IMPORTANT
The information contained herein is based on the experience and knowledge relating to
the subject matter gained by Ortho Clinical Diagnostics prior to publication.
Always consult the current Product Instructions for Use (IFU) for complete directions.
Ortho Clinical Diagnostics reserves the right to change this information without notice.
All product and company names are trademarks™ or registered® trademarks of their respective
holders.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – iii
4.5 Host Query .................................................................................................................................................... 26
4.5.1 Basic Functionality ................................................................................................................................... 26
4.5.2 Host Query Timer ..................................................................................................................................... 26
4.5.3 Sending and Receiving Messages ............................................................................................................. 27
4.5.4 Host Query Priority .................................................................................................................................. 27
4.6 Inventory Query............................................................................................................................................. 28
4.7 Asynchronous Notifications .......................................................................................................................... 29
4.7.1 Automation Status Updates ...................................................................................................................... 29
4.7.2 Error Messages......................................................................................................................................... 29
4.7.3 Download Messages ................................................................................................................................. 29
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
iv – 2017-12-19
Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – v
5.5.14 Query Identification Details (QID)........................................................................................................... 88
5.5.15 Query Parameter Definition Details (QPD) ............................................................................................. 89
5.5.16 Response Control Parameter Details (RCP) ............................................................................................ 90
5.5.17 Specimen Container Details (SAC)........................................................................................................... 90
5.5.18 Specimen (SPM) ....................................................................................................................................... 92
5.5.19 Z-Segment Extended Results (ZER) .......................................................................................................... 94
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
vi – 2017-12-19
Laboratory Information System (LIS) Guide Introduction
1.1 Scope
This document contains all specifications to aid in developing communications between the
VITROS 5600 Integrated System, the VITROS® XT 7600 Integrated System, the VITROS
4600 Chemistry System, the VITROS® XT 3400 Chemistry System, or the VITROS 3600
Immunodiagnostic System and the Laboratory Computer. Supported communication protocols
include ASTM and HL7.
Change bars in the margins of the document pinpoint the exact locations of the most recent
changes. Deletions are not noted.
Issued Section Section Title Revision Details
Number
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 1
Introduction Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2 – 2017-12-19
Laboratory Information System (LIS) Guide Introduction
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 3
Introduction Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
4 – 2017-12-19
Laboratory Information System (LIS) Guide Introduction
5.4.4.1 HL7 OUL Unsolicited Added two examples for suppressed mean
Observation Message result.
Examples
5.4.5.1 HL7 QPB Host Query Message Added metering point type in the QPD
Example record example.
5.4.6.1 HL7 RSP Host Query Added metering point type in the QPD
Response Message Example record example.
5.4.6.2 HL7 RSP Host Query Added metering point type in the QPD
Response Message Example record example.
(No Program Found)
5.5.9 Observation Result (OBX) For field #s 5,6, and 8, added information in
the Notes column.
5.5.15 Query Parameter Definition Added metering reference point information
Details to field #10 Location, which is now
supported.
5.5.19 Z-Segment Extended Results For field #5, added "G" to the Notes
(ZER) column.
Appendix F Result Codes Updated the explanations of codes CE, EC,
M2, and RE.
Appendix G Sample Indices Flags Sent to Edited the name of this appendix and
the LIS included only the 2-character flags sent to
the LIS.
Previous Versions
Issued Section Revision Details
2008-08-29 4.3.2 Reworded: Explanation of attempt to edit a program in process (26).
4.3.3 Reworded: Note regarding sample programs in process.
4.3.3 Added: An explanation of a new sample program downloaded for a sample that
already has a program in process.
4.3.4 Changed: Reference to download message 27.
4.4 Reworded: Explanation of results records and categorization.
4.4 Added: Statement about qualitative text comments and result comment record.
5.1.7 Added: Reprocessing Type table.
5.2.2 Added: [C] Comment line in the Result Upload Message Structure.
5.2.2.1 Added: A second Results Upload Message Example.
5.3.6 Changed: #4 and #5 of ASTM C Comment Record.
5.3.8 Added: #8 Reprocessing Type to the ASTM M/X Extended Result Record.
5.4.4 Added: [NTE] line to HL7 OUL Unsolicited Observation Message Structure.
5.4.4.1 Added: A second example of HL7 OUL Unsolicited Observation Message.
5.5.7 Added: Information to #3 and #4 of Notes and Comments (NTE).
5.5.19 Added: #5 Reprocessing Type to Z-Segment Extended Results (ZER).
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 5
Introduction Laboratory Information System (LIS) Guide
1.3 Definitions
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
6 – 2017-12-19
Laboratory Information System (LIS) Guide Introduction
MLLP Minimal Low-Level Protocol commonly used for HL7 communication over a TCP/IP
network connection
MSA HL7 message acknowledge segment
MSH HL7 message header segment
NAK ASTM Negative Acknowledgement. Can be ASCII character 21.
NDS HL7 Notification Detail segment
NTE HL7 Notes and Comments segment
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 7
Introduction Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
8 – 2017-12-19
Laboratory Information System (LIS) Guide Introduction
Restricted Assay A test requiring more stringent handling than other assays. Assays that have some
parameters, defined on the Assay Data Disk (ADD), that cannot be changed.
RSP_ZOS HL7 Host Query Response Message. This is a proposed message format for the upcoming
HL7 release.
RTS ASTM “Request To Send” RS-232 control signal
Rx Receive(r).
SAC HL7 Specimen Container Detail Segment
Sample The discrete portion of a body fluid or tissue taken for examination, study, or analysis of one
or more quantities or characteristics to determine the character of the whole
Sample Program Test(s) requested on sample.
SI Sample Indices (Sample Index)
Specimen HL7 term for Sample.
Specimen ID Sample ID. Can be manually entered from the GUI or Bar coded
SPM HL7 Specimen segment
STX ASTM Start of Text. ASCII character 2. This control character denotes the beginning of the
data or “text” information block.
System The software and hardware that together make up the Analyzer.
Table The table attribute of the data field definition specifies the HL7 identifier for a set of coded
values
TCP/IP TCP/IP is a network protocol designed around a simple four-layer scheme. It does omit
some features found under the OSI model. Also it combines the features of some adjacent
OSI layers and splits other layers apart. The four network layers defined by TCP/IP model
are as follows.
Layer 1 - Link
Layer 2 - Network
Layer 3 - Transport
Layer 4 - Application
Trigger events An event in the real world of healthcare creates the need for data to flow among systems.
The real-world event is called the trigger event. This term is used in the context of HL7.
Tx Transmit, Transmission, or Send(er).
Upload Data sent from the System to the Lab Computer
UTF8 Multi byte character set
Z Segment HL7 All message types trigger event codes, and segment ID codes beginning with Z are
reserved for locally defined messages.
ZER Custom HL7 segment representing the extended results
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 9
General Description Laboratory Information System (LIS) Guide
1.4 References
These documents form the basis for many of the specifications in this LIS Guide. The first two
listed are particularly important.
NCCLS LIS1-A Standard Specification for Low-Level Protocol to Transfer Messages Between
(Formerly ASTM Clinical Laboratory Instruments and Computer Systems
E1381-02)
NCCLS LIS2-A Standard Specification for Transferring Information Between Clinical
(Formerly ASTM Laboratory Instruments and Computer Systems
E1394-97)
N/A HL7 2.5 Specifications, 2003
N/A HL7 2.3 Implementation Guidelines
IEEE 802.3i Ethernet Standard 10BaseT
IEEE 802.3u Ethernet Standard 100BaseT
2. General Description
Laboratory Information System refers to the interface that communicates between the System and
the Laboratory Computer System (Lab Computer). This communication can occur using either the
HL7 or ASTM protocols.
With the ASTM protocol, communication can occur over either a serial line from the System to the
Lab Computer or through the user’s network. With the HL7 protocol, the communication only
occurs over the user's network. The communications are bi-directional and include downloading
sample programs from the Lab Computer to the System, uploading result records from the System
to the Lab Computer, querying sample programs from the System to the Lab Computer, querying
inventory from the Lab Computer to the System, and sending asynchronous notifications from the
System to the Lab Computer. This document covers each of these aspects from the perspective of
the System.
The specifications in this document are grouped into the following sections:
• Communications – This includes the physical interfaces and specific communications
processing called for by the HL7 and ASTM protocols.
• Record Processing - This specifies the content of the messages (e.g., sample programs, results
records patient information) and their components to be implemented in this application. The
format and hierarchy are specified in Section 5.2 ASTM Messages and Section 5.4 HL7
Messages.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
10 – 2017-12-19
Laboratory Information System (LIS) Guide General Description
2.1 Functions
• Provide a method to receive remotely programmed patient and quality control sample
programs from another system. (Download from Lab Computer)
• Provide a method to send patient results, quality control results, patient and physician
demographics to another system. (Upload Results to Lab Computer)
• Provide the ability to request sample programs from another system in order to process
sample. (Host Query)
• Provide indication to the operator the status of communication between the system and the
Lab Computer. (Status)
• Provide the ability to process requests for onboard inventory levels. (Inventory Query)
• Provide a method to send Analyzer status information to another system. (Error
Messages/Automation Status Updates/Download Messages)
The following overview shows the general way the interactions occur between the LIS and System.
There are other modes including error scenarios that are further detailed in the specifications.
• Sample programs containing specimen ID, patient information, and tests to be run are entered
into the Lab Computer.
• The Lab Computer transfers the sample program and patient information to the System.
• Information is entered in the System's Sample Database.
• The sample is presented at metering.
• The Sample Program database is queried for a sample program that matches the specimen ID
or the tray cup position.
• The program is found and the sample is metered.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 11
General Description Laboratory Information System (LIS) Guide
• The System posts a condition code (for some task other than LCI).
• The System transfers information about the condition code (severity, mod-err, short text).
The System software is compatible with ASTM specifications LIS1-A and LIS2-A.
The System software is compatible with HL7 v2.5 specifications, chapters 2, 3, 4, 7 and 13. This
HL7 version does not include provisions for Host Query functionality so a custom Z segment is
used.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
12 – 2017-12-19
Laboratory Information System (LIS) Guide Communication Specifications
• The system’s firewall needs to be configured to allow communications to/from the LIS when
using TCP/IP.
3. Communication Specifications
The following specifications are for the layers responsible for transporting messages between the
System and the LIS. The content of these messages is not relevant for these layers, which only
ensure that messages are transferred in an error-free manner between systems.
The System acts as a sender
• When a host query is initiated
• When results need to be uploaded to the Lab Computer
• When sending an asynchronous message
• When responding to an inventory query
The System acts as a receiver
• When the Lab Computer responds to a host query
• When sample programs are to be downloaded
• When sending an inventory query
Upon System startup, the LCI module tests communications.
If a connection to an LIS cannot be established, the LCI communicates an error condition to the
operator.
Note: See Appendix I Mechanical/Electrical Interfaces for details about these interfaces.
The ASTM and HL7 protocols support the use of TCP/IP communication for the physical
transport layer. The two protocols share the same specifications for this layer.
The System establishes a connection to the Lab Computer at a configured IP address and port.
Note: This interface is bidirectional but only initiated by the System.
The System
• Uses 1 start bit
• Supports 1 or 2 stop bits
• Supports EVEN, ODD, and NONE parity
• Uses 8 data bits
• Supports the following baud rates:
• 9600
• 19200
• 38400
• 57600
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 13
Communication Specifications Laboratory Information System (LIS) Guide
• 115200
Note: All referenced sections can be found in Standard Specification for Low-Level Protocol to
Transfer Messages Between Clinical Laboratory Instruments and Computer Systems,
NCCLS LIS1-A (Formerly ASTM E1381-02).
The ASTM layers involved in transferring data from the System to the Lab Computer and the Lab
Computer to the System are divided into these components:
• Physical Layer - Comprised of the actual hardware and software configuration used to
communicate between the two systems. For serial communications, this corresponds to
Section 5. For Ethernet communications, this corresponds to Section 7 and 8.2.1.1. For
specifications, refer to the appropriate Interface (Ethernet or Serial).
• Datalink Layer - Responsible for the logical data frames comprised of the raw data exchanged
with the physical layer. It ensures that packets can be transferred in an error-free manner
between machines. For serial communications this corresponds to the Datalink layer specified
in Section 6. For Ethernet communications, this corresponds to Section 8.
The System supports a configurable maximum frame size that supersedes values specified in
Standard Specification for Low-Level Protocol to Transfer Messages Between Clinical Laboratory
Instruments and Computer Systems, NCCLS LIS1-A (Formerly ASTM E1381-02).
Note: All referenced sections can be found in Standard Specification for Low-Level Protocol to
Transfer Messages Between Clinical Laboratory Instruments and Computer Systems,
NCCLS LIS1-A (Formerly ASTM E1381-02).
The System supports the Establishment Phase (Link Connection) as specified in Section 6.2. This
includes establishment and contention.
The System supports the Transfer Phase as specified in Section 6.3. This includes:
• Frame Format, Section 6.3.1
• Frame Numbering, Section 6.3.2
• Frame Checksums, Section 6.3.3
• Frame Acknowledgements, Section 6.3.4
• Frame Receiver Interrupts, Section 6.3.5
The System supports the Termination Phase (Link Release) as specified in, Section 6.4.
The System supports Error Recovery as specified in, Section 6.5. This includes:
• Detecting and handling defective frames, Section 6.5.1
• Timeouts, Section 6.5.2
The System supports the Restricted Message Characters requirement specified in Section 6.6.
Note: All referenced sections can be found in Standard Specification for Low-Level Protocol to
Transfer Messages Between Clinical Laboratory Instruments and Computer Systems,
NCCLS LIS1-A (Formerly ASTM E1381-02).
The System supports the Establishment Phase (Link Connection) as specified in Section 8.2. This
includes establishment and contention.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
14 – 2017-12-19
Laboratory Information System (LIS) Guide Communication Specifications
The System supports the Transfer Phase as specified in Section 8.3. This includes:
• Frame Format, Section 8.3.1
• Frame Numbering, Section 8.3.2
• Frame Checksums, Section 8.3.3
• Frame Acknowledgements, Section 8.3.4
• Frame Receiver Interrupts, Section 8.3.5
The System supports the Termination Phase (Link Release) as specified in Section 8.4
The System supports Error Recovery as specified in Section 8.5. This includes:
• Detecting and handling defective frames, Section 8.5.1
• Timeouts, Section 8.5.2
The System supports the Restricted Message Characters requirement specified in Section 8.6.
This layer is responsible for implementing the MLLP protocol. This protocol, defined by
HL7 2.3.1 standard, is essentially a half-duplex protocol, where a new message is sent only after
receiving the acknowledgement for the previous message. The System implements MLLP as
described in HL7 2.3 Implementation Guidelines. The System supports Error Recovery as
specified in Section C.2.6. This includes Connect Retries, Section C.2.6.1; Connect Pause, Section
C.2.6.3; and Receive Timeout Errors, Section C.2.6.2.
