Invacare Comet Service Manual: Edition: 28.03.12
Invacare Comet Service Manual: Edition: 28.03.12
Invacare Comet Service Manual: Edition: 28.03.12
SERVICE MANUAL
Edition: 28.03.12
SERVICE MANUAL Invacare ® - Comet
Service addresses
2
Invacare ® - Comet SERVICE MANUAL
3
SERVICE MANUAL Invacare ® - Comet
Tillverkare: MÖLNDAL
Deutschland
LANDSKRONA
(: +46 (0)418 2 85 40
Fax: +46 (0)418 1 80 89
@: [email protected]
OSKARSHAMN
(: +46 (0)491 1 01 40
Fax: +46 (0)491 1 01 80
@: [email protected]
4
Invacare ® - Comet SERVICE MANUAL
Table of Contents
Chapter Page
1 INTRODUCTION 7
1.3 Definition and representation of information and safety information in this manual 8
3 TIGHTENING TORQUES 13
6 OPERATIONAL FAULTS 17
7 REPAIR WORK 20
5
SERVICE MANUAL Invacare ® - Comet
8 ACCESSORIES 63
6
Invacare ® - Comet SERVICE MANUAL
1 Introduction
1.1 General information
· Service and maintenance work must be carried out taking this service manual into account.
· It is imperative that you observe safety information.
· Information about operation or about general maintenance and care work on the mobility aid
should be taken from the operating manual.
· You can find information about ordering spare parts in the spare parts catalogue.
· Only use original Invacare® spare parts. The guarantee will become invalid if other spare
parts are used!
· We reserve the right to make any alterations on the grounds of technical improvements.
· The mobility aid may only be maintained and overhauled by qualified personnel.
· The minimum requirement for service technicians is suitable training, such as in the cycle or
orthopaedic mechanics fields, or sufficiently long-term job experience.
- Experience in the use of electrical measuring equipment (multimeters) is also a requirement.
- Special Invacare® training is recommended.
· Alterations to the mobility aid which occur as a result of incorrectly or improperly executed
maintenance or overhaul work lead to the exclusion of all liability on the side of INVACARE.
· If you have any problems or questions please contact Invacare® Service.
7
SERVICE MANUAL Invacare ® - Comet
HAZARD!
The signal word "HAZARD!" refers to immediate hazards.
· The following lines in italics refer to actions which serve to avoid such hazards.
WARNING!
The signal word "WARNING!" refers to possibly-occurring hazards which can lead to
death or serious injuries if they are not avoided.
· The following lines in italics refer to actions which serve to avoid such hazards.
ATTENTION!
The signal word "ATTENTION!" refers to possibly-occurring hazards which can lead to
minor injuries and/or material damage if they are not avoided.
· The following lines in italics refer to actions which serve to avoid such hazards.
CAUTION!
The signal word "CAUTION!" refers to hazards which could lead to material damage if
they are not avoided.
· The following lines in italics refer to actions which serve to avoid such hazards.
Note
The signal word "Note" is used to denote general information which simplifies the handling of
your product and refers to special functions.
8
Invacare ® - Comet SERVICE MANUAL
General hazards
This symbol warns you of general hazards!
· Always follow the instructions to avoid injury to the user or damage to the product!
BURN HAZARD!
This symbol warns you of the danger of chemical burns, for example due to the discharge
of battery acids!
· Always follow the instructions to avoid injury to the user or damage to the product!
DANGER OF CRUSHING!
This symbol warns you of crushing hazards due to inattentive working with heavy
components.
· Always follow the instructions to avoid injury to the user or damage to the product!
EXPLOSION HAZARD!
This symbol warns you of an explosion hazard, which can be caused by excessive tyre
pressure in a pneumatic tyre.
· Always follow the instructions to avoid injury to the user or damage to the product!
Note
This symbol identifies general information which is intended to simplify working with your product
and which refers to special functions.
Requirements:
· This symbol identifies a list of various tools, components and items which you will need in
order to carry out certain work. Please do not attempt to carry out the work if you do not
have the listed tools available.
9
SERVICE MANUAL Invacare ® - Comet
10
Invacare ® - Comet SERVICE MANUAL
Safety shoes
The mobility device, and some of its components, are very heavy. These parts can result
in injuries to the feet if they are allowed to drop.
· Wear standardised safety shoes during all work.
Eye protection
It is possible that battery acid can be discharged when working on defective batteries or
when handling batteries improperly.
· Always wear eye protection when working on any defective or possibly defective batteries.
Safety gloves
It is possible that battery acid can be discharged when working on defective batteries or
when handling batteries improperly.
· Always wear acid-proof safety gloves when working on any defective or possibly defective
batteries.
WARNING!
Injury hazard if the vehicle starts moving unintentionally during repair work!
· Switch the power supply off (ON/OFF key)!
