Diamond Installation Instructions
Diamond Installation Instructions
Diamond Installation Instructions
Installation Instructions
11/2002 86501-4AB
Copyright © 2002 by Instrumentarium Imaging
Documentation, trademark and the software are copyrighted with all rights
reserved. Under the copyright laws the documentation may not be copied,
photocopied, reproduced, translated, or reduced to any electronic medium or
machine readable form in whole or part, without the prior written permission
of Instrumentarium Imaging.
This manual applies to units with software version 1.02.5 and up.
Table of Contents
1 Notes, cautions and warnings ............................................................................ 1
2 Preinstallation instructions................................................................................ 3
2.1 Unit measurements and space requirements............................................................ 3
2.2 Electrical supply requirements ................................................................................ 4
2.2.1 Environmental requirements ......................................................................................4
2.2.2 Necessary installation tools........................................................................................5
2.3 Test equipment needed for installation.................................................................... 5
2.4 Transportation of an assembled unit........................................................................ 6
2.5 External warning light connection........................................................................... 7
3 Unpacking the unit ............................................................................................. 9
3.1 Unpacking the crate ................................................................................................. 9
3.2 Unpacking the transportation wheels (optional).................................................... 11
3.3 Unpacking the counterweight tiles ........................................................................ 11
4 Installation procedures..................................................................................... 13
4.1 Lifting up the column - minimum 2 people........................................................... 13
4.2 Insertion of the counterweight tiles ....................................................................... 15
4.3 Installation of the control panel ............................................................................. 21
4.3.1 Wall installation .......................................................................................................21
4.3.2 Left-hand installation ...............................................................................................24
4.3.3 Right-hand installation.............................................................................................26
4.4 Attachment of the pedals ....................................................................................... 26
5 Testing the control functions ........................................................................... 27
6 Start-up procedures.......................................................................................... 31
6.1 Checking the cassette holders................................................................................ 31
6.2 Powering up........................................................................................................... 32
6.3 Anode run voltage check ....................................................................................... 32
6.4 kV-regulation check .............................................................................................. 33
6.5 Preheat and ma-regulation check........................................................................... 35
6.5.1 Measuring the preheat values...................................................................................35
6.5.2 Preheat charts ...........................................................................................................35
6.5.3 Adjusting the preheat and mA-values......................................................................36
6.6 Attaching the covers .............................................................................................. 37
6.6.1 Installation of the tube head cover ...........................................................................37
6.6.1.1Checking the optoswitches ................................................................................................... 37
6.6.1.2Attaching top tube head cover .............................................................................................. 37
6.6.2 Left side cover..........................................................................................................38
6.6.3 Right clam shell cover .............................................................................................39
6.6.4 Attaching the counterweight protection sheet..........................................................41
7 Checklist prior to use ....................................................................................... 43
7.1 Initial settings ........................................................................................................ 43
7.1.1 Date/Time (normal mode)........................................................................................43
7.1.2 Using service mode interface...................................................................................43
7.1.3 Physicist mode .........................................................................................................44
7.1.4 Printer/Port setup .....................................................................................................44
7.1.5 Connecting film marker (ID II) to Diamond unit ....................................................45
7.2 Check items ........................................................................................................... 45
7.2.1 Light field intensity..................................................................................................45
NOTE
Only engineers trained by Instrumentarium are allowed to install the unit or
stand by it during the installation or service work. Parts installed by non-
trained personnel carry no warranty.
CAUTION
Always check for proper seating on all the connectors or external components
located in the areas you have worked. Equipment damage can result if power
is applied or exposures made with improperly seated connectors or external
components!
CAUTION
Do not install the unit in a mobile environment. Please contact the
manufacturer for availability of van installation kit.
CAUTION
When handling circuit boards, always utilize ground wrist straps or
grounding kits to prevent damage to the “mos” type chips. Always keep
circuit boards in a static proof bag.
WARNING
Always install the proper fuses into the unit. Failure to do so jeopardizes
patient, operator and equipment safety.
WARNING
This unit has floating grounds. Always make all scope measurements
differentially and always be careful when servicing the unit.
WARNING
Always disconnect the unit from line supply by removing the plug from wall
outlet before opening the covers.
WARNING
The x-ray tube utilizes a beryllium window. Do not perform any operations
on beryllium which produce dust or fumes, such as grinding, grit blasting or
acid cleaning. Beryllium dust or fumes are highly toxic and breathing them
can result in serious personal injury or death.
WARNING
This x-ray unit may be dangerous to patient and operator unless safe exposure
factors and instructions are observed.
WARNING
Pay attention during the installation not damage optoswitches, C-Arm
mounting and other parts. Check the unit thoroughly before handing it to the
client.
2 Preinstallation instructions
NOTE
Two people are required through the mechanical installation, one person
thereafter.
NOTE
Calibration instructions are for reference only. Country specific regulations
must be observed and kept.
WARNING
Pay attention to the safety of your fingers when handling the counterweight
tile package.
The unit must be installed on a level surface in such a way that there is
enough (70cm recommended) space between the C-Arm and various objects
(walls, furniture).
NOTE
Operating Range 207 to 253 VAC. Line voltages measuring less than 208
VAC may require a line boost transformer.
NOTE
Storage and transportation in dry cool place.
CAUTION
Do not install the unit in a mobile environment. Please contact the
manufacturer for availability of van installation kit.
For transportation in the car, the unit has to be repacked. If the transportation
is short and smooth, the unit may remain in upright position in following
conditions:
1 x-ray unit ON
2 exposure ON
In order to hook up the cable, check how far it has to reach and how it affects
the positioning of the unit.
Crate:
– 2110 x 1320 x 740 mm (83” x 52” x 29 1/8”)
– wooden
– weight approximately 470 kg (1 035 lbs.)
– column with carriage and standard accessories
NOTE
With transportation wheels, the unit can be pre-installed and calibrated in
another room and then transported to the final place of use.
NOTE
The crate includes the packing list. Make sure that you have received all
items.
After removing the covers, the crate looks like this (the transportation wheels
in the image are optional):
NOTE
In case the unit has to be carried through the steps or tight spaces, remove the
base plate to make the system lighter to carry.
4 Installation procedures
WARNING
As C-Arm movement is strong, the device must be installed so that the
operator will not get caught between C-Arm and wall/ table while operating
regular or stereo C-Arm movement. After the installation, check that there are
no obstructions to free C-Arm rotation, also when in parked position. The
movement can also be canceled by pressing one of the emergency buttons or
any button on Diamond unit, except on the Control Panel.
WARNING
Check that counterweight assembly and the counterweight cables are OK so
that the system would not become unbalanced and fall on operator or patient.
3 Remove the upper column support. Hold the wooden crate base steady.
5 Remove the base of the crate and lay it somewhere where it does not
disturb the rest of installation process.
6 Open up two screws to remove the tube support.
WARNING
Get assistance and be careful when erecting the column. As counterweight
tiles are inserted later, the column is now off the balance.
CAUTION
Remember to place star washers underneath the bolt to attain proper
grounding.
The next step is to insert the counterweight tiles as described in the following
chapter.
CAUTION
Check that the assembly pins are inserted through the holes at both sides of
the column and fixed with a bolt.
1 Check that the C-Arm assembly pin is inserted. This will hold the C-
Arm steady while adding counterweight tiles.
WARNING
Do not release the C-Arm assembly pin before it is instructed to do so.
Movement of the uncovered counterweight cage poses risk to fingers.
2 Slide in counterweight tiles into the counterweight cage one by one. The
longer side of the tile is inserted to the front. Start with the smaller ones
(more space for fingers). Insert the lead weight (painted white) in the
middle of the counterweight stack.
WARNING
Be careful with the fingers.
In order to avoid the tiles from touching the inside of the column, push them
as much forward as possible so that they are tightly lined up against the back
of the counterweight cage.
3 Plug in Diamond Control Panel and ECS pedal. Refer to Chapter 4.4,
Attachment of the pedals for connection box figure.
4 When all counterweight tiles have been inserted, attach the support rod
that in order to keep the tiles in place. In order to do this, switch on the
unit and then release the C-Arm assembly pin.
NOTE
When the power is switched on, the tube head runs from an upside-down
position to an upright position.
CAUTION
At this stage, the counterweight tiles are not yet secured, but the
counterweight cage can be driven up or down.
WARNING
Do not mechanically force the C-Arm assembly past the vertical movement
sensor pieces in lower and upper parts of the column (on both sides). The
sensors have in this case a very high risk to get broken.
5 Drive the the carriage so that the opening underneath the counterweight
tiles and the screw that secures the lower part of the tile support rod are
in level. Then, loosen up the screw a bit.
CAUTION
Do not loosen up the bolt too much so that it might hinder the movement of
the carriage or drop down between column and the counterweight cage as
into the column.
6 Attach the support rod so that screws align with the slanted side.
7 Slide the support rod in (A) so that the screw does not let it any further.
Then drive the C-Arm so that the slanted side appears (B). Then tighten
up the screw (C) to secure the lower end of the rod.
8 Align the support rod vertically straight. Check that the support rod is
steadily fixed at the lower end and does not move alone, except with the
counterweight cage. Then secure the upper end of the support rod.
9 Run the C-Arm to the lowest and highest position and listen to any
abrasive or scratchy sounds. If any, then rearrange some key tiles that
seem cause the sound.
CAUTION
Check that the counterweight cable is not wrinkled or anyhow damaged and
that it can operate without abrasive contact(s) that may cause serious damage
to cable.
WARNING
The counterweight cover hatch is instructed to be inserted in Chapter 6.6.4.
Meanwhile, the area will remain open so be very careful that no one has his/
her fingers there when various movements are done to check for noises.
10 Insert the bolt to the base plate and tighten up a bit. Later on, if
necessary, the bolt can be tightened or slackened with a screwdriver
(right image).
12 Attach the rod and lift up the base plate by tightening up the nut-bolt
connection.
NOTE
Lift the base plate evenly up. Do not tighten up one nut all the way before
starting with the next one.
13 When all four nuts have been fixed, the base plate with Diamond unit is
ready for local transportation. After the final location for the Diamond
has been determined and the unit positioned there, repeat steps 1-15 in
reverse order and slackening instead of tightening.
14 If the floor surface is not level, adjust with a support piece with a
screwdriver. Then attach the plastic cover piece. Diamond base plate can
be adjusted at each corner.
3 Place a lock piece in place and secure the rest support with two screws,
but leave some slack so that the control panel can be adjusted for angle.
4 Lay the cord into notch and secure the base plate to the wall with two
screws.
5 Slide the rest piece into lower end of the opening of the control panel.
Then slide the control panel downwards until it fastens up and will not
go any further. Be careful, so that the control panel would not fall off as
it is not yet steady.
6 Adjust the position of the control panel. When the required angle has
been found, secure lock piece tightly.
7 Plug in the control panel cord to Diamond (at the back of the Diamond
column). The control panel is now ready for use (after the power
switched on after completion of all other installation work).
Note that these blocks are not of rectangular shape. For left-hand installation
the block should lean on left. Secure the blocks with screws.
3 Insert the control panel cord through the opening in the counterpiece.
Then insert the moving block so that the thicker section (between the
arrow and the line in the right image below) points to the same direction
as the support blocks.
4 Insert two screws through holes aside. Tighten up a bit so that the screws
enter the holes in the moving block and start pulling it towards the side.
This keeps the block steady until it is secured to the radiation shield. The
other end of the cable has a black control panel rest with a hole. Lay the
rest on the round end of support rod.
5 Place a lock piece in place and secure the rest support with two screws,
but leave some slack so that the control panel can be adjusted for angle.
