4373e 90B01 PDF
4373e 90B01 PDF
4373e 90B01 PDF
Component Analyzer
User's Guide
Notices
© Keysight Technologies 2018 disclose the Software. The EULA and the NOT BE LIABLE FOR ERRORS OR FOR
license set forth therein, does not require INCIDENTAL OR CONSEQUENTIAL
No part of this manual may be reproduced
or permit, among other things, that Key- DAMAGES IN CONNECTION WITH THE
in any form or by any means (including
sight: (1) Furnish technical information FURNISHING, USE, OR PERFORMANCE OF
electronic storage and retrieval or transla-
related to commercial computer software THIS DOCUMENT OR ANY INFORMATION
tion into a foreign language) without prior
or commercial computer software docu- CONTAINED HEREIN. SHOULD KEYSIGHT
agreement and written consent from
mentation that is not customarily provided AND THE USER HAVE A SEPARATE
Keysight Technologies as governed by
to the public; or (2) Relinquish to, or other- WRITTEN AGREEMENT WITH WARRANTY
United States and international copyright
wise provide, the government rights in TERMS COVERING THE MATERIAL IN THIS
laws.
excess of these rights customarily provided DOCUMENT THAT CONFLICT WITH THESE
Manual Part Number to the public to use, modify, reproduce, TERMS, THE WARRANTY TERMS IN THE
release, perform, display, or disclose com- SEPARATE AGREEMENT WILL CONTROL.
4373E-90B01 mercial computer software or commercial
computer software documentation. No Safety Notices
Edition additional government requirements
Edition 1.0, June 2018 beyond those set forth in the EULA shall
apply, except to the extent that those CAUTION
Keysight Technologies Deutschland GmbH terms, rights, or licenses are explicitly
Herrenberger Strasse 130, required from all providers of commercial A CAUTION notice denotes a hazard. It
71034 Böblingen, Germany computer software pursuant to the FAR and calls attention to an operating proce-
the DFARS and are set forth specifically in dure, practice, or the like that, if not
Technology Licenses writing elsewhere in the EULA. Keysight
correctly performed or adhered to,
The hardware and/or software described in shall be under no obligation to update,
revise or otherwise modify the Software.
could result in damage to the product
this document are furnished under a
With respect to any technical data as or loss of important data. Do not pro-
license and may be used or copied only in
accordance with the terms of such license. defined by FAR 2.101, pursuant to FAR ceed beyond a CAUTION notice until
12.211 and 27.404.2 and DFARS 227.7102, the indicated conditions are fully
the U.S. government acquires no greater understood and met.
U.S. Government Rights than Limited Rights as defined in FAR
The Software is “commercial computer 27.401 or DFAR 227.7103-5 (c), as appli-
software,” as defined by Federal Acquisition cable in any technical data.
Regulation (“FAR”) 2.101. Pursuant to FAR WARNING
12.212 and 27.405-3 and Department of Warranty A WARNING notice denotes a hazard. It
Defense FAR Supplement calls attention to an operating proce-
THE MATERIAL CONTAINED IN THIS
(“DFARS”) 227.7202, the U.S. government dure, practice, or the like that, if not
DOCUMENT IS PROVIDED "AS IS," AND IS
acquires commercial computer software
SUBJECT TO BEING CHANGED, WITHOUT
correctly performed or adhered to,
under the same terms by which the soft- could result in personal injury or death.
NOTICE, IN FUTURE EDITIONS. FURTHER,
ware is customarily provided to the public. Do not proceed beyond a WARNING
TO THE MAXIMUM EXTENT PERMITTED BY
Accordingly, Keysight provides the Soft- notice until the indicated conditions
APPLICABLE LAW, KEYSIGHT DISCLAIMS
ware to U.S. government customers under are fully understood and met.
ALL WARRANTIES, EITHER EXPRESS OR
its standard commercial license, which is
IMPLIED WITH REGARD TO THIS MANUAL
embodied in its End User License Agree-
AND ANY INFORMATION CONTAINED
ment (EULA), a copy of which can be found
HEREIN, INCLUDING BUT NOT LIMITED TO
at http://www.keysight.com/find/sweula.
THE IMPLIED WARRANTIES OF
The license set forth in the EULA represents
MERCHANTABILITY AND FITNESS FOR A
the exclusive authority by which the U.S.
PARTICULAR PURPOSE. KEYSIGHT SHALL
government may use, modify, distribute, or
4 Making Measurements
Starting the Lightwave Component Analyzer 52
Calibrating the Network Analyzer before Measurements 53
About measurements 53
Electrical calibration for single-ended measurements 53
Electrical calibration for differential OE measurements 54
Electrical calibration for differential EO measurements 56
Aborting a Measurement 82
Measuring Optical Power 83
6 Measurement Concepts
General Measurement Techniques and Considerations 100
EO measurements (lasers, modulators) 102
OE measurements (photodiodes) 103
OO measurements 103
7 Maintenance
Cleaning 118
Safety Precautions 118
Why is it important to clean optical devices? 119
What do I need for proper cleaning? 119
Additional Cleaning Equipment 122
Preserving Connectors 124
Cleaning Instrument Housings 125
Which Cleaning Procedure should I use? 125
How to clean connectors 126
How to clean connector interfaces 127
How to clean bare fiber adapters 128
How to clean lenses 129
How to clean instruments with a fixed connector interface 129
How to clean instruments with a physical contact interface 130
How to clean instruments with a recessed lens interface 131
How to clean optical devices which are sensitive to mechanical stress and
pressure 132
8 Troubleshooting
Checking the Operation of the Lightwave Component
Analyzer 136
Event Log 136
10 Definition of Terms
Definition of Terms 142
11 Specifications
Specifications 150
Definitions 150
Specifications (guaranteed) 150
Typical Values (characteristics) 150
General Characteristics 150
Measurement Conditions 151
13 Ordering Information
Ordering Information 174
Power Cords 176
Index
1 Introduction
Documentation / 10
General Safety Considerations / 11
Optical Test Set Front and Rear Panels / 18
1 Introduction
Documentation
This product has been designed and tested in accordance with the
standards listed on the manufacturer’s Declaration of Conformity (see
Declaration of Conformity on page -167), and has been supplied in a safe
condition. The documentation contains information and warnings that
must be followed by the user to ensure safe operation and to maintain the
product in a safe condition.
Safety symbols
Instrument markings
The recycling symbol indicates the general ease with which the
instrument can be recycled.
Install the instrument so that the power switch is readily identifiable and
CAUTION is easily reached by the operator. This is the instrument disconnecting
device. It disconnects the mains circuit from the mains supply before
other parts of the instrument. Alternatively, an externally installed switch
or circuit breaker (which is readily identifiable and is easily reached by
the operator) may be used as a disconnecting device.
This instrument has autoranging line voltage input. Be sure the supply
CAUTION voltage is within the specified range.
Laser Safety
Table 1
N4373E
* CW output power is defined as the highest possible optical output power that the laser source
can produce at the output connector
To avoid the possibility of injury or death, you must observe the following
WARNING precautions before switching on the instrument.
• Insert the power cable plug only into a socket outlet provided with a
protective earth contact. Do not use an extension cord without a
protective conductor. Using an extension cord without a protective
conductor means the instrument is not earthed.
• Do not interrupt the protective earth connection intentionally.
