Operator’s Manual
Version 1.0
URISYS 2400
Revision History
English 03184315.018
Intended use The URISYS 2400 is a fully automated urinalysis system intended for in vitro
qualitative or semi-quantitative determination of urine analytes, including
pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, and
erythrocytes.
, URISYS 2400 are registered trademarks of the Roche group. All other
trademarks that are mentioned in this manual are the property of the
respective trademark holders: KOVA® and Monovette®.
Instrument approvals The URISYS 2400 analyzer meets the requirements stated in
Directive 98⁄79⁄EC of the European Parliament and the Council of the
European Union (EU) on in vitro diagnostic medical devices. Furthermore,
the URISYS 2400 analyzer is manufactured and tested according to
International Standard IEC 1010-1, "Safety requirements for electrical
equipment for measurement, control, and laboratory use, Part 1: General
requirements". This International Standard is equivalent to the standards UL
3101-1 for the USA, CSA C 22.2 No. 1010.1 for Canada, and DIN EN 61010-1
for Germany.
The URISYS 2400 fulfils the EMC immunity requirements for laboratory use
equipment in industrial areas, according to the EMC standards EN 50082-2
and EN 61326-A1. According to EN 55011, the URISYS 2400 analyzer is a
device of limit class A.
Roche Diagnostics
ii Operator’s Manual · Version 1.0
URISYS 2400
Contact addresses
Manufacturer:
Authorized Representative
Warranty
Refer to the URISYS 2400 analyzer purchase agreement for warranty
conditions. Contact your local Roche Technical Support for further
information.
Roche Diagnostics
Operator’s Manual · Version 1.0 iii
URISYS 2400
Roche Diagnostics
iv Operator’s Manual · Version 1.0
URISYS 2400
Table of contents
Roche Diagnostics
Operator’s Manual · Version 1.0 v
URISYS 2400
Roche Diagnostics
vi Operator’s Manual · Version 1.0
URISYS 2400
Preface
Roche Diagnostics
Operator’s Manual · Version 1.0 vii
URISYS 2400
a Procedural step
o List item
e Cross-reference
Abbreviation Definition
BIL Bilirubin
CLA Clarity
COL Color
EN European Standard
ERY Erythrocytes
GLU Glucose
Roche Diagnostics
viii Operator’s Manual · Version 1.0
URISYS 2400
Abbreviation Definition
ID Identification number
KET Ketones
LEU Leukocytes
NEG Negative
NIT Nitrite
nm Nanometer
NORM Normal
ORANG Orange
POS Positive
PRO Protein
SG Specific Gravity
SI Standard International
TS Test strip
TURBD Turbid
UBG Urobilinogen
YELLO Yellow
Roche Diagnostics
Operator’s Manual · Version 1.0 ix
URISYS 2400
Safety classifications
Before operating with the URISYS 2400 it is essential that the warnings,
cautions, and safety requirements contained in this manual are read and
understood by the user. This section explains how precautionary information
is formatted in the manual.
The safety precautions and important user notes are classified according to
ANSI Z535 standards. Familiarize yourself with the following meanings and
icons:
Warning
Indicates a possibly hazardous situation which, if not avoided, may result in death or
serious injury.
Examples of a “serious injury” include loss of eyesight, burn (high temperature, low
temperature), electric shock, bone fracture, or poisoning. These injuries require
medical assistance.
Biohazard
Samples containing material of human origin must be treated as potentially
infectious. The relevant laboratory guidelines on safe use must be observed.
Caution
Indicates a possibly hazardous situation which, if not avoided, may result in slight or
minor injuries, and/or damage to equipment.
“Minor injury” refers to injuries that may require medical assistance.
“Equipment” refers to extended damage to buildings, furniture, and so on.
Spillage
Avoid all liquid spills on the analyzer.
Roche Diagnostics
x Operator’s Manual · Version 1.0
URISYS 2400
Safety information
Electrical safety
Connect the analyzer to grounded power outlets only (protection class 1). All
peripheral devices that are connected to the URISYS 2400 must comply with safety
standard IEC 950 (UL 1950) for information technology equipment, or with IEC
1010 (UL 3101) for laboratory use instruments.
Instrument in use
Keep hands or other objects away from the sampling probe area to prevent personal
injury while the instrument is in use. Do not touch parts of the analyzer other than
those specified.
User qualification
Only appropriately trained operators are qualified to operate the analyzer.
Correct use
Any disregard of the instructions in the Operator’s Manual may result in a safety risk.
Use the URISYS 2400 analyzer to analyze urine samples only. It is not intended for
any other application.
Transport
At least two persons must carry the analyzer, by holding the baseplates on the left
side and right side of the analyzer. Be careful not to hurt your hands or fingers when
putting the analyzer in place.
Environmental conditions
The URISYS 2400 is approved for indoor use only.
Biological safety
Liquid waste and strip waste are potentially biologically hazardous. Always wear
gloves if handling those materials. Do not touch parts of the analyzer other than
those specified. Consult your laboratory protocol for handling biohazardous
materials.
Roche Diagnostics
Operator’s Manual · Version 1.0 xi
URISYS 2400
Radio interference
URISYS 2400 is a class A device. In residential areas it may cause radio
interference. The user must take precautions as required.
Installation
Follow the specified installation conditions carefully. Otherwise, inaccurate results,
or damage to the analyzer may occur.
Waste
Handle liquid waste and used test strips properly, according to legislation on water
pollution, and on the treatment of drainage and waste matter.
Electromagnetic waves
Devices that emit electromagnetic waves may affect measured data, or cause the
analyzer to malfunction. Do not operate the following devices in the same room
where the analyzer is installed: mobile phone, transceiver, cordless phone, other
electrical devices that generate electromagnetic waves.
Roche Diagnostics
xii Operator’s Manual · Version 1.0
URISYS 2400
C
B
A
E
Roche Diagnostics
Operator’s Manual · Version 1.0 xiii
URISYS 2400
Roche Diagnostics
xiv Operator’s Manual · Version 1.0
Overview A
Roche Diagnostics
Operator’s Manual · Version 1.0 A-1
1 Introduction to the system URISYS 2400
Description of the URISYS 2400 analyzer
Measuring principle
The URISYS 2400 is a fully automated reflectance photometer for in vitro
semi-quantitative measurements of urine test strips. The light sources (LED’s
[light-emitting diodes]) and reading times are optimized for the reaction
chemistry and color development that occur on the test pads.
The measuring head of the URISYS 2400 contains LED’s of three different
wavelengths. The test strip remains stationary at the measuring position as the
measuring head moves over each test pad. Measurement starts at the reference
plate, which is used to test the optical system.
The URISYS 2400 verifies that the test strip is properly positioned under the
measuring head, by measuring the reflected light. If a test strip is not properly
positioned under the measuring head, the URISYS 2400 displays a message.
Roche Diagnostics
A-2 Operator’s Manual · Version 1.0
URISYS 2400 1 Introduction to the system
Description of the URISYS 2400 analyzer
C D E
F
B
Optical transmission Electronic transmission
Figure A-1 Process of measuring
A LED D Analog-to-digital converter
B Test pad E Microprocessor
C Photodiode detector F Concentration result
The results are stored in the memory and can be printed, saved to a diskette,
or sent to another computer.
Each of the three LED’s transmits light of a different wavelength. The use of
different wavelengths improves the results that are obtained from measuring
each of the test strip parameters. The wavelengths used to measure the
different parameters are listed in the table below.
Roche Diagnostics
Operator’s Manual · Version 1.0 A-3
1 Introduction to the system URISYS 2400
Description of the URISYS 2400 analyzer
pH 555, 620
Leukocytes 555
Nitrite 555
Protein 555
Glucose 555
Ketones 555
Urobilinogen 555
Bilirubin 555
Roche Diagnostics
A-4 Operator’s Manual · Version 1.0
URISYS 2400 1 Introduction to the system
Description of the URISYS 2400 analyzer
When a sample has been pipetted onto the test pads of a test strip, and the
specific gravity and clarity have been measured, the remaining urine is ejected
into the rinse station. The needle and the flow cell are washed with water
before the next sample is taken.
Roche Diagnostics
Operator’s Manual · Version 1.0 A-5
1 Introduction to the system URISYS 2400
Description of the URISYS 2400 analyzer
Roche Diagnostics
A-6 Operator’s Manual · Version 1.0
System description B
Hardware
Roche Diagnostics
Operator’s Manual · Version 1.0 B-1
2 Hardware URISYS 2400
Hardware description
Hardware description
The URISYS 2400 analyzer consists of several major components:
o Rack transport system
o Liquid handling system
o Test strip cassette compartment
o Automated test strip processing area
o 3-wavelength reflectance photometer
o Combined, flow-through refracto-/turbidimeter
o User interface screen
o Inbuilt computer
o Battery-buffered memory
o Diskette drive (for MS-DOS compatible diskettes)
Roche Diagnostics
B-2 Operator’s Manual · Version 1.0
URISYS 2400 2 Hardware
Hardware description
C D
E
B
A
J I H G F
Figure B-1 The URISYS 2400 analyzer
A Diskette drive F Rack output line
B URISYS 2400 test strip cassette G STAT sampling position
compartment
H Rack cross-feed line
C Touch screen
I Barcode reader
D Pipette mechanism
J Rack input line
E Wash and sample syringes
Roche Diagnostics
Operator’s Manual · Version 1.0 B-3
2 Hardware URISYS 2400
Hardware description
Standard accessories
The standard accessories that accompany the URISYS 2400 analyzer are:
o Water supply tank with a liquid level detector
o Water supply tube with filter
o Two rack trays
o Liquid waste tube 5 m (16.4 ft) length
o Strip waste container
o Standard maintenance kit (O-rings, seals)
o Fuses
o Power cord
o Three software installation diskettes
o Operator’s Manual (English version)
Water supply tank with liquid The water supply tank that supplies deionized water to the analyzer is located
level detector externally. It has a maximum capacity of 5 L (5.3 qts). This volume is
sufficient to process approximately 1000 samples.
Rack trays Two identical trays are supplied with the analyzer. Each tray holds up to 15
racks at one time. One tray is used to load the racks onto the rack input line.
The second tray is used as a stationary rack output tray during sample
processing and serves to remove racks from the rack output line.
Strip waste container Used test strips are deposited in a strip waste container that has a minimum
capacity of 420 test strips. (The capacity is calculated on the basis of the
contents of one URISYS 2400 cassette plus a safety margin.)
