Kruss Techdata bp100 en PDF

The document summarizes the specifications of the BP100 bubble pressure tensiometer, which can measure dynamic surface tension versus surface age or at constant surface age with a resolution of 0.01 mN/m. It has a maximum pressure of 3000 Pa and can measure samples with minimum volume of 15 mL from -10 to 130 °C with precision of 0.05 °C. The BP100 has a brushless DC servo motor, optional simple platform and thermostated jacket, and interfaces with a PC via USB to measure bubble pressure and analyze adsorption/diffusion constants using selected models.

Uploaded by

Travis H
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
476 views4 pages

Kruss Techdata bp100 en PDF

The document summarizes the specifications of the BP100 bubble pressure tensiometer, which can measure dynamic surface tension versus surface age or at constant surface age with a resolution of 0.01 mN/m. It has a maximum pressure of 3000 Pa and can measure samples with minimum volume of 15 mL from -10 to 130 °C with precision of 0.05 °C. The BP100 has a brushless DC servo motor, optional simple platform and thermostated jacket, and interfaces with a PC via USB to measure bubble pressure and analyze adsorption/diffusion constants using selected models.

Uploaded by

Travis H
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

BUBBLE PRESSURE TENSIOMETER – BP100

SPECIFICATIONS
Product group specifications BP100
Pressure measurement
Maximum pressure 3000 Pa
Precision 0.05 %
Accuracy 1.3 %
Measurement rate 20 kHz
Sample stage
Travel distance >110 mm
Resolution 20 µm
Travel speed 0.1 to 500 mm/min
Type of motor brushless DC servo motor
Simple platform optional
Thermostated jacket 70 mm (100 mm) 1)
Software
LabDesk bubble pressure

Measurement specifications BP100


Bubble pressure
Results dynamic surface tension vs. surface age dynamic surface tension at constant surface age
Surface tension range 15 to 100 mN/m
Resolution 0.01 mN/m
Precision 1%
Surface age range 5 to 200000 ms
Adsorption/diffusion
Results adsorption constant diffusion constant
Models Ward & Tordai Joos & Rillaerts

1)
optional
General specifiations BP100
Sample dimensions
Minimum recommended sample volume 15 mL 2)
Temperature control
Type liquid
Range -10 to 130 °C
Flow-through thermostat optional
Temperature measurement
Range -10 to 130 °C
Resolution 0.01 °C
Precision ±0.05 °C
Accuracy ±0.5 °C
Internal sensor sample stage
External sensor sample vessel 1)
Housing and peripherals
Built-in bubble level yes
Glass windshield doors yes
Stainless steel measuring compartment yes
Control pad yes
Environment
Operating temperature 15 to 30 °C
Humidity without condensation
Instrument dimensions
Footprint 300 mm × 390 mm (W × D)
Height 585 mm
Weight (without accessories) 13 kg
Power supply
Voltage 100 to 240 VAC
Power consumption 40 W
Frequency 47 to 63 Hz
Compressed air 100 to 400 kPa 1)
Interfaces
PC USB 2.0
Auxiliary RS232
Thermostat quick lock fittings
External gas 1) quick lock fittings

1)
optional
2)
sample vessel SV23 recommended