Icap Q: Preinstallation Requirements Guide
Icap Q: Preinstallation Requirements Guide
Icap Q: Preinstallation Requirements Guide
iCAP Q
Preinstallation Requirements
Guide
Revision A - 1288020
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use
in the product operation. This document is copyright protected and any reproduction of the whole or
any part of this document is strictly prohibited, except with the written authorization of Thermo
Fisher Scientific Inc.
Assistance
For technical support and ordering information, visit us on the Web:
www.thermoscientific.com/ms
www.unitylabservice.com
Thermo Fisher Scientific recommends that you register with the site as
soon as the instrument is installed. To register, visit
register.thermo-bremen.com/form/cis and fill in the registration form.
Once your registration has been finalized, you will receive confirmation
by e-mail.
28199 Bremen
Germany
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) i
Contacting Us
Suggestions to the Manual
ii iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Contents
Chapter 1 Introduction................................................................................1-1
Chapter 2 Site...............................................................................................2-1
Entrance Requirements ..................................................... 2-2
Dimensions and Weights .................................................. 2-2
Placing the MS System...................................................... 2-5
Floor Space for the System ............................................. 2-5
Placing the Fore Vacuum Pump..................................... 2-6
Placing the Data System................................................. 2-8
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) iii
Contents
iv iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Chapter 1 Introduction
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 1-1
Introduction
1-2 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Chapter 2 Site
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 2-1
Site
Entrance Requirements
Entrance Requirements
To allow moving a packed iCAP Q instrument, the entrance to your
facility and the width of all hallways, elevators, etc., should have a
minimum width of 89 cm (35.1 in). Also allow additional room for
maneuvering the system around corners, into elevators, or through
doorways.
2-2 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Site
Entrance Requirements
1030
948
707
748 771
665
772
Top View
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 2-3
Site
Entrance Requirements
1030
948
615
707
1363 771
665
772
Top View
Ensure that the Thermo Fisher Scientific field service engineer has access
to the left and right of the instrument.
2-4 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Site
Placing the MS System
On the left side panel, free access to the mains switch and circuit breaker
is needed to allow shutting off the instrument in an emergency at all
times.
Allow at least 70 to 100 cm of clear space left of the system for clearance
of the gas lines, electrical connections, as well as for the exhaust line and
vacuum hose of the fore vacuum pump.
6
5
4
772 1
15
1000
767
748
2
700 - 1000 1200
3
Labeled Components: 1=monitor, 2=mouse, 3=keyboard, 4=the iCAP Q instrument, 5=wall, 6=standard workbench
Figure 2-3. Required floor space for the iCAP Q instrument without autosampler
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 2-5
Site
Placing the MS System
Figure 2-4 shows the typical required floor space for the system with
autosampler. All values are given in mm:
6 7
5
1
4
15
1015
1000
773 2
1391 3
700 - 1000 1400
2100 - 2400
3500 - 3800
Labeled Components: 1=monitor, 2=mouse, 3=keyboard, 4=the iCAP Q instrument, 5=wall, 6=autosampler housing,
7=computer table
Figure 2-4. Required floor space for the iCAP Q instrument with autosampler
2-6 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Site
Placing the MS System
Install the fore vacuum pump on the floor beneath the workbench, see
Figure 2-5.
power
connection
exhaust
connection
vacuum
hose
noise reduction
cover
fore
vacuum
pump
If no space for the pump is available beneath the workbench, you can
place the pump near the left side of the bench. In this case, the left side
of the MS ideally aligns with the left side of the workbench.
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 2-7
Site
Placing the MS System
monitor
computer
communication
cables USB and
Ethernet
wall outlet for data system components
120 or 230 V AC
Table 2-2 lists space requirements and weights of the typical data system
hardware components. The actual values depend upon your equipment.
The USB communication cables between the computer and the mass
spectrometer must be no longer than 3 m each. The workbench that
holds the data system must be located next to the workbench that holds
the mass spectrometer. For a connection length of more than 3 m, an
active hub must be placed between the instrument and the computer.
