Gamma and Xray Detectors Rev1 200507 PDF
Gamma and Xray Detectors Rev1 200507 PDF
Gamma and Xray Detectors Rev1 200507 PDF
1026 Route 518 , Rocky Hill, NJ 08553 , Tel.: 609-924-7310 • Fax: 609-924-1729
http://www.pgt.com
PRINCETON GAMMA-TECH
Reproduction or publication of the contents in any manner without the written permission of Princeton Gamma-Tech,
Inc. is prohibited.
No liability with regard to improper operation of equipment or to the use of the information in this manual is assumed
by Princeton Gamma-Tech, Inc.
PGT Detector Systems User Manual
Table of Contents
Section 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.1.1 Available Detector Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.1.2 Choosing the Right Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1.1.3 Special Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Section 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.1 Cryostats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.2 Liquid Nitrogen Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2.3 Instructions for Filling Dewars . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2.3.1 Dipstick Cryostat Models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2.3.2 Elbow Dipstick Models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
2.3.3 Unitary Cryostats (Bucket Dewars) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
2.3.4 MPS Cryostats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
2.3.5 Gravity-Fed Portable Detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
2.4 Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Section 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1 Germanium Detectors with RG-11, RG-15 or 352 Preamplifiers . . . . . . . . . . . . 3-3
3.1.1 Needed Equipment and Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.1.2 Detector Cool-down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.1.3 Cable Connections to the Preamplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3.2 Germanium Detectors with PO-12, PO-14 or TR-14 Preamplifiers . . . . . . . . . . 3-9
3.2.1 Needed Equipment and Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3.2.2 Detector Cool-down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3.2.3 Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3.3 SpecificationMeasurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.4 Manual Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
i
User Manual PGT Detector Systems
Section 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.1 Detector Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.1.1 Photon Interactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.1.1.1 Photoelectric Effect. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.1.1.2 Compton Scattering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.1.1.3 Pair Production . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.1.1.4 Linear Attenuation Coefficients. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
4.2 Spectral Performance Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
4.2.1 Efficiency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4.2.2 Peak Shape and Energy Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
4.2.3 Noise Contributions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4.2.4 Peak-to-Compton Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4.2.5 Peak-to-Background Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4.2.6 Timing Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4.3 Other Detector Geometries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
4.3.1 Well Detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
4.3.2 Marinelli Beakers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4.4 Other Factors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4.4.1 Temperature Cyclability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4.4.2 Window Thickness Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
4.4.3 Transmission/Absorption Nomogram . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Section 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.1 Preamplifiers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.1.1 Preamplifier Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.2 Preamplifier Layouts and Circuit Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
5.2.1 Resistive Feedback Preamplifiers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
5.2.2 Replacement of Resistive Feedback Preamplifiers . . . . . . . . . . . . . . . . . 5-20
5.2.3 Replacement of In-Line Preamplifiers . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
5.2.4 Pulsed Optical Preamplifiers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
5.2.5 Transistor Reset Preamplifiers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28
5.2.6 Replacement of PO and TR Preamplifiers . . . . . . . . . . . . . . . . . . . . . . . 5-36
ii
PGT Detector Systems User Manual
Section 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
6.1.1 Liquid Nitrogen Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
6.1.2 Neutron Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
6.2 Fault Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6.2.1 Resistive Feedback Preamplifier Detectors . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6.2.2 Pulsed Optical or Transistor Reset Detectors . . . . . . . . . . . . . . . . . . . . . . 6-9
Section 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.1 HPGe Detector Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
7.2 Si(Li) Detector Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
7.3 Detector Repair Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
7.4 Packing Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
List of Tables
iii
User Manual PGT Detector Systems
IMPORTANT
Please notify your shipping/receiving personnel that all packaging must be thoroughly inspected before
signing any document presented by the delivery carrier. If damage is detected, notify the carrier and
PGT immediately. It is the customer’s responsibility to submit a claim; PGT will assist you in whatever
way it can. Take Polaroid pictures of damages to substantiate any claim that might be made by you or
on your behalf. Remember: examine carefully before you sign. This prudent procedure can save you
both time and bother.
Be sure to save the plastic protective endcap and packing box for use if the detector needs to be re-
turned to PGT Service at some future date. The boxes are custom fitted for the detector and are
expensive to replace.
CAUTION!!
THESE SYSTEMS INVOLVE THE USE OF HIGH VOLTAGES.
BE SURE YOUR SYSTEM IS PROPERLY GROUNDED AND
AVOID CONTACT WITH THE HIGH VOLTAGE CIRCUITS.
iv
Section 1
1
Introduction
Detector Options
Table 1-1: PGT Gamma and X-ray Detectors
Detector Energy Resolution Peak
Series Crystal Geometry Sizes
Crystal Range Shape
1-2
PGT Detector Systems Detector Options
1-3
Detector Options PGT Detector Systems
1.1.2.2. Efficiency
The size of the detector should be optimal. A detector that is too big not
only wastes money, but may also create problems due to high count
rates. An undersized detector may be cheaper in price, but expensive
in time.
1-4
PGT Detector Systems Detector Options
PGT QUIET ONES are available with P- and N-type germanium coaxial
detectors which use the RG-11B Preamplifier in any of the standard
PGT cryostat configurations. In this application, the preamplifier layout
is similar to that for in-line preamplifiers (see Fig. 5-6).They are of par-
ticular use for low-activity measurements, where well-defined peaks
are needed, for high resolution analysis of complex spectra with multi-
ple isotopes, for environmental analysis with portable or truck-installed
detectors, and for nuclear plants and other industrial sites where vibra-
tions from heavy machinery are a problem.
1-5
Detector Options PGT Detector Systems
1-6
Section 2
Cryostat Configurations 2
Safety Precautions
Liquid Nitrogen Fill Procedures
PGT Detector Systems Cryostats/Safety Precautions
2.1: Cryostats
The basic cryostat configurations are shown on the following pages,
Tables 2-1 through 2-3, and the preamplifiers are described in detail in
Section 5. PGT will manufacture a system for your particular needs.
The Data Sheet included with your detector system specifies the com-
ponents and their performance characteristics. Please keep this with
your records; it is needed during set-up of your system to verify opera-
tion and in case of problems in the future.
After you have identified your system components, go to page 2-7 for
safety precautions for handling liquid nitrogen and directions for filling
the dewar and cooling the system. Then proceed to the set-up instruc-
tions in Section 3 for your particular preamplifier.
2-3
Cryostats/Safety Precautions PGT Detector Systems
fill tube
overflow LN2 fill tube
Insulation &
molecular sieve
32 liter capacity
end
cap Coaxial Germanium Endcap
in-line end diameter: 74 mm (2-15/16”)
in-line preamp cap
pre- length: 114 mm (4-1/2”)
amp larger endcaps available for large crystals
fill tube
Standard Dewar
2-4
PGT Detector Systems Cryostats/Safety Precautions
endcap
preamp
end
cap
endcap
diameter: 67 mm (2-5/8”)
detector size: 1000 mm2
406 mm (16”) std. diameter: 81 mm (3-3/16”)
or as specified detector size:1500-2000 mm2
endcap
preamp Coaxial Ge Endcap Dewar volumes
diameter: 76 mm (3”) 15 liter
endcap length: 203 mm (8”) 30 liter
preamp
92 mm (3-5/8”) (Cryoflex only)
2-5
Cryostats/Safety Precautions PGT Detector Systems
Dewar volumes
1.5 liters
3 liters
5 liters
7.5 liters
with 50-liter LN2 pressured fill dewar
Dewar sizes:
variable, per customer specifications
2-6
PGT Detector Systems Cryostats/Safety Precautions
2.2: Liquid
Nitrogen
Precautions
Liquid nitrogen (LN2) is safely used every day in laboratories and fac-
tories around the world. However, proper procedures must be followed
to prevent injury to person or damage to the detector system. Knowl-
edge of the hazards involved is a prerequisite to safety.
The use of LN2 poses three potential hazards: build-up of high pres-
sures, damage to materials due to the extremely low temperatures, and
personal injury. The compression ratio of nitrogen is 1:692. Therefore,
the expansion from liquid to gas can result in the build up of high pres-
sures as the LN2 warms up if the container does not have adequate
venting or pressure relief provisions. Some materials become brittle
and fracture when exposed to liquid-nitrogen temperatures (77°K). For
advice when selecting materials for use in storing and transferring LN2,
contact your liquid nitrogen supplier.
When ordering liquid nitrogen from your supplier, be sure to specify that
the dewar be pressurized only with dry nitrogen gas.
2-7
Cryostats/Safety Precautions PGT Detector Systems
2.3: Instructions
for Filling It is absolutely necessary to prevent the electronics and the outside of
Dewars the cryostat from becoming excessively cold. Avoid spilling LN2 on
these components and vent cold gas away from the system being filled.
Although the detector may be filled with LN 2 while it is still strapped in
its shipping pallet, all insulating packing materials, especially foam,
must be removed first to prevent damage to the system. Damage can
also occur any time a cold detector is packed in thermally insulating
materials. Always keep such materials away from the detector. Follow
these instructions to prevent damage to vacuum seals or to system
electronics due to excessive chilling.
The dipstick enters the dewar through a white Teflon fill collar which is
both insulator and shock mount. The Teflon fill collar contains a stain-
less steel tube which is used for filling and gas venting. A Teflon tube
extends down into the dewar neck. In filling, the liquid nitrogen enters
through the extended steel tube and the opposite channel serves as an
exhaust vent. Fig. 2-1 shows a dipstick cryostat connected to a PGT au-
tomatic fill system.
Prepare for filling the dewar by connecting the supply hose to the fill
tube and an exhaust hose (optional) to the vent tube in the Teflon collar.
The exhaust hose should be approximately six (6) feet of plastic tubing
or flexible stainless steel which carries the cold gas and liquid overflow
away from the cryostat and electronics. The connection to the liquid ni-
trogen supply hose can be made by a short length of plastic tube, but a
flexible metal hose with a Swagelok fitting is preferable, since the cold
will eventually crack the plastic. The liquid nitrogen can be supplied un-
der pressure from a supply dewar that was correctly pressurized by the
LN2 vendor or from a supply dewar correctly pressurized by the user
from a dry nitrogen pressurization outlet. Recommended pressure is
10-15 psi. A metal funnel can also be used if the dewar can be gravity
fed.
2-8
PGT Detector Systems Cryostats/Safety Precautions
fill collar
LN2 fill tube
overflow
Swagelok fitting
Overhead view
Standard Dewar
LN2 Source
Vent valve
Pressure build-up Halo ring
regulator
Pressure build-
up valve
Pressure gauge
Solenoid
power line Evacuation valve &
bursting disk
Fig. 2-1 Filling a dipstick dewar or large bucket dewar with a fill collar, using the PGT Auto-Fill supply
dewar (above). The optional Kadel 92008 withdrawal device is shown below.
