Preparing and Evaluating Specimens For Automatic Inclusion Assessment of Steel
Preparing and Evaluating Specimens For Automatic Inclusion Assessment of Steel
Preparing and Evaluating Specimens For Automatic Inclusion Assessment of Steel
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
Copyright by ASTM Int'l (all rights reserved); Thu Mar 12 00:51:15 EDT 2015
1
Downloaded/printed by
(Higher Education Commission) pursuant to License Agreement. No further reproductions authorized.
E768 − 99 (2010)
plane of polish. In many cases, the preparation of specimens TABLE 1 Comparison of ANSI (CAMI) versus FEPA versus
for inclusion rating is more readily performed after the Median Diameter of Grit Size in Micrometers
specimens have been hardened by a suitable heat treatment ANSIA (CAMI)B FEPAC Approximate Median
Diameter
procedure (austenize, quench to fully martensitic structure, (µm)
temper at a relatively low temperature). 60 P60 250
5.1.2 Cleanliness is an important consideration in all stages 80 P80 180
of specimen preparation. 100 P100 150
120 P120 125
5.1.2.1 Heat-treated specimens should be wire brushed or 150 P150 90
shot blasted or have the surface ground to remove adherent 180 P180 75
220 P220 63
scale. 240 P240 58.5
5.1.2.2 After completing the grinding steps and before P280 52.2
performing the polishing steps, the specimens and specimen 280 P320 46.2
320 P360 40.5
holders must be cleaned to prevent contamination of the next P400 35.0
preparation step. Cleaning the specimens and specimen holders 360 P500 30.2
between each grinding step can eliminate contamination of 400 P600 25.8
P800 21.8
coarse abrasives to the following finer preparation step. 500 P1000 18.3
5.1.2.3 After the preparation is complete, swab the surface 600 P1200 15.3
800 P2400 8.4
carefully with cotton and a water/soap solution containing a 1200 P4000D 3.0
corrosion inhibitor such as a machine coolant or ethyl alcohol A
ANSI - American National Standards Institute
solution to remove any films or other debris that would B
CAMI - Coated Abrasives Manufacturers Institute
interfere with the inclusion rating. C
FEPA - Federation of European Abrasive Producers
D
Not a FEPA designation
5.1.2.4 It is advisable to perform the inclusion analysis as
soon as possible after preparation to minimize staining or other
problems that can affect the analysis.
5.1.3 The two methods that follow have been found to be
reliable procedures for retaining inclusions in steel and achiev- 5.2.2.1 Low pressures are recommended to ensure the
ing the desired results when evaluated by DIC. There are other retention of an assortment of inclusion types found in a variety
methods that will result in a quality specimen as revealed by of steel grades. The relatively low pressures suggested in this
DIC. Each laboratory should develop preparation procedures procedure will not necessarily result in a satisfactory polish for
for their materials so that the prepared surfaces meet the etching and the further evaluation of the specimen’s general
requirements presented in 6 of this guide. microstructure. The pressures used in the following rigid disk
5.1.4 As described in Prractice E3, the specimens may be procedure are more likely to result in a surface more satisfac-
sectioned and mounted to ease handling during preparation. It tory for revealing general microstructures.
is advisable to use a mounting medium that is hard enough to 5.2.3 Grind the specimens on ANSI 80 grit (P80 FEPA)
preserve edges and maintain flatness. silicon carbide paper to ensure all sectioning artifacts and
5.1.5 Abrasive grit size designations in this guide are deformation damage have been removed and the entire speci-
expressed in the American National Standards Institute (ANSI) men surface is co-planar to the grinding surface.
or Coated Abrasives Manufacturers Institute (CAMI) system
NOTE 2—If the sectioning method resulted in a smooth face and little
units with the corresponding Federation of European Abrasive deformation damage, and if after securing the specimens in a fixture for
Procedure (FEPA) numbers in parentheses. Table 1 provides a polishing, the entire surface of interest is co-planar to the grinding surface,
correlation between these two systems and the approximate then finer grit papers, such as ANSI 180 to 240 (P180 to P240 FEPA) can
median particle diameter for a given grit size in micrometres. be used for the initial grinding step.
