Gasman II Instruction Manual

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 19

Gasman II Instruction Manual

Gasman II Flammable Rechargeable Batteries


Upon receipt we recommend that Flammable instruments are fully charged for 16 hours
prior to use.
This product has been tested and found to comply with the European Directive
89/3361EEC
WARNING: Substitution of components may impair Intrinsic Safety.
Issue 5 Page 2 11/99
CROWCON

Content s Page
1. Quick Operating Guide 4
2. Unpacking 4
3. Introduction
4. Specification 6
4.1 Instrument Settings g
4.2 Battery Lifetime g
4.3 General Assembly 10
5. Operation 11
5.1 Switch-on sequence 11
5.2 Display Symbol Guide 12
6. Batteries 15
6.1 Recharging 15
6.2 Changing the batteries 16
6.3 Charger Supplies 16
7. Alarm Types 16
8. Keypad Adjustments 18
8.1 Viewing settings 18
8.2 Adjusting settings 19
8.3 Zeroing 2p
8.4 Calibration 20
9. Accessories 21
9.1 Calibration Adaptor 21
9.2 Universal Clip 21
9.3 Rubber Boot 22
10. Routine Checking 22
10.1 General 22
10.2 Sensor filter 22
10.3 Zero & Calibration 23
10.4 Sensor Replacement 23
11. Troubleshooting Guide 24
11.1 Error Codes 25
12. Datalogging and Datacomms 26
12.1 Hardware Interface 27
13. Sensor Limitations 27
14. Accessories and Spare Parts 28
14.1 Accessories List 28
14.2 Spare Parts List 29

Issue 5 Page 3 11 /99


Gasman II Instruction Manual
1. Quick Operating Guide Switch on: Press large button; alarm test and LCD test occurs,
followed by battery level display.
Normal condition: Gas level displayed on LCD*. Intermittent 'confidence' signal
(if selected).
Alarm condition: Fast flashing red LEDs and sounder.
Alarm reset: Press large button when hazard has passed. Activate backlight: Press and
hold large button.
Switch Instrument Press both buttons Off: together & release. Zero instrument: When unit
is off, press small button & hold. The symbol will appear. Keeping the small button
pushed in, press the large button.
For oxygen models, the symbol appears, and pressing the large button sets the reading to
20.9%.
* reading may take a few seconds to settle after switch-on.
2. Unpacking
Check the shipment against the packing list and carton labels. Instruments and
accessories may be packed separately. Smaller
items may be located under cardboard partitions, so check carefully for contents. If you
are missing any items please contact Crowcon immediately. Be prepared to quote the
instrument's serial number which can be found at the bottom of the label on the rear of the
instrument.
Ensure batteries are fresh/charged prior to use. See 'Switch on Sequence' (Section 5.1) for
details of the battery level indicator.
3. Introduction
The Crowcon Gasman II is a personal gas detector which can continuously monitor the
level of oxygen or the presence of a single toxic or flammable gas. It is designed to be
worn by individuals working in hazardous environments such as confined spaces and will
give aloud audible and bright visual alarm warning when preset concentrations of gas are
exceeded.
It is a simple instrument to use, designed to provide reliable service for many years. The
shape of the instrument has been designed to fit comfortably against the body and pose the
least restriction to movement. It can fit in a pocket or on a belt.
Datalogging is standard on all models providing approximately 1 week capacity at a 1
minute sampling rate. This capacity can be extended by reducing the sampling rate (see
Section 12). Toxic and oxygen versions are fitted with alkaline batteries which offer a life
of 1 year*. Flammable versions are fitted with re-chargeable nickel metal hydride
batteries which have a life of 10-12 hours* and up to 1000 charge/discharge cycles.
* depending on alarm incidence.
4. Specification
Weight 210g (7.4oz)
Dimensions 130h X 60w X 30d mm
(5.1h X 2.4w X 1.2d inches;
Typical measuring 0-100% LEL flammable gas,
ranges 0-25% volume O2,
0-50ppm H2S, NH3,
0-500ppm CO,
0-5 ppm Cl2, PH3,
0-lOppm SO2, NO2,
0-25ppm HCN,
0-999PPM H2.
Typical alarm 20 & 40% LEL methane,
levels 19% and 23% oxygen,
10 & 20ppm
hydrogen sulphide,
50 & 100ppm
carbon monoxide.
Typical TWA alarm levels for toxic 15ppm STEL, 10ppm
gas version LTEL (H2S),
300ppm STEL, 50ppm
LTEL (CO).
Typical response times (T) methane 20 seconds,
oxygen 10 seconds,
toxic gas 20 seconds
Visual alarm indicators Ultra bright flashing red LED pair in alarm.