Hose barb
Pipe plug
Vent valve
Withdrawal valve
Pipe
plug
2-9
Cryostats/Safety Precautions PGT Detector Systems
Step 1. To fill/refill the dewar, open the vent valve to allow any
pressurized gas to escape. After venting and when the
pressure gauge reads zero (0), loosen the quick clamp
and remove the upper flange of the withdrawal device by
pulling it straight up out of the dewar. The fill hose from the
bulk tank may then be inserted through the opening in the
lower flange, and transfer may begin.
Step 2. When the dewar is full, remove the transfer hose and re-
place the upper flange and quick clamp. Close both the
withdrawal valve and the vent valve if they are not already
closed. Evaporating gases within the dewar will self-pres-
surize the vessel to a satisfactory working pressure deter-
mined by a relief valve (typically 10 psi). An external inert
gas source may be used to pressurize the vessel through
the vent valve, but caution should be exercised to main-
tain the internal pressure below that specified above.
Remove the dewar fill cap, insert a large metal funnel, and carefully
pour in the liquid nitrogen from the top. Take care that LN2 is not spilled
on the cryostat. If the cold liquid contacts a flange which contains a vac-
uum seal, it is possible that the seal may be breached. Use of the large
funnel will help avoid such damage.
2-10
PGT Detector Systems Cryostats/Safety Precautions
For manual filling, you need a special fill funnel, a small dewar or ther-
mos, and a rubber stopper. Follow this procedure:
Step 1. Screw the fill funnel onto the front fill hole.
Step 3. Place the rubber stopper (or other suitable stopper) over
the funnel so that the boiling of the liquid nitrogen pressur-
izes the funnel and forces the LN2 into the dewar.
Step 6. When the dewar is full, LN2 will start to shoot out of the
vent hole. At this point, stop filling. After waiting the rec-
ommended time for cooling the detector, refill the dewar to
top it off. Unscrew the fill funnel and operate the detector.
The steps in filling an MPS cryostat with a PGT liquid nitrogen filling
system or its equivalent are as follows:
Step 2. Screw the Swagelok fitting onto the front fill hole.* Use Te-
flon tape for a better seal.
Step 3. Connect the fill hose to the Swagelok fitting. Tighten light-
ly with a wrench.
Step 4. Fill the MPS cryostat until liquid nitrogen comes out of the
vent hole. NOTE: Keep the pressure under 20 psi.
Step 5. Top off the MPS cryostat after a few hours of cooling. Re-
fill at smaller intervals than the specified holding time (16
hours for most cryostats).
2-11
Cryostats/Safety Precautions PGT Detector Systems
Valve handle
Valve stem
LN2
Bayonet
Valve Knurled knob
preamp
endcap
Step 1. Unscrew the knurled knob on the back end of the detector
dewar and remove the plastic rod.
Step 3. Open the valve handle at the top of the mother dewar. To
verify that the pressure relief valve is operating and that
LN2 is flowing freely, open the brass plunger just below
the knurled knob for approximately 1 second. You will also
see water vapor at the top of the mother dewar where ni-
trogen gas is escaping.
2-12
PGT Detector Systems Cryostats/Safety Precautions
Step 4. When the dewar is filled and the detector cooled, it can be
removed from the mother dewar. Close the valve handle
at the top of the mother dewar. Be sure to support the de-
tector while unscrewing the knob. Carefully lower the
detector. LN2 may splash from the full dewar! Replace
the plastic tube and tighten down the knob securely.
The reason for this lies in how warm-up occurs: The molecular sieve
material in the dewar and cold finger begins to outgas at the early stag-
es of warm-up. These impurities will condense on the crystal, which is
still cold. If the system is cooled down at that stage, these impurities on
the crystal will cause system performance to degrade significantly. By
allowing the system to warm up completely, the impurities will eventu-
ally volatilize off the crystal, and the system can then be recooled.
2-13
Cryostats/Safety Precautions PGT Detector Systems
2.4: Safety
Precautions
Your detector system uses high voltage (up to 5000V) to bias the de-
tector element. The standard PGT bias supplies made for use with pho-
ton detectors are capable of delivering only very low current. Your de-
tector system, however, has a high voltage filter with capacitors capa-
High Voltage ble of delivering dangerously high current for a brief time while being
discharged (even if the bias supply has been disconnected). As a gen-
eral rule, you should never open the electronics shield or module boxes
except when following the explicit instructions of an authorized PGT
representative.
If the cryostat develops a leak while cold and then warms up after the
molecular sieve has absorbed a large quantity of gas, significant inter-
nal pressure build-up is possible. A pressure-relieving device is incor-
Internal Pressure porated to prevent this. Nevertheless, if the cold cryostat shows evi-
dence of poor vacuum (e.g., an unusually cold cryostat or endcap,
Build-up moisture condensation or “sweating” or, in the extreme case, an out-
ward bulge to the end window on warm-up), place the original plastic
cover over the beryllium window. Cover any hole in the window with
tape and keep the detector cold. Disconnect the system from external
electronics. Do not warm up such a cryostat or take additional ac-
tion except on instruction from PGT.
Standard PGT well detector systems have an aluminum well in the end-
cap. Surrounded by the detector element, this well allows high efficien-
cy counting of a source which is placed inside it. To reduce attenuation
Well Detectors at lower energies, the aluminum well is relatively thin. Penetration of
this well would cause loss of vacuum, resulting in ice condensation on
the sensitive detector surfaces and electronic components. Therefore,
be careful not to insert sharp objects or force oversized samples into
the well. If reasonable care is exercised when placing samples in the
well, no problems should arise.
2-14
Section 3
3
System Set-up
Calibration
Voltmeter Oscilloscope
BIAS
Preamp TEST OUTPUT
(expanded)
VDC
1 preamp output
POWER 2 main amp output
1 2
TP Clipper
preamp preamp (see Fig. 3-2 on p. 3-6)
power output
ADC
Multichannel Analyzer
... . .
ADC
PGT Model 326 Pulser
LN2 Level PGT Model 312, PGT Model
Controller 314 or 315 347 Amplifier
(optional) High Voltage
Power Supply
TP
RG15 or OUTPUT OUTPUT
352 in-line
Preamps
optional LED
BIAS TEST
POWER
Fig. 3-1 Test connections to the RG-11 or 352 preamplifier. The in-line RG-15 and 352 front
panel is illustrated below.
3-2
PGT Detector Systems Set Up/Calibration
3.1: Germanium
Detectors with
RG-11, RG-15
or 352
3.1.1. Needed Equipment and Precautions Preamplifiers
On receipt of the detector, examine the detector and the dewar care-
fully to determine if damage has occurred during shipment. If any dam-
age is observed, call the factory at once at (609) 924-7310, ext. 229.
Do not attempt to operate the unit.
IMPORTANT!!
Use a high voltage supply with maximum current of 1 mAmp be-
tween the output poles and between each pole and ground.
The cooling time, usually 8 hours, is printed on the label supplied with
the detector. This length of time must elapse between the initial contact
with LN2 and any application of the high voltage (HV). Overnight cool-
ing is recommended if time permits.
3-3
Set Up/Calibration PGT Detector Systems
Make certain, also, that the multiwire Amphenol connector for the
preamplifier power at the back of the main amplifier delivers the correct
voltages (see operator manual for the amplifier).
pin 1: ground pin 4: +12 V
pin 9: -12 V
pin 7: +24 V
pin 6: -24 V
3.1.3.1. Procedure
Remove all radioactive sources from near the detector. Then follow
these steps:
3-4
PGT Detector Systems Set Up/Calibration
Step 5. Again verify that the HV has been turned down completely
to 0 V and is at the correct polarity: (+) for IGC/IGW and
(-) for IGP/NIGC/NIGP/LS. Check the data sheet supplied
with the detector or the identification label on the cryostat.
WARNING!!
APPLYING SUDDEN REVERSE-POLARITY HIGH VOLTAGES TO
A DETECTOR CAN DESTROY THE FET INSIDE THE CRYOSTAT
AND POSSIBLY DAMAGE THE PREAMPLIFIER.
Step 6. Turn the HV control to 100 V. When the TP voltage has re-
turned to a steady value, the peak-to-peak amplitude of
the noise on the oscilloscope should be considerably less
than before. Make a note of the TP voltage and the noise.
Step 8. The set of test points and noise values should be similar
to that shown in the following table, obtained for an IGC
whose recommended operating voltage is +3000V
3-5
Set Up/Calibration PGT Detector Systems
to amplifier to scope
10 kΩ
IN914
white
noise
(A) (C)
(B)
undercompensated well-set P/Z: overcompensated
undershoot, sometimes long tail:
with overshoot: PGT 346/347 may have
small undershoot P/Z too far clockwise
P/Z too far counterclockwise
PGT 340 may have
small overshoot
Fig. 3-2 Main amplifier Pole Zero adjustment and related pulse shapes.
Step 9. Optimize the Pole Zero cancellation (P/Z) of the main am-
plifier by placing a ~10 µCi source, such as 60Co, some
10-30 cm in front of the detector. Observe the shape of
the resulting pulses on the oscilloscope after adjusting the
trigger status. Adjust the main amplifier Pole Zero poten-
tiometer so that the pulses return to the baseline in as
short a time as possible (see Fig. 3-2). The three pulse
shapes are interpreted as indicated.
3-6
PGT Detector Systems Set Up/Calibration
(D) (D)
(E) (E)
(F) (F)
N-type Detectors
P-type Detectors
(D) correct
(E) undercompensated
P/Z too far counter clockwise
(F) overcompensated
P/Z too far clockwise
Fig. 3-3 Pulse shapes for the preamplifier P/Z adjustment.
Step 10. The Pole Zero of the preamplifier is normally preset be-
fore delivery. As a check, or when the preamplifier has
been changed for some reason, the P/Z can be adjusted
as follows:
Step 11. You are now ready to measure the detector specifications
and determine resolution. Go to Section 3.3: on page 14.
3-7
Set Up/Calibration PGT Detector Systems
Oscilloscope
BIAS OUTPUT
PO-12 TEST
PO-14 INHIBIT
TR-14 # 1 preamp output
POWER
2 main amp output
Preamps 1 2
(expanded)
preamp preamp
power output
ADC
Multichannel Analyzer
... . .
ADC
Pulser
PGT Model 326
LN2 Level PGT Model 312, PGT Model
314 or 315 347 Amplifier
Controller
(optional) High Voltage
Power Supply
# NOTE: The BNC connection
from the preamplifier inhibit is
either to the “preamp inhibit” at the back of the amp
or to the anticoincidence connection
at the front of the ADC unit.
PO and TR
in-line models
Output inhibit on PO and TR amps
optional LED
HV Test
Fig. 3-4 Connections for pulsed optical and transistor reset preamplifiers; in-line preamp
front panel is illustrated below.
3-8
PGT Detector Systems Set Up/Calibration
3.2: Germanium
Detectors with
PO-12, PO-14
or TR-14
3.2.1. Needed Equipment and Precautions Preamplifiers
On receipt of the detector, examine the detector and the dewar care-
fully to determine if damage has occurred during shipment. If any dam-
age is observed, call the factory at once at (609) 924-7310. Do not at-
tempt to operate the unit.
