FCO560 Users Guide Issue 2a
FCO560 Users Guide Issue 2a
Calibrator
Users Guide
Manufactured by:
Email: [email protected]
Web Site: www.furness-controls.com
Table of Contents
Table of Contents .................................................................................................................................. 2
Safety Notes And Recommendations ................................................................................................. 4
Introduction ............................................................................................................................................ 5
General Description ............................................................................................................................. 5
Front Panel .......................................................................................................................................... 6
Rear Panel ........................................................................................................................................... 7
Enclosure ............................................................................................................................................. 8
Battery Operation ................................................................................................................................. 8
Basic Operation ..................................................................................................................................... 9
Main Display ........................................................................................................................................ 9
Menu System ..................................................................................................................................... 10
Menu Navigation ................................................................................................................................ 11
Menu Navigation Diagram ................................................................................................................. 13
Global Settings Menu .......................................................................................................................... 15
Instrument Definitions Menu .............................................................................................................. 18
Differential Pressure .......................................................................................................................... 19
Custom Differential Pressure ............................................................................................................. 20
Pressure Sensor ................................................................................................................................ 22
Laminar Flow Element ....................................................................................................................... 23
Pitot Static Tube ................................................................................................................................. 26
Calibration Profiles Menu ................................................................................................................... 29
Pressure Sources Menu...................................................................................................................... 30
Temperature Sources Menu ............................................................................................................... 31
Laminar Flow Elements Menu ............................................................................................................ 32
Fbus Settings Menu ............................................................................................................................ 33
Serial Port Settings Menu ................................................................................................................... 34
Data Logger Menu ............................................................................................................................... 35
Warning System Menu ........................................................................................................................ 37
Security Menu ...................................................................................................................................... 38
Time & Date Menu ............................................................................................................................... 39
Utilities Menu ....................................................................................................................................... 40
Calibration Due Date.......................................................................................................................... 40
D.P. Generation ................................................................................................................................. 40
Data Logger ....................................................................................................................................... 40
Print Data ........................................................................................................................................... 40
Copy Data .......................................................................................................................................... 40
Initialise Data ..................................................................................................................................... 40
Adjust FCO3xx calibration ................................................................................................................. 41
Service Centres ................................................................................................................................. 41
System Information ............................................................................................................................ 41
Flash Update ...................................................................................................................................... 41
Calibrate Mode Scenarios................................................................................................................... 42
DP Transmitter Calibration................................................................................................................. 43
Leak Detector DP Calibration ............................................................................................................ 44
Flow Calibration ................................................................................................................................. 45
High Pressure Calibration .................................................................................................................. 46
Multiple sensor configurations allow different external references to be used quickly and simply. When
combined with laminar flow elements the FCO560 may be used as a flow calibrator covering flows
from less than 1 ml/min to over 30000 l/min. An external pressure sensor may also be used as a
reference for calibrating pressures much higher than the FCO560’s differential pressure range.
When used for site calibrations, the FCO560 may be preconfigured with information for up to 99
different instruments to be calibrated. For bench use, these configurations are useful for rapidly
changing between commonly used setups.
In ‘measurement mode’, the FCO560 may be used to measure and data-log a variety of signals.
Features:
Push Buttons
Differential pressure (d.p.) Equalise/Read. The d.p. reference will
display O=O when equalised and the ports will be connected together.
Fn Functions Menu. This is context sensitive - the functions offered vary
according to the instruments other settings.
Backup/escape from menus or functions
Store calibration points or trigger the measurement mode data logger
Change the calibration point and navigation within menus
Change the main display content and navigation within menus
ENTER The centre button is used to enter the menu system and enter data
Digital I/O
& Analogue OP
Supply Voltage
12Vdc, 24VA
Pressure Pump
Pump Head
Tube Element
The FCO560 is supplied with 2 sizes of pipe depending on the range. Each pipe is colour coded for
easy identification. Size 1 pipe can be used for low flow small volumes and the size 2 pipe for larger
flow larger volumes.
Note: the pump is rated for ±30kPa maximum, if the calibrator is used with pressures
exceeding ±30kPa the pipe must be removed and blanking caps fitted to the pneumatic ports
Carry Handle
The supplied handle locks in two positions 0º and 180º. The 90º position is for carrying and does not
fully lock. The handle can also be folded back over the top when not in use.
Battery Operation
When the optional rechargeable battery pack is fitted it provides a running time of >8 hours of normal
operation from a full charge. An icon on the display shows the battery charge level.
Charging
The battery charges when a 12V power supply is connected; this may be the universal mains adaptor
that is supplied with the instrument, or an in-car adapter.
Charging has two stages: The first stage is a fast-charge that takes approximately two hours from fully
discharged or less time if the battery is partially charged. The second stage charges at a slower rate
for three hours to achieve maximum capacity. While charging, an icon on the display indicates the
charge stage. Once it is fully charged the battery will be given a small maintenance charge to ensure
that it remains ready for battery operation. The mains adaptor may be left connected for bench use.
Calibration Screen
The main calibration screen allows comparison of the reference and the Equipment Under
Test (EUT), lets you store the calibration points, and displays other useful information.
The content of the top two lines at the left of the display is user-selectable. This example
shows the description/title of the instrument being calibrated and the FCO560 differential
pressure reference.
The symbol to the right of the description shows the measurement response as undamped
(step) or damped (filtered).
The next symbol is only shown for instruments with automatic pressure generation. The
symbol shows if pressure control is active or turned off.
The right side of the screen has three windows, the contents of which are user-selectable.
The upper large value is the reference with appropriate engineering units. For calibrations
where d.p. is used for the reference, this will change to indicate 0=0 when the d.p. ports are
equalised (linked) by an internal valve.
The lower large value is the signal read from the EUT in the same engineering units as the
reference. This is typically a 4-20mA signal scaled to the required engineering units.
The left side shows the calibration point being taken and how many points are needed for this
calibration. The up/down buttons select the calibration point. A tick is shown when the point
has been logged.The target value for this point is shown beneath.
The bottom left corner shows the battery status, if a battery is fitted.
The bottom of the display has a bar graph showing the current reference as a percentage of
EUT full scale. The bar can be displayed as centre zero or left-hand zero.
Measurements Screen
The left side of the screen provides feedback from the measurement mode data logger.
