Hudson RCI
Hudson RCI
User’s Manual
88924-07 06/07
English (EN)
REF NO. DESCRIPTION
425-91 ConchaTherm Neptune Power Cord: Europe, Brazil, South Korea, Chile
2. Symbols
4. Principles of Operations
a. ConchaTherm Basics
b. Concha-Column Descriptions
5. Specifications
a. ConchaTherm Neptune Performance Characteristics
b. Concha-Column Performance Characteristics
c. General System Characteristics
8. Applications
11. Appendix
a. Electromagnetic Emissions Table
1
1. Introduction
Intended Use
The ConchaTherm® Neptune™ (also referred to herein as ‘humidifier’) is a respiratory humidifier
designed to heat and humidify respiratory gases delivered via endotracheal tubes, nasal cannula or
face masks to adult, pediatric and infant/neonatal patients. This system may be used with either
conventional (non-heated wire) breathing circuits or compatible (21-volt) Hudson RCI heated-wire
circuits.
The ConchaTherm Neptune can be used with ventilators, continuous flow systems, oxygen diluters
and blenders, adjustable nebulizer adapters for aerosol therapy, or non-flammable anesthesia gases
to help maintain patient body temperature.
Product Features
• Automatic operation with factory defaults or previous use selections for mode, proximal
airway temperature, temperature gradient and pause time, or fully adjustable, if desired.
• User selectable Supra- and Sub-Glottic airway modes
• May be used with heated-wire or non-heated wire circuits
• Automatic detection of adult, pediatric and infant heated-wire circuits
• Digital temperature display
• Heater-to-patient temperature gradient for control of condensation
• Enhanced temperature gradient to accommodate higher inlet gas temperatures
• Responsive membrane front panel
• Internationally accepted Graphic Icon Front Panel
• LED prompts and audio-visual alarms permit rapid detection and correction of problems
• Temperature tracking alarms at 2°C above or below the selected airway temperature.
• Low Airway tracking alarm not activated until system reaches Airway Set-Point (within 2°C)
• Fixed high-temperature alarm at 41.0°C
• Disconnected/dislodged probe indicators
• Heated-wire disconnect alarm when using heated-wire circuits
• User adjustable pause time from 5 to 60 minutes.
• Continuous self-diagnostic evaluation of hardware and software
• Low Humidity Advisory
Setup Diagrams
FROM VENTILATOR HEATED WIRE EXPIRATORY HEATED WIRE
(YELLOW) (IF USED) BREATHING CIRCUIT
PLACE COLUMN
TO VENTILATOR PROBE HERE FROM VENTILATOR
TO VENTILATOR
CONVENTIONAL CONCHA COLUMN PLACE COLUMN
BREATHING CIRCUIT PROBE HERE
CONCHA 1650ml
RESERVOIR BOTTLE CONCHA COLUMN
Figure 1: The CochaTherm Neptune Using a Figure 2: The CochaTherm Neptune Using a
Conventional Circuit Heated-Wire Circuit
2
2. Symbols
The following symbols are noted on the device labeling. (See Controls and Indicators section of this
manual for descriptions of Front Panel and other Icons used in this manual)
Symbol Description
A Amperes
HZ Hertz
3
3. General Warnings and Cautions
Warning: For safe operation, only use Hudson RCI Branded Concha Columns, Water Reservoirs,
Temperature Probes and 21-Volt Heated Wire Circuits, as listed in the Hudson RCI Product Catalog.
Warning: Fire Hazard. Do not use the humidifier in the presence of flammable anesthetic gases.
Warning: Burn Hazard. The heating element of the humidifier is hot (as high as 140°C). Allow the
humidifier to cool before handling.
Warning: Burn Hazard. The Concha-Column is hot (as high as 140°C) and may contain liquid at
temperatures up to 80°C. Allow the column to cool before handling.
Warning: Electrical Shock Hazard. Refer all servicing to qualified, trained personnel only.
Warning: This humidifier requires 100-120V ~, (nominal) 60 Hz sine wave output. If used in a
transport vehicle requiring an inverter, do not use square or pulse width modulated sine wave output.
To do so may result in overheating.
Warning: The temperature probes must be properly placed in the breathing circuit before operating
the humidifier. Failure to properly connect the temperature probes will cause the unit to alarm and
shut down.
Warning: Always verify gas flow and airway temperature before connecting the humidifier to the
patient.
Warning: Maintain adequate gas flow through the column and breathing circuit. This will prevent
overheating the column and breathing circuit.
Warning: Always monitor tidal volume, minute volume, respiratory rate and all pressures and ensure
that all monitoring alarms are appropriately set and functioning before connecting patient to the gas
delivery system (ventilator, humidifier and circuit).
Warning: Do not leave the humidifier ON, do not turn the humidifier ON and do not exit the pause
mode until there is regulated gas flow through the system. To do so may result in heat buildup,
causing a bolus of hot air to be delivered to the patient. Circuit tubing may significantly soften under
these conditions. Turn the humidifier OFF or place the unit in pause and allow the system to cool
before stopping gas flow.
Warning: The probe alarm is not intended to replace routine physical inspection of the temperature
probe placement. Be sure both probes are in place during operation of heater.
Warning: Do not use a Low-Compliance Concha-Column in continuous flow CPAP applications with
gas flows greater than 50 LPM. There is potential for flooding of the breathing circuit and airway.
Warning: Only use the Comfort-Flo Concha-Column in continuous High Flow Oxygen applications.
Do not exceed 40 LPM in this mode. There is potential for flooding of the breathing circuit and
airway.
Warning: Sudden changes in circuit pressure may cause the water in the column to rise temporarily
and may affect compressible volume.
4
Warning: Use of the RS-232 service connector by anyone other than qualified, factory-trained per-
sonnel may damage the instrument and result in harm to the patient. No connection to this port is to
be made when device is operational and connected to a patient. Patient injury may occur.