IMPORTANT!!
Use a high voltage supply with maximum current of 1 mAmp be-
tween the output poles and between each pole and ground.
The cooling time, usually 8 hours, is printed on the label supplied with
the detector. This length of time must elapse between the initial contact
with LN2 and any application of the high voltage (HV). Overnight cool-
ing is recommended if time permits.
3-9
Set Up/Calibration PGT Detector Systems
Make certain, also, that the multiwire Amphenol connector for the
preamplifier power at the back of the main amplifier delivers the correct
voltages (see operator manual for the amplifier).
pin 1: ground pin 4: +12 V
pin 9: -12 V
pin 6: -24 V
pin 7: +24 V
3.2.3.1. Procedure
Remove all radioactive sources from near the detector. Then follow
these steps:
3-10
PGT Detector Systems Set Up/Calibration
For IGC
initial within 1 minute
3-11
Set Up/Calibration PGT Detector Systems
2 0.1 0.2 0.3 millisec 2 0.1 0.2 0.3 millisec 0.1 0.2 0.3 millisec
2
4 4 4
6 6 6
8 8 8
10 10 10
12 12 12
3-12
PGT Detector Systems Set Up/Calibration
Step 11. The spectrometer is now ready for a resolution test. Con-
nect the inhibit output of the preamplifier to the anticoinci- CAREFUL!!
dence input of the MCA or to the “preamp reset” input of
the amplifier. Remove the endcap plastic cover and place THE BERYLLIUM OR VESPEL
an appropriate radioactive source in front of the detector. WINDOW IN FRONT OF THE
DETECTOR IS EXTREMELY
THIN AND SHOULD NEVER
BE TOUCHED!
Note: When operating near or above 100 keV, it is recom-
mended that the gain of the preamplifier be decreased so
as to reduce by half the number of reset pulses. It might,
however, cause a slight deterioration in energy resolution
at low gain energies, depending on the main amplifier
used. Gain can be lowered for the PO-14B and TR-14B
systems by “jumpering” R 18 (voltage divider). See the ap-
propriate preamplifier layout in Section 5.
3-13
Set Up/Calibration PGT Detector Systems
3.3: Specification
For planar detectors (IGP), the specifications are usually the resolu-
Measurements tions at 5.9 keV (55 Fe) or 6.4 keV (57Co) and 122 keV (57Co). For co-
axial detectors (IGC, NIGC), the specifications normally include the
resolutions at 122 keV (57Co) and 1332 keV (60Co).
Table 3-2 lists the standard radioactive sources for testing the perfor-
mance of germanium detectors:
3-14
PGT Detector Systems Set Up/Calibration
3.4: Manual
Calibration
In calibrating the detector manually, it is necessary to estimate the Full
Width at Half Maximum (FWHM). Spectrum collection should continue
for 5—10 minutes until at least 3000 counts have accumulated in the
channel corresponding to the center of the peak in order to avoid errors
associated with the counting statistics. Refer to your analyzer manual
for autocalibration of your system. The following is a schematic descrip-
tion for manually calibrating the detector.
∆E
P = peak count E1
P FWHM
= half maximum
2
count
E2
Channels
C1 ∆C C2
∆E E2 - E1
Calibration: = eV/ch
∆C C2 - C1
3-15
Set Up/Calibration PGT Detector Systems
P = peak count
δ1= A-H
A C A-B
FWHM H = P/2
δ2= C - H
C-D
B
D
δ1 δ2
(X1 -1) X1 X2 (X2+1) Channels
As seen in Fig. 3-8, the FWHM is the number of channels from channel
X1 to channel X2 plus two partial channels δ1 and δ2 as determined by
proportionality.
Multiply the FWHM by the energy calibration (in eV/ch or keV/ch) to de-
termine the resolution of the detector for the appropriate energy. Com-
pare this value with that shown on the Data sheet supplied with the de-
tector.
2000 – 1800
Then, δ 1 = ------------------------------ = 0.2
2000 – 1000
2300 – 1800
and δ 2 = ------------------------------ = 0.33
2300 – 800
∴FWHM = X 2 – X 1 + δ 1 + δ 2 = 6.53channels
3-16
PGT Detector Systems Set Up/Calibration
It should be noted that most peaks are Gaussian curves, and the use
of simple linear interpolation will produce an estimate for the FWHM
which can be significantly greater than the true value, especially when
the peak contains a very small number of channels. Similarly, if the cen-
ter of the peak lies between two channels, then the maximum count P
can be significantly smaller than the true peak height, leading to a slight
over-estimation of the FWHM. In general, to avoid such errors use the
calibrations listed in the following table.
Approx. (eV/ch)
Peak Energy Position of Peak in
Detector Source Needed to Measure
keV MCA (channel no.)
Resolution
55
Si(Li) Fe 5.9 ~1000 6
small planar
55
large planar Fe 5.9 ~400 15(a)
N-type coaxial
57
planar, Co 122 ~2500 50
coaxial
60
large planar, Co 1332 ~7500(b) 200
coaxial
(a)
With large planars and N-type coaxial detectors, the two peaks (5.9 and 6.4 keV) are not re-
solved. It is possible to measure the energy calibration using the 6.4 and 14.4 keV peaks from a
57
Co source, measured separately.
(b)
For multichannel analyzers with only 4K channel display, it is usually possible to use an 8K
ADC conversion gain, together with a 4K digital offset, to obtain the necessary calibration for
60
Co. Analyzers with less than 4K channels are not recommended.
3-17
Set Up/Calibration PGT Detector Systems
3-18
Section 4
Detector Theory
4
Parameters
PGT Detector Systems Detector Parameters
4.1: Detector
Theory
4-3
Detector Parameters PGT Detector Systems
Compton effect:
E’
E
E θ E ′ = ------------------------------------------------------
1 + E ( 1 – cos θ ) ⁄ 511
E
T CE = -------------------------------
1 + 511 ⁄ 2E
4-4
PGT Detector Systems Detector Parameters
10
PE
Linear absorbance coefficients (cm4)
PE = photoelectric effect
C = Compton scattering
1 PP = pair production
10-1 Germanium
Silicon
PP
10-2
10-3
10-2 10-1 1 10 102
Energy (MeV)
4-5
Detector Parameters PGT Detector Systems
Fig. 4-2 shows the pulse height spectral distribution for monoenergetic
gamma rays incident on a large coaxial germanium crystal, with the
contributions of the various events labelled. The full energy peak is the
sum of photoelectric events, totally absorbed multiply scattered Comp-
ton events, and totally absorbed pair events.
102
θ < 90° from active volume θ > 90° from active volume
θ > 90° from inactive volume θ < 90° from inactive volume
0 500 1000
Eγ (keV)
Fig. 4-2 Spectral distribution of a monoenergetic gamma ray (1.4 MeV)
incident on coaxial germanium.
4.2: Spectral
Performance
Parameters
The most important parameters characterizing a radiation detector are
efficiency, energy resolution, and peak-to-Compton ratio. The following
discussion of these items is based on the ANSI/IEEE Standards 645-
1977 and 325-1986. Please consult also ANSI Standard N42.14-1978
and the ANSI/IEEE Standards 680-1978, 300-1982, 759-1984B, 1131-
1987, 301-1988, and their updates, as well as any other ANSI/IEEE
standards which may relate to your applications. NIST Standard Refer-
ence Material (SRM) Standards and US Regulatory Guides for fluores-
cence applications also contain valuable information.
4-6
PGT Detector Systems Detector Parameters
4.2.1. Efficiency
The counting efficiency of a detector is a function of the active volume
and shape (geometry) of the detector crystal, the source/detector ge-
ometry, and interactions in the materials in the immediate vicinity of the
detector. The tests described in ANSI/IEEE 325-1986 call for a spectral
peak at least 6 channels wide at half-maximum, with the total number
of counts within the FWHM of at least 50,000.
Ea = A ⁄ Ns
where Ns is the total number of such photons emitted by the source dur-
ing the live counting time.
E
rel
= A ⁄ ANaI
Since the 60Co source used decays at a rate of about 1.1% per month,
the source intensity Is depends on the age of the source, i.e., time
elapsed since it was itself calibrated. This is determined from the fol-
lowing formula:
Is = I0 e-0.693∆t/T
4-7
Detector Parameters PGT Detector Systems
100
20% N-type
70
Ge absorption edge
50
30
20
20% P-type
10
Intrinsic Detector Efficiency (percent)
7
Beryllium
window
5 thickness
2 5 10 20
(mils)
3
0.7
30
0.5
0.3
0.2
Detector 15
Thickness 13
(mm)
5 7 10
0.1
1 2 3 5 7 10 20 30 50 70 100 200 300 500 700 1000 2000 3000
Photon Energy (keV)
Fig. 4-3 Intrinsic Detector Efficiency for various crystal thicknesses and types, determined for a
point source approximately 25 cm. from the crystal face.
where “area” is the active area of the detector crystal and r is the dis-
tance from the source.
4-8
PGT Detector Systems Detector Parameters
where ∆Es is the total spectral energy resolution at energy E 1 and ∆Ns
is the number of channels in the peak at FWHM.
4-9
Detector Parameters PGT Detector Systems
4-10
PGT Detector Systems Detector Parameters
Background
drawn at 100× for clarity
FWHM
Si(Li):
138 eV
small planar IG:
down to 125 eV
5.9 keV
PGT Si(Li) detectors have P/B ratios from 1500:1 to 10,000:1, depend-
ing on window material, with a typical value of 8000:1. For IG detectors,
P/B ranges from 200:1 to 10,000:1, with 1000:1 typical, depending on
the window material.
4-11
Detector Parameters PGT Detector Systems
4.3: Other
Detector
Geometries Two special detector geometries are designed for samples that are ei-
ther very small or very large: A well detector has a reentrant cavity in
the endcap and is used for very small samples. A Marinelli beaker is an
inverted well beaker that fits over the endcap. It is used for larger sam-
ples of low activity.
For 60Co, the 1332.5 keV line is coincident with the 1173.2 keV line.
The sum peak is therefore at 2505.7 keV. The in-well efficiency (W) is
then
A + As
W 60 = --------------- where A = area of the 1332.5 keV peak
Co N s As = area of the 2505.7 keV sum peak
Ns = total number of 1332.5 keV photons
emitted by the source during the live time
NOTE: PGT also determines the absolute efficiency (Ea) of the 40 keV
line of 129I to give a measure of the absorption of the sample hole ma-
terial and the detector internal dead layer:
A 40keV
W 129 = ------------------ where A40 keV = area of the 40 keV peak
I Ns Ns = total number of 40 keV photons
emitted by the source during the
live time
4-12
PGT Detector Systems Detector Parameters
Marinelli beaker
end
cap
Centering support
dewar
Two beaker sizes are standard: 450 mL for detectors of diameter ≤76.5
mm and 1000 mL for larger detectors. These are filled (to ±2 mL of the
stated volume) with a solid or liquid carrier containing uniformly
dispersed radioactive material. The source is either a certified MBSS
(Marinelli Beaker Standard Source) or calibrated MBSS.
