XT Errorcode
XT Errorcode
Error codes
The following list of error codes and corresponding information was edited to remove many that
are not relevant to XT/LT instruments, however some of these codes may not apply to XT/LT.
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,2/55
1.
Shut down the
software.
2. Restart the run.
1 11 Error encountered HASP Hardware Unable to 1. Reinstall the HASP
reading hardware Key Errors retrieve the drivers.
key. database version
2. If that does not clear it
from the
up, replace the HASP
database version
key.
field in the HASP
Key.
1 12 Error encountered HASP Hardware Unable to 1. Reseat the HASP key.
reading hardware Key Errors retrieve the
2. If that does not work,
key. database version
reinstall the HASP
from the
drivers and try again.
database version
field in the HASP 3. If that does not work, a
Key. bad HASP key could be
the problem or a bad
parallel port. Replace as
needed.
2 1 Bad Touch Product Unable to 1. Verify that the wand
Memory Button Registration retrieve data works with another
from the touch button.
memory button.
2. If the wand works
properly, replace the
button. If not, replace
the wand.
2 2 Data Retrieval Product Error when 1. Verify that the wand
Error Registration retrieving data works with another
from the touch button.
memory button.
2. If the wand works
properly, replace the
button. If not, replace
the wand.
2 3 Seal has already Product The touch The button was already
been burned by Registration memory button burned to a different host.
another NexES was registered User error. If the user did
Host %d. (%d by another Host not burn the button, send
being a Host ID system. them a new button.
value)
2 4 Seal was burned Product The touch The buttons give you a 10-
by this NexES Registration memory button day window to re-register if
system too long was registered the touch memory button
ago. for more than 10 needs to be re-registered.
days and the Send the user a new
user is trying to button, if prudent.
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,3/55
reregister it.
2 4 Seal has already The button on Buttoned packages cannot
been burned by the package was be registered more than
this NexES system. already one time. The customer
registered on should purchase more
this host product from Ventana.
computer.
2 4 The touch memory The operator has Ensure that the operator
has been changed switched keeps the probe on the
after reading its buttons during button until the current
contents. the process. operation is completed.
(Each button has
a unique
number, and the
software
detected that
this changed.)
2 5 The touch memory Product The touch Send a new button.
has been changed Registration memory button
after reading its was changed
contents. after the
contents were
read, but not
burned.
3 1 Communication Communication The remote code 1. Try the download again.
failure attempting with the form was closed
2. If the download fails,
to upgrade Instrument before the
contact Ventana.
software, Retry. download was
completed.
3 3 The new staining Communication The remote code 1. Try the download again.
instrument with the checksum does
2. If the download still
software was not Instrument not match the
fails, contact Ventana.
loaded correctly, checksum
Retry. returned from
the staining
instrument,
preventing
switching to the
current version
of the remote
software.
3 4 Instrument Not Communication The run is 1. Check the
Communicating, with the terminated communication
Run Terminated. Instrument because the host between the host and
cannot the instrument by
communicate running test functions.
with the
2. If there is no
instrument
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,4/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,5/55
represents port Instrument port could not Reconfigure and test the
number). Error be opened for connectivity.
number %d (%d communications
represents error) to the remote
device.
3 13 Failed to setup Communication Unable to Check the communications
buffer sizes. Error with the initialize the path.
number %d (%d Instrument communications
represents error) parameters for a
communications
device.
3 14 Failed to Set Com Communication Unable to Check the communications
States. Error with the configure a path.
number %d (%d Instrument communications
represents error) device according
to the
specifications in
a device control
block.
3 15 Failed to Set Com Communication Unable to set Check the communications
Timeouts. Error with the the time-out path.
number %d (%d Instrument parameters for
represents error) all read-and-
write operations
on a specified
communications
device.
3 16 Failed to Set RTS. Communication Unable to set Check the communications
Error number %d with the the Request to path.
(%d represents Instrument Send signal.
error)
3 17 Failed to Set DTR. Communication Unable to set Check the communications
Error number %d with the the Data path.