5.1.6 Most preparation systems apply pressure on the speci- 5.2.3.1 An adequate flow of water should remove all loos-
mens being processed. The best pressure to be used for each ened abrasive and grinding debris from the paper during the
preparation step should be determined experimentally. Conver- grinding procedure. The flow of water should ensure the
sions between applied force and pressure are discussed in the specimen is kept cool during grinding.
Appendix X1. 5.2.4 Continue grinding through the sequence of silicon
carbide papers listed in Table 2. It may be necessary to clean
NOTE 1—Care must be taken to protect the polished specimen surface
from scratches or contaminants when using a specimen leveling device.
the samples between every grinding step to prevent contami-
nation of the next preparation step.
5.2 Silicon Carbide Procedure: 5.2.5 After completing the entire grinding operation, clean
5.2.1 A summary of the silicon carbide procedure can be the specimens thoroughly, using ethyl alcohol and cotton, then
found in Table 2. rinse and dry. Ultrasonic cleaning can be used.
5.2.2 When using a semi-automatic polishing equipment, 5.2.5.1 The use of a soap and water solution in an ultrasonic
grinding and polishing should be performed using approxi- cleaner can attack non-metallic inclusions in some carbon and
mately 18 kPa pressure per specimen. (For a specimen holder low alloy steels, leading to an exaggeration in the inclusion’s
containing six 32-mm mounts, a force of approximately 87 N apparent size. Adding an inhibitor (such as that used as a
must be applied (see X1.4.2). machining coolant) may reduce this size exaggeration.
Copyright by ASTM Int'l (all rights reserved); Thu Mar 12 00:51:15 EDT 2015
2
Downloaded/printed by
(Higher Education Commission) pursuant to License Agreement. No further reproductions authorized.
E768 − 99 (2010)
TABLE 2 Preparation Method I Silicon Carbide Abrasive Paper Grinding
Surface Coolent/ Abrasive Size/Type Time ForceC Surface Relative Rotation
Lubricant ANSIA [FEPA]B (seconds) newtons [lbs] Speed
RPM
Planar Grinding
Paper Water 80 [P80] grit SIC 60 14 [2] 300 ComplementaryD
Fine Grinding
Paper Water 120 [P120] grit SIC 14 [2] 300 Complementary
Paper Water 240 [P240] grit SIC 60 14 [2] 300 Complementary
Paper Water 320 [P500] grit SIC 60 14 [2] 300 Complimentary
Paper Water 400 [P600] grit SIC 60 14 [2] 300 Complementary
Paper Water 600 [P1200] grit SIC 60 14 [2] 300 Complimentary
Paper Water 800[P2400] grit SIC 60 14 [2] 300 Complementary
Rough Polishing
NAP free cloth Extending fluid 3 µm diamond 50 14 [2] 150 Complementary
Final Polishing
Low nap cloth Extending fluid 1 µm diamond 50 14 [2] 150 Complementary
Optional Polishing (manual)
Low nap cloth Extending fluid 1 µm diamond 50 relatively high NA NAE
A
American National Standards Institute (ANSI) designation of grit size.
B
Federation of European Producers of Abrasives (FEPA) designation of grit size.
C
Force per 3-mm (1.25-in.) diameter specimen.
D
Complementary rotation, surface and specimens spin in same direction.
E
Since this optional step is completed manually, these two categories are not applicable.