The function of the lower large display value is user-programmable in addition to the three
windows down the right-hand side.
Menu System
The FCO560’s settings may be viewed or set using the front panel buttons via a menu system. The
main menu consists of:
Settings may be protected by a security code. When security is enabled it is possible to view the
settings but you will be prompted to enter the security code if you try to alter settings.
If you select a menu with multiple configurations, such as Instrument Definitions or Calibration Profiles,
you must first choose which to use. Use the UP or DOWN buttons to select from the list and press
ENTER.
When you press ENTER a list of parameters is displayed with their settings, e.g.
The text to edit is within the double-lined box with its title shown above.
Initially the first character in the edit box blinks to show that it is selected. The up, down, left, and right
buttons move a blinking cursor around the characters shown at the bottom of the display. Press the
ENTER button to replace the edit box character with the new selection from below. The edit box
selection automatically advances to the next character. Any of the blanks in the selection characters
may be used as a space.
Note that the and Fn buttons have secondary and usage for text editing. These are
used to move the selection point in the edit box without changing the text. Unwanted text can be
removed by entering blank spaces.
SELECT
INSTRUMENT INSTRUMENT INSTRUMENT ID
DEFINITIONS (1 to 99) DESCRIPTION
SERIAL NUMBER
CALIBRATION PROFILE * * Calibration mode only
DIFFERENCE *
INPUT SOURCE *
(EUT VOLTS / mA) (COMMS) (MANUAL)
INPUT MINIMUM COMMS PROTOCOL (NO FURTHER OPTIONS)
INPUT MAXIMUM COMMS BAUD RATE
COMMS DATA-PARITY
COMMS FBUS ID
REFERENCE SOURCE
(D.P.) (CUSTOM D.P.) (PRESSURE SENSOR) (LFE) (PITOT)
D.P. UNITS D.P. UNITS DECIMAL POINT POS. DECIMAL POINT POS. DECIMAL POINT POS.
D.P. MINIMUM D.P. MINIMUM DISPLAY MINIMUM DISPLAY MINIMUM DISPLAY MINIMUM
D.P. MAXIMUM D.P. MAXIMUM DISPLAY MAXIMUM DISPLAY MAXIMUM DISPLAY MAXIMUM
OUTPUTS DECIMAL POINT POS. PRESSURE SENSOR LFE K FACTOR
DISPLAY MINIMUM PRESSURE UNITS JET FLOW TEST GAS
DISPLAY MAXIMUM PRESSURE TYPE TEST GAS
CUSTOM UNITS OUTPUTS (CUSTOM)
RESPONSE (CUSTOM) REL. DENSITY
OUTPUTS REL. VISCOSITY FLOW TYPE
REL. DENSITY VOL. FLOW UNITS
FLOW TYPE FLOW @
VOL. FLOW UNITS MASS FLOW UNITS
FLOW @ VELOCITY UNITS
MASS FLOW UNITS POWER UNITS
VELOCITY UNITS CALORIFIC VALUE
POWER UNITS AREA UNITS
CALORIFIC VALUE DUCT AREA
AREA UNITS PRESSURE SENSOR
DUCT AREA PRESSURE UNITS
PRESSURE SENSOR TEMPERATURE SENSOR
PRESSURE UNITS TEMPERATURE UNITS
TEMPERATURE SENSOR STD. PRESSURE
TEMPERATURE UNITS STD. TEMPERATURE
STD. PRESSURE OUTPUTS
STD. TEMPERATURE
SELECT OUTPUTS
CALIBRATION PROFILE DESCRIPTION
PROFILES No. SAMPLES
(1 to 20)
No. CAL. POINTS
POINT 1..POINT 21
SERIAL PORT
PORT
SETTINGS BAUD RATE
PARITY
AUTO LINE FEED
FORMAT
CSV CHARACTERS
WARNING
LATEST VALUES
SYSTEM LIMITS
SECURITY LOCK
PASS CODE
Operating Mode
The instrument can be used in two distinctly different modes.
Calibrate: The calibration features are enabled to allow calibration data to be logged.
Measure: The calibration features are replaced by a measurements data logger.
Atmospheric Pressure
The atmospheric pressure may be required for a number of uses.
Select from a list of pressure sources. This allows the atmospheric pressure to be
entered manually as a constant value or to be measured by an internal or external
absolute sensor.
Filter
Any values derived from the FCO560 differential pressure measurement may be digitally
filtered to provide a damped response.
Adjustable up to a maximum filter time constant of 60.0 seconds.
Damping can be simply switched on/off using the Fn button menu.
FCO560 ID.
Assign an identifying description to this FCO560.
This is used when exporting calibration data and when differential pressure is the
reference and one of the Info Lines below is set to “Reference”. It defaults to “FCO560”
followed by its serial number but can be changed according to the user’s preference.
Note that this value does not get sent or received by Fbus Communications as part of
global settings. It can only be set via this menu.
Info Line 1, 2
Choose from a list of items to be displayed at the top left of the main display screen.
None: the display is left blank.
Instrument definition ID.
Instrument definition description.
Instrument definition serial number.
Reference description / ID.
Test gas.
Date/time.
Window 1, 2, 3
Choose what to display in each window of the main display screen.
None: the display is left blank.
Difference: in calibration mode, the difference between reference and EUT.
Std. Temperature: as defined in the instrument definition menu.
Temperature: as defined in the instrument definition menu.
Aux. Temperature: .as defined in this menu.
Std. Pressure: as defined in the instrument definition menu.
Atmospheric Pressure: as defined in this menu.
Absolute Pressure: based on the pressure sensor defined in the instrument definition
menu. If a gauge sensor is used then the absolute value is derived by adding the
atmospheric pressure.
Gauge Pressure: based on the pressure sensor defined in the instrument definition
Main Display 2
Measurement mode only.
Choose what to display on the lower main display position. The choices are the same as
the window options.
Aux. Temp.
This may be used by a window or the data logger to read a temperature from any source.
Select from a list of temperature sources.
Aux. Press.
This may be used by a window or the data logger to read a pressure from any source.
Select from a list of pressure sources.
Persistent Menus
Enable or disable the persistent menu feature.
No: Each time the menu is entered the first menu entry is displayed.