Every HL7 message is enclosed by special characters to form a block formatted as
<SB>dddd<EB><CR>, where:
<SB> - start block character (1 byte), ASCII VT = 0x0B
dddd - Data (variable number of bytes). This is the HL7 data content of the block. The data
can contain any single-byte values greater than 0x1F and the ASCII carriage return character,
<CR>.
<EB> - end block character (1 byte), ASCII FS = 0x1C
<CR> - carriage return (1 byte) = 0x0D
Messages received with incorrect delimitation characters are ignored.
The System
• Supports a timeout of 30 seconds. This timeout is used by the connect and receive timeout.
• Supports a retry count of 5. The value is used by the connect retry.
• Supports a retry action pause of 0 seconds.
• Accepts messages up to the configured frame size (in bytes). Messages longer than this value
are treated as a timeout and logged as a bad frame.
• Goes into a neutral state after attempting to reconnect to the LIS a configured number of times
(5 times), pauses a configured amount of time (0 seconds), and then waits a configured
timeout for each connection attempt (30 seconds).
• Requires a manual operation to restart the connection process after reaching the neutral state.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 15
Processing Specifications Laboratory Information System (LIS) Guide
4. Processing Specifications
The following specifications apply to all protocols.
The transfer of data between the LIS and the System occurs when:
• The System is at least at ready state and configured to send and/or receive.
• The hardware required is connected.
The software supports the processing of samples while sample programs are downloaded to the
System.
In the case of contention between the System sending or requesting data and the Lab Computer
sending unsolicited data, the System activity takes priority.
For the HL7 protocol, the System expects to receive an ACK for each message it sends to the LIS.
If the System does not receive an ACK before the send timeout expires, then the System assumes
that the RSP^ZOS message is lost, reestablishes the connection, and sends the message again.
The System acknowledges all HL7 messages (except OML) received from an LIS by sending an
ACK. See Section 5.4.1 HL7 ACK Acknowledgement Message.
If the protocol is set to HL7, the System produces an ORL - general laboratory order response for
each OML (event O21) and each RSP^ZOS instead of an ACK message.
If the message cannot be acted upon, the System includes an error segment in the
acknowledgement.
Messages that are not supported by the System result in an ACK message with an error condition
code 200 – Unsupported message type.
ACK messages that are not expected by the System are discarded.
The System supports
• A timeout of 30 seconds. This timeout is used by the send.
• A retry count of 5. The value is used by the send retry.
• A retry action pause of 0 seconds
The following specifications are for the download of sample program messages. See Section 4.3
LIS Append for a complete set of requirements regarding the subsequent processing that occurs
when downloaded sample program is received.
The System uses the following format for the Universal Test ID: value=^^^LocalOrMfctrCode
where LocalOrMfctrCode = ManualDilution + TestCode + TestDilution~TestCode +
TestDilution~… where TestCode + TestDilution is repeated for each assay in the program.
If the protocol is set to ASTM, the System processes an Order Download message when received
from an LIS. See Section 5.2.1 Order Download Message Structure for the structure.
If the protocol is set to ASTM, the System discards any comment records that appear after an
Order Record on a download.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
16 – 2017-12-19
Laboratory Information System (LIS) Guide Processing Specifications
If the protocol is set to ASTM, the System accepts only General comments pertaining to patient
demographics (those that appear after a Patient Record).
If the protocol is set to HL7, the System processes an OML message, Laboratory Order Request
(event O21) when received from a LIS. See Section 5.4.2 HL7 OML Laboratory Order Message
Structure.
Note: In relationship to triggers O21, O33, and O35, this message/trigger (O21) is used when an
order with multiple samples and optionally multiple containers per order item are to be
communicated.
If the protocol is set to HL7, the System sends the application acknowledgement (ORL) message
indicating whether the order is accepted or rejected by the System. See Section 5.4.3 HL7 ORL
Laboratory Order Response Message Structure.
Note: 1) This ORL message indicates that the message was parsable and passed an initial set of
edits. The order may still fail Sample Programming audits and return another message
(Download Message) back to the LIS indicating the subsequent failure. 2) Manually (on
System) entered orders and updates to orders originally received from the LIS will not
create an ORL. 3) Other test requests (e.g., derived tests or their test components) may
appear in the ORL when they were not requested by the LIS-OML. These do not need to
be created by the LCI since they will be automatically generated by the System and
associated with the order when the order (OML) is processed by the System.
If the protocol is set to HL7, the System only processes the first SAC segment for an SPM in each
OML message while ignoring the remaining SAC segments if they are present.
If the protocol is set to HL7, the System accepts the first NTE segment that contains a comment
type of “Generic” while ignoring the remaining NTE segments.
The System supports the following sample programming fields downloaded from an LIS.
• Sample ID (up to 15 printable characters)
• Tray ID (up to 2 alphanumeric characters)
• Cup Position (up to 2 numbers)
• (Test) Dilute (up to 4 numbers and a decimal point)
• Manual Dilution (4 numbers and a decimal point. Precision after the decimal is 4
numbers.)
• Test Code (3 numbers identify an Assay)
• Specimen Descriptor (1 number identifies a Body Fluid)
The System supports the following patient demographic fields downloaded from an LIS:
• Patient ID (up to 20 characters)
• Patient Last Name (up to 20 characters)
• Patient First Name (up to 15 characters)
• Patient Middle Initial (1 character)
• Patient Address (two input fields each with up to 20 characters)
• Patient Birth Date
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 17
Processing Specifications Laboratory Information System (LIS) Guide
After stripping any leading and trailing blanks, if any of the following fields in a downloaded
program contains more than their maximum number of characters, the System truncates the fields
and no download message is generated.
• The demographic “Physician Last Name” field
• The demographic “Physician First Name” field
• The demographic “Physician Middle Initial” field
• The demographic “Patient Address” field
• The demographic “Patient Room Number” field
• The demographic “Comments” field
• The demographic “Physician ID” field
After stripping any leading and trailing blanks, if any of the following fields in a downloaded
program contains more than their maximum number of characters, download message #2 is
generated and the program is rejected.
• The demographic “Patient ID” field
• The demographic “Patient Last Name” field (part of a key field)
• The demographic “Patient First Name” field (part of a key field)
• The demographic “Patient Middle Initial” field (part of a key field)
• The demographic “Patient Birth Date” field (System date audits apply)
• The demographic “Collection Date” field (System date audits apply)
• The demographic “Collection Time” field (System time audits apply)
• The sample program “Sample ID” field (other audits apply)
• The sample program “Tray” field (other audits apply)
• The sample program “Cup” field (other audits apply)
• The sample program “Dilute” field (other audits apply)
• The sample program “Manual Dilution” field (other audits apply)
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
18 – 2017-12-19
Laboratory Information System (LIS) Guide Processing Specifications
Downloaded sample programs that contain errors can be dispositioned in a variety of ways.
Serious errors result in the rejection of the entire sample program and cause a download message
to be generated. Less serious errors could cause the rejection of a single assay but would also
generate a download message. The least serious conditions result in processing the valid portion of
the sample program while ignoring the invalid portion. For example, if a dilution factor is specified
for a derived test, the test is performed but the dilution factor is not applied.
Download messages are generated to represent the detection of the following types of sample
programming and validation errors. Except when explicitly noted, all the messages indicate the
rejection of an entire sample program. (Code numbers are included in parenthesis. Item numbers
displayed in brackets do not pertain to this System; they are included in this list for completeness.)
Note: See Appendix H Download Messages for a list of the messages and the recommended
action for each message.
• Missing sample ID (1) – The sample program has a blank sample ID.
• Invalid data in field (2) – The sample program has problems with data in one or more
fields. That could include invalid characters, fields that were too long, a body fluid
mismatch for an append program, or an unsuccessful lookup of sex or range attribute.
• Tray name or cup missing (4)
o The tray name was specified but the sample program does not have an assigned
sample position.
o The sample program has a sample position but no tray name specified.
• Sample/patient name mismatch (6) – The sample program has the same sample ID as a
program already in the sample DataBase, but the patient names (last, first and MI) do not
match.
• Sample program/cup mismatch (7) – The sample program has a position that has already
been assigned to another sample program.
• Sample manually edited (12) – An attempt was made to edit the sample program from the
Lab Computer after it had been edited using the Sample Programming screen.
• No assays requested (13) – The sample program was downloaded with no assay requests.
This applies only if the sample program does not exist to begin with.
• Invalid assay requested (14) – An assay was requested which is currently not supported
by the System. The program is accepted but the unsupported assay is deleted from the
program.
• Derived test replicated (15) – The sample program included a request to replicate a
derived test. The program is accepted and the requested derived test is calculated.
• Too many assays (16) – The sample program has more than the maximum assays or
replicates requested. This includes volume checks.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 19
Processing Specifications Laboratory Information System (LIS) Guide
• Sample/tray program changed (17) – The sample program has been assigned to another
tray. The sample program is accepted as long as that tray is not currently in the metering
area.
• Sample program taken off tray (18) – The sample program has become unassigned. The
sample program is accepted. If the matching sample is in the buffer, automatic
assignment will assign the unassigned program; therefore, this audit code may only be
generated if the matching sample is not in the metering area.
• No assays: sample deleted (19) – The sample program did not include any assays and was
deleted by the Lab Computer.
• Dilution out of range (20)
o The manual dilution factor for this sample program is not between 0.0001 – 9999.0.
o The product of Test Dilution factor (i.e., Operator Requested Dilution) and Standard
Dilution factor is not a valid dilution factor.
o The assay is not dilutable and the test dilution factor is not 1.0.
o The assay is not configured for a diluent fluid and the Test Dilution factor (i.e.,
Operator Requested Dilution) is not 1.0
• Body fluid unknown (22) – The sample program included a body fluid that the System
does not currently support.
• Program pretreated assays separately (23) – The sample program included pretreated and
non-pretreated assays or included pretreated assays from multiple pretreatment groups.
This restriction does not apply to QC programs.
• Assays in progress could not be deleted (25) – An attempt to delete the program was
made. The sample program is accepted but not all tests could be deleted because they
were already in process. These tests will be completed, but they will be prevented from
reprocessing.
• Attempt to edit a program in process (26) – An attempt was made to append to a program
while it was in process but program information was changed that cannot be changed
when the program is in process. The program is rejected.
• An append program was downloaded but not all the assays could be added to the program
(27). For example, a derived test was requested, but one of the components was already
in the original program. The program is saved.
The System allows local sample programming to occur simultaneously with LIS downloaded
sample programming or host query.
The System supports the ability to add a new sample program received from the LIS.
A sample program being downloaded from the LIS overwrites a saved sample program when all of
the following conditions are met:
• The Sample ID of a saved sample program matches the ID of the downloaded sample
program.
• The patient first name, last name, and middle initial on the saved program are the same as
on the downloaded program.
• The saved sample program was previously downloaded from the LIS.
• The saved sample program has not been modified through the System’s GUI.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
20 – 2017-12-19
Laboratory Information System (LIS) Guide Processing Specifications
The LIS can explicitly request an unequal number of Na+ or K+ tests in urine.
For each Na+ rep in the downloaded sample program, the System ensures there is an associated K+
rep to serve as a blank in the following manner:
• If there is the same number of K+ reps as Na+ reps in the download, the System adds the
requested number of Na+ and K+ reps to the sample program. (There is no need to add
blank K+ reps since K+ reps are already requested.)
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 21
Processing Specifications Laboratory Information System (LIS) Guide
• If there are more Na+ reps than K+ reps in the download, the System adds the appropriate
number of “blank” K+ reps to the sample program, so that the number of Na+ and K+
reps are equal.
The dilution factor for an Na+ rep and its associated K+ rep is the same and is set to the dilution
factor specified for the Na+ rep.
Bu and Bc tests are performed using the same slide. Although the user can explicitly request an
unequal number of Bu and Bc tests, the System programs and reports an equal number of Bu and
Bc tests.
If there are more Bu reps than Bc reps in the downloaded sample program, the System adds
enough Bc reps to make the rep counts equal. The added Bc reps are labeled "adjunct" and use the
same dilution as the corresponding Bu rep.
If there are more Bc reps than Bu reps in the downloaded sample program, the System adds
enough Bu reps to make the rep counts equal. The added Bu reps are labeled "adjunct" and use the
same dilution as the corresponding Bc rep.
If the dilution factor for a Bu rep and matching Bc rep is not the same, the sample program is
rejected with an audit code 2.
If a downloaded program contains a derived test and the program already includes one or more
assays that are components of the derived test, additional replicates for the included assays are not
programmed.
If a downloaded program contains a derived test and a dilution factor is requested, the dilution
factor is not applied to the test. No download message is generated.
Note: The screen Options and Configuration – Configure Assays allows users to specify
whether or not a derived test result should be calculated and reported if that derived test’s
components have been programmed.
If the user requests the results for a derived test reported when all the components are requested in
a downloaded program, the derived test is added (and considered directly selected) only if all
components are directly selected to the sample program to the sample program.
Note: For example, If only Bc is requested in a program, Bu will be added but will not be
considered directly selected. In this case, the derived test NBIL (which has two
components, Bu and Bc) will not be added.
If a downloaded program contains a derived test but not any of the derived test's components,
when the components are added, they shall be considered directly selected.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
22 – 2017-12-19
Laboratory Information System (LIS) Guide Processing Specifications
The LIS may request to add tests to an existing program. This section outlines the behavior for
dealing with these append programs.
An append program being downloaded from the LIS finds a match with an existing saved program
to modify when all of the following conditions are met:
• The Sample ID of a saved sample program matches the ID of append program.
• The patient first name, last name, and middle initial on the saved program are the same as
on the append program.
• The body fluid of the saved program is the same as on the append program.
• The saved sample program was previously downloaded from the LIS.
• The saved sample program has not been modified through the System’s GUI.
An append program is treated as a regular downloaded program if no match is found.
Not In Process: An append program is composed with the matching saved program that is not in
process in the following ways to create a resultant program:
• Any tests specified in the append program not existing in the saved program are added
along with any related tests (derived, blanks, two-result, components).
• Any tests specified in the append program that exist in the saved are added, overwriting
the existing tests. Any non-existing related tests will also be added (derived, blanks,
components).
• Non-test program information from the append program will overwrite existing
information.
Note: Assays in the saved program but not in the append program remain in the resultant
program.
In Process: An append program is composed with the matching saved program that is in process
in the following ways to create a resultant program:
• Any tests specified in the append program not existing in the saved program are added
along with any related tests (derived, blanks, two-results, components) provided that all
related tests also do not exist in the saved program.
Note: Derived tests will not be automatically added if one of the components is already in the
program and the second component is appended
• Any tests specified in the append program that exist in the saved program are ignored.
• All other non-test program information and patient demographics from the append
program overwrite existing information.