• Do not remove protective covers. Operating personnel must not
remove instrument covers. Component replacement and internal
adjustments must be made only by qualified service personnel.
• Instruments that appear damaged or defective should be made
inoperative and secured against unintended operation until they can
be repaired by qualified service personnel.
• Defective, damaged, or malfunctioning instruments must be returned
to an Keysight Technologies Service Center.
Do not operate the instrument in the presence of flammable gases or
fumes. Operation of any electrical instrument in such an environment
constitutes a definite safety hazard.
Environmental information
10 9
1 2 3 4 5 6 7
For further information on the front panel of the network analyzer, please
refer to the documentation supplied with the network analyzer.
3 Laser switch Press the switch to turn the laser on (LED on),
and off (LED off).
The laser can also be switched on and off from
the LCA Measurement Setup Macro (in the LCA
System Settings menu), or remotely.
This switch overrides other.
Pressing this switch during a measurement will
invalidate your results.
9 11
10
12
1 2 3 4 5 6 7
For further information on the front panel of the network analyzer, please
refer to the documentation supplied with the network analyzer.
3 Laser switch Press the switch to turn the laser on (LED on),
and off (LED off).
The laser can also be switched on and off from
the LCA Measurement Setup Macro (in the LCA
System Settings menu), or remotely.
This switch overrides other.
Pressing this switch during a measurement will
invalidate your results.
Rear panel
1 2 3
For information on the rear panel of the network analyzer, please refer to
the documentation supplied with the network analyzer.
2 Setting Up the
Instrument
Unpacking the Lightwave Component Analyzer / 26
Mounting the Test Set and the Network Analyzer / 28
Setting Up the Lightwave Component Analyzer / 34
2 Getting Started
1 Use a Torx T15 screwdriver and the screws 0515-2317 to mount the
left and right bracket adapter N4373-25271 to the rear side of the test
head.
2 Use the screws 0515-0433 and a Torx T20 screwdriver to attach the
rear brackets N4373-25290 and N4373-25291 to the right and left
bracket adapter.
3 Turn the test head on its side and use a Pozidrive size 2 screwdriver to
mount the adjustable feet 0403-1166 to the left and right brackets.
5 Turn the Network Analyzer on its side. Unlock the feet to remove them
from the bottom of the network analyzer.
6 Using a Torx T20 screwdriver, remove the 2 lower rear feet from the
rear panel of the network analyzer
7 Put the network analyzer on top of the test head and align it with the
mounting slots of the rear brackets.
8 Use the Torx T20 screwdriver and the screws 0515-1619 to connect
the network analyzer and left and right rear brackets.
9 Use a Torx T20 screwdriver to attach the side bars 83427-21274 with
the screws 0515-1269 to the left and right sides of the two
instruments.
10 Attach the adhesive trim covers to the left and right side bars to cover
the screws.
11 Use the 13mm wrench to level the adjustable feet to the other feet of
the test head. Use the upper nuts to lock the feet in place.
Before you move the system or return it to a service center, make sure
NOTE you
• put the rigid cables back into the horizontal position and
• reattach the cable guard.
3 Remove the dust cover from Port B on the optical test set.
4 Attach the female to female adapter (part number 85058-60121) to
Port B.
5 Screw the adapter finger-tight to Port B without rotating it.
6 Holding the adapter with the spanner wrench from the calibration kit
(part number 08513-20014), use the 20mm torque wrench from the
calibration kit (0.90 N.m/8 lb.in, part number 8710-1764) to tighten
the adapter.
Always rotate the nut and never the female to female adapter.
CAUTION
Rotating the adapter can damage the electrical interface of the adapter
and the connector on the optical test set.
For information on handling using and maintaining the flexible test port
NOTE cables, please refer to the Operating and Service Manual, Keysight
N4697E/F NMD-1.85 mm -f- to 1.85 mm Flexible Test Port Return
Cables (part no. N4697-90001).
7 Remove the dust cover from Port 2 of the 2-port network analyzer, or
from Port 4 of the 4-port network analyzer. Remove the dust cover
from both ends of the flexible test port cable (part number
N4697-60030 or 85133-60043).
8 Connect the test port cable finger-tight to Port B on the optical test set
without rotating the female to female adapter. Connect the other end
to Port 2 of the 2-port network analyzer or, respectively, to Port 4 of
the 4-port network analyzer.
9 Holding the adapter with the spanner wrench, use the 20 mm torque
wrench from the calibration kit to tighten both connections.
Always use optical patch cords to connect to your DUT. This protects the
CAUTION connectors of the optical test set, by minimizing the number of connector
changes.
Make sure to connect only matching connector types to the optical test
CAUTION set: connecting a straight connector to an angled port, or vice versa, will
damage both interfaces.
15 On the rear, connect the network analyzer to the optical test set using
the supplied USB cable.
3 Performance Quick
Check
Preparation / 40
Verifying the Optical Connections and Optical CW Operation / 43
Verifying the Electrical Connection and Electro-optical Operation / 46
Preparation
Passing this test ensures that the electrical calibration is within expected
performance.
1 Follow the steps in the sections Preparation on page -40 and Network
analyzer settings and electrical calibration on page -40.
2 Connect the N5520B "1.85 mm Coaxial Adapters" (or 85056-60006
"2.4 mm Coaxial Adapters") supplied between the two N4697-60200
"1.85 mm Flexible Cables for Test Ports" (or 85133-60016 "2.4 mm
Flexible Cables for Test Ports") supplied.
3 Press the [Macro/Local utility] button until the Macro 1 group is
6 Click .
7 Make a corrected measurement with the network analyzer.
8 Inspect the results.
• Electrical return loss S11 should be similar to S22.
• Both curves should be below -20 dB over the full measurement
range, with only low frequency dependence.
Passing these tests ensures that the transmitter and receiver are within
expected optical CW performance.
1 Follow the steps in the sections Preparation on page -40 and Network
analyzer settings and electrical calibration on page -40.
2 If you have option -021:
Connect the two straight-angled optical patch cords (N4373-87907) to
the optical output and input 1 (“+7 dBm”) or input 2 (“+17 dBm”). Note
that the angled connector must be attached to the optical test set.
Use the FC/PC adapter (1005-0256) to connect the straight ends of
the optical patch cords.
If you have option -022:
Connect the angled-angled optical patch cords (N4373-87906) to the
optical output and input 1 (“+7 dBm”) or input 2 (“+17 dBm”).
Use the FC/APC through adapter (1005-1027) to connect the ends of
the optical patch cords.
3 Press the [Macro/Local] utility button until the Macro 1 group is
The following tests ensure that the internal transmitter and receiver of the
Lightwave Component Analyzer operate as expected.
1 Follow the steps in the sections Preparation on page -40 and Network
analyzer settings and electrical calibration on page -40.
2 On the network analyzer make the following settings:
• IF bandwidth: 100 Hz
• Select “Reduce IF BW at Low Frequencies”.
3 If you have option -021:
Connect the two straight-angled optical patch cords (N4373-87907) to
the optical output and input 1 (“+7 dBm”) or input 2 (“+17 dBm”). Note
that the angled connector must be attached to the optical test set.
Use the FC/PC through adapter (1005-0256) to connect the straight
ends of the optical patch cords.
If you have option -022:
Connect the angled-angled optical patch cords (N4373-87906) to the
optical output and input 1 (“+7 dBm”) or input 2 (“+17 dBm”).
