing CyDye NHS-esters in solution can lead to rapid
loss of reactivity. As these reactive dyes were originally developed for
protein labelling, the quantities provided commercially were adjusted
for this application. This necessitated aliquoting of reactive dyes before
their use for microarray sample labelling and frequently resulted in
decreased activity as a consequence of handling and storage. The
availability of individually foil-packed, pre-dispensed, and freeze-dried
aliquots of Cy3 and Cy5 NHS-esters for microarray analysis has
removed this problem. Furthermore, the quality of the CyDye NHS-ester
in CyDye Post-Labelling Reactive Dye Packs is higher than available
otherwise.
● 65
M I C R O A R R AY
A A A A A A A A A
A A A A A A A A A
NTPs
Double-stranded
T7 IVT templates
(T7 RNA polymerase + NTPs)
Single-stranded
amplified RNA
(T7 RNA polymerase + NTPs)
hb 63-0048-49 ● 66
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
A A A A A A A A A
A A A A A A A A A
NTPs
Double-stranded
T7 IVT templates CyFye-NTPs
(T7 RNA polymerase + NTPs)
Single-stranded
amplified RNA
(T7 RNA polymerase + Cy-dUTP
+ unlabelled NTPs)
● 67
M I C R O A R R AY
hb 63-0048-49 ● 68
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
● 69
M I C R O A R R AY
hb 63-0048-49 ● 70
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
100
Cy3
80
250 ng Cy5
Total signal from
spot set 60
100 ng 40
20
50 ng
0
a) b) 500 ng mRNA 250 ng mRNA 100 ng mRNA 50 ng mRNA
Probe reaction template
150
Cy3
10 µg of total RNA Cy5
100
Relative signal
50
5 µg of total RNA
0
10 µg 5 µg 2.5 µg
c) d)
Amount of total RNA
Fig 43. The use of CyScribe First-Strand cDNA Labelling Kit with small
template amounts. The indicated amounts of mRNA and total RNA
templates were labelled with Cy3 and Cy5 in duplicate reactions, purified,
and all of the recovered probe was used in a microarray hybridization with
duplicate slides. Panels A and C show part of microarray images obtained
with mRNA probes and total RNA probes, respectively. Panels B and D
show quantified Cy3 and Cy5 signals from these arrays. As the amounts of
probes were not adjusted before hybridization, a high amount of variation
in signal intensities was observed in this experiment.
● 71
M I C R O A R R AY
hb 63-0048-49 ● 72
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
● 73
M I C R O A R R AY
Normalized
log ratio
Fig 44. Identification of muscle-specific 2.75
Gene number
hb 63-0048-49 ● 74
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
Average
normalized
log ratio
Fig 45. Differential gene expression determined 2.5
2.0
with CyScribe First-Strand cDNA Labelling Kit 1.5 CyScribe Post-Labelling Kit
and CyScribe Post-Labelling Kit for 31 randomly 1.0 CyScribe First-Strand
0.5 cDNA Labelling Kit
chosen genes in skeletal muscle and placental 0
-0.5
mRNA samples. Data from two replica slides, -1.0
each of which contained two identical spot sets -1.5
-2.0
for each of the labelling methods, is shown. 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31
Gene number
● 75
M I C R O A R R AY
CyDye
Reagent
G A U T C G C C A A U C A G G A G C U C A G A U G C A A A A A A
G A U T C G C C A A U C A G G A G C U C A G A U G C A A A A A A
hb 63-0048-49 ● 76
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
Quantify pmoles of
CyDye per labelling
Hybridize
Hybridize
● 77
M I C R O A R R AY
6.15.2 Application
Labelling reactions performed with CyScribe Direct mRNA Labelling
Kit can be scaled up or down to accommodate different amounts of
template: as little as 250 ng or as much as 5 µg of mRNA can be labelled
in a single reaction. The ratio of the CyDye Direct Labelling Reagent to
mRNA determines the labelling density achieved with the kit. Extending
the labelling time beyond 1 h can result in higher labelling efficiency
than shorter times. The length of transcripts is an important factor in
determining the intensity of hybridization signals from mRNA probes
labelled with the CyScribe Direct mRNA Labelling Kit. As seen in
Figure 48, the expression patterns obtained with the direct labelling
method differ considerably from those generated with the first-strand
labelling method. This evidence suggests that the direct labelling method
may be advantageous for analyzing gene expression from transcripts
that are difficult templates for reverse transcription.
The CyScribe Direct mRNA Labelling Kit has been developed for use
with purified mRNA that is free from contaminating DNA, proteins,
or nucleotides. The kit requires the use of purified mRNA as template
because the chemical labelling reaction cannot discriminate between
transcripts and other species of RNA. Labelling of total RNA will
result in low signal to noise values from microarray experiment.
CyScribe
15 pmol dye/slide
Fig 48. Hybridization of microarrays with
CyScribe Direct
varying amounts of mRNA labelled with Cy3 1 µg mRNA/slide
using CyScribe Direct mRNA Labelling Kit.
Results from a DNA probe generated with CyScribe Direct
10.5 µg mRNA/slide
CyScribe First-Strand cDNA Labelling Kit
are shown for comparison. CyScribe Direct
0.25 µg mRNA/slide
hb 63-0048-49 ● 78
C H A P T E R 6 : S A M P L E L A B E L L I N G F O R G E N E E X P R E S S I O N A N A LY S I S
● 79
M I C R O A R R AY
hb 63-0048-49 ● 80
C H A P T E R 7 : C H A R A C T E R I Z AT I O N O F L A B E L L E D M I C R O A R R AY P R O B E S
Chapter 7
C H A R A C T E R I Z AT I O N O F L A B E L L I N G M I C R O A R R AY P R O B E S
7.0 Introduction
When performing gene expression microarrays, it is important to
characterize the labelled probes in order to avoid experimental error
derived from variation between the probes. Even under carefully
controlled conditions, some differences in the amounts of nucleic acid
samples, labelling success, and recovery of material from the purification
system will occur. In order to account for these artifacts and to ensure
that these variations are not carried through to hybridization, it is highly
recommended that the properties of the labelled probes are determined
before microarray hybridization.
