Dynamic Mechanical Analyzer Operator's Manual: Thermal Analysis & Rheology AS W C
Dynamic Mechanical Analyzer Operator's Manual: Thermal Analysis & Rheology AS W C
Dynamic Mechanical Analyzer Operator's Manual: Thermal Analysis & Rheology AS W C
DMA 2980
Operator’s Manual
Notice
Notes, Cautions,
and Warnings .............................................. xvi
Chapter 1:
Introducing the DMA 2980 ...................... 1-1
Introduction................................................. 1-3
Specifications............................................ 1-18
Chapter 2:
Installing the DMA 2980 .......................... 2-1
Unpacking/Repacking
the DMA ..................................................... 2-6
Connecting Cables
and Gas Lines ........................................... 2-10
Chapter 3:
Calibrating the DMA 2980 ....................... 3-1
Introduction................................................. 3-3
Clamp Mass
Calibration (Step 1) .............................. 3-7
Force Calibration
(Balance and Weight) ......................... 3-14
Performing
Temperature Calibration .................... 3-22
Dynamic Temperature
Calibration Procedure ........................ 3-23
Chapter 4:
Running an Experiment ........................... 4-1
Introduction................................................. 4-5
Aligning the
Thermocouples ......................................... 4-10
Selecting the
Operating Mode ........................................ 4-12
VI TA INSTRUMENTS DMA 2980
DMA Multifrequency Mode ..................... 4-14
Creating a DMA
Multifrequency Method ..................... 4-21
Creating a DMA
Multistrain Method ............................ 4-26
Creating a DMA
Creep Method ..................................... 4-31
Creating a DMA
Controlled Force Method ................... 4-46
Temperature-Based
Experiments ................................. 4-46
Time-Based Experiments
(Manual Creep Measurements) ... 4-47
Force Ramp Experiments ............ 4-48
Measuring
Sample Length ................................... 4-58
Using the Telescoping Gauge ...... 4-59
Introduction................................................. 5-5
Tensioning/Nontensioning
Clamps ........................................................ 5-6
3-Point Bending
Sample Preparation ............................ 5-18
Mounting a Liquid
or Gel Sample ..................................... 5-34
Operating Range
of the Tension Clamps ........................ 5-62
Introduction................................................. 6-3
Comparison to
Other Techniques ................................. 6-4
Dynamic Mechanical
Analysis Testing ................................. 6-11
Calculations Based
on Clamp Type .......................................... 6-20
Clamping Factors
(Compression Clamps Only) .................... 6-36
Introduction................................................. 7-3
Index............................................................ I-1
TA INSTRUMENTS DMA 2980 XV
Notes, Cautions,
and Warnings
Safety:
EN 60110-1/1993 and A2/1995 Installation
Category II
EN 61010-2-010/1994
Emissions:
EN 55022: 1995, CISPR 22:1993
Class B (30-1000 MHz) Radiated
EN 55022: 1995, CISPR 22:1993
Class B (0.15-30 MHz) Conducted
Immunity:
EN 50082-1: 1992, Electromagnetic
CompatibilityGeneric immunity standard
Part 1. Residential, commercial, and light
industry.
IEC 801-2: 1991, 8 kV air discharge,
direct. No change of state.
IEC 801-3: 1984, 27-500MHz, 3V/m.
No response above 0.5°C temperature.
IEC 801-4: 1988 Fast transients com-
mon mode 1kV AC power. No change
of state.
TA INSTRUMENTS DMA 2980 XVII
Instrument Symbols
The following labels are displayed on the DMA
2980 instrument for your protection:
Symbol Explanation
Thermal Safety
During an experiment, the furnace, sample, and
clamp can become very hot or very cold to the
touch.
Potential Asphyxiant
Sample Decomposition
The DMA 2980 is capable of heating samples to
600°C. Many materials may decompose during
the heating, which can generate hazardous
byproducts.
Introduction
The TA Instruments Dynamic Mechanical
Analyzer (DMA) 2980 is an analytical instru-
ment used to test the physical properties of many
different materials. The sample is mounted on a
clamp and then subjected to changes in stress/
strain while undergoing a change in temperature.
The DMA measures the modulus (stiffness) and
damping (energy dissipation) of the sample.
Components
Your instrument consists of two major parts, the
DMA cabinet and the DMA assembly (see
Figure 1.1 on the next page). The following
components make up the DMA assembly:
Instrument Instrument
Clamp Display Keypad
Assembly
Sample
Thermocouple
Mechanical Cabinet
Section
Enclosure
Figure 1.1
The DMA 2980
Instrument
* CHROMEL® and ALUMEL® are registered trademarks
of the Hoskins Manufacturing Company.
Standby 23.25°C
DimChg 0.135µm
▲ ▲
ZERO FLOAT CLAMP FURNACE
MEASURE
CLAMP LOCK ▼ ▼
Figure 1.2
DMA 2980
Keypad and Display
(table continued)
Table 1.1
DMA 2980 Keypad
Functions
(continued)
Key/Function Explanation
STOP If an experiment is
running, this key ends the
method normally, as
though it had run to
completion; i.e., the
method-end conditions go
into effect and the data
that has been generated is
saved. This is the same
function as Stop on the
Instrument Control
software.
(table continued)
Table 1.1
DMA 2980 Keypad
Functions Key/Function Explanation
(continued)
STOP If an experiment is not
running (the instrument is
in a standby or method-
end state), the STOP key
halts any activity (air cool,
all mechanical motion,
etc.).
REJECT If an experiment is
running, SCROLL-STOP
ends the method normal-
(Hold down ly, as though it had run to
SCROLL and completion; i.e., the
press STOP) method-end conditions go
into effect and the data
that has been generated is
discarded. This is the
same function as Reject
on the controller.
If an experiment is not
running, SCROLL-STOP
halts any activity as
described for STOP.
HEATER Switch
The HEATER on/off switch turns the power to
the instrument heater on and off (see Figure
1.2). The switch should be in the ON position
before you start an experiment. If the HEATER
switch is off, the method will not start.
POWER Switch
Accessories
The DMA 2980 can perform experiments on
different types of samples using both standard
and optional accessories.
Standard Accessories
The accessory kit supplied with the DMA 2980
contains the following:
Clamps
The DMA 2980 utilizes several different types of
clamps. These clamps can be classified as
either tensioning or nontensioning clamps.
Tensioning clamps require that a positive force
(preload force) be placed on the sample at all
times.
3-point bending
film tension
fiber tension
compression
penetration
submersion compression
submersion film/fiber.
single cantilever
dual cantilever
shear sandwich.
Fixed
Clamp
Figure 1.3
DMA Dual
Cantilever Clamp
Optional Accessories
You can purchase extra items to use with your
TA Instruments DMA. Some of the additional
options available for use with the DMA are:
Air Compressor
Accessory (ACA)
The Air Compressor Accessory (ACA) is a
compact unit that connects directly to the air
filter regulator. It supplies compressed air, to the
air filter regulator for the DMA air bearing gas
supply.
Fuse
Power
Switch
Air tubing
Power
connection
Cable
Connection
To Air Filter
Regulator
Figure 1.4
Air Compressor Accessory
Optional Clamps
These optional clamps will allow you to perform
experiments on different types of samples:
Specifications
Tables 1.2, 1.3 and 1.4 contain information about
the DMAs specifications and temperature
control. Table 1.5 contains specifications for
the ACA. Only values with tolerances or limits
are guaranteed data. Values without tolerances
are for information only.
Table 1.2
Dimensions Depth: 17 in. (43.2 cm)
DMA 2980 Instrument Width: 26 in. (66 cm)
Specifications Height:
(furnace open)
28 in. (71 cm)
(furnace closed)
22 in. (56 cm)
Sample Thickness 5 to 10 mm
Loading 0.001 to 18 N
Table 1.3
Temperature Control
Specifications
Temperature -145 to 600°C
Range
Temperature +/-2oC
Reproducibility
Table 1.4
Experimental
Specifications
Modulus 103 Pa to 3 x 10 12 Pa
Range
Modulus +/- 1%
Precision
Maximum 18 N
Force
Minimum 0.001 N
Preload
Force
Table 1.4
Experimental
Specifications (cont'd)
Strain 1 nanometer
Resolution
Table 1.5
ACA Specifications
DMA 2980
Output Values
You can obtain the following output data from
running experiments on the DMA 2980.
Storage modulus
Loss modulus
Tan delta
Complex viscosity
Dynamic viscosity
Storage compliance
Loss compliance
Stress
% Strain
Amplitude
Frequency
Temperature
Position
Static force (preload force)
Time.
Creep Mode:
Creep compliance
% Strain
% Strain Recovery
Stress
Static force (preload force)
Temperature
Time
Decay time
Displacement
Recoverable compliance.
Stress Relaxation
Dimension change
Static force (preload force)
Position
Stress
% Strain
Temperature
Time.
DMA Isostrain
Dimension change
Position
Static force (preload force)
Strain
Stress
Temperature
Time.
Installing
the Instrument
Before shipment, the DMA 2980 instrument is
tested and calibrated both electrically and
mechanically, so that it is ready for operation
after it has been installed. Installation involves
the procedures described in this chapter:
Choosing
a Location
Because of the sensitivity of DMA experiments,
it is important to choose a location for the
instrument using the following guidelines:
In . . . a temperature-controlled area.
... a clean environment.
... an area with ample working and
ventilation space around the instru-
ment. (Refer to the technical speci-
fications in Chapter 1 for the instru-
ments dimensions.)
Away
from. . . dusty environments.
... exposure to direct sunlight.
... direct air drafts (fans, room air ducts).
... poorly ventilated areas
... flammable material that may come in
contact with the furnace assembly.
Unpacking/Repacking
the DMA
You may wish to retain all of the shipping
hardware, the plywood, and boxes from the
instrument in the event you wish to repack and
ship your instrument.
Inspecting
the System
When you receive your DMA, look over the
instrument and shipping container carefully for
signs of shipping damage. Check the parts
received against the enclosed shipping list.
Figure 2.1
Shipping Boxes
Figure 2.2
Removing the
Plywood Board
Connecting Cables
and Gas Lines
In order to connect the cables and gas lines, you
will need to have access to the instruments rear
panel.*
GPIB Cable
1. Locate the GPIB connector on the right rear
of the instrument (see Figure 2.4).