Yes: Upon entering the menu, the display starts at the position last used.
Display Brightness
Adjust the display for different viewing conditions. High brightness will slightly reduce the
operating time when operating from the battery.
Enter a value from 1 (dim) to 16 (bright).
Display Contrast
Adjust the display for different viewing conditions.
Enter a number from 1 to 16.
Description
May be used for additional text relating to the instrument. Up to 16 characters.
Serial number
The serial number of the instrument to be calibrated. Up to 16 characters.
Calibration Profile
Calibration Mode only.
Choose a named calibration profile. The calibration profile defines the number of calibration
points that are required and at what %FS each point is required.
Difference
Calibration Mode only.
When calibrating, the difference between EUT and reference may be displayed as either
%Reading or %Full Scale.
See the Window options in Global Settings.
Input Source
Calibration Mode only.
Select the signal source to calibrate against the reference.
EUT mA: via the front panel Test connector.
EUT Volts: via the front panel Test connector.
Communications: read the EUT by digital communications via the USB-A port. This may
be directly connected to Furness Controls instruments that have a USB-B port, or via an
FCO301-USB adaptor to the FCO300 series of transmitters, or via a USB-RS232 adaptor
to instruments having an RS232 port. Note that only USB-RS232 adaptors using an FTDI
chipset are supported.
Manual: for analogue gauges etc, the value may be entered from the keypad.
Input Minimum
Only displayed if the input source is Voltage or mA.
Enter the minimum value for the input signal, e.g. 4mA.
Input Maximum
Only displayed if the input source is Voltage or mA.
Enter the maximum value for the input signal, e.g. 20mA.
Protocol
Only displayed if the input source is Communications.
Fbus: Furness Control instruments such as leak detectors and flow meters.
FCO3xx: the display value of any of the FCO300 series, e.g. FCO332. Note that normally
the loop signal would be calibrated for transmitters.
FCO5xx: FCO510, FCO560, or PPC500.
FRS4: normally only used by Furness Controls or its agents.
Parity
Only displayed if the input source is Communications and the protocol is not FCO3xx.
8, NONE: 8 data bits and no parity
7, ODD: 7 data bits plus odd parity
7, EVEN: 7 data bit plus even parity
8, ODD: 8 data bits plus odd parity
8, EVEN: 8 data bits plus even parity
Fbus ID
Only displayed if the input source is Communications and protocol is Fbus.
Enter the instrument address from 1 to 99.
Reference Source
Select the measurement source to be used as the calibration reference.
DP: Differential Pressure
Custom DP: Differential Pressure scaled to custom units
Pressure Sensor
Laminar Flow Element
Pitot Static Tube
The remaining settings depend upon the reference source selected above and are detailed for
each reference type…
Differential Pressure
The calibration reference is the FCO560’s differential pressure measurement. Pressure may be
automatically generated by the FCO560 if the pressure generation option is fitted, or provided from an
external pressure source.
D.P. Units
Select the differential pressure units of the EUT.
Pa: Pascals
kPa: kiloPascals
ubar: microbar
mbar: millibar
mmWC: millimetres Water Column
Thou: thousands of an inch Water Column
inWC: inches Water Column
mmHg: millimetres mercury
inHg: inches mercury
-2
NM : Newtons per square metre (Pa)
PSI: Pounds per square inch
PSF: Pounds per square foot
D.P. Minimum
Enter the differential pressure that corresponds with the minimum value for the input. This is
required even if the input source is Communications or Manual so that an appropriate display
resolution can be set and so that calibration errors can be expressed as %FS.
Outputs
Digital outputs may be set for each instrument definition, e.g. for external valve selection.
Each of the four outputs may be selected on (1) or off (0), e.g. to turn outputs 1 and 3 on you
would set “1010”. Note that the digital inputs/outputs must be configured in the global setting
menu.
Other values that are not used for D.P. calibration may be seen if display windows have been
configured that need additional settings.
D.P. Units
Select the differential pressure units of the EUT.
Pa: Pascals
kPa: kiloPascals
ubar: microbar
mbar: millibar
mmWC: millimetres Water Column
Thou: thousands of an inch Water Column
inWC: inches Water Column
mmHg: millimetres mercury
inHg: inches mercury
-2
NM : Newtons per square metre (Pa)
PSI: Pounds per square inch
PSF: Pounds per square foot
D.P. Minimum
Enter the Differential Pressure that corresponds with the Display Minimum value below.
D.P. Maximum
Enter the Differential Pressure that corresponds with the Display Maximum value below.
Display Minimum
Enter the display value that corresponds with the minimum differential pressure and the
minimum value for the input signal.
Custom Units
The custom units list matches the FCO300 series custom units. These are for display
purposes only and have no effect on the FCO560’s measurement. It is the user’s responsibility
to ensure that this selection makes sense with the DP units that have been selected.
Pa, kPa, mbar, mmH20, inH20, mmHg, inHg, PSI, m/s, mph, ft/s, ft/m, kmh, l/s, l/m, l/h,
CFS, CFM, CFH, CMS, CMM, CMH, LBS, LBM, LBH, KGS, KGM, KGH, MA, PC, ‘ ‘,
CCS, CCM, CCH
Response
Define the response of the EUT to applied differential pressure.
Linear: the differential pressure is scaled directly to the custom scaling.
Square law: differential pressure is square-rooted before it is scaled. Note that when
square law is selected the calibration profile relates to the differential pressure rather than
the square law values.
Outputs
Digital outputs may be set for each instrument definition, e.g. for external valve selection.
Each of the four outputs may be selected on (1) or off (0), e.g. to turn outputs 1 and 3 on you
would set “1010”. Note that the digital inputs/outputs must be configured in the global setting
menu.
Other values that are not used for Custom D.P. calibration may be seen if display windows
have been configured that need additional settings.
Pressure Units
Select the pressure units that the EUT uses.
Display Minimum
Enter the display value that corresponds with the minimum pressure value for the input.
Display Maximum
Enter the display value that corresponds with the maximum pressure value for the input.
Pressure sensor
Select from a list of pressure sources for use as the reference. It is possible to use either
absolute or gauge references. When either the reference or the EUT is gauge the Fn button
menu offers a tare facility that operates as follows:
Reference EUT Operation
Absolute Absolute Tare not available.