■ The quantity of CyDye and nucleic acid in the labelled probe can be
determined using spectrophotometry.
● 81
M I C R O A R R AY
❶
Sample preparation
■ Dilute an aliquot of purified probe with water. The required volume
depends on the size of available measuring cells. It is recommended
to use the smallest cells possible. Small volume disposable measuring
cells are available, and they offer the added advantage that a separate
cell can be used for each sample, thus minimizing cross contamination
from one sample to another. However, it is also possible to use regular
100 µl glass cells, but these need to be cleaned thoroughly by rinsing
with sterile water between samples.
❷
Measurement
■ Measure the absorbance spectrum of the sample from 200 to
700 nm against a blank. Although only the absorbance values at
the absorbance peaks are needed to calculate the amount of CyDye
in the sample, the shape of the absorption spectra can give additional
information about the quality of the labelled sample. The absorption
spectrum of CyDye contains two peaks, of which the second should be
of higher intensity (Fig 49). The first peak indicates the presence
of intermolecular interactions between different dye molecules. If both
peaks are of nearly similar intensity, this is a sign of quenching, i.e.
loss of fluorescent signal because absorbed light energy is spent
on intermolecular interactions. This is usually associated with over-
labelling of the sample.
Normalized
Fig 49. Quantification of the amount Absorbance
Aggregation
of CyDye in a labelled probe with 1.0 quenching
spectrophotometry: examples of 0.9
hb 63-0048-49 ● 82
C H A P T E R 7 : C H A R A C T E R I Z AT I O N O F L A B E L L E D M I C R O A R R AY P R O B E S
❸
Calculation
■ The amounts of Cy3 and Cy5 incorporated into probes can be
calculated from their respective extinction coefficients and using
the following equations:
Extinction coefficients
150 000 M–1 cm –1 at 550 nm for Cy3 and
250 000 M–1 cm –1 at 650 nm for Cy5
Calculation equations
pmol Cy3 in purified sample =
(A550 / 150 000) × dilution factor × (z µl) × (w cm) × 1012
where
A550 = absorbance at 550 nm
z µl = the volume of sample after purification
w cm = optical path in cuvette
● 83
M I C R O A R R AY
7.2.1 UV spectrophotometry
Two methods can be used to determine the amount of labelled nucleic
acid in the sample: spectrophotometry and radioactive spiking. Nucleic
acids absorb at 260 nm, and absorption measurement at this wavelength
can be used for quantitative purposes. However, several purification
systems release other materials absorbing near or at this wavelength and,
depending on the amounts of these compounds, most of the absorbance
at 260 nm will be derived from something other than nucleic acid.
Therefore absorption spectra from 200 to 400 nm should be measured,
and only if the peak at 260 nm is clearly distinguishable from absorption
at near wavelengths, should estimation of nucleic acid amount be made
(Fig 50).
hb 63-0048-49 ● 84
C H A P T E R 7 : C H A R A C T E R I Z AT I O N O F L A B E L L E D M I C R O A R R AY P R O B E S
● 85
M I C R O A R R AY
❶
Preparation
■ Prepare a 20 × 20 cm glass-backed PEI cellulose chromatography
plate (available from Merck) for use by cutting a thin linear groove
into the PEI cellulose layer at 1 cm distance from the top edge of
the plate. Mark sample positions along a line that is 3 cm from the
bottom edge of the plate.
❷
Titration
■ Pipet 0.5 µl samples of labelling reactions in duplicate along the
marked line. The sample spots should be about 1 cm apart. Do
not damage the PEI cellulose layer with the tip.
❸
Separation
■ Place the plate in a rectangular chromatography tank so that the
bottom of the plate is immersed 2 cm deep in 1 M K2HPO4. Make
sure that the level of the buffer is below the level of the marked
sample line. Cover the tank and let sample separation take place.
Newly synthesized nucleic acid will not move far from the sample
line whereas free nucleotides will move progressively towards the
top of the plate.
■ When the buffer has reached the top groove, i.e. when the
samples have migrated the full available length of the 20 cm
plate, remove the plate from the tank and let it air dry.
❹
Imaging
■ Wrap the plate in cling-film and expose it to a phosphor screen for
1 – 6 h. Take care not to overexpose the phosphor screen as it will
saturate the signal. Some trial and error may be needed to determine
correct exposure time.
hb 63-0048-49 ● 86
C H A P T E R 7 : C H A R A C T E R I Z AT I O N O F L A B E L L E D M I C R O A R R AY P R O B E S
➎
Analysis
■ Calculate the proportion of the radioactive nucleotide that is
associated with the synthesized nucleic acid. This is the incorporation
percentage. For example, the ImageQuantTM software can be used for
this purpose.
Yield of cDNA =
2 × 10-9 mol × 20% × 4 × 330 g/mol = 528 ng
● 87
M I C R O A R R AY
Once the amount of CyDye incorporated into nucleic acid and the
amount of the nucleic acid are known, labelling density can be
calculated.
hb 63-0048-49 ● 88
C H A P T E R 7 : C H A R A C T E R I Z AT I O N O F L A B E L L E D M I C R O A R R AY P R O B E S
● 89
M I C R O A R R AY
❶
Preparation
■ PAGE gels suitable for analysis of CyDye labelled samples can be
prepared from standard 6% or 8% (w/v) sequencing gel mix such
as RapidGel-XL-6%. The gel can be cast using equipment designed
for sequencing, but since single base pair resolution is not required,
slab gel instruments can be used. These provide the added benefit of
thicker and deeper wells that simplify sample loading.
