IR400 Manual
IR400 Manual
MANIR400
ii
Model IR400
Table of Contents
MODEL IR400 ................................................................................................................................. I
TABLE OF FIGURES................................................................................................................ IV
iii
Model IR400
Table of Figures
Figure 1: IR400 with 31305 Junction Box .......................................................................................................... 1
Figure 2: Wiring Diagram from IR400 to Control Equipment .............................................................................. 3
Figure 3: Model IR400 ...................................................................................................................................... 5
Figure 4: Model IR400 with a 31305 Junction Box ............................................................................................ 8
Figure 5: Dimensions for IR400 mounted with S5000 Gas Monitor ...................................................................10
Figure 6: Wiring Diagram from IR400 to Control Equipment .............................................................................16
Figure 7: IR400 Zero Switch / LED During Zeroing and Calibration ..................................................................20
Figure 8: IR400 Optical Window Locations .....................................................................................................26
Figure 9: Outline and Dimensional Drawing, IR400, Inch Measurements..........................................................37
Table of Tables
Table 1: Wiring Chart ....................................................................................................................................... 2
Table 3: IR400 without HART Option ...............................................................................................................19
Table 4: IR400 with HART Option ....................................................................................................................19
Table 5: Locations ...........................................................................................................................................30
Table 6: Analog Current Output .......................................................................................................................35
Table 7: Maximum Distance between IR400 and Power Source ......................................................................35
Table 8: Maximum Distance between IR400 and Analog Input Device .............................................................35
iv
Model IR400
Quick-Start Guide
Mounting and Orientation
The Model IR400 should be mounted horizontally (Figure 1) to reduce the likelihood of dirt and
dust build-up on the windows. For optimum performance, the splashguards should be located
on the top and bottom as shown in Figure 1. Apply the supplied thread lubricant/sealant to all
conduit entries before use to prevent binding.
1
Model IR400
Wiring Connections
Power Connections
To supply power to the IR400, connect the red lead from the IR400 to the +24 VDC terminal on
the power supply. Connect the black lead from the IR400 to the power supply Common. Refer
to the manual of the power source being used, for more detailed instructions.
NOTE: If the Model IR400 is being used with a +24 VDC power supply and an industrial analog
to digital (A/D) converter, then the negative supply (Common) of all three must be
connected.
Applying Power
Before applying power to the system for the first time, all wiring connections should be checked
for correctness. Upon initial power-up or after a fault condition has been corrected, the unit will
enter a start-up mode for 2 minutes before normal operation. IR400 Ethylene start-up can take
up to 10 minutes. As a unit warms up, output may briefly be negative.
After power is applied, the IR400 should be allowed to stabilize for approximately 60 minutes
while the unit attains the proper operating temperature. After stabilization, it is recommended
that the IR400 be zeroed, per the procedure in section 4.1 (step 1 only). A gas check should
then be performed to ensure that the unit is operating properly. Use the General Monitors Gas
Check Kit (P/N 32548) to perform this check.
2
Model IR400
GREEN/YELLOW 2 FG
8 MOD 2+
9 MOD 2-
10
BLACK
RED
General Monitors Zero
BROWN Switch
3
Model IR400
1.0 Introduction
1.1 Protection for Life
General Monitors’ mission is to benefit society by providing solutions through industry-leading
safety products, services and systems that save lives and protect capital resources from the
dangers of hazardous flames, gases and vapors.
The safety products you have purchased should be handled carefully and installed, calibrated
and maintained in accordance with this instruction manual. Remember, these products are for
your safety.
4
Model IR400
5
Model IR400
6
Model IR400
3.0 Installation
Receipt of Equipment
All equipment shipped by General Monitors is packaged in shock absorbing containers, which
provide considerable protection against physical damage. The contents should be carefully
removed and checked against the packing list. If any damage has occurred or there is any
discrepancy in the order, please notify General Monitors as soon as possible. All subsequent
correspondence with General Monitors must specify the equipment part number and the serial
number.
Detector Location Considerations
There are no standard rules for detector placement, since the optimum detector location varies
with the application. The customer must evaluate conditions at the facility to make this
determination. If practical, the Model IR400 infrared point detector should be easily accessible
for occasional integrity checks. The unit should be mounted horizontally so that dirt and dust do
not build-up on the windows. Although the IR400 is EMI/RFI resistant, it should not be mounted
close to radio transmitters, high magnetic or electrical fields or similar equipment. Proper wiring
and grounding is essential for optimum performance. Snap-on ferrite filters can provide
additional resistance to electromagnetic and radio frequency interference.