Warning: The humidifier has been designed and tested for use with Hudson RCI Branded acces-
sories and breathing circuits. Use of non-approved circuits or accessories with this unit may cause
an unsafe condition. See the current issue of the Hudson RCI Catalog or go to the Hudson RCI
Website at www.hudsonrci.com for compatible circuits and accessories validated for use with the
CONCHATHERM Heated Humidifiers.
Warning: Electrical Short Hazard: Do not immerse this device in water or other fluids for cleaning.
Do not autoclave, gas sterilize or pasteurize. Severe damage will result.
Warning: Only connect this device to a Hospital Grade Grounded 120 V ~ outlet, with the power
cord supplied in the original packaging. This device has been developed to operate over the voltage
ranges noted on the name plate label for use in foreign countries. If device is used in areas outside
the United States, ensure a Hospital Grade, grounded compatible power cord is utilized. Please call
Teleflex Medical’s customer service to obtain Foreign Country power cord specification for your appli-
cation.
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Caution: Always operate the humidifier in a vertical position with the water reservoir properly
mounted.
Caution: Temperature patterns within the system associated with normal operation will be disrupted
if there is insufficient water to pass through the system, causing a high temperature alarm.
Caution: Do not operate the humidifier with a dry or disconnected reservoir or without a column
installed. Severe damage may result.
Caution: Hudson RCI Branded circuits and accessories are Single Patient Use only. With the
exception of the Temperature Probes, they are not cleanable or re-useable. Accessories should be
changed after a maximum of 14 days of use, or when noticeably contaminated by the patient or
excessive leakage is observed.
Caution: Replace the power fuse only with one of the same rating.
Caution: Monitor the Inspiratory and Expiratory Circuits periodically for excessive condensation
(rainout).
Caution: In the Invasive, or Sub-glottic Mode, the humidifier is designed to deliver a minimum fluid
output of 30 mg H2O/L at 30º C, for the proximal airway temperatures at zero gradient and with the
recommended Flow Rates listed in the Specification Section of this manual.
• Use of Positive or "Sun" gradient settings in some conditions can result in humidity outputs
at the patient of less than 30mg H2O/L at 30º C.
• Visually inspect for condensation droplets in the inspiratory circuit after one hour of
operation, and periodically thereafter.
Caution: Some Ventilators produce elevated gas temperatures at the patient inhalation port which
can reduce the amount of relative humidity in the circuit. Visually inspect for condensation droplets
in the inspiratory circuit after one hour of operation, and periodically thereafter.
Caution: Monitor the Inspiratory and Expiratory Circuits periodically for excessive condensation
(rainout). Use water traps to collect rainout when using non-heated wire breathing circuits.
5
4. Principles of Operation
ConchaTherm Basics
The servo-controlled heated humidifier continuously monitors the proximal airway temperature and
regulates the heat supplied to the column by the heating element, and if used, the heated-wire
circuit.
The Concha-Column humidification cartridge is inserted into the heater where it is surrounded by a
heating element. Sterile water flows via a gravity-feed system from the reservoir into the
Concha-Column, wetting an internal absorbent wick. The heating element conducts heat through the
column's metal wall, heating the wick and vaporizing the sterile water. Ventilatory gas passes
through the column where it picks up the moisture in molecular form. The heated, humidified gas is
delivered to the patient through a breathing circuit attached to the top of the column.
Concha-Columns are available in four configurations: Standard, Low-Compliance, Minimal
Compliance and Comfort-Flo. All columns are labeled for easy identification and operate on a
reliable, gravity-feed system for uninterrupted water flow.
Concha-Column Description
The Standard Concha-Column (see Figure 1) is designed for use with ventilatory parameters most
commonly used with adults. The water in the column will be at the same level as that in the
reservoir. The water level in both the reservoir and the column drops as water is consumed. As the
water level in the column drops, the compressible volume of the system increases and the system
compliance changes.
Saturated Wick
CONCHA
Prefilled. Disposable
Sterile Water Reservoir
6
The Low-Compliance Concha-Column is recommended for applications requiring a more constant
compressible volume and compliance, (see Figure 2). The Low-Compliance Concha-Column
maintains a constant water level in the column resulting in a more stable compressible volume. The
system's compliance equals the column’s compliance plus the amount of fluid transferred to the
reservoir during each inspiratory phase. Some fluid, in the form of either water vapor or water,
always transfers during each cycle. In addition, because of the smaller mass of water contained
inside the column, the Low-Compliance Column will heat and cool faster as compared to the
Standard Concha-Column.
Saturated Wick
CONCHA
Prefilled. Disposable
Sterile Water Reservoir
The Comfort Flo Concha-Column is for use with continuous high flow oxygen applications. This
column is specially designed to maintain a constant water level in the column under high flow oxygen
applications.
Caution: The Comfort-Flo Column should be used exclusively with continuous high flow
oxygen applications. Use of any other column during high flow oxygen applications may
cause flooding of the circuit.
Note: The amount of fluid transferred to the reservoir varies slightly depending on venti-
lator settings. Several factors increase the amount of fluid transfer: high peak pressure,
lower breath rates, reduced water volume in reservoir, I/E ratios approaching 1:1.4, and
square pressure waveforms.
Note: The ConchaTherm Neptune requires the use of a dual temperature probe at
all times.
7
5. Specifications
ConchaTherm Neptune Performance Characteristics:
Tracking Alarm: ± 2.0°C of the selected airway temperature, after the set temperature has been
reached.
Warm-up time: (Warm up is defined as patient temperature within 2°C of set point)
Heated-Wire 10 minutes
Non-Heated Wire 35 minutes
*The Invasive Mode of the Concha Neptune was designed to provide 100% RH , as evidenced by moisture con-
densation, or “rainout” in the circuit at the recommended flow rates over the settable proximal airway tempera-
ture settings at zero gradient. In some cases, the use of positive gradient settings “ ” or flow rates outside
the recommended ranges may result in increased warm-up times or lower than recommended humidity outputs.