❷
Denature
■ Denature all samples by boiling for 2 min at 95 °C. Snap cool
on ice before loading on to the gel.
❸
Electrophoresis
■ Load samples to a gel that has been pre-electrophoresed for
15 – 30 min. Perform electrophoresis according to the instructions
provided with your equipment. Use 1× TBE as buffer. Protect the
samples from light during the electrophoresis.
■ Remove one of the gel plates before scanning and make sure that
the back of the remaining plate is clean. Do not let the gel dry
before scanning.
hb 63-0048-49 ● 90
C H A P T E R 7 : C H A R A C T E R I Z AT I O N O F L A B E L L E D M I C R O A R R AY P R O B E S
❹
Imaging
■ Scan the gel on a Typhoon Variable Mode Imager. Detect Cy3 by
excitation with 532 nm laser and using emission filter 555 BP 20.
Detect Cy5 by excitation with 633 nm laser and using emission filter
670 BP 30. Set PMT to 800 V, focal plane to +3 mm, and use normal
sensitivity. The PMT values may need to be adjusted to account for
different amounts of sample in the gel.
● 91
M I C R O A R R AY
hb 63-0048-49 ● 92
C H A P T E R 8 : M I C R O A R R AY H Y B R I D I Z AT I O N
Chapter 8
M I C R O A R R AY H Y B R I D I Z AT I O N
8.0 Introduction
The process of hybridization is typically performed in order to identify
and quantitate nucleic acids within a larger sample. Generally, it involves
annealing a single-stranded nucleic acid to a target complementary
strand. Southern blotting is one well-established hybridization method.
In this technique, samples of genomic DNA—the target—are attached
to a membrane and incubated within a solution of fluorescently labelled
DNA—the probe (54). Binding of probe molecules to the sample on the
membrane highlights complementary sequences, and the intensity of
signal is proportional to the amount of immobilized sample.
■ Pre-hybridization
■ Hybridization conditions
■ Hybridization buffer
■ Stringency washes
● 93
M I C R O A R R AY
❷
Hybridization
Hybridization buffer containing a known amount of labelled sample
DNA—often referred to as probe— is then placed on the slide surface.
A coverslip can then be carefully placed on top of the slide.
❸
Washing
Once the hybridization is complete, the slides are washed, and buffer and
probe of little or no homology to the target will be washed away, leaving
the labelled probe of high homology attached to the target and available
for detection.
hb 63-0048-49 ● 94
C H A P T E R 8 : M I C R O A R R AY H Y B R I D I Z AT I O N
8.2 Pre-hybridization
Pre-hybridization consists of incubating the spotted slide in a buffer
in the absence of probe. Different slide chemistries require slightly
different pre-hybridization protocols that vary in the type of buffer
used. Consult manufacturers’ recommendations to find out what is the
best procedure for each slide. Pre-hybridization prepares the microarray
for hybridization in the following two ways:
8.3 Hybridization
There are several widely used methods for carrying out the hybridization,
either using automated instruments or performing the procedure manually.
In this chapter we describe the manual hybridization method, which is
the most widely used. General properties of manual hybridization will be
discussed, followed by advice for choosing a suitable hybridization method
for different microarray slide types.
● 95
M I C R O A R R AY
humid chamber with slide trays and a reservoir, to a plastic box lined
with wet tissues at the base (Fig 53). In either case, the atmosphere
within the chamber must remain at >95% humidity throughout the
16-h hybridization. Equally important is not to allow the slide to come in
contact with the water, which may dilute the probe or cause water marks
on the slide surface.
a)
hb 63-0048-49 ● 96
C H A P T E R 8 : M I C R O A R R AY H Y B R I D I Z AT I O N
600000
Fig 55. When increased amounts of
Cy3
labelled probe are used, there is an 500000
Cy5
increase in background levels. This 400000
decreases the amount of background-
300000
corrected signal detected from the
microarray slide. 200000
100000
0
3 pM 8 pM 25 pM 75 pM
Probe concentration
● 97
M I C R O A R R AY
hb 63-0048-49 ● 98
C H A P T E R 8 : M I C R O A R R AY H Y B R I D I Z AT I O N
❶
Pre-hybridization
■ Store spotted slides in a desiccator until use.
■ Read through the pre-treatment and hybridization protocols thoroughly
before use, as buffers often need preparing and preheating before use.
■ Pretreat the number of slides required for the experiment. Some
manufacturers say that pre-treated slides can be stored before use,
but not all, so it is worth checking before doing a large batch.
■ During the pre-treatment stage, prepare the probes. This will often
involve drying down equivalent amounts of the two probes of interest
together and reconstituting them in the manufacturer’s recommended
buffer. Some protocols require heating the probe before use; the probe
prepared by reverse transcription will be single-stranded and therefore
should not require denaturing before use.
● 99
M I C R O A R R AY
❷
Hybridization
■ Once the probe is prepared, lay the spotted slide, DNA side up, on a
clean surface. Absorbent tissue that does not release any fibers is a
good choice of surface. This is important, as most slides are glass, and
dirt on the rear of the slide will affect the result on the front of the
slide (this is obviously not an issue with opaque or mirrored slides).
■ Using a pipette, transfer the required amount of hybridization
buffer/probe mixture onto the slide. Avoid touching the slide surface
with the pipette tip. Try to deposit the mixture along the short side of
the slide, away from the spotted area.
■ Take a clean coverslip and place it on the slide near the probe
mixture, and allow surface tension to speed the buffer along the
coverslip. Then gently lower the coverslip, avoiding trapping air
bubbles underneath. There are several ways of lowering the coverslip,
two of which are illustrated in Figure 56.
Fig 56. Lowering the coverslip.
■ If air bubbles become trapped beneath a coverslip (Fig 57), do not
move the coverslip to try and remove them. Movement of the
coverslip will result in damage to the targets themselves. Most small
air bubbles will disperse once the slide is transferred to hybridization
temperature. Larger bubbles can be ‘encouraged’ to move by gently
pressing on the surface of the coverslip with a pipette tip.
