0% found this document useful (0 votes)
281 views

ITP

The inspection and test plan outlines 15 points of acceptance criteria for pressure gauges with diaphragm seal assemblies. Key points include material certification, weld inspection, dimensional checks, accuracy testing, and environmental protection certification. The plan specifies documents to reference for each test point and identifies roles for WIKA, NPCC, and ADMA companies to hold, witness, or approve tests.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
281 views

ITP

The inspection and test plan outlines 15 points of acceptance criteria for pressure gauges with diaphragm seal assemblies. Key points include material certification, weld inspection, dimensional checks, accuracy testing, and environmental protection certification. The plan specifies documents to reference for each test point and identifies roles for WIKA, NPCC, and ADMA companies to hold, witness, or approve tests.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

INSPECTION AND TEST PLAN

Customer : National Petroleum Construction Company Job No.: SO042459

Customer P.O No.: 2207-2013-31713-AC Date: 05.06.2018

Description : Pressure Gauges with Diaphragm Seal Assembly Page No.: 2 of 2


WIKA NPCC / TPI ADMA/TPA
Point Acceptance
Activity Ref. Documents Project Spec
No. Criteria H W R H W R H W R

Receipt Inspection of existing Pressure Gauge prior to taking up


1 PO - - X
modification
Material Test Certificate EN10204 3.1
2 3.1 for Pressure Retaining Parts MTC 3.1 EN10204 X X X
2.2 for Non-Pressure parts
3 NACE MR0175 Compliance Certificate Compliance Certificate MR0175 X X X
PR: 2207-13-
4 Review of WPS & PQR 2207-31713A-X007-001 31713 - X X X
Hydro Test Report / Over Pressure Protection; (As per
5 2207-31713A-X003-001 ASME B16.5 X X X
approved Hydrotest procedure)
6 LPI Test Report of the Welds on Diaphragm Seal WIKA Standard Procedure - X X X
AD162-562-G-
7 PMI based on Heat Numbers (As per approved PMI Procedure) 2207-31713A-X008-001 00009 Rev.0 ASME Standard X X X
PMI Spec
Visual Inspection; (100%-W: WIKA; 10%-W: ADMA/NPCC) 2207-31713A-J005-001
8 Marking, Tagging, Identification & Packing, Type, size, Model WIKA Standard X X X
2207-31713A-J036-001
Number, Mounting, diaphragm seal.
Dimensional Inspection; (100%-W: WIKA; 10%-W: 2207-31713A-J005-001
9 2207-31713A-J036-001
Approved GAD X X X
ADMA/NPCC)
Function Test including check of Accuracy (Calibration Test),, PR: 2207-13- Approved Calibration
10 2207-31713A-J009-001 31713 Procedure
X X X
Repeatability: (100%-W: WIKA; 10%-W: ADMA/NPCC)
11 Painting (Epoxy Coating) on Gauge Case (Visual Check only) WIKA Standard Procedure Epoxy Coating X X X
Check of Identification, Packing and Marking (Name Plate & 2207-31713A-J005-001
12 2207-31713A-J036-001
X X X
Tagging)
13 IP 66 Certificate X X X
14 Inspection Release Note - X X X
15 Review of MRB 2207-31713A-A012-001 - X X X

H: Hold R: Review W: Witness A: Approval

eyadc-932217-932018165143_review.pdf 3 5/23/2019 11:48:32 AM