Use the FC/APC through adapter (1005-1027) to connect the ends of
the optical patch cords.
4 Connect the two supplied flexible test port cables to the RF ports of
the Lightwave Component Analyzer test set.
5 Press the [Macro/Local] utility button until the Macro 1 group is
4 Making
Measurements
Starting the Lightwave Component Analyzer / 52
Calibrating the Network Analyzer before Measurements / 53
Measuring Opto-Electrical (OE) devices / 58
Measuring Electro-Optical (EO) Devices / 68
Returning to Electrical Measurements / 74
Measuring Optical-Optical (OO) Devices / 75
Aborting a Measurement / 82
Measuring Optical Power / 83
4 Making Measurements
About measurements
To ensure the specified performance of the LCA, you need to perform a full
two-port electrical calibration. For this purpose we recommend using the
automated electronic calibration kit N4694A #00F or N4694A #00A.
To ensure the specified performance of the LCA, you should perform a full
three- port electrical calibration. See the Keysight website for applicable
calibration kits.
If you want to perform an OE user calibration (see page 61 for details), we
recommend executing a full four-port calibration, or a two-port calibration
between network analyzer ports 1 and 4, in addition to the three-port
calibration (see Electrical calibration for single-ended measurements on
page -53). Connect an additional RF cable (N4697-60035) to port 4.
Calibrate before you start using the Lightwave Component Analyzer
application. Alternatively, return to EE measurement mode, as described in
Returning to Electrical Measurements on page -74, to calibrate.
and select the appropriate Cal Kit for each port, or “ECal Thru as
Unknown”.
7 Make sure “Do orientation” is selected, and click [Next >].
8 Follow the prompts to complete the calibration:
• Connect the ECal Module to the RF Cables connected to ports 1
and 2 of the network analyzer. Click "Measure".
and select the appropriate Cal Kit for each port, or “ECal Thru as
Unknown”.
7 Make sure “Do orientation” is selected, and click [Next>].
8 Follow the prompts to complete the calibration:
• Connect the ECal Module to the RF Cables connected to ports 2
and 4 of the network analyzer. Click "Measure".
• Connect the ECal Module to the RF Cables connected to ports 3
and 4 of the network analyzer. Click "Measure".
9 After the calibration has been completed, save the calibration file for
later reuse.
For OE measurements, the LCA sets up traces for S22, S21, S33, S31 by
default.
For more information, see Differential Measurements (applies to 4-port
PNA only) on page -110.
Before you make your first opto-electrical measurement, you must
initialize the OE measurement. You can see which measurement is
currently initialized at the top left of the trace screen.
If you have already initialized the OE measurement, you can connect your
1 Press .
2 Select the “OE Measurement” tab.
If you already have an .s2p file for your device, set "Deembedding
Data" to "File" and enter or select the .s2p file name.
c Click to proceed.
d Choose whether you want to save the calibration data for later use.
For EO measurements the LCA sets up traces for S22, S42, S33, S43 by
default. For more information, see
Differential Measurements (applies to 4-port PNA only) on page -110.
Before you make your first electro-optical measurement, you must switch
to EO measurement mode. You can see which measurement is currently
initialized at the top left of the trace screen.
If you have already initialized the EO measurement, you can connect your
.
Please refer to the user’s guide and online help of the network analyzer for
further information.
1 Press .
2 Select the “EO Measurement” tab.
If you already have an .s2p file for your device, set "Deembedding
Data" to "File" and enter or select the .s2p file name.
8 If you are using a 75 Ohm minimum loss pad, enable the RF Path
Deembedding, and make sure 75 Ohm is selected.
Enter or select the file with the characterization of the minimum loss
pad.
Before you measure purely electrical devices with the network analyzer, or
perform an electrical calibration, you need to switch to "EE Measurement"
mode.
Initializing the EE Measurement as described here removes all the fixturing
information in the network analyzer that is specific to the optical test set.
Also, the RF power on all ports is set to the power specified for “Reverse
RF Power [dBm]"
1 Press .
2 Make sure the “EE Measurement” tab is selected.
3 Click .
If you have already initialized the OO measurement, you can connect your
.
Please refer to the user’s guide and online help of the network analyzer for
further information.
1 Click .
2 Select the “OO Measurement” tab.
you include a “User Calibration” in your setup (see the section “User
Calibration” under LCA System Settings on page -92) if you are not
using “Continuous” modulator bias optimization.
The system is calibrated with a default setting. You can determine the
default setting by going to the “Tools” page, choosing “Default”, and
pressing “Run”.
7 If you need to deembed optical components, such as attenuators, fiber
optic cables or splitters, located between the laser output and your
device under test, open and enable the “Optical Path Deembedding”.
Here you can enter the parameters of the optical path directly.
If you already have an .s2p file for your device, set "Deembedding
Data" to "File" and enter or select the .s2p file name.
c Click to proceed.
d Choose whether you want to save the calibration data for later use.
Making OO measurements
Aborting a Measurement
1 Press .
2 Select Input 2 as the Optical Input under Receiver on the “Tools” tab.
3 Under Receiver on the “Tools” tab, specify the receiver wavelength.
5 User Interface
Reference
The LCA Function Keys / 86
Default and Advanced Measurement Parameters / 88
Saving and Loading Setups / 89
The Measurement Parameters / 91
5 User Interface Reference
1 If the LCA server is running, but the function keys are not displayed,
press the [Macro/Local] Utility key until Macro 1 group is shown:
Saving setups
A setup contains the current state of the LCA user interface and optionally
also the current state of the network analyzer. Setups are stored as
XML-files or as compressed XML-files.
The settings on the “Tools” tab are not saved in the setup, since the
settings shown on this tab always reflect the current state of the LCA
hardware. After loading the setup the LCA system is configured to press
“Start” from the LCA user interface.
There are two ways of saving setups.
First saving method
Click to save the current setup.
• If user calibration mode is set to “Measure” (only possible for OE and
OO measurements), you are asked to specify the electrical calibration
set to be saved with the setup.
This is the electrical calibration set to be used during user calibration
measurements.
Only a link to the electrical calibration set is saved, not the calibration
data itself.
For more details, see Measuring Opto-Electrical (OE) devices on
page -58 or Measuring Optical-Optical (OO) Devices on page -75.
• If User Calibration is set to “Load” (only possible for OE and OO
measurements), you will be asked to specify which user calibration file
should be loaded.
The filename of the calibration data is saved with the setup, not the
calibration data itself.
For more details, see Measuring Opto-Electrical (OE) devices on
page -58 or Measuring Optical-Optical (OO) Devices on page -75.
Second saving method
After initializing with the User Calibration mode set to “Measure”, you are
asked if you want to save the user calibration data and/or the LCA Setup.
Check “Also save the LCA Setup” if you want to save the LCA setup. Then
choose between “Load” and “Measure” to specify the user calibration
mode to save with the setup.
• If you selected “Load”, the “No” button is disabled to force you to store
the measured user calibration data.
After you press “Yes”, you are asked to specify the filename for the user
calibration data and the filename of the LCA setup.
Only the filename of the user calibration data is saved with the setup.
• If you selected “Measure”, it is also possible to press the “No” button. In
this case only the LCA setup is saved.