Cooling
Power Accessory Ready Reset Address GPIB Connector
Cord Connector Light Button Selector
Dial
Figure 2.4
DMA Connector Panel
5. Use the GPIB address selector dial to set an
address for the DMA. Choose an address
that is not being used by other modules
connected to the same controller. The figure
seen on the next page shows an instrument
address of 7.
Figure 2.5
Address Selector Dial
Pressure
Gauge
Electronic
Timer
Solenoid Dial
Manual
Override Button
Valve
Water Open Light
Drainage Tube Power Light
Figure 2.6
Air Filter Regulator
TA INSTRUMENTS DMA 2980 2–13
Installation
Power
Switch
Air line
to the
regulator
Figure 2.7
Air Compressor
Accessory
5. Locate the air bearing gas fitting on the right
rear of the DMA instrument.
Connect to
Air Cool Hose
Connect to Air
Filter Regulator
Figure 2.8
Fitting on DMA
for Air Bearing Gas
Figure 2.9
DMA Air Cool
Fitting
Cooling Hose
Accessory Inlet
Figure 2.10
Cooling Hose
Accessory Inlet
Power Cable
Use the following steps to connect the power
cable to the instrument.
Off On
Standby 23.25°C
DimChg 0.135µm
▲ ▲
ZERO FLOAT CLAMP FURNACE
MEASURE
CLAMP LOCK ▼ ▼
Figure 2.11
Front Panel of DMA
Showing POWER Switch
Shutting Down
the DMA
Before you decide to power down your DMA
instrument, consider the following:
Removing the
Shipping Material
The DMA 2980 is shipped with several screws
and a bracket to hold the instrument firmly in
place and prevent damage. This shipping
material must be removed as follows before the
instrument can be used.
Jackscrews
Figure 2.12
Location of Screws
TA INSTRUMENTS DMA 2980 2–25
Installation
Figure 2.13
Furnace Shipping
Nut and Bolt
3. Remove any packing material between the
rear of the furnace and the back cabinet of
the instrument.
Hex Screws
Dovetail
Drive Shaft
Mounting Posts
Figure 2.14
The Shipping
Bracket
9. Loosen the four hex screws holding the
shipping bracket in place (see the figure
above).
Dovetail Holder
Drive Shaft
Figure 2.15
Lifting Off the
Shipping Bracket
13. Install the desired clamp. See page 2-29 for
installation of the standard dual cantilever
clamp.
Hex Key
Moveable Clamp
Drive Shaft
Mounting
Posts
Figure 2.16
Installing the Single/Dual
Cantilever Moveable Clamp
Support Bar
Moveable
Jaws
Moveable Jaws
Sample
Fixed Clamp
Hex Screws
Figure 2.17
Single/Dual Cantilever
Clamp Installed (Shown
with Sample Mounted)
Clamp Mass
Calibration (Step 1) .............................. 3-7
Force Calibration
(Balance and Weight) ......................... 3-14
Temperature
Transition .................................................. 3-16
Performing
Temperature Calibration .................... 3-22
Dynamic Temperature
Calibration Procedure ........................ 3-23
Introduction
Calibration of the DMA 2980 is accomplished
through the use of the Thermal Solutions/
Advantage Instrument Control software.
Determining
When to Calibrate
Each DMA must be calibrated for accuracy,
using the following basic guidelines.
Clamp Calibration
When you install a clamp for the first time or
change from one clamp to another on the DMA
2980, you will need to perform the following
steps:
t CAUTION: Make sure that you have installed both the fixed
and moveable clamps before performing clamp
mass calibration.
Clamp Mass
Calibration (Step 1)
This type of calibration is performed to allow
the instrument to compensate for the mass of a
specific clamp to ensure accurate force measure-
ments. Clamp Mass Calibration is Step 1 in the
Thermal Solutions/Advantage DMA clamp
calibration procedure and is performed for all
clamps.
Position Calibration
Position calibration must be done when the
DMA is moved or at least once a month. The
purpose of this calibration is to calibrate the
absolute position of the drive shaft (and slide) as
read by the optical encoder.
Clamp Check
You can run the following check to see whether
your clamp position needs calibration by per-
forming a small test. Press the FLOAT key to
release (float) the clamp.
Instrument Calibrations
The following types of calibrations need to be
done when the DMA is moved or at least once a
month:
Electronics calibration
Force calibration (balance and weight)
Dynamic calibration.
Electronics Calibration
This procedure calibrates the instruments
electronics and drive motor over the entire
frequency range of the instrument. It automati-
cally removes the air to lock the slide (drive
shaft) and applies a static force (preload force)
to the motor as a calibration reference. The
instrument then uses a series of frequencies to
perform the calibration.
Dovetail Holder
Drive Shaft
Figure 3.1
Positioning the
Shipping Bracket
4. Press downwards on the shipping bracket to
position it against the mounting posts, as
shown in the figure above. The drive shaft
(slide) should move easily, DO NOT
FORCE IT. (You may need to reposition
the dovetail in the dovetail holder in order to
align the shipping bracket with the posts.)
Hex Screws
Dovetail
Drive Shaft
Mounting Posts
Figure 3.2
Mounting the
Shipping Bracket
7. Follow the instructions found in the Thermal
Solutions/Advantage documentation to
calibrate the DMA electronics.
Force Calibration
(Balance and Weight)
Force calibration is used to adjust the force
exerted by the clamp on the sample and the
force registered by the instrument as the experi-
ment proceeds. There are two steps in the force
calibrationbalance (Step 1) and weight (Step
2) calibration. This calibration must be done in
the following situations:
or
Dynamic Calibration
Dynamic calibration is used to measure several
samples of known stiffness and loss to charac-
terize the dynamic performance of the instru-
ment. Dynamic calibration must be done in the
following situations:
or
Temperature
Transition
DMA is a very popular and powerful technique
for measuring transitions in materials because it
is sensitive to side-chain or main-chain motions
and local mode relaxations in polymers that can
not be detected in DSC (differential scanning
calorimetry). Transitions in materials can be
defined by the extrapolated onset temperature of
E, or as the peak temperatures in E or tan d
curves (see Figure 3.3). Therefore, DMA
experiments primarily involve monitoring the
changes in the viscoelastic parameters of E, E,
and/or tan d, with changing temperature. Be-
cause transitions in materials are defined by the
temperature at which they are observed, it is
important to understand the dynamics of tem-
perature in the DMA 2980 when conducting
dynamic mechanical measurements.
Figure 3.3
DMA Transitions
3–16 TA INSTRUMENTS DMA 2980
Temperature Transition
Step-and-Hold Test
To perform a step-and-hold test, you should set
up a method with a start temperature, end
temperature, and temperature step size. Hold
the sample isothermally at each step for a
defined length of time, prior to conducting the
measurement.
Temperature
Ramp Test
To perform a temperature ramp test, you should
set up a method with a start temperature, end
temperature, and rate of temperature change
(e.g. 2°C/min). The instrument simultaneously
ramps the temperature and sweeps the
frequency(ies).
Testing Considerations
When conducting dynamic mechanical measure-
ments using either temperature profile test
already described, the following factors should
be considered:
Figure 3.4
Effect of Frequency
DMA Temperature
Calibration
When you analyze samples using the DMA, it is
important that the measured sample temperature
be as accurate as possible. Like any other
thermal analyzer, the accuracy of the tempera-
ture measurement relies heavily on the tempera-
ture calibration. Temperature calibration must
be performed routinely to ensure reproducible
results.
Performing
Temperature Calibration
Perform these steps for absolute temperature
calibration of the DMA sample thermocouple:
Dynamic Temperature
Calibration Procedure
If you are going to perform temperature ramp
experiments on the DMA 2980, with the single/
dual cantilever or 3-point bending clamps, you
can use the following steps to compensate for
temperature lag. However, similar temperature
calibration standards are not currently available
for the other DMA clamps. (NOTE: ASTM
Standard Test Method E186797, Standard
Test Method for Temperature Calibration of
Dynamic Mechanical Analyzer1 describes
another method for dynamic temperature
calibration based on a polymer encapsulated
metal standard.)
1
You may obtain ASTM E186797 from: ASTM,
100 Bar Harbor Drive, West Conshohocken, PA
19428-2959, (610) 832-9500.
Creating a DMA
Multifrequency Method ...................... 4-21
Creating a DMA
Multistrain Method ............................. 4-26
Creating a DMA
Stress Relaxation Method ................... 4-37
Introduction
Most DMA 2980 experiments follow the same
basic set of steps. Use the following instructions
as a guide when you perform DMA experiments:
Table 4.1
Clamps and Samples
Sample Size
Figure 4.1
Dual Cantilever Clamp
Aligning the
Thermocouples
The sample thermocouples should be close to,
but not touching the sample when it is loaded on
the clamp. (The sample thermocouple is the
taller one and is normally located on the right
side.) You may find it necessary to realign one
or both of the thermocouples periodically,
should they become bent or misaligned, or when
a new clamp is installed. Follow these steps:
Figure 4.2
View From Top
with Clamp Removed
Selecting the
Operating Mode
There are six possible operating modes for the
DMA 2980. Each mode reflects a different
class of experiment that can be performed.
Before you can begin an experiment, you need to
select an operating mode.
Mode Experiments
DMA Multifrequency
Mode
In the DMA multifrequency mode, constant
amplitude DMA (oscillatory stress and strain)
experiments are performed on samples as a
function of time, temperature, and frequency of
oscillation. Storage and loss modulus, tan delta,
viscosities, and, with some clamps, displacement
and static force (preload force) are the results of
experiments in this mode. Using the DMA
multifrequency mode, you can perform three
types of experiments: single frequency (with
temperature ramps, holds, and combinations of
the two), multifrequency temperature ramps, and
multifrequency isothermal holds. This is a
general purpose DMA mode, which is most
often used to measure transition temperatures
and the effect of frequency on transitions.