The most widely used line for specifying efficiency and resolution is the
1332 keV γ ray of 60Co; for low energies, the 88 keV γ ray of 109Cd is
preferred.
4.4: Other
Factors
4-13
Detector Parameters PGT Detector Systems
During the heating cycle, outgassed vapors may contaminate the de-
tector. If this happens, the detector must be returned to the factory for
reprocessing. To minimize the possibility of detector contamination,
lower temperatures with correspondingly longer annealing times
should be used. Pumping the cryostat during the heat cycle is also
highly recommended. A high vacuum system with an LN2 cold trap is
best, although a mechanical vacuum pump which can produce an ulti-
mate vacuum of better than 20 microns may be suitable. The LN2 cold
trap is absolutely necessary to prevent backstreaming of oil.
Note: after several years, the molecular sieve in the cryostat may be
loaded to the point that a positive pressure is generated in the cryostat
when it is warmed above room temperature. It is then necessary to
pump the cryostat during the whole procedure.
4-14
PGT Detector Systems Detector Parameters
100
60 0.55 mm
% Transmission
50
40
30
20
10
0
1 2 5 10 20 50 100 200
Energy (keV)
Fig. 4-6 shows the percent transmission for beryllium, aluminum, and
copper entrance windows of various thicknesses.
4-15
Detector Parameters PGT Detector Systems
( I0 – I )
---------------- = ( 1 – e –µ d ) (for absorbed γ rays)
I0
Step 1. Find the value of the attenuation coefficient (µ) for the se-
lected energy from the curve labelled “Total” in the graph
of the chosen crystal material.
Step 4. Draw a straight line connecting these two points and read
off the transmission/absorption value where the connect-
ing line intersects the center line of the nomogram.
For example, at 0.2 MeV, a 2-cm. Ge crystal has a total linear attenu-
ation coefficient of 1. The line connecting 1 on the left scale to 2 on the
right scale crosses the center line at 0.90 absorption (0.10 transmis-
sion), as shown by the dotted line on the nomogram.
4-16
PGT Detector Systems Detector Parameters
Transmission/Absorption Nomogram
100 100
Transmission Absorption
50 50
20 20
10 10
5 5
0.001 0.999
2 0.01 0.99 2
0.05 0.95
0.10 0.90
0.20 0.80
Linear attenuation coefficients (cm-1)
1 0.30 0.70 1
0.40 0.60
0.50 0.50 Crystal length (cm)
0.60 0.40
.5 0.70 0.30 .5
0.80 0.20
0.85 0.15
0.90 0.10
.2 0.95 0.05
0.96 0.04 .2
0.97 0.03
0.98 0.02
.1 0.99 0.01 .1
0.995 0.005
.05 .05
0.999 0.001
.02 .02
.01 .01
4-17
Detector Parameters PGT Detector Systems
Silicon
1000
500
200
100
50
20
10
Linear Attenuation Coefficients (cm-1)
Total
1
0.5
Compton
0.2
0.1
0.005
0.001
0.01 0.02 0.05 0.1 0.2 0.5 1 2 5 10
Energy (MeV)
4-18
PGT Detector Systems Detector Parameters
Germanium
1000
500
200
100
50
20
Linear Attenuation Coefficients (cm-1)
10
Total
2
0.5
Compton
0.2
0.1
0.05
Pair
Photoelectric Production
0.02
0.01
0.01 0.02 0.05 0.1 0.2 0.5 1 2 5 10
Energy (MeV)
4-19
Detector Parameters PGT Detector Systems
Beryllium
100
50
20
10
2
Total
1
Linear Attenuation Coefficients (cm-1)
0.5
0.2
Compton
0.1
0.05
0.02
Photoelectric
0.01
0.005
0.002
0.001
1 2 5 10 20 50 100 200 500 1000
Energy (MeV)
4-20
Section 5
Preamplifiers 5
Resistive Feedback
Lower-Power
Pulsed Optical
Transistor Reset
Table 5-1: PGT PREAMPLIFIERS
RG-11 or
352 PO-14B TR
RG-15
Connectors
Input Hard wired Hard wired Hard wired Hard wired
Detector Bias SHV SHV SHV SHV
Timing Output BNC BNC -*- -*-
Signal Output BNC BNC BNC BNC
Test Pulse BNC BNC BNC BNC
Power 9-Pin Amphenol 9-Pin Amphenol 9-Pin Amphenol 9-Pin Amphenol
5.1: Preamplifiers
See Sec. 5.2.1 for description, layouts, and circuit diagrams of the re-
sistive feedback preamps.
5-3
Preamplifiers PGT Detector Systems
reset
decay time time
Fig. 5-1 Decay/Reset patterns for resistive feedback (left) and pulsed-reset preamplifiers (right).
Further, if the time between events is less than 1.5× the peaking time
of the main amplifier, regardless of the type of preamplifier, the pulses
pile up and cannot be processed. Main amplifier and ADC (analog-to-
digital converter) dead times are the major contributors to system dead
time.
5-4
PGT Detector Systems Preamplifiers
1Q – 19
--------------- × ( 1.6 × 10 Coulomb ⁄ Q ) × Mev ⁄ sec = I D
2.98ev
ID × R f = V TP
detector feedback test-point
current resistor voltage
These relationships are more easily seen in the following example for
a detector with a resistive-feedback preamplifier: In a germanium crys-
tal there is one charge carrier produced for every 2.98 ev of energy de-
posited in the crystal. At 1.6 × 10-19 coulomb per charge carrier, 1 MeV
of energy falling on the detector produces a charge of 0.53 × 10-13 cou-
lombs. Thus, if 1 GeV/sec is detected, the current generated is 0.053
nAmps. This current flowing through a 10GΩ feedback resistor raises
the voltage at the preamplifier test point by 0.53V. As the energy rate
is increased to 44 GeV/sec, the test point voltage increases to >23.3 V,
at which point the preamplifier shuts down. Therefore, for this example,
44 GeV/sec is the maximum energy rate (ERmax) that the detector/
preamp assembly can handle. The maximum count rate is determined
by dividing this value by the mean energy of the spectrum. In this par-
ticular system, the maximum count rate for 57Co would thus be approx-
imately 440,000 counts/sec.
It should be noted that the count rates are those observed in the total
spectrum. Once the maximum energy rate of a given system is known,
count rates for other spectra can be calculated. For example, in the sys-
tem above with a maximum energy rate of 60 GeV/sec, the maximum
total count rate for 137Cs (662 keV line) with a mean energy of ≈0.5
MeV is 120,000 counts/sec.
5-5
Preamplifiers PGT Detector Systems
Using a lower value feedback resistor can yield a higher maximum en-
ergy rate, but with a loss in resolution. For high count rates and high
energies, the use of a transistor reset preamp is much better.
IGC/NIGC RG-11/15 70
IGP RG-11/15 6
(<500 mm2)
IGP RG-11/15 12
(>500 mm2)
5-6
PGT Detector Systems Preamplifiers
Nopt
Noise2
1/ ƒ
Ns 2
Np 2
Topt
1.0 10 100
Pulse Processing Time (µsec)
Fig. 5-2 shows how the parallel and series noise contributions, and the
total noise, in a resistive feedback preamplifier vary with the pulse pro-
cessing time. Minimum electronic noise (Nopt) occurs at Topt, when
these two components are equal. The 1/ƒ contribution, indicated by the
dashed line, does not change the shape or position of the minimum on
the time axis, but only its height (noise value). The variation of detector
resolution with pulse processing time essentially coincides with the
curve for NTot2.
Decreasing the shaping time to 0.5 Topt can double the count rate ca-
pability of a system while increasing the noise, and therefore resolution,
only about 12%.
The optimum shaping time (Topt ) for a given germanium detector sys-
tem is typically around 4 µsec Gaussian shaping time, corresponding
to approximately 10 µsec pulse processing time. (See also Table 5-3.)
5-7
Preamplifiers PGT Detector Systems
These figures are meant as a general guide and assume the use of a suitable main
amplifier, pile-up rejector, and multichannel analyzer.
5-8
PGT Detector Systems Preamplifiers
10
N2 = Quadratic sum of
parallel and series noise
Np = Parallel noise
Ns = Series noise
Noise2
N2
Ns2
Np 2
Topt
0.1
1.0 10 100
For a given shaping time, high count rates can lead to the possibility of
pulse pile-up in which pulses arrive at the multichannel analyzer (MCA)
almost simultaneously. The MCA treats such a double pulse as a single
pulse of greater amplitude (and therefore, energy). This leads to distor-
tion in the spectrum: extra high-energy counts, fewer counts in the pho-
topeak, and increase in the FWHM for the full energy peak. This
spectrum degradation becomes significant when the product of input
count rate and pulse processing time exceeds 0.5, and it is a dominant
factor when this value is >1.
5-9
Preamplifiers PGT Detector Systems
5-10
PGT Detector Systems Preamplifiers
Ns 2
R2
R2
Noise2
Noise2
Np 2
FWTM
FWTM
BD2
BD2
Ns2 Np 2
Topt Topt
Fig. 5-5 Combined effect of noise components and ballistic deficit on system resolution.
5-11
Preamplifiers PGT Detector Systems
5.2: Preamplifier
Layouts and
Circuit
Diagrams
Layouts of the standard Models RG-11 B/C, RG-15, and 352 are shown
in Fig. 5-6 on page 5-13, and their circuit diagrams and component lay-
outs follow. Directions for replacing the preamplifier are also included.
The RG-11 B/C is commonly used for gamma counting between 5 keV
and 10 MeV. It has a standard energy rate product of 50,000 MeV/sec,
but can be modified to increase this value.
The Pulsed Optical Preamplifiers PO-12 and PO-14B use an LED to re-
set the storage capacitor. Their performance is close to that of an ideal
charge-sensitive preamp, and they are particularly suited for such low-
noise applications as the measurement of soft gamma or X rays with
Si(Li) and small-to-medium intrinsic germanium detectors. Energy res-
olution remains optimized, almost independently of count rate, up to
the performance limit of the main amplifier. The PO-14B has a maxi-
mum energy rate product of 250,000 MeV/sec. An internal delay (up to
120 µsec) after the reset threshold is reached prevents loss of the trig-
ger event and thereby avoids preferential loss of high energy pulses.
Fig. 5-7 on page 5-23 shows the layouts of the PO-12 and PO-14B
Preamplifiers. with their circuit diagrams, component layouts, and re-
placement procedures on the following pages.