Additionally, some environmental conditions, such as warm ventilator outlet gas temperatures and/or high ambi-
ent temperatures, may require the use of the negative gradient setting ” ” to reach 100% RH, or visible
moisture droplets in the circuit.
Humidity Output:
Invasive Mode > 33 mg H2O/L at 32º C proximal airway temperatures when set at zero
gradient at the recommended Flow Rates listed above.
• Use of Positive or "Sun" gradient settings in some conditions can result in
humidity outputs at the patient of less than 33mg H2O/L at 32º C. Visually
inspect for condensation droplets in the inspiratory circuit after one hour of
operation, and periodically thereafter.
Non-Invasive Mode > 10 mg/L at 32º C proximal airway temperatures when set at zero
gradient at the recommended Flow Rates listed above.
• This mode should only be used when the patient airways ARE NOT
bypassed.
8
Concha-Column Performance Characteristics
Water Capacity:
Standard 193 ± 10 ml with full reservoir
Low Compliance 40 ± 5 ml
Comfort-Flo Low-Compliance: 40 ± 5 ml
Minimal Compliance column: 40 ± 5 ml
Column Compliance:
Standard: 1.12 ml/cm H2O with full reservoir
2.51 ml/cm H2O with reservoir at “replace” mark
Low-Compliance: 0.25 ml/cm H2O
Comfort-Flo Low-Compliance: 0.25 ml/cm H2O
Minimal Compliance: 0.2 ml/cm H2O
Column Compressible Volume:
Standard: 103 ml with full reservoir
268 ml with reservoir at “replace” mark
Low-Compliance: 248 ml ± 12 ml
Comfort-Flo Low-Compliance: 248 ml ± 12 ml
Minimal Compliance: 109 ml
9
Physical Characteristics:
Size: 17.3 cm (H*) x 14.0 cm (W) x 20.6 cm (D)
6.8” (H*) x 5.5” (W) x 8.1” (D)
Weight: 2.8 pounds/1.3 Kg (approximate)
10
6. Controls and Indicators
6 7 4
5
10
3
1 2
2. Invasive Mode - Icon illuminates when the humidifier is in the SUB-GLOTTIC or Invasive
or Intubated mode.
4. Airway Temperature - Icon illuminates green during normal operation. This indicator
flashes red when there is a temperature probe problem or an over temperature condition at
the patient airway.
5. Column Temperature - Icon illuminates green when the column button is pressed and the
column temperature is displayed on the temperature display LED. This indicator flashes red
when there is a temperature probe problem or high temperature condition at the column.
6. Inspiratory Wire Indicator - Arrow icon illuminates green when the inspiratory wire of a
heated-wire circuit is connected to the humidifier. This arrow will flash red when the
inspiratory wire status has changed (connected or disconnected) during operation. This
arrow will flash green if the status (connected/disconnected) is changed during “Confirmation
of Settings or Set-up”.
11
7. Expiratory Wire Indicator - Arrow icon illuminates green when the expiratory wire of a
heated-wire circuit is connected to the heater. This arrow will flash red when the expiratory
wire status has changed (connected or disconnected) during operation. This arrow will flash
green if the status (connected/disconnected) is changed during “Confirmation of Settings
or Set-up”.
8. Pause Indicator - When the humidifier has been placed in pause mode, the icon will
illuminate and the LED display will count down the remaining pause time.
9. Service Indicator - If the heater detects an internal problem during operation, the icon
and the Emergency Alarm Indicator will flash and an audible alarm will sound. If this hap-
pens, turn the heater off and let it cool. If the condition persists, take the unit out of service
and return to Teleflex Medical for servicing.
10. Temperature Gradient Indicator - Indicates the temperature gradient set point. Indicator is
active when using a heated wire circuit and will be illuminated during “Set-up” and
“Check Settings” modes.
STATUS INDICATORS
2. YELLOW - A Yellow flashing icon in conjunction with a low volume, 3-tone audio signal
indicates a cautionary alarm condition (also see “Alarm Conditions” in the “Troubleshooting”
section of this manual).
3. RED - A red flashing icon in conjunction with a high volume, 5-tone audio signal indicates
an emergency alarm condition requiring immediate action by the clinician (see “Alarm
Conditions” in the “Troubleshooting” section of this manual).
12
i
a
c
h
d
b e
g f
b. Mode Selector - This key selects the operating mode (Invasive or Non-Invasive).
c. Temperature and Pause Time Control - The Increase and Decrease Buttons allow
the user to adjust the proximal patient airway temperature and pause time settings.
d. Temperature Gradient - Sun and Raindrop - These keys increase or decrease the
temperature gradient between the column and the patient end of the heated-wire circuit.
e. Gradient Display - Blue LED between the Sun and Raindrop indicates the actual
temperature gradient setting. The example below shows a setting of one degree towards
the raindrop. This setting would increase humidity in the circuit.
Example:
f. Accept Button - Used to check settings, access adjustable settings and accept the
displayed parameters and store settings into memory.
g. Pause Button - Used to enter the Pause Mode for a user adjustable period of 5-60
minutes.
h. Alarm Mute Button - Used to silence a Cautionary Alarm for two minutes.
i. Column Button - When pressed, the LED displays the temperature of the gas at the outlet
of the column. The column temperature will remain displayed as long as the key is pressed.
13
7. Directions for Use
Read this entire section, including all warnings and cautions, before operating the ConchaTherm
Neptune. See the “Applications” section for tips on using the ConchaTherm Neptune in specific clini-
cal situations.
System Setup
1. Installation:
Mount the humidifier to a pole, ventilator or rail system using the appropriate mounting
bracket. Contact your Teleflex Medical representative for available mounting brackets.
Mounting instructions are supplied with the bracket.
Warning: Fire Hazard: Do not use humidifier in the presence of flammable anesthetic
gases.
Warning: This unit requires 120 V ~ (nominal) 60 Hz sine wave output. If used in a
transport vehicle requiring an inverter, do not use square wave or pulse width modulated
sine wave output. To do so may result in overheating.
