OPERATING MANUAL
Model UT71A/B/C/D/E: OPERATING MANUAL
TABLE OF CONTENTS
CHAPTER TITLE PAGE
1. Before You Start 5
Overview 5
Unpacking Inspection 6
Safety Information 7
Rules For Safe Operation 7
International Electrical Symbols 9
2. Getting Acquainted 10
Turning the Meter On 10
Battery Considerations 10
Automatic Power Off 10
Automatic Backlight Off 10
Low Battery Indication 11
The Meter Structure 11
Rotary Switch 12
Functional Buttons 15
The Meter Functions Vs Displays 21
Selecting the Range 24
Understanding the Display 24
1
Model UT71A/B/C/D/E: OPERATING MANUAL
2
Model UT71A/B/C/D/E: OPERATING MANUAL
3
Model UT71A/B/C/D/E: OPERATING MANUAL
4
Model UT71A/B/C/D/E: OPERATING MANUAL
Chapter 1
Before You Start
Overview
This Operating Manual covers information on safety and cautions. Please read the relevant information carefully and
observe all the Warnings and Notes strictly.
Warning
To avoid electric shock or personal injury, read the ìSafety Informationî and ìRules for Safe Operationî carefully
before using the Meter.
Digital Multimeter UT71A/B (hereafter referred to as ìthe Meterî) is a 20000 counts and 4 1/2 digits with steady
operations; Digital Multimeter UT71C/D/E (hereafter referred to as ìthe Meterî) is a 40000 counts and 4 3/4 digits
with steady operations, fashionable structure and auto ranging instrument. They all not only can measure AC voltage
and current, DC voltage and current, Resistance, Capacitance, Temperature, Frequency, Diodes, Continuity, 4~20mA
Loop, Max/Min, Relative Mode but also has Data Store, Data Recall, AC True RMS, AC+DC,Low Battery Display,
Double Display Backlight, Data Hold, Automatic Power Off and full overload protection.
UT71B/C/D/E has extra Temperature, 4~20mA Loop, Data Store and Data Recall feature; UT71E has extra Power
Measurement feature.
5
Model UT71A/B/C/D/E: OPERATING MANUAL
Unpacking Inspection
Open the package case and take out the Meter. Check the items shown on Table 1-1 carefully to see any missing
or damaged part:
Table 1-1. Unpacking Inspection
Item Description Qty
1 English Operating Manual 1 piece
2 Test Lead 1 pair
3 K-Type (nickel chromium ~ nickel silicon) Point Contact Temperature Probe (It is 1 piece
only suitable for measuring temperature under 230°Ê
4 Alligator Clip 1 piece
5 Test Clip 1 pair
6 USB interface cable 1 piece
7 CD-ROM (Installation Guide & Computer Interface Software) 1 piece
8 Carrying Bag 1 piece
9 Power Adaptor (UT71E only) 1 piece
10 9V Battery (NEDA 1604, 6F22, 006P) 1 piece
In the event you find any missing or damage, please contact your dealer immediately.
6
Model UT71A/B/C/D/E: OPERATING MANUAL
7
Model UT71A/B/C/D/E: OPERATING MANUAL
l When the Meter working at an effective voltage over l When servicing the Meter, use only the same model
60V in DC or 30V rms in AC, special care should be number or identical electrical specifications
taken for there is danger of electric shock. replacement parts.
l Use the proper terminals, function, and range for l The internal circuit of the Meter shall not be altered
your measurements. at will to avoid damage of the Meter and any accident.
l If the value to be measured is unknown, use the l Soft cloth and mild detergent should be used to
maximum measurement position. clean the surface of the Meter when servicing. No
l Do not use or store the Meter in an environment of abrasive and solvent should be used to prevent the
high temperature, humidity, explosive, inflammable surface of the Meter from corrosion, damage and
and strong magnetic field. The performance of the accident.