· Engage the drive!
· Before raising the vehicle, secure the wheels by blocking them with wedges!
ATTENTION!
Fire and burn hazard due to electrical short-circuit!
· The mobility device must be completely switched off before removal of voltage-carrying
components! To do this, remove the batteries.
· Avoid short-circuiting the contacts when carrying out measurements on voltage-carrying
components!
11
SERVICE MANUAL Invacare ® - Comet
CAUTION!
Danger of burns from hot surfaces on the motor!
· Allow the motors to cool down before commencing work on them.
ATTENTION!
Injury hazard and danger of damage to vehicle due to improper or incomplete
maintenance work!
· Use only undamaged tools in good condition.
· Some moving parts are mounted in sockets with PTFE coating (Teflon™). Never grease
these sockets!
· Never use "normal" nuts instead of self-locking nuts.
· Always use correctly-dimensioned washers and spacers
· When reassembling, always replace any cable ties which were cut during dismantling.
· After completing your work / before renewed start-up of the mobility device, check all
connections for tight fitting.
· After completing your work / before renewed start-up of the mobility device, check all parts for
correct locking.
· Only operate the vehicle with the approved tyre pressures (see technical data).
· Check all electrical components for correct function. Please note that incorrect polarity can
result in damage to the electronic.
· Always carry out a trial run at the end of your work.
Note
Mark all current settings for the mobility aid (seat, armrests, backrest etc.), and the associated
cable connecting plugs, before dismantling. This makes reassembly easier.
All plugs are fitted with mechanical safety devices which prevent release of the connecting plugs
during operation. To release the connecting plugs the safety devices must be pressed in. When
reassembling ensure that these safety devices are correctly engaged.
WARNING!
Any changes to the drive program can affect the driving characteristics and the tipping
stability of the vehicle!
· Changes to the drive program may only be carried out by trained Invacare® specialist
dealers!
· Invacare® supplies all mobility aids with a standard drive program ex-works. Invacare® can
only give a warranty for safe vehicle driving behaviour - especially tipping stability - for this
standard drive program!
12
Invacare ® - Comet SERVICE MANUAL
3 Tightening torques
The tightening torques stated in the following list are based on the thread diameter for the nuts and
bolts for which no specific values have been determined. All values assume dry and de-greased
threads.
CAUTION!
Damage can be caused to the mobility device due to improperly tightened screws, nuts or
plastic connections.
· Always tighten screws, nuts etc to the stated tightening torque.
· Only tighten screws or nuts which are not listed here fingertight.
13
SERVICE MANUAL Invacare ® - Comet
DS112KB02 110 A
DS162KD01 160 A
DS120 120 A
DS160 160 A
14
Invacare ® - Comet SERVICE MANUAL
15
SERVICE MANUAL Invacare ® - Comet
16
Invacare ® - Comet SERVICE MANUAL
6 Operational Faults
6.1 Operational faults
Please proceed as follows if you have any problems:
· First assess the possible cause of the problem using the following table.
· Check the drive console status display. Evaluate the blink error code.
· Carry out the necessary checks and repairs as recommended in the following table.
17
SERVICE MANUAL Invacare ® - Comet
3 Battery voltage too Stops driving · The battery voltage is too high. If the
high battery charger is connected, disconnect
it from the Scooter.
· The electronic system charges the
batteries when running downhill and
when braking. This fault is caused when
the battery voltage becomes too high
during this process. Switch the Scooter
off and on again.
18
Invacare ® - Comet SERVICE MANUAL
19
SERVICE MANUAL Invacare ® - Comet
7 Repair Work
7.1 General warning information on installation work
CAUTION!
Risk of damage to the vehicle! Collisions can be caused if shim rings are removed from
the drive wheels during installation work!
Shim rings are frequently placed between the drive shaft and the wheel hub to compensate
tolerances. Collisions can be caused if these shim rings are removed and not re-installed!
· Install all shim rings in exactly the same positions they were in before dismantling.
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
NOTE:
The plugs on the electronic module cannot be wrongly connected because all plugs have a
different size and only fit in one socket.
Requirements:
· Open spanner, 10 mm.
· Open spanner, 17 mm.
· oblique pliers
· Rubber hammer
· Phillips screwdriver
· Supporting wooden block
20
Invacare ® - Comet SERVICE MANUAL
21
SERVICE MANUAL Invacare ® - Comet
Rhino1:
DS112KB02
DS162KD01
Rhino 2:
DS120:
DS160
22
Invacare ® - Comet SERVICE MANUAL
23
SERVICE MANUAL Invacare ® - Comet
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
Requirements:
· Open spanner, 32 mm.
· Phillips screwdriver
· Supporting wooden block
24
Invacare ® - Comet SERVICE MANUAL
25
SERVICE MANUAL Invacare ® - Comet
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
NOTE:
The drive motor needs to be removed before you can replace the left-hand shock absorber. If the
right-hand shock absorber is to be replaced, you do not need to remove the drive motor.
