Wika TT
Wika TT
Wika TT
further approvals
see page 10
Applications
Special features
Description
Fig. left: Resistance thermometer, model TR34
Resistance thermometers of this series are used as universal Fig. right: M12 x 1 adapter to DIN EN 175301-803
thermometers for the measurement of liquid and gaseous angular connector
media in the range -50 ... +250 °C (-58 ... +482 °F). The
instruments are intrinsically safe and suitable for use in
hazardous areas.
The resistance thermometers of the series TR34 are designed for direct installation into the
process. Using it in an additional thermowell is not advisable.
Specifications
Thermometer with direct sensor output with Pt100 (model TR34-x-Px) and Pt1000 (model TR34-x-Sx)
Temperature range
■■ Class A Without neck tube -30 ... +150 °C (-22 ... +302 °F)
With neck tube -30 ... +250 °C (-22 ... +482 °F)
■■ Class B Without neck tube -50 ... +150 °C (-58 ... +302 °F)
With neck tube -50 ... +250 °C (-58 ... +482 °F)
Temperature at the connector Max. 85 °C (185 °F)
Measuring element ■■ Pt100 (measuring current: 0.1 ... 1.0 mA)
■■ Pt1000 (measuring current: 0.1 ... 0.3 mA)
Connection method ■■ 2-wire The lead resistance is recorded as an error in the measurement.
■■ 3-wire With a cable length of 30 m or longer, measuring deviations can occur.
■■ 4-wire The lead resistance can be ignored.
Tolerance value of the measuring element ■■ Class A
per IEC 60751 ■■ Class B at 2-wire
Electrical connection M12 x 1 circular connector (4-pin)
Explosion protection Intrinsically safe to Ex i (ATEX) gas/dust
(for further information see “Further specifications for explosion-protected version”)
For detailed specifications for Pt sensors, see Technical information IN 00.17 at www.wika.com.
Case
Material Stainless steel
Ingress protection
■■ Case with connected connector IP67 and IP69 per IEC/EN 60529, IP69K per ISO 20653
The stated ingress protection only applies when plugged in using mating
connectors that have the appropriate ingress protection.
■■ Coupler connector, not connected IP67 per IEC/EN 60529
Weight in kg Approx. 0.2 ... 0.7 (depending on version)
Dimensions See “Dimensions in mm”
Marking:
Hazardous gas Temperature Ambient temperature Maximum surface temperature (Tmax) at
atmosphere class range (Ta) the sensor or thermowell tip
II 1G Ex ia IIC T1 - T6 Ga T6 -40 ... +45 °C TM (medium temperature) + self-heating (15 K)
II 1/2G Ex ia IIC T1 - T6 Ga/Gb Pay attention to the specific conditions for safe use.
T5 -40 ... +60 °C
II 2G Ex ia IIC T1 - T6 Gb
T4 -40 ... +85 °C
T3 -40 ... +85 °C
T2 -40 ... +85 °C
T1 -40 ... +85 °C
Safety-related maximum values for the current loop circuit (+ and - connections):
Parameters Hazardous gas Hazardous dust
atmosphere atmosphere
Terminals +/- +/-
Voltage Ui DC 30 V DC 30 V
Current li 120 mA 120 mA
Power Pi 800 mW 750/650/550 mW
Effective internal capacitance Ci 29.7 nF 29.7 nF
Effective internal inductance Li Negligible Negligible
Maximum self-heating at the sensor or thermowell tip 15 K 15 K
Marking:
Marking Temperature Ambient temperature Maximum surface temperature (Tmax) at
class range (Ta) the sensor or thermowell tip
II 1G Ex ia IIC T1 - T6 Ga T6 -50 ... +80 °C TM (medium temperature) + self-heating
II 1/2G Ex ia IIC T1 - T6 Ga/Gb Pay attention to the specific conditions for safe use.
T5 -50 ... +85 °C
II 2G Ex ia IIC T1 - T6 Gb
T4 -50 ... +85 °C
T3 -50 ... +85 °C
T2 -50 ... +85 °C
T1 -50 ... +85 °C
Safety-related maximum values for the current loop circuit (connections in accordance with pin assignment 1 - 4):
Parameters Gas applications Dust applications
Terminals 1-4 1-4
Voltage Ui DC 30 V DC 30 V
Current li 550 mA 250 mA
Power Pi 1.500 mW 750/650/550 mW
Effective internal capacitance Ci Negligible Negligible
Effective internal inductance Li Negligible Negligible
Maximum self-heating at the sensor or thermowell tip (Rth) = 335 K/W (Rth) = 335 K/W
140069565.02
without process
connection
X
Legend:
A (U1) Insertion length
N (MH) Neck length
1) At a process temperature of > 150 °C (302 °F), a neck length N (MH) of 70 mm is necessary, X Height process connection
otherwise N (MH) selectable (55, 65 or 70 mm). Ød Sensor diameter
140069565.02
X
Legend:
A (U2) Insertion length
N (MH) Neck length
1) At a process temperature of > 150 °C (302 °F), a neck length N (MH) of 70 mm is necessary, X Height process connection
otherwise N (MH) selectable (55, 65 or 70 mm). Ød Sensor diameter
Adapter cable M12 to PU-548 Adapter cable for the connection of a model TR34 resistance thermometer to the 14003193
model PU-548 programming unit
M12 x 1 transmitter adapter Adapter for the connection of a resistance thermometer with a DIN EN 175301-803 14069503
to angular connector form A angular connector with a 4 ... 20 mA output signal (data sheet AC 80.17)
DIN EN 175301-803 (yellow female Case: PA
connector element) M12 x 1 Angular
Ambient temperature: -40 ... +115 °C
connector connector
4 ... 20 mA
Union nut: zinc diecast
Contacts: copper-zinc alloy, tin-plated
Dielectric strength: 500 V
Ingress protection: IP65
mA
4 ... 20 mA
M12 x 1 Pt adapter to angular Adapter for the connection of the resistance thermometer with a 14061115
connector DIN EN 175301-803 DIN EN 175301-803 form A angular connector with direct resistance output signal
(black female connector element) (data sheet AC 80.17)
mA
Case: PA
M12 x 1 Angular
connector connector Ambient temperature: -40 ... +115 °C
Union nut: zinc diecast
Contacts: copper-zinc alloy, tin-plated
Dielectric strength: 500 V
Ingress protection: IP65
Sealing for for use with angular connector DIN EN 175301-803-A 11437902
angular connector EPDM, brown
M12 connection cable Cable socket straight, 4-pin, ingress protection IP67 Cable length 2 m 14086880
■■ Temperature range -20 ... +80 °C
Cable length 5 m 14086883
Angled socket, 4-pin, ingress protection IP67 Cable length 2 m 14086889
■■ Temperature range -20 ... +80 °C
Cable length 5 m 14086891
M12 connector Female angled, 4-pin, ingress protection IP67 14136815
Screw connection for conductor cross-section 0.25 … 0.75 mm² (24 …18 AWG)
Cable gland Pg7, cable outer diameter 4 … 6 mm
■■ Temperature range -40 ... +80 °C
■■ Suitable for hazardous areas
14004919.01
TR34
Electrical connection
Load diagram
The permissible load depends on the loop supply voltage. 1083
0 10 24 30 36
Voltage UB in V
Certificates (option)
Explosion protection
Ordering information
Model / Explosion protection / Output signal / Transmitter temperature unit / Process temperature / Transmitter initial value /
Transmitter end value / Process connection / Sensor diameter / Insertion length A (U1) or A (U2) / Neck length N (MH) /
Accessories / Certificates
© 11/2014 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specifications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modifications to the specifications and materials.