Meter may deteriorate after dampened. l The Meter is suitable for indoor use.
l When using the test leads, keep your fingers behind l Turn the Meter off when it is not in use and take out
the finger guards. the battery when not using for a long time.
l Disconnect circuit power and discharge all high- l Constantly check the battery as it may leak when it
voltage capacitors before testing resistance, has been using for some time, replace the battery
continuity, diodes. as soon as leaking appears. A leaking battery will
l Before measuring current, check the Meterís fuses damage the Meter.
and turn off power to the circuit before connecting
the Meter to the circuit.
l Replace the battery as soon as the battery indicator
appears. With a low battery, the Meter might
produce false readings that can lead to electric
shock and personal injury.
8
Model UT71A/B/C/D/E: OPERATING MANUAL
9
Model UT71A/B/C/D/E: OPERATING MANUAL
Chapter 2
Getting Acquainted
Turning the Meter On The automatic power off is preset to 10 minutes. From
To turn the Meter on, turn the rotary switch from OFF the Setup menu (see Chapter 5), you could specify a
to any switch setting. time (10 minutes, 20 minutes, 30 minutes or OFF). If
you set to OFF, the Meter retains on until you turn the
rotary switch to OFF or the battery becomes too weak.
Battery Considerations
The Meter uses one 9V Battery (NEDA 1604, 6F22,
006P). The following paragraphs describe several Automatic Backlight Off
techniques used to conserve battery power. Press LIGHT button to turn the backlight on and press
LIGHT again to turn it off. Press EXIT to exit the feature.
Automatic Power Off
The display blanks and the Meter goes into a ìsleepî Press LIGHT to select the backlight level (low or high).
mode if you have not changed the rotary switch position In Setup menu (see Chapter 5), you could specify a
or pressed a button for a set period. While in Sleep time to automatically turn off the backlight (10 seconds,
mode, pressing the blue button or turning the rotary 20 seconds, 30 seconds or OFF). If the period is set
switch could turn the Meter on. The Meter then returns to OFF, the backlight feature is disabled.
to the display for the function selected with the rotary
switch; all previously activated button features are
discarded.
10
Model UT71A/B/C/D/E: OPERATING MANUAL
Warning
To avoid false readings, which could lead to possible
electric shock or personal injury, replace the battery as
soon as the battery icon ( ) appears.
1. LCD Display
2. Functional Buttons
3. Rotary Switch
4. Input Terminals
11
Model UT71A/B/C/D/E: OPERATING MANUAL
Rotary Switch
Turn the Meter on by selecting any measurement function. The Meter presents a standard display for that function.
The display may also be influenced by some of the choices made in Setup.
Use the blue button to select any rotary switch alternate function (labeled in blue letters).
When you turn the rotary switch from one function to another, a display for the new function appears. Button choices
made in one function do not carry over into another function.
12
Model UT71A/B/C/D/E: OPERATING MANUAL
13
Model UT71A/B/C/D/E: OPERATING MANUAL
Rotary
Rotary Switch Function Blue Key Function
Switch Position
Hz % Centigrade temperature measurement l Fahrenheit temperature measurement
∫C ∫F l Frequency measurement
(UT71E only) l Frequency signal duty cycle measurement
µA AC or DC current measurement (400µA ,4000µA) Toggle between AC or DC current
mA AC or DC current measurement (40mA ,400mA) Toggle between AC or DC current
(4~20mA)
% 4~20mA loop current as % reading
A AC or DC current measurement (10A) Toggle between AC or DC current
14
Model UT71A/B/C/D/E: OPERATING MANUAL
Functional Buttons
The buttons activate features that augment the function selected with the rotary switch. The buttons are shown
in Table 2-2.
RECALL
STORE
Press the button once to access the main feature (e.g. STORE).
To access the first additional feature of the button (e.g. RECALL), press and hold the button for over 1 second to
access this additional feature. This additional feature appears right above or on the left hand side of the appropriate
keys.