(%d represents Instrument Terminal Ready
error) signal.
3 18 Message timed Communication The command Check the communications
out. Address: %0:d with the on the path. This issue can cause a
Host Command: % Instrument instrument did power loss on the
1:s ( %2:d ) not return within instrument while the host
milliseconds: % the set time software is running.
3:d' limit.
3 19 Unknown Tagged Communication Tag values sent Code error. Contact
Value. Identifier: with the from the Host PC Ventana.
%0:d [ %1:d ] Instrument are not defined
on the
Instrument
Remote Code.
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,6/55
4 2 %0:s is not a valid Serial Numbers An invalid serial Reset the serial number.
Staining number for the
Instrument serial staining
number. (%0:s instrument is
represents version detected.
number) Call
Ventana technical
support before
starting another
run on this
Staining
Instrument
5 2 Slide number is Software The slide count 1. Verify the slide count.
invalid, Please Re- Application that was entered
2. Clean off the barcodes if
enter did not reflect
needed.
the number of
slides read on 3. Try again.
the instrument.
5 3 One or more of Software You did not Complete the actions in the
the checklist Application check off the list, then check the boxes.
questions have required boxes
not been checked. on the pre-run
checklist.
5 6 The Starting Date Software An invalid start Enter a valid start date.
is invalid! Application date is entered
for the Product
Usage Report.
5 7 The Ending Date is Software An invalid Enter a valid ending date.
invalid! Application ending date was
entered for the
Product Usage
Report.
5 8 The Ending Date Software The ending date Reenter the dates.
precedes Starting Application entered for the
Date! Product Usage
Report is earlier
than the start
date.
5 10 At least one Software No instruments Make sure that at least one
Staining Application are connected to instrument is connected to
Instrument must the host. the host.
be connected.
5 17 You must select at Software No product was Select a product.
least one product! Application selected on the
Product Usage
form.
5 18 Product usage is Software No records are Enter a valid date range.
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,7/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,8/55
5 81 The following Software The bulk fluid Fill the listed bottles or click
Automated Application bottles are not Yes to proceed.
Fluidics full when you
Instrument bottles start a test
are not full: [list]. function.
Proceed Anyway?
5 82 Case: %s has Software The case that Enter a unique case.
already been Application you are logging
registered for %s was already
(%s represents registered.
Case Number)
Waste level is too
high.
5 86 Invalid Date Software The date range is Enter a valid date range.
Range! Dates must Application invalid.
be between
January 1, 1997
and the current
date. The start
date must not be
after the end date.
5 88 Waste level is too Software When a test Empty the waste container.
high. Please empty Application function is
it. started, the
waste bottle is
full.
5 102 The Archive Software The archive 1. Make sure that no
Procedure has Application process failed to staining runs are in
failed. complete. progress.
2. Repeat the archive
process.
3. Contact Ventana if the
problem persists.
5 108 The Response Software While 1. Get a valid
code you entered Application attempting to Authorization code.
was invalid. change the
2. Try again.
staining
instrument
personality, an
unauthorized
remote response
message NAK is
received from
the remote
software. The
Authorization
Code was
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,9/55
invalid.
5 108 Staining module is Software While 1. Put the instrument into
in an invalid state. Application attempting to Sleep mode and retry.
change the
2. Reboot the Host PC and
staining
the instrument.
instrument
personality, an 3. If the problem persists,
unauthorized contact Ventana.
remote response
message NAK is
received from
the remote
software.
5 111 The Archive Software The user Run the archive process
Procedure Application cancelled the again.
canceled. archive process.
5 117 The Starting Date Software An invalid start Enter a valid start date.
is invalid! Application date is entered
for the Bulk
Usage Report.
5 118 The Ending Date is Software An invalid Enter a valid ending date.
invalid! Application ending date is
entered for the
Bulk Usage
Report.