■ The probe mixture is light sensitive, so once the coverslip is on, place
the slide in the humid chamber and incubate overnight in the dark.
hb 63-0048-49 ● 100
C H A P T E R 8 : M I C R O A R R AY H Y B R I D I Z AT I O N
The stringency washes will affect the amount of labelled probe retained
on the slide for the final analysis. While it is obviously important to
remove all the loosely bound probe, it is important to not strip the
bound probe. Generally, stringency washes are carried out in a SSC/SDS
solutions of different concentrations, with the primary washes often
being carried out at the same temperature as the hybridization. Primary
wash solutions have a high salt content (typically 1–2× SSC/0.1% SDS
buffer), and they remove most of the hybridization buffer components.
The secondary washes are performed with low salt buffer (typically
0.1× SSC/0.1% SDS), and they will remove the loosely bound probe from
the blot. It also serves to remove any remaining salt from the primary
washes. Failing to warm solutions thoroughly before use will lower their
effectiveness and may lead to increases in background noise. Conversely,
warming solutions too much (as often happens if a microwave oven is
used) or using too low a salt concentration in the buffers, will strip
precious signal from the blot. Check the manufacturer’s protocols for
exact wash dilution volumes. Manufacturers often suggest a water dip
prior to drying the slides to prevent smearing. Check the protocols
provided with the buffer components for instructions on performing the
water dip.
● 101
M I C R O A R R AY
hb 63-0048-49 ● 102
CHAPTER 9 : LUCIDEA SLIDEPRO HYBRIDIZER
Chapter 9
LUCIDEA SLIDEPRO HYBRIDIZER
9.0 Introduction
A complex system, microarray analysis is affected by a number of
experimental factors, including target preparation, physical deposition of
the targets, slide chemistry, probe chemistry, hybridization, and detection
of the fluorescent signal (37). In order to improve the efficiency of
microarray analysis, each source of variation must be eliminated or
minimized. Control of environmental conditions during hybridization in
particular, is critical in producing and maintaining a consistent fluorescent
signal. Lucidea SlidePro was designed to overcome the problems associated
with hybridization variability.
Fig 58. Lucidea SlidePro Hybridizer. 9.1 Features of Lucidea SlidePro Hybridizer
Lucidea SlidePro Hybridizer (Fig 58) has a modular format that consists
of a base control unit, up to four additional modules, and control
software on a laptop computer. Each unit contains six individually
temperature-controlled chambers, each of which holds a standard
microscope slide. With all five modules a total of 30 slides can be
processed. The multi-module protocol software allows each module to
be started at different times and different experiments can be conducted
on each module, thereby increasing user flexibility. Also, each module
has its own pump, which allows a faster processing time.
Heating block
Slide Lucidea SlidePro is capable of automating a variety of chemical
O-ring seal O-ring seal
PEEK chamber Port and biochemical techniques in which incubation and wash steps are
(100 µm recess)
performed at varying temperatures. It is primarily used in microarray
Outlet port analysis to automate the pre-hybridization, hybridization, and washing
of microarray slides. It has been designed to be used in conjunction with
the Lucidea range of microarray instruments and reagents.
Hinged clamp
In Lucidea SlidePro, each slide is held in a chamber sealed with a patented
O-ring. Pre-treatment and wash solutions are drawn into the chamber,
from up to five reservoirs, and deposited into a waste bottle. Each
Inlet port module can run off one set of wash botttles, or multiple modules can run
off the same set of wash bottles. Hybridization samples are injected
Septum
injection port through a septum port at the lower end of the chamber (Fig 59).
Side view Bottom view
of chamber
● 103
M I C R O A R R AY
■ Reduced demands on user time. The user needs only to load slides and
inject the probe. Following automated hybridization, washing, and
drying, the slides are removed from the instrument ready to scan.
hb 63-0048-49 ● 104
CHAPTER 9 : LUCIDEA SLIDEPRO HYBRIDIZER
● 105
M I C R O A R R AY
RFU
800000
Cy3 manual
Cy3 Lucidea SlidePro
600000
Cy5 manual
Cy5 Lucidea SlidePro
400000
200000
0
1 2 3 4 5 6 7 8 9 10 11 12
Slide
0.6
0.4
0.2
0
1 2 3 4 5 6 7 8 9 10 11 12
Slide
hb 63-0048-49 ● 106
CHAPTER 9 : LUCIDEA SLIDEPRO HYBRIDIZER
6.E+06
5.E+06
1.E+06
0.E+00
2 50 100 150 200
● 107
M I C R O A R R AY
Total Cy3 signal intensity with Total Cy5 signal intensity with
hybridization time hybridization time
500000 500000
400000 500000
0 0
3h 6h 15 h 3h 6h 15 h
hb 63-0048-49 ● 108
CHAPTER 9 : LUCIDEA SLIDEPRO HYBRIDIZER
-1
-2
3h 6h 15 h
Hybridization
Menu RFU
10000000
1000000
Cy3 mean
Cy5 mean
100000
10000
1000
3 6 9 12 15 3 6 9 12 15 3 6 9 12 15 15
Control
DR2 DR3 DR5
(20 pmol)
Hybridization time (hr)
Fig 68. Effect of hybridization time. Total mean Cy3 and Cy5
signals of dynamic range controls.
● 109
M I C R O A R R AY
Lucidea SlidePro has been optimized for use with Amersham Biosciences
version 2 hybridization buffer and microarray slides. Lucidea SlidePro
can also be used to hybridize other manufacturer’s slides (Fig 69).
