Man 4020 PDF
Man 4020 PDF
Man 4020 PDF
MODEL 4020
Standard Software Version
P/N M74661
4/24/15
DANGER
Toxic and/or flammable gases or liquids may be present in this monitoring system.
Personal protective equipment may be required when servicing this instrument.
Hazardous voltages exist on certain components internally which may persist for
a time even after the power is turned off and disconnected.
Only authorized personnel should conduct maintenance and/or servicing. Before
conducting any maintenance or servicing, consult with authorized
supervisor/manager.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from
defects of material and of construction, and that our liability shall be limited to replacing or
repairing at our factory (without charge, except for transportation), or at customer plant at
our option, any material or construction in which defects become apparent within one year
from the date of shipment, except in cases where quotations or acknowledgements provide
for a shorter period. Components manufactured by others bear the warranty of their
manufacturer. This warranty does not cover defects caused by wear, accident, misuse,
neglect or repairs other than those performed by Teledyne or an authorized service center.
We assume no liability for direct or indirect damages of any kind and the purchaser by the
acceptance of the equipment will assume all liability for any damage which may result from
its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus,
and to make any alterations in the dimensions, shape or weight of any parts, in so far as
such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by which
valuable data can be gathered. The information provided by the instrument may assist the user
in eliminating potential hazards caused by his process; however, it is essential that all
personnel involved in the use of the instrument or its interface be properly trained in the
process being measured, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent
danger, it has no control over process conditions, and it can be misused. In particular, any
alarm or control systems installed must be tested and understood, both as to how they
operate and as to how they can be defeated. Any safeguards required such as locks, labels,
or redundancy, must be provided by the user or specifically requested of Teledyne at the
time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser
is responsible for the training of personnel, for providing hazard warning methods and
instrumentation per the appropriate standards, and for ensuring that hazard warning devices
and instrumentation are maintained and operated properly.
The instrument for which this manual was supplied may incorporate
one or more options not supplied in the standard instrument. Commonly
available options are listed below, with check boxes. Any that are
incorporated in the instrument for which this manual is supplied are
indicated by a check mark in the box.
Safety Messages
Table of Contents
4.6.17.1 Disabled 61
4.6.17.2 User Select 62
4.6.17.3 Default (By Function) 63
4.6.17.4 Observing the gas Stream Concentrations 64
4.6.18 Standby 65
4.7 Advanced User Functions 66
4.7.1 Zero Offset Adjustment 66
4.7.2 Background Gas Selection 67
Maintenance & Troubleshooting ................................................ 69
5.1 Measuring Circuit Electrical Checks 70
5.1.1 Loss of Zero Control 70
5.1.2 Anode Voltage Check 71
5.1.3 Electronic Stability 72
5.1.4 Printed Circuit Board Replacement 72
5.1.5 Collector Cable 72
5.2 Temperature Control Electronic Check 73
5.3 Ignition and/or Flame Guard Circuit Checks 74
5.4 Sampling System 75
5.5 Printed Circuit Board Descriptions 75
5.5.1 Flame Guard and Anode Power Supply PCB 75
5.5.3 Proportional Temperature Controller PCBError! Bookmark not defined.
5.5.4 Electrometer-Amplifier PCB 76
Appendix ...................................................................................... 79
Specifications 79
Application Data 80
Recommended Spare Parts List 81
Drawing List 82
List of Figures
Figure 1-1: Gas Sample System with Optional AutoCal Valves ...... 4
Figure 1-2: Gas Sample System Without Optional AutoCal Valves . 5
Figure 1-3: Flame Ionization Cell ..................................................... 6
Figure 2-1: View Inside Cabinet .................................................... 10
Figure 3-1: Gas Connections......................................................... 15
Figure 3-2: Equipment Interface Connector Pin Arrangement ....... 18
Figure 4-1: Front Panel View of Regulator and Gages .................. 32
Figure 4-2: Typical VFD Screen Layout ........................................ 35
Figure 4-3: Setup Mode Functions ................................................ 39
Figure 4-4: AUTOCAL Screens ..................................................... 42
List of Tables
DANGER
COMBUSTIBLE GAS USAGE
WARNING
1. Switch off the power to the 4020 and remove the mains (line)
power cord from the 4020.