5-12
PGT Detector Systems Preamplifiers
black yellow
Feedthrough Connections (rear panel):
red white
HV Feedthrough
green
HV HV HV
.002 .002
µF µF
HV HV HV
TP
BIAS Output Output
TEST OUTPUT
POWER
Front Panel
(RG-11 or 352 external)
Bias Test
+12 V
5 4 3 2 1
Front Panel
(RG-15 or 352 in-line)
9 6
-12 V 8 7
+24 V -24 V
thermistor output (352 only) Fig. 5-6 Layout of Models RG-11, RG-15 and 352
Preamplifiers.
9-pin Amphenol (expanded)
5-13
RG-11B Circuit Diagram
5-14
TP1 TEST
WHITE WHITE
TEST J3 BNC
R7
R33
Preamplifiers
JACK TEST
POINT J6
R34
RED TP4 5 CURRENT
FB SENSE
L2 RED
DETECTOR 7 +24V
+
CR1 R6 C12
R25
R4 Q2 R22
Q8 P1
1 GND 9 PIN SUB-D
C13
R5 MALE
Q9
+
R8 L3
C4 ORANGE
4 +12V
+
* * R9 C14
C9 R26
R35
+
C17
FRONT
Q5 R20 PANEL
VIOLET R18
TP2 *R13 Q10 Q11
Q3
Q1 D J4
R31
3 2 R11 R24 R27
1 C8 C10 Q6 Q7
4 TP3
* OUTPUT
S Q4
YELLOW R12 R15 R28 BNC
+
R19 C15
+
C7 R32
R16 * C11 R23 J5
C6 R17 Q13
Q12
R21 C18
R10
+
CR2
R29
L4 VIOLET
X-TAL 9 -12V
C16
+
R30
L5 BLUE
6 -24V
C20
+
(STANDOFF)
GREEN HV For room temp. FET use this configuration J2
(OPTION A)
BIAS
R1 SHV
*
For cold FET use this configuration
C1
*Selected (OPTION B)
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
PGT Detector Systems Preamplifiers
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
J3
(solder side) white
To TP1 TEST BNC
(solder side)
DETECTOR To TP4 R33
J6 JACK TEST POINT
TEST To TP1
(white) R34 5 CURRENT SENSE
C12
C4
L2 red
C9
R5
CR1
L3 7 +24V
DRAIN To TP2
R22
(violet) orange
R6
R8
4
R35
R7 R25 +12V FRONT
R26
C8
PANEL
C14
R13
SOURCE To TP3 C17
(yellow) TP1 Q5 Q2
R4
R20
TP2 Q8 Q9 Q10 Q11
GND Q3 TP4
R9
To TP3 J4
R11
(black) C13
R32
R27
C15
C10
FEEDBACK
R28
To TP4 Q7
R18
Q4 Q6 R24 R31
(red)
TP3 J5
R16
R17
THERMISTOR
(orange) To Pin 8 Q12 Q13
(amphenol) R12
1 GND
R30
HV
R23
C18
C6
(green)
R29
C11
R10
C7
R15 violet
R21
CR2
C1 R19 9 -12V
ARCO L4
C16
blue
C20
L5 6 -24V
R1 BIAS SHV
J2
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
Components
5-15
RG-15 Circuit Diagram
5-16
TP1 TEST
WHITE WHITE
TEST J3 BNC
R7
R33
Preamplifiers
JACK TEST
POINT J6
R34
RED TP4 5 CURRENT
FB SENSE
L2 RED
DETECTOR 7 +24V
+
CR1 R6 C12
R25
R4 Q2 R22
Q8 P1
1 GND 9 PIN SUB-D
C13
R5 MALE
Q9
+
R8 L3
C4 ORANGE
4 +12V
+
* * R9 C14
C9 R26
R35
+
C17
FRONT
Q5 R20 PANEL
VIOLET R18
TP2 *R13 Q10 Q11
Q3
Q1 D J4
R31
3 2 R11 R24 R27
1 C8 C10 Q6 Q7
4 TP3
* OUTPUT
S Q4
YELLOW R12 R15 R28 BNC
+
R19 C15
+
C7 R32
R16 * C11 R23 J5
C6 R17 Q13
Q12
R21 C18
R10
+
CR2
R29
L4 VIOLET
X-TAL 9 -12V
C16
+
R30
L5 BLUE
6 -24V
C20
+
(STANDOFF)
GREEN HV For room temp... FET use this configuration J2
(OPTION A)
BIAS
R1 SHV
*
For cold FET use this configuration
*Selected (OPTION B)
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
PGT Detector Systems Preamplifiers
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
CR1
C12 C9 red
DETECTOR C4 L2 7 +24V
+
R5
orange
+ + L3 4 +12V
TEST BNC
R6
To TP1 black
R35
(white) R7 R8 R25 1 GND
R22
C8
R13
C14
R26
DRAIN TP2 Q5 Q2
To TP2 R4
(violet) Q8 Q9 Q10 Q11
TP1 FRONT
R20
Q3 R33 PANEL
R11
R9
SOURCE R34 C13 C17
To TP3 R32
(yellow)
R27
C15
TP4
C10
R28
Q4 Q6 Q7 R24 R31
GND To TP3 TP3
R18
(black)
R17
C18
R16
FEEDBACK To TP4 R12
R30
Q12 Q13
R21
R23
(red)
HV
R29
C6
To SHV
C11
R10
(green) C7
connector R15
CR2
R19 violet
L4 9 -12V
C20 blue
C16 L5 6 -24V
+ +
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
Components
5-17
Model 352 Low Power ±12 V Preamplifier Circuit Diagram
5-18
Preamplifiers
TRIMPOT. R4
L2
C8 R13
+
+12 V
+
R14 PIN 4
R11 R8 C4
R9
R21
Q4 Q3 C6 A ONLY = Negative Bias Detectors
+
Q5 B ONLY = Positive Bias Detectors
C12
R5 R3 Positive Bias
B B
+
R10 Q2
+
R17 C9
TEST
C1 R1
TP 3 C2
TP 1 OUTPUT 1
D Q1
+
C R19
R12 C7 +
OUTPUT 2
.5 pf A A
Negative Bias
INPUT C5
TP 4 R20
+
2 GΩ 2 GΩ .5 pf R16 TEST POINT
E Q6 G
TP2 GROUND
+
L1 PIN 1
R15 C10 R6 R2
R7
-12 V
PIN 9
C3
+
GE(LI) HV FILTER
F R18
BIAS
C11
For component specifications, see Model 352 Lower Power Preamplifier Component Layout (page 5-19).
PGT Detector Systems
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
PGT Detector Systems Preamplifiers
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
Model 352 Lower Power Preamplifier
C4
DETECTOR
C8
C7
R12
Orange
L2
R11
C6
+12 V POWER
R14
TEST
R8
pin 4
R9
To TP3
R5
(white)
R4
R21
R13
DRAIN To TP1 Q5 Q4 Q2
(violet) Q3 OUTPUT GND
R3
TP1
C9
SOURCE C12
To TP2 TP2 C2
(yellow) TP4 R20 TEST POINT
C5
GND White
To TP2 OUTPUT 1
(black)
R6
TP3 Q6 Q1 Black POWER GND pin 1
R1
Yellow
FEEDBACK TEST
To TP4
(red) C1
R7
R15
R10
R2
C10
R17
R16
C3
HV C11 L1
(green) R19 OUTPUT 2
Violet
-12 V POWER
pin 9
R18 BIAS
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
Components
5-19
Preamplifiers PGT Detector Systems
Step 3. Resolder the wires: the green wire to the HV post, and the
red, black, violet, yellow, and white wires to the pins on the
preamplifier (again, see Fig. 5-6 and be sure to use a heat
sink on the capacitor).
5-20
PGT Detector Systems Preamplifiers
5.2.2.3. Optimization
Step 1. Be sure the detector is properly cooled!! Then connect the
preamplifier power, HV. and signal cables, as shown in
Section 3.1.3. Connect an oscilloscope to the output of the
main amplifier and observe the high noise level.
Step 4. Place a 60Co source next to the detector and turn the pole
zero (P/Z) potentiometers on the preamplifier and on the
main amplifier to obtain the right pulse shape (see Fig. 3-
2 and Fig. 3-3).
Step 4. Unsolder the red, black, violet, yellow and white wires at
the preamplifier end and the green wire at the HV mount-
ing post (Teflon insulator in board).
Step 7. Remove the two mounting blocks from the old preamp by
removing the two 4-40×5/16" flat-head phillips screws.
5-21
Preamplifiers PGT Detector Systems
Step 3. Align the mounting block holes with the mounting holes in
the flange. Make sure the wires of the preamp are on the
opposite side of the SHV connector. Screw in the socket-
head 4-40×3/8" screws.
5-22
PGT Detector Systems Preamplifiers
IFET IFET
Teflon Gain
2 GΩ mounting 2 GΩ
posts
Feedback
(red)
VFET
HV HV
current
sense +12 V
BIAS OUTPUT
TEST 5 4 3 2 1
INHIBIT
PO-12
POWER -12 V 9 8 7 6
+24 V
current
Front Panel sense +12 V
thermistor output (optional) 5 4 3 2 1
9 8 7 6
PO-14B
-12 V +24 V -24 V
5-23
PO-14B Circuit Diagram
5-24
CRI R9 +18V VOUT Q1 VIN J1
R2 Amphenol
+ 1 3
17-20090
R1 C3
Preamplifiers
3 + R10 C2 2
2 R4 R5 R6 C1 R8
R3 R7 J2 SIGNAL +24V
1 OUTPUT 7
R11
Q2 +
C7
+ C4 CR3
VOUT Q3
VIN 4 +12V
3
R12 R13 CR2 Q4 Q5 +5V 1 GND
2
1
C6 +
R47† C5 R14 C9
Q6 R19 R20
CR4 C8
R15
R18 15
CR5 14
R21 14
1 7 R16 R17 3 8 2 E2 Q 13 13 11
CR6 7 10 8
D 2 Q8 Q9 3 E1 E3
12 E3
2 Q7 R22 1 4 1 C10 7 9
FET 6 R24 2 120µS
XTAL S + 1 Q10 3 - 4
CR8 5
R23 CLR Q
3 C11 C12
R26 R28 CR7 R31 +
LED- R25 5 6
R30 C13 2
+ 3
R32 E3
R29 E3 4
R33 1
VOUT
†
COMPONENTS FOR CIRCUIT OPTIONS (for special applications; normally omitted)
R36: Gain potentiometer option to fine-tune gain. This might be used in detector arrays where matched gains are required. This pot. (typically 5K) then replaces the 2.7K.
For the component specifications, see the PO-14B Component Layout (page 5-25).