If you press “Yes”, you are asked for the filename of the LCA setup. The
name of the electrical calibration set just used for this user calibration
measurement will be saved in the LCA setup.
Before the LCA saves the setup, it asks you whether you want to save the
network analyzer state with your LCA setup.
Loading setups
Measure Mode
Differential
If you select this mode, means the LCA uses a differential electrical input
or output for this measurement.
Whether the differential signal is the input to the DUT or the output from
the DUT, this uses ports 2 and 3 of a four port network analyzer.
Single-Ended
If you select this mode, the LCA uses a single-ended electrical input or
output for this measurement.
The electrical input to the DUT is Port 1 of the network analyzer.
The electrical output from the DUT is Port 2 on a two-port network
analyzer or, respectively, Port 4 on a four-port network analyzer.
Wavelength (nm)
• If you have not selected to use the optional External Input, select the
wavelength of the laser source and the optical power meter use for
measurements.
• If you have selected to use the External Input, you can enter the
wavelength of the external source here. This wavelength is used by the
optical power meter.
Power [dBm]
Set the optical output power of the internal laser source.
If you select to use the optional External Input, enter the average optical
output power at the LCA transmitter output connector.
Laser On
This control, which overrides the default behavior of the Lightwave
Component Analyzer, is only enabled in Advanced mode and if you have
not selected to use the optional External Input.
Select to turn the laser source on for the following LCA measurement.
Deselect to turn the laser source off for the following LCA measurement.
External Input
This control is only active for lightwave component analyzers equipped
with an optional external optical input.
• Select to use a laser source connected to the external input (on the rear
of the optical test set.
• Deselect to use the laser source in the optical test set.
Optical Input
Select the input to which your device under test is connected.
• For input signals up to +7 dBm (calibrated up to+5 dBm), use Input 1.
• For input signals up to +17 dBm (calibrated up to+15 dBm), use
Input 2.
User Calibration
Select the calibration data you want to use.
• None uses the factory calibration data only (no additional user
calibration data is used).
• Measure runs an extra calibration measurement step during the
measurement setup.
Before running the first measurement, you need to ensure the
instrument has a valid electrical calibration, as described in Calibrating
the Network Analyzer before Measurements on page -53.
At the end of this calibration measurement, you can save the
calibration data.
This calibration is only used for the measurement setup for which it is
specified.
• Load prompts you to load data from a file saved as part of a previous
user calibration measurement.
Modulator Optimization
Select to how the modulator bias is to be optimized.
• Select “Once” to automatically optimize the modulator bias at the first
measurement after the measurement parameters have been set.
This is sufficient for most measurements unless the tester is subject to
large changes in environmental conditions.
• Select “EverySweep” to optimize at the start of everymeasurement
sweep.
This increases accuracy, but also increases measurement times.
This is the default modulator bias optimization mode, and is used to
calibrate and verify the instrument.
• Select “Continuous” to optimize while the measurement is running.
With “Continuous” modulator bias optimization, a 10 kHz pilot tone is
applied to the signal. This improves the short and long term signal
stability of the modulator working point.
The pilot tone will affect measurements with IFBW settings of the
network analyzer >1.5 kHz.
For IFBW settings of the network analyzer >3 kHz were commended
you do not use “Continuous” modulator bias optimization, Use "Every
Sweep" or "Once" instead.
Forward RF Power [dBm]
Sets the RF power level for the source port(s).
Enable
Select if you want the LCA to compensate for the optical connections to
the DUT.
Deembedding Data
Select how you want to specify the characteristics of the optical path.
• Parameter lets you characterize the source or receiver optical path by
its length and refractive index, and attenuation.
• Length (m) and Refractive Index can only be used together. That is,
you cannot give a value just for the refractive index or just the length.
The value for the Refractive Index of a single-mode fiber is typically
1.467 at 1310 nm and 1.468 at 1550 nm.
The value for the Refractive Index of a graded index multi-mode fiber is
1.49 for 62.5 μm at 850 nm, 1.475 for 50 μm at 850 nm and 1.465 for
50 μm at 1300 nm.
• Attenuation (dB) can be used alone, or with the values for length and
refractive index.
• File lets you use the results of a measurement of the optical path to
characterize the source or receiver optical path.
RF Path Deembedding
Enable
Select if you want the LCA to compensate for the RF connections to the
DUT.
Use the results of a measurement of the RF path to characterize the
source or receiver RF path.
Tools
Some of the controls that can be selected on the Tools tab are also
available on the measurement tabs. The controls on the measurement tabs
are activated when you start the next measurement setup. The controls on
the Tools tab are activated immediately.
The Tools tab always showing the current settings of the instrument. These
may be different to what is selected on the other tabs.
Laser Source
Receiver
Optical Input
Select the input to which your device under test is connected.
• For input signals up to +7 dBm (calibrated up to+5 dBm), use Input 1.
• For input signals up to +17 dBm (calibrated up to+15 dBm), use
Input 2.
Receiver Wavelength (nm)
Set the wavelength of to be used by the optical power meter.
RF Power
Reset UI
The "Reset UI" button resets all LCA settings to default settings.
6 Measurement
Concepts
General Measurement Techniques and Considerations / 100
Signal Relationships in Opto-Electric Devices / 104
Single-Ended Measurements / 109
Differential Measurements (applies to 4-port PNA only) / 110
Triggering Concepts / 114
Measuring at Different Wavelengths and with the External Optical Source
Input / 115
6 Measurement Concepts
OE measurements (photodiodes)
The LCA measures the input optical modulation power and output
modulation current and displays the ratio of the two in Amps/Watt.
OO measurements
where:
and
Single-Ended Measurements
Single port measurements are made on devices when the signal on the
electrical port is referenced to ground. This port is described by
S-parameters.
Differential OE measurements
For OE measurements, the LCA sets up traces for S22, S21, S33, S31 by
default.
Differential EO measurements
For EO measurements the LCA sets up traces for S22, S42, S33, S43 by
default.
Differential results
Calibration Concept
Triggering Concepts
Measuring at Different Wavelengths and with the External Optical Source Input
Delete all other wavelength values. (Note: LCA interpolates for missing
wavelengths.) Optionally repeat for input 2 and enter values in 3rd
column.
We do not recommend to change the values in the "LCA_TxCorr.csv" file.
Instead we recommend you measure with user calibration for
measurements using an external input.
EO measurements
The wavelength box on the EO tab is an editable, drop down list. This
means you can select one of the default wavelengths, or enter a different
wavelength.
If you select a wavelength other than the default wavelengths, the LCA
checks for the closest default wavelength. The calibration data for this
wavelength is used for the LCA measurements.
Next the LCA checks if the "LCA_RxCorr.csv" file is available.
• If not, no wavelength correction is applied.
• If the file is found, the LCA performs a wavelength correction.
• If the wavelength you entered is not covered by the"LCA_RxCorr.csv"
file, an error message will be shown.
If the selected wavelength is covered by the correction file, the correction
factor is calculated from the values found in the file. This correction factor
is applied to the values from the internal optical power meter and to the
calibration data used for the LCA measurements.
OE measurements
On the OE tab of the LCA user interface, if you check the external (laser)
input check box, you can edit the wavelength. The principles described for
EO measurements are applied to the transmitter side, using the
wavelength correction data in the file: "LCA_TxCorr.csv".
The LCA expects the power value in the "Power" box to be the optical
power at the Optical Output of the LCA test head.