These protocols can be found on the Amersham Biosciences web site
(www.amershambiosciences.com).
a) d)
Fig 69. Lucidea SlidePro hybridization
with various commercially available
slides.
b) e)
c)
T28 direct labeled
Cy3 heart / Cy5 muscle
hb 63-0048-49 ● 110
C H A P T E R 1 0 : F L U O R E S C E N C E I M A G I N G S Y S T E M S I N M I C R O A R R AY A N A LY S I S
Chapter 10
F L U O R E S C E N C E I M A G I N G S Y S T E M S I N M I C R O A R R AY
A N A LY S I S
10.0 Introduction
All fluorescence imaging systems require the following key elements
(Fig 70):
■ Excitation source
Sample
Emission filter
Excitation source
Detection and
amplification
● 111
M I C R O A R R AY
Light energy is essential to fluorescence. Light sources fall into two broad
categories—wide-area, broad-wavelength sources, such as UV and xenon
arc lamps, and line sources with discrete wavelengths, such as lasers
(Fig 71). Broad-wavelength excitation sources are used in fluorescence
spectrometers and camera imaging systems. Although the spectral output
300 500 700 900 1100 of a lamp is broad, it can be tuned to a narrow band of excitation light
Wavelength (nm) with the use of gratings or filters. In contrast, lasers deliver a narrow
beam of collimated light that is predominantly monochromatic. In most
camera systems, excitation light is delivered to the sample by direct
Fig 71. Spectral output of light from a xenon
illumination of the imaging field, with the excitation source positioned
lamp and Nd:YAG laser. The “relative output”
axis is scaled arbitrarily for the two light sources. either above, below, or to the side of the sample.
The 532-nm line of the Nd:YAG laser is shown
Laser-based imaging systems, on the other hand, use more sophisticated
in green.
optical paths, comprising mirrors and lenses, to direct the excitation
beam to the sample. Some filtering of the laser light may also be required
before the excitation beam is directed to the sample. For microarray
applications, laser-based instruments are substantially favored, therefore
CCD scanners will not be discussed.
hb 63-0048-49 ● 112
C H A P T E R 1 0 : F L U O R E S C E N C E I M A G I N G S Y S T E M S I N M I C R O A R R AY A N A LY S I S
Digital resolution defines the ability to resolve two signals with similar
intensities. Since only a limited number of intensity levels are available, it
is unavoidable that this conversion process introduces a certain amount
of error. To allow ample discrimination between similar signals and to
keep the error as low as possible, the distribution of the available
intensity levels should correspond well to the linear dynamic range of a
detector. There are two methods of distributing intensity levels. A linear
(even) distribution has the same spacing for all the intensity levels,
allowing measurement across the dynamic range with the same absolute
accuracy. However, relative digitization error increases as signals become
smaller. A non-linear distribution (e.g. logarithmic or square root
functions) divides the lower end of the signal range into more levels
while combining the high end signals into fewer intensity levels. Thus,
the absolute accuracy decreases with higher signals, but the relative
digitization error remains more constant across the dynamic range.
● 113
M I C R O A R R AY
Lasers
There are several commonly used types of lasers.
hb 63-0048-49 ● 114
C H A P T E R 1 0 : F L U O R E S C E N C E I M A G I N G S Y S T E M S I N M I C R O A R R AY A N A LY S I S
Sample
● 115
M I C R O A R R AY
excitation beam does not illuminate the sample from the same angle in
every position, a parallax effect can result. The term parallax here refers
to the shift in apparent position of targets, predominately at the outer
boundaries of the scan area. Additionally, the arc of excitation light
created by the galvanometer mirror produces some variations in the
effective excitation energy reaching the sample at different points across
the arc. These effects can be minimized with an f-theta lens, but when the
angle of incident excitation light varies over the imaging field, some
spatial distortion can still occur in the resulting image.
Sample
Glass platen
Mirror
Laser
hb 63-0048-49 ● 116
C H A P T E R 1 0 : F L U O R E S C E N C E I M A G I N G S Y S T E M S I N M I C R O A R R AY A N A LY S I S
Sample
Glass platen
Detector
● 117
M I C R O A R R AY
Emission filter
hb 63-0048-49 ● 118
C H A P T E R 1 0 : F L U O R E S C E N C E I M A G I N G S Y S T E M S I N M I C R O A R R AY A N A LY S I S
After the fluorescent emission has been filtered and only the desired
wavelengths remain, the light is detected and quantified. Because the
100 intensity of light at this stage is very small, a PMT must be used to detect
it. In the PMT, photons of light hit a photocathode and are converted
into electrons which are then accelerated in a voltage gradient and
multiplied from 106 to 107 times. This produces a measurable electrical
Cathode radiant sensitivity (mA/W)
10
signal that is proportional to the number of photons detected. The
response of a PMT is typically useful over a wavelength range of
300–800 nm (Fig 76). High-performance PMTs extend this range to
1
200–900 nm.
0.1
10.6 System performance
The performance of a laser scanner system is described in terms of system
resolution, linearity, uniformity, and sensitivity.
0.01 Resolution can be defined in terms of both spatial and amplitude
100 200 300 400 500 600 700 800 900 1000
Wavelength (nm)
resolution. Spatial resolution of an instrument refers to its ability to
distinguish between two very closely positioned objects. It is a function
Fig 76. An example of the response of a of the diameter of the light beam when it reaches the sample and the
PMT versus wavelength. distance between adjacent measurements. Spatial resolution is dependent
on, but not equivalent to, the pixel size of the image. Spatial resolution
improves as pixel size is reduced. Systems with higher spatial resolution
can not only detect smaller objects, but can also discriminate more
accurately between closely spaced targets. However, an image with a
100-µm pixel size will not have a spatial resolution of 100 µm. The
pixel size refers to the collection sampling interval of the image.
According to a fundamental sampling principle, the Nyquist Criterion,
the smallest resolvable object in an image is no better than twice the
sampling interval (59). Thus, to resolve a 100-µm sample, the sampling
interval must be at most 50 µm. Amplitude resolution, or gray-level
quantification, describes the minimum difference that is distinguishable
between levels of light intensity (or fluorescence) detected from the
sample (60). For example, an imaging system with 16-bit digitization can
resolve and accurately quantify 65 536 different values of light intensity
from a fluorescent sample.
