FORCE TENSIOMETER – K100

SPECIFICATIONS
Product group specifications K100C K100 K100SF
Force measurement
Maximum load 120 g 210 g 6g
Resolution 100 µg 10 µg 0.1 µg
Measurement rate 50 Hz
Adjustment automated, external weight fully automated fully automated
Adjustment weight CP0503: 100 g 1) internal weight internal weight
Locking mechanism automatic
Sample stage
Travel distance >110 mm >110 mm >110 mm
Resolution 20 µm 0.1 µm 0.1 µm
Travel speed 0.1 to 500 mm/min
Type of motor brushless DC servo motor
Optical height sensor – yes yes
Simple platform optional
Thermostated jacket 50 mm, 70 mm or 100 mm 1)
Inverse CMC cone-shaped vessel 1)
Integrated sample stage yes
Software
ADVANCE surface tension (SFT)/interfacial tension (IFT) 1)
contact angle 1)
critical micelle concentration (CMC) 1)
liquid density 1)
solid density 1) solid density 1) –
- sedimentation/penetration 1) –

1)
optional
Measurement specifications K100C K100 K100SF
Du Noüy ring
Results surface tension (SFT)/interfacial tension (IFT)/critical micelle concentration (CMC)
Range 1 to 2000 mN/m 1 to 2000 mN/m 1 to 400 mN/m
Resolution 0.01 mN/m 0.001 mN/m 0.0001 mN/m
Correction methods Harkins-Jordan, Huh-Mason, Zuidema-Waters, linear correction, no correction
Rod method
Results SFT/IFT/CMC
Range 1 to 2000 mN/m
Resolution 0.2 mN/m 0.02 mN/m 0.002 mN/m
Wilhelmy plate 2)

Result(s) SFT/IFT/CMC contact angle (CA) SFT/IFT/CMC contact angle (CA) SFT/IFT/CMC contact angle (CA)
Range 1 to 2000 mN/m 0 to 180° 1 to 2000 mN/m 0 to 180° 1 to 700 mN/m 0 to 180°
Resolution 0.02 mN/m 0.01° 0.002 mN/m 0.01° 0.0002 mN/m 0.01°
Types – advancing, receding 3) – advancing, receding 4) – advancing, receding 5)
Washburn
Result contact angle (CA)
Range 0 to 90°
Resolution 0.01°
Type advancing
Surface free energy of solids
Result surface free energy
Models equation of state, Zisman, Fowkes, Wu, Owens-Wendt-Rabel-Kaelble, extended Fowkes, acid-base theory
Liquid density
Range 1 to 2200 kg/m³
Resolution 1 kg/m³
Precision ±3 kg/m³
Solid density
Range 1000 to 20000 kg/m³ 1000 to 20000 kg/m³ –
Resolution 1 kg/m³ 1 kg/m³ –
Precision ±3 kg/m³ ±3 kg/m³ –
Sedimentation
Result – graph: mass vs. time –
Penetration
– graph: mass vs. time –

2)
general & single side Wilhelmy plate method possible
3)
>200 µm
4)
>20 µm
5)
>5 µm
General specifications K100C K100 K100SF
Temperature control

Technical changes are reserved. K100-20181029


Types liquid, electrical, Peltier liquid, Peltier
Range -15 6) to 300 °C 7) -15 to 50 °C 6)
Temperature controlled measurement
Range -60 to 450 °C
Resolution 0.01 °C
Precision ±0.05 °C
Accuracy ±0.5 °C
Internal sensor sample stage
External sensor sample vessel 1)
Housing and peripherals
Built-in and software-controlled ionizer – yes yes
Built-in bubble level yes
Glass windshield doors yes
Stainless steel measuring compartment yes
Control pad yes
Touch panel optional optional –
Environment
Operating temperature 15 to 30 °C
Humidity > 30% without condensation
Instrument dimensions
Footprint 300 mm × 390 mm (W × D)
Height 585 mm
Weight (without accessories) 19 kg 23 kg 24 kg
Power supply
Voltage 100 to 240 VAC
Power consumption 40 W
Frequency 47 to 63 Hz
Interfaces
PC USB 2.0
Auxiliary RS232
Thermostat external (optional)
Inert gas yes

6)
with TJ50 Peltier temperature control unit
7)
with TJ60 high temperature control unit

kruss-scientific.com