To set the right power value if you are using an external source:
1 Use a short patch cord to connect the Optical Output to the Optical
Input.
2 Switch to the "Tools" tab of the user interface and set the receiver
wavelength to the external laser wavelength and check the "External
Input" check box.
OO measurements
7 Maintenance
Cleaning / 118
Replacing the Fuse / 133
Cleaning
Safety Precautions
Before you can start your cleaning procedure you need the following
standard equipment:
• Dust and shutter caps
• Isopropyl alcohol
• Cotton swabs
• Soft tissues
• Pipe cleaner
• Compressed air
Isopropyl alcohol
Cotton swabs
We recommend that you use swabs such as Q-tips or other cotton swabs
normally available from local distributors of medical and hygiene products
(for example, a supermarket or a chemist's shop). You may be able to
obtain various sizes of swab. If this is the case, select the smallest size for
your smallest devices.
Ensure that you use natural cotton swabs. Foam swabs will often leave
behind filmy deposits after cleaning.
Use care when cleaning, and avoid pressing too hard onto your optical
device with the swab. Too much pressure may scratch the surface, and
could cause your device to become misaligned. It is advisable to rub gently
over the surface using only a small circular movement.
Swabs should be used straight out of the packet, and never used twice.
This is because dust and dirt in the atmosphere, or from a first cleaning,
may collect on your swab and scratch the surface of your optical device.
Soft tissues
These are available from most stores and distributors of medical and
hygiene products such as supermarkets or chemists' shops.
We recommend that you do not use normal cotton tissues, but
multi-layered soft tissues made from non-recycled cellulose. Cellulose
tissues are very absorbent and softer. Consequently, they will not scratch
the surface of your device over time.
Use care when cleaning, and avoid pressing on your optical device with
the tissue. Pressing too hard may lead to scratches on the surface or
misalignment of your device. Just rub gently over the surface using a small
circular movement.
Use only clean, fresh soft tissues and never apply them twice. Any dust
and dirt from the air which collects on your tissue, or which has gathered
after initial cleaning, may scratch and pollute your optical device.
Pipe cleaner
Compressed air
Ultrasonic bath
Only use water if you are sure that there is no other way of cleaning your
optical device without corrosion or damage. Do not use hot water, as this
may cause mechanical stress, which can damage your optical device.
Ensure that your liquid soap has no abrasive properties or perfume in it.
You should also avoid normal washing-up liquid, as it can cover your
device in an iridescent film after it has been air-dried.
Some lenses and mirrors also have a special coating, which may be
sensitive to mechanical stress, or to fat and liquids. For this reason we
recommend you do not touch them.
If you are not sure how sensitive your device is to cleaning, please contact
the manufacturer or your sales distributor.
Some special lens cleaning papers are not suitable for cleaning optical
devices like connectors, interfaces, lenses, mirrors and so on. To be
absolutely certain that a cleaning paper is applicable, please ask the
salesperson or the manufacturer.
Polymer film
Take care never to look into the end of a fiber or any other optical
WARNING component, when they are in use. This is because the laser can seriously
damage your eyes.
Preserving Connectors
Listed below are some hints on how best to keep your connectors in the
best possible condition.
Making Connections
Before you make any connection you must ensure that all cables and
connectors are clean. If they are dirty, use the appropriate cleaning
procedure.
When inserting the ferrule of a patch cord into a connector or an adapter,
make sure that the fiber end does not touch the outside of the mating
connector or adapter. Otherwise you will rub the fiber end against an
unsuitable surface, producing scratches and dirt deposits on the surface of
your fiber.
Be careful when replacing dust caps after use. Do not press the bottom of
the cap onto the fiber as any dust in the cap can scratch or dirty your fiber
surface.
When you have finished cleaning, put the dust cap back on, or close the
shutter cap if the equipment is not going to be used immediately.
Keep the caps on the equipment always when it is not in use.
All of Keysight Technologies’ lightwave instruments and accessories are
shipped with either laser shutter caps or dust caps. If you need additional
or replacement dust caps, contact your nearest Keysight Technologies
Sales/Service Office.
Use a dry and very soft cotton tissue to clean the instrument housing and
the keypad. In the case of heavy dirt, you can moisten the cotton tissue in
water.
Light dirt
If you just want to clean away light dirt, observe the following procedure
for all devices:
• Use compressed air to blow away large particles.
• Clean the device with a dry cotton swab.
• Use compressed air to blow away any remaining filament left by the
swab.
Heavy dirt
If the above procedure is not enough to clean your instrument, follow one
of the procedures below. Please consult Cleaning on page -118 for the
procedure relevant for this instrument.
If you are unsure of how sensitive your device is to cleaning, please contact
the manufacturer or your sales distributor.
Never look into the end of an optical cable that is connected to an active
WARNING source.
To assess the projection of the emitted light beam you can use an infrared
sensor card. Hold the card approximately 5 cm from the output of the
connector. The invisible emitted light is project onto the card and becomes
visible as a small circular spot.
Preferred Procedure
Use this procedure particularly when there is greasy dirt on the connector:
1 Moisten a new cotton-swab with isopropyl alcohol.
2 Clean the connector by rubbing the cotton-swab over the surface
using a small circular movement.
3 Take a new, dry soft-tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular
movement.
4 Blow away any remaining lint with compressed air.
An Alternative Procedure
Be careful when using pipe-cleaners, as the core and the bristles of the
CAUTION pipe-cleaner are hard and can damage the interface.
Preferred Procedure
Use this procedure particularly when there is greasy dirt on the interface:
1 Moisten a new pipe-cleaner with isopropyl alcohol.
2 Clean the interface by pushing and pulling the pipe-cleaner into the
opening. Rotate the pipe-cleaner slowly as you do this.
3 Moisten a new cotton-swab with isopropyl alcohol.
4 Using a new, dry pipe-cleaner, and a new, dry cotton-swab remove the
alcohol, any dissolved sediment and dust.
5 Blow away any remaining lint with compressed air.
Bare fiber adapters are difficult to clean. Protect from dust unless they are
in use.
Never use any kind of solvent when cleaning a bare fiber adapter as
CAUTION solvents can damage the foam inside some adapters.
They can deposit dissolved dirt in the groove, which can then dirty the
surface of an inserted fiber.
Preferred Procedure
Use this procedure particularly when there is greasy dirt on the adapter:
1 Clean the adapter by pushing and pulling a new, dry pipe-cleaner into
the opening. Rotate the pipe-cleaner slowly as you do this.
Be careful when using pipe-cleaners, as the core and the bristles of the
CAUTION pipe-cleaner are hard and can damage the adapter.
2 Clean the adapter by pushing and pulling a new, dry pipe cleaner into
the opening. Rotate the pipe cleaner slowly as you do this.
3 Clean the adapter by rubbing a new, dry cotton-swab over the surface
using a small circular movement.
4 Blow away any remaining lint with compressed air.
Some lenses have special coatings that are sensitive to solvents, grease,
liquid and mechanical abrasion. Take extra care when cleaning lenses with
these coatings.
Lens assemblies consisting of several lenses are not normally sealed.
Therefore, use as little alcohol as possible, as it can get between the lenses
and in doing so can change the properties of projection.
Preferred procedure
You should only clean instruments with a fixed connector interface when it
is absolutely necessary. This is because it is difficult to remove any used
alcohol or filaments from the input of the optical block.