Note: Assays in the saved program but not in the append program remain in the resultant
program.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 23
Processing Specifications Laboratory Information System (LIS) Guide
The following audits are performed on the resultant program as well as the append program:
• Remove Sample Indices (SI) assays for programs with mixed SI and non-SI assays.
o If the program does not exist in the database, SI tests will be removed from the
resultant program and no download message will be generated.
o If the program exists in the database, and has not started, SI tests will be removed
from the resultant program and no download message will be generated.
o If the program exists in the database, and is in process:
o If the original program was SI only, all Non-SI assays will be ignored and
download message 27 will be generated. (Refer to Appendix H Download
Messages.)
o If the original program was not SI only, all SI assays will be ignored and no
download message will be generated.
• Program pretreated assays separately. (Download message 23. Refer to Appendix H
Download Messages.)
• Too many assays. (Download message 16. Refer to Appendix H Download Messages.)An
append program with no tests indicated updates non-test program information and patient
demographics only.
The System provides a method to upload test results, patient, and physician demographics.
The System uploads result records to the Lab Computer provided all of the following conditions
are met:
• The Lab Computer is able to accept transmissions.
• No downloads are occurring at the time. (Not Applicable for HL7)
• No host queries are in process.
• Result records exist and are ready to be uploaded.
• No inventory queries are in process.
If the System is configured for the ASTM protocol, results are sent to the LIS with the Result
Upload message. See Section 5.2.2 for the message structure.
If the System is configured for the HL7 protocol, results are sent to the LIS with an OUL message,
Unsolicited Specimen Container Oriented Observation, (event R23). See Section 5.4.4 HL7 OUL
Unsolicited Observation Message Structure.
Note: The OUL^R23 message event is relatively new in HL7, specifically for lab automation
equipment container-oriented test results. Even though the message is titled Unsolicited
Specimen Container Oriented Observation, there is very little that distinguishes a solicited
result from an unsolicited result so there is only one result type message for both. See
Chapter 7.3.8 of the HL7 2.5 Specification, 2003.
Results with multiple replicates have each replicate uploaded along with the mean of the replicates.
Mean results are indicated by an MN code in the abnormal flags field.
For results records where the mean concentration is suppressed, the numerical value and units are
suppressed for a mean result record.
For semi-quantitative/qualitative assays the numerical value and units are suppressed in a result
record if the assay is not configured to display value with the result.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
24 – 2017-12-19
Laboratory Information System (LIS) Guide Processing Specifications
Results with 0 (e.g., Derived Tests) or 1 replicate record have only the test record sent.
For results records that have the following:
• a qualitative categorization
• a numerical result outside the reportable range
the categorization takes priority and the following occurs:
• the numerical value and units are suppressed if the system is not configured to include
reportable range flags in the result
• above/below reportable range flag are suppressed.
The System uploads an extended result record for each replicate result record only when
configured to send extended result information. The following extended results information will
be sent:
• Reagents
• Calibration
• Quality Control (for quality control samples)
• Diluents
If an attempt to upload results occurs during the download of sample programs, the current sample
program being downloaded is allowed to complete.
Once the sample program currently being downloaded is complete, the upload of results occurs.
The system waits for all tests in a sample to complete until results are uploaded when the upload
mode is configured to "Sample." The system uploads results as each test completes when upload
mode is configured to "Result."
When the system is configured with an upload mode of "Group," results are uploaded using the
following rules:
• Measured tests are eligible for upload if they are complete and all the other tests in the
result record that share the same upload group are also complete.
• Derived tests are eligible for upload if all of their components are eligible for send or
have already been sent.
Sample Indices tests have their own result records like any other assay. These result records will be
uploaded with the first result message for a sample regardless of the upload mode.
The Sample Indices result records contain any Hemolysis, Icterus or Turbidity codes specific to
the sample.
The Sample Indices results are created whether Sample Indices is enabled or not.
If the System is configured to include reportable range flags in the result field, the System
includes:
• A ">" in front of the result value when the result is above the reportable range
• A "<" in front of the result value when the result is below the reportable range
QC samples always upload the numerical result and units for semi-quantitative/qualitative assays.
Note: This overrides the “Display value with result” selection on the "Review/Edit
Configuration" screen in Options & Configuration.
If configured to include qualitative text comments, the system uploads a result comment record
containing the qualitative text for result records with a qualitative categorization.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 25
Processing Specifications Laboratory Information System (LIS) Guide
The user can configure the Host Query feature to be either on or off.
If the System is configured for the ASTM protocol, a host query is sent to the LIS with a Host
Query message. See Section 5.2.3 Host Query/Query Cancel Message Structure for the message
structure.
If the System is configured for the HL7 protocol, a host query is sent to the LIS with a QBP_ZOS
message, Host Query Message. See Section 5.4.5 HL7 QBP_ZOS Host Query Message Structure.
The System requests sample program information based on the barcode of the sample.
The System processes all downloaded sample programs (either requested or unsolicited) while a
query is pending.
If the System is configured for the HL7 protocol, a host query response is received from the LIS
with a RSP_ZOS message, Host Query Response. See Section 5.4.6 HL7 RSP_ZOS Host Query
Response Message Structure.
If the System is configured for the ASTM protocol, a host query response is received from the LIS
with an Order Download message for successful queries and a Failed Host Query Response
message for unsuccessful queries. See Sections 5.2.1 Order Download Message Structure and
5.2.4 Failed Host Query Response Message Structure.
The user sets a timer that determines how long the System waits for a response to a query. The
default timer value is 2.4 seconds, but this can be changed to values between 1.9 and 9.9 seconds.
After a configuration change, a timer change goes into effect with the next query.
The Host Query timer starts when the query record/segment is sent to the LIS.
The Host Query timer stops when any one of the following conditions occurs:
• A matching specimen ID with sample program is downloaded.
• No sample program is available for the query. (For ASTM, the indication is an “I” in the
termination code field of the Message Termination Record. For HL7, the indication is the
Query Response Status value of “NF” that is returned in the QAK segment of the
RSP^ZOS.)
• An error has occurred. (For ASTM, the indication is a “Q” in the termination code field
of the Message Termination Record. For HL7, the indication is the Query Response
Status value of “AE” or “AR” that is returned in the QAK segment of the RSP^ZOS.)
• The timer has expired.
Note: An append program can satisfy a Host Query.
If a query is requested and the timer expires before a response to the query, an attention error is
posted with text substitution displaying the timer value and the sample ID.
Once a query has been requested and a response is pending, no results are uploaded.
If a query request is sent and the timer expires, a query cancel message is sent. For HL7 protocol, a
QCN message is sent. See Section 5.4.7 HL7 QCN Host Query Cancel Message Structure. For
ASTM protocol, a Host Query Cancel message is sent. See Section 5.2.3 Host Query/Query
Cancel Message Structure.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
26 – 2017-12-19
Laboratory Information System (LIS) Guide Processing Specifications
If the protocol is set to ASTM and a Host Query request needs to be sent to the LIS while
messages are being downloaded, the download of the current message is allowed to complete.
The query is allowed to start and end (including cancel) before the System initiates any other
requests.
If the protocol is set to ASTM, messages are being uploaded, and a query is requested; the upload
stops on a frame boundary unless the next frame to send is the last.
If the protocol is set to ASTM and an upload is interrupted due to a query request, the System
sends a T (sender cancel) in the Termination Code field of the Message Termination record.
If the protocol is set to HL7, messages are being uploaded, and a query is requested; the upload
stops on a message boundary.
If the protocol is set to ASTM and the upload of results for a patient record was interrupted due to
a host query request, all previously transmitted results for that patient (ASMT Patient record) are
resent along with those results not sent once the upload resumes.
Note: Downloaded messages include sample programs and inventory queries. Uploaded
messages include results, inventory replies, and asynchronous notifications.
The messages upload when no additional queries are pending.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 27
Processing Specifications Laboratory Information System (LIS) Guide
The System accepts incoming inventory queries when no inventory queries are in process.
If the System is configured for the ASTM protocol, an inventory query is received by the LIS with
an Inventory Query message. See Section 5.2.5 Inventory Query Message Structure.
If the System is configured for the HL7 protocol, an inventory query is received from the LIS with
an INR Equipment Inventory Request message. See Section 5.4.8 HL7 INR Equipment Inventory
Request Message Structure.
The System sends an inventory response message when:
• No downloads are occurring at the time (Not Applicable to HL7)
• No host queries are in process
• An inventory query was received from the Lab Computer
If the System is configured for the HL7 protocol, an inventory update is sent to the LIS with an
INU Equipment Inventory Update message. See Section 5.4.9 HL7 INU Equipment Inventory
Update Message Structure.
If the System is configured for the ASTM protocol, an inventory update is sent to the LIS with an
Inventory Response message. See Section 5.2.7 Inventory Response Message Structure.
An inventory response message contains Reagent Inventory Records if the Inventory Type code in
the Inventory Query record is either R (Reagents) or A (All inventory types).
An inventory response message shall contain Diluent Inventory Records if the Inventory Type
code in the Inventory Query record is either D (Diluents) or A (All inventory types).
An inventory response message contains a Bulk Consumables Inventory Record if the Inventory
Type code in the Inventory Query record is either B (Bulk Consumables) or A (All inventory
types).
If the protocol is set to ASTM and an inventory query cannot be processed, the System responds
with a message containing only a Header and Trailer record. The Termination Code of the Trailer
record is Q (Error In Last Request). See Section 5.2.6 Failed Inventory Query Response Message
Structure.
If the protocol is set to HL7 and an inventory query cannot be processed, the System responds with
an Acknowledgement with an error segment in it.
If the protocol is set to ASTM and inventory is not on board for the requested inventory type, the
System responds with a message containing only a Header and Trailer record. The Termination
Code of the Trailer record is I (No Information Available). See Section 5.2.6 Failed Inventory
Query Response Message Structure.
If the protocol is set to HL7 and inventory is not on board for the requested inventory type, the
System responds with an INU Equipment Inventory Update message with no INV segments.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
28 – 2017-12-19
Laboratory Information System (LIS) Guide Processing Specifications
The System sends an automation status update message when all the following conditions are met:
• The System is configured to send asynchronous messages.
• No downloads are in process. (Not Applicable to HL7)
• No host queries are in process.
• No inventory queries are in process.
• A user has changed automation configuration settings.
If the System is configured for the HL7 protocol, an automation status update is sent to the LIS
with an Automated Equipment Notification (EAN) message. See Section 5.4.10 HL7 EAN
Equipment Notification Message Structure.
If the System is configured for the ASTM protocol, an automation status update is sent to the LIS
with an Automation Configuration Notification message. See Section 5.2.8 Automation
Configuration Notification Message Structure.
The System sends an error message when all the following conditions are met:
• The System is configured to send asynchronous messages.
• No downloads are in process. (Not Applicable to HL7)
• No host queries are in process.
• No inventory queries are in process.
• One or more condition codes have been posted.
If the System is configured for the HL7 protocol, an error message is sent to the LIS with an
Automated Equipment Notification (EAN) message. See Section 5.4.10 HL7 EAN Equipment
Notification Message Structure.
If the System is configured for the ASTM protocol, an error message is sent to the LIS with an
Error Notification message. See Section 5.2.9 Error Notification Message Structure.
The System excludes the following types of condition codes from transmission:
• Condition codes with a Transient severity
• Condition codes posted by the LIS interface task
The System sends a message containing a Download Messages when all the following conditions
are met:
• The System is configured to send asynchronous messages.
• No downloads are in process. (Not Applicable to HL7)
• No host queries are in process.
• No inventory queries are in process.
• One or more download messages have been posted.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 29
Message Specifications Laboratory Information System (LIS) Guide
If the System is configured for the HL7 protocol, Download Messages are sent to the LIS with an
Automated Equipment Notification (EAN) message. See Section 5.4.10 HL7 EAN Equipment
Notification Message Structure.
If the System is configured for the ASTM protocol, Download Messages are sent to the LIS with
an Download Message Notification message. See Section 5.2.10 Download Message Notification
Message Structure.
5. Message Specifications
The following specifications apply to all protocols.
In order to provide consistency with Sample Programming and to meet database storage
requirements, the field size limits described in Section 4.3.1 Supported Fields are used.
Not supported fields for all records are ignored on a download.
The System uploads null values for any field listed as Not supported for all records.
The System transmits and receives data streams using Unicode/UTF-8.
The System transmits and receives data streams using the ISO8859/1 Extended ASCII character
set.
The character encoding is configurable.
Note: Only one character encoding is supported at a given time.
The System logs an error if it receives a message using a character set different from its configured
character set.
The System transmits unmappable characters as a tilde (~) for the ASTM protocol and a pound
sign (#) for the HL7 protocol.
Note: Not all UTF-8 characters can be mapped to the Extended ASCII set.
The System does not allow the “+” character to be used as an encoding character. The “+”
character is reserved as a separator for the Ortho-Clinical Diagnostic universal test identifier.
5.1.1 General
The record and segment specifications in Sections 5.3 ASTM Records and 5.5 HL7 Message
Segments include fields that are not be supported by the interface. These fields are shaded in the
tables that follow and the words “Not Supported” appear in the Notes column for these fields.
Flag Description
0 No error
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
30 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
Flag Description
A The value of the QC result is greater than two but less than or equal to three baseline standard
deviations below the baseline mean. (-2S).
B The value of the QC result is greater than two but less than or equal to three baseline standard
deviations above the baseline mean. (+2S)
C The value of the QC result is greater than three baseline standard deviations below the baseline
mean. (-3S)
D The value of the QC result is greater than three baseline standard deviations above the baseline
mean. (+3S)
E The values of two consecutive QC results are both greater than two standard deviations above or
below the baseline mean. (22s)
F The values of two consecutive QC results have a change of at least four standard deviations. (R4s)
G The values of four consecutive QC results are all greater than one standard deviation above or
below the baseline mean. (41s)
H The values of ten consecutive QC results are all above or all below the baseline mean. (10x)
J The result of this Sample Indices test exceeds the threshold for this assay.
Status Description
N Normal Calibration
U User Calibrated
M User Modified
B User Modified/Calibrated
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 31
Message Specifications Laboratory Information System (LIS) Guide
Status Description
null Not Supported / Not Calibrated: This fluid is not supported or there are no passing calibrations for
this lot/fluid.
A N Attention
N W Action
M S Malfunction
S C Shutdown
Type Description
D Diluents
Type Description
R Retesting
D Reflex dilution/reduction
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
32 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
Type Description
Type Description
An empty string that identifies a specimen not positioned at an external metering point
Note: All referenced sections can be found in Standard Specification for Transferring
Information Between Clinical Laboratory Instruments and Computer Systems, NCCLS
LIS2-A (Formerly ASTM E1394-97) unless otherwise noted.
The LIS processes the record types and fields defined in Standard Specification for Transferring
Information between Clinical Laboratory Instruments and Computer Systems, NCCLS LIS2-A
(Formerly ASTM E1394-97) with the exception of Scientific (S) records. In the record field
descriptions, all values are text and meet the ASTM Common Field Types specification,
Section 6.6. If a field value has multiple components, they are differentiated by the “^” character.
If the System protocol is set to ASTM, the following Messages are utilized.
The System supports the hierarchical message structure as defined in Sections 5.1.8
through 5.1.11.
The System supports the Logical Information Storage and Logical Transmission Error Recovery
Requirements defined in Sections 5.2.1 and 5.2.2.