Warning: The temperature probes must be properly placed in the breathing circuit
before operating the humidifier. Failure to properly connect the temperature probes may
cause the unit to alarm and shut down.
6. Connect the heated-wire pigtails (not applicable when using conventional circuits).
• Connect the inspiratory limb heated-wire (blue) pigtail to the blue
heated-wire connector on the humidifier cable. Connect the expiratory limb
heated-wire (yellow) pigtail to the yellow heated-wire connector on the
humidifier cable.
• Verify heated-wires are connected to the humidifier before turning the
unit on.
Note: The humidifier will automatically enter heated-wire operation when it senses the
inspiratory heated-wire from a Hudson RCI circuit.
Warning: If using a heated-wire circuit with this humidifier, use only Hudson RCI 21-volt-
circuits.
14
Warning: When using Hudson RCI Heated-Wire Circuits:
• Observe the minimum minute volumes recommended in this manual.
• Do not cover the circuit with sheets, blankets, towels, clothing or other materials.
• Do not apply excessive tension to the heated-wire harness — DO NOT "MILK"
TUBING.
• Do not allow the circuit to rest on the patient's bare skin.
8. For next steps proceed to the Automatic Operation and/or User Adjustable Operation
sections of this manual.
FROM VENTILATOR
PLACE COLUMN
PROBE HERE
TO VENTILATOR
CONCHA 1650ml
RESERVOIR
BOTTLE CONCHA HEATED
HUMIDIFIER
PLACE PATIENT
AIRWAY TEMPERATURE
PROBE HERE
HEATED WIRE
CABLES
(NOT USED)
15
Automatic Operation
1. Set up and turn ON the ventilator or gas source (see the Ventilator Manufacturer Instructions
for Use to determine ventilator operation and monitoring requirements).
Warning: Always monitor tidal volume, minute volume, respiratory rate and all pres-
sures and ensure that all monitoring alarms are appropriately set and functioning before
connecting patient to the gas delivery system (ventilator, humidifier and circuit).
Once activated, the humidifier will perform a self-check (a scrolling dash will be seen on the
LED display). When the self-check is complete, the unit will issue an audible alert. This
indicates the humidifier is ready for operation. At this point the user can choose to confirm
or change the settings.
Warning: If the heater detects a fault during the self-check, the icon will flash and
an audible alarm will sound. If this happens, take the unit out of service and return to
Teleflex Medical for servicing.
Warning: Do not leave the humidifier on, do not turn the humidifier on and do not exit
the pause mode until there is regulated gas flow through the system. To do so may
result in heat buildup, causing a bolus of hot air to be delivered to the patient. Circuit
tubing may significantly soften under these conditions. Turn the humidifier power switch
OFF, or place the unit in pause and allow the system to cool before stopping gas flow.
Warning: Always verify gas flow and airway temperature before connecting the humidifer
to the patient.
Note: The humidifier will not begin to heat until the settings have been confirmed.
Note: The humidifier will automatically detect Hudson RCI adult, pediatric and neonatal
heated-wire circuits.
Note: Confirm the heated-wire icons illuminate green when the appropriate heated-wires
are connected to the unit.
PLACE COLUMN
PROBE HERE
CONCHA 1650ml
RESERVOIR
CONCHA COLUMN
BOTTLE
HEATED WIRE
(BLUE)
PLACE PATIENT INSPIRATORY
AIRWAY TEMPERATURE
PROBE HERE
CONCHA HEATED
HUMIDIFIER
TO PATIENT
HEATED WIRE
CABLE (BLUE)
HEATED WIRE
CABLE (YELLOW)
DUAL TEMPERATURE PROBE
16
3. Confirm Settings.
The accept button will flash and the humidifier will automatically display the stored
settings for mode (invasive or non-invasive) and patient airway temperature. When using a
heated-wire circuit, the heated-wire and temperature gradient indicators will illuminate.
Patient airway temperature and temperature gradient will default to the last used values for
the present configuration (mode and breathing circuit type). If the configuration has not
been used previously, factory default settings will apply.
To begin operation using the displayed settings, press and release the accept
button. The heater will begin warm-up and then normal operation.
Note: Power is not applied to the heater or heated-wires until the settings have been
confirmed.
Note: If the user does not press the accept button to confirm these settings or press
and hold the accept button to enter the User Adjustable Operation within 5 minutes, the
unit will alarm. To clear this alarm, press the accept button .
FROM REGULATED
PLACE COLUMN
OXYGEN SOURCE
PROBE HERE
CONCHA COLUMN
PLACE PATIENT
AIRWAY TEMPERATURE
PROBE HERE
HEATED WIRE
INSPIRATORY (BLUE)
CONCHA 1650ml
RESERVOIR BOTTLE
CONCHA
HEATED HUMIDIFIER
HEATED WIRE
EXPIRATORY (YELLOW)
(NOT USED)
TEMPERATURE PROBE
4. Warm-Up.
During warm-up, the °C will flash and the LED display will show the actual proximal patient
airway temperature. Warm-Up ends when the patient airway temperature reaches within
2°C of the set airway temperature. Once warm-up is complete the °C will stop flashing.
Note: Warm-up times consider that warm-up is achieved when the patient airway tem-
perature is within 2°C of the patient airway set temperature.
17
5. Normal Operation.
After warm-up, the humidifier will convert to normal operation. Figure 8 shows a typical
display for the ConchaTherm Neptune during normal operation.
Warning: Do not leave the humidifier on, do not turn the humidifier on and do not exit
the pause mode until there is regulated gas flow through the system. To do so may
result in heat buildup, causing a bolus of hot air to be delivered to the patient. Circuit
tubing may significantly soften under these conditions.
6. Pause Mode
To facilitate activities such as circuit changes and nebulizer treatments, the ConchaTherm
Neptune features a pause mode. During this period, humidifier settings are maintained and
automatically re-activated when the designated pause time ends.