5 119 The Ending Date Software The ending date Enter valid dates.
precedes Starting Application specified for the
Date! Bulk Usage
Report is earlier
than the starting
date.
6 1 Rule operand: Run This should not This error code could
XXXXXX only valid Compilation normally occur. indicate database
in a recipe step. corruption or an internal
software error.
6 2 Rule operand Run This error should This error message could
argument: XXXXX Compilation not normally indicate database
missing from occur. corruption or an internal
Argument software error.
database.
6 3 Rule operand: Run This error should This error message could
XXXX missing from Compilation not normally indicate database
Argument occur. corruption or an internal
database. software error.
6 4 Rule operator: %s Run This error should This error message could
missing from Compilation not normally indicate database
Argument occur. corruption or an internal
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,10/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,11/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,12/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,13/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,14/55
2. Save again.
6 50 Unbalanced Run Incompatible Check the protocols and
manual Compilation timing with the slides to make them
application or Titration slides compatible.
titration protocols. and Non-
Titration slides.
6 51 Macro step %0:s Run The macro step This message could indicate
has an invalid Compilation function is database corruption or an
macro function outside a valid internal software error.
number: %1:d (% range (0 to 162). 1. Reinstall VSS.
0:s represents the
2. Try again.
macro step name
and %1:d
represents the
macro step
function number)
6 52 Macro step %s has Run The MacrFunc This error message could
no match in the Compilation database table indicate database
macro function does not contain corruption or an internal
database. (%s the a record that software error.
represents the matches the 1. Reinstall VSS.
macro step name) macro step.
2. Try again.
6 55 Missing argument Run When compiling This error message could
for Function %s. Compilation macros for a indicate database
(%s represents the staining run, the corruption or an internal
macro function) macro function software error.
does not have an 1. Reinstall VSS.
associated
2. Try again.
argument in the
Argument
database table.
6 56 Macro %s is Run Compiling sub This error message could
missing in the Compilation macros for a indicate database
Macros file. (%s staining run and corruption or an internal
represents the sub the sub macro is software error.
macro name) not included in 1. Reinstall VSS.
the Macros
2. Try again.
database table.
6 58 Maximum staining Run The run contains A limited number of steps
run size exceeded. Compilation too many steps can be downloaded to an
for the staining instrument in a run. Report
module to the message to Ventana,
perform. including the protocols in
the run.
6 59 Function : %0:s Run An invalid This error message could
contains function Compilation function is indicate database
# %1:d, which is identified in the corruption or an internal
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,15/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,16/55
reagent) remaining to
stain the slides
that are loaded
on the
instrument.
6 68 Reagent %s Run A reagent 1. Put the missing reagent
missing from tray. Compilation needed for the on the carousel.
(%s represents the staining run is
2. Restart the run.
reagent) not on the
reagent
carousel.
6 69 Dispensers %0:s, # Run Reagents 1. Ensure that the
%1:d and %2:s, # Compilation required to be in dispensers noted in the
%3:d must be a specific order error message are
beside each other. are not loaded beside each other.
(%0:s represents on the reagent
2. Restart the run.
the reagent name, carousel in the
%1:d represents proper order.
the reagent serial
number, %2:s
represents the
reagent name, and
%3:d represents
the reagent serial
number)
6 70 Errors in recipe Run Error creating This error message could
constructor rule Compilation rule block. indicate database
block : %s corruption or an internal
software error. Contact
Ventana.
6 71 Errors in recipe Run An error is This error message could
sync rule block : % Compilation indicated while indicate database
s creating recipe corruption or an internal
synchronization software error. Contact
rules. Ventana.
6 72 Macro 0 Run The instrument 1. Put the instrument in
generating object Compilation is not in Ready Ready mode.
not ready to start mode.
2. Try again.
now.
6 74 Blank instrument Run The macro 0 is 1. Ensure that the
name Compilation being compiled instrument name is
and the staining assigned and is not
instrument does blank.
not have a
2. Restart the run.
defined
instrument
name.