Selecting
Instrument Parameters
After you have decided on the clamp, mode, and
type of experiment, you need to select the
instrument parameters to be used. This section
provides some guidelines for selecting the
instrument parameters to use with the single/dual
cantilever clamp (for information on the other
types of clamps, see Chapter 5). The instrument
parameters are selected using the Thermal
Solutions/Advantage DMA Instrument Control
software. See the User Reference Guide for
information.
t = 3/F
Oscillation Amplitude
Table 4.2
Recommended Frequency Number of Time
Frequencies and Time Cycles Needed
0.1 7 70 sec
0.11 77 12 min
0.12 42 6 min
0.3 21 70 sec
0.32 112 6 min
0.33 231 12 min
1 7 7 sec
1.1 77 70 sec
1.2 42 35 sec
1.5 21 14 sec
10 70 7 sec
100 700 7 sec
Creating a DMA
Multifrequency Method
After you have entered the instrument param-
eters based on the type of clamp and experiment
desired, you need to create and edit a prepro-
grammed series of instrument instructions for
your experiment. These instrument instructions
are called segments. A series of segments make
up a method. This section provides basic
guidelines to use when setting up your methods.
For details on methods, refer to the Thermal
Solutions/Advantage User Reference Guide.
1 Equilibrate at -145°C
2 Data Storage Off
3 Isothermal 5 minutes (The instrument
can cool faster than the specimen; time
is needed for equilibration.)
4 Data Storage On
5 Ramp 3°C/min to 150°C
1 Equilibrate 35°C
2 Data Storage Off (Do not collect data
during the initial isothermal segment.)
Selecting
Instrument Parameters
After you have decided on the clamp, mode, and
type of experiment, you need to select the
instrument parameters to be used. This section
provides some guidelines for selecting the
instrument parameters to use with the single/dual
cantilever clamp (for information on the other
types of clamps, see Chapter 5). The instrument
parameters are selected using the Thermal
Solutions/Advantage DMA Instrument Control
software. See the User Reference Guide for
information.
t = 7/F
Frequency
All experiments are performed at a single
frequency.
Creating a DMA
Multistrain Method
After you have entered the instrument param-
eters based on the type of clamp and experiment
desired, you need to create and edit a prepro-
grammed series of instrument instructions for
your experiment. These instrument instructions
are called segments. A series of segments make
up a method. This section provides basic
guidelines to use when setting up your methods.
For details on methods, refer to the Thermal
Solutions/Advantage User Reference Guide.
1 Equilibrate 35°C
2 Data Storage Off (Do not collect data
during the initial isothermal segment.)
3 Isothermal 3 min (Hold at temperature
to ensure sample equilibrium.)
4 Amplitude Sweep (This segment turns
the data storage on, steps through the
amplitude table, and turns off data
storage after the final amplitude.)
5 Increment 3°C
6 Repeat from segment 3 until 151°C
(This segment creates a loop that will
end when the temperature is met or
exceeded.)
Selecting
Instrument Parameters
After you have decided on the clamp, mode, and
type of experiment, you need to select the
instrument parameters to be used. This section
provides some guidelines for selecting the
instrument parameters to use with the single/dual
cantilever clamp. The instrument parameters
are selected using the Thermal Solutions/
Advantage DMA Instrument Control software.
See the User Reference Guide for information.
Stress
The stress represents the amount of force
applied to the sample. Select the stress value as
follows, based on the type of experiment that you
want to run:
Equilibrium Criteria
Creating a DMA
Creep Method
After you have entered the instrument param-
eters based on the type of clamp and experiment
desired, you need to create and edit a prepro-
grammed series of instrument instructions for
your experiment. These instrument instructions
are called segments. A series of segments make
up a method. This section provides basic
guidelines to use when setting up your methods.
For details on methods, refer to the Thermal
Solutions/Advantage User Reference Guide.
1 Equilibrate 50°C
2 Data Storage OFF
3 Isothermal 5 minutes
4 Displace 10 min, Recover 10 min
(This segment lets the sample
equilibrate, turns the data storage on,
applies the selected stress for 10
minutes, then removes the stress and
lets the sample recover for 10 minutes.
5 Increment 5°C
6 Repeat from segment 2 until 152°C
(This segment creates a loop that will
end when the temperature is met or
exceeded.)
DMA Stress
Relaxation Mode
In DMA stress relaxation mode, an instanta-
neous strain is applied to the sample and the
force required to maintain the strain is measured
as a function of time (strain step). When you
perform an experiment, the primary result in this
mode is stress relaxation modulus during the
strain step.
Selecting
Instrument Parameters
After you have decided on the clamp, mode,
and type of experiment, you need to select the
instrument parameters to be used. This section
provides some guidelines for selecting the
instrument parameters to use with the single/dual
cantilever clamp. The instrument parameters
are selected using the DMA Instrument Control
software. See the User Reference Guide for
information.
Strain %
The percent strain represents the amount of
deformation applied to the sample. Select the
strain value as follows, based on the type of
experiment that you want to run:
Equilibrium Criteria
Stress relaxation experiments are run isother-
mally. To ensure that the sample has reached
thermal equilibrium before the strain is applied, a
soak time (typically 5 to 10 minutes) is recom-
mended (unless the temperature of interest is
ambient). Data collection should be initiated
immediately prior to straining the material. (See
method on page 4-37.)
Creating a DMA
Stress Relaxation Method
After you have entered the instrument param-
eters based on the type of clamp and experiment
desired, you need to create and edit a prepro-
grammed series of instrument instructions for
your experiment. These instrument instructions
are called segments. A series of segments make
up a method. This section provides basic
guidelines to use when setting up your methods.
For details on methods, refer to the Thermal
Solutions/Advantage User Reference Guide.
1 Equilibrate 50°C
2 Data Storage OFF
3 Isothermal 5 minutes
4 Displace 10 min, Recover 0 min (This
segment lets the sample equilibrate,
turns the data storage on, applies the
selected strain for 10 minutes, then
releases the sample.)
5 Increment 5°C
6 Repeat from segment 2 until 152°C
(This segment creates a loop that will
end when the temperature is met or
exceeded.)
Selecting
Instrument Parameters
This section provides some guidelines for
selecting the instrument parameters to use with
the film or fiber clamps when operating in the
DMA isostrain mode. The instrument param-
eters are selected using the Thermal Solutions/
Advantage DMA Instrument Control software.
See the User Reference Guide for information.
NOTE: You should not use the Auto Zero function when
programming the strain in the instrument
parameters. When you want to use the Auto
Zero function, you need to program the strain
as a method segment.
Strain %
The percent strain represents the amount of
deformation applied to the sample. The strain
chosen should be small enough to ensure that the
measurement is occuring within the linear
viscoelastic region. Typical values are 0.01 to
0.1%. (To determine the actual linear viscoelas-
tic region, use the procedure described for stress
relaxation.)
1 Isostrain at 1%
1 Equilibrate at -145°C
DMA Controlled
Force Mode
The DMA controlled force mode is used to
measure the displacement of a sample as a
function of time, temperature, and applied force.
Force can be applied in one of three manners:
constant force, stepped force, or continuous
force ramp. DMA controlled force can be used
to measure glass transition, softening point,
static modulus, etc. Creep measurements can
also be made in this mode, although time or
position is not reset between the applications of
force.
Selecting
Instrument Parameters
After you have decided on the clamp, mode, and
type of experiment, you need to select the
instrument parameters to be used. This section
provides some guidelines for selecting the
instrument parameters to use with the single/dual
cantilever clamp (for information on the other
types of clamps, see Chapter 5). The instrument
parameters are selected using the Thermal
Solutions/Advantage DMA Instrument Control
software. See the User Reference Guide for
information.
Creating a DMA
Controlled Force Method
After you have entered the instrument param-
eters based on the type of clamp and experiment
desired, you need to create and edit a prepro-
grammed series of instrument instructions for
your experiment. These instrument instructions
are called segments. A series of segments make
up a method. This section provides basic
guidelines to use when setting up your methods.
For details on methods, refer to the Thermal
Solutions/Advantage User Reference Guide.
Temperature-Based
Experiments
You can use temperature ramps to measure
thermal expansion, glass transitions, and soften-
ing points of materials, depending on the type of
clamping accessory used. The penetration
clamps are recommended for temperature-based
experiments since these clamps provide the best
results (low noise and relatively featureless, low
drift baseline).
1 Equilibrate at -145°C
Time-Based Experiments
(Manual Creep Measurements)
Time-based experiments most often involve
stepping the static force (preload force) level up
and down to view creep response and recovery.
Usually, creep experiments are better performed
in the DMA creep mode. (Creep Time-Tem-
perature Superposition [TTS] experiments should
always be performed in the DMA creep mode.)
An example experiment in the DMA controlled
force mode is shown on the next page.
1 Equilibrate 30°C
2 Ramp Force 2 N/min to 18 N
Preparing and
Mounting Samples
Sample preparation is one of the most important
factors in achieving accurate and reproducible
modulus values. Each type of clamp requires a
specific method for sample preparation.
t CAUTION: Make sure that the sample does not touch the
furnace or the heating element.
Single Cantilever:
1 L3 L
GF = + 2S(1 + ν )
F 12I A
Dual Cantilever:
1 L3 L
GF = + S(1 + ν)
F 192I 2A
where:
10 12
10 11
10 10
10 9
10 8
10 7
10 6
10 3
10 -1 10 0 10 1 10 2 10 3 10 4 10 5
Geometry Factor (1/mm)
Figure 4.3
Operating Range of the
Single Cantilever Clamps
Modulus (Pa)
10 13
10 12
10 11
1010
10 9
108
10 7
6
10
10 5
16 mm long 35 mm long 35 mm long 8 mm long
10 mm wide 12.5 mm wide 12.5 mm wide 12.5 mm wide
4
10 4 mm thick 3.2 mm thick 1.75 mm thick 0.1 mm thick
10 3
1 2
10 -1 10 0 10 10 10 3 10 4 10
5
Use of Operating
Range Figures
Figures 4.3 and 4.4 can help to determine the
proper sample shape for a given material, or to
determine if a sample of a particular size can be
measured as is. To determine the proper sample
size for a specific material requires some
knowledge of the approximate behavior (modu-
lus) of the material to be tested. For example, a
given material may have a modulus approxi-
mately 3x109 Pa at room temperature and will go
to 107 Pa above Tg. From Figures 4.3 and 4.4, it
can be seen that most geometry factors can
accommodate the material at room temperature,
but only a limited range of geometries (100 to
102 1/mm) can accommodate the sample above
glass transition.