The System processes Receiver Interrupt requests from the LIS. The System re-establishes a
session as sender at the end of the LIS message transmission or after the 15-second timeout. At that
time, the System sends the records specified by the Logical Information Storage and Logical
Transmission Error Recovery Requirements defined in Section 5.2.2
The ASTM Order Download message uses the following message structure:
H -- HEADER (See Section 5.3.1 in this LIS Guide.)
{
P -- PATIENT (See Section 5.3.3 in this LIS Guide.)
[C] -- COMMENT (See Section 5.3.6 in this LIS Guide.)
{O} -- ORDER (See Section 5.3.4 in this LIS Guide.)
}
L -- TRAILER (See Section 5.3.2 in this LIS Guide.)
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 33
Message Specifications Laboratory Information System (LIS) Guide
H|\^&|||LITT||||||||LIS2-A|20061214093913
P|1|U000856|||ORR^ABIGAIL^G||19780407|F||843 TALL OAKS
DR^HAILVILLE, MD 45831|||RASHAMDRA^SANJAY^V|S|||||||||||U7
C|1|N|Patient is complaining of shortness of breath and chest
pain.|G
O|1|S4331009704||^^^1.0+300+1.0\300+2.0\001+1.0|R||20061214093913|
|||N||||5||||||||||O
O|2|U4331009704||^^^1.0+301+1.0|R||20061214093913||||N||||3|||||||
|||O
L|1|N
5.2.2 Result Upload Message Structure
The ASTM Result Upload message uses the following message structure:
H -- HEADER (See Section 5.3.1 in this LIS Guide.)
P -- PATIENT (See Section 5.3.3 in this LIS Guide.)
[C] -- COMMENT (See Section 5.3.6 in this LIS Guide.)
O -- ORDER (See Section 5.3.4 in this LIS Guide.)
{
R -- RESULT (See Section 5.3.5 in this LIS Guide.)
[C] – COMMENT (See Section 5.3.6 in this LIS Guide)
[M/X] -- EXTENDED RESULT
(See Section 5.3.8 in this LIS Guide.)
}
L -- TRAILER (See Section 5.3.2 in this LIS Guide.)
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
34 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 35
Message Specifications Laboratory Information System (LIS) Guide
R|7|^^^1.0000+952+1.0|No<SP>Result|||^0^NR||V|||20080822093518|200
80822093518|43000224
L|1|N
This is an example of a suppressed mean result when three replicates of the assay are run:
H|\^&|||qnxa152||||||||LIS2-A|20090402152556<CR>
P|1|PID12345|||Doe^John^X||19850105|M||123<SP>Main<SP>St<SP>Apt<SP
>4^Anytown<SP>USA|||Jones^Robert^H|1|||||||||||S8
C|1|I|This<SP>is<SP>a<SP>patient<SP>co|G
O|1|test3^1^1||^^^1.0000+019+1.0\019+1.0\019+1.0\019+1.0\019+1.0\0
19+1.0\950+1.0\951+1.0\952+1.0|R||||||N||||5||||||||||F<CR>
R|1|^^^1.0000+019+1.0||||^0^UCFRMN\^6^ES\^6^ES\^6^ES||V|||20090402
073138|20090402074353|qnxa0152<CR>
C|1|I|See<SP>Replicates|I<CR>
R|2|^^^1.0000+019+1.0||||^S^REEMUC\^6^ES\^6^ES\^6^ES||V|||20090402
073142|20090402074218|qnxa0152<CR>
C|1|I|Negative|I<CR>
R|3|^^^1.0000+019+1.0||||^S^REEMUC\^6^ES\^6^ES\^6^ES||V|||20090402
073201|20090402074237|qnxa0152<CR>
C|1|I|Negative|I<CR>
R|4|^^^1.0000+019+1.0||||^S^REEMUC\^6^ES\^6^ES\^6^ES||V|||20090402
073220|20090402074256|qnxa0152<CR>
C|1|I|Negative|I<CR>
R|5|^^^1.0000+019+1.0||||^S^REEMUC\^6^ES\^6^ES\^6^ES||V|||20090402
073239|20090402074315|qnxa0152<CR>
C|1|I|Negative|I<CR>
R|6|^^^1.0000+019+1.0||||^S^REEMUC\^6^ES\^6^ES\^6^ES||V|||20090402
073258|20090402074334|qnxa0152<CR>
C|1|I|Negative|I<CR>
R|7|^^^1.0000+019+1.0||||^S^REEMUC\^6^ES\^6^ES\^6^ES||V|||20090402
073317|20090402074353|qnxa0152<CR>
C|1|I|Negative|I<CR>
R|8|^^^1.0000+950+1.0|No<SP>Result|||^6^ES||V|||20090402073138|200
90402073331|qnxa0152<CR>
R|9|^^^1.0000+951+1.0|No<SP>Result|||^6^ES||V|||20090402073138|200
90402073331|qnxa0152<CR>
R|10|^^^1.0000+952+1.0|No<SP>Result|||^6^ES||V|||20090402073138|20
090402073331|qnxa0152<CR>
L|1|N<CR>
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
36 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
This is an example of a suppressed mean result when only one replicate of the assay is run:
H|\^&|||qnxa152||||||||LIS2-A|20090402152556<CR>
P|1|PID12345|||Doe^John^X||19850105|M||123<SP>Main<SP>St<SP>Apt<SP
>4^Anytown<SP>USA|||Jones^Robert^H|1|||||||||||S8
C|1|I|This<SP>is<SP>a<SP>patient<SP>co|G
O|1|OneRepOnly^8^1||^^^1.0000+019+1.0\950+1.0\951+1.0\952+1.0|R|||
|||N||||5||||||||||F<CR>
R|1|^^^1.0000+019+1.0|2.75|My<SP>Units||^S^REEMUC\^6^ES\^6^ES\^6^E
S||V|||20090402080046|20090402081126|qnxa0152<CR>
C|1|I|Negative|I<CR>
R|2|^^^1.0000+950+1.0|No<SP>Result|||^6^ES||V|||20090402080046|200
90402080104|qnxa0152<CR>
R|3|^^^1.0000+951+1.0|No<SP>Result|||^6^ES||V|||20090402080046|200
90402080104|qnxa0152<CR>
R|4|^^^1.0000+952+1.0|No<SP>Result|||^6^ES||V|||20090402080046|200
90402080104|qnxa0152<CR>
L|1|N<CR>
The ASTM Host Query/Query Cancel message uses the following message structure:
H -- HEADER (See Section 5.3.1 in this LIS Guide.)
Q -- REQUEST (See Section 5.3.7 in this LIS Guide.)
L -- TRAILER (See Section 5.3.2 in this LIS Guide.)
5.2.3.1 Host Query/Query Cancel Example
Host Query
H|\^&|||qnxa224||||||||LIS2-A|20061214091313
Q|1|^100987654321||ALL||||||A||O
L|1|N
Host Query Cancel
H|\^&|||qnxa224||||||||LIS2-A|20061214091316
Q|1|^100987654321||ALL||||||A||A
L|1|T
5.2.4 Failed Host Query Response Message Structure
The ASTM Failed Host Query Response message uses the following message structure:
H -- HEADER (See Section 5.3.1 in this LIS Guide.)
[R] -- REQUEST (See Section 5.3.7 in this LIS Guide.)
L -- TRAILER (See Section 5.3.2 in this LIS Guide.)
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 37
Message Specifications Laboratory Information System (LIS) Guide
H|\^&|||LITT||||||||LIS2-A|20061214091313
L|1|I
5.2.5 Inventory Query Message Structure
The ASTM Inventory Query message uses the following message structure:
H -- HEADER (See Section 5.3.1 in this LIS Guide.)
M/I -- INVENTORY QUERY (See Section 5.3.10 in this LIS
Guide.)
L -- TRAILER (See Section 5.3.2 in this LIS Guide.)
5.2.5.1 Inventory Query Message Example
H|\^&|||LITT||||||||LISH2-A|20061213142107
M|1|I|A
L|1|N
5.2.6 Failed Inventory Query Response Message Structure
The ASTM Failed Inventory Query Response message uses the following message structure:
H -- HEADER (See Section 5.3.1 in this LIS Guide.)
L -- TRAILER (See Section 5.3.2 in this LIS Guide.)
5.2.6.1 Failed Inventory Query Response Message Example
H|\^&|||qnxa224||||||||LISH2-A|20061213093820
L|1|I
5.2.7 Inventory Response Message Structure
The ASTM Inventory Response message uses the following message structure:
H -- HEADER (See Section 5.3.1 in this LIS Guide.)
[{M/R}] -- REAGENT INVENTORY (See Section 5.3.11 in this LIS
Guide.)
[{M/D}] -- DILUENT INVENTORY (See Section 5.3.12 in this LIS
Guide.)
[{M/B}] -- BULK CONSUMABLES INVENTORY (See Section 5.3.13 in
this LIS Guide.)
L -- TRAILER (See Section 5.3.2 in this LIS Guide.)
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
38 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
H|\^&|||qnxa224||||||||LISH2-A|20061213093820
M|1|R|001||1600^20071213093820^40^^^^^^^^^
M|2|D|2601|0001||20071213093821|40
M|3|D|2603|0516||20071213093821|40
M|4|B|Lot1^125|Lot2^25|Infinite|45|10|3500|4|22|49|1^596000|2645
L|1|N
5.2.8 Automation Configuration Notification Message Structure
The ASTM Automation Configuration Notification message uses the following message structure:
H -- HEADER (See Section 5.3.1 in this LIS Guide.)
{M/A} -- AUTOMATION CONFIGURATION (See Section 5.3.9 in
this LIS Guide.)
L -- TRAILER (See Section 5.3.2 in this LIS Guide.)
5.2.8.1 Automation Configuration Notification Message Example
H|\^&|||qnxa224||||||||LIS2-A|20061214091933
M|1|A|D\D
L|1|N
5.2.9 Error Notification Message Structure
The ASTM Error Notification message uses the following message structure:
H -- HEADER (See Section 5.3.1 in this LIS Guide.)
{M/E} -- ERROR (See Section 5.3.14 in this LIS Guide.)
L -- TRAILER (See Section 5.3.2 in this LIS Guide.)
5.2.9.1 Error Notification Message Example
H|\^&|||qnxa224||||||||LIS2-A|20061214091744
M|1|E|PWE|005|A|Hemolysis coded 'ES' - Turbidity may be
interfering|20061214091744
L|1|N
5.2.10 Download Message Notification Message Structure
The ASTM Download Message Notification uses the following message structure:
H -- HEADER (See Section 5.3.1 in this LIS Guide.)
{M/M} -- DOWNLOAD MESSAGE (See Section 5.3.15 in this LIS
Guide.)
L -- TRAILER (See Section 5.3.2 in this LIS Guide.)
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 39
Message Specifications Laboratory Information System (LIS) Guide
H|\^&|||qnxa224||||||||LIS2-A|20061214092212
M|1|M|13|13-No assays requested.|20060130044311|3705000004^2^1
L|1|N
5.3 ASTM Records
ASTM Records are the components that make up ASTM Messages. In the tables that show the
field sequences, shaded rows with Not Supported in the Notes column indicate fields that are not
supported by the System. No data is sent in these field positions. The Column Headings of "D" and
"U" indicate the optionality of the fields in the record when Downloading to the System (D) and
Uploading to the LIS (U). The column Heading of “R” indicates the Repeatability. Use the table
below as a key for the values in these columns:
D U R
The System ignores incoming Result, Request Information, Scientific, and Manufacturer
Information record types with the exception of Manufacturer Information records defined in this
document.
The software accepts and ignores the following escape sequences:
• &H& Start highlighting text
• &N& Normal text (end highlighting)
• &Zcccc& Local ( manufacturer) defined escape sequence
The software accepts and supports the following escape sequences:
• &F& Embedded field delimiter character
• &S& Embedded component field delimiter character
• &R& Embedded repeat field delimiter character
• &E& Embedded escape delimiter character
• &Xhhhh& hexadecimal data (e.g., &XA& is a linefeed character)
All date fields are formatted as specified by Dates and Times defined in Section 6.6.2 of Standard
Specification for Transferring Information Between Clinical Laboratory Instruments and
Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
40 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following unshaded fields of the ASTM Header (H) Record as defined in
Section 7.1 of Standard Specification for Transferring Information Between Clinical Laboratory
Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 41
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following fields of the ASTM Trailer (T) Record as defined in Section
13.1 of Standard Specification for Transferring Information Between Clinical Laboratory
Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
2 R A Sequence number 1 =1
The System supports the following unshaded fields of the ASTM Patient (P) Record as defined in
Section 8.1 of Standard Specification for Transferring Information Between Clinical Laboratory
Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
42 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 43
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following unshaded fields of the ASTM Order (O) Record as defined in
Section 9.4 of Standard Specification for Transferring Information Between Clinical Laboratory
Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
6 R A Priority 1 = S, A, R, C, P
S and A considered STAT priority
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
44 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
12 O A Action Code 1 = C, N, A, P, L, X, Q
P, L, X, and Q are treated as N.
On Download:
N (normal) adds a program for this
order
A (append) adds tests to the matching
program overwriting any existing tests
and demographics. For orders with no
tests, only demographics will be
updated. See Section 4.3.4 LIS
Append for details.
C (cancel) deletes the matching
program.
On upload System always sends “N”
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 45
Message Specifications Laboratory Information System (LIS) Guide
26 N A Report Types 1 = O, P, F
O for downloading
P for uploading when some tests are
still in progress
F for uploading when all tests in
sample are complete
Other values not supported
The System supports the following unshaded fields of the ASTM Result (R) Record as defined in
Section 10.1 of Standard Specification for Transferring Information Between Clinical Laboratory
Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
46 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 47
Message Specifications Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
48 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following fields of the Comment (C) Record as defined in Section 11.1 of
Standard Specification for Transferring Information Between Clinical Laboratory Instruments
and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 49
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following unshaded fields of the Request (Q) Record as defined in
Section 12.1 of Standard Specification for Transferring Information Between Clinical Laboratory
Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
50 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following fields of the Extended Result (M/X) Record as defined in
Section 15.1 of Standard Specification for Transferring Information Between Clinical Laboratory
Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
2 - A Sequence number 1 =1
4 - A Reagents 12^14^1 =
4^5^5^4 ReagentLot^ReagentExpDate^Reagent
LoadDate^ ERFLot^IWFLot^SRLot
Note: SRLot is left blank if no SR was
used for this result.