NOTE: The Model IR400 cannot detect hydrogen (H2) gas.
7
Model IR400
8
Model IR400
5. Fasten the IR400 wires to a wiring connector inside the junction box, S5000, IR4000M or
IR4000S enclosure. Once the two units are attached, you may replace the cover on the
enclosure attached to the IR400 using the four captive screws, or leave it off until additional
cabling from the enclosure is completed.
9
Model IR400
10
Model IR400
11
Model IR400
Screw Earth/Stopping Plug (P/N 311001-1) into Top Entry of 31421 Housing.
Fully tighten using a suitable hand wrench.
Do not over tighten such that the O’Ring is damaged
Screw Continuity Hex Nut (P/N 928-814) on the inside of the 31421 Housing to the
Earth/Stopping Plug (P/N 311001-1).
Tighten to insure it is secure and making good contact with internal continuity plate.
If removing the Earth/Stopping Plug (P/N 311001-1), unscrew Continuity Hex Nut (P/N
928-814) on the inside of the 31421 Housing from the Earth/Stopping Plug.
Unscrew the Earth/Stopping Plug (P/N 311001-1) from the Top Entry of 31421 Housing.
12
Model IR400
The maximum distance between the IR400 and the power source is specified in Section 9.3.3.
3.3.1 Power Connections
To supply power to the Model IR400 connect the red lead from the IR400 to the +24 VDC
terminal on the power supply. Connect the black lead from the IR400 to the power supply
Common. Refer to the manual of the power source being used, for more detailed instructions.
NOTE: If the IR400 is being used with a +24 VDC power supply and an industrial analog to
digital (A/D) converter, then the negative supply (Common) of all three must be
connected.
An internal diode protects the system in the event of inadvertent supply reversal.
3.3.2 4- 20 mA Output
A 4 to 20 mA output signal is provided by the Model IR400 and can be sent to a General
Monitors’ readout/relay display module or any industrial device that can accept a 4 to 20 mA
signal for computer based multi-point monitoring. The Analog Output connection provides a
signal for use in displaying current LEL readings, special operation or fault conditions. The
maximum distance between the IR400 and the device connected to the Analog Output signal is
specified in Section 9.3.5.
To access the 4-20 mA signal, connect the white lead from the IR400 to the signal-in terminal
of the input unit. Refer to the manual of the display or other device being used for detailed
instructions.
Connect the black lead from the IR400 to the device Common. The Common connection
serves both the analog signal and the power connections.
13
Model IR400
14
Model IR400
15
Model IR400
GREEN/YELLOW 2 FG
8 MOD 2+
9 MOD 2-
10
BLACK
RED
General Monitors Zero
BROWN Switch
16
Model IR400
NOTE: Entering Gas Check, Zeroing, or Calibration mode sends a 1.5 mA output signal that
disables the S5000 Warning and Alarm relay circuits.
General Monitors gas check kit with portable purge calibrator equipment.
CAUTION – The S5000 and IR4000S turn off the Warn and Alarm relays during Gas Check,
but the current loop from the IR400 will still transmit the gas concentration level. Make
sure that any equipment monitoring the analog output is in alarm bypass mode before
performing a gas check with the S5000 or IR4000S.
The portable purge calibrator is a compact, accurate and safe system containing a non-
explosive gas concentration. The lecture bottle is filled with a standard 50% LEL mixture of
gas/air. Using a known gas/air mixture reduces the likelihood of error in field calibration.
The hose and cup adapter that are included allow for quick calibrations and gas checks.
Pre-mixed calibration gases at approximately 50% LEL are available in lecture bottles.
Butane C4H10 Ethane C2H6
Ethylene C2H4 Hexane C6H14
Methane CH4 Pentane C5H12
Propane C3H8
Spare bottles containing these gases may be ordered. Methane lecture bottles may be returned
to General Monitors for refilling with the standard 2.5% by volume methane (50% LEL per ISO
10156 and NFPA 325).