Audible alarms Programmable from five different tones,


85dBA at 1 metre. (92dBA at 1 ft). Intermittent
blip for discharged battery. Selectable
'confidence' blip. Fault warning.
Display Custom 7 segment high contrast LCD giving
gas reading. Symbols for low battery, peak
hold, zero mode, cal mode and alarm
annunciation. LED backlight.
Operating temperature range -20° to 50°C (-4°F to 122°F)
Humidity 0-95% RH, non-condensing.
Expected sensor life Pellistor 5 years, toxic gas 3 years, oxygen 1
year.
Batteries 1.1Ah nickel metal hydride rechargeable: 10-12
hours between recharge. AA size alkaline
replaceable: Typical life 1 year.
Aspirator Manual bulb type with 2m of hose, non-
absorbent, hydrocarbon resistant.
Approval code (Europe) Toxic/oxygen: EEx ia IIC T4
(Tamb=55°C)
Flammable:
EEx iad IIC T3 (rechargeable), T4 (non
rechargeable)

(USA & Canada) Class 1, Division 1, Groups A


B,C & D.
Standards EN50014, EN50020,
(Europe) EN50018
(USA & Canada) UL913
RF Immunity Tested to IEC801-3 and meets EN50082-2
RF Radiation Tested to EN55022 Class B
and meets EN50081-2
Battery charging Standard charge rate 16 hours, trickle charge
will not
damage batteries. Charger
Input 11-29V DC. Mains
adaptors have 110 or 230V
AC inputs. Batteries must be
charged in an ambient
temperature between 0°C
and 40°C.

4.1 Instrument Settings


Unless otherwise specified, instruments will be shipped from the factory with default
settings. These cover alarm tones, logging period, and detailed operation of each alarm
point. Full details are printed on the Configuration Report supplied with each instrument.
Those settings which are changeable from the instrument's keypad are described in section
8.
All configuration settings may be viewed and changed using a PC running the
SetPortable interface program (Section 12).
4.2 Battery Lifetime
 Rechargeable flammable instruments' batteries last 12 hours on a full charge.
 Non-rechargeable flammable instruments using alkaline batteries last 24 hours.
 Toxic and Oxygen instruments last 1 year using alkaline batteries, under normal
usage conditions.

4.3 General Assembly


Key:
1 Gas Sensor
2 Fixing screw/charging
contact
3 Small button
4 Calibration adjuster
cover
5 Alarm LED lens
6 Alarm sounder
7 Top moulding
8 Gas legend
9 Display window
10 Large button
11 Base moulding
12 Universal clip
13 Certification labels

5. Operation 5.1 Switch-on Sequence


Make sure you are in clean air, then switch on the instrument by pressing the large
button.
The instrument tests all LCD segments, red alarm LEDs and sounder for 5 seconds, then
displays the battery condition for 5 seconds before entering monitoring mode. The
battery condition is indicated by the battery symbol and a number representing the
percentage battery capacity remaining.

NOTE: A reading below 20 indicates that replacement or recharging is due. For


the first 4-5 charge cycles, rechargeable batteries cannot hold a full charge. Initial
operation time is therefore around 7 1/2 hours.
The instrument is now ready to be used:
• The gas level will be shown numerically on the display.
• The instrument may be configured to display nothing (alarm only application).
• The display backlight is activated by pressing and holding the large button.
• If the display shows an E code (e.g. E2) instead of the gas level, this indicates a fault
condition as defined in Section 11.1.
Figure 5.1 Typical monitoring display.
Gasman II Instruction Manual
NOTE: It is possible that the instrument may display an alarm condition upon entering
the monitoring mode (indicated by a flashing alarm symbol). This is due to the sensor
taking time to settle after switch on. It is NOT a fault. Press the large button to cancel the
alarm after a few seconds.

If configured to do so, the instrument will give an intermittent audible blip and/or visual
red flash to indicate correct functioning of the instrument. Selection of these 'confidence'
signals will affect battery life according to the table in Section 4.2. Pressing both buttons
together will switch the instrument off. The instrument may be configured in such a way
that it is not possible to turn it off via the push buttons. In this case, only a supervisor with
a computer link can switch the instrument off
5.2 Display Symbol Guide

Figure 5.2 Display symbols


The gas level is continuously shown on the display. The surrounding symbols will only
be displayed when the appropriate function or warning is active.

Peak hold
Indicates that the reading displayed is the highest level recorded since the instrument
was switched on or reset. For oxygen, Peak Hold records the lowest value encountered.
Peak Hold values can be toggled on and off by pressing the small button when the unit
is on. The symbol will come on if Peak Hold is operational. The large button is used to
reset the peak value when the symbol is displayed.