● 119
M I C R O A R R AY
Linearity of a laser scanner is the signal range over which the instrument
yields a linear response to fluorochrome concentration and is therefore
useful for accurate quantification. The linear dynamic range can be
defined in at least 3 ways:
A scanning system with a wide dynamic range can detect and accurately
quantify signals from both very low- and very high-intensity targets in
the same scan. The linear dynamic range of most laser scanner
instruments is between 104 and 105.
Uniformity across the entire scan area is critical for reliable quantitation.
A given fluorescent signal should yield the same measurement at any
position within the imaging field. Moving-head scanners, in particular,
deliver flat-field illumination and uniform collection of fluorescent
emissions across the entire scan area.
hb 63-0048-49 ● 120
C H A P T E R 1 0 : F L U O R E S C E N C E I M A G I N G S Y S T E M S I N M I C R O A R R AY A N A LY S I S
Long-pass (LP) filters pass light that is longer than a specified wavelength
and reject all shorter wavelengths. A good quality long-pass filter is
100 characterized by a steep transition between rejected and transmitted
Transmission (%)
80 wavelengths (Fig 77a). Long-pass filters are named for the wavelength
60 at the midpoint of the transition between the rejected and transmitted
cutoff point
40 light (cutoff point). For example, the cutoff point in the transmission
20 spectrum of a 560 LP filter is 560 nm, where 50% of the maximum
0 transmittance is rejected. The name of a long-pass filter may also
550 560 570 580 590 600
a) Wavelength (nm)
include other designations, such as OG (orange glass), RG (red glass),
E (emission), LP (long-pass), or EFLP (edge filter long-pass). OG and RG
are colored glass absorption filters, whereas E, LP, and EFLP filters are
100 coated interference filters. Colored glass filters are less expensive and
Transmission (%)
40
cutoff point Short-pass (SP) filters reject wavelengths that are longer than a specified
20
value and pass shorter wavelengths. Like long-pass filters, short-pass
0
filters are named according to their cutoff point. For example, a 526 SP
500 510 520 530 540 550
filter rejects 50% of the maximum transmittance at 526 nm (Fig 77b).
b) Wavelength (nm)
Band-pass (BP) filters allow a band of selected wavelengths to pass
Fig 77. Transmission profiles for a (a) 560-nm
through, while rejecting all shorter and longer wavelengths. Band-pass
long-pass and a (b) 526-nm short-pass filter. filters provides very sharp cutoffs with very little transmission of the
The cutoff points are noted. rejected wavelengths. High-performance band-pass filters are also
referred to as Discriminating Filters (DF). The name of a band-pass filter
is typically made up of two parts:
60
FWHM
[655–685 nm] with an efficiency equal to or greater than half
40 the maximum transmittance of the filter).
20
Band-pass filters with an FWHM of 20–30 nm are optimal for most
0
650 660 670 680 690 700
fluorescence applications, including multi-label experiments. Filters
Wavelength (nm) with FWHMs greater than 30 nm allow collection of light at more
wavelengths and give a higher total signal; however, they are less able
Fig 78. Transmission profile for a band-pass to discriminate between closely spaced, overlapping emission spectra
(670 BP 30) filter. The full-width at half- in multichannel experiments. Filters with FWHMs narrower than
maximum (FWHM) transmission of 30 nm 20 nm transmit less signal and are most useful with fluorochromes
is indicated by the arrows.
with very narrow emission spectra.
● 121
M I C R O A R R AY
hb 63-0048-49 ● 122
C H A P T E R 1 0 : F L U O R E S C E N C E I M A G I N G S Y S T E M S I N M I C R O A R R AY A N A LY S I S
•
maximum correlates closely with the excitation wavelength of the imaging
system. However, the absorption profiles of most fluorochromes are
• rather broad, and some fluorochromes have a second (or additional)
300 350 400 450 500 550 600 absorption peak or a long “tail” in their spectra. It is not mandatory that
Wavelength (nm) the fluorochrome’s major absorption peak matches exactly the available
excitation wavelength for efficient excitation. For example, the absorption
Fig 79. Excitation of fluorescein (green) and maxima of the fluorescein and Cy3 fluorochromes are 490 nm and 552
Cy3 (orange) using 532-nm laser light. The
nm respectively (Fig 79). Excitation of either dye using the 532-nm
absorption spectra of Cy3 and fluorescein are
wavelength line of the Nd:YAG laser may seem to be inefficient, since
overlaid with the 532-nm wavelength line of
the Nd:YAG laser. the laser produces light that is 40 nm above the absorption peak of
fluorescein and 20 nm below that of Cy3. In practice, however, delivery
of a high level of excitation energy at 532 nm does efficiently excite both
fluorochromes. For emission, selecting a filter that transmits a band at or
near the emission peak of the fluorochrome generally improves the
sensitivity and linear range of the measurement. Figure 80 shows
560 LP
collection of Cy3 fluorescence using either a 580 BP 30 or a 560 LP
emission filter.
Emission
580 BP 30
10.9.2 Multicolor imaging
Multicolor imaging allows detection and resolution of multiple targets
500 550 600 650 700
using fluorescent labels with different spectral properties. The ability to
Wavelength (nm)
multiplex or detect multiple labels in the same experiment is both time
and cost-effective and improves accuracy for some assays. Analysis using
Fig 80. Filtering of Cy3 fluorescence using
a single label can require a set of experiments or many repetitions of the
either a 580 BP 30 (dark gray area) or a
560 LP filter (light and dark gray areas). same experiment to generate one set of data. For example, single-label
analysis of gene expression from two different tissues requires two
separate hybridization to different gene arrays, or consecutive
hybridization to the same array with stripping and re-probing. With a
dual-label approach, however, the DNA probes from the two tissue types
are labelled with different fluorochromes and used simultaneously with
the same gene array. In this way, experimental error is reduced because
only one array is used, and hybridization conditions for the two probes
are identical. Additionally, by using a two-channel scan, expression data
is rapidly collected from both tissues, thus streamlining analysis.