6 Lay the cord into notch, align the control panel base plate with the
groove in the radiation shield column. Adjust for the height, then secure
the base plate with two screws that force the piece in the middle of the
base plate to the radiation shield. Check that the base plate is steadily
fixed to bear the weight of the control panel.
7 Slide the rest piece into lower end of the opening of the control panel.
Then slide the control panel downwards until it fastens up and will not
go any further. Be careful, so that the control panel would not fall off as
it is not yet steady.
8 Adjust the position of the control panel and then tighten up the screws
that secure the lock piece.
9 Plug in the control panel cord to Diamond (at the back of the Diamond
column).
10 The control panel is now ready for use (after the power switched on after
completion of all other installation work).
WARNING
Place the pedals so that the patient is not likely to press pedals while
positioned for imaging. Advice also the customer about this.
Diamond ports:
1 COM1 for software upload
2 COM2 for ID II film marking
or label printing
3 Pedal
4 Pedal
5 Diamond Control Panel port
6 Delta 32 port
7 Warning light port
WARNING
Remember to unplug the override piece when you finish with the system
checking. When the override piece is not used, it hangs on a ponduit.
WARNING
Bear in mind that the emergency stop buttons are now inoperational. In case
an emergency stop is necessary, use the user power switch or the mains
switch at the back of the unit.
Also, when turning on the system, the Diamond LCD screen displays the
current software version installed into the unit.
CAUTION
Do not press the exposure button until all necessary calibrations are
performed.
4 With the kV+ and kV- buttons, increase kV to 35, decrease to 23 kV, and
then return to the default kV setting.
5 With the density+ and density- buttons, increase density to +7, decrease
to -7 and then return the setting to 0.
6 Change the filter and notice that the filter does change from
molybdenum to rhodium. Notice that the timer display changes from
Mo to Rh. Select molybdenum (Mo) filter.
7 Push the Mode selector key (32) until the manual mode is entered, the
AutoPoint light (if active) and density display are switched off and the
mAs display will indicate 16 mAs. Also the detector lights on
compression arm are switched off.
8 Select 25 kV. With the mAs+ and mAs-buttons, increase the mAs value
to 500 and then decrease the display back down to 80.
NOTE
Press the vertical drive control down and remove the assembly pin.
9 Drive the C-arm to CC and verify that it stops at 0°. If not, then refer to
Diamond Service Manual (86502) for instructions.
10 Test that the C-Arm moves nicely down and up in CC and LAT
positions. Verify also that up and down movements of the carriage start
slowly and then speed up as the switch is held down. If not, check that
the counterweight tiles are inserted correct and causing no friction.
Check also that 4 rollers in carriage running in column groove are
correctly adjusted.
WARNING
Do not mechanically force the C-Arm assembly past the vertical movement
sensor pieces in lower and upper parts of the column (on both sides). The
sensors have in this case a very high risk to get broken.
13 Check that the breast thickness display and the compression force
changes when driving the compression down.
14 Check that the field light is switched ON when the ‘down’ pedal is
pushed and goes OFF automatically after approximately 40 seconds.
15 Drive the tube head into the front position and verify that the field light
is automatically turned on.
16 Small and large Buckies are delivered with grid block that prevents
movement of the grid during transportation from manufacturer to
customer. Remove the grid block from both Buckies before inserting
them to Diamond.
17 Insert a small bucky and check that the Bucky 18x24 appears on the
LCD display. If not, check that the sensors are sensitive enough and that
the magnet is located at the right place.
18 Repeat previous step for large Bucky, magnification tunnel (with all
magnification factors), small and large paddle.
19 Check with both Diamond buckies and magnification tunnel that they
recognize the customer’s cassettes. If not, adjust from sense
potentiometer on Casopto Board.
20 Check that the cassettes move to the front and back correct. If not, check
that there is no friction is cassette tray movement. Try this also in CC
and LAT positions.
21 Turn the C-Arm and check that the reading in the angle display on the C-
Arm changes.
22 Set C-arm rotation speed offset value at 3 in service program PrE02,
drive the C-arm between LAT-positions and check that it slows down as
it approaches CC-position. After CC-position it will ramp up speed.
After the check, restore the previous speed offset value.
23 For proper filtration, check that the filter is in place and moves when
filter selection is changed.
6 Start-up procedures
Turn the power OFF from the main power switch.
CAUTION
The purpose of this procedure is to ensure that before applying full power to
the generator, all vital signals and adjustments are correct. Use large focus.
Fig 6.1. Patient identification window in a DIN standard cassette allows the use of Alpha ID
II or other identification system to print the patient data onto the film.
6.2 Powering up
Turn the unit ON from the mains switch (16). The following screen should
appear:
Check that the POWER Board LEDs H1, H4, H6, H10, H13, H16 are lit (6
pcs), on the GENERATOR Board LEDs H1, H2 and H3 are lit.
NOTE
All testing should be performed in the service mode. Use manual exposure
values and service diaphragm.
WARNING
AMD Board has floating ground of 230 volts.
2 Make an exposure. After the exposure check that the anode motor plate
stops running (listen to the sound of the rotating anode) and that no error
messages are displayed on Control Panel LCD-screen.
Immediately after pressing the exposure button, the testpoints TP1 and TP2
located at the edge of the AMD board (X4 pin 2 and X5 pin 2) show voltage
of 230V. With a true RMS (Digital Volt Meter) the readout should be 20-
30Vac (around 30Vac with a non-true RMS meter) after 1.5 seconds (run
voltage).
If not, adjust the anode running voltage with trimmer potentiometer R100.
WARNING
X-rays!
kV is set at the factory. For calibration use calibrate meter, which measures
from x-ray. Unfors, RTI, PMX-I/M or similar can be used in calibration.
2 Make sure that the exposure is long enough for the DVM reading to
stabilize). One volt is equal to 10 kV. Testpoint TP20 should have a
reading of 2.8V ±0.15V. If not, adjust from trimmer potentiometer R68.
NOTE
Preheat setting can be set separately for every kV and both focus sizes.
Preheat values all kVs/focuses have been set at the manufacturer before
shipment.
WARNING
X-rays!
For large focus adjustment check that bucky or cassette holder is on. For
small focus adjustment, check that the magnification tunnel is attached.
CAUTION
Too slow a rise time will result in an error message.
CAUTION
If the connector is attached the wrong way, not only the buttons will not
work, but also the system is at risk of getting damaged.
2 Lower the rear end of the top cover (A). Then, lay down the front part
smoothly on the lower cover (aligns between the align piece and the
edge of the top cover (B)). Check that top cover lies smoothly (C) before
securing it.
3 Secure the top cover with a hex pin driver. If the screw holes of the top
tube head do not match with the support inside the cover around the X-
ray tube (e.g. if something has bent the support), lift up the top cover and
adjust the support so that the holes align again.
WARNING
Remember to unplug the override piece when you finish with the system
checking.
1 Connect the emergency stop cable coming from the left clam shell to the
one on the right clam shell. Make sure that the cable is well in place and
during movement such as the ECS, does not get caught in any of the
mechanical parts of the system.
2 Plug the emergency stop cable into the X15 on Power Board. Push it in
all the way until it locks up.
NOTE
Mechanical wear transforms the structure of the grease used in C-arm
rotation system to slightly more liquid and this physical phenomenon may
produce an oil drip inside, on the bottom of the clam shell. Although this drip
does not affect the operation of the unit, the right clam shell contains a patch
that captures the drip.
3 Align openings of the clam shell with the carriage and secure it.
4 Before attaching the back cover, pass the power cord through the rear
box opening.
Set the correct time and time, then Exit the screen by pressing the SET-key.
You can enter into the service mode by pressing kV- and Den-buttons down
simultaneously for 3 seconds until the LCD-display shows the word
SERVICE.
You can exit the service mode by pressing kV- and Den-buttons down
simultaneously for 3 seconds.
Physicist mode collimation settings for paddle and holder do not override the
actual setting for the holder. For example, if you have attached a small bucky
and you select large bucky and paddle from Physicist Set-Up, collimation is
what is defined for small bucky. Collimation works the same way as in
PaddleLogic so that the smaller field overrides the bigger one.
NOTE
Any changes made in the physicist mode settings are not stored and the
system default settings are restored when you exit the service mode.
Enter into the service mode by pressing kV- and Den-buttons down
simultaneously. Then check for and adjust various system settings in service
program A:System Setup/Communication setup (PrA02).
The effect of the ‘no parity bit’ and the stop bit is a total of three stop bits, so
receiving end must be configured correspondingly.
ID II film marking or label printing is connected to COM2 serial port. Use the
cable that is supplied with Diamond. For data input, refer to ID II User’s
Manual (1462 B), Chapter 2, Customizing the label.
Adjust the light intensity at the top of a cassette holder centers of the 4
quadrants to a minimum average of 160 LUX (ISLC’s). Take note of the
effect of the examination room lighting unless it can be totally switched off.
Outside USA, if a LUX meter is not available, set the collimator lamp
voltage to 11 -11.5 VDC, measured at the Collimator board TP9 to TP12 or at
conductor stands (marked by arrows in the image below).
The halogen lamp lifetime increases when voltage is decreased and vice
versa.
WARNING
Check that the mirror or lamp in the collimator has not been damaged during
transportation and can be used.
In the service mode, select the manual mode with 25 kV and 20 mAs.
Place a loaded cassette on top of the cassette holder so that the cassette
overhangs 20 mm on the chest wall edge of the holder.
20mm
ruler
cassette
Bucky
Mark the chest wall edge of the film with a ruler placed 20 mm back from the
chest wall edge of the film. (The ruler now shows the actual chest wall edge).
Make an exposure and develop the film. The exposed area between the ruler
and the chest wall edge of the film must be less than 6 mm and more than 1
mm in USA, less than 3 mm and more than 0 mm in Europe.
Check separately for each cassette holder (little Bucky, large Bucky,
Magnification, Contact etc.)
The x-ray field adjustment must be checked and aligned prior to the light
field alignment. Place a loaded cassette in the holder and position the
corresponding diaphragm.
Enter the service mode by pressing the kV- and density- buttons
simultaneously and holding them the service screen appears.
1 Place a sheet of white paper on the cassette holder so that the light field
can be seen clearly.
2 Drive the tube head forward to turn on the field light.
3 Place some coins at the corners of the illuminated area on the cassette
holder.
4 Make an exposure and develop the film.
The coins outlined on the film should not be more than 5 mm out of the
exposed field and sum of misalignments on opposite sides not more than 10
mm.
If the light field needs adjustment, refer to Diamond Service Manual (86502).
The connection between the compression paddle acrylic and the paddle
holder may be adjusted with 1mm/2mm shim plates. Regular paddle (A) and
spot magnification (B) and sizes (1mm/2mm) of the shim plates are available.
A
B
A
B
1 Turn ON the film processor and let it warm up for at least half an hour.
2 Run a clean-up film through the processor. This is just a piece of regular
film, which will pick up dirt, etc., which would appear on your film.
3 Obtain a sheet of the same film type to be used in your Diamond. Select
the color on the sensitometer that matches the color emitted by the
intensifying screen used.
4 Ensure that the emulsion side of the film is towards the light source on
the sensitometer. Expose the two opposite long edges of the film,
waiting 10 seconds between exposures, and run it through the processor.
5 Measure the film density with densitometer from the unexposed part of
film. This is the base fog and O.D. should be less than 0.20.
NOTE
Always run the film through the processor in the same orientation and
location.