Battery low
When this symbol flashes and the sounder emits an intermittent blip, it is an indication
that there is less than 10% of battery life remaining. A
toxic/oxygen instrument may run for weeks in this condition allowing time to schedule a
battery replacement service. A flammable instrument will run for 'h hour at most and
should therefore be recharged immediately.

Zero
Indicates unit is in adjust zero mode. See Section 8.3.

Calibrate
Indicates unit is in adjust calibration mode. See Section 8.4.

Alarm 1

When the first alarm level is reached, this symbol will flash and the visual and audible
alarms will be activated. See Section 7, Alarm types.

Alarm 2
When the second alarm level is reached, this symbol will flash and the visual and audible
alarms will be activated. See Section 7, Alarm types.

15 Minute TWA Alarm (STEL)


Toxic versions have TWA alarms in addition to the two levels of instantaneous alarm. If
the 15 minute exposure level is reached, this symbol will flash and the visual and audible
alarms will be activated. The minimum averaging time before a STEL alarm can be
triggered is 1 minute.

8 hour TWA Alarm (LTEL)


If the 8 hour exposure level is reached, this symbol will flash and the visual and audible
alarms will be activated. The minimum averaging time before a LTEL alarm can be
triggered is 15 minutes.
Personnel should not continue to work after a TWA alarm occurs as this indicates the
maximum safe cumulative exposure has been reached. Workers should remove
themselves to safety or use breathing apparatus.
6. Batteries
WARNING - Battery charging or replacement should only take place in a SAFE
AREA, i.e. no explosive risk.
WARNING - to ensure Intrinsic Safety, the charger must only be used with DC
supplies as specified in Section 6.3.

6.1 Recharging
Flammable gas versions are normally supplied with rechargeable batteries, though a non-
rechargeable type using alkaline batteries is available.
To recharge, connect power to the charger and place the instrument in the cradle. The
charger's red LED will light while the instrument is being charged. Full recharge from flat
takes 15 hours, with the instrument switched off. When fully charged, the green LED will
illuminate on the charger. Once the charge cycle is complete, the red LED will
occassionally light for a minute or so to ensure the batteries are kept topped up.
If the instrument is not in use, it should be charged once per fortnight, or it can be stored
permanently in its charger without damage. If it is left uncharged for 3 weeks or more, the
batteries may self-discharge to a point where the battery charger cannot supply enough
charge current to restore the batteries to a fully charged state. If this occurs, a "boost-
charge" can be provided by pressing the button which is mounted on the rear of the
charger. This can be accessed through the top-left mounting hole with a blunt object such
as a ball-point pen.
The charger also acts as a comms interface for all instrument types. Non-rechargeable
units will not be recharged in this instance.
The charger makes contact with the instrument via the two side fixing screws. As these
are electronically protected, no harm will come from these contacts being short circuited
during use.
6.2 Changing the batteries
Toxic gas and oxygen versions are supplied with alkaline batteries which should power
the instrument for a year.
• Disassemble in a clean, dry environment and observe anti-static handling precautions.
• Remove side fixing screws
• Remove calibration adjuster cover.
• Slide lower case section from top section.
• Remove batteries from clips.
• Replace batteries using only approved types* ensuring correct polarity orientation.
• Refit lower case section and screws ensuring seals are correctly fitted in grooves.
*Duracell MN1500 LR6
Ever Ready Energizer LR6BA

6.3 Charger Supplies


These must not exceed 29 volts DC (11 to 16V is ideal). They can be either:
(a) a mains DC power supply such as those in Section 14.2, "Spare Parts" meeting
EN60742 insulation class II (Europe) or UL1310 insulation class II (United States) or
(b) a battery or other isolated DC source.