● 123
M I C R O A R R AY
hb 63-0048-49 ● 124
C H A P T E R 1 0 : F L U O R E S C E N C E I M A G I N G S Y S T E M S I N M I C R O A R R AY A N A LY S I S
Scanning area: 35 × 43 cm
Maximum resolution: 10 µm
Scanning area: 35 × 43 cm
Maximum resolution: 10 µm
● 125
M I C R O A R R AY
hb 63-0048-49 ● 126
C H A P T E R 1 1 : D E S I G N , C O N T R O L S A N D D ATA A N A LY S I S O F M I C R O A R R AY E X P E R I M E N T S
Chapter 11
D E S I G N , C O N T R O L S A N D D ATA A N A LY S I S O F
M I C R O A R R AY E X P E R I M E N T S
11.0 Introduction
Methods for the analysis of microarray data are still evolving, and
there is no standard experimental design or method of data analysis for
microarray experiments at the present time. However, some efforts are
being made to set a common annotation and standards for microarray
data in order to create public databases for microarray results (61, 62,
63). Meanwhile, in this chapter some important considerations for
analyzing microarray data are discussed.
● 127
M I C R O A R R AY
Replicate slides
Image analysis
Visualize data
Cluster data
9 4 7
3 Store data in database
6 1
2 2
hb 63-0048-49 ● 128
C H A P T E R 1 1 : D E S I G N , C O N T R O L S A N D D ATA A N A LY S I S O F M I C R O A R R AY E X P E R I M E N T S
■ Image analysis
■ Examination of controls
■ Data normalization
● 129
M I C R O A R R AY
■ Manual: This method involves first dividing the grid into several
subgrids, and then visually aligning each smaller subgrid with the
corresponding area of the image by adjusting the position of circles.
hb 63-0048-49 ● 130
C H A P T E R 1 1 : D E S I G N , C O N T R O L S A N D D ATA A N A LY S I S O F M I C R O A R R AY E X P E R I M E N T S
The next step in image analysis is to determine the signal present inside
and outside the spot. The background signal is then subtracted from the
spot signal to give background-corrected signal. Whereas spot signal is
calculated from within the positioned grid, the background signal is
determined by calculating the average pixel intensity in a user-defined
region. Although mean or median background signal can be used,
median values are more resistant to variation in background caused, for
example, by fluorescent speckles. Some of the various regions in which
the background can be calculated are illustrated in Figure 83. It is
important to use a background correction method such that any pixels
from the spot do not get included in the background. This can easily
occur if the spots are close together.
Around spot groups Corner between spots User defined areas Around individual spots User defined spots
Fig 83. Some of the background region options available to the user
in the ArrayVision™ Image Analysis Software. The green represents the
spots enclosed by the grid while the blue encloses the background
region. Image analysis quality metrics calculated by analysis software
are increasingly used to highlight data that may be unreliable and
should be omitted from further analysis. Typical causes of suspect
data include dust speckles over spots, poor spot morphology, very low
or very high signal.
● 131
M I C R O A R R AY
■ background subtraction
11.4.2 ArrayVision
ArrayVision software is a semi-automated software used for performing
image analysis (Fig 85). Some of its features include:
■ quality metrics
■ reporting tools and data export
hb 63-0048-49 ● 132
C H A P T E R 1 1 : D E S I G N , C O N T R O L S A N D D ATA A N A LY S I S O F M I C R O A R R AY E X P E R I M E N T S
● 133
M I C R O A R R AY
1 2 3 4 1
hb 63-0048-49 ● 134
C H A P T E R 1 1 : D E S I G N , C O N T R O L S A N D D ATA A N A LY S I S O F M I C R O A R R AY E X P E R I M E N T S
● 135
M I C R O A R R AY
1 700 2
1.5
0.5 600
1
0 500
0.5
-0.5 400
0
-1 300
-0.5
-1.5 200
-1
-2 100 -1.5
-2.5 0 -2
a) 2 3 4 5 6 7 8 b) 2 3 4 5 6 7 8
hb 63-0048-49 ● 136
C H A P T E R 1 1 : D E S I G N , C O N T R O L S A N D D ATA A N A LY S I S O F M I C R O A R R AY E X P E R I M E N T S
11.7 Normalization
In order to compare ratio data from one microarray slide to another
microarray slide, the ratio data needs to be normalized to correct for
experimental variation. The reason for this is that from one slide to
another there will be differences between the relative Cy3 and Cy5
signals due to one or more of the following:
■ the amounts of mRNA used in the Cy3 and Cy5 labelling reaction
■ use signal ratios from spikes which have been added to the labelling
reaction in equal amounts
■ use total signal, which is the summation of all the Cy3 signals and all
the Cy5 signals
The three methods discussed above assume that the normalization factor
is constant over the whole signal range, which in most cases is not.