4.1.4 Calibration Guidelines
Calibrating the detector corrects any errors that may be affecting the % LEL measurement that
is taking place. You should use the General Monitors gas check kit for calibration. Section
4.1.3.1 describes the gas check kit equipment in more detail. You may need to calibrate the
IR400 detector under several circumstances.
If the gas check test indicates that the detector needs adjustment.
If you are reconfiguring the IR400 to detect a different type of gas.
If you are moving the detector to a higher altitude location (every 1000 feet difference in
altitude requires recalibration).
General Monitors configures the IR400 with three calibration input options: magnetic switch
(default), manual solenoid, and automatic remote gas calibration device (ARGC). To use the
18
Model IR400
manual solenoid or ARGC, purchase a factory configured IR400 or change the calibration input
via Modbus or HART.
4.1.5 IR400 Analog Output (AO) Interpretation
19
Model IR400
2 Seconds
Zeroing Step 1
Zero Complete
Gas Present
Fault
Re-Calibration
Complete
20
Model IR400
Review the general guidelines in Section 4.1 on Zeroing, Gas Check Tests, and Calibration. Make
sure the windows are clean and there is nothing blocking the optical beam.
4.2.2 Zero Only Calibration
The junction box used with the IR400 is fitted with a zero switch which eliminates any background
gas fluctuation.
Zero Switch
Zero the Unit. Apply the General Monitors magnet that was included with the unit to the Zero
Switch / LED for approximately three seconds. The LED in the switch will light to show proper
placement.
Remove the magnet and the LED will flash on for one second and off for one second to
indicate that the unit is attaining a zero value (Zeroing in the figure).
When the unit has finished zeroing, the LED will turn on and flash off quickly once per
second for 30 seconds (Zero Complete in Figure 7).
NOTE: If an error occurs during the zeroing/calibration sequence, the LED will flash on and off
rapidly (Fault).
NOTE: If an error occurs during the zeroing/calibration sequence, the LED will flash on and off
rapidly (Fault).
21
Model IR400
2. Return to Normal Operation or Start Calibration. If the magnet is not applied again, the
unit will return to normal operation. To continue on and calibrate, apply the magnet again
and the unit will enter the calibration mode.
The LED will flash off quickly once every half-second while the unit is waiting for gas to be
applied (Waiting for Gas).
3. Apply Gas. Apply a 50% LEL gas using a Gas Check Kit with portable purge calibrator
equipment (or 50% by volume of the gas being detected in nitrogen for units designed for
monitoring 0 to 100% by volume).
When the unit detects the gas, the LED will flash on for a half second every one and
one-half seconds (Gas Present).
If the unit does not detect the gas, the LED will flash off quickly once every half-second
while the unit is still waiting for gas to be applied (No Gas Error).
Once calibration has been completed, the LED will turn on and flash off once every
second (Calibration Complete).
4. Return to Normal Operation. Remove the gas and the unit will return to normal operation
once the gas has fallen below 5% of full scale.
22
Model IR400
3. Once the detector is placed in gas check mode and the gas is applied, monitor the % LEL
reading for the detector to see if it is functioning properly using the S5000 or IR4000S
display or Modbus/HART commands. When the reading is stabilized, it should be 50%
LEL if the gas from the check kit is applied.
4. Remove the gas.
NOTE: If the manual solenoid calibration input is configured, send the Modbus or HART
command to turn off the solenoid.
The unit will return to normal operation when the concentration drops below 5% full-scale.
Since the IR400 zeros before calibrating, you must remove the gas completely before going to
calibration mode.
The products offered by General Monitors are not for the purpose of testing response time, but
as a method to allow a user to check that the unit is responding to gas and/or that final
response is within tolerance. If it’s required to demonstrate a reading of 50% LEL at site this
can be achieved using the calibration cup, however it should be noted that you need to apply
the test gas for approximately three minutes to get a reading of 50% LEL. This time is due to
the ambient air located in the optical path of the detector having to be replaced progressively by
the test gas. This replacement of ambient air is quicker at the beginning but longer for the last
percentage because it is linked to the slow replacement of air dilution by the test gas. This test
is only to indicate a gas level of 50% LEL within the calibration cup and is not a reflection of the
response time of the detector.
24
Model IR400
5.0 Maintenance
The Model IR400 is calibrated at the factory and is fail-to-safe; once it is correctly installed and
calibrated upon start-up, it requires little maintenance other than periodic cleaning, gas checks,
zeroing and recalibration to ensure system integrity. Integrity checks can be performed using
General Monitors’ Gas Check Kit (P/N 32548).