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,17/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,18/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,19/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,20/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,21/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,22/55
blocks when no
preregistered
control tissue is
available.
7 17 There are no Database Attempting to Register new dispensers.
Active dispensers refill a
to fill dispensers when
no active
dispensers are
available.
7 18 There are no Database Attempting to Register antibodies.
preregistered log antibodies
antibodies and no pre-
registered
antibodies are
available.
7 19 There are no runs Database An attempt is Enter the control results
to QC at this time made to enter when there are executed
control results, staining runs.
and there are no
executed
staining runs.
7 21 Backup was Database User cancels the Run the backup again.
cancelled prior to backup before it
completion. is completed.
7 24 Incorrect Database The Select the correct
Frequency in User maintenance frequency.
Maintenance frequency is not
Notes Daily, Monthly,
or Quarterly.
7 27 Inventory Database There are no Register the product into
Database is Empty product records inventory.
in the inventory
database and an
attempt is made
to generate an
Inventory report.
7 48 Database Integrity Database An error is Try to run the database
Error: %s (%s encountered integrity check again.
represents data while Different actions are
integrity error performing a required, depending on
message) database which tables get errors.
integrity check. Send a screenshot to GTS
and the error log for
further analysis.
7 58 There is no system Database There is no There is no system data to
data to archive staining run data archive.
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,23/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,24/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,25/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,26/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,27/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,28/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,29/55
opened during
the run” was not
signed off by the
operator within
the required 5-
minute interval.
12 10 Power failed Remote The power to 1. Ensure that the power
during run Software the instrument is requirements are met
cut off during a from the lab's
staining run. standpoint.
2. Check the power supply
on the instrument.
3. Ensure that the power
switch is on.
12 11 Slide heat Remote A slide heater on This error could be caused
exceeded the Software the instrument by a lack of LCS on the slide
absolute exceeds the at incubation or a faulty
maximum maximum slide heater.
temperature. allowable
temperature.
12 12 Slide heat Remote A slide heater The sensor could be
temperature Software sensor on the unplugged from the banana
sensor is bad or instrument board, or a faulty heater.
not connected. cannot be
detected.
12 15 Pressure exceeded Remote Air pressure has
the absolute Software increased
maximum limit beyond the
upper range of
the sensor (14.2
psi). Check:
regulator, valves
(VRMIX and
BCBLO),
pressure sensor
12 16 Nozzle plate failed Remote The home sensor Faulty home sensor which
to find home Software fails to detect probably needs to be
position during the replaced.
homing of the
nozzle plate.
12 16 Slide tray failed to
find home
position.
12 17 Reagent tray failed Remote The home sensor Faulty home sensor which
to find home Software fails to detect probably needs to be
position during the replaced.
homing of the
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,30/55
reagent tray
carousel.
12 17 Reagent tray failed Seek for home
to find home position failed.
position or tray is Possible
missing. problems with
motor, motor
controller, or
home sensor.
Also check belts
and bearing
pads.
12 18 Slide heat is too Slide
low. temperature
remained below
normal (-3° C)
for over 5
minutes. Check
the slide heater
sensor, and fan.
12 19 Fluid in tub Remote The excess fluid Drain could be clogged, or
exceeded Software level exceeds the the sensor is faulty.
maximum level sensor level
located on the
internal tub.
12 21 Depressurizing
time exceeded.
12 23 User aborted run Remote The User ends a Could be that the user
before normal Software staining run aborted the run manually,
completion prior to its or a VSS failed to detect the
natural HASP key during the start
completion. of a run while other slides
were in progress
12 24 Nozzle plate Remote The nozzle plate 1. Restart the runs from
position incorrect Software becomes Sleep mode.
during the run misaligned.
2. If the problem persists,
reboot the instrument.
3. Try again.
12 24 Slide tray position
incorrect during
the run
12 25 Reagent tray Remote The reagent 1. Restart the runs from
position incorrect Software carousel became Sleep mode.
during the run misaligned.