Support Bar
Moveable
Jaws
Moveable Jaws
Sample
Fixed Clamp
Figure 4.5
Dual Cantilever
Clamp with Sample
t CAUTION: Make sure that the sample does not touch the
furnace or the heating element.
Measuring
Sample Length
To obtain accurate modulus values, it is impor-
tant that the sample dimensions be measured
accurately.
Spring-Loaded Tips
Handle
Kurled Knob
Figure 4.6
Telescoping Gauge
3. Insert the gauge sideways between the
clamps, just above the sample.
Performing Experiments
Once you have set the appropriate instrument
and experimental parameters and have mounted
a sample on the DMA 2980, you are ready to
run the experiment.
Starting an Experiment
To start an experiment on the DMA 2980, follow
these steps:
Stopping an Experiment
You can stop an experiment before the end of
the selected method when necessary. Use one
of the following actions to stop an experiment in
progress:
Removing Samples
When the experiment has run to completion,
remove the sample from the single/dual cantile-
ver clamp as follows:
Installing and
Calibrating the Clamp ......................... 5-81
Operating Range of the Submersion
Tension Film/Fiber Clamp ................... 5-88
Mounting a Sample ............................. 5-90
Running an Experiment ...................... 5-92
Removing a
Sample and Clamp .............................. 5-94
Introduction
The TA Instruments DMA 2980 can be used
with various accessory clamps. The
35-mm dual cantilever clamp is the standard
clamp that comes with the instrument. This
chapter provides details on the optional clamps
(clamps other than the dual cantilever clamp).
Information such as how to install the optional
clamps, mount a sample, and remove the optional
clamps is provided on the next several sections.
To decide which clamp to use for your samples,
turn to Chapter 4 in this manual to learn the
differences between the clamps and choose the
appropriate type of clamp.
Tensioning/Nontensioning
Clamps
There are two classes of clamps for the DMA
2980tensioning and nontensioning. The 3-point
bend, tension/film, tension/fiber, compression and
penetration clamps are tensioning clamps, while
the single/dual cantilever and shear sandwich
clamps are nontensioning.
Constant Force
To apply a constant force to the sample, set up
your instrument parameters as follows:
Sample Sizes
Figure 5.1
Large 3-Point
Bending Clamp
Figure 5.2
Small 3-Point
Bending Clamp
Installing the
Large Clamps
To install the 20 or 50 mm 3-point bending clamp
on the DMA 2980, follow these steps (refer to
Figure 5.3 on the next page for identification of
the parts):
Moveable Clamp
Fixed Clamp
Thermocouples
Mounting Posts
Figure 5.3
Large 3-Point Bending
Clamp Installed
Installing the
Small Clamps
To install the 5, 10, or 15 mm 3-point bending
clamps on the DMA 2980, follow these steps
(refer to Figure 5.4 on the next page for identifi-
cation of the parts):
Moveable Clamp
Dovetail
Mounting Posts
Drive Shaft
Figure 5.4
Small 3-Point Bending
Clamp Installed
3-Point Bending
Sample Preparation
Sample preparation is one of the most important
factors in achieving accurate and reproducible
modulus values. Each type of clamp requires a
specific method for sample preparation.
L3
[12 t
( )]
2
GF = 48I 1+ 5 (1 + n) L
where:
L = sample length (mm)
I = geometric moment (mm4) = 1/12 t3 W for
rectangular samples
t = sample thickness (mm)
W = sample width (mm)
n = Poissons ratio (nominally 0.44).
Modulus (Pa)
10 13
10 12
10 11
10 10
10 9
10 8
10 7
6
10
20 mm long 50 mm long
50 mm long 12.5 mm wide
10 5 12.5 mm wide
12.5 mm wide 1 mm thick
2.5 mm thick 3.2 mm thick
10 4
3 4
10 0 10 1 10 2 10 10 10 5
Geometry Factor (1/mm)
Figure 5.5
Operating Range of the
3-Point Bending Clamps
Mounting a Sample
on the 3-Point
Bending Clamps
To mount the rectangular sample on the 3-point
bending clamps, follow these steps:
Running an Experiment
When you use any of the 3-point bending clamps
for experiments, you will need to set up the
experimental parameters using the Instrument
Control software.
F = 2/3 *s (T2W/L)
where:
F = force (N)
s = stress (MPa)
T = thickness (mm)
W = width (mm)
L = length (mm) = 50 mm
Removing Samples
When the experiment has run to completion,
remove the sample from the 3-point bending
clamp as follows:
Sample Size
Plates: 10 mm x 10 mm
Thickness: up to 4 mm each side
Figure 5.6
Shear Sandwich Clamp
Top Sockethead
Capscrew Samples Setscrew Knurled Knob
Hex Screw
Fixed Clamp
Moveable Clamp
Dovetail
Dovetail Holder
Drive Shaft
Mounting Posts
Figure 5.7
Shear Sandwich Clamp
(Shown with Sample Mounted)
3 T
GF =
10 A
where:
T = sample thickness (mm)
(= side A+side B)
A = sample cross sectional area (mm2 ).
Modulus (Pa)
10
10
9
10
10 8
7
10
6
10
5
10
4
10
3
10
each piece each piece
2 mm thick 4 mm thick
2
10 10 mm square 5 mm square
1
10
-3 -2 -1 0
10 10 10 10
Geometry Factor (1/mm)
Figure 5.8
Operating Range of
the Shear Sandwich Clamp
NOTE: The shear sandwich clamp will not work very well
with high modulus materials such as polymers
below Tg.
Mounting a
Solid Sample
You will need to use two samples of equal size
and shape for the shear sandwich clamp. Each
sample can be up to 10 mm square and up to 4
mm thick. To mount the samples on the shear
sandwich clamp, follow these steps:
Mounting a Liquid
or Gel Sample
You can also analyze certain types of liquid or
gel samples on the DMA 2980.
Running an Experiment
When you use the shear sandwich clamp for
experiments, you will need to set up the experi-
mental parameters using the Instrument Control
software. Follow these general guidelines when
using the shear sandwich clamp:
Removing Samples
To remove the samples from the shear sandwich
clamp, loosen the two top setscrews and reverse
the steps used to mount the samples.
Compression Clamp
Compression is used to measure the properties
of low to medium modulus materials, including
gels and weak elastomers. The sample must
support a static (preload) force during testing.
The sample should have as high a thickness-to-
diameter ratio as is possible, depending on the
sample preparation and instrument limits.
Sample Size
Diameter: up to 15 mm and 40 mm
Thickness: up to 10 mm
Figure 5.9
Compression Clamp
Moveable Clamp
Lower
Compression Disk Hex Screw
Fixed Clamp
Dovetail Setscrews
Dovetail
Dovetail Holder
Figure 5.10
Lower Compression
Disk and Fixed Clamp
in Place
Upper
Allen Screw
Moveable
Clamp
Washers (2)
Upper Shims
Compression
Disk
Dovetail
Setscrew
Figure 5.11
Adding Shims to the
Compression Clamp for
Alignment
Operating Range of
the Compression Clamp
The equation below can help you determine
which sample clamps to use and the appropriate
sample size. This equation can also help deter-
mine if the properties of a sample of a particular
size can be measured or if the sample dimen-
sions will have to be changed.
T
GF = Fe
A
where:
Modulus (Pa)
10
10
9
10
8
10
7
10
6
10
5
10
4
10
3
10
1 mm thick 2 mm thick
2 6 mm thick
40 mm diameter
10 20 mm diameter 10 mm diameter
1
10
-4 -3 -2 -1 0
10 10 10 10 10
Geometry Factor (1/mm)
Figure 5.12
Operating Range of
the Compression Clamps
Mounting a Sample
To mount a sample on the compression clamp,
follow these directions:
Figure 5.13
Compression Clamp
(Shown with Sample
Mounted)
Running an Experiment
When you use the compression clamp for
experiments, you will need to set up the experi-
mental parameters using the instrument control
software.
Removing Samples
To remove a sample from the compression
clamp follow these guidelines.
Penetration Clamp
The penetration clamp is only used for TMA
penetration analysis on the DMA 2980. Infor-
mation such as glass transition or melting points
can be obtained. Almost any type of material
can be used as a sample with the penetration
clamp installed.
Sample Size
Diameter: up to 15 mm and 40 mm
Thickness: up to 10 mm
Figure 5.14
Penetration
Clamp
Figure 5.15
Penetration
Clamp in Place
Mounting a Sample
After the penetration clamp has been installed on
the DMA, you can mount the sample using the
following steps:
Running an Experiment
When you use the penetration clamp for experi-
ments, you will need to set up the experimental
parameters using the Thermal Solutions/
Advantage software.
Sample Size
Length: 5 to 30 mm
Width: up to 6.5 mm
Thickness: up to 2 mm
Figure 5.16
Film Tension
Clamp
Fixed Clamp
Clamping Screw
Moveable Clamp
Clamping Screw
Figure 5.17
Aligning the Fixed and
Moveable Clamps of
the Film Tension Clamp
8. Tighten the setscrew while holding the
moveable clamp in place.
Operating Range
of the Tension Clamps
The equation below can help you determine
which sample clamps to use and the appropriate
sample size. This equation can also help deter-
mine if the properties of a sample of a particular
size can be measured or if the sample dimen-
sions will have to be changed.
L
GF =
A
where:
12
10
10 11
10 10
9
10
108
107
106
10 mm long 20 mm long
wide 20 mm long
105 5 mm long
thick
0.2 mm long
4 mm wide
0.1 mm thick 0.1 mm diameter
4
10
10
-1
10
0
10 1 102 103 104 105
Mounting a Sample
After the film tension clamp has been installed
on the DMA, you can mount the sample using
the following steps:
Fixed Clamp
Sample
Moveable Clamp
Clamping Screw
Dovetail
Dovetail Holder
Figure 5.19
Film Tension Clamp
(Shown with Sample
Mounted)
TA INSTRUMENTS DMA 2980 5–65
Using Your Options
Running an Experiment
When you use the film tension clamp for experi-
ments, you will need to set up the experimental
parameters using the instrument control soft-
ware.