It is important, therefore, to keep dust caps on the equipment at all times,
except when your optical device is in use.
Only use clean, dry compressed air. Make sure that the air is free of dust,
CAUTION water, and oil. If the air that you use is not clean and dry, this can lead to
filmy deposits or scratches on the surface of your connector interface.
This will degrade the performance of your transmission system.
Never try to open the instrument and clean the optical block by yourself,
because it is easy to scratch optical components, and cause them to be
misaligned.
Remove any connector interfaces from the optical output of the instrument
before you begin the cleaning procedure.
Cleaning interfaces is difficult as the core diameter of a single-mode fiber
is only about 9 mm. This generally means you cannot see streaks or
scratches on the surface. To be certain of the degree of pollution on the
surface of your interface and to check whether it has been removed after
cleaning, you need a microscope.
Never look into an optical output, because this can seriously damage
WARNING your eyesight.
To assess the projection of the emitted light beam you can use an infrared
sensor card. Hold the card approximately 5 cm from the interface. The
invisible emitted light is projected onto the card and becomes visible as a
small circular spot.
Preferred procedure
For instruments with a deeply recessed lens interface (for example the
Keysight Technologies 81633A and 81634A Power Sensors) do NOT follow
this procedure. Alcohol and compressed air could damage your lens even
further.
Keep your dust and shutter caps on when your instrument is not in use.
This should prevent it from getting too dirty.
If you must clean such instruments, please refer the instrument to the
skilled personnel of the Keysight service team.
Preferred procedure
How to clean optical devices which are sensitive to mechanical stress and pressure
Preferred procedure
Alternative procedure
For these types of optical devices you can often use an ultrasonic bath
with isopropyl alcohol. Only use the ultrasonic bath if you are sure that it
won’t cause any damage to any part of the device.
1 Put the device into the bath for at least three minutes.
2 Blow away any remaining liquid with compressed air.
If there are any streaks or drying stains on the surface, repeat the cleaning
procedure.
Bare fiber ends are often used for splices or, together with other optical
components, to create a parallel beam.
The end of a fiber can often be scratched. You make a new cleave. To do
this:
1 Strip off the cladding.
2 Take a new soft tissue and moisten it with isopropyl alcohol.
3 Carefully clean the bare fiber with this tissue.
4 Make your cleave and immediately insert the fiber into your bare fiber
adapter in order to protect the surface from dirt.
(1)
1 On the rear of the optical test set, use a flat-tipped screwdriver to lever
open the fuse holder from underneath (1).
2 Pull the fuse holder and fuse clear of the back panel.
3 Take the fuse out of the fuse cap and note the fuse rating.
4 Make sure the replacement fuse has the same rating (T1A, 250V,
Keysight part number 2110- 0007).
5 Put the fuse into the fuse holder.
6 Put the fuse and fuse cap into the receptacle in the rear of the case and
press it into place.
8 Troubleshooting
Checking the Operation of the Lightwave Component Analyzer / 136
Event Log / 136
Event Log
9 Reinstalling and
Updating the
Software
Backing Up your LCA Data / 138
Reinstalling the LCA Software / 139
9 Reinstalling and Updating the Software
The information for your particular configuration of the LCA and the factory
calibration data are kept in the “Table” and “Information” subdirectories of
the directory in which the software is installed.
These two directories are backed up under the "LCA" directory on the "D"
drive of your network analyzer. These directories are not overwritten when
performing the network analyzer's system recovery process.
Please refer to the user documentation of the network analyzer for
information on backing up calibration data.
1 Before updating the software, you must remove the existing software.
a In Windows’ Settings, go to the Control Panel and select Add or
Remove Programs.
b Select the LCA application from the list of programs.
c Remove the program.
2 Start the LCA setup software LCAInstaller.msi.
3 Follow the instructions on the screen.
We recommend you use the default settings.
When starting the “LCA Measurement Setup” with no server running, the
NOTE server is started automatically.
4 Press .
10 Definition of Terms
Definition of Terms / 142
10 Definition of Terms
Definition of Terms
Any change in the cabling after calibration add to the specified absolute
NOTE frequency response uncertainty.
where maxt{} and mint{} are the maximum and minimum value over time t.
Conditions: Time period as specified after modulator auto bias
(“modulator optimization”) is performed with the selected wavelength.
Other conditions as specified.
Measurement: Using an average power meter with averaging time set
to 1 s.
Average output power uncertainty (optical)
The maximum difference between actual (true) Average output power
(optical) and the nominal (selected) average output power.
or
For O/E and E/O converters with linear relation between optical power
NOTE and electrical current, optical ratios in dBo have half the magnitude of
the electrical ratios in dBe.
For E/O and O/E, group delay uncertainty excludes a constant group
NOTE delay offset (see Phase uncertainty). Other conditions as specified.
Applying more than the specified maximum safe input power may
CAUTION damage the LCA.
Responsivity values are converted from Decibel (dBm, dBo, dBe) values
NOTE to linear values by: linear = 10decibel/20. The average is converted back to
decibels by: decibel = 20*log10(linear).
For E/O and O/E, phase uncertainty excludes the effect of a (constant but
NOTE unknown) Group delay (GD) offset (resulting from a path length
uncertainty in the LCA system). A group delay offset Δ GD at modulation
frequency fmod causes a phase change of Δ φ = 360° - Δ GD - fmod
expressed in degrees.
Changes in the cabling after user e-cal add to the specified relative
NOTE frequency response uncertainty.
References
(1) “Guide to the Expression of Uncertainty in Measurement” (“GUM”),
BIPM, IEC, ISO et al. (1993)
11 Specifications
Specifications / 150
Transmitter and Receiver Specifications / 152
Specifications for Electro-Optical Measurements at 1310 nm (E/O Mode) /
154
Specifications for Electro-Optical Measurements at 1550 nm (E/O Mode) /
155
Specifications for Opto-Electrical Measurements at 1310 nm (O/E Mode) /
156
Specifications for Opto-Electrical Measurements at 1550 nm (O/E Mode) /
157
Specifications for Optical to Optical Measurements at 1310 nm (O/O Mode) /
158
Specifications for Optical to Optical Measurements at 1550 nm (O/O Mode) /
159
Specifications for Electrical-Electrical Measurements (E/E Mode) / 160
General Characteristics / 161
Mechanical Outline Drawings, options 2xx (dimensions in mm) / 163
Mechanical Outline Drawings, options 4xx (dimensions in mm) / 164
11 Specifications
Specifications
Definitions
Specifications (guaranteed)
General Characteristics
Give additional information for using the instrument. These are general
descriptive terms that do not imply a level of performance.
Specifications require an angled connector at the source output and at
optical input ports 1 and 2 of the optical receiver.
Angled contact connectors help you to control return loss.
The contact connector on your Lightwave Component Analyzer is angled,
use only cables with angled connectors. The Lightwave Component
Analyzer input requires angled connectors. Do not use a cable with a flat
connector on either the angled input connector or on the angled output
connector.
The angled connector symbol is typically colored green.