Absolute Gauge Tare the reference to display as gauge pressure
Tare the reference to remove any zero offset.
Atmospheric pressure is added to the gauge reference to give the
Gauge Absolute
absolute pressure. Atmospheric pressure may be measured using the
internal absolute sensor or entered as a constant value if know.
Gauge Gauge Tare the reference to remove any zero offset.
Pressure Units
Select the pressure units of the EUT.
kPa, mbar, bar, PSI, mmHg, inHg, mmWC, inWC.
Pressure Type
Pressure calibrations may be either absolute pressure or gauge pressure. Select the type of
pressure that the EUT displays. The letter ‘A’ or ‘G’ is appended to the pressure units as a
reminder.
Outputs
Digital outputs may be set for each instrument definition, e.g. for external valve selection.
Each of the four outputs may be selected on (1) or off (0), e.g. to turn outputs 1 and 3 on you
would set “1010”. Note that the digital inputs/outputs must be configured in the global setting
menu.
Other values that are not used for pressure calibration may be seen if display windows have
been configured that need additional settings.
Display Minimum
Enter the display value that corresponds with the minimum value for the input.
Display Maximum
Enter the display value that corresponds with the maximum value for the input.
Jet Flow
When Jet Flow is selected the flow measurements are corrected for gas jets. Jet flow has the
following calculation applied in addition to the standard flow corrections for temperature,
pressure and viscosity:
Test Gas
LFE flow calculations require a value for relative viscosity, and mass flow calculations require
a value for relative density. Select a gas type from the list to provide these values, or select
‘Custom’ to allow relative viscosity, relative density to be entered manually.
Custom, Air, Ammonia, Argon, Butane-I, Butane-n, Carbon Dioxide, Carbon Monoxide,
Ethane, Ethylene, G20 test gas, G25 test gas, Helium, Hydrogen, Methane, Neon,
Nitrogen, Nitrous Oxide, Oxygen, Propane, Sulphur Dioxide.
Relative Viscosity
Only shown when the gas type is custom.
The laminar flow element differential pressure output is dependent on the viscosity of the
gas. This value is the gas viscosity relative to air, so air is 1.000. When using compatible
gases other than air enter the appropriate value for the gas.
The range of values is from 0.100 to 3.000.
Flow Type
The basic measurement from the laminar flow element is volumetric flow but the FCO560 can
derive other values from this.
Volume Flow.
Mass Flow: the gas type or relative density must be set.
Velocity: the duct area must be set.
Power: the calorific value of the gas must be set.
Flow @
Used when volume flow is displayed.
Select from Standard or Actual.
The FCO560 normally calculates flow at standard conditions. If the actual flow at working
conditions is required then it can be selected here. When Actual flow is being displayed
the flow units will be prefixed with the letter ‘A’, for example AL/m.
Velocity Units
Used when velocity is displayed.
m/s, mph, ft/s, ft/m, km/h, knots
Power Units
Used when power is displayed.
Watts, kW, Btu/m, Btu/h, kBtu/h.
Calorific Value
Used when power is displayed.
3
Enter a value in MJ/m .
Area Units
Used when velocity is displayed.
2 2 2 2
cm , m , in , ft
Temperature Sensor
The temperature of the gas in the laminar flow element is required for flow calculations. This
would typically be from a temperature transmitter connected to the front panel T connector
Select from a list of temperature sources.
Standard Pressure
The standard pressure is the reference absolute pressure used for flow calculations.
Select from a list of pressure sources.
To work to standard conditions you can configure one of the pressure sources to use a
constant value such as 1013.25mb and name it “Standard Pressure”.
It is also possible to show flow at other conditions using either a constant value or a
pressure transducer.
Standard Temperature
The standard temperature is the reference temperature for flow calculations.
Select from a list of temperature sources.
To work to standard conditions you can configure one of the temperature sources to use
a constant value, for example 15°C, and name it “Standard Temperature 15°C”.
It is also possible to show flow at other conditions using either a constant value or a
temperature transducer.
Outputs
Digital outputs may be set for each instrument definition, e.g. for external valve selection.
Each of the four outputs may be selected on (1) or off (0), e.g. to turn outputs 1 and 3 on you
would set “1010”. Note that the digital inputs/outputs must be configured in the global setting
menu.
Other values that are not used for Flow calibration may be seen if display windows have been
configured that need additional settings.
Display Minimum
Enter the display value that corresponds with the minimum value for the input.
Display Maximum
Enter the display value that corresponds with the maximum value for the input.
K Factor
The K factor is a pitot tube scaling factor. Different designs of pitot tubes give different
differential pressures for the same flow.
The standard Furness Controls pitot has a K factor of 1.0.
The K factor limits are from 0.5 to 3.0.
Test Gas
Pitot calculations require a value for relative density. Select a gas type from the list to provide
this value, or select ‘Custom’ to allow relative density to be entered manually.
Custom, Air, Ammonia, Argon, Butane-I, Butane-n, Carbon Dioxide, Carbon Monoxide,
Ethane, Ethylene, G20 test gas, G25 test gas, Helium, Hydrogen, Methane, Neon,
Nitrogen, Nitrous Oxide, Oxygen, Propane, Sulphur Dioxide.
Relative Density
Only shown when the gas type is ‘Custom’.
The relative density is used when measure velocity and for mass flow calculations.
The range of values is from 0.100 to 3.000.
The relative density for air is 1.000.
Flow Type
The basic measurement from the pitot static tube is velocity but the FCO560 can derive other
values from this.
Volume Flow: the duct area must be set.
Mass Flow.
Velocity.
Power: the calorific value of the gas must be set.
Velocity Units
Used when velocity is displayed.
m/s, mph, ft/s, ft/m, km/h, knots
Power Units
Used when power is displayed.
Watts, kW, Btu/m, Btu/h, kBtu/h.
Calorific Value
Used when power is displayed.
3
Enter a value in MJ/m .
Area Units
Used when volume flow, mass flow, or power is displayed.
2 2 2 2
cm , m , in , ft
Pressure Sensor
The absolute pressure of the gas at the pitot is required. This would typically be from an
absolute pressure transmitter connected to the front panel P connector.
Select from a list of pressure sources.
Temperature Sensor
The temperature of the gas at the pitot is required. This would typically be from a temperature
transmitter connected to the front panel T connector
Select from a list of temperature sources.
