Kruss Techdata k100 en PDF
Kruss Techdata k100 en PDF
Kruss Techdata k100 en PDF
SPECIFICATIONS
Product group specifications K100C K100 K100SF
Force measurement
Maximum load 120 g 210 g 6g
Resolution 100 µg 10 µg 0.1 µg
Measurement rate 50 Hz
Adjustment automated, external weight fully automated fully automated
Adjustment weight CP0503: 100 g 1) internal weight internal weight
Locking mechanism automatic
Sample stage
Travel distance >110 mm >110 mm >110 mm
Resolution 20 µm 0.1 µm 0.1 µm
Travel speed 0.1 to 500 mm/min
Type of motor brushless DC servo motor
Optical height sensor – yes yes
Simple platform optional
Thermostated jacket 50 mm, 70 mm or 100 mm 1)
Inverse CMC cone-shaped vessel 1)
Integrated sample stage yes
Software
ADVANCE surface tension (SFT)/interfacial tension (IFT) 1)
contact angle 1)
critical micelle concentration (CMC) 1)
liquid density 1)
solid density 1) solid density 1) –
- sedimentation/penetration 1) –
1)
optional
Measurement specifications K100C K100 K100SF
Du Noüy ring
Results surface tension (SFT)/interfacial tension (IFT)/critical micelle concentration (CMC)
Range 1 to 2000 mN/m 1 to 2000 mN/m 1 to 400 mN/m
Resolution 0.01 mN/m 0.001 mN/m 0.0001 mN/m
Correction methods Harkins-Jordan, Huh-Mason, Zuidema-Waters, linear correction, no correction
Rod method
Results SFT/IFT/CMC
Range 1 to 2000 mN/m
Resolution 0.2 mN/m 0.02 mN/m 0.002 mN/m
Wilhelmy plate 2)
Result(s) SFT/IFT/CMC contact angle (CA) SFT/IFT/CMC contact angle (CA) SFT/IFT/CMC contact angle (CA)
Range 1 to 2000 mN/m 0 to 180° 1 to 2000 mN/m 0 to 180° 1 to 700 mN/m 0 to 180°
Resolution 0.02 mN/m 0.01° 0.002 mN/m 0.01° 0.0002 mN/m 0.01°
Types – advancing, receding 3) – advancing, receding 4) – advancing, receding 5)
Washburn
Result contact angle (CA)
Range 0 to 90°
Resolution 0.01°
Type advancing
Surface free energy of solids
Result surface free energy
Models equation of state, Zisman, Fowkes, Wu, Owens-Wendt-Rabel-Kaelble, extended Fowkes, acid-base theory
Liquid density
Range 1 to 2200 kg/m³
Resolution 1 kg/m³
Precision ±3 kg/m³
Solid density
Range 1000 to 20000 kg/m³ 1000 to 20000 kg/m³ –
Resolution 1 kg/m³ 1 kg/m³ –
Precision ±3 kg/m³ ±3 kg/m³ –
Sedimentation
Result – graph: mass vs. time –
Penetration
– graph: mass vs. time –
2)
general & single side Wilhelmy plate method possible
3)
>200 µm
4)
>20 µm
5)
>5 µm
General specifications K100C K100 K100SF
Temperature control
6)
with TJ50 Peltier temperature control unit
7)
with TJ60 high temperature control unit
kruss-scientific.com