Attach Electrodes To Device Peel Off Plastic From Electrodes

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Short manual

1 Attach electrodes to device 2 Peel off plastic from electrodes

3 Turn on device
Press button

4 Attach device to patient


- Slow green flashes indicate C3 is correctly attached and recording Please note:
- The C3 will keep recording for 24h - 48h C3 is NOT
waterproof

Place C3 on midsternal line


Please note:
In case of chest hair,
shaving may be required

Support: Website: www.cortrium.com I Email: [email protected]


Cortrium C3 Holter Monitor - Short Manual - Document Version: 0.4 EN - Date: 2018-09-04
Additional instructions
A Device notification lights
Green flashes
- Fast: ON / Event button pressed
- Slow: C3 is recording

Blue flashes
- C3 does not have proper
contact to the body

Yellow flashes
- Slow: Low battery
- Fast: Non-critical Error (Consult Manual)

White flashes
- Critical Error (Contact Supplier)

B Contents of package NB Discomfort!


If sensing symptoms
or discomfort, press button
- Button will flash green
quickly for 10 seconds.

Cortrium C3 unit USB-cable

Instructions for Use

C Charging the C3
Please note:
Charge for at least
one hour

USB Plug
Micro
USB socket
USB Cable

Main power outlet


The back of the C3

Support: Website: www.cortrium.com I Email: [email protected]


Cortrium C3 Holter Monitor - Short Manual - Document Version: 0.4 EN - Date: 2018-09-04