The strip waste container is located on the right side of the analyzer, behind
the door panel. The container is designed to minimize the contact an operator
has with potentially infectious and toxic materials.
When the strip waste container is full, a message is displayed on the touch
screen.
Roche Diagnostics
B-4 Operator’s Manual · Version 1.0
URISYS 2400 2 Hardware
Installation
Additional items
Items that are necessary or optional for the installation and operation of the
URISYS 2400 analyzer, but are not supplied with the analyzer, are:
o Liquid waste tank (with or without an optional overflow sensor)
o Racks
o Sample tubes
o Washing solution
o URISYS 2400 cassette
o URISYS 2400 calibration strip
o External printer and standard printer cable (parallel/Centronics interface)
o Waste container cartons
o Serial interface (host) cable
o 1.44 MB diskette for storage of data
Installation
Read this manual carefully before installation, to ensure correct operation of
the analyzer.
If you have any questions after initial installation, contact Technical Support.
Basic requirements
Surface and space The following surface and space requirements must be met for the use of the
measurements URISYS 2400 analyzer, and the proper ventilation of the analyzer when it is in
use:
o 5 cm (2.0 inches) or more (rear)
o 20 cm (7.9 inches) or more (left side)
o 30 cm (11.8 inches) or more (right side)
Use a table that is able to support the mass of the analyzer, which is
approximately 85 kg (187 lbs). The table must be level. Any inclination must
be less than 1.5 degrees in every direction.
Roche Diagnostics
Operator’s Manual · Version 1.0 B-5
2 Hardware URISYS 2400
Installation
Power supply A standard 230 V/50 Hz or 115 V/60 Hz power line is required to operate the
analyzer. It must be connected to a grounded power supply socket (protection
class 1).
Although the maximum power consumption of the instrument is 200 VA, the
external power supply should fulfill the following minimum requirements to
cover for peaks at switching on.
Electric power supply capacity:
o Minimum 500 VA
o Minimum 5 Amp, 3-pin power supply socket
The power switch and power supply socket are located on the right side of the
analyzer.
A B
Water supply An external water supply tank for deionized water is supplied with the
URISYS 2400. The deionized water must be germ free and have an electric
conductance value that is less than or equal to 1 µS/cm.
Roche Diagnostics
B-6 Operator’s Manual · Version 1.0
URISYS 2400 2 Hardware
Installation
Set up the water supply The external water supply tank provides water to the analyzer for washing.
tank The tank has a volume of approximately 5 liters (5.3 quarts), which allows for
the measurement of about 1000 samples.
B C
A D
Figure B-3 Connecting the water supply tank and liquid waste tank to the analyzer
A Water inlet connection C Connector for water level detector
B Liquid waste tube connection D Connector for liquid waste level detector
If the water is contaminated by the growth of bacteria or algae, it may result in faulty
liquid detection and rinsing in the pipetting mechanism and rinse circuits of the
analyzer.
Roche Diagnostics
Operator’s Manual · Version 1.0 B-7
2 Hardware URISYS 2400
Installation
Liquid waste tank The liquid waste tank is located outside the analyzer. The capacity of the tank
depends on the specific tank that is used. The tank must have a volume of at
least 5 L (5.3 qts). An overflow sensor is available as an option.
A 5 m (16.4 ft) liquid waste tube is supplied for the connection of the liquid
waste tank to the analyzer.
1 Ensure the liquid waste tank is empty before use. The liquid waste tank
must be positioned well below the level of the analyzer, to allow for the
reliable flow of the liquid waste into the tank.
2 Connect the liquid waste tube to the drain socket on the analyzer. If
necessary, shorten the liquid waste tube to ensure smooth and even flow.
3 If an overflow sensor is used, connect it to the drain level connector on the
analyzer and place it into the liquid waste tank.
TS transfer base
1 Open the strip waste container door on the right side of the analyzer.
2 Insert the TS transfer base smoothly to its end position, and ensure that it
is positioned correctly.
Figure B-4 Inserting the TS transfer base into the URISYS 2400 analyzer
3 Turn the screw on the TS transfer base until it locks to secure the base.
4 Insert a waste box carton into the strip waste container and place it into its
correct position.
Roche Diagnostics
B-8 Operator’s Manual · Version 1.0
URISYS 2400 2 Hardware
Installation
External printer
An external printer can be connected via a standard parallel interface
(Centronics type).
Recommended printers are:
Country Printer
Roche Diagnostics
Operator’s Manual · Version 1.0 B-9
2 Hardware URISYS 2400
Installation
Environmental conditions
The URISYS 2400 analyzer must be used in an environment that meets the
following conditions:
o Free from excessive dust
o Well-ventilated area
o Not exposed to direct sunlight
o Floor that is level (with an inclination of less than 1.5 degrees in every
direction) and firm
o Table that can support the mass of the instrument, which is approximately
85 kg (187 lbs)
o Ambient temperature ranging from 15-30°C (59-86°F), with an acceptable
variation of ±2°C (±3°F) during measurement
o Relative humidity ranging from 20% to 80%, without moisture
condensation
o Free from vibrations
o Availability of a 3-pin power supply socket within 3 m (9.8 ft) of the
analyzer
o Free from abrupt mains voltage fluctuation (±10 V)
o Well-distanced from a machine generating a high frequency voltage (for
example, a centrifuge)
o Free from electromagnetic wave interference
Roche Diagnostics
B-10 Operator’s Manual · Version 1.0
URISYS 2400 3 Software
Table of contents
Software
Roche Diagnostics
Operator’s Manual · Version 1.0 B-11
3 Software URISYS 2400
Installation
Installation
The software supplied with the analyzer supports all functions of the
URISYS 2400 analyzer. Specific functions are password-protected to avoid
accidental modifications.
The English software is pre-installed and ready for use on delivery of the
analyzer.
In case re-installation of the software is necessary, proceed as follows.
1 Save the system parameters, test parameters, routine sample results, STAT
sample results, control results and the alarm trace onto a diskette as safety
backup.
Step 1
Roche Diagnostics
B-12 Operator’s Manual · Version 1.0
URISYS 2400 3 Software
Installation
Basic screen composition Generally, the status of the analyzer is indicated in the upper left of the yellow
line of the main screen. There are three main status types. For information on
the different status types displayed on the touch screen, see:
Roche Diagnostics
Operator’s Manual · Version 1.0 B-13
3 Software URISYS 2400
Installation
Immediately below the line indicating the status of the analyzer, the name of
the current screen is displayed. For example, the name of the screen displayed
above is [Routine Monitor].
Press <Home> on the initial [Copyright] screen to display the [Routine Monitor]
screen.
There are a number of buttons that consistently appear on the screens. The
following table summarizes these buttons and their functions.
Button Explanation
<PgUp> Page up. Press this button to return to the previous screen.
<CLR> The data entered is cleared. The CLR button only works
before the Enter button is pressed.
<Cancel> Cancels any data that has been entered and returns to the
previous screen without having made any changes. The
Cancel button only works before the Enter button is pressed.
< > Enter. Press this button to confirm the newly entered data.
Roche Diagnostics
B-14 Operator’s Manual · Version 1.0
URISYS 2400 3 Software
Installation
As the operator navigates through the different screens, there are various
windows that can appear. These serve different functions, such as
confirmation of a selection. The figure shows an example:
Figure B-7 Example of a window that can appear on the touch screen
Some of the buttons for selecting settings work with a toggle function.
Roche Diagnostics
Operator’s Manual · Version 1.0 B-15
3 Software URISYS 2400
Installation
Parameters
Units Password
Figure B-8 System hierarchy chart of the progression through the different
screens of the Parameters menu
Roche Diagnostics
B-16 Operator’s Manual · Version 1.0
URISYS 2400 3 Software
Installation
Test parameters
Sieve and abnormal Use the sieve settings to define the criteria for a flag (S) on the corresponding
values test parameter result. These flags can be used to identify urine samples to be
examined by additional methods, for example, sediment microscopy.
Use the abnormal settings to define the criteria that indicate potentially
pathological values. An asterisk (*) flags the corresponding test parameter
results.
The concentration values which are specified on the screen represent the
lowest concentration ranges from which on the test parameter results are
flagged.
For more information, see:
Roche Diagnostics
Operator’s Manual · Version 1.0 B-17
3 Software URISYS 2400
Installation
1 Press <Sieve & Abnormal> on the [Test Parameters] screen. The [Sieve &
Abnormal] screen is displayed listing the sieve criteria for each test.
2 Press <Select> to change the display of the items from sieve to abnormal
criteria and back.
3 Press <Set> to change the criteria. The [Password] window appears.
4 Enter the 4-digit password using the keyboard on the screen.
5 Press <Enter>. The [Sieve & Abnormal Setting] screen is displayed with
both the sieve and abnormal values listed for each test parameter,
beginning with specific gravity (SG).
6 Press <Test> to select another test parameter.
7 Press “<” or “>” to select sieve or abnormal settings.
8 Press <+> or <-> to select the desired concentration range.
9 Press <Enter> after each selection to accept the chosen values.
10 Press <PgUp> to return to the [Sieve & Abnormal] screen and check if the
settings correspond to your requirements.
11 Press <PgUp> to return to the [Test Parameters] screen.
Print order The operator can define the order of the test parameters on the screen and on
the result printout, in order from 1 to 12. If an item is not to be printed, enter
zero.
Roche Diagnostics
B-18 Operator’s Manual · Version 1.0
URISYS 2400 3 Software
Installation
1 Press <Print Order> on the [Test Parameters] screen. The [Print Order]
screen is displayed.
2 Enter the test parameter order if the requirements of the laboratory differ
from the default settings.
3 Press “<” or “>” to move through the different fields.
4 Press <Enter> after each entry.
5 Press <PgUp> to return to the [Test Parameters] screen.
If the 5-digit number of the chosen rack is not known, read and print the rack ID via
the Barcode Check function (see “Checks” on page C-41).
Roche Diagnostics
Operator’s Manual · Version 1.0 B-19
3 Software URISYS 2400
Installation
Roche Diagnostics
B-20 Operator’s Manual · Version 1.0
URISYS 2400 3 Software
Installation
Range table The operator can view the concentration ranges, reflectance values,
wavelengths and compensation data for each test parameter. The operator
gains access to the range and reflectance values by using the password that is
supplied with the analyzer.