Standard Pressure
The standard pressure is the reference absolute pressure used in flow calculations.
Select from a list of pressure sources.
To work to standard conditions you can configure one of the pressure sources to use a
constant value such as 1013.25mb and name it “Standard Pressure”.
It is also possible to show flow at other conditions using either a constant value or a
pressure transducer.
Outputs
Digital outputs may be set for each instrument definition, e.g. for external valve selection.
Each of the four outputs may be selected on (1) or off (0), e.g. to turn outputs 1 and 3 on you
would set “1010”. Note that the digital inputs/outputs must be configured in the global setting
menu.
Other values that are not used for Flow calibration may be seen if display windows have been
configured that need additional settings.
Description
16 characters. This is used to identify the profile when it is selected for use within the
instrument definition menu.
No. Samples
Each calibration point may be a single reading or an average of a number of samples taken at
400ms intervals.
The number of samples can be set from 1 to 50.
Point 1 … Point 21
Enter a percentage of Full Scale for each calibration point to be used.
Positive and negative values can be entered up to 150% of the EUT’s full scale. Note that
values over 100% only apply when the EUT’s full scale is less than the FCO560’s full
scale.
The calibration screen displays a target value based on this percentage but you are not
obliged to calibrate at the suggested target. Then Fn menu may be used to override the
suggested value.
Description
16 characters, for example a reference number, manufacturer, type, and range.
Input Type
Define the type of pressure source.
Constant Value
External Gauge Sensor (front panel P connector)
External Absolute Sensor (front panel P connector)
Internal Absolute Sensor (only if option is fitted)
Pressure units
Select the units for the definition of this source.
kPa, mbar, bar, PSI, inHg, inHg, mmWC, inWC
These units only apply to this menu to set the scaling of this source. The FCO560 will
convert from these units into other pressure units as required when this source is used.
Constant
Only when the input type is set to Constant Value.
This may be used for setting a fixed value if a sensor is not fitted or for setting the
standard pressure.
Input @ 4mA
Only shown if an external sensor selected above
Enter the calibrated value for the sensor being used at an output of 4mA.
For the internal absolute sensor this should normally be set to zero.
Input @ 20mA
Only shown if an external sensor selected above
Enter the calibrated value for the sensor being used at an output of 20mA.
For the internal absolute sensor this should be normally set to 160kPa.
Description
16 characters. For example a reference number, manufacturer, type, and range.
Input Type
Define the type of temperature source.
Constant Value
External Sensor (front panel T connector)
Pt100 Input
Temperature units
Select the units for the definition of this source.
°C, °F, K
These units only apply to this menu to set the scaling of this source. The FCO560 will
convert from these units into other temperature units as required when this source is
used.
Constant
Only when the input type is set to Constant Value.
This may be used for setting a fixed value if a sensor is not fitted or for setting the
standard temperature.
Input @ 4mA
Only shown if external sensor selected above
Enter the calibrated value for the sensor being used at an output of 4mA.
Input @ 20mA
Only shown if external sensor selected above
Enter the calibrated value for the sensor being used at an output of 20mA.
Wire Configuration
Only shown if Pt100 is selected.
Select 3-wire or 4-wire.
STANDARD R0 A B
-3 -7
IEC 751 100.000 Ohms 3.9083 x 10 -5.7750 x 10
-3 -7
DIN 43760 100.000 Ohms 3.9080 x 10 -5.8019 x 10
-3 -7
AMERICAN 100.000 Ohms 3.9692 x 10 -5.8495 x 10
-3 -7
ITS-90 100.000 Ohms 3.9848 x 10 -5.8700 x 10
When using an LFE its calibration data must be entered. This is often just the full scale differential
pressure and flow values. For greater accuracy the FCO560 can also correct for non-linearity by
interpolating between up to fifteen known calibration points.
Description
A 16 character description of the flow element can be entered here, typically indicating the
nominal flow range of the LFE.
Select the Fbus communication port and configure it to match the computer that the FCO560 is
connected to.
Port
Select a communication port to use.
Disabled: Fbus communications is not available.
RS232: Uses the 9 way D rear panel connector.
USB: Uses the front panel USB-B (square) connector. It may be necessary to install a
USB driver on the computer – see the Furness Controls product CDROM.
I/D number
Assign a unique identification number to this instrument for Fbus communications in the range
01 to 99.
Baud
The baud rate is the speed that data is sent via the serial ports. The baud rate must be the
same on the FCO560 and device being connected. This is not used for USB.
300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200.
Parity
Select the number of data bits and type of parity error checking to use when sending data via
the RS232 port. Parity is not used for the USB port. The parity setting must be the same on
the FCO560 and the computer being used.
8,NONE: 8 data bits and no parity.
7,ODD: 7 data bits plus odd parity.
7,EVEN: 7 data bit plus even parity.
8,ODD: 8 data bits plus odd parity.
8,EVEN: 8 data bits plus even parity.
Note that extended symbols/characters require 8 data bits. European characters are a
subset of code page 1252. Cyrillic characters are a subset of code page 1251.
Generally, 8 data bits with no parity is preferred. Only use 7 data bits when compatibility
with older equipment is necessary and parity is required and you do not use extended
characters.
Port
Select a serial port to use.
Disabled: the serial port is not available.
RS232: Uses the 9 way D rear panel connector.
USB: Uses the front panel USB-B (square) connector. Note that this will not drive a USB
printer because the FCO560 is a USB peripheral device, however it may be useful to
print-out to a PC. It may be necessary to install a USB driver on the computer – see the
Furness Controls product CDROM.
Baud
The baud rate is the speed that data is sent via the serial ports. The baud rate must be the
same on the FCO560 and device being connected. This is not used for USB.
300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200.
Parity
Select the number of data bits and type of parity error checking to use when sending data via
the RS232 port. Parity is not used for the USB port. The parity setting must be the same on
the FCO560 and the computer being used.
8,NONE: 8 data bits and no parity.
7,ODD: 7 data bits plus odd parity.
7,EVEN: 7 data bit plus even parity.
8,ODD: 8 data bits plus odd parity.
8,EVEN: 8 data bits plus even parity.
