Sl. Quantity Measured / Instrument Range/Frequency Calibration Measurement Capability ( ) Remarks
Sl. Quantity Measured / Instrument Range/Frequency Calibration Measurement Capability ( ) Remarks
Sl. Quantity Measured / Instrument Range/Frequency Calibration Measurement Capability ( ) Remarks
ELECTRO-TECHNICAL CALIBRATION
I. SOURCE
1. Temperature
Simulation$
(Temperature Indicator /
Controller / Recorder/
Scanner)
RTD Type (-) 200°C to 800°C 0.56°C Using Process Calibrator
‘J’ Type Thermocouple (-) 200°C to 1200°C 0.8°C By Direct Method
‘K’ Type Thermocouple (-) 200°C to 1350°C 0.81°C
100°C to 1750°C 1.14°C
‘R’ Type Thermocouple 100°C to 200°C 1.1°C
200°C to 1750°C 0.82°C
‘S’ Type Thermocouple 600°C to 1820°C 0.94°C
B’ Type Thermocouple (-)200°C to 400°C 0.59°C
T’ Type Thermocouple (-) 200°C to 0°C 0.43°C
0°C to 1300°C 0.37°C
‘N’ Type Thermocouple (-) 200°C to 1000°C 0.29°C
‘E’ Type Thermocouple
2. Time Interval (Measure) 1 sec to 1 min 0.08 to 0.13 sec Using Digital Timer
1 min to 24 hr 0.13 to 22 sec by Comparison Method
3. Temperature
Simulation♣
(Temperature Indicator /
Controller / Recorder/
Scanner )
RTD’ Type -200°C to 800°C 0.82°C Using Multifunction
‘J’ Type Thermocouple -200°C to 1200°C 1.45°C Calibrator by Direct
‘K’ Type Thermocouple -200°C to 1350°C 0.96°C Method
‘R’ Type Thermocouple 0°C to 1750° 2.81°C
‘S’ Type Thermocouple 0°C to 1750°C 2.32°C
4. Time Interval (Measure) 1 sec to 1 min 0.2 to 0.23 sec Using Digital Timer
1 min to 24 hr 0.23 to 22 sec by Comparison Method
MECHANICAL CALIBRATION
1. Calipers$
(Vernier/Dial/Digital)
L.C:0.01mm 0 to 450mm 18.2 µm Using Slip gauge Set
450 mm to 600mm 18.5 µm Grade ‘0’, Caliper
600 mm to 1000mm 19.5 µm Checker, Length bar &
1000 mm to 1500mm 21.3 µm Slip gauge Accessories.
Comparison Method
L.C:0.02mm 0 to 2000mm 20.2 µm IS :3651(Part 1 & 2)
2. External Micrometer/
Ball Micrometer/
Blade Micrometer/
Disc
Micrometer/Flange
Micrometer/Groove
Micrometer/
Spline Micrometer/
Sheet Metal
Micrometer/
Point
Micrometer/Pitch
Micrometer$
L.C.:0.001mm 0 to 25mm 2.0 µm Using Slip Gauge Set
25 mm to 150mm 5.8 µm Grade ‘0’, Optical Flat &
150 mm to 300mm 6.2 µm Parallels, length bar.
300 mm to 600mm 8.5 µm Comparison Method
600 mm to 1000mm 9.2 µm IS:2967
3. Depth Micrometer$
L.C:0.01mm 0 to 150mm 5.82 µm Using Slip Gauge Set
150 mm to 300 mm 8.2 µm Grade ‘0’, Caliper
Checker & Gauge Block
Accessories by
Comparison Method
BIS:6468
4. Internal Micrometer/
Stick Micrometer$
L.C:0.01mm 0 to 600mm 6.7 µm Using Slip Gauge Set
600 mm to 1000mm 8.5 µm Grade ‘0’, Gauge Block
1000 mm to 2000mm 20.0 µm Accessories Set & Length
bar by Comparison
Method as per IS:2966
14 Pistol Caliper$
L.C:0.1mm 0 to 100mm 58.0 µm Using Slip gauge Set
Grade ‘0’ by
Comparison method
IS:2092
25. Flush Pin Gauge$ Up to 200 mm 2.8 µm Using Slip Gauge Set
Grade ‘0’ , Digital
Micrometer and Plunger
Dial gauge by
Comparison Method
28. Setting Rod / Setting Up to 50mm 3.5 µm Using Slip Gauge Set
Piece$ 50 mm to 100mm 3.5 µm Grade ‘0’, Length bar &
100 mm to 200mm 3.5 µm Electronic comparator by
200 mm to 1000mm 4.7 µm Comparison Method
31. Ultrasonic
Thickness gauge$
L.C.: 0.1mm 0 to 100mm 70.2 µm Using slip gauges set
grade“0” by
