Chapter 21
Chapter 21
Chapter 21
Dosimetry
Slide set of 69 slides based on the chapter authored by
J. C. Hourdakis and R. Nowotny
of the IAEA publication (ISBN 978-92-0-131010-1):
IAEA
International Atomic Energy Agency
Chapter 21. TABLE OF CONTENTS
21.1. Introduction
21.2. Radiation detectors and dosimeters
21.3. Ionization chambers
21.4. Semiconductor dosimeters
21.5. Other dosimeters
21.6. Dosimeter calibration
21.7. Instruments for measuring tube voltage and time
21.8. Instruments for occupational and public
exposure measurements
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 2
21.1. INTRODUCTION
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 3
21.1. INTRODUCTION
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 4
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.1. General characteristics of radiation detectors
Dosimeter
• is an instrument that measures dose of ionizing radiation
• usually comprises a measuring assembly - electrometer and one
or more detector assemblies which may or may not be an
integral part of the measuring assembly
• can be classified as:
active or passive
displays the dose cannot display the dose value directly, but
value directly record a signal which must be subsequently
retrieved and converted to dose (or air kerma)
by a reading device
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 5
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.1. General characteristics of radiation detectors
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 6
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.1. General characteristics of radiation detectors
Other instruments are needed to measure:
• X ray tube voltage (kV meter)
• exposure time (timer)
They can be used without direct connection into the
electrical circuits of the X ray units
There are also a variety of devices used for:
• occupational
dose assessment
• public
including
• ionization chambers for direct measurements
• TLD, OSL and film for indirect use as either personal
dosimeters or area monitors
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 7
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2. Properties of diagnostic radiology dosimeters
Many types of
ionization chambers
diagnostic radiology dosimeter
solid state detectors
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 8
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2. Properties of diagnostic radiology dosimeters
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 9
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2. Properties of diagnostic radiology dosimeters
K = M Q kTP N K ,Q0 kQ ∏ k j
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 10
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2. Properties of diagnostic radiology dosimeters
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 11
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2.1. Sensitivity
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 12
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2.1. Sensitivity
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 13
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2.2. Linearity
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 14
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2.3. Energy dependence
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 15
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2.3. Energy dependence
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 16
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2.4. Directional dependence
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 17
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2.4. Directional dependence
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 18
21.2. RADIATION DETECTORS AND DOSIMETERS
21.2.2.5. Leakage current
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 19
21.3. IONIZATION CHAMBERS
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 21
21.3. IONIZATION CHAMBERS
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 22
21.3. IONIZATION CHAMBERS
P0T
The air density correction factor kTP =
PT0
should always be applied to the dosimeter’s readings
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 23
21.3. IONIZATION CHAMBERS
21.3.1. Clinical application of ionization chambers
21.3.1.1. Chambers for air kerma (dose) measurements
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 25
21.3. IONIZATION CHAMBERS
21.3.1.2. Cylindrical pencil type chambers
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 26
21.3. IONIZATION CHAMBERS
21.3.1.3. KAP chambers
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 27
21.3. IONIZATION CHAMBERS
21.3.2. Application hints for ionization chambers
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 28
21.3. IONIZATION CHAMBERS
21.3.2. Application hints for ionization chambers
• Due to their small size and rigidness, they are convenient for use in
many applications
• MOSFETs often require a connection to a bias voltage during irradiation
• They are mainly used in patient dosimetry
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 31
21.4. SEMICONDUCTOR DOSIMETERS
21.4.1. Theory of operation
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 32
21.4. SEMICONDUCTOR DOSIMETERS
21.4.1. Theory of operation
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 33
21.4. SEMICONDUCTOR DOSIMETERS
21.4.2. Application hints for semiconductors
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 34
21.4. SEMICONDUCTOR DOSIMETERS
21.4.2. Application hints for semiconductors
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 35
21.4. SEMICONDUCTOR DOSIMETERS
21.4.2. Application hints for semiconductors
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 36
21.5. OTHER DOSIMETERS
21.5.1. Film dosimetry: radiographic film and radiochromic film
21.5.1.1. Radiographic film
Radiographic film still finds application as a dosimeter in personal
radiation monitoring using film badges
• The emulsion in a film dosimeter directly absorbs ionizing
radiation and can be correlated to the optical density of the
developed film
• The sensitometric curve is very different from that for screen-film
systems
• A radiographic emulsion is far from tissue equivalent and the
energy response of a film badge is modified by addition of several
filters
• The provision, processing and analysis of such dosimeters are
the task of specialized departments and companies, and
commonly not within the duties of a medical physicist
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 37
21.5. OTHER DOSIMETERS
21.5.1.2. Radiochromic film
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 38
21.5. OTHER DOSIMETERS
21.5.1.2. Radiochromic film
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 39
21.5. OTHER DOSIMETERS
21.5.1.2. Radiochromic film
• Their energy response and other properties can differ and the
specifications should be collected from the supplier or from literature
Sensitivity ranges from ~1 mGy to ~50 Gy depending on film type
The sensitometric response is not linear and suitable calibration curves
need to be applied
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 40
21.5. OTHER DOSIMETERS
21.5.2. Thermoluminescent dosimetry (TLD)
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 41
21.5. OTHER DOSIMETERS
21.5.2. Thermoluminescent dosimetry (TLD)
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 42
21.5. OTHER DOSIMETERS
21.5.2. Thermoluminescent dosimetry (TLD)
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 43
21.5. OTHER DOSIMETERS
21.5.2. Thermoluminescent dosimetry (TLD)
typical glow curve for LiF:Mg,Cu,P If a linear temperature ramp is applied the
TL signal (glow-curve) shows various peaks at
TL signal (a.u.)
