iVA - Cable & Antenna Analyzer: VSWR, Return Loss Measurement & Distance To Fault
iVA - Cable & Antenna Analyzer: VSWR, Return Loss Measurement & Distance To Fault
PRODUCT FEATURES
Reinventing site certification sweep testing, dramatically reducing test time on site
Directly Measure insertion loss and isolation when using multiple iVA’s. Measure calculated insertion loss with a single iVA and
an RF short
Accurately measure swept VSWR/return loss and Distance-to-Fault in RF path
Can be used in conjunction with an iPA Series PIM Analyzer to directly measure insertion loss and isolation
Connect directly to the device under test; Eliminates the need for a phase stable cable in most cases
With the Kaelus iPA controlling the iVA, your RL data can be combined with your PIM data into a single report. Reports are
combined and completed on-site with no post-processsing required
Uses the Kaelus customer-proven iPA reporting workflow & taggin features to facilitate a faster, simpler and more efficient
workflow
Simple to operate, highly intuitive software user interface with the unique abilityn to generate and complete the test report
on-site
Geotag each test point, insert a Google Maps® snapshot directly into the report
Smart high power input protection when used with a Kaelus iPA
Return Loss
VSWR
iVA Analysis Modes
Cable Loss1
Distance-to-Fault (DTF)
Frequency Range 690 - 2700 MHz
Minimum Frequency Increment 1 kHz
Sweep Speed 4 ms per frequency point
Number of Measurement Points 2 to 2001
RF Output Power +5 dBm ± 3 dB
Return Loss Dynamic Range 0 – 40 dB
VSWR Dynamic Range 1 – 100
Cable Loss Dynamic Range1 0 – 20 dB
Applies over the temperature range −10°C to +45°C, with less than 5°C deviation
Return Loss Measurement Accuracy
from calibration temperature.
0 – 10 dB ± 0.3 dB
10 – 20 dB ± 0.6 dB
20 – 30 dB ± 1.5 dB
30 – 40 dB ± 4.0 dB
Calibrated Directivity 43 dB (typ)
Interference Immunity +10 dBm at 500 kHz offset from stimulus frequency
System Impedance 50 ohms
1.
Cable loss can be measured either as a 1-port measurement, with the far end of the cable terminated in an open or short circuit, or
directly measured for increased accuracy as a 2-port measurement using a second iVA
1.
The iVA offers a novel multi-port S-parameter test capability using multiple iVAs. Up to 7 units can be connected simultaneously via
Bluetooth, while up to 32 can be connected via USB. As an example, 6 iVAs could be used to perform measurements on a multi-port
antenna. This configuration would cover all 36 transmission pathways (6x6), including the return loss at each port (6 measurements),
and the transmission loss between every possible pair of ports (30 measurements). Return loss measurements made by the iVA
contain both both magnitude and phase information, while transmission loss measurements are limited to magnitude only.
1.
Preamp on.
2.
Damage can occur at input levels above +23dBm.
User interface USB or Bluetooth supported user device with iVA Application Software installed
iPA Portable PIM Analyzer
Tablet computer (iOS & Android)
Supported Devices
Smartphone (iOS & Android)
PC, Windows 7 & 8 running .NET verison 4 or later
Communications Interface to iVA Bluetooth and USB 2.0
Bluetooth Antenna Integrated into housing
RF +23dBm max
DC Voltage ±30V
ELECTRICAL
DC Power Consumption
Return Loss Mode 4.7W
Transmission Mode 4.7W
Spectrum Monitor Mode 3.7W
Standby (Idle) 0.6W
Battery Lithium-Ion 3.6V, 2350 mAh, 8.5Wh
Battery Charging Method USB-compatible power source connected to USB port of iVA
Battery Operating Time 8 Hours at typical usage factor