Measurement Conditions
+5 dBm @ 1550 nm
¥G%PWRG%P#QP
Ě¥G%:$ ± 0.03 dBe ± 0.03 dBe ± 0.11 dBe ± 0.4 dBe ± 0.8 dBe
Ě¥G%:$ ± 0.03 dBe ± 0.03 dBe ± 0.6 dBe ± 1.3 dBe ± 2.2 dBe
Minimum measurable frequency response (noise ¥G%:$ ¥G%:$ ¥G%:$ ¥G%:$ ¥G%:$
floor) 2, 4, 5
Phase uncertainty DUT response – – – – –
(typical) 3
Ě¥G%:$1 ± 3.5° ± 3.0° ± 2.7° ± 3.7° ± 5.5°
Group delay uncertainty Derived from phase uncertainty, see section “Group delay uncertainty”.
Example: ± 2.0° ĵ8 ps (1 GHz aperture)
)RU'87UHVSRQVHPD[¥G%:$
2. IFBW = 10 Hz.
([FHSWSKDVHZUDSDOLDVLQJ([DPSOH$'87JURXSGHOD\RIQVPFDEOHOHQJWKUHTXLUHVDIUHTXHQF\VWHSVL]HRIę*+]WRDYRLGSKDVHZUDSV
Excluding a constant group delay offset of < ± 0.3 ns typical. (Cable length uncertainty < ± 0.06 m). A constant group delay offset leads to a phase offset
đŠ ¡§đ*'§IPRGLQGHJ
4. Average value over frequency range.
5. In reverse coupler configuration, for normal configuration add typically 35 dB (0.05 GHz to 0.2 GHz), 12 dB (0.2 GHz to 0.7 GHz), 8 dB (> 0.7 GHz).
Ě¥G%:$ ± 0.02 dBe ± 0.02 dBe ± 0.02 dBe ± 0.3 dBe ± 0.5 dBe
Ě¥G%:$ ± 0.02 dBe ± 0.02 dBe ± 0.1 dBe ± 1 dBe ± 2.0 dBe
Minimum measurable frequency response ¥G%:$ ¥G%:$ ¥G%:$ ¥G%:$ ¥G%:$
(noise floor) 2, 4, 5
Phase uncertainty DUT response – – – – –
(typical) 3 Ě¥G%:$1 ± 3.5° ± 3.0° ± 2.3° ± 3.2° ± 4.5°
Ě¥G%:$ ± 5.5° ± 3.5° ± 2.3° ± 4.2° ± 6.5°
Group delay uncertainty Derived from phase uncertainty, see section “Group delay uncertainty”.
Example: ¡ĵSV*+]DSHUWXUH
)RU'87UHVSRQVHPD[¥G%:$
2. IFBW = 10 Hz.
([FHSWSKDVHZUDSDOLDVLQJ([DPSOH$'87JURXSGHOD\RIQVPFDEOHOHQJWKUHTXLUHVDIUHTXHQF\VWHSVL]HRIę*+]WRDYRLGSKDVH
wraps). Excluding a constant group delay offset of < ± 0.3 ns typical. (Cable length uncertainty < ± 0.06 m). A constant group delay offset leads
WRDSKDVHRIIVHWđŠ ¡§đ*'§IPRGLQGHJ
4. Average value over frequency range.
5. In reverse coupler configuration, for normal configuration add typically 35 dB (0.05 GHz to 0.2 GHz), 12 dB (0.2 GHz to 0.7 GHz), 8 dB (> 0.7 GHz).
Ě¥G%$: ± 0.03 dBe ± 0.03 dBe ± 0.15 dBe ± 0.5 dBe ± 0.7 dBe
Ě¥G%$: ± 0.03 dBe ± 0.03 dBe ± 0.3 dBe ± 0.5 dBe ± 0.8 dBe
Minimum measurable frequency response (noise ¥G%$: ¥G%$: ¥G%$: ¥G%$: ¥G%$:
floor) 2, 3, 8, 9
Phase uncertainty DUT response – – – – –
(typical) 2, 4
Ě¥G%$:1 ± 3.5° ± 3.0° ± 2.7° ± 4.4° ± 6.0°
Ě¥G%$: ± 5.5° ± 3.5° ± 2.7° ± 4.9° ± 7.5°
Group delay uncertainty Derived from phase uncertainty, see section “Group delay uncertainty”.
Example: ± 2.0° ĵ ± 8 ps (1 GHz aperture)
'87UHVSRQVHPD[¥G%$:
2. For +4 dBm average output power from LCA optical output.
3. IFBW = 10 Hz.
([FHSWSKDVHZUDSDOLDVLQJ([DPSOH$'87JURXSGHOD\RIQVPFDEOHOHQJWKUHTXLUHVDIUHTXHQF\VWHSVL]HRIę*+]WRDYRLGSKDVHZUDSV
Excluding a constant group delay offset of < ± 0.3 ns typical. (Cable length uncertainty < ± 0.06 m). A constant group delay offset leads to a phase
RIIVHWđŠ ¡§đ*'§IPRGLQGHJ
5. After CW responsivity and user calibration with external source.
6. Requires option -100 or -102.
7. Typical with internal source.
8. Average value over frequency range.
9. In reverse coupler configuration, for normal configuration add typically 35 dB (0.05 GHz to 0.2 GHz), 12 dB (0.2 GHz to 0.7 GHz), 8 dB (> 0.7 GHz).
Ě¥G%$: ± 0.02 dBe ± 0.02 dBe ± 0.02 dBe ± 0.5 dBe ± 0.7 dBe
Ě¥G%$: ± 0.02 dBe ± 0.02 dBe ± 0.06 dBe ± 0.5 dBe ± 0.8 dBe
Minimum measurable frequency response (noise ¥G%$: ¥G%$: ¥G%$: ¥G%$: ¥G%$:
floor) 2, 3, 8, 9
Phase uncertainty DUT response – – – – –
(typical) 2, 4
Ě¥G%$:1 ± 3.5° ± 3.0° ± 2.4° ± 3.2° ± 5.0°
Group delay uncertainty Derived from phase uncertainty, see section “Group delay uncertainty”.
Example: ± 2.0° ĵ ± 8 ps (1 GHz aperture)
)RU'87UHVSRQVHPD[¥G%$:
2. For +5 dBm average output power from LCA optical output.
3. IFBW = 10 Hz.
([FHSWSKDVHZUDSDOLDVLQJ([DPSOH$'87JURXSGHOD\RIQVPFDEOHOHQJWKUHTXLUHVDIUHTXHQF\VWHSVL]HRIę*+]WRDYRLGSKDVHZUDSV
Excluding a constant group delay offset of < ± 0.3 ns typical. (Cable length uncertainty < ± 0.06 m). A constant group delay offset leads to a phase
RIIVHWđŠ ¡§đ*'§IPRGLQGHJ
5. After CW responsivity and user calibration with external source.
6. Requires option -101 or -102.
7. Typical with internal source.
8. Average value over frequency range.
9. In reverse coupler configuration, for normal configuration add typically 35 dB (0.05 GHz to 0.2 GHz), 12 dB (0.2 GHz to 0.7 GHz), 8 dB (> 0.7 GHz).