The wavelength data (W) that are displayed are fixed values. These values
cannot be changed.
The compensation data (C) designate factors for the compensation of the
intrinsic urine color, for example, 0 means no color compensations. Higher
numbers mean stronger compensation.
Changing ranges leads to result reports which differ from the default
concentration ranges as shown in:
The concentration ranges and reflectance values can be reset to the default
values using the <Default> button.
Roche Diagnostics
Operator’s Manual · Version 1.0 B-21
3 Software URISYS 2400
Installation
1 Press <Range Table> on the [Test Parameters] screen. The [Range Table]
screen is displayed.
2 Press <Test> to select the test parameter to be viewed. For each test
parameter, the range and reflectance values can be changed.
3 Press <Set Range> to change the range. The [Password] window appears.
4 Enter the 4-digit password.
5 Press <Enter>. When the correct password is entered, the [Set Range]
screen is displayed.
6 Press <Test> to select a test parameter.
7 Enter the range values. Press “<” or “>” to move through the different
fields.
8 Press <Enter> after each value is entered.
9 Repeat steps 6-8 if the ranges of other test parameters need to be changed.
10 Press <PgUp> to return to the [Range Table] screen.
o The screen allows for eight ranges to be set for each parameter. However, not all
of the ranges are used by the default settings.
o If ranges are changed for conventional concentration units, these changes are
not transferred to the SI concentration units memory, and vice versa. To be
consistent, corresponding changes must be made in both range tables. (See
Table F-1 on page F-5.)
If arbitrary units (1+, 2+, 3+) should be used instead of concentration values
or in addition to concentration values, proceed as follows:
1 Call up the [Set Range] screen.
2 Press <Test> to select a test parameter.
3 Enter the arbitrary unit in the corresponding field and delete the
concentration value by pressing <Space> if you want to replace the
conventional or SI units.
4 Enter the arbitrary unit in the corresponding field without deleting the
concentration value if you want to used combined units, for example,
“2+ 25”.
If combined units are used for the parameter Glucose, the entry of “4+1000” is not
possible because of the limitation of 5 digits per field. Use “4+999” instead.
Roche Diagnostics
B-22 Operator’s Manual · Version 1.0
URISYS 2400 3 Software
Installation
1 Press <Set Reflect>. on the [Range Table] screen. The [Password] window
appears.
2 Enter the 4-digit password.
3 Press <Enter>. When the correct password is entered, the [Set Reflectance]
screen is displayed.
4 Press <Test> to select a test parameter.
5 Enter the new reflectance values between the indicated values. Apply
moderate changes and validate the effect to the evaluation sensitivity.
6 Press <Enter> after each value is entered.
7 Repeat steps 4-6 if reflectance values of other test parameters need to be
changed.
8 Press <PgUp> to return to the [Range Table] screen.
o The screen allows for eight values to be set for each parameter. However, not all
of the values are used by the default settings.
o Reflectance values that have been changed will be marked with a “#” flag on
the [Range table] screen as well as in the Parameter Settings printout. This flag
will also appear with the corresponding test parameter on the result printout.
Roche Diagnostics
Operator’s Manual · Version 1.0 B-23
3 Software URISYS 2400
Installation
Units
The operator can select the units for reporting results (conventional or SI).
1 Press <Units> on the [Test Parameters] screen. The [Units] screen is
displayed. The units currently in use are displayed.
2 Press <Units> to change the type of units to be displayed on the screen and
on the printouts, and to be sent to the host computer. The [Confirmation]
window appears.
3 Press <OK> to change units.
When the units are changed, results are reported in the newly chosen units.
Therefore, it is important to set units before starting measurements. The change of
units also applies when re-printing previously measured results.
System parameters
The operator can select six system parameter settings. They are:
o Sample barcode
o Printer (includes the paper length, the number of samples printed per
page, and the page heading)
o Communication
o Date and time
o Password
o Registration of the wash rack
Roche Diagnostics
B-24 Operator’s Manual · Version 1.0
URISYS 2400 3 Software
Installation
Sample barcode
3 Press <Code39 C/D>, <NW7 C/D>, and <ITF C/D> to enable or disable
the use of a check digit function for each type of barcode. A [Warning]
window will alert the operator of the increased probability for misreads
when check digits are disabled.
4 Press <PgUp> to return to the [System Parameters] screen.
If using an ITF barcode type, Roche/Hitachi recommends its use only with a check
digit function.
Printer
Roche Diagnostics
Operator’s Manual · Version 1.0 B-25
3 Software URISYS 2400
Installation
COMM. parameter
Changing the settings is only possible when the analyzer is set to “Host Off-line”
(see “Start conditions before sampling” on page C-5).
These settings must match the corresponding settings of the host computer. Only
specialists or trained supervisors should change these settings.
Date/Time The operator can change the format of the date and time, and set the correct
local dates, which are displayed on the touch screen and reported with the
measurement results and in the alarm messages.
Roche Diagnostics
B-26 Operator’s Manual · Version 1.0
URISYS 2400 3 Software
Installation
Password The operator can change the password used to gain access to the
[Sieve & Abnormal Setting], [Set Range], [Set Reflectance], and [Default]
screens. The password is identical in each case. Only numerical characters can
be used.
Register the wash rack The operator can define the ID of the rack to be used for washing. If the
washing procedure is performed using a rack on the rack input line, the
analyzer recognizes the specified rack and performs the washing procedure.
If the 5 digit number of the chosen rack is not known, read and print the rack ID via
the Barcode Check function (see “Checks” on page C-41).
Mark the specified rack as a wash rack. Do not use the wash rack for routine or
control samples; otherwise, the samples will not be measured.
Roche Diagnostics
Operator’s Manual · Version 1.0 B-27
3 Software URISYS 2400
Installation
Parameter media
The operator can select three media functions. They are to Load, Save, and
Print the parameter settings.
Load
Save
Roche Diagnostics
B-28 Operator’s Manual · Version 1.0
URISYS 2400 4 Cassette
Table of contents
Cassette
This chapter describes installation and handling of the URISYS 2400 cassette.
Roche Diagnostics
Operator’s Manual · Version 1.0 B-29
4 Cassette URISYS 2400
First-time installation
First-time installation
Unpack and install the cassette
The cassette is delivered in an airtight polyethylene-coated aluminum bag,
which is tightly placed in an outer package box. This secure packaging ensures
the shelf life of the cassette in the unopened package as printed on the package
box.
When the aluminum bag is opened, the cassette must be placed into the cassette
compartment of the instrument and the compartment door firmly closed within
three minutes to prevent damage of the test strips by humidity or nitrous gases in
the air.
The test strips are stable within the tightly closed cassette compartment for
fourteen days. After this time, the cassette must be replaced by a new one.
Carefully unpack the URISYS 2400 cassette to prevent damage to the test
strips. Use the [Change Cassette] screen in the Maintenance menu to install
the cassette for the first time. The instrument must be in Standby.
Roche Diagnostics
B-30 Operator’s Manual · Version 1.0
URISYS 2400 4 Cassette
First-time installation
Roche Diagnostics
Operator’s Manual · Version 1.0 B-31
4 Cassette URISYS 2400
First-time installation
Handling precautions When handling test strips, the following precautions must be observed:
o Only use a URISYS 2400 cassette that has not exceeded its expiration date.
o Store cassette packages at a temperature between 2-30°C (36-86°F).
o Refrigerated cassettes must be at room temperature for at least two hours
before opening the package and using the cassette.
o Check the package for damage.
o After removing the cassette from the packaging, ensure the test strip layers
are correctly aligned and the test pads are not unusually discolored.
o Do not directly touch the test pads of the test strip.
o Do not let the strips come in contact with water or with any volatile,
strongly acidic, or alkaline chemicals.
o Install the cassette into the analyzer within 3 minutes of opening the seal.
o Keep the cassette in the firmly locked cassette compartment until the test
strips are completely used up.
o Do not open the cassette compartment door as long as the cassette is not
empty or the on-board stability has not expired.
o Follow the instructions provided with the URISYS 2400 cassette by the
manufacturer and read carefully all information on the package insert.
Roche Diagnostics
B-32 Operator’s Manual · Version 1.0
Operation C
Daily operation
Roche Diagnostics
Operator’s Manual · Version 1.0 C-1
5 Daily operation URISYS 2400
Roche Diagnostics
C-2 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Initial preparations
Initial preparations
Before switching on the power to operate the analyzer, perform the following
tasks:
o Fill the water supply tank with deionized water and place the liquid level
detector in the water supply tank.
o Ensure that the TS transfer base is inserted correctly.
o Empty the strip waste container and ensure that it is positioned correctly.
o Empty the liquid waste tank.
o Place an empty rack tray on the rack output line.
o Remove all air bubbles from the syringes after switching on the analyzer.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-3
5 Daily operation URISYS 2400
System start
System start
The power switch is located on the right side of the analyzer. The OFF position is
represented by Ο.
1 Turn the power switch to the ON (|) position. The [Please Wait] screen
appears. Loading and initializing of the software takes approximately 5
minutes. After the first loading stage, the progress is indicated on-screen.
2 Check and remedy any alarm messages. Clear the alarm messages.
Roche Diagnostics
C-4 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
System start
For input of numbers or change of the host and printer settings, the analyzer must
be in Stand-By.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-5
5 Daily operation URISYS 2400
System start
The “Input Area” indicates the sequence number range that follows the displayed
sequence number and is not yet occupied by sample results. The range end number
is the number of the next occupied sequence number minus five.
An example of the numbering 122 samples have been measured starting with Seq. No. 0001. The displayed
ranges would be as follows: Routine Seq. No. is 0123. No samples with higher sequence numbers have
been measured. In this case, an input range of 0123-0996 is displayed.
For example, the next occupied (measured) sequence number is 0001. Due to
the ring buffer structure of the result memory, this is equivalent to 1001. The
end of the input range is calculated as 1001 - 5 = 996. Input of 996 allows the
measurement of one additional rack with 5 samples before reaching the next
occupied sequence number. For alarm messages when reaching the next
occupied sequence number, see:
When using the “Overwrite Mode”, be aware that the sequence number will swap to
1 automatically after the sequence counter has reached 1000 (or 200 for STAT). In
this case, the instrument starts to overwrite results with sequence numbers starting
at 1 again.
If the “Input Area” is indicated as 0000 - 0000, delete the results of samples
with sequence numbers higher than the displayed Seq. No., (see “Delete
routine or STAT sample data” on page C-27). Otherwise, analysis of samples
cannot continue.