6 Find the step, which has the density closest to 1.0. This step number is
the SPEED INDEX. Record the average density and step number on the
film and in tables below.
7 Proceed one step above and one step below the step of Speed Index.
Measure and record both densities.
8 Subtract the density value of the step below, from the density value of
the step above the speed index step. This result is the CONTRAST
INDEX. Record this on the film and in tables below.
DATE: TIME:
FILM TYPE: SCREEN TYPE:
WATER TEMP: DEV TEMP:
GROSS FOG: (max 02) SPEED INDEX:
CONTRAST INDEX:
Table 7.1 Film processor evaluation
Notice that the typical values depend on the film/screen combination and
consult the corresponding manufacturer’s instructions.
Run C-Arm to various angles and verify if the angles are correct. If not, enter
the service program PrE01 and calibrate the C-Arm angle:
In case the calibration was not successful, the LCD screen then indicates
‘Fail’.
NOTE
The C-arm rotation in autodrive mode is also disabled without any paddle, so
paddle (except a biopsy paddle) must be attached before running C-arm.
Remind also the customer about this feature.
Mirror angle
Check with bathroom scale that if the compression force display values are
correct. If not, enter the service program PrG04 and calibrate the compression
force display:
NOTE
The maximum manual (mechanical) compression force must not exceed
300N by IEC-standard.
NOTE
The maximum motorized (electrical) compression force is limited to 200N by
IEC-standard.
If it becomes necessary to test kV, mAs and exposure time, refer to Diamond
Service Manual, Chapter 5.1.6, kV, mAs and exposure time test.
The grid is aligned and tested at the factory and should normally not be
adjusted at the site. In case of grid lines, however, it is recommended to test
the grid movement with short exposures. If alignment is necessary, however,
follow the grid alignment instructions in Diamond Service Manual (86502).
The internal voltages can be checked according to the list in Diamond Service
Manual (86502), Chapter 4.13, Fuses and voltages.
This measures the dose based on the exposure, not on a sensitometer. Default
values are 28 kV with Molybdenum filter.
Enter into the service mode by pressing kV- and Den- buttons down
simultaneously for 3 seconds.
Testing procedure:
NOTE
Carry out this test once a week using the same procedure all the time.
The gain adjustment affects the operation of the Diamond detectors. One
detector (C1) acts as a benchmark and it is adjusted first. Then the other
detectors are adjusted to equalize the output of all detectors in relation to C1.
The graph below describes the detector numbers. It is useful to know the
detector numbers when the exposure data is examined through hyperterminal.
4=C4
5=C3
7=
L2
R
6=C2
3=
2
0=L1 1=C1 2=R1
Fig 7.2. Diamond detectors
1 Ask the customer how dark images they want. Light images have less
information and modern viewboxes can display dark images better than
before. Typical O.D. values are for example 1.4-1.6
2 Enter the service program PrD02 and check that the overall AEC gain
parameter ‘g’ is set to 1.00. (In order to start from the very beginning,
the thickness compensation and kV compensation values can be set to 0
and r1, r2 and detector-specific values to 1.)
3 Make an exposure in Auto Time mode using the clinically typical kV
(e.g. 28kV) with a Molybdenum filter and a 40 mm acrylic. Use a
standard compression paddle and a Bucky with a loaded cassette in it.
Detector C1 must be used. Also, remember to place the white plastic
cover before exposure, if that has been removed for some reason.
acrylic
cass. holder top
cassette
image receptor
base
Fig 7.3. Placement of the acrylics (and phantoms!) onto the cassette holder
4 Develop the film and mark the optical density of the film (=O.D) to the
graph next page. The higher the O.D., the darker the image and vice
versa.
5 Alter the AEC Gain parameter ‘g’ in the service program PrD02 to 1.10
(10 % increase), reload the cassette with a new film and make a new
exposure. Develop the film and mark the optical density of the film
(=O.D) to the graph below.
6 Draw a line between the two points and read the corresponding GAIN
for the wanted optical density (greater O.D. values correspond to darker
images). Notice that the object is to find the right gain, so the initial gain
value does not have to be 1.00. For example, if 1.00 falls below and 1.10
surpasses the desired O.D., then the correct gain should be some value
in-between. Draw new lines and estimate the correct gain.
7 Alter the AEC Gain parameter ‘g’ in the service program PrD02) to
obtain the desired O.D. (for example 1.6). Diminishing ‘g’ will increase
the O.D. and vice versa.
8 Make one control exposure with Auto Time-mode the clinical kV (e.g.
26kV) Molybdenum filter with 40 mm acrylic and measure the O.D. to
check the adjustment. Make note of the exposure time for the thickness
compensation in Chapter 7.3.6.
In case the values for individual detectors must be adjusted, follow the
instructions in Diamond Service Manual (86502), Chapter, 5.2.3, Automatic
AEC detector gain adjustment.
Every detector should be checked and adjusted to each film. Films can be
different.
For example the detector C4 would have the following notations (x marks the
O.D. reading):
6 If the acquired O.D. is not within the limits, adjust the values as
instructed in Diamond Service Manual (86502), Chapter 5.2.4,
Adjustment of an individual detector.
Object thickness compensation adjusts linearity that affects the curve form
1.90
1.80
1.70
1.60
1.50
1.40
1.30
1.20
1.10
mm 20 40 60 70
4 Mark the points to the O.D vs. exposure time graph. Draw a line through
the three points. They should form a horizontal line within ±0.15 O.D.
If the O.D. values for 20mm, 40mm and 60mm thicknesses fall within these
limits, no thickness compensation is required.
For example, if O.D. for 40mm thickness is 1.6, other O.D. values (for 20mm
and 60mm thicknesses) should fall between 1.45 and 1.75. If not, then
adjustment is required.
If the optical density with 20mm acrylic or 60mm acrylic are different than
the optical density with than 40 mm acrylic, then thickness compensation
parameters have to be adjusted as instructed in next chapter.
Values from the O.D. vs. exposure time graph in previous chapter
2 Look for the exposure time with which the required O.D. with 40mm
acrylic was obtained (in Chapter 7.3.3). Write this time on the line
below or to the AEC Calibration table in the Appendix section:
3 Look for the O.D. for the shorter exposure time (for example the 20 mm
Acrylic exposure time). Divide the required O.D. with this value. This
value will provide the direction of the adjustment. Test always for exact
values, because the film, processors and processing conditions vary.
4 Look for the O.D. for the longer exposure time (for example the 60 mm
Acrylic exposure time). Divide the required O.D. with this value.
CAUTION
Perform the thickness compensation with both filters (Mo/Rh) and all
cassette tunnels (Bucky/cassette holder/magnification tunnel) separately
(especially the combination that the customer starts to use). DO NOT change
the gain parameter anymore. Adjust the density correction in the service
program PrD03 to bring O.D. to same level with other filter/cassette holder
combinations.
NOTE
Fine tuning is optional and may only have minor effect to compensation. It is
not necessary to perform if the sensor adjustment for r1 and r2 is available.
1 Return to the service mode and make two exposures with Auto Time-
mode 27kV so that very long exposure time (four times longer than
exposure time for 40 mm Acr = for example 4-5 sec.) and very short
exposure time (3 of exposure time for 40 mm Acr = for example 250-
300 ms) are achieved. Write all exposure data to the AEC Calibration
table in the Appendix section.
2 Develop the films and record the optical density of the films (=O.D).
3 Select the adjustment direction of ‘S1’ (affects the compensation slope
at exposures shorter than rotation point ‘P’, when a very thin breast is
examined) and ‘S2’ (affects the compensation slope at exposures longer
than rotation point ‘P’ when a very thick breast is examined) from the
table below.
‘S1’ and ‘S2’ can have values 0, 50, 100, 150...up to 700. The maximum
compensation is 140% or 60% of O.D. in ‘P’ (rotation Point).
When small thickesses are used, the O.D. may end up too high. This may
happen if exposure time saturates (exposure time shorter than 0.05s cannot be
achieved with AEC), but tube current can be lowered down to 50% if small
thicknesses are used. Lower mA values can be used with Buckies and AEC
mode only.
If you want to use low mA mode for low thicknesses you need to adjust
preheat values for low mA (LF) as follows:
If the limit has been set, for example, at 2 cm, then lower mA values are used
for thicknesses from 0 to 2 cm. If the limit value stands at 0 cm, low mA
mode is inactive.
7.3.7 kV compensation
1 Check the linearity with a couple of control kVs (e.g. low (24kV),
medium (28kV) and high (32kV), using Molybdenum and Rhodium fil-
ter if available). Use Auto Time-mode and thick enough an acrylic to
produce a clinically typical exposure time at each kV (appr. 1 sec.).
2 Mark the received O.Ds for Bucky to the table below. x = molybdenum
filter, o = rhodium filter
4 Connect the received O.D.:s with a line. If the line shows a regular
pattern close to the required O.D. line in that chart (one of the horizontal
lines at regular intervals) then the O.D. is alright. If not, measure optical
density for each kV, using Molybdenum and Rhodium filter if available.
5 Plot the received O.D.:s on the chart and check which ones need
adjustment (graph is for reference only). If the line is not linear, check
all kVs.
If the required O.D. is for example 1.6 at 26kV and 1.3 at 23kV, increase in
service program PrD03 (23kV selected), the factor to 1.2. Exposure again and
check the O.D. Test every other kV values (...23, 25, 27, 29...) and then
estimate factors for values in between (...24, 26, 28...).
Adjust density compensations for each filter and cassette holder combination
in Auto kV mode. Set the new values in the service program PrD03. Usually
for the user the density step change is 10% or 12%, but in the service program
the step is half of this.
If initial kV is badly off, the necessary change in kV might be more than the
kV change limit in service program PrB06 allows. In this case Diamond will
use the kV that is a sum of the initial kV and the maximum allowed change
(+/-) in kV. The initial kV adjustment instructions are in Chapter 7.3.10,
Initial kV tables. The kV change limit can adjusted in service program PrB06.
The curve slope is defined with two thickness points, one at 2cm and another
one at 6cm. Target times can be set in service program PrD05. As the used
target time depends on the country, film and film developer, the parameters
have to adjusted according to local conditions and requirements. For
reference only, some users have used values from 0.4s to 0.5s for lower limit
and from 1.0s to 1.6s for upper limit. If the lower limit is less than 0.2s, the
exposure is too quick and grid lines appear in the image.
The graph below describes how changes in target time limit affect the curve
slope. If the target time limits are set at the maximum values, the result is the
steep curve A. When the lower limit is raised up the and higher limit lowered
down, the curve becomes flatter (B and C) and vice versa.
NOTE
If lowered tube current is used (refer to Chapter 7.3.6.1, Lowering mA for
small breasts), Diamond automatically compensates for it. Select target times
always with full tube current.
The initial kV is the first kV that Diamond uses for a certain thickness. Initial
kV tables can be set in service program PrD06 and the kV change limit in
PrB06. Each cassette holder can have its own Initial kV for various
thicknesses (the same for small and large Bucky). The ideal case would be
that initial kV-values would conform to target time curve values and that
there would no need for kV changes.
1 Expose in Auto kV-mode and check the used kV-values at 2cm, 4cm,
6cm and 8cm thicknesses.
2 In service program PrD06, select Bucky, Magnification or Contact.
3 Scroll through thicknesses (at 0.5cm intervals) and set initial kV for
each one of them. Set the values from step 1 and make a linear estimate
for other thicknesses.