The length of the USB cable between the iCAP Q instrument and the
2-8 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Site
Placing the MS System
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 2-9
Site
Placing the MS System
2-10 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Chapter 3 Operating Environment
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 3-1
Operating Environment
Temperature
Temperature
NOTICE After arrival the iCAP Q instrument must be stored as packed
by Thermo Fisher Scientific for at least 24 hours at an environmental
condition of 15 to 35 °C and a relative humidity of 20 to 80% before
connecting to power supply. ▲
3-2 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Operating Environment
Exhaust System
Exhaust System
An exhaust system is needed for the instrument to remove gases that
may contain ozone and other noxious substances. Two exhaust ports are
situated at the top of the instrument, see Figure 3-1.
The plasma exhaust 1 in Figure 3-1 is used to remove gases and heat
generated by the plasma. A flexible exhaust duct must be connected
during operation of the instrument.
NOTICE Thermo Fisher Scientific will only install your iCAP Q system
when an adequate exhaust system is present and functioning. Exhaust
gas venting must comply with all local environmental codes. ▲
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 3-3
Operating Environment
Exhaust System
The velocity v at the end of the duct with an inner diameter (d) of
63 mm connected to the exhaust port of the iCAP Q instrument can be
measured with an anemometer (values in m/s) and can then be
converted to an exhaust flow f (m3/h) by the following equation:
d 2
f = --- v 3600
2
0.063 m 2 m3
f = 3.14 ------------------- 6 ---- 3600 67 ----
m
2 s h
NOTICE The extraction system must be tested for leakage before you
connect it to the iCAP Q instrument. ▲
3-4 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Operating Environment
Heat Generation and Dissipation of System
For estimated values for the average heat generation of the mass
spectrometer and other heat sources during analysis, see Table 3-3.
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 3-5
Operating Environment
Heat Generation and Dissipation of System
Table 3-4. Heat dissipation for a typical iCAP Q system and chiller
Module Heat dissipation [W] at 50 Hz
iCAP Q instrument, with active plasma 800
exhaust removal
iCAP Q instrument, with active plasma 450
and heat exhaust removal
Fore vacuum pump approx. 1100
Recirculating chiller 2200 (2500 at 60 Hz)
Monitor 35
Computer 470
Total approx. 4.6 kW
with active plasma exhaust removal (4.9 kW at 60 Hz)
Total approx. 4.3 kW
with active plasma and heat exhaust (4.6 kW at 60 Hz)
removal
3-6 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Operating Environment
Heat Generation and Dissipation of System
Table 3-5. Heat dissipation with house water supply or chiller not placed
in laboratory
Module Heat dissipation [W] at 50 Hz
iCAP Q instrument, with active plasma 800
exhaust removal
iCAP Q instrument, with active plasma 450
and heat exhaust removal
(chiller connected to house water supply -
or not placed in laboratory)
Fore vacuum pump approx. 1100
Monitor 35
Computer 470
Total approx. 2.4 kW
with active plasma exhaust removal
Total approx. 2.1 kW
with active plasma and heat exhaust
removal
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 3-7
Operating Environment
Heat Generation and Dissipation of System
Table 3-6. Heat dissipation of the iCAP Q instrument and computer only
Module Heat dissipation [W] at 50 Hz
iCAP Q instrument, with active plasma 800
exhaust removal
iCAP Q instrument, with active plasma 450
and heat exhaust removal
(chiller connected to house water supply -
or not placed in laboratory)
(fore vacuum pump not in laboratory) -
Monitor 35
Computer 470
Total approx. 1.3 kW
with active plasma exhaust removal
Total approx. 1.0 kW
with active plasma and heat exhaust
removal
3-8 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Operating Environment
Air Conditioning
Air Conditioning
The air conditioning system must be capable of maintaining a constant
temperature as specified in “Temperature” on page 3-2 in the immediate
vicinity of the system.
For the approximate load for the air conditioning in a room containing
one instrument see “Heat Generation and Dissipation of System” on
page 3-5.
NOTICE Do not locate the iCAP Q instrument under an air duct, near
windows, or near heating and cooling sources. ▲
Humidity
The relative humidity of the operating environment must be between
20 and 80%, with non-condensing and non-corrosive atmosphere. It is
recommended that your laboratory be equipped with a temperature and
humidity monitor to insure that your laboratory is always within the
required temperature and humidity specifications.
Altitude
The iCAP Q instrument is designed for indoor use at an altitude of up
to 2000 m (6562 ft) above sea level. For altitudes above 2000 m contact
Thermo Fisher Scientific.