Roche Diagnostics
C-6 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
System start
Using the [Start Condition] screen, the operator can choose to activate or
deactivate the connection to a host computer system. Ensure that the analyzer
and the host computer are connected via an interface cable.
1 Press <Start Condition> on the [Routine Monitor] screen. The [Start
Condition] screen is displayed.
2 Press <Host On/Off> to activate or deactivate the connection.
3 Press <PgUp> to return to the [Routine Monitor] screen.
Using the [Start Condition] screen, the operator can select to have the result
data printed in real time. In this case, the result is printed as soon as the
measurement of a sample is completed. The categories of results to be printed
can also be selected.
The options are:
o All
o Abnormal
o Sieve
o Normal
o None
1 Press <Start Condition> on the [Routine Monitor] screen. The
[Start Condition] window is displayed.
2 Press <Print Mode> to select the category of results to be printed.
3 Press <PgUp> to return to the [Routine Monitor] screen.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-7
5 Daily operation URISYS 2400
Sample preparation
Sample preparation
Analyze samples within two hours from the time that the urine samples were
obtained.
Do not expose samples to direct sunlight.
Do not add any preservatives to the samples.
Take care that sample material does not contain air bubbles or foam.
If a syringe sample tube is used, for example Urine Monovette, ensure that the
plunger is at its lowest position. Otherwise, the sample probe may be
damaged.
Roche Diagnostics
C-8 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Sample preparation
If the sample tubes are labeled with barcodes, ensure that the barcode labels
are correctly aligned with the barcode reader. For more information about the
specifications of usable barcodes and the placement of labels on sample tubes,
see:
Figure C-3 Loading racks onto URISYS 2400 analyzer with the rack tray
Roche Diagnostics
Operator’s Manual · Version 1.0 C-9
5 Daily operation URISYS 2400
Sample preparation
5 Ensure that the rack tray is aligned horizontally with the rack input line.
Move the racks with care in order not to spill the samples.
C
D
Figure C-4 Loading a rack directly into the guide rail of the rack input line
A Rack C Guide rail
B Sample tube D Rack input line
a To load racks directly onto the rack input line, without a rack tray
Roche Diagnostics
C-10 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Termination of analysis
“To enter the routine sequence number, STAT sequence number and operator ID” on
page C-6
Delete the results of samples with sequence numbers higher than the actual
sequence number.
Do not open the cassette compartment door or the strip waste container door
during analysis. Otherwise, the analyzer stops and moves to Stand-By status. When
<Start> is pressed again, the subsequent reset clears the racks still present in the
cross-feed line without sampling them, and removes test strips still present on the
TS transfer base without measuring them.
Termination of analysis
Completion of analysis
After measuring the last test strip, the analyzer changes to Stand-By.
To continue sampling, place more racks with sample tubes on the input line,
and press <Start> to begin analysis again. The analyzer continues sampling
and measurement. For more information on the processing of additional
samples, see:
Interrupting analysis
Using the S. Stop button on the [Routine Monitor] screen, the operator can
interrupt analysis. Racks in the cross-feed line are still processed; then the
analyzer turns to Sampling Stop status. Racks in the input line remain in their
current position. Samples that have already been pipetted onto test strips are
measured.
a To interrupt analysis
Roche Diagnostics
Operator’s Manual · Version 1.0 C-11
5 Daily operation URISYS 2400
Remove processed samples
1 Place the next racks directly into the rack input line.
2 Press <Start> on the [Routine Monitor] screen. The analyzer continues
sampling and measurement without an intermediate reset.
When <Start> is pressed again, the subsequent reset clears the racks still present in
the cross-feed line without sampling them, and removes test strips still present on
the TS transfer base without measuring them.
Automatic termination
If there are no more racks on the rack input line, sampling is automatically
terminated. The analyzer moves to Sampling Stop status. The processing of
test strips continues until all samples that have been pipetted onto test strips
have been measured.
Roche Diagnostics
C-12 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Process additional samples
Operation status
1 Place the racks on the rack input line, either directly or using a rack tray.
2 When the rack end of the previous rack is detected, the rack pushers
automatically return to the front of the rack input line to search for
additional racks, and move them to the cross-feed line for sampling.
Stand-By status
1 Place the racks on the rack input line, either directly or using a rack tray.
2 Press <Start> on the [Routine Monitor] screen. The analyzer performs a
reset; then the rack pushers automatically move the racks to the cross-feed
line for sampling.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-13
5 Daily operation URISYS 2400
Process STAT (emergency) samples
STAT sample results are stored in a separate STAT sample memory. STAT samples
are marked on the result printout with an “E” preceding the sequence number.
Sample ID and rack ID are not read in the STAT position of the analyzer. To
enter a STAT sample ID manually, the barcode reader at the cross-feed line
must be set to “Enable”.
Roche Diagnostics
C-14 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Process STAT (emergency) samples
Roche Diagnostics
Operator’s Manual · Version 1.0 C-15
5 Daily operation URISYS 2400
Process STAT (emergency) samples
2 Press <Start> on the [Routine Monitor] screen. The status of the analyzer
changes to Operation.
3 The STAT sample is pipetted and measured. Then the status of the analyzer
returns to Sampling Stop.
If <OK> is pressed during the rack end status, the sample will not be
measured. Wait until the analyzer changes to Sampling Stop, then proceed
as described in “To process a STAT sample in Sampling Stop status” on page
C-16.
Roche Diagnostics
C-16 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Process control samples
Do not use the control rack for routine samples, otherwise the routine sample
results will be stored in the control data memory
Roche Diagnostics
Operator’s Manual · Version 1.0 C-17
5 Daily operation URISYS 2400
Replace an empty URISYS 2400 cassette
When the new cassette is inserted, the [Routine Monitor] screen indicates the
number of remaining strips as “400” and the new loading date.
Roche Diagnostics
C-18 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Replace an expired URISYS 2400 cassette
The analyzer must be in the Stand-By status. If it is not, press <STOP> on the
[Routine Monitor] screen and wait for the Stand-By status to appear.
When the new cassette is inserted, the [Routine Monitor] screen indicates the
number of remaining strips as “400” and the new loading date. The analyzer is now
ready for further processing.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-19
5 Daily operation URISYS 2400
Re-insert a used URISYS 2400 cassette
2 Check if all remaining test strip layers in the cassette are correctly aligned.
If necessary, remove loose strips from the cassette.
3 Remove loose strips from the cassette compartment, if present.
4 Re-insert the cassette and close the compartment door firmly.
5 Press <Used> in the [Change Cassette] screen. A window “Now searching
next TS” appears.
6 After completion, press <OK> to return to the [Change Cassette] screen.
7 Press <Home> to return to the [Routine Monitor] screen. The display
shows the number of remaining test strips and the date of the first-time
loading of this cassette.
The analyzer is now ready to continue analysis. If the failure occurs again, contact
technical support.
Roche Diagnostics
C-20 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Analysis and documentation of results
Roche Diagnostics
Operator’s Manual · Version 1.0 C-21
5 Daily operation URISYS 2400
Analysis and documentation of results
Display the result data The operator can access result data in the same way for all sample types.
Search sample data Use the sequence number (Seq. No.) or the sample ID number (ID) to search
for data for routine and STAT samples. The result data can be searched at any
analyzer status – Operation, Sampling Stop, or Stand-By.
Results cannot be retrieved using rack number, position number, date and time, or
operator ID.
Roche Diagnostics
C-22 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Analysis and documentation of results
a To search for routine and STAT sample data using the sequence
number
1 Press <Type> in the [Routine Data Monitor] screen to select the type of
data to be displayed (routine or STAT).
2 Press <Seq.No.> to enter the requested sequence number. The [Search
Seq.No.] window appears.
3 Enter the sequence number.
4 Press <Enter>. The result data for the requested sample are displayed.
5 Press <Prev> or <Next> on the [Routine Data Monitor] screen, to display
prior or subsequent results.
a To search for routine and STAT sample data using the sample ID
number
1 Press <Type> on the [Routine Data Monitor] screen to select the type of
data to be displayed.
2 Press <ID>. The [Search ID] window appears.
3 Enter the requested sample ID number, using alphanumeric characters.
4 Press <Enter>. The result data of the requested samples are displayed.
5 Press <Prev> or <Next> on the [Routine Data Monitor] screen, to display
prior or subsequent results of the same sample ID.
Entering the sample ID If a sample ID number has not been read by the barcode reader, or if the
number sample ID must be added after measurement (for example, a STAT sample),
the analyzer must be in Stand-By status and the barcode reader must be set to
“Enabled”.
Entering the sample ID number is only possible if no ID number is already stored for
this sequence number.
1 Press <Type> in the [Routine Data Monitor] screen to select the type of
data to be displayed (routine or STAT).
2 Press <Input ID> to enter the sample ID number. The [Input ID] screen is
displayed.
3 Enter the sample ID number using alphanumeric characters.
4 Press <Enter> to confirm the entry.
5 Press <OK>. A confirmation window appears.
6 Press <Yes> to return to the [Routine Data Monitor] screen.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-23
5 Daily operation URISYS 2400
Analysis and documentation of results
When result data are edited, they are printed with an “!” edit mark.
Print routine and STAT This print option is available for batch printing. For more information on
sample data printing in real time, see:
“To select real time printouts and specify the result data to be printed” on page C-7
1 Press <Type> on the [Routine Data Monitor] screen to select the type of
data to be displayed (routine or STAT).
2 Press <Media>. The [Routine Data Monitor Media] screen is displayed.
3 Enter the range of sequence numbers to be printed, using numbers
between 1 and 1000 for routine sample data, or between 1 and 200 for
STAT sample data. Enter 0 to print all sample data.
4 Press <Print>. The [Confirmation] window appears.
5 Press <Data Select> to select the category of results to be printed (All,
Abnormal, Sieve, Normal).
6 Press <OK> to start printing.
Roche Diagnostics
C-24 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Analysis and documentation of results
Presentation of results The operator can specify the print order of the test parameters on the
on the printout printout, and the number of samples to be printed on each page.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-25
5 Daily operation URISYS 2400
Analysis and documentation of results
Send result data to the This option is available for batch sending.
host computer To send data, the host connection must be activated.
1 Press <Type> on the [Routine Data Monitor] screen to select the type of
sample data to be sent (routine or STAT).