PGT Detector Systems
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
PGT Detector Systems Preamplifiers
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
Output BNC (J2)
C3
Q1 R34 red
C2
CR1 TP7 +24V pin 7 (J1)
C5
+LED R9 green
To TP1
C4
Leakage current
R5
8
R8
7
CR3
R1
(yellow) Q2 R37 *
R4
C7
C6 monitor pin 5 (J1)
C20
R2
R10
R12
CR2 orange
GND R41 R38 E3 +12V pin 4 (J1) *
R11
To TP2
R14
(black)
8
Q4 Q5 black
*
R13
Q8 Q3 Ground pin 1 (J1)
Q6
R6
R7
DRAIN To TP3 R3 CR5 R39
R15 E2
14
(violet)
1
CR4
C9
CR6 R16
L1 or R47 CR7 C10
SOURCE R24 R19
To TP4 R40 Q10 Q7
(blue) R18
R21
R29
R25
16
C8
1
R36
1
Q9
8
FEEDBACK
R31
CR8
To TP5 E1
R28
R20
R23
(red) *
R30
violet -12V pin 9 (J1)
C12
R42
R26
HV R17
R44
5
To SHV
4
(green) connector R22 blue *
R45
C11
C15 -24V pin 6 (J1)
C13 R32
THERMOCOUPLE To TP6
R33
C21 C22
C16
C17
(orange) Q11 R46 Test BNC (J4)
Components
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
DIODES RESISTORS (all are 1/8W, 5%, unless otherwise specified;
CR1-8 IN914 values in ohms)
R1 680
IC’s R2,R11 47
E1 LM311N R3,R13,R14,R26,R41 100
E2 74LS123 R4,R5,R6,R7,R17,R22,R37 10K
E3 74LS00 R8,R29,R30 1K
R12 15K
TRANSISTORS R15,R16 4.7
Q1 MC78L18CP R18 4.7K
Q2,Q4,Q5,Q7,Q9 2N4403 R19 3.9K
Q3 MC78L05 R20 47K
Q6,Q10 2N4401 R21 470
Q8 LM394 R23 33K
Q11 MC79L18CP R24 5K
R25 22K
CAPACITORS (all in µf, unless otherwise specified R28,R44 2.2K
C1,C15 47, 20V R36,R40 2.7K
C2,C4,C7,C12,C16 4.7, 25V R38 6.8K
C3,C17 1.0, 50V R39 100K
C5,C10 330 pf R45 10K
C6 33 pf R9 100, 1/4W
C8 .005, 50V R10,R31 470, 1/4W
C9,C13 8.2, 15V R32,R33 220, 1/4W
C20 .01, 25V R34,R46 47, 1/4W
C21,C22 .002, 3kV R35 1K, 1/4W, 1%
R42 2G, 1/2W
INDUCTOR (for PO preamp, Si(Li) only) R43 100M, 1/4W, 10%
L1 56mH R47 (for TR preamp, Si(Li)) 1K
or (for IG applications) 1K or 1.2K, 1/4W
5-25
PO-12B and PO-12B8 Pulsed Optical Preamplifier Circuit Diagram
5-26
Preamplifiers
R1
R3 7 +24V
+ R32
CR1 R6 R7 C3
R15
C1 4 +12V
R2 R4 Q2 +5V
CR3 5
Leakage Current
R16 Monitor
R5 + C4
C2 + + 9 -12V
CR2 C12 R22 C14
Q5 R24
1 Ground
R35 R17 4 11
DET. Q1 C8 +
* C11 9 9 8
* E1 * R25
Q3 R34 10 E2
CR7 R10 3 2
Q6 Q7 6 INHIBIT
13 11
2 3
R14 E2
C9 C13 R31
C15 R26 1 E2 12
4 6
Q4
+ R9 + E2 Q10
LED
* C5 C6 R13 R18 5 CR5
R8
* R27
Q8 R20 R21
Q9
R19 R28
C10 CR4 R30
CR6
R11 R12 + R29
C7
+ R33 OUTPUT
NOTES: All components designated by are “Selected.” Portion within dotted lines is in cryostat.
PO12B: R18 is used; R34 is not used.
*
PO12B8: R18 is replaced by jumper; R34 is 3.9K, 1/8W.
Please see the PO-12B and PO-12B8 Pulsed Optical Preamplifier Component Layout (page 5-27) for the list of components.
PGT Detector Systems
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
PGT Detector Systems Preamplifiers
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
white
OUTPUT BNC
C12 red
C3
DETECTOR +24V PI -Pin 7
orange
R32 +12V PI-Pin 4
R7
CR1
TEST R16
C4
TP1
CR2
To TP1
(white) R30 GND To BNC LUGS
R15 C55
R6
(OUTPUT, INHIBIT)
CR3
Q5 Q7
R10
Q6
R29
LED C11 R27
To T2 TP2
R35
(yellow) R4 Q3
R14
Q2 C8 Q10 R24
CR6
C9
DRAIN
R20
To TP3 R28 C14
(violet) Q4 R26
TP3 To INHIBIT BNC
*
14
1
R13
14
R34
1
C5
R5
GND To TP4
R31
C6
(black) R9 E1 E2
CR7
TP4
R8
R17 UA
Q8 Q9 74HOO
C15
FEEDBACK To TP5 710PC PI-Pin 5
(red) TP5
GND PI-Pin 1
TP R18 TP
*
7
R2 To TEST BNC
7
THERMISTOR
8
To Pin 8
R3
(orange) R22
R21
R19
R11
R23 R25
HV C10
C2
C13
(green) To SHV C1 R12
connector CR4 violet
C7
R33 -12V PI-Pin 9
R1 To BIAS BNC
* PO-12B: R18 “Selected”
R34 not used
PO-12B8: R18 replaced with jumper
R34 3.9K, 1/8W
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
Components
5-27
Preamplifiers PGT Detector Systems
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
violet
black yellow
HV Feedthrough red white (10Ω resistor)
green
blue green (emitter)
(base)
HV HV
red blue
IFET IFET
green
+18V Q IN Gain -18V Q IN orange
Gain
black black
GND Feedback GND +5V green
Feedback -18V blue red
(orange) B E (orange) B E +18V
FB orange FB orange
green
green
Ground blue
Ground blue
VFET (black) VFET (black)
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
ILED Inhibit ILED
Inhibit
(orange) (orange)
HV HV
current
sense +12 V
BIAS OUTPUT
TEST INHIBIT thermistor output 5 4 3 2 1
(optional)
POWER 9 8 7 6
5-28
TR-14B Transistor Reset Preamplifier Circuit Diagram
CRI R9 +18V VOUT VIN J1
R2 Q1 Amphenol
+ 1 3
17-20090
R1 C3
TR Negative 3 + R10 2
2 C1 C2 7
R3 R4 R5 R6 R7 R8 J2 SIGNAL +24V
1 OUTPUT
TP3 TP4 R11
+5V +18V Q2 +
C7
C4 R7 TP5 + CR3
VOUT VIN 4 +12V
C4 3
R2 R4 FB R12 R13 +5V 1 Q3 1
CR2 Q4 Q5 GND
PGT Detector Systems
2
D1 TP6 C6† +
E + R14
C5 Q6 C9
+ R47 R19 C8 R20
CR4
*
C1 R6 R15
B R18 15
TR CR5 14
+ TP7 1 7 R17 2 R21 14
R16 3 8 E2 Q 13 13 11
C3 CR6 10 8
D 2 Q8 3 E1 7 E3
JP1 Q9 12
E3
TP8 2 Q7 R22 1 4 1 C10 9
FET 6 R24 2 120µS 7
GND S +
TP2 Q1 Q2 1 Q10 3 - 4
R1 R5 CR8 5
R23 CLR Q
3 C11 C12
R26 R28 CR7 R31 +
LED- 6
+ R25 C13 2 5
Q IN R30 + 3
C2 R32 E3
R29 E3 4
D2 R33 1
R3 VOUT
R42
J1
R41
J5
BIAS
C22 R43
THC C21 R44 J4
TEST
R46
†
Value of this capacitor to be determined at time of assembly.
*
denotes special component inside cryostat.
5-29
Preamplifiers
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
TR-14B Transistor Reset Preamplifier Circuit Diagram
5-30
CRI R9 +18V VOUT VIN J1
R2 Q1 Amphenol
+ 1 3
17-20090
R1 C2 C3
3 + R10 2
TR Positive 2 R4 R5 R6 C1 R8
7
R3 R7 J2 SIGNAL +24V
Preamplifiers
1 OUTPUT
R11
TP3 TP4 Q2 +
C7
JP2 -18V + CR3
VOUT VIN 4 +12V
C4 3
C4 R7 TP5 R12 R13 CR2 +5V 1 Q3 1
Q4 Q5 GND
R2 R4 FB 2
-5V D1 TP6 C6†
+ +
E R47 R14 C9
D3 C5 Q6 R19
C1 C8 R20
CR4
*
R6 R15 R18
+ 15
B TR CR5 14
+ 1 7 R17 2 R21 14
TP7 R16 3 8 E2 Q 13 13 11
C3 CR6 10 8
D 2 Q8 3 E1 7 E3
Q9 12
E3
TP8 2 Q7 R22 1 4 1 C10 9
FET 6 R24 2 120µS 7
GND S +
TP2 Q1 Q2 1 Q10 3 - 4
R1 R5 CR8 3 5
R23 CLR Q
C11 C12
XTAL R26 R28 CR7 R31 +
LED- R25 6
+ C13 2 5
Q IN JP1 R30 + 3
+3 C2 R32 E3
R29 E3 4
D2 R33 1
0
R3 VOUT
+18V -18V
C15 C16 VIN
+ + Q11
3 2
TP1 1 9 5
R34 J3
E2 Q INHIBIT
10
INH
† 11 CLR Q 12
R36 C17
30-
3
R36 300µS INH
2 * R37 7 6
C .002 FBK 1 C20
R39 5 LEAKAGE
1 R38 CURRENT
2GΩ * For 2X gain, remove jumper. (20µS Min)
R40 3 MONITOR
2
R45 +5V 9 -12V
2 6
Q IN 3
-24V
1
R42
J1
R41
J5
BIAS
C22 R43
THC C21 R44 J4
TEST
R46
†
Value of this capacitor to be determined at time of assembly.
* denotes special component inside cryostat.
For Component specifications, see TR-14B Transistor Reset Preamplifier Component Layouts (page 5-31).
PGT Detector Systems
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
PGT Detector Systems Preamplifiers
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
TP3 TP4 -18V TP3 TP4
+5V +18V
blue green red
TP2 orange JP2
R2 orange orange TP5
Q IN TP5 TP2 orange R2
R1 Feedback Feedback
Q IN R1
Q1 JP1 Q1
C4
D1
R4
R7
D1
C4
R4
D3
R7
JP1
R6
Q2
C1
Q2
R6
TP1 red
TP1 blue
C1
+18V
-18V
D2
green TP6
D2
Emitter green TP6
Emitter
C3
R5
R3
C2
blue TP7
R5
R3
blue
C2
TP7
C3
Base Base
black
TP8 Ground TP8 black
Ground
Components
TR Positive Components TR Negative Ccomponents
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
CAPACITORS C1,C2,C3 15, 10V
C1,C2,C3 15, 10V C4 22 pf
C4 22 pf
5-31
TR-1 Transistor Reset Preamplifier Circuit Diagram
5-32
Preamplifiers
TP8
THERMISTOR
Thermistor
R1 R2
HV-OUT HV_IN
C1 C2 TP1 TP2
‘IDS’
GND GND FET
CURRENT
ADJUST
L1 R6 Q1 R8 R9
1 3
+12 V
CW
2
C3
C4
GND D7 D1
R7 C10
GND
TP3 GND
VDRAIN + 7
3 GND
+ U1
R14
6 OPAMP_OUT
AD829 PREAMP_OUT
LOW OUTPUT IMP.