7. Alarm Types
There are two instantaneous alarms; level 1 and 2. For toxic gas instruments, there are
also two time weighted average (TWA) alarms which operate at short term (15 minute)
and long term (8 hour) exposure limits. All alarm thresholds can
be adjusted from the instrument keypad (see Section 8) by means of a screwdriver. Alarm
thresholds may also be set via the computer interface (see Section 12) and the following
parameters may be set for instantaneous alarms: • Normally triggered on a rising level,
alarms
may be set to trigger on a falling level as with oxygen deficiency alarms.
• Alarms may be latching, requiring manual reset, or self clearing when the hazard has
passed.
• Normal alarms cannot be cleared whilst the gas hazard persists. It is possible to set level
1 so it may be silenced (the red LEDs will continue to flash). This allows the user to set a
'pre-alarm' warning at a low gas level and a non-silenceable alarm at a higher level.
• Unused alarms may be deactivated.
• The level 1 alarm may be set to be visual only (not oxygen units).
If any alarm event is triggered, the red LEDs will flash and the sounder will emit aloud
tone. Each alarm type has a different tone sequence which can be selected from a menu if
using
SetPortable software (Section 12). Alarm 2's tone takes precedence over alarm 1's tone.
The alarm type will be indicated by a flashing symbol on the LCD (see Section 5.2). If
either of the TWA (time weighted average) symbols appear on the
display, this indicates that the cumulative exposure to toxic gas has been reached for an 8-
hour shift or a 15 minute 'window'. UK Health & Safety regulations require that workers
should leave the hazardous area once they have exceeded this dose. A TWA alarm cannot
be
Gasman ll Instruction Manual
manually reset. The 15 minute TWA alarm will self clear when the average exposure
over the previous 15 minutes falls below the set threshold. The 8 hour TWA alarm will
self clear when the average exposure over the previous 8 hours falls below the set
threshold. Average exposure values are cleared when the instrument is switched off. A
worker should not resume work where the risk of gas exposure persists until the Safety
Officer declares it safe.
The large button may be used to silence or clear certain instantaneous alarms as the
configuration permits.
8. Keypad Adjustments
When the instrument is off, calibration and alarm settings can be viewed and adjusted if
required. 8.1 Viewing settings

• Press both buttons to turn instrument off.


• Press and hold small button to view the zero level. The "lenyíl" symbol appears. The
numbers will flash until the reading has stabilised. The reading cannot be adjusted
until the digits stop flashing.
Note. The zeroing facility is not available on oxygen units.
• Release small button then press and hold again to view the calibration level. The
"ÉK" symbol appears. Unless calibration gas
is being applied to the sensor, the display should read zero, or 20.9 for oxygen.
• Release small button then press and hold again to view Alarm 1. The "csengô1"
symbol appears and the Alarm 1 level is displayed.
• Release small button then press and hold again to view Alarm 2. The "csengô2"
symbol appears and the Alarm 2 level is displayed.
• Release small button, then press and hold again to view the 15 minute STEL. The p
symbol appears and the 15 minute STEL alarm level is displayed.
• Release small button then press and hold again to view the 8 hour TWA. The
symbol appears and the 8 hour TWA alarm level is displayed.
• Release small button, then press and hold again to bring the display back to zero
mode. If the small button is released and not pressed again for 1 minute or more the
sequence automatically starts back to zero.
8.2 Adjusting Settings
• Rotate the calibration adjuster cover to line up the mark with the mark on the case.
• Unplug the cover. It is held captive and should not be fully removed.
• Insert the small screwdriver supplied into the adjustment potentiometer.
• Select the setting to be adjusted with the small button as described in Section 8.1. The
current value is displayed and can be
adjusted by rotating the screwdriver.
• Only one potentiometer covers all functions adjustments are stored by the instrument.
• Potentiometer is only active while small button is held down.
• If the potentiometer is turned to one of its end stops, the reading continues to advance
in the relevant direction until backed off. It is
advisable to set the pot back to its central position after releasing the small button in
readiness for the next adjustment.
8.3 Zeroing
The reading cannot be adjusted until the digits stop flashing. This is to allow the sensor
to stabilise after energising. A 'short-cuY zero is provided by pressing the large button
whilst the small button is held down in zero mode. Only perform this operation in fresh
air.
Note: The zeroing facility is not provided on oxygen units as the reading should be
calibrated to 20.9% in fresh air.

8.4 Calibration
As with zeroing, the reading will flash until the sensor has stabilised. Additionally, time
must be allowed for the reading to level off once calibration gas has been applied. The
final reading can then be adjusted. A 'short-cut' calibration to 20.9% on oxygen units is
provided by pressing the large button whilst the small button is held down in calibrate
mode.
Gas connection for calibration requires the use of the Calibration Adaptor (Section 9.1).
The sample tube should be connected to a gas bag filled with calibration gas. Use the
hand bulb to draw the gas over the sensor. If using a cylinder of calibration gas, connect
it via a pressure regulator and flowmeter to the adaptor inlet. Disconnect the hand bulb
and allow waste gas to vent to atmosphere or a collection bag. A flowrate of 0.5 - 1 I/min
is required for an accurate calibration. Gas readings can be adjusted via the keypad
(Section 8.4) or with the SetPortable computer interface (Section 12).