Requirements:
· Open spanner, 10 mm.
· Open spanner, 17 mm.
· Phillips screwdriver
· 2 Allen keys 5m
· Supporting wooden block
26
Invacare ® - Comet SERVICE MANUAL
27
SERVICE MANUAL Invacare ® - Comet
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
NOTE:
The front axle can only be replaced on the 3-wheel version. The 4-wheel version does not have a
replaceable axle.
Requirements:
· Open spanner, 17 mm.
· Open spanner, 19 mm.
· Supporting wooden block
28
Invacare ® - Comet SERVICE MANUAL
29
SERVICE MANUAL Invacare ® - Comet
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
Requirements:
· Open spanner, 10 mm.
· Open spanner, 17 mm.
· Open spanner, 19 mm.
· Open spanner, 32 mm.
· Phillips screwdriver
· Supporting wooden block
30
Invacare ® - Comet SERVICE MANUAL
31
SERVICE MANUAL Invacare ® - Comet
32
Invacare ® - Comet SERVICE MANUAL
33
SERVICE MANUAL Invacare ® - Comet
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
NOTE:
The front axle can only be replaced on the 3-wheel version. The 4-wheel version does not have a
replaceable axle.
Requirements:
· Open spanner, 12 mm.
· Open spanner, 13 mm.
· Philips screwdriver
· Allen key 5 mm
· Supporting wooden block
34
Invacare ® - Comet SERVICE MANUAL
3-wheel version:
· Slacken brake cable (see chapter entitled
"Replacing brake cable" on page 37).
· Disconnect the brake cable at the wheel (1).
· Remove screws (3) on spacer (2).
· Remove screws (6) on shock absorber (5).
· Loosen axle bolts (4) simultaneously on both
sides.
· Pull front axle slowly out.
4-wheel version:
· Slacken both brake cables.
· Disconnect brake cable at affected wheel.
· Loosen wheel bolt.
· Pull wheel and brake off shaft.
35
SERVICE MANUAL Invacare ® - Comet
36
Invacare ® - Comet SERVICE MANUAL
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
Requirements:
· Open spanner, 10 mm.
· Phillips screwdriver
37
SERVICE MANUAL Invacare ® - Comet
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
Requirements:
· Open spanner, 14 mm.
· Phillips screwdriver
Replacing lifter:
· Remove batteries.
· Disconnect lifter cable at wiring harness
connecting plug in motor enclosure.
· Remove screws (1) on lifter (2).
· Replace lifter.
38
Invacare ® - Comet SERVICE MANUAL
39
SERVICE MANUAL Invacare ® - Comet
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
Requirements:
· Open spanner, 10 mm.
· Open spanner, 12 mm.
· Open spanner, 13 mm.
· Phillips screwdriver
· Circlip pliers
· Supporting wooden block
40
Invacare ® - Comet SERVICE MANUAL
41
SERVICE MANUAL Invacare ® - Comet
Note:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
Note
The following seat support tubes are available for the seat suspension:
Requirements:
· 2 open spanners, 17 mm
42
Invacare ® - Comet SERVICE MANUAL
43
SERVICE MANUAL Invacare ® - Comet
ATTENTION!
Burn and scald hazard if power cable is short-circuited!
· Turn off the power completely before removing any power supply components of the electric
vehicle! Therefore take out the batteries.
· Avoid a bridging of the contacts during measurements on live electrical components!
NOTE:
The plugs on the operating console cannot be wrongly connected because all plugs have a
different size and only fit in one socket.
Requirements:
· Phillips screwdriver
44
Invacare ® - Comet SERVICE MANUAL
45
SERVICE MANUAL Invacare ® - Comet
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
Requirements:
· Phillips screwdriver
· Open spanner, 32 mm.
46
Invacare ® - Comet SERVICE MANUAL
Adjusting:
· Set steering to straight ahead.
· Align sensor straight using regulating unit.
· Take hold of the sensor and tighten the fork
screw.
Check:
· Turn the steering completely to the right and
then the left.
· When the steering is turned to its limit, the 3
regulating unit switches should be visible.
47
SERVICE MANUAL Invacare ® - Comet
ATTENTION!
Burn and scald hazard if power cable is short-circuited!
· Turn off the power completely before removing any power supply components of the electric
vehicle! Therefore take out the batteries.
· Avoid a bridging of the contacts during measurements on live electrical components!
Requirements:
· Phillips screwdriver
· Digital multimeter
· Allen key 2m
48
Invacare ® - Comet SERVICE MANUAL
49
SERVICE MANUAL Invacare ® - Comet
50
Invacare ® - Comet SERVICE MANUAL
NOTE:
You can replace the headlight without removing the chassis. If, however, the working space is too
small, you can remove the chassis as described in chapter entitled "Replacing steering
column/fork 3-wheel" on page 30.