Temperature (oC)
The reading cycle of a TLD is divided into preheat, signal integration and
annealing
• During preheat the dosimeter is maintained for some seconds at a constant
temperature sufficient to remove all low temperature signals
• Then the temperature is raised up to the maximum temperature
• Finally, the dosimeter is annealed in a dedicated oven to remove all remaining
signals, resetting the dosimeter to zero
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 44
21.5. OTHER DOSIMETERS
21.5.2. Thermoluminescent dosimetry (TLD)
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 45
21.5. OTHER DOSIMETERS
21.5.3. Optically stimulated luminescence (OSL)
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 46
21.5. OTHER DOSIMETERS
21.5.3. Optically stimulated luminescence (OSL)
Typical OSL curve • The intensity of the emitted light is related to
the rate at which the system returns to
equilibrium, resulting in a characteristic
luminescence – time curve
• In a typical measurement using an OSL
dosimeter, the sample material is illuminated with
an appropriate light source
• The emitted light is passed through an optical
filter to suppress unwanted light and then
detected with a photomultiplier tube
• The arrangement of a OSL reader is similar to a
TLD reader
Adapted from E. Yoshimura, 2007
• An improvement in signal to noise can be
achieved by pulsing the stimulating light
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 47
21.5. OTHER DOSIMETERS
21.5.3. Optically stimulated luminescence (OSL)
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 48
21.5. OTHER DOSIMETERS
21.5.3. Optically stimulated luminescence (OSL)
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 49
21.5. OTHER DOSIMETERS
21.5.4. Dosimetric applications of TLD and OSL
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 50
21.5. OTHER DOSIMETERS
21.5.4. Dosimetric applications of TLD and OSL
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 51
21.6. DOSIMETER CALIBRATION
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 52
21.6. DOSIMETER CALIBRATION
21.6.1. Standard free air ionization chamber
Free air ionization chambers are
often used by PSDLs as the primary
standard for the determination of air
kerma against which the secondary
standard chambers from SSDLs are
calibrated
The charge liberated by X rays in the
mass of the air inside the chamber
volume is measured
dEtr dQWair
The air kerma is deduced according to its definition K= =
dm dm
from measurements of basic physical quantities (charge and mass) and
applying physical constants and relative correction factors
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 53
21.6. DOSIMETER CALIBRATION
21.6.2. SSDL calibration
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 54
21.6. DOSIMETER CALIBRATION
21.6.2. SSDL calibration
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 55
21.6. DOSIMETER CALIBRATION
21.6.2. SSDL calibration
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 57
21.6. DOSIMETER CALIBRATION
21.6.2. SSDL calibration
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 58
21.6. DOSIMETER CALIBRATION
21.6.2. SSDL calibration
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 59
21.6. DOSIMETER CALIBRATION
21.6.3. Field calibration
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 60
21.6. DOSIMETER CALIBRATION
21.6.3. Field calibration
M Qfield M Qfield
• ‘field’ and ‘ref’ refer to the field and the reference instruments, respectively
• M values are readings of the reference and the field instruments and have
been corrected for the influence of all quantities except beam quality
• Since the calibration coefficient refers to a specific beam quality, the cross-
calibration should be performed at the whole range of beam qualities that are
used in the hospital
• It is important to note that other essential elements of traceability of
measurement, such as uncertainty evaluation, evidence of competence,
documentation, etc. should be taken into account and be declared for cross-
calibrations
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 61
21.7. INSTRUMENTS FOR MEASURING TUBE VOLTAGE AND TIME
Measurement of the X ray tube voltage and exposure Depending on the model, the
duration (often referred as “exposure time”, are kV-meter measures the
usually performed with non-invasive, portable,
• absolute peak voltage, (the maximum
electronic devices, often called kV-meters and timers value of the voltage during the exposure -
circled point)
• average peak voltage
• average voltage
• effective peak voltage, (the voltage that
Voltage (kV)
Time (ms)
Typical X ray tube voltage waveform from a three phase six pulse PPV has been proposed as the standard
generator operating at 80 kV tube voltage and 165 ms exposure time quantity for the X ray tube voltage
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 62
21.7. INSTRUMENTS FOR MEASURING TUBE VOLTAGE AND TIME
• The kV-meter is positioned in the primary X ray beam and measures the
X ray tube voltage with methods based on attenuation measurements
• Such instruments usually incorporate two (or more) detectors covered with filters (usually
made of copper) of different thickness
• The detectors, when exposed to radiation, produce different signals, due to the different
attenuation of the X ray beam by the filters
• The signal ratio (or any other relationship of the signals) is a function of the incident X ray
energy and consequently of the tube voltage
• During the initial calibration of the kV-meter at the factory, the signal output and/or the
reading is appropriately adjusted to the ‘correct’ tube voltage value
• Many kV-meters digitize, process and store their detector signals and can supply voltage
and/or exposure waveforms
• The kV-meter detectors’ long geometrical axis should be positioned perpendicular to the
tube anode – cathode direction to eliminate the influence of the “heel” effect on the kV
measurement
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 63
21.7. INSTRUMENTS FOR MEASURING TUBE VOLTAGE AND TIME
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 64
21.7. INSTRUMENTS FOR MEASURING TUBE VOLTAGE AND TIME
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 65
21.8. INSTRUMENTS FOR OCCUPATIONAL AND
PUBLIC EXPOSURE MEASUREMENTS
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 66
21.8. INSTRUMENTS FOR OCCUPATIONAL AND
PUBLIC EXPOSURE MEASUREMENTS
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 67
21.8. INSTRUMENTS FOR OCCUPATIONAL AND
PUBLIC EXPOSURE MEASUREMENTS
IAEA
Diagnostic Radiology Physics: a Handbook for Teachers and Students – chapter 21, 68
BIBLIOGRAPHY