9. Accessories

9.1 Calibration Adaptor


The calibration adaptor is a push-fit device for connecting calibration gas or a remote gas
sample to the sensor. It is supplied with a 2m length of sample tubing and a rubber hand
bulb. The sample tube can be lowered into a manhole, for example, to enable pre-entry
checks to be made. The hand bulb should be squeezed in a consistent manner (every 2-3
seconds) for at least a minute to draw a reliable sample up to the sensor.
9.2 Universal Clip
The combined belt and pocket clip can be used with a trouser belt or clipped onto a
pocket. The
instrument may be worn inside the pocket, providing the sensor protrudes. A shoulder
strap is available for fixing to the top of this clip.
9.3 Rubber Boot
This is a protective cover for use in the most arduous environments. It can remain in
place for battery charging.

10. Routine checking 8 Maintenance Gasman II is designed to operate with little


maintenance under most conditions. The following routine maintenance is
recommended.
WARNING - Substitution of components may impair intrinsic safety.

10.1 General
If used in dirty conditions, a regular wipe with a damp cloth will help in viewing the
display and maintaining operation of the pushbuttons. This is also recommended for the
side fixing screws to maintain good contact for battery charging or data communication.
10.2 Sensor Filter
Dirt should be removed from the sensor apertures as this may impede diffusion. For
access to the sensor filter, remove the fixing screws and top moulding (Section 10.4).
10.3 Zero and Calibration
Because all gas sensors change their performance characteristics over time, the
following checks are recommended:
• Zero Every 1 month • Gas response' Every 1 month • Calibration Every 6 months
Instructions on how to make zero and calibration
adjustments from the instrument keypad are in Section 8. Refer to the manual supplied
with SetPortable software for use of a PC to make adjustments (see Section 12).
*Due to the safety nature of gas detection instruments, it is recommended that
sensors are regularly gas checked. Instrument sensors may be adversely affected by
certain environments (Section 13). A monthly test with a known gas mixture will
ensure that the instrument responds with appropriate indications and alarms.
10.4 Sensor Replacement
Required when instrument fails to calibrate due to low sensor output.

Key
:
1 Top moulding
2 Fixing screw
3 Gas sensor
4 Sealing ring
5 Sounder `O'
ring
6 Main `O' rings

Figure 10.1
On toxic and oxygen instruments, it is practical to replace the alkaline batteries at the
same time (Section 6.2).

• Disassemble in a clean, dry environment and observe anti-static handling precautions.


• Switch off instrument
• Remove side fixing screws • Remove top moulding
• Unplug sensor
• Replace sensor with same type only • Re-use sensor sealing ring
• Ensure main body 'O' rings and sounder 'O' ring are in place before refitting top
moulding. • Secure with screws
• Zero and calibrate instrument
11. Troubleshooting Guide Instrument may be customer configured such that:
display is blanked
• no 'confidence' signals given • cannot switch off
• adjustment potentiometer not active
These features may be individually re-enabled if required (Section 12).

If the instrument is reading low levels of gas in fresh air, it needs to be zeroed (Section
8.3). Replace or recharge batteries if unit will not switch on. If rechargeable instruments
do not seem to be charging, force some charge into
them by using the hidden button at the back of the charger, then reset the instrument by
removing and re-inserting a battery. (If rechargeable batteries are left uncharged for
many weeks, they gradually self-discharge until the normal charging methods do not
work).

11.1 Error codes


The instrument may display the following warning codes:
EO : Calibration due: FATAL (accept by pressing large button). Do not ignore because in
4 weeks it will become...
E01: Calibration Expired: FATAL. The period between regular calibrations has
elapsed. Re-zero and calibrate (Section 8).
E02: Zero Failed: FATAL. Results from attempting to Zero a channel exposed to gas
(including an oxygen channel in air). The function request is aborted without effect.
Re-zero (Section 8).
E03: Calibration Failed: WARNING. The gain is being set out of range indicating
sensor failure or incorrect calibration gas.
Re-zero and calibrate (Section 8).
Change sensor if necessary (Section 10.4) E04: Sensor Overrange: Warning will
clear when restored to fresh air. Gas response should be re-checked.
E05: High/low Temperature: Gas readings may not be reliable at these extremes.
Warning will clear when temperature is in range.
E06: Electronic Hardware Failure: FATAL. The instrument should be sent for repair.
E07: Configuration Invalid: FATAL. Down load
original configuration (see Section 12) or return for service.
If a FATAL error is rectified, a lower priority error code may then be displayed.
E0, E01, E02, E03 are normally non fatal but may be configured as fatal using
SetPortable.
An instrument with a FATAL error may not be used but may only be switched off or
rectified. The instrument will raise an audible alarm in the event of a fault.
12. Datalogging and Datacommunications Gasman II has an integral datalogging
feature which takes regular readings and stores them in memory. When the memory is
full, oldest data is overwritten so the instrument will always contain the most recently
accumulated data.
As standard, a reading is taken every minute which will give a minimum logging capacity
of
30 hours. If gas readings are zero or unchanging, this capacity is extended by compression
software to as much as 3000 hours. The basic capacity can be altered by changing the
sample rate.
Note: If batteries fully discharge, logged data may be lost.
To change sample rate, upload the log, clear
the log or alter any other instrument settings, it is necessary to use a PC running
SetPortable software. To view the log, use Logmanager. Historical time is calculated
relative to current PC time, so ensure the date and time are set
correctly on your computer before uploading logged data.
SetPortable and Logmanager software is supplied with its own manual, Crowcon
stock number M07-119.