To activate the PAUSE Mode, press and hold the pause button . The timer icon will
illuminate and the LED display will count down the remaining time in pause. The factory
default pause time is 30 minutes.
To exit PAUSE, press the flashing pause button or allow the pause time to expire. Both
exit methods will return the humidifier to warm-up and normal operation.
Warning: Do not exit the pause mode until there is regulated gas flow through the system.
To do so may result in heat buildup, causing a bolus of hot air to be delivered to the patient.
Circuit tubing may significantly soften under these conditions.
18
7. Changing the breathing circuit.
When replacing the breathing circuit with the same type of breathing circuit (heated-wire
being replaced with heated-wire), place the heater in pause mode before changing the
circuit. If replacing the breathing circuit with a different type of breathing circuit (Infant/Adult
or Heated-wire/Non-Heated wire), turn the humidifier OFF before changing the circuit. This
will allow the heater to auto-detect the new circuit without alarming.
Note: The humidifier will give a heated-wire change alarm if PAUSE is exited and the
heated-wire configuration has been changed.
Note: If the circuit type is changed the humidifier will revert to the remembered settings
the last time that circuit was used.
Warning: Do not leave the humidifier ON, do not turn the humidifier ON and do not exit
the pause mode until there is regulated gas flow through the system. To do so may
result in heat buildup, causing a bolus of hot air to be delivered to the patient. Circuit
tubing may significantly soften under these conditions. Turn the humidifier power switch
OFF, or place the unit in pause and allow the system to cool before stopping gas flow.
To turn the humidifier OFF, press and hold the button until the display goes blank.
Allow the unit to cool before stopping gas flow.
Warning: If the humidifier detects an internal problem during operation, the service icon
and the Priority Alarm Indicator will flash and an audible alarm will sound. If this hap-
pens, turn the humidifier OFF and let it cool. If the condition persists, take the unit out of
service and return to Teleflex Medical for servicing.
19
User Adjustable Operation
The ConchaTherm Neptune will arrive with the following factory default settings: invasive
mode, patient airway temperature 37.0° C, temperature gradient 0° C, and pause interval 30
minutes. All of these parameters may be adjusted by the clinician.
To change the displayed settings, press and hold the accept button . Once desired
settings are entered, press the button to accept changes. The heater will begin
Warm-Up and then normal operation.
While the heater is operating you can view the “set” parameters by pressing and releasing
the accept button . The humidifier will display the set mode, airway temperature, and
the temperature gradient, if applicable, for 5 seconds.
Adjustable Parameters
To change the patient airway mode, press and hold the accept button. The currently
selected patient airway mode icon and the mode selector button will be illuminated.
To accept the new mode, press the accept button. The unit will return to
warm-up and then normal operation. The selected mode will be retained in
memory and the appropriate mode icon will remain illuminated during warm-up
and normal operation.
Note: When the mode is changed, the humidifier will remember the last settings used
with that mode and circuit type, or will default to factory settings if used for the first time.
To access the temperature for adjustment, press and hold the accept button .
The LED display will show the set Proximal Patient Airway Temperature and the increase
and decrease buttons will be illuminated.
To increase the airway temperature, press the increase button once for each
degree desired.
To decrease the airway temperature, press the decrease button once for each
degree desired.
To accept the new temperature, press the accept button. The unit will return to
Warm-Up and normal operation.
Note: With each adjustment, allow adequate time for the humidifier to reach the selected
temperature. This will vary and is dependent on flow rate, type and length of tubing,
ventilating volume, respiratory rates and Concha-Column.
20
3. Temperature Gradient (only active if using heated-wire circuits)
When using the ConchaTherm Neptune with heated-wire circuits, the temperature gradient
may be adjusted to allow for the proximal patient airway temperature to be up to 3.0 ºC
warmer or cooler than the temperature at the column outlet. Adjustment of the temperature
gradient allows for better control of condensation or “rainout” within the circuit. A positive
temperature gradient results in an airway temperature that is warmer than the temperature of
the gas at the column. A negative temperature gradient results in an airway temperature that
is cooler than the temperature of the gas at the column. This column-to-patient temperature
gradient is displayed as a blue LED when viewing the “set” parameters.
The ConchaTherm Neptune will arrive from the factory with the column-to-patient tempera-
ture gradient set at 0 ºC. A temperature gradient of 0 ºC allows the temperature of the gas,
and therefore, the relative humidity (RH) of the gas, to be constant throughout the length of
the heated-wire circuit. Figure 9 shows a typical circuit response and LED display when the
airway temperature is set at 37 ºC and the temperature gradient is set at 0 ºC.
To view the set temperature gradient, quickly press and release the accept button.
The temperature gradient the will be displayed for five seconds and will be
represented by a the blue LED between the Sun and Raindrop.
To adjust the temperature gradient, press and hold the accept button. The zero
temperature gradient icon , the blue temperature gradient LED, and the sun
and rain icons will be illuminated.
To set a more positive temperature gradient, press the sun button for each
additional positive degree desired.
Important: A positive temperature gradient may result in a humidity deficit to the patient,
depending on the temperatures and ventilator parameters used.
21
To set a more negative temperature gradient, press the rain button for each
additional negative degree desired.
Note: A negative temperature gradient increases the potential for rainout within the
heated-wire circuit.
To accept the new temperature gradient settings, press the accept button. The
unit will return to warm-up and normal operation. The temperature gradient selected
will be retained in memory.
Note: During normal operation, the temperature gradient LED and buttons are not
displayed.
The AARC Clinical Practice Guideline1 for Humidification during Mechanical Ventilation
states that the humidification method chosen should provide a minimum of 30 mg H2O/L of
delivered gas at 30 ºC. The ConchaTherm Neptune is designed to advise the user of low-
humidity conditions (when the humidifier is in the Invasive mode and the set airway
temperature and gradient result in a column temperature equal to or less than 31 ºC). The
humidifier will advise the user of this condition by displaying a flashing proximal patient air-
way temperature in the “Set-Up” or “Confirm Settings” Mode. An audible alarm will not
sound. In order to address this low humidity situation, the user should either increase the
patient airway temperature or move the gradient closer to the Raindrop.