5 - S Calibration 14^1^14 =
CalibrationDate^CalibrationStatus^Cal
ExpDate (See 5.1.3 Calibration Status
in this LIS Guide.) Only included if
the associated calibration record still
exists in the database.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 51
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following fields of the Automation Configuration (M/A) Record as
defined in Section 15.1 of Standard Specification for Transferring Information Between Clinical
Laboratory Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
2 - A Sequence number 1 =1
The System supports the following fields of the Inventory Query (M/I) Record as defined in
Section 15.1 of Standard Specification for Transferring Information Between Clinical Laboratory
Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
2 R - Sequence number 1 =1
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
52 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following fields of the Reagent Inventory (M/R) Record as defined in
Section 15.1 of Standard Specification for Transferring Information Between Clinical Laboratory
Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
2 - A Sequence number 1 =1
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 53
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following fields of the Diluent (M/D) Inventory Record as defined in
Section 15.1 of Standard Specification for Transferring Information Between Clinical Laboratory
Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
2 - A Sequence number 1 =1
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
54 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following fields of the Bulk Consumables Inventory (M/B) Record as
defined in Section 15.1 of Standard Specification for Transferring Information Between Clinical
Laboratory Instruments and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
Note: Values for any items that are not present on VITROS 3600 or VITROS® XT 3400 are
set to 0.
2 - A Sequence number 1 =1
7 - A MicroTips 3 = NumberRemaining
8 - A Cuvettes 3 = NumberRemaining
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 55
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following fields of the Error (M/E) Record as defined in Section 15.1 of
Standard Specification for Transferring Information Between Clinical Laboratory Instruments
and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
2 - A Sequence number 1 =1
Note: For automation samples at the MicroSlide POR, the tray is “A#” and cup will be 1.
Note: For automation samples at the MicroIA POR, the tray is “B#” and cup will be 1.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
56 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following fields of the Error (M/M) Record as defined in Section 15.1 of
Standard Specification for Transferring Information Between Clinical Laboratory Instruments
and Computer Systems, NCCLS LIS2-A (Formerly ASTM E1394-97).
2 - A Sequence number 1 =1
The System supports the HL7 Version 2.5 messaging protocol specifications. The System
implements HL7 message structures and formats as described in HL7 2.5 Specifications, 2003,
specifically Clinical Laboratory Automation messaging as detailed and referenced in Chapter 13
and all referenced chapters of the HL7 messaging document. If the System protocol is set to HL7,
the following messages are utilized.
In defining the HL7 message structures, the optional HL7 segments that are not relevant to the
System are not included in the requirements for the HL7 interface.
Note: All referenced sections can be found in HL7 2.5 Specifications, 2003 unless otherwise
noted.
The HL7 ACK message uses the following message structure as defined in Section 13.3 of
HL7 2.5 Specifications, 2003.
MSH -- Message Header (See Section 5.5.5 in this LIS Guide.)
MSA -- Message Acknowledgment (See Section 5.5.4 in this LIS
Guide.)
[{ERR}] -- Error (See Section 5.5.1 in this LIS Guide.)
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 57
Message Specifications Laboratory Information System (LIS) Guide
MSH|^&~\|||||20080128111742.315||ORL^O22^ORL_O22|20080128111742.31
5|P|2.5||||||UNICODE<SP>UTF-8|||
MSA|AR|tBTTPzcded||||
ERR||PID^^5^^2^|102|E|||Field<SP>exceeds<SP>maximum<SP>number<SP>o
f<SP>characters,<SP>chars=16,<SP>max=15|||||
5.4.2 HL7 OML Laboratory Order Message Structure
MSH|^&~\|LITT||||20071022093057||OML^O21^OML_O21|tWcKE<SP>QBK5|P|2
.5||||||UNICODE<SP>UTF-8|||
PID|||1234567890||Doe^John^M||19791203000000|M|||100<SP>Indigo<SP>
Creek<SP>Dr^Rochester,<SP>NY<SP>14626||||||||||||||||||||||||||||
NTE||L|Comment<SP>about<SP>patient|G
PV1||N|^RM<SP>2MS7||||8651847^Robert^Smith^M|||||||||||S||||||||||
||||||||||||||||||||||||
ORC|NW|||||||||||||||||||||||||||||
OBR||||^^^1.0+001+1.0|R||20071022092920||||S||||||||||||||F|||||||
|||||||||||||||||
SPM||||5|||||||||||||||||||||||||
SAC|||SID54981||||||||||1|1||||||||||||||||||||||||||||||
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
58 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The HL7 ORL message uses the following message structure as defined in Section 4.4.7 of
HL7 2.5 Specifications, 2003.
MSH -- Message Header (See Section 5.5.5 in this LIS Guide.)
MSA -- Message Acknowledgment (See Section 5.5.4 in this LIS
Guide.)
[{ERR}] -- Error (See Section 5.5.1 in this LIS Guide.)
[ -- RESPONSE begin
[ -- PATIENT begin
PID -- Patient Identification (See Section 5.5.11 in this
LIS Guide.)
[{ -- ORDER begin
ORC -- Common Order (See Section 5.5.10 in this LIS
Guide.)
[ -- OBSERVATION_REQUEST begin
OBR -- Observation Request (See Section Section
5.5.8 in this LIS Guide.)
[{ -- SPECIMEN begin
SPM -- Specimen (See Section 5.5.18 in this
LIS Guide.)
[{SAC}] -- Specimen Container Details (See Section
5.5.17 in this LIS Guide.)
}] -- SPECIMEN end
] -- end OBSERVATION_REQUEST
}] -- ORDER end
] -- PATIENT end
] -- RESPONSE end
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 59
Message Specifications Laboratory Information System (LIS) Guide
MSH|^&~\|||||20071022093016.332||ORL^O22^ORL_O22|20071022093016.33
2|P|2.5||||||UNICODE<SP>UTF-8|||
MSA|AA|tWcKE<SP>QBK5||||
PID|||1234567890||Doe^John^M||19791203000000|M|||100<SP>Indigo<SP>
Creek<SP>Dr^Rochester,<SP>NY<SP>14626||||||||||||||||||||||||||||
ORC|NW|||||||||||||||||||||||||||||
OBR||||^^^1.0+001+1.0|R||20071022092920||||S||||||||||||||F|||||||
|||||||||||||||||
SPM||||5|||||||||||||||||||||||||
SAC|||SID54981||||||||||1|1||||||||||||||||||||||||||||||
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
60 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The HL7 OUL message uses the following message structure as defined in Section 7.3.8 of
HL7 2.5 Specifications, 2003.
Note: The ZER, Extended Results segment appears when Extended Results upload is enabled.
MSH -- Message Header (See Section 5.5.5 in this LIS Guide.)
[ PID ] –- Patient Identification (See Section 5.5.11 in this LIS
Guide.)
[{NTE}] -- Notes and Comments for Patient ID (See Section 5.5.7 in
this LIS Guide.)
[ PV1 ] -- Patient Visit (See Section 5.5.12 in this LIS Guide.)
{ -- SPECIMEN BEGIN
SPM -- Specimen information (See Section 5.5.18 in this LIS
Guide.)
{SAC} -- Container information (See Section 5.5.17 in this LIS
Guide.)
{ -- ORDER begin
OBR (See Section 5.5.8 in this LIS Guide.)
[ORC] (See Section 5.5.10 in this LIS Guide.)
[{ -- RESULT Begin
OBX (See Section 5.5.9 in this LIS Guide.)
[NTE] – Notes and Comments for Observation (See Section
5.5.7 in this LIS Guide.)
[ZER] – Extended Results (See Section 5.5.19 in this LIS
Guide.)
}] -- RESULT End
} -- ORDER end
} -- SPECIMEN end
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 61
Message Specifications Laboratory Information System (LIS) Guide
62 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
OBX|||^^^1.0000+523+1.0||500|ng/mL||^R^EP~^0^NR~^0^NR~^0^NR|||F|||
||||43000224|20080822093936
NTE||I|Positive|I
OBX|||^^^1.0000+525+1.0||>1000|ng/mL||^R^OREP~^0^NR~^0^NR~^0^NR|||
F|||||||43000224|20080822094001
NTE||I|Positive|I
OBX|||^^^1.0000+950+1.0||No<SP>Result|||^0^NR~^^~^^~^^|||F|||||||4
3000224|20080822093518
OBX|||^^^1.0000+951+1.0||No<SP>Result|||^0^NR~^^~^^~^^|||F|||||||4
3000224|20080822093518
OBX|||^^^1.0000+952+1.0||No<SP>Result|||^0^NR~^^~^^~^^|||F|||||||4
3000224|20080822093518
This is an example of a suppressed mean result when six replicates of the assay are run:
MSH|^&~\|qnxa0152||||20090402151403.275||OUL^R23^OUL_R23|200904021
51403.275|P|2.5||||||UNICODE<SP>UTF-8|||<CR>
PID|||||^^||||||^||||||||||||||||||||||||||||<CR>
NTE||||<CR>
PV1|||^||||^^^|||||||||||||||||||||||||||||||||||||||||||||<CR>
SPM||||5|||||||||||||||||||||||||<CR>
SAC|||test3||||||||||1|1||||||||||||||||||||||||||||||<CR>
OBR||||^^^1.0000+019+1.0~019+1.0~019+1.0~019+1.0~019+1.0~019+1.0~9
50+1.0~951+1.0~952+1.0|R||||||||||||||||||||F|||||||||||||||||||||
|||<CR>
ORC|NW|||||||||||||||||||||||||||||<CR>
OBX|||^^^1.0000+019+1.0|||||^0^UCFRMN~^6^ES~^6^ES~^6^ES|||F|||||||
qnxa0152|20090402074353<CR>
NTE||I|See<SP>Replicates|I<CR>
OBX|||^^^1.0000+019+1.0|||||^S^REEMUC~^6^ES~^6^ES~^6^ES|||F|||||||
qnxa0152|20090402074218<CR>
NTE||I|Negative|I<CR>
OBX|||^^^1.0000+019+1.0|||||^S^REEMUC~^6^ES~^6^ES~^6^ES|||F|||||||
qnxa0152|20090402074237<CR>
NTE||I|Negative|I<CR>
OBX|||^^^1.0000+019+1.0|||||^S^REEMUC~^6^ES~^6^ES~^6^ES|||F|||||||
qnxa0152|20090402074256<CR>
NTE||I|Negative|I<CR>
OBX|||^^^1.0000+019+1.0|||||^S^REEMUC~^6^ES~^6^ES~^6^ES|||F|||||||
qnxa0152|20090402074315<CR>
NTE||I|Negative|I<CR>
OBX|||^^^1.0000+019+1.0|||||^S^REEMUC~^6^ES~^6^ES~^6^ES|||F|||||||
qnxa0152|20090402074334<CR>
NTE||I|Negative|I<CR>
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 63
Message Specifications Laboratory Information System (LIS) Guide
OBX|||^^^1.0000+019+1.0|||||^S^REEMUC~^6^ES~^6^ES~^6^ES|||F|||||||
qnxa0152|20090402074353<CR>
NTE||I|Negative|I<CR>
OBX|||^^^1.0000+950+1.0||No<SP>Result|||^6^ES~^^~^^~^^|||F|||||||q
nxa0152|20090402073331<CR>
OBX|||^^^1.0000+951+1.0||No<SP>Result|||^6^ES~^^~^^~^^|||F|||||||q
nxa0152|20090402073331<CR>
OBX|||^^^1.0000+952+1.0||No<SP>Result|||^6^ES~^^~^^~^^|||F|||||||q
nxa0152|20090402073331<CR>
This is an example of a suppressed mean result when only one replicate of the assay is run:
MSH|^&~\|qnxa0152||||20090402151404.343||OUL^R23^OUL_R23|200904021
51404.343|P|2.5||||||UNICODE<SP>UTF-8|||<CR>
PID|||||^^||||||^||||||||||||||||||||||||||||<CR>
NTE||||<CR>
PV1|||^||||^^^|||||||||||||||||||||||||||||||||||||||||||||<CR>
SPM||||5|||||||||||||||||||||||||<CR>
SAC|||OneRepOnly||||||||||8|1||||||||||||||||||||||||||||||<CR>
OBR||||^^^1.0000+019+1.0~950+1.0~951+1.0~952+1.0|R||||||||||||||||
||||F||||||||||||||||||||||||<CR>
ORC|NW|||||||||||||||||||||||||||||<CR>
OBX|||^^^1.0000+019+1.0||2.75|My<SP>Units||^S^REEMUC~^6^ES~^6^ES~^
6^ES|||F|||||||qnxa0152|20090402081126<CR>
NTE||I|Negative|I<CR>
OBX|||^^^1.0000+950+1.0||No<SP>Result|||^6^ES~^^~^^~^^|||F|||||||q
nxa0152|20090402080104<CR>
OBX|||^^^1.0000+951+1.0||No<SP>Result|||^6^ES~^^~^^~^^|||F|||||||q
nxa0152|20090402080104<CR>
OBX|||^^^1.0000+952+1.0||No<SP>Result|||^6^ES~^^~^^~^^|||F|||||||q
nxa0152|20090402080104<CR>
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
64 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
MSH|^&~\|||||20071022103351.228||QBP^ZOS^QBP_ZOS|20071022103351.22
8|P|2.5||||||UNICODE<SP>UTF-8|||
QPD|ZOS^Lab<SP>Order<SP>Specimen<SP>Query|20071022103351.228|SID12
345|||||||A||
RCP|I||R||||
5.4.6 HL7 RSP_ZOS Host Query Response Message Structure
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 65
Message Specifications Laboratory Information System (LIS) Guide
MSH|^&~\|LITT||||20071022104336||RSP^ZOS^RSP_ZOS|LJMnnU1ZIY|P|2.5|
|||||UNICODE<SP>UTF-8|||
QAK|20071022104255.318|OK|ZOS^Lab<SP>Order<SP>Specimen<SP>Query|1|
|
QPD|ZOS^Lab<SP>Order<SP>Specimen<SP>Query|20071022104255.318|SID12
345|||||||A||
PID|||PID123456||Doe^John^M||20071203000000|M|||100<SP>Indigo<SP>C
reek<SP>Dr^Rochester,<SP>NY<SP>14626||||||||||||||||||||||||||||
NTE||P|patient<SP>comments|G
PV1||N|^2M<SP>S7||||8651847^Robert^Smith^M|||||||||||S||||||||||||
||||||||||||||||||||||
SPM||||5|||||||||||||||||||||||||
SAC|||SID12345|||||||||||||||||||||||||||||||||||||||||
ORC|NW|||||||||||||||||||||||||||||
OBR|||qTY7a|^^^1.000+300+1.0|R||20060609000000||||S||||||||||||||F
||||||||||||||||||||||||
5.4.6.2 HL7 RSP Host Query Response Message Example (No Program Found)
MSH|^&~\|LITT||||20071022103433||RSP^ZOS^RSP_ZOS|gpyLLEG<SP>vz|P|2
.5|||||||||
QAK|20071022103351.228|NF|ZOS^Lab<SP>Order<SP>Specimen<SP>Query|0|
|
QPD|ZOS^Lab<SP>Order<SP>Specimen<SP>Query|20071022103351.228|SID12
345|||||||A||
5.4.7 HL7 QCN Host Query Cancel Message Structure
MSH|^&~\|||||20071022103354.230||QCN^J01^QCN_J01|20071022103354.23
0|P|2.5||||||UNICODE<SP>UTF-8|||
QID|20071022103351.228|ZOS^Lab<SP>Order<SP>Specimen<SP>Query
5.4.8 HL7 INR Equipment Inventory Request Message Structure
The HL7 INR message uses the following message structure as defined in Section 13.3.6 in
HL7 2.5 Specifications, 2003.