Note that extended symbols/characters require 8 data bits. European characters are a subset
of code page 1252. Cyrillic characters are a subset of code page 1251.
Generally, 8 data bits with no parity is preferred. Only use 7 data bits when compatibility with
older equipment is necessary and parity is required and you do not use extended characters.
Data Format
Set the output format for the measurement mode logger when logging to the serial port.
Table: Each item is output on a new line.
CSV: Each record is output on a new line. Each item is separated by the list separator
character.
CSV Characters
The decimal character and list separator may be chosen to suit your location. These are only
used for output sent to the serial port.
., decimal point and comma list separator
,; decimal comma and semicolon list separator
The measurement mode data logger can store data to internal memory and/or a serial port. Data can
be logged on demand from the front panel store button, from the serial port, at timed intervals, or from
an external switch or logic signal.
Data logged to the serial port can be in table format or CSV format.
Data logged to the internal memory can be exported in CSV format to the serial port, or extracted to a
CSV file using Fbus Communications and the Fbus Utility.
Data logged to the internal memory is stored in a compressed form. The maximum number of log
records is dependent upon the data types and the number of items logged for each record. For
example if just the main reading is logged without a title then >350000 records may be stored; if all
possible options are logged with titles this may be reduced to around 25000 records. When the
internal memory fills the logger will overwrite the oldest data. The warning system can be used to get
notice in advance of this happening if desired.
Trigger
Data is stored when a trigger event occurs. Select the source of the trigger.
Front Panel: the Store button
Serial Port: trigger command from serial port
Timed: built-in timer
External Input: programmable digital / switch input
Interval
Set the data logging interval when the trigger source is set to Timed.
From 0.1s to 3600.0s
Log Data To
Choose where the data gets logged to.
Memory
Serial Port
Memory+Serial
The following menu entries have Yes/No options to enable/disable whether the value is logged or not.
Titles
Logging the titles of each value is clearly helpful when looking at the stored data; however,
since they are stored with every logged record they may consume significant resources. It
may be useful to disable the logging of titles, particularly if it is necessary to log a large
number of records to internal memory.
Time
Enable or disable logging of the time.
Date
Enable or disable logging of the date.
Main Display 1
Enable or disable logging of the top main display value.
Window 1
Enable or disable logging of the top display window.
Window 2
Enable or disable logging of the middle display window.
Window 3
Enable or disable logging of the bottom display window.
Aux. Temp.
Enable or disable logging of the Auxiliary Temperature value. See Global Settings.
Aux. Press.
Enable or disable logging of the Auxiliary Pressure value. See Global Settings.
Latest Values
Any warnings that are currently active will flash.
Logged Data: This may be used to warn that the data logger memory is nearly full so that the
data may be off-loaded.
Calibration Due: It is recommended that the calibration of this instrument is checked
annually, or as often as your quality system demands. This value is a display of the number of
days left before calibration is due. If the value is negative then calibration is over-due.
DP Zero: If an overload situation occurs the instrument will attempt to protect the transducer
by equalising the ports. However, the DP transducer is sensitive and its zero point may shift if
it is badly overloaded. The instrument is able to tolerate this to a large degree, but eventually
may stop with a fault.
Limits
Logged Data: set as a percentage of the memory capacity.
Calibration Due: how many days’ notice is required in advance of the calibration due date.
The calibration due date is set in the utilities menu.
DP Zero: set as a percentage of the fault level.
Lock
Lock or unlock data entry. When locked you will be asked for a security code before changing
data.
No: unlocked.
Supervisor: all settings are locked.
Operator: all settings are locked except those that an operator requires.
Security Code
Allows the user to set a 4 digit personal security code.
Time
The time is set in 24 hour format as hh:mm:ss
D.P. Generation
Only displayed if the pressure generation option is fitted.
This allows direct manual control of the pressure generation.
A value is displayed in the d.p. units of the currently selected instrument definition. The
target value is changed in the same way as numbers are changed within the menus.
Use the Fn menu to equalise/read, turn control on/off and turn damping on/off.
If the instrument is powered down in this mode it will return to this mode when power is
restored.
Data Logger
The FCO560 has two independent data loggers. This menu applies to either the calibration
mode data logger or the measurement mode data logger depending upon the mode currently
set in the Global Settings menu.
Erase Data
You will be prompted for confirmation before all logged data for the current mode is
erased.
Print Data
This allows menu data to be sent to the serial port. Choose what to output. For menus with
multiple data sets you can choose the range of data sets to print.
Copy Data
For menus with multiple data sets, this allows data to be copied from one entry to another to
minimise the amount of setting up. This includes instrument definitions, calibration profiles,
pressure sources, temperature sources and laminar flow elements.
Initialise Data
Restore all settings back to factory defaults. You will be prompted for confirmation before any
action is taken.
Service Centres
Lists contact details for key service centres.
System Information
Lists system details for this instrument, including firmware version, range, options etc.
Flash Update
Initiate a product firmware upgrade. You will be prompted for confirmation before any action is
taken.
It is recommended that all instrument settings are saved using the Fbus Utility before updating
firmware. Major firmware updates normally initialise all settings to factory defaults.
Once the procedure is started the upgrade process is commanded from the PC using either
the RS232 or USB-B port, using the Fbus Utility program from Furness Controls. To cancel
flash update mode, press or command it from the Fbus Utility.
Note that flash update is disabled if either the calibration mode or measurement mode data
loggers have stored data. You should export the logged data and then erase it in both modes
before updating the instrument firmware.
A 4-20mA or voltage output process transmitter or instrument that the FCO560 can measure.
A Furness Controls leak detector or flow meter that can be read digitally via USB, or a USB to
RS232 adaptor.
An analogue gauge or other device using manual keyboard entry of the devices readings.
Notes:
The percentage error can be displayed by setting one of the windows to show “Difference”.
The Fn button gives a context sensitive menu to allow selection of the instrument to calibrate,
as well as other options such as re-sampling and damping on/off.
The displayed target value is based on the active calibration profile but can be manually
overridden via the Fn button.
Press the ‘Store’ button to log the data for the current point.
When the calibration profile is set for a number samples the CalPt is replaced with a count-
down as the samples are being taken. You can abort sampling process by pressing before
completion.