2 Press <Media>. The [Routine Data Monitor Media] screen is displayed.
3 Enter the range of sequence numbers to be sent, using numbers between 1
and 1000 for routine sample data, or between 1 and 200 for STAT sample
data. Enter 0 to send all sample data.
4 Press <Send>. The [Confirmation] window appears.
5 Press <OK>.
6 Press <Quit> to terminate the transfer of data at any time.
Write/read routine or This option allows for data to be saved on a diskette or for data to be read
STAT sample data to/ from a diskette.
from a diskette
1 Press <Type> on the [Routine Data Monitor] screen to select the type of
sample data to be written/read (routine or STAT).
2 Press <Media>. The [Routine Data Monitor Media] screen is displayed.
3 Enter the range of sequence numbers to be written/read to/from a diskette,
using numbers between 1 and 1000 for routine sample data, or between 1
and 200 for STAT sample data. Enter 0 to write/read all sample data.
4 Press <FD R/W>. The [Confirmation] window appears.
5 Press <File Type> to select the format of the files to be written/read (ASCII
or binary).
6 Insert appropriate diskette.
Roche Diagnostics
C-26 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Analysis and documentation of results
7 Press <Write> to write data to a diskette, or press Read to read data from a
diskette.
If result data are read from a diskette, the data actually stored in the analyzer are
deleted before the data on diskette are read.
Delete routine or STAT Sample data can be cleared from the memory either completely or selectively,
sample data within a range of sequence numbers.
1 Press <Type> on the [Routine Data Monitor] screen to select the type of
sample data to be deleted (routine or STAT).
2 Press <Media>. The [Routine Data Monitor Media] screen is displayed.
3 Enter the range of sequence numbers to be deleted using numbers between
1 and 1000 for routine sample data, or between 1 and 200 for STAT sample
data. Enter 0 to delete all sample data.
4 Press <Delete>. The [Confirmation] window appears.
5 Press <OK>.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-27
5 Daily operation URISYS 2400
Analysis and documentation of results
If no results are available for a sample, a message, for example “Sample short” is
displayed on the screen and printed on the printout.
The operator can access result data in the same way for all control samples.
1 Press <Menu> on the [Routine Monitor] screen. The [Menu] screen is
displayed.
2 Press <Data Monitor>. The [Data Monitor] screen is displayed.
3 Press <Control Data Monitor>. The [Control Data Monitor] screen is
displayed. The concentration results of the most recently tested control
sample are displayed.
Roche Diagnostics
C-28 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Analysis and documentation of results
Search for control Use the sequence number and the level category of the controls to search for
sample data data of control samples.
Results cannot be retrieved using the control name, lot number, date and
time, or operator ID.
1 Press <Level> on the [Control Data Monitor] screen to select the level of
controls to be displayed.
2 Press <Seq. No.> to enter the requested sequence number.
3 The [Control Seq. No.] window appears.
4 Enter the sequence number, using numbers between 1 and 100.
5 Press <Enter>. The result data of the entered sequence number are
displayed.
6 Press <Prev> or <Next> to display previous or subsequent control samples
data.
7 Press <Concentr.> or <Reflectance> to select concentration data or
reflectance data.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-29
5 Daily operation URISYS 2400
Analysis and documentation of results
Print control sample data This option allows for batch printing of control sample data of the selected
level.
1 Press <Level> on the [Control Data Monitor] screen to select the level of
controls.
2 Select <Concentration> or <Reflectance>.
3 Press <Media>. The [Control Data Monitor Media] screen is displayed.
4 Enter the range of sequence numbers to be printed between 1 and 100.
Enter 0 to print all data.
5 Press <Print>. The [Confirmation] window appears.
6 Press <OK>.
7 Press <Stop> to terminate printing at any time.
Presentation of result The print order of the test parameters results in concentration values is
data on the printout identical to the order selected for routine samples.
Roche Diagnostics
C-30 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Analysis and documentation of results
Roche Diagnostics
Operator’s Manual · Version 1.0 C-31
5 Daily operation URISYS 2400
Analysis and documentation of results
Send control sample This option is available for batch sending. The host connection must be
data to the host activated.
computer
Sample data are transferred only for the control level selected.
1 Press <Level> on the [Control Data Monitor] screen to select the level of
controls.
2 Select <Concentr.> or <Reflectance>.
3 Press <Media>. The [Control Data Monitor Media] screen is displayed.
4 Enter the range of sequence numbers between 1 and 100 to be sent to the
host computer. Enter 0 to send all data to the host computer.
5 Press <Send>. The [Confirmation] window appears.
6 Press <OK>.
7 Press <Quit> to terminate the transfer of data at any time.
To send control sample data to the host, the analyzer must be in Stand-By status.
If concentration is selected, only the concentration results are sent to the host.
If reflectance is selected, both concentration and reflectance results are sent to
the host.
Write/read control This option allows for data to be saved to a diskette or for data to be read from
sample data to/from a a diskette. Control sample data are written to a diskette only for the selected
diskette level. A DOS formatted 1.44 MB diskette is needed.
1 Press <Level> on the [Control Data Monitor] screen to select the level of
controls.
2 Press <Media>. The [Control Data Monitor Media] screen is displayed.
3 Enter the range of sequence numbers to be written to a diskette, using
numbers between 1 and 100. Enter 0 to write all sample data of the selected
level.
4 Press <FD R/W>. The [Confirmation] window appears.
5 Press <File Type> to select the format of the files to be written/read (ASCII
or binary).
6 Insert diskette.
Roche Diagnostics
C-32 Operator’s Manual · Version 1.0
URISYS 2400 5 Daily operation
Analysis and documentation of results
When reading control sample data from a diskette, all data of the selected level are
read. Currently stored data with identical sequence numbers are overwritten.
Delete control sample Sample data can be cleared from the memory either completely or selectively,
data within a range of sequence numbers.
Control sample data are deleted only for the level selected in the [Control Data
Monitor] screen.
1 Press <Level> on the [Control Data Monitor] screen to select the level of
controls.
2 Press <Media>. The [Control Data Monitor Media] screen is displayed.
3 Enter the range of sequence numbers to be deleted using numbers between
1 and 100. Enter 0 to delete all sample data.
4 Press <Delete>. The [Confirmation] window appears.
5 Press <OK>.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-33
5 Daily operation URISYS 2400
Switch off the analyzer
Roche Diagnostics
C-34 Operator’s Manual · Version 1.0
URISYS 2400 6 Special operations
Table of contents
Special operations
Roche Diagnostics
Operator’s Manual · Version 1.0 C-35
6 Special operations URISYS 2400
Calibration
Calibration
The operator can calibrate the photometer and the SG (specific gravity) meter
using the [User Calibration] screen.
Photometer calibration is based on the measurement of a calibration strip
with known reflectance specifications. This must only be performed using the
URISYS 2400 calibration strip manufactured by Roche Diagnostics.
Specific gravity calibration is based on the measurement of calibration
solutions of known refractive indices.
Calibration procedures are only possible when the analyzer is in Stand-By status.
Under routine conditions, calibrate the photometer and the SG meter every
four weeks.
Photometer calibration
The calibration values measured for each elevated area of the calibration strip
are compared with the previous user calibration values measured and the
factory calibration values.
If the calibration values are within ±2% of previous values, and within ±5% of
the factory values, “Calibration OK” is displayed on the [Photometer
Calibration] screen and the previous user calibration values are replaced.
If the calibration values differ by more than ±2% from previous values, or
more than ±5% from factory values, “Calibr. Failed” is displayed on the
[Photometer Calibration] screen and the calibration must be repeated.
Handle and store the URISYS 2400 calibration strips according to the
instructions in the calibration strip package insert.
The calibration strip is made of grey plastic material of constant reflectance
characteristics. It has nine elevated areas on the upper side, and three feet on
the underside. The two outer feet fix the strip on the lateral sides of the TS
Roche Diagnostics
C-36 Operator’s Manual · Version 1.0
URISYS 2400 6 Special operations
Calibration
basetransfer base, while the middle one fits into a slit at th e measuring
position of the TS transfer base..
Be aware that the TS transfer base, strip waste container and liquid tank may be
contaminated with potentially biological hazardous material. Always wear gloves
when handling these parts.
A D
Figure C-11
Make sure that the feet of the strip are locked in the correct place, the
middle foot being set to the end of the slit.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-37
6 Special operations URISYS 2400
Calibration
A
C
Figure C-12 Correct position of the calibration strip on the TS transfer base
A Calibration strip C Slit
B Holding grid
Roche Diagnostics
C-38 Operator’s Manual · Version 1.0
URISYS 2400 6 Special operations
Calibration
used for measurement (B = blue 470 nm, G = green 555 nm, O = orange
620 nm).
If the calibration failed, an alarm message appears. Repeat the calibration with
another calibration strip from the vial.
In case of “Calibr. Failed”, the values out of range are displayed with an
asterisk on the screen and the printout.
Insert the new calibration strip correctly and press <Calibrate>, followed by
<Execute> in the confirmation screen.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-39
6 Special operations URISYS 2400
Calibration
Roche Diagnostics
C-40 Operator’s Manual · Version 1.0
URISYS 2400 6 Special operations
Checks
Checks
The operator can perform a variety of checks by using the [Tools] screen.
These include a barcode check, printer check, and program check. The
analyzer must be in Stand-By status.
Barcode check
Using this option, barcodes of racks and sample tubes can be read and the
barcode numbers can be printed out. To use this function, the barcode reader
must be set to “Enabled”.
Printer check
This option checks the appearance of the printout of characters.
Program check
This option checks the actual software version installed on the analyzer.
Roche Diagnostics
Operator’s Manual · Version 1.0 C-41
6 Special operations URISYS 2400
Checks
Alarm trace
For more information, see:
Communication trace
This option allows activation of the communication trace. The HOST
connection must be deactivated.
Compact memory
This option allows for compacting the memory of the analyzer.
Roche Diagnostics
C-42 Operator’s Manual · Version 1.0
URISYS 2400 6 Special operations
Checks
FD utilities
This option allows for formatting a diskette and cleaning the disk drive.
a To use FD utilities
Roche Diagnostics
Operator’s Manual · Version 1.0 C-43
6 Special operations URISYS 2400
Checks
Roche Diagnostics
C-44 Operator’s Manual · Version 1.0
Maintenance D
This chapter explains general maintenance procedures for the URISYS 2400.