2. Remove all external voltages from the connections to the
alarm contacts.
3. Wait one minute.
4. Discharge the anode supply voltage.
a. Connect one end of an insulated (to 1000 VDC or more)
clip lead to 4020 chassis ground (the standoff for the
upper right corner of the mother PCB).
b. Put one end of a 500V rated 1000 ohm resistor in the
other end of the clip lead.
c. Check the voltage between chassis ground (the standoff
for the upper right corner of the mother PCB) and the top
side of R2 at PCB number B74671. It should be between
-5VDC and +5VDC. If is in that range, the inaccessible
hazardous voltage removal procedure is completed, if not
repeat steps 4.a and 4.b.
If it is absolutely necessary to work inside the instrument with power
on, use the ONE HAND RULE:
Work with one hand only.
Keep the other hand free without contacting any other object. This
reduces the possibility of a ground path through the body in case of
accidental contact with hazardous voltages.
Introduction
1.4 Applications
Monitoring the purity of oxygen, argon, nitrogen and other
gases in the manufacture of semiconductors and related
equipment.
Monitoring hydrocarbon contamination in air liquefaction
and other gas production processes.
Gas purity certification.
Detecting trace hydrocarbons in ambient air.
Detecting atmospheric pollutants.
Cryogenics.
Monitoring for fuel leakage or toxic solvents.
Monitoring hydrocarbons in a wide range of process streams.
Operational Theory
2.1 Introduction
The Model 4020 Total Hydrocarbon Analyzer is composed of three
subsystems:
1. Sample System
2. Detector Cell
3. Electronic Signal Processing, Display and Control
located in line with the sample port. See Figure 1-1 for a flow schematic
for instruments equipped with AutoCal valves and Figure 1-2 for
instruments without the AutoCal option.
The thermistor acts as the sensor in the flame guard circuit. Its
ambient temperature resistance is in the 100 K ohms region. When the
flame is ignited, its resistance is reduced by a factor of 100. The
thermistor is coupled to a semiconductor control circuit on the anode-
flame guard power supply PC board, which will be described in a
following section.
The cell electrodes of both the anode-igniter and flame guard
thermistor are connected to the electronics chassis by means of a plug-in
cable.
The electrode section of the cell may be removed for inspection by
turning off the power, disconnecting the electrode lead plug, and
removing the screws which retain the electrode assembly in the sensor
body.
2.3.1 Electrometer-Amplifier
The collector cable is coupled directly to a coaxial fitting located on
the electrometer-amplifier PC board. The PC board is located on the side
panel next to but outside of the isothermal chamber. See Figure 2-1. and
consists of an electrometer amplifier and an operational amplifier. This
circuit is a very high-gain, current-to-voltage converter circuit, having
chamber rises above 70°C (160°F) a cutoff switch will turn off the
power to the heating element.
Installation
Sensor Vent:
All the gases introduced into the detection cell are vented from a
single fitting labeled SENSOR VENT at the rear of the analyzer.
1. TAI recommends that the cell be permitted to vent directly to
atmosphere wherever possible.
2. If the vent line is required, the installation must include a
drop-out pot to collect any water that is formed by the
burning of the hydrogen fuel.
3. The vent line must be constructed so that the water and dirt
cannot collect in it.
4. Sensor vent line must slope downward so condensed water
will freely drain.
Pin Function
3 + Range ID, 4-20 mA, floating
4 – Range ID, 4-20 mA, floating
5 + % Range, 4-20 mA, floating
6 – % Range, 4-20 mA, floating
8 + Range ID, 0-1 VDC
Examples:
The analog output signal has a voltage which depends on gas
concentration relative to the full scale of the range. To relate the signal
output to the actual concentration, it is necessary to know what range the
instrument is currently on, especially when the analyzer is in the
autoranging mode.
The signal output for concentration is linear over the currently
selected analysis range. For example, if the analyzer is set on a range
that was defined as 0-10 ppm CH4, then the output would be as shown in
Table 3-2.
To provide an indication of the range, the Range ID analog outputs
are used. They generate a steady preset voltage (or current when using
the current outputs) to represent a particular range. Table 3-3 gives the
range ID output for each analysis range.
