LSM1106 Practical Manual 20182019
LSM1106 Practical Manual 20182019
LSM1106 Practical Manual 20182019
LSM 1106
Molecular Cell Biology
AY 2018/2019
Semester II
CONTENTS
Introduction
Some points on the laboratory course .................................................................................... 3
Pipettors ................................................................................................................................. 8
Pipette Controller ................................................................................................................. 10
Spectrophotometry ............................................................................................................... 12
1
MINIMAL SAFETY REQUIREMENTS
Lab coats must be worn at all times in the lab, but removed before leaving
the laboratory
Do not eat, drink or apply cosmetics in the lab (especially nail polish –
flammable solvents)
If you have an abrasion or cut on your hands prior to the class, cover it
before you begin work
Wash your hands thoroughly with soap and water before leaving the lab
2
INTRODUCTION
It is important to approach the practical in the right spirit, for an early feel for the way biological
substances, cells and tissues are handled, will stand you in good stead. You should aim at doing
as many of the manipulative procedures as possible with your own hands.
To gain the maximum benefit, it is important to discuss any difficulties with your demonstrator
or lecturer.
Try to plan your work ahead as much as possible; this means reading the text of the practical
and anticipating the various steps you have to perform.
3
EQUIPMENT AND GLASSWARES PROVIDED
Do not crowd round another worker, both for safety reasons and to avoid distraction.
Keep any article liable to be damaged off the bench. This includes textbooks and clothing.
Keep raucous behaviour for outside the laboratory.
Never pour excess reagents back into the original bottles to avoid possible contamination.
Take small quantities for your individual use.
Replace stoppers from common reagent bottles immediately.
Put everything back where you find it.
Dispose of waste appropriately. The following are some general guidelines on the disposal
of different types of wastes:
1. Corrosive wastes - into the specific liquid waste bin indicated.
2. Organic wastes - into the specific liquid waste bin indicated.
3. Solids should go into bins, provided that they are not classified as biological or
sharps waste.
4. Biological waste will be disposed of into specially marked plastic bags. Whereas
the liquid biological waste is disposed into the specific liquid waste bin indicated.
5. Sharps (broken glass, blades and needles) should go into a specially-labelled sharp
disposal box.
4
UNITS AND SIGNIFICANT FIGURES
*In print, the upper case “L” is often used to avoid confusion between one (“1”) and lower case “l”.
In addition, concentration is sometimes still expressed as g/100 mL (see table above). “%”
also means quantity of substance dissolved in 100 mL solution (not solvent). The concentration
M for mol/L is still often used, especially in scientific journals. The formal is never used for
the constituents of body fluids but can be used for lists of reagents, for the sake of brevity, as
is done in this manual. Do not make up your own units.
w/v and w/w are self-explanatory terms giving the relative proportions of solids and liquids in
reagents.
In particular, never put a plural for a unit, nor a full stop, e.g. mol/L (not mol./L. or mols/L or
any similar combination).
As important as the correct use of units is the choice of significant figures appropriate to the
accuracy and precision of the method. [Note, accuracy refers to how near the result is to the
“true” result; the true result is only ascertainable by a reference method and is a concept of
minimal importance in much of biochemistry, where comparisons over time and/or between
groups or individuals are more important. Precision refers to the repeatability of a result; that
is’ the closeness of a number of measurements on the same material.] If you report a serum
glucose result as say 5.2 mmol/L you are reporting two significant figures, which are the digits
required to show the precision with which a measurement can be made. To report 5.2 implies
a result nearer 5.2 than 5.1 or 5.3, or more exactly, you believe the result is between 5.15 and
5.25. Were you to report, say, 5.23 mmol/L, this would imply a result between 5.225 and 5.23,
which already seems optimistic for a biological assay. Zeros to the right of the decimal point
are significant but those on the left may or may not be. For example, in 0.003 and 0.0030 the
zeros are meaningful but in 3000, the digit 3 may be the only significant figure, implying a
5
result somewhere between 2,500 and 3,500. This difficulty is removed by using exponents, for
example 3.0 x 103 specifies two significant figures.
The most skilled operator dealing with biological substances can rarely achieve repeatability
better than 3% - 5%. Therefore, your choice of significant figures should be appropriate to this.
This may be a difficult concept at first but at a minimum it is important not to be deceived by
the capacity of electronic calculators. Lists of recommended significant figures for substances
in human body fluids are published by authorities in the field and can always be consulted if
need be. An example is the excerpt below.