MSH -- Message Header (See Section 5.5.5 in this LIS Guide.)
EQU -- Equipment Detail (See Section 5.5.2 in this LIS Guide.)
INV -- Inventory Detail (See Section 5.5.3 in this LIS Guide.)
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
66 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The LIS queries the chemistry analyzer 0001 for status of all packages of the substance
(id=MF01239)
MSH|^&~\|LITT||||20071022093918||INR^U06^INR_U06|6Js5<SP>CHv7A|P|2
.5||||||UNICODE<SP>UTF-8|||
EQU|J43000001|20071022093918|||
INV|A|~~~~~~~~||||||||||||||||||
5.4.9 HL7 INU Equipment Inventory Update Message Structure
MSH|^&~\|||||20071022093836.369||INU^U05^INU_U05|20071022093836.36
9|P|2.5||||||UNICODE<SP>UTF-8|||
EQU|00000000|20071022093836|||
INV|001|~~~~~~~~|SR||||||100||COUNT|20071211075549|||001|2050||||
INV|300|~~~~~~~~|SR||||||60||COUNT|20071029092641|||300|00492619||
||
INV|317|~~~~~~~~|MR||||||60||COUNT|20071105082644|||317~327|023659
12||||
INV|327|~~~~~~~~|MR||||||60||COUNT|20071105082644|||327~317|023659
12||||
INV|500|~~~~~~~~|SR||||||50||COUNT|20071114075505|||500|1500040001
||||
INV|1001|~~~~~~~~|DI||||||63||mL|20071212075505||||8000010001||||
INV|2601|~~~~~~~~|DI||||||100||COUNT|20071211075549||||9998||||
INV|ERF|~~~~~~~~|LI||||||800||COUNT|||||100||||
INV|IWF|~~~~~~~~|LI||||||300||COUNT|||||100||||
INV|VT|~~~~~~~~|SC||||||Infinite||COUNT|||||||||
INV|MT|~~~~~~~~|SC||||||384||COUNT|||||||||
INV|CUV|~~~~~~~~|SC||||||300||COUNT|||||||||
INV|SW|~~~~~~~~|SW||||||3780||COUNT|||||||||
INV|CW|~~~~~~~~|SW||||||0||%|||||||||
INV|TW|~~~~~~~~|SW||||||0||%|||||||||
INV|LW|~~~~~~~~|LW||||||96||%|||||||||
INV|SR|~~~~~~~~|LI||||||393||COUNT|||||1||||
INV|UWR|~~~~~~~~|LI||||||5198||mL|||||||||
INV|MCA|~~~~~~~~|SC||||||1|COUNT|||||||||
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 67
Message Specifications Laboratory Information System (LIS) Guide
The HL7 EAN message uses the following message structure as defined in Section 13.3.9 in
HL7 2.5 Specifications, 2003.
MSH -- Message Header (See Section 5.5.5 in this LIS Guide.)
EQU -- Equipment Detail (See Section 5.5.2 in this LIS Guide.)
{ -- Notification Begin
NDS -- Notification Detail (See Section 5.5.6 in this LIS
Guide.)
} -- Notification End
5.4.10.1 HL7 EAN Equipment Notification Message Example (Error)
MSH|^&~\|||||20071022094305.929||EAN^U09^EAN_U09|20071022094305.92
9|P|2.5||||||UNICODE<SP>UTF-8|||
EQU|00000000|20071022094305|||
NDS|329|20071022093708|N|PWE-107^MicroSensor<SP>Ref<SP>and<SP>Dark
<SP>readings<SP>have<SP>same<SP>value:<SP>ID<SP>SID54981,<SP>Index
<SP>&SID54981&1&1
5.4.10.2 HL7 EAN Equipment Notification Message Example (Automation Configuration)
MSH|^&~\|||||20071022094437.945||EAN^U09^EAN_U09|20071022094437.94
5|P|2.5||||||UNICODE<SP>UTF-8|||
EQU|00000000|20071022094437|||
NDS|0|20071022094437|N|AutoConfig-0^Disabled
NDS|1|20071022094437|N|AutoConfig-1^Enabled
5.4.10.3 HL7 EAN Equipment Notification Message Example (Download Message)
MSH|^&~\|||||20071022101519.271||EAN^U09^EAN_U09|20071022101519.27
1|P|2.5||||||UNICODE<SP>UTF-8|||
EQU|00000000|20071022101519|||
NDS|4|20071022100658|N|Audit-17^17-Sample/tray<SP>program<SP>chang
ed.&SID54981&2&1
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
68 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The following section defines the HL7 message segments that are used. All segments are
terminated with a Carriage Return <CR> (hex 0D0A). This designation cannot be changed.
The delimiter values are given in MSH-1 and MSH-2 and used throughout the message.
Applications must use agreed-upon delimiters to parse the message segments. The recommended
delimiters for laboratory messages are listed in the first two fields of the MSH segment. The
System employs the delimiters for all upload message segments. This is not configurable.
Escape sequences for field separator, component separator, subcomponent separator, repetition
separator, and escape character are also valid within an ST (String) data field. No escape sequence
may contain a nested escape sequence.
The following escape sequences are used in the System:
\F\ field separator
\S\ component separator
\T\ subcomponent separator
\R\ repetition separator
\E\ escape character
\Xdddd...\ hexadecimal data
The following message segments are used for HL7 messages. In the tables that show the field
sequences, shaded rows with Not Supported in the Notes column indicate fields that are not
supported by the System. No data is sent in these field positions. The Column Headings of “D” and
“U” indicate the optionality of the fields in the record when Downloading to the System (D) and
Uploading to the LIS (U). The column Heading of “R” indicates the Repeatability. The System
uses only the first repetition when it expects a single value. All other repetitions are ignored. Use
the table below as a key for the values in these columns:
D U R
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 69
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following unshaded fields of the Error (ERR) Segment as defined in
Section 2.15.5 of HL7 2.5 Specifications, 2003.
4 R A Severity 1 W or E
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
70 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following unshaded fields of the Equipment Detail (EQU) Segment as
defined in Section 13.4.1 of HL7 2.5 Specifications, 2003.
The System supports the following unshaded fields of the Inventory Detail (INV) Segment as
defined in Section 13.4.4 of HL7 2.5 Specifications, 2003.
Note: Values for any items that are not present on VITROS 3600 or VITROS® XT 3400 are
set to 0.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 71
Message Specifications Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
72 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following unshaded fields of the Message Acknowledgement (MSA)
Segment as defined in Section 2.15.8 of HL7 2.5 Specifications, 2003.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 73
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following unshaded fields of the Message Header (MSH) Segment as
defined in Section 2.15.9 of HL7 2.5 Specifications, 2003.
12 R A Version ID 3 “2.5”
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
74 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 75
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following fields of the Notification Detail (NDS) Segment as defined in
Section 13.4.7 of HL7 2.5 Specifications, 2003.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
76 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following unshaded fields of the Notes and Comments (NTE) Segment as
defined in Section 2.15.10 of HL7 2.5 Specifications, 2003.
The System supports the following unshaded fields of the Observation Request (OBR) Segment as
defined in Section 7.4.1 of HL7 2.5 Specifications, 2003.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 77
Message Specifications Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
78 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following unshaded fields of the Observation Result (OBX) Segment as
defined in Section 7.4.2 of HL7 2.5 Specifications, 2003.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 79
Message Specifications Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
80 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 81
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following unshaded fields of the Common Order (ORC) Segment as
defined in Section 4.5.1 of HL7 2.5 Specifications, 2003.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
82 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following unshaded fields of the Patient ID (PID) Segment as defined in
Section 3.4.2 of HL7 2.5 Specifications, 2003.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 83
Message Specifications Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
84 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 85
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following unshaded fields of the Patient Visit (PV1) Segment as defined
in Section 3.4.3 of HL7 2.5 Specifications, 2003.
1 N N Set ID - PV1
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
86 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 87
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following unshaded fields of the Query Acknowledgement Details (QAK)
Segment as defined in Section 5.5.2 of HL7 2.5 Specifications, 2003.
The System supports the following fields of the Query Identification Details (QID) Segment as
defined in Section 5.5.3 of HL7 2.5 Specifications, 2003.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
88 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
The System supports the following unshaded fields of the Query Parameter Definition Details
(QPD) Segment as defined in Section 5.5.4 of HL7 2.5 Specifications, 2003.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 89
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following unshaded fields of the Response Control Parameter Details
(RCP) Segment as defined in Section 5.5.6 of HL7 2.5 Specifications, 2003.
The System supports the following unshaded fields of the Specimen Container Details (SAC)
Segment as defined in Section 13.4.3 of HL7 2.5 Specifications, 2003.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
90 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 91
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following unshaded fields of the Specimen (SPM) Segment as defined in
Section 7.4.3 of HL7 2.5 Specifications, 2003.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
92 – 2017-12-19
Laboratory Information System (LIS) Guide Message Specifications
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 93
Message Specifications Laboratory Information System (LIS) Guide
The System supports the following fields of the Z Segment Extended Results (ZER) Segment.
1 - A Reagents 12^14^1 =
4^5^5^4 ReagentLot^ReagentExpDate^Reagent
LoadDate^ERFLot^IWFLot^SRLot
Note: SRLot is left blank if no SR was
used for this result.
2 - S Calibration 14^1^14 =
CalibrationDate^CalibrationStatus^Cal
ExpDate (See 5.1.3 Calibration Status
in this LIS Guide.) Sent only if
calibration info still in database. Date
in the format of
YYYYMMDDHHMMSS
4 - S Y Diluents 12 = DiluentLot
Repeated for each diluent lot used
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
94 – 2017-12-19
Laboratory Information System (LIS) Guide Notes
Appendix A Notes
Delayed Response
When an Assay Data Disk (ADD) is being loaded onto the System (about 10 minutes), the System
cannot process any incoming requests from the LIS. This delayed response in the communications
between the LIS and the System may cause the LIS to time out. Communication is reestablished
automatically when the ADD load has been completed.
Example: If the LIS sends an Inventory Query to the System during an ADD load, the query cannot
be completed and the LIS may time out. The Inventory Query takes approximately 1 minute to
process.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 95
Analyte Codes Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
96 – 2017-12-19
Laboratory Information System (LIS) Guide Analyte Codes
Analyte Report Assay Name Body VITROS VITROS® VITROS VITROS VITROS®
Code Name Fluid 5600 XT 7600 3600 4600 XT 3400
Body Fluid Key: 2 CSF, 3 Urine, 4 Whole Blood, 5 Serum, 6 Plasma, 7 Amniotic Fluid, 8 Reserved Fluid 1,
9 Reserved Fluid 2, 10 Reserved Fluid 3, UD User Defined
014 Testo Testosterone 6 √ √ √ NA NA
015 AFP AFP 5 √ √ √ NA NA
015 AFP AFP 6 √ √ √ NA NA
015 AFP AFP 7 √ √ √ NA NA
016 CEA CEA 5 √ √ √ NA NA
016 CEA CEA 6 √ √ √ NA NA
017 HBsAg HBsAg 5 √ √ √ NA NA
017 HBsAg HBsAg 6 √ √ √ NA NA
017 HBsAg HBsAg ES 5 √ √ √ NA NA
017 HBsAg HBsAg ES 6 √ √ √ NA NA
018 aHBs anti-HBs 5 √ √ √ NA NA
018 aHBs anti-HBs 6 √ √ √ NA NA
019 aHBc Anti-HBc 5 √ √ √ NA NA
019 aHBc Anti-HBc 6 √ √ √ NA NA
020 HBc M Anti-HBc IgM 5 √ √ √ NA NA
020 HBc M Anti-HBc IgM 6 √ √ √ NA NA
021 HBeAg HBeAg 5 √ √ √ NA NA
022 HAV M Anti-HAV IgM 5 √ √ √ NA NA
022 HAV M Anti-HAV IgM 6 √ √ √ NA NA
023 aHCV Anti-HCV 5 √ √ √ NA NA
023 aHCV Anti-HCV 6 √ √ √ NA NA