To access the second additional feature of the button (e.g. ), press once the button again while the Meter has
already entered the first additional feature (e.g. RECALL). The second additional feature appears on the right hand
side above the appropriate keys.
The RANGE and EXIT buttons has only one additional feature.
15
Model UT71A/B/C/D/E: OPERATING MANUAL
16
Model UT71A/B/C/D/E: OPERATING MANUAL
18
Model UT71A/B/C/D/E: OPERATING MANUAL
19
Model UT71A/B/C/D/E: OPERATING MANUAL
Use the blue button to select any rotary switch alternate function Press the button once
(labeled in blue letters).
Press and hold the Button while turning on the Meter to toggle
Blue Button
to 4000 counts for all functions.
It is faster when the Meter is at 4000 counts measurement
mode.
After the Meter is resuming from Automatic Power Off or turn
on and off again, the Meter will back to normal measurement
mode (40000 counts).
20
Model UT71A/B/C/D/E: OPERATING MANUAL
ACV The tested AC voltage value The tested frequency value: Full range: 4, 40, 400, 1000
45.00Hz~ 100.0kHz
DCmV The tested DCmV value No display Full range 400
The tested resistance value No display Full range: 400, 4, 40, 400, 4, 40
The tested resistance value No display Full range value: 400
The tested resistance value No display Full range 4
Hz The tested frequency value No display Full range: 40, 400, 4, 40, 400, 4, 40, 400
The tested capacitance value No display Full range: 40, 400, 4, 40, 400, 4, 40
∫C The tested ∫C value No display 1000
∫F The tested ∫F value No display 1832
DCµA The tested DCµA value No display Full range: 400, 4000
21
Model UT71A/B/C/D/E: OPERATING MANUAL
22
Model UT71A/B/C/D/E: OPERATING MANUAL
23
Model UT71A/B/C/D/E: OPERATING MANUAL
24
Model UT71A/B/C/D/E: OPERATING MANUAL
25
Model UT71A/B/C/D/E: OPERATING MANUAL
26
Model UT71A/B/C/D/E: OPERATING MANUAL
27
Model UT71A/B/C/D/E: OPERATING MANUAL
28
Model UT71A/B/C/D/E: OPERATING MANUAL
Analogue Bar Graph the right secondary display. The primary display shows
The bar graph provides an analogue indication of the the maximum value.
measured input. For most measurement functions, the
bar graph updates 10 times per second. Press MAX MIN the third time, the current measurement
reading is shown on the left secondary display. The
maximum reading and MAX are shown on the right
Using MAX MIN secondary display. The primary display shows the
The MAX MIN mode stores minimum (MIN) and minimum value.
maximum (MAX) input values. When the input goes
below the stored minimum value or above the stored Each subsequent press of MAX MIN steps through the
maximum value, the Meter beeps and stores the new above three modes.
value.
To exit MAX MIN mode, press EXIT.
Press MAX MIN to enter MAX MIN mode. The sampling
time is every 2 seconds. The maximum reading and Press HOLD to stop the Meter updating reading.
MAX are shown on the left secondary display. The
minimum reading and MIN are shown on the right MAX MIN mode can only be used under MANUAL
secondary display. The primary display shows the ranging mode.
current measurement reading.
Chapter 3
Making Measurement
Introduction
Chapter 3 explains how to make measurements. Most
measurement functions can be selected by using the
rotary switch.
A. Measuring Voltages
Warning
To avoid harms to you or damages to the Meter
from electric shock, please do not attempt to
measure voltages higher than 1000V, although
readings may be obtained.