Roche Diagnostics
Operator’s Manual · Version 1.0 D-1
7 General maintenance actions URISYS 2400
Maintenance procedures
Maintenance procedures
The URISYS 2400 analyzer requires proper care and attention. Regular
maintenance ensures that excessive contamination of the analyzer is avoided
and that the analyzer works reliably.
Handle liquid waste and used test strips properly, and according to legislation on
water pollution, and treatment of drainage and waste matter.
Carefully follow the procedures specified in the Operator’s Manual for the operation
and maintenance of the analyzer. Leave maintenance of other areas to the Roche
Technical Support.
Every day
Be aware that the TS transfer base, strip waste container and liquid waste tank may
be contaminated with potentially hazardous material. Always wear gloves when
handling these parts.
Roche Diagnostics
D-2 Operator’s Manual · Version 1.0
URISYS 2400 7 General maintenance actions
Maintenance procedures
Every week
The following are maintenance procedures to be performed weekly:
o Check the water supply filter for growth of bacteria or algae.
o Clean the rack input line.
o Clean the rack trays.
Every month
The following are maintenance procedures to be performed monthly:
o Clean the outer parts (housing) of the analyzer.
o Clean or change the water supply filter.
Roche Diagnostics
Operator’s Manual · Version 1.0 D-3
7 General maintenance actions URISYS 2400
Maintenance menu
Maintenance menu
Using the [Maintenance] screen, the operator can perform general
maintenance tasks, such as washing the liquid handling system, exchanging
the URISYS 2400 cassette, resetting the analyzer, and removing air bubbles.
Maintenance procedures are only possible when the analyzer is in Stand-By
status.
1 Register the wash rack ID number to be used for the wash procedure.
Do not use the registered wash rack for routine or control samples.
If the washing cycle is interrupted, carry out the wash procedure again from the
beginning. Otherwise, the remaining detergent in the flow path may damage the
instrument or cause incorrect data.
When the <Stop> button on the [Routine Monitor] screen is pressed during the
wash procedure, a [Warning] window alerts the operator to the risks of interrupting
the washing process.
Roche Diagnostics
D-4 Operator’s Manual · Version 1.0
URISYS 2400 7 General maintenance actions
Maintenance menu
When handling the washing solution, ensure that you wear gloves and safety
goggles to avoid direct contact with the solution. If the solution contacts the body,
wash immediately with a large volume of water. Consult your laboratory protocol for
handling biohazard materials. If a sample is spilled on the analyzer, wipe up
immediately and apply disinfectant.
Do not remove the rack with the washing solution until the wash procedure is
finished.
If the washing cycle is interrupted, carry out the wash procedure again from the
beginning. Otherwise, the remaining detergent in the flow path may damage the
instrument or cause incorrect data.
When the <Stop> button on the [Executing] window is pressed, a [Warning] window
alerts the operator to the risks of interrupting the washing process.
Roche Diagnostics
Operator’s Manual · Version 1.0 D-5
7 General maintenance actions URISYS 2400
Maintenance menu
Do not use acid or alkaline detergents. Do not scratch the surface of the TS transfer
base.
Do not open the strip waste container door during analysis. Otherwise, the analyzer
stops and moves to Stand-By status.
Opening of the strip waste container door resets the counter of processed strips to
zero.
Roche Diagnostics
D-6 Operator’s Manual · Version 1.0
URISYS 2400 7 General maintenance actions
Maintenance menu
Do not use volatile solvents such as benzene. Do not scratch the surface of the
touch screen.
Air purge
The air purge function is used to remove air bubbles from the syringes, flow
cell, internal tubing, and sample needle. During first time installation, this
function is used to fill the water supply tube and the liquid handling system
with deionized water.
In addition, use the air purge function after refilling the water supply tank and
if air bubbles are present in the syringes.
Roche Diagnostics
Operator’s Manual · Version 1.0 D-7
7 General maintenance actions URISYS 2400
Exchange user-replaceable parts
Any component that comes into contact with sample fluids is potentially
biohazardous, and should be handled in an appropriate manner. Wear clean
protective gloves when changing Syringe seals.
m Absorbent towels
The gauze absorbs liquid that drains from the tubing as it is disconnected.
Roche Diagnostics
D-8 Operator’s Manual · Version 1.0
URISYS 2400 7 General maintenance actions
Exchange user-replaceable parts
Do not open the syringe cover while the power switch is turned on. Otherwise your
fingers may be snagged with moving parts.
1 Loosen the screw of the syringe cover below the W/S syringes and remove
it.
2 Disconnect the tubing at the top of the syringe by turning the small,
colorless tube-retaining nut counterclockwise.
3 Place a dry absorbent gauze under the small, colorless tube-retaining nut
that is connected to the side of the lower blue part of the syringe and
disconnect this side-tube (some water will flow out of the side connection
of the syringe).
Roche Diagnostics
Operator’s Manual · Version 1.0 D-9
7 General maintenance actions URISYS 2400
Exchange user-replaceable parts
4 Loosen the blue knurled locking screw at the top of the syringe assembly
that fixes the glass barrel of the syringe by turning it counterclockwise.
5 For the W-syringe, loosen and remove the locking screw completely to
prevent damage to the glass barrel.
6 Pull the plunger down sufficiently so that you can lift out the glass barrel.
Then remove the glass assembly.
7 Carefully pull the syringe assembly out of its mounting block. Support the
glass barrel with one hand; with the other hand, grasp the syringe holder,
lifting it up and out to remove the assembly.
Roche Diagnostics
D-10 Operator’s Manual · Version 1.0
URISYS 2400 7 General maintenance actions
Exchange user-replaceable parts
8 Remove the glass syringe barrel and remove the O-ring that fits below the
syringe barrel. Place both components in a safe place.
1 Turn the syringe holder upside down and expose the plunger retaining
screw.
2 Apply a spanner wrench to the plunger retaining screw and turn the screw
counterclockwise until it is loose.
3 Turn the syringe holder right side up and carefully lift the syringe holder
from the plunger. Leave the plunger intact.
Roche Diagnostics
Operator’s Manual · Version 1.0 D-11
7 General maintenance actions URISYS 2400
Exchange user-replaceable parts
1 Remove the seal piece, press piece, spring, and retaining screw from the
plunger. Discard the seal piece.
2 Wipe the plunger with a gauze square soaked in 70% isopropyl alcohol to
remove any debris, followed by a gauze square soaked in distilled or
deionized water.
3 Ensure that the retaining screw, spring and press piece are properly
positioned on the plunger. Place the new seal on the plunger and check
that the rounded end is facing downwards.
4 Place the plunger back in the syringe holder with the syringe vertically
oriented. Tighten the retaining screw with the spanner wrench until the
screw is snug with the syringe holder.
Roche Diagnostics
D-12 Operator’s Manual · Version 1.0
URISYS 2400 7 General maintenance actions
Exchange user-replaceable parts
E
F
G
H
Roche Diagnostics
Operator’s Manual · Version 1.0 D-13
7 General maintenance actions URISYS 2400
Exchange user-replaceable parts
1 Place the syringe holder O-ring into the syringe holder. If the O-ring
appears damaged or worn, replace it.
2 Inspect the glass syringe barrel for chips or cracks at the top and bottom. If
the syringe barrel is etched or damaged in any way, replace it. Ensure that
the barrel and the syringe holder are dry. Place the syringe barrel over the
plunger and onto the syringe holder.
3 Return the syringe holder onto its mounting block by tilting the top of the
syringe towards the instrument. Ensure that the syringe holder is in the
recess on the top surface of the mounting block.
If the syringe holder is not properly seated in its mounting block recess, damage to
the syringe assembly may occur.
4 Rotate the syringe holder until the side tube port is aligned to accept the
syringe side-tubing. Check if the blue syringe base correctly fits into the
well of the metal support plate.
5 Tighten the top knurled locking screw with your fingers. Ensure that the
syringe assembly is seated correctly and that the glass barrel O-ring is
properly positioned.
6 Reconnect the top tubing of the syringe and tighten the retaining nut with
your fingers.
7 Reconnect the side tubing and tighten the retaining nut with your fingers.
8 Turn the analyzer on.
Roche Diagnostics
D-14 Operator’s Manual · Version 1.0
URISYS 2400 7 General maintenance actions
Exchange user-replaceable parts
9 Perform “Air Purge”, if necessary more than once, to fill the syringe with
water and to get rid of any remaining air bubbles.
The top and side tubing retaining nuts can be damaged if excessive force is applied
when replacing them. Take care not to cross-thread these nuts.
Looseness of the retaining nut may result in incorrect data. Tighten the retaining nut
securely by hand while being careful about the packing orientation.
Use only fuses that are specified in this Operator Manual, (see “Technical
specifications” on page F-2). The use of other fuses is a breach of the warranty.
Roche Diagnostics
Operator’s Manual · Version 1.0 D-15
7 General maintenance actions URISYS 2400
Exchange user-replaceable parts
Roche Diagnostics
D-16 Operator’s Manual · Version 1.0
Troubleshooting E
Part E lists and describes the meaning of the alarm flags which are shown by
the URISYS 2400.
URISYS 2400 8 Data alarms (flags)
Table of contents
This chapter describes the data alarms that appear on printouts, and explains
how to handle alarms and reset the analyzer..
Roche Diagnostics
Operator’s Manual · Version 1.0 E-1
8 Data alarms (flags) URISYS 2400
Flags on result data printout
Appearance
Flag Data alarm Explanation
on printout
S Sieve Test parameter result corresponds After the value
to sieve criteria settings
* Abnormal Test parameter result corresponds Before the
to abnormal criteria settings value
O Specific gravity Result is greater than 1.050 or less Instead of SG
(SG) out of range than 1.000 value
I SG value invalid Refraction position is not definite Instead of SG
value
T SG error Hardware error Instead of SG
value
# Changed range Initial range table is changed After the value
value
! Edit Data have been edited After the value
Appearance
Flag Sample alarm Explanation
on printout
E Sample empty No sample detected Instead of
value
N Sample short Sample volume below limit Instead of
value
R Test strip Test strip error Instead of
value
Roche Diagnostics
E-2 Operator’s Manual · Version 1.0
URISYS 2400 8 Data alarms (flags)
Handle alarms
Handle alarms
The [Alarm monitor] screen displays up to four alarms.
When more than four messages are generated, press <Alarm Trace> on the
[Tools] screen to print the alarm codes and the times at which they occurred.