024 aHIV Anti HIV 1/2 5 √ √ √ NA NA
024 aHIV Anti HIV 1/2 6 √ √ √ NA NA
025 Rub G Rubella IgG 5 √ √ √ NA NA
025 Rub G Rubella IgG 6 √ √ √ NA NA
026 Rub M Rubella IgM 5 √ √ √ NA NA
026 Rub M Rubella IgM 6 √ √ √ NA NA
027 Tox G Toxoplasma IgG 5 √ √ √ NA NA
027 Tox G Toxoplasma IgG 6 √ √ √ NA NA
028 Tox M Toxoplasma IgM 5 √ √ √ NA NA
028 Tox M Toxoplasma IgM 6 √ √ √ NA NA
029 CK-MB CK-MB 5 √ √ √ NA NA
029 CK-MB CK-MB 6 √ √ √ NA NA
030 TropI Troponin I ES 5 √ √ √ NA NA
030 TropI Troponin I ES 6 √ √ √ NA NA
031 Ferr Ferritin 5 √ √ √ NA NA
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 97
Analyte Codes Laboratory Information System (LIS) Guide
Analyte Report Assay Name Body VITROS VITROS® VITROS VITROS VITROS®
Code Name Fluid 5600 XT 7600 3600 4600 XT 3400
Body Fluid Key: 2 CSF, 3 Urine, 4 Whole Blood, 5 Serum, 6 Plasma, 7 Amniotic Fluid, 8 Reserved Fluid 1,
9 Reserved Fluid 2, 10 Reserved Fluid 3, UD User Defined
031 Ferr Ferritin 6 √ √ √ NA NA
032 B12 Vitamin B12 5 √ √ √ NA NA
032 B12 Vitamin B12 6 √ √ √ NA NA
033 Fol Red Cell Folate 4 √ √ √ NA NA
033 Fol Folate 5 √ √ √ NA NA
033 Fol Folate 6 √ √ √ NA NA
034 Cort Cortisol 3 √ √ √ NA NA
034 Cort Cortisol 5 √ √ √ NA NA
034 Cort Cortisol 6 √ √ √ NA NA
036 PSA PSA 5 √ √ √ NA NA
036 PSA PSA 6 √ √ √ NA NA
037 fPSA Free PSA 5 √ √ √ NA NA
037 fPSA Free PSA 6 √ √ √ NA NA
038 CA125 CA 125 II 5 √ √ √ NA NA
038 CA125 CA 125 II 6 √ √ √ NA NA
039 CA153 CA 15-3 5 √ √ √ NA NA
039 CA153 CA 15-3 6 v √ √ NA NA
040 CA199 CA 19-9 5 √ √ √ NA NA
040 CA199 CA 19-9 6 √ √ √ NA NA
042 INS Insulin 5 √ √ √ NA NA
042 INS Insulin 6 √ √ √ NA NA
043 aHBe Anti-HBe 5 √ √ √ NA NA
044 NTx NTx 3 √ √ √ NA NA
045 - (Reserved for √ √ √ NA NA
Internal Use)
049 HBCon HBsAg Confirm 5 √ √ √ NA NA
049 HBCon HBsAg Confirm 6 √ √ √ NA NA
050 C-pep* C-peptide 5 √ √ √ NA NA
050 C-pep* C-peptide 6 √ √ √ NA NA
051 Myog Myoglobin 5 √ √ √ NA NA
051 Myog Myoglobin 6 √ √ √ NA NA
055 FT3II Free T3 II 5 √ √ √ NA NA
055 FT3II Free T3 II 6 √ √ √ NA NA
058 - (Reserved for √ √ √ NA NA
Internal Use)
061 HAV T Anti-HAV Total 5 √ √ √ NA NA
061 HAV T Anti-HAV Total 6 √ √ √ NA NA
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
98 – 2017-12-19
Laboratory Information System (LIS) Guide Analyte Codes
Analyte Report Assay Name Body VITROS VITROS® VITROS VITROS VITROS®
Code Name Fluid 5600 XT 7600 3600 4600 XT 3400
Body Fluid Key: 2 CSF, 3 Urine, 4 Whole Blood, 5 Serum, 6 Plasma, 7 Amniotic Fluid, 8 Reserved Fluid 1,
9 Reserved Fluid 2, 10 Reserved Fluid 3, UD User Defined
062 CMV G CMV IgG 5 √ √ √ NA NA
062 CMV G CMV IgG 6 √ √ √ NA NA
063 CMV M CMV IgM 5 √ √ √ NA NA
063 CMV M CMV IgM 6 √ √ √ NA NA
064 aHBs Anti-HBs 5 √ √ √ NA NA
065 NTBNP NT-proBNP 5 √ √ √ NA NA
065 NTBNP NT-proBNP 6 √ √ √ NA NA
069 HIV c HIV Combo 5 √ √ √ NA NA
069 HIV c HIV Combo 6 √ √ √ NA NA
070 iPTH Intact PTH 5 √ √ √ NA NA
070 iPTH Intact PTH 6 √ √ √ NA NA
073 Syph Syphilis TPA 5 √ √ √ NA NA
073 Syph Syphilis TPA 6 √ √ √ NA NA
074 tVitD VITAMIN D 5 √ √ √ NA NA
TOTAL
075 tPSA Total PSA II 5 √ √ √ NA NA
075 tPSA Total PSA II 6 √ √ √ NA NA
080 TIMP2 TIMP-2 3 √ √ √ NA NA
081 IGFBP IGFBP-7 3 √ √ √ NA NA
300 GLU Glucose 2 √ √ NA √ √
300 GLU Glucose 3 √ √ NA √ √
300 GLU Glucose 5 √ √ NA √ √
300 GLU Glucose 6 √ √ NA √ √
301 TP Total Protein 5 √ √ NA √ √
301 TP Total Protein 6 √ √ NA √ √
302 URIC Uric Acid 3 √ √ NA √ √
302 URIC Uric Acid 5 √ √ NA √ √
302 URIC Uric Acid 6 √ √ NA √ √
303 ALB Albumin 5 √ √ NA √ √
303 ALB Albumin 6 √ √ NA √ √
304 TRIG Triglycerides 5 √ √ NA √ √
304 TRIG Triglycerides 6 √ √ NA √ √
305 CHOL Cholesterol 5 √ √ NA √ √
305 CHOL Cholesterol 6 √ √ NA √ √
306 AMYL Amylase 3 √ √ NA √ √
306 AMYL Amylase 5 √ √ NA √ √
306 AMYL Amylase 6 √ √ NA √ √
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 99
Analyte Codes Laboratory Information System (LIS) Guide
Analyte Report Assay Name Body VITROS VITROS® VITROS VITROS VITROS®
Code Name Fluid 5600 XT 7600 3600 4600 XT 3400
Body Fluid Key: 2 CSF, 3 Urine, 4 Whole Blood, 5 Serum, 6 Plasma, 7 Amniotic Fluid, 8 Reserved Fluid 1,
9 Reserved Fluid 2, 10 Reserved Fluid 3, UD User Defined
307 CI- Chloride 3 √ √ NA √ √
307 CI- Chloride 5 √ √ NA √ √
307 CI- Chloride 6 √ √ NA √ √
308 K+ Potassium 3 √ √ NA √ √
308 K+ Potassium 5 √ √ NA √ √
308 K+ Potassium 6 √ √ NA √ √
309 Na+ Sodium 3 √ √ NA √ √
309 Na+ Sodium 5 √ √ NA √ √
309 Na+ Sodium 6 √ √ NA √ √
310 ECO2 Enzymatic CO2 5 √ √ NA √ √
310 ECO2 Enzymatic CO2 6 √ √ NA √ √
311 PHOS Phosphorus 3 √ √ NA √ √
311 PHOS Phosphorus 5 √ √ NA √ √
311 PHOS Phosphorus 6 √ √ NA √ √
312 LAC Lactate 6 √ √ NA √ √
314 CREA Creatinine 3 √ √ NA √ √
314 CREA Creatinine 5 √ √ NA √ √
314 CREA Creatinine 6 √ √ NA √ √
315 UREA Urea Nitrogen 3 √ √ NA √ √
315 UREA Urea Nitrogen 5 √ √ NA √ √
315 UREA Urea Nitrogen 6 √ √ NA √ √
317 Bu Unconjugated 5 √ √ NA √ √
Bilirubin
317 Bu Unconjugated 6 √ √ NA √ √
Bilirubin
318 Ca Calcium 3 √ √ NA √ √
318 Ca Calcium 5 √ √ NA √ √
318 Ca Calcium 6 √ √ NA √ √
319 TBIL Total Bilirubin 5 √ √ NA √ √
319 TBIL Total Bilirubin 6 √ √ NA √ √
320 AST Aspartate 5 √ √ NA √ √
Aminotransferase
320 AST Aspartate 6 √ √ NA √ √
Aminotransferase
321 ALKP Alkaline 5 √ √ NA √ √
Phosphatase
321 ALKP Alkaline 6 √ √ NA √ √
Phosphatase
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
100 – 2017-12-19
Laboratory Information System (LIS) Guide Analyte Codes
Analyte Report Assay Name Body VITROS VITROS® VITROS VITROS VITROS®
Code Name Fluid 5600 XT 7600 3600 4600 XT 3400
Body Fluid Key: 2 CSF, 3 Urine, 4 Whole Blood, 5 Serum, 6 Plasma, 7 Amniotic Fluid, 8 Reserved Fluid 1,
9 Reserved Fluid 2, 10 Reserved Fluid 3, UD User Defined
322 ALT Alanine 5 √ √ NA √ √
Aminotransferase
322 ALT Alanine 6 √ √ NA √ √
Aminotransferase
323 LDH Lactate 5 √ √ NA √ √
Dehydrogenase
323 LDH Lactate 6 √ √ NA √ √
Dehydrogenase
324 CK Creatine Kinase 5 √ √ NA √ √
324 CK Creatine Kinase 6 √ √ NA √ √
325 LIPA Lipase 5 √ √ NA √ √
325 LIPA Lipase 6 √ √ NA √ √
326 GGT Gamma 5 √ √ NA √ √
Glutamyltransferase
326 GGT Gamma 6 √ √ NA √ √
Glutamyltransferase
327 Bc Conjugated 5 √ √ NA √ √
Bilirubin
327 Bc Conjugated 6 √ √ NA √ √
Bilirubin
328 THEO Theophylline 5 √ √ NA √ √
328 THEO Theophylline 6 √ √ NA √ √
329 CKMB CKMB 5 √ √ NA √ √
330 Mg Magnesium 3 √ √ NA √ √
330 Mg Magnesium 5 √ √ NA √ √
330 Mg Magnesium 6 √ √ NA √ √
331 Fe Iron 5 √ √ NA √ √
331 Fe Iron 6 √ √ NA √ √
333 PROT CSF Protein 2 √ √ NA √ √
334 SALI Salicylate 5 √ √ NA √ √
334 SALI Salicylate 6 √ √ NA √ √
335 ALC Alcohol 5 √ √ NA √ √
335 ALC Alcohol 6 √ √ NA √ √
336 AMON Ammonia 6 √ √ NA √ √
337 CHE Cholinesterase 5 √ √ NA √ √
337 CHE Cholinesterase 6 √ √ NA √ √
338 AcP Acid Phosphatase 5 √ √ NA √ √
338 AcP Acid Phosphatase 6 √ √ NA √ √
340 Li Lithium 5 √ √ NA √ √
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 101
Analyte Codes Laboratory Information System (LIS) Guide
Analyte Report Assay Name Body VITROS VITROS® VITROS VITROS VITROS®
Code Name Fluid 5600 XT 7600 3600 4600 XT 3400
Body Fluid Key: 2 CSF, 3 Urine, 4 Whole Blood, 5 Serum, 6 Plasma, 7 Amniotic Fluid, 8 Reserved Fluid 1,
9 Reserved Fluid 2, 10 Reserved Fluid 3, UD User Defined
340 Li Lithium 6 √ √ NA √ √
341 DGXN Digoxin 5 √ √ NA √ √
341 DGXN Digoxin 6 √ √ NA √ √
342 PHBR Phenobarbital 5 √ √ NA √ √
342 PHBR Phenobarbital 6 √ √ NA √ √
343 PHYT Phenytoin 5 √ √ NA √ √
343 PHYT Phenytoin 6 √ √ NA √ √
344 CRP C Reactive 5 √ √ NA √ √
Protein
344 CRP C Reactive 6 √ √ NA √ √
Protein
345 CRBM Carbamazepine 5 √ √ NA √ √
345 CRBM Carbamazepine 6 √ √ NA √ √
347 ACET Acetominophen 5 √ √ NA √ √
347 ACET Acetominophen 6 √ √ NA √ √
348 UPRO Urine Protein 3 √ √ NA √ √
353 CRPJ CRPJ 5 √ √ NA √ √
353 CRPJ CRPJ 6 √ √ NA √ √
354 ALTJ ALTJ 5 √ √ NA √ √
354 ALTJ ALTJ 6 √ √ NA √ √
355 ASTJ ASTJ 5 √ √ NA √ √
355 ASTJ ASTJ 6 √ √ NA √ √
356 dHDL Direct HDLC 5 √ √ NA √ √
356 dHDL Direct HDLC 6 √ √ NA √ √
357 ALTV Alanine 5 √ √ NA √ √
Aminotransferase
357 ALTV Alanine 6 √ √ NA √ √
Aminotransferase
358 ALT2 Alanine 5 √ √ NA √ √
Aminotransferase
358 ALT2 Alanine 6 √ √ NA √ √
Aminotransferase
500 TRFRN Transferrin 5 √ √ NA √ NA
500 TRFRN Transferrin 6 √ √ NA √ NA
501 IgG IgG 5 √ √ NA √ NA
501 IgG IgG 6 √ √ NA √ NA
502 IgM IgM 5 √ √ NA √ NA
502 IgM IgM 6 √ √ NA √ NA
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
102 – 2017-12-19
Laboratory Information System (LIS) Guide Analyte Codes
Analyte Report Assay Name Body VITROS VITROS® VITROS VITROS VITROS®
Code Name Fluid 5600 XT 7600 3600 4600 XT 3400
Body Fluid Key: 2 CSF, 3 Urine, 4 Whole Blood, 5 Serum, 6 Plasma, 7 Amniotic Fluid, 8 Reserved Fluid 1,
9 Reserved Fluid 2, 10 Reserved Fluid 3, UD User Defined
503 IgA IgA 5 √ √ NA √ NA
503 IgA IgA 6 √ √ NA √ NA
504 PALB Prealbumin 5 √ √ NA √ NA
504 PALB Prealbumin 6 √ √ NA √ NA
505 mALB Microalbumin 3 √ √ NA √ NA
506 ApoA1 ApoA1 5 √ √ NA √ NA
506 ApoA1 ApoA1 6 √ √ NA √ NA
507 ApoB ApoB 5 √ √ NA √ NA
507 ApoB ApoB 6 √ √ NA √ NA
508 hsCRP high sensitivity 5 √ √ NA √ NA
CRP
508 hsCRP high sensitivity 6 √ √ NA √ NA
CRP
509 C3 C3 5 √ √ NA √ NA
509 C3 C3 6 √ √ NA √ NA
510 C4 C4 5 √ √ NA √ NA
510 C4 C4 6 √ √ NA √ NA
511 VALP Valproic Acid 5 √ √ NA √ NA
511 VALP Valproic Acid 6 √ √ NA √ NA
512 GENT Gentamicin 5 √ √ NA √ NA
512 GENT Gentamicin 6 √ √ NA √ NA
513 TOBRA Tobramycin 5 √ √ NA √ NA
513 TOBRA Tobramycin 6 √ √ NA √ NA
514 VANC Vancomycin 5 √ √ NA √ NA
514 VANC Vancomycin 6 √ √ NA √ NA
515 CAFFN Caffeine 5 √ √ NA √ NA
515 CAFFN Caffeine 6 √ √ NA √ NA
517 dLDL direct LDL 5 √ √ NA √ NA
517 dLDL direct LDL 6 √ √ NA √ NA
520 RF Rhematoid Factor 5 √ √ NA √ NA
520 RF Rhematoid Factor 6 √ √ NA √ NA
521 AMPH Amphetamines 3 √ √ NA √ NA
522 BARB Barbiturates 3 √ √ NA √ NA
523 BENZ Benzodiazepines 3 √ √ NA √ NA
524 THC Cannabinoids 3 √ √ NA √ NA
525 COCM Cocaine 3 √ √ NA √ NA
Metabolite
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 103
Analyte Codes Laboratory Information System (LIS) Guide
Analyte Report Assay Name Body VITROS VITROS® VITROS VITROS VITROS®
Code Name Fluid 5600 XT 7600 3600 4600 XT 3400
Body Fluid Key: 2 CSF, 3 Urine, 4 Whole Blood, 5 Serum, 6 Plasma, 7 Amniotic Fluid, 8 Reserved Fluid 1,
9 Reserved Fluid 2, 10 Reserved Fluid 3, UD User Defined
526 METD Methoadone 3 √ √ NA √ NA
527 OP-LO Opiates-Low 3 √ √ NA √ NA
Cutoff
528 PCP Phencyclidine 3 √ √ NA √ NA
530 dTIBC Direct TIBC 5 √ √ NA √ NA
531 ASO Anti-streptolysin O 5 √ √ NA √ NA
531 ASO Anti-streptolysin O 6 √ √ NA √ NA
533 HPT Haptoglobin 5 √ √ NA √ NA
533 HPT Haptoglobin 6 √ √ NA √ NA
534 AAT Alpha-1- 5 √ √ NA √ NA
antitrypsin
535 dHA1c Hemoglobin A1c 4 √ √ NA √ NA
536 dHb Hemoglobin 4 √ √ NA √ NA
537 OP-HI Opiates-High 3 √ √ NA √ NA
Cutoff
539 A1c Hemoglobin A1c 4 √ √ NA √ NA
540 Hb Hemoglobin 4 √ √ NA √ NA
542 HCY2 Homocysteine 5 √ √ NA √ NA
542 HCY2 Homocysteine 6 √ √ NA √ NA
911 LDLmt Derived LDL 5 √ √ NA √ NA
911 LDLmt Derived LDL 6 √ √ NA √ NA
950 HEM Hemolysis 2 √ √ √ √ √
950 HEM Hemolysis 5 √ √ √ √ √
950 HEM Hemolysis 6 √ √ √ √ √
951 ICT Icterus 2 √ √ √ √ √
951 ICT Icterus 5 √ √ √ √ √
951 ICT Icterus 6 √ √ √ √ √
952 TUR Turbidity 2 √ √ √ √ √
952 TUR Turbidity 5 √ √ √ √ √
952 TUR Turbidity 6 √ √ √ √ √
980 UDA01 (User-Defined) UD √ √ NA √ NA
981 UDA02 (User-Defined) UD √ √ NA √ NA
982 UDA03 (User-Defined) UD √ √ NA √ NA
983 UDA04 (User-Defined) UD √ √ NA √ NA
984 UDA05 (User-Defined) UD √ √ NA √ NA
985 UDA06 (User-Defined) UD √ √ NA √ NA
986 UDA07 (User-Defined) UD √ √ NA √ NA
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
104 – 2017-12-19
Laboratory Information System (LIS) Guide Analyte Codes
Analyte Report Assay Name Body VITROS VITROS® VITROS VITROS VITROS®
Code Name Fluid 5600 XT 7600 3600 4600 XT 3400
Body Fluid Key: 2 CSF, 3 Urine, 4 Whole Blood, 5 Serum, 6 Plasma, 7 Amniotic Fluid, 8 Reserved Fluid 1,
9 Reserved Fluid 2, 10 Reserved Fluid 3, UD User Defined
987 UDA08 (User-Defined) UD √ √ NA √ NA
988 UDA09 (User-Defined) UD √ √ NA √ NA
989 UDA10 (User-Defined) UD √ √ NA √ NA
990 UDA11 (User-Defined) UD √ √ NA √ NA
991 UDA12 (User-Defined) UD √ √ NA √ NA
992 UDA13 (User-Defined) UD √ √ NA √ NA
993 UDA14 (User-Defined) UD √ √ NA √ NA
994 UDA15 (User-Defined) UD √ √ NA √ NA
995 UDA16 (User-Defined) UD √ √ NA √ NA
996 UDA17 (User-Defined) UD √ √ NA √ NA
997 UDA18 (User-Defined) UD √ √ NA √ NA
998 UDA19 (User-Defined) UD √ √ NA √ NA
999 UDA20 (User-Defined) UD √ √ NA √ NA
*Product in development.