INITIAL KV TABLE
Mo and Rh
Mo and Rh
0.5 23 25
1.0 23 25
1.5 23 25
2.0 23 25
2.5 24 25
3.0 24 25
3.5 24 25
4.0 25 26
4.5 25 26
5.0 25 26
5.5 26 27
6.0 26 27
6.5 26 28
INITIAL KV TABLE
Mo and Rh
Mo and Rh
7.0 27 28
7.5 27 28
8.0 27 29
8.5 28 29
9.0 28 30
9.5 29 30
10.0 30 30
Table 7.3 Initial kV table
4 Repeat the procedure for each cassette holder and filter combination.
5 Adjust the kV change limit in service program PrB06.
If the initial kV-values are badly off, the target times are not reached because
the required kV change exceeds the adjustment range for kV change limit.
Diamond will make the exposure but the exposure time will be different than
what target time curve suggests. High kV-values tend to reduce contrast, but
it is worthwhile to check if film manufacturer recommends higher initial kV-
values.
Diamond user might want to change of the extent of a change when a density
key of the Diamond Control Panel is pressed. This can be adjusted in service
program PrD03, Densities. The range for increase/decrease steps range from
5% to 20% (values 5 to 20 in PrD03 screen). Default setting is 10% (value
10)
Before performing any AEC adjustment work, make sure that kV values are
alright.
CAUTION
The AEC gain is adjusted for one filter and one cassette holder, normally for
the molybdenum filter and with the bucky. The right optical densities for the
rest of the cassette holders with Molybdenum filter and for all cassette
holders with rhodium filter are done as described in Chapter 7.3.
For AEC and detector-specific data, write param aecdump in the command
prompt and make an exposure.
7.3.14 AutoPoint
With software 1.02.1 and up, the AutoPoint-feature uses two detectors (if so
desired). The selection of the detector is adjusted through the Auto Point
Coefficient. Software versions 1.01.3 and older have always used one
detector for this measurement.
NOTE
If detector selection (AutoPoint coefficient < 1.0) is used to increase density
of less dense sections of the image, it will also increase the density of the
most dense areas of the breast.
Parameter settings
A AutoPoint limit value
The Auto Point limit value can be set in service program PrD01 within range
of 50–2500.
The limit adjusts the behavior of the unit on selection of the detector. High
values will make the unit less likely to select a detector under pectoral muscle
(near the chest wall edge), but may select a detector too far if the dense area is
located under C1. Low values may cause the unit to select the detector
covered by pectoral muscle. Value of 700 has proved to be a good
compromise, but this may depend on the local practices (exposure values,
positioning, population).
With software 1.02.1, the default Auto Point limit value stands at 700. In
earlier software versions the range for this value was 700-1500. For small
breasts 700 (default) is recommended
B AutoPoint Coefficient
After Diamond has selected the primary detector, it can be used alone to
determine the required exposure or together with a secondary detector. The
coefficient is a figure that defines what percentage of the measurement is
achieved with the primary detector. Value 1.0 corresponds to 100% use of the
primary detector.
Generally a value other than 1.0 (most often 0.7) is recommended only if
users would object to exposing a small dense area of the breast to target O.D.
NOTE
Diamond software does not allow the use of the detectors that are not
completely within the collimated X-ray field.
Diamond detects the paddle and the cassette holder in use. The field size with
different paddle/cassette holder combinations can be checked with the field
light.
Paddle related collimation can be adjusted in service programs PrF01 for rear
plates and in service program PrF02 for side plates. Adjust step values
according to paddle size.
NOTE
If the paddle and the cassette holder suggest different X-ray field sizes,
Diamond will always select the smaller field size.
Diamond can select most usable detectors according to the paddle type.
Detectors used with the paddle can be adjusted in service program PrD09,
Paddle Logic. For example certain detectors that are not completely under a
small size paddle can be eliminated from use.
When Diamond is switched on, name of one of the four modes is displayed in
LCD screen of Control Panel. Set the initial mode in service program PrB01.
Note that the modes are numerically listed, so check for the corresponding
PreProg number in service programs PrB06-09.
NOTE
The following three items disconnects all power from the generator, but the
system can be still operated. This can be done for example when installing for
an exhibition, where X-rays are not allowed.
8 Appendices
The Diamond software block diagrams, factory default settings list, a
mechanical checklist, AEC detector location figure and an AEC calibration
list are attached as an appendix at the end of this document.
Set
QA - MENU - 01
Modes Setup
(Quality Assurance)
QA MODES SETUP
SET
Discard changes
CANCEL
Accept
- DISPLAY - 33
Brightness: 5 Save changes
Contrast: 5
SET
Down Next Up
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
Set
Scroll
Scrolland
andselect
selectone
oneofof Scroll and select one of Scroll and select one of Scroll and select one of Scroll and select one of
the
thefollowing
followingprocedures
procedures the following procedures the following procedures the following procedures the following procedures
- -COLLIMATOR
COMPRESSION - - ?00
G00 - FILTER - H00 - GENERATOR - I00 - TUBE MOVEMENT- J00 - STEREO MOVEMENT- K00
5:Force
Angle Calibration
Limits 1:AutoFilt set 1:Gener Set 2: Autodrive 1:Speeds
> >Rotation
1:Compress
Speed
Set< <Pr## > 2:Filt Calibr < > 2:Preh Cal LF < > 1:Speeds < > 2:Auto drive <
upDown Accept
Accept
down Up Down Accept Up Down Accept Up Down Accept Up Down Accept Up
Pr301
G01: Rear
Compression
Plate Settings H01: AutoFilter setting I01: Generator Setup J01: Speeds K01: Speeds
Pr302
G02: Thickness
Side PlatesPaddles H02: Filter Calibration I02: Preheat Calibration LF J02: Autodrive K02: Auto drive
Pr303
G03: Thickness
Collimator Cassette
Light hold I03: Preheat Calibration SF
G04: Force Calibration I04: Preheat Compensation LF
G05: Force Limits I05: Preheat Compensation SF
G06: Up Force Limit
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
Button1 Button3
Button2
Set
- SYSTEM - A00
2:Comm setup
> 1:Sys params <
Down Accept Up
System Parameters A01 Hardware Configuration A02
CANCEL/SET
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
Button1 Button3
Button2
PROGR 1 Bucky 18x24 PROGR 1 Bucky 18x24 PROGR 1 Bucky 18x24
12:00 SERVICE 12:00 SERVICE 12:00 SERVICE
Auto Auto Auto
Rel Rel Rel
Set
Accept Set/Cancel
Bucky Settings
Mode: AutokV/AutoTime/Manual
Default Filter: Mo/Rh/Alu PREPROG4 MODE SETTINGS:
Auto Dose: On/Off
Contact Settings PREPROG3 MODE SETTINGS:
Autopoint: On/off
Default detector: 0-8 Mode: AutokV/AutoTime/Manual PREPROG2 MODE SETTINGS:
Def kV Auto Time Mo:23-35 Default Filter: Mo/Rh/Alu
Def kV Auto Time Rh:23-35 Auto Dose: On/Off QA MODE SETTINGS:
Def kV Manual Mo :23-35 Autopoint: On/off PREPROG1 MODE SETTINGS: QA- mode: None/Exp
Def kV Manual Rh :23-35 Default detector: 0-8 Density: -7 -+7 Default kV: 23-35
Default mAs:0-33 Def kV Auto Time Mo:23-35 Target Time: 1/2 Default Filter: Mo/Rh
Def kV Auto Time Rh:23-35 Auto kV Initial kV: -3- +3
Def kV Manual Mo :23-35 Menu Text: PROGR 1
Def kV Manual Rh :23-35 kV Change Limit:0-5
Default mAs:0-33 AutoPnt Coefficient: 0.50-1.00
Mag Settings
Mode: AutokV/AutoTime/Manual
Default Filter: Mo/Rh/Alu
Auto Dose: On/Off
Autopoint: On/off
Default detector: 0-8
Def kV Auto Time Mo:23-35
Def kV Auto Time Rh:23-35
Def kV Manual Mo :23-35
Def kV Manual Rh :23-35
Default mAs:0-33
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
PROGR 1 Bucky 18x24 PROGR 1 Bucky 18x24 PROGR 1 Bucky 18x24 AEC gains
12:00 SERVICE 12:00 SERVICE Scroll and select one of AEC Setup
12:00 SERVICE
Auto Auto Auto the following procedures
Detector adjustment
Rel Rel Rel Accept - AEC gains- D02
- AEC Setup - D01
- AEC - D00 AEC gain 1.15
LowmA lim cm 0
2:AEC gain >DetectorL1 1.02 <
> Target O.D 1.50 < -DETECT ADJUST-D08
> 3:Densities < Down Next Up
Down Next Up Calc new gains based
Down Accept Up
on recent exposure?
Cancel Pr301 Check Inputs NO YES
Set Pr301 Check Inputs Parameters: Parameters:
Pr302 Check Bucky
D01Check
Pr302 AEC setup
Bucky Target O.D: 1.00 - 2.00 AEC gain': 0.5 - 2.0
Pr303
D02 AEC gains Target lim l: 0.20-1.00 Pr301 Check Inputs
Pr303 Detector L1:1.00
D03 Densities Target lim h: 1.00-3.00 Pr302 Check Bucky
Detector C1:1.00 Calc new gains based
Pr303
D04 Detector compensation Autopoint lim: 50-2500 Detector R1:1.00 on recent exposure?