Pin Contact
45 Threshold Alarm 1, normally closed contact
28 Threshold Alarm 1, moving contact
46 Threshold Alarm 1, normally open contact
42 Threshold Alarm 2, normally closed contact
44 Threshold Alarm 2, moving contact
43 Threshold Alarm 2, normally open contact
36 System Alarm, normally closed contact
20 System Alarm, moving contact
37 System Alarm, normally open contact
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The remote valve connections (described below) provides
signals to ensure that the zero and span gas valves will be
controlled synchronously.
pin # Description
41 Calibration Contact
42 Alarm 2 NC Contact
43 Alarm 2 NO Contact
44 Alarm 2 C Contact
45 Alarm 1 NC Contact
46 Alarm 1 NO Contact
47
48 Exhaust Solenoid Return
49 Zero Solenoid Hot
50 Sample Solenoid Return
Input: The input functions using RS-232 that have been implemented
to date are described in Table 3-8.
3.3.5.2 EFFLUENT
All the gases introduced into the detection cell vent from one fitting
at the rear of the analyzer. TAI recommends that the cell be permitted to
vent directly to the atmosphere wherever possible.
If a vent line is required, the installation must include a drop-out pot
to collect the water that is formed by the burning of the hydrogen fuel.
The vent line must be constructed so that water and dirt cannot collect in
it.
Operation
This section of the manual describes how to setup and operate the
Model 4020 Analyzer. Sections 4.1 through 4.5 describe preliminary
steps and equipment needed to operate the analyzer. Beginning with
Section 4.6, the actual operation of the analyzer is described along with
descriptions of the display prompts, messages and options available to
the user within a menu or sub menu.
4.1 Equipment
The following supporting gases and hardware will be required to
operate the (standard) analyzer:
1. Fuel: A cylinder containing a 40% hydrogen, 60% nitrogen
composition will be required to supply the fuel for the flame
ionization burner. The cylinder is to be equipped with an oil-free
metallic diaphragm regulator (dual stage).
Note: When hydrogen is the parent gas in the application, an
auxiliary source of fuel may not be required. See the
Appendix: Specifications for specific recommendations.
2. Blanket Air: A cylinder of water-pumped (low hydrocarbon)
compressed air will be required to maintain the proper
atmosphere within the cell. The cylinder is to be equipped with
an oil-free, dual stage, metallic diaphragm regulator. Less than
1% of the full scale reading (in the selected measurement range)
of hydrocarbon contamination is advisable, if the instrument‘s
accuracy is to be realized.
3. Zero Gas: A cylinder of the parent (or background) gas,
containing less than 10% full scale (of the narrowest range)
hydrocarbon impurity, will be required to zero standardize the
analyzer. For example, if 10 PPM is the narrowest range of the
instrument, then the zero gas impurities are to be less than 1
PPM. The cylinder is to be equipped with an oil-free, dual stage,
metallic diaphragm regulator.
4.4.1 Air
1. Set the air tank regulator to 30 psig.
2. Adjust the instrument air regulator until the air pressure
gauge reads the recommended air pressure.
After the air is flowing through the sensor and warm-up time has
been completed, activate the following gases:
4.4.4 Fuel
1. Open the main valve on the tank and set the fuel tank pressure
regulator to 30 psig (or to match sample pressure).
4.6.5.1 ENTER
The ENTER key is used in several context-sensitive ways.
1. When the selected option is a function on the MAIN MENU
screen, the function name appears with an arrow next to it.
The ENTER key is used to activate the function. The next
screen for that function or sub function will appear on the
VFD.
2. If the selected option is a modifiable item, the UP or DOWN
keys are used to increment or decrement the item to the value or
action you want. The ENTER key is then used to accept the
value and move you to the next field to continue programming.
3. When the last field is entered, ENTER takes you to the next
screen in the process, or if the process is completed, ENTER
takes you back to the ANALYZE screen.
4.6.5.2 ESCAPE
Pressing the ESCAPE key takes you back to the previous screen.
If you do not wish to continue a function, you can abort the session
by escaping. Escaping a function takes the analyzer back to the previous
screen, or to the ANALYZE Function, depending on the nesting level of
screens in the function you are moving out of.