12.1 Hardware Interface


Connection to the PC is via the charger unit which has a computer 'D' socket at the rear.
• Use the Crowcon supplied cable to connect between the charger and the PC serial
port. Check the labelling of this lead as it only works one way round.
• Ensure power is connected to the charger unit.
• The instrument should be switched on if it is to be zeroed or calibrated.
13. Sensor Limitations
Gasman II uses a catalytic flammable gas sensor which operates in the presence of
oxygen. Prior to entering a confined space, it is advisable to check oxygen
concentration in addition to combustible gas concentration. Oxygen levels below 10%
will reduce a flammable gas reading.
The performance of catalytic sensors may be permanently degraded if exposed to
silicones, sulphur containing gases (such as H2S), lead or chlorine compounds
(including chlorinated hydrocarbons).
The unit is not suitable for use in ambient temperatures above 50°C (122°F) and electro
chemical toxic gas sensors may be degraded at these levels.
Water should not be allowed to collect on the sensor as this may impede gas diffusion.
Use with care in wet or humid environments, and check response after use.
Persistent exposure to high levels of toxic gas will shorten the life of the sensor. If the gas
is corrosive (e.g. hydrogen sulphide) damage may occur to metal components.
Sensors may be cross sensitive to other gases. If unsure, contact Crowcon or your local
agent.
Use of high power radio transmitters in close proximity to the instrument may exceed
rated RFI immunity levels and cause erroneous indications. If such problems are
experienced, remove antennae to a reasonable separation from the instrument (e.g. 30cm).
14. Accessories & Spare Parts 14.1 Accessories List
C01613 Gasman II Charger 12V DC input C01628 Gasman II Charger with 230V UK
power supply
C01629 Gasman II Charger with 230V European power supply
C01630 Gasman II Charger with 110V US power supply
C01631 Gasman II Charger with 230V power supply - no plug
C01632 Gasman II Charger with 110V power supply - no plug
C0164 Multi-charger ribbon cable
7
C0132 Comms lead
7
C0131 SetPortable software 3 1/2"
5 diskette
C0129 Vehicle charging lead
6
C0139 Shoulder strap
4
C0145 Waist belt
1
C0161 Aspirator complete with flow
4 adaptor,tubing, bulb & pouch

C0109 Aspirator probe


7
C0151 Alarm earphone with 1.5m of
1 tubing
C0155 Autocalibrator II (requires
2 PC &
charger)

14.2. Spare Parts List


C0100 2.5mm Allen key
9
C0134 2mm Allen key
7
DE01 Rechargeable battery (3
02 required) -
Flammable instrument
E0145 AA size alkaline battery, 3
1 required -
toxic & oxygen
instruments
DE01 Oxygen sensor
11
DE01 Carbon monoxide sensor
12
DE01 Hydrogen sulphide sensor
13
E0163 Flammable gas sensor
1
M044 Moulding set
27
M044 Universal pocket/belt Gip
28
M015 Pocket clip spring
58
M036 Fixing screw - (2 per
03 instrument)
M044 Sensor sealing ring
29
M044 Calibration cover (requires
51 'O' ring)
M043 Calibration cover 'O' ring
37
M044 Main 'O' ring (2 per inst.)
30
M044 Sounder 'O' ring
61
M044 Rubber protective boot
37
S0170 Oxygen pcb assembly
8
S0170 Toxic pcb assembly
7
S0170 Flammable pcb assembly
9
S0171 Display pcb assembly
0
SERVICE RECORD
Ser.No: Location:
Date Cal. Zero Span Comme
Due Date nts

Crowcon may also be relied upon for


• Portable multi-gas monitors for simultaneous detection of up to four gases.
• Fixed equipment for continuous plant monitoring.
• Calibration and repair at our instrument workshops.
• On site service and maintenance.