● 137
M I C R O A R R AY
If the data from the slide meets the criteria set by the researcher then the
next step is to look at the variation between replicates. It is recommended
that each experiment be repeated several times, and there are statistical
hb 63-0048-49 ● 138
C H A P T E R 1 1 : D E S I G N , C O N T R O L S A N D D ATA A N A LY S I S O F M I C R O A R R AY E X P E R I M E N T S
● 139
M I C R O A R R AY
hb 63-0048-49 ● 140
C H A P T E R 1 1 : D E S I G N , C O N T R O L S A N D D ATA A N A LY S I S O F M I C R O A R R AY E X P E R I M E N T S
● 141
M I C R O A R R AY
Scierra MA-LWS has been designed to mirror the workflow of the typical
microarray lab (Fig 89). It allows users to perform many tasks. First,
users can organize large numbers of samples and experiments based on
projects. Users can also manage and track a large variety of samples and
reagents, including:
■ arrayed slides
Scierra MA-LWS allows users to manage and track every activity in the
microarray workflow, including:
hb 63-0048-49 ● 142
C H A P T E R 1 1 : D E S I G N , C O N T R O L S A N D D ATA A N A LY S I S O F M I C R O A R R AY E X P E R I M E N T S
Sample
Definition
Experimental
Design
Biological Samples Array
Sample
Submission
Sample Spotting plate
Receiving Receiving
mRNA
Preparation Quantification
Fig 89. Scierra MA-LWS mirrors
Sample Normalize and
the microarray workflow. Labeling Aliquot
Sample
Quantification Spotting
Sample Treatment
Fragmentation Post-spotting
Hybridization Pre-arrayed
slides
Array
Scanning
Spotfinding
Normalization
● 143
M I C R O A R R AY
hb 63-0048-49 ● 144
C H A P T E R 1 2 : T R O U B L E S H O O T I N G M I C R O A R R AY E X P E R I M E N T S
Chapter 12
T R O U B L E S H O O T I N G M I C R O A R R AY E X P E R I M E N T S
12.0 Introduction
Microarray analysis is a complicated multistep process consisting of
discrete steps as shown in Figure 90. Each of these steps is critical in
determining whether the experiment is successful, and problems
encountered at any stage will be detrimental to the quality of data
obtained. Troubleshooting microarray experiments needs to consider
all these steps.
Glass slide
Array scanner
Scanned slide
9 4 7
Biological data
3 1
generated by microarrays 6
2 2
● 145
M I C R O A R R AY
Labelling
density
Number of target
molecules in sample
b) Cy5 volume
hb 63-0048-49 ● 146
C H A P T E R 1 2 : T R O U B L E S H O O T I N G M I C R O A R R AY E X P E R I M E N T S
system optimization is to find the best overall protocol for the system
and to determine what is the standard performance of the system.
Yellow experiments only require one type of RNA sample, but provide
information on all aspects of the microarray process. In contrast,
experiments in which two different RNA species are used have the added
complication that differential gene expression will be present in unknown
quantities. In reverse-color experiments, in which both samples are
labelled separately with two colors and hybridizations are performed
with both possible combinations, any inherent variation between the
quality of the two arrays can complicate the optimization process and
result in wrong conclusions. For example, high and uneven Cy3
background on one of the slides in a reverse-color experiment can give
the appearance of unbalanced labelling with one fluorescent dye. For
‘real’ experiments in which information about gene expression is being
sought, reverse color experiments are useful.
● 147
M I C R O A R R AY
hb 63-0048-49 ● 148
C H A P T E R 1 2 : T R O U B L E S H O O T I N G M I C R O A R R AY E X P E R I M E N T S
● 149
M I C R O A R R AY
hb 63-0048-49 ● 150
C H A P T E R 1 2 : T R O U B L E S H O O T I N G M I C R O A R R AY E X P E R I M E N T S
● 151
M I C R O A R R AY
hb 63-0048-49 ● 152
C H A P T E R 1 2 : T R O U B L E S H O O T I N G M I C R O A R R AY E X P E R I M E N T S
● 153
M I C R O A R R AY
a) b) c)
d) e) f)
g) h) i)
hb 63-0048-49 ● 154
C H A P T E R 1 2 : T R O U B L E S H O O T I N G M I C R O A R R AY E X P E R I M E N T S
j) k) l)
m) n) o)
p)
● 155
M I C R O A R R AY
hb 63-0048-49 ● 156
REFERENCES
References
References cited in text
1. Baldwin, D. et al. A comparison of gel-based, nylon filter and
microarray techniques to detect differential RNA expression in
plants. Curr Opinions in Biol. 2, 96–103 (1999).
14. Hu, G. K. et al. Predicting splice variant from DNA chip expression
data. Genome Research 11, 1237–1245 (2001).
● 157
M I C R O A R R AY
23. Li, F. and Stormo, G. D. Selection of optimal DNA oligos for gene
expression arrays. Bioinformatics 17, 1067–1076 (2001).
hb 63-0048-49 ● 158
REFERENCES
31. Knight, J. When the chips are down. Nature 410, 860–861 (2001).
41. Yu, H. et al. Cyanine dye dUTP analogs for enzymatic labeling of
DNA probes. Nucleic Acids Res. 22, 3226–32 (1994).
● 159
M I C R O A R R AY
hb 63-0048-49 ● 160
REFERENCES
61. Brazma A. et al. One-stop shop for microarray data. Nature. 403,
699–700 (2000).
62. Brazma, A. and Vilo, J. Gene expression data analysis. FEBS Letters
480, 17–24 (2000).
● 161
M I C R O A R R AY
hb 63-0048-49 ● 162
O R D E R I N G I N F O R M AT I O N
Bioinformatics products
Scierra Microarray Laboratory Workflow System
Electrophoresis systems
Ready-to-Run Separations Unit 1 80-6460-95
Spectrophotometry products
Ultraspec 3300 pro UV/Visible Spectrophotometer 1 80-2112-33
Other
Hybond-N+ Membrane 50 RPN82B
Vistra Green Nucleic Acid Stain 500 ml RPN5786
RapidGel-XL - 6% 100 ml US75861
RapidGel-XL - 8% 100 ml US75862
ALFexpress Sizer 50-500 50 27-4539-01
TBE Buffer Pre-mixed Powder 10 6 bottles US70454
● 163
M I C R O A R R AY
Index
bioinformatics 141
E
hb 63-0048-49 ● 164
INDEX
● 165
M I C R O A R R AY
S
Y
saturation 38, 98, 122, 136
yellow experiment 55, 147
scanners 39, 81, 89, 111-116,
119-124
hb 63-0048-49 ● 166