Note however, that the instrument can still be used for analysis or for
initiating a self-test without entering the password. To defeat security
the password must be changed back to TAI.
Note: If you use password security, it is advisable to keep a copy
of the password in a separate, safe location.
To log in, enter the correct password as described in the next section. To
logout, use the LOGOUT function as described in Section 4.6.8.
Pressing <Enter> will logout the current user and protect the system
from unauthorized changes to parameters. All users will still have access
to the SELF-TEST function and the instrument will function normally in
Analyze Mode. If you press <Escape>, the current password will still be
in effect and the user has not logged off.
Note: Make sure that the proper calibration gas is flowing into the
analyzer before initiating this function. Depending on your
sample system, you may have to flow calibration gas for
several minutes to purge the gas line of sample gas.
Pressing ENTER, the 4020 will begin to capture the individual gain
for the specific ranges that have been set. The first of three screens will
appear, one for high gain, one for medium gain and the last for low
gain. The user can set the offset by pressing ENTER when the
appropriate offset appears on the screen.
The screens appear as follows:
The defeat alarm mode is incorporated into the alarm circuit so that
maintenance can be performed under conditions which would normally
activate the alarms.
Note: If you are using password protection, you will need to enter
your password to access the alarm functions. Follow the
instructions in section 4.6.7.1 to enter your password.
Once you have clearance to proceed, you can enter the
ALARM function.
To configure or set the alarms:
Activate or Defeat Alarms
<Enter> Use the <Enter> key to move to the value you want to
change to define the range you want to work with.
<▲▼> Use the UP/DOWN keys to cycle the numbers up or
down for each digit.
<Enter> Use the <enter> key to move to the next character
placeholder in the number field.
<Enter> When the last digit has been entered, press <Enter> again
to accept the value or <Escape> to abort the numerical
entry.
This same procedure is used to select and define all ranges.
amplifier saturation, the over range readings are accurate UNLESS the
application uses linearization over the selected range.
To setup automatic ranging:
<▲▼> Scroll to the Range function using the UP/DOWN keys.
<Enter> Press <Enter> to activate the RANGE function.
Contacts
Zero Span Range
Low Low Autorange mode
Low High R3
High Low R2
High High R1
2. User Select . The user can select which port opens to the
analyzer.
3. By Function: In this option, a specific gas port opens to the
analyzer depending on the current mode of the analyzer. For
instance, when performing a zero calibration, the ―ZERO‖
port would be open while ―SAMPLE‖ and ―SPAN‖ ports
would be closed.
To enable/disable or select a specific valve selection mode:
<▲▼> From the MAIN MENU, scroll to the VALVE
SELECTIONS function using the UP/DOWN keys.
<Enter> Press <Enter> to bring up the valve selection setup
screen.
4.6.17.1 DISABLED
When the valve selection mode ―Disabled‖ is selected, the gas
connected to the sample port will be fed to the analyzer, i.e. the
―SAMPLE‖ port will always be open as the default stream fed to the
analyzer. During all modes of operation (Analyze, Span, or Zero), the
valve position will not be indicated on the screen as shown below.
or
▲▼> Use the <▲▼> keys to toggle between the options of:
―ALWAYS SAMPV‖ (sample valve)
―ALWAYS SPAN‖ (span valve)
―ALWAYS ZERO‖ (zero valve)
<Enter> Press <Enter> when the appropriate valve selection is
displayed on screen.
<Escape> Use Escape to move back to the Analyze Mode. Note that
the analyzer will now switch gas to the analyzer from the
selected port regardless of whether the analyzer is in
analyze or calibration mode.
with a similar display for a zero calibration except the zero valve will be
indicated as open (―ZERO ZEROV‖).
Or
4.6.18 Standby
This function allows you to place the instrument in STANDBY.
CAUTION: STANDY SHUTS DOWN POWER TO THE DISPLAYS
ONLY. INTERNAL CIRCUITS ARE STILL ENERGIZED
AND ELECTRICAL SHOCK HAZARD STILL EXISTS.
<Enter> Use the <Enter> key to move over to the Offset field.
<▲▼> Use the UP/DOWN keys to change each digits value.
<Enter> Use the <Enter> key to accept a digit and move to the
next character position. The final enter will accept the
value and move the arrow to the function which allows
you to set the gain.