5.2.6 Polish the specimens using 3-µm diamond abrasive on 5.3.1 A summary of the rigid grinding disk procedure can be
a low nap cloth, such as woven wool, for 50 s. Clean and dry found in Table 3.
the specimens as described in 5.2.5. 5.3.2 Semi-automatic grinding equipment is required, with
5.2.7 Polish the specimens using a 1-µm diamond abrasive specimens contained in a circular fixture for the entire proce-
on a high nap cloth for 50 s. Clean and dry the specimens as dure. For 5.3.2-5.3.5, a pressure of 42 kPa should be applied
described in 5.2.5. per specimen. (For a specimen holder containing six 32-mm
5.2.8 A final manual polishing step may be added, using circular mounts, a force of approximately 203 N must be
0.25-µm diamond abrasive on a low nap cloth for 10 to 20 s,
applied (see X1.4.3).
using relatively high pressure. Clean and dry the specimens as
described in 5.2.5. 5.3.3 Grind for at least 15 s after achieving a co-planar
5.2.8.1 Ultrasonic cleaning is not recommended after the condition, using a 150-grit alumina grinding stone at 1450 rpm
final manual or automatic polishing step. Ultrasonic cleaning or 80-grit (P80 FEPA) alumina/zirconia grinding paper at 300
can cause cavitation damage, ultimately distorting the inclu- rpm.
sion sizes. 5.3.3.1 An adequate flow of water should remove all loos-
NOTE 3—Depending on the material, it may be advisable to avoid water ened abrasive and grinding debris from the paper during the
after polishing. grinding procedure. The flow of water should ensure the
5.3 Rigid Grinding Disk Procedure: specimen is kept cool during grinding.
Copyright by ASTM Int'l (all rights reserved); Thu Mar 12 00:51:15 EDT 2015
3
Downloaded/printed by
(Higher Education Commission) pursuant to License Agreement. No further reproductions authorized.
E768 − 99 (2010)
5.3.3.2 It may be necessary to clean the samples between condition, using a first order gypsum (1l) plate, is the most
every grinding step to prevent contamination of the next useful mode of DIC illumination (plane areas magenta; inclu-
preparation step. sions dark).
5.3.4 Finish grind using 9-µm diamond on a rigid grinding 6.2 Observation at 100x will reveal whether residual
disk for 3 m at 150 rpm, using an alcohol/glycol lubricant. scratches are present and whether background relief is evident.
Clean the specimens and fixture thoroughly (ultrasonic clean- 6.3 At 500x in the DIC sensitive tint condition, all of the
ing is helpful) with alcohol and cotton, then dry with forced air. edges of inclusion particles should be sharp and recognizable
5.3.4.1 The use of a soap and water solution in an ultrasonic as sharp edges. Narrow bright lines indicating narrow ditches
cleaner can attack non-metallic inclusions in some carbon and at some inclusion edges may, however, still be seen. A test may
low alloy steels, leading to an exaggeration in the inclusion’s be made to determine whether these ditches will be detected by
apparent size. Adding an inhibitor (such as that used as a the image analysis system.
machining coolant) may reduce this size exaggeration. 6.3.1 The polarizer is slowly rotated away from the full DIC
5.3.5 Rough polish using 3-µm diamond on a napless cloth position and the inclusion edges watched for apparent motion.
(silk or non-woven chemotextile) or a low nap cloth (woven There could be a critical trough configuration beyond which
wool) for 4 min at 150 rpm using alcohol/glycol lubricant. the bright troughs turn black at some point in the transition to
Clean as in 5.3.4. bright field illumination, and the inclusions appear to become
5.3.6 Final polish using 1-µm diamond on a high nap cloth larger. This condition indicates insufficient preparation. (The
for 2 min at 150 rpm, using alcohol/glycol lubricant. Clean as use of a filar eyepiece may be helpful at this point to determine
in 5.3.4. whether the inclusion through has widened)
5.3.6.1 Ultrasonic cleaning is not recommended after the 6.3.2 Narrow troughs can be accepted as harmless if they
final polishing step. Ultrasonic cleaning can cause cavitation vanish into the background when the polarizer is opened a
damage, ultimately distorting the inclusion sizes. small part of its range.