For equipment that cannot be read directly such as analogue gauges, it is possible to enter
the value manually. In this case the lower reading will be replaced with a prompt “Manual
EUT”. When you press the ‘Store’ button you will be prompted to enter the reading from the
gauge via the key pad.
Once a point has been stored it is protected from accidentally being overwritten; if you need to
re-take the point you must select Enable Re-sampling from the Fn menu. Re-sampling is
deactivated when a new instrument is selected, or when power save operates or the FCO560
is powered down.
The bar at the bottom of the display can be displayed as centre zero or left-hand zero – see
the global settings menu. When the EUT is split scale, such as -50 to +150Pa, or offset scale
such as 500 to 1000Pa then the bar scale ±100% points are simply ± the largest value, i.e.
±150Pa or ±1000Pa in these examples.
Typical configuration showing connections for both an analogue output and communications
from a FCO332.
Test
Adapter
FCO332 Adaptor
FCO332 FCO332
- + COM - + COM
FCO301
Instrument
Reference source:
DP or Custom DP
Typical configuration using the USB port on a FCO750 to record the leak detector reading.
ABS-P FCO560
Note: + To -
Due to leak
+ -
FCO750
USB-B
When using the USB port on a leak detector, the FCO560 input source must be set to
“communications” and the protocol set to “Fbus”. The baud rate, parity and Fbus I/D must match the
settings in the leak detector.
Instrument
Reference source:
DP or Custom DP
ABS-P FCO560
Test
Adapter
FCO301
FCO332
- + COM
P
Tx
T
T
Tx
Controlled
Flow + - + - Atm.
Source Ref. E.U.T
Instrument
ABS-P FCO560
ABS-P FCO560
Test
Adapter
P
Tx
Pressure
Transmitter
Reg Reg
Out In
+ P
Tx FCO750
Pressure
Source USB-B
When using the USB port on a leak detector, the FCO560 input source must be set to
“communications” and the protocol set to “Fbus”. The baud rate, parity and Fbus I/D must match the
settings in the leak detector.
Instrument
Notes:
The Fn button gives a context sensitive menu to allow selection of the instrument that
determines the measurement mode, as well as other options such as damping on/off.
The bar at the bottom of the display can be displayed as centre zero or left-hand zero – see
the global settings menu.
When timed logging is in progress the word “Logging” is displayed. When another trigger type
is used the word “Logged” appears briefly when a trigger occurs.
DP Measurement
Typical configurations showing differential pressure measurement.
ABS-P FCO560
To Process
Instrument
Reference source:
DP or Custom DP
ABS-P FCO560
P
Tx T
Tx
Flow + - Flow
Ref.
Flow measurement
Instrument
Reference source:
Laminar Flow
Element
ABS-P FCO560
P
Tx
+
T
Tx
-
Instrument
Reference source:
Pitot
Temperature (T)
1 M1852 Cable P5
P1 +24V Red
5 P2 mAin Blue +
2
P3 0V Green Temperature
3 4 P4 0V Yellow 4 to 20mA
-
P5 +15V White
P2
Pressure (P)
M1852 Cable P5
1 P1 +24V Red
P2 mAin Blue +
2 5
P3 0V Green Pressure
P4 0V Yellow 4 to 20mA
3 4 P5 +15V White -
P2
Pt100
P1 P1
P4
M1851 Cable
1 4
P1 V+ Red
P2 V- Blue
3 P3 I- Green
2 P2 P2
P4 I+ Yellow
P3 P3
4-Wire 3-Wire
Battery Charging
The 12Vdc supply is normally used to charge the internal batteries.
An optional in-car charging lead is also available (plug into cigarette lighter socket).
Input/Output
Pin Function Function Pin
3 Input/Output 3 +5V 7
4 Input/Output 4 0V 8
Pins 1-4 are digital input or outputs depending on their assignment in Global Settings.
As outputs they are active high 5V logic levels with approximately 1K in series.
As inputs they have an internal weak pull-up to +5V allowing them to be used with a
simple volt-free contact to 0V. Alternatively, a 5V logic level may be used with the active
state being low.
The 0 to 10V analogue output. is configured in Global Settings as 0 to +FS, 0 to –FS or –FS to
+FS when 5V equates to zero signal. Full Scale is the larger value of DP Min/Max (or Display
Min/Max depending on reference type), i.e. it matches the analogue bar display. The output is
updated every 50ms (20/sec). If damping is enabled then the analogue output is damped.
Pin 7 is connected to +5V via a 100 resistor.
Pin 8 is 0V that may be used with volt-free contacts for inputs.
Note: the digital i/o is 5V only. Application of more than 12V may damage your instrument.
The RS232 drivers employ a power-saving shutdown feature such that the output signals are only
active when a valid RS232 input signal is detected.
Note: RS232 connection to the instrument must be made using screened cable with the screen
connected to a metallic shell of the 'D' connector at the instrument end.
Test Adaptor
Circular Screw
Plug Terminal
P4 +24V Terminal block
(M1839) P1 0V suitable for stranded
P2 +15V or solid wire of
P8 0V Ø1.5mm (16 to
P6 -15V 28AWG) maximum.
P5 +Vin
P3 -Vin
P7 mAin
-Vin and +Vin are differential inputs. –Vin should normally be connected to 0V, ideally at the EUT but
at the test adaptor is normally acceptable. mAin has a burden resistor to 0V giving a voltage loss of
less than 0.5V.
4 to Isolated 4 Isolated
8-Way
20mA Voltage Output to 20mA Voltage
Screw
O/P O/P O/P
Terminal
2-Wire 3-Wire 4-Wire 4-Wire 4-Wire
+24V 1 1
0V 2 2
+15V 1 1 1
0V 3 3 3 3
-15V 2
FCO318 Adaptor +Vin 4 4 4
(M1845) -Vin 3 3 3
mAin 2 4
4 to Isolated 4 Isolated
8-Way
20mA Voltage Output to 20mA Voltage
Screw
O/P O/P O/P
Terminal
2-Wire 3-Wire 4-Wire 4-Wire 4-Wire
+24V 2 2
0V 1 1
+15V 2 2 2
0V
FCO332 Adaptor -15V 1
(M1846) +Vin 3 3 3
-Vin
mAin 1 3
Isolated
4 to Isolated
8-Way 4 to
20mA Voltage Output Voltage
Screw 20mA
O/P O/P
Terminal O/P
2-Wire 3-Wire 4-Wire 4-Wire 4-Wire
+24V 6 6
0V 5 5
+15V 6 6 6
0V 4 4 4 4
-15V 5
FCO342 (Panel Mount) Adaptor +Vin 3 3 3
(M1848) -Vin 4 4 4
mAin 5 3
4 to
8-Way
20mA Voltage Output
Screw
O/P
Terminal
2-Wire 3-Wire 4-Wire
+24V
0V
+15V 3 3 3
FCO352 Adaptor 0V 1 1
(M1849) -15V 4
+Vin 2 2
-Vin 1 1
mAin 4
8-Way 4 to 20mA
Screw O/P
Terminal 2-Wire
+24V
0V
+15V 1
FCO354 Adaptor 0V
(M1850) -15V
+Vin
-Vin
mAin 2
When the FCO560 is in Measure Mode, commands S and T work in the same way as the FCO510.