D05 Target Times Low mA limit cm: 0-4 Detector L2:1.00
Scroll and select one of D06 initial kV-tables
NO YES
A: System setup Accept Detector C4:1.00
the following groups D07 QA- coefficient history
B: Defaults Detector C3:1.00
D08 Detector adjustmen Accept Detector C2:1.00
C: Diagnostics
Cancel D09 PaddleLogic YES
- GROUPS - 000 D: AEC Detector R2:1.00
Cancel Cancel
C:Diagnostics E: C-arm Rotation
> D:AEC < F: Collimator
Down Accept Up G: Compression NO
H: Filter New adjusted detector
I: Generator gains
J: Park Back
K: Stereo movement Accept/Cancel NEW DTR GAINS-
Accept DetectorL1 0.98
L: Physicist
DetectorC1 1.00 <
Down Accept Up
-7.0 Bucky
Rh Button1
.. Button1
+7.0
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
Button1 Button3
Button2
Set
Pr301 Check Inputs
Parameters:
Pr302 Check Bucky
Autodrive
Pr303 Off/On/Pdl_up
Fail
- ANGLE CALIBR- E01
Pass
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
Button1 Button3
Button2
Set
Button2
Rear Plate F01 Rear Plate F01 Side Plates F02 Side Plates F02
F01: Rear Plate
F02: Side Plates
F03: Collimator Light - SIDE PLATES - F02 - SIDE PLATES - F02
-REAR PLATE- F01 -REAR PLATE- F01 Stand small 12345
Bucky 18x24: 12345 Stand small 12345 Bucky 18x24: 12345
Bucky 24x30: 12345 >Stand large 12345 <
>Bucky 24x30: 12345< >Stand large 12345< Down Next Up
Down Next Up Down Next Up Down Next Up
Pr301 Check Inputs Pr301 Check Inputs Pr301 Check Inputs Pr301 Check Inputs
Parameters:
Pr302 Check Bucky Parameters:
Pr302 Check Bucky Parameters:
Pr302 Check Bucky Parameters:
Pr302 Check Bucky
Front Plate : 39002
Pr303 No Paddle : 500
Pr303 Front Plate : 0
Pr303 No Paddle : 100
Pr303
Bucky 18x24: 9500 Stand small : 1000 Bucky 18x24: 4800 Stand small : 300
Colimator Bucky 24x30: 1000 Stand large : 1000 Bucky 24x30: 300 Stand large : 300
light Mag 1.6 : 9500 Mag spot : 10550 Mag 1.6 : 4800 Mag spot : 5950
Unknown : 10000 Stand Mag : 9500 Unknown : 4100 Stand Mag : 4800
Cont 18x24 : 9500 Implant : 19600 Cont 18x24 : 4800 Implant : 3450
Cont 24x30 : 1000 Mag implant : 8700 Cont 24x30 : 300 Mag implant : 1500
Mag 1.8 : 9500 Delta32 : 25000 Mag 1.8 : 4800 Delta32 : 15000
SpotMag 1.6: 9500 Spot : 23750 SpotMag 1.6: 4800 Spot : 13450
Stereo dig : 27000 Mag spt enh : 7450 Stereo dig : 17500 Mag spt enh : 5300
Stereo film : 10000 Biopsy D : 9500 Stereo film : 4100 Biopsy D : 4800
Collimator light F03 Mag2.0 : 9500 Biopsy SQ : 5750 Mag2.0 : 4800 Biopsy SQ : 10600
reserved : 10000 Spot enhanc : 16850 reserved : 4100 Spot enhanc : 12000
reserved : 10000 Biopsy spot : 10550 reserved : 4100 Biopsy spot : 5950
reserved : 10000 Reserved14 : 500 reserved : 4100 Reserved14 : 500
-COLLIM LIGHT - F03 reserved : 10000 Reserved15 : 500 reserved : 4100 Reserved15 : 500
Time St/Bi 80sec
>Intensity 3 <
Down Next Up
Cancel/Set
Parameters:
Intensity : 0-10(7)
Time normal: 20 - 60 secs(40)
Time Stereo/Biopsy: 20-90 sec(80) Cancel/Set
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
Button1 Button3
Button2
Auto kV mode Auto Time mode Manual mode
Set
Scroll
Scrolland
andselect
selectone
oneofof
Compression Setup G01 Paddle Calibration G02 Cass Holder Calibration G03 Compression Force Limits G05
COMPRESSION FORCE the
thefollowing
followingprocedures
procedures
CALIBRATION PROCEDURE BUTTON2
BUTTON2
BUTTON2
Pass
- UP FORCE LIM - G06
Adjust force 4
Down Accept Up
BUTTON2
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
Button1 Button3
Button2
Set
Parameters: Parameters:
Change Thickness: 1- 15 cm(5) Molybden : 1950
Autofilter On/Off (Off) Rhodium : 150
Aluminum : 3150
CANCEL/SET
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
Button1 Button3
Button2
Set
PREHEAT ADJUSTMENT
PROCEDURE
CANCEL
- PREH. ADJUST. - I02
Low mA
Press EXP button to Standard mA
start automatic Generator setup
calibration
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
Button1 Button3
Button2
Set
Cancel/Set
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
Button1 Button3
Auto kV mode Auto Time mode Manual mode Button2
Set
Ramp in 10ms
CANCEL
Speed
Ramp up Time
Ramp
down
Speed
Speed offset = +2
Time
Legend
not used in program
Diamond Control Panel User Interface Specifications (SW) v1.13 18.6.02 / KPa
Button1 Button3
Button2
Set
-PHYSICIST SETT-L00
Grid Off
>Focus Small<
Down Next Up
Parameters:
Grid: Off/System
Focus: Large/Small/System
Paddle: Any paddle/System
Holder: Any holder/System
kV limit: Off/System
System is default for all.
CANCEL
Legend
not used in program
A sample of Battery RAM contents, valid from software version 1.02.5 and up (for details, refer to
Diamond Service Manual (86502), Chapter 3.2, Service mode overview):
Serial#=20150
Country=1
Efficiency Mo=160
Efficiency Rh=120
Efficiency Al=350
HVL Mo=33
HVL Rh=38
HVL Al=47
Label out=0
Label type=1
Data out=0
COM1 Baud=1
COM2 Baud=1
Startup:Preprog=1
Autorel=1
Autoparkback=0
Bucky:mode=0
B filter=0
B autodose=0
B autopoint=1
B detector=1
B autotime kV Mo=28
B autotime kV Rh=28
B manual kV Mo=28
B manual kV Rh=28
B def MAS=5
Mag:mode=0
M filter=0
M autodose=0
M autopoint=1
M detector=1
M autotime kV Mo=28
M autotime kV Rh=28
M manual kV Mo=28
M manual kV Rh=28
M def MAS=5
Contact:mode=0
C filter=0
C autodose=0
C autopoint=1
C detector=1
C autotime kV Mo=28
C autotime kV Rh=28
C manual kV Mo=28
C manual kV Rh=28
C def MAS=5
QA-mode=1
QA def kV=28
QA filter=0
Preprog1:density=7
PR1 targettime=1
PR1 auto kV initial=10
PR1 menutext=80 82 79 71 82 32 49 32
PR1 AkV change limit=3
PR1 AutoPoint coef=100
Preprog2:density=7
PR2 targettime=1
PR2 auto kV initial=10
PR2 menutext=80 82 79 71 82 32 50 32
PR2 AkV change limit=3
PR2 AutoPoint coef=100
Preprog3:density=7
PR3 targettime=1
PR3 auto kV initial=10
PR3 menutext=80 82 79 71 82 32 51 32
PR3 AkV change limit=3
PR3 AutoPoint coef=100
Preprog4:density=7
PR4 targettime=1
PR4 auto kV initial=10
PR4 menutext=80 82 79 71 82 32 52 32
PR4 AkV change limit=3
PR4 AutoPoint coef=100
Image counter=0 0
Target O.D.=160
Target lim l=20
Target lim h=300
AutoPnt lim=700
Low mA limit (cm)=0
AEC gain=115
DetectorL1=100
DetectorC1=100
DetectorR1=100
DetectorL2=100
DetectorC4=100
DetectorC3=100
DetectorC2=100
DetectorR2=100
DEN_AUTOKV_BUCKY_MO=14
DEN_AUTOKV_BUCKY_RH=14
DEN_AUTOKV_BUCKY_AL=14
DEN_AUTOKV_MAG_MO=14
DEN_AUTOKV_MAG_RH=14
DEN_AUTOKV_MAG_AL=14
DEN_AUTOKV_CONTACT_MO=14
DEN_AUTOKV_CONTACT_RH=14
DEN_AUTOKV_CONTACT_AL=14
DEN_AUTOTIME_BUCKY_MO=14
DEN_AUTOTIME_BUCKY_RH=14
DEN_AUTOTIME_BUCKY_AL=14
DEN_AUTOTIME_MAG_MO=14
DEN_AUTOTIME_MAG_RH=14
DEN_AUTOTIME_MAG_AL=14
DEN_AUTOTIME_CONTACT_MO=14
DEN_AUTOTIME_CONTACT_RH=14
DEN_AUTOTIME_CONTACT_AL=14
THC_COMP_P_BUCKY_MO=30
THC_COMP_R1_BUCKY_MO=100
THC_COMP_R2_BUCKY_MO=100
THC_COMP_S1_BUCKY_MO=2500
THC_COMP_S2_BUCKY_MO=2500
THC_COMP_P_BUCKY_RH=30
THC_COMP_R1_BUCKY_RH=100
THC_COMP_R2_BUCKY_RH=100
THC_COMP_S1_BUCKY_RH=2500
THC_COMP_S2_BUCKY_RH=2500
THC_COMP_P_BUCKY_AL=100
THC_COMP_R1_BUCKY_AL=100
THC_COMP_R2_BUCKY_AL=100
THC_COMP_S1_BUCKY_AL=2500
THC_COMP_S2_BUCKY_AL=2500
THC_COMP_P_MAG_MO=100
THC_COMP_R1_MAG_MO=100
THC_COMP_R2_MAG_MO=100
THC_COMP_S1_MAG_MO=2500
THC_COMP_S2_MAG_MO=2500
THC_COMP_P_MAG_RH=100
THC_COMP_R1_MAG_RH=100
THC_COMP_R2_MAG_RH=100
THC_COMP_S1_MAG_RH=2500
THC_COMP_S2_MAG_RH=2500
THC_COMP_P_MAG_AL=100
THC_COMP_R1_MAG_AL=100
THC_COMP_R2_MAG_AL=100
THC_COMP_S1_MAG_AL=2500
THC_COMP_S2_MAG_AL=2500
THC_COMP_P_CONTACT_MO=100
THC_COMP_R1_CONTACT_MO=100
THC_COMP_R2_CONTACT_MO=100
THC_COMP_S1_CONTACT_MO=2500
THC_COMP_S2_CONTACT_MO=2500
THC_COMP_P_CONTACT_RH=100
THC_COMP_R1_CONTACT_RH=100
THC_COMP_R2_CONTACT_RH=100
THC_COMP_S1_CONTACT_RH=2500
THC_COMP_S2_CONTACT_RH=2500
THC_COMP_P_CONTACT_AL=100
THC_COMP_R1_CONTACT_AL=100
THC_COMP_R2_CONTACT_AL=100
THC_COMP_S1_CONTACT_AL=2500
THC_COMP_S2_CONTACT_AL=2500
KV_COMP_BUCKY_MO(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 10
0 100 100
KV_COMP_BUCKY_RH(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 10