5.3.7 An optional oxide-polishing step, using a porous 6.4 Some fine scratches from final polishing will usually be
synthetic cloth with either alkaline colloidal silica or acidic seen under DIC at 500x. These normally also vanish on
colloidal alumina, may be used if necessary to eliminate fine opening the polarizer and, therefore, should not affect quanti-
scratches. The conditions are 90 s at 150 rpm, 100 N, with tative measurement at lower magnification.
water substituted for the colloidal oxide during the last 20 s. NOTE 4—While highly instructive, the DIC examination remains
Clean as in 5.3.4. subjective with respect to the question of whether a reproducible and
accurate surface preparation has been obtained. Tests by actual measure-
6. DIC Evaluation Method5 ment are necessary to determine that errors in specimen preparation were
not contributory to outlier observations.
6.1 A microscope equipped with DIC illumination at mag-
nifications of about 100x and 500x should be used while 7. Keywords
developing the preparation procedure to verify the true surface 7.1 automatic image analysis; differential interference con-
topography of properly prepared specimens. The sensitive tint trast; inclusions; specimen preparation; steel
APPENDIX
(Nonmandatory Information)
X1.1 Automatic preparation machines commonly display and display the applied load accordingly.
applied force in either pound-force (lbf) or newtons (N). The
ability to convert from one unit to another may be necessary X1.3 Caution should be taken when using automated ma-
when trying to interpret a documented procedure. chines that display pressure in pound-force per square inch
(psi). Typically, the machine is displaying the air pressure
X1.1.1 To convert from pound-force to newton, multiply the within the loading cylinder and not the true pressure applied to
pound-force value by 4.5. either the specimen holder or each individual specimen.
X1.1.2 To convert from newton to pound-force, multiply the
X1.4 When converting from a force to a pressure, the
newton value by 0.225.
surface area of the specimen must be determined. The value of
X1.2 When multiple specimens are held in a holder, the force is then divided by the contact area to determine the
applied force must be divided by the number of specimens in required pressure.
the holder to determine the load per specimen. Some auto- X1.4.1 Example 1—What is the force in pound-force if 87 N
mated machines apply the load individually to each specimen is applied?
Copyright by ASTM Int'l (all rights reserved); Thu Mar 12 00:51:15 EDT 2015
4
Downloaded/printed by
(Higher Education Commission) pursuant to License Agreement. No further reproductions authorized.
E768 − 99 (2010)
~ 87 N ! 3 ~ 0.225! 5 19.6 lbf (X1.1) X1.4.3 Example 3—What is the applied load, in newtons,
X1.4.2 Example 2—What is the applied load, in newtons, required to apply a pressure of kPa to an individual specimen
required to apply a pressure of 18 kPa to an individual in a holder containing six 32-mm mounts?
specimen in a holder containing six 32-mm circular mounts? Pressure = (Force)/(Area), therefore . . . Force = (Pressure)
Pressure = (Force)/(Area), therefore . . . Force = (Pressure) × (Area)
× (Area) Area = P r2 = P (16)2 = 804.25 mm2
Area = P r2 = P (16)2 = 804.25 mm2 Total Area = (6 mounts) × (804.25 mm2 = 4825.5 mm2
Total Area = (6 mounts) × (804.25 mm2) = 4825.5 mm2
18 kPa = 18 000 Pa = 18 000 N/m2 = 0.018 N/mm2 42 kPa = 42 000 Pa = 42 000 N/m2 = 0.042 N/mm2
Therefore, Therefore,
Force = (0.018 N/mm2) × 4825.5 mm2) = 86.9 N Force = (0.042 N/mm2) × 4825.5 mm2) = 203 N
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/
COPYRIGHT/).
Copyright by ASTM Int'l (all rights reserved); Thu Mar 12 00:51:15 EDT 2015
5
Downloaded/printed by
(Higher Education Commission) pursuant to License Agreement. No further reproductions authorized.