SetPortable page20
Like Alarm 1 and Alarm 2 above, the noise generated when the alarm is activated can be selected. There
is also the option to try the sound on the computer.
Range and Units These are factory settings used for labelling data In LogManager and setting calibration
Iimits. They correspond to the sensor and gas label fitted to the Instrument. Changes should only be made
if a sensor type is changed. Such an operation may only be carried out by a Crowcon approved service
centre.
Click Finished when alarm configuration has been completed
5.3.2 Edit/Optlons
This screen allows SetPortable to be customized. It controls what operations are possible. This is useful,
for example, to allow a site office to upload logged data, whilst preventing them from accidentally
changing instrument configurations.
The options available are self explanatory, and no further description will be given. (See Fig 5.7)
6.4 Adjust Menu
This menu consists of the following options:
Zero A dialog box (s displayed allowing you zero the gas sensor This operation should only be
performed when the instrument is in clean air. See Fig 5.9.
A confirmation message will report the outcome of the operation.
Callbrate This allows the instrument's span to be calibrated. See Fig 5.10. Type fn the calibration gas
value; this should be given on the gas cylinder label. When the gas has been flowing for at least 30
seconds, and the reading has stabiIized, click OK. A confirmation message will report the outcome of the
operation.
Note: Refer to the instrument's manual for more Information on the Zero and Calibrate functions. Clear
Log A warning box Is displayed before the data log is cleared. Acknowledge that the data log is to be
cleared by clicking on Yes. Cal Due This allows you to override the calibration due date in the
Instrument. Enter the number of days before you want the cal due
warning to appear on the instrument, Only use this facility if you know the instrument responds correctly
to gas, or if you employ other systems to ensure that instruments are regularly calibrated. See Fig 5.11.
8 CUSTODIAN
6.1 Files menu Fig 6.1 This menu allows data to be saved and restored from the computer's disk, and
printing out of instrument configurations.
New This allows a new setup file to be created. To do this select New, then Edit from the main menu
bar. When editing is complete, select Save As... from the files menu, and type in a file name, making sure
the .set file extension is saved.
Open... This allows you to open an existing file already saved on the PC. This can be edited, or copied to
another instrument.
Save As... This allows you to save edited files with a different name or to a different drive or directory.
Print This produces a hard copy record
of the setup Information created in Edit (this will be similar to the Configuration Report originally
supplied with the instrument). This will be printed on the default printer connected to the PC.
Quality Print If a high quality font (similar to the original Configuration Report) is required for a better
presentation, click on Quality Print before selecting
Print. Whilst Quality Print is selected there will be a tick (pipa ) next to the quality Print
option.
B.2 Datalink Menu Fig 6.2 The Datalink menu manages communications between the computer and
instrument.
Upload Transfers setup data from the Instrument to the PC. This data can then be edited to modify the
way in which the Instrument operates.
Download This allows edited data to be transferred from the PC to the instrument.
Read Gas This produces a "snap shot" of the Instrument gas readings This can be used, for example, to
check that the sensors have returned to zero after a calibration.
Show Values This is like a continuous Read Gas option. Select Read Gas to stop the continuous update
of the display.
Next Cal This indicates the number of days left before the next calibration is due. After calibration, the
instrument will count down in days from a pre-set time (normally 26 weeks) and when the time has
elapsed the instrument will display the Calibration Due warning during the switch on routine.

Upload Log This transfers logged data stored In the instrument to the PC for viewing In LogManager
(see Section e). The Save File screen prompts you to name the log file. Either create a name by typing it
into the File Name box (replacing the asterisk) or by selecting an existing file (this will overwrite the old
tile). Click OK to proceed. The log should be saved with the file name extension of .log.
B.3 Edit Menu
The edit menu allows the instrument configuration (which has been loaded Into SetPortable either via the
Files or Datalink menu) to be modified, and for SetPortable options to be set.
Config This allows instrument configuration to be modified. See below and Fig 6.4.
Optlons This screen allows SetPortable to be customized. See below and Fig 6.8.
6.3.1 Edit/Config
This brings up the SetParametere dialog (Fig 6.4). In brief,
the check boxes and drop down menus select the following options. Consult the instrument manual for
details.
Alarm Sound
A menu of alarm tones is available for each alarm. A "silent" alarm (visual only 'pre alarm') is available
only on level 1 alarms for safety reasons. Click Test for a computer simulation of the sound selected for
level 1.
Display mode
Select Full or No display. Selecting No Display will cause the LCD to remain blank at all times. Note: if
No Display is selected It is advised that the Confidence blip Is selected on: With both these options
selected as Off the user has only the
flashing green LED, under normal conditions, to indicate that the instrument is operating.
Password 1 and 2 These are used to prevent unauthorized adjustments being made to the instrument
(refer to the instrument manual for details on entering passwords). The adjustments which may be
protected are zero, calibrate, alarm levels and instrument switch off. Check the boxes for which the
relevant password is required.
The default passwords are:
To change the
password, click
Note that a single button can be clicked on more than once.