Ě¥G%H ± 0.03 dBe ±0.03 dBe ± 0.1 dBe ± 0.4 dBe ± 0.8 dBe
Ě¥G%R
Ě¥G%H ± 0.03 dBe ±0.03 dBe ± 0.1 dBe ± 1 dBe ± 1.5 dBe
Ě¥G%R
Minimum measurable frequency response (noise ¥G%HW\SLFDO ¥G%H ¥G%H ¥G%H ¥G%H
floor) 1, 2, 7, 8 ¥G%R ¥G%R ¥G%R ¥G%R ¥G%R
Phase uncertainty DUT response – – – – –
(typical) 2, 3
Ě¥G%H4 ± 3.5° ± 3.0° ± 2.2° ± 2.7° ± 3.5°
Ě¥G%R
Ě¥G%H ± 5.5° ± 3.5° ± 2.2° ± 3.3° ± 4.0°
Ě¥G%R
Group delay uncertainty Derived from phase uncertainty, see section “Group delay uncertainty”.
Example: ± 2.0° ĵ ± 8 ps (1 GHz aperture)
1. IFBW = 10 Hz.
2. For +4 dBm average output power from LCA optical output.
([FHSWSKDVHZUDSDOLDVLQJ([DPSOH$'87JURXSGHOD\RIQVPFDEOHOHQJWKUHTXLUHVDIUHTXHQF\VWHSVL]HRIę*+]WRDYRLGSKDVHZUDSV
4. For DUT response max. +6 dBe (+3 dBo) gain.
5. After CW responsivity and user calibration with external source.
6. Requires option -100 or -102.
7. Average value over frequency range.
8. In reverse coupler configuration, for normal configuration add typically 35 dB (0.05 GHz to 0.2 GHz), 12 dB (0.2 GHz to 0.7 GHz), 8 dB (> 0.7 GHz).
Ě¥G%H ± 0.02 dBe ± 0.02 dBe ± 0.02 dBe ± 0.3 dBe ± 0.5 dBe
Ě¥G%R
Ě¥G%H ± 0.02 dBe ± 0.02 dBe ± 0.1 dBe ±1.0 dBe ± 2.0 dBe
Ě¥G%R
Minimum measurable frequency response (noise ¥G%HW\SLFDO ¥G%H ¥G%H ¥G%H ¥G%H
floor) 1, 2, 7, 8 ¥G%R ¥G%R ¥G%R ¥G%R ¥G%R
Phase uncertainty DUT response – – – – –
(typical) 2, 3
Ě¥G%H4 ± 3.5° ± 3.0° ± 2.2° ± 2.6° ± 3.0°
Ě¥G%R
Ě¥G%H ± 5.5° ± 3.5° ± 2.2° ± 3.0° ± 3.5°
Ě¥G%R
Group delay uncertainty Derived from phase uncertainty, see section “Group delay uncertainty”.
Example: ± 2.0° ĵ ± 8 ps (1 GHz aperture)
1. IFBW = 10 Hz.
2. For +5 dBm average output power from LCA optical output.
([FHSWSKDVHZUDSDOLDVLQJ([DPSOH$'87JURXSGHOD\RIQVPFDEOHOHQJWKUHTXLUHVDIUHTXHQF\VWHSVL]HRIę*+]WRDYRLGSKDVHZUDSV
4. For DUT response max. +6 dBe (+3 dBo) gain.
5. After CW responsivity and user calibration with external source.
6. Requires option 101 or -102.
7. Average value over frequency range.
8. In reverse coupler configuration, for normal configuration add typically 35 dB (0.05 GHz to 0.2 GHz), 12 dB (0.2 GHz to 0.7 GHz), 8 dB (> 0.7 GHz).
All specifications of the 1$%1$%1$% option 200, 201, 219, 400, 401, or 419 Network Analyzer
apply depending on selected LCA option -x4z, -x5z, -x7z. Please see the corresponding Network Analyzer data sheet
and User’s Guide.
Group delay
Group delay is computed by measuring the phase change within a specified aperture (for aperture see below):
Aperture
Determined by the frequency span and the number of points per sweep:
GD Range
The maximum group delay is limited to measuring no more than ±180 degrees of phase change within the selected
aperture (see Equation 1).
General Characteristics
ShiSSLQJ&ontents
43.5/50 GHz LCA 67 GHz LCA
N5224/5A% NA according to ordered option N5227A% NA according to ordered option
3x 85133-600 f-m flexible test port MW cable (4-port network analyzer) or 3x N4697-6000 f-m flexible test port MW cable (4-port network analyzer) or
2x 85133-600 f-m flexible test port MW cable (2-port network analyzer) 2x N4697-600 f-m flexible test port MW cable (2-port network analyzer)
Connectivity
LCA electrical input LCA electrical output
1.85 mm (m) 1.85 mm (m)
LCA optical input 1 LCA optical input 2
PVLQJOHPRGHDQJOHGZLWK.H\VLJKWXQLYHUVDODGDSWHU PVLQJOHPRGHDQJOHGZLWK.H\VLJKWXQLYHUVDODGDSWHU
LCA external source input (Option -050 only) LCA optical output
PSRODUL]DWLRQPDLQWDLQLQJVLQJOHPRGHDQJOHGZLWK.H\VLJKWXQLYHUVDO PVLQJOHPRGHDQJOHGZLWK.H\VLJKWXQLYHUVDODGDSWHU
adapter
Humidity
15% to 80% relative humidity, non-condensing
Altitude (Operating)
0 ... 2000 m
g p
12 Regulatory and
Warranty Information
Regulatory Information / 166
Declaration of Conformity / 167
Warranty / 168
Phone or Fax / 170
Keysight Online Information / 171
12 Regulatory and Warranty Information
Regulatory Information
LpA < 50 dB
Operator position
Normal operation
Declaration of Conformity
Warranty
All system warranties and support agreements are dependent upon the
integrity of the Keysight N4373E Lightwave Component Analyzer. Any
modification of the system software or hardware will terminate any
obligation that Keysight Technologies may have to the purchaser. Please
contact your local Keysight field engineer before embarking in any
changes to the system.
System
Get the latest information on the products and applications you select.
www.keysight.com/find/emailupdates
myKeysight
Quickly choose and use your test equipment solutions with confidence.
www.keysight.com/find/mykeysight
Keysight Open
Phone or Fax
United States:
(tel) 800 829 4444
(fax) 800 829 4433
Canada:
(tel) 877 894 4414
(fax) 800 746 4866
China:
(tel) 800 810 0189
(fax) 800 820 2816
Europe:
(tel) 31 20 547 2111
Japan:
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Korea:
(tel) (080) 769 0800
(fax) (080) 769 0900
Latin America:
(tel) (305) 269 7500
Taiwan:
(tel) 0800 047 866
(fax) 0800 286 331
Other Asia Pacific Countries: (tel) (65) 6375 8100 (fax) (65) 6755
0042Email: [email protected]
13 Ordering Information
Ordering Information / 174
Power Cords / 176
13 Ordering Information
Ordering Information
The 1( consists of an optical test set and an electrical network analyzer which are mechanically connected. To protect
your network analyzer investment, Keysight offers the integration of an already owned PNA/PNA-X with the optical test set
as listed below.
2.For detailed ordering requirements for software optionsrefer to the LCA configuration guide.
. )RULQIRUPDWLRQDERXWRWKHUVRIWZDUHRSWLRQVUHIHUWRWKHQHWZRUNDQDO\]HUFRQILJXUDWLRQJXLGH
Power Cords
Power Cords
OE measurements, 103
W
F OO measurements, 103
Optical Connections, 43 Warranty, 168
Fuse, 133 Optical CW Operation, 43
Opto-electric Devices, 104
www.keysight.com