When the analyzer is switched on and in Stand-by status, any displayed alarm
must be checked, solved and cleared before next samples are processed.
If any errors have occurred during the setting of parameters, the message does
not appear until <Start> is pressed.
If the analyzer comes to an emergency stop due to an error, execute the reset
function on the [Maintenance] screen after resolving the alarm cause.
Alarm levels
The main alarm levels are WARNING, S.STOP, STOP, and E.STOP.
WARNING Although this warning is issued, the analyzer continues to measure samples.
After completion of the current sampling, check the alarm condition, and take
the action outlined.
S.STOP Only the sampling sequence is stopped. For samples in process, measurement
continues and the result data are generated. Check the alarm condition, take
the action outlined, and restart the operation.
STOP The analyzer stops at the end of its current, 15-second instrument cycle. For
samples in process, measurement, result data printing and host
communication stop immediately. Check the alarm condition, and take the
action outlined. Check the results for the last samples that were processed
before the alarm message appeared. Reset all samples that were processed but
not measured/ printed/ sent to host and repeat the processing of these
samples.
E.STOP The analyzer stops instantly (15-second instrument cycle is NOT finished).
For samples in process, measurement, result data printing and host
communication stop immediately. Check the alarm condition and take the
action outlined. Execute the reset function on the [Maintenance] screen after
resolving the alarm cause. Check the results for the last samples that were
processed before the alarm message appeared. Reset all samples that were
processed but not measured/ printed/ sent to host and repeat processing of
these samples.
Roche Diagnostics
Operator’s Manual · Version 1.0 E-3
8 Data alarms (flags) URISYS 2400
Handle alarms
Roche Diagnostics
E-4 Operator’s Manual · Version 1.0
URISYS 2400 8 Data alarms (flags)
Handle alarms
Alarm menu
Using the <Alarm Trace> button on the [Tools] screen, the operator can
perform the following procedures:
o Print a list of the alarms for that day (last alarm printed first).
o Print a list of cumulative alarms (current day printed first).
o Download alarm lists from the analyzer to a diskette.
o Clear alarms.
Roche Diagnostics
Operator’s Manual · Version 1.0 E-5
8 Data alarms (flags) URISYS 2400
Handle alarms
C D E F
| |
| DAILY ALARM TRACE 02.04.1999 09:51 |1
| |2
B | 09:15 7 10 |3
| A ***-***-*** ***-***-*** ***-***-*** ***-***-*** |4
| 09:12 2 1 |5
| R ***-***-*** |6
A | 09:10 43 1 |7
| A ***-***-*** ***-***-*** |8
| 08:59 2 1 |9
| R ***-***-*** |10
| 08:55 2 1 |11
| A ***-***-*** |12
| 08:52 2 1 |13
| A ***-***-*** |14
| |15
I H G
Figure E-1 Example of a printout of daily alarm list
A Identification flag (A: Alarm, R: Retry) D Generation frequency G Generation time in cycle time
(ms x 100)
B Generation time E Printing date
H Sub category code or retry frequency
C State of instrument F Printing time
I General category code or retry code
Roche Diagnostics
E-6 Operator’s Manual · Version 1.0
URISYS 2400 8 Data alarms (flags)
Handle alarms
D E F
| |
| CUMULATIVE ALARM TRACE 04.04.2001 17:02 |1
| |2
C | 04.04.2001 15:20 - |3
B | 15:56 |4
| A ***-***-*** ***-***-*** ***-***-*** ***-***-*** |5
A | ***-***-*** ***-***-*** ***-***-*** ***-***-*** |6
| R ***-*** ***-*** |7
| 03.04.2001 08:30 - 15:30 |8
| 11:08 |9
| A ***-***-*** ***-***-*** ***-***-*** ***-***-*** |10
| ***-***-*** |11
| R ***-*** |12
| |13
| |14
| |15
L K J I H G
Figure E-2 Example of a printout of cumulative alarm list
A Identification flag (A: Alarm, R: Retry) E Printing date I Generation time
B Generation time F Printing time J Power off time
C Alarm date G Generation frequency K Retry code
D Power on time H Sub category code L Retry frequency
Roche Diagnostics
Operator’s Manual · Version 1.0 E-7
8 Data alarms (flags) URISYS 2400
Handle alarms
a To clear alarms
o The maximum number of daily alarms that can be stored is 255 logs or 10 types.
The maximum number of cumulative alarms that can be stored is 255 logs or 20
types.
o Switching the power supply on or off is counted once.
o When the date is changed or the power supply is switched off, the daily alarm
log is cleared.
Roche Diagnostics
E-8 Operator’s Manual · Version 1.0
URISYS 2400 8 Data alarms (flags)
Reset the analyzer
Start
If the analyzer does not switch on at all, check that:
o The main power is on.
o The power cable is connected properly.
Contact Technical Support if the analyzer still does not respond.
Roche Diagnostics
Operator’s Manual · Version 1.0 E-9
8 Data alarms (flags) URISYS 2400
Reset the analyzer
Roche Diagnostics
E-10 Operator’s Manual · Version 1.0
URISYS 2400 9 Instrument alarms (messages)
Table of contents
Roche Diagnostics
Operator’s Manual · Version 1.0 E-11
9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
Alarms
Roche Diagnostics
E-12 Operator’s Manual · Version 1.0
URISYS 2400 9 Instrument alarms (messages)
Instrument alarms (messages)
Roche Diagnostics
Operator’s Manual · Version 1.0 E-13
9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
Roche Diagnostics
E-14 Operator’s Manual · Version 1.0
URISYS 2400 9 Instrument alarms (messages)
Instrument alarms (messages)
Roche Diagnostics
Operator’s Manual · Version 1.0 E-15
9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
Roche Diagnostics
E-16 Operator’s Manual · Version 1.0
URISYS 2400 9 Instrument alarms (messages)
Instrument alarms (messages)
Roche Diagnostics
Operator’s Manual · Version 1.0 E-17
9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
Roche Diagnostics
E-18 Operator’s Manual · Version 1.0
URISYS 2400 9 Instrument alarms (messages)
Instrument alarms (messages)
Roche Diagnostics
Operator’s Manual · Version 1.0 E-19
9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
Roche Diagnostics
E-20 Operator’s Manual · Version 1.0
URISYS 2400 9 Instrument alarms (messages)
Instrument alarms (messages)
Roche Diagnostics
Operator’s Manual · Version 1.0 E-21
9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
Roche Diagnostics
E-22 Operator’s Manual · Version 1.0
URISYS 2400 9 Instrument alarms (messages)
Instrument alarms (messages)
Roche Diagnostics
Operator’s Manual · Version 1.0 E-23
9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
Roche Diagnostics
E-24 Operator’s Manual · Version 1.0
Appendix F
Technical specifications
Roche Diagnostics
Operator’s Manual · Version 1.0 F-1
10 Technical specifications URISYS 2400
Technical specifications
Technical specifications
Measurement methods
Roche Diagnostics
F-2 Operator’s Manual · Version 1.0
URISYS 2400 10 Technical specifications
Barcode specifications
Barcode specifications
Item Specification
NW7 Modulus 16
Code 39 Modulus 43
Number of ID digits
Characters usable
NW7 0 to 9, -, /, $, +
Code 39 0 to 9, space, A to Z, -, ., /, $, %, +
ITF 0 to 9
If using an ITF barcode system, Roche/Hitachi recommends its use only with a
check digit function.
Roche Diagnostics
Operator’s Manual · Version 1.0 F-3
10 Technical specifications URISYS 2400
Barcode specifications
C
B
D
A
A B
C
N
D
M L E
K F
I G
J H
Figure F-2 Sample barcode labels
A Sample barcode standard size C 17 mm I 13 mm
B Sample barcode actual label size D 72 mm J 16 mm
E 50 mm K 5 mm
F 5 mm L 50 mm
G 13 mm M 60 mm
H 16 mm N 5 mm
Roche Diagnostics
F-4 Operator’s Manual · Version 1.0
URISYS 2400 10 Technical specifications
Concentration ranges
Concentration ranges
SG 1.000-1.050 1.000-1.050
PH 5.0 5.0
6.0 6.0
6.5 6.5
7.0 7.0
8.0 8.0
9.0 9.0
Roche Diagnostics
Operator’s Manual · Version 1.0 F-5
10 Technical specifications URISYS 2400
Concentration ranges
Roche Diagnostics
F-6 Operator’s Manual · Version 1.0
Glossary and index G
URISYS 2400 Glossary
Analyte - Flow cell
Glossary
CG EG
Calibration To adjust the reflectance photometer Expiration date For test strip cassettes, the date
to a gray standard (calibration strip) with known after which the contents must not be used in
reflectance values. analytical tests. Also applies to calibration strips and
Calibration strip Gray plastic strip used for to controls.
calibrating the photometer of the
URISYS 2400analyzer.
FG
Charge-coupled device (CCD) Photoelectrical
detector used in the URISYS 2400 to determine the Flow cell A component in the URISYS 2400
angle of total reflection of light for the measurement analyzer used to measure specific gravity and clarity.
of the specific gravity of the urine sample.
Check digit (C/D) Barcode verification number.
Clarity (CLA) Feature of urine samples
determined by measurement of the percentage value
of light transmission. Results can be CLEAR,
LIGHT TURBID, or TURBID.
Roche Diagnostics
Operator’s Manual · Version 1.0 G-1
Glossary URISYS 2400
Host computer - Qualitative determination
HG OG
Host computer A central computer that Operation status The analyzer processes samples
communicates with the URISYS 2400 receiving in this status. Additional samples can be loaded and
sample results from the analyzer. measured without an intermediate stop of the
Host connection The connection (via a cable) analyzer.
between the analyzer and the host computer. The
operator can either activate or deactivate this
connection, as required. PG
Parallel interface The interface (Centronics type)
through which the analyzer can be connected to an
IG external printer.
ITF A barcode type for sample tubes and racks that Parameters System and analysis configurations
can be read by the barcode reader. that can be set individually (and stored) according
to customer requirements if they deviate from the
default factory settings.
LG Parameter-specific reflectance values
Light-emitting diode (LED) A semiconductor Reflectance ranges stored in the analyzer that are
that emits light of a defined wavelength. used to convert the reflectance values of each test
pad into semi-quantitative concentration results.
Password-protected See Special functions.