0 100 100
KV_COMP_BUCKY_AL(36-39kV)=100 100 100 100
KV_COMP_MAG_MO(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 100
100 100
KV_COMP_MAG_RH(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 100
100 100
KV_COMP_MAG_AL(36-39kV)=100 100 100 100
KV_COMP_CONT_MO(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 100
100 100
KV_COMP_CONT_RH(20-35kV)=100 100 100 100 100 100 100 100 100 100 100 100 100 100
100 100
KV_COMP_CONT_AL(36-39kV)=100 100 100 100
Density step % =10
DETECTOR_COMPENSATION=100 100 100 100 100 100 100 100
No Paddle =255
Stand small =255
Stand large =255
Mag spot =66
Stand Mag =255
Implant =239
Mag implant =255
Delta32 =66
Spot =66
Mag spt enh =66
Biopsy D =255
Biopsy SQ =250
Spot enhanc =66
Biopsy spot =66
Reserved14 =255
Reserved15 =255
Reserved16 =255
Reserved17 =255
Reserved18 =255
Reserved19 =255
Reserved20 =255
Reserved21 =255
Reserved22 =255
Reserved23 =255
Reserved24 =255
Reserved25 =255
Reserved26 =255
Reserved27 =255
Reserved28 =255
Reserved29 =255
Reserved30 =255
Reserved31 =255
Speed offset=0
Lateral angle=270
Oblique angle=225
Mirror angle=1
Auto drive=1
no holder =39002
Bucky 18x24 =9500
Bucky 24x30 =1000
Mag 1.6 =9500
Unknown =10000
Cont 18x24 =9500
Cont 24x30 =1000
Mag 1.8 =9500
SpotMag 1.6 =9500
Stereo dig =27000
Stereo film =10000
Mag 2.0 =9500
Reserved =10000
Reserved =10000
Reserved =10000
Reserved =10000
No Paddle =500
Stand small =1000
Stand large =1000
Mag spot =10550
Stand Mag =9500
Implant =19600
Mag implant =8700
Delta32 =25000
Spot =23750
Mag spt enh =7450
Biopsy D =9500
Biopsy SQ =5750
Spot enhanc =16850
Biopsy spot =10550
Reserved14 =500
Reserved15 =500
Reserved16 =500
Reserved17 =500
Reserved18 =500
Reserved19 =500
Reserved20 =500
Reserved21 =500
Reserved22 =500
Reserved23 =500
Reserved24 =500
Reserved25 =500
Reserved26 =500
Reserved27 =500
Reserved28 =500
Reserved29 =500
Reserved30 =500
Reserved31 =500
No Paddle =100
Stand small =300
Stand large =300
Mag spot =5950
Stand Mag =4800
Implant =3450
Mag implant =1500
Delta32 =15000
Spot =13450
Mag spt enh =5300
Biopsy D =4800
Biopsy SQ =10600
Spot enhanc =12000
Biopsy spot =5950
Reserved14 =500
Reserved15 =500
Reserved16 =500
Reserved17 =500
Reserved18 =500
Reserved19 =500
Reserved20 =500
Reserved21 =500
Reserved22 =500
Reserved23 =500
Reserved24 =500
Reserved25 =500
Reserved26 =500
Reserved27 =500
Reserved28 =500
Reserved29 =500
Reserved30 =500
Reserved31 =500
Intensity=7
Time normal=40
Time St/Bi=80
Slow speed=4
High speed=6
Release speed=5
AR travel=15
Sp threshold=50
no holder =19
Bucky 18x24 =19
Bucky 24x30 =19
Mag 1.6 =19
Unknown =19
Cont 18x24 =19
Cont 24x30 =19
Mag 1.8 =10
SpotMag 1.6 =19
Stereo dig =7
Stereo film =19
Mag 2.0 =3
Reserved =19
Reserved =19
Reserved =19
Reserved =19
THICKNESS_BASE_POINT_BR=10
Force limit:mode=0
Force limits=70 90 110 130 150 170 200
Negat. Force limit=4
FILTER_MO_CALIBRATION=1950
FILTER_RH_CALIBRATION=150
FILTER_AL_CALIBRATION=3150
Autofilter=0
Thickness=50
Anode Brake=0
Preheat=102
LF kV(range:15-39kV)=54 55 55 55 56 56 56 56 57 57 57 56 55 54 54 53 53 53 52 52
51 48 48 48 48
Preheat(low mA)=100
LF kV(range:15-39kV)(low mA)=50 51 51 51 52 52 52 52 52 52 52 52 51 51 51 50 50
50 49 49 48 47 47 47 47
Preheat=101
SF kV(range:15-39kV)=48 48 49 49 49 49 49 49 50 49 49 48 48 47 47 47 46 46 46 45
45 45 45 45 45
Speeds=2 5
Exp drive=1
AutoPark=0
CHECKSUM_BR=12473
ANGLE_CALIBRATION_BR=81
FORCE_CALIBRATION_BR=100 40
CPULSE_BR=100 1005 0
STEREOARMPULSES_BR=0
STEREOARMPULSEPERANGLE=27
STEREOARMZEROANGLE=1800
STEREOARMLEFTLIMANGLE=1650
STEREOARMRIGHTLIMANGLE=1950
SW_VERSION_BR=17201 2 5 21553 0 0
Mirror angle timeout (sec)=60
MGX / Diamond 18.6.2002
AUTO_KV_BUCKY_BR
DEN_AKV_BUCKY_MO_BR Density- Mo (-7.0-+7.0) 0,0 0 28
DEN_AKV_BUCKY_RH_BR Density - RH (-7.0-+7.0) 0,0 0 28
DEN_AKV_BUCKY_AL_BR Density- Al (-7.0-+7.0) 0,0 0 28
AUTO_KV_MAG_BR
DEN_AKV_MAG_MO_BR Density- Mo (-7.0-+7.0) 0,0 0 28
DEN_AKV_MAG_RH_BR Density - RH (-7.0-+7.0) 0,0 0 28
DEN_AKV_MAG_AL_BR Density- Al (-7.0-+7.0) 0,0 0 28
AUTO_KV_CONTACT_BR
DEN_AKV_CONTACT_MO_BR Density- Mo (-7.0-+7.0) 0,0 0 28
DEN_AKV_CONTACT_RH_BR Density - RH (-7.0-+7.0) 0,0 0 28
DEN_AKV_CONTACT_AL_BR Density- Al (-7.0-+7.0) 0,0 0 28
AUTO_TIME_BUCKY_BR
DEN_AT_BUCKY_MO_BR Density- Mo (-7.0-+7.0) 0,0 0 28
DEN_AT_BUCKY_RH_BR Density - RH (-7.0-+7.0) 0,0 0 28
DEN_AT_BUCKY_AL_BR Density- Al (-7.0-+7.0) 0,0 0 28
AUTO_TIME_MAG_BR
DEN_AT_MAG_MO_BR Density- Mo (-7.0-+7.0) 0,0 0 28
DEN_AT_MAG_RH_BR Density - RH (-7.0-+7.0) 0,0 0 28
DEN_AT_MAG_AL_BR Density- Al (-7.0-+7.0) 0,0 0 28
AUTO_TIME_CONTACT_BR
DEN_AT_CONTACT_MO_BR Density- Mo (-7.0-+7.0) 0,0 0 28
DEN_AT_CONTACT_RH_BR Density - RH (-7.0-+7.0) 0,0 0 28
DEN_AT_CONTACT_AL_BR Density- Al (-7.0-+7.0) 0,0 0 28
THICK_COMP_BUCKY_BR
Thick_comp for Mo
THICK_COMP_P_BUCKY_MO_BR Mo-P 0,30 30 200
THICK_COMP_R1_BUCKY_MO_BR Mo-r1 1,00 50 200
THICK_COMP_R2_BUCKY_MO_BR Mo-r2 1,00 50 200
THICK_COMP_S1_BUCKY_MO_BR Mo-S1 2500 0 7000
THICK_COMP_S2_BUCKY_MO_BR Mo-S2 2500 0 7000
Thick_comp for Rh
THICK_COMP_P_BUCKY_RH_BR Rh-P 0,30 30 200
THICK_COMP_R1_BUCKY_RH_BR Rh-r1 1,00 50 200
THICK_COMP_R2_BUCKY_RH_BR Rh-r2 1,00 50 200
THICK_COMP_S1_BUCKY_RH_BR Rh-S1 2500 0 7000
THICK_COMP_S2_BUCKY_RH_BR Rh-S2 2500 0 7000
Thick_comp for Al
THICK_COMP_P_BUCKY_AL_BR Al-P 1,00 30 200
THICK_COMP_R1_BUCKY_AL_BR Al-r1 1,00 50 200
THICK_COMP_R2_BUCKY_AL_BR Al-r2 1,00 50 200
THICK_COMP_S1_BUCKY_AL_BR Al-S1 2500 0 7000
THICK_COMP_S2_BUCKY_AL_BR Al-S2 2500 0 7000
THICK_COMP_MAG_BR
D - AEC Thick_comp for mo
THICK_COMP_P_MAG_MO_BR Mo-P 1,00 30 200
THICK_COMP_R1_MAG_MO_BR Mo-r1 1,00 50 200
THICK_COMP_R2_MAG_MO_BR Mo-r2 1,00 50 200
THICK_COMP_S1_MAG_MO_BR Mo-S1 2500 0 7000
THICK_COMP_S2_MAG_MO_BR Mo-S2 2500 0 7000
Thick_comp for Rh
THICK_COMP_P_MAG_RH_BR Rh-P 1,00 30 200
THICK_COMP_R1_MAG_RH_BR Rh-r1 1,00 50 200
THICK_COMP_R2_MAG_RH_BR Rh-r2 1,00 50 200
THICK_COMP_S1_MAG_RH_BR Rh-S1 2500 0 7000
THICK_COMP_S2_MAG_RH_BR Rh-S2 2500 0 7000
Thick_comp for Al
THICK_COMP_P_MAG_AL_BR Al-P 1,00 30 200
THICK_COMP_R1_MAG_AL_BR Al-r1 1,00 50 200
THICK_COMP_R2_MAG_AL_BR Al-r2 1,00 50 200
THICK_COMP_S1_MAG_AL_BR Al-S1 2500 0 7000
THICK_COMP_S2_MAG_AL_BR Al-S2 2500 0 7000
THICK_COMP_CONTACT_BR
D - AEC Thick_comp for mo
THICK_COMP_P_CONTACT_MO_BR Mo-P 1,00 30 200
THICK_COMP_R1_CONTACT_MO_BR Mo-r1 1,00 50 200
THICK_COMP_R2_CONTACT_MO_BR Mo-r2 1,00 50 200
THICK_COMP_S1_CONTACT_MO_BR Mo-S1 2500 0 7000
THICK_COMP_S2_CONTACT_MO_BR Mo-S2 2500 0 7000
Thick_comp for Rh
THICK_COMP_P_CONTACT_RH_BR Rh-P 1,00 30 200
THICK_COMP_R1_CONTACT_RH_BR Rh-r1 1,00 50 200
THICK_COMP_R2_CONTACT_RH_BR Rh-r2 1,00 50 200
THICK_COMP_S1_CONTACT_RH_BR Rh-S1 2500 0 7000
THICK_COMP_S2_CONTACT_RH_BR Rh-S2 2500 0 7000
Thick_comp for Al
THICK_COMP_P_CONTACT_AL_BR Al-P 1,00 30 200
THICK_COMP_R1_CONTACT_AL_BR Al-r1 1,00 50 200
THICK_COMP_R2_CONTACT_AL_BR Al-r2 1,00 50 200
THICK_COMP_S1_CONTACT_AL_BR Al-S1 2500 0 7000
THICK_COMP_S2_CONTACT_AL_BR Al-S2 2500 0 7000
KV_COMP_BUCKY_BR
KV_COMP_BUCKY_MO_BR kV-comp Bu for Mo
20 1.00 70 130
21 1.00 70 130
22 1.00 70 130
23 1.00 70 130
24 1.00 70 130
25 1.00 70 130
26 1.00 70 130
27 1.00 70 130
28 1.00 70 130
29 1.00 70 130
30 1.00 70 130
31 1.00 70 130
32 1.00 70 130
33 1.00 70 130
34 1.00 70 130
35 1.00 70 130
KV_COMP_BUCKY_RH_BR kV-comp Bu for Rh
20 1.00 70 130
21 1.00 70 130
22 1.00 70 130
23 1.00 70 130
24 1.00 70 130
25 1.00 70 130
26 1.00 70 130
27 1.00 70 130
28 1.00 70 130
29 1.00 70 130
30 1.00 70 130
31 1.00 70 130
32 1.00 70 130
33 1.00 70 130
34 1.00 70 130
35 1.00 70 130
kV-comp Bu for Al
36 1.00 70 130
37 1.00 70 130
38 1.00 70 130
39 1.00 70 130
KV_COMP_MAG_BR
kV-comp Mag for Mo
20 1.00 70 130
21 1.00 70 130
22 1.00 70 130
23 1.00 70 130
24 1.00 70 130
25 1.00 70 130
26 1.00 70 130
27 1.00 70 130
28 1.00 70 130
29 1.00 70 130
30 1.00 70 130
31 1.00 70 130
32 1.00 70 130
33 1.00 70 130
34 1.00 70 130
35 1.00 70 130