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 3-9
Operating Environment
Air Conditioning
Vibration
Floors must be free of vibration caused, for example, by equipment in
adjoining locations.
NOTICE The fore vacuum pump must not have any mechanical contact
to the mass spectrometer with exception of the vacuum hose during
operation. The vibration of the fore vacuum pump might impede the
performance of the instrument. Install the fore vacuum pump on the
floor beneath the mass spectrometer. Do not install the fore vacuum
pump near the system on the workbench. ▲
The mean value for the emission sound pressure level of the iCAP Q
instrument with and without noise reduction cover on the fore vacuum
pump are listed in Table 3-7:
Table 3-7. Mean values for emission sound pressure of the iCAP Q
instrument
Emission sound specification of the iCAP Q instrument Mean value
Fore vacuum pump without noise reduction cover 55.9 dB(A)
Fore vacuum pump with noise reduction cover 53.1 dB(A)
Radio Frequencies
The iCAP Q system withstands the electric fields and frequency ranges
listed in Table 3-8 without any influence to operation:
3-10 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Operating Environment
Magnetic Fields
Magnetic Fields
The instrument site must be free of interfering magnetic fields. The
maximum acceptable field amplitude (AC) for any frequency is
5 x 10-6 T.
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 3-11
Operating Environment
Magnetic Fields
3-12 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Chapter 4 Line Power
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 4-1
Line Power
Electrical Power Requirements
4-2 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Line Power
Delta-To-Y Conversion Transformer
(Delta configuration)
(Y configuration)
400 V AC between phases
shield
230 V AC between phases and neutral
Line OUT
Line IN
L1
L2
L3
neutral
ground (earth)
transformer (buck/boost)
15 KVA, transformation from delta to Y
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 4-3
Line Power
Power Connections
Power Connections
The electrical wall outlet for the main power of the iCAP Q instrument
should be located at the wall near the intended location of the
instrument, ideally at the left side of the instrument.
The mains power cable for the iCAP Q mass spectrometer is a 5-wire
cable, length 5 m. A CECON plug, 5 poles, certified for 400 V, 16 A, is
mounted to the cable, see lower right photo in Figure 4-2. This power
cable and plug is already mounted to the instrument.
The lower left photo in Figure 4-2 shows the wall receptacle required for
the iCAP Q instrument. The receptacle is provided by Thermo Fisher
Scientific as part of the Preinstallation Kit.
Figure 4-2. Wall mounting of receptacle and plug of mains power cable
Local codes in your area may require another type of plug and receptacle
to be installed. The local Thermo Fisher Scientific organization (or
authorized dealer) will help you with further questions.
Power cables and connectors for the options are standard equipment
delivered by the manufacturers.
4-4 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Line Power
Power Connections
The iCAP Q instrument provides a power output socket for the fore
vacuum pump and a power output socket for the optionally shipped the
autosampler.
At the left side, the iCAP Q instrument provides a power output socket
for the fore vacuum pump of the instrument, see 2 in Figure 4-3:
Power cables and connectors for the options are standard equipment
delivered by the manufacturers.
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 4-5
Line Power
Auxiliary Wall Outlets
NOTICE Single-phase auxiliary wall outlets should use the same ground
as the instrument. ▲
Quality of Power
The iCAP Q instrument complies with the requirements listed in
Table 4-3:
4-6 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Chapter 5 Cooling Water
Figure 5-1. Cooling water connections at the left of the iCAP Q system
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 5-1
Cooling Water
Technical Data Recirculating Chiller
water in
iCAP Q
water out
water in Chiller
water out
Table 5-2 lists the technical specifications for the NESLAB ThermoFlex
2500 recirculation chiller optionally shipped with the iCAP Q system:
Table 5-2. Chiller specifications
Chiller component Specification
Cooling capacity 2.2 kW (50 Hz), 2.5 kW (60 Hz)
(at 20 °C ambient temperature)
Ambient Temperature 10 to 40 °C
Water temperature 5 to 40 °C
Stability of temperature ±0.1 °C
regulation
Reservoir Volume 7.2 L
Flow rate
Turbine pump (60 Hz) 13.2 L/min at 0.41 MPa (4.1 bar)
Turbine pump (50 Hz) 9.5 L/min at 0.41 MPa (4.1 bar)
Unit dimensions (H × W × L) 73.6 cm × 43.6 cm × 67.3 cm
Unit weight 79.6 kg
5-2 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Chapter 6 Gas Supply
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 6-1
Gas Supply
Gas Supply Connections
Labeled Components: 1=CCT 1 gas inlets (for example for standard CCT gas
He), 2=CCT 2 gas inlets (for example H2/He), 3=argon in, 4=additional gas 1
inlet, 5=additional gas 2 inlet
Figure 6-1. Gas connection at the left of the iCAP Q instrument
6-2 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Gas Supply
Gas Supply Connections
Table 6-2 lists the connectors for the gases to be provided by the
customer:
pressure
regulator
iCAP Q
(left side) pressure gauge
argon inlet
shut-off valve
gas source
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 6-3
Gas Supply
Gas Supply Connections
CCT Gas
Collision Cell Technology (CCT) allows the selective removal of
polyatomic interferences. If an iCAP Q model including CCT is
ordered (iCAP Q Qc or iCAP Q Qs), CCT operation at the system is
possible. Thermo Fisher Scientific supplies 2 × 3 meters of gas tubing
with the iCAP Q model including CCT.