It is essential to record as you go along and not rely on memory. There is scope for creativity
in all laboratory work but at this stage it lies mainly in your personal report.
It should however always be headed with the title of the experiment and the date. The practical
instructions should NOT be copied down in the report. Instead you should note your results,
showing clearly how you worked them out, prepare charts where appropriate, and relate any
difficulties or comments.
You are free to make suggestions as to improvements or even to question the rationale,
objectives and conduct of the session.
Experimental data are best presented in the form of a table and/or graph.
You are strongly encouraged to learn to use software (e.g. Excel) to plot the graphs.
7
PIPETTORS
In clinical and research laboratories, mouth pipettes have mostly been replaced by specially
designed pipettors. These pipettors are designed to measure and transfer liquids accurately and
safely. They are expensive, which is why you still have to use transfer pipettes for much of
your work. Currently, the Gilson or Eppendorf Pipettors, P10, P20, P200 and P1000 are
provided in the biochemistry laboratory (Figure 1). These numbers represent the maximum
volumes of respectively 10, 20, 200 and 1000 l intended for use on each pipettor. This is
clearly labelled on the push-button.
Pay careful attention to the use of pipettors. You must know how to operate the pipettors
correctly, as it will affect all your subsequent experiments.
OPERATION
1. Ensure that you have chosen the correct volume pipettor, e.g. P10 for 1 to 10 L, P20 for
2 to 20 L, P200 for 20 to 200 L, P1000 for 100 to 1000 L.
2. Setting the volume: The pipettor consists of two or three numbers and is read from top to
bottom. The numbers indicate the volume selected, e.g. 100 is 100 L. The volume is set
by turning the adjustment ring slowly to increase or decrease the volume, making sure not
to overshoot the mark.
3. Press a plastic tip firmly on the shaft of the pipettor to ensure a positive, airtight seal.
4. Aspiration: Press the push-button to the first positive stop. Holding the pipette vertically,
immerse the tip into the sample liquid. Next, release the push-button slowly and smoothly
to aspirate the sample. Wipe the outside of the tip, without touching the orifice.
5. Dispensing: Place the end of the tip against the inside wall of the vessel at an angle. Press
the push-button to the first stop to release the sample and if necessary, press to the second
stop to expel any remaining liquid. Remove the pipettor and release the push-button.
6. Eject the tip by pressing the tip ejector button.
8
PRECAUTIONS AND CARE
9
PIPETTE CONTROLLER
CAUTION: Always use a pipette controller whenever you are working with biological,
caustic or toxic samples. Do not pipette with your mouth.
1. The pipette controller (Figure 2) serves to facilitate the pipetting of liquids with graduated
pipettes, volumetric pipettes and blow-out pipettes of either glass or plastic composition
within the volume range of 0.1 mL to 100 mL. If the instrument is correctly used, the liquid
comes into contact with the pipette only.
2. The pipette controller must not be used with vaporous liquids that are incompatible with
polypropylene (PP), polytetrafluorethylene (PTFE) and silicone.
3. Pipetting:
Compress the suction bellows – Before attaching the pipette,
squeeze the suction bellows.
10
4. Attach the pipette:
Hold the pipette as near to its upper end as possible, and
carefully insert it into the adapter until it fits tightly.
Attention: Never use force. Thin pipettes are particularly
liable to break. Avoid the risk of injury!
Once the pipette has been securely attached, always hold the
pipette controller in a vertical position, tip down.
Dispense:
Hold the collecting vessel in an inclined position. Place the
pipette tip against the inner vessel wall. Press the pipetting lever
down to dispense.
8. After pipetting:
Hold the pipette at its extreme upper end, and gently twist and
pull it out of the adapter.
11
SPECTROPHOTOMETRY
QUANTITATIVE LAWS OF LIGHT ABSORPTION
When light passes through a solution of a substance the chance of a quantum of light being
absorbed is proportional to the number of quanta entering the solution, the number of molecules
absorbing substance in the light and to the chance of a quantum being absorbed if it “hits” a
molecule.
This proportionality explains the two empirical laws of light absorption, Beer’s and Lambert’s
Laws. Together they give the Combined Law (Beer-Lambert Law):
log10(I0/I) = A = × c × l
I0 is the incident light, I is the emergent or transmitted light, A is called the absorbance
(alternatively extinction or optical density), (the Greek letter epsilon) is the molar absorption
coefficient, c is the molar concentration (mol.L-1), l is the path length in cm.