Figure 3-1. Voltages Measurement
To measure voltages, set up the Meter as Figure 3-1
and do the following:
30
Model UT71A/B/C/D/E: OPERATING MANUAL
1. Insert the red test lead into the V terminal and the Note
black test lead into the COM terminal. l When measuring voltage, the Meter acts around a
2. Set the rotary switch to V or V or mV ; 10M ( V and V ) or 2.5G ( mV ) impedance in
Set the rotary switch to or V or mV (UT71E). parallel with the circuit. This loading effect can cause
The default is DC voltage measurement, press BLUE measurement errors in high impedance circuits. In
button to switch to AC voltage measurement mode. most cases, the error is negligible (0.1% or less) if
3. Connect the test leads across with the object being the circuit impedance is 10k or less.
measured. l Special care should be taken when measuring high
4. The measured value shows on the display. voltage.
AC measurement displays the True RMS value. l When voltage measurement has been completed,
DC measurement displays the effective value of sine disconnect the connection between the testing leads
wave (mean value response). and the circuit under test and remove testing leads
away from the input terminals of the Meter.
When a ACV function is selected, you can press the
Yellow Button to view the AC + DC True RMS value
in the primary display. To exit, please EXIT button.
31
Model UT71A/B/C/D/E: OPERATING MANUAL
B. Measuring Currents
Warning
If the fuse burns out during measurement, the Meter
may be damaged or the operator himself may be hurt.
32
Model UT71A/B/C/D/E: OPERATING MANUAL
1. Turn off power to the circuit. Discharge all high- When a ACA function is selected, you can press the
voltage capacitors. Yellow Button to view the AC + DC True RMS value
2. Insert the red test lead into the mAµA or A terminal in the primary display. To exit, please EXIT button.
and black test lead into the COM terminal.
3. If you are using the A terminal, set the rotary switch Note
to A . If you are using mAµA terminal, set the
l If the value to be measured is unknown, use the
rotary switch to µA for currents below 40000µA
maximum measurement position and reduce the
(UT71A is 20000µA), or mA for current above
range step by step until a satisfactory reading is
40000µA (UT71B is 20000µA).
obtained.
4. DC measurement is default, press blue button to
l When the measured current is 5A, continuous
select AC measurement.t
measurement is allowed.
5. Open the circuit path to be tested. Touch the red
l When the measured current is between 5A-10A,
testing leads to the more positive side of the break;
continuous measurement 10 seconds and interval
touch the black probe to the more negative side of
more than 15 minutes.
the bread. Reversing the leads will produce a
l When current measurement has been completed,
negative reading, but will not damage the Meter.
disconnect the connection between the testing leads
6. Turn on power to the circuit; then read the display.
and the circuit under test and remove testing leads
AC measurement displays the True RMS value.
away from the input terminals of the Meter.
DC measurement displays the effective value of sine
wave (mean value response)
7. Turn off power to the circuit and discharge all high-
voltage capacitors. Remove the Meter and restore
33
Model UT71A/B/C/D/E: OPERATING MANUAL
C. Measuring Resistance
Warning
To avoid harms to you, please do not attempt to input
voltage higher than 60V DC or 30V rms AC.
1. Insert the red test lead into the terminal and the
black test lead into the COM terminal.
2. Set the rotary switch to ; press BLUE button
to select measurement mode.
3. Connect the test leads across with the object being
measured.
The measured value shows on the display.
The BLUE button cycles among resistance, continuity, Figure 3-3. Resistance Measurement
and diode.
34
Model UT71A/B/C/D/E: OPERATING MANUAL
35
Model UT71A/B/C/D/E: OPERATING MANUAL
Note
l Open circuit voltage around ñ1.2V and range is 400
measurement range.
l When continuity testing has been completed,
disconnect the connection between the testing leads
and the circuit under test and remove the test leads
away from the input terminals.
36
Model UT71A/B/C/D/E: OPERATING MANUAL
E. Testing Diodes
Warning
To avoid harms to you, please do not attempt to input
voltages higher than 60V DC or 30V rms AC.
1. Insert the red test lead into the terminal and the
black test lead into the COM terminal.
Figure 3-5. Diode Test
2. Set the rotary switch to ; and press BLUE
button to select measurement mode.