kV-comp Mag for Rh
20 1.00 70 130
21 1.00 70 130
22 1.00 70 130
23 1.00 70 130
24 1.00 70 130
25 1.00 70 130
26 1.00 70 130
27 1.00 70 130
28 1.00 70 130
29 1.00 70 130
30 1.00 70 130
31 1.00 70 130
32 1.00 70 130
33 1.00 70 130
34 1.00 70 130
35 1.00 70 130
kV-comp Mag for Al
36 1.00 70 130
37 1.00 70 130
38 1.00 70 130
39 1.00 70 130
KV_COMP_CONTACT_BR
kV-comp contact for Mo
20 1.00 70 130
21 1.00 70 130
22 1.00 70 130
23 1.00 70 130
24 1.00 70 130
25 1.00 70 130
26 1.00 70 130
27 1.00 70 130
28 1.00 70 130
29 1.00 70 130
30 1.00 70 130
31 1.00 70 130
32 1.00 70 130
33 1.00 70 130
34 1.00 70 130
35 1.00 70 130
kV-comp contact for Rh
20 1.00 70 130
21 1.00 70 130
22 1.00 70 130
23 1.00 70 130
24 1.00 70 130
25 1.00 70 130
26 1.00 70 130
27 1.00 70 130
28 1.00 70 130
29 1.00 70 130
30 1.00 70 130
31 1.00 70 130
32 1.00 70 130
33 1.00 70 130
34 1.00 70 130
35 1.00 70 130
kV-comp Contact for Al
36 1.00 70 130
37 1.00 70 130
38 1.00 70 130
39 1.00 70 130
DEN_STEP_BR Density step % 10 5 20
DETECTOR_COMPENSATION_BR
Detector 0 100
Detector 1 100
Detector 2 100
Detector 3 100
Detector 4 100
Detector 5 100
Detector 6 100
Detector 7 100
D - AEC TARGET_TIMES1_BR
TARGET_TIMES1_BUCKY_MO_BR Bucky Mo a 175
Bucky Mo b 50
Bucky Mo s 1
TARGET_TIMES1_BUCKY_RH_BR Bucky Rh a 175
Bucky Rh b 50
Bucky Rh s 1
TARGET_TIMES1_MAG_MO_BR Mag Mo a 175
Mag Mo b 50
Mag Mo s 1
TARGET_TIMES1_MAG_RH_BR Mag Rh a 175
Mag Rh b 50
Mag Rh s 1
TARGET_TIMES1_CONTACT_MO_BR Contact Mo a 175
Contact Mo b 50
Contact Mo s 1
TARGET_TIMES1_CONTACT_RH_BR Contact Rh a 175
Contact Rh b 50
Contact Rh s 1
TARGET_TIMES2_BR
TARGET_TIMES2_BUCKY_MO_BR Bucky Mo a 175
Bucky Mo b 50
Bucky Mo s 1
TARGET_TIMES2_BUCKY_RH_BR Bucky Rh a 175
Bucky Rh b 50
Bucky Rh s 1
TARGET_TIMES2_MAG_MO_BR Mag Mo a 175
Mag Mo b 50
Mag Mo s 1
TARGET_TIMES2_MAG_RH_BR Mag Rh a 175
Mag Rh b 50
Mag Rh s 1
TARGET_TIMES2_CONTACT_MO_BR Contact Mo a 175
Contact Mo b 50
Contact Mo s 1
TARGET_TIMES2_CONTACT_RH_BR Contact Rh a 175
Contact Rh b 50
Contact Rh s 1
INITIAL_KV_TABLE_BUCKY_MO_BR
Thickness: 0,5 cm 25 23 35
Thickness: 1,0 cm 25 23 35
Thickness: 1,5 cm 25 23 35
Thickness: 2,0 cm 25 23 35
Thickness: 2,5 cm 26 23 35
Thickness: 3,0 cm 26 23 35
Thickness: 3,5 cm 27 23 35
Thickness: 4,0 cm 27 23 35
Thickness: 4,5 cm 27 23 35
Thickness: 5,0 cm 28 23 35
Thickness: 5,5 cm 28 23 35
Thickness: 6,0 cm 28 23 35
Thickness: 6,5 cm 29 23 35
Thickness: 7,0 cm 29 23 35
Thickness: 7,5 cm 29 23 35
Thickness: 8,0 cm 30 23 35
Thickness: 8,5 cm 30 23 35
Thickness: 9,0 cm 30 23 35
Thickness: 9,5 cm 30 23 35
Thickness: 10 cm 30 23 35
INITIAL_KV_TABLE_MAG_MO_BR
Thickness: 0,5 cm 25 25 35
Thickness: 1,0 cm 25 25 35
Thickness: 1,5 cm 25 25 35
Thickness: 2,0 cm 25 25 35
Thickness: 2,5 cm 25 25 35
Thickness: 3,0 cm 25 25 35
Thickness: 3,5 cm 25 25 35
Thickness: 4,0 cm 26 25 35
Thickness: 4,5 cm 26 25 35
Thickness: 5,0 cm 26 25 35
Thickness: 5,5 cm 27 25 35
Thickness: 6,0 cm 27 25 35
Thickness: 6,5 cm 28 25 35
Thickness: 7,0 cm 28 25 35
Thickness: 7,5 cm 28 25 35
Thickness: 8,0 cm 29 25 35
Thickness: 8,5 cm 29 25 35
Thickness: 9,0 cm 30 25 35
Thickness: 9,5 cm 30 25 35
Thickness: 10 cm 30 25 35
INITIAL_KV_TABLE_BUCKY_RH_BR
Thickness: 0,5 cm 25 23 35
Thickness: 1,0 cm 25 23 35
Thickness: 1,5 cm 25 23 35
Thickness: 2,0 cm 25 23 35
Thickness: 2,5 cm 26 23 35
Thickness: 3,0 cm 26 23 35
Thickness: 3,5 cm 27 23 35
Thickness: 4,0 cm 27 23 35
Thickness: 4,5 cm 27 23 35
Thickness: 5,0 cm 28 23 35
Thickness: 5,5 cm 28 23 35
Thickness: 6,0 cm 28 23 35
Thickness: 6,5 cm 29 23 35
Thickness: 7,0 cm 29 23 35
Thickness: 7,5 cm 29 23 35
Thickness: 8,0 cm 30 23 35
Thickness: 8,5 cm 30 23 35
Thickness: 9,0 cm 30 23 35
Thickness: 9,5 cm 30 23 35
Thickness: 10 cm 30 23 35
INITIAL_KV_TABLE_MAG_RH_BR
Thickness: 0,5 cm 25 25 35
Thickness: 1,0 cm 25 25 35
Thickness: 1,5 cm 25 25 35
Thickness: 2,0 cm 25 25 35
Thickness: 2,5 cm 25 25 35
Thickness: 3,0 cm 25 25 35
Thickness: 3,5 cm 25 25 35
Thickness: 4,0 cm 26 25 35
Thickness: 4,5 cm 26 25 35
Thickness: 5,0 cm 26 25 35
Thickness: 5,5 cm 27 25 35
Thickness: 6,0 cm 27 25 35
Thickness: 6,5 cm 28 25 35
Thickness: 7,0 cm 28 25 35
Thickness: 7,5 cm 28 25 35
Thickness: 8,0 cm 29 25 35
Thickness: 8,5 cm 29 25 35
Thickness: 9,0 cm 30 25 35
Thickness: 9,5 cm 30 25 35
Thickness: 10 cm 30 25 35
QA_COEF_HISTORY_BR
QA coeff0 0
QA coeff1 0
QA coeff2 0
QA coeff3 0
QA coeff4 0
QA coeff5 0
QA coeff6 0
QA coeff7 0
QA coeff8 0
QA coeff9 0
QA coeff10 0
QA coeff11 0
QA coeff12 0
QA coeff13 0
QA coeff14 0
QA coeff15 0
QA coeff16 0
QA coeff17 0
QA coeff18 0
QA coeff19 0
QA_COUNTER_BR 0 0 20
E - C-arm Rotation ROTATION_SPEED_BR
CARM_SPD_OFFSET_BR Speed offset 0 0 3
PROJECTIONS_BR
LATERAL_ANGLE_BR Lateral Angle 270 0 359
OBLIQUE_ANGLE_BR Oblique angle 225 0 359
MIRROR_ANGLES_BR Mirror Angle On/off On 0 1
CARM_BUTTONS_BR
CARM_CONTINUOUS_DRV_BR Continuous drive (On/Off) On 0 1
REAR_PLATE_BR
RPLATE_CASH0_BR Front plate 39002 0 50000
RPLATE_CASH1_BR Bucky 18x24 9500 0 50000
RPLATE_CASH2_BR Bucky 24x30 1000 0 50000
RPLATE_CASH3_BR Mag 1.6 9500 0 50000
RPLATE_CASH4_BR Unknown 10000 0 50000
RPLATE_CASH5_BR Cont 18x24 9500 0 50000
RPLATE_CASH6_BR Cont 24x30 1000 0 50000
RPLATE_CASH7_BR Mag 1.8 9500 0 50000
RPLATE_CASH8_BR SpotMag 1.6 9500 0 50000
RPLATE_CASH9_BR Stereo dig 27000 0 50000
RPLATE_CASH10_BR Stereo film 10000 0 50000
RPLATE_CASH11_BR Mag 2.0 9500 0 50000
RPLATE_CASH12_BR Reserved 10000 0 50000
RPLATE_CASH13_BR Reserved 10000 0 50000
RPLATE_CASH14_BR Reserved 10000 0 50000
RPLATE_CASH15_BR Reserved 10000 0 50000
F - Collimator SIDE_PLATES_BR
SPLATE_CASH0_BR Front plate 0 0 50000
SPLATE_CASH1_BR Bucky 18x24 4800 0 50000
SPLATE_CASH2_BR Bucky 24x30 300 0 50000
SPLATE_CASH3_BR Mag 1.6 4800 0 50000
SPLATE_CASH4_BR Unknown 4100 0 50000
SPLATE_CASH5_BR Cont 18x24 4800 0 50000
SPLATE_CASH6_BR Cont 24x30 300 0 50000
SPLATE_CASH7_BR Mag 1.8 4800 0 50000
SPLATE_CASH8_BR SpotMag 1.6 4800 0 50000
SPLATE_CASH9_BR Stereo dig 17500 0 50000
SPLATE_CASH10_BR Stereo film 4100 0 50000
SPLATE_CASH11_BR Mag 2.0 4800 0 50000
SPLATE_CASH12_BR Reserved 4100 0 50000
SPLATE_CASH13_BR Reserved 4100 0 50000
SPLATE_CASH14_BR Reserved 4100 0 50000
SPLATE_CASH15_BR Reserved 4100 0 50000
F - Collimator COLLIMATOR_LIGHT_BR Light Intensity 7 0 10
COL_LIGHT_TIME_BR Time Normal 20-60sec 40 20 60
COL_LIGHT_TIME_STEREO_BR Time Stereo/biopsy 20-90sec 80 20 90
COMPRESSION_SETUP_BR
COMP_SLOW_SPD_BR Slow Speed:1-5 4 2 7
COMP_HIGH_SPD_BR High Speed:1-5 6 2 7
COMP_REL_SPD_BR Release Speed: 1-5 5 1 7
COMP_AUTOREL_TRAVEL_BR AutoRelease travel: 0 - 20 cm 15 0 20
COMP_SPD_THRESHOLD_BR Speed Threshold: 0-150N 50 0 150
G - Compression THICKNESS_CASS_HOLDER_BR
THICKNESS_CASS_HOLDER0_BR No holder 19 0 40
THICKNESS_CASS_HOLDER1_BR Bucky 18x24 19 0 40
THICKNESS_CASS_HOLDER2_BR Bucky 24x30 19 0 40
THICKNESS_CASS_HOLDER3_BR Mag 1.6 19 0 40
THICKNESS_CASS_HOLDER4_BR Unknown 19 0 40
THICKNESS_CASS_HOLDER5_BR Cont 18x24 19 0 40
THICKNESS_CASS_HOLDER6_BR Cont 24x30 19 0 40
THICKNESS_CASS_HOLDER7_BR Mag 1.8 10 0 40
THICKNESS_CASS_HOLDER8_BR Mag 1.8 spot 19 0 40
THICKNESS_CASS_HOLDER9_BR Stereo digital 19 0 40
THICKNESS_CASS_HOLDER10_BR Stereo film 7 0 40
THICKNESS_CASS_HOLDER11_BR Mag 2.0 3 0 40
THICKNESS_CASS_HOLDER12_BR Reserved 19 0 40
THICKNESS_CASS_HOLDER13_BR Reserved 19 0 40
THICKNESS_CASS_HOLDER14_BR Reserved 19 0 40
THICKNESS_CASS_HOLDER15_BR Reserved 19 0 40
G - Compression THICKNESS_BASE_POINT_BR Thickness base point 10
FORCE_LIMITS_BR
Mode: 0/1 0 0 1
FORCE_LIMITS_STOP_BR
Position 1 : 0-255N 70 0 255
Position 2 : 0-255N 90 0 255
Position 3 : 0-255N 110 0 255
Position 4 : 0-255N 130 0 255
Position 5 : 0-255N 150 0 255
Position 6 : 0-255N 170 0 255
FORCE_LIMIT_DELTA_BR
Max delta compression 200
NEGATIVE_FORCE_LIMIT_BR Negative force limit 4 1 10
H - Filter FILTER_CALIBRATION_BR
FILTER_MO_CALIBRATION_BR Mo 1950 0 20000
FILTER_RH_CALIBRATION_BR Rh 150 0 20000