Removing a Sample
After you have finished your experiment, you
can remove the sample from the film tension
clamp as follows:
Sample Size
Length: 5 to 30 mm
Diameter: up to 1.25 mm
Figure 5.20
Fiber Tension Clamp
Sample
Hex Screws
Moveable
Clamp
Dovetail
Mounting Posts
Dovetail Holder
Drive Shaft
Figure 5.21
Fiber Tension
Clamp (Shown
with Sample
Mounted)
Mounting a Sample
After the fiber tension clamp has been installed
on the DMA, you can mount the samples
according to the diameter of the fiber: large
diameter (high denier) monofilaments or small
diameter filaments and fiber bundles. The fiber
clamp can hold samples from 5 denier (20 µm)
up to 1/32 in. (0.8 mm) in diameter. See the
following sections for instructions.
Large Diameter
(High Denier) Monofilaments
Follow these steps to mount large diameter
monofilaments:
Running an Experiment
When you use the fiber tension clamp for
experiments, you will need to set up the experi-
mental parameters using the Thermal Solutions
software.
Removing Samples
To remove the fiber samples from the fiber
tension clamp after an experiment, take off the
knurled vise collars and then remove the sample
from clamps. If necessary, scrape sample out of
pin vise in the center of the vise collar.
Submersion
Film/Fiber Clamp
This clamp is used to evaluate thin films or fibers
while they are submerged in an ambient tem-
perature fluid.
Sample Size
Yoke
Splash Guard
with Rim Up
Figure 5.22
Submersion Film/Fiber Tension
Clamp (with Sample Loading Fixture in Place)
5–80 TA INSTRUMENTS DMA 2980
Submersion Film/Fiber Tension Clamps
Installing and
Calibrating the Clamp
NOTE: The submersion film/fiber tension clamp
requires the clamp mass calibration to be
performed prior to complete installation of the
clamp. The following instructions will outline the
installation, calibration, and sample loading of
the submersion film/fiber fixture.
Mounting Thermocouple-Mounting
Posts Brackets Thermocouples
Figure 5.23
View from the Top
10. Install the top bar onto the yoke and tighten
the two hex screws. (See the figure below.)
Top Bar
Yoke
Drive Shaft
Figure 5.24
Yoke and Top Bar Installed
on Drive Shaft
Fluid Container
Hex Screws
Mounting Posts
Figure 5.25
Fluid Container Mounted
Crossbar Loading
Fixture
Upper Clamp
Lower
Sample U-Clamp
Figure 5.26
Sample-Loading Clamp Screw
Assembly Lower Clamp
Top Bar
Fluid Container
Tub
Lower U-Clamp
Figure 5.27
Sample Loading
Assembly in Place
Sample
Top Bar
Fluid
Container
Tub
Lower
U-Clamp
Figure 5.28
Sample in Place on
Submersion Film Clamp
Operating Range of
the Submersion
Tension Film/Fiber Clamp
The geometry factor, GF, in the equation below
can help you determine the optimum size for the
sample material or determine if a sample of a
particular size can be measured.
GF = L/A
where:
GF = 15/A
for the Submersion Film Tension Clamp
Modulus = Ks x GF, or
Modulus = Ks x (15/A)
where the measurable stiffness range of the
instrument is a fixed quantity between 100 to
10,000,000 N/m.
Ks = Modulus/GF, or
Ks = (Modulus x A)/15
Since the modulus of the material is an intrinsic
propertyi.e., independent of the sample
dimensionsit can be seen that the sample
dimensions must be selected to yield stiffness
values within the allowable measurable range of
the instrument.
Mounting a Sample
After the submersion film/fiber tension clamp
has been installed and calibrated, you can mount
the sample using the following steps:
Upper Clamp
Top Sample
Clamp Screw
Sample
Bottom Sample
Clamp Screw
Figure 5.29 Lower Clamp
Running an Experiment
When you use the submersion film tension clamp
for experiments, you will need to set up the
experimental parameters using the Instrument
Control software.
Removing a
Sample and Clamp
After you have finished your experiment, you
can remove the sample from the submersion film
clamp as follows:
Submersion
Compression Clamp
Compression is used to measure the properties
of low to medium modulus materials, including
gels and weak elastomers. These clamps are
used to to test low to medium modulus samples
while they are submerged in an ambient tem-
perature fluid. The sample must support a
preload force, so the samples should not flow.
The sample should have as high a thickness-to-
diameter ratio as is possible, depending on the
sample preparation and instrument limits.
Samples that are in the stiffness range of 100 N/
m to 10,000,000 N/m can be tested.
Sample Size
Diameter: up to 15 mm and 30 mm
Thickness: up to 5 mm
Figure 5.30
Submersion Compression Clamp
Mounting Thermocouple-Mounting
Posts Brackets Thermocouples
Figure 5.31
View from the Top
6. Straighten the thermocouples, then push
them down to their lowest position.
Setscrew
Drive Shaft
Figure 5.32
Yoke Installed
on Drive Shaft
Captive Screws
Stages
Mounting Posts
Figure 5.33
Stages Installed
Hex Screws
Stage
Fluid Cup
Rim up on
Splash Guard
Figure 5.34
Fixed Clamp
Fully Installed
Hex Screws
Top Bar
Compression Disk
Yoke
Figure 5.35
Submersion Compression
Clamp Fully Mounted
Operating Range
of the Submersion
Compression Clamp
The geometry factor, GF, in the equation below
can help you determine the optimum sample size
for the sample material or determine if a sample
of a particular size can be measured.
GF = Fe(T/A)
where:
Modulus = Ks x GF or
Modulus = Ks x Fe(L/A)
where the measurable stiffness range of the
instrument is a fixed quantity between 100 to
10,000,000 N/m.
Ks = Modulus/GF or
Ks = (Modulus x A)/Fe x L
Since the modulus of the material is an intrinsic
propertyi.e., independent of the sample
dimensionsit can be seen that the sample
dimensions must be selected to yield stiffness
values within the allowable measurable range of
the instrument.
Mounting a Sample
After the submersion compression clamp has
been installed and calibrated on the DMA, you
can mount the sample using the following steps:
Running an Experiment
When you use the submersion compression
clamp for experiments, you will need to set up
the experimental parameters using the Instru-
ment Control software.
Removing a
Sample and Clamp
After you have finished your experiment, you
can remove the sample from the submersion
compression clamp as follows:
Solids ....................................................6-5
Liquids ..................................................6-7
Dynamic Mechanical
Analysis Testing .................................. 6-11
Calculations Based
on Clamp Type .......................................... 6-20
Clamping Factors
(Compression Clamps Only) ..................... 6-36
Introduction
This chapter provides information regarding the
theory and applications of the DMA 2980
instrument and the Dynamic Mechanical Analy-
sis technique.
theory of operation
comparison to other techniques
viscoelasticity
modes of operation
sample stiffness and modulus calculations
calculations based on clamp type.
Theory of
Operation
The Dynamic Mechanical Analyzer 2980 is a
precision instrument designed to measure
viscoelastic properties, such as modulus (stiff-
ness) and damping (energy dissipation), of solid
and soft solid materials. Samples of various
shapes and sizes are fixed into position using a
variety of stainless steel clamping arrangements.
A mechanical deformation is applied to the
sample via the instruments drive motor. The
deformation can be applied sinusoidally, in a
constant (or step) fashion, or under a fixed rate.
The combination of temperature and deforma-
tion profile are programmed using the Thermal
Solutions Instrument Control software.
Comparison to
Other Techniques
DMA and DEA (Dielectric Analysis) are two of
the techniques that can bridge the disciplines of
thermal analysis and rheology as seen in the
figure below.
Liquids Gels Solids
Molten Polymers Soft Solids
DMA
DEA
Fluids Rheology
Figure 6.1
Correlation to Other Techniques
6–4 TA INSTRUMENTS DMA 2980
Defining Viscoelasticity
Defining
Viscoelasticity
The concept of viscoelasticity comes from the
fact that most materials do not exhibit purely
elastic (ideal solids) or purely viscous (ideal
liquids) behavior but a combination of both. By
definition, viscoelasticity simply means having
both viscous and elastic properties. This section
presents a discussion of solids and liquids and
their behavior as it applies to the DMA.
Solids
Materials are often referred to as solids or
liquids, depending on whether or not they retain
their shape under the force of gravity. An ideal
solid is a material that is purely elastic. Elastic-
ity is simply a materials ability to store deforma-
tional energy. A purely elastic material (an ideal
solid) will change its shape when deformed and
regain its original shape when the deformation is
removed. All the energy (stress) applied to the
material is stored by the material during the
deformation.
Hookes Law:
Where:
Elastic Deformation
J = 1/G
γ
Figure 6.2
Elastic Deformation
Liquids
An ideal liquid, in contrast, is a material that has
no elasticity. An ideal fluid stores no deforma-
tional energy. An ideal liquid is not rigid but
conforms to the shape of its container and finds
its own level under gravity. When an ideal fluid
is deformed, it changes its shape and, when the
deformation is removed, it does not regain its
original shape but remains in its new deformed
shape. An example of an ideal fluid (ideal
viscous fluid) is water. Consider a glass of
water sitting on your table. The water takes on
the shape of glass. If you spill the glass of
water, when you pick up the glass the water does
not find its way back into the glass. This
behavior is referred to as viscous flow and is
described by Newtons Law which linearly
relates the stress to the rate of shear:
Newtons Law:
τ = ηdγ/dt
Viscous Flow
η
γ
Figure 6.3
Viscous Flow
Viscoelastic Behavior
As stated earlier, the concept of viscoelasticity
comes from the fact that most materials do not
exhibit purely elastic (ideal solids) or purely
viscous (ideal liquids) behavior but a combina-
tion of both. When as stress is applied to a
viscoelastic material, it will show time-depen-
dent deformation. Any viscoelastic material,
given enough time, will flow under an applied
stress. When the stress is removed the material
will not fully recover. The portion of strain that
is recovered represents the energy stored or the
elastic portion of the materials response. The
portion of the strain that is not recovered repre-
sents the energy dissipated or viscous portion of
the materials response. Viscoelastic behavior
of a material is shown graphically in Figure 6.4
below.