Use the same procedure to set the gain of the instrument. After the
last digit is entered, the final <Enter> press will accept the gain value.
The value you enter for the offset to zero or gain will be
automatically added to the reading. Thus, if you entered -0.10 for zero
offset, the display will show -0.10 for zero concentration.
During calibration, when the instrument enters the zero mode in
AUTO, the instrument will do the work of bringing the reading back to
zero plus the offset value that was entered. If you chose MANual zero
mode, then you must adjust the zero of the instrument to read zero plus
the value entered as offset.
<Enter> Use the <Enter> key to move over to the gas field (O2 or
N2).
<▲▼> Use the UP/DOWN keys to toggle between the two
options..
<Enter> Use the <Enter> key to accept the appropriate
background gas when it is displayed.
2. Then connect the DVM leads across tantalum capacitor C3. Note
the positive polarity is marked on the PCB. The Plus side
connects to ground. Measured voltage should be -15±1 volt.
If correct readings are not obtained at both points, power supply has
failed, and that printed circuit card must be replaced. If correct readings
are noted, the electrometer-amplifier has failed, and that printed circuit
card must be replaced.
Note: TAI recommends that spares for each of the printed
circuits be kept on hand so that service can be restored
immediately. The faulty circuit card can be returned to the
factory for repair. Unless personnel knowledgeable in
electronics are available, do not tamper with the circuits.
intermittent open circuit. If there is, replace the cable. Do not attempt to
repair the cable, as special tooling is required to disassemble and
reassemble the cable plugs.
Appendix
Specifications
Fuel: Per application:
- Mixture of 40% hydrogen, 60% nitrogen
- Hydrogen
Accuracy: 1% of full scale
Electronic response time: 90% within one minute
Noise: Less than ±0.5% of full scale
Drift: Less than 1% of full scale per day
Ranges: 3 ranges user defined from 0 ppm to 6%
(methane equivalent)
Sensitivity: As low as 0-1 PPM methane equivalent
full scale
Output: Linear signal; meter readout
0-1 VDC plus 4-20 mA analog output
Ambient temperature: 40 °F to 110 °F (5 °C to 43 °C)
Flow rate: 100-400 ml per minute of sample
100-200 ml per minute of fuel (200 cu. ft.
cylinder lasts about 1 month)
Power requirement: 115 VAC, 60 Hz, single phase
Power consumption: 140 W maximum
Pressure rating: Up to 50 psi
Dimensions: 8.75 " x 19 " x 15.5 " (H x W x D)
222 mm x 483 mm x 394 mm
Weight: 55 lbs.
Application Data
Model Number:
Serial Number:
Ranges Of Analysis:
Output Signal:
Start-Up Span Setting:
Control Panel Pressure Settings:
Comments:
Blanket Air:
Fuel:
Sample:
Bypass Flowrate: A minimum of 0.25 SCFH is indicated on
the integral flowmeter
Recommended Support Gases: The accessory gases should all be
laboratory analyzed and certified as to
their hydrocarbon content.
Blanket Air: Water pumped, hydrocarbon free,
compressed air.
Zero Gas: Water pumped, hydrocarbon free
(certified)
Span Gas: ppm methane in (certified)
The span gas contains a known amount of methane impurity. The
span gas must be certified as to its methane equivalent impurity.
IMPORTANT: For the best possible results, zero and span
gases that are representative of the parent (or
background) gas should always be used.
This is particularly true in applications where the analyzer is being used
to analyze a number of different samples with varying concentrations of
oxygen. The geometry and temperature of the flame will be somewhat
dependent on the oxygen content of the sample. Alterations in the flame
configuration will lead to differences in output readings. For the best possible
results, TAI recommends that zero and span gases representative of each of
these samples be used to standardize the analyzer.
IMPORTANT: Orders for replacement parts should include the part number
and the model and serial number of the system for which the parts are
intended.
Drawing List
Assemblies
D79158 Assembly, Sensor
PC Board Assemblies
B74671 PCB Assy, Anode Power Supply, Flame Guard
B79153 PCB Assy, Electrometer
C75825A PCB Microprocessor
C65507A PCB Power Supply