(See instructions in the User Adjustable Operation section.)
Note: Setting a positive gradient may invoke this advisory condition. For example, a
selected airway temperature of 33 ºC and a Gradient setting of +3 ºC would result in a
column temperature of 30 ºC causing the humidifier to invoke the low-humidity advisory.
Note: The Low Humidity Advisory function is not active in the Non-Invasive mode.
5. Pause Time
To view or adjust the set pause time, press and hold the accept button. The
pause button will illuminate.
1 Branson, RD, Campbell SC, Chatburn RL, Covington J. AARC Clinical Practice Guideline: Humidification during
Mechanical Ventilation. Respiratory Care 1992; 37:887-890. 22
Press the pause button. The set pause time will be shown on the LED Display and
the increase and decrease buttons and the pause indicator icon will
illuminate. The pause time can now be adjusted from 5 to 60 minutes, in 5-minute
increments.
To increase the pause time, press the increase button once for each 5-minute
increase desired.
To decrease the pause time, press the decrease button once for each 5-minute
decrease desired.
To accept the new pause time, press the accept button. The unit will return to
warm-up and normal operation.
Some ventilators generate high internal temperatures during operation, therefore heating the
gas that is delivered to the humidifier. A humidifier responds to this high incoming gas tem-
perature by assuming that less heat is required. As a result less water vapor is created. To
compensate for this situation, the ConchaTherm Neptune features a High Output Mode (see
figure 12). This mode allows the humidifier to respond to high incoming gas temperatures
by increasing the column-to-patient temperature gradient. By increasing this gradient, the
column will generate more heat and therefore produce more water vapor.
Caution: When using the High Output Mode to compensate for high inlet gas tempera-
tures from the ventilator, visually inspect for condensation droplets in the inspiratory
circuit after one hour of operation, and periodically thereafter to ensure adequate
humidity to the patient.
To access and display the High Output mode, press and hold the accept
button to enter the set-up mode,
& To check the current gradient mode, press the sun and rain buttons
simultaneously. If normal gradient mode is active, the LED display will show
three dashes (---). If High Output Mode is active, the LED display will read “HI.”
23
& To change the gradient mode, press and hold the sun and the rain buttons
simultaneously for 3 seconds. The LED display will indicate the mode selected
(--- for Normal and HI for High).
To set a positive temperature gradient, press the sun icon for each positive
degree desired.
To set a negative temperature gradient, press the rain icon for each negative
degree desired.
Note: Adjust the gradient until adequate humidity is visible in the breathing circuit.
To accept the Temperature Gradient selection, press the accept button. The
humidifier will return to warm-up and normal operation. (See Figure 13 for
examples of high gradient settings)
When High Output Mode is active, the gradient icon will flash.
The ConchaTherm Neptune is intended to heat and humidify respiratory gases at a proximal
patient airway temperature selected by the clinician. In addition, the ConchaTherm Neptune
will allow the use of heated-wire circuits to minimize the condensation or "rainout" within the
circuit. Please note that certain environmental conditions will not allow the humidifier to
maintain both the proximal patient airway temperature and the temperature gradient at the
levels entered by the user. Under these conditions, the humidifier will automatically adjust
the temperature gradient in order to maintain the proximal patient airway temperature within
2°C of the selected temperature. Once the environmental conditions allow, the humidifier will
automatically return to utilizing the selected temperature gradient.
Note: In some conditions, the use of sun (or positive) gradient settings can result in
humidity outputs at the patient of less than 33mg H2O/L. Visually inspect for condensation
droplets in the inspiratory circuit after one hour of operation.
24
Other Operating Features
Display Settings
During operation, the humidifier displays the mode, actual proximal patient airway
temperature and the airway temperature and heated-wire icon(s), if applicable.
To display the airway temperature and temperature gradient settings, press and
release the accept button. The humidifier will display the set parameters for 5
seconds (the temperature gradient settings are only applicable when using heat-
ed wire circuits).
Column Temperature
During operation, the humidifier displays actual proximal patient airway tempera-
ture. To display the temperature of the gas at the column, press the column
button. The temperature probe LED will move to the column location and the
column temperature will be displayed as long as the button is pressed. Once
the key is released, the display will resume showing the actual patient airway
temperature.
Alarm Mute
Press the alarm mute button to silence the cautionary audio alarm. The alarm
will be silenced for approximately two minutes, or until the alarm deactivates.
When muted, the alarm remains activated and the alarm mute and cautionary
alarm icons are illuminated. The alarm will deactivate if the alarm condition self-
corrects. Pressing the alarm mute button before the end of the mute period, or
before the condition self-corrects, will reactivate the audible alarm.
Power Interruption
If the electrical power is interrupted while the unit is in operation the humidifier will restore
the settings that were in effect prior to the interruption.
Warning: For safe operation, only use Hudson RCI labeled Concha Columns, Water
Reservoirs, Temperature Probes and 21-Volt Heated Wire Circuits, as listed in the
Hudson RCI Product Catalog.
25
8. Applications
The ConchaTherm Neptune heated humidification system is designed to heat and humidify
respiratory gases delivered to adult, pediatric, and infant patients. The following information
is offered regarding the use of the humidifier in specific applications.
The ConchaTherm Neptune has been tested to operate with ventilators designed to support
neonatal, pediatric, and adult patients using commonly applied ventilation modalities. These
ventilators may incorporate some or all of the following performance characteristics:
Anesthesia
The ConchaTherm Neptune may be used with non-flammable anesthetic gases with either
non-heated wire or heated-wire breathing circuits at the minimum recommended flows for
adults and infants (see Specifications). As flow through the circuit nears the lower limits, the
use of a heated-wire circuit and a Low-Compliance Concha-Column is recommended in
order to maintain circuit temperature and avoid a tracking alarm. The humidifier is NOT
RECOMMENDED in anesthesia applications requiring flows of less than those stated in the
"Specifications". For additional information, see "Low-Flow Applications" later in this section.