• gas
Helium
For the iCAP Q model including CCT, a CCT installation test is
performed with helium gas to pressurize the CCT cell. See Table 6-3 for
details of the helium gas requirements:
6-4 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Gas Supply
Gas Supply Connections
Each country has a different standard outlet fitting for its gas supplies
and therefore each country will require a different regulator. Examples
for regulators are given by region in Table 6-4. If your region is not
listed, contact your local Thermo Fisher Scientific service office for
advice.
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 6-5
Gas Supply
Additional Gas
Additional Gas
Depending on the application, a variety of additional gases may be
employed. The use of mass flow controllers is required when working
with additional gases.
Oxygen
For the analysis of organic solvents the use of oxygen may be required.
The gas requirements are detailed in Table 6-5.
Helium
For laser ablation analysis, helium gas may be needed. The gas
requirements are detailed in Table 6-6.
Table 6-6. Additional gas requirements for helium
Gas Requirements Specification
Purity 99.996%
Maximum water content < 5 ppmv
Supply rate max. 1 L/min
Pressure regulator output 0.6 MPa (6 bar)
6-6 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Gas Supply
Additional Gas
Argon
For some applications, the use of additional argon gas is beneficial. The
gas requirements are detailed in Table 6-7.
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 6-7
Gas Supply
Additional Gas
6-8 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Chapter 7 Cleaning Agents
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 7-1
Cleaning Agents
7-2 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Chapter 8 General Preinstallation Information
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 8-1
General Preinstallation Information
Instrument Arrival
Instrument Arrival
When your lab site preparation is completed, and the system is
delivered, call your local Thermo Fisher Scientific office to arrange for
an installation date.
Transportation Risk
Transportation risk depends on the terms of delivery agreed. The terms
of shipment determine who has responsibility for filing a claim against
the carrier if the system is damaged in transit.
8-2 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
General Preinstallation Information
Installation
Installation
It is the policy of Thermo Fisher Scientific that the customer should not
unpack the system or accessory items prior to installation of the system.
NOTICE Consumables sent with the system are intended for use by the
Thermo Fisher Scientific field service engineer during the installation. ▲
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 8-3
General Preinstallation Information
Installation
Key Operator
Experience has shown that the maximum benefit can be derived from a
scientific instrument if there is one person, a key operator, who has
major responsibility for that instrument. It is recommended that you
designate a key operator to oversee the operation and maintenance of
the system in your laboratory. This person will also be the key figure in
the communication between your laboratory and Thermo Fisher
Scientific.
NOTICE Do not plan to use your new system for sample analysis until
the installation is complete and the Acceptance Form has been signed. ▲
Germany
Preventive Maintenance
Routine and preventive maintenance of mass spectrometer is in the
responsibility of the customer. Included in this category are replacement
of the cone, exchange of pump oil, replacement of filters, etc. on a
regular basis. Refer also to the manufacturers manuals delivered with the
instrument – especially for the maintenance of mechanical pumps and
turbopumps.
8-4 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
General Preinstallation Information
Installation
Operating Environment
These general specifications for your the operating environment help
ensuring continued high performance of the system.
Particulate Matter
Particulate matter may contaminate the samples and the ion source and
may limit the background level of the instrument.