FILTER_AL_CALIBRATION_BR Al 3150 0 20000
AUTO_FILTER_SET_BR
AUTO_FILTER_MODE_BR On/Off 0 0 1
AUTO_FILTER_THICK_BR Thickness of filter change point 50
I - Generator GENERATOR_SETUP_BR
ANODE_BREAK_BR Anode Brake On/Off 0
Use high mAs On/Off 0
Min kV 23
Max kV 39
PREHEAT_COMPENSATION_LF_BR
PREH_COMP_LF_BASE_BR Preheat Calib_LF (-20 - +20) 102 95 105
PREH_COMP_LF_KV_BR Preheat_LF 15kV 54 20 235
Preheat_LF 16kV 55 20 235
Preheat_LF 17kV 55 20 235
Preheat_LF 18kV 55 20 235
Preheat_LF 19kV 56 20 235
Preheat_LF 20kV 56 20 235
Preheat_LF 21kV 56 20 235
Preheat_LF 22kV 56 20 235
Preheat_LF 23kV 57 20 235
Preheat_LF 24kV 57 20 235
Preheat_LF 25kV 57 20 235
Preheat_LF 26kV 56 20 235
Preheat_LF 27kV 55 20 235
Preheat_LF 28kV 54 20 235
Preheat_LF 29kV 54 20 235
Preheat_LF 30kV 53 20 235
Preheat_LF 31kV 53 20 235
Preheat_LF 32kV 53 20 235
Preheat_LF 33kV 52 20 235
Preheat_LF 34kV 52 20 235
Preheat_LF 35kV 51 20 235
Preheat_LF 36kV 48 20 235
Preheat_LF 37kV 48 20 235
Preheat_LF 38kV 48 20 235
Preheat_LF 39kV 48 20 235
I - Generator PREHEAT_COMPENSATION_SF_BR
PREH_COMP_SF_BASE_BR Preheat Calib_SF (0.5-1.50) 100 95 105
PREH_COMP_SF_KV_BR Preheat_SF 15kV 48 20 235
Preheat_SF 16kV 48 20 235
Preheat_SF 17kV 49 20 235
Preheat_SF 18kV 49 20 235
Preheat_SF 19kV 49 20 235
Preheat_SF 20kV 49 20 235
Preheat_SF 21kV 49 20 235
Preheat_SF 22kV 49 20 235
Preheat_SF 23kV 50 20 235
Preheat_SF 24kV 49 20 235
Preheat_SF 25kV 49 20 235
Preheat_SF 26kV 48 20 235
Preheat_SF 27kV 48 20 235
Preheat_SF 28kV 47 20 235
Preheat_SF 29kV 47 20 235
Preheat_SF 30kV 47 20 235
Preheat_SF 31kV 46 20 235
Preheat_SF 32kV 46 20 235
Preheat_SF 33kV 46 20 235
Preheat_SF 34kV 45 20 235
Preheat_SF 35kV 45 20 235
Preheat_SF 36kV 45 20 235
Preheat_SF 37kV 45 20 235
Preheat_SF 38kV 45 20 235
Preheat_SF 39kV 45 20 235
J - Tube Park SPEEDSTUBE_BR
Fast Speed 1-5 5 1 5
Slow Speed 2 1 5
TUBE_EXP_DRIVE_BR Exp drive 1 0 1
TUBE_AUTOPARK_BR AutoPark 0 0 1
STEREO_LOW_SPD_LIMIT_BR Slow Speed 2
STEREO_HIGH_SPD_LIMIT_BR High Speed 5
STEREO_SPD_OFFSET_BR Speeds table offset 10 8 12
STEREO_BREAK_SPD_BR Breaking speed 3
STEREO_RAMP_DELAY_BR Accel. ramp speed chg.delay 25
STEREOARM_CONTINUOUS_DRV_BR
Continuous drive (On/Off) 0 0 1
BRIGHTNESS_BR Brightness 5 1 5
CONTRAST_BR Contrast 5 1 5
LABEL_BR
ANGLE_CALIBRATION_BR
FORCE_CALIBRATION_BR
Force slope 100 0 65535
Force offset 40 0 65535
CPULSE_BR
STEREOARMPULSES_BR
STEREOARMPULSEPERANGLE_BR
STEREOARMZEROANGLE_BR
&KHFN/LVWIRU0DLQWHQDQFHDQG,QVWDOODWLRQ$SSHQGL[WR'LDPRQG,QVWDOODWLRQ,QVWUXFWLRQV
(&6029(0(17
• collimator light will light up when Paddle is pressed
• verify good synchronization with compression; vary paddle speed by pressing with hand
• check for smooth, quiet operation at 180°, +90°, and -90° C-Arm angles.
• press 5HO-button of the foot pedal, listen that C-arm movement will not crash on the
mechanical stop. If it does:
67(5(2029(0(17
• stops exactly at middle, right and left positions
• not interrupted during movement from one position to another
- C-arm in CC-position
- C-arm in left LATERAL-position
- C-arm in right LATERAL-position
/DVWXSGDWH(3R 3DJH
&KHFN/LVWIRU0DLQWHQDQFHDQG,QVWDOODWLRQ$SSHQGL[WR'LDPRQG,QVWDOODWLRQ,QVWUXFWLRQV
78%(3$5.
• mechanically stops exactly at exposure position (front position)
- mechanical stop can be heard
- will not slide or bounce back (belt tension OK, not too loose)
• verify that the safety clutch does NOT disengage when the tube is in the fully-extended
position
• moves freely all the way to front and back position at following angles
a) CC
b) left LAT position, -90°
c) right LAT position, +90°
• safety stop intervenes and stops the movement when force in CC-position exceeds the force
set by manual clutch spring adjustment
- limit 5 kg
• no noise of scraping covers or rubbing cables during movement
• if finger protection sheets are noisy, lubricate sliding surfaces
&2035(66,21
• force ok with 6 different force settings
- normal paddle/bucky
- spot/magn paddle/magn table
• high speed ( 8.5 -9.5 cm/5 sec.), drive down with paddle to test
• slow speed (6 -7.5 cm/5 sec, drive with ECS paddle to test
• during the ECS movement the paddle should move up by max 25 mm
• not noisy
• will not generate force upside down position
• no error message 24 appears when C-arm is rotated upside down (to 180 degrees). If it
appears:
a) check Up Force Limit parameter in service program PrG06
b) calibrate pre-tension
/DVWXSGDWH(3R 3DJH
&KHFN/LVWIRU0DLQWHQDQFHDQG,QVWDOODWLRQ$SSHQGL[WR'LDPRQG,QVWDOODWLRQ,QVWUXFWLRQV
&$66(77(+2/'(5&$66(77(6
• when setting cassette inside, cassette will slide forward to the front position
• forward and backward movement in following C-arm positions is OK:
Small bucky
a) CC
b) left LAT position
c) right LAT position
Large bucky
a) CC
b) left LAT position
c) right LAT position
Magnification
a) CC
b) left LAT position
c) right LAT position
Magnification
- will display 1.6, 1.8 and 2.0 positions
- cassette moves freely in and out
• verify proper connection between grid and drive mechanism. Check spring-pin tension,
check for pin damage
• check for worn sliding rails and roller bearings. Clean/Lubricate
• check for sharp edges, especially near patient contact surfaces
• check in following positions:
a) CC
b) left LAT position
c) right LAT position
When a cassette is inserted into the cassette holder, it starts to slide forward to the front
position. If not, adjust cassette optosensor sensitivity with trimmer on Casopto Board.
/DVWXSGDWH(3R 3DJH
&KHFN/LVWIRU0DLQWHQDQFHDQG,QVWDOODWLRQ$SSHQGL[WR'LDPRQG,QVWDOODWLRQ,QVWUXFWLRQV
3$''/(6
- all paddles are easily inserted and released (manually)
3DGGOHGHWHFWLRQ
• test with the paddles delivered to the customer with the original unit delivery
• paddle front edge less than 1mm away from the cassette holder front edge
- Bucky with different paddles
- special magnification table with magnification paddles
- use shim, if necessary
• check that paddle distance display is correct with all different paddles
- use for example 4 cm acrylic and check that 4 cm is displayed
- adjustment from service software compression PrG03
&2//,0$725',$3+5$*06
• back and side diaphragms are driven by motor, slide freely and open small focus front plate
• in order to test, change from Bucky to Magnification holder
• check that X-ray field conforms to your local regulations
- small, large Bucky and Magnification
/DVWXSGDWH(3R 3DJH
&KHFN/LVWIRU0DLQWHQDQFHDQG,QVWDOODWLRQ$SSHQGL[WR'LDPRQG,QVWDOODWLRQ,QVWUXFWLRQV
),/7(5
• filter runs easily between Molybdenum - Rhodium - Aluminum
=029(0(17
• run this test first without covers
• attach covers and run this test again
• Z-direction has strong enough friction to keep the C-arm in its position, when patient is
leaning to Bucky (screw adjustment in the column)
• movement responds to up and down micro switches
• movement is not noisy and moves freely in following C-arm positions
a) CC
b) left LAT position
c) right LAT position
6$)(7<)($785(6
6DIHW\VZLWFK
$'',7,21$/
• unit is clean and with no scratches
• covers
• Buckies and Magnification
• face shield
• rotate C-arm 180° in both directions, there should be no loose parts inside the covers that
might cause noise or short-circuits
,19(1725<
• check out the packing list
/DVWXSGDWH(3R 3DJH
Diamond AEC detector locations
for detector-specific film O.D. measurement
Film marker text
R1
R2
C4
C3
C2
C1
L2
L1
Instructions:
Tel. +1 800 558 6120, +1 414 747 1030 20090 Segrate (MI), Italy 4181 Latham Street Riverside, CA 92501, USA
Fax +1 414 481 8665 Tel. +39 02 21 30 28 1 · Fax +39 02 21 30 28 60 Tel. +1 909 781 6660 · Fax +1 909 781 6457
[email protected] [email protected] [email protected]
Instrumentarium Imaging is constantly improving its products and reserves the right to
change these specifications without notice. Distributor:
www.InstrumentariumImaging.com