When setting up the humidifier for use with a radiant warmer or incubator, place the proximal
patient airway temperature probe away from lamps or other sources of heat so as not to
affect the temperature reading.
Low-Flow Applications
The ConchaTherm Neptune may be used with a heated-wire breathing circuit at the
following minimum continuous flows:
For example, with the humidifier in the adult mode, a volume of three liters of gas must pass
the proximal patient airway temperature probe every minute while the unit is operating. In
low-flow applications, the use of a Low-Compliance Concha-Column is recommended in
order to maintain circuit temperature and avoid a tracking alarm. When using the humidifier
in low-flow applications, the warm-up period may be longer due to the slower heat transfer
through the breathing circuit.
For non-heated wire applications in low flow conditions, the following minimum flows must be
observed to ensure maintenance of the patient airway temperatures. If lower flows are
desired, please use a heated-wire breathing circuit:
26
Transport Vehicles
Warning: This unit requires 120 V ~ (nominal) 60 Hz sine wave output. If used in a
transport vehicle requiring an inverter, do not use square wave or pulse width modulated
sine wave output. Overheating could result.
Operating the humidifier in environments near the extremes of its operating environment
temperature range may result in alarms as the unit attempts to control the airway
temperature as it is affected by heat transfers along the circuit path.
27
9. Troubleshooting and Maintenance
ALARMS
The ConchaTherm Neptune has two alarms conditions: priority and cautionary. Priority alarms are
indicated by a red status light and an audible alarm. Clinician involvement is required to reset a
priority alarm. To reset a priority alarm, turn the unit’s power OFF, correct the alarm condition and
turn the unit’s power ON. Cautionary alarms are indicated by yellow status light and an audible
alarm. The cautionary audible alarm can be silenced by pressing the alarm mute button. The alarm
will deactivate if the alarm condition self-corrects.
A. HIGH TEMP:
A1. Patient temp >43°C (Patient Temp Indicator Flashes Red):
• Airway temperature improperly set for ambient and ventilatory conditions.
Reconfirm setings are appropriate for ventilation and ambient conditions.
A2. High temperature at column outlet (Column Temp Indicator Flashes Red):
• Low/no flow through column or reservoir, correct low/no flow condition.
• Low water level in column or reservoir, replace water reservoir and/or column.
D. SERVICE
• Unit failed self-diagnostic check. Turn power OFF, let the humidifier cool
and turn the power back ON. If the condition persists, remove the unit
from service.
A2 B A1 E
C F
CAUTIONARY ALARMS (Yellow Status Indicator Light and Audible Alarm):
Note: Press the alarm mute button to silence the Cautionary Audio Alarms for
approximately two minutes. When muted, the alarm remains activated, and the ALARM
MUTE button and Cautionary Alarm Icon are illuminated. The alarm will automatically
reset if the alarm condition “self-corrects.”
28
E. AIRWAY TEMP (Patient Temp Indicator Flashes Green):
• Airway temperature >41ºC, power is disconnected to humidifier and
wires, power returns when temp falls back below 40.5ºC and alarm will
deactivate, check temperature setting for ambient and/or ventilator
conditions
• Patient Temp > +/- 2ºC from set point after warm-up is complete (°C stops
flashing), check for proper usage for current ambient and ventilatory
conditions.
Should the humidifier exhibit any of the symptoms listed below, perform the indicated corrective
action before removing the unit from service.
• Internal malfunction, usually accompanied by a service alarm. Return the unit for
service.
Proximal patient airway temperature will not set below 32 °C or above 39 °C in the Invasive
mode or with Heated-Wire circuit in use.
• Settings outside the humidifier operating range, select a temperature within the
operating range (see Specifications)
Proximal patient airway temperature will not set below 30 °C or above 37 °C in the
Non-Invasive mode with a Non-Heated Wire circuit in use.
• Settings outside the humidifier operating range, select a temperature within the
operating range (see Specifications)
29
Temperature probe not properly connected to humidifier.
• Turn unit off for 60 seconds. Properly connect the temperature probe to the
humidifier and turn unit back on.
• Program desired settings for proximal airway, gradient and pause time. Press
accept button to return unit to warm-up and normal operation. If problem
persists after reprogramming, return to factory for service.
• Replace cord if necessary, press power ON/OFF button and the humidifier will
display user defined settings for mode, proximal patient airway temperature,
temperature gradient, and pause time. Press accept button to return to warm-up
and program memory with displayed settings.
Note: If the humidifier was in the pause mode prior to the power failure, restart the sys-
tem as noted above. Unit cannot be placed in pause until completion of the Warm-up
cycle.
Routine Checks
1. Observe the condensation levels in the breathing circuit with every ventilator check. Drain
as necessary.
Warning: Burn Hazard: The metal surfaces of the column and heater may be HOT (as
high as 140 °C). Allow the column and heater to cool before handling.
3. Check the water reservoir for adequate sterile water. Water level at or below the replace-
ment line may cause erratic temperature fluctuations.
30
• Open all clamps on the column and squeeze the reservoir to initiate flow into the
column.
Caution: Proper operation of the humidifier requires proper installation of the water
reservoir. Follow the reservoir installation instructions exactly.
4. Check the digital temperature display on the humidifier whenever making any adjustments to
the setup.
5. Check the LED Icons whenever an alarm condition occurs and note the digital temperature
display.
Cleaning
Use a 3% hydrogen peroxide solution or sodium hypochlorite to disinfect the outer surfaces. DO
NOT use alcohol or solvent on the unit.
Caution: Never autoclave, gas sterilize (ETO), irradiate, pasteurize or submerge the unit
in solution.
To clean the temperature probe, see cleaning instructions included with the probe.
Routine Maintenance
• Immediately remove from service any humidifier which shows signs of over
heating or electrical malfunction.
It may be necessary to replace the power fuse should it blow. To replace the power fuse:
• Remove the fuse tray from the “Power Entry Module” at the rear of the unit.