2 5
- ‘VDS’ 1
- 4 FET R13
VOLTAGE GAIN
ADJUST CW
ADJUST CW 1 ‘GAIN’
C5 2 R16
2 R11
C6
GND 3
GND 3
D2
+ R15
C7
R12
TP5 GND
FEEDBACK GND
TP6 -12 V
VSUBSTRATE R19
R17 CW 1 R18
2 D3
TP7 C8 C9 ‘SUB’
FET
GND + + 3 SUBSTRATE
ADJUST
GND
PGT Detector Systems
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
TR-1 Transistor Reset Preamplifier Circuit Diagram
(Continued)
R23 +12 V
C13 C14 +5 V
R30
GND U3
R25 R24
7 RC R31 U4
C15 5
PGT Detector Systems
6 Q & U4
R29 C 12 I1 &
+ 8 9 11 9 I1
U2 A 13 O1 8
2 + 10 B I2 10 O1
12 I2
5 Q2 11 CLR QN
6
LM311
OPAMP_OUT R21 C16
3 1
- - 4 GND U4
&
R22 -12 V +5 V 4 I1
O1 6
GND 5 I2
D4 D5 C11 C12
U4
GND 1 &
I1 3
R27 R26 2 I2 O1 NC1
1
CW GND
2 R28
‘RAIL’
LOWER RESET
THRESHOLD 3
ADJUST
GND R36 Q5 R39
+5 V
U3
D6 15
C17 RC R40
R35 R37 13
14 INHIBIT
C Q
1
A
2 4
GND B NC2
3 QN
CLR
R32 R38
Q3
+5 V
R34
Q4 +12 V
Q6
IN OUT
OPAMP_OUT R33
GND
TP4 C22
RESET C18 C19 C20 C21
-12 V
C23
C20 for U3
5-33
Preamplifiers
C21 for U4
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
5-34
TR-1 Transistor Reset Preamplifier Circuit Diagram
(Continued)
Preamplifiers
J1
12 11
HV_IN
1 6 Q7
+24 V
2 7
IN OUT -12 V
THERMISTOR 3 8
4 9
GND
5 10
+
INHIBIT 20 19 C24
17 16
PREAMP_OUT
GND GND Q8
124 V
IN OUT -12 V
GND
C25
+
GND
PGT Detector Systems
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
PGT Detector Systems Preamplifiers
PROPRIETARY INFORMATION: This document contains information considered proprietary to PRINCETON GAMMA-TECH, INC., and is issued for maintenance purposes only. Reproduction of all or any part is strictly
DRAIN R13 R6 C3
TP3 TP2 TP1 R9
R15
yellow D6
Q1
L1 R26
R16
OUT
C23 R8
TP5 C6 GAIN C21
FDBK
D7
C22
VDS
R7
GND
Q3 Q6
D1
R32
C18
Q5
TP10
RST
red
D2 R28
INH
U1 D5 R21 U4
C7
RAIL
Q4
TP8 R33 R35
R23
D4
TH
R12
R38
+24 V
R34 R27 R36 C13
C5
TP7
C8
R37 C20
R39
GND
black R19
-24 V
C9
U2 C24
C17
R18
TP6
R25
SUB
R17 SUB
GND
blue
Q7
Q2 C25
R1 U3 C16
C19
C11
TH
D3
C12 Q8
R29
R30
R24
C14
C15
C1 C2 R22 R31
R40 C10
prohibited without the written consent of PRINCETON GAMMA-TECH, INC., 1200 State Road, Princeton, NJ 08540
Components
DIODES INDUCTOR
D1, D4-7 1N914 L1 56mH
D2 1N5234
D3 1N5239 RESISTORS (values in ohms)
R1 2G
IC’s R2 100M
U1 AD829 R6,R8,R22,R23 100
U2 LM311 R7,R18,R26,R29,R38 10K
U3 74LS123 R9 33
U4 74HCT00 R11, R28 5K
R12 1.8K
TRANSISTORS R13, R27 2.2K
Q1, Q3, Q5 2N4403 R14, R40 75
Q2 BS170 R15 2.7K
Q4 2N4401 R16 20K
Q6 78L05 R17 100K
Q7 7812 R19 1K
Q8 79L12 R21 3.9K
R24, R33, R39 39K
CAPACITORS R25 12K
C1,C2, C23 2.2 nf R30 15K
C3 4.7µf R31 470
C4, C5, C11, C13, C20, C21 220 nf R32, R35 5.6K
C6, C22 47 pf R34, R36 8.2K
C7, C8, C9, C10, C12, C14, C19 8.2 µf R37 22K
C15 5 nf
C16 330 pf
C17 10 nf
C18, C24, C25 2.2µf
5-35
Preamplifiers PGT Detector Systems
Step 6. Remove the saddle block from the old preamplifier by un-
screwing the two socket-head 4-40 screws, using the
same wrench.
Step 3. Resolder the wires: the green wire to the HV post and the
red, black, violet, yellow and white to the pins on the
preamplifier in that order. (Again, see the appropriate Fig.
5-7 or Fig. 5-8 for your system.)
5-36
PGT Detector Systems Preamplifiers
5-37
Preamplifiers PGT Detector Systems
5-38
Section 6
6
Troubleshooting
PGT Detector Systems Troubleshooting
6.1: Maintenance
It is normal for the cryostat to be cold and wet when in an open area at
low temperature and high humidity. However, the condensed water
should soon evaporate when the cryostat is brought to normal ambient
temperature (20°C). A cold, wet cryostat at room temperature may in-
dicate a vacuum leak. If that is the case, DO NOT TRY TO OPERATE
THE DETECTOR. CONTACT PGT.
6-3
Troubleshooting PGT Detector Systems
log Counts
Energy (MeV)
To delay repair and get significant additional use from an N-type detec-
tor which shows evidence of neutron damage, observe the following
precautions:
When unacceptable degradation has occurred, PGT can repair the de-
tector promptly. An annealing kit is also available (see “Annealable De-
tectors” on page 4-14.
6-4
PGT Detector Systems Troubleshooting
6.2: Fault
Diagnosis
Each PGT detector is guaranteed to meet all of the ordered perfor-
mance specifications. If there appears to be a problem with the system,
use the following troubleshooting procedures to identify and eliminate
the problem.
The following pages describe various faults, their possible causes, and
some remedial steps to take.
As a first step, always thoroughly check the main amplifier, cables, and
multichannel analyzer (MCA) independently of the detector in which a
fault is suspected. If possible, do this with another detector or with a
pulser unit.
6-5
Troubleshooting PGT Detector Systems
Table 6-1: Troubleshooting Guide for 352, RG-11 and RG-15 Preamplifier Detectors
MCA controls improp- If pulses are present at the input, see the MCA manual and
erly set. Possible check the control settings. Be sure that the amplifier output
improper amplifier set- is positive. You may be misled if the oscilloscope is acciden-
ting. tally set to “invert.”
2. MCA registers Pulses corresponding Use an oscilloscope to examine the signal which is going to
background counts, but no to spectral peaks have the MCA. Adjust the amplifier gain to achieve pulse heights
peaks are visible. voltage heights out- within the input range of the MCA. (Make sure that you know
side the range of the this range, which may change with MCA settings, e.g., digital
pulse height analyzer. offset or conversion gain.)
3. No amplifier output for Amplifier not receiving Check preamplifier output with the oscilloscope set on 50
gamma ray source or pulse pulses. mV/cm vertical gain. If no output, see items 4 and 5.
generator.
Amplifier faulty or Check amplifier settings and connections. Consult the ampli-
maladjusted; faulty fier manual. Replace the amplifier, if necessary. Replace
cables. cable set or check for continuity, including preamp power
cables.
4. No preamplifier output at Excessive detector Decrease detector bias in steps of 100 V until output pulses
recommended bias leakage current bias- are obtained. If the detector does not perform properly at a
voltages. ing off the preamplifier lower bias and the following results are observed, contact
input stage. PGT:
RG-11 or RG-15: Test point normal with zero HV, but
+24 V at higher HV
6-6
PGT Detector Systems Troubleshooting
Table 6-1: Troubleshooting Guide for 352, RG-11 and RG-15 Preamplifier Detectors
Other checks:
IGC TP voltage: +23.3 V = blown FET
-23.3 V = high LC
-23.3 V @ 0 V bias = RF shorted
0 V @ all bias = RF shorted,
TP not connected, or
FB not connected, or
no PA power
6. Poor gamma ray resolu- Microphonics Place the system dewar on a foam pad if vibrations through
tion. High noise (pulser res- the floor are a problem. Arrange other suitable shock mount-
olution). 100-10000 Hz base ing. Try using a shorter time constant. If a shorter time con-
oscillations can be easily stant is used, do not forget to readjust the P/Z.
induced or made much
worse by nearby noise or by
tapping the cryostat lightly.
Such oscillations persist in
response to normal room
noise and vibrations. Oscil-
lations may disappear if HV
or MCA is disconnected.
7. Poor gamma ray resolu- Faulty voltage bias Turn HV to zero (0), wait 5 mins., then thoroughly clean (with
tion and high noise; spark- supply or cable. ethanol) the HV parts in the preamp. Resolder HV joints if
ing/highvoltage noise; Humidity on HV line in any sharp points are evident. Avoid humid atmosphere.
negative pulses; test point or preamp. Breakdown of Check HV supply unit and cable. Call PGT if problem per-
reset rate unstable. filter capacitor or HV sists.
feedthrough.
8. Poor gamma ray resolu- Power line noise, RF Eliminate AC power line noise by isolation or filtration. Check
tion and high noise; 20 kHz pickup, ground loop ground connections in preamplifier; avoid grounding cryo-
oscillation at main amplifier stat/dewar; loop HV cable through ferrite ring; ground MCA
output which may disappear to NIM chassis.
if HV or MCA is discon-
nected.
9. Poor gamma ray resolu- Breakdown across the Be sure that the bias voltage is not more than specified.
tion and high noise. A very surface of the detector Reduce detector bias in steps of 100 V until base line
ragged base line with one or or an insulator returns to normal. If the system meets all performance
more of the following types requirements, it may be usable at the lower bias. If not, con-
of anomalous pulses: tact PGT. In addition, lower shaping time may be used.
positive spikes,
negative spikes,
square-shaped positive
and negative pulses
6-7
Troubleshooting PGT Detector Systems
Table 6-1: Troubleshooting Guide for 352, RG-11 and RG-15 Preamplifier Detectors
11. Poor gamma ray resolu- Neutron (or similar If the detector is warmed up, the resolution will get much
tion at 1.33 MeV or higher radiation) damage worse. DO NOT WARM UP THE DETECTOR. Neutron dam-
energy. Normal noise as age can be repaired. Call PGT.
measured by pulse genera-
tor resolution. Resolution at
122 keV or lower energy is
much less degraded than at
higher energy.