WARNING: Disconnect or inhibit external devices such as Trip Amplifiers, PLC’s, or DCS
systems before performing any maintenance.
NOTE: If an optical fault still occurs after cleaning and re-zeroing of an IR400 detector is
complete, then you must return the unit to the factory for service. The system’s full two-
year warranty will be voided if customer personnel or third parties damage the system
during repair attempts or maintenance activities. Gassing into the screened
splashguard will not provide a stable or accurate reading.
25
Model IR400
NOTE: Do not clean the windows while an IR400 unit is zeroing or in recalibration mode. The
unit must be re-zeroed after cleaning.
5.4 Storage
The Model IR400 Gas Detector and S5000 Gas Monitor should be stored in a clean, dry area,
and within the temperature and humidity ranges noted for environmental specifications in
Section 9.3.3 Electrical Specifications for the IR400, and the separate user manual for the
S5000. Insert the red dust caps into any vacant cable entry holes while the unit is stored.
26
Model IR400
6.0 Troubleshooting
The IR400 will alert the operator that there are problems in a number of ways. The table on the
following page shows the value for the analog output, the value read through Modbus register 2
(if Modbus is being used), and the fault code, shown as F and a number, on the S5000 display
(if connected).
If the IR400 is connected to a zero switch, then the LED in the zero switch will blink at a rate of
5 flashes per second to show that the detector needs attention.
HART Troubleshooting
27
Model IR400
MODBUS FAULT
FAULT
FLAG REGISTER DESCRIPTION POSSIBLE CAUSE CORRECTIVE ACTION
CODE ANALOG OUTPUT (mA)
2
HART non-HART
1)Fouling of the detector windows or
2mA for 30s 2mA for 30s Gas Concentration is
F0 Bits 3 & 14 beam path is excessive and requires 1) Clean detector beam path and windows.
then 1.25mA then 0mA Excessively Negative
attention
Gas Concentration is 1) Detector windows or beam path are
F1 Bits 0 & 1 2 2 1) Clean detector beam path and windows.
Negative beginning to be obscurred.
1) Cal bottle empty and time-out occurred 1) Obtained filled CAL bottle and re-CAL.
4) Attempted to CAL with too much wind 4) Calibrate at a much less windy time or
resulting in unstable signal. shield the unit and CAL cup from the wind.
CAL Check Period 1) Test gas still present after gas check
F9 Bit 8 1.25 0 1) Remove the gas
Exceeded completed.
28
Model IR400
29
Model IR400
UNITED STATES
26776 Simpatica Circle Phone: +1-949-581-4464
Lake Forest, CA 92630 Email: [email protected]
IRELAND
Ballybrit Business Park
Galway Phone: +353-91-751175
Republic of Ireland Email: [email protected]
MIDDLE EAST
P.O. Box 54910
Dubai Airport Freezone Phone: +971-4-294-3640
United Arab Emirates Email: [email protected]
SINGAPORE
35 Marsiling Industrial Estate, Road 3
#04-01 Phone: +65-6748-3488
Singapore 739257 Email:
Table 5: Locations
Additional locations can be found on our web site, www.MSAsafety.com
30
Model IR400
9.0 Appendix
9.1 Warranty
General Monitors warrants the Model IR400 to be free from defects in workmanship or material
under normal use and service within two (2) years from the date of shipment.
General Monitors will repair or replace without charge any such defective equipment found to
be defective during the warranty period. Full determination of the nature of, and responsibility
for, defective equipment will be made by General Monitors’ personnel.
Defective or damaged equipment must be shipped prepaid to General Monitors’ plant or
representative from which shipment was made. In all cases this warranty is limited to the cost
of the equipment supplied by General Monitors. The customer will assume all liability for the
misuse of this equipment by its employees or other personnel.
NOTE: The Model IR400 Infrared Point Detector is easy to install; however, you should read and
understand this manual before attempting to install or operate the device. It includes
important safety information.
All warranties are contingent upon proper use in the application for which the product was
intended and do not cover products which have been modified or repaired without General
Monitors’ approval, or which have been subjected to neglect, accident, improper installation or
application, or on which the original identification marks have been removed or altered.