37
Model UT71A/B/C/D/E: OPERATING MANUAL
3. For forward voltage drop readings on any l When diode testing has been completed, disconnect
semiconductor component, place the red test lead on the connection between the testing leads and the
the componentís anode and place the black test lead circuit under test and remove the test leads away
on the componentís cathode. The red test lead polarity from the input terminals.
is ì+î while the black test lead polarity is ì- ì.
The measured value shows on the display.
The BLUE button cycles among resistance, continuity,
and diode.
Note
l In a circuit, a good diode should still produce a
forward voltage drop reading of 0.5V to 0.8V; however,
the reverse voltage drop reading can vary depending
on the resistance of other pathways between the
probe tips.
l Connect the test leads to the proper terminals as
said above to avoid error display.
l The LCD will display OL indicating either open circuit
or wrong polarity connection.
l The unit of diode is volt (V), displaying the positive-
connection voltage-drop value.
l O p e n c i r c u i t v o l t a g e a p p r o x i m a t e 2 . 8 V.
38
Model UT71A/B/C/D/E: OPERATING MANUAL
F. Measuring Capacitance
Warning
To ensure accuracy, the Meter inside is discharged
against the tested capacitor. ìDIS.Cî will be shown
on the display when it is under discharging, this
process will be quite slow.
39
Model UT71A/B/C/D/E: OPERATING MANUAL
2. Set the rotary switch to measurement mode, l When capacitance measurement has been
the Meter may display a fixed reading which is a completed, disconnect the connection between the
internal distributed capacitor value. For testing less testing leads and the circuit under test and remove
than 10nF capacitor, the tested value must subtract the test leads away from the input terminals of the
the internal distributed capacitor value to maintain Meter.
the accuracy.
To improve the measurement accuracy of small value
capacitors (less than 10nF), press REL with the
test leads open to subtract the residual capacitance
of the Meter and leads.
3. It is recommended to use test clip to carry out
measurement to reduce the effect of internal
distributed capacitor.
Note
l The LCD displays OL indicating the tested capacitor
is shorted or it exceeds the maximum range.
l Capacitors larger than 400µF take longer time. The
analogue bar graph shows the time left before
finishing the measurement.
40
Model UT71A/B/C/D/E: OPERATING MANUAL
Warning
To avoid harms to you, please do not attempt to input
voltage higher than 30V rms.
To measure frequency and duty cycle, connect the
Meter as Figure 3-7 and do the following:
1. Insert the red test lead into the Hz terminal and
the black test lead into the COM terminal.
2. Set the rotary switch to (UT71A), (UT71B/C/D)
or ∫C ∫F (UT71E) and press BLUE button to select
the Hz measurement mode for frequency
measurement or % for duty cycle measurement.
The BLUE button cycles among mV , frequency
and duty cycle for UT71C and UT71E.
3. Connect the test leads across with the object being
measured.
The measured value shows on the primary display.
41
Model UT71A/B/C/D/E: OPERATING MANUAL
Note
l The requirement of Input amplitude ìaî is as follows:
When 10Hz~40MHz: 200 mV a 30Vrms;
> 40MHz: Un-specified
l When Hz measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test and remove the test leads
away from the input terminals.
42
Model UT71A/B/C/D/E: OPERATING MANUAL
H. Measuring Temperature
Warning
To avoid harms to you, please do not attempt to
input voltages higher than 60V DC or 30V rms AC.
To measure temperature, set up the Meter as shown
in Figure 3-8 and proceed the following.
1. Set the rotary switch to ∫C ∫F, the display shows OL.
Short circuit the test leads to show the room
temperature. The Meter is default to Celsius ∫C ∫C ∫F
degree unit, you can change units by press the BLUE
button once you have selected the temperature
function.
2. Insert the point contact temperature probe into the
Meter as figure 10.
3. Place the temperature probe to the object being
measured.
The measured value shows on the display after
several seconds.