Je0
Figure 6.4
Viscoelastic Response
(The Voigt Model)
Energy lost
Viscous response
Dropped Bounce
back
Energy stored
Elastic response
Figure 6.5
Viscoelastic Behavior
Modes of
Operation
The viscoelastic properties of a material can be
characterized on the DMA 2980 using three
experimental testing modes. The next sections
discuss these modes:
Dynamic Mechanical
Analysis Testing
During dynamic testing, an oscillatory (sinusoi-
dal) strain (or stress) is applied to the material
and the resulting stress (or strain) developed in
the material is measured. For an ideal solid
material, which obeys Hookes law, the resulting
stress will be proportional to the amplitude of
the applied strain. The stress and strain waves
will be in phase or, put another way, the phase
shift (phase angle δ) between the stress and
strain is 0° (see Figure 6.6A on the next page).
For an ideal fluid which obeys Newtons law,
the stress will be proportional to the strain rate.
The stress signal will lead the strain signal by
90° (see Figure 6.6B on the next page).
o
δ=0 δ = 90
o
Stress
Strain
A B
Figure 6.6
Ideal Solid and Fluid
DMA Stress and Strain
For a viscoelastic material, the phase angle will
lie somewhere between 0° and 90° (Figure 6.7).
E’’ E*
Strain δ
E’ Stress axis
Complex Modulus:
E* = σ*/ε
or
E* = E + iE
(Tension, Compression or Bending)
G* = τ*/γ
or
G* = G+ iG
(Shear)
Storage Modulus:
E = σ/ε
or
E = E* cos δ
(Tension, Compression or Bending)
G = τ/γ
or
G = G* cos δ
(Shear)
Loss Modulus:
E = σ/ε
or
E = E* sin δ
(Tension, Compression or Bending)
G = τ/γ
or
G = G* sin δ
(Shear)
Relaxation Recovery
Step Step
J
Stress s
σ η
γ
Figure 6.8
Viscoelastic Response
of the Maxwell Model
Dynamic
Measurements
In dynamic experiments, the DMA 2980 mea-
sures the raw signals of force, amplitude of
deformation, and phase angle. Instrument
calibration constants are applied to the raw
signals. Force and amplitude are used with the
phase angle to calculate the storage and loss
stiffness (K and K). Tan δ is calculated
Transient
Measurements
In transient measurements, the DMA 2980
measures the raw signals of force and amplitude
of deformation. Stress and strain are derived
from the raw force and amplitude, normalized
for the geometry factor.
Calculations Based
on Clamp Type
Since a materials modulus is independent of its
geometry, equations relating the sample stiffness
to the modulus depend on the type of clamps
used, the sample shape, and the mode of defor-
mation. This section contains stiffness calcula-
tions for the different clamp types along with
comments on appropriate correction factors.
Also included are stress and strain equations,
which can be used as a general guideline for
calculations made from the force and amplitude
of deformation. The equations for stress and
strain assume linear viscoelastic behavior.
Modulus calculations based on the equations
may not agree with values computed by the
DMA 2980 program, especially if the sample
stretches or shrinks significantly during an
experiment.
Dual Cantilever
When running experiments using the dual
cantilever clamp, the equations found in this
section are applied to obtain your results.
Modulus Equation
The stiffness model equation for a rectangular
cross section sample, analyzed on the dual
cantilever clamp, is as follows. (Similar calcula-
tions were performed for cylindrical samples but
are not detailed here.)
24 . E. I
K
2
3 12 . t
L . 1 (1 ν ).
5 L
Where:
K = stiffness or spring constant
E = elastic modulus
L = sample length (one side)
t = sample thickness
I = sample moment of inertia
ν = Poisons ratio
3
w. t
I
12
Where:
t = sample thickness
I = sample moment of inertia
w = sample width
Ks
Fc
K
Where:
K = stiffness or spring constant
Ks = measured stiffness
Fc = clamping correction factor
Fc = 0.7616 0.02713 x
L
t
+ 0.1083 ln ()
L
t
Where:
E = elastic modulus
L = sample length (one side)
ln = natural log
Ks = measured stiffness
t = sample thickness
I = sample moment of inertia
ν = Poisons ratio
Fc = clamping correction factor
E=2(1+ν)G
3 . P. L
σx
2
w. t
3 .δ. t . Fc
εx
2
2 12 . t
L. 1 (1 ν ).
5 L
Where:
σx = stress
εx = strain
P = 1/2 applied force
δ = amplitude of deformation
L = sample length (one side)
t = sample thickness
w = sample width
Fc = clamping correction factor
ν = Poisons ratio
Single Cantilever
When running experiments using the single
cantilever clamp, the equations found in this
section are applied to obtain your results.
Modulus Equation
The stiffness model equation for a rectangular
cross section sample, analyzed on the single
cantilever clamp, is as follows. (Similar calcula-
tions were performed for cylindrical samples but
are not detailed here.)
Ks L 3 12 . t
2
E . . 1 (1 ν ).
Fc 12 . I 5 L
Fc = 0.7616 0.02713 x
L
t
+ 0.10831n ()
L
t
Where:
E = elastic modulus
L = sample length
t = sample thickness
I = sample moment of inertia
ν = Poisons ratio
Ks = measured stiffness
Fc = clamping correction factor
3-Point Bending
When running experiments using the 3-point
bending clamp, the equations found in this
section are applied to obtain your results.
Modulus Equation
The stiffness model equation for a rectangular
cross section sample, analyzed on the 3-point
bending clamp, is as follows. (Similar calcula-
tions were performed for cylindrical samples but
are not detailed here.)
3 2
L 12 . t
E Ks . . 1 (1 ν ).
6. I 5 L
Where:
E = elastic modulus
L = sample length (one side)
t = sample thickness
I = sample moment of inertia
ν = Poisons ratio
Ks = measured stiffness
PLt 3P L
σx = =
8I 2 w t2
6 . δ. t
εx 2
2 6 . t
L. 1 (1 ν) .
10 L
0
Where:
σx = stress
εx = strain
P = applied force
δ = amplitude of deformation
L = sample length (span)
t = sample thickness
w = sample width
ν = Poisons ratio
I = moment of inertia
Shear Sandwich
When running experiments using the shear
sandwich clamp, the equations found in this
section are applied to obtain your results.
Modulus Equation
The stiffness model equation for a rectangular
cross section sample, analyzed on the shear
sandwich clamp, is as follows. (Similar calcula-
tions were performed for cylindrical samples but
are not detailed here.)
5. G. w. h
K
3. t
Where:
K = stiffness or spring constant
G = shear modulus
w = sample width, i.e. horizontal dimension
h = sample height, i.e. vertical dimension
t = sample thickness, between clamp faces
3. t
G Ks .
5 . w. h
Where:
G = shear modulus
w = sample width, i.e. horizontal dimension
h = sample height, i.e. vertical dimension
t = sample thickness, between clamp faces
Ks = measured stiffness
P
τ0
2. A
δ
γ0
T
Where:
σ0 = stress
γ0 = strain
P = applied force
δ = amplitude of deformation
A = sample cross-sectional area
T = separation of clamp surfaces
Compression
When running experiments using the compres-
sion clamp, the equations found in this section
are applied to obtain your results.
Modulus Equation
The stiffness model equation for a sample,
analyzed on the compression clamp, is as
follows.
A. E
K
t
Where:
K = stiffness or spring constant
E = elastic modulus
A = sample cross-sectional area
t = sample thickness
Ks
Fe
K
Where:
K = stiffness or spring constant
Ks = measured stiffness
Fe = correction factor. See Clamping
Factors in Chapter 6 to determine the
value of Fe.
Ks . t
E
Fe A
Where:
E = elastic modulus
t = sample thickness
Ks = measured stiffness
Fe = correction factor. See Clamping
Factors in Chapter 6 to determine the
value of Fe.
A = sample cross-sectional area
P
σ0 Fe .
A0
δ
γ0
T
Where:
σ0 = stress
γ0 = strain
P = applied force
δ = amplitude of deformation
A = sample cross-sectional area
T = separation of clamp surfaces
Fe = correction factor that accounts for
sample dilatation effects. See Clamp-
ing Factors in Chapter 6 to determine
the value of Fe.
Modulus Equation
The stiffness model equation for a sample,
analyzed on the film or fiber tension clamp, is as
follows.
A. E
K
L
Where:
K = stiffness or spring constant
E = elastic modulus
A = sample cross-sectional area
L = sample length
L
E Ks .