2. If the problem persists,
reboot the instrument.
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,31/55
3. Try again.
12 36 LCS reservoir
failed to prime.
12 40 LCS reservoir
failed to prime.
12 41 Optional reservoir
failed to prime.
12 42 Vacuum trap is full
of fluid
12 44 LCS reservoir
failed to recharge
itself in time.
12 46 Optional reservoir
failed to recharge
itself in time.
12 48 LCS reservoir
failed to recharge
itself in time.
12 49 Optional reservoir
failed to recharge
itself in time.
12 50 Waste container is
full.
12 51 Slide heater error.
12 53 Nozzle plate check Remote The nozzle plate Contact Ventana to align
at end of run Software is slightly out of and check hardware.
indicated minor alignment.
misalignment
12 53 Slide tray check at
end of run
indicated minor
misalignment
12 54 Nozzle plate check Remote The nozzle plate Contact Ventana to align
indicated major Software is seriously out and check hardware.
misalignment of alignment.
12 54 Slide tray check at
end of run
indicated major
misalignment
12 55 Reagent tray Remote The reagent tray Contact Ventana to align
check at end of Software is slightly out of and check hardware.
run indicated alignment.
minor
misalignment.
12 55 Reagent tray
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,32/55
check indicated
major
misalignment
12 56 Reagent tray Remote The reagent tray Contact Ventana to align
check at end of Software is seriously out and check hardware.
run indicated of alignment.
major
misalignment.
12 56 Reagent tray
check indicated
major
misalignment
12 59 Priming operation
completed
12 60 Cleaning operation Remote A cleaning Go to the next step in the
completed. Software function has cleaning operation or
completed. resume using the
instrument.
12 61 Pressurizing time The standard 13
was exceeded psi air pressure
could not be
achieved at
startup in under
60 seconds.
Check
compressor,
relay, leakage
from fluid
bottles, etc.
12 63 Run stopped due Remote The slide drawer Replace the heater.
to heater(s) Software heaters are too
abnormally hot. If hot for slide
this situation processing.
persists, call for
service.
12 65 High pressure Remote The high Dispatch an FSE for service.
exceeded Software pressure
exceeds the
maximum
allowable limit
for a specified
amount of time.
12 99 AFM left board is Remote The AFM control Could be a bad board or a
not responding Software board is not connection came loose.
normally to responding to Contact Ventana.
commands. If this commands.
situation persists,
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,33/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,34/55
Please notify
Ventana.
12 116 Inactive slide Remote A slide requests Transfer the dataset for
attempted to start Software a submacro step analysis.
a sub macro. This while the nozzle
typically indicates plate is at
a software error stations 1, 7, 8,
or 9, or an
inactive slide
requests the
macro step.
12 117 Inactive slide Remote An inactive slide Transfer the dataset for
attempted move Software or a slide that is analysis.
the reagent tray. not under the
This typically nozzle plate
indicates a station 6
software error. requests a
reagent carousel
move.
12 118 Slide heater error.
If this situation
persists, call for
service.
12 122 Slide barcode is Remote The instrument 1. Make sure that the slide
unreadable, check Software cannot read the barcode is readable.
slide and retry. barcode on the
2. Restart the slide.
indicated slide.
12 125 High system air
pressure was too
low for an
extended period
of time.
12 126 Low system air
pressure was too
low for an
extended period
of time.
12 130 Slide #21 Remote The slide The error message could be
temperature Software position cannot as a result of a bad slide
control error (too achieve the heater. Call the FSE to run
low) setpoint tests.
temperature in
the timeout
duration.
12 131 Slide #22 Remote The slide The error message could be
temperature Software position cannot as a result of a bad slide
control error (too achieve the heater. Call the FSE to run
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,35/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,36/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,37/55
maximum
allowable.
12 146 Slide #27 Remote The slide The error message could be
temperature Software position as a result of a bad slide
control error (too temperature heater. Call the FSE to run
high) exceeds the tests.
maximum
allowable.