Normal (table) and CSV formats are supported. The data logger trigger must be set to ‘Serial Port’ to
use the S and T commands. The returned data is similar to the FCO510, but not identical: notably the
date format, and spelling/abbreviation of engineering units may be different. Use the Data Logger
menu to select what data to log. Note that for compatibility with the FCO510, data is only returned
when the instrument is in read mode (not equalised).
If any of the three windows is used its contents is included after the usual PPC500 data.
The returned data is not dependent on the format option in the Serial Port Settings menu – it is always
CSV format.
Measurement inputs
3 x current inputs: 4-20 mA (0.001mA resolution)
Ranges
1 x Voltage input: ± 20 VDC (0.1mV resolution)
4 to 20 mA input: < 0.025% + 0.002 mA
Accuracy @ 20°C
±20 VDC Input: < 0.025% + 0.2 mV
PT100 input Measurement range: 0 to 50 °C (0.01°C resolution) Accuracy <0.025% + 0.004Ω
Absolute Pressure Range: 600 to 1600 mbar absolute. (0.01 mbar resolution) Accuracy < 0.5%FS
Electrical
12 VDC ± 10% 2A (Universal mains power supply included)
Supply Voltage
2.1mm power socket
EUT: 8-pin socket ( voltage + current)
External Pressure Sensor: 5-pin socket (4-20mA)
Measurement inputs
External Temperature Sensor: 5-pin socket (4-20mA)
External PT100 Sensor: 4-pin socket (3 or 4-wire connection)
RS232: 9-pin D plug
Communication
USB: Type A connector for connection to EUT or memory stick
Interfaces
USB: Type B connector for connection to PC
Digital I/O: 4 inputs/outputs (5V logic)
Interface
Analogue output: 0 to 10VDC
General
Enclosure Aluminium enclosure with tilt feet
Display Blue backlit graphic display.
Data logger capacity Storage of approximately 35,000 single display readings.
Dimensions W 250mm x H 105mm x D 265mm (without handle)
Operating temperature +5°C to +45°C
Storage Temperature -10°C to +70°C
Battery life > 8 hours of normal use
Weight <4.5 kg
Long Term Drift (span) < 0.25% reading per year
Overload 20 times fsd without damage
Media Compatibility Clean dry air or non-corrosive gas
Without internal pump: -1 bar to + 4 bar.
Static pressure:
With internal pump: -500mbar to + 500 mbar.
Test Ports: Push-on tube fittings for 6mm o/d, 4mm i/d flexible tube.
Pneumatic Connections
Absolute sensor port: M5 female thread.
Basic unit (Without pressure generator or battery)
Automatic pressure generator
Internal Absolute pressure sensor
Options:
Internal battery pack
Soft carry case
Laminar flow elements and pitot static tubes for flow / velocity measurement.
All information in this document is provisional and is subject to change without notice.
272
115
256
A
E
Absolute Sensor, 30
Adaptor Cables, 53 Erase Data, 40
Adjust FCO3xx calibration, 41
Analogue Bar Display, 17
Analogue Output, 51 F
Analogue Output Scale, 17
Area Units, 24, 27 Fbus ID, 19
Atmospheric Pressure, 15 FCO560 ID., 15
Auto Line Feed, 34 Filter, 15
Automatic Zero Time, 17 Flash Update, 41
Aux. Pressure, 16, 36 Flow @, 24, 27
Aux. Temperature, 15, 16, 36 Flow Point, 32
Flow Type, 24, 26
Flow, Actual, 24, 27
B Flow, Standard, 24, 27
Format, 34
Battery Charging, 8, 51 Front Panel Connectors, 50
Baud Rate, 19, 33, 34
Brightness, 17
G
C Gauge Pressure, 15, 22
Gauge Sensor, 30
Calibration Due Date, 40
Calibration Points, 29
Calibration Profile, 18 I
Calibration Screen, 9
Calorific Value, 24, 27 I/D number, 33
Carry Handle, 8 Info Line, 15
Constant Value, 30, 31 Initialise Data, 40
Contrast, 17 Input @ 20mA, 30, 31
Copy Data, 40 Input @ 4mA, 30, 31
Cross Section Area, 24, 27 Input Maximum, 18
Custom Differential Pressure, 20 Input Minimum, 18
Custom Units, 21 Input Source, 18
Input Type, 30, 31
Input/Output Connector, 51
D Instrument ID., 18
Interval, 35
D.P. Generation, 40
D.P. Maximum, 20
D.P. Minimum, 19, 20 J
D.P. Point, 32
D.P. Units, 19, 20 Jet Flow, 23
Damping, 15, 42, 47
Data Logger, 35, 40 K
Datalog Interval, 35
Datalog Trigger, 35 K Factor, 26
Date, 35, 39
DC Supply, 4, 8, 51
Decimal Point Position, 20, 22, 23, 26 L
Description, 18, 29, 30, 31, 32
Description, General, 5 Laminar Flow Element, 23
Difference, 18 Language, 15
Differential Pressure, 19 LFE calibration temperature, 32
Digital I/O, 16, 17, 51 LFE calibration units, 32
Digital Outputs, 20, 21, 22, 25, 28 Lock, 38
Display Maximum, 21, 22, 23, 26 Log Data To, 35
R
W
Rear Panel Connectors, 51
Reference Source, 19 Window, 15, 36