Comparison Method
34. Tri Square$ Up to 500 x 500mm 4.4 µm Using Slip Gauge Set
(Squareness) Up to 500 x 500mm 4.0 µm Grade ‘ 0’, Master
(Parallelism) Square, Plunger dial,
Surface plate by
Comparison Method as
per IS :2103
41. Microscope♣
Linear
L.C: 0.0001mm 0 to 300 mm 1.1 m Using Graticules scale by
Angle 0 to 360° 3’ Comparison Method
42. Bench Centre Centre height:300 mm 6.0 m Using Mandrel & Dial
Parallelity Admit between Indicator as per
Co- axiality of Centre:750mm IS:5980:2000
centre♣
1. Tachometer / RPM$
Contact Type 150 rpm to 2500 rpm 0.82% Using RPM Indicator with
Non contact type 150 rpm to 50,000 rpm 0.87% Electric Motor by
Comparison Method
2. Indicator of
Centrifuge / RPM♣
Meter 60 to 25,000 rpm 0.06% of rdg Using Digital Tachometer
Comparison Method
III. ACCOUSTICS
IV. WEIGHTS
1. Weighing Balance
Calibration of
Electronic Weighing
Balance$
Class II and coarser
Readability : 0.01mg Up to 200 g 0.04 mg Using E1 Class standard
weights, as per OIML R
76-1
2. Calibration of
Electronic Weighing
Balance♣
Class I and coarser
Readability : 0.001 Up to 6.1 g 0.01 mg Using E1 Class Standard
mg Weights as per OIML R
76-1
Readability : 20 g Up to 200 kg 41.0 g
VI. VOLUME
1. Shore Hardness
Tester$
Shore A 0-100 Shore A 0.6 shore A Using Load cell with
Shore D 0-100 Shore D 0.6 shore D Indicator based on ASTM
D 2240-05
3. Pressure (-)350 mbar to 350 mbar 0.24 % rdg Using Digital low pressure
Magnehelic Calibrator by Comparison
Gauges / method as per DKD-R6-1
Differential & DKD-R6-2
Pressure Gauge#
5. Vacuum Gauge / (-) 0.1bar to (-) 0.85 bar 0.95 % rdg Using Digital Pressure
Vacuum Indicator/ Calibrator by Comparison
Switch / Method as per DKD-R6-1
Compound Gauge / & ISO: 3567
Transmitters and
Transducer with
Indicator♣
1. Calipers:
(Venire / Dial /
Digital)
L.C:0.01mm 0 to 1000 mm 15.0 µm Using Slip gauge Set
grade ‘0’, Caliper
Checker, Length bar &
Slip gauge Accessories
IS:3651(Part I & II)
2. External
Micrometer/ Ball
Micrometer/ Blade
Micrometer /Disc
Micrometer /Flange
Micrometer/
Groove
Micrometer/Spline
Micrometer /Sheet
Metal
Micrometer/Point
Micrometer/Pitch
Micrometer
L.C:0.001mm 0 to 100mm 1.0 µm Using Slip Gauge Set
100 mm to 600mm 4.5 µm Grade ‘0’, Optical Flat &
Parallels, Length Bar,
Gauge Block Accessories
IS:2967
3. Depth Micrometer
L.C:0.01mm 0 to 300 mm 11.15 µm Using Slip Gauge Set
Grade ‘0’, Caliper
Checker, Gauge Block
Accessories
BIS:6468
4. Internal
Micrometer/ Stick
Micrometer
L.C:0.01 mm 0 to 600 mm 7.1 µm Using Slip Gauge Set
Grade ‘0’, Length Bar &
Gauge Block Accessories
IS:2966
15. Flush Pin Gauge Up to 200 mm 2.89 m Using Slip Gauge Set
Grade ‘0’ , Digital
micrometer & Plunger dial
by Comparison Method
18. Measuring Pin 1mm to 20mm 2.83 m Using Slip Gauge Set
Grade ‘ 0’ Electronic
Comparator Stand
IS 11103
THERMAL CALIBRATION
1. TEMPERATURE
1. Humidity Indicator 15% to 95% @ 25°C 1.4%RH Using Single Point Probe
of Humidity with Digital Indicator
Chamber / By Comparison Method,
Environment as per DKD – R-5-7
Chamber♣
(Single point)
2. Humidity Chamber / 15% to 95% @ 30°C 1.75%RH to 2.83%RH Using Humidity Sensors
Environment with Digital Indicator
Chamber♣ (Multi point)
(Multi point) By Comparison Method,
as per DKD- R-5-7