The air in your laboratory must not have excessive dust, smoke, or other
particulate matter. For reference, the air should contain fewer than
3500000 particles per cubic meter (100000 particles per cubic foot) in
excess of 5 μm.
Dust can clog the air filters, causing a reduction in airflow around
electronic components. Dust will also form a layer on electronic
components that will act as an insulating blanket and thus reduce the
transfer of heat from the components to the surrounding air.
Quality of Power
The quality of power supplied to your system is very important. The
quality of line voltage must be stable and within the specifications listed
in this manual. The line voltage must be free of fluctuations due to slow
changes in the average voltage, surges, sags, or transients.
• Sags and surges are sudden changes in average RMS voltage level,
with typical durations between 50 μs and 2 s.
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 8-5
General Preinstallation Information
Installation
Technical Assistance
8-6 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
General Preinstallation Information
Installation
Electrostatic Discharge
Electrostatic discharge (ESD) can damage the electronic components of
your MS system. Thermo Scientific instruments are designed to
withstand electrostatic discharges (ESD) up to 4 kV (air discharge) and
4 kV (contact discharge) with all panels in place. However, if the panels
are removed and the PCBs are handled without proper precautions, the
electronic components might be damaged or fail prematurely. Static
electricity can develop in a variety of ways. A few examples of how
electrostatic charge can develop are as follows:
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) 8-7
General Preinstallation Information
Installation
8-8 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Index
A F
acceptance form 8-4 filtering solvents 7-1
advanced training courses 8-4 floor space
air conditioning 3-9 with autosampler 2-6
airborne noise emission 3-10 without autosampler 2-5
fore vacuum pump
connecting 2-7
exhaust 2-7
C heat generation 3-5
cleaning agents 7-1 location 2-6
collision cell technology gas 6-4 power connection 2-7
computer
heat generation 3-5
placing the data system 2-8
H
connecting
cooling water 5-1 heat dissipation
fore vacuum pump 2-7 complete system 3-6
cooling capacity 5-2 without chiller 3-7
without chiller or fore vacuum pump 3-8
cooling water
connections 5-1 heat generation
computer 3-5
specifications 5-1
fore vacuum pump 3-5
CTT gas 6-4
monitor 3-5
system 3-5
water chiller 3-5
D humidity 3-9
data system 2-8
dimensions 2-8
placing 2-5 K
workbench 2-8
key operator 8-4
dimensions and weights 2-2
dimensions, of packed units 2-2, 2-8
distilled water
as cleaning agent 7-1 L
as cooling fluid 5-1 laboratory
floor space 2-5
temperature 3-2
E
electrostatic discharge 8-7
emission sound pressure level 3-10 M
entrance requirements 2-2 magnetic fields 3-11
exhaust monitor
fore vacuum pump 2-7 heat generation 3-5
port location 3-3 location 2-8
system 3-3
Thermo Fisher Scientific iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) I-1
Index: N–W
N T
noise emission 3-10 technical assistance 8-6
noise reduction cover 2-7 temperature
cooling water 5-1
laboratory 3-2
training courses 8-4
O
transportation risk 8-2
operator responsibilities 2-1, 4-1
V
P
vacuum hose 2-7
particulate matter, in air 3-10
vibration 3-10
pH value cooling water 5-1
power
electrical quality 4-6
fore vacuum pump 2-7 W
outlets 2-7 wall outlet
quality 4-6 chiller 4-6
pressure computer 4-6
cooling water 5-1 water chiller
two-stage gas regulator 6-4 heat generation 3-5
preventive maintenance 8-4 technical data 5-2
weight
equipment 2-2, 2-8
instrument 2-2
Q
quality electrical power 4-6
R
radio frequencies 3-10
recirculating chiller 5-2
responsibilities operator 2-1, 4-1
S
solid residual in cooling water 5-1
solvents for cleaning 7-1
specifications 5-2
supply rate cooling water 5-1
system
dimensions and weights 2-2
heat generation 3-5
I-2 iCAP Q Preinstallation Requirements Guide (P/N 1288020, Revision A) Thermo Fisher Scientific
Thermo Fisher Scientific Inc.
81 Wyman Street
P.O. Box 9046
Waltham, Massachussetts 02454-9046
United States
www.thermoscientific.com