A
Where:
E = elastic modulus
A = sample cross-sectional area
L = sample length
Ks = measured stiffness
P
σ0
A0
∆L
ε0
L0
Where:
σ0 = stress
γ0 = strain
P = applied force
∆L = cumulative change in sample length
L0 = initial sample length
A0 = initial sample cross-sectional area
Clamping Factors
(Compression Clamps Only)
Table 6.1
Clamping Factors (Fe) Thickness OD ID Fe
for 1 mm Ring Sample (mm) (mm) (mm)
1 5 4 0.7669
1.5 5 4 0.8026
2 5 4 0.8207
2.5 5 4 0.8314
3 5 4 0.8383
3.5 5 4 0.8430
4 5 4 0.8464
4.5 5 4 0.8488
5 5 4 0.8507
1 10 9 0.7669
1.5 10 9 0.8026
2 10 9 0.8207
2.5 10 9 0.8314
3 10 9 0.8383
3.5 10 9 0.8430
4 10 9 0.8464
4.5 10 9 0.8488
5 10 9 0.8507
1 15 14 0.7669
1.5 15 14 0.8026
2 15 14 0.8207
2.5 15 14 0.8314
3 15 14 0.8383
3.5 15 14 0.8430
(table continued )
Table 6.1
Clamping Factors (Fe)
for 1 mm Ring Sample
(continued)
Thickness OD ID Fe
(mm) (mm) (mm)
4 15 14 0.8464
4.5 15 14 0.8488
5 15 14 0.8507
1 20 19 0.7669
1.5 20 19 0.8026
2 20 19 0.8207
2.5 20 19 0.8314
3 20 19 0.8383
3.5 20 19 0.8430
4 20 19 0.8464
4.5 20 19 0.8488
5 20 19 0.8507
1 25 24 0.7669
1.5 25 24 0.8026
2 25 24 0.8207
2.5 25 24 0.8314
3 25 24 0.8383
3.5 25 24 0.8430
4 25 24 0.8464
4.5 25 24 0.8488
5 25 24 0.8507
1 30 29 0.7669
1.5 30 29 0.8026
2 30 29 0.8207
2.5 30 29 0.8314
3 30 29 0.8383
3.5 30 29 0.8430
4 30 29 0.8464
4.5 30 29 0.8488
(table continued )
Table 6.1
Clamping Factors (Fe)
for 1 mm Ring Sample
(continued)
Thickness OD ID Fe
(mm) (mm) (mm)
5 30 29 0.8507
1 35 34 0.7669
1.5 35 34 0.8026
2 35 34 0.8207
2.5 35 34 0.8314
3 35 34 0.8383
3.5 35 34 0.8430
4 35 34 0.8464
4.5 35 34 0.8488
5 35 34 0.8507
1 40 39 0.7669
1.5 40 39 0.8026
2 40 39 0.8207
2.5 40 39 0.8314
3 40 39 0.8383
3.5 40 39 0.8430
4 40 39 0.8464
4.5 40 39 0.8488
5 40 39 0.8507
Table 6.2
Clamping Factors (Fe)
for 2 mm Ring Sample
Thickness OD ID Fe
(mm) (mm) (mm)
1 5 3 0.67337
1.5 5 3 0.73322
2 5 3 0.76688
2.5 5 3 0.78811
3 5 3 0.80256
3.5 5 3 0.81293
4 5 3 0.82069
4.5 5 3 0.82666
5 5 3 0.83137
1 10 8 0.67337
1.5 10 8 0.73322
2 10 8 0.76688
2.5 10 8 0.78811
3 10 8 0.80256
3.5 10 8 0.81293
4 10 8 0.82069
4.5 10 8 0.82666
5 10 8 0.83137
1 15 13 0.67337
1.5 15 13 0.73322
2 15 13 0.76688
2.5 15 13 0.78811
3 15 13 0.80256
3.5 15 13 0.81293
4 15 13 0.82069
4.5 15 13 0.82666
5 15 13 0.83137
1 20 18 0.67337
1.5 20 18 0.73322
(table continued)
Table 6.2
Clamping Factors (Fe)
for 2 mm Ring Sample
(continued)
Thickness OD ID Fe
(mm) (mm) (mm)
2 20 18 0.76688
2.5 20 18 0.78811
3 20 18 0.80256
3.5 20 18 0.81293
4 20 18 0.82069
4.5 20 18 0.82666
5 20 18 0.83137
1 25 23 0.67337
1.5 25 23 0.73322
2 25 23 0.76688
2.5 25 23 0.78811
3 25 23 0.80256
3.5 25 23 0.81293
4 25 23 0.82069
4.5 25 23 0.82666
5 25 23 0.83137
1 30 28 0.67337
1.5 30 28 0.73322
2 30 28 0.76688
2.5 30 28 0.78811
3 30 28 0.80256
3.5 30 28 0.81293
4 30 28 0.82069
4.5 30 28 0.82666
5 30 28 0.83137
1 35 33 0.67337
1.5 35 33 0.73322
2 35 33 0.76688
2.5 35 33 0.78811
3 35 33 0.80256
(table continued)
Table 6.2
Clamping Factors (Fe)
for 2 mm Ring Sample
(continued)
Thickness OD ID Fe
(mm) (mm) (mm)
3.5 35 33 0.81293
4 35 33 0.82069
4.5 35 33 0.82666
5 35 33 0.83137
1 40 38 0.67337
1.5 40 38 0.73322
2 40 38 0.76688
2.5 40 38 0.78811
3 40 38 0.80256
3.5 40 38 0.81293
4 40 38 0.82069
4.5 40 38 0.82666
5 40 38 0.83137
Table 6.3
Clamping Factors (Fe)
for 3 mm Ring Sample
Thickness OD ID Fe
(mm) (mm) (mm)
1 5 2 0.6013
1.5 5 2 0.6734
2 5 2 0.7173
2.5 5 2 0.7464
3 5 2 0.7669
3.5 5 2 0.7820
4 5 2 0.7935
4.5 5 2 0.8026
5 5 2 0.8098
1 10 7 0.6013
1.5 10 7 0.6734
2 10 7 0.7173
2.5 10 7 0.7464
3 10 7 0.7669
3.5 10 7 0.7820
4 10 7 0.7935
4.5 10 7 0.8026
5 10 7 0.8098
1 15 12 0.6013
1.5 15 12 0.6734
2 15 12 0.7173
2.5 15 12 0.7464
3 15 12 0.7669
3.5 15 12 0.7820
4 15 12 0.7935
4.5 15 12 0.8026
5 15 12 0.8098
1 20 17 0.6013
1.5 20 17 0.6734
2 20 17 0.7173
(table continued)
Table 6.3
Clamping Factors (Fe)
for 3 mm Ring Sample
(continued)
Thickness OD ID Fe
(mm) (mm) (mm)
2.5 20 17 0.7464
3 20 17 0.7669
3.5 20 17 0.7820
4 20 17 0.7935
4.5 20 17 0.8026
5 20 17 0.8098
1 25 22 0.6013
1.5 25 22 0.6734
2 25 22 0.7173
2.5 25 22 0.7464
3 25 22 0.7669
3.5 25 22 0.7820
4 25 22 0.7935
4.5 25 22 0.8026
5 25 22 0.8098
1 30 27 0.6013
1.5 30 27 0.6734
2 30 27 0.7173
2.5 30 27 0.7464
3 30 27 0.7669
3.5 30 27 0.7820
4 30 27 0.7935
4.5 30 27 0.8026
5 30 27 0.8098
1 35 32 0.6013
1.5 35 32 0.6734
2 35 32 0.7173
2.5 35 32 0.7464
3 35 32 0.7669
3.5 35 32 0.7820
(table continued)
Table 6.3
Clamping Factors (Fe)
for 3 mm Ring Sample
(continued)
Thickness OD ID Fe
(mm) (mm) (mm)
4 35 32 0.7935
4.5 35 32 0.8026
5 35 32 0.8098
1 40 37 0.6013
1.5 40 37 0.6734
2 40 37 0.7173
2.5 40 37 0.7464
3 40 37 0.7669
3.5 40 37 0.7820
4 40 37 0.7935
4.5 40 37 0.8026
5 40 37 0.8098
Table 6.4
Clamping Factors (Fe)
for 4 mm Ring Sample
Thickness OD ID Fe
(mm) (mm) (mm)
1 5 1 0.7752
1.5 5 1 0.8539
2 5 1 0.8572
2.5 5 1 0.8390
3 5 1 0.8150
3.5 5 1 0.7903
4 5 1 0.7669
4.5 5 1 0.7452
5 5 1 0.7254
1 10 6 0.7752
1.5 10 6 0.8539
2 10 6 0.8572
2.5 10 6 0.8390
3 10 6 0.8150
3.5 10 6 0.7903
4 10 6 0.7669
4.5 10 6 0.7452
5 10 6 0.7254
1 15 11 0.7752
1.5 15 11 0.8539
2 15 11 0.8572
2.5 15 11 0.8390
3 15 11 0.8150
3.5 15 11 0.7903
4 15 11 0.7669
4.5 15 11 0.7452
5 15 11 0.7254
1 20 16 0.7752
1.5 20 16 0.8539
2 20 16 0.8572
(table continued)
Table 6.4
Clamping Factors (Fe)
for 4 mm Ring Sample
(continued)
Thickness OD ID Fe
(mm) (mm) (mm)
2.5 20 16 0.8390
3 20 16 0.8150
3.5 20 16 0.7903
4 20 16 0.7669
4.5 20 16 0.7452
5 20 16 0.7254
1 25 21 0.7752
1.5 25 21 0.8539
2 25 21 0.8572
2.5 25 21 0.8390
3 25 21 0.8150
3.5 25 21 0.7903
4 25 21 0.7669
4.5 25 21 0.7452
5 25 21 0.7254
1 30 26 0.7752
1.5 30 26 0.8539
2 30 26 0.8572
2.5 30 26 0.8390
3 30 26 0.8150
3.5 30 26 0.7903
4 30 26 0.7669
4.5 30 26 0.7452
5 30 26 0.7254
1 35 31 0.7752
1.5 35 31 0.8539
2 35 31 0.8572
2.5 35 31 0.8390
3 35 31 0.8150
3.5 35 31 0.7903
(table continued)
Table 6.4
Clamping Factors (Fe)
for 4 mm Ring Sample
(continued)
Thickness OD ID Fe
(mm) (mm) (mm)
4 35 31 0.7669
4.5 35 31 0.7452
5 35 31 0.7254
1 40 36 0.7752
1.5 40 36 0.8539
2 40 36 0.8572
2.5 40 36 0.8390
3 40 36 0.8150
3.5 40 36 0.7903
4 40 36 0.7669
4.5 40 36 0.7452
5 40 36 0.7254
Table 6.5
Clamping Factors (Fe)
for Square Sample
Thickness Length Fe
(mm) (mm)
1 5 0.4647
1.5 5 0.5550
2 5 0.6326
2.5 5 0.6937
3 5 0.7404
3.5 5 0.7759
4 5 0.8032
4.5 5 0.8246
5 5 0.8417
1 10 0.3784
1.5 10 0.4193
2 10 0.4647
2.5 10 0.5108
3 10 0.5550
3.5 10 0.5958
4 10 0.6326
4.5 10 0.6651
5 10 0.6937
1 15 0.3558
1.5 15 0.3784
2 15 0.4050
2.5 15 0.4342
3 15 0.4647
3.5 15 0.4956
4 15 0.5259
4.5 15 0.5550
5 15 0.5827
1 20 0.3465
1.5 20 0.3610
2 20 0.3784
(table continued)
Table 6.5
Clamping Factors (Fe)
for Square Sample
(continued)
Thickness Length Fe
(mm) (mm)
2.5 20 0.3980
3 20 0.4193
3.5 20 0.4417
4 20 0.4647
4.5 20 0.4879
5 20 0.5108
1 25 0.3415
1.5 25 0.3519
2 25 0.3643
2.5 25 0.3784
3 25 0.3940
3.5 25 0.4107
4 25 0.4282
4.5 25 0.4463
5 25 0.4647
1 30 0.3386
1.5 30 0.3465
2 30 0.3558
2.5 30 0.3665
3 30 0.3784
3.5 30 0.3913
4 30 0.4050
4.5 30 0.4193
5 30 0.4342
1 35 0.3366
1.5 35 0.3429
2 35 0.3503
2.5 35 0.3588
3 35 0.3682
3.5 35 0.3784
(table continued)
Table 6.5
Clamping Factors (Fe)
for Square Sample
(continued)
Thickness Length Fe
(mm) (mm)
4 35 0.3894
4.5 35 0.4010
5 35 0.4131
1 40 0.3352
1.5 40 0.3404
2 40 0.3465
2.5 40 0.3534
3 40 0.3610
3.5 40 0.3694
4 40 0.3784
4.5 40 0.3880