Requirements:
· Phillips screwdriver
3-wheel version
· Remove screw (1) on front headlight under
front cladding.
· Disconnect direction indicator cable (2) at
wiring harness behind steering column
cladding.
51
SERVICE MANUAL Invacare ® - Comet
4-wheel version
· Remove screws (1) on front headlight under
front cladding.
· Disconnect direction indicator cable (2) at
wiring harness behind steering column
cladding.
· Remove headlight.
· Remove screws (1) on headlight glass.
· Remove headlight glass.
· Replace bulb (2) in headlight.
52
Invacare ® - Comet SERVICE MANUAL
Requirements:
· Phillips screwdriver.
3-wheel version
· Remove direction indicator screw (1).
· Disconnect direction indicator cable (2) from
wiring harness.
· Replace direction indicator.
· Reassembly of the direction indicators takes
place in reverse order
· Check functions.
4-wheel version
· Remove direction indicator screw (1) on
housing.
· Disconnect direction indicator cable (2) from
wiring harness.
· Remove screws (3) on indicator glass.
· Remove indicator glass.
· Loosen screws (4) on orange-coloured cap
(5) and remove cap.
· Replace lightbulb.
· Reassembly of the direction indicators takes
place in reverse order
· Check functions.
53
SERVICE MANUAL Invacare ® - Comet
ATTENTION!
Burn and scald hazard if power cable is short-circuited!
· Turn off the power completely before removing any power supply components of the electric
vehicle! Therefore take out the batteries.
· Avoid a bridging of the contacts during measurements on live electrical components!
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
Requirements:
· Phillips screwdriver
54
Invacare ® - Comet SERVICE MANUAL
55
SERVICE MANUAL Invacare ® - Comet
ATTENTION!
Burn and scald hazard if power cable is short-circuited!
· Turn off the power completely before removing any power supply components of the electric
vehicle! Therefore take out the batteries.
· Avoid a bridging of the contacts during measurements on live electrical components!
NOTE:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
Requirements:
· Phillips screwdriver
56
Invacare ® - Comet SERVICE MANUAL
57
SERVICE MANUAL Invacare ® - Comet
ATTENTION!
Burn and scald hazard if power cable is short-circuited!
· Turn off the power completely before removing any power supply components of the electric
vehicle! Therefore take out the batteries.
· Avoid a bridging of the contacts during measurements on live electrical components!
WARNING!
Any changes to the drive program can affect the driving characteristics and the tipping
stability of the mobility aid!
· Changes to the drive program may only be carried out by trained Invacare® specialist
dealers!
· Invacare® can only give a warranty for safe mobility aid driving behaviour - especially tipping
stability - for unaltered standard drive programs!
Note:
The electronic system is supplied with a standard drive program. If you have carried out
customer-specific modifications to the drive program, you will have to make these changes again
after installing the new electronic module.
Note:
When disassembling, note the position of small parts such as screws and washers. Put small
parts down so that they can be reassembled in the right sequence.
Note:
The plugs on the electronic module cannot be wrongly connected because all plugs have a
different size and only fit in one socket.
Requirements:
· Open spanner, 10 mm.
· Open spanner, 17 mm.
· Rubber hammer
· Phillips screwdriver
· Supporting wooden block
· To adapt the drive program: programming software or hand programming device and
electronic system installation manual, available from Invacare®.
58
Invacare ® - Comet SERVICE MANUAL
59
SERVICE MANUAL Invacare ® - Comet
Rhino1:
DS112KB02
DS162KD01
Rhino 2:
DS120:
DS160:
60
Invacare ® - Comet SERVICE MANUAL
61
SERVICE MANUAL Invacare ® - Comet
NOTE:
The electronic system is supplied with a standard drive program. If you have carried out
customer-specific modifications to the drive program, you will have to make these changes again
after installing the new electronic module.
WARNING!
Any changes to the drive program can affect the driving characteristics and the tipping
stability of the mobility aid!
· Changes to the drive program may only be carried out by trained Invacare® specialist
dealers!
· Invacare® can only give a warranty for safe mobility aid driving behaviour - especially tipping
stability - for unaltered standard drive programs!
Requirements:
· Dynamic® Wizard software
· Operating manual for Wizard software
· For further information on other requirements - such as the minimum system configuration of
the PC to be used for programming, necessary programming cables - see the user manual of
the Wizard software. You find the latest version of the user manual in the download area on
http://www.dynamiccontrols.com/.
62
Invacare ® - Comet SERVICE MANUAL
8 Accessories
8.1 Installing the windshield
Requirements:
· 13 mm open-ended wrench
· 14 mm open-ended wrench
63
SERVICE MANUAL Invacare ® - Comet
64