12 147 Slide #28 Remote The slide The error message could be
temperature Software position as a result of a bad slide
control error (too temperature heater. Call the FSE to run
high) exceeds the tests.
maximum
allowable.
12 148 Slide #29 Remote The slide The error message could be
temperature Software position as a result of a bad slide
control error (too temperature heater. Call the FSE to run
high) exceeds the tests.
maximum
allowable.
12 149 Slide #30 Remote The slide Send service rep to
temperature Software position investigate.
control error (too temperature is
high) out-of-range or
disconnected.
12 150 Slide #21
temperature
sensor bad or
disconnected
12 151 Slide #22 Remote The slide Send a service rep to
temperature Software position investigate.
sensor bad or temperature is
disconnected out-of-range or
disconnected.
12 152 Slide #23 Remote The slide Send a service rep to
temperature Software position investigate.
sensor bad or temperature is
disconnected out-of-range or
disconnected.
12 153 Slide #24 Remote The slide Send a service rep to
temperature Software position investigate.
sensor bad or temperature is
disconnected out-of-range or
disconnected.
12 154 Slide #25 Remote The slide Send a service rep to
temperature Software position investigate.
sensor bad or temperature is
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,38/55
disconnected out-of-range or
disconnected.
12 155 Slide #26 Remote The slide Send a service rep to
temperature Software position investigate.
sensor bad or temperature is
disconnected out-of-range or
disconnected.
12 156 Slide #27 Remote The slide Send a service rep to
temperature Software position investigate.
sensor bad or temperature is
disconnected out-of-range or
disconnected.
12 157 Slide #28 Remote The slide Send a service rep to
temperature Software position investigate.
sensor bad or temperature is
disconnected out-of-range or
disconnected.
12 158 Slide #29 Remote The slide Send a service rep to
temperature Software position investigate.
sensor bad or temperature is
disconnected out-of-range or
disconnected.
12 159 Slide #30 Remote The slide Send a service rep to
temperature Software position investigate.
sensor bad or temperature is
disconnected out-of-range or
disconnected.
12 160 Slide #1 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 161 Slide #2 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 162 Slide #3 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,39/55
duration.
12 163 Slide #4 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 164 Slide #5 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 165 Slide #6 Remote The slide Send a service rep to
temperature Software positioncannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 166 Slide #7 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 167 Slide #8 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 168 Slide #9 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 169 Slide #10 Remote The slide Send a service rep to
temperature Software positioncannot investigate.
control error (too achieve the
low). setpoint
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,40/55
temperature in
the timeout
duration.
12 170 Slide #11 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 171 Slide #12 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 172 Slide #13 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 173 Slide #14 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 174 Slide #15 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 175 Slide #16 Remote The slide Send a service rep to
temperature Software position cannot investigate.
control error (too achieve the
low). setpoint
temperature in
the timeout
duration.
12 176 Slide #17 Remote The slide Send a service rep to
temperature Software position cannot investigate.
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,41/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,42/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,43/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,44/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,45/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,46/55
disconnected.
12 218 Slide #19 Remote The slide Send a service rep to
temperature Software position investigate.
sensor bad or temperature is
disconnected. out-of-range or
disconnected.
12 219 Slide #20 Remote The slide Send a service rep to
temperature Software position investigate.
sensor bad or temperature is
disconnected. out-of-range or
disconnected.
12 220 Waste container
was full during
run.
12 225 Initializing
12 226 Draining Ventana
Fluids
12 227 Waiting for
Decontamination
Fluid
12 229 Incubating with
Decontamination
Fluid
12 230 Draining
Decontamination
Fluid
12 231 Waiting for DI
Water
12 233 Draining DI Water
12 234 Waiting for
Ventana Fluids
12 236 Priming lines with
Ventana Fluids
12 237 Disconnect tubing
manifold and
insert filled Bulk
bottles into
instrument.
12 241 Cleaning with
Disinfectant
12 243 Press button on
staining module to
continue the run.