Except for the express warranty stated above, General Monitors disclaims all warranties with
regard to the products sold, including all implied warranties of merchantability and fitness and
the express warranty stated herein are in lieu of all obligations or liabilities on the part of
General Monitors for damages including, but not limited to, consequential damages arising out
of/or in connection with the use or performance of the product.
31
Model IR400
Control Electronics
The Model IR400 operates from an unregulated +24 VDC (nominal) input, which is fed to an
onboard power-supply that produces all of the necessary voltages within the unit. The
microprocessor constantly monitors the infrared wavelengths and performs mathematical
operations on these values in conjunction with values obtained during the factory set-up
process.
The microprocessor generates output information and feeds it to the digital analog converter to
produce a 4 to 20 milliampere (mA) signal that is proportional to the 0 to 100% LEL (or
methane % by volume) concentration of gas at the sensor. The microprocessor program also
monitors other conditions such as the supply voltage and the optical path integrity.
The Model IR400 provides a two-wire RS-485 addressable communications link conforming to
the Modbus protocol that is used to monitor the IR400’s status and settings in order to simplify
installation and maintenance.
32
Model IR400
9.3 Specifications
9.3.1 System Specifications
FM required statement: “This Approval does not include or imply Approval of apparatus to
which the subject instrumentation may be connected. In order to maintain an FM Approved
system, the apparatus to which this instrument is connected, must also be Approved by FM
Approvals.”
1
DC110 is not approved for use in ATEX installations
2
Rev E S/W and up for EU installations
33
Model IR400
Analog Signal:
Range 0 - 21.7 mA
Load (max. resistance) 600 Ω
Current Level (mA): Meaning
0 Startup mode and critical fault for non HART unit
1.25* Startup mode and critical fault for HART unit
1.5* Zero, Calibration and Gas Check Mode
2* Dirty Optics
4 – 20 0 – 100% LEL
20.1 – 21.7 Over range
* HART units can be configured to never output current less than 3.5 mA if the host equipment
is incapable of working below this level.
Area Classification:
IR400
Class I, Divisions 1 & 2, Groups B, C and D (Ta = -40°C to
+75°C), (Ta = -40°C to +60°C for ethylene only), IP66, Type
4X
Ex db, IIB+H2 T5 Gb, IP66 (Ta = -60°C to +75°C)
Ex tb IIIC T100°C Db
34
Model IR400
35
Model IR400
9.5 Communications
9.5.1 RS-485 Interface
The Model IR400 has built-in serial communications in the form of a half duplex RS-485 digital
serial interface designed to conform to EIA-485 specifications. The default settings are 9600
baud with 1 start bit, 8 data bits, 1 stop bit and no parity. The interface supports Modbus RTU
communications.
9.5.2 Modbus RTU
A programming manual is available from the MSA website that gives details on all the available
Modbus RTU commands.
9.5.3 HART
The IR400 HART Field Device Specification provides details on HART commands. The
specification is available from the MSA website.
36
Model IR400
37
Model IR400
38
Model IR400
9.7.4 Accessories
31305-1 Junction Box with magnetic switch, CSA/FM
31305-2 Junction Box without magnetic switch, CSA/FM
31305-3 Junction Box with magnetic switch, SST, CSA/FM
31305-4 Junction Box without magnetic switch, SST, CSA/FM
31305-5 Junction Box without magnetic switch, Painted, Flameproof,
ATEX/IECEx
31305-6 Junction Box without magnetic switch, Flameproof, ATEX/IECEx
31305-7 Junction Box without magnetic switch, Flameproof, SST,
ATEX/IECEx
31421-3 Junction Box with magnetic switch, ATEX/IECEx
31421-4 Junction Box without magnetic switch, ATEX
32554-1 Calibration Cup / Flow Block
32545-1 Splash Guard standard
32545-2 Splash Guard for use with remote cal
32545-3 Splash Guard with no screen for areas with occasional moisture
32545-4 Splash Guard with no screen for remote cal
31306-1 Duct Mount Junction Box
32548-Specify Gas Gas bottle with regulator and calibration cup
31420-1 Flow Block for gas sampling system
31545-1 Rain Guard Assembly
39
Model IR400
ADDENDUM
All other countries or states: please dispose of in accordance with existing federal, state and local
environmental control regulations.
40