Figure 3-8. Temperature Measurement
43
Model UT71A/B/C/D/E: OPERATING MANUAL
44
Model UT71A/B/C/D/E: OPERATING MANUAL
Warning
To avoid damages to the Meter, please do not attempt
to input higher than 250V from outlet altogether reading
may be obtained.
Switch off the power before the Meter and the object
being measured connect to the circuit.
Note
l The current of the object being measured must >
10A.
5A continuous measurement is allowed.
l 5A~10A, only 10 seconds continuous measurement
is allowed and the interval between each
measurement must be greater than 15 minutes.
When power measurement has been completed,
first switch off the power, then disconnect the
connection between the adaptor and the outlet.
46
Model UT71A/B/C/D/E: OPERATING MANUAL
Chapter 4
Using Store, Recall & Send Features
47
Model UT71A/B/C/D/E: OPERATING MANUAL
48
Model UT71A/B/C/D/E: OPERATING MANUAL
Chapter 5
Changing the Default Setting
Introduction
The Meter allows you to change the default operating configuration of the Meter by changing setup options made
at the factory.
These settings are stored and can be changed in the Setup mode using the procedure described in this chapter.
In the Setup mode, each press of SETUP button steps to the next Selection. Each press of ñ or + button decrement
or increment an Option.
Each Setup Selection and Option appears in the primary display in the sequence shown in Table 5-1.
49
Model UT71A/B/C/D/E: OPERATING MANUAL
50
Model UT71A/B/C/D/E: OPERATING MANUAL
51
Model UT71A/B/C/D/E: OPERATING MANUAL
Chapter 6
Maintenance
This chapter provides basic maintenance information including battery and fuse replacement instruction.
Warning
Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration,
performance test, and service information.
A. General Service
l Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
l To clean the terminals with cotton bar with detergent, as dirt or moisture in the terminals can affect readings.
l Turn the Meter to OFF when it is not in use.
l Take out the battery when it is not using for a long time.
l Do not use or store the Meter in a place of humidity, high temperature, explosive, inflammable and strong magnetic
field.
52
Model UT71A/B/C/D/E: OPERATING MANUAL
Warning
To avoid electrical shock or arc blast, or personal
injury or damage to the Meter, use specified fuses
ONLY in accordance with the following procedure.
Follow Figure 6-1 and proceed as follows to replace
the Meterís fuse:
l Turn the rotary switch to OFF and remove all
connections from the terminals.
l Remove the 5 screws from the case bottom.
l Remove the fuse by gently prying one end loose,
then take out the fuse from its bracket.
l Install ONLY replacement fuses with the identical
type and specification as follows and make sure the
fuse is fixed firmly in the bracket.
Fuse 1: 0.5A, 250V, fast type fuse, 5◊20mm
Fuse 2: 10A, 250V, fast type fuse, 5◊20mm
l Rejoin the case bottom and case top, and install the
5 screws.
Figure 6-1. Fuse Replacement
Replacement of the fuses is seldom required. Burning
of a fuse always results from improper operation.
53
Model UT71A/B/C/D/E: OPERATING MANUAL
Warning
To avoid false readings, which could lead to possible
electric shock or personal injury, replace the battery
as soon as the battery indicator ì î appears.
54
Model UT71A/B/C/D/E: OPERATING MANUAL
Chapter 7
Specifications
55
Model UT71A/B/C/D/E: OPERATING MANUAL
Physical Specifications
Display (LCD) Digital: 40000 counts on primary display; updates 2-3
times / second.
4000 counts on two secondary displays.
Analog: 40 segments; updates 10 times / second.
Operating Temperature 0∫C~40∫C (32∫F~104∫F)
Storage Temperature -10∫C~50∫C (14∫F~122∫F)
Relative Humidity 75% @ 0∫C~30∫CÊ below;
50% @ 30∫C~40∫C:
Battery Type 9V NEDA 1604 or 6F22 or 006P.