MG
Photodiode detector A photometer component
Maintenance procedures Procedures that must that receives the reflected light.
be performed on a regular basis for example daily, Photometer An optical device (measuring head)
weekly, monthly, or every three months to secure that measures the intensity of reflected light of each
reliable operation of the analyzer. test pad.
Menus Parts of the software program that serve for Print mode Software function in the [Start
dedicated user operations, for example, setting of Condition] screen to select the category of results
parameters. (all, abnormal, sieve, or normal) to be printed in
Message (alarm) Information on the touch real time.
screen of sample or test strip errors, or reasons for
the analyzer stopping.
QG
Qualitative determination The measurement of
NG
analytes or features with descriptive (qualitative)
Nomogram Correlation function between result classifications such as negative/positive, clear/
quantities. In the URISYS 2400 analyzer, the turbid, or yellow, amber and so on.
refraction/gravity function used to convert the
measured refraction values into specific gravity
results.
NW7 A barcode type for sample tubes that can be
read by the barcode reader.
Roche Diagnostics
G-2 Operator’s Manual · Version 1.0
URISYS 2400 Glossary
Racks - TS transfer base
Roche Diagnostics
Operator’s Manual · Version 1.0 G-3
Glossary URISYS 2400
Urinalysis system - Water supply tank
UG
Urinalysis system A system consisting of an
analyzer and reagents (test strips), designed for the
chemical analysis of urine samples.
Urine analytes The constituents in the urine
sample whose concentrations are to be determined.
URISYS 2400 cassette An innovative cassette
manufactured by Roche Diagnostics for the most
convenient loading of test strips onto the
URISYS 2400 analyzer, supporting its fully
automated operations.
User interface screen See Touch screen.
WG
Water supply tank This tank is located externally
and supplies deionized water to the analyzer. It has a
maximum capacity of 5 liters (5.3 quarts). This is
enough to process approximately 1000 samples.
Roche Diagnostics
G-4 Operator’s Manual · Version 1.0
URISYS 2400 Index
Abbreviations - Dimensions
Index
Checks
AG
– barcode check, C-41
Abbreviations, viii – communication trace, C-42
Accessories, B-4 – compact memory, C-42
Alarms – FD utilities, C-43
– alarm levels, E-3 – printer check, C-41
– alarm list, E-12 – program check, C-41
– data alarms (flags), E-2 Clarity
– handling of, E-3 – measurement, A-4
– instrument alarms, E-11 – result ranges, F-6
– strip waste container, D-6 Color
– termination of analysis, C-12 – measurement, A-4
Analyzer memory – result ranges, F-6
– control samples, C-28 Communication parameters, B-26
– routine and STAT samples, C-21 Communication trace, C-42
Assembly, syringe, D-14 Compact memory, C-42
Compensation
– compensation of intrinsic urinary color, A-4, B-21
BG – compensation pad, A-4, C-39
Barcode Concentration ranges
– check, C-41 – range values, F-5
– settings, B-25 – setting default concentration ranges, B-23
– specifications, F-3 Concentration units, B-24
– types, B-25, F-3 Control samples
Bilirubin, concentration ranges, F-6 – display data, C-28
Biohazard, x – processing, C-17
– results, C-28
– search for data, C-29
CG – specification of, B-19
Cordless phone, xii
Calibration, C-36
– calibration strip, C-36
– photometer calibration, C-36 DG
– SG calibration, C-40
Cassette Data alarms, E-2
– first-time installation, B-30 Date/time setting, B-26
– handling precautions, B-32 Default setting, B-23
– on-board stability, B-30, C-19 Deletion of result data
– re-inserting a used cassette, C-20 – control sample data, C-33
– replacement of an empty cassette, C-18 – routine and STAT samples, C-27
– replacement of an expired cassette, C-19 Detector, measurement methods, A-3
– unpacking the cassette, B-30 Detector, water supply tank, B-7
Detergent, D-6, D-7
Dimensions, F-2
Roche Diagnostics
Operator’s Manual · Version 1.0 G-5
Index URISYS 2400
Diskette - Intrinsic urinary color
Diskette
HG
– download alarm list, E-8
– formatting, C-43 Hardware
– load and save parameter settings, B-12, B-28 – additional items, B-5
– software installation disk, B-4, B-12 – basic requirements, B-5
– write/read control sample data, C-32 – environmental conditions, B-10
– write/read routine or STAT sample data, C-26 – hardware description, B-2
Diskette drive, B-2 – installation, B-5
Display of result data – rack trays, B-4
– control samples, C-28 – standard accessories, B-4
– routine and STAT samples, C-22 – URISYS 2400 analyzer, B-2
– waste container, B-8
– water supply, B-6
EG Host
– activating/deactivating the connection to, C-7
Edit
– communication, B-26, C-26, C-32
– routine and STAT samples, C-24
– connection, B-9, C-7
Electromagnetic compatibility, ii, xii
– interface, B-5
Emergency sample, C-14
– sending control sample data to, C-32
Emission, xii
– sending routine or STAT sample data to, C-26
Empty liquid waste tank, D-6
Empty strip waste container, D-6
Environmental conditions, B-10 IG
Erythrocytes, concentration ranges, F-6
Exchanging fuses, D-15 ID numbers
Expiration date of cassette, B-32 – routine samples, input of, C-23
– STAT samples, input of, C-15, C-23
Incubation time, A-2, A-4, F-2
FG Installation
– assembling the parts, B-7
FD utilities, C-43
– basic requirements, B-5
Flow cell, A-4
– cassette, B-30
– air purge, D-7
– external host, B-9
– cleaning, D-4
– external printer, B-9
Fuses, B-4
– hardware, B-2, B-5
– liquid waste tank, B-8
– software, B-12
GG
– TS transfer base, B-8
Glossary, G-1 – water supply tank, B-7
Glucose, concentration ranges, F-5 Interfaces
– communication parameters, B-26
– host connection, B-9
– printer connection, B-9
Intrinsic urinary color, A-4, B-21
Roche Diagnostics
G-6 Operator’s Manual · Version 1.0
URISYS 2400 Index
Ketones - Power supply
KG NG
Ketones, concentration ranges, F-5 Nitrite, result ranges, F-5
Nomogram, A-4
LG
OG
LED (light emitting diodes), A-2, F-2
Leukocytes, concentration ranges, F-5 Operation
Level detection – analysis of results, C-21
– liquid waste tank, B-5, B-8 – automatic termination, C-12
– samples, alarms, E-2 – daily operation, C-1
– water supply tank, B-4, B-7 – display the result data, C-22, C-28
Liquid waste tank – documentation of results, C-21
– detector, B-8 – editing result data, C-24
– emptying, D-6 – host on/off, C-7
– setting up, B-8 – initial preparations, C-3
– interrupting analysis, C-11
– preparation of sample racks and tubes, C-8
MG – presentation of results on the printout, C-25, C-30
– printing, C-24, C-30
Maintenance
– processing additional samples, C-13
– air purge, D-7, D-15
– processing control samples, C-17
– cleaning of outer parts, D-7
– processing samples, C-10
– cleaning of TS transfer base, D-6
– processing STAT (emergency) samples, C-14
– daily, D-2, D-4
– removing processed samples, C-12
– emptying liquid waste tank, D-6
– sample preparation, C-8
– emptying strip waste container, D-6
– start conditions before sampling, C-5
– every three months, D-3
– starting analysis, C-10
– exchanging main fuses, D-15
– status type, B-13
– exchanging user-replaceable parts, D-8
– switching off the analyzer, C-34
– general maintenance actions, D-1
– system start, C-4
– maintenance menu, D-4
– termination due to an alarm, C-12
– maintenance procedures, D-2
– termination of analysis, C-11
– monthly, D-3
– replacing syringe seals, D-8
– washing solution, D-4, D-5
PG
– washing the liquid handling system, D-4
– weekly, D-3 Parameters
Measuring head, A-2, F-2 – parameter media, B-28
Measuring principles, A-2 – system parameter setting, B-24
Microprocessor, A-3 – test parameter setting, B-17
Mobile phone, xii Password
– changing the password, B-27
– password protected functions, B-27
Peripheral device connection, xi
Photodiode detector, A-3
Photometer calibration, C-36
pH, result ranges, F-5
Pipetting mechanism, A-2, B-3
Power supply, B-6
Roche Diagnostics
Operator’s Manual · Version 1.0 G-7
Index URISYS 2400
Print mode - Termination of analysis
Roche Diagnostics
G-8 Operator’s Manual · Version 1.0
URISYS 2400 Index
Test parameters - Wavelengths
Test parameters
WG
– concentration ranges, F-5
– printout, B-28 Warning label, x
– setting of, B-17 Warranty, ii, iii, D-15
– wavelengths, A-4 Wash rack
Test strips, A-2, B-2, B-30 – registering the wash rack, B-27
Toggle function, B-15 – using on the input line, D-4
Touch screen Washing procedure
– basic screen composition, B-13 – detergent, D-4
– main buttons, B-14 – liquid handling system, D-4
– window, B-15 Waste container, see Strip waste container
Transceiver, xii Waste liquid tank, see Liquid waste tank
Tray, B-4, C-3, C-9 Waste liquid, disposal of, D-6
Trouble shooting Water filter, maintenance, D-3
– alarm levels, E-3 Water supply tank
– alarm log, E-4, E-8 – liquid level detector, B-7
– clearing alarms, E-8 – setting up, B-7
– data alarms (flags), E-1 Wavelengths, A-2, A-4
– FD write, E-8
– handling alarms, E-3
– instrument alarms, E-12
– printing cumulative alarm, E-7
– printing daily alarm, E-6
– switch the alarm buzzer on and off, E-4
– test strip empty alarm, C-18, E-14
TS cassette, see Cassette
TS transfer base
– cleaning, D-6
– setting up, B-8
UG
Units
– arbitrary units, B-22
– selecting concentration units, B-24
Urinalysis, A-2
Urine analytes, A-2
Urine sample, see Samples
Urine test parameters, see Test parameters
Urine test strips, see Test strips
URISYS 2400 analyzer, A-2, B-3
URISYS 2400 cassette, see Cassette
Urobilinogen, concentration ranges, F-5
Used test strips, D-2
User interface screen, see Touch screen
Roche Diagnostics
Operator’s Manual · Version 1.0 G-9
Index URISYS 2400
Test parameters - Wavelengths
Roche Diagnostics
G-10 Operator’s Manual · Version 1.0