12. Low- or high-energy tail- Incorrect P/Z cancella- See appropriate section on detector installation and optimi-
ing and poor resolution. tion zation.
13. Wandering peaks or Unstable electronics Check each electronic component for proper operation.
multiple peaks observed. (especially main ampli- Repair or replace faulty units. Check all cables and connec-
fier) tors (including panel connectors) for intermittent signal or
ground connections.
14. High loss rate of liquid Degradation of cryo- Measure LN2 loss rate by weighing (0.807 kg/liter). For an
nitrogen. Excessively cold stat vacuum accurate measurement, the dewar must not be moved dur-
cryostat with moisture con- ing the 2 hrs. before the initial measurement until after the
densation; increase in leak- final measurement. The normal LN2 loss rate depends on
age current. cryostat/dewar configuration. For most systems, a loss rate
of >2 liters/day is excessive. For a dipstick system, the dewar
loss rate may be subtracted from the system loss rate (which
should not be >1.5 liters/day).
15. High loss rate of liquid Degradation of dewar Check system loss rate as in item #14. If cryostat is a dip-
nitrogen. Excessively cold vacuum stick model, place it in another dewar and measure the
dewar with moisture conden- dewar loss rate alone. Place a stopper in the white Teflon
sation. Cryostat temperature collar and measure the loss by weighing (as in item #14). A
normal. 32-liter dewar with a collar assembly should have a loss rate
<0.7 liter/day.
Note: Instructions for replacing resistive feedback preamplifiers are given in Secs. 5.2.2. and 5.2.3.
6-8
PGT Detector Systems Troubleshooting
Table 6-2: Troubleshooting Guide for Pulsed Optical and Transistor Reset Preamplifier
Detectors
MCA controls improp- If pulses are present at the input, see the MCA manual and
erly set. Possible check the control settings. Be sure that the amplifier output
improper amplifier set- is positive. You may be misled if the oscilloscope is acciden-
ting. tally set to “invert.”
2. MCA registers Pulses corresponding Use an oscilloscope to examine the signal which is going to
background counts, but no to spectral peaks have the MCA. Adjust the amplifier gain to achieve pulse heights
peaks are visible. voltage heights out- within the input range of the MCA. (Make sure that you know
side the range of the this range, which may change with MCA settings, e.g., digital
pulse height analyzer. offset or conversion gain.)
3. No amplifier output for Amplifier not receiving Check preamplifier output with the oscilloscope set on 2V/
gamma ray source or pulse pulses. cm vertical gain. Observe reset at 50 mV/cm. Observe steps
generator. on ramp. If no output, see items 4 and 5.
Amplifier faulty or Check amplifier settings and connections. Consult the ampli-
maladjusted; faulty fier manual. Replace the amplifier, if necessary. Replace
cables. cable set or check for continuity, including preamp power
cables.
4. No preamplifier output at Excessive detector Decrease detector bias in steps of 100 V until output pulses
recommended bias leakage current bias- are obtained. If the detector does not perform properly at a
voltages. ing off the preamplifier lower bias and the following results are observed, contact
No reset output. input stage. PGT:
(dc level positive for N-type PO-12 or PO-14: Reset normal at 50-100 V HV, but
and planar detectors; extremely fast if HV is increased and the DC level
negative for P-types) reaches +12 V. Test pulse gives signal at zero HV.
6-9
Troubleshooting PGT Detector Systems
Table 6-2: Troubleshooting Guide for Pulsed Optical and Transistor Reset Preamplifier
Detectors
6. Poor gamma ray resolu- Microphonics Place the system dewar on a foam pad if vibrations through
tion. High noise (pulser res- the floor are a problem. Arrange other suitable shock mount-
olution). 100-10000 Hz base ing. Try using a shorter time constant. Try the high restorer
oscillations observed on the mode, if available.
amplifier can be easily
induced or made much
worse by nearby noise or by
tapping the cryostat lightly.
Such oscillations persist in
response to normal room
noise and vibrations. Oscil-
lations may disappear if HV
or MCA is disconnected.
7. Poor gamma ray resolu- Faulty voltage bias Turn HV to zero (0), wait 5 mins., then thoroughly clean (with
tion and high noise; spark- supply or cable. ethanol) the HV parts in the preamp. Resolder HV joints if
ing/highvoltage noise; Humidity on HV line in any sharp points are evident. Avoid humid atmosphere.
negative pulses; test point or preamp. Breakdown of Check HV supply unit and cable. Call PGT if problem per-
reset rate unstable. filter capacitor or HV sists.
feedthrough.
8. Poor gamma ray resolu- Power line noise, RF Eliminate AC power line noise by isolation or filtration. Check
tion and high noise; 20 kHz pickup, ground loop ground connections in preamplifier; avoid grounding cryo-
oscillation at main amplifier stat/dewar; loop HV cable through ferrite ring; ground MCA
output which may disappear to NIM chassis.
if HV or MCA is discon-
nected.
9. Poor gamma ray resolu- Breakdown across the Be sure that the bias voltage is not more than specified.
tion and high noise. A very surface of the detector Reduce detector bias in steps of 100 V until base line
ragged base line with one or or an insulator returns to normal. If the system meets all performance
more of the following types requirements, it may be usable at the lower bias. If not, con-
of anomalous pulses tact PGT. In addition, lower shaping time may be used.
observed on the amplifier:
positive spikes,
negative spikes,
square-shaped positive
and negative pulses
6-10
PGT Detector Systems Troubleshooting
Table 6-2: Troubleshooting Guide for Pulsed Optical and Transistor Reset Preamplifier
Detectors
11. Poor gamma ray resolu- Neutron (or similar If the detector is warmed up, the resolution will get much
tion at 1.33 MeV or higher radiation) damage worse. DO NOT WARM UP THE DETECTOR. Neutron dam-
energy. Normal noise as age can be repaired. Call PGT.
measured by pulse genera-
tor resolution. Resolution at
122 keV or lower energy is
much less degraded than at
higher energy.
12. Low- or high-energy tail- Incorrect P/Z cancella- See appropriate section on detector installation and optimi-
ing and poor resolution. tion zation.
13. Wandering peaks or Unstable electronics Check each electronic component for proper operation.
multiple peaks observed. (especially switches on Repair or replace faulty units. Check all cables and connec-
the main amplifier) tors (including panel connectors) for intermittent signal or
ground connections.
14. High loss rate of liquid Degradation of cryo- Measure LN2 loss rate by weighing (0.807 kg/liter). For an
nitrogen. Excessively cold stat vacuum accurate measurement, the dewar must not be moved from
cryostat with moisture con- 2 hrs. before the initial measurement until after the final mea-
densation; increase in leak- surement. The normal LN2 loss rate depends on cryostat/
age current. dewar configuration. For most systems, a loss rate of
>2 liters/day is excessive. For a dipstick system, the dewar
loss rate may be subtracted from the system loss rate (which
should not be >1.5 liters/day).
15. High loss rate of liquid Degradation of dewar Check system loss rate as in item #14. If cryostat is a dip-
nitrogen. Excessively cold vacuum stick model, place it in another dewar and measure the
dewar with moisture conden- dewar loss rate alone. Place a stopper in the white Teflon
sation. Cryostat temperature collar and measure the loss by weighing (as in item #14). A
normal. 32-liter dewar with a collar assembly should have a loss rate
<0.7 liter/day.
6-11
Troubleshooting PGT Detector Systems
6-12
Section 7
Warranties
Return Instructions 7
PGT Detector Systems Warranty
7.1: HPGe
Detector
Warranty
Princeton Gamma-Tech, Inc. guarantees that its HPGe (Intrinsic Ger-
manium) Detectors are free from all defects in materials and workman-
ship and they will not fail in service for a period of one year after
delivery, provided they are properly cared for.
7-3
Warranty PGT Detector Systems
7-4
PGT Detector Systems Warranty
7.3: Detector
Repair
Warranty
Princeton Gamma-Tech guarantees that the specific repairs done on
the HPGe detectors will not fail in service for 90 days, provided the de-
tectors are stored properly, are kept at all times at liquid nitrogen tem-
peratures, with bias applied, and are not subjected to doses of fast neutrons
or other damaging radiation. It is the customers responsibility to monitor the
LN 2 level continuously.
When the detector is first received, please note carefully the packing
method and save the packing materials for possible future use. The
shipping containers are specifically designed to protect the crystal from
damage due to rough handling.
7-5
Warranty PGT Detector Systems
7.4: Packing
Instructions
If for any reason it becomes necessary to return a photon detector sys-
tem for repair or replacement, please note the following instructions
and precautions regarding shipment.
The instructions and precautions stated here are to ensure proper at-
tention in handling shipment of an expensive radiation detector. Do not
be apprehensive because of these precautions; detectors are success-
fully shipped all over the world when proper care is exercised.
7-6
PGT Detector Systems Warranty
PGT, Inc.
7-7
Warranty PGT Detector Systems
7-8
PGT Detector Systems Index
Index
Numerics resolution
352 preamplifier, see preamplifiers count rate effects 5-11
resistive feedback selection criteria 1-3
temperature cyclability 4-13
A types 4-3
annealable detectors 4-14
E
B efficiency 4-6, 4-12
beryllium window energy resolution 4-9
precautions 2-14
rupture 2-14 F
Full Width at Half Maximum, see FWHM
C FWHM
circuit diagrams 5-14–5-31 determination 3-16
Compton scattering 4-4
counting efficiency 4-7 I
internal pressure build-up 2-14
D intrinsic germanium 4-3
detector
cool down 3-3, 3-9 L
incomplete cool-down 2-13 liquid nitrogen
partial warm up 2-13 filling bucket dewars 2-10
detector calibration filling dipstick dewars 2-8
manual 3-15 filling portable cryostats 2-11
detector specifications gravity-fed portables 2-12
measurement 3-14 safety precautions 2-7
resolution 3-17 liquid nitrogen monitor 6-3
detectors low background systems 1-5
annealing 4-14
crystal geometries 4-3 M
performance parameters main amplifier
efficiency 4-6 pole zero cancellation 3-6
energy resolution 4-9 reset pulses 3-12
noise contributions 4-10 shaping time 5-6
peak-to-background ratio 4-10 Marinelli reentrant beaker
peak-to-Compton ratio 4-10 efficiency 4-13
timing resolution 4-11 maximum energy rate 5-5
photon interactions 4-4 MCA, see multichannel analyzer
i
Index PGT Detector Systems
R
replacement 5-20
resolution
count rate 5-11
RG-11 circuit diagram 5-14
component layout 5-15
RG-11 preamplifier, see preamplifiers
resistive feedback
ii