12 245 Priming, please
wait
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,47/55
12 247 Press Ventana Bug Remote A function test Press the button to resume
button on staining Software has stopped and the test.
module to must be
continue test continued.
function
12 251 Press stainer Remote The system has 1. Complete the
button to open, Software paused for a application.
perform manual manual
2. Resume the staining
application, press application.
run.
and hold button
when complete
13 2 No slides found,
run terminated
13 3 Barcode read
Aborted with Error
Code %d, Please
Retry.
13 4 Barcode read Instrument Tray The instrument Restart the run.
timed out, Please Scan attempted to
Retry. read barcodes
during the run
process but the
barcode
information is
formatted
incorrectly.
13 5 Barcode read
Aborted with Bad
Format error,
Please Retry.
13 6 Barcode Reader Instrument Tray Misread of the Restart the run.
Did Not Initialize, Scan slide label.
Please Retry.
13 7 Slide Barcode Instrument Tray Misread of the Restart the run.
Check Digit Error, Scan slide label.
Restart Run.
13 8 Invalid Barcode at Instrument Tray Trying to read Restart the run.
position %d. Scan reagent
barcodes during
the run process
and the reagent
barcode cannot
be decrypted.
13 9 (%d represents the Instrument Tray Misread of slide Restart the run.
reagent position Scan label.
on the reagent
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,48/55
carousel).
13 10 %0:s with Serial
#%1:d at Position
%2:d is a duplicate
product, please
remove from tray.
13 11 Barcode Trigger Instrument Tray Misread of Restart the run.
Failed, Please Scan Reagent label.
Retry.
13 12 Barcode read Error Instrument Tray Misread of Restart the run.
Code %d, Scan Reagent label.
Retrying… Barcode
read Timed Out,
Retrying…
13 13 Barcode read Bad Instrument Tray Misread of Restart the run.
Format, Retrying… Scan Reagent label.
13 14 Unknown Bar
Code Reader Type,
run terminated
13 15 Keycode Bar Code Instrument Tray The keycode Check the connectivity
on Slide at Scan embedded in the between the other hosts in
Position %d could slide barcode the replicated environment
not be Resolved. could not be for the ability to replicate
read correctly case slides.
during the
staining run
process.
13 17 Unknown error
text, but the error
is defined
13 18 (%d represents Instrument Tray Case slide 1. Wait until the other
slide position Scan ownership could NexES Host system is
Could not acquire not be acquired finished with the case
Ownership of all from another slide.
Case Slide(s). Host system
2. Force the ownership
during the
acquisition with an
staining run
access code from
process.
Ventana.
13 21 [#] Slide Run Informational Information Information Only.
Resumed after Message Only.
being paused for 3
Mins 32 Secs
13 23 [#] Reagent Access Informational Information Information Only.
occurred during Message Only.
run.
14 3 Slide at position Staining Run The selected Replace the heater.
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,49/55
#XXXX is on a protocol to be
faulty heater. used during the
run process was
not created for
the instrument
type.
14 4 Protocol #%0:s at Staining Run Communication 1. Create a protocol.
slide position #% with an
2. Restart the staining run.
1:d is not defined instrument fails
for this instrument during a staining
type. (%0:s run when
represents the information is
protocol number being
and %1:d downloaded.
represents slide
position on the
carousel)).
14 5 Slide Barcode The number of Possible reasons:
Count Mismatch - slides specified • An incorrect count was
%d expected, but in the pre-run input in the checklist
%d acquired. checklist did not
• The barcode reader is
match the
not functioning properly
number found.
• The barcode blow-off is
not functioning
• The barcode reader is
not adjusted properly
14 6 Run Aborted. Not Staining Run The staining Check the communications
acknowledged by instrument path from the instrument
remote. Error encountered a to the host.
Code %d, Please timeout
Retry. (%d condition during
represents the a staining run.
error code).
14 7 Run download Staining Run During the run Restart the staining run.