Derived Assays
Analyte Report Assay Name Body VITROS® VITROS® VITROS® VITROS® VITROS®
Code Name Fluid 5600 XT 7600 3600 4600 XT 3400
Body Fluid Key: 2 CSF, 3 Urine, 4 Whole Blood, 5 Serum, 6 Plasma, 7 Amniotic Fluid, 8 Reserved Fluid 1,
9 Reserved Fluid 2, 10 Reserved Fluid 3
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 105
Analyte Codes Laboratory Information System (LIS) Guide
Analyte Report Assay Name Body VITROS® VITROS® VITROS® VITROS® VITROS®
Code Name Fluid 5600 XT 7600 3600 4600 XT 3400
Body Fluid Key: 2 CSF, 3 Urine, 4 Whole Blood, 5 Serum, 6 Plasma, 7 Amniotic Fluid, 8 Reserved Fluid 1,
9 Reserved Fluid 2, 10 Reserved Fluid 3
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
106 – 2017-12-19
Laboratory Information System (LIS) Guide Diluent Codes
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 107
Result Classification Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
108 – 2017-12-19
Laboratory Information System (LIS) Guide Result Classification
Analyte Assay Name Result Class 1 Result Class 2 Result Class 3 Result
Code (Q) (R) (S) Class 4 (T)
524 THC Negative Positive
525 COCM Negative Positive
526 METD Negative Positive
527 OP-LO Negative Positive
528 PCP Negative Positive
537 OP-HI Negative Positive
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 109
Standard and Conventional Units for Assays Laboratory Information System (LIS) Guide
A1c g/dL - - √ √ NA √ NA
AAT mg/dL - g/L √ √ NA √ NA
ACET μg/mL mg/dL μmol/L √ √ NA √ √
AcP U/L ηkat/L U/L √ √ NA √ √
AFP ng/mL IU/mL - √ √ √ NA NA
aHBc - - - √ √ √ NA NA
aHBs IU/L mIU/mL - √ √ √ NA NA
aHCV - - - √ √ √ NA NA
aHIV - - - √ √ √ NA NA
ALB g/dL μmol/L g/L √ √ NA √ √
ALC mg/dL g/L mmol/L √ √ NA √ √
ALKP U/L μkat/L U/L √ √ NA √ √
ALT U/L μkat/L U/L √ √ NA √ √
ALTV U/L μkat/L U/L √ √ NA √ √
ALTJ U/L μkat/L U/L √ √ NA √ √
ALT2 U/L μkat/L U/L √ √ NA √ √
AMON μmol/L μg/dL μmol/L √ √ NA √ √
AMPH ng/mL - ug/L √ √ NA √ √
AMYL U/L μkat/L U/L √ √ NA √ √
ApoA1 mg/dL - g/L √ √ NA √ NA
ApoB mg/dL - g/L √ √ NA √ NA
ASO lU/mL KIU/L IU/mL √ √ NA √ NA
AST U/L μkat/L U/L √ √ NA √ √
ASTJ U/L μkat/L U/L √ √ NA √ √
B12 pg/mL - - √ √ √ NA NA
BARB ng/mL - ug/L √ √ NA √ NA
Bc mg/dL mg/L μmol/L √ √ NA √ √
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
110 – 2017-12-19
Laboratory Information System (LIS) Guide Standard and Conventional Units for Assays
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 111
Standard and Conventional Units for Assays Laboratory Information System (LIS) Guide
Folat ng/mL - - √ √ √ NA NA
FSH IU/L mIU/mL - √ √ √ NA NA
FT3 pmol/L pg/mL - √ √ √ NA NA
FT3 II nmol/L ng/mL - √ √ √ NA NA
FT4 pmol/L ng/dL - √ √ √ NA NA
FT4 II nmol/L µg/dL - √ √ √ NA NA
GENT μg/mL mg/L μmol/L √ √ NA √ NA
GGT U/L μkat/L U/L √ √ NA √ √
GLU mg/dL g/L mmol/L √ √ NA √ √
HAV M - - - √ √ √ NA NA
HAV T - - - √ √ √ NA NA
Hb g/dL - - √ √ NA √ NA
HBc M - - - √ √ √ NA NA
HBCon % reduction - - √ √ √ NA NA
HBeAg - - - √ √ √ NA NA
HBsAg - - - √ √ √ NA NA
HCY2 umol/L - - √ √ NA √ NA
HEM - - - √ √ √ √ √
HIV c - - - √ √ √ NA NA
HPT mg/dL - g/L √ √ NA √ √
hsCRP mg/L - mg/L √ √ NA √ NA
ICT - - - √ √ √ √ √
IgA mg/dL - mg/L √ √ NA √ NA
IGFBP ng/ml - ug/L √ √ √ NA NA
IgG mg/dL - g/L √ √ NA √ NA
IgM mg/dL - mg/L √ √ NA √ NA
INS uIU/mL pmol/L - √ √ √ NA NA
iPTH pg/mL pmol/L - √ √ √ NA NA
K+ mmol/L - mmol/L √ √ NA √ √
LAC mmol/L mg/dL mmol/L √ √ NA √ √
LDH U/L μkat/L U/L √ √ NA √ √
LH IU/L mIU/mL - √ √ √ NA NA
Li mmol/L mEq/L mmol/L √ √ NA √ √
LIPA U/L μkat/L U/L √ √ NA √ √
mALB mg/L - - √ √ NA √ NA
METD ng/mL - ug/L √ √ NA √ NA
Mg mg/dL mEq/L mmol/L √ √ NA √ √
Na+ mmol/L - mmol/L √ √ NA √ √
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
112 – 2017-12-19
Laboratory Information System (LIS) Guide Standard and Conventional Units for Assays
NCHEK - - - √ √ √ NA NA
NTx nM BCE - - √ √ √ NA NA
OP-HI ng/mL - ug/L √ √ NA √ NA
OP-LO ng/mL - ug/L √ √ NA √ NA
PALB mg/dL - mg/L √ √ NA √ NA
PCP ng/mL - ug/L √ √ NA √ NA
PHBR μg/mL mg/L μmol/L √ √ NA √ √
PHOS mg/dL mg/L mmol/L √ √ NA √ √
PHYT μg/mL mg/L μmol/L √ √ NA √ √
Prog nmol/L ng/mL - √ √ √ NA NA
Prol mIU/mL ng/mL - √ √ √ NA NA
PROT mg/dL g/L mg/L √ √ NA √ √
PSA ng/mL µg/L - √ √ √ NA NA
RF IU/mL - kU/L √ √ NA √ NA
RUB G IU/mL - - √ √ √ NA NA
RUB M - - - √ √ √ NA NA
SALI mg/dL mg/L mmol/L √ √ NA √ √
T3U % uptake Unit - √ √ √ NA NA
value
TBIL mg/dL mg/L μmol/L √ √ NA √ √
Testo nmol/L ng/mL - √ √ √ NA NA
THC ng/mL - ug/L √ √ NA √ NA
THEO μg/mL - μmol/L √ √ NA √ √
TIBC μg/dL mg/L μmol/L NA NA NA √ √
TIMP2 ng/ml - ug/L √ √ √ NA NA
TOBRA μg/mL mg/L μmol/L √ √ NA √ NA
Tox G IU/mL - - √ √ √ NA NA
Tox M - - - √ √ √ NA NA
TP g/dL - g/L √ √ NA √ √
TRFRN mg/dL - g/L √ √ NA √ NA
TRIG mg/dL g/L mmol/L √ √ NA √ √
Trop ng/mL µg/L - √ √ √ NA NA
TSH µIU/L µIU/mL - √ √ √ NA NA
TT3 nmol/L ng/mL - √ √ √ NA NA
TT4 nmol/L µg/dL - √ √ √ NA NA
TUR - - - √ √ √ √ √
UPRO mg/dL mg/L g/L √ √ NA √ √
UREA mg/dL mg/dL(A) mmol/L √ √ NA √ √
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 113
Standard and Conventional Units for Assays Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
114 – 2017-12-19
Laboratory Information System (LIS) Guide Standard and Conventional Units for Assays
Derived Assays
Assay Conventional Alternate International VITROS® VITROS® VITROS® VITROS® VITROS®
Units Units Units 5600 XT 7600 3600 4600 XT 3400
%A1c %NGSP - - √ √ NA √ NA
%MB % % % √ √ NA √ √
%SAT % - - √ √ NA √ NA
A/G - - - √ √ NA √ √
AGP mmol/L - mmol/L √ √ NA √ √
AGPK mmol/L - mmol/L √ √ NA √ √
C/H - - - √ √ NA √ √
d%A1c %NGSP - - √ √ NA √ NA
DBIL mg/dL mg/L μmol/L √ √ NA √ √
DELB mg/dL mg/L μmol/L √ √ NA √ √
FT3I - - - √ √ √ NA NA
FT4I - - - √ √ √ NA NA
GLOB g/dL - g/L √ √ NA √ √
HbA1c mmol/mol - - √ √ NA √ NA
L/F - - - √ √ √ NA NA
LDL mg/dL g/L mmol/L √ √ NA √ √
mmA1 mmol/mol - - √ √ NA √ NA
c
NBIL mg/dL mg/L μmol/L √ √ NA √ √
OSMO mosm/kg - mosm/kg √ √ NA √ √
T3/T4 - - - √ √ √ NA NA
U/CR - - - √ √ NA √ √
VLDL mg/dL g/L mmol/L √ √ NA √ √
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 115
Result Codes Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
116 – 2017-12-19
Laboratory Information System (LIS) Guide Result Codes
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 117
Result Codes Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
118 – 2017-12-19
Laboratory Information System (LIS) Guide Result Codes
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 119
Result Codes Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
120 – 2017-12-19
Laboratory Information System (LIS) Guide Result Codes
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 121
Result Codes Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
122 – 2017-12-19
Laboratory Information System (LIS) Guide Result Codes
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 123
Result Codes Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
124 – 2017-12-19
Laboratory Information System (LIS) Guide Sample Indices Flags Sent to the LIS
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 125
Download Messages Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
126 – 2017-12-19
Laboratory Information System (LIS) Guide Download Messages
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 127
Download Messages Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
128 – 2017-12-19
Laboratory Information System (LIS) Guide Mechanical/Electrical Interfaces for Serial Communications
Cable
The user supplies the standard interface cable. Cable configurations are determined by the
interface to the laboratory computer. Most computer systems have an RS-232-compatible serial
port, and emulate DTE or DCE (Data Communications Equipment). A null-modem cable is used if
the laboratory computer is a DTE emulator. A straight-through cable is needed if the laboratory
computer is a DCE emulator. For hardware flow control, the laboratory computer output signal is
connected to the System’s Clear to Send (CTS) input. On a System, a cable length of no more than
50 feet (5.24 meters) is recommended to maintain electrical signal characteristics defined by
standard EIA RS-232C.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 129
Mechanical/Electrical Interfaces for Serial Communications Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
130 – 2017-12-19
Laboratory Information System (LIS) Guide Mechanical/Electrical Interfaces for Serial Communications
CTS Hardware Flow Not Functional (DTS/DSR Handshake and CTS/RTS Loopback)
Either System can sense when the other goes on-line or off-line.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 131
Mechanical/Electrical Interfaces for Serial Communications Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
132 – 2017-12-19
Laboratory Information System (LIS) Guide Mechanical/Electrical Interfaces for Serial Communications
Electrical Interface
The System operates interface signals according to the voltage levels and electrical characteristics
defined by EIA Standard RS-232 (August 1969), which are +5 V to +25 V for a SPACE (logic 0)
and –5 V to –25 V for a MARK (logic 1). The System uses the NRZ encoding technique with
signal transitions between +12 V (logic 0) and –12 V (logic 1).
The DSR signal is used as the on-line indicator from the laboratory computer. If the DSR goes
OFF during transmission, an error is reported. No further communication is attempted until the
DSR signal is active. If DSR goes ON, the communication will resume with the Header Record.
When DSR switches to OFF during communication, a single ATTENTION level condition is
reported.
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
2017-12-19 – 133
Network Communications through Ethernet Laboratory Information System (LIS) Guide
VITROS 5600 and XT 7600 Integrated, VITROS 4600 and XT 3400 Chemistry, and VITROS 3600 Immunodiagnostic Systems
134 – 2017-12-19
Ortho-Clinical Diagnostics