Logging Period This sets the sample rate at which the instrument stores logged data. The log operates on
a rolling basis, so that when ft becomes full the oldest data in the log will be deleted and replaced by the
current data. See the instrument manual for further details on data logging.

on Password 1 or Password 2, as required. This will bring up a picture of the instrument (Fig 6.6). Click
on four buttons in the password sequence required.
Next calibrate This defines the period between an instrument warning, and the generation of the
Calibration Expired warning (normally 26 weeks). Changing the value in this box will have no effect
until the instrument Is calibrated.
Functions timeout When this option is selected the instrument Is locked In Run mode 2 minutes after
switch on. This is to prevent accidental adjustments being made during normal operation.
Confidence blip This switches the audible confidence blip off and on.
Display alarm When selected, this causes the Instrument to lock its display on the first channel to go Into
alarm.
To configure the instrument channel, click on the Channels button. Once all configuration has been
completed, click on the Finish button.
When Channels is clicked on, the dialog in Fig 6.7 appears. Select the channel to be viewed or edited by
clicking on the channel tab at the top of the screen.
For each channel, the following options are available:
Gas, Range and Units These are selected from pull down menus. Note: These are factory settings used
for labeling data in LogManager and setting calibration limits. They correspond to the sensor and upper
label fitted to the Instrument. Changes should only be made if a sensor type is changed. Such an
operation may only be carried out by a Crowcon approved service center.
Time Weighted Averages The TWA levels only apply to toxic gas channels. The Short term level is the
15 minute cumulative exposure level and the Long term level is the 8+ hour exposure level. These alarm
levels are factory set to reflect the Health and Safety requirements of the country to which the instrument
Is delivered.
Alarm 1, 2 and 3 These determine the operating modes of each of the instantaneous gas alarms
available for each channel. Each alarm has several options which may be toggled off/on and an alarm
value:
Active This determines whether the alarm is active (i.e. is in use) or not. Rising alarm Always selected
On, except for oxygen deficiency alarms, which are triggered on falling gas concentrations.
Latching When this option is selected R requires the alarm to be manually reset, by pressing the
Main button, once the alarm condition has passed.
Can mute When this is elected an audible alarm to be silenced whilst the alarm condition is still present.
Value This sets the gas concentration at which the alarm will trip. Note: The number entered In the
Value box should be between 5% and 100% of the number in the Range box and can be entered to one
decimal place. (e.g. If the Range is 0-50,
the Value can be between 2.5 and 50.0).
It is not advisable to set alarm Values higher than 60.0 for flammable channels. Only Alarm 1 and Alarm
2 values can be accessed from the instrument front panel. For oxygen units, the enrichment alarm (rising)
is therefore set as Alarm 2.
Click on the Finish button when sensor channel configuration has been completed. 6.3.2 Edit/Options
This screen allows SetPortabie to be customized. It controls what operations are possible. This is useful,
for example, to allow a site office to upload logged data
whilst preventing them from accidentally changing instrument configurations. The options available are
self explanatory, and no further description will be given. See Fig 6.8
6.4 Adjust Menu
This menu consists of the following options: Zero A dialog box Is displayed which has a drop down
menu, allowing one channel (selected by name) or all channels to be selected for zeroing. This operation
should only be carried out In clean air. A confirmation message will report the outcome of the operation.
See Fig 6.10.

Calibrate This allows an Individual channel in instrument to be calibrated. See Fig 6.11. Select the gas
channel to be calibrated from the drop down menu. Type
in the gas value; this should be given on the gas cylinder label. When the gas has been flowing for at
least 30 seconds and the reading has stabilized, click OK. A confirmation message will report the
outcome of the operation.
Note: Refer to the instrument's manual for
more information on the Zero and Calibrate functions. Fig 6.11 Clear Log A warning box is
displayed before the data log is cleared. Acknowledge that the data log is to be cleared by clicking on
Yea.
Cal Due This allows you to override the calibration due date in the Instrument. See Fig 6.12. Enter
the number of days before you want the calibration due waning to appear on the instrument. Only use this
facility if you know the instrument responds correctly to gas, or if you employ other systems to ensure
that the instruments are calibrated regularly.

You might also like