Electromagnetic Compatibility l In a radio field of 1 V/m below:
Overall Accuracy = Specified Accuracy + 5% of Range
l In a radio field of 1 V/m above:
No assigned accuracy is specified.
Dimensions (H x W x L) 177 x 85 x 40mm.
Weight Approx.340g (including battery)
56
Model UT71A/B/C/D/E: OPERATING MANUAL
General Specifications
Range Auto
Polarity Auto
Overloading Display OL
(except at 4~20mA Loop range which display HI or LO)
Battery Deficiency Display
57
Model UT71A/B/C/D/E: OPERATING MANUAL
Feature Summary
Tri Digital Displays Primary: 40,000 counts
Analogue Bar Graph Left Secondary: 4000 counts
Right Secondary: 4000 counts
Bar Graph: 40 segments, updates 10 times / second
Backlight with 2 brightness levels Bright backlight for clear readings in poorly lighted areas.
Autorange The Meter automatically selects best range
AC+DC True RMS, AC RMS Choices for AC only or AC+DC readings
Data Hold Holds readings on display
Continuity Beeper sounds for resistance readings below threshold.
Bar Graph 40 segments
Duty Cycle Measure signal on or off time in %.
MAX MIN Mode Record maximum and minimum
Battery Access Door Battery replaceable.
58
Model UT71A/B/C/D/E: OPERATING MANUAL
Basic Specifications
Function Ranges / Description
DC Voltage 0 to 1000V
AC Voltage, True RMS 0 to 1000V, 100kHz bandwidth
Basic Accuracy DC Voltage: 0.025%
AC Voltage: 0.4%
DC Current 0 to 10A (5~10A for 10 seconds, interval 15 minutes)
AC Current, True RMS 0 to 10A (5~10A for 10 seconds, interval 15 minutes)
Resistance 0 to 40M (UT71A/B is 0 to 20M )
Capacitance 0 to 40mF(UT71A/B is 0 to 20mF)
Frequency 0~400MHz (UT71A/B is 0 to 200MHz)
Temperature -40∫C~1000∫C (-40∫F~1832∫F)(UT71B/C)
STORE Readings Up to 100 readings for UT71C and UT71E or 9999
readings for UT71D may be saved by the user in a
memory. These readings may be viewed by using Recall
feature.
59
Model UT71A/B/C/D/E: OPERATING MANUAL
A. DC Voltage
60
Model UT71A/B/C/D/E: OPERATING MANUAL
61
Model UT71A/B/C/D/E: OPERATING MANUAL
Remarks:
l Input Impedance: Approx 10M .
l Overload Protection: 1000V.
l Display:
a) True rms are valid from 10% of range to 100% of range
b) AC crest factor can be up to 3.0 except 1000V where it is 1.5.
c) A residual reading of 80 digits with test leads shorted, will not affect stated accuracy.
d) When frequency is lower than 100kHz, the accuracy guarantee range 10%-100%.
e) When making AC+DC measurment, the accuray need to add (1%+ 35 digits) of reading based on the above table.
62
Model UT71A/B/C/D/E: OPERATING MANUAL
C. DC Current
Range Accuracy
Resolution Overload Protection
UT71A/B UT71C/D/E UT71A UT71B UT71C/D/E
200µA 400µA 0.01µA 0.5A, 250V,
(0.1%+15)
2000µA 4000µA 0.1µA (0.2%+20) (0.15%+20) fast type fuse,
20mA 40mA 0.001mA (0.15%+15) 5◊20mm
200mA 400mA 0.01mA
10A 10A 0.001A (0.8%+30) (0.7%+30) (0.5%+30) 10A, 250V, fast type fuse,
5◊20mm
Remarks:
At 10A range:
l When the measured current is 5A, continuous measurement is allowed.
l When the measured current is between 5A-10A, continuous measurement 10 seconds and interval more than
15 minutes.