5 40 0.3980
Table 6.6
Clamping Factors (Fe)
for Solid Circular Sample
Thickness Diameter Fe
(mm) (mm)
1 5 0.4871
1.5 5 0.5890
2 5 0.6708
2.5 5 0.7319
3 5 0.7771
3.5 5 0.8114
4 5 0.8385
4.5 5 0.8612
5 5 0.8814
1 10 0.3842
1.5 10 0.4334
2 10 0.4871
2.5 10 0.5400
3 10 0.5890
3.5 10 0.6327
4 10 0.6708
4.5 10 0.7037
5 10 0.7319
1 15 0.3570
1.5 15 0.3842
2 15 0.4162
2.5 15 0.4511
3 15 0.4871
3.5 15 0.5226
4 15 0.5568
4.5 15 0.5890
5 15 0.6187
1 20 0.3459
1.5 20 0.3632
2 20 0.3842
(table continued)
Table 6.6
Clamping Factors (Fe)
for Solid Circular Sample
(continued)
Thickness Diameter Fe
(mm) (mm)
2.5 20 0.4079
3 20 0.4334
3.5 20 0.4601
4 20 0.4871
4.5 20 0.5139
5 20 0.5400
1 25 0.3401
1.5 25 0.3523
2 25 0.3672
2.5 25 0.3842
3 25 0.4030
3.5 25 0.4230
4 25 0.4440
4.5 25 0.4655
5 25 0.4871
1 30 0.3367
1.5 30 0.3459
2 30 0.3570
2.5 30 0.3699
3 30 0.3842
3.5 30 0.3997
4 30 0.4162
4.5 30 0.4334
5 30 0.4511
1 35 0.3345
1.5 35 0.3417
2 35 0.3504
2.5 35 0.3605
3 35 0.3718
3.5 35 0.3842
(table continued)
Table 6.6
Clamping Factors (Fe)
for Solid Circular Sample
(continued)
Thickness Diameter Fe
(mm) (mm)
4 35 0.3974
4.5 35 0.4114
5 35 0.4260
1 40 0.3329
1.5 40 0.3388
2 40 0.3459
2.5 40 0.3541
3 40 0.3632
3.5 40 0.3733
4 40 0.3842
4.5 40 0.3957
5 40 0.4079
Maintaining the
Air Filter Regulator ..................................... 7-6
Introduction
When you have been using the TA Instruments
DMA 2980 for a while, some maintenance
procedures that need to be done in order to keep
your instrument running as it should. Several
steps are recommended:
Inspection
Examine the instrument for good condition as
follows:
Cleaning
Cleaning your instrument regularly helps to
increase its longevity and maintain its efficiency.
Maintaining the
Air Filter Regulator
The air filter regulator supplies clean air bearing
gas to the DMA. It must be checked regularly
to maintain optimum air-cleaning capability.
Air Supply
Valve
From ACA
or Lab
Air Supply To DMA
Pressure
Gauge
Electronic
Timer
Solenoid Dial
Manual
Override Button
Valve
Water Open Light
Drainage Tube Power Light
Figure 7.1
Air Filter
Regulator Assembly
Changing a Filter
Follow these instructions to change a filter:
Error Messages
As you perform experiments using your DMA
2980, you may see error messages appearing on
the display or on the controller.
Diagnosing
Power Problems
Fuses
The DMA contains internal fuses that are not
user serviceable. If any of the internal fuses
blows, a hazard may exist. Call your TA Instru-
ments service representative.
Fuse 1 Fuse 2
Figure 7.2
Fuse Locations
Heater
Indicator Light
The indicator light in the HEATER switch on the
front panel of the DMA 2980 glows whenever
the power control circuitry is enabled. If the light
does not come on when the method is started,
the indicator light may be defective or a hard-
ware problem may exist in the instrument (call
your TA Instruments Service Representative).
Power Failures
A power failure caused by a temporary drop in
line voltage results in one of two responses by
the DMA instrument:
DMA 2980
Test Functions
The DMA 2980 has three levels of test and
diagnostic functions:
Table 7.1
DMA Confidence Test
Test Number Area Being Tested
CPU logic
30 CMOS RAM
4n Program memory
5n CPU board I/O
functions
6n DRAM data storage
memory
70 GPIB test
82 Keypad test
90 RS-232 test
An Analog board tests
Bn Drive board tests
D0 Saved memory
checksum
Parts List
The following parts may be ordered for use with
the DMA 2980.
Table 7.2
DMA 2980 Parts List
Part Number Description
Table 7.2
DMA 2980 Parts List
(cont'd)
Part Number Description
(table continued)
Table 7.2
DMA 2980 Parts List
(cont'd)
Part Number Description
TA Instruments, Inc.
109 Lukens Drive
New Castle, DE 19720
Telephone: 1-302-427-4000 or 1-302-427-4040
Fax: 1-302-427-4001
HELPLINEU.S.A.
For technical assistance with current or
potential thermal analysis applications,
please call the Thermal Analysis Help Desk
at 1-302-427-4070.
SERVICEU.S.A.
For instrument service and repairs,
please call 1-302-427-4050.
TA Instruments Ltd.
Europe House, Bilton Centre
Cleeve Road
Leatherhead, Surrey KT22 7UQ
England
Telephone: 44-1372-360363
Fax: 44-1372-360135
TA Instruments GmbH
Max-Planck-Strasse 11
D-63755 Alzenau
Germany
Telephone: 49-6023-9647-0
Fax: 49-6023-9647-77
TA Instruments Benelux
Ottergemsesteenweg 461
B-9000 Gent
Belgium
Telephone: 32-9-220-79-89
Fax: 32-9-220-83-21
TA Instruments Japan
No. 5 Koike Bldg.
1-3-12 Kitashinagawa
Shinagawa-Ku, Tokyo 140
Japan
Telephone: 813/3450-0981
Fax: 813/3450-1322
TA Instruments France
B.P. 608
78056 Saint-Quentin-Yvelines
Cedex
France
Telephone: 33-1-30-48 94 60
Fax: 33-1-30-48 94 51
TA Instruments Spain
Waters Cromatografía, S.A.
División TA Instruments
Avda. Europa, 21. Pta. Baja
28108 Alcobendas
Madrid, Spain
Telephone: 34-91-661-8448
Fax: 34-91-661-0855
TA Instruments Australia
Unit 3
38-46 South Street
Rydalmere NSW 2116
Autstralia
Telephone: 61-29-9331-705
Fax: 61-29-8981-455
TA Instruments Italy
Division of Waters SpA
via Achille Grandi 27
20090 Vimodrone (MI), Italy
Telephone: 39-02-27421-1
Fax: 39-02-250-1827
Printed in U.S.A.
A–2 TA INSTRUMENTS DMA 2980
Index
Index
G I
Gas Cooling Accessory (GCA) 1- inspection 2-6, 7-3
15, 2-19
instrument
gas lines air cooling 2-19
connecting 2-10 calibration 3-3 to 3-6, 3-11 to 3-
coonecting 2-10 13
installation 2-10 cleaning 7-4
display 1-7
geometry factors (GF) 5-19, 5- dynamic calibration 3-9
31, 5-88 grounding 2-5
heater switch 1-12
glass transition (Tg) 3-23 inspection 7-3
installing 2-3
glass transitions 4-46 keys 1-8
lifting xxiv
GPIB location 2-4
cable 2-10 mounting feet installation 2-9
output values 1-21
GPIB interface 1-6 power requirements 2-4
power switch 1-12
removal of shipping material 2-25
H to 2-26
hazardous products xxiv repacking 2-28
shutting down 2-24
heat deflection temperature specifications 1-18 to 1-21
(HDT) 5-24
instrument parameters 4-14 to 4-
heater indicator light 7-12 23, 4-23 to 4-28, 4-28 to 4-
33, 4-33 to 4-39, 4-44 to 4-49
HEATER switch 1-12, 7-11, 7-12
introduction 1-3 to 1-6
heating/cooling rate 3-19
isostrain 4-39
Hookes law 6-6
K mode
DMA creep 4-28 to 4-33
keypad 1-5, 1-8 to 1-12 DMA isostrain 4-39
cleaning 7-4 DMA multifrequency 4-15
DMA multistrain 4-23, 4-26, 4-
31, 4-37
L DMA stress relaxation 4-33 to
liquid nitrogen xx 4-39
handling xx TMA controlled force 4-44, 4-
thermal shock xx 44 to 4-49
M moisture 2-5
maintenance 7-3
monofilaments 5-72
manual
mounting feet
using xxv
installation 2-9
materials
multifrequency
transitions 3-16
creating tables 4-18
mode 4-14 to 4-23
Maxwell Model 6-16
multistrain
mechanical section enclosure 1-3
mode 4-23 to 4-28
method
amplitude sweep 4-26, 4-32, 4- N
37, 4-38
creating 4-21, 4-26, 4-31, 4- Newtons Law 6-7, 6-11
37, 4-42, 4-46
creep 4-31 nontensioning 1-13
examples 4-48
TMA controlled force 4-46
S SCROLL 1-8
techniques thermocouple
comparison 6-4 absolute temperature 3-21
absolute temperature calibration 3-
telescoping gauge 4-58, 4-59 22
measuring 4-60 correction 3-21
position 3-19
temperature positioning 2-31
lags 3-17 temperature lags 3-21
compensation 3-23
profiles 3-17, 3-19 thermocouples
transition 3-16 aligning 4-10
viscosity
V coefficient of 6-7
valve
air supply (air filter regulator) 2- Voigt Model 6-9, 6-15
17
solenoid 2-17 voltage xix
viscoelastic W
parameters 3-16
work surface 2-4
viscoelastic limit 4-23