Timed Out, Restart process, the run
Run. download
information is
formatted
incorrectly.
14 8 Run Download Staining Run A slide is seated Restart the staining run.
Aborted with Bad on a faulty
Format Code %d, heater during
Please Retry. (%d the staining run.
represents the
format code).
14 9 The following slide Staining Run Communication Contact Ventana to replace
heater(s) are between the faulty heaters.
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,50/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,51/55
Continue anyway.
Press [ Ignore ] to
Continue anyway.
Press [ Abort ] to
abort Test
Function. Press
[ Ignore ] to
Continue anyway.
14 20 Run exceeds
desired delayed
start time. Run
starting
immediately.
14 21 Run exceeds
desired delayed
completion time.
Run starting
immediately.
14 22 Database version Staining Run The dispense Send attempt to fix code.
fields are out of record’s
date. Please database version
shutdown and field is different
restart NexES. than the
database version
in the HASP key.
14 23 Database Staining Run The ownership Send attempt to fix code.
ownership version database version
fields are out of field of the
date. Please dispense record
shutdown and is different than
restart NexES. the database
version field in
the dispense
record.
14 25 Slide Barcode
Check Digit Error,
Please re-label
slide at position: %
d
14 27 No reagents found
on reagent
carousel. Do you
wish to continue
anyway?
14 29 Protocol "%0:s" at
slide position #%
1:d is missing
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,52/55
Processing Steps.
14 31 %0:s dispenser #%
1:d missing from
Dispenser
database.
14 32 Duplicate %0:d #%
1:d dispensers on
tray.
14 33 %0:d dispenser #%
1:d missing from
Package.
14 34 %0:s, #%1:d
dispenser missing
from tray
14 35 %0:s, Serial #%1:d
is owned by host
#%2:d and can not
be used
14 36 Dispensers %0:s,
#%1:d at position
%2:d and a
dispenser of "%
3:s" must be
beside each other.
14 37 Dispensers %0:s,
#%1:d and %2:s,
#%3:d must be
beside each other.
14 38 Not enough of
reagent %s to
perform run
14 39 %0:s, #%1:d
dispenser missing
from tray
14 39 Reagent %s
missing from tray
14 40 Reagent missing in
Argument
database, product
code: %d.
14 41 %0:s, Serial #%1:d
has expired and
can not be used
14 42 %0:s, Serial #%1:d
has been
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,53/55
Deactivated and
can not be used
14 43 %0:s, Serial #%1:d
has no reagent in
it and can not be
used
14 44 %0:s, Serial #%1:d
is not enabled for
this instrument
type and can not
be used
14 45 %0:s, Serial #%1:d
is currently in use
on another
instrument
14 46 Continuing with
dispenser %0:s, #%
1:d missing from
tray
14 47 %0:s dispenser
missing from tray,
required adjacent
to dispenser "%
1:s", #%2:d at
position %3:d.
14 48 Inadequate waste
capacity to
continue.
14 49 Inadequate %s
bulk volume to
perform slide run.
14 51 A batch run is
starting on
instrument: %s.
For those slides,
the internal bulk
capacity must be
supplemented by
the external
bottles. Verify that
the following
amounts are
available in these
external bottles:
14 52 Protocol #%0:s at Staining Run The protocol is 1. Check the protocol
slide position #% defined for a names and numbers.
1:d is not defined different
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,54/55
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8
BenchMark XT_LT - Error codes 页码,55/55
the host
application.
99 7 The Zip file already Attempting to Rename the old file if you
exists. Overwrite?. create a zip want to keep it. Otherwise,
archive file and click Yes to automatically
the zip file overwrite the file.
already exists.
99 8 Error reading data Unable to read 1. Try again.
from hardware the HASP key
2. Reinstall the HASP
key. when creating a
drivers.
zip archive.
3. Reseat the HASP key.
4. Replace the key if
necessary.
https://grips.dia.roche.com/isdocdata/K2/ServiceManual/html/3101057035.htm 2016/3/8