Check and replace blown fuses if detected, replace fuse tray.
• Power Fuses:
Two, 250 V ~, 4 amp, 5 x 20mm time delayed
The ConchaTherm Neptune has no user serviceable components inside and requires no annual cali-
brations. Component part lists and circuit diagrams are confidential and not commercially available.
Please contact the Teleflex Medical Customer Service Department at 1-866-246-6990 if you have
special needs or questions with this policy.
If your humidifier requires service and repairs, contact the Teleflex Medical Customer Service
Department at 1-866-246-6990.
31
Repair of leased heaters:
A Teleflex Medical customer service representative will assist you in obtaining a replacement heater
for your ConchaTherm Neptune. User-performed repairs of leased heaters should not be attempted.
Such repairs shall be considered user caused damage and may result in the cancellation of the
lease.
32
10. Limited Warranty Statement
WHAT THE WARRANTY COVERS AND FOR HOW LONG
Teleflex Medical ("Teleflex") warrants the Cat. No. 425-00 CONCHATHERM NEPTUNE ("Product")
against defects in material and workmanship under normal use and service for a period of one (1)
year from the date of purchase.
Teleflex Medical, at its option, will at no charge either repair the Product (with new or reconditioned
parts) or replace it (with new or reconditioned Product), during the warranty period provided that it is
returned in accordance with the terms of this warranty. Replaced parts or replaced Product is war-
ranted for the balance of the original warranty period. All replaced parts or Product shall become the
property of Teleflex. This expressed limited warranty is extended by Teleflex to the original end user
purchaser only and is not assignable or transferable to any other party. This is the complete warranty
for the Product manufactured by Teleflex. Teleflex assumes no obligations or liability for additions or
modifications to this warranty unless made in writing and signed by an officer of Teleflex. Teleflex
does not warrant the installation, maintenance or service of the Product. Teleflex cannot be respon-
sible in any way for any ancillary equipment not furnished by Teleflex, which is attached to or used in
connection with the Product. Because each system, which may use the Product, is unique, Teleflex
disclaims liability for range, coverage, or operation of the system as a whole under this warranty.
GENERAL PROVISIONS
THIS WARRANTY IS THE EXCLUSIVE REMEDY AND SETS FORTH THE FULL EXTENT OF
TELEFLEX MEDICAL’S RESPONSIBILITY. THIS WARRANTY IS IN LIEU OF ALL OTHER
EXPRESS WARRANTIES. IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIM-
ITED TO THE DURATION OF THIS LIMITED WARRANTY.
This warranty gives you specific legal rights; you may also have other rights, which vary, from state
to state.
You must provide proof of purchase (bearing the date of purchase and Product item serial number)
in order to receive warranty service and, also, deliver or send the Product transportation and insur-
ance prepaid, to an authorized service location. Warranty services will be provided by Teleflex
Medical or through an authorized Teleflex Medical service center. If you first contact the company
that sold you the Product (e.g. Teleflex Medical Distributor), they can facilitate your obtaining warran-
ty service. You can also call Teleflex Medical at 1-866-246-6990 for warranty service information.
33
WHAT THIS WARRANTY DOES NOT COVER
a. Defects or damage resulting from use of the Product in other than its normal and
customary manner.
g. Scratches or other cosmetic damage to Product surfaces that does not affect the operation
of the Product.
88924-07 06/07
© 2007, Teleflex Incorporated
Hudson RCI, Concha, Conchatherm, and Concha-Column are registered trademarks of Teleflex Medical Incorporated.
34
Appendix I
Electromagnetic Emissions Tables
Recommended separation distances between portable and mobile RF communications equipment and the
CONCHATHERM NEPTUNE Humidifier.
The CONCHATHERM NEPTUNE Humidifier is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the CONCHATHERM NEPTUNE
Humidifier can help prevent electromagnetic interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and the CONCHATHERM NEPTUNE Humidifier as
recommended below, according to the maximum output power of the communications equipment.
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
35
Guidance and manufacturer's declaration - electromagnetic immunity
The CONCHATHERM NEPTUNE Humidifier is intended for use in the electromagnetic environment specified
below. The customer or the user of the CONCHATHERM NEPTUNE Humidifier should assure that it is used
in such an environment.
Electrical fast ±2 kV for power ±2 kV for power Mains power quality should be that of a
transient/burst supply lines supply lines typical commercial or hospital environ-
ment.
IEC 61000-4-4 ±1 kV for ±1 kV for
input/output lines input/output lines
<5% UT <5% UT
(> 95% dip in UT) (> 95% dip in UT)
for 5 sec for 5 sec
IEC 61000-4-6
NOTE UT is the a.c. mains voltage prior to application of the test level.
36
Guidance and manufacturer's declaration - electromagnetic immunity
The CONCHATHERM NEPTUNE Humidifier is intended for use in the electromagnetic environment specified below. The
customer or the user of the CONCHATHERM NEPTUNE Humidifier should assure that it is used in such an environment
d = [3.5/V1]√P
NOTE 1: At 80 MHz and 800 MHZ, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electro-
magnetic site survey should be considered. If the measured field strength in the location in which the CONCHATHERM NEPTUNE Humidifier is used exceeds the
applicable RF compliance level above, the CONCHATHERM NEPTUNE Humidifier should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the CONCHATHERM NEPTUNE Humidifier.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
37
Guidance and manufacturer's declaration - electromagnetic emissions
The CONCHATHERM NEPTUNE Humidifier is intended for use in the electromagnetic environment specified
below. The customer or the user of the CONCHATHERM NEPTUNE Humidifier should assure that it is used in
such an environment.
CISPR 11
The CONCHATHERM NEPTUNE Humidifier is suitable for use in
Harmonic emissions Class A
all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply net-
IEC 61000-3-2 work that supplies buildings used for domestic purposes.
Voltage fluctuations/ Complies
flicker emissions
IEC 61000-3-3
38