Note that A is the logarithm of the ratio of two quantities and it has no units. has the units M-
1
.cm-1 or mol-1.L.cm-1. Note that you also need to specify the wavelength, in nm, which can be
done with a subscript, e.g. A405 is the absorbance at 405 nm.
12
SETTING UP A SPECTROPHOTOMETER
Practical points:
Never:
Stand bottles or cuvettes on top of a spectrophotometer.
Overfill the cuvettes.
Mix different types of cuvette in one set of measurements.
Scratch the optical surfaces.
Hold the cuvette by the optical surfaces.
Always:
Check that you have the correct cuvettes:
o Plastic cuvette for visible (320-800 nm) or near UV light. Some plastic
cuvettes (e.g. Bio-Rad trUView cuvettes) are made of material with low UV
absorbance. These cuvettes are suitable for use at both UV and visible
regions.
o Quartz cuvette for far UV (180-320 nm).
for the Shimadzu UV-1800 Spectrophotometer.
Check that the outside of the cuvette is clean and dry before measuring.
Wipe it gently with kimwipe.
Make sure you put the optical, polished faces of the cuvette in the light beam.
13
Practical 1: pH and Buffers
1. OBJECTIVES
2. INTRODUCTION
Weak acids such as CH3COOH or KH2PO4 are only partially dissociated in solution.
When titrated with a strong base such as NaOH, the pH of the solution changes as a
function of the amount of base added as shown in the figure below.
14
At any point along the titration, the equilibrium can be defined by the equation
Ka
HA H+ + A-
The protons are mainly hydrated in solution and exist as H3O+, but for simplicity we
shall refer to them as H+. Since by definition, pH = - log10[H+] and pKa = - log10[Ka],
the above equation may be rearranged and expressed in the form of the Henderson-
Hasselbalch equation
It can be seen from this relationship between pH and pKa that when the concentration
of acid [HA] and conjugate base [A-] are equal, pH = pKa.
In addition, you will note that on either side of the pKa, the acid-base mixture resists
change in pH upon addition of acid or base, i.e. it acts as a buffer. Buffer solutions are
therefore mixtures of weak acids and their conjugate base.
The pH meter
E = 1n RT [H3O+]A
ZF [H3O+]B
Besides this temperature-dependent variation, the electrode may also vary in efficiency
giving less or even greater potential than that predicted by theory. The scale must
therefore be adjusted to read correctly over the range of pH values being measured.
From this you will realize that the pH meter needs to be calibrated regularly before use.
This can be done using standard buffers of known pH.
Read carefully the instructions on the use of the pH meter which are available near the
instrument. The glass bulb of the pH electrode is extremely fragile and should be
handled with care at all times. Adhere to the following precautions when using the pH
meter.
1. Ensure that the electrodes are always left with the bulb immersed in distilled
water or a suitable buffer.
2. Both before and after use, rinse the electrodes carefully by dipping into a beaker
of distilled water and then squirting with distilled water from a wash bottle. To
prevent diluting the sample, remove excess water from the electrode by blotting
lightly (but not wiping) with a tissue.
3. When taking measurements, check that the bulb of the electrode is completely
immersed in the solution.
4. Remember to calibrate the pH meter with the standard buffers provided before
use. The steps for calibrating the pH meter are as follows: enter the calibration
mode and place the electrode in standard buffer pH 4. Standardize the pH meter
to read exactly 4. Repeat this process to re-standardize the pH meter with
buffers of pH 7 and pH 10 respectively. Exit the calibration mode before using
the pH meter for your experiments.
16
3. EXPERIMENTS
You will perform the titration of histidine solution with NaOH. The titration
curve obtained will be used to determine the pKa and effective buffering range
of histidine.
3.1.1 Materials
Burette
Retort stand
Glass funnel
pH meter
Magnetic stirrer
3.1.2 Reagents
Histidine monohydrochloride (given as dry salt,
MW = 209.63)
0.05 M NaOH
3.1.3 Procedure
1. Each group will be provided with a 50 mL falcon tube containing
histidine monohydrochloride salt. The mass of the salt is indicated
on the tube. Calculate the volume of deionized water required to
prepare 20 mM histidine monohydrochloride solution in the tube.
Vortex the contents until the salts are fully dissolved.