63
Model UT71A/B/C/D/E: OPERATING MANUAL
E. Resistance
Range Accuracy Overload
Resolution
UT71A/B UT71C/D/E UT71A UT71B UT71C/D/E Protection
200 400 0.01 (0.5%+20)+test leads (0.4%+20)+test leads (0.3%+8)+test leads
short circuit value short circuit value short circuit value
2k 4k 0.0001k 1000V
(0.5%+20) (0.4%+20) (0.3%+8)
20k 40k 0.001k
200k 400k 0.01k (1%+20) (0.8%+20) (0.5%+20)
2M 4M 0.0001M (1%+40) (1%+40) (1%+40)
20M 40M 0.001M (1.5%+40) (1.5%+40) (1.5%+40)
F. Continuity Test
Range Resolution Overload Protection
0.01 1000V
Remarks:
l Open circuit voltage approximate -1.2V.
l The buzzer does not sound when the test resistance is > 60 .
l The beeper comes on continuously for open conditions, that is test resistance is 40 .
65
Model UT71A/B/C/D/E: OPERATING MANUAL
G. Diode Test
Range Resolution Overload Protection
0.0001V 1000V
Remarks:
l Open circuit voltage approximate 2.8V.
l A good silicon junction drops between 0.5V and 0.8V.
H. Capacitance
Range Accuracy Overload
Resolution
UT71A/B UT71C/D/E UT71A UT71B UT71C/D/E Protection
20nF 40nF 0.001nF (1.5%+20)+ (1.2%+20)+ (1%+20)+
capacitance value of capacitance value of capacitance value of
open circuit test leads open circuit test leads open circuit test leads
200nF 400nF 0.01nF
(1.5%+20) (1.2%+20) (1%+20)
2µF 4µF 0.0001µF 1000V
20µF 40µF 0.001µF (1.2%+40)
(1.5%+40)
200µF 400µF 0.01µF (1.5%+40) (1.2%+20)
2mF 4mF 0.0001mF (5%+40) (5%+40) (5%+20)
20mF 40mF 0.001 mF Not specified Not specified Not specified
66
Model UT71A/B/C/D/E: OPERATING MANUAL
I. Frequency
Range Accuracy
Resolution Overload Protection
UT71A/B UT71C/D/E UT71A/B UT71C/D/E
20Hz 40Hz 0.001Hz
200Hz 400Hz 0.01Hz
2kHz 4kHz 0.0001kHz
20kHz 40kHz 0.001kHz (0.1%+15) (0.01%+8) 1000V
200kHz 400kHz 0.01kHz
2MHz 4MHz 0.0001MHz
20MHz 40MHz 0.001MHz
200MHz 400MHz 0.01MHz Not Specified Not Specified
Remarks:
l Input amplitude ìaî as follows; (DC electric level is zero)
When 10Hz~40MHz : 200mV a 30Vrms;
When > 40MHz : Not specified
67
Model UT71A/B/C/D/E: OPERATING MANUAL
J. Temperature (UT71B/C/D/E)
Degrees Celsius
Range Resolution Accuracy Overload Protection
-40∫C~40∫C (3%+30)
40∫C~400∫C 0.1∫C (1%+30) 1000V
400∫C~1000∫C 2.5%
Fahrenheit
Range Resolution Accuracy Overload Protection
-40∫F~32∫F (4%+50)
32∫F~752∫F 0.1∫F (1.5%+50) 1000V
752∫F~1832∫F 3%
Remarks:
l Included is a K-Type (nickel chromium ~ nickel silicon) point contact temperature probe which could only measure
temperature below 230∫C. If you want to measure temperature higher than 230∫C, you must use the rod contact
temperature probe.
68
Model UT71A/B/C/D/E: OPERATING MANUAL
69
Model UT71A/B/C/D/E: OPERATING MANUAL
Manufacturer:
Uni-Trend Technology (Dongguan) Limited
Dong Fang Da Dao
Bei Shan Dong Fang Industrial Development District
Hu Men Town, Dongguan City