17
2. Design an experiment for the titration of histidine with 0.05 M
NaOH. You will use 20 mL of 20 mM histidine
monohydrochloride prepared in step 1 for this purpose. Construct a
table to record your data. The table must include a column for the
values of mole NaOH/mole histidine.
3. Titrate at 0.5 mL increments of NaOH. The initial pH of the
histidine monohydrochloride solution should be around 4. Stop the
titration when the pH of the solution reaches between pH 11 to
11.5.
3.2.1 Materials
Test tubes
Disposable Pasteur pipettes
Universal pH indicator
3.2.2 Reagents
0.01 M potassium phosphate buffer, pH 7.0 (pKa 6.8)
0.01 M Tris-HCl, pH 7.0 (pKa 8.1)
0.05 M HCl
0.05 M KOH
3.2.3 Procedure
The buffer solutions provided are commonly used in the biochemistry
laboratory.
3.2.4 Questions
1. What conclusions can you draw from your experiments?
2. Which of these two buffers would you use if you were studying
the properties of a phosphatase which functions optimally at pH
7.2? Explain your choice.
19
3.3 Effect of temperature on the pH of a buffer
3.3.1 Materials
Ice box
pH meter
3.3.2 Reagents
0.01 M potassium phosphate buffer, pH 7.0
0.01 M Tris-HCl, pH 7.0
3.3.3 Procedure
Measure the pH of the 0.01 M Tris-HCl and 0.01 M potassium
phosphate buffer provided at room temperature and after cooling to 4C
by placing the solutions in an ice bath.
3.3.4 Questions
What effect does temperature have on the pH of the Tris-buffer and
potassium phosphate buffer? Give an explanation for your observation.
20
Practical 2: Quantitative Protein Estimation
1 OBJECTIVES
1.1 To record a protein spectrum and to estimate the protein concentration using a
direct spectrophotometric method.
1.2 To estimate the protein concentration using a dye-binding (“Bradford”) method.
2 INTRODUCTION
Proteins are ubiquitous components of all living tissues, whether of animal, plant or
bacterial origin. They serve indispensable functions in cellular architecture, catalysis,
regulation, contractile processes and immune defence. Proteins are intimately
concerned with virtually all physiological events.
Protein estimation is therefore very important in every investigation in biochemistry.
For example, laboratory practice in protein purification often requires a rapid and
sensitive method for the quantitation of protein. Presently, a variety of methods are
available for the determination of protein concentration of a given sample. The methods
employ different principles, and may be sensitive to interferences by certain salts,
buffer components and solvents. Each method therefore has certain unique and useful
characteristics as well as limitations. In the biochemical laboratory, the most widely
used methods often employ photometric and/or colorimetric analyses as these methods
are simple, rapid and have the required sensitivities.
3 EXPERIMENTS
3.1.1 Materials
5 mg/mL bovine serum albumin (BSA)
Two samples of unknown concentration
3.1.2 Procedure
1. Prepare 10 mL of 1.0 mg/mL BSA from the 5 mg/mL BSA stock,
using water as the diluent.
2. Construct absorption spectra for the 1.0 mg/mL BSA and the two
unknowns by measuring their absorbances at 10 nm intervals over
21
the range 250-360 nm (start measurement from the longer
wavelength first).
3. Calculate the concentrations of all 3 samples using the respective
absorbance values at 280 nm.
A rough conversion is
3.1.3 Questions
1. Calculate the protein concentration of the prepared BSA and
unknown samples using E1% values of 10.0 and 6.67.
2. What can you surmise by comparing between the absorption spectra
of the 1 mg/mL BSA and the two unknowns?
3.2.1 Materials
5 mg/mL bovine serum albumin (BSA)
Samples of unknown concentration
Dye reagent concentrate
3.2.2 Procedure
1. The linear range of the assay for BSA is 0.1 to 0.8 mg/mL.
2. Prepare five (5) dilutions of BSA within this range from the 1.0
mg/mL BSA prepared earlier, using water as diluent.
3. At the same time, also dilute the unknown samples to within this
linear range by referring to their concentrations estimated in part 1.
4. Pipette 50 L of each standard and unknown sample solution into a
clean, dry tube. Protein solutions are normally assayed in duplicate
or triplicate.
5. For the blank or control, pipette 50 L of water into the tube instead.
22
6. Dilute 1 part Dye Reagent with 4 parts distilled water. Add 2.5 mL
of diluted dye reagent to each tube and vortex. Incubate at room
temperature for at least 5 minutes and not more than one hour.
7. Measure the absorbance at 595 nm, and plot a standard calibration
curve of absorbance against protein (BSA).
8. Estimate the protein concentration of the unknown samples from
their absorbances at 595 nm using the calibration curve.
3.2.3 Questions
1. Construct a protein standard curve based on the dye-binding
(“Bradford”) method, and establish the protein concentration of the
unknown solutions provided.
2. Compare the protein concentration of the unknowns obtained from
the two methods. Comment on the advantages and disadvantages of
the two methods in estimating the protein concentration of an
unknown sample. Which method will provide a more reliable
estimate of the protein concentration in our experiment? Use the
experimental data obtained to support your argument.
4. Calculate the molar extinction coefficient of BSA if a 1.0 mg/mL
solution in buffer has an absorbance of 0.667, and the molecular
weight of BSA is 66,000.
23
Practical 3: Enzyme Kinetics
1. OBJECTIVES
2. INTRODUCTION
3. EXPERIMENTS
3.1.1 Materials
0.1 mM solution of NAD+, pH 7.5
0.1 mM NADH, pH 9.0.
3.1.2 Procedure
Record the absorbance of each solution between 240 to 400 nm at
intervals of 20 nm. With every change in wavelength setting, zero the
24
instrument with a blank containing buffer alone. Why is it necessary to
zero the instrument?
3.2.1 Materials
1 mM NADH in 0.05 M sodium pyrophosphate buffer, pH 9.0
0.05 M sodium pyrophosphate buffer, pH 9.0
3.2.2 Procedure
Prepare a series of tubes containing final concentrations of 0.01 mM to
0.2 mM NADH from 1 mM NADH solution, using the 0.05M sodium
pyrophosphate buffer, pH 9.0 as the diluent. The final volume of each
diluted solution should be 3 mL. Read the absorbance of NADH for each
concentration at 340 nm.
3.2.3 Report
1 Tabulate your data in the form of a table.
2 Plot a graph to show the relationship between absorbance and
concentration of NADH.
3 Calculate the molar (or millimolar) extinction coefficient of NADH,
applying the Beer-Lambert equation.
3.3.1 Materials
200 mM sodium lactate
10 mM NAD+
2 units/mL LDH
0.05 M sodium pyrophosphate buffer, pH 9.0
3.3.2 Procedure
Pipette 0.1 mL NAD+ and 30 l of sodium lactate into a cuvette and
topped it up with buffer to a final volume of 2.8 mL. Zero the
spectrophotometer. Start the reaction by adding 0.2 mL of LDH. Cover
the cuvette with a small piece of parafilm and mix rapidly and
thoroughly by inversion. Monitor the progress of the reaction by
recording A340 at 0.5 min intervals for 4 min. Repeat the above assay
with different volumes (up to 300 L) of sodium lactate, keeping other
25
conditions unchanged. You should have 5 different concentrations of
sodium lactate (final lactate concentration ranges from 2 to 20 mM).
3.3.3 Report
1 Plot the increasing absorbance against the reaction time for each
substrate concentration. Find the initial velocity from the curves
obtained (note: that the value obtained here will be in absorbance
units per min)
2 By using the information from 3.2.3, convert the initial velocity to
moles per min.
3 Plot the Michaelis-Menten curve and Lineweaver-Burk plot (note:
in both plots, the initial velocity should be in moles per min).
Compare and comment on the Km and Vmax values obtained from
these two procedures.
4 What are the advantages and disadvantages of the Michaelis-
Menten curve and the Lineweaver-Burk plot?
26
Practical 3: Enzyme Kinetics
Time A340
(min) Tube 1 Tube 2 Tube 3 Tube 4 Tube 5
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
27
Supplementary Information
for
Practical 3: Enzyme Kinetics
Coenzymes:
NAD+: Nicotinamide adenine dinucleotide, oxidized form
NADH: Nicotinamide adenine dinucleotide, reduced form
NADPH: Nicotinamide adenine dinucleotide phosphate, reduced form
28
An illustration of how NADPH is utilised in metabolism.
The free energy released during metabolism is conserved by the synthesis of ATP from ADP
and phosphate or by the reduction of the coenzyme NADP+ to NADPH. ATP and